Waters Reagent Manager Operator s Guide

Size: px
Start display at page:

Download "Waters Reagent Manager Operator s Guide"

Transcription

1 Waters Reagent Manager Operator s Guide / Revision A Copyright Waters Corporation 2010 All rights reserved

2 Copyright notice 2010 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND IN IRELAND. ALL RIGHTS RESERVED. THIS DOCUMENT OR PARTS THEREOF MAY NOT BE REPRODUCED IN ANY FORM WITHOUT THE WRITTEN PERMISSION OF THE PUBLISHER. The information in this document is subject to change without notice and should not be construed as a commitment by Waters Corporation. Waters Corporation assumes no responsibility for any errors that may appear in this document. This document is believed to be complete and accurate at the time of publication. In no event shall Waters Corporation be liable for incidental or consequential damages in connection with, or arising from, its use. Trademarks Waters is a registered trademark of Waters Corporation, and THE SCIENCE OF WHAT S POSSIBLE. is a trademark of Waters Corporation. Other registered trademarks or trademarks are the sole property of their owners. ii

3 Customer comments Waters Technical Communications department invites you to tell us of any errors you encounter in this document or to suggest ideas for otherwise improving it. Please help us better understand what you expect from our documentation so that we can continuously improve its accuracy and usability. We seriously consider every customer comment we receive. You can reach us at Contacting Waters Contact Waters with enhancement requests or technical questions regarding the use, transportation, removal, or disposal of any Waters product. You can reach us via the Internet, telephone, or conventional mail. Waters contact information Contacting medium Information Internet The Waters Web site includes contact information for Waters locations worldwide. Visit Telephone and fax From the USA or Canada, phone HPLC, or fax For other locations worldwide, phone and fax numbers appear in the Waters Web site. Conventional mail Waters Corporation 34 Maple Street Milford, MA USA iii

4 Safety considerations Some reagents and samples used with Waters instruments and devices can pose chemical, biological, and radiological hazards. You must know the potentially hazardous effects of all substances you work with. Always follow Good Laboratory Practice, and consult your organization s safety representative for guidance. When you develop methods, follow the Protocol for the Adoption of Analytical Methods in the Clinical Chemistry Laboratory, American Journal of Medical Technology, 44, 1, pages (1978). This protocol addresses good operating procedures and the techniques necessary to validate system and method performance. Considerations specific to the Reagent Manager Pinch-point hazard Warning: To avoid hazards associated with the reciprocating or rotating parts in the source, keep clear of the regions marked with yellow and gray labels. High temperature hazard Warning: To avoid burn injuries, avoid touching the source enclosure with your hand when operating or servicing the instrument. High voltage hazard Warning: To avoid electric shock, do not remove the instrument s protective panels. The components they cover are not user-serviceable. To avoid nonlethal electric shock when the instrument is in Operate mode, avoid touching the areas marked with the high voltage warning symbol. To touch those areas, first put the instrument in Standby mode. iv

5 Safety advisories Consult Appendix A for a comprehensive list of warning and caution advisories. Environmental information This product complies with the European Union RoHS Directive 2002/95/EC. Operating this device When operating this device, follow standard quality-control (QC) procedures and the guidelines presented in this section. Applicable symbols Symbol Definition Authorized representative of the European Community Confirms that a manufactured product complies with all applicable European Community directives ABN Australia C-Tick EMC Compliant Confirms that a manufactured product complies with all applicable United States and Canadian safety requirements This product has been tested to the requirements of CAN/CSA-C22.2 No , second edition, including Amendment 1, or a later version of the same standard incorporating the same level of testing requirements v

6 Audience and purpose This guide is intended for use by individuals who need to operate, maintain, and/or troubleshoot the Waters Reagent Manager. You should be familiar with HPLC terms, practices, and basic HPLC system operations such as connecting tubing. Intended use of the Reagent Manager Waters designed the Reagent Manager as a dependable metering pump for general laboratory or industrial use. It is a single-piston, pulse-dampened pumping system for post-column reagent delivery. ISM classification ISM Classification: ISM Group 1 Class B This classification has been assigned in accordance with CISPR 11 Industrial Scientific and Medical (ISM) instruments requirements. Group 1 products apply to intentionally generated and/or used conductively coupled radio-frequency energy that is necessary for the internal functioning of the equipment. Class B products are suitable for use in both commercial and residential locations and can be directly connected to a low voltage, power-supply network. vi

7 EC Authorized Representative Waters Corporation (Micromass UK Ltd.) Floats Road Wythenshawe Manchester M23 9LZ United Kingdom Telephone: Fax: Contact: Quality manager vii

8 viii

9 Table of Contents Copyright notice... ii Trademarks... ii Customer comments... iii Contacting Waters... iii Safety considerations... iv Considerations specific to the Reagent Manager... iv Safety advisories... v Environmental information... v Operating this device... v Applicable symbols... v Audience and purpose... vi Intended use of the Reagent Manager... vi ISM classification... vi ISM Classification: ISM Group 1 Class B... vi EC Authorized Representative... vii 1 Overview Introduction Features Polymeric flow path Automatic seal wash Front control panel Priming and purging Filters Spill protection Remote Start/Stop feature Fluid-handling components Flow paths Pump cycle Table of Contents ix

10 Pulse damper Electronic components Remote Start/Stop feature Installation Installation tasks Site requirements Unpacking Making electrical connections AC Voltage input Connecting to the power supply Connecting for remote start/stop operation Making fluidic connections Connecting the solvent reservoir to the pump inlet Connecting the pump outlet to a device Connecting the overflow tubing Operating the Waters Reagent Manager Using the front control panel Digital display Operation keys Status LEDs Preparing the Reagent Manager Solvent guidelines Priming the pump Purging the pump Priming the seal wash tubing Operating as a stand-alone pump Before operating Operating the pump After operating Routine tasks x Table of Contents

11 Remote start/stop operation Using the Waters ACQUITY binary solvent manager Using the Waters 2695 separations module Using the Waters 600E controller Maintaining the Waters Reagent Manager Maintenance Considerations Safety and handling Proper operating procedures Spare parts Routine maintenance procedures schedule Contacting Waters Technical Service Filters Replacing the inlet filter Replacing the outlet filter Replacing the high-pressure inline filter Check valves Cleaning the check valves Replacing the pump head check valves Check valve and pressure ramp test Pump head, piston, and components Removing the pump head Cleaning and inspecting the piston Replacing the seal-wash components Installing the piston seal Installing the pump head Prime/purge valve Inspecting the prime/purge valve Removing the prime/purge valve Replacing the prime/purge valve seal Installing the prime/purge valve Table of Contents xi

12 Restrictor coil Fuse Fan Troubleshooting Overview Contacting Waters Technical Service General troubleshooting Troubleshooting fluidics Troubleshooting leaks Troubleshooting electronics Identifying and correcting audible noises A Safety Advisories... A-1 Warning symbols... A-2 Task-specific hazard warnings... A-2 Specific warnings... A-3 Caution symbol... A-4 Warnings that apply to all Waters instruments... A-5 Electrical and handling symbols... A-11 Electrical symbols... A-11 Handling symbols... A-12 B Specifications... B-1 C Spare Parts... C-1 D Solvent Considerations... D-1 Solvent preparation... D-2 Solvent outgassing and sparging... D-2 Solvent quality... D-2 Using buffers... D-3 xii Table of Contents

13 Solvent miscibility... D-3 Solvent viscosity... D-6 Solvent degassing... D-6 Gas solubility... D-6 Solvent degassing methods... D-7 Preventing cavitation... D-8 Filtration... D-9 Index... Index-1 Table of Contents xiii

14 xiv Table of Contents

15 1 Overview Topic Page Introduction 1-2 Features 1-3 Fluid-handling components 1-5 Electronic components 1-10 Remote Start/Stop feature

16 Introduction The Waters Reagent Manager is a high-performance pump designed for use as a dependable metering pump for general laboratory or industrial use. As a single-piston, pulse-dampened pumping system for post-column reagent delivery, the Waters Reagent Manager combines the most important aspects of solvent delivery: High precision and accuracy Reliability Smooth solvent flow Ease of use The reagent manager is designed for optimum performance, operational convenience, and safety. The following figure shows the front view of a Waters Reagent Manager. Waters Reagent Manager Front control panel Mode Fault Run ml/min PSI Hi Pr Lo Pr PRIME/PURGE Pump head with seal wash Prime Run/ Stop FILTER/OUTLET Waters Reagent Manager Power switch I/O TP Overview

17 Features The reagent manager includes a full array of important features, such as pulse damping, a nonmetallic flow path, automatic washing of the pump piston seal, and an easy-to-use front control panel. Other features include a convenient priming/purging system, a high-pressure inline filter, spill protection, and a remote start/stop feature. The Waters Reagent Manager operates from 0.05 to 2.00 ml/min up to 2500 psi. The flow rate can be set in 0.01-mL increments with a precision of 0.5%. The pump automatically shuts down at user-settable high- or low-pressure limits. The pump contains a pulse damper to minimize detector baseline noise. Low flow modulation of the reciprocating, single-piston pump is achieved with an advanced electronic rapid-refill design, programmed stepper motor acceleration, and a pulse damper. The rapid refill design reduces pulsations in the pump cycle and provides constant, stable flow from the pump at high pressure. Polymeric flow path The materials that make up the fluidic pathway of the Waters Reagent Manager are nonmetallic and biocompatible. They include: polyetheretherketone (PEEK) polytetrafluoroethylene (PTFE) Tefzel sapphire ruby polyethylene fluoroethylene propylene copolymer (FEP) ultra-high molecular weight polyethylene (UHMWPE) Wetted surfaces in the pump head include ruby, sapphire, and polymer (PEEK). Features 1-3

18 Automatic seal wash During pump operation, the pump head is continuously washed in the region behind the primary high pressure seal. The seal washing action helps to reduce seal wear by washing away the microscopic layer of solvent that wets the area between the piston and the seal. If you use buffered solvents, the solvent layer can evaporate behind the seal and deposit crystals on the piston. Crystals on the piston can abrade the pump seal and cause premature seal failure, leakage, and damage to the pump. The seal-washing pump head washes away the film of solvent before crystals can precipitate. Front control panel A keypad and LED display on the front panel allow you to quickly enter parameters before and during a run. The solvent-resistant membrane keypad has touch keys, while the display provides status and monitoring information for the pump: The four-digit LED front panel display shows the flow rate or pressure. The status LED indicators show the operational status of the pump. Priming and purging Filters The prime/purge valve and the Prime key on the front panel provide quick and easy solvent introduction and rapid solvent changeover using the priming and purging feature. To help maintain the solvent quality through the pump flow path, the reagent manager uses the following filters: An inlet filter In the solvent reservoir (external, slip-on type) A high-pressure inline filter Between the pulse damper and the restrictor coil An outlet filter At the filter/outlet port 1-4 Overview

19 Spill protection The reagent manager has internal spill protection in case of leaks in the flow path. In the event of a leak, solvent drips to the drip pan and is routed to the waste fluid drain connector on the bottom right side of the pump. Recommendation: Provide a solvent bottle tray for the top of the pump if you decide to place solvents or any instrument with fluidic lines on top of the pump. For example, a solvent bottle tray can provide storage for solvent reservoirs and a seal wash reservoir, with leak protection for a spill. Remote Start/Stop feature You can easily integrate the reagent manager into your HPLC system using remote stop/start inputs and stop output capabilities. To accomplish this, the reagent manager accepts remote run/stop input (contact closure) signals from other instruments. It also generates a remote stop output signal to an instrument when a high- or low-pressure fault is triggered. This capability requires a remote start/stop cable, which is included in the startup kit. Fluid-handling components The fluid-handling components of the Waters Reagent Manager are shown in the following figure. Fluid-handling components 1-5

20 Pump fluid-handling components Pump outlet Pump head Seal wash outlet Seal wash inlet Pump inlet Prime/purge valve Filter/outlet valve TP01563 Caution: To prevent solvent leaks from the pump outlet, position the solvent reservoir below the pump inlet before you connect the pump outlet fitting. The following table describes the functions of the major fluidics components. Fluid-handling components Component Pump head Pump inlet and outlet Function Draws in and delivers solvent. Maintains flow direction of solvent through the pump using check valves that open in one direction only. 1-6 Overview

21 Fluid-handling components (Continued) Component Prime/purge valve Filter/outlet valve Seal wash inlet and outlet Function When the valve is open, allows priming and purging of the pump. Considerations: Connecting a syringe to the valve allows you to draw solvent through the pump for priming. Operating the pump allows solvent to flow through the pump and out the valve for purging. When the valve is closed, high-pressure flow is directed to the filter/outlet valve. Routes solvent out of the pump to a fitting, injector, column, or detector. Washes the piston seal with seal wash solvent using check valves to maintain flow of seal wash in one direction. Flow paths The flow path of a solvent through the pump is as follows: During priming and purging Solvent flows from the solvent reservoir through the pump head to the pulse damper and out the prime/purge valve. The flow rate ramps up to 10 ml/min during priming and purging. During normal operation Solvent flows from the inlet filter in the solvent reservoir through the pump head to the pulse damper, prime/purge valve, inline filter, restrictor coil, and out the filter/outlet. Tip: As shown in the following figure, the seal wash solvent (dotted lines) continuously flushes the pump in the area behind the piston seal. Fluid-handling components 1-7

22 Flow paths of solvents through pump Reagent manager Pulse damper Pump Head Seal wash reservoir Outlet check valve for seal wash Solvent outlet check valve Prime/purge valve High-pressure inline filter Fluid waste Solvent reservoir with filter Inlet check valve for seal wash Solvent inlet check valve Restrictor coil Filter/outlet Next device Pump cycle The pump cycle consists of two phases, the delivery phase and the refill phase. During the delivery phase, the pump piston moves forward at a constant linear speed, and is eccentrically driven by the motor. This results in a constant, stable flow from the pump at high pressure. During the refill phase, the piston quickly retracts, refilling the pump head with solvent. Although the pump head output flow completely stops during refill, the pulse damper transiently maintains some flow. In addition, the motor speed is adjusted by the microprocessor to facilitate an efficient refill phase. The combination of increased motor speed and the rapid refill design of the cam generates a refill time of less than 5% of the pump cycle at 1 ml/min. 1-8 Overview

23 Pulse damper The patented diaphragm pulse damper helps to maintain the flow rate and pressure. The pulse damper consists of a compressible fluid (isopropanol) held in an isolated cavity by a flexible diaphragm. During the delivery phase, the pressure of the solvent displaces the diaphragm, compressing the fluid in the cavity and storing energy. During the refill phase, the pressure of the pumped solvent on the diaphragm is reduced and the compressed fluid expands, releasing its stored energy and continuing to pressurize the solvent path. Because the amount of solvent in contact with the pulse damper is small (0.9 ml at 2500 psi), and the flow past the damper is completely swept, solvent changeover is rapid. The pulse damper requires a back pressure of approximately 500 psi or greater. A length of small-bore, restrictor tubing (the restrictor coil) between the pulse damper and the filter/outlet ensures that this minimum back pressure is maintained. Fluid-handling components 1-9

24 Electronic components The Waters Reagent Manager electronics are illustrated in the following figures. Pump electronics front view LED display Fault ml/min PSI Hi Pr Lo Pr Status LEDs Keypad Mode Prime Run Run/ Stop Waters Reagent Manager Power switch I/O TP Overview

25 Pump electronics back view The following table describes the functions of the electronic components. Electronic components Component Location Function LED display Front Provides pump status and monitoring information. Keypad with operation keys Front Use to select pump operating parameters and values. Status LEDs Front Provides status information. Power switch Front Powers the pump on and off. Power entry module Back Provides the power cord connector and a receptacle for a fuse. Cooling fan Back Exhausts air to cool the internal electronics. Terminal strip Back Provides connection through a remote start/stop cable to an external control device for remote start/stop signal control. Also provides a remote stop output on pressure fault. See Connecting for remote start/stop operation on page 2-5 for more information. RS-232 connector Back Reserved for future use. Electronic components 1-11

26 Remote Start/Stop feature When the reagent manager is configured for remote start/stop operation, the external control device triggers the pump on or off by contact closure signals sent over a cable. If a fault occurs in the pump during operation, the reagent manager stops running and sends a signal to the external device. The reagent manager startup kit contains the remote start/stop cable for remote start/stop operation. See Connecting for remote start/stop operation on page 2-5, for instructions on how to install the remote start/stop cable Overview

27 2 Installation This chapter describes how to unpack and install your Waters Reagent Manager. Topic Page Installation tasks 2-2 Site requirements 2-2 Unpacking 2-3 Making electrical connections 2-5 Making fluidic connections

28 Installation tasks The following figure outlines the installation tasks. Flowchart for installing the reagent manager Start installation Make signal and power connections Select a site Make fluidic connections Unpack Prime the pump Inspect Installation complete Site requirements Install the reagent manager on a stable, flat, clean surface at a site that meets the specifications indicated in the following table. Warning: To avoid overheating and to provide clearance for cable connections, ensure that the rear panel of the reagent manager has at least 3 inches (7.6 cm) of clearance. Installation site requirements Factor Temperature Relative humidity Requirement 5 to 35 C (41 to 95 F) 20% to 90%, noncondensing 2-2 Installation

29 Installation site requirements (Continued) Factor Size Clearance Input voltage ranges Input voltage frequency Electromagnetic fields Requirement Width: 9.1 in. (23.1 cm) Depth: 17.4 in. (44.2 cm) Height: 6.5 in. (16.5 cm) At least 3 in. (7.6 cm) at rear for ventilation and cable connections 100 to 240 VAC Universal Input - No fuse change required 50 to 60 Hz No nearby source of electromagnetic noise, such as arcing relays or electric motors Note: The unit is only for indoor use, at a maximum altitude of 2000m. The unit is intended for use in a POLLUTION DEGREE 2 ENVIRONMENT. Unpacking The Waters Reagent Manager is shipped in one carton that contains the following items: Waters Reagent Manager Reagent Manager Startup Kit (see Appendix C for a list of contents) Prior to opening the shipping container, inspect it for damage or evidence of mishandling. If it has been damaged or mishandled, notify the carrier before opening the container. See the Waters Licenses, Warranties, and Support Services document (on the documentation CD) for more information. Tip: Save the shipping container for future use. To unpack the reagent manager 1. Open the carton and remove the startup kit and other items from the slot at one end of the carton. 2. Using both hands, lift the pump and its foam packing material out of the carton. Unpacking 2-3

30 3. Carefully set the pump down, then remove the foam packing material from the ends of the pump. 4. Check the contents of the startup kit against the startup kit parts list to confirm all items are included. 5. Verify that the serial number on the back of the pump matches the serial number on the Certificate of Functional Performance. Keep the Certificate of Functional Performance for future reference. 6. For more information about the warranty, see the Waters Licenses, Warranties, and Support Services document (on the documentation CD). 7. Inspect all items for damage. Immediately report any shipping damage to both the shipping company and your Waters representative. Contact Waters Technical Service at , U.S. and Canadian customers only. Other customers, call your local Waters subsidiary or Technical Service Representative, or consult the Waters web site ( for worldwide phone and fax numbers. See the Waters Licenses, Warranties, and Support Services document for complete information on reporting shipping damages and submitting claims. 2-4 Installation

31 Making electrical connections For proper operation, the reagent manager requires a grounded AC power supply with no abrupt voltage fluctuations. AC Voltage input The reagent manager allows voltage input from 100 to 240 VAC at 50 to 60 Hz. No fuse change is required. Connecting to the power supply The reagent manager automatically adjusts for AC input voltage and frequency. To connect the power supply 1. Ensure that the power switch on the front panel is in the Off (O) position. 2. Insert the power cord into the power connector on the back of the pump. 3. Insert the other end of the power cord into a grounded power outlet. Connecting for remote start/stop operation For optional remote start and stop operation, the remote start/stop cable is required. This cable is included in the startup kit. This cable connects the Waters Reagent Manager to a Waters ACQUITY Binary Solvent Manager, a Waters 2695 Separations Module, a Waters 600E Controller, or other external device equipped with TTL-logic or contact closure outputs. Making electrical connections 2-5

32 The following table describes the connections on the reagent manager terminal strip. Reagent Manager terminal strip connections Term Number Description STOP OUT - NO 8 Stop output signal, normally open; contact closure signal from reagent manager to external device a STOP OUT - NC (None) Stop output signal, normally closed; contact closure signal from reagent manager to external device b OUTPUT COM 6 Ground for output connection RUN INPUT 3 Contact closure signal from external device to start pumping INPUT COM 1 Ground for input connection a. Waters devices are normally open b. Other devices may require normally closed for TTL closure To connect the reagent manager to a Waters ACQUITY binary solvent manager 1. Remove the terminal strip from the back of the reagent manager. 2. Attach the leads on the end of the start/stop cable to the reagent manager terminal strip as shown in Figure Connecting to a Waters ACQUITY binary solvent manager on page 2-8. Use the Table titled Reagent Manager terminal strip connections on page 2-6 as a guide. 3. Reinstall the terminal strip on the back of the reagent manager. 4. Remove the terminal strip from the Waters ACQUITY binary solvent manager. 2-6 Installation

33 5. Connect the leads on the end of the start/stop cable to the terminal strip of the Waters ACQUITY binary solvent manager as shown in Figure Connecting to a Waters ACQUITY binary solvent manager on page Reinstall the terminal strip on the Waters ACQUITY binary solvent manager. Making electrical connections 2-7

34 Connecting to a Waters ACQUITY binary solvent manager Remote Start/Stop Cable Waters Reagent Manager Terminal Strip Green Blue White Red STOP OUT - NO 7 STOP OUT - NC 6 OUTPUT COM RUN INPUT 2 STOP INPUT 1 INPUT COM Waters ACQUITY Binary Solvent Manager Stop Flow - Stop Flow Green Blue Run Stopped Out Run Stopped Out White Red I II 2-8 Installation

35 To connect the reagent manager to a Waters 2695 separations module 1. Remove the terminal strip from the back of the reagent manager. 2. Attach the leads on the end of the start/stop cable to the reagent manager terminal strip as shown in Figure Connecting to a Waters 2695 separations module on page Use the Table titled Reagent Manager terminal strip connections on page 2-6 as a guide. 3. Reinstall the terminal strip on the back of the reagent manager. 4. Remove the terminal strip from the Waters 2695 separations module. 5. Connect the leads on the end of the start/stop cable to the terminal strip of the Waters 2695 separations module as shown in Figure Connecting to a Waters 2695 separations module on page Reinstall the terminal strip on the Waters 2695 separations module. Making electrical connections 2-9

36 Connecting to a Waters 2695 separations module Remote Start/Stop Cable Waters Reagent Manager Terminal Strip Green Blue White Red STOP OUT - NO 7 STOP OUT - NC 6 OUTPUT COM RUN INPUT 2 STOP INPUT 1 INPUT COM Waters 2695 Separations Module Green Blue White Red + Stop Flow - Stop Flow Switch 1 Switch Connector B Connector A 2-10 Installation

37 To connect the Reagent Manager to a Waters 600E Controller 1. Remove the terminal strip from the back of the reagent manager. 2. Attach the leads on the end of the start/stop cable to the reagent manager terminal strip as shown in Figure Connecting to a Waters 600E controller on page Use the Table titled Reagent Manager terminal strip connections on page 2-6 as a guide. 3. Reinstall the terminal strip on the back of the reagent manager. 4. Remove the terminal strip from the Waters 600E Controller. 5. Connect the leads on the end of the start/stop cable to the terminal strip of the Waters 600E Controller as shown in Figure Connecting to a Waters 600E controller on page Reinstall the terminal strip on the back of the Waters 600E Controller. Making electrical connections 2-11

38 Connecting to a Waters 600E controller Reagent Manager Remote Start/Stop Cable Terminal Strip Green Blue White Red STOP OUT - NO 7 STOP OUT - NC 6 OUTPUT COM RUN INPUT 2 STOP INPUT 1 INPUT COM White: Switch 1 Green: Stop Terminal Strips Waters 600E Controller Blue: Ground Red: Ground 2-12 Installation

39 Making fluidic connections As shown in the Figure Connecting the tubing on page 2-14, making fluidic connections to the Waters Reagent Manager involves: Connecting the solvent reservoir to the pump inlet Connecting the pump outlet to a device Installing the seal wash tubing that washes the piston seal on the pump head Connecting tubing from the waste fluid drain under the pump to a waste container Making fluidic connections 2-13

40 Connecting the tubing Note: Ensure that the reservoirs are at least 4 in. (10 cm) above the pump head during operation. Reagent Reservoir Reagent Tubing Seal Wash Reservoir Waters Reagent Manager Seal Wash Tubing Priming Syringe Outlet Tubing to Next Device Seal Wash Tubing to Waste Container TP01565e Recommendation: Degas, filter, and cover all solvents. See Appendix D for solvent considerations. Required Materials Startup Kit Inlet filter, inlet ferrule, and inlet nut (one each) Outlet ferrule and outlet nut (three each) Inlet tubing, 48 in. (122 cm), inch i.d. 1/8-inch o.d Installation

41 Seal wash tubing, PTFE, 72 in. (183 cm), 1/8-inch i.d. 1/4-inch o.d. Seal wash fittings (two each) Priming syringe (30 ml) Outlet tubing, PEEK, 60 in. (152 cm), inch i.d. 1/16-inch o.d. Overflow tubing, PTFE, 72 in. (183 cm, cut in two 36-in [91.44 cm] lengths), 1/4-inch i.d. 3/8-inch o.d. Polymeric tubing cutter Wrench, 5/16-in. open-end Reservoir containing filtered, degassed solvent Reservoir containing filtered seal wash solvent (20% (v/v) methanol in water, or 20% (v/v) acetonitrile in water) Waste container Connecting the solvent reservoir to the pump inlet Caution: To avoid a solvent leak from the pump outlet during installation, position the reservoir below the pump inlet until the pump outlet is connected to the system. During pump operation, position the bottom of the reservoir about 4 inches (10 cm) higher than the pump head. If you use high-viscosity solvents, you may need to elevate the reservoirs as much as 18 to 24 inches (46 to 61 cm) above the inlet to prevent bubble formation as the pressure drops during the refill phase. Making fluidic connections 2-15

42 To connect the solvent reservoir to the pump solvent inlet 1. Measure the length of inlet tubing required to connect the solvent reservoir to the pump inlet. The inlet tubing should be as short as practical to minimize pressure losses on the inlet side of the pump during the refill phase (to prevent bubble formation). 2. Use the tubing cutter to cut the PEEK tubing to the measured length. Make sure the cut end is straight and free from burrs. 3. Slide a nut and then a ferrule over one end of the tubing, as shown in the following figure. Attaching a ferrule and nut to tubing Nut Tubing Ferrule Tubing End (straight and smooth) Caution: To avoid damaging the ferrule, do not overtighten the nut. 4. Firmly seat the tubing end into the inlet on the pump, then finger-tighten the nut. 5. Insert the free end of the inlet tubing through the cap of the solvent reservoir. 6. Cut a 1.5-inch (3.8-cm) section of outlet tubing and slide half of its length over the end of the inlet tubing. 7. Insert the tubing fitting of the inlet filter into the outlet tubing sleeve at the free end of the inlet tubing. 8. Install the cap onto the solvent reservoir (containing filtered, degassed solvent) and push the outlet tubing through the cap until the filter reaches the bottom of the reservoir Installation

43 Connecting the pump outlet to a device To connect the pump outlet to a device 1. Measure the length of outlet tubing required to connect the pump outlet to a device. Use a polymeric tubing cutter to cut the tubing to the measured length. Ensure that the tubing is cut squarely with no burrs or crimps and that the center hole is open. 2. Remove the fitting plug from the pump filter/outlet fitting. 3. Slide an outlet nut and then a ferrule over one end of the tubing (see the above figure). 4. Firmly seat the tubing into the filter/outlet fitting, then finger-tighten the nut. 5. Unscrew the assembled fitting and verify that the length of tubing extending beyond the ferrule is 0.12 inch (3 mm). 6. Finger-tighten the nut again. 7. Repeat steps 3 through 6 for the other end of the tubing only after you have primed the pump and flushed the system, as described in Priming the pump on page 3-6. Installing and priming the seal wash tubing The pump head seal wash requires 250 ml to 500 ml of seal wash solvent. The seal wash solvent drains to a fluid waste container below the pump. Use an aqueous seal wash solvent with enough organic content to inhibit bacterial growth. To install the seal wash tubing 1. Cut the seal wash tubing at its midpoint. 2. Install seal wash fittings on one end of each length of seal wash tubing. 3. Screw the fitting on one length of tubing to the seal wash inlet as shown in the following figure. Ensure that the fitting and tubing fit tightly to prevent leaks. Making fluidic connections 2-17

44 Connecting the seal wash tubing Tubing from seal wash outlet to waste container Tubing from seal wash reservoir to seal wash inlet TP Install the other end of the tubing in the seal wash reservoir. Secure the tubing in the reservoir to ensure that the tubing stays immersed during pump operation. Position the reservoir at least 4 in. (10 cm) higher than the pump head. 5. Screw the fitting on the other length of tubing to the seal wash outlet. Ensure that the fitting and tubing fit tightly. 6. To prime the seal wash tubing, connect the seal wash fitting to the priming syringe, then connect the syringe to the free end of the seal wash outlet tubing. 7. Draw 10 ml to 20 ml of seal wash solvent through the seal wash outlet. Ensure that the tubing is filled with solvent and that no bubbles remain in the lines. 8. Disconnect the syringe, then install the end of the seal wash outlet tubing in a waste container located at a level below the pump so that fluid drains by gravity. Secure the tubing to ensure that the tubing stays in the container during pump operation Installation

45 Connecting the overflow tubing To install the overflow tubing 1. Connect one end of the overflow tubing to the drip tray connector underneath the reagent manager. 2. Connect the other end of the tubing to a waste container located at a level below the pump. Making fluidic connections 2-19

46 2-20 Installation

47 3 Operating the Waters Reagent Manager This chapter describes how to prepare and operate the Waters Reagent Manager either as a stand-alone pump using the front control panel or under remote start/stop operation. Topic Page Using the front control panel 3-2 Preparing the Reagent Manager 3-5 Operating as a stand-alone pump 3-8 Remote start/stop operation

48 Using the front control panel The front control panel of the Reagent Manager consists of a four-digit LED display, the operation keys, and status LEDs, as shown in the following figure. Reagent Manager front control panel Display Fault ml/min PSI Hi Pr Lo Pr Status LEDs Operation Keys Mode Prime Run Run/ Stop Waters Reagent Manager Power Switch To power-on the pump, locate the power switch on the lower left front panel and set it to the On (I) position. Digital display The four-digit LED display can show the pump flow rate (ml/min), system pressure (psi), or the high- or low-pressure settings (psi). You use the Mode key to choose which parameter to display. Operation keys I/O TP01560 The operation keys provide access to all pump operating parameters. Use the Mode, Up, Down, and Run/Stop operation keys to navigate through the parameters. 3-2 Operating the Waters Reagent Manager

49 The following table describes the functions of the operation keys on the Reagent Manager control panel. Functions of operation keys Key Mode Function Cycles through the four (parameter) modes: flow rate in ml/min, pressure in psi, high-pressure limit, and low-pressure limit. The status LEDs to the right of the digital display indicate the active mode. Up - increases the value for the selected mode (parameter). Down - decreases the value for the selected mode (parameter). Prime Initiates priming at 10 ml/min. To stop priming, press any key. Run/ Stop When the unit is not pumping, starts the pump at all current parameter settings. The maximum flow rate during operation is 2.0 ml/min. When the unit is pumping, immediately stops the pump. Tip: Also clears any high- or low-pressure faults that may be occurring within the Reagent Manager. Functions of and details on all Status LED parameters are provided in Table titled Status LEDs on page 3-4. Tip: You can display the software version of the Reagent Manager using the Run/Stop and Up arrow keys. To display the software version, prior to turning on the instrument, press and hold both the Run/Stop and Up arrow keys, power-on the Reagent Manager, and then release the two keys when User appears in the LED display. The Reagent Manager software version appears in the LED display (in decimal point format). To exit this test mode, press the Run/Stop key. Using the front control panel 3-3

50 Status LEDs The status LEDs display the real-time status of the pump. This table describes all Status LEDs. Status LEDs Parameter Status LED Description Parameter value Flow Rate ml/min Lights when the flow rate is displayed. Pressure PSI Lights when the current system pressure is displayed. High- Pressure Limit Low- Pressure Limit Hi Pr Lo Pr Lights when the user-selected high- pressure limit (in psi) is displayed. Lights when the user-selected low- pressure limit (in psi) is displayed. Fault Fault Lights when a fault occurs and stops the pump. Run Run Lights when the pump is running. The flow rate in ml/min at which the pump is currently operating. The flow rate range is 0.05 ml/min to 2.00 ml/min. The pressure in psi at which the pump is currently set to operate. The pressure range is 0 psi to 2500 psi. The pressure is a read-only value. Select this mode to track the pressure during priming, operating, and troubleshooting tasks. The pressure at or above which the pump is to shut down. The Fault LED also lights. You can set the Hi Pr in 10-psi increments. The maximum value is 2500 psi. The minimum value is the low-pressure limit plus 100 psi. The pressure at or below which the pump is to shut down. The Fault LED also lights. You can set the Lo Pr in 10-psi increments. The maximum value is the upper limit minus 100 psi. The minimum value is 0 psi. On indicates a problem that must be cleared before operation. Off indicates that the pump is ready to operate. On indicates that the pump is running or priming. Off indicates that the pump is not running. 3-4 Operating the Waters Reagent Manager

51 Preparing the Reagent Manager The procedure to prepare the Reagent Manager for operation depends on the status of the pump. If the pump is dry, you must prime the pump (see Priming the pump on page 3-6). If you are changing solvent, restarting the pump after a long idle time, or adding fresh solvent, and the pump contains fluid, you must purge the pump (see Purging the pump on page 3-7). Solvent guidelines Required Materials Priming syringe Solvents for the system, filtered and degassed Waste tubing Waste container Seal wash solvent Selection and preparation of solvents to use in the Waters Reagent Manager are critical factors for successful operation. See Appendix D for detailed solvent consideration information. Caution: Do not use any of the following solvents in the Waters Reagent Manager because they damage PEEK components. Methylene chloride Tetrahydrofuran (THF) at greater than 10% (v/v) Freon Concentrated sulfuric acid (H 2 SO 4 ) Concentrated nitric acid (HNO 3 ) Preparing the Reagent Manager 3-5

52 Also keep in mind the following solvent guidelines: Use methanol to prime a new pump or a pump that has been stored for an extended period. Use an intermediate solvent when you change between solvents that are not miscible. Always keep the solvent level in the reservoir at least 4 inches (10 cm) higher than the pump inlet. Do not place solvent reservoirs on top of the Reagent Manager. Use only filtered, degassed solvents. After using a buffer, purge the pump with HPLC-grade, distilled, or deionized water to prevent crystallization and damage to the piston or seal. Know the properties of the solvents under different pressures. Dedicate a solvent supply tube and filter assembly for each solvent. If this is not possible, purge the tube and filter of any remaining solvent before you use a new solvent. To prevent cross-contamination of solvents, dedicate a set of glassware for preparing, storing, and supplying each solvent. Once a container has been used for a solvent, do not use it for any other solvent unless it has been thoroughly cleaned. Warning: To avoid chemical hazards and prevent injury, always observe Good Laboratory Practices when you handle solvents. Know the physical and chemical properties of the solvents you use. See the Material Safety Data Sheets for the solvents in use. For more information on solvent properties, see Appendix D. Priming the pump To prime the pump 1. Power on the pump. 2. Raise the reservoir so that its base is above the level of the pump inlet. 3. Open the prime/purge valve by turning it fully counterclockwise. 4. Connect the fully discharged priming syringe to the prime/purge valve (see the following figure). 3-6 Operating the Waters Reagent Manager

53 Priming the Reagent Manager Prime/Purge Valve Priming Syringe 5. Press Prime to run the pump. The pump automatically ramps up to 10 ml/min. 6. Draw at least 20 ml of solvent (and any air bubbles) through the pump and into the syringe. 7. Press Run/Stop or any key to stop priming. 8. Close the prime/purge valve by turning it fully clockwise, and remove the syringe. Dispose of solvent properly. 9. Connect the free end of the outlet tubing to the inlet fitting of the next device in your HPLC system and finger-tighten the nut. Purging the pump Purging ensures that all solvent in the fluidic path is replaced with fresh solvent. Caution: When you change between two solvents that are not totally miscible, follow the purging procedure described below, first using an intermediate solvent, then the final solvent. See Appendix D for more information. Before you press the Prime key, be sure the prime/purge valve is open. To prevent damage to the pump, always draw fluid from the prime/purge valve. Do not inject fluid into the prime/purge valve with a syringe. To prevent solvent flow through the pump head under gravity pressure, never leave the pump unattended with the prime/purge valve in the open position. Preparing the Reagent Manager 3-7

54 To purge the pump 1. Perform steps 1 through 9 in Priming the pump on page Press Run/Stop to start the pump. 3. Let the pump run for about one minute to verify that the pump pressure is normal for the selected solvent and flow rate. If the pump pressure is not correct, repeat steps 1 and 2, as necessary. 4. Press Run/Stop to stop the pump. Priming the seal wash tubing If you see air bubbles in the seal wash tubing, prime the seal wash tubing as described in Installing and priming the seal wash tubing on page Operating as a stand-alone pump You operate the Reagent Manager as a stand-alone pump using only the operation keys on the front control panel. Before operating Before you operate the Reagent Manager, ensure that: You have already completed any required priming and purging, as described in Preparing the Reagent Manager on page 3-5. Solvent reservoirs are filled with filtered, degassed solvent. Caution: To avoid chemical hazards and prevent injury, always observe Good Laboratory Practices when you handle solvents. See the Material Safety Data Sheets for the solvents you use. For more information on solvent properties, see Appendix D. Operating the pump The operation keys provide access to all pump operating parameters. Use the Mode, Up, Down, and Run/Stop operation keys to navigate through the parameters. You can change pump parameters while the pump is running. 3-8 Operating the Waters Reagent Manager

55 To operate the Reagent Manager 1. Press Mode to display the flow rate (as indicated by the illuminated ml/min status LED). Press the Up or Down arrow key to increase or decrease the value. 2. Press Mode to display the high-pressure limit (as indicated by the illuminated Hi Pr status LED). Press the Up or Down arrow key to increase or decrease the value. 3. Press Mode to display the low-pressure limit (as indicated by the illuminated Lo Pr status LED). Press the Up or Down arrow key to increase or decrease the value. 4. Once all parameters values are correctly set, press Mode to display the operating pressure. 5. Press Run/Stop to start the pump. Operation continues until you press Run/Stop to stop the pump. Tip: To immediately stop the pump, press Run/Stop. After operating After operating the Reagent Manager, you can power-off the pump for storage. The procedures to follow before you power-off the pump depend on the status of your chromatography system and the expected length of time until the next use. Use Table titled Pump storage recommendations on page 3-10 as a guide for determining when and in what sequence to perform certain storage procedures. Purging the pump on page 3-7 describes the purging procedure. Caution: To avoid damaging the pump, ensure that the pump always has solvent. Do not allow the pump to run dry. Operating as a stand-alone pump 3-9

56 Pump storage recommendations Status Purge with Water Purge with 20% Methanol Purge with Isopropanol For overnight storage For short-term storage For long-term storage For shipping Short-term storage Required Materials HPLC-grade, distilled, or deionized water 20% (v/v) methanol/water To prepare for short-term storage (up to a week) 1. Use HPLC-grade, distilled, or deionized water to purge the pump (see Purging the pump on page 3-7). 2. If the pump is to be idle for more than one day, use 20% (v/v) methanol/water as the final solvent to purge the pump. If 20% methanol/water is not fully miscible with the current solvent, purge with an intermediate solvent before switching to 20% methanol/water. 3. Press Run/Stop, then turn the power switch to the Off (O) position. Tip: All Reagent Manager parameter values are saved at their current settings when power is turned off Operating the Waters Reagent Manager

57 Long-term storage Required Materials HPLC-grade, distilled, or deionized water 20% (v/v) methanol/water Isopropanol Priming syringe Fitting plug or plastic film Small plastic bag and elastic band To store the pump for more than a week 1. If you use buffers, purge the pump with HPLC-grade, distilled, or deionized water (see Purging the pump on page 3-7). 2. Purge the pump with 20% (v/v) methanol/water. If 20% methanol/water is not fully miscible with the current solvent, purge with an intermediate solvent before switching to 20% methanol/water. 3. Disconnect the outlet tubing from the pump. 4. Place the inlet filter in isopropanol. 5. Open the prime/purge valve, then use the priming syringe to draw a minimum of 50 ml of isopropanol through the pump. 6. Pump a minimum of 5 ml of solvent through the filter/outlet. 7. Press Run/Stop, then turn the power switch to the Off (O) position. 8. With the inlet tubing connected to the pump, place the inlet filter in a small plastic bag and seal the bag against the tubing with an elastic band. 9. Plug the outlet port with the fitting plug, leave a length of outlet tubing on the pump, or cover the outlet port with plastic film. Operating as a stand-alone pump 3-11

58 Routine tasks Perform the following tasks on a routine basis to keep your Reagent Manager running smoothly: Replenish the seal wash solvent weekly or as needed. If the exterior requires cleaning, wipe the surface with a lint-free cloth moistened with water. Wipe up leaks and spills immediately. Empty and rinse the waste fluid containers when they are full or at least weekly. Remote start/stop operation For optional remote start and stop operation, the remote start/stop cable is required. This cable is included in the startup kit. See Connecting for remote start/stop operation on page 2-5 for instructions on how to install the remote start/stop cable. You can remotely start and stop the Reagent Manager using the Waters ACQUITY Binary Solvent Manager, the Waters 2695 Separations Module, or the Waters 600E Controller. Using the Waters ACQUITY binary solvent manager To start the Reagent Manager 1. Power-on the Waters Reagent Manager. 2. On the Waters ACQUITY binary solvent manager front panel, press Menu/Status. 3. Press Switch Press 1, and then press Enter. To stop the Reagent Manager 1. On the Waters ACQUITY binary solvent manager front panel, press Menu/Status. 2. Press Switch Operating the Waters Reagent Manager

59 3. Press 0, and then press Enter. 4. Power-off the Waters Reagent Manager. For more information, see the Waters ACQUITY UPLC System Operator s Guide. Using the Waters 2695 separations module To start the Reagent Manager 1. Power-on the Waters Reagent Manager. 2. On the Waters 2695 Separations Module front panel, press Menu/Status. 3. Press Switch Press 1, and then press Enter. To stop the Reagent Manager 1. On the Waters 2695 Separations Module front panel, press Menu/Status. 2. Press Switch Press 0, and then press Enter. 4. Power-off the Waters Reagent Manager. For more information, see the Waters 2695 Separation Module Operator s Guide. Using the Waters 600E controller To start the Reagent Manager 1. Power-on the Waters Reagent Manager. 2. On the Waters 600E controller front panel, press Direct Screen. 3. Press Switch Press 1, and then press Enter. Remote start/stop operation 3-13

ACQUITY Advanced Polymer Chromatography

ACQUITY Advanced Polymer Chromatography ACQUITY Advanced Polymer Chromatography Isocratic Solvent Manager Overview and Maintenance Guide 715004020 / Revision B Copyright Waters Corporation 2014 All rights reserved ii October 13, 2014, 715004020

More information

PATROL UPLC Process Analysis System for Production

PATROL UPLC Process Analysis System for Production PATROL UPLC Process Analysis System for Production The PATROL UPLC Process Analysis System for Production is an integrated system solution designed and engineered to perform online and atline reaction

More information

High Pressure Piston Pump for HPLC and High Performance Metering

High Pressure Piston Pump for HPLC and High Performance Metering High Pressure Piston Pump for HPLC and High Performance Metering Operator Manual 903034 REV K Teledyne SSI 349 N. Science Park Road State College, PA 16803 https://ssihplc.com Phone: 814-234-7311 Fax:

More information

Passive Check Valve Installation Instructions for H-Class Systems / Revision A Copyright Waters Corporation 2014 All rights reserved

Passive Check Valve Installation Instructions for H-Class Systems / Revision A Copyright Waters Corporation 2014 All rights reserved Passive Check Valve Installation Instructions for H-Class Systems 75004569 / Revision A Copyright Waters Corporation 04 All rights reserved General information Copyright notice 04 WATERS CORPORATION. PRINTED

More information

Product Guide: Series III Pump Control Board Set (RoHS)

Product Guide: Series III Pump Control Board Set (RoHS) revised 04/08/10 Description: The Series III Pump Control Board Set provides motor drive and pump control for a wide assortment of pumps from Scientific Systems, Inc. The assembly consists of two circuit

More information

VERITY Purification Systems Build the perfect system to fit your research

VERITY Purification Systems Build the perfect system to fit your research VERITY Purification Systems Build the perfect system to fit your research SPEC SHEET PURIFICATION VERITY PURIFICATION SYSTEMS WITH VERITY 3011 ISOCRATIC PUMP MAKE YOUR LAB LIFE EASIER From automated GPC

More information

Post Column Reactor. Dual Channel. Operator s Manual REV T

Post Column Reactor. Dual Channel. Operator s Manual REV T Post Column Reactor Dual Channel Operator s Manual 90-2508 REV T Scientific Systems, Inc. 349 N. Science Park Road State College, PA 16803 www.ssihplc.com Phone: 800-441-4752 Fax: 814-238-7532 Email: sales@ssihplc.com

More information

Chromatography. Liquid. HPLC Accessories SupelPRO Valves SUPELCO

Chromatography. Liquid. HPLC Accessories SupelPRO Valves SUPELCO 8 SupelPRO Valves Order:.800..00 Technical Service:.800.9.0 Web: www.sigma-aldrich.com/supelco SupelPRO Automated Fluidics Instruments Supelco s SupelPRO series are precision, electronically-controlled,

More information

ACQUITY UPLC I-Class System with 2D Technology

ACQUITY UPLC I-Class System with 2D Technology ACQUITY UPLC I-Class System with 2D Technology The Waters ACQUITY UPLC I-Class System with 2D Technology allows chemists to increase sensitivity and selectivity, eliminate unwanted interferences, characterize

More information

Application Note. General Use Inlet Solvent Filters. Inlet Solvent Filters. Filters and Degassers. New!

Application Note. General Use Inlet Solvent Filters. Inlet Solvent Filters. Filters and Degassers. New! Inlet Solvent Filters Filters and Degassers General Use Inlet Solvent Filters Large Surface Areas disposable µm, 0 µm, and 0 µm Pore Sizes Available General Use and Prep Filters for Higher Flow Applications

More information

Series II Pump. Operator's Manual rev M

Series II Pump. Operator's Manual rev M Series II Pump Operator's Manual 90-2051 rev M SAFETY SYMBOLS EARTH GROUND CAUTION - REFER TO MANUAL CAUTION HIGH VOLTAGE Scientific Systems, Inc. 349 N Science Park Rd., State College PA 16803 Phone:

More information

PICO Cleaning Instructions

PICO Cleaning Instructions PICO Cleaning Instructions How to clean PICO piezoelectric metering valves IMPORTANT! Save this Sheet. Forward to Maintenance or Tool Crib Supervisors Electronic pdf files of EFD manuals are also available

More information

HPLC Model 210D. UV detector lamp switch. Rheodyne 7725i injection valve. Prime/purge valve. Column. compartment. thumbscrews. Storage.

HPLC Model 210D. UV detector lamp switch. Rheodyne 7725i injection valve. Prime/purge valve. Column. compartment. thumbscrews. Storage. 1. Overview The front panel houses the UV detector lamp switch, Rheodyne injection valve, prime/purge valve, pressure gauge, and the storage compartment. In the storage compartment are the LED display

More information

Varian Dual Chamber Dynamic Mixer Instruction Manual

Varian Dual Chamber Dynamic Mixer Instruction Manual Varian Dual Chamber Dynamic Mixer Instruction Manual Installation Category II Pollution Degree 2 Safety Class 1 (EN 61010-1) Publication no. 0391467400 Issue 5 February 2009 Varian, Inc. Serving Industries

More information

ReciPro Liquid Metering Pumps. Reproducible flow rates High pressure capabilities Corrosion resistant Remote control options

ReciPro Liquid Metering Pumps. Reproducible flow rates High pressure capabilities Corrosion resistant Remote control options ReciPro Liquid Metering Pumps Reproducible flow rates High pressure capabilities Corrosion resistant Remote control options ReciPro HP Series Eldex metering pumps can be used wherever precise flow rate

More information

Waters 1500-Series HPLC Pump and Options Operator s Guide

Waters 1500-Series HPLC Pump and Options Operator s Guide Waters 1500-Series HPLC Pump and Options Operator s Guide 715002013/Revision C Copyright Waters Corporation 2009-2015 All rights reserved Page ii General Information Copyright notice 2009-2015 WATERS CORPORATION.

More information

XLR-8 60 Extended Rack Autosampler

XLR-8 60 Extended Rack Autosampler XLR-8 60 Extended Rack Autosampler Quick Installation Guide Manual Part Number 480226 Rev 0 COPYRIGHT 2015 Teledyne Technologies Incorporated. All rights reserved. 480226 Rev 0, October, 2015 Printed in

More information

Quick Reference Guide

Quick Reference Guide Quick Reference Guide 3 Spike Set Cassette Inflow Connector Cassette Outflow Connector Return Line Latch Handle Patient Line Air Trap Keypad Cassette Customer Service Toll Free: 1-866-944-9992 Office:

More information

Auxiliary Solvent Manager Overview and Maintenance Guide

Auxiliary Solvent Manager Overview and Maintenance Guide Auxiliary Solvent Manager Overview and Maintenance Guide 715004026 / Revision A Copyright Waters Corporation 2014 All rights reserved ii February 6, 2014, 715004026 Rev. A General Information Copyright

More information

PrimeLine TM HPLC Pumps

PrimeLine TM HPLC Pumps PrimeLine TM HPLC Pumps Isocratic Model 501 Gradient System 500G PrimeLine TM Isocratic and Gradient HPLC Pumps Feature: Self-priming pump head requires no manual priming Patented Floating Pump Seal design

More information

2311FA. Pressure Tester USER INSTRUCTION MANUAL BARFIELD M/N 2311FA Revision C April 12, 2011 BARFIELD, INC. Corporate Headquarters

2311FA. Pressure Tester USER INSTRUCTION MANUAL BARFIELD M/N 2311FA Revision C April 12, 2011 BARFIELD, INC. Corporate Headquarters 2311FA Pressure Tester USER INSTRUCTION MANUAL BARFIELD M/N 2311FA 56-101-00212 Revision C April 12, 2011 BARFIELD, INC. Corporate Headquarters 4101 Northwest 29th Street Miami, Florida 33142 www.barfieldinc.com

More information

Economical Piston Pump for HPLC and High Performance Metering

Economical Piston Pump for HPLC and High Performance Metering Economical Piston Pump for HPLC and High Performance Metering Operator Manual 903030 REV I Scientific Systems, Inc. 349 N. Science Park Road State College, PA 16803 www.ssihplc.com Phone: 814-234-7311

More information

Agilent Dynamax Analytical HPLC Column Systems 4.6 mm id Data Sheet

Agilent Dynamax Analytical HPLC Column Systems 4.6 mm id Data Sheet Agilent Dynamax Analytical HPLC Column Systems 4.6 mm id Data Sheet 1. Description The Agilent Dynamax modular HPLC column system uses a unique technology, Dynamic, to maintain the structure of the packed

More information

ACQUITY QDa Detector Overview and Maintenance Guide

ACQUITY QDa Detector Overview and Maintenance Guide ACQUITY QDa Detector Overview and Maintenance Guide 715003956 Revision B Copyright Waters Corporation 2014 All rights reserved ii 13 November, 2014, 715003956 Rev. B General Information Copyright notice

More information

Instrument Accessories & Parts, Tubing & Connections

Instrument Accessories & Parts, Tubing & Connections Instrument Accessories & Parts, Tubing & Connections ASI Piston Seals for Agilent HPLC Systems Over time the piston seal in every HPLC system becomes worn and must be replaced. Our piston seals are constructed

More information

OFX7 Operating Maintenance and Troubleshooting Manual L-4219

OFX7 Operating Maintenance and Troubleshooting Manual L-4219 OFX7 Operating Maintenance and Troubleshooting Manual L-4219 L-2999 Created 9.2012 User Manual OFX7 L-4219 - Table of Contents - Preface. 4 Customer Service. 4 Modifications to the Product 4 Warranty.

More information

Agilent 1200 Series Manual Injector

Agilent 1200 Series Manual Injector Agilent 1200 Series Manual Injector User Manual 1200 Series MI User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2006, 2008 No part of this manual may be reproduced in any form or by

More information

High Pressure Gradient System. User Manual. Version 1.0 ( )

High Pressure Gradient System. User Manual. Version 1.0 ( ) S 9432 HPLC Pump System High Pressure Gradient System User Manual Version 1.0 (2014-05-26) 2 Table of Contents Contents 1. Introduction 5 1.1 How to Use this Manual 5 1.1.1 Symbols 5 1.2 Safety Information

More information

Dilutor Module for TriPlus RSH

Dilutor Module for TriPlus RSH Thermo Scientific Dilutor Module for TriPlus RSH Instruction Manual Additional section of the TriPlus RSH User Guide and Hardware Manual 31709661 Revision A December 2015 2015 Thermo Fisher Scientific

More information

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS MS Identification of Trace level Impurities from a Non-MS Compatible Mobile Phase Using ACQUITY UPLC System with 2D Technology by Heart-cutting and Online Sample Concentration Bronsky Gopinadh, Dilshad

More information

Hybex Microsample Incubator

Hybex Microsample Incubator www.scigene.com Hybex Microsample Incubator USER MANUAL Cat. #1057-30-0, 1057-30-2 FOR RESEARCH USE ONLY 470 Lakeside Dr, Ste F, Sunnyvale, CA 94085-4720 USA Tel 408-733-7337 Fax 408-733-7336 techserv@scigene.com

More information

BroadBand PowerShield. 20 AHr Battery. User Manual

BroadBand PowerShield. 20 AHr Battery. User Manual BroadBand PowerShield 20 AHr Battery User Manual 990-1316A 10/2004 Chapter 1 General Information The PowerShield provides a power source for broadband telephony applications. Important Safety Instructions

More information

98 동방하이테크상사 T.(02) F.(02)

98 동방하이테크상사   T.(02) F.(02) 98 동방하이테크상사 http://www.allpump.co.kr T.(0)57-69 F.(0)57-69 00--7 troduction to Upchurch Scientific Valves Valves troduction Valves jection and Selection Valves for Pressures to,000 psi (69 bar) Switching

More information

GE Healthcare. Valve INV-917. Instructions

GE Healthcare. Valve INV-917. Instructions GE Healthcare Valve INV-917 Instructions INV-917 2 1 3 7 4 5 6 Important user information All users must read this entire manual to fully understand the safe use of Valve INV-917. Important! Valve INV-917

More information

Constant Flow/Constant Pressure Dual Piston Pump

Constant Flow/Constant Pressure Dual Piston Pump Constant Flow/Constant Pressure Dual Piston Pump Operator s Manual 90-2476 REV U Scientific Systems, Inc. 349 N. Science Park Road State College, PA 16803 www.ssihplc.com Phone: 800-441-4752 Fax: 814-238-7532

More information

Agilent SD-1 Purification System ml/min

Agilent SD-1 Purification System ml/min Agilent SD-1 Purification System - 200 ml/min Introduction Installation of the Capillary Kit This technical note shows you how to install the Stainless Steel Capillary Kit (SD- 1 Purification System -

More information

GlycoChrom TM HPLC Analyzer. Hardware Installation and System Test

GlycoChrom TM HPLC Analyzer. Hardware Installation and System Test GlycoChrom TM HPLC Analyzer Hardware Installation and System Test PIN 410-5004 For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-8 00-424-6723) ,, Table of Contents

More information

F-4600 INLINE ULTRASONIC FLOW METER Installation and Operation Guide

F-4600 INLINE ULTRASONIC FLOW METER Installation and Operation Guide F-4600 INLINE ULTRASONIC FLOW METER Installation and Operation Guide 11451 Belcher Road South, Largo, FL 33773 USA Tel +1 (727) 447-6140 Fax +1 (727) 442-5699 1054-7 / 34405 www.onicon.com sales@onicon.com

More information

GX-271 ASPEC Gilson, Inc. All Rights Reserved. LT

GX-271 ASPEC Gilson, Inc. All Rights Reserved. LT GX-271 ASPEC GX-271 ASPEC World Headquarters Gilson, Inc. 3000 Parmenter Street P.O. Box 620027 Middleton, WI 53562-0027 USA Telephone: 608-836-1551 Fax: 608-831-4451 Gilson S.A.S. 19, avenue des Entrepreneurs

More information

10 Maintenance. Agilent Technologies. This chapter describes the maintenance of the module Infinity Binary Pump User Manual

10 Maintenance. Agilent Technologies. This chapter describes the maintenance of the module Infinity Binary Pump User Manual 1260 Infinity Binary Pump User Manual 10 Maintenance Introduction to Maintenance and Repair 148 Warnings and Cautions 149 Cleaning the Module 151 Overview of Maintenance and Simple Repairs 152 Maintenance

More information

Operators Manual. Recirculating Chiller /06/08

Operators Manual. Recirculating Chiller /06/08 Operators Manual Recirculating Chiller 110-197 11/06/08 Table of Contents Section 1. General Information 1.1 Warranty 1.2 Unpacking 1.3 Package Contents 1.4 Description of the Recirculating Chiller 1.5

More information

ACQUITY UPLC I-Class System Specifications

ACQUITY UPLC I-Class System Specifications ACQUITY UPLC I-Class System s Revision A Copyright Waters Corporation 2011 All rights reserved Copyright notice 2011 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND IN IRELAND. ALL RIGHTS

More information

850 Professional IC. AnCat MCS Manual EN

850 Professional IC. AnCat MCS Manual EN 850 Professional IC AnCat MCS 2.850.3030 Manual 8.850.8052EN Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com 850 Professional IC AnCat

More information

Centrifuge Operator / Service Manual

Centrifuge Operator / Service Manual 3000 Centrifuge Centrifuge Operator / Service Manual cat.# 26230 & 26231 The Q-sep 3000 centrifuge complies with all requirements of UL standard 3101 20, Can/CSA C22.2 No. 1010.1, and Can/CSA C22.2 No.

More information

Agilent 210/218 - Binary Pump

Agilent 210/218 - Binary Pump Agilent 210/218 - Binary Pump Installation of the Stainless Steel Tubing Kit Introduction This technical note shows you how to install the Stainless Steel Tubing Kit for the binary pump of the Agilent

More information

Agilent 1260 Infinity Manual Injector

Agilent 1260 Infinity Manual Injector Agilent 1260 Infinity Manual Injector User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2006-2011, 2012 No part of this manual may be reproduced in any form or by any means (including

More information

DIONEX ELUENT GENERATOR CARTRIDGE DIONEX EGC 500 K 2 CO 3 HIGH PRESSURE ELECTROLYTIC PH MODIFIER (DIONEX EPM 500) Quick Start

DIONEX ELUENT GENERATOR CARTRIDGE DIONEX EGC 500 K 2 CO 3 HIGH PRESSURE ELECTROLYTIC PH MODIFIER (DIONEX EPM 500) Quick Start DIONEX ELUENT GENERATOR CARTRIDGE DIONEX HIGH PRESSURE ELECTROLYTIC PH MODIFIER (DIONEX EPM 500) Quick Start 1. Preparation for Installation of Dionex If installing a Thermo Scientific Dionex and Thermo

More information

Instrumentation of HPLC. HPLC Solvent Pumping Systems

Instrumentation of HPLC. HPLC Solvent Pumping Systems Instrumentation of HPLC HPLC Solvent Pumping Systems i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference

More information

Pack In A Box Dual Piston Pump Operator s Manual

Pack In A Box Dual Piston Pump Operator s Manual Pack In A Box Dual Piston Pump Operator s Manual cat.# 26408 Chromatography Products www.restek.com 1 www.restek.com 1-800-356-1688 1-814-353-1300 Table of Contents 1. Introduction 1.1 Description of the

More information

RIGOL Installation Reference

RIGOL Installation Reference RIGOL Installation Reference L-3000 HPLC System Oct. 2014 RIGOL Technologies, Inc. RIGOL General Safety Summary Please review the following safety precautions carefully before putting the instrument into

More information

ACQUITY UPLC Column Manager - Active and Column Manager - Auxiliary

ACQUITY UPLC Column Manager - Active and Column Manager - Auxiliary ACQUITY UPLC Column Manager - Active and Column Manager - Auxiliary Overview and Maintenance Guide 715005054 Revision A Copyright Waters Corporation 2016 All rights reserved General information Copyright

More information

PUMP SYSTEM OPERATING MANUAL:

PUMP SYSTEM OPERATING MANUAL: OPERATING MANUAL: PUMP SYSTEM The MASTERFLEX C/L Pump Systems are small peristaltic pumps with integral drive motors and are intended for use with a series of tubing sizes that provide flow rates in the

More information

Seeker HL Source Transmitter. User s Guide

Seeker HL Source Transmitter. User s Guide Seeker HL Source Transmitter User s Guide Notice Every effort was made to ensure that the information in this manual was accurate at the time of printing. However, information is subject to change without

More information

FastPrep FP120 Cell Disrupter INSTRUCTION MANUAL

FastPrep FP120 Cell Disrupter INSTRUCTION MANUAL FastPrep FP120 Cell Disrupter INSTRUCTION MANUAL Applies to FP100, FP120, and FP220 Models Qbiogene, Inc. (North America) 2251 Rutherford Road Carlsbad, CA 92008 Phone (800) 424-6101 Fax (760) 918-9313

More information

Deans Switch Microfluidics

Deans Switch Microfluidics TRACE 1300 and TRACE 1310 Gas Chromatographs Deans Switch Microfluidics Installation Guide 31709740 Revision A June 2014 2014 Thermo Fisher Scientific Inc. All rights reserved. TRACE 1300, and TRACE 1310

More information

Waters SQ Detector Operator s Guide

Waters SQ Detector Operator s Guide Waters SQ Detector Operator s Guide Revision F Copyright Waters Corporation 2006 2010 All rights reserved Copyright notice 2006 2010 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND IN IRELAND.

More information

Series I Pump. Operator's Manual rev D

Series I Pump. Operator's Manual rev D Series I Pump Operator's Manual 90-2517 rev D SAFETY SYMBOLS EARTH GROUND CAUTION - REFER TO MANUAL CAUTION HIGH VOLTAGE Scientific Systems, Inc. 349 N Science Park Rd., State College PA 16803 Phone: 800-441-4752

More information

RENA AF371Feeder Operating Manual. Feeder. Operating Manual. Manual Part #: M AF371 Operations Rev

RENA AF371Feeder Operating Manual. Feeder. Operating Manual. Manual Part #: M AF371 Operations Rev Manual Part #: M-3022 Feeder AF371 Operations Rev. 3-16-04 1 RENA AF371 Feeder YOUR RENA AF371 IS DISTRIBUTED BY RENA SYSTEMS INC. SERVICE AND SUPPORT FOR THIS PRODUCT IS PROVIDED BY YOUR RENA DEALER.

More information

SCXI Power Supply/Float Charger Installation Guide

SCXI Power Supply/Float Charger Installation Guide SCXI -1383 Power Supply/Float Charger Installation Guide Conventions This guide describes how to install and use the SCXI-1383 power supply/float charger. The SCXI-1383 is a 13.8 VDC, 4 A power supply/float

More information

nual k Voltage Stabilizer N) manual New Brunswick Voltage Stabilizer Operating manual

nual k Voltage Stabilizer N) manual New Brunswick Voltage Stabilizer Operating manual nual k Voltage Stabilizer N) manual New Brunswick Voltage Stabilizer Operating manual Copyright Copyright 2014 Eppendorf AG, Germany. No part of this publication may be reproduced without the prior permission

More information

Register your instrument! HeatSealer S100. Operating manual

Register your instrument!   HeatSealer S100. Operating manual 00N) manual Register your instrument! www.eppendorf.com/myeppendorf Operating manual Copyright 2015 Eppendorf AG, Germany. All rights reserved, including graphics and images. No part of this publication

More information

PrepStar SD-2 Solvent Delivery Module

PrepStar SD-2 Solvent Delivery Module Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek, CA 94598-1675/USA PrepStar SD-2 Solvent Delivery Module Operation Manual Varian, Inc. 2001 03-914903-00:Rev. 1 Table of Contents Introduction...

More information

Instructions for Use SeQuant CARS Continuous Auto-Regeneration System for SeQuant SAMS Suppressors

Instructions for Use SeQuant CARS Continuous Auto-Regeneration System for SeQuant SAMS Suppressors Instructions for Use SeQuant CARS Continuous Auto-Regeneration System for SeQuant SAMS Suppressors Instructions for Use SeQuant CARS Continuous Auto-Regeneration System CONTENTS General Information...

More information

SAFETY INFORMATION & INSTALLATION GUIDE RMX Programmable Power Supplies. About the RMX Safety Information & Installation Guide

SAFETY INFORMATION & INSTALLATION GUIDE RMX Programmable Power Supplies. About the RMX Safety Information & Installation Guide SAFETY INFORMATION & INSTALLATION GUIDE RMX Programmable Power Supplies RMX-4120/4121/4122/4123/4124/4125/4126/4127 About the RMX Safety Information & Installation Guide This guide is intended for users

More information

Refinery Supply Company, Inc.

Refinery Supply Company, Inc. Refinery Supply Company, Inc. OPERATING INSTRUCTIONS Super Pressure Dead Weight Tester Catalog No. 35260 Serving the Oil & Gas Industry since 1923 9133-A East 46th Street Tulsa, Oklahoma 74145-4823 Voice

More information

Techcon Systems TS7000 Series Interchangeable Material Path Rotary Valve

Techcon Systems TS7000 Series Interchangeable Material Path Rotary Valve Techcon Systems TS7000 Series Interchangeable Material Path Rotary Valve User Guide Copyright OK International CONTENTS Page number 1. Specifications 3 2. Unpacking and inspection 4 3. Description 4 4.

More information

StepWave ion guide: cleaning instructions addendum

StepWave ion guide: cleaning instructions addendum StepWave ion guide: cleaning instructions addendum StepWave ion guide: cleaning instructions addendum This document, an addendum to the following documents, describes an improved cleaning procedure for

More information

Bubble King. User Manual

Bubble King. User Manual Bubble King User Manual TABLE OF CONTENTS 1. Before You Begin... 3 What Is Included... 3 Unpacking Instructions... 3 Claims... 3 Text Conventions... 3 Icons... 3 Document Information... 3 Product at a

More information

LC20 CHROMATOGRAPHY ENCLOSURE OPERATOR S MANUAL Dionex Corporation

LC20 CHROMATOGRAPHY ENCLOSURE OPERATOR S MANUAL Dionex Corporation LC20 CHROMATOGRAPHY ENCLOSURE OPERATOR S MANUAL 1993 Dionex Corporation Document No. 034859 Revision 02 September 1993 1993 Dionex Corporation All rights reserved worldwide Printed in the United States

More information

FITTINGS, LOOPS, AND VALVE ACCESSORIES

FITTINGS, LOOPS, AND VALVE ACCESSORIES FITTINGS, LOOPS, AND VALVE ACCESSORIES Fittings Rheodyne offers a variety of fittings. They differ in material of construction, tubing size, and pressure rating. Materials/Biocompatibility. 316 stainless

More information

Installation Operation Maintenance ADDENDUM

Installation Operation Maintenance ADDENDUM Installation Operation Maintenance ADDENDUM Bulletin #: PS-IOM-HYPO2-0111-A FACTORY SERVICE POLICY If you are experiencing a problem with your Pulsafeeder pump, first review the IOM, and consult the troubleshooting

More information

Operation Manual BOTTLE TOP DISPENSER

Operation Manual BOTTLE TOP DISPENSER Operation Manual BOTTLE TOP DISPENSER TABLE OF CONTENTS Page No. Intended Use Of The Instrument 1 Safety Instructions 1 Functions and Limitations of Use 2 Operating Exclusions 3 Storage Conditions 3 Chemical

More information

Automated Liquid Handling Station

Automated Liquid Handling Station Automated Liquid Handling Station Quick Installation Guide Manual Part Number 32-0406-048 Rev 0 COPYRIGHT 2018 Teledyne Technologies Incorporated. All rights reserved. 32-0406-048 Rev 0, May, 2018 Printed

More information

Paraffin Dispenser Catalog # PD-120, PD-220 Operator s Manual Version 1.5b, Jan 2007

Paraffin Dispenser Catalog # PD-120, PD-220 Operator s Manual Version 1.5b, Jan 2007 Paraffin Dispenser Catalog # PD-120, PD-220 Operator s Manual Version 1.5b, Jan 2007 Be certain to read this manual thoroughly before proceeding with unpacking and installation. TABLE OF CONTENTS Topic

More information

User Guide. Lubricus Lubrication System LUB-D1/LUB-D2/LUB-D3/LUB-D4 (24 VDC)

User Guide. Lubricus Lubrication System LUB-D1/LUB-D2/LUB-D3/LUB-D4 (24 VDC) User Guide Lubricus Lubrication System LUB-D1/LUB-D2/LUB-D3/LUB-D4 (24 VDC) version 04/2013 Content General Information 3 Warning 3 Scope of Supply 3 Overview 3 General safety details 4 Intended use 4

More information

M OTO R. The electric motor. DC - direct current voltage.

M OTO R. The electric motor. DC - direct current voltage. TMD-5 Lubricator Automatic Cyclic Before installing or operating this lubricator, please read this manual carefully. Failure to follow instructions can result in damage to the product and/or serious bodily

More information

NANOPAC-300 & 500 Power Supply. Instruction manual NANOPAC-300 & NANOPAC-500

NANOPAC-300 & 500 Power Supply. Instruction manual NANOPAC-300 & NANOPAC-500 NANOPAC-300 & 500 Power Supply Instruction manual NANOPAC-300 & NANOPAC-500 Version 01C Feb 5th, 2014 1 Packing list NANOPAC-300 or 500-1x NANOPAC-300 Power Supply or NANOPAC-500-1x Power Cord - 1x Instruction

More information

Filters & Frits. IDEX Health & Science introduces Upchurch Scientific precolumn and inline very high pressure (VHP) filters on page 159

Filters & Frits. IDEX Health & Science introduces Upchurch Scientific precolumn and inline very high pressure (VHP) filters on page 159 Using a scale to count Upchurch Scientific frits in Oak Harbor, WA. IDEX Health & Science introduces Upchurch Scientific precolumn and inline very high pressure (VHP) filters on page 159 156 Inlet Solvent

More information

Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by:

Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by: Storage System Ultrasound Probe Cabinet Operator's Manual Manufactured by: CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA 319.248.6757 / 800.445.6741 WWW.CIVCO.COM Copyright 2018 All

More information

AIS and 40 kva 400 V. Site Preparation and Installation Manual

AIS and 40 kva 400 V. Site Preparation and Installation Manual AIS 3100 20 and 40 kva 400 V Site Preparation and Installation Manual www.apc.com AIS 3100 20 and 40 kva 400 V Site Preparation and Installation Manual www.apc.com IMPORTANT SAFETY INSTRUCTIONS SAVE THESE

More information

Kontes Brand ULTRA-WARE Five Valve Recirculation / Filtration Caps Kontes Article Nos Four Valve Filtration / Delivery Caps

Kontes Brand ULTRA-WARE Five Valve Recirculation / Filtration Caps Kontes Article Nos Four Valve Filtration / Delivery Caps Kontes Brand ULTRA-WARE Five Valve Recirculation / Filtration Caps Kontes Article Nos. 953983-0047 953983-6347 Four Valve Filtration / Delivery Caps Kontes Article Nos. 953981-0047 953981-6347 Filter Size:

More information

Lower Polymer Block Cleaning Kit

Lower Polymer Block Cleaning Kit Lower Polymer Block Cleaning Kit For Applied Biosystems 3730/3730xl DNA Analyzers and Applied Biosystems 3130/3130xl Genetic Analyzers Protocol November 15, 2004 2:23 pm, CleanKit_Title.fm Copyright 2004

More information

Operating Manual for. Dispenser. Read this manual thoroughly before operating the Labmax Bottle Top Dispenser!

Operating Manual for. Dispenser. Read this manual thoroughly before operating the Labmax Bottle Top Dispenser! Operating Manual for Dispenser Read this manual thoroughly before operating the Labmax Bottle Top Dispenser! General Safety Precautions When using the Labmax please observe the following safety precautions

More information

c-go 24V/6A 24V/8A 24V/12A

c-go 24V/6A 24V/8A 24V/12A c-go 24V/6A 24V/8A 24V/12A Battery charger GB Instruction manual 1 Index 1. Product description... 2 2. Safety advices... 3 3. Quick start guide... 4 4. Operation... 4 5. Problem solving... 6 6. Specifications...

More information

seal, needle seat, needle assembly, seat cap kit Includes: rotor seal, needle assembly, needle seat 1100 G kit 25271

seal, needle seat, needle assembly, seat cap kit Includes: rotor seal, needle assembly, needle seat 1100 G kit 25271 Restek Replacement Parts for Agilent HPLC Systems Meet or exceed OEM performance. Agilent part # qty. cat.# price Preventive Maintenance Kit Includes: rotor seal, needle seat, needle assembly, seat cap

More information

Bubble Machine. User Manual

Bubble Machine. User Manual Bubble Machine User Manual TABLE OF CONTENTS 1. Before You Begin... 3 What Is Included... 3 Unpacking Instructions... 3 Claims... 3 Text Conventions... 3 Icons... 3 Document Information... 3 Product at

More information

Prospenser. User Manual Mode d emploi Bedienungsanleitung Manual Usuario. For capacities of 2.5 ml 5 ml 10 ml 30 ml and 50 ml

Prospenser. User Manual Mode d emploi Bedienungsanleitung Manual Usuario. For capacities of 2.5 ml 5 ml 10 ml 30 ml and 50 ml Prospenser User Manual Mode d emploi Bedienungsanleitung Manual Usuario For capacities of 2.5 ml 5 ml 10 ml 30 ml and 50 ml Prospenser User Manual...1 Mode d emploi... 11 Bedienungsanleitung... 21 Manual

More information

Agilent 1260 Infinity Quaternary Pump VL

Agilent 1260 Infinity Quaternary Pump VL Agilent 1260 Infinity Quaternary Pump VL User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2010-2012, 2013 No part of this manual may be reproduced in any form or by any means (including

More information

Autosealer Hernon Manufacturing Autosealer

Autosealer Hernon Manufacturing Autosealer Autosealer 2650 110-128 121 Tech Drive Sanford, FL 32771 www.hernon.com Phone: (407) 322-4000 Fax: (407) 321-9700 August 2015 1 P a g e About This Manual Thank you for selecting Hernon's Autosealer 2650

More information

STI OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

STI OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY Extended range HPLC has been investigated for many years; however, progress was delayed because of

More information

Miniature Inert PTFE Isolation Valves. Physical Properties Electrical Performance Characteristics

Miniature Inert PTFE Isolation Valves. Physical Properties Electrical Performance Characteristics 2-Way and 3-Way Solenoid Valves Typical Applications Reagent Control Solvent Management Aggressive Liquid Control Product Specifications Valve Type: Diaphragm Isolation Valve Valve Configuration: 3-Way

More information

Welker Adjustable Probe with Check Valve and Welker Automatic Insertion Probe with Check Valve

Welker Adjustable Probe with Check Valve and Welker Automatic Insertion Probe with Check Valve a Installation, Operation, & Maintenance Manual Welker Adjustable Probe with Check Valve and Welker Automatic Insertion Probe with Check Valve Models AP-3MI & AIP-3MI The information in this manual has

More information

Pressure chlorine changeover unit C 7520

Pressure chlorine changeover unit C 7520 BW 2 24 04 / 1 Content 1. Scope of delivery 2. Device description 3. Installation 4. Operation 5. Shutdown 6. Maintenance 7. Troubleshooting 1 Scope of delivery The chlorine gas changeover unit C 7520

More information

USER MANUAL. Tides La Mer Treatment Table Rev A 3/28/19. *all tables shown with optional accessories and/or finishes

USER MANUAL. Tides La Mer Treatment Table Rev A 3/28/19. *all tables shown with optional accessories and/or finishes USER MANUAL Tides La Mer Treatment Table *all tables shown with optional accessories and/or finishes 1-400-0041 Rev A 3/28/19 Living Earth Crafts 2019, All Rights Reserved Legal Notice The information

More information

Chrom Tech your chromatography specialists. IDEX Injection Valves Manual. IDEX Injection Valves Actuated. 1/4-28 Flat-Bottom Injection Port Adapter

Chrom Tech your chromatography specialists. IDEX Injection Valves Manual. IDEX Injection Valves Actuated. 1/4-28 Flat-Bottom Injection Port Adapter HPLC ACCESSORIES IDEX Injection Valves IDEX Injection Valves Manual IDEX MANUAL INJECTION VALVES DESCRIPTION POS. SENSING SWITCH MAX PSI FITTING HIGH PRESSURE, ANALYTICAL (accepts a 22 gauge needle) FITS

More information

MOTOR. The electric motor. DC - direct current voltage.

MOTOR. The electric motor. DC - direct current voltage. TMD-5 Lubricator Automatic Cyclic Before installing or operating this lubricator, please read this manual carefully. Failure to follow instructions can result in damage to the product and/or serious bodily

More information

Techcon Systems TS5620 Series Diaphragm Valve. User Guide

Techcon Systems TS5620 Series Diaphragm Valve. User Guide Techcon Systems TS5620 Series Diaphragm Valve User Guide CONTENTS Page number 1. Specifications - Horizontal Valve. 3 2. Specifications -Vertical Valve.. 4 3. Unpacking and Inspection 5 4. Description........5

More information

2 Model 826 Dispensing Valve User Guide

2 Model 826 Dispensing Valve User Guide 2 Model 826 Dispensing Valve User Guide About Dymax UV/Visible light-curable adhesives. Systems for light curing, fluid dispensing, and fluid packaging. Dymax manufactures industrial adhesives, light-curable

More information

FlexChem Model 404 Rotating Oven

FlexChem Model 404 Rotating Oven www.scigene.com FlexChem Model 404 Rotating Oven USER MANUAL Cat. #1052-20-1, 1052-20-2, 1052-22-1 & 1052-22-2 FOR RESEARCH USE ONLY Not for Use in Diagnostic Procedures SciGene 306 Potrero Ave, Sunnyvale,

More information

Agilent 210/218 - Isocratic Pump

Agilent 210/218 - Isocratic Pump Agilent 210/218 - Isocratic Pump Installation of the Stainless Steel Tubing Kit Introduction This technical note shows you how to install the Stainless Steel Tubing Kit for the isocratic pump of the Agilent

More information