SPECIFICATION. Bench top triple quadrupole Mass Spectrometry System coupled to a gas performance Chromatography (GC-MS/MS)

Size: px
Start display at page:

Download "SPECIFICATION. Bench top triple quadrupole Mass Spectrometry System coupled to a gas performance Chromatography (GC-MS/MS)"

Transcription

1 SPECIFICATION Bench top triple quadrupole Mass performance Chromatography (GC-MS/MS) 1. Scope These specifications describe the requirements for a bench top triple quadrupole mass spectrometry system coupled to capillary gas chromatography (GC-MS/MS) for quantitative and qualitative persistent organic contaminants (POPs) analysis and other emerging organic contaminants in marine environmental matrices (water, sediments and biota). The instrument shall provide robustness, high sensitivity and selectivity to meet the EU guidelines for POPs and other emerging contaminants. The GC-MS/MS System will strongly assist in the implementation of the current Agency regular environment programs at IAEA Environmental Laboratories. 2. Definitions, Acronyms, and Abbreviations The following definitions, acronyms, and abbreviations shall apply throughout this Specification unless defined otherwise hereinafter: MS shall mean Mass Spectrometer; GC shall mean Gas Chromatography; MRM shall mean Multiple Reaction Monitoring; µm shall mean Micrometer; m/z shall mean Mass-to-charge Ratio; AMU shall mean Atomic Mass Unit; EI: electron impact CI shall mean Chemical Ionization; CID shall mean Collision Induced Dissociation; and RSD shall mean Relative Standard Deviation. IDL shall mean instrument detection limit pg: picogram fg: femtogram 3. Requirements 3.1. Functional and Performance Requirements The GC-MS/MS shall meet the following functional and performance requirements: The tandem quadrupole mass spectrometer shall be compatible with capillary GC columns for high sensitivity trace level analysis and be able to work in EI and CI; The tandem quadrupole mass spectrometer shall be able to quickly switch between MRM and full scan MS acquisition mode to allow precise quantification and qualitative analysis in a single run; The GC-MS/MS shall provide high sensitivity and selectivity to meet the EU guidelines for pesticide and other emerging contaminants; femtogram sensitivity Page 1 of 7

2 Performance specification with capillary injector using helium as carrier gas shall be as given below or better: Mode Test Specification EI full scan 1 pg octafluoronaphtalene (OFN) from m/z 50 S/N 300 :1 to 300 for m/z 272 EI MRM 100 fg OFN for m/z IDL calculated for at least 8 consecutive S/N 7000 :1 IDL < 4fg injection of 10 fg OFN EI MRM precision 10 replicate (sequential) injections of 100 fg Peak area RSD<8% OFN in EI MRM mode (m/z ) PCI MRM 5 pg Benzophenone (BZP) for m/z S/N 2500 :1 NCI SIM 100 fg OFN for m/z 272 S/N 2000: The gas chromatography system (GC) of the GC-MS/MS shall be suitable for operation to the above MS/MS system and compatible with the software controlling the MS/MS system The cleaning and maintenance of the ion source shall be simple The GC system shall come with a backflush technology to remove high-boiling compounds to reduce GC and mass spectrometry maintenance The system shall come with an auto injector and carousel. 3.2 Technical Requirements The MS/MS detector of the GC-MS/MS shall meet the following requirements: The GC-MS/MS shall be equipped with a triple quadrupole detector to perform quantitative and qualitative analysis in a single run; The mass spectrometer shall include an EI source, and a chemical ionization (CI) to work in positive and negative chemical ionization The quadrupole detectors shall have a robust design with appropriate mechanical tolerances and minimum coefficient of thermal expansion to ensure highest mass stability with realigned or no pre-filters to ensure excellent focusing of ions into all the quadrupoles for high sensitivity and resolution; The mass spectrometry shall be capable of simultaneous scan and MRM. The ability to switch between MS (full scan) and MS/MS (including MRM, product ion scanning, parent ion scanning and neutral loss scanning) acquisition in some few seconds to ensure full compatibility for quantitative and qualitative applications in the same run shall be included; Page 2 of 7

3 Analyzer MS-MS specification The minimal working mass range (m/z) shall be from 10 amu up to 1050 amu or more with the following operating scan modes: Full scan MS, selected ion monitoring in both Q1 and Q3, precursor ion scan, product ion scan, neutral loss scan, selected ion monitoring (SIM), multiple reaction monitoring (MRM), high sensitivity product ion scan for library confirmation at trace level, scan for structural elucidation and unambiguous confirmation; Ionization modes: The ionization source shall include electronic ionization (EI) and positive and negative chemical ionization (CI); switching back and forth between CI and EI sources shall be relatively fast within a half day The standard ion source (highly efficient electron multiplier) shall have the highest sensitivity and resolution while giving the tightest and comprehensive specifications for linearity (5 orders of dynamic range) and stability; The ion source shall be computer controlled, supporting manual and automatic tuning; source temperature up to 350 C or more with quick user change over for EI/CI mode (within ½ day) and Auto aligned with dual filament for EI Mass Analyzer/Filters: It shall be based on Quadrupole ideally without any pre and post filters and high ion transmission efficiency to ensure the stability performance and high sensitivity. It shall be easier to tune and shall stay tuned for longer time and shall offer efficient transmission from q1 to q2 and q2 to q The quadrupole shall be sufficiently robust to operate at high temperature (at least 150C) to allow cleaning of any contaminants The ion optics heated to 135 deg C, to ensure no cold spots Collision cell: linear configuration and axial acceleration design for high speed MS/MS analysis, with a preferential use of helium quench gas stream, to reduce the neutral noise. MS/MS capability shall be available via collision induced dissociation (CID) and the collision cell shall be capable of doing MRM analysis with shortest dwell time (minimum 1 msec) without sacrificing sensitivity and eliminate cross-talk to enable multiple MRM transition studies within a single run. The system shall be able to measure at least 750 MRM in one second; The collision energy shall be selectable at least up to 60 ev Resolution: Better than 0.7 amu with mass axis stability of ±0.1 amu over 24 hour period or better. The minimum scan speed shall be equal to or more than 6,000 amu/s, with user selectable scan speed to allow rapid and high quality multicomponent MRM data; Electron energy: 10 to 300 Electron Volt (EV) users selectable or higher Detector: Extended dynamic range Electron Multiplier or off-axis high-energy detector with configuration to direct the charged ion of interest away from the neutrals Page 3 of 7

4 Independently GC heated GC/MS interface Auto-tuning in all ionization modes and manual tuning shall be allowed; the system shall remain tuned for longer time avoiding tuning for better system performance All source parameters shall be adjustable through software; Vacuum system: The turbo molecular pump shall be dual stage air cooled with high capacity (at least for a total gas flow up to 250 ml/min) for fast vacuum and easy and quick change over from EI to CI; the vacuum pump shall be equipped with filter kit systems to trap the exhaust gases from the pump, avoiding to be vented to the atmosphere outside the laboratory; and The Contractor shall certify support by the manufacturer of the pumping installations (in particular of the turbo molecular type) for at least 10 years Fore line pump: dual-stage rotary vane; voltage 230V (same as GC voltage). Gas chromatography system: Systems shall be installed with a split/splitless injector with a pressure range of psi and a programmable temperature vaporizer (PTV) injector GC shall have provision of Electronic Pressure Control (EFC) for all the pneumatic parameters with pressure range of 150 psi for use of wide variety of capillary columns The GC shall be equipped with the capillary flow backflush technology to minimize chemical noise and reduce contamination of the mass spectrometry source. The backflush shall be available in pre column, mid-column or post-column mode The injection port septa and liners shall be able to be changed without venting the MS. Replacement timing shall be kept by the software The GC shall be able to work with both, constant flow and constant pressure mode. A system providing constant retention time, even after column change or maintenance, will be an asset The mass spectrometer shall support the twin-line option of having 2 capillary columns installed with two different injectors, with the option of selecting the column connected to the ion source without venting the MS, and without loss of sensitivity The GC-MS/MS system shall have a saving mode for reduced consumption of helium gas and change to nitrogen carrier gas during system standby. Autosampler: The auto sampler of the GC-MS/MS shall meet the following requirements: Page 4 of 7

5 The auto-sample injector shall have syringe speed options of fast and slow and plunger speed options of slow, medium and fast to allow for injection of samples in a wide range of viscosity It shall have reproducibility lower than 0.4 % RSD with variable injection volumes between 0.1 to 5 µl, in 0.1 µl increments with excellent sample recovery The auto-sampler shall enable samples in conical bottom (tapered) vials of 10 X 32 mm At least 100 sample vials of 1.8 ml capacity with programmable needle wash before and after injection, and sample cooling facility shall be included The carousel shall hold up great solvent capacity (> 4ml-solvent) and waste vials for thorough syringe rinsing between samples The auto injector controlling program shall be flexible and allow for temporary change of sample sequence In addition to making traditional injections, the auto injector shall be able to perform injection in solvent flush mode (air-sample-air-solvent) or solvent flush with a second solvent mode (air-sample air-internal standard-air-second solvent). These modes shall be supported to allow for analysis of samples containing a wide boiling point range and for the addition of internal standard from a vial on the carrousel The samples in the auto-sampler shall be kept at constant temperature, with a temperature programmable setting ranging from 4 to 20 o C The needle height control shall be programmable in 0.1 mm increments; and The auto-sampler and the carousel shall be automated controlled through GC-MS/MS software Additional items: A GC-MS/MS start-up kit as standards for EI and CI, including, the essential spares and consumables for the ion source, analyzer, gas chromatography system, column for pesticide analysis, vials; Also a complete kit of tools for handling and servicing of any part of the GC-MS/MS (spanner, Allen keys, drivers); PC based data system with all required software, monitor, color laser printer and an English language keyboard. The GC-MS/MS shall be supplied with easy to use control of all the modules of GC and MS and acquisition software, and at least one library (original NIST library) for identifications of pesticides and their metabolites and emerging contaminants using MS data. The software shall have new compound acceptance criteria as latest EU directives and automated quantitative data processing features. It shall also include self-diagnosis and auto-tuning. The shelf-diagnosis feature of the software shall enable diagnosis of the mass spectrometer and all connected GC units. To facilitate the export of the data, the software shall be compatible with Windows. The files of acquisition of the data shall have to be usable for the EXCEL spread sheet; and Page 5 of 7

6 Free and unlimited number of software licenses for Windows Desirable features for the GC-MS/MS: Pre-configured and optimized methods, with reference standards, for analysis of pesticide residues and other emerging contaminants; and A MRM list of at least 1000 compounds with a minimum of 4 different MRM for each compounds to allow better adaptation of the methods in different matrixes. 4. Marking The GC-MS/MS shall have all safety markings in the English (mandatory) and French (optional) languages. 5. Quality Requirements The GC-MS/MS shall be manufactured, shipped and installed in accordance with the Contractor s ISO quality assurance system or an equivalent quality assurance system. 6. Testing and Acceptance The GC-MS/MS, prior to shipment, shall be tested for conformance of the System with manufacturer s performance specifications and the minimum requirements specified herein; The GC-MS/MS, after installation, shall be tested by the Contractor together with the End-User to demonstrate that the performance meets the manufacturer s performance specifications and the minimum requirements specified herein as determined by the IAEA and the End-User; The results of the above tests on the installed GC-MS/MS shall be documented by the Contractor in an acceptance protocol that shall be signed by the IAEA. If the tests results after installation do not meet the minimum requirements as specified, the manufacturer shall resolve the problem by correcting the installation of the equipment, or if necessary, change the equipment part responsible for the error. 7. Installation and Training 7.1. The Contractor shall install and demonstrate the system at the Marine Environment Laboratories in Monaco; 7.2. The Manufacturer shall complete the installation and validate that the GC-MS/MS system meet all the above specifications and requirements The Contractor shall provide the required training of three (3) IAEA staff in the routine operation and maintenance of all instruments and modules supplied immediately after the installation of the System. Additional on-site advanced training for IAEA staff after six (6) months of the installation is also required. 8. Deliverable Data Items The Contractor shall provide complete sets of operation and servicing manuals with circuit diagrams and circuit descriptions in the English (mandatory) and French (optional) languages, along with Page 6 of 7

7 diagnostic tools including interface cards and software, plus any specialized mechanical tools required for instrument servicing and repair; 9. On-Site Technical Support and (optional) Additional Warranty 9.1. The Contractor shall demonstrate the availability of a good and competent technical support in France or Monaco, and, when a problem is reported The Contractor shall provide free telephone support for the lifetime of the instrument, regardless of whether it is covered under warranty of service contract The instrument model shall be recent enough to guarantee that the part supplies are available for a minimum of ten (10) years after installation The Contractor shall provide the IAEA with all manufacturer s warranties but not limited to a warranty for a period of twelve (12) months from the date after the equipment has been commissioned If required by the IAEA, the Contractor shall provide warranty for the complete equipment, including independent accessories, application, technical support, and an annual maintenance visit for three (3) years following the expiration of the twelve (12) months manufacturer s warranties. Page 7 of 7

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585 GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc. 850 Centerville

More information

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS) Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph-mass spectrometer (GC-MS). 2.0 Scope

More information

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel Application Note Energy and Chemicals Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel Applying the Energy Institute Method IP 8 with an Agilent J&W

More information

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Application Note Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Authors Kelly Beard and James McCurry Agilent Technologies, Inc. Abstract An Agilent

More information

Technical Procedure for Gas Chromatography (GC-FID)

Technical Procedure for Gas Chromatography (GC-FID) Technical Procedure for Gas Chromatography (GC-FID) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph (GC- FID). 2.0 Scope This procedure applies to all

More information

PRODUCT SPECIFICATION. Product Specification Crystal 9000 Gas Chromatograph. Chromatec Crystal Laboratory Gas Chromatography System

PRODUCT SPECIFICATION. Product Specification Crystal 9000 Gas Chromatograph. Chromatec Crystal Laboratory Gas Chromatography System Chromatec Crystal 9000 Laboratory Gas Chromatography System GC Chromatec-Crystal 9000 Crystal 9000 is a leading gas chromatograph in Chromatec GC product line. Crystal 9000 has highest performances and

More information

Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS)

Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS) Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the pyrolysis-gas chromatograph-mass spectrometer

More information

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI) Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Bill Wilson and Chin-Kai Meng Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

Agilent 7693A Automated Liquid Sampler

Agilent 7693A Automated Liquid Sampler Agilent 7693A Automated Liquid Sampler Specifications Overview The Agilent 7693A is a state-of-the-art sample handling and injection system that provides the highest levels of precision and reliability

More information

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE AN HPLC THAT WORKS AS HARD AS YOU DO Confidence in your results comes with high performance, day-to-day reliability, and overall system robustness.

More information

Agilent 7693A Automated Liquid Sampler

Agilent 7693A Automated Liquid Sampler IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7693A Automated Liquid Sampler

More information

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS FLASH VAPORISATION INJECTION Split Splitless On-Column COOL INJECTION Large Volume Injection (LVI) On-Column On-Column-SVE (with solvent vapour

More information

Large Volume Injection of Polycyclic Aromatic Hydrocarbons

Large Volume Injection of Polycyclic Aromatic Hydrocarbons JSB is an authorised partner of Large Volume Injection of Polycyclic Aromatic Hydrocarbons Application Note - Environmental #113 Author Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Abstract

More information

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System Application Note Authors Frank David Research Institute for Chromatography, Pres.

More information

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors Model 8610C Gas Chromatograph Mounts up to Six Detectors and Five Injectors Ambient to 400 C Temperature Programmable Column Ovens Dimensions: 19 wide x 13.5 high x 14.5 deep Implement virtually any EPA

More information

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D759 Application Petrochemical Author James D. McCurry Agilent Technologies 285 Centerville

More information

DFS High Resolution GC/MS High Performance Magnetic Sector GC/MS

DFS High Resolution GC/MS High Performance Magnetic Sector GC/MS mass spectrometry DFS High Resolution GC/MS High Performance Magnetic Sector GC/MS The DFS High Resolution GC/MS is the highest performance magnetic sector mass spectrometer ever built for target compound

More information

Agilent 6850 Series II Network GC System G2630A Performance Specifications

Agilent 6850 Series II Network GC System G2630A Performance Specifications IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Agilent 6850 Series II Network GC

More information

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application Petroleum Authors ChunXiao Wang Agilent Technologies (Shanghai) Co.,Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai

More information

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS Rev 1 3/6/2004 Selerity Technologies Inc. www.selerity.com SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS What is supercritical fluid chromatography? A chromatographic technique in

More information

Agilent Multimode Inlet for Gas Chromatography

Agilent Multimode Inlet for Gas Chromatography Agilent Multimode Inlet for Gas Chromatography Technical Note Agilent Multimode Inlet for the 7890A GC Designed to give you ease of use and maximum flexibility, the Agilent Multimode Inlet does everything

More information

GCMS-QP2010 Series. Consumables. Shimadzu Gas Chromatograph/Mass Spectrometer

GCMS-QP2010 Series. Consumables. Shimadzu Gas Chromatograph/Mass Spectrometer GCMS-QP2010 Series Consumables Shimadzu Gas Chromatograph/Mass Spectrometer Diag. # 1) Injection Port - Septum Nut 2) Injection Port - Needle Guide 3) Septum 4) O-ring for Glass Insert 5) Glass Insert

More information

Application Note. Abstract. Authors. Environmental Analysis

Application Note. Abstract. Authors. Environmental Analysis High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID using the Agilent Low Thermal Mass (LTM II) System Application Note Environmental Analysis Authors Frank David and Karine Jacq Research

More information

Simulated Distillation Analyzers, Software, Standards, Consumables, Training

Simulated Distillation Analyzers, Software, Standards, Consumables, Training Simulated Distillation Analyzers, Software, Standards, Consumables, Training www.separationsystems.com Offering the Fullest Range of Optimized Solutions Simulated distillation (SimDis) has been used to

More information

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector Detection of Sulfur Compounds in Natural Gas According to ASTM D554 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector Application Note Author Rebecca Veeneman Abstract Sulfur compounds in natural

More information

Setting up SilFlow for BackFlush in your GC

Setting up SilFlow for BackFlush in your GC Setting up SilFlow for BackFlush in your GC What is backflush and why use it? The BackFlush system eliminates the need to bake heavy sample fractions off the capillary column. Oils, tars and other semivolatile

More information

URB '-carbamoylbiphenyl-3-yl cyclohexylcarbamate. DEA Reference Material Collection. Form Chemical Formula Molecular Weight Melting Point ( o C)

URB '-carbamoylbiphenyl-3-yl cyclohexylcarbamate. DEA Reference Material Collection. Form Chemical Formula Molecular Weight Melting Point ( o C) O NH O NH O. GENERAL INFORMATION IUPAC Name: 3'-carbamoylbiphenyl-3-yl cyclohexylcarbamate CAS#: 56-08-6 Synonyms: Source: Appearance: UV max (nm): KDS-03 DEA Reference Material Collection White powder

More information

From Helium to Hydrogen: GC-MS Case Study on SVOCs in Water

From Helium to Hydrogen: GC-MS Case Study on SVOCs in Water From Helium to Hydrogen: GC-MS Case Study on SVOCs in Water Jessie Butler Alexander N. Semyonov December 13, 2012 Helium? It s gone! Not only at Disneyland Resort 2 Helium shortage grounds Mickey at Tokyo

More information

S 1125 HPLC Pump System

S 1125 HPLC Pump System Routine S 170 EXPERT HPLC HPLC System SYSTEMS 500 The Sykam ROUTINE HPLC System S 500 is intended for all routine analysis. The system configuration is highly variable and several upgrade options make

More information

Rapid Determination of 54 Kinds of Pesticide Residues in Vegetables by PTV-GC-MS/MS

Rapid Determination of 54 Kinds of Pesticide Residues in Vegetables by PTV-GC-MS/MS PO-CON1517E Rapid Determination of 54 Kinds of Pesticide Residues in Vegetables by PTV-GC-MS/MS Pittcon 15 22-11P Shiheng Luo, Feifei Tian, Xizhi Wang, Guixiang Yang, Taohong Huang, Shin-ichi Kawano, Yuki

More information

Analysis of Petroleum Fractions by ASTM D2887

Analysis of Petroleum Fractions by ASTM D2887 Analysis of Petroleum Fractions by ASTM D2887 Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 2582 Key Words Simulated distillation, D2887, TRACE TR-SimDist Abstract ASTM

More information

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS MS Identification of Trace level Impurities from a Non-MS Compatible Mobile Phase Using ACQUITY UPLC System with 2D Technology by Heart-cutting and Online Sample Concentration Bronsky Gopinadh, Dilshad

More information

Oil Free Vacuum Pumps for Mass Spectrometry

Oil Free Vacuum Pumps for Mass Spectrometry Oil Free Vacuum Pumps for Mass Spectrometry Technical Overview Agilent is seeing additional impetus for the switch to dry pump technology from these departments within enduser customer organizations: Scientific

More information

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Application Note Petrochemicas High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Author James D. McCurry, Ph.D. Agilent Technologies, Inc. Abstract An Agilent

More information

Using the PSD for Backflushing on the Agilent 8890 GC System

Using the PSD for Backflushing on the Agilent 8890 GC System Application Note Petrochemicals Using the PSD for Backflushing on the Agilent 889 GC System Author Brian Fitz Agilent Technologies, Inc. Wilmington, DE, USA. Abstract An Agilent 889 series GC equipped

More information

The Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over

The Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over The Agilent 12 Series high performance autosampler SL: Area precision, injection volume linearity, imum accessible volume, carry-over Technical Note 8 6 4 2.5 1 Introduction A main performance criterion

More information

Fausto Munari e Andrea Cadoppi ThermoFisher - Italy

Fausto Munari e Andrea Cadoppi ThermoFisher - Italy The world leader in serving science Ultra Fast GC Determination of Total Hydrocarbons (C7-C40) and BTEX in Water and Soils through Direct Resistively Heated capillary columns and Robotic Autosampler. Fausto

More information

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis Introduction Petroleum & Petrochemical Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis By Katarina Oden, Barry Burger, and Amanda Rigdon Crude oil consists of thousands of different

More information

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis Page 1 Alternate Carrier Gas Considerations and Faster GC Analysis Faster GC Total Analytical Cycle Times A Variety of Approaches Pre-Run ALS Set-Up Chromatographic Run Post-Run Bake-Out Post-Run Cool-Down

More information

Application. Gas Chromatography June 1995

Application. Gas Chromatography June 1995 Determining Oxygenates in Gasoline: ASTM Method D Application Gas Chromatography June 99 Authors Michael J. Szelewski Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-60 USA Matthew S. Klee

More information

Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides

Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides Application Note Food Testing & Agriculture Authors Limian Zhao and Chin-Kai Meng Agilent Technologies, Inc. 85 Centerville

More information

Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter

Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter Cd The Chrom Doctor Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands. In Part 1 of this series we focused

More information

2-D GC/MS and the Deans Switch: Theory and Applications. Fred Feyerherm, Agilent Technologies Inc. Houston, Tx

2-D GC/MS and the Deans Switch: Theory and Applications. Fred Feyerherm, Agilent Technologies Inc. Houston, Tx 2-D GC/MS and the Deans Switch: Theory and Applications Fred Feyerherm, Agilent Technologies Inc. Houston, Tx Introduction Simplified 2-D GC can be used to accomplish the separation of specific analytes

More information

C2, C3, C4 Monomer Analysis

C2, C3, C4 Monomer Analysis C2, C3, C4 Monomer Analysis Malgorzata Sierocinska Agilent Technologies Waldbronn Page 1 Why Analyze Monomers? To Insure Consistent Production of High Quality Polymer Protect against food contamination

More information

NEW HPLC - SYSTEM LINE

NEW HPLC - SYSTEM LINE NEW HPLC - SYSTEM LINE S c h a m b e c k S F D G m b H Drieschweg 13A D-53604 Bad Honnef/Germany Phone: +49 2224 9239-0 E-Mail: info@schambeck-sfd.com Internet: www.schambeck-sfd.com S 9425 HPLC solvent

More information

Method Development for Capillary GC Systems. Slide 1

Method Development for Capillary GC Systems. Slide 1 Method Development for Capillary GC Systems Slide 1 AREAS TO OPTIMIZE Injector Carrier gas Column temperature Slide 2 COMMON INJECTOR MODES Vaporization Injection Modes Megabore Direct Split Splitless

More information

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard 7890 GC Standard Agilent Preventive Maintenance provides factory recommended service for your analytical systems to assure reliable operation and the accuracy of your results. Delivered by highly-trained

More information

MSD EI and CI Source Cleaning and Installation. Technical Overview. Introduction

MSD EI and CI Source Cleaning and Installation. Technical Overview. Introduction MSD EI and CI Source Cleaning and Installation Technical Overview Charles Thomson, Max Ruemler, Mickey Freed, Dave Peterson, and Harry Prest Introduction During the course of operation, eventually GC-MSD

More information

Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut

Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut Application Note Environmental Author Ken Lynam Agilent Technologies, Inc. Abstract Specially designed self-tightening

More information

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC Technical Overview Trace gas analysis is a challenge in today s world. The ability to analyze lower component

More information

Application Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER

Application Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D7423-09 using AC OXYTRACER Fast Analysis in

More information

PATROL UPLC Process Analysis System for Production

PATROL UPLC Process Analysis System for Production PATROL UPLC Process Analysis System for Production The PATROL UPLC Process Analysis System for Production is an integrated system solution designed and engineered to perform online and atline reaction

More information

ACQUITY UPLC I-Class System with 2D Technology

ACQUITY UPLC I-Class System with 2D Technology ACQUITY UPLC I-Class System with 2D Technology The Waters ACQUITY UPLC I-Class System with 2D Technology allows chemists to increase sensitivity and selectivity, eliminate unwanted interferences, characterize

More information

Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology

Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology Electronic Noses An electronic nose produces a recognizable response based

More information

Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS

Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS Application Environmental Authors Wei Luan and Chuanhong Tu Agilent Technologies (Shanghai) Co., Ltd. 412 Ying

More information

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc.

More information

White Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information

White Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Abstract High Temperature Simulated Distillation (High Temp SIMDIS) is one of the most frequently used techniques to determine

More information

TSHR 6000 series TN/TS/TX analyzer

TSHR 6000 series TN/TS/TX analyzer TSHR series TN/TS/TX analyzer Versatile, Accurate and Reliable Trace Level Nitrogen, Sulfur, Chlorine Analysis Petrochemicals Refineries Analytical Services Research Analysis The TSHR series TN/TS/TX Analyzer

More information

Agilent Multimode Inlet

Agilent Multimode Inlet Agilent Multimode Inlet Large Volume Injection Tutorial Agilent Technologies Notices Agilent Technologies, Inc. 2009 No part of this manual may be reproduced in any form or by any means (including electronic

More information

The PMAC Dynamic Scale Loop

The PMAC Dynamic Scale Loop The Dynamic Scale Loop Original System for evaluation of scaling tendency and inhibitor performance The System is an invaluable research and development tool for technologists responsible for the design,

More information

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels James D. McCurry Senior Scientist Agilent Technologies Wilmington, DE USA Page 1 Application Summary There is a need to measure

More information

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector Authors: The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector Introduction Specification of the hydrocarbon composition of LPG is required as traces

More information

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI) Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Brian Fitz and Bill Wilson Agilent Technologies, Inc. 285 Centerville Road Wilmington,

More information

GC Method Compliance and Large Valve Oven Application

GC Method Compliance and Large Valve Oven Application GC Method Compliance and Large Valve Oven Application Solution Focused Kelly Beard June 20, 2016 Compliant Methods Compliant Methods What are our options? Analyzers and Analyzer Kits High Volume Well Known

More information

Technical Procedure for the Examination of Arson Evidence

Technical Procedure for the Examination of Arson Evidence Technical Procedure for the Examination of Arson Evidence 1.0 Purpose This technical procedure shall be followed for the examination of arson (fire debris) evidence. 2.0 Scope This procedure applies to

More information

Pipetted Liquid Sampling. Automated sample preparation platform for online LC-MS/MS bioanalysis

Pipetted Liquid Sampling. Automated sample preparation platform for online LC-MS/MS bioanalysis Pipetted Liquid Sampling Automated sample preparation platform for online LC-MS/MS bioanalysis Integrated Sample Processing SCAP PLS system components mounted on a robotic platform Integrated Sample Processing

More information

Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D

Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Instructions Agilent Valve Kit G4231A/C and G4232C/D - Instructions This technical note describes the installation and application of the Agilent

More information

Agilent and ASTM. Update on Recent Activities. Page 1

Agilent and ASTM. Update on Recent Activities. Page 1 Agilent and ASTM Update on Recent Activities Page 1 ASTM Committees for Refining, Fuels, Petroelum Products ASTM International D2 Petroleum products & lubricants D3 Gas fuels (natural gas) gas fuels D16

More information

LIABLE RESULTS AND RELIABLE RESULTS AND EXCEPTIONAL REPRODUCIBILITY

LIABLE RESULTS AND RELIABLE RESULTS AND EXCEPTIONAL REPRODUCIBILITY STATIC HEADSPACE SAMPLER SHS HISTORICALLY HISTORICALLY THE THE BEST BUST A ROBUST AND FLEXIBLE SYSTEM TO SYSTEM MEET COMPLEX TO AND VERSATILE MEET NEEDS COMPLEX AND LIABLE RESULTS AND RELIABLE RESULTS

More information

GC Inlets. An Overview. Simon Jones GC Applications Engineer

GC Inlets. An Overview. Simon Jones GC Applications Engineer GC Inlets An Overview Simon Jones GC Applications Engineer Types of Inlets Purged Packed Split / Splitless Cool On Column Programmable Temperature Vaporization Volatiles Interface Multi Mode Inlet Where

More information

Liquid Chromatography

Liquid Chromatography SCINTOMICS Liquid Chromatography Always the right system for your needs. SCINTOMICS: Always the right system for your needs Advanced Research Daily QC Routine and beyond High Performance analytical/semi-preparative

More information

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract. Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers Application Note Abstract Versa With the rising prices of fossil fuels, more emphasis is being put on renewable resources

More information

Choose for Your Instrument. GC Supplies - Agilent Instrument Quick Pick Guide

Choose for Your Instrument. GC Supplies - Agilent Instrument Quick Pick Guide Choose for Your Instrument GC Supplies - Agilent Instrument Quick Pick Guide Syringes Septa Inlet Liners O-rings and Sealing Rings SilTite FingerTite Ferrules Ferrules Your chromatography analysis does

More information

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584 Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN 14105 and ASTM D6584 Introduction With today s increasing concern for the environment and the depletion of fossil

More information

TSHR 7000 series TN-TS-TX analyzer

TSHR 7000 series TN-TS-TX analyzer TSHR 7000 series TN-TS-TX analyzer Enhanced Capabilities with Powerful Performance for Trace Level Nitrogen, Sulfur, Chlorine Analysis Petrochemicals Refineries Analytical Services Research Analysis The

More information

C196-E081C. Ultra High Performance Liquid Chromatograph Nexera X2. Specifications

C196-E081C. Ultra High Performance Liquid Chromatograph Nexera X2. Specifications C196-E081C Ultra High Performance Liquid Chromatograph Nexera X2 Specifications Specifications System Controllers CBM-20A / 20Alite CBM-20A (228-45012-XX) CBM-20Alite (228-45011-XX) Connectable units Solvent

More information

Introducing the SCIEX Triple Quad and QTRAP LC-MS/MS System with optional SelexION + Technology

Introducing the SCIEX Triple Quad and QTRAP LC-MS/MS System with optional SelexION + Technology RUO-MKT-11-2491 For Research Use Only. Not for use in diagnostic procedures Introducing the SCIEX Triple Quad and QTRAP 6500 + LC-MS/MS System with optional SelexION + Technology SCIEX, Concord, Ontario

More information

Phenomenex Exclusive! Inlet Base Seals. Agilent 5000, 6000 and. Guaranteed fit for Series GCs. Easy Seals

Phenomenex Exclusive! Inlet Base Seals. Agilent 5000, 6000 and. Guaranteed fit for Series GCs. Easy Seals Phenomenex Exclusive! Easy Seals Inlet Base Seals Guaranteed fit for Agilent 5000, 6000 and 7000 Series GCs The Easy Seals Gold Standard No Strained Hands No Troublesome Washers or Adaptors Continuing

More information

Online sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve

Online sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve Online sample cleanup on the Agilent 129 Infinity LC using a built in 2-position/6-port valve Analysis of Sudan red compounds in paprika powder Application Note Food Authors Angelika Gratzfeld-Huesgen

More information

AKVdiesel. Automated kinematic Viscometer key features. Fully complies with ASTM D445. Two measuring positions. Remove cap and pour in sample

AKVdiesel. Automated kinematic Viscometer key features. Fully complies with ASTM D445. Two measuring positions. Remove cap and pour in sample Suitable for measuring gas oils, distillate fuels and diesel fuels, as specified in ASTM D975, EN590 and ISO 8217 (marine fuels). The RHEOTEK AKV-diesel is a fully automated viscometer system suitable

More information

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector Application Note Hydrocarbon Processing Authors ChunXiao Wang Agilent Technologies

More information

HPLC SYSTEM S 500 SERIES

HPLC SYSTEM S 500 SERIES HPLC SYSTEM S 500 SERIES The Sykam HPLC System S 500 Series is intended for all routine analysis as well as for the ambitious analyst. The system configuration is highly variable and several upgrade options

More information

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column Cory S. Fix, Director of Application Development cory.fix@vgcchromatography.com Willie Steinecker,

More information

Liquid Injection Techniques in GC and GC-MS

Liquid Injection Techniques in GC and GC-MS Liquid Injection Techniques in GC and GC-MS Inge De Dobbeleer Regional Marketing Manager GC and GC-MS, EMEA Thermo Fisher Scientific, Breda/The Netherlands The world leader in serving science Content Hot

More information

HIGH PRESSURE SWITCHING VALVES

HIGH PRESSURE SWITCHING VALVES HIGH PRESSURE SWITCHING VALVES Rheodyne offers high pressure switching valves to simplify procedures and improve the speed, resolution, and sensitivity of HPLC analysis. The switching valves are available

More information

Perfluorinated compounds by LC MS

Perfluorinated compounds by LC MS Analysis, fate and risks of organic contaminants in river basins under water scarcity Valencia 7-8 ebruary 2010 Perfluorinated compounds by LC MS Marinella arré, Marta Llorca, rancisca Pérez Yolanda Picó,

More information

Accuracy Starts Here. Agilent Captiva Filtration Selection Guide

Accuracy Starts Here. Agilent Captiva Filtration Selection Guide Accuracy Starts Here Agilent Captiva Filtration Selection Guide TAbLe of CoNTeNTS Tips & Tools Learn the core concepts surrounding Solid Phase Extraction and best practices for Sample Prep. View the video

More information

one filter having K between 1.5 and 2.5 m -1

one filter having K between 1.5 and 2.5 m -1 CHAPTER IV : CONFORMTY OF PRODUCTION PROCEDURE FOR TESTING OF SMOKE METER. 1 Physical Check It shall consist of checking - 1) Suitability and label on the instrument for the intended use. 2) Identification

More information

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard Agilent Preventive Maintenance provides factory recommended service for your analytical systems to assure reliable operation and the accuracy of your results. Delivered by highly-trained and certified

More information

Stabinger Viscometer SVM 3001

Stabinger Viscometer SVM 3001 Stabinger Viscometer SVM 3001 Expect More Welcome to New Viscometry Measuring the viscosity of oils and fuels is all about certainty and compliance. Two approaches get you there: the traditional glass

More information

Thermo Scientific Dionex UltiMate 3000 Thermostatted Column Compartment

Thermo Scientific Dionex UltiMate 3000 Thermostatted Column Compartment chromatography Thermo Scientific Dionex UltiMate 3000 Thermostatted Column Compartment Product Specifications Thermo Scientific Dionex UltiMate 3000 products are UHPLC compatible by design, establishing

More information

Liquefied Gas Injector. Solution for the Sampling and Analysis of Liquefied Gases

Liquefied Gas Injector. Solution for the Sampling and Analysis of Liquefied Gases Liquefied Gas Injector Solution for the Sampling and Analysis of Liquefied Gases Safe and Representative Sampling of Liquefied Gases The analysis of impurities and contaminants in liquefied gases is an

More information

Jaap de Zeeuw Restek Middelburg, The Netherlands. Copyrights: Restek Corporation

Jaap de Zeeuw Restek Middelburg, The Netherlands. Copyrights: Restek Corporation Liner Selection in GC Jaap de Zeeuw Restek Middelburg, The Netherlands What is a liner? A liner is an inert glass tube that is positioned inside the injection port of the GC The sample solution is introduced

More information

Application Note. Author. Introduction. Energy and Fuels

Application Note. Author. Introduction. Energy and Fuels Analysis of Free and Total Glycerol in B-100 Biodiesel Methyl Esters Using Agilent Select Biodiesel for Glycerides Application Note Energy and Fuels Author John Oostdijk Agilent Technologies, Inc. Introduction

More information

Quantification and Characterization of Sulfur in Low-Sulfur Reformulated Gasolines by GC-ICP-MS Application

Quantification and Characterization of Sulfur in Low-Sulfur Reformulated Gasolines by GC-ICP-MS Application Quantification and Characterization of Sulfur in Low-Sulfur Reformulated Gasolines by GC-ICP-MS Application Authors Steven M. Wilbur and Emmett Soffey Agilent Technologies 338 146th Place SE Bellevue,

More information

CrossLab. Supplies for Thermo Scientific GC Systems. work for you non-agilent GCs. Put Agilent innovation and quality to. Agilent

CrossLab. Supplies for Thermo Scientific GC Systems. work for you non-agilent GCs. Put Agilent innovation and quality to. Agilent works with BRUKER PERKINELMER SHIMADZU THERMO SCIENTIFIC AND MORE Put innovation and quality to work for you non- GCs is a growing portfolio of supplies critical to instrument performance and productivity.

More information

SUPPRESSION IN ANALYSIS OF ARTIFICIAL SWEETENERS WITH

SUPPRESSION IN ANALYSIS OF ARTIFICIAL SWEETENERS WITH APPROACHES TO DECREASE ION SUPPRESSION IN ANALYSIS OF ARTIFICIAL SWEETENERS WITH UPLC-TQ MS Noora Perkola Finnish Environment Institute 3rd Nordic MS Symposium September 12 2012, Jurmala, Latvia ARTIFICIAL

More information

Elemental Analysis Advanced

Elemental Analysis Advanced better analysis counts Elemental Analysis Advanced Rapid sulfur measurement plus 12 elements from P to Zn. ASTM D4294 ISO 8754 IP 336 Advanced Software and Data Management Petra offers the latest software

More information

Technical Offer. Antek Model 6200 Analyzer Determine Total Sulfur and Total Nitrogen in a single analyzer

Technical Offer. Antek Model 6200 Analyzer Determine Total Sulfur and Total Nitrogen in a single analyzer Technical Offer Antek Model 6200 Analyzer 2015.4 Determine Total Sulfur and Total Nitrogen in a single analyzer Meet sulfur regulations and control your process with continuous sulfur and/or nitrogen measurements

More information

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application 7890A/7890B GC Overview October 29, 2014 1 Definitions Carrier Gas Pressurized gas used to transport the sample through the system.

More information