Fausto Munari e Andrea Cadoppi ThermoFisher - Italy
|
|
- Joel George
- 5 years ago
- Views:
Transcription
1 The world leader in serving science Ultra Fast GC Determination of Total Hydrocarbons (C7-C40) and BTEX in Water and Soils through Direct Resistively Heated capillary columns and Robotic Autosampler. Fausto Munari e Andrea Cadoppi ThermoFisher - Italy
2 Summary Official regulations for Total Hydrocarbons: GC as replacement of IR Ultra Fast GC and Triplus Autosampler ISO with Ultra Fast GC ISO (Mod.) with Ultra Fast GC 2
3 IR and GC methods Infra Red GC Method: EPA ISO TNRCC 1005 EPA 8015 ISO/TR ISO/TR (for Oil in soil and sediments) Extraction Solvent: CCl 4, Freon n-pentane, (banned) n-hexane TTCE Analysis Time: around 2 min 20 min 2 min by Ultra Fast GC 3
4 How fast is UFGC? UFGC (direct resistive heating) Rate ( C/min) Rate ( C/s) Conventional fast oven Final oven temperature ( C) 4
5 Ultra Fast GC: Requirements Ultra Fast heating capability ( C/min) Short capillary columns (typically 2-5m narrow-bore; or 0.32mm i.d. when separation power is not an issue) High pressure pneumatic control capability Fast Detector (High Frequency Response) 5
6 Direct column heating technology The heating element and the T sensor are wrapped around the column The column is mounted in a cage ULTRA FAST Column Module Dedicated heated interfaces for injector and detector NO oven heating is required Heater Capillary column Column assembly Ceramic fiber Temperature sensor The column module includes both heating and temperature sensing elements 6
7 The column module in the oven Column modules are housed in the GC oven and connected to inlet (SSL or PTV) and detector as a conventional capillary column Configurations used for TPH analysis: Additional heated block FID Injector SSL / FID PTV / FID Trace GC Ultra oven 7
8 Trace GC Ultra Column Module FEATURES Assemblies of 2, 5, 10 m ( mm i.d.) Up to 20 C/s (1200 C/min) temperature programming rates Broad temperature range (from 35 to 370 C) Cooling from 350 C to 50 C in about 1 min If needed, also cryo cooling is available (0 C minimum temp) Significant implications over conventional GC Faster analysis times Higher column lifetime Higher sensitivity: capability to handle trace analysis 8
9 Increased sensitivity - Peak compression effect 1.6E E E+05 nc20 (about 1ng) area (uv.s) 300 C/min 0.1 mm id column PW 1/2 = 0.10 s FID signal (uv) 1.0E E E E E C/min 0.32 mm id column PW 1/2 = 1.6 s Narrower peaks result in better signal to noise ratio 0.0E time (s) 9
10 E7 E8 E20 E9 E17 E18 E19 E10 E16 E15 E11 E12 E13 E14 E7 E8 E20 E19 E9 E18 E17 E10 E16 E15 E11 E14 E12 E13 Ultra Fast Analysis of FAME Classical Analysis 30 m, 0.32 mm id, 0.25 um Column TR-5 10 C/min 20 min 1 min Ultra Fast GC Analysis 5m, 0.10 mm id, 0.10 um Column TR C/min 1min 10
11 Retention Time repeatability study 10 C/s Carrier gas mode: constant flow (1 ml/min H2) Column 5m x 0.1 mm id, 0.1 um TR-5 70 to 330 C Split 1:200 Components 35 s 5 C/s 5 C/s nc10 nc12 nc nc20 nc22 nc23 4 nc16 9 nc25 60 s 5 5 nc C/s 2 C/s 110 s Time (s) 11
12 Retention Time repeatability Clear Identification: SD << peak width Rate ( C/s) Component Retention times Mean (s) SD (s) Peak widths at ½ h at 5% h 6 SD window* C C12 C C C C12 C C C C12 C C * In a ± 3 SD span there is a probability of 99.7 % for a measure to fall in 12
13 ISO ISO Norm Extraction of 900ml of water with 50 ml of n-hexane 2. Purification of the extract through Florisil, to remove interferences 3. Concentration of the extract from 50 to 1ml 4. GC analysis (FID detector) Solvent: n-hexane Range of hydrocarbons: C10-C40 13
14 Hydrocarbon analysis in water and soils Soil extract (50 mg/l tot. hydroc.) C10 C40 Splitless injection Total hydrocarbons 2 min time (min) Column :TR- 5m, 0.32 mm id, 0.25 um FT, T prog.: 40 (20s), 3.3 C/s, 350 (30s), Sample: Soil extract (50 ng/ul tot hyd.), Injection: 1 ul SPLITLESS, Splitless time: 20 s, Carrier: He at 5 ml/min CF 14
15 Hydrocarbon analysis in water and soils High linearity from 5 to 500 mg/l C40/C20 area ratio > 0.9 (required 0.8) RSD% on group peak absolute area 0.8% Area counts (*10^6) Calibration curve mg/l R Concentration (mg/l) 15
16 ISO SSL - UFM - Fast FID Trace GC Ultra Compliance with the requirements of the official methods Sampling Run Time Cooling Time 5 min High sample/day throughput! 100 samples in about 8h (a work-day)! 16
17 ISO QC sample (50 ng/µl) Column: 5 m x 0.32 mm id, 0.25 µm TR-5 Temperature Program: 40 C (0.3 ), 200 C/min to 350 C (0.5 ) Injection volume: 1 µl - splitless (mvolt) C10 Ultra Fast GC C Gain in speed Faster run Faster cooling Gain in sensitivity Peaks compression: peaks are narrower and consequently higher (mvolt) (sec) C Conventional GC (sec) 2 min C40 20 min 17
18 No more extract re-concentration! Ultra Fast GC LIQUID EXTRACTION SAMPLE CLEAN-UP SOLVENT EVAPORATIVE CONCENTRATION Conventional GC The sample reconcentration step can be by-passed! 18
19 ISO (Mod.) 2003: OSPAR proposed a new reference method ISO (Mod.), to get in force from January 2007 in North Sea area Solvent: n-pentane Range: C7-C40 BTEX havetobeseparatedaswell! Is it still possible to fulfil these specs at high speed of analysis? YES! using PTV Injector and Triplus Autosampler 19
20 Optimized PTV Injector Split Programmable Temperature Vaporizer (PTV) Cool Injection Prevents in-needle Discrimination During Transfer from the Needle to the Injector Higher Boiling Point Compounds Operate PTV independently of Column at Optimum Cool Temperature for Injection Sample evaporation zone Cold Permanently Heated Block at Base Avoids Cold Spots for the Sample Transfer to the Column Sample transfer line Hot 20
21 TriPlus Trio: AS, HS, and SPME Flexible Sampling Solutions Standard and Advanced Liquid sampling: From nano-volumes to Large Volume techniques (up to 450 µl) Accurate Headspace sampling: Zero bench space system based on gas tight syringe approach SPME (*) technique: Unprecedented fiber handling robustness to cut sample prep costs (*) under license from Supelco 21
22 TriPlus Technical Overview Main Features TriPlus AS TriPlus HS - Designed to fully exploit the potentiality of the GC Injectors - AS-HS convertibility; - Easy installation; - Multiple GCs configurations; Control software 22
23 Optimization of Injection : UFGC with Triplus Autosampler Robotic autosampler Optimum injection to avoid sample introduction discrimination over wider bp range Improved precision Cooling of syringe before sampling Avoid losses of light solvents Needle solvent Wash technique Depth of injection control Injection speed control 23
24 Ultra Fast GC : PTV Split injection of BTEX in n-pentane Column 5m x 0.1 mm id, 0.1 um TR TRACE GC-Channel 1 Fluorobenzene + BTEX Mix_New_40C Col temp Fluorobenzene + BTEX Mix_New_40C Col temp Name Retention Time If the analysis is performed at 40 C Initial Col Temp a loss of resolution is obtained! Millivolts Benzene Fluorobenzene Toluene Ethylbenzene m/p-xylene O-Xylene Millivolts Seconds 24
25 Co-solvent addition during injection What is a co-solvent? A liquid with: a lower volatility a different polarity (in some cases) Co-solvent addition: to tune solvent trapping effect to better managing difficult solvents and difficult solutes to reduce the total amount of solvent transferred to column and detector It is a momentary retention power that disappear by evaporation! 25
26 TriPlus: Software utilities Needle Solvent wash: selection of solvent vial position and solvent vial depth 26
27 Co-solvent addition trough the Autosampler syringe during PTVLVI injection at two different liner level Co-solvent Sample 27
28 Variation of injection speed and injection level during LVI Liner/precolumn conditions are different at the beginning and during injection Injection speed can be high at the beginning of the injection and reduced for the rest of the injection 28
29 PTV-BKF with Co-solvent addition during sampling PAH s in n-hexane with iso-octane ul in Hexane only ul Hexane +15ul Iso-octane
30 ON-Column LVI with co-solvent addition during injection 30
31 Ultra Fast GC : PTV Split injection of BTEX in n-pentane Column 5m x 0.1 mm id, 0.1 um TR TRACE GC-Channel 1 Fluorobenzene + BTEX Mix_New_40C Col temp Fluorobenzene + BTEX Mix_New_40C Col temp Name Retention Time If the analysis is performed at 40 C Initial Col Temp a loss of resolution is obtained! Millivolts Benzene Fluorobenzene Toluene Ethylbenzene m/p-xylene O-Xylene Millivolts Seconds 31
32 Injection and separation of BTEX (10ng/ul) + Fluorobenzene (20ng/ul) in Pentane using the cryo option for the oven Column 5m x 0.1 mm id, 0.1 um TR TRACE GC-Channel 1 Fluorobenzene + BTEX Mix_New 002_Fluorobenzene and BTEX Mix_New Name Fluorobenzene 5 C Initial Column Temp m/p_xylene Millivolts Benzene Toluene Ethylbenzene Millivolts 7000 O-Xylene Seconds 32
33 Co-Solvent addition applied during PTV injection Perfect solvent effect for BTEX at 40 C Column 5m x 0.1 mm id, 0.1 um TR TRACE GC-Channel 1 Fluorobenzene + BTEX Mix_New_40C Col temp_2ul with 1 MTBE Fluorobenzene + BTEX Mix_New_40C Col temp_2ul with 1 MTBE Name ul of MTBE added to 2 ul of Standard solution in n-pentane m/p_xylene C Initial Column Temperature! Millivolts Benzene Fluorobenzene Toluene Ethylbenzene O-Xylene Millivolts Seconds 33
34 Standard Mix BTEX and Hydrocarbons Column: 5 m x 0.1 mm id, 0.1 µm TR-5 Temperature Program: 38 C (0.8 ), 100 C/min to 350 C (0.5 ) Injection volume: 3µl split flow 20ml/min C40/C20 > 0.9 < 5 min 34
35 BTEX and Hydrocarbons separation 35
36 Mix Real Sample and BTEX in Pentane PTV injection with Co-solvent (MTBE) Millivolts Millivolts Name TRACE GC-Channel 1 2ul Real Sample in Pentane plus 1ul of MTBE 002_2ul Real Sample in Pentane plus 1ul of MTBE Minutes Seconds Benzene Fluorobenzene Toluene Ethylbenzene m/p_xylene O-Xylene Millivolts Millivolts Column 5m x 0.1 mm id, 0.1 um TR-5 TRACE GC-Channel 1 MIX_ BTEX_Real sample_ in C5_ 5C Col Temp 001_MIX_ BTEX_Real sample_ in C5_ 5C Col Temp Name Ethylbenzene Benzene Toluene m/p_xylene O-Xylene
37 Reference Gasoil Reference Gasoil - 50 ppm RSD% = 2.2% (10 repetitions) 37
38 Software: Chrom-Card A specific application of Chrom-Card allows automatic integration and reporting The group peak area C7-C40 is calculated The BTEX are individually integrated Benzene Toluene Ethyl-Benzene m,p-xylene o-xylene 38
39 ISO (Mod.): Linearity Test Gasoil Reference Mix - from 2 to 200 ppm 39
40 Sensitivity Reference Gasoil 3µl only! 40
41 Robustness vs dirty matrices System Robustness? 41
42 Robustness vs dirty matrices RSD% Group Peak Real Water Extract After 100 real water extract injections is (10 runs) = 2.3% 42
43 ISO (Mod.): Conventional capillary GC with LVOC Inj Large Volume On Column Injection 50µl injected into a pre-column Automatic OC LV Ultra Fast analysis is 10-fold more productive! 40 min!!! 43
44 Conclusions The analysis of Total Hydrocarbons in water/soil is carried out by Ultra Fast GC in 10-fold shorter run time, still mantaining compliance with ISO (Mod.) and other official methods The Ultra Fast method allows to: - Maximize productivity - Grant top level of repeatability, sensitivity and ruggedness 44
High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System
High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System Application Note Authors Frank David Research Institute for Chromatography, Pres.
More informationApplication Note. Abstract. Authors. Environmental Analysis
High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID using the Agilent Low Thermal Mass (LTM II) System Application Note Environmental Analysis Authors Frank David and Karine Jacq Research
More informationGAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS
GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS FLASH VAPORISATION INJECTION Split Splitless On-Column COOL INJECTION Large Volume Injection (LVI) On-Column On-Column-SVE (with solvent vapour
More informationFast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC
Fast GC SPEEDY GC Slide 4 Optimization Goals Primary: Minimize time for the separation of a given number of peaks Secondary: Maximize the number of peaks separated by a given column Slide 5 HOW CAN THESE
More informationAnalysis of Petroleum Fractions by ASTM D2887
Analysis of Petroleum Fractions by ASTM D2887 Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 2582 Key Words Simulated distillation, D2887, TRACE TR-SimDist Abstract ASTM
More informationGC Inlets. An Overview. Simon Jones GC Applications Engineer
GC Inlets An Overview Simon Jones GC Applications Engineer Types of Inlets Purged Packed Split / Splitless Cool On Column Programmable Temperature Vaporization Volatiles Interface Multi Mode Inlet Where
More informationAchieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)
Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Brian Fitz and Bill Wilson Agilent Technologies, Inc. 285 Centerville Road Wilmington,
More informationMethod Development for Capillary GC Systems. Slide 1
Method Development for Capillary GC Systems Slide 1 AREAS TO OPTIMIZE Injector Carrier gas Column temperature Slide 2 COMMON INJECTOR MODES Vaporization Injection Modes Megabore Direct Split Splitless
More informationOne-Minute Essential Oils Characterization by Gas Chromatography through Nanovolume Injection
Application Note: 10100 One-Minute Essential Oils Characterization by Gas Chromatography through Nanovolume Injection Thermo Fisher Scientific Inc., Milan, Italy Key Words 1-minute Essential Oils Analysis
More informationDetermination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105
Application Note: 10215 Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105 Fausto Munari, Daniela Cavagnino, Andrea Cadoppi, Thermo Fisher Scientific, Milan,
More informationImpact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter
Cd The Chrom Doctor Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands. In Part 1 of this series we focused
More informationAppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS
AppNote 6/26 Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) Andreas Hoffmann GERSTEL GmbH & Co.KG, Eberhard-Gerstel-Platz 1,
More informationANALYSIS OF GASOLINE RANGE HYDROCARBONS ON BP1-PONA
PET 01 - PETROLEUM ANALYSIS OF GASOLINE RANGE HYDROCARBONS ON BP1-PONA GASOLINE RANGE HYDROCARBONS Column Part No.: 054950 BP1, PONA 50 m x 0.15 mm ID Initial Temp.: 30 C, 5 min hold Rate 1: 2 C/min Temp.
More informationAgilent Multimode Inlet for Gas Chromatography
Agilent Multimode Inlet for Gas Chromatography Technical Note Agilent Multimode Inlet for the 7890A GC Designed to give you ease of use and maximum flexibility, the Agilent Multimode Inlet does everything
More informationGas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593
Application Note Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Authors Kelly Beard and James McCurry Agilent Technologies, Inc. Abstract An Agilent
More informationLarge Volume Injection of Polycyclic Aromatic Hydrocarbons
JSB is an authorised partner of Large Volume Injection of Polycyclic Aromatic Hydrocarbons Application Note - Environmental #113 Author Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Abstract
More informationJaap de Zeeuw Restek Middelburg, The Netherlands. Copyrights: Restek Corporation
Liner Selection in GC Jaap de Zeeuw Restek Middelburg, The Netherlands What is a liner? A liner is an inert glass tube that is positioned inside the injection port of the GC The sample solution is introduced
More informationOxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels
Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels James D. McCurry Senior Scientist Agilent Technologies Wilmington, DE USA Page 1 Application Summary There is a need to measure
More informationASTM D 6730 Detailed Hydrocarbon Analysis
ASTM D 6730 Detailed Hydrocarbon Analysis Jaap de Zeeuw, Jan Pijpelink and Barry Burger Restek Corporation ASTM D 6730-01(2006)e1 Determination of Individual Components in Spark Ignition Engine Fuels as
More informationFast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application
Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application Petroleum Authors ChunXiao Wang Agilent Technologies (Shanghai) Co.,Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai
More information6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR
6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR A15836 This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use
More informationComplete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges
Complete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges Alexandria Pavkovich Jason Thomas Trent Sprenkle Outline Background EPA Method Requirements Background
More informationDetection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology
Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology Electronic Noses An electronic nose produces a recognizable response based
More informationOptimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE
Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System Jim McCurry Roger Firor Agilent Technologies Wilmington, DE Page 1 Scope of the Analyzer Designed for ASTM D7798 For
More informationAgilent and ASTM. Update on Recent Activities. Page 1
Agilent and ASTM Update on Recent Activities Page 1 ASTM Committees for Refining, Fuels, Petroelum Products ASTM International D2 Petroleum products & lubricants D3 Gas fuels (natural gas) gas fuels D16
More informationC2, C3, C4 Monomer Analysis
C2, C3, C4 Monomer Analysis Malgorzata Sierocinska Agilent Technologies Waldbronn Page 1 Why Analyze Monomers? To Insure Consistent Production of High Quality Polymer Protect against food contamination
More informationThe Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector
Authors: The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector Introduction Specification of the hydrocarbon composition of LPG is required as traces
More informationAlternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis
Introduction Petroleum & Petrochemical Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis By Katarina Oden, Barry Burger, and Amanda Rigdon Crude oil consists of thousands of different
More informationChoose for Your Instrument
Choose for Your Instrument GC Supplies - Instrument Quick Pick Guide Agilent Technologies Bruker/Scion (Varian) PerkinElmer Shimadzu Thermo Scientific GC Supplies - Instrument Quick Pick Guide Choose for
More informationAgilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass. James D. McCurry, Ph.D. Senior Scientist
Agilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass James D. McCurry, Ph.D. Senior Scientist Agilent Solutions for the Analysis of Ethanol-Based Fuels Derived From Biomass Presented
More informationASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column
ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column Cory S. Fix, Director of Application Development cory.fix@vgcchromatography.com Willie Steinecker,
More informationUsing a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application
Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D759 Application Petrochemical Author James D. McCurry Agilent Technologies 285 Centerville
More informationDual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector
Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector Application Note Hydrocarbon Processing Authors ChunXiao Wang Agilent Technologies
More informationTechnical Procedure for Gas Chromatography (GC-FID)
Technical Procedure for Gas Chromatography (GC-FID) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph (GC- FID). 2.0 Scope This procedure applies to all
More informationThermo Scientific TRACE GC Ultra Multi-channel gas chromatograph. Performance Beyond Limits
c h r o m a t o g r a p h y Thermo Scientific TRACE GC Ultra Multi-channel gas chromatograph Performance Beyond Limits Innovative advances for unprecedented reliability and productivity Ultra in system
More informationPage 1. Alternate Carrier Gas Considerations and Faster GC Analysis
Page 1 Alternate Carrier Gas Considerations and Faster GC Analysis Faster GC Total Analytical Cycle Times A Variety of Approaches Pre-Run ALS Set-Up Chromatographic Run Post-Run Bake-Out Post-Run Cool-Down
More informationCP-3800 GAS CHROMATOGRAPH
CP-3800 GAS CHROMATOGRAPH P/N 03-914689-00 Rev 4 TABLE OF CONTENTS I. GC OPERATION... 2 Overview... 2 Instrument Setup... 3 Building a Method... 4 Automation... 5 Activating the GC Method... 5 Making a
More informationIncreased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel
Application Note Energy and Chemicals Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel Applying the Energy Institute Method IP 8 with an Agilent J&W
More informationMET-Biodiesel Capillary GC Columns
MET-Biodiesel Capillary GC Columns Product Specifications Product Features & Benefits Chromatograms FAQs Related Products Updated: February 2, 2009 Product Specifications 2 Product Specifications What
More informationThermo Scientific. TRACE GC Ultra. Gas Chromatograph Spare Parts Catalog. PN , Revision October 2011
Thermo Scientific TRACE GC Ultra Gas Chromatograph Spare Parts Catalog PN 31709210, Revision October 2011 TRACE GC Ultra Spare Parts Catalog October 2011 Edition Part Number 317 092 10 2007-2011 Thermo
More informationAgilent Multimode Inlet
Agilent Multimode Inlet Large Volume Injection Tutorial Agilent Technologies Notices Agilent Technologies, Inc. 2009 No part of this manual may be reproduced in any form or by any means (including electronic
More informationHigh-Performing Restek PAL SPME Fibers
Restek Sample Handling High-Performing Restek PAL SPME Fibers Suitable for a wide range of analyte chemistries and sample matrices. Reliable performance meets or exceeds other brands. Robust aluminum hub
More informationHigh Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph
Application Note Petrochemicas High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Author James D. McCurry, Ph.D. Agilent Technologies, Inc. Abstract An Agilent
More informationAchieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)
Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Bill Wilson and Chin-Kai Meng Agilent Technologies, Inc. 2850 Centerville Road Wilmington,
More informationPerforming ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software
Install a capillary column in the oven of the SRI GC. The ASTM method suggests a 12 meter.32mm id narrow-bore column coupled with a 2.5 meter guard column but permits the use of any column which exhibits
More informationSet Your Sights on Superior Performance
Restek Sample Handling Set Your Sights on Superior Performance Rugged stainless steel construction ensures longer lifetimes. Faster extraction means higher sample throughput. Better sensitivity allows
More informationApplication Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER
Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D7423-09 using AC OXYTRACER Fast Analysis in
More informationGC Method Compliance and Large Valve Oven Application
GC Method Compliance and Large Valve Oven Application Solution Focused Kelly Beard June 20, 2016 Compliant Methods Compliant Methods What are our options? Analyzers and Analyzer Kits High Volume Well Known
More informationSetting up SilFlow for BackFlush in your GC
Setting up SilFlow for BackFlush in your GC What is backflush and why use it? The BackFlush system eliminates the need to bake heavy sample fractions off the capillary column. Oils, tars and other semivolatile
More informationPractical Steps in GC Troubleshooting
Practical Steps in GC Troubleshooting Techniques, Tips, and Tricks Mark Sinnott Application Engineer GC Columns & Supplies Page 1 Everything was just fine and then this happened! How do I go about TROUBLESHOOTING?
More informationThe Analysis of Total Petroleum Hydrocarbons. Hazel Davidson Technical Marketing Manager Derwentside Environmental Testing Services
The Analysis of Total Petroleum Hydrocarbons Hazel Davidson Technical Marketing Manager Derwentside Environmental Testing Services Derwentside Environmental Testing Services DETS is an independent testing
More informationBeverage Grade Carbon Dioxide
Analysis by Gas Chromatography Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION The Challenge Carbon dioxide, used in the production of carbonated soft drinks and other beverages,
More informationTechniques for Making Your GC Analysis More Repeatable, Reproducible and Robust
Techniques for Making Your GC Analysis More Repeatable, Reproducible and Robust Mark Sinnott Application Engineer Columns and Consumables June 20, 2008 Primary Areas of Concern Sample Auto-Injector Inlet
More informationGC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585
GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc. 850 Centerville
More informationGC Best Practices & Troubleshooting. Group/Presentation Title Agilent Restricted Month ##, 200X
GC Best Practices & Troubleshooting Group/Presentation Title Agilent Restricted Month ##, 200X Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what it is not)
More informationSet Your Sights on Superior Performance
Restek Sample Handling Set Your Sights on Superior Performance Restek PAL SPME Arrow Rugged stainless steel construction ensures longer lifetimes. Faster extraction means higher sample throughput. Better
More informationApplication. Gas Chromatography June 1995
Determining Oxygenates in Gasoline: ASTM Method D Application Gas Chromatography June 99 Authors Michael J. Szelewski Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-60 USA Matthew S. Klee
More informationTechniques for Improving the Reproducibility of GC Analysis
Techniques for Improving the Reproducibility of GC Analysis Primary Areas of Concern Sample Auto-Injector Inlet Column Detector Sample Extract Handling and Care It is critical that the sample extract be
More informationAgilent 9000 Intuvo GC Applications. Matthew Giardina, Ph.D. James McCurry, Ph.D. Rebecca Veeneman, Ph. D.
Agilent 9000 Intuvo GC Applications Matthew Giardina, Ph.D. James McCurry, Ph.D. Rebecca Veeneman, Ph. D. Summary of Intuvo Applications by Industry Environmental EPA Method 8270 Semivolatile Organic Compounds
More informationAgilent 6850 Series II Network GC System G2630A Performance Specifications
IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Agilent 6850 Series II Network GC
More informationGCMS-QP2010 Series. Consumables. Shimadzu Gas Chromatograph/Mass Spectrometer
GCMS-QP2010 Series Consumables Shimadzu Gas Chromatograph/Mass Spectrometer Diag. # 1) Injection Port - Septum Nut 2) Injection Port - Needle Guide 3) Septum 4) O-ring for Glass Insert 5) Glass Insert
More informationThe Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over
The Agilent 12 Series high performance autosampler SL: Area precision, injection volume linearity, imum accessible volume, carry-over Technical Note 8 6 4 2.5 1 Introduction A main performance criterion
More informationLive Crude Oil Volatility
Live Crude Oil Volatility Dan Wispinski : Alberta Innovates Technology Futures Bob Falkiner : Imperial Oil Engineering Services CCQTA/COQA October 31, 2014 Food-Agriculture Environment Health Pipeline
More informationDANI Transformer Oil Gas Analyzer
DANI Transformer Oil Gas Analyzer APPLICATION NOTE - AN169 Introduction Transformers are electrical devices used for energy transfer by electromagnetic induction between two or more circuits. Large oil
More informationRestek Corporation 110 Benner Circle, Bellefonte, Pa Barry L. Burger, Neil Johansen, Valerie Gamble, Donald Rhoades
High Speed PONA Analysis For Detailed Hydrocarbon Analysis Extended (DHAX) Using Hydrogen Carrier Gas For The Determination of Individual Components In Spark Ignition Fuels Barry L. Burger, Neil Johansen,
More informationAgilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011
Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc.
More informationCrossLab. Supplies for Thermo Scientific GC Systems. work for you non-agilent GCs. Put Agilent innovation and quality to. Agilent
works with BRUKER PERKINELMER SHIMADZU THERMO SCIENTIFIC AND MORE Put innovation and quality to work for you non- GCs is a growing portfolio of supplies critical to instrument performance and productivity.
More information2-D GC/MS and the Deans Switch: Theory and Applications. Fred Feyerherm, Agilent Technologies Inc. Houston, Tx
2-D GC/MS and the Deans Switch: Theory and Applications Fred Feyerherm, Agilent Technologies Inc. Houston, Tx Introduction Simplified 2-D GC can be used to accomplish the separation of specific analytes
More informationLiquefied Gas Injector. Solution for the Sampling and Analysis of Liquefied Gases
Liquefied Gas Injector Solution for the Sampling and Analysis of Liquefied Gases Safe and Representative Sampling of Liquefied Gases The analysis of impurities and contaminants in liquefied gases is an
More informationAgilent 7693A Automated Liquid Sampler
Agilent 7693A Automated Liquid Sampler Specifications Overview The Agilent 7693A is a state-of-the-art sample handling and injection system that provides the highest levels of precision and reliability
More informationTechniques for Making Your GC Analysis More Repeatable, Reproducible and Robust
Techniques for Making Your GC Analysis More Repeatable, Reproducible and Robust Mark Sinnott Application Engineer Columns and Consumables June 20, 2008 Primary Areas of Concern Sample Auto-Injector Inlet
More informationMethanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.
Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers Application Note Abstract Versa With the rising prices of fossil fuels, more emphasis is being put on renewable resources
More informationTechnical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)
Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph-mass spectrometer (GC-MS). 2.0 Scope
More informationHigh-Temperature Simulated Distillation System Based on the 6890N GC Application
High-Temperature Simulated Distillation System Based on the 6890N GC Application Petroleum Authors ChunXiao Wang Agilent Technologies (Shanghai) Co., Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai
More informationProductivity Improvements for Dual Column Applications: USP <467> and Blood Alcohol Analysis
Productivity Improvements for Dual Column Applications: USP and Blood Alcohol Analysis Rebecca Veeneman, PhD Applications Chemist Agilent Technologies The GC Usability Gap Usability lags features
More informationDetermination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC
Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC Application Note Energy and Fuels Authors Detlef Wilhelm Anatox GmbH & Co. KG Fürstenwalde, Germany Udo Huber Agilent
More informationThe next level of True Blue Performance
Restek Supplies & Accessories The next level of True Blue Performance Topaz GC inlet liners feature revolutionary technology and inertness to deliver the next level of: Deactivation Reproducibility Productivity
More informationDetailed Hydrocarbon Analysis Featuring Rtx -1 PONA Columns
Detailed Hydrocarbon Analysis Featuring Rtx -1 PONA Columns Compatible with hydrogen, for 50% faster run times. Improved resolution between oxygenates and hydrocarbons, for more accurate reporting. Individually
More informationModel 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors
Model 8610C Gas Chromatograph Mounts up to Six Detectors and Five Injectors Ambient to 400 C Temperature Programmable Column Ovens Dimensions: 19 wide x 13.5 high x 14.5 deep Implement virtually any EPA
More informationSulfur Detection at ppb Levels in Light Hydrocarbon Streams
Sulfur Detection at ppb Levels in Light Hydrocarbon Streams Based on a New Super Permeable PLOT Column Agilent Select Low Sulfur Johan Kuipers Channel Training Specialist Oct 12 th, 2010 1 October 18,
More informationRapid Qualitative GC-TOFMS Analysis of a Petroleum Refinery Reformate Standard
Rapid Qualitative GC-TFMS Analysis of a Petroleum Refinery Reformate Standard LEC Corporation; Saint Joseph, Michigan USA Key Words: GC-TFMS, Petrochemical, Deconvolution 1. Introduction Analyses of petroleum
More informationInstalling a Cool On-Column Inlet
Agilent 6850 Series II Network GC System Accessory G3344B This kit contains: Description Quantity Machine screws, M4 x 0.7 12 mm 6 Cable ties,.062.625 diameter 6 Ship kit* 1 T-20 Torx screw, M4 x 8 mm
More informationProof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut
Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut Application Note Environmental Author Ken Lynam Agilent Technologies, Inc. Abstract Specially designed self-tightening
More informationPRODUCT SPECIFICATION. Product Specification Crystal 9000 Gas Chromatograph. Chromatec Crystal Laboratory Gas Chromatography System
Chromatec Crystal 9000 Laboratory Gas Chromatography System GC Chromatec-Crystal 9000 Crystal 9000 is a leading gas chromatograph in Chromatec GC product line. Crystal 9000 has highest performances and
More informationLiquid Injection Techniques in GC and GC-MS
Liquid Injection Techniques in GC and GC-MS Inge De Dobbeleer Regional Marketing Manager GC and GC-MS, EMEA Thermo Fisher Scientific, Breda/The Netherlands The world leader in serving science Content Hot
More informationSPECIFICATION. Bench top triple quadrupole Mass Spectrometry System coupled to a gas performance Chromatography (GC-MS/MS)
SPECIFICATION Bench top triple quadrupole Mass performance Chromatography (GC-MS/MS) 1. Scope These specifications describe the requirements for a bench top triple quadrupole mass spectrometry system coupled
More informationGroup-Type Analysis (PiPNA) in Diesel and Jet Fuel by Flow Modulated GCxGC FID.
Group-Type Analysis (PiPNA) in Diesel and Jet Fuel by Flow Modulated GCxGC FID. Dedicated PiPNA + FAME For (Bio-)Diesel and Jet Fuels Robust System, Easy to use No Cryogenic coolant Required Keywords:
More informationUsing the PSD for Backflushing on the Agilent 8890 GC System
Application Note Petrochemicals Using the PSD for Backflushing on the Agilent 889 GC System Author Brian Fitz Agilent Technologies, Inc. Wilmington, DE, USA. Abstract An Agilent 889 series GC equipped
More informationGC Instruments GC Columns GC Accessories
GC GC Instruments................. 200 212 GC Systems & Data Collection........... 200 205 Gas Generators Hydrogen......................... 206 207 Air............................ 208 209 Nitrogen............................
More informationRefinery Gas. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.
Refinery Gas Analysis by Gas Chromatography Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Refinery Gas Analysis Reliability Placing refinery gas analyzers in the field for over
More informationAgilent Portfolio. Luca Godina. Technical Marketing Specialist GC-GCMS
Agilent Portfolio Luca Godina Technical Marketing Specialist GC-GCMS 1 Summary Introduction Sample Prep Sample Introduction Inlets 7890B LTM CFT Detectors Special Solutions 13/11/12 Tirana Page 2 Agilent
More informationAZ Chrom s.r.o.
AZ Chrom s.r.o. azetchrom@hplc.sk www.azetchrom.sk ÙÝ Í ó Ê ²ñÞ «µ» ² «³»² Ï«½µ Ð ½µ Ù«¼» Í ²¹ ²¹» Í» ²» Ô ²» Ñó ²¹ Ú» Ú ²¹» Ì» Ú» Ako nás možno kontaktova : AZ Chrom s.r.o. Robotnícka 10 831 03 Bratislava
More informationState of Texas UST Monitoring visit us online at The latest TNRCC 1005/1006 UST methods used by Texas.
At-a-Glance Product Information from Restek State of Texas UST Monitoring To help laboratories comply with and use these analytical procedures, Restek has been active in following the state guidance. Based
More informationFast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC
Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC Technical Overview Trace gas analysis is a challenge in today s world. The ability to analyze lower component
More informationDETECTORS Photo Ionization Detector - PID. Overview
Overview DETECTORS The Photo Ionization Detector (PID) responds to all molecules whose ionization potential is below 10.6eV, including aromatics and molecules with carbon double bonds. The PID is nondestructive,
More informationSame Column and Gas Type Try Different/Faster Velocities. Same Column, Switch He to H2 Carrier Then Try Faster Velocities
: What Are You Willing To Do? Daron Decker Chromatography Technical Specialist Page 1 Agilent Restricted 3 Options Same Column and Gas Type Try Different/Faster Velocities Same Column, Switch He to H2
More informationCrossLab. Supplies for PerkinElmer GC Systems. to work for you non-agilent GCs. Put Agilent innovation and quality. Agilent. Quick Reference Guide
works with BRUKER PERKINELMER SHIMADZU THERMO SCIENTIFIC AND MORE Put innovation and quality to work for you non- GCs is a growing portfolio of supplies critical to instrument performance and productivity.
More informationFrom Helium to Hydrogen: GC-MS Case Study on SVOCs in Water
From Helium to Hydrogen: GC-MS Case Study on SVOCs in Water Jessie Butler Alexander N. Semyonov December 13, 2012 Helium? It s gone! Not only at Disneyland Resort 2 Helium shortage grounds Mickey at Tokyo
More informationQuantification and Characterization of Sulfur in Low-Sulfur Reformulated Gasolines by GC-ICP-MS Application
Quantification and Characterization of Sulfur in Low-Sulfur Reformulated Gasolines by GC-ICP-MS Application Authors Steven M. Wilbur and Emmett Soffey Agilent Technologies 338 146th Place SE Bellevue,
More informationS 1125 HPLC Pump System
Routine S 170 EXPERT HPLC HPLC System SYSTEMS 500 The Sykam ROUTINE HPLC System S 500 is intended for all routine analysis. The system configuration is highly variable and several upgrade options make
More information