Evaluating and Determining Precision and Accuracy for Nanoliter Volumes of Viscous Solutions on an Automated Liquid Handling Instrument
|
|
- Simon Blake Wells
- 5 years ago
- Views:
Transcription
1 Evaluating and Determining Precision and Accuracy for Nanoliter Volumes of Viscous Solutions on an Automated Liquid Handling Instrument Application Note 231 Niki Popp, Tim Hegeman, Joan Stevens, Ph.D. and Greg Robinson (Gilson, Inc.) Introduction The Constellation 1200 is a precise, small-volume liquid handler. It consists of a twelve-channel, noncontact dispenser capable of nanoliter to microliter sample delivery. It is capable of spotting to virtually any format or surface, and can be used for applications ranging from microarray and MALDI spotting to crystallography. This study will examine the instrument s precision and accuracy for dispensing nanoliter volumes of solutions used in crystallography applications. The five solutions being evaluated in this application vary in viscosity, composition, concentration and their overall physical characteristics, and include: 1M HEPES, 30% 2-methyl-2,4-pentanediol, 20% PEG 4000/ 20% IPA, 0.1M MgCl 2 and 30% PEG Photo 1. Six 100-nL drops on a dime. June 2005 Page 1 of
2 Materials & Methods Constellation 1200 Hardware Characteristics X/Y/Z robot with a large bed capacity for holding twelve microplates. Aspirates from 96- and 384-well microplates using syringe technology. Dispenses from 50 nl to 500 µl volumes into 96-, 384- and 1536-well plates employing solenoids for small volumes and syringes for large volumes. (Versions of the instrument are available for higher density plates and volumes down to 5 nl.) Contains two syringes and two selector valves. Syringe fluid pathways are independent of solenoids. Dual wash system allows for both flow-through and nonflow-through rinses. External pressure source maintains a constant, low pressure (10 psi) on the solenoids, assuring the longevity of the valves. True volume aspirating and dispensing. Air gap eliminates any sample diffusion. Solenoids can be individually calibrated to a specific application. Noncontact fluid transfer. Sample never passes through the solenoids, eliminating the possibility of contamination and plugging. Photo 2. Constellation 1200 Liquid Handler. June 2005 Page 2 of
3 Smart Tip Technology Ensures precision in the tips by checking the alignment and positioning of each tip in relation to each other and to the bed Accuracy to ±30 µm Precision to ±2 µm Out of position probes will provide an on-screen alert instructing the user to make appropriate adjustments Photo 3. Closeup of the SmartTip Technology. 735 Sampler Software Software allows design of each zone individually. Drag-and-drop tasks allow quick-and-easy method development and optimization changes. Option for incorporating variables into fields increases flexibility. Custom racks allow unique formats and geometries, including nonstandard targets. Figure 1. Example of a plate type with unique zones denoted by the various colored wells. June 2005 Page 3 of
4 Figure 2. Example of a task and its available data fields. Purpose The purpose of this experiment is to determine and evaluate the Constellation s precision and accuracy in dispensing variable volumes of five unique crystallography solutions including: 1M HEPES 30% 2-methyl-2,4-pentanediol 20% PEG 4000/20% IPA 0.1M MgCl 2 30% PEG 8000 Precision and accuracy are defined as follows: Precision = The quality of being reproducible in all replicate transfers. Coefficient of variance will be used to measure precision. Accuracy = The quality of conforming exactly to a requested delivery volume. Average volume will be used to measure accuracy. Method The precision and accuracy of the Constellation 1200 was evaluated by determining the CVs and average volumes of the crystallography solutions for each of the 12 probes at test volumes of 50, 100 and 300 nl. Each volume was tested in triplicate, and all of the solutions were dispensed using a neartouch-off method with the same solenoid calibration. The CVs and averages were calculated based on spectrophotometric analysis of a standard curve (ranging from nl) for each solution. See tables 1-10 for precision and accuracy data for all 12 probes and three plates. June 2005 Page 4 of
5 Photo 4. Near-touch-off method. The probe was placed in very close proximity (~0.1mm) to the plate during the dispense cycle. Upon dispense, the deformation of the fluid at the probe orifice is sufficient to form a drop in the wells of the plate. 735 Sampler Software Task 1) Aspirated sample from source plate utilizing an air gap and three predispenses. 2) Dipped probes in flowing rinse station to eliminate any sample on the outside surface of the probes. 3) Dispensed sample/methylene blue solution into three destination plates and manually brought to volume with water. 4) Performed spectrophotometric analysis on solutions using Molecular Devices Vmax Reader. 5) Calculated precision and accuracy based on solution s standard curve. Photo 5.Closeup of air gaps. June 2005 Page 5 of
6 Table 1: Avg Precision and Accuracy Data for Plates 1 3 Solution CV (%) 50 nl 100 nl 300 nl CV (%) CV (%) 0.1M HEPES % 2-methyl- 2,4-pentanediol 20% PEG 4000/20% IPA M MgCl % PEG Table 2: Precision and Accuracy Data of 12 Probes at 100 nl Solution Probe M HEPES 30% 2- methyl-2,4- pentanediol 20% PEG 4000/ 20% IPA CV (%) CV (%) CV (%) M MgCl CV (%) % PEG 8000 CV (%) June 2005 Page 6 of
7 Summary As a means to determine the precision and accuracy of an automated liquid handling system, it is useful to determine both the CVs and average volumes of the data sets being evaluated. The Constellation 1200 is capable of producing plate CVs (grouping all five solutions as a single data set) that range from % at 50 nl, % at 100 nl and % at 300 nl and average volumes that include nl at 50 nl, nl at 100 nl and nl at 300 nl. Quantitatively, there was no significant difference in either the precision or accuracy among the five solutions. In addition to determining the precision and accuracy of each plate, the CVs and average volumes were also calculated for the 12 probes at each volume. This data is available upon request. Conclusion The results of the experiment demonstrate that the Constellation 1200 is an effective liquid handling instrument for aspirating and dispensing sub microliter volumes of crystallography solutions. The near touch-off method and solenoid calibration that was used in this study proved to produce both accurate and precise results even when the solutions differed in viscosity, composition, concentration and physical properties. These values coupled with hardware characteristics that include independent syringe and solenoid pathways, a dual wash system, ease of solenoid calibration, SmartTip Technology and plate flexibility, make the Constellation 1200 a logical choice for dispensing nanoliter to microliter volumes of solutions for applications that include both microarray and MALDI spotting and crystallography. Gilson, Inc. World Headquarters P.O. Box , Middleton, WI USA Telephone: (1) or (1) Fax: (1) Gilson S.A.S. 19, avenue des Entrepreneurs, F VILLIERS LE BEL France Telephone: (33-1) Fax: (33-1) sales@gilson.com, service@gilson.com, training@gilson.com June 2005 Page 7 of
235/235P Autoinjector User's Guide
235/235P Autoinjector User's Guide LT2401/ 2003 Gilson, Inc. All rights reserved. January 2003 Table of Contents Declaration of Conformity 1 Introduction 2 Installation Safety Precautions... 1-2 Unpacking...
More informationGX-271 ASPEC Gilson, Inc. All Rights Reserved. LT
GX-271 ASPEC GX-271 ASPEC World Headquarters Gilson, Inc. 3000 Parmenter Street P.O. Box 620027 Middleton, WI 53562-0027 USA Telephone: 608-836-1551 Fax: 608-831-4451 Gilson S.A.S. 19, avenue des Entrepreneurs
More informationThe Efficient Solution for Pipetting up to 100 ml at once! USER'S GUIDE
The Efficient Solution for Pipetting up to 100 ml at once! This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the User s Guide, may
More information4 - SETTING UP THE MACROMAN CHARGING POSITION 4 Wall mounted 4 Workbench set-up 4
User's Guide Contents Page 1 - SAFETY SYMBOLS USED 3 2 - PROPER USE 3 3 - SAFETY PRECAUTIONS 3 Possible dangers if the safety precautions are not observed 3 Safety at work 3 Safety aspects for the user
More informationThe Importance of Liquid Handling Details and Their Impact on your Assays
The Importance of Liquid Handling Details and Their Impact on your Assays European Lab Automation Hamburg, Germany May 30, 2012 John Thomas Bradshaw, Ph.D. Artel, Inc. Automation is My Friend Automated
More informationProgress made simple!
Progress made simle! ELECTRONIC Piette TRUSTED RESULTS EASY & INTUITIVE USE ADVANCED ERGONOMICS ESSENTIAL ETTING FEATURES 4 essential ietting modes: FORWARD ietting, retive, mix and reverse. TRUSTED RESULTS
More informationProgress made simple!
Progress made simple! A d j u s ta b l e - V o l u m e P i p e t t e EASY TO USE ADVANCED ERGONOMICS TRUSTED RESULTS EASY TO USE ADVANCED ERGONOMICS TRUSTED RESULTS Progress made simple! PIPETMAN M is
More information819 Injection Module User s Guide
819 Injection Module User s Guide LT2533/ 2004 Gilson, Inc. All rights reserved August 2004 Table of Contents Declaration of Conformity 1 Introduction 2 Installation 3 Operation 4 Maintenance Unpacking...
More informationMANUAL DISPENSING REDEFINED CERTUS HANDY LIQUID DISPENSER
MANUAL DISPENSING REDEFINED CERTUS HANDY LIQUID DISPENSER REVOLUTIONARY FUSION OF MANUAL AND AUTO- MATED DISPENSING CERTUS HANDY unlocks a new dimension of manual digital dispensing systems for unmatched
More information845Z Injection Module User s Guide
845Z Injection Module User s Guide LT2575/ 2004 Gilson, Inc. All rights reserved May 2004 Table of Contents Declaration of Conformity 1 Introduction 2 Installation 3 Operation 4 Maintenance Unpacking...
More informationGilson, Inc. Table of Contents. HPLC: Liquid Handlers/Injectors GX-271 Analytical Liquid Handler...43 ISO 9001 certified.
Gilson, Inc. Table of Contents Designed for purification and engineered for science, Gilson, Inc. provides high quality, dependable solutions for today s demanding liquid chromatography, solid phase extraction,
More informationVERITY Purification Systems Build the perfect system to fit your research
VERITY Purification Systems Build the perfect system to fit your research SPEC SHEET PURIFICATION VERITY PURIFICATION SYSTEMS WITH VERITY 3011 ISOCRATIC PUMP MAKE YOUR LAB LIFE EASIER From automated GPC
More informationIncreased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel
Application Note Energy and Chemicals Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel Applying the Energy Institute Method IP 8 with an Agilent J&W
More informationAgilent 7693A Automated Liquid Sampler
Agilent 7693A Automated Liquid Sampler Specifications Overview The Agilent 7693A is a state-of-the-art sample handling and injection system that provides the highest levels of precision and reliability
More informationMicro 215 Liquid Handler User's Guide
Micro 215 Liquid Handler User's Guide LT2537/ 2000 Gilson, Inc. All rights reserved April 2000 Table of Contents 1 Introduction 2 Installation Safety Precautions... 1-2 Unpacking... 1-3 Standard Equipment...
More informationAgilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011
Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc.
More informationGilson, Inc. Table of Contents. Solid Phase Extraction Systems HPLC: Systems 2 9. Liquid Handlers HPLC: Liquid Handlers/Injectors 10 14
Gilson, Inc. Table of Contents Designed for purification and engineered for science, Gilson, Inc. provides high quality, dependable solutions for today s demanding liquid chromatography, solid phase extraction,
More informationSpotting Blocking Hybridization Washing
Reagents & kits Nexterion reagents The quality of the results from a microarray experiment is dependent on many factors, the substrate utilized for printing, the printing buffer, the labeling method employed,
More informationAgilent 7693A Automated Liquid Sampler
IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7693A Automated Liquid Sampler
More informationA Comparison of Boat Introduction and Direct Injection using the Thermo Scientific ipro 5000 Series Combustion Analyzer
Application Note: 42165 A Comparison of Boat Introduction and Direct Injection using the Thermo Scientific ipro 5000 Series Combustion Analyzer Kristian J. Hoffman, Angela Seipel, Application Specialists,
More informationNEW INSTRUMENT. 08th November / MKT. We announce the launching of the new member of the Biokit Elisa Instrumentation family: DS2
08th November 2010 293/ MKT NEW INSTRUMENT We announce the launching of the new member of the Biokit Elisa Instrumentation family: DS2 The DS2 instrument is the new Biokit automated ELISA processor. The
More information306 Pump User s Guide
306 Pump User s Guide 306 Pump User s Guide LT801175G/ 2002 Gilson SAS All rights reserved September 2002 Table of Contents Safety 1 Introduction 2 Description 3 Installation 4 Operation Using this manual...
More informationJoint Oil Analysis Program Spectrometer Standards VHG Labs Inc. Qualification Report For D19-0, D3-100 and D12-XXX Series Standards
Joint Oil Analysis Program Spectrometer Standards VHG Labs Inc. Qualification Report For D19-0, D3-100 and D12-XXX Series Standards NF&LCFT REPORT 441/13-010 Prepared By: MICHAEL PERETICH, PhD Oil Analysis
More informationGC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585
GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc. 850 Centerville
More informationBIAcore: Real-Time Biomolecule Interaction Analysis
BIAcore: Real-Time Biomolecule Interaction Analysis Biacore T-100 奇異亞洲醫療設備股份有限公司林怡欣產品專員 Understanding molecular recognition a fundamental phenomenon of life How strong => Affinity How fast => Kinetics
More informationSPECIFICATION. Bench top triple quadrupole Mass Spectrometry System coupled to a gas performance Chromatography (GC-MS/MS)
SPECIFICATION Bench top triple quadrupole Mass performance Chromatography (GC-MS/MS) 1. Scope These specifications describe the requirements for a bench top triple quadrupole mass spectrometry system coupled
More informationAppendix A.1 Calculations of Engine Exhaust Gas Composition...9
Foreword...xi Acknowledgments...xiii Introduction... xv Chapter 1 Engine Emissions...1 1.1 Characteristics of Engine Exhaust Gas...1 1.1.1 Major Components of Engine Exhaust Gas...1 1.1.2 Units Used for
More informationMultichannel ENGLISH
PIPETMAN Ultra Multichannel ENGLISH PIPETMAN Ultra Multichannel CONTENTS page 1 - INTRODUCTION 2 2 - PARTS CHECK LIST 3 3 - DESCRIPTION 4 4 - SWITCHING ON 5 5 - VOLUME INDICATOR - LCD WINDOW 6 6 - SETTING
More informationCLASSIC. User s Guide. Do not lubricate this pipette. The use of lubricant cancels the warranty of this pipette. NOTICE
CLASSIC User s Guide EN NOTICE Do not lubricate this pipette. The use of lubricant cancels the warranty of this pipette. TABLE OF CONTENTS INTRODUCTION 3 Parts Checklist 3 GLP Features 3 Description 4
More informationGENERAL CATALOG L.TEX CORPORATON
GENERAL CATALOG L.TEX CORPORATON 1/8 동방하이테크상사 http://www.allpump.co.kr T.(02)457-6292 F.(02)457-6293 13-12-01 L. Tex Corporation was established in March 2001 to design, develop and manufacture the versatile
More informationMethanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.
Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers Application Note Abstract Versa With the rising prices of fossil fuels, more emphasis is being put on renewable resources
More informationDxC 700 AU Job Aid Booklet
DxC 700 AU Job Aid Booklet For Training Purposes Only These job aids are shortened versions of procedures found in the references below. Information in the job aid is correct as of the date published.
More informationQuickSplit TM Adjustable Flow Splitters Analytical Splitters Analytical range, 0.1 ml/min. to 5 ml/min. input flow, calibration flow rate 1.0 ml/min.
Adjustable Flow ters Applications Post-column splitting is fairly straight forward. Like pre-column flow splitting, any significant additional pressure (resistance) down stream from the splitter may affect
More informationHigh Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph
Application Note Petrochemicas High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Author James D. McCurry, Ph.D. Agilent Technologies, Inc. Abstract An Agilent
More informationQualification of a precision pattern dispenser
Qualification of a precision pattern dispenser Raf Van Ammel, Katarzyna Sobiech-Matura, Uwe Wätjen 2016 EUR 28268 EN This publication is a Technical report by the Joint Research Centre, the European Commission
More informationUsing a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application
Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D759 Application Petrochemical Author James D. McCurry Agilent Technologies 285 Centerville
More information2F MEMS Proportional Pneumatic Valve
2F MEMS Proportional Pneumatic Valve Georgia Institute of Technology Milwaukee School of Engineering North Carolina A&T State University Purdue University University of Illinois, Urbana-Champaign University
More informationMulti-Sample In Situ Filtration System from Deep Sea Thermal Vents
Multi-Sample In Situ Filtration System from Deep Sea Thermal Vents Authors: Paul Spencer (Delaware Tech. Mechanical Design) Tyler Davidson (Delaware Tech. Electrical Engineering Technology) Advisors: George
More informationTotally Automated Method for the Determination of Sudan Dyes in Food via On-Line Filtration, SPE and HPLC Analysis
Totally Automated Method for the Determination of Sudan Dyes in Food via On-Line Filtration, SPE and HPLC Analysis Joan M Stevens, Ph.D., Mark Crawford and Greg Robinson Single Presenter Name Abstract:
More informationAutomated Liquid Handling Station
Automated Liquid Handling Station Quick Installation Guide Manual Part Number 32-0406-048 Rev 0 COPYRIGHT 2018 Teledyne Technologies Incorporated. All rights reserved. 32-0406-048 Rev 0, May, 2018 Printed
More informationLarge Volume Injection of Polycyclic Aromatic Hydrocarbons
JSB is an authorised partner of Large Volume Injection of Polycyclic Aromatic Hydrocarbons Application Note - Environmental #113 Author Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Abstract
More informationQuality Control in Mineral Exploration
Quality Control in Mineral Exploration Controlling the Quality of Information from Field to Data Base Not to be reproduced without written permission Quality Control in Mineral Exploration There many goals
More informationTSHR 7000 series TN-TS-TX analyzer
TSHR 7000 series TN-TS-TX analyzer Enhanced Capabilities with Powerful Performance for Trace Level Nitrogen, Sulfur, Chlorine Analysis Petrochemicals Refineries Analytical Services Research Analysis The
More informationTHERMOELECTRIC SAMPLE CONDITIONER SYSTEM (TESC)
THERMOELECTRIC SAMPLE CONDITIONER SYSTEM (TESC) FULLY AUTOMATED ASTM D2983 CONDITIONING AND TESTING ON THE CANNON TESC SYSTEM WHITE PAPER A critical performance parameter for transmission, gear, and hydraulic
More informationRegards from the Team at Rheology Solutions.
Rheology Solutions is the sole Australian distributor of this product range and we welcome the opportunity of discussing your application requirements. We hope the information you are seeking is contained
More informationmulti Mid. vol. Max. vol. Min. vol. < 3.0% 1) < 1.0% < 0.6% < 0.6% < ± 1.5% < ± 0.8% < ± 0.7% < ± 0.8% < ± 1.5% < ± 0.8% < ± 0.7% < 1.0% < 0.
multichannel manual micropipettes Built with the users in mind, these pipettes combine high-tech materials with exceptional, user-friendly ergonomics. They extend pipetting possibilities in 96-well microplates
More informationApplication. Gas Chromatography June 1995
Determining Oxygenates in Gasoline: ASTM Method D Application Gas Chromatography June 99 Authors Michael J. Szelewski Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-60 USA Matthew S. Klee
More informationHBCDs AND TBBP-A IN INDOOR AIR AND DUST IN BIRMINGHAM, UK: IMPLICATIONS FOR HUMAN EXPOSURE
SUPPORTING INFORMATION FOR ABDALLAH ET AL, 2008 HBCDs AND TBBP-A IN INDOOR AIR AND DUST IN BIRMINGHAM, UK: IMPLICATIONS FOR HUMAN EXPOSURE Mohamed Abou-Elwafa Abdallah, Stuart Harrad, Adrian Covaci 13
More informationStabinger Viscometer SVM 3001
Stabinger Viscometer SVM 3001 Expect More Welcome to New Viscometry Measuring the viscosity of oils and fuels is all about certainty and compliance. Two approaches get you there: the traditional glass
More informationSpark Holland introduces ALIAS for ÄKTA pure All-round, state-of-the-art
BETTER SAMPLE CARE Spark Holland introduces ALIAS for ÄKTA pure All-round, state-of-the-art autosampler for ÄKTA pure Spark Holland introduces ALIAS for ÄKTA pure All-round, state-of-the-art autosampler
More informationRheoStress 600. The Modular High-Tech Rheometer
RheoStress 600 The Modular High-Tech Rheometer What counts in Rheometry... VC-Motor Powerful, dynamic and extremely fast The proven VC motor of the RS300 has been further improved and sets new standards
More informationAnalysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES
Analysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES Application note Petrochemical Author Neli Drvodelic Agilent Technologies Melbourne, Australia Introduction
More informationPATROL UPLC Process Analysis System for Production
PATROL UPLC Process Analysis System for Production The PATROL UPLC Process Analysis System for Production is an integrated system solution designed and engineered to perform online and atline reaction
More informationGC MERGE OF LIGHT ENDS WITH ASTM D7169 BOILING POINT DISTRIBUTION
GC MERGE OF LIGHT ENDS WITH ASTM D7169 BOILING POINT DISTRIBUTION Crude Oil Quality Association St. Louis, MO June 8, 2017 Arden Strycker, Ph.D. SGS North America Providing Solutions Across the Value Chain
More informationQuickSplit TM Flow Splitters
Quick TM Flow ters Quick TM Flow ter Features: Fluid resistor technology eliminates tedious adjustments to capillary tubing for split ratio optimization ratios are stable and reproducible, and not affected
More informationEDP 1. Multichannel Electronic Pipette with LTS LiteTouch Tip Ejection. Advanced Multichannel Pipette
EDP 1 Multichannel Electronic Pipette with LTS LiteTouch Tip Ejection Advanced Multichannel Pipette Models: EP8-10 EP8-20 EP8-50 EP8-200 EP8-300 EP8-1200 EP12-10 EP12-20 EP12-50 EP12-200 EP12-300 EP12-1200
More informationWHEATON. SOCOREX Acura electro Pipettes Your Electronic Choice. A World of Pipetting Expertise with SOCOREX.
WHEATON SOCOREX Acura electro Pipettes A World of Pipetting Expertise with SOCOREX www.wheaton.com Precision. Economical. Trusted. Easy To Use. WHEATON 800.225.1437 (U.S. & Canada Only) 856.825.1100 856.825.1368
More informationPIPETMAN ultra. EN ENGLISH User's Guide
PIPETMAN ultra EN ENGLISH User's Guide PIPETMAN ultra Contents Page 1 - INTRODUCTION 2 2 - PARTS CHECK LIST 3 3 - DESCRIPTION 4 4 - SWITCHING ON 5 5 - VOLUME INDICATOR - LCD WINDOW 6 6 - SETTING THE VOLUME
More informationGC-101. Testing of Biodiesel Using Gas Chromatography. by Bob Armantrout. Presented at: Local Biodiesel: A Biodiesel Coop Conference
GC-101 Testing of Biodiesel Using Gas Chromatography by Bob Armantrout (Image Research Machines plc) Presented at: Local Biodiesel: A Biodiesel Coop Conference Strategies For Community-based Biodiesel
More informationWhite paper: Originally published in ISA InTech Magazine Page 1
Page 1 Improving Differential Pressure Diaphragm Seal System Performance and Installed Cost Tuned-Systems ; Deliver the Best Practice Diaphragm Seal Installation To Compensate Errors Caused by Temperature
More informationPerformance of Batteries in Grid Connected Energy Storage Systems. June 2018
Performance of Batteries in Grid Connected Energy Storage Systems June 2018 PERFORMANCE OF BATTERIES IN GRID CONNECTED ENERGY STORAGE SYSTEMS Authors Laurie Florence, Principal Engineer, UL LLC Northbrook,
More informationElectric. Efficient. Compact & precise. ENGEL e-mac
Electric. Efficient. Compact & precise. be the first. All-electric. Best-in-class efficiency and precision to the max. All of the s movements are performed by servo-electric drives. The all-electric drive
More informationIRBT 2005: Medium Track Platform Overview
ABB Robotics, November 2015 IRBT 2005: Medium Track Platform Overview Slide 1 IRBT 2005 Agenda Overview Differentiated Value Proposition Targeted Applications Customer Benefits Design Options Technical
More informationTABLE OF CONTENTS INTRODUCTION. Chapter 1. Parts Check List INTRODUCTION TABLE OF CONTENTS NOTICE INTRODUCTION 3
User s Guide EN G TABLE OF CONTENTS TABLE OF CONTENTS INTRODUCTION 3 Parts CheckList 3 GLP Features 4 Description Specifications 6 SETTING THE VOLUME 8 PIPETTING 9 Fit the Tips 9 Pre-rinse the Tips 0 Aspirate
More informationF U L LY A U TO M AT I C D I S P E N S I N G S Y S T E M OF THE NEXT GENERATION. mid-vol green C=85 M=15 Y=100 K=5. low-vol yellow C=0 M=22 Y=100 K=0
prototyping brown C=30 M=65 Y=100 K=30 low-vol yellow C=0 M=22 Y=100 K=0 mid-vol green C=85 M=15 Y=100 K=5 high-vol blue C=96 M=80 Y=0 K=0 customized red C=15 M=100 Y=100 K=5 solar oran C=0 M=60 prototyping
More informationUltrasonic and Magnetic Particle Testing of New Railway Wheels
19 th World Conference on Non-Destructive Testing 2016 Ultrasonic and Magnetic Particle Testing of New Railway Wheels Wolfram A. Karl DEUTSCH 1, Wolfgang WEBER 1, Klaus MAXAM 1, Mathias RAZENG 1, Frank
More information402 Dilutor Dispenser User s Guide
402 Dilutor Dispenser User s Guide 402 Dilutor Dispenser User s Guide LT801310/C - 2003 Gilson SAS All rights reserved November 2003 Contents Table of Contents 0 Safety 1 Introduction 2 Description 3
More informationThe PMAC Dynamic Scale Loop
The Dynamic Scale Loop Original System for evaluation of scaling tendency and inhibitor performance The System is an invaluable research and development tool for technologists responsible for the design,
More informationJoint Oil Analysis Program Spectrometer Standards SCP Science (Conostan) Qualification Report For D19-0, D3-100, and D12-XXX Series Standards
Joint Oil Analysis Program Spectrometer Standards SCP Science (Conostan) Qualification Report For D19-0, D3-100, and D12-XXX Series Standards NF&LCFT REPORT 441/15-008 Prepared By: MICHAEL PERETICH, PHD
More informationPursuing high precision, speed, and ease of use
Carbon/Sulfur Analyzers for Solid Materials EMIA series Pursuing high precision, speed, and ease of use Carbon / Sulfur Analyzer High-frequency induction furnace type EMIA-920V2/320V2 Series New EMIA-V2/FA
More informationPhone: Fax: More information:
PZ HTL S.A. is a modern and dynamically growing enterprise which is part of an international group. We have our own construction and technology department and an experienced, highly-qualified personnel.
More informationFixed Flow Splitter *
Fixed Flow * *Patent Pending Analytical Scientific Instruments Fluid resistor technology eliminates tedious adjustments to capillary tubing to change split ratios. Easy to use interchangeable fluid resistors
More informationPipetted Liquid Sampling. Automated sample preparation platform for online LC-MS/MS bioanalysis
Pipetted Liquid Sampling Automated sample preparation platform for online LC-MS/MS bioanalysis Integrated Sample Processing SCAP PLS system components mounted on a robotic platform Integrated Sample Processing
More informationFrom Patient to Plate
From Patient to Plate Development of Highly Sensitive Clinical PK Assays Andrew Macintyre M.D., Ph.D. Director of Research and Development Sarah Christensen B.S. Research Scientist II Custom Assay Service
More informationModel LE-300C/LH-300C/LZ-300C Coating Thickness Tester. Operating Manual
Model LE-300C/LH-300C/LZ-300C Coating Thickness Tester Operating Manual Contents The Model LE-300C / LH-300C / LZ-300C Coating Thickness Testers 2 LE-300C / LZ-300C l Description 3 ll Specifications 5
More informationDilutor Module for TriPlus RSH
Thermo Scientific Dilutor Module for TriPlus RSH Instruction Manual Additional section of the TriPlus RSH User Guide and Hardware Manual 31709661 Revision A December 2015 2015 Thermo Fisher Scientific
More informationACQUITY UPLC I-Class System with 2D Technology
ACQUITY UPLC I-Class System with 2D Technology The Waters ACQUITY UPLC I-Class System with 2D Technology allows chemists to increase sensitivity and selectivity, eliminate unwanted interferences, characterize
More informationAutoplex Specifications
Autoplex Specifications Autoplex G1 w/heating coil Stainless steel tip holds up to 150µl of liquid Probe tubing holds up to 3.18ml of liquid (length=88in, diameter=0.673mm/0.053in, 36.1µl/inch-high surface
More informationS Pyatt. UV glue dispensing studies on the Dima Dotmaster
UV glue dispensing studies on the Dima Dotmaster 1 The study involves the testing of 3 UV cure glues. The candidates are: Loctite 3525 Dymax 6-621 Dymax 3013 Two dot pattern types have been defined, a
More informationPICO Pµlse Contact Dispense Valve
PICO Pµlse Contact Dispense Valve The Latest Advance in Precision Contact Dispensing Technology The PICO Pµlse contact dispense valve delivers faster, more precise dispensing with less turbulence for greater
More informationCOMPUTATIONAL FLOW MODEL OF WESTFALL'S 2900 MIXER TO BE USED BY CNRL FOR BITUMEN VISCOSITY CONTROL Report R0. By Kimbal A.
COMPUTATIONAL FLOW MODEL OF WESTFALL'S 2900 MIXER TO BE USED BY CNRL FOR BITUMEN VISCOSITY CONTROL Report 412509-1R0 By Kimbal A. Hall, PE Submitted to: WESTFALL MANUFACTURING COMPANY May 2012 ALDEN RESEARCH
More informationA unique density, viscosity sensor. High technology brick. Field oriented. For light & medium petroleum products. Related norms
For light & medium petroleum products A unique density, viscosity sensor DEVIL for light fuels is the only density + viscosity + temperature sensor available on the market, dedicated to light refined cuts.
More informationMicron-S Precision Dispensing Tips
Description Micron-S Precision Dispensing Tips are high accuracy dispensing tips for small deposit fluid dispensing. Micron-S Precision Dispensing Tips have a Luer lock thread and can be used with any
More informationVehicle Simulation for Engine Calibration to Enhance RDE Performance
Vehicle Simulation for Engine Calibration to Enhance RDE Performance IPG Apply & Innovate 2018 11st and 12nd of September, Karlsruhe, Germany Dr. Yutaka Murata Yui Nishio Dr. Yukihisa Yamaya Masato Kikuchi
More informationand Direct Write Authors: Ludwing A. Hernandez, Brent Stucker: Utah State University.
Integration & Process Planning for Combined Ultrasonic Consolidation and Direct Write Authors: Ludwing A. Hernandez, Brent Stucker: Utah State University. Abstract A research project is underway to integrate
More informationPremium Design - Premium Performance - Premium Quality
WITOPET premium NEW Premium Design - Premium Performance - Premium Quality Looking for the latest pipetting innovation at an attractive price? WITOPET premium is just the right pipette for you. The ergonomic
More informationMicro Dyne Motor Testing System
Data Sheet Motor Testing System Features designed specifically for miniature and micro s Torque: Easily convertible from 2.0 mn m to 4.0 mn m (0.28 oz in to 0.57 oz in) Speed: up to 100,000 rpm Power:
More informationHyperShear TM HPLC and UHPLC Mixers
HyperShear TM HPLC and UHPLC s HyperShear TM HPLC and UHPLC Features: Micro Flow, Low Flow, Analytical Flow, High Flow and Combo Series Reduce baseline noise, increase sensitivity and improve gradient
More informationAnalysis of Petroleum Samples Using the Teledyne Leeman Labs
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1802 ICP Petroleum Analysis of Petroleum Samples Using the Teledyne Leeman Labs Prodigy Plus ICP-OES John Condon, Applications Chemist and Bruce MacAllister, Applications
More informationProduct Description Sheet Posi-Link Volumetric Dispensing System See Item Numbers listed below North American Engineering Center
Posi-Link Volumetric Dispensing System North American Engineering Center PRODUCT DESCRIPTION The Loctite Posi-Link Volumetric Dispensing System is a programmable positive displacement dispenser specifically
More informationEfficiency in variable pipetting
Efficiency in variable pipetting Microliter pipettes with variable volumes VITLAB micropipette VITLAB specialises in accurate dispensing, titration and pipetting, and offers a variety of highly developed
More informationenglish Series 1 The versatile rheometer concept
english Series 1 The versatile rheometer concept Three Instruments... RheoStress 1 The RheoStress 1 is a universal laboratory rheometer with the features of a research grade instrument. These excellent
More informationStandard Test Method for Pour Point of Petroleum Products (Automatic Air Pressure Method) 1
Designation: D 6749 02 An American National Standard Standard Test Method for Pour Point of Petroleum Products (Automatic Air Pressure Method) 1 This standard is issued under the fixed designation D 6749;
More informationIMPROVING SULFURIC ACID PLANT PERFORMANCE THROUGH NEW SHAPE & HIGHER ACTIVITY CATALYSTS
IMPROVING SULFURIC ACID PLANT PERFORMANCE THROUGH NEW SHAPE & HIGHER ACTIVITY CATALYSTS BY: TIMOTHY R. FELTHOUSE, Ph.D; MARIO P. DIGIOVANNI, P.E.; JOHN R. HORNE AND SARAH A. RICHARDSON PRESENTED AT: THE
More informationEmbedded Torque Estimator for Diesel Engine Control Application
2004-xx-xxxx Embedded Torque Estimator for Diesel Engine Control Application Peter J. Maloney The MathWorks, Inc. Copyright 2004 SAE International ABSTRACT To improve vehicle driveability in diesel powertrain
More informationGas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593
Application Note Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Authors Kelly Beard and James McCurry Agilent Technologies, Inc. Abstract An Agilent
More informationApplication Note Original Instructions Development of Gas Fuel Control Systems for Dry Low NOx (DLN) Aero-Derivative Gas Turbines
Application Note 83404 Original Instructions Development of Gas Fuel Control Systems for Dry Low NOx (DLN) Aero-Derivative Gas Turbines Woodward reserves the right to update any portion of this publication
More informationThe sensor for every occasion: VA-300
The sensor for every occasion: VA-300 VA-300H-ST with flange 2.5 2500# and special non-active-extension (NAE) for operation in polymer melts at 280 C and 175 bar in hazardous area The ViscoScope VA-300
More informationAppNote 1/2010 KEYWORDS ABSTRACT. Biodiesel, Automation, ASTM D
AppNote 1/2010 Full Automation of ASTM Method D6584-07 Standard Test Method for the Determination of Free and Total Glycerin in B-100 Biodiesel Methyl Esters by Gas Chromatography using a GERSTEL Dual
More informationLabSmith uprocess System Microfluidic Automation
LabSmith uprocess System Microfluidic Automation Automated fluid delivery and rou ng Innova ve so ware for coordinated controlol Constant-pressure or constant-rate flow Accurate temperature control and
More information