The Importance of Liquid Handling Details and Their Impact on your Assays

Size: px
Start display at page:

Download "The Importance of Liquid Handling Details and Their Impact on your Assays"

Transcription

1 The Importance of Liquid Handling Details and Their Impact on your Assays European Lab Automation Hamburg, Germany May 30, 2012 John Thomas Bradshaw, Ph.D. Artel, Inc.

2 Automation is My Friend Automated liquid handlers (Mr. Robot) can be powerful tools when their performance is Measured and Controlled.

3 What Do You WANT and EXPECT From Your Liquid Handler? Simple goals for your liquid handler. It should Reproducibly deliver the target volume with high accuracy & precision Behave it should have consistent, predictable behavior according to the programmed methods and scripts (volume transfer, washing, mixing, alignment with labware, etc.)

4 Assessing Liquid Handler Performance How is performance of an automated liquid handler evaluated, monitored and/or maintained? MVS Gravimetric methods Fluorescence/Absorbance methods Combination Other home-brew methods Accuracy and precision should be assessed on a tipby-tip basis some of the above methods do not capture this important information without a lot of time, skill and effort

5 Overview Overview of MVS Dual-Dye Photometric Method

6 Output Report: Performance Well-by-well and Tip-by-tip Tip-by-tip data Well-by-well data Failed wells Failed tip

7 Precision and Accuracy: The Impact of Not Knowing

8 Monitoring Liquid Handler Performance at 10 ml: Precision Only All experiments A G seem relatively similar Precision only measurements the amount of volume, may or may not be correct, but all the volume transfers are repeatable; all tips are performing the same.

9 Monitoring Liquid Handler Performance at 10 ml: Precision and Accuracy By measuring accuracy, reagent concentration can be determined tip-by-tip and well-by-well and the assay s characteristics are more defined, enabling a more quantitative methodology.

10 Monitoring Liquid Handler Performance at 10 ml: Knowing Accuracy, and Therefore, Concentration 9.97 ml 6.80 ml 10 ml of critical compound (2 mm), 40 ml total working volume Method D (6.80 ml)(2 mm)/(40 ml) = 0.34 mm Method G (9.97 ml)(2 mm)/(40 ml) = mm % difference in compound concentration = 35.2% (if experimental total working volume equals theoretical)

11 Liquid Handler Optimization

12 Optimizing Liquid Handler Performance After Sequential Parameter Adjustment Sequential Parameter Adjustment Albert & Bradshaw, J. Assoc. Lab. Autom., 2007, 12,

13 Optimizing Liquid Handler Performance After Sequential Parameter Adjustment By tweaking and measuring (iteratively), liquid handler methods can easily be optimized when there is a standardized yard stick to compare results

14 Facilitating a Bench-to-Automation Transfer of a PCR Assay

15 Facilitating Method Transfer: Assay Transfer and Liquid Handling Steps Target volumes delivered to the working assay = Target volumes delivered to the transferred assay Who or what is used for the volume transfer steps? How are the volumes measured? Quantified? Are the volumes equal or comparable?

16 Troubleshooting After Transferring a Manual PCR Assay to Automation: ACCURACY MATTERS (1 of 2) Manual Method Automated Method 10 µl Sample In Excess; not a critical volume transfer step 15 µl Master Mix CRITICAL STEP; accuracy required Thermocycler for 1.5 hours

17 Accuracy Matters: Transferring & Troubleshooting a Manual PCR Assay to Automation (2 of 2) Manual PCR assay works Automated PCR assay does not work MVS used for liquid handler optimization, automation accurately dispenses critical target volume (15 ml) Automated PCR assay still does not work MVS used to assess technician pipetting proficiency The technician was over-pipetting master mix (15.67 ml) The automation was adjusted to deliver the same volume as the manual method Automated PCR assay worked SUCCESS! The automation was not to blame

18 Measuring CV After Mixing: Aspirate/Dispense Cycles vs. MVS Shaker 1-tip, 16 reps, three different target volumes; 96-well plate After each dispense, 3 common aspirate/dispense cycles used to mix Read plate with MVS (1) Re-read plate with MVS reader (2) Shake Again (3) Read plate a third time Shaken on MVS Plate Shaker Same plate

19 Measuring CV After Mixing: Aspirate/Dispense Cycles vs. MVS Shaker The original mixing method (aspirate/dispense steps) was not efficient. Customer had to notify other sister sites using same method. Note diffusion time helped the mixing process for the 2-uL data in the original mixing method.

20 Identifying a Bad-tip-in-the-box

21 Identifying A Bad Tip In the Box-of-Tips Situation: 8-tip tool on a Biomek Disposable generic tips 1-use tips wet dispense 1 ml or 15 ml 12 reps across 96-w plate Generic Tip Box 96-well destination plate 1. Pick up tips 2. Aspirate target 3. Wet-dispense target into buffer 4. Dispose of tips

22 Repeatedly Identifying A Bad Tip (8-tip robot, wet dispense, 12 reps across 96-w plate, HEAT MAP) Plate 1, Box 1 well 9D off Plate 2, Box 2 tip 9D did not deliver properly, again Plate 3, Box 3 (place rotated) 180 deg rotation of plate, showing Plate 3, well Box3 4E = 9D well location well 4E off Plate 4, Box 4 tip 9D swapped with tip 1D before run.

23 Monitoring Sequential Dispenses Uncovering Systematic Error

24 Target Volume Accuracy Decreases After 5 th Replicate Across Plate After 5 th rep, the mean volume decreases for the next 5 reps 70 ul 110 ul

25 Target Volume Accuracy Decreases After 5 th Replicate Across Plate Drop in accuracy after 5 th replicate

26 Standardized Platform for Volume Verification

27 Directly Comparing Identical Robots & Methods at Multiple Locations - Three different forensics laboratories (in three different cities) - Same robot models (Biomek 3000) - Same head and tip types - Exact methods and techniques employed in each lab 2 ml (P20 tool) 10 ml (P200L tool)

28 Directly Comparing Operator Pipetting Proficiency * Tech 1 Tech 2 J. Assoc. Lab. Autom., 2007, 12, * Two different technicians dispensing 20 ml of same reagents in the same laboratory with the same 8-tip, electronic pipette within minutes of each other

29 Conclusions: Importance of Knowing Liquid Handler Performance & Behavior Liquid handlers are powerful tools, but need to be Measured and Controlled. Measuring volume accuracy and precision for the assay of interest will help build confidence in the capabilities of your liquid handler. Not Measuring = Not Knowing

30 Thank you 25 Bradley Drive Westbrook, Maine USA Tel: (207) Fax: (207)

From Patient to Plate

From Patient to Plate From Patient to Plate Development of Highly Sensitive Clinical PK Assays Andrew Macintyre M.D., Ph.D. Director of Research and Development Sarah Christensen B.S. Research Scientist II Custom Assay Service

More information

Environmental Impact on Liquid Handling

Environmental Impact on Liquid Handling Environmental Impact on Liquid Handling Pascal Silversmet, GSK Katie Haith, Tecan George Rodrigues, Artel 1 GSK : Who we are Pharmaceuticals We develop and make medicines to treat a range of conditions

More information

Evaluating and Determining Precision and Accuracy for Nanoliter Volumes of Viscous Solutions on an Automated Liquid Handling Instrument

Evaluating and Determining Precision and Accuracy for Nanoliter Volumes of Viscous Solutions on an Automated Liquid Handling Instrument Evaluating and Determining Precision and Accuracy for Nanoliter Volumes of Viscous Solutions on an Automated Liquid Handling Instrument Application Note 231 Niki Popp, Tim Hegeman, Joan Stevens, Ph.D.

More information

NEW INSTRUMENT. 08th November / MKT. We announce the launching of the new member of the Biokit Elisa Instrumentation family: DS2

NEW INSTRUMENT. 08th November / MKT. We announce the launching of the new member of the Biokit Elisa Instrumentation family: DS2 08th November 2010 293/ MKT NEW INSTRUMENT We announce the launching of the new member of the Biokit Elisa Instrumentation family: DS2 The DS2 instrument is the new Biokit automated ELISA processor. The

More information

MB3600-CH30 Laboratory FT-NIR analyzer for biodiesel applications Suitable for production optimization and product quality assessment

MB3600-CH30 Laboratory FT-NIR analyzer for biodiesel applications Suitable for production optimization and product quality assessment Measurement & Analytics Measurement made easy MB3600-CH30 Laboratory FT-NIR analyzer for biodiesel applications Suitable for production optimization and product quality assessment FT-NIR optimizing productivity

More information

ASTM B117 Testing Quality Control

ASTM B117 Testing Quality Control ASTM B117 Testing Quality Control ASTM B117 Testing, also known as a Standard Practice for Operating Salt Spray (Fog), is used to analyze relative corrosion for specimens of metals and coated metals exposed

More information

TopPette Mechanical Pipettes

TopPette Mechanical Pipettes TopPette Mechanical Pipettes Features for TopPette - Lightweight, ergonomic, low force design - Digital display clearly reads volume setting - The pipettes cover volume range of 0.1μl to 10ml - Easy to

More information

Biodiesel Analysis Utilizing Mini-Scan - Handheld Analyzer V.C. Gordon PhD, Bonanza Labs

Biodiesel Analysis Utilizing Mini-Scan - Handheld Analyzer V.C. Gordon PhD, Bonanza Labs Biodiesel Analysis Utilizing Mini-Scan - Handheld Analyzer V.C. Gordon PhD, Bonanza Labs Overview According to the National Biodiesel Board, biodiesel production in the United States reached 450 million

More information

SmartChip MyDesign Kit User Manual

SmartChip MyDesign Kit User Manual SmartChip MyDesign Kit User Manual Cat. Nos. 640032, 640036 (102017) 1290 Terra Bella Avenue, Mountain View, CA 94043, USA U.S. Technical Support: techus@takarabio.com United States/Canada 800.662.2566

More information

QUALITY ASSURANCE & LAB ACCREDITATION

QUALITY ASSURANCE & LAB ACCREDITATION TOP 9 QUALITY ASSURANCE & LAB ACCREDITATION Control Charts Accreditation of laboratories 6./7.Dec 2011 Workshop ANKARA TOP 9 Quality 1 TOP 9 QUALITY ASSURANCE & LAB ACCREDITATION Control Charts Accreditation

More information

Yesterday, today and tomorrow

Yesterday, today and tomorrow Yesterday, today and tomorrow Why Scientific has consistently been the leading producer of transfer pipets for over 35 years. Design Grips easily with latex gloves no slipping Clear graduated markings

More information

Single and 8-channel Electronic Pipettes. Operation Manual Version: 2.9

Single and 8-channel Electronic Pipettes. Operation Manual Version: 2.9 Single and 8-channel Electronic Pipettes Operation Manual Version: 2.9 Table of Contents 1. Safety Precautions...1 1.1. Operation Safety Precautions...1 1.2. Battery Safety...1 2. General Description...3

More information

DURAN GROUP DISPENSER & DURAN GROUP DISPENSER PRO. A safe and precise dispensing system

DURAN GROUP DISPENSER & DURAN GROUP DISPENSER PRO. A safe and precise dispensing system DURAN GROUP DISPENSER & DURAN GROUP DISPENSER PRO A safe and precise dispensing system REMOVAL OF LIQUID MEDIA MADE EASY DURAN laboratory glassware ensures outstanding safety and precision when working

More information

8.0 pipet4u pipets General Information

8.0 pipet4u pipets General Information 8.0 pipet4u pipets General Information An incredible combination of accuracy, precision and value High Comfort, Comfortable and Low Effort Pipetting The new design with soft grip allows you to use less

More information

Breath Alcohol Simulator Solution Analysis No Summary Report

Breath Alcohol Simulator Solution Analysis No Summary Report Collaborative Testing Services, Inc FORENSIC TESTING PROGRAM Breath Alcohol Simulator Solution Analysis No. 18-568 Summary Report Each sample pack consisted of two bottles of solution which participants

More information

Discovery of Design Methodologies. Integration. Multi-disciplinary Design Problems

Discovery of Design Methodologies. Integration. Multi-disciplinary Design Problems Discovery of Design Methodologies for the Integration of Multi-disciplinary Design Problems Cirrus Shakeri Worcester Polytechnic Institute November 4, 1998 Worcester Polytechnic Institute Contents The

More information

TDG-F-113 CEC New Test Development Proposal for a New Engine Fuels Test Procedure

TDG-F-113 CEC New Test Development Proposal for a New Engine Fuels Test Procedure TDG-F-113 CEC New Test Development Proposal for a New Engine Fuels Test Procedure DISI (Direct Injection spark ignited engine) Injector fouling Test 1. Demonstrated need- The proposed test will address

More information

Northville Product Services and Colonial Pipeline B5 Soy Test Data and Report February Soy blended low sulfur diesel

Northville Product Services and Colonial Pipeline B5 Soy Test Data and Report February Soy blended low sulfur diesel Scope of Work: Bio Diesel Trial NPS-73-441 Soy blended low sulfur diesel Colonial Pipeline Company (CPL) conducted a test movement of 74 grade (500 ppm highway diesel) blended with 5% biodiesel (soy oil).

More information

Dispensette. Testing Instructions (SOP) 1. Introduction. May 2009

Dispensette. Testing Instructions (SOP) 1. Introduction. May 2009 Dispensette Testing Instructions (SOP) May 2009 1. Introduction The standard ISO DIS 8655 describes both the design and the testing of the bottle-top dispenser. The following Testing Instructions describe

More information

Objectives. WP7: On-engine aftertreatment systems. WP Leader: Jukka Leinonen. Partners:

Objectives. WP7: On-engine aftertreatment systems. WP Leader: Jukka Leinonen. Partners: WP7: On-engine aftertreatment systems Objectives Integration of SCR (Selective Catalytic Reduction) with the existing strong Miller cycle 4-stroke diesel engine and combining it with particulate emission

More information

multi Mid. vol. Max. vol. Min. vol. < 3.0% 1) < 1.0% < 0.6% < 0.6% < ± 1.5% < ± 0.8% < ± 0.7% < ± 0.8% < ± 1.5% < ± 0.8% < ± 0.7% < 1.0% < 0.

multi Mid. vol. Max. vol. Min. vol. < 3.0% 1) < 1.0% < 0.6% < 0.6% < ± 1.5% < ± 0.8% < ± 0.7% < ± 0.8% < ± 1.5% < ± 0.8% < ± 0.7% < 1.0% < 0. multichannel manual micropipettes Built with the users in mind, these pipettes combine high-tech materials with exceptional, user-friendly ergonomics. They extend pipetting possibilities in 96-well microplates

More information

Perfection in Liquid Handling PRECISE AND CONVENIENT PIPETTING

Perfection in Liquid Handling PRECISE AND CONVENIENT PIPETTING Perfection in Liquid Handling PRECISE AND CONVENIENT PIPETTING 21 VITLAB micropipette The VITLAB piston-operated pipettes are the ideal manual pipettes for demanding laboratory applications, and have all

More information

Thermo Scientific Finnpipette Performance Specifications

Thermo Scientific Finnpipette Performance Specifications 1 Thermo Scientific Finnpipette Performance Salla Ruosaari, Product Manager, Thermo Fisher Scientific, Vantaa, Finland. Application Notes Key words ISO 8655 - Calibration standard for the piston operated

More information

3/E3x ual Repeater E3/E3x

3/E3x ual Repeater E3/E3x 3/E3x ual Repeater E3/E3x N) manual Register your instrument! www.eppendorf.com/myeppendorf Multipette E3/E3x Repeater E3/E3x Operating manual Copyright 2016 Eppendorf AG, Germany. All rights reserved,

More information

Sekisui Virotech GmbH ELISA System Diagnostic Error Sources and Approaches

Sekisui Virotech GmbH ELISA System Diagnostic Error Sources and Approaches Sekisui Virotech GmbH ELISA System Diagnostic Error Sources and es The following questionnaire shall help to identify the error source and provide the approach as quickly as possible. It is not assay specific

More information

DxC 700 AU Job Aid Booklet

DxC 700 AU Job Aid Booklet DxC 700 AU Job Aid Booklet For Training Purposes Only These job aids are shortened versions of procedures found in the references below. Information in the job aid is correct as of the date published.

More information

MANUAL DISPENSING REDEFINED CERTUS HANDY LIQUID DISPENSER

MANUAL DISPENSING REDEFINED CERTUS HANDY LIQUID DISPENSER MANUAL DISPENSING REDEFINED CERTUS HANDY LIQUID DISPENSER REVOLUTIONARY FUSION OF MANUAL AND AUTO- MATED DISPENSING CERTUS HANDY unlocks a new dimension of manual digital dispensing systems for unmatched

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 NORTHERN BALANCE AND SCALE, INC. 9556 W. Bloomington Freeway Bloomington, MN 55431 Joey Kaluser Phone: 952 881 7716 CALIBRATION Valid

More information

Autoplex Specifications

Autoplex Specifications Autoplex Specifications Autoplex G1 w/heating coil Stainless steel tip holds up to 150µl of liquid Probe tubing holds up to 3.18ml of liquid (length=88in, diameter=0.673mm/0.053in, 36.1µl/inch-high surface

More information

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment FAST SHIPPING AND DELIVERY TENS OF THOUSANDS OF IN-STOCK ITEMS EQUIPMENT DEMOS HUNDREDS OF MANUFACTURERS SUPPORTED

More information

Cold Vapor Mercury Apparatus* Purpose Figure 1 Cold Vapor Mercury Apparatus Assembly Overview

Cold Vapor Mercury Apparatus* Purpose Figure 1 Cold Vapor Mercury Apparatus Assembly Overview 26744-88 Cold Vapor Mercury Apparatus* Purpose This instruction sheet shows how to assemble the apparatus set (26744-00) necessary for performing the Hach Cold Vapor Mercury Method (method 10065). Figure

More information

Agilent 7693A Automated Liquid Sampler

Agilent 7693A Automated Liquid Sampler IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7693A Automated Liquid Sampler

More information

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc.

More information

Eppendorf Repeater Pro Instruction Manual Manual de Instrucciones

Eppendorf Repeater Pro Instruction Manual Manual de Instrucciones Eppendorf Repeater Pro Instruction Manual Manual de Instrucciones Technical Data / Datos técnicos Technical Data / Datos técnicos See chapter 4 / ver apartado 4 Systematic error/ Error sostemàtico (Inaccuracy/Incorreción)

More information

economics and performance Spark Holland introduces OPTIMAS The optimum balance of BETTER SAMPLE CARE

economics and performance Spark Holland introduces OPTIMAS The optimum balance of BETTER SAMPLE CARE BETTER SAMPLE CARE Spark Holland introduces OPTIMAS The optimum balance of economics and performance Spark Holland introduces OPTIMAS The optimum balance of economics and performance OPTIMAS The optimum

More information

Armands Senfelds, Leonids Ribickis, Ansis Avotins, Peteris Apse-Apsitis

Armands Senfelds, Leonids Ribickis, Ansis Avotins, Peteris Apse-Apsitis Development of 600V Industrial DC Microgrid for Highly Automated Manufacturing Applications: Factory and Laboratory Infrastructure Experience Armands Senfelds, Leonids Ribickis, Ansis Avotins, Peteris

More information

PRIEMONIŲ IŠPARDAVIMAS

PRIEMONIŲ IŠPARDAVIMAS PRIEMONIŲ IŠPARDAVIMAS INOSPECTRA, UAB Breslaujos g. 3B, 44403 Kaunas Tel. +370 37 280702 info@inospectra.com www.inospectra.com Pipette OL 154 65 58% With the ergonomic shape and the stylish colour the

More information

N) manual. Register your instrument! Eppendorf Reference 2. Operating manual

N) manual. Register your instrument!  Eppendorf Reference 2. Operating manual erence ual 2 N) manual Register your instrument! www.eppendorf.com/myeppendorf Operating manual Copyright 2016 Eppendorf AG, Germany. All rights reserved, including graphics and images. No part of this

More information

Register your instrument! Eppendorf Reference 2. Operating manual

Register your instrument!  Eppendorf Reference 2. Operating manual Register your instrument! www.eppendorf.com/myeppendorf Operating manual Operating Eppendorf see on Fig. Tab. p. p. manual Reference 2 Copyright 2016 Eppendorf AG, Germany. All rights reserved, including

More information

Quality Control in Mineral Exploration

Quality Control in Mineral Exploration Quality Control in Mineral Exploration Controlling the Quality of Information from Field to Data Base Not to be reproduced without written permission Quality Control in Mineral Exploration There many goals

More information

Qualification of a precision pattern dispenser

Qualification of a precision pattern dispenser Qualification of a precision pattern dispenser Raf Van Ammel, Katarzyna Sobiech-Matura, Uwe Wätjen 2016 EUR 28268 EN This publication is a Technical report by the Joint Research Centre, the European Commission

More information

DG CLIMA studies on CO2 emissions from vehicles

DG CLIMA studies on CO2 emissions from vehicles DG CLIMA studies on CO2 emissions from vehicles KICK-OFF MEETING of ERMES Group 22-23 June 2010 Bruxelles Fabio Dalan DG CLIMA, Unit C2 Current areas of activities Passenger cars - Regulation (EC) 443/2009

More information

Progress made simple!

Progress made simple! Progress made simple! A d j u s ta b l e - V o l u m e P i p e t t e EASY TO USE ADVANCED ERGONOMICS TRUSTED RESULTS EASY TO USE ADVANCED ERGONOMICS TRUSTED RESULTS Progress made simple! PIPETMAN M is

More information

3/E3x ual Repeater E3/E3x

3/E3x ual Repeater E3/E3x 3/E3x ual Repeater E3/E3x N) manual Register your instrument! www.eppendorf.com/myeppendorf Multipette E3/E3x Repeater E3/E3x Operating manual Copyright 2017 Eppendorf AG, Germany. All rights reserved,

More information

Thermo Scientific Matrix Manual Pipette User Manual

Thermo Scientific Matrix Manual Pipette User Manual Thermo Scientific Matrix Manual Pipette User Manual Thermo Scientific Matrix Manual Pipette User Manual Trademarks and Copyright 2007 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are

More information

Ion PGM System with Reagent and Wash Bottles attached

Ion PGM System with Reagent and Wash Bottles attached Ion PGM System with Reagent and Wash Bottles attached Product information A Ion PGM System B C D I G E H F Label A B C D E F G H I Component Touchscreen Chip clamp Grounding plate Power button Reagent

More information

Chem(Bio) Week 10 Bomb Calorimetry of Biodiesel

Chem(Bio) Week 10 Bomb Calorimetry of Biodiesel Lab Overview: Water/ check status plants (at some point) Biodiesel Workup (may not be necessary) Bomb Calorimetry of biodiesel (calorimeter calibration will be necessary) IR spectroscopy of biodiesel (possibly

More information

Determining the Ethanol Content of Denatured Fuel Ethanol Using Near Infrared. Gulf Coast Conference Patrick Ritz PAC LP

Determining the Ethanol Content of Denatured Fuel Ethanol Using Near Infrared. Gulf Coast Conference Patrick Ritz PAC LP Determining the Ethanol Content of Denatured Fuel Ethanol Using Near Infrared Gulf Coast Conference Patrick Ritz PAC LP Global Ethanol Use Consumption of fuel-grade ethanol is on the rise Produced from

More information

Siemens PLM Software develops advanced testing methodologies to determine force distribution and visualize body deformation during vehicle handling.

Siemens PLM Software develops advanced testing methodologies to determine force distribution and visualize body deformation during vehicle handling. Automotive and transportation Product LMS LMS Engineering helps uncover the complex interaction between body flexibility and vehicle handling performance Business challenges Gain insight into the relationship

More information

Dilutor Module for TriPlus RSH

Dilutor Module for TriPlus RSH Thermo Scientific Dilutor Module for TriPlus RSH Instruction Manual Additional section of the TriPlus RSH User Guide and Hardware Manual 31709661 Revision A December 2015 2015 Thermo Fisher Scientific

More information

Infinite F50 family INNOVATIVE EIGHT CHANNEL ABSORBANCE READERS WITH ADVANCED LED TECHNOLOGY. 21 CFR Part 11* 98/79/EC IVD-D*

Infinite F50 family INNOVATIVE EIGHT CHANNEL ABSORBANCE READERS WITH ADVANCED LED TECHNOLOGY. 21 CFR Part 11* 98/79/EC IVD-D* Infinite F50 family INNOVATIVE EIGHT CHANNEL ABSORBANCE READERS WITH ADVANCED LED TECHNOLOGY 21 CFR Part 11* 98/79/EC IVD-D* 02 Innovative. Sustainable. Smart The Infinite F50 is Tecan s next generation,

More information

elektronik Designing vehicle power nets A single simulation tool from initial requirements to series production

elektronik Designing vehicle power nets A single simulation tool from initial requirements to series production www.atzonline.de elektronik 04 April 2013 Volume 8 Offprint from ATZelektronik 4/2013 Springer Automotive Media Springer Fachmedien Wiesbaden GmbH for Bosch Engineering Designing vehicle power nets A single

More information

3D101 Part 2: Keep Calm and Culture On: InSphero s easy to use GravityPLUS Hanging Drop System and GravityTRAP TM ULA Plate

3D101 Part 2: Keep Calm and Culture On: InSphero s easy to use GravityPLUS Hanging Drop System and GravityTRAP TM ULA Plate 3D101 Part 2: Keep Calm and Culture On: InSphero s easy to use GravityPLUS Hanging Drop System and GravityTRAP ULA Plate www.insphero.com GravityTRAP TM ULA Plate 1 A complete portfolio of 3D cell culture

More information

Correlating TBP to Simulated Distillations. COQA Long Beach, CA

Correlating TBP to Simulated Distillations. COQA Long Beach, CA Correlating TBP to Simulated Distillations COQA Long Beach, CA 2017-10-05 Maxxam Analytics Overview Maxxam Analytics A Bureau Veritas Group Company 69,000 1,400 3 Maxxam Analytics Part of the Bureau Veritas

More information

OMA I, MLA/MLT I, LLA I

OMA I, MLA/MLT I, LLA I Is NIST SRM2806b Responsible for the Sudden Increase in Particle Counts you have Seen on your Oil Analysis Reports? Alistair Geach, CLS, OMA I, MLA/MLT I, LLA I CINRG Systems Inc., Burlington, ON, CANADA

More information

Operator s Guide for Mechanical Volume-Set Pipettes (Ovation M)

Operator s Guide for Mechanical Volume-Set Pipettes (Ovation M) Operator s Guide for Mechanical Volume-Set Pipettes (Ovation M) L070-0001-001 Rev. B TABLE OF CONTENTS Introduction Product Description & Specifications 1-1 Pipette Components 1-2 BioNatural Pipetting

More information

Vivacon 2 Ultrafiltration devices desalting, concentrating

Vivacon 2 Ultrafiltration devices desalting, concentrating Vivacon 2 Ultrafiltration devices desalting, concentrating Reproducible DNA sample desalting and concentration Vivacon 2 centrifugal concentrators offer the optimal solution for DNA concentration and buffer

More information

Draft Project Deliverables: Policy Implications and Technical Basis

Draft Project Deliverables: Policy Implications and Technical Basis Surveillance and Monitoring Program (SAMP) Joe LeClaire, PhD Richard Meyerhoff, PhD Rick Chappell, PhD Hannah Erbele Don Schroeder, PE February 25, 2016 Draft Project Deliverables: Policy Implications

More information

5 Levels. CVC 123 Calibration Verification Controls. Set: Level 1: Level 2: Level 3: Level 4: Level 5: INTENDED USE

5 Levels. CVC 123 Calibration Verification Controls. Set: Level 1: Level 2: Level 3: Level 4: Level 5: INTENDED USE CVC 123 5 Levels CVC 123 Calibration Verification Controls Set: Level 1: Level 2: Level 3: Level 4: Level 5: 212319 14116 11522 11621 11722 14217 Set: CVC 123 Level 1: Level 2: For In Vitro Diagnostic

More information

Lab Session #1 Initiation Systems and Types of Explosives

Lab Session #1 Initiation Systems and Types of Explosives Lab Session #1 Initiation Systems and Types of Explosives The main goal of this laboratory session is to provide a practical experience in the use of different type of explosives and initiations systems.

More information

Accelerated Testing of Advanced Battery Technologies in PHEV Applications

Accelerated Testing of Advanced Battery Technologies in PHEV Applications Page 0171 Accelerated Testing of Advanced Battery Technologies in PHEV Applications Loïc Gaillac* EPRI and DaimlerChrysler developed a Plug-in Hybrid Electric Vehicle (PHEV) using the Sprinter Van to reduce

More information

4 ual Repeater M4. N) manual. Register your instrument! Multipette M4 Repeater M4. Operating manual

4 ual Repeater M4. N) manual. Register your instrument!   Multipette M4 Repeater M4. Operating manual 4 ual Repeater M4 N) manual Register your instrument! www.eppendorf.com/myeppendorf Operating manual Copyright 2018 Eppendorf AG, Germany. All rights reserved, including graphics and images. No part of

More information

Improving Analog Product knowledge using Principal Components Variable Clustering in JMP on test data.

Improving Analog Product knowledge using Principal Components Variable Clustering in JMP on test data. Improving Analog Product knowledge using Principal Components Variable Clustering in JMP on test data. Yves Chandon, Master BlackBelt at Freescale Semiconductor F e b 2 7. 2015 TM External Use We Touch

More information

Excessive Waste. Some of the grease is used to supplement feed farms but majority of it ends up in landfills

Excessive Waste. Some of the grease is used to supplement feed farms but majority of it ends up in landfills Excessive Waste According to the Environmental Protection Agency (EPA), hotels and restaurants in the U.S. generate at least 3 billion gallons of waste vegetable oil annually * Note: this figure excludes

More information

Crude Evaluation Best Practices

Crude Evaluation Best Practices Crude Evaluation Best Practices IDTC 2017 May 16 & 17 Dubrovnik, CROATIA Malek Masri www.haverly.com Crude Evaluation Optimized Process IDTC 2017 May 16 & 17 Assay accuracy and update Importance of crude

More information

OPTIMAL BATCH DISTILLATION SEQUENCES USING ASPEN PLUS

OPTIMAL BATCH DISTILLATION SEQUENCES USING ASPEN PLUS OPTIMAL BATCH DISTILLATION SEQUENCES USING ASPEN PLUS Bradley H. Cook, Matthew J. Engel, and Anthony J. Zehnder Air Products and Chemicals, Inc. 7201 Hamilton Blvd. Allentown, PA 18195-1501 Craig S. Landis

More information

Tier 2 Advanced Power Strips Revisiting Behavior Based Models for Estimation of Savings in Laboratory and Field Trial Evaluations

Tier 2 Advanced Power Strips Revisiting Behavior Based Models for Estimation of Savings in Laboratory and Field Trial Evaluations Tier 2 Advanced Power Strips Revisiting Behavior Based Models for Estimation of Savings in Laboratory and Field Trial Evaluations Michael J. Klopfer, Joy Pixley, Klint Segarra, G.P. Li California Plug

More information

Steady-State Engine Modeling for Calibration: A Productivity and Quality Study

Steady-State Engine Modeling for Calibration: A Productivity and Quality Study Steady-State Engine Modeling for Calibration: A Productivity and Quality Study MathWorks Automotive Conference 2007 Hyatt Regency, Dearborn, MI Ulrike Schoop John Reeves Satoru Watanabe Ken Butts IAV GmbH

More information

University Of California, Berkeley Department of Mechanical Engineering. ME 131 Vehicle Dynamics & Control (4 units)

University Of California, Berkeley Department of Mechanical Engineering. ME 131 Vehicle Dynamics & Control (4 units) CATALOG DESCRIPTION University Of California, Berkeley Department of Mechanical Engineering ME 131 Vehicle Dynamics & Control (4 units) Undergraduate Elective Syllabus Physical understanding of automotive

More information

The Mutagenic Effects of Crude Oil Fuels on Cell Mutation. Michael Bushnell Pittsburgh Central Catholic High School 9th Grade

The Mutagenic Effects of Crude Oil Fuels on Cell Mutation. Michael Bushnell Pittsburgh Central Catholic High School 9th Grade The Mutagenic Effects of Crude Oil Fuels on Cell Mutation Michael Bushnell Pittsburgh Central Catholic High School 9th Grade The Question Do common crude oil fuels have significant mutagenic properties,

More information

Reli-a-Flex TM Reliance

Reli-a-Flex TM Reliance Reli-a-Flex TM Reliance Precision Mechatronics LLP www.myostat.ca Reli-a-Flex - Accurate transfer of motion between two shafts Patented: - UK No. 2316735 - US No. 6,203,437 B1 - European No. EP 0922168

More information

GENERAL CATALOG L.TEX CORPORATON

GENERAL CATALOG L.TEX CORPORATON GENERAL CATALOG L.TEX CORPORATON 1/8 동방하이테크상사 http://www.allpump.co.kr T.(02)457-6292 F.(02)457-6293 13-12-01 L. Tex Corporation was established in March 2001 to design, develop and manufacture the versatile

More information

Totally Automated Method for the Determination of Sudan Dyes in Food via On-Line Filtration, SPE and HPLC Analysis

Totally Automated Method for the Determination of Sudan Dyes in Food via On-Line Filtration, SPE and HPLC Analysis Totally Automated Method for the Determination of Sudan Dyes in Food via On-Line Filtration, SPE and HPLC Analysis Joan M Stevens, Ph.D., Mark Crawford and Greg Robinson Single Presenter Name Abstract:

More information

Finnpipette Novus. Single Channel & Multichannel. User Manual. Instructions for Use

Finnpipette Novus. Single Channel & Multichannel. User Manual. Instructions for Use User Manual Finnpipette Novus Single Channel & Multichannel Instructions for Use Thermo Fisher Scientific Thermo Scientific Novus SCP/MCP User Manual 1 CONTENT PRODUCT DESCRIPTION...............................

More information

Antibody List. Array Name: Tyrosine Kinase Adaptor Phosphorylation Antibody Array Cat. #: PTK098 Total Antibodies: 98 Revision: A1

Antibody List. Array Name: Tyrosine Kinase Adaptor Phosphorylation Antibody Array Cat. #: PTK098 Total Antibodies: 98 Revision: A1 Array Name: Tyrosine Kinase Adaptor Phosphorylation Antibody Array Cat. #: PTK098 Total Antibodies: 98 Revision: A1 Antibody List ID Antibody Name A Beta actin E Empty G GAPDH N Negative control P Positive

More information

electronic micropipettes

electronic micropipettes electronic micropipettes Pipetting has never been so simple and safe. Precision designed in every aspect, the Acura electro yields professional results of the highest quality, while also offering the ultimate

More information

Synthesis of Optimal Batch Distillation Sequences

Synthesis of Optimal Batch Distillation Sequences Presented at the World Batch Forum North American Conference Woodcliff Lake, NJ April 7-10, 2002 107 S. Southgate Drive Chandler, Arizona 85226-3222 480-893-8803 Fax 480-893-7775 E-mail: info@wbf.org www.wbf.org

More information

Picus Electronic Pipette

Picus Electronic Pipette Picus Electronic Pipette Picus Electronic Pipettes. The Most Sophisticated and Ergonomic Pipette Ever! Picus, the winner of the Red Dot design award and the Fennia Prize Honorary Mention in 2012, is Sartorius

More information

Fibertec 8000 Fully automated Crude and Detergent Fibre analysis

Fibertec 8000 Fully automated Crude and Detergent Fibre analysis Fibertec 8000 Fully automated Crude and Detergent Fibre analysis The Fibertec 8000 is a fully automated system for determination of crude fibre and detergent fibre and related parameters according to standard

More information

Technical Report. 10 th Round Robin Test for Multi-Capillary Pressure Drop Calibration Standards (2014)

Technical Report. 10 th Round Robin Test for Multi-Capillary Pressure Drop Calibration Standards (2014) Physical Test Methods Sub-Group Technical Report 10 th Round Robin Test for Multi-Capillary Pressure Drop Calibration Standards (2014) August 2016 Prepared by: James Vincent Cerulean, Milton Keynes, United

More information

Ballotini Grade Solid Soda Glass Balls $53.33 KG Laboratory glass filters

Ballotini Grade Solid Soda Glass Balls $53.33 KG Laboratory glass filters Part Number Item Group Description Price Unit Description 136-001 Ballotini Grade 5216 Solid Soda Glass Balls $50.49 KG Laboratory glass filters 136-002 Ballotini Grade 5215 Solid Soda Glass Balls $50.00

More information

S Pyatt. UV glue dispensing studies on the Dima Dotmaster

S Pyatt. UV glue dispensing studies on the Dima Dotmaster UV glue dispensing studies on the Dima Dotmaster 1 The study involves the testing of 3 UV cure glues. The candidates are: Loctite 3525 Dymax 6-621 Dymax 3013 Two dot pattern types have been defined, a

More information

Human Sensitivity in Forced Feedback Systems as a Function of Frequency and Amplitude of Steering Wheel Vibrations

Human Sensitivity in Forced Feedback Systems as a Function of Frequency and Amplitude of Steering Wheel Vibrations Human Sensitivity in Forced Feedback Systems as a Function of Frequency and Amplitude of Steering Wheel Vibrations Peter Thomas Tkacik Assistant Professor Mechanical Engineering The University of North

More information

Three Phase Hermetic Protector Application Process

Three Phase Hermetic Protector Application Process Purdue University Purdue e-pubs International Compressor Engineering Conference School of Mechanical Engineering 1996 Three Phase Hermetic Protector Application Process J. Petraitis Texas Instruments Follow

More information

eline Electronic Pipette

eline Electronic Pipette eline Electronic Pipette 1 eline electronic pipettes Proven performance and ergonomics The electronic Biohit eline pipette family has been providing high precision and comfort for years. This award-winning

More information

Article: Sulfur Testing VPS Quality Approach By Dr Sunil Kumar Laboratory Manager Fujairah, UAE

Article: Sulfur Testing VPS Quality Approach By Dr Sunil Kumar Laboratory Manager Fujairah, UAE Article: Sulfur Testing VPS Quality Approach By Dr Sunil Kumar Laboratory Manager Fujairah, UAE 26th September 2017 For over a decade, both regional ECA and global sulphur limits within marine fuels have

More information

Transferpette. Testing Instructions (SOP) 1. Introduction. October 2013

Transferpette. Testing Instructions (SOP) 1. Introduction. October 2013 Transferpette Testing Instructions (SOP) October 2013 1. Introduction The standards DIN EN ISO 8655 describe both the design and the testing of the piston operated pipette Transferpette. The following

More information

SERVICE MANUAL ACROSS Pro Pipette

SERVICE MANUAL ACROSS Pro Pipette SRVIC MNUL SRVIC MNUL CROSS Pro Pipette Ohaus Corporation, 7 Campus Drive, Suite 310, Parsippany, NJ 07054 (973) 377-9000 SRVIC MNUL CROSS Pro Pipette The information contained in this manual is believed

More information

White Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information

White Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Abstract High Temperature Simulated Distillation (High Temp SIMDIS) is one of the most frequently used techniques to determine

More information

National comparison on verification of fuel dispensers

National comparison on verification of fuel dispensers 16 th International Congress of Metrology, 06011 (2013) DOI: 10.1051/ metrology/201306011 C Owned by the authors, published by EDP Sciences, 2013 National comparison on verification of fuel dispensers

More information

Laboratory Roll Mills Type W 110 E / W 150 E Testing Roll Mills Type 110 P / W 150 P Automated Testing Roll Mills Type W 110 AP / W 150 AP

Laboratory Roll Mills Type W 110 E / W 150 E Testing Roll Mills Type 110 P / W 150 P Automated Testing Roll Mills Type W 110 AP / W 150 AP ROLL MILLS, CALENDERING EQUIPMENT Laboratory machines for the processing of polymers Laboratory Roll Mills Type W 110 E / W 150 E Testing Roll Mills Type 110 P / W 150 P Automated Testing Roll Mills Type

More information

N) manual. Register your instrument! Eppendorf Research plus. Operating manual

N) manual. Register your instrument!   Eppendorf Research plus. Operating manual earch ual plus N) manual Register your instrument! www.eppendorf.com/myeppendorf Operating manual Copyright 2018 Eppendorf AG, Germany. All rights reserved, including graphics and images. No part of this

More information

The Proven Industry Leader in Ergonomics and Accuracy. The VWR UHP Pipettor was superior in all of the test categories.

The Proven Industry Leader in Ergonomics and Accuracy. The VWR UHP Pipettor was superior in all of the test categories. The Proven Industry Leader in Ergonomics and Accuracy In 2010, in order to determine the ergonomic benefits of the VWR Ultra High-Performance Pipettor, United States Ergonomics was hired to conduct an

More information

CEMS FOR COAL FIRED POWER STATIONS DEVICE CALIBRATION AND MAINTENANCE

CEMS FOR COAL FIRED POWER STATIONS DEVICE CALIBRATION AND MAINTENANCE CEMS FOR COAL FIRED POWER STATIONS DEVICE CALIBRATION AND MAINTENANCE Deborah Padwater Strategic Industry Manager Corporate Solution Center Process Automation March 2016 CONTINUOUS CEMS FOR THE POWER INDUSTRY

More information

Real-time Bus Tracking using CrowdSourcing

Real-time Bus Tracking using CrowdSourcing Real-time Bus Tracking using CrowdSourcing R & D Project Report Submitted in partial fulfillment of the requirements for the degree of Master of Technology by Deepali Mittal 153050016 under the guidance

More information

Premium Design - Premium Performance - Premium Quality

Premium Design - Premium Performance - Premium Quality WITOPET premium NEW Premium Design - Premium Performance - Premium Quality Looking for the latest pipetting innovation at an attractive price? WITOPET premium is just the right pipette for you. The ergonomic

More information

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract. Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers Application Note Abstract Versa With the rising prices of fossil fuels, more emphasis is being put on renewable resources

More information

VITLAB Dispenser. Standard Operating Procedure (SOP)

VITLAB Dispenser. Standard Operating Procedure (SOP) VITLAB Dispenser Standard Operating Procedure (SOP) VITLAB GmbH Linus-Pauling-Str. 63762 Grossostheim Germany tel: +49 6026 97799-0 fax: +49 6026 97799-30 info@vitlab.com www.vitlab.com Table of contents.

More information

Callisto 100. Cold Filter Plugging Point Tester

Callisto 100. Cold Filter Plugging Point Tester Callisto 100 Cold Filter Plugging Point Tester Cold Filter Plugging Point The Cold Filter Plugging Point (CFPP) method is used to determine the low-temperature operability of diesel fuel, biodiesel, blends

More information

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009 Transferpette Testing Instructions (SOP) May 2009 1. Introduction The standards DIN EN ISO 8655 describe both the design and the testing of the piston operated pipette Transferpette. The following Testing

More information