RIGOL Installation Reference

Size: px
Start display at page:

Download "RIGOL Installation Reference"

Transcription

1 RIGOL Installation Reference L-3000 HPLC System Oct RIGOL Technologies, Inc.

2

3 RIGOL General Safety Summary Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument and any product connected to it. To prevent potential hazards, please use the instrument only specified by this manual. Use Proper Power Cord. Only the power cord designed for the instrument and authorized for use within the local country could be used. Ground the Instrument. The instrument is grounded through the Protective Earth lead of the power cord. To avoid electric shock, it is essential to connect the earth terminal of the power cord to the Protective Earth terminal before connecting any inputs or outputs. Connect the Probe Correctly. If a probe is used, do not connect the ground lead to high voltage since it has isobaric electric potential as the ground. Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and markers on the instrument and check your manual for more information about ratings before connecting the instrument. Use Proper Overvoltage Protection. Make sure that no overvoltage (such as that caused by a thunderstorm) can reach the product, or else the operator might be exposed to the danger of electrical shock. Do Not Operate Without Covers. Do not operate the instrument with covers or panels removed. Do Not Insert Anything Into the Holes of Fan. Do not insert anything into the holes of the fan to avoid damaging the instrument. Use Proper Fuse. Please use the specified fuses. Avoid Circuit or Wire Exposure. Do not touch exposed junctions and components when the unit is powered. Do Not Operate With Suspected Failures. If you suspect damage occurs to the instrument, have it inspected by RIGOL authorized personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel. L-3000 Installation Reference I

4 RIGOL Keep Well Ventilation. Inadequate ventilation may cause an increase of instrument temperature which would cause damage to the instrument. So please keep the instrument well ventilated and inspect the intake and fan regularly. Do Not Operate in Wet Conditions. In order to avoid short circuiting to the interior of the device or electric shock, please do not operate the instrument in a humid environment. Do Not Operate in an Explosive Atmosphere. In order to avoid damage to the device or personal injuries, it is important to operate the device away from an explosive atmosphere. Keep Product Surfaces Clean and Dry. To avoid the influence of dust and/or moisture in the air, please keep the surface of the device clean and dry. Electrostatic Prevention. Operate the instrument in an electrostatic discharge protective environment to avoid damage induced by static discharges. Always ground both the internal and external conductors of cables to release static before making connections. Proper Use of Battery. If a battery is supplied, it must not be exposed to high temperature or in contact with fire. Keep it out of the reach of children. Improper change of battery (note: lithium battery) may cause explosion. Use RIGOL specified battery only. Handling Safety. Please handle with care during transportation to avoid damage to buttons, knob interfaces and other parts on the panels. II L-3000 Installation Reference

5 Content RIGOL Content General Safety Summary... I Chapter 1 Site preparation Chapter 2 Unpacking Overall Appearance checking Unpacking steps Chapter 3 Stack the Modules L3500/L3520 Detector L3400 Column Oven L3320 Autosampler L3200 series Pump L3100 series Solvent Organizer Chapter 4 Tube connection Preparation Tube connection steps Tube connection from the solvent bottle to inlet check valve Tube connection of autosampler Tube connections from column oven to flow cell in detector Tubeconnections from flow cell outlet to the waste Chapter 5 Power Requirement... 错误! 未定义书签 Chapter 6 Connecting power cable Chapter 7 Clarity installation Chapter 8 Rinse the system L-3000 Installation Reference III

6

7 Chapter 1 Site preparation RIGOL Chapter 1 Site preparation Prepare installation site in accordance to RIGOL L-3000 Pre-Installation Guide. L-3000 Installation Reference 1-1

8

9 Chapter 2 Unpacking RIGOL Chapter 2 Unpacking Overall Appearance checking Please complete the following procedure before installation. 1. Make sure all the boxes were delivered in accordance to your order. Note: A label on the box indicates the model, serial number and manufacturing date. Solvent bottle, column and other accessories are packed in those boxes. 2. Check the packages carefully, ensure no apparent damage was observed. You shall inform the carrier and fill out a claim if anything abnormal found. Unpacking steps 1. Shear the bandages for protection on the outer packing and open the boxes. Note: Keep the packing materials properly for future transportation purpose. 2. Take out all accessories, check each item in the box against to the packing list, contact RIGOL in case of anything missing. 3. Take out the instrument and remove the protection foam. Note: The way in which the protection foam and accessories are placed in the box varies for each module, pay attention to those parts and don t miss any of them. 4. Take out the instrument carefully and place it on the workbench. Note: The module could be too heavy and sized to handle by single person, team work is essential to avoid injury. 5. Remove the fixing tape on the module, check the appearance. Note: Inform RIGOL and do not connect the instruments when some damages are found. 6. Self-check: connect the power cable of the module, turn it on and check the checking result. L-3000 Installation Reference 2-1

10

11 Chapter 3 Stack the Modules RIGOL Chapter 3 Stack the Modules Unpacking the boxes in following order: L3500 UV-Vis detector/l3520 DAD, L3400 column oven, L3320 autosampler, L3210/L3220/L3240/L3225/L3245 high pressure pump, L3100/L3120/L3140 solvent organizer. The number and variety of the module may differ with different system configurations, but should remain the installation sequence. L-3500/L-3520 Detector 1. According to the requirement of Unpacking Procedure, complete the inspection and unpacking. 2. The two transportation screws on the bottom of L3500/L3520 detector depicted in following pictures are intended for transportation purpose, and must be removed before installation. You can find them easily at the bottom plate of detector by the guidance of red arrow beside them. Note: Save those transportation screws properly for future usage. These two screws are larger than T20, remove them with proper driver or hex key. Plate 1 Remove the transportation screws before installation 3. Put the module on the workbench carefully, connect the rubber waste pipe to the drain outlet at the lower right corner of the front panel (under the leak collection tray). Note: It is import to connect the waste pipe now because the waste outlet is hardly accessible after the whole stack was settled. Plate 2 Connect drain pipe L-3000 Installation Reference 3-1

12 RIGOL Chapter 3 Stack the Modules L-3400 Column Oven 1. According to the requirement of Unpacking Procedure, complete the inspection and unpacking. 2. Put L-3400 column oven on the detector. The supporting leg under the column oven can be put in the dents above the detector, which made the stack relative stable. This works for each module. Note: L3400 does not have transportation screws. L-3320 Autosampler 1. According to the requirement of Unpacking Procedure, complete the inspection and unpacking. 2. Put L-3320 autosampler on the column oven. Note: L-3320 does not have transportation screws. Note: If the system is equipped with manual injection valve (7725i) instead of autosampler, it should be fixed on the right of the pump, you can see three screw holes on the pump reserved for this purpose. In this case, pump should be unpacked before manual valve. Plate 3 Holes reserved for manual valve L-3200 series Pump 1. According to the requirement of Unpacking Procedure, complete the inspection and unpacking. 2. Put L3200 pump on the L3320 auto sampler. Note: There are two transportation screws for L-3210/L-3240/L-3245 while four are there for L-3220/L-3225 (two for each pump body), remove them before stacking with a T20 screw driver. Note: Save those transportation screws properly for future usage. 3-2 L-3000 Installation Reference

13 Chapter 3 Stack the Modules RIGOL L-3210/3240/3245 L-3220/3225 Plate 4 Transportation screws on L-3210/3240/3245 and L-3220/3225 L-3100 series Solvent Organizer 1. According to the requirement of Unpacking Standard, complete the inspection and unpacking. 2. Put L-3100 solvent organizer on L-3200 pump. 3. Put the solvent bottle (s) in the tray of solvent organizer. Note: L-3100 does not have transportation screws. The ready set will be like this: L-3100 Solvent Organizer L-3200 Pump L-3320 Autosampler L-3400 Column Oven L-3500 UV-Vis Detector Plate 5 All the modules in stack L-3000 Installation Reference 3-3

14

15 Chapter 4 Tube connection RIGOL Chapter 4 Tube connection Preparation Note: Put on gloves before starting. A. Remove all the adhesive tapes (intended for transportation) on the modules before starting. B. The front door of each module was connected to the housing by magnets on the left and can be removed. In case to do this, open the door from the right, lift slightly and pull to disconnect from the left. Plate 6 Open the front door, lift and pull to disconnect it C. Tubes on the pump was already fixed, other tubes between different modules, solvent bottle and the waste require user s work. These tubes are pre-cut, shaped and can be used directly. Find these tubes in the accessory boxes for each module. Note: Do not over-tighten the nuts. For stainless steel screws, tighten it first by hand. The screw should go into the connector smoothly, some resistance can be caused by improper angle between screw and connector, remove it and try again in this case. When the screw reach the end and cannot be turned by finger anymore, tighten it no more than 90 degrees with a wrench. Usually 90 degrees is enough, sometimes a little more is required to stop leaking. But NEVER THIGHTEN THE STAINLESS STEEL NUT MORE THAN 180 DEGREES WITH WRENCH. Plate 7 Angle to tighten the stainless steel nuts. Green: Safe; Yellow: Be careful; Red: Max. L-3000 Installation Reference 4-1

16 RIGOL Chapter 4 Tube connection D. Connect all tubes and column as the following guidance service video, and following the steps bellow: Note: The video is narrated in Chinese but we are sure you can understand it. 10% isopropyl alcohol solution Solvent bottle L-3120 Solvent Organizer Solvent with good solubility for the sample. Usually 50% methanol aqueous solution. L-3220Pump Seal wash waste tube L-3320 Autosampler Purge waste tube Autosampler washing waste tube L-3400 Column oven Autosampler needle wash waste tube L-3500 UV-Vis Detector Waste bottle Figure 1 The sketch map of connecting tubes for binary system with autosampler 4-2 L-3000 Installation Reference

17 Chapter 4 Tube connection RIGOL 10% isopropyl alcohol solution Solvent bottle L-3120 Solvent Organizer L-3220Pump L-3400 Column oven Purge waste tube L-3500 UV-Vis Detector The waste bottle Figure 2 The sketch map of connecting tubes for binary system with manual injection valve Connected with the input on the column Connected with the output at the right side of purge valve The waste Figure 3 Partial enlarged drawing of connecting tubes for manual injection valve observed from the back side. L-3000 Installation Reference 4-3

18 RIGOL Chapter 4 Tube connection 10% isopropyl alcohol solution Solvent bottle Pure organic solvent used in the mobile phase L-3245 Pump L-3320 Autosampler Purge waste tube Seal wash waste tube Autosampler washing waste tube L-3400 Column oven Autosampler needle wash waste tube L-3500 UV-Vis Detector The waste bottle Figure 4 The sketch map of connecting tubes for quaternary system with autosampler Tube connection steps Tube connection from the solvent bottle to inlet check valve A. Binary high-pressure gradient pump: 1. Remove the front door of solvent organizer and pump. 2. Make the transparent FEP tubes through the bottle caps and connect them to the frits (metal filter), connect the other side with PEEK nuts to online degasser in L3100 solvent organizer. 3. Use the shorter FEP tubes (which have PEEK nuts on both end) to connect online degasser with inlet check valves under the right pump head. Note: Use transparent FEP tubes to connect solvent bottles and inlet check valves under the right pump head directly if the system not configured with online degasser. Note: The blank nuts used for sealing the system should be kept properly for future usage. Note: Pay attention to channel tag on the tube and the label on pump. For binary system, the pump on the right side was designated as pump A and should be connected to tubes with tag A on them. 4-4 L-3000 Installation Reference

19 Chapter 4 Tube connection RIGOL B. Quaternary low-pressure gradient pump 1. Remove the front door of solvent organizer and pump. 2. Make the transparent FEP tubes through the bottle caps and connect them to the frits (metal filter), connect the other side with PEEK nuts to online degasser in L-3140 solvent organizer. Note: For system with L-3245 quaternary pump, online degasser is integrated in the pump module, in this case L-3140 is not necessarily exit. 3. Use the shorter FEP tubes (which have PEEK nuts on both end) to connect online degasser and four channel proportion valve. Note: These tubes were pre-install for L-3245 quaternary pump, not need for user s work. Note: Pay attention to channel tag on the tube and the label on pump. For quaternary pump, channel label can be found on four channel proportion valve. C. Seal wash tube and purge drain tube 1. Take out three transparent FEP tubes which are different in length, use them properly. 2. Use these three tubes to connect washing bottle to peristaltic pump, pump, A pump and washing bottle respectively. 3. Connect the FEP tube to the outlet port of purge valve (below the valve), the other end should be plug into a suitable reservoir for waste. Note: RIGOL recommend aqueous solution with 10% isopropanol as seal wash solvent, due to its good solubility to salt. The solution should be freshly made and replaced every day. Tube connection of autosampler 1. Remove the front door of autosampler. 2. Connect purge valve outlet (on the right of the purge valve, where an online filter can be found) to Port 3 on six ports valve of autosampler with a pre-shaped stainless steel tube. Note: If manual injection valve was used instead of autosampler, connect purge valve outlet to Port 2 on manual injection valve. 3. Connect Port 2 on six port valve to the connector on column oven. Note: Port 3 should be connected to connector on column oven in case of manual injection valve was used. 4. Put out washing tube in L3320 autosampler accessory kit, connect the PEEK nut end to washing port on four ways valve in autosampler, and connet the other end with the frit, put it in adequate washing buffer (Normally use methanol for the first time due its ability to remove bubbles. Solvent with good solubility to sample should be used in future usage. 50% methanol aqueous solution is a good choice if you are not sure about the solubility). 5. Find the waste tube in L3320 autosampler accessory kit (it look exactly like the washing tube), connect the PEEK end to the right port on four ways valve, let the other end go into the waste tank. 6. Connect the bold rubber pipe to the right waste port on the autosampler, put the other end into waste. L-3000 Installation Reference 4-5

20 RIGOL Chapter 4 Tube connection Tube connections from column oven to flow cell in detector 1. Install column. Note: Check the flow direction, mobile phase go through the column oven from left to right, columns should be installed in accordance with this flow direction. Note: The highest sustainable pressure of PEEK nut on the left side of chromatographic column is 2000 psi, you should use stainless steel fitting(in the accessory kit) instead of PEEK one when the system pressure is higher than the limit. Note: The system cannot work steadily except that the front door of column oven is closed tightly. Pay more attention to the groove on the right side of the front door, the tube inlet and outlet of the column oven must be compatible with it, otherwise the alarm will be triggered once the door not closed tightly. 2. Connect column outlet and flow cell inlet in detector with the pre-cut PEEK tube in accessory kit. Note: Elutes go into flow cell from the lower inlet, the higher one is outlet. Tube connections from flow cell outlet to the waste Connect the white outlet tube to the outlet port of flow cell (upper one), the tube is about 80cm long (0.17mm inner diameter) to provide essential backpressure for the cell. 4-6 L-3000 Installation Reference

21 错误! 未找到引用源 错误! 未找到引用源 RIGOL Chapter 5 Communication Cables L-3000 HPLC modules (except for L3320 autosampler) communicate with each other by CAN interface. The whole system (except for L3320 autosampler) communicates with workstation (computer) by RS232 cable. The detector act as a hub in the system, all other modules should be connected in series to detector with CAN cable, and detector was connected to computer by RS232 cable. L-3320 communicate with workstation (computer) separately. Connecting cable steps: 1. Make sure the power switch is on OFF mode. 2. Connect all modules (without autosampler) in series by CAN cables (with the round ends). 3. Connect all modules to power supply by power cords. 4. Connect the system and computer by RS-232 cable. The following steps are for autosampler: 5. Connect autosampler and computer with USB cable supplied in accessory kit. L-3000 Installation Reference 5-1

22 RIGOL 错误! 未找到引用源 错误! 未找到引用源 Those blue lines represent CAN cable with round connectors on both side L3120 This orange line represents USB cable with two different connectors L3220 USB interface on computer L3320 RS232 interface on computer L3400 L3500 This purple line represents color RS232 cable CAN interface USB RS232 interface Figure 4 Communication cable settings for system with autosampler 5-2 L-3000 Installation Reference

23 错误! 未找到引用源 错误! 未找到引用源 RIGOL Those blue lines represent CAN cable with round connectors on both side L3120 L3220 RS232 interface on computer L3400 L3500 This purple line represents RS232 cable This green line represents black color trigger cable with two different connectors Trigger cable interface Figure 6 Communication cable schematic for system with manual injection valve L-3000 Installation Reference 5-3

24 RIGOL 错误! 未找到引用源 错误! 未找到引用源 This orange line represents USB cable with two different connectors L3245 Those blue lines represent CAN cable with round connectors on both side USB interface on computer L3320 RS232 interface on computer L3400 L3500 This purple line represents RS232 cable Figure 7 Communication cable schema for quaternary system 5-4 L-3000 Installation Reference

25 错误! 未找到引用源 错误! 未找到引用源 RIGOL Chapter 6 Connecting power cable 1. Put all the power switch off. Note: The power switch of L3100 and L3400 is near the power cable at the rear side of the module, other modules power switches are at the upper left corner of the front panel. 2. Connect one side of the power cord to the power strip at the rear position of the module, another side is connected to the power strip. Note: RIGOL recommends that all modules are connected to the same power strip, to ensure the same power supply. 3. Put all the power switches on, and ensure the self-inspection are successfully operated. Note: Please contact us if the self-inspection failed. 4. Put all the power switches off, tidy up the power cable at the rear position of the modules, and keep them clean. L-3000 Installation Reference 6-1

26

27 错误! 未找到引用源 错误! 未找到引用源 RIGOL Chapter 7 Clarity installation Note: Install RIGOL CWS before connecting any device and dongle. GET YOUR COMPUTER READY Before installing workstation, make sure your computer comply with the Pre-Installation Guide, and has the following software installed: Microsoft windows XP SP3 or above, or Windows7. Microsoft.Net Framework 3.5 or above. NI-VISA or above. You can download NI-VISA here: Note: NI-VISA must be installed before RIGOL CWS/Clarity, or the L3320 autosampler driver will be unavailable. INSTALL WORKSTATION Install Clarity with the CD enclosed in the package. 1. Install Clarity before connecting any device. 2. Connect dongle key to USB port, the driver should be installed automatically. 3. Start Clarity, complete configuration process. 4. Log into Clarity and check the connection, if everything goes well, you can use RIGOL HPLC system now. L-3000 Installation Reference 7-1

28

29 Chapter 8 Rinse the system RIGOL Chapter 8 Rinse the system Get the solvent ready, connect a syringe at the end of purge waste tube, open purge valve a little bit and draw to clean all the bubbles in each channel. Note: For binary pump, the solvent in the bottle should go slowly into the system and rinse the air out for the two channels simultaneously. If this doesn t happen, there could be something wrong with the check valve, an ultra-sonic bath for the check valve will solve this problem. Note: If a quaternary pump is installed, solvent will not go through the tube because the proportion valve is closed by default. In this case the system must be wired up and communicates with workstation, set a flow for each channel (so the pump will open the corresponding channel in proportional valve) and then draw, clean all four channels one by one. The following steps can only be done with the help of workstation: 1. Rinse all channels one by one with methanol at flow rate 3mL/min. 2. Do initial washing for autosampler. Get ready for use Now you can purge the system, equilibrate the column and start you chromatography experience with RIGOL L-3000 HPLC. L-3000 Installation Reference 8-1

30 RIGOL Chapter 8 Rinse the system Reference Hardware Requirements Type Specification Comments Weight 6.6kg (L3100); 8kg (L3120/L3240) 17kg(L3210/L3240); 22kg (L3220);19kg (L3245) 17kg 7.6kg L3100 series L3200 series L3320 L3400 L kg Dimensions(H W D) 90 x 370 x 464 mm 143 x 367 x 462 mm; 244 x 359 x 453 mm (L3245) 244 x 359 x 453 mm 90 x 370 x 464 mm L3100 series L3200 series L3320 L3400 L x 367 x 462 mm Power Plug Single-phase three-hole socket L-3000 system Line Voltage Line Frequency Power Consumption Temperature Humidity Solvents Solvent Waste Container VAC L-3000 system 50 60HZ L-3000 system 60 Wmax 150 Wmax 110 Wmax 160 Wmax 90 Wmax % to 80% RH Clean/HPLC grade solvents Approved and labeled L3100 series L3200 series L3320 L3400 L3500 L3100 series, L3200 series L3100 series, L3200 series Source and sample waste 8-2 L-3000 Installation Reference

High Pressure Gradient System. User Manual. Version 1.0 ( )

High Pressure Gradient System. User Manual. Version 1.0 ( ) S 9432 HPLC Pump System High Pressure Gradient System User Manual Version 1.0 (2014-05-26) 2 Table of Contents Contents 1. Introduction 5 1.1 How to Use this Manual 5 1.1.1 Symbols 5 1.2 Safety Information

More information

XLR-8 60 Extended Rack Autosampler

XLR-8 60 Extended Rack Autosampler XLR-8 60 Extended Rack Autosampler Quick Installation Guide Manual Part Number 480226 Rev 0 COPYRIGHT 2015 Teledyne Technologies Incorporated. All rights reserved. 480226 Rev 0, October, 2015 Printed in

More information

Agilent SD-1 Purification System ml/min

Agilent SD-1 Purification System ml/min Agilent SD-1 Purification System - 200 ml/min Introduction Installation of the Capillary Kit This technical note shows you how to install the Stainless Steel Capillary Kit (SD- 1 Purification System -

More information

HPLC Model 210D. UV detector lamp switch. Rheodyne 7725i injection valve. Prime/purge valve. Column. compartment. thumbscrews. Storage.

HPLC Model 210D. UV detector lamp switch. Rheodyne 7725i injection valve. Prime/purge valve. Column. compartment. thumbscrews. Storage. 1. Overview The front panel houses the UV detector lamp switch, Rheodyne injection valve, prime/purge valve, pressure gauge, and the storage compartment. In the storage compartment are the LED display

More information

GlycoChrom TM HPLC Analyzer. Hardware Installation and System Test

GlycoChrom TM HPLC Analyzer. Hardware Installation and System Test GlycoChrom TM HPLC Analyzer Hardware Installation and System Test PIN 410-5004 For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-8 00-424-6723) ,, Table of Contents

More information

Automated Liquid Handling Station

Automated Liquid Handling Station Automated Liquid Handling Station Quick Installation Guide Manual Part Number 32-0406-048 Rev 0 COPYRIGHT 2018 Teledyne Technologies Incorporated. All rights reserved. 32-0406-048 Rev 0, May, 2018 Printed

More information

RIGOL. Service Guide. L-3200 Series Pump. Nov RIGOL Technologies, Inc.

RIGOL. Service Guide. L-3200 Series Pump. Nov RIGOL Technologies, Inc. RIGOL Service Guide L-3200 Series Pump Nov. 2012 RIGOL Technologies, Inc. RIGOL Guaranty and Declaration Copyright 2012 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL is a registered

More information

Agilent 1200 Series Nano Pump

Agilent 1200 Series Nano Pump Agilent 1200 Series Nano Pump Service Manual Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

TECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure

TECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure TECHNICAL REPORT Modifying Agilent 1100 HPLC Systems to Achieve UHPLC-like performance with HALO Fused-Core Columns Agilent 1100 HPLC system With a few modifications, an Agilent 1100 HPLC can produce UHPLC-like

More information

Agilent 1100 Series Capillary LC System. System Manual

Agilent 1100 Series Capillary LC System. System Manual Agilent 1100 Series Capillary LC System System Manual s1 Notices Agilent Technologies, Inc. 2002 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval

More information

Switching DC Power Supply

Switching DC Power Supply 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Model 1693, 1694 Switching DC Power Supply INSTRUCTION MANUAL 1 Safety Summary

More information

System Manual. Agilent 1260 Infinity Capillary LC System

System Manual. Agilent 1260 Infinity Capillary LC System System Manual Agilent 1260 Infinity Capillary LC System Notices Agilent Technologies, Inc. 2010, 2012 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

Application Note. General Use Inlet Solvent Filters. Inlet Solvent Filters. Filters and Degassers. New!

Application Note. General Use Inlet Solvent Filters. Inlet Solvent Filters. Filters and Degassers. New! Inlet Solvent Filters Filters and Degassers General Use Inlet Solvent Filters Large Surface Areas disposable µm, 0 µm, and 0 µm Pore Sizes Available General Use and Prep Filters for Higher Flow Applications

More information

Agilent 1260 Infinity Quaternary Pump VL

Agilent 1260 Infinity Quaternary Pump VL Agilent 1260 Infinity Quaternary Pump VL User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2010-2012, 2013 No part of this manual may be reproduced in any form or by any means (including

More information

Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D

Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Instructions Agilent Valve Kit G4231A/C and G4232C/D - Instructions This technical note describes the installation and application of the Agilent

More information

Agilent InfinityLab Quick Change Valve G4237A

Agilent InfinityLab Quick Change Valve G4237A Agilent InfinityLab Quick Change Valve G4237A Instructions Agilent G4237A Valve Kit Instructions This technical note describes the installation and application of the Agilent InfinityLab Quick Change 4

More information

HOYMILES MICRO-IVERTER MI-250

HOYMILES MICRO-IVERTER MI-250 HOYMILES MICRO-IVERTER MI-250 TECHNICAL MANUAL CONTENTS INTRODUCTION... 3 SAFETY... 4 SYMBOL ILLUSTRATION... 4 INSTALLATION WARNINGS... 5 PREPARE FOR INSTALLING... 6 TRANSPORT AND INSPECT... 6 CHECK INSTALLATION

More information

G1160A Agilent 1100 Series Multiple Purpose Switching Valve

G1160A Agilent 1100 Series Multiple Purpose Switching Valve G1160A Agilent 1100 Series Multiple Purpose Switching Valve Description: The G1160A Agilent 1100 Series Multiple Purpose Switching Valve is a stainless steel 12-position, 13-port motorized valve operated

More information

Technical Information Manual

Technical Information Manual Technical Information Manual Revision n. 4 18 June 2008 MOD. A3000NF 3-PHASE NOTCH FILTER MANUAL REV. 4 NPO: 00120/04:3000N.MUTx/04 CAEN will repair or replace any product within the guarantee period if

More information

USER MANUAL MPS 1-5KW SERIES

USER MANUAL MPS 1-5KW SERIES USER MANUAL MPS 1-5KW SERIES Low Series Frequency Solar and Wind Power Charging Inverter CONTENTS 1 WWW.NEXTGENNRG.COM 1. Introduction... 3 2. Profile Structure 4 2.1. Working Principle 4 2.2.Product Features..5

More information

LC20 CHROMATOGRAPHY ENCLOSURE OPERATOR S MANUAL Dionex Corporation

LC20 CHROMATOGRAPHY ENCLOSURE OPERATOR S MANUAL Dionex Corporation LC20 CHROMATOGRAPHY ENCLOSURE OPERATOR S MANUAL 1993 Dionex Corporation Document No. 034859 Revision 02 September 1993 1993 Dionex Corporation All rights reserved worldwide Printed in the United States

More information

HPLC System S 600 Series

HPLC System S 600 Series HPLC System S 600 Series The Sykam HPLC System S 600 Series is intended for the ambitious analyst. The system configuration is highly variable and several upgrade options make this system suitable for

More information

Agilent G5632 Valve Kit Instructions

Agilent G5632 Valve Kit Instructions Agilent G5632 Valve Kit Instructions Agilent G5632 Valve Kit Instructions Contents Typical Applications 2 Alternating Column Regeneration (2 position/10 port valves only) 2 Dual column selection 3 Sample

More information

Agilent 210/218 - Isocratic Pump

Agilent 210/218 - Isocratic Pump Agilent 210/218 - Isocratic Pump Installation of the Stainless Steel Tubing Kit Introduction This technical note shows you how to install the Stainless Steel Tubing Kit for the isocratic pump of the Agilent

More information

ÄKTA system Training Guide

ÄKTA system Training Guide ÄKTA system Training Guide Purifier 00 / 0 amersham bioscience amersham bioscience amersham bioscience amersham bioscience ÄKTApurifier Buffers Injector Box Monitor Conductivity ph On-line Filter UV Detector

More information

1260 Infinity II Degasser User Manual. 5 Maintenance

1260 Infinity II Degasser User Manual. 5 Maintenance 1260 Infinity II Degasser User Manual 5 Maintenance Warnings and Cautions 40 Simple Repairs - Maintenance 42 Clean the Instrument 43 Assemble the Main Cover 44 Check and Change the Power Fuses 45 This

More information

Agilent G4239C Valve Kit - Instructions

Agilent G4239C Valve Kit - Instructions Agilent G4239C Valve Kit - Instructions Agilent G4239C Valve Kit - Instructions Technical Information about the Agilent Valve Kit G4239C. Contents Typical Applications 2 Multicolumn Selection 2 Method

More information

Vanquish. Pumps VH-P10. Operating Manual

Vanquish. Pumps VH-P10. Operating Manual Vanquish Pumps VH-P10 Operating Manual 4820.4401-EN Revision 2.0 November 2015 Copyright 2015 Thermo Fisher Scientific Inc. All rights reserved. Trademarks MP35N is a registered trademark of SPS Technologies.

More information

Combustion ion chromatography with a Dionex Integrion HPIC System

Combustion ion chromatography with a Dionex Integrion HPIC System TECHNICAL NOTE Combustion ion chromatography with a Dionex Integrion HPIC System Authors Daniel Khor,1 Jeff Rohrer,1 Adelon Austin,2 Mark Manahan2 1 Thermo Fisher Scientific, Sunnyvale, CA, USA; 2COSA

More information

GX-271 ASPEC Gilson, Inc. All Rights Reserved. LT

GX-271 ASPEC Gilson, Inc. All Rights Reserved. LT GX-271 ASPEC GX-271 ASPEC World Headquarters Gilson, Inc. 3000 Parmenter Street P.O. Box 620027 Middleton, WI 53562-0027 USA Telephone: 608-836-1551 Fax: 608-831-4451 Gilson S.A.S. 19, avenue des Entrepreneurs

More information

Agilent 218 Purification Solution

Agilent 218 Purification Solution Agilent 218 Purification Solution System User Guide Purification Solution System User Guide Agilent Technologies Notices Agilent Technologies, Inc. 2012, 2013 No part of this manual may be reproduced in

More information

10 Maintenance. Agilent Technologies. This chapter describes the maintenance of the module Infinity Binary Pump User Manual

10 Maintenance. Agilent Technologies. This chapter describes the maintenance of the module Infinity Binary Pump User Manual 1260 Infinity Binary Pump User Manual 10 Maintenance Introduction to Maintenance and Repair 148 Warnings and Cautions 149 Cleaning the Module 151 Overview of Maintenance and Simple Repairs 152 Maintenance

More information

Agilent 1290 Infinity Quaternary Pump

Agilent 1290 Infinity Quaternary Pump Agilent 1290 Infinity Quaternary Pump User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2012-2015, 2016 No part of this manual may be reproduced in any form or by any means (including

More information

260 Solvent Vapor Exit Accessory

260 Solvent Vapor Exit Accessory Theory of operation The Solvent Vapor Exit (SVE) is a GC accessory for performing large volume injections with a Cool On-Column inlet (COC). At the start of the run, the SVE solenoid valve is open as the

More information

Please exercise care when using solvents close to the instrument. It could damage the vapor sensor present in the ELSD.

Please exercise care when using solvents close to the instrument. It could damage the vapor sensor present in the ELSD. Infinity II - Standard Preventive Maintenance Service Agilent Preventive Maintenance (PM) provides factory recommended service for your systems to assure reliable operation and the accuracy of your results.

More information

S 1125 HPLC Pump System

S 1125 HPLC Pump System Routine S 170 EXPERT HPLC HPLC System SYSTEMS 500 The Sykam ROUTINE HPLC System S 500 is intended for all routine analysis. The system configuration is highly variable and several upgrade options make

More information

HIGH PRESSURE SWITCHING VALVES

HIGH PRESSURE SWITCHING VALVES HIGH PRESSURE SWITCHING VALVES Rheodyne offers high pressure switching valves to simplify procedures and improve the speed, resolution, and sensitivity of HPLC analysis. The switching valves are available

More information

TBI /2012 TRAUMATIC BRAIN INJURY DEVICE

TBI /2012 TRAUMATIC BRAIN INJURY DEVICE USER MANUAL TBI 0310 6/2012 TRAUMATIC BRAIN INJURY DEVICE Page 1 of 26 Setting up the TBI 0310 Head Impactor The TBI 0310 Head Impactor when fully assembled has the following components: 1. Control box

More information

MULTIVOLTAGE PORTABLE BATTERY CHARGER MVM

MULTIVOLTAGE PORTABLE BATTERY CHARGER MVM _ M MULTIVOLTAGE PORTABLE BATTERY CHARGER MVM User's MANUAL Code: MVM Version: 01-BF Date: OCT 2005 Page 1/10 _ 1. INTRODUCTION Before starting to use your Energic plus MVM battery charger, please take

More information

NEW HPLC - SYSTEM LINE

NEW HPLC - SYSTEM LINE NEW HPLC - SYSTEM LINE S c h a m b e c k S F D G m b H Drieschweg 13A D-53604 Bad Honnef/Germany Phone: +49 2224 9239-0 E-Mail: info@schambeck-sfd.com Internet: www.schambeck-sfd.com S 9425 HPLC solvent

More information

- Keep children and other people away while operating the rivet tool. Distractions can cause you to lose control of the tool.

- Keep children and other people away while operating the rivet tool. Distractions can cause you to lose control of the tool. PLEASE SPEND 5 MINUTES READING THESE INSTRUCTIONS BEFORE USING YOUR NEW BATTERY RIVET TOOL. TRUST US, IT WILL SAVE YOU TIME AND INCONVENIENCE IN THE LONG RUN. Please read the following safety information

More information

Installation Note for Solvent Selection Valve Upgrade Kit (G1360A)

Installation Note for Solvent Selection Valve Upgrade Kit (G1360A) s1 Installation Note for Solvent Selection Valve Upgrade Kit (G1360A) This note describes how to install the solvent selection valve upgrade kit into an Agilent 1100 Series binary pump. General Information

More information

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors Model 8610C Gas Chromatograph Mounts up to Six Detectors and Five Injectors Ambient to 400 C Temperature Programmable Column Ovens Dimensions: 19 wide x 13.5 high x 14.5 deep Implement virtually any EPA

More information

XS400 Load Bank. Read all instructions before using the load bank. Contents

XS400 Load Bank. Read all instructions before using the load bank. Contents Read all instructions before using the load bank Contents 1) Components... 3 Total Assembly... 3 2) Specifications... 4 3) Receiving... 4 4) Safety... 5 a) Grounding cam... 6 b) Power connections... 6

More information

P680 HPLC Pumps. Operating Instructions. Revision: 2.0 Date: April Dionex. Doc.: P680_OI_E_V2_0.doc

P680 HPLC Pumps. Operating Instructions. Revision: 2.0 Date: April Dionex. Doc.: P680_OI_E_V2_0.doc P680 HPLC Pumps Operating Instructions Revision: 2.0 Date: April 2004 2004 Dionex Doc.: P680_OI_E_V2_0.doc Declaration of Conformity Product: Type: HPLC Pump P680 Dionex GmbH herewith declares conformity

More information

Maintaining Your Agilent 1100 Series HPLC System

Maintaining Your Agilent 1100 Series HPLC System Operate your lab at peak performance Troubleshoot difficult problems Select the right Agilent supplies For detailed information about supplies from Agilent, please contact your local Agilent sales representative,

More information

PATROL UPLC Process Analysis System for Production

PATROL UPLC Process Analysis System for Production PATROL UPLC Process Analysis System for Production The PATROL UPLC Process Analysis System for Production is an integrated system solution designed and engineered to perform online and atline reaction

More information

Agilent 210/218 - Binary Pump

Agilent 210/218 - Binary Pump Agilent 210/218 - Binary Pump Installation of the Stainless Steel Tubing Kit Introduction This technical note shows you how to install the Stainless Steel Tubing Kit for the binary pump of the Agilent

More information

Asepsis 21 Delivery Unit

Asepsis 21 Delivery Unit INSTALLATION INSTRUCTIONS for the Console Mounted TM Asepsis 21 Delivery Unit SECTION I - REQUIREMENTS 1. PHYSICAL REQUIREMENTS... 1 2. ELECTRICAL REQUIREMENTS... 1 3. WATER SUPPLY REQUIREMENTS... 1 4.

More information

INSTRUCTION MANUAL ANGLE GRINDER PT W

INSTRUCTION MANUAL ANGLE GRINDER PT W INSTRUCTION MANUAL ANGLE GRINDER PT50360 4½ INCHES 120V 60Hz 600W 5A 12,000 rpm C US Note : Before operating this tool, read this manual and follow all safety rules and operating instructions. This electric

More information

PRODUCT OPERATING MANUAL

PRODUCT OPERATING MANUAL PRODUCT OPERATING MANUAL PANBLAST TM CS37 SUCTION BLAST CABINET Manual Number: ZVP PC 0069 00 SECTION 1. GENERAL INFORMATION 2. ASSEMBLY & INSTALLATION INSTRUCTIONS 3. OPERATING INSTRUCTIONS 4. MAINTENANCE

More information

Liquid Chromatography

Liquid Chromatography SCINTOMICS Liquid Chromatography Always the right system for your needs. SCINTOMICS: Always the right system for your needs Advanced Research Daily QC Routine and beyond High Performance analytical/semi-preparative

More information

889 IC Sample Center

889 IC Sample Center 889 IC Sample Center Manual 8.889.8001EN Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com 889 IC Sample Center Manual 8.889.8001EN 02/2010

More information

HPLC SYSTEM S 500 SERIES

HPLC SYSTEM S 500 SERIES HPLC SYSTEM S 500 SERIES The Sykam HPLC System S 500 Series is intended for all routine analysis as well as for the ambitious analyst. The system configuration is highly variable and several upgrade options

More information

OPERATORS MANUAL FOR USE WITH MAGNETIC/HEATED STIRRER MG200

OPERATORS MANUAL FOR USE WITH MAGNETIC/HEATED STIRRER MG200 OPERATORS MANUAL FOR USE WITH MAGNETIC/HEATED STIRRER MG200 ISSUE 4 April 2007 TABLE OF CONTENTS Page 1.0. GENERAL INFORMATION 2 1.1. DEFINITION 2 1.2. PRINCIPE OF OPERATION 2 2.0. TECHNICAL DESCRIPTION

More information

Installing an Auxiliary Pressure Control Module

Installing an Auxiliary Pressure Control Module Installing an Auxiliary Pressure Control Module Agilent 6850 GCs Accessory G3349B These instructions are divided into two parts: Part 1 to prepare the Auxiliary Electronic Pressure Control ("AuxEPC") module

More information

c-go 24V/6A 24V/8A 24V/12A

c-go 24V/6A 24V/8A 24V/12A c-go 24V/6A 24V/8A 24V/12A Battery charger GB Instruction manual 1 Index 1. Product description... 2 2. Safety advices... 3 3. Quick start guide... 4 4. Operation... 4 5. Problem solving... 6 6. Specifications...

More information

FITTINGS, LOOPS, AND VALVE ACCESSORIES

FITTINGS, LOOPS, AND VALVE ACCESSORIES FITTINGS, LOOPS, AND VALVE ACCESSORIES Fittings Rheodyne offers a variety of fittings. They differ in material of construction, tubing size, and pressure rating. Materials/Biocompatibility. 316 stainless

More information

1. INTRODUCTION SYSTEM DESCRIPTION Front Panel CONNECTION AND OPERATION TROUBLESHOOTING...8

1. INTRODUCTION SYSTEM DESCRIPTION Front Panel CONNECTION AND OPERATION TROUBLESHOOTING...8 Contents : 1. INTRODUCTION...1 2. IMPORTANT SAFETY INSTRUCTIONS...2 3. SYSTEM DESCRIPTION...4 3.1 Front Panel...4 4. CONNECTION AND OPERATION...6 5. TROUBLESHOOTING...8 6. MAINTENANCE...9 6.1 Operation...9

More information

Nickel Catalyst Tube Accessory 19205A

Nickel Catalyst Tube Accessory 19205A Installation and Operating Guide Nickel Catalyst Tube Accessory 19205A Agilent Technologies 2000 All Rights Reserved. Reproduction, adaptation, or translation without permission is prohibited, except as

More information

Agilent 1200 Series Manual Injector

Agilent 1200 Series Manual Injector Agilent 1200 Series Manual Injector User Manual 1200 Series MI User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2006, 2008 No part of this manual may be reproduced in any form or by

More information

3000W HF/PFC Battery Charger

3000W HF/PFC Battery Charger 3000W HF/PFC Battery Charger Description Advanced high frequency switching design with 92% typical efficiency Fully sealed enclosure providing improved reliability in demanding environments > 0.98 Power

More information

IMPORTANT! DO NOT THROW AWAY THE SHIPPING CARTON AND PACKING MATERIAL

IMPORTANT! DO NOT THROW AWAY THE SHIPPING CARTON AND PACKING MATERIAL Operator s Manual IMPORTANT! DO NOT THROW AWAY THE SHIPPING CARTON AND PACKING MATERIAL ii Table of Contents Operator Safety... 1 Introduction... 2 Unpacking and Setup... 3 Unpacking... 3 Setup... 4 ROCKET

More information

Operations and Service Manual. X30208 Load Bank

Operations and Service Manual. X30208 Load Bank Operations and Service Manual Read all instructions before using the load bank Contents 1. Components... 3 Total Assembly... 3 2) Specifications... 4 a)... 4 3) Receiving... 5 4) Safety... 5 a) Ground

More information

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve Technical Note This Technical Note describes the purpose, installation, and configuration and use of a diverter valve

More information

Agilent 1290 Infinity Quaternary Pump

Agilent 1290 Infinity Quaternary Pump Agilent 1290 Infinity Quaternary Pump User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2012-2014, 2015 No part of this manual may be reproduced in any form or by any means (including

More information

ACQUITY UPLC I-Class System with 2D Technology

ACQUITY UPLC I-Class System with 2D Technology ACQUITY UPLC I-Class System with 2D Technology The Waters ACQUITY UPLC I-Class System with 2D Technology allows chemists to increase sensitivity and selectivity, eliminate unwanted interferences, characterize

More information

Installing a Purged Packed (PP) Inlet

Installing a Purged Packed (PP) Inlet Installing a Purged Packed (PP) Inlet Agilent 6850 Gas Chromatograph Accessory G2651A These instructions are divided into two parts: Part 1 to prepare the Electronic Pressure Control ("EPC") module for

More information

VERITY Purification Systems Build the perfect system to fit your research

VERITY Purification Systems Build the perfect system to fit your research VERITY Purification Systems Build the perfect system to fit your research SPEC SHEET PURIFICATION VERITY PURIFICATION SYSTEMS WITH VERITY 3011 ISOCRATIC PUMP MAKE YOUR LAB LIFE EASIER From automated GPC

More information

ASTROAI USER MANUAL DIGITAL MULTIMETER. NOTE: Fully read and understand this manual before using this instrument.

ASTROAI USER MANUAL DIGITAL MULTIMETER. NOTE: Fully read and understand this manual before using this instrument. ASTROAI USER MANUAL DIGITAL MULTIMETER NOTE: Fully read and understand this manual before using this instrument. WARNING: To avoid possible electric shock or personal injury, and to avoid possible damage

More information

Agilent Valve Kit G4231A/B and G4232A/B/C

Agilent Valve Kit G4231A/B and G4232A/B/C Agilent Valve Kit G4231A/B and G4232A/B/C Instructions Contents Typical Applications 2 Dual column selection (G4231A/B or G4232A/B/C) 2 Sample enrichment and sample cleanup (G4231A/B or G4232A/B/C) 2 Alternating

More information

LC30 CHROMATOGRAPHY OVEN OPERATOR S MANUAL Dionex Corporation

LC30 CHROMATOGRAPHY OVEN OPERATOR S MANUAL Dionex Corporation LC30 CHROMATOGRAPHY OVEN OPERATOR S MANUAL 1995 Dionex Corporation Document No. 034860 Revision 03 January 1995 1995 Dionex Corporation All rights reserved worldwide. Printed in the United States of America.

More information

A B 0 0 C D E 6 7 G F F H 8 9 K M O O L N I J 1

A B 0 0 C D E 6 7 G F F H 8 9 K M O O L N I J 1 1 2 1 5 4 3 2 2 1 6 3 8 7 1 9 4 C A B 5 0 0 D E 6 G 7 F F H 8 K 9 M O O I J L N 1 GENERAL OPERATIONAL PRECAUTIONS 1. Keep work area clean. Cluttered areas and benches invite accidents. 2. Avoid dangerous

More information

Agilent InfinityLab Quick Change Valve G4235A Instructions

Agilent InfinityLab Quick Change Valve G4235A Instructions Agilent InfinityLab Quick Change Valve G4235A Instructions Contents Typical Application 2 Delivery Checklist 3 Specifications 4 Installing the Valve Heads 5 Agilent Technologies Typical Application This

More information

PRODUCT MANUAL. CARBONATE REMOVAL DEVICE 200 CRD 200 (4 mm) (P/N ) CRD 200 (2 mm) (P/N )

PRODUCT MANUAL. CARBONATE REMOVAL DEVICE 200 CRD 200 (4 mm) (P/N ) CRD 200 (2 mm) (P/N ) CRD 200 Product Manual Page 1 of 18 PRODUCT MANUAL FOR CARBONATE REMOVAL DEVICE 200 CRD 200 (4 mm) (P/N 062983) CRD 200 (2 mm) (P/N 062986) Dionex Corporation, 2006 Document No. 065068 Revision 04 November

More information

Instructions for Use SeQuant CARS Continuous Auto-Regeneration System for SeQuant SAMS Suppressors

Instructions for Use SeQuant CARS Continuous Auto-Regeneration System for SeQuant SAMS Suppressors Instructions for Use SeQuant CARS Continuous Auto-Regeneration System for SeQuant SAMS Suppressors Instructions for Use SeQuant CARS Continuous Auto-Regeneration System CONTENTS General Information...

More information

Trouble Shooting Guide for Hubbell Systems

Trouble Shooting Guide for Hubbell Systems Trouble Shooting Guide for Hubbell Systems NOTE: Hubbell Special Products discontinued operations in 2008, so there is a very limited number of repair parts available for motor repair. In some ways, this

More information

Thank you for purchasing a kit from SparKIT. Your support is greatly appreciated.

Thank you for purchasing a kit from SparKIT. Your support is greatly appreciated. SparKIT Instructions Thank you for purchasing a kit from SparKIT. Your support is greatly appreciated. First of all we would like to go through a few basic precautions. 1. This kit includes small parts

More information

Waters Reagent Manager Operator s Guide

Waters Reagent Manager Operator s Guide Waters Reagent Manager Operator s Guide 715002216 / Revision A Copyright Waters Corporation 2010 All rights reserved Copyright notice 2010 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND

More information

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard Agilent Preventive Maintenance provides factory recommended service for your analytical systems to assure reliable operation and the accuracy of your results. Delivered by highly-trained and certified

More information

Keystone Butterfly valves ParaSeal Installation and maintenance instructions

Keystone Butterfly valves ParaSeal Installation and maintenance instructions Before installation these instructions must be fully read and understood Please read these instructions carefully Hazard potentials: disregarding of instructions improper use of product insufficiently

More information

RENA AF371Feeder Operating Manual. Feeder. Operating Manual. Manual Part #: M AF371 Operations Rev

RENA AF371Feeder Operating Manual. Feeder. Operating Manual. Manual Part #: M AF371 Operations Rev Manual Part #: M-3022 Feeder AF371 Operations Rev. 3-16-04 1 RENA AF371 Feeder YOUR RENA AF371 IS DISTRIBUTED BY RENA SYSTEMS INC. SERVICE AND SUPPORT FOR THIS PRODUCT IS PROVIDED BY YOUR RENA DEALER.

More information

Agilent 1200 Series Standard and Preparative Autosamplers

Agilent 1200 Series Standard and Preparative Autosamplers Agilent 1200 Series Standard and Preparative Autosamplers User Manual Agilent 1200 Series Autosampler User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2006,2008 No part of this manual

More information

Single Stage Rotary Vane Vacuum Pump Installation and Operation Manual RX-10 RX-21 RX-25

Single Stage Rotary Vane Vacuum Pump Installation and Operation Manual RX-10 RX-21 RX-25 V acuum Pumps Single Stage Rotary Vane Vacuum Pump Installation and Operation Manual RX-10 RX-21 RX-25 www.republicsales.com Revised 10.14 2014 Republic Sales & Manufacturing Single Stage Rotary Vane Vacuum

More information

Agilent 1200 Series Valves

Agilent 1200 Series Valves Agilent 1200 Series Valves User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or by any means (including electronic storage and

More information

850 Professional IC. AnCat MCS Manual EN

850 Professional IC. AnCat MCS Manual EN 850 Professional IC AnCat MCS 2.850.3030 Manual 8.850.8052EN Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com 850 Professional IC AnCat

More information

Nickel Catalyst Tube on a 6820 GC, Accessory G4318A

Nickel Catalyst Tube on a 6820 GC, Accessory G4318A Nickel Catalyst Tube on a 6820 GC, Accessory G4318A Installation Guide The Nickel Catalyst tube (NCT) is used with the 6820 Gas Chromatograph (GC) for trace analysis of CO and CO 2 using a Flame Ionization

More information

Installation Guide. Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A

Installation Guide. Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A Installation Guide Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A Agilent Technologies 2001 All Rights Reserved. Reproduction, adaptation, or translation without permission is prohibited,

More information

FSG175 FENCE STAPLE GUN

FSG175 FENCE STAPLE GUN Kencove Farm Fence Supplies 344 Kendall Rd Blairsville, PA 15717 1-800-KENCOVE sales@kencove.com www.kencove.com OPERATING MANUAL FSG175 FENCE STAPLE GUN To reduce the risk of possible injury, read the

More information

DISCHARGER-ANALYZER BDX USER'S MANUAL

DISCHARGER-ANALYZER BDX USER'S MANUAL BATTERY DISCHARGER-ANALYZER BDX USER'S MANUAL OWM-BDX-261006 Page 1/10 1. INTRODUCTION Before starting to use your BDX battery discharger/analyzer, please take the time to read these instructions carefully.

More information

QuickSplit TM Flow Splitters

QuickSplit TM Flow Splitters Quick TM Flow ters Quick TM Flow ter Features: Fluid resistor technology eliminates tedious adjustments to capillary tubing for split ratio optimization ratios are stable and reproducible, and not affected

More information

CHROMATOGRAPHY. SYKAM S 6000 Valve Series

CHROMATOGRAPHY. SYKAM S 6000 Valve Series CHROMATOGRAPHY SYKAM S 6000 Valve Series Version: March 2011 SYKAM S 6000 Valve Series Rotor Seals Sykam offers a variety of different Rotor Seals materials to satisfy any possible application. All materials

More information

PF3100 TROUBLESHOOTING SOLUTIONS TO COMMON PROBLEMS. v1.1 Revised Nov 29, 2016

PF3100 TROUBLESHOOTING SOLUTIONS TO COMMON PROBLEMS. v1.1 Revised Nov 29, 2016 PF3100 TROUBLESHOOTING SOLUTIONS TO COMMON PROBLEMS v1.1 Revised Table of Contents 1 Common Alarms and Warnings... 1 2 Common Issues... 6 2.1 Communication problems... 6 2.1.1 Controller communication

More information

BRAVER UPS. (Uninterruptible Power System) User s Manual

BRAVER UPS. (Uninterruptible Power System) User s Manual BRAVER UPS (Uninterruptible Power System) User s Manual Safety CAUTION! This UPS utilizes voltages that may be hazardous. Do not attempt to disassemble the unit. The unit contains no user replaceable parts.

More information

TomTom-Tools GmbH Wiesenstrasse Baden Switzerland. Phone 1: Phone 2: VAT ID:

TomTom-Tools GmbH Wiesenstrasse Baden Switzerland. Phone 1: Phone 2: VAT ID: TomTom-Tools GmbH Wiesenstrasse 15 5400 Baden Switzerland www.tomtom-tools.com Phone 1: +41 79 774 06 42 Phone 2: +41 79 774 06 44 VAT ID: 698 468 Info@tomtom-tools.com User Manual: (Draft Version) OVALITY

More information

Porsche 928 with 16v LH-Jetronic Fuel System

Porsche 928 with 16v LH-Jetronic Fuel System Porsche 928 with 16v LH-Jetronic Fuel System Toll-Free Tech Hot Line: 877-FOR-928M 877-367-9286 Please do not copy this manual and give copies to your friends. Our ability to bring you this supercharger

More information

Agilent Valve Kit G4231A/B/C and G4232B/C/D

Agilent Valve Kit G4231A/B/C and G4232B/C/D Agilent Valve Kit G4231A/B/C and G4232B/C/D Instructions Agilent Valve Kit G4231A/B/C and G4232B/C/D - Instructions This Technical Note describes the installation of the Agilent G4231A/B/C and G4232B/C/D

More information

WATERFLUX 3070 Quick Start

WATERFLUX 3070 Quick Start WATERFLUX 3070 Quick Start Battery powered electromagnetic water meter Electronic Revision ER 4.3.0_ up to ER 4.3.4_ (SW.REV 4.2.2_ up to 4.2.5_) KROHNE CONTENTS WATERFLUX 3070 1 Safety instructions 4

More information

Protein Electrophoresis Unit

Protein Electrophoresis Unit A Geno Technology, Inc. (USA) brand name Protein Electrophoresis Unit Cat. No. BT201 1-800-628-7730 1-314-991-6034 info@btlabsystems.com Before using the instrument, please read the operation instruction

More information

PFC W HF/PFC Battery Charger

PFC W HF/PFC Battery Charger PFC 5000 5000W HF/PFC Battery Charger Description Advanced high frequency switching design with 92% typical efficiency Fully sealed enclosure providing improved reliability in demanding environments >

More information