Practical Steps in GC Troubleshooting
|
|
- Ferdinand Parker
- 6 years ago
- Views:
Transcription
1 Practical Steps in GC Troubleshooting Techniques, Tips, and Tricks Mark Sinnott Application Engineer GC Columns & Supplies Page 1
2 Everything was just fine and then this happened! How do I go about TROUBLESHOOTING? Page 3
3 Everything was just fine and then this happened! Logic = Something changed (slowly or sudden) = Something is different Track Events log book -Changed column, liner, septum, syringe, etc. -Injected samples, other method, etc. -Did maintenance, cut column, inlet flush, etc. Page 4
4 Logical Troubleshooting Troubleshooting Starts with Isolating the problem There are 5 basic areas from where the problem arises INJECTOR FLOW COLUMN DETECTOR ELECTRONICS But of course it can always be some COMBINATION Knowing what can & can t cause the symptom is the key Page 5
5 Typical Problems of Optimized Methods becoming Unoptimized and the Reason Why. Peak Tailing Flow Path or Activity Bonus Peaks In Sample or Back Flash (Carry Over) Split Peaks Injector Problems, Mixed Solvent No Peaks Wasn t Introduced, Wasn t Detected Response Changes Activity, Injector Discrimination, Detector Problem Peak Fronting Overload or Solubility Mismatch, Injector Problems Shifting Retention Leaks, Column Aging, Contamination or Damage Loss of Resolution Separation Decreasing, Peak Broadening Baseline Disturbances Column Bleed, Contamination, Electronics Noisy or Spiking Baseline Electronics or Contaminated Detector Quantitation Problems Activity, Injector or Detector Problems Page 6
6 Peak Tailing INJECTOR or COLUMN is Active -Reversible adsorption of active compounds (-OH, -NH, -SH) FLOW problem - dead volume, obstruction, poor installation, or severe column contamination Miscellaneous - overloading of PLOT columns, co-elution, polarity mismatch between phase, solute or solvent, and some compounds always tail *Tip = Inject a light hydrocarbon, should not tail unless flow path problem. Page 7
7 Bonus Peaks W h = Column: J&W P/N: Oven: DB-5 30 m x 0.53 mm I.D., 1.5 µm 60 C for 1 min C at 20 /min 300 C for 3 min Carrier Gas: Helium at 36 cm/sec Injector: Split 1:100, 250 C sample C 7 - C 20 Detector: FID, 250 C W h = Wh = Page 8
8 Bonus Peaks or Ghost Peaks Contamination in INJECTOR, COLUMN or FLOW (carrier gas) -Carry-over from a backflash or previous sample -Bad tank of gas or traps have expired -Septum bleed *TIP = Run a blank run it should be blank! Page 9
9 Page 10 Bonus Siloxane Peaks
10 GC Column Bleed Ions Page 11
11 Column Bleed is Influenced by: 1.3e4 1.2e4 1.1e4 Phase type Temperature Column dimensions 1.0e DB M x.53mm I.D., 3.0µm 24 pa / 260 C DB-1 30m x.32mm I.D.,.25µm 12 pa / 320 C Time (min.)
12 Split Peaks INJECTOR (poor sample introduction) -Injecting the sample twice (some how?) -Mixed sample solvent (polarity difference) -Sample in syringe needle (manual inject) INJECTOR (activity) -Breakdown (not really a split peak, 2 peaks) -Sample degradation in injector VOLATILITY High boilers dropping out on Cold Spots -Transfer line temps -Unions or fittings not tracking column temp Page 13
13 No Peaks DETECTOR (not on or not operational) INJECTOR (not working) -Plugged syringe/plunger not moving -Wrong injector (or detector) -Huge leak (older systems) -No carrier gas flow NOT the COLUMN Unless -Broken column or No column Page 14
14 Peak Response All Change in Size INJECTOR -Leaky syringe -Split ratio set incorrectly -Wrong purge activation time -Septum purge flow too high -Injector temperature too low* DETECTOR (response problem) -Settings or flows changed -Electronics failing *Tip = Ask is it all of them or some of them, if all then injector or detector Page 15
15 Peak Response Some Change in Size INJECTOR or COLUMN is active/contaminated -Irreversible adsorption of active compounds (-OH, -NH, -SH) -Decomposition of sample -Temperature Change Discrimination -Evaporation from sample *Tip = If only some change, then ask which ones? If active compounds then activity. If tracks volatility then cold spots or inlet discrimination. Page 16
16 Peak Fronting Shark Fin Shaped or Just Slight COLUMN (contaminated) -Overload (More pronounced with large solute and phase polarity differences) INJECTOR -Column installation -Compound very soluble in injection solvent (need retention gap) -Mixed sample solvent OTHER -Co-elution -Breakdown Page 17
17 Retention Time Shift INJECTOR -Leak in the septum -Change in injection solvent -Large change in sample concentration FLOW -Change in gas velocity COLUMN -Contamination -Damaged stationary phase -Loss of stationary phase -Change in temperature Page 18
18 Effect of Sample Overload on Retention Time and Peak Shape 1400 ng 7 ng min Page 19
19 Loss of Resolution Separation Peak Width Resolution is a function of separation and peak width Page 20
20 Loss of Resolution - Separation Decrease COLUMN -Different column temperature -Contamination (more phase?) Separation -Matrix components co-eluting -Different column phase? Peak Width Page 21
21 Loss of Resolution - Peak Broadening FLOW -Change in carrier gas velocity -Make-up gas Separation COLUMN -Contamination -Phase degradation Peak Width INJECTOR (efficiency) -Settings, Liner, Installation, etc. Page 22
22 Baseline Disturbances Sudden Changes, Wandering, or Drifting WANDER DRIFT COLUMN or DETECTOR -Not fully conditioned or stabilized (electronics) -Contamination FLOW -Changes in carrier and/or detector gas flows -Valves switching, leaks Page 23
23 Noisy Baseline MILD SEVERE FLOW -Contaminated gas -Incorrect detector settings COLUMN -Bleed if at high temperature -In detector flame (poor installation) DETECTOR -Air leak - ECD, TCD -Electronics malfunction Page 24
24 Spiking Baseline DETECTOR -Particles entering the detector -Random: poor connection -Regular: nearby "cycling" equipment (electronics) Page 25
25 Quantitation Problems DETECTOR -Poor stability (electronics) or Baseline disturbances (contamination) -Outside detector's linear range or wrong settings Activity (adsorption) in INJECTOR or COLUMN OTHER INJECTOR -Technique, settings, conditions -Syringe worn -Co-elution -Matrix effects -Sample evaporation leaky vials -Sample decomposition Page 26
26 What is NOT caused by a Column??? Peaks!! Any reproducible, sharp chromatographed peak! Siloxanes Degradation product peaks: Endrin Aldehyde, Endrin Ketone, DDE, DDD. Carryover of sample compounds Splitting of peaks Page 27
27 Troubleshooting Tools Bleed Profile: baseline problems Inject a non-retained peak: peak shape problems Test mix: all problems Isolate the components: all problems Condensation Test: baseline problems Jumper Tube Test: baseline problems Page 28
28 Generating a Bleed Profile 1.3e4 1.2e4 1.1e4 Produce when the column is new (for future reference) when there is a baseline problem 1.0e Time (min.) *DB-1 30m x.32mm I.D.,.25µm Temperature program // 40 C, hold 1 min // 20 /min to 320 C, hold 10 min. Page 29
29 Non-Retained Peak Shapes Used to Check Flowpath Good Installation Improper Installation or Injector Leak Potential problems: Injector or septum leak Too low of a split ratio Liner problem (broken, leaking, misplaced) Column position in injector and detector Page 30
30 Test Mix Used to determine how good the column is or if the problem is related to the chemical properties of the analytes. Test Conditions Inlet: Split (250 C) Detector: FID (320 C) Carrier Gas: Hydrogen Flow: 37.3 cm/sec (1.8 ml/min) Holdup Compound: Methane (0.671-min) Temperature Program: Isothermal at 110 C Page 31
31 Test Mixture Components Compounds Hydrocarbons Alcohols FAME s, PAH s Acids Bases Purpose Efficiency Retention Activity Retention Acidic Character Basic Character Page 32
32 Own Test Mixture More specific to your application Selective detectors Concentrations specific to your application Use same instrument conditions Easiest to simply inject a calibration standard Store for future measure of column performance Group/Prese Page 33
33 Isolate the Components Simplify the system: - example - Direct injection instead of P&T sample introduction Put in a known good column Move column to a different GC, inlet or detector Page 34
34 Condensation Test Used* to isolate the cause of: - Erratic baselines - Ghost peaks or carryover *Use when problems are worse after periods of GC non-use Page 35
35 Condensation Test Procedure Leave GC at C for > 8 hours Blank run Repeat a blank run immediately after the first blank run is complete Compare the two blank runs Page 36
36 Condensation Test Results First blank run is worse: - Contaminants (from injector, lines, traps or carrier gas) carried into the column Blank runs the same: contaminants are not strongly focused on the front of the column Page 37
37 Jumper Tube Test Purpose Helps to locate the source of contamination or noise Isolates GC components Page 38
38 Jumper Tube Test Isolate the Detector Remove column from the detector Cap detector and turn on Blank run Page 39
39 Jumper Tube Test Isolation of Detector - Results Detector OK Detector is the problem Page 40
40 Jumper Tube Test Isolate the Injector Connect the injector and detector meters deactivated fused silica tubing Turn on carrier gas Blank run Page 41
41 Jumper Tube Test Isolate the Injector - Results Injector OK Injector, lines or carrier gas contaminated Page 42
42 Jumper Tube Test Isolate the Column Reinstall the column Setup as before Blank Run Page 43
43 Jumper Tube Test Isolate the Column - Results Problem returns: It s the column Problem gone: Previous leak, solid debris, or installation problem Page 44
44 Page 45 And Now Let s do Some
45 Page 46 Troubleshooting-Example #1
46 A Real Troubleshooting Example 30 No Peaks Page 47
47 Logical Steps Taken to Find Peaks (most of our problems are leaks and plugs) Is the flame Lit? put glass piece over FID outlet----answer in this case, Water condenses look at output in instrument guage-- is the digital value greater than 0.0? Answer in this case is approximately 16.2 pico amps Is there flow through the column? disconnect column from detector and measure flow with bubble solution or meter Answer in this case was YES THERE IS FLOW Assess the observations Flame is lit and we have flow from end of column Hypothesis: Sample not getting on column-syringe plugged? Take syringe out and make injection manually on a dry paper towel Answer towel stayes dry (Syringe was clogged with septum) Pull plunger out top, add solvent and replace plunger will usually dislodge septum particle (should hear a little pop) If you can t dislodge plug, Replace syringe Reassemble the Injector & Re-inject Page 48
48 Peaks!! Page 49
49 Page 50 Troubleshooting-Example #2
50 8.0e4 7.0e4 6.0e4 5.0e DB-624 COLUMN 1. 1,2-Dichloropropane 2. Octane 3. Tetrachloroethylene 4. Chlorobenzene 5. Nonane 7.43 QC Test Mix Column: DB m x 53mm I.D., 3.0µm Carrier: Helium at 40 cm/sec measured at 35 C Injector: Mega Direct, 260 C Detector: FID, 300 C Oven: 35 C for 1.50 min 30 /min to 65 for 10 min 4.0e4 3.0e e4 1.0e Time (min.) Page 51
51 Example of Column Contamination 1.5e e4 1.3e4 1.2e DB-624 QC Test Mix* After 75 Injections of Oily Sample 1.1e4 1.0e Time (min.) *Temperature program// 35 C hold 1.50 min // 30 /min to 65 C, hold 10 min Page 52
52 Column and Liner Contamination Inlet coil of column Page 53
53 1.2e4 1.1e4 Example of Column Contamination 2.80 Removed 1 1/2 m from injector end * 1.0e Time (min.) *Before Column rinse and bake Temperature program // 35 C hold 1.50 min // 30 /min to 65 C, hold 10 min Page 54
54 Page 55 Looks Fixed Doesn't it?
55 Example of Column Contamination 1.3e4 1 1/2 mtrs removed* QC Test mix to Upper Temperature Limit 1.2e4 1.1e4 1.0e Time (min.) *Before Column rinse and bake. Temperature program // 35 C, hold 1.50 min // 30 /min to 65 C, hold 15 min // 20 /min to 260, hold 50 min Page 56
56 Backflush Column Rinse with 10ml each: Methanol, Methylene Chloride, Hexane Capillary column Special connector and ferrule Beaker for solvent collection 1/16" flexible teflon line to regulated pressure source Flexible teflon tubing Cap Capillary column Special adapter Vial Page 57
57 Jumper Tube Test Used to Isolate Source of Contamination. Cap off the detector and establish normal gas flows and temperature.. Plot the baseline using a temperature program. If flat.... Connect 1 meter of deactivated tubing between the injector and detector. Plot the baseline using a temperature program. If flat.... Install the column.. Plot the baseline using a temperature program. Page 58
58 1.0e4 Contaminated Inlet Jumper Tube Test* Time (min.) *1/2 mtr length of.53 mm I.D. deactivated tubing Temperature program // 35 C, hold 1.50 min // 30 /min to 65 C for 15 min Page 59
59 Rinsing Injector Carrier gas line GC Oven Injector body MeCl 2 C6 Page 60
60 Troubleshooting Tips 1. Isolate the problem. (Blank Run, Inject Un-retained Compound, Jumper Tube Test) 2. Change only one variable at a time. 3. Compare before/after chromatograms. (Peak shape, response, retention, baseline rise, background, look for trends, etc.) 4. Utilize Technical Support. Page 61
61 Remember Complete system = Carrier Gas + Injector + Column + Detector + Data System Multiple cause and effect Do not change too many variables at once Page 62
62 Contact Agilent Chemistries and Supplies Technical Support Option 3, Option 3: Option 1 for GC/GCMS Columns and Supplies Option 2 for LC/LCMS Columns and Supplies Option 3 for Sample Preparation, Filtration and QuEChERS Option 4 for Spectroscopy Supplies Available in the USA 8-5 all time zones gc-column-support@agilent.com lc-column-support@agilent.com spp-support@agilent.com spectro-supplies-support@agilent.com
GC Best Practices & Troubleshooting. Group/Presentation Title Agilent Restricted Month ##, 200X
GC Best Practices & Troubleshooting Group/Presentation Title Agilent Restricted Month ##, 200X Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what it is not)
More informationTypical Gas Chromatographic System
Page 72 Air Hydrogen Carrier Gas Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector Electrometer Cylinders or Generators Flow Controller Column Recorder/
More informationMethod Development for Capillary GC Systems. Slide 1
Method Development for Capillary GC Systems Slide 1 AREAS TO OPTIMIZE Injector Carrier gas Column temperature Slide 2 COMMON INJECTOR MODES Vaporization Injection Modes Megabore Direct Split Splitless
More informationGas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593
Application Note Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Authors Kelly Beard and James McCurry Agilent Technologies, Inc. Abstract An Agilent
More informationGAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS
GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS FLASH VAPORISATION INJECTION Split Splitless On-Column COOL INJECTION Large Volume Injection (LVI) On-Column On-Column-SVE (with solvent vapour
More informationIncreased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel
Application Note Energy and Chemicals Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel Applying the Energy Institute Method IP 8 with an Agilent J&W
More informationTechniques for Making Your GC Analysis More Repeatable, Reproducible and Robust
Techniques for Making Your GC Analysis More Repeatable, Reproducible and Robust Mark Sinnott Application Engineer Columns and Consumables June 20, 2008 Primary Areas of Concern Sample Auto-Injector Inlet
More informationUsing a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application
Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D759 Application Petrochemical Author James D. McCurry Agilent Technologies 285 Centerville
More informationGC Inlets. An Overview. Simon Jones GC Applications Engineer
GC Inlets An Overview Simon Jones GC Applications Engineer Types of Inlets Purged Packed Split / Splitless Cool On Column Programmable Temperature Vaporization Volatiles Interface Multi Mode Inlet Where
More informationJaap de Zeeuw Restek Middelburg, The Netherlands. Copyrights: Restek Corporation
Liner Selection in GC Jaap de Zeeuw Restek Middelburg, The Netherlands What is a liner? A liner is an inert glass tube that is positioned inside the injection port of the GC The sample solution is introduced
More informationFast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC
Fast GC SPEEDY GC Slide 4 Optimization Goals Primary: Minimize time for the separation of a given number of peaks Secondary: Maximize the number of peaks separated by a given column Slide 5 HOW CAN THESE
More informationTroubleshooting Tips & Tricks for your GC Analyzer & CFT Application
Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application 7890A/7890B GC Overview October 29, 2014 1 Definitions Carrier Gas Pressurized gas used to transport the sample through the system.
More informationTechniques for Making Your GC Analysis More Repeatable, Reproducible and Robust
Techniques for Making Your GC Analysis More Repeatable, Reproducible and Robust Mark Sinnott Application Engineer Columns and Consumables June 20, 2008 Primary Areas of Concern Sample Auto-Injector Inlet
More informationMET-Biodiesel Capillary GC Columns
MET-Biodiesel Capillary GC Columns Product Specifications Product Features & Benefits Chromatograms FAQs Related Products Updated: February 2, 2009 Product Specifications 2 Product Specifications What
More informationPage 1. Alternate Carrier Gas Considerations and Faster GC Analysis
Page 1 Alternate Carrier Gas Considerations and Faster GC Analysis Faster GC Total Analytical Cycle Times A Variety of Approaches Pre-Run ALS Set-Up Chromatographic Run Post-Run Bake-Out Post-Run Cool-Down
More informationUsing the PSD for Backflushing on the Agilent 8890 GC System
Application Note Petrochemicals Using the PSD for Backflushing on the Agilent 889 GC System Author Brian Fitz Agilent Technologies, Inc. Wilmington, DE, USA. Abstract An Agilent 889 series GC equipped
More informationTechnical Procedure for Gas Chromatography (GC-FID)
Technical Procedure for Gas Chromatography (GC-FID) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph (GC- FID). 2.0 Scope This procedure applies to all
More informationAgilent Multimode Inlet for Gas Chromatography
Agilent Multimode Inlet for Gas Chromatography Technical Note Agilent Multimode Inlet for the 7890A GC Designed to give you ease of use and maximum flexibility, the Agilent Multimode Inlet does everything
More informationImpact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter
Cd The Chrom Doctor Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands. In Part 1 of this series we focused
More informationAchieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)
Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Bill Wilson and Chin-Kai Meng Agilent Technologies, Inc. 2850 Centerville Road Wilmington,
More informationTechniques for Improving the Reproducibility of GC Analysis
Techniques for Improving the Reproducibility of GC Analysis Primary Areas of Concern Sample Auto-Injector Inlet Column Detector Sample Extract Handling and Care It is critical that the sample extract be
More informationSetting up SilFlow for BackFlush in your GC
Setting up SilFlow for BackFlush in your GC What is backflush and why use it? The BackFlush system eliminates the need to bake heavy sample fractions off the capillary column. Oils, tars and other semivolatile
More informationASTM D 6730 Detailed Hydrocarbon Analysis
ASTM D 6730 Detailed Hydrocarbon Analysis Jaap de Zeeuw, Jan Pijpelink and Barry Burger Restek Corporation ASTM D 6730-01(2006)e1 Determination of Individual Components in Spark Ignition Engine Fuels as
More informationAchieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)
Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Brian Fitz and Bill Wilson Agilent Technologies, Inc. 285 Centerville Road Wilmington,
More informationChoose for Your Instrument
Choose for Your Instrument GC Supplies - Instrument Quick Pick Guide Agilent Technologies Bruker/Scion (Varian) PerkinElmer Shimadzu Thermo Scientific GC Supplies - Instrument Quick Pick Guide Choose for
More informationHigh Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph
Application Note Petrochemicas High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Author James D. McCurry, Ph.D. Agilent Technologies, Inc. Abstract An Agilent
More informationInstalling the 6850 Direct Valve Column Connector Kit
Use this kit to bypass the inlet and connect your 6850 gas or liquid sampling valve directly to a column. To use this kit, your 6850 GC must have a purged packed inlet. This kit contains: Installing the
More informationASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column
ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column Cory S. Fix, Director of Application Development cory.fix@vgcchromatography.com Willie Steinecker,
More informationTechnical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)
Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph-mass spectrometer (GC-MS). 2.0 Scope
More informationSplit and Splitless Injection in Capillary Gas Chromatography
Split and Splitless Injection in Capillary Gas Chromatography With Some Remarks on PTV Injection 3rd enlarged and revised Edition By Konrad Grob Hüthig Buch Verlag Heidelberg 1993 A Split Injection 1 Introduction
More informationAgilent Multimode Inlet
Agilent Multimode Inlet Large Volume Injection Tutorial Agilent Technologies Notices Agilent Technologies, Inc. 2009 No part of this manual may be reproduced in any form or by any means (including electronic
More informationAlternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis
Introduction Petroleum & Petrochemical Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis By Katarina Oden, Barry Burger, and Amanda Rigdon Crude oil consists of thousands of different
More informationC2, C3, C4 Monomer Analysis
C2, C3, C4 Monomer Analysis Malgorzata Sierocinska Agilent Technologies Waldbronn Page 1 Why Analyze Monomers? To Insure Consistent Production of High Quality Polymer Protect against food contamination
More informationProof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut
Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut Application Note Environmental Author Ken Lynam Agilent Technologies, Inc. Abstract Specially designed self-tightening
More informationAnalysis of Petroleum Fractions by ASTM D2887
Analysis of Petroleum Fractions by ASTM D2887 Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 2582 Key Words Simulated distillation, D2887, TRACE TR-SimDist Abstract ASTM
More informationFree and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584
Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN 14105 and ASTM D6584 Introduction With today s increasing concern for the environment and the depletion of fossil
More informationDetection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector
Detection of Sulfur Compounds in Natural Gas According to ASTM D554 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector Application Note Author Rebecca Veeneman Abstract Sulfur compounds in natural
More informationPerforming ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software
Install a capillary column in the oven of the SRI GC. The ASTM method suggests a 12 meter.32mm id narrow-bore column coupled with a 2.5 meter guard column but permits the use of any column which exhibits
More informationApplication Note. Abstract. Authors. Environmental Analysis
High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID using the Agilent Low Thermal Mass (LTM II) System Application Note Environmental Analysis Authors Frank David and Karine Jacq Research
More informationOptimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE
Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System Jim McCurry Roger Firor Agilent Technologies Wilmington, DE Page 1 Scope of the Analyzer Designed for ASTM D7798 For
More informationTechnical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS)
Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the pyrolysis-gas chromatograph-mass spectrometer
More informationSame Column and Gas Type Try Different/Faster Velocities. Same Column, Switch He to H2 Carrier Then Try Faster Velocities
: What Are You Willing To Do? Daron Decker Chromatography Technical Specialist Page 1 Agilent Restricted 3 Options Same Column and Gas Type Try Different/Faster Velocities Same Column, Switch He to H2
More informationCP-3800 GAS CHROMATOGRAPH
CP-3800 GAS CHROMATOGRAPH P/N 03-914689-00 Rev 4 TABLE OF CONTENTS I. GC OPERATION... 2 Overview... 2 Instrument Setup... 3 Building a Method... 4 Automation... 5 Activating the GC Method... 5 Making a
More informationYour GC Analysis More. and Robust. Mark Sinnott. May 11, Agilent Restricted
Techniques for Making Your GC Analysis More Repeatable, Reproducible and Robust Mark Sinnott GC Column Specialist May 11, 2010 Primary Areas of Concern Sample Auto-Injector Inlet Column Detector Sample
More informationFausto Munari e Andrea Cadoppi ThermoFisher - Italy
The world leader in serving science Ultra Fast GC Determination of Total Hydrocarbons (C7-C40) and BTEX in Water and Soils through Direct Resistively Heated capillary columns and Robotic Autosampler. Fausto
More informationDetermination of Free and Total Glycerin in B100 Biodiesel
Page 1 of 5 Page 1 of 5 Return to Web Version Determination of Free and Total Glycerin in B100 Biodiesel By: Michael D. Buchanan, Katherine K. Stenerson, and Vicki Yearick, Reporter US Vol 27.1 techservice@sial.com
More informationAgilent G2855A Deans Switching System
Agilent G2855A Deans Switching System Installation and Operation Agilent Technologies Notices Agilent Technologies, Inc. 2003 No part of this manual may be reproduced in any form or by any means (including
More informationSimple Heart Cutting with Deans. Capillary Flow Technolgy. Simon Jones Applications Engineer
Simple Heart Cutting with Deans Switch and Backflushing with Capillary Flow Technolgy Simon Jones Applications Engineer GC or GC/MS Analysis in Complex Matrices In complex sample matrices, there are often
More informationAnalysis and. Separation of Oxygenates in Hydrocarbon Matrices. Simon Jones Application Engineer Folsom, CA
Analysis and Chromatographic Separation of Oxygenates in Hydrocarbon Matrices Simon Jones Application Engineer Folsom, CA August 20, 2009 Agenda WCOT vs. PLOT columns OxyPlot A Unique stationary phase
More informationAgilent 7890 Series Gas Chromatograph
Agilent 7890 Series Gas Chromatograph Troubleshooting Agilent Technologies Notices Agilent Technologies, Inc. 2013 No part of this manual may be reproduced in any form or by any means (including electronic
More informationModel 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors
Model 8610C Gas Chromatograph Mounts up to Six Detectors and Five Injectors Ambient to 400 C Temperature Programmable Column Ovens Dimensions: 19 wide x 13.5 high x 14.5 deep Implement virtually any EPA
More informationDetailed Hydrocarbon Analysis Featuring Rtx -1 PONA Columns
Detailed Hydrocarbon Analysis Featuring Rtx -1 PONA Columns Compatible with hydrogen, for 50% faster run times. Improved resolution between oxygenates and hydrocarbons, for more accurate reporting. Individually
More informationSulfur Detection at ppb Levels in Light Hydrocarbon Streams
Sulfur Detection at ppb Levels in Light Hydrocarbon Streams Based on a New Super Permeable PLOT Column Agilent Select Low Sulfur Johan Kuipers Channel Training Specialist Oct 12 th, 2010 1 October 18,
More information6890 Installing ¼ Glass Columns
6890 Installing ¼ Glass Columns A15891 This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this material. The information
More informationHigh Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System
High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System Application Note Authors Frank David Research Institute for Chromatography, Pres.
More informationAnalysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction
Analysis of Glycerin and Glycerides in Biodiesel (B1) Using ASTM D68 and EN11 Application HPI/Petrochemicals/Polymers Author James D. McCurry Agilent Technologies, Inc. 8 Centerville Road Wilmington, DE
More informationThe next level of True Blue Performance
Restek Supplies & Accessories The next level of True Blue Performance Topaz GC inlet liners feature revolutionary technology and inertness to deliver the next level of: Deactivation Reproducibility Productivity
More informationRefinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography
Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography Dr. Robert Lorenz Analytical Chemist Chevron Energy Technology Company Gulf Coast Conference Galveston, TX October
More informationApplication. Gas Chromatography June 1995
Determining Oxygenates in Gasoline: ASTM Method D Application Gas Chromatography June 99 Authors Michael J. Szelewski Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-60 USA Matthew S. Klee
More informationOptimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides
Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides Application Note Food Testing & Agriculture Authors Limian Zhao and Chin-Kai Meng Agilent Technologies, Inc. 85 Centerville
More informationSELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS
Rev 1 3/6/2004 Selerity Technologies Inc. www.selerity.com SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS What is supercritical fluid chromatography? A chromatographic technique in
More informationApplication Note. Author. Introduction. Energy and Fuels
Analysis of Free and Total Glycerol in B-100 Biodiesel Methyl Esters Using Agilent Select Biodiesel for Glycerides Application Note Energy and Fuels Author John Oostdijk Agilent Technologies, Inc. Introduction
More informationBeverage Grade Carbon Dioxide
Analysis by Gas Chromatography Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION The Challenge Carbon dioxide, used in the production of carbonated soft drinks and other beverages,
More informationAgilent and ASTM. Update on Recent Activities. Page 1
Agilent and ASTM Update on Recent Activities Page 1 ASTM Committees for Refining, Fuels, Petroelum Products ASTM International D2 Petroleum products & lubricants D3 Gas fuels (natural gas) gas fuels D16
More informationOxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels
Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels James D. McCurry Senior Scientist Agilent Technologies Wilmington, DE USA Page 1 Application Summary There is a need to measure
More informationAgilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass. James D. McCurry, Ph.D. Senior Scientist
Agilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass James D. McCurry, Ph.D. Senior Scientist Agilent Solutions for the Analysis of Ethanol-Based Fuels Derived From Biomass Presented
More informationChromatography. Liquid. HPLC Accessories SupelPRO Valves SUPELCO
8 SupelPRO Valves Order:.800..00 Technical Service:.800.9.0 Web: www.sigma-aldrich.com/supelco SupelPRO Automated Fluidics Instruments Supelco s SupelPRO series are precision, electronically-controlled,
More informationDetection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology
Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology Electronic Noses An electronic nose produces a recognizable response based
More informationGC Analysis of PCBs in Fish Oil by Direct Injection Without Any Sample Cleanup. Philip L. Wylie, Ph.D. Sr. Applications Chemist Agilent Technologies
GC Analysis of PCBs in Fish Oil by Direct Injection Without Any Sample Cleanup Philip L. Wylie, Ph.D. Sr. Applications Chemist Agilent Technologies Pittsburgh Conference, 2008 Paper 1120-1 Tuesday, 8:30
More informationChoose for Your Instrument. GC Supplies - Agilent Instrument Quick Pick Guide
Choose for Your Instrument GC Supplies - Agilent Instrument Quick Pick Guide Syringes Septa Inlet Liners O-rings and Sealing Rings SilTite FingerTite Ferrules Ferrules Your chromatography analysis does
More informationDANI Transformer Oil Gas Analyzer
DANI Transformer Oil Gas Analyzer APPLICATION NOTE - AN169 Introduction Transformers are electrical devices used for energy transfer by electromagnetic induction between two or more circuits. Large oil
More informationMethanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.
Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers Application Note Abstract Versa With the rising prices of fossil fuels, more emphasis is being put on renewable resources
More informationNickel Catalyst Tube Accessory 19205A
Installation and Operating Guide Nickel Catalyst Tube Accessory 19205A Agilent Technologies 2000 All Rights Reserved. Reproduction, adaptation, or translation without permission is prohibited, except as
More informationHigh-Temperature Simulated Distillation System Based on the 6890N GC Application
High-Temperature Simulated Distillation System Based on the 6890N GC Application Petroleum Authors ChunXiao Wang Agilent Technologies (Shanghai) Co., Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai
More informationDetermination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105
Application Note: 10215 Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105 Fausto Munari, Daniela Cavagnino, Andrea Cadoppi, Thermo Fisher Scientific, Milan,
More informationOptimized Method for Analysis of Commercial and Prepared Biodiesel using UltraPerformance Convergence Chromatography (UPC 2 )
Optimized Method for Analysis of Commercial and Prepared Biodiesel using UltraPerformance Convergence Chromatography (UPC 2 ) Mehdi Ashraf-Khorassani, 1 Giorgis Isaac, 2 and Larry T. Taylor 1 1 Department
More informationREFINED RESTEK. Detailed Hydrocarbon Analysis. innovative petrochemical solutions. Featuring Rtx -DHA Columns
RESTEK REFINED innovative petrochemical solutions Detailed Hydrocarbon Analysis Featuring Rtx -DHA Columns Individually tested to meet DHA method criteria; guaranteed column-to-column reproducibility.
More informationDetermination of Sudan Dyes I IV in Curry Paste
Determination of Sudan Dyes I IV in Curry Paste Suparerk Tukkeeree and Jeffrey Rohrer 2 Thermo Fisher Scientific, Bangkok, Thailand; 2 Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 23 Key
More informationDETECTORS Photo Ionization Detector - PID. Overview
Overview DETECTORS The Photo Ionization Detector (PID) responds to all molecules whose ionization potential is below 10.6eV, including aromatics and molecules with carbon double bonds. The PID is nondestructive,
More informationApplication Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER
Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D7423-09 using AC OXYTRACER Fast Analysis in
More informationThe Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector
Authors: The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector Introduction Specification of the hydrocarbon composition of LPG is required as traces
More informationAgilent 7693A Automated Liquid Sampler
Agilent 7693A Automated Liquid Sampler Specifications Overview The Agilent 7693A is a state-of-the-art sample handling and injection system that provides the highest levels of precision and reliability
More informationCustomer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard
Agilent Preventive Maintenance provides factory recommended service for your analytical systems to assure reliable operation and the accuracy of your results. Delivered by highly-trained and certified
More information260 Solvent Vapor Exit Accessory
Theory of operation The Solvent Vapor Exit (SVE) is a GC accessory for performing large volume injections with a Cool On-Column inlet (COC). At the start of the run, the SVE solenoid valve is open as the
More informationAZ Chrom s.r.o.
AZ Chrom s.r.o. azetchrom@hplc.sk www.azetchrom.sk ÙÝ Í ó Ê ²ñÞ «µ» ² «³»² Ï«½µ Ð ½µ Ù«¼» Í ²¹ ²¹» Í» ²» Ô ²» Ñó ²¹ Ú» Ú ²¹» Ì» Ú» Ako nás možno kontaktova : AZ Chrom s.r.o. Robotnícka 10 831 03 Bratislava
More informationperformance productivity reliability
Solutions that meet your demands for: performance productivity reliability Excellent Choices for Global Hydrocarbon Processing Applications Crude Oil & Natural Gas Natural Gas > Return to Table of Contents
More informationPhenomenex Exclusive! Inlet Base Seals. Agilent 5000, 6000 and. Guaranteed fit for Series GCs. Easy Seals
Phenomenex Exclusive! Easy Seals Inlet Base Seals Guaranteed fit for Agilent 5000, 6000 and 7000 Series GCs The Easy Seals Gold Standard No Strained Hands No Troublesome Washers or Adaptors Continuing
More information2-D GC/MS and the Deans Switch: Theory and Applications. Fred Feyerherm, Agilent Technologies Inc. Houston, Tx
2-D GC/MS and the Deans Switch: Theory and Applications Fred Feyerherm, Agilent Technologies Inc. Houston, Tx Introduction Simplified 2-D GC can be used to accomplish the separation of specific analytes
More informationLiquefied Gas Injector. Solution for the Sampling and Analysis of Liquefied Gases
Liquefied Gas Injector Solution for the Sampling and Analysis of Liquefied Gases Safe and Representative Sampling of Liquefied Gases The analysis of impurities and contaminants in liquefied gases is an
More information6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR
6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR A15836 This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use
More informationLiquid Injection Techniques in GC and GC-MS
Liquid Injection Techniques in GC and GC-MS Inge De Dobbeleer Regional Marketing Manager GC and GC-MS, EMEA Thermo Fisher Scientific, Breda/The Netherlands The world leader in serving science Content Hot
More informationGC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585
GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc. 850 Centerville
More informationMicrosep Low-Bleed PTFE-Lined Septa. XLB Septa
Septa Chromatogram Cross Reference Guide Actual Size Drawing Measurement 1/4" (6mm) Shimadzu*, SRI* Chromatographs 3/8" (9.5mm) 7/16" (11mm) Antek, Finnigan 9100, 9600, Gow- Mac, Agilent 5700, 5880 & 5900,
More informationAppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS
AppNote 6/26 Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) Andreas Hoffmann GERSTEL GmbH & Co.KG, Eberhard-Gerstel-Platz 1,
More informationGC Method Compliance and Large Valve Oven Application
GC Method Compliance and Large Valve Oven Application Solution Focused Kelly Beard June 20, 2016 Compliant Methods Compliant Methods What are our options? Analyzers and Analyzer Kits High Volume Well Known
More informationRestek Corporation 110 Benner Circle, Bellefonte, Pa Barry L. Burger, Neil Johansen, Valerie Gamble, Donald Rhoades
High Speed PONA Analysis For Detailed Hydrocarbon Analysis Extended (DHAX) Using Hydrogen Carrier Gas For The Determination of Individual Components In Spark Ignition Fuels Barry L. Burger, Neil Johansen,
More informationSTI OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY
OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY Extended range HPLC has been investigated for many years; however, progress was delayed because of
More informationProductivity Improvements for Dual Column Applications: USP <467> and Blood Alcohol Analysis
Productivity Improvements for Dual Column Applications: USP and Blood Alcohol Analysis Rebecca Veeneman, PhD Applications Chemist Agilent Technologies The GC Usability Gap Usability lags features
More informationFast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC
Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC Technical Overview Trace gas analysis is a challenge in today s world. The ability to analyze lower component
More information[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS
MS Identification of Trace level Impurities from a Non-MS Compatible Mobile Phase Using ACQUITY UPLC System with 2D Technology by Heart-cutting and Online Sample Concentration Bronsky Gopinadh, Dilshad
More information