Small hydrocarbon flow calibration facilities at NMIJ

Size: px
Start display at page:

Download "Small hydrocarbon flow calibration facilities at NMIJ"

Transcription

1 APMP/TCFF Workshop Small hydrocarbon flow calibration facilities at NMIJ Yoshiya Terao

2 Original presentations Expansion of Calibration Flow Range of Small Liquid Hydrocarbon Flow Facility at NMIJ by Kar-Hooi Cheong, Ryouji Doihara Takashi Shimada and Yoshiya TERAO Development of prototype low-flow rate test rig applicable for highly volatile liquids by Ryouji Doihara, Takashi Shimada, Kar-Hooi Cheong and Yoshiya Terao at 9 th International Symposium on Fluid Flow Measurement (ISFFM2015), April 2015

3 LPG Gasoline Kerosene Light oil Heavy oil Flowrate (m 3 /h) 1.E+04 1.E+03 1.E+02 1.E+01 1.E+00 1.E-01 NATIONAL METROLOGY INSTITUTE OF JAPAN Primary standards for hydrocarbon at NMIJ Petroleum terminal Petroleum refinery Tanker Chemical plant Hydraulic machinery 100 L/h Primary standard for hydrocarbon 1.E-02 1.E-03 1.E-04 Fuel efficiency measurement Fuel mixer 1 L/h Small Hydrocarbon Flow Facility 1.E L/h 1.E Viscosity (mpa s)

4 JC08 mode fuel efficiency (km/l) Background and Motivation NATIONAL METROLOGY INSTITUTE OF JAPAN Top 10 car models with best fuel efficiency in Japan hybrid vehicle mini / light vehicle Car models Source:

5 Automotive industry Main targets NATIONAL METROLOGY INSTITUTE OF JAPAN Fuel flowmeter Engine test bench Semiconductor industry Liquid mass flowmeter Vaporizer for manufacturing equipment

6 Automotive industry Fuel flowmeter Measurement conditions: Real gasoline (high volatility), Light oil, 200 L/h 0.02 L/h, Wide flow range (over 1000:1), Wide temperature range, (High pressure over 200MPa). Semiconductor industry Water, Alcohols, Solvents, Toxic liquids, 10 L/h L/h, Wide temperature range (10 85 ), Liquid mass flowmeter

7 Measurement principles Gravimetric calibration: Advantage to accuracy and to simple traceability Disadvantage to evaporation error at the low flow rate range with volatile liquid Water is mostly used as test liquid. Volumetric calibration: SVPs have been adopted in many calibration facilities for oil (including real gasoline), because it can make a closed pipe line system. Some of them have a linear encoder to output many pulses. Syringe pump: Syringe pumps are widely used in medical and analysis fields for ultra-low flow rate. Fluctuation has been observed as a result of the stepper motor driving mechanism.

8 Cheong Kar-Hooi s gravimetric system

9 Calibration Facility Weighing section (liquid collection) Test section (DUT mounting) Flow generation section (tank, pump)

10 Calibration Facility Weighing section (liquid collection) Test section (DUT mounting) Flow generation section (tank, pump) thermostatic chamber (for better stability of liquid temperature controllable at any value between 15 ~35 ) Calibration method Gravimetric (static weighing) Standing-start-and-finish 2kg weighing system (liquid collection for 1~100 L/h) 100g weighing system (liquid collection for 0.02~1 L/h) Measurement point of temperature, T tm and pressure, p tm Test DUT meter Pulse signal Weighing vessel Weighing scale, m s Bypass to storage tank Pulse counter, Number of pulses, I p

11 Uncertainty Budget NATIONAL METROLOGY INSTITUTE OF JAPAN Uncertainty budget for 1 L/h ~ 100 L/h using 2 kg weighing system Uncertainty sources Light oil Kerosene Pulse count Estimation of liquid density Measurement of liquid mass Dead volume effect Relative combined standard uncertainty Uncertainty budget for 0.02 L/h ~ 1 L/h using 100 g weighing system Uncertainty sources Light oil Kerosene Pulse count Estimation of liquid density Measurement of liquid mass Dead volume effect Relative combined standard uncertainty

12 Uncertainty Budget (Kerosene) NATIONAL METROLOGY INSTITUTE OF JAPAN estimation of liquid density 0.030% pulse count 0.008% dead volume effect 0.001% measurement of liquid mass 0.003% estimation of liquid density 0.030% pulse count 0.008% measurement of liquid mass 0.018% dead volume effect 0.013% 2 kg weighing system (1 L/h~100 L/h) 100 g weighing system (0.02 L/h~1 L/h)

13 Uncertainty budget for 0.02 L/h ~ 1 L/h using 100 g weighing system estimation of liquid density 0.030% pulse count 0.008% measurement of liquid mass 0.018% dead volume effect 0.013% Dead volume effect The value is evaluated in relative to the smallest amount of liquid collection which is 10 g. 10 g of liquid collection is performed at 0.02 L/h and it takes about 40 minutes for one collection. Some measures to cut down the dead volume effect: Dead volume is made as small as possible. Temperature variation of liquid in the dead volume is controlled and stabilized (±0.025 o C in one hour duration).

14 Contributing factors to uncertainty of liquid mass measurement NATIONAL METROLOGY INSTITUTE OF JAPAN estimation of liquid density 0.030% pulse count 0.008% measurement of liquid mass 0.018% dead volume effect 0.013% leakage possibility 0.009% correction factor of weighing scale 0.003% reading of weighing scale 0.001% liquid evaporation 0.015% buoyancy effect correction 0.001%

15 Evaluation of liquid evaporation NATIONAL METROLOGY INSTITUTE OF JAPAN Conditions for liquid collection Liquid temperature:15 ~35. Smallest liquid collection: 10 g (about 40 minutes at 0.02 L/h ) Method of evaluation An evaluation method thought to give the closest value to the actual condition was performed. Collection 100 g discharge Measurement of evaporation for 60 minutes Results At 35, the evaporation rate is g per hour for kerosene and g per hour for light oil. In relative to 10 g of liquid collection, this produces (0.015 %) in terms of relative uncertainty.

16 Calibration and Measurement Capability Ultimate uncertainty of calibration (Calibration and measurement capability in italic, coverage factor: k=2) Flow range, Q / Weighing system Volumetric flowrate Mass flowrate 1 L/h Q 100 L/h (2 kg weighing system) % (0.064 %) % (0.020 %) 0.02 L/h Q < 1 L/h (100 g weighing system) % (0.078 %) % (0.050 %) Note: (ultimate standard uncertainty) 2 = (uncertainty due to facility) 2 + (uncertainty due to flowmeter) 2 Uncertainties due to flowmeter are as follows: Response of flowmeter to standing-start and finish Repeatability

17 Doihara s weighing and syringe system

18 Schematic diagram of the prototype test rig

19 Schematic diagram of the prototype test rig

20 Calibration procedure NATIONAL METROLOGY INSTITUTE OF JAPAN Calibration The syringe pump is calibrated using the gravimetric weighing tank.

21 Calibration procedure Then a DUT flowmeter is calibrated using the syringe pump. At a DUT calibration stage, the flow rig is able to make a closed pipe line system. NATIONAL METROLOGY INSTITUTE OF JAPAN Calibration

22 Standing method and flying method Standing method The syringe pump is calibrated with a standing start and stop method. What the weighing tank needs to calibrate is only the syringe pump (over pulses). Characteristic of the syringe pump is essentially linear to flow rate. Thus the standing method is not significant problem in this syringe pump calibration. Calibration

23 DUT flowmeter is calibrated at a constant flow rate with a flying start and finish method. NATIONAL METROLOGY INSTITUTE OF JAPAN If the number of pulses of the flowmeter would be small, the pulse counting error can be reduced using a double chronometry pulse interpolation technique in the flying method. Calibration Flying method

24 Relative deviation from nominal value [ % ] Calibration results of the syringe They all agree within ±0.02%. Influence of the standing method was enough small Flow rate [ L/h ] NATIONAL METROLOGY INSTITUTE OF JAPAN Light oil: Jan K SY P V SY SY ±0.02%. [Pulse/mL]

25 Evaporation effect and correction The weighing tank system has some evaporation controls to reduce evaporation. The weighing tank was surrounded by small liquid pools. The weighing tank has a lid with small holes to insert tubes. The changing of weight from an initial weight was measured in several conditions for one hour (Static). Condition Tank lid with Evaporation Evaporation rate small holes trap pool [ mg/h ] Static A 0.6 Static B 3 Static C 4 Static D 6

26 Relative deviation from nominal value [ % ] NATIONAL METROLOGY INSTITUTE OF JAPAN The syringe pulse factor was increased at low flow rate. Results with evaporation correction (Static A) were similar Without evaporation correction With evaporation correction (Static A) Flow rate [ L/h ]

27 Evaporation effect and correction The atmosphere inside the wind shield was ventilated when the working liquid in the weighing tank was discharged. The actual evaporation rate was measured in the calibration sequence. The measured evaporation rate (actual condition E) was between static A and static C. Condition Tank lid with Evaporation Evaporation rate small holes trap pool [ mg/h ] Static A 0.6 Static B 3 Static C 4 Static D 6 Actual E 2 4 Since actual evaporation rates were not constant, the evaporation effect was corrected at each calibration.

28 Relative deviation from nominal value [ % ] NATIONAL METROLOGY INSTITUTE OF JAPAN Corrected results (Actual E) showed almost constant value over full flow rate range Without evaporation correction With evaporation correction (Actual E) Flow rate [ L/h ]

29 Relative deviation from nominal value [ % ] NATIONAL METROLOGY INSTITUTE OF JAPAN Position dependency of syringe pulse factor The standing stop position was varied from the cylinder end position to 35 ml, keeping the calibration collection volume of 5 ml Standing stop position from cylinder end [ ml ] 0 5mL 5mL

30 Relative deviation from nominal value [ % ] NATIONAL METROLOGY INSTITUTE OF JAPAN Position dependency of syringe pulse factor Volume domain of the cylinder has been fixed. Standing stop position: 5m L, Collection volume: 10 ml mL collection (light oil) 10mL collection (lightl oil) 0 5mL 10mL Standing stop position from cylinder end [ ml ]

31 Relative deviation from nominal value [ % ] NATIONAL METROLOGY INSTITUTE OF JAPAN Calibration results of the syringe Difference depending on kinds of liquids has not been observed Light oil: Jan Flow rate [ L/h ] Industrial gasoline: Feb Industrial gasoline: Mar. 2015

32 Preliminary uncertainty budget Uncertainty source Standard uncertainty Combined standard uncertainty Expanded uncertainty (k=2) Syringe pulse count 0.002% Mass measurement 0.008% Density at syringe 0.010% Change of mass in dead volume 0.007% Random effect of syringe calibration 0.008% Calibration for syringe pulse factor 0.017% 0.034% Linearity and long term reproducibility of reference syringe pulse factor 0.017% Density at syringe 0.010% Change of mass in dead volume 0.015% Random effect of DUT calibration 0.005% Calibration for mass flow rate 0.030% 0.060% Density at DUT 0.016% Calibration for volume flow rate 0.034% 0.068%

33 Flow rate: Specification 60 L/h 0.02 L/h (1000 ml/min 0.3 ml/min). Working liquid (as the first step): Light oil (diesel oil), Industrial gasoline (Flash point 40 ) Calibration principle: Gravimetric weighing tank, Syringe pump. Temperature and pressure (as the first step): 20 (Stability ±0.08 ) MPa (at test section) Preliminary expanded uncertainty (k=2): % (for volumetric flow), % (for mass flow).

34 Summary Cheong s gravimetric system 0.02 L/h to 100 L/h Kerosene and light oil Liquid temperature: 15 to 35 Open for public calibration service Will be peer-reviewed in 2017 Doihara s gravimetric and syringe system 0.02 L/h to 60 L/h Applicable to volatile liquids Liquid temperature: 20 Flexible structure Used for research purpose

35 Thank you for your attention. NATIONAL METROLOGY INSTITUTE OF JAPAN

Research on the micro liquid flow standard system and inter-comparison

Research on the micro liquid flow standard system and inter-comparison Proposal of 2018 APMP TC Initiative Project Research on the micro liquid flow standard system and inter-comparison Type of Project: Comparison Technique January 1, 2019 - December 31, 2019 Seok Hwan LEE

More information

EI HMC Report Spring API Committee on Petroleum Measurement Standards Meeting

EI HMC Report Spring API Committee on Petroleum Measurement Standards Meeting EI HMC Report 2017 Spring API Committee on Petroleum Measurement Standards Meeting STAC EI Technical Committee Structure STAC Health Technical Committee Environmental Management Group Safety Management

More information

Calibration of Pipe Provers in the field

Calibration of Pipe Provers in the field Calibration of Pipe Provers in the field Erik Smits, Senior Metrologist Liquid Flow & Volume at VSL March 30 th 2016 All pictures have been removed for the pdf version! Content - Where and when are pipe

More information

AVL PLU 131U FLOW METER

AVL PLU 131U FLOW METER CONSUMPTION MEASUREMENT TECHNOLOGIES AVL PLU 131U FLOW METER AVL PLU131 UREA FLOW MEASUREMENT SENSORS Function Summary The AVL PLU 131U-020 Flow Sensor has been designed for transient flow rate measurement

More information

9900 Series Oval Flowmeter/Transmitter with SMART METER MANAGER

9900 Series Oval Flowmeter/Transmitter with SMART METER MANAGER Design Specifications DS-9900 Revision 04 9900 Series Oval Flowmeter/Transmitter with SMART METER MANAGER 2-Wire loop powered flowmeter/transmitter 316 Stainless steel measuring unit Automatic temperature

More information

LBTC Series Positive Displacement Rotary Vane Flow Meters

LBTC Series Positive Displacement Rotary Vane Flow Meters LBTC Series Positive Displacement Rotary Vane Flow Meters Summary LBTC Series positive displacement rotary vane flow meters are independently designed and manufactured by our company on the base of absorbing

More information

Traceability overview

Traceability overview Traceability overview Weighing scale and traceable calibration weights from ISO17025 accredited supplier Pipe prover calibration by VSL (NMI Netherlands) Turbine master meter - - - - - - - - - - - - -

More information

Flowmeters. The Satam flowmeter range is used for the custody transfer measurement of liquid hydrocarbons, liquid gas and liquid chemicals.

Flowmeters. The Satam flowmeter range is used for the custody transfer measurement of liquid hydrocarbons, liquid gas and liquid chemicals. Countum Group Metering Solutions Flowmeters The Satam range is used for the custody transfer measurement of liquid hydrocarbons, liquid gas and liquid chemicals. Consisting of four distinct families, the

More information

AVL FUEL BALANCE & FUEL TEMPERATURE CONTROL

AVL FUEL BALANCE & FUEL TEMPERATURE CONTROL CONSUMPTION MEASUREMENT AVL FUEL BALANCE & FUEL TEMPERATURE CONTROL Description The combination of AVL Fuel Balance and AVL Fuel Temperature Control is a high precise fuel consumption measurement and conditioning

More information

Intercomparison Between Trapil, France and NEL, UK Using Oil Flow Turbine Meters

Intercomparison Between Trapil, France and NEL, UK Using Oil Flow Turbine Meters Intercomparison Between Trapil, France and NEL, UK Using Oil Flow Turbine Meters A Report for NMSPU, DTI 151 Buckingham Palace Road London SW1W 9SS Project No: OI/3 (OSDC40300) Report No: 305/99 Date:

More information

Technical Datasheet. HM R Series. Turbine Flow Meters. with Female Inch Threads

Technical Datasheet. HM R Series. Turbine Flow Meters.   with Female Inch Threads www.kem-kueppers.com info@kem-kueppers.com Technical Datasheet HM R Series Turbine Flow Meters with Female Inch Threads Certified according to DIN EN ISO 9001 2 HM R Series Turbine Flow Meters Description

More information

Standard Calibration Certificate

Standard Calibration Certificate Mettler Toledo Service Business Unit Laboratory 1900 Polaris Parkway Columbus, OH 43240 1-800-METTLER Customer Standard Calibration Certificate Company: ABC Pharma Incorporated Address: 1432 East 9th Street

More information

Annex to the Accreditation Certificate D-PL according to DIN EN ISO/IEC 17025:2005

Annex to the Accreditation Certificate D-PL according to DIN EN ISO/IEC 17025:2005 Deutsche Akkreditierungsstelle GmbH Annex to the Accreditation Certificate D-PL-17435-01-00 according to DIN EN ISO/IEC 17025:2005 Period of validity: 30.09.2016 to 08.12.2018 Holder of certificate: k3works

More information

Standard Calibration Certificate

Standard Calibration Certificate Mettler Toledo Service Business Unit Laboratory 1900 Polaris Parkway Columbus, OH 43240 1-800-METTLER Customer Standard Calibration Certificate Company: Address: Allied Food Processing 3433 West High Street

More information

Coriolis Density Error Compensating for Ambient Temperature Effects

Coriolis Density Error Compensating for Ambient Temperature Effects Coriolis Density Error Compensating for Ambient Temperature Effects Presented by Gordon Lindsay Oil & Gas Focus Group December 2018 Contents Project aims and objectives Experiment Setup Phase 1 Exploratory

More information

Calibration, Fluid Property Effects Reproducibility and Long Term Stability. All in 20 minutes! Dr Gregor Brown Cameron

Calibration, Fluid Property Effects Reproducibility and Long Term Stability. All in 20 minutes! Dr Gregor Brown Cameron Calibration, Fluid Property Effects Reproducibility and Long Term Stability. All in 20 minutes! Dr Gregor Brown Cameron Agenda Calibration process Cameron calibration laboratory Reproducibility/calibration

More information

LIQUID MEASUREMENT STATION DESIGN Class No

LIQUID MEASUREMENT STATION DESIGN Class No LIQUID MEASUREMENT STATION DESIGN Class No. 2230.1 Michael Frey Systems Sales Manager Daniel Measurement & Control, Inc. 5650 Brittmoore Rd. Houston, Texas 77041 INTRODUCTION The industry continues to

More information

Field test for the comparison of LNG static and dynamic mass measurement methods

Field test for the comparison of LNG static and dynamic mass measurement methods 31 st International North Sea Flow Measurement Workshop 22-25 October 2013 Field test for the comparison of LNG static and dynamic mass measurement methods Tore Mortensen, Justervesenet Henning Kolbjørnsen,

More information

BAYLAN AUTOMATED TEST BENCH

BAYLAN AUTOMATED TEST BENCH BAYLAN AUTOMATED TEST BENCH At this test bench, the electromagnetic flow meters are used with the measurement method. Calibrations of flowmeters are made with the help of the water collection tank and

More information

PERFORMANCE OF THE NEW PRIMARY TORQUE STANDARD MACHINE OF INMETRO, BRAZIL

PERFORMANCE OF THE NEW PRIMARY TORQUE STANDARD MACHINE OF INMETRO, BRAZIL PERFORMANCE OF THE NEW PRIMARY TORQUE STANDARD MACHINE OF INMETRO, BRAZIL R. Oliveira 1, L. Cabral 1, U. Kolwinski 2, D. Schwind 2 1 INMETRO National Institute of Metrology, Xerem, Brazil 2 GTM Gassmann

More information

Legal Metrology Control of a Liquefied Petroleum Gas Transfer Standard and Field Practice

Legal Metrology Control of a Liquefied Petroleum Gas Transfer Standard and Field Practice Legal Metrology Control of a Liquefied Petroleum Gas Transfer Standard and Field Practice Kun-Chih Tsai, Win-Ti Lin, Jiunn-Haur Shaw, Cheng-Tsair Yang, Chun-Min Su Center for Measurement Standards, CMS/ITRI

More information

VITLAB Dispenser. Standard Operating Procedure (SOP)

VITLAB Dispenser. Standard Operating Procedure (SOP) VITLAB Dispenser Standard Operating Procedure (SOP) VITLAB GmbH Linus-Pauling-Str. 63762 Grossostheim Germany tel: +49 6026 97799-0 fax: +49 6026 97799-30 info@vitlab.com www.vitlab.com Table of contents.

More information

MIXPAC TM Technical Services Your partner for successful solutions. Sulzer Chemtech

MIXPAC TM Technical Services Your partner for successful solutions. Sulzer Chemtech MIXPAC TM Technical Services Your partner for successful solutions Sulzer Chemtech Turning ideas into products efficiently Technical Services to help you move forward The success of your single or multi-component

More information

Product Loss During Retail Motor Fuel Dispenser Inspection

Product Loss During Retail Motor Fuel Dispenser Inspection Product Loss During Retail Motor Fuel Dispenser Inspection By: Christian Lachance, P. Eng. Senior Engineer - ment Engineering and Laboratory Services ment Canada Date: Product Loss During Retail Motor

More information

Future Powertrain Conference 24 th February C 2016 HORIBA Ltd. All rights reserved.

Future Powertrain Conference 24 th February C 2016 HORIBA Ltd. All rights reserved. Recent and Future Developments In The Legislation and Measurement of Particle Number for Type Approval, In Service Conformity and Real Driving Emissions Future Powertrain Conference 24 th February 2016

More information

USV Ultra Shear Viscometer

USV Ultra Shear Viscometer USV Ultra Shear Viscometer A computer controlled instrument capable of fully automatic viscosity measurements at 10,000,000 reciprocal seconds Viscosity measurement background Accurate measurement of dynamic

More information

Guideline No.M-10(201510) M-10 Marine Diesel Engine Fuel Injection Pump

Guideline No.M-10(201510) M-10 Marine Diesel Engine Fuel Injection Pump Guideline No.M-10(201510) M-10 Marine Diesel Engine Fuel Injection Pump Issued date: 20 th October, 2015 China Classification Society Foreword This Guideline is a part of CCS Rules, which contains technical

More information

M-11 MARINE DIESEL ENGINE FUEL INJECTOR

M-11 MARINE DIESEL ENGINE FUEL INJECTOR Guideline No.M-11(201510) M-11 MARINE DIESEL ENGINE FUEL INJECTOR Issued date: 20 October 2015 China Classification Society Foreword This Guideline constitutes the CCS rules, and establishes the applicable

More information

Test Report Project Nr. 2013/020

Test Report Project Nr. 2013/020 PISTON Ltd. 1033 Budapest Szőlőkert u. 4./B HUNGARY Tel: Fax: http: Email: (361)-275-0033 (361)-275-0034 www.pistonmedical.com info@pistonmedical.com Test Report Project Nr. 2013/020 Name and address of

More information

RENOVATION OF NMIJ'S 5 MN HYDRAULIC AMPLIFICATION TYPE FORCE STANDARD MACHINE

RENOVATION OF NMIJ'S 5 MN HYDRAULIC AMPLIFICATION TYPE FORCE STANDARD MACHINE Measurement of Mass, Force and Torque (APMF 213) International Journal of Modern Physics: Conference Series Vol. 24 (213) 13611 (7 pages) The Authors DOI: 1.1142/S211945136112 RENOVATION OF NMIJ'S 5 MN

More information

Flow Research and Test Center. Creating resource technology for today and tomorrow.

Flow Research and Test Center. Creating resource technology for today and tomorrow. Flow Research and Test Center Creating resource technology for today and tomorrow. "The Measurement Solutions Flow Research and Test Center in Erie is certainly one of the gems of our corporation s technological

More information

FUELS. Product Specifications

FUELS. Product Specifications FUELS Product Specifications This Lesson Is Designed: To provide you with an understanding of specifications and terms used in the marketing of fuels. To show you how to use the information contained in

More information

产品技术规范. Product Name: Product Technical Specification SMARTSTONE. No.: CPFL0002 Version: A Issued date: Page 1 of 16

产品技术规范. Product Name: Product Technical Specification SMARTSTONE. No.: CPFL0002 Version: A Issued date: Page 1 of 16 Product Name: TMAP Part Number: CPF00002 Drawing Number: CPFJ0002 Page 1 of 16 CONTENTS 1. General Data 1.1. Description 1.2. Installation Guide 1.3. Maximum ratings of pressure sensor 1.4. Operating characteristics

More information

DaimlerChrysler Alternative Particulate Measurement page 1/8

DaimlerChrysler Alternative Particulate Measurement page 1/8 DaimlerChrysler Alternative Particulate Measurement page 1/8 Investigation of Alternative Methods to Determine Particulate Mass Emissions Dr. Oliver Mörsch Petra Sorsche DaimlerChrysler AG Background and

More information

Appendix A.1 Calculations of Engine Exhaust Gas Composition...9

Appendix A.1 Calculations of Engine Exhaust Gas Composition...9 Foreword...xi Acknowledgments...xiii Introduction... xv Chapter 1 Engine Emissions...1 1.1 Characteristics of Engine Exhaust Gas...1 1.1.1 Major Components of Engine Exhaust Gas...1 1.1.2 Units Used for

More information

CONTOIL VZFA/VZOA , versions for higher requirements / applications

CONTOIL VZFA/VZOA , versions for higher requirements / applications CONTOIL VZF/VZO 4...50, versions for higher requirements / applications For applications requiring an increased accuracy of ±0.5 % or better, such as: Measurement of L heating fuel or diesel in testing

More information

Vapor Pressure Measurement Of Gasoline, Crude Oil & LPG With New ASTM D5191 & D5188 Speed Test

Vapor Pressure Measurement Of Gasoline, Crude Oil & LPG With New ASTM D5191 & D5188 Speed Test Vapor Pressure Measurement Of Gasoline, Crude Oil & LPG With New ASTM D5191 & D5188 Speed Test Philipp Jordan eralytics TM GmbH www.eralytics.com Welcome to a new era of petroleum testing PERFORMANCE YOU

More information

AVL SHOT TO SHOT PLU 131 UREA FLOW MEASURING SYSTEM

AVL SHOT TO SHOT PLU 131 UREA FLOW MEASURING SYSTEM UREA INJECTION TESTING AVL SHOT TO SHOT PLU 131 UREA FLOW MEASURING SYSTEM Function Summary The AVL Shot To Shot (STS) PLU 131 UREA flow measuring system is based on the well-proven PLU 131 flow metering

More information

Chapter 4 ANALYTICAL WORK: COMBUSTION MODELING

Chapter 4 ANALYTICAL WORK: COMBUSTION MODELING a 4.3.4 Effect of various parameters on combustion in IC engines: Compression ratio: A higher compression ratio increases the pressure and temperature of the working mixture which reduce the initial preparation

More information

FLOWMETER (SMALL CAPACITY) METER SIZE 80, 81

FLOWMETER (SMALL CAPACITY) METER SIZE 80, 81 UF- FLOWMETER (SMALL CAPACITY) METER SIZE, GENERAL SPECIFICATION GS.No.GBL--E GENERAL The UF-II Series is a new positive displacement flowmeter developed specifically to serve as the heart in loading facilities

More information

White Paper.

White Paper. The Advantage of Real Atmospheric Distillation Complying with the ASTM D7345 Test Method in the Distillation Process Introduction / Background In the past, refiners enjoyed a constant supply of the same

More information

ADU 5. Automatic Distillation Unit

ADU 5. Automatic Distillation Unit ADU 5 Automatic Distillation Unit Atmospheric Distillation Distillation tests are used to characterize the volatility of petrochemical products. These volatility characteristics are a critical measurement

More information

MidFlow _ 135. Sliding Vane Meters DN (1-2 ) Model PT / Fuel Oil Flowmeters135. Product Bulletin

MidFlow _ 135. Sliding Vane Meters DN (1-2 ) Model PT / Fuel Oil Flowmeters135. Product Bulletin MidFlow Model PT / Fuel Oil Flowmeters135 Sliding Vane Meters DN 25-50 (1-2 ) _ 135 Product Bulletin WWW.VAF.NL 1 Introduction VAF Instruments MidFlow Model PT positive displacement sliding vane type liquid

More information

Technical Data. Series Oval Flowmeter. Model 9400

Technical Data. Series Oval Flowmeter. Model 9400 Technical Data Series Oval Flowmeter Model 9400 General The Brodie 9400 Series Oval Flowmeters, Models 9402 through 9459 (sizes 1/2-3 ), are mechanical, positive displacement meters designed for use in

More information

KEM Automotive Test Stands DESIGN PRODUCTION DISTRIBUTION

KEM Automotive Test Stands DESIGN PRODUCTION DISTRIBUTION www.kem-kueppers.com KEM Automotive Test Stands DESIGN PRODUCTION DISTRIBUTION 2 3 CONTENT KEM Küppers in Auto test stands... 04 Overview of applications... 06 KEM Küppers oil measurement applications...

More information

Dispensette. Testing Instructions (SOP) 1. Introduction. May 2009

Dispensette. Testing Instructions (SOP) 1. Introduction. May 2009 Dispensette Testing Instructions (SOP) May 2009 1. Introduction The standard ISO DIS 8655 describes both the design and the testing of the bottle-top dispenser. The following Testing Instructions describe

More information

Hybrid torque standard machine for 1 kn m developed in CENAM

Hybrid torque standard machine for 1 kn m developed in CENAM IMEKO 20 th TC3, 3 rd TC16 and 1 st TC22 International Conference Cultivating metrological knowledge 27 th to 30 th November, 2007. Merida, Mexico. Hybrid torque standard machine for 1 kn m developed in

More information

CEE 452/652. Week 6, Lecture 1 Mobile Sources. Dr. Dave DuBois Division of Atmospheric Sciences, Desert Research Institute

CEE 452/652. Week 6, Lecture 1 Mobile Sources. Dr. Dave DuBois Division of Atmospheric Sciences, Desert Research Institute CEE 452/652 Week 6, Lecture 1 Mobile Sources Dr. Dave DuBois Division of Atmospheric Sciences, Desert Research Institute Today s topics Read chapter 18 Review of urban atmospheric chemistry What are mobile

More information

Technical Datasheet. HM AC Series. Turbine Flow Meters. for High Pressure Applications up to 1,400 bar

Technical Datasheet. HM AC Series. Turbine Flow Meters.   for High Pressure Applications up to 1,400 bar www.kem-kueppers.com info@kem-kueppers.com Technical Datasheet HM AC Series Turbine Flow Meters for High Pressure Applications up to 1,400 bar Certified according to DIN EN ISO 9001 2 HM AC Series Turbine

More information

Determination of Volume Correction Factors for FAME and FAME / Mineral-diesel blends

Determination of Volume Correction Factors for FAME and FAME / Mineral-diesel blends H & D Fitzgerald Ltd. Determination of Volume Correction Factors for FAME and FAME / Mineral-diesel blends Carried out for the Energy Institute by H&D Fitzgerald Ltd. Cefn Du Tremeirchion St. Asaph LL17

More information

For the National Voluntary Laboratory Accreditation Program

For the National Voluntary Laboratory Accreditation Program SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 Florida Department of Agriculture Metrology Laboratory 3125 Conner Blvd. Lab #2 Tallahassee, FL 32399-1650 Ms. Amy Smith Phone: 850-921-1570 Fax: 850-921-1593

More information

Flow Research and Test Center

Flow Research and Test Center SALES BULLETIN Flow Research and Test Center CREATING RESOURCE TECHNOLOGY FOR TODAY AND TOMORROW The key to FMC Technologies success has been its ability to provide the latest technology in products and

More information

Annex to the Accreditation Certificate D-PL according to DIN EN ISO/IEC 17025:2005

Annex to the Accreditation Certificate D-PL according to DIN EN ISO/IEC 17025:2005 Deutsche Akkreditierungsstelle GmbH Annex to the Accreditation Certificate D-PL-17435-01-00 according to DIN EN ISO/IEC 17025:2005 Period of validity: 06.12.2017 to 08.12.2018 Holder of certificate: k3works

More information

ProFlow TO BE REALLY SURE. Vane Meters. Experience in flow measurement. Introduction

ProFlow TO BE REALLY SURE. Vane Meters. Experience in flow measurement. Introduction ProFlow N142 P R O D U C T B U L L E T I N Vane Meters Special versions This brochure comprises only VAF Instruments standard delivery program. Special flowmeter variants can be offered as tailor-made

More information

FEASIBILITY STYDY OF CHAIN DRIVE IN WATER HYDRAULIC ROTARY JOINT

FEASIBILITY STYDY OF CHAIN DRIVE IN WATER HYDRAULIC ROTARY JOINT FEASIBILITY STYDY OF CHAIN DRIVE IN WATER HYDRAULIC ROTARY JOINT Antti MAKELA, Jouni MATTILA, Mikko SIUKO, Matti VILENIUS Institute of Hydraulics and Automation, Tampere University of Technology P.O.Box

More information

Technical Datasheet. HM U Series. Turbine Flow Meters. for lubricating liquids

Technical Datasheet. HM U Series. Turbine Flow Meters.   for lubricating liquids www.kem-kueppers.com info@kem-kueppers.com Technical Datasheet HM U Series Turbine Flow Meters for lubricating liquids Certified according to DIN EN ISO 9001 2 HM U Series Turbine Flow Meters Description

More information

Standard Test Method for Determination of Vapor Pressure of Crude Oil: VPCR x (Expansion Method) 1

Standard Test Method for Determination of Vapor Pressure of Crude Oil: VPCR x (Expansion Method) 1 Designation: D6377 14 Standard Test Method for Determination of Vapor Pressure of Crude Oil: VPCR x (Expansion Method) 1 This standard is issued under the fixed designation D6377; the number immediately

More information

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009 Transferpette Testing Instructions (SOP) May 2009 1. Introduction The standards DIN EN ISO 8655 describe both the design and the testing of the piston operated pipette Transferpette. The following Testing

More information

VALVE TIMING DIAGRAM FOR SI ENGINE VALVE TIMING DIAGRAM FOR CI ENGINE

VALVE TIMING DIAGRAM FOR SI ENGINE VALVE TIMING DIAGRAM FOR CI ENGINE VALVE TIMING DIAGRAM FOR SI ENGINE VALVE TIMING DIAGRAM FOR CI ENGINE Page 1 of 13 EFFECT OF VALVE TIMING DIAGRAM ON VOLUMETRIC EFFICIENCY: Qu. 1:Why Inlet valve is closed after the Bottom Dead Centre

More information

Internal Combustion Engines

Internal Combustion Engines Thermochemistry & Fuels Lecture 4 1 Outline In this lecture we will discuss the properties and characteristics of diesel fuels: Cetane number and index Viscosity and cold behaviour Flash point Sulphur

More information

CHAPTER 3 EXPERIMENTAL SET-UP AND TECHNIQUES

CHAPTER 3 EXPERIMENTAL SET-UP AND TECHNIQUES 37 CHAPTER 3 EXPERIMENTAL SET-UP AND TECHNIQUES 3.1 EXPERIMENTAL SET-UP The schematic view of the experimental test set-up used in the present investigation is shown in Figure 3.1. A photographic view

More information

Mark Shamley Managing Director Elster Kent Metering South Africa

Mark Shamley Managing Director Elster Kent Metering South Africa Mark Shamley Managing Director Elster Kent Metering South Africa Reducing Non-Revenue Water by Selecting Good Quality Meters IWA Water Balance Billed Water Exported Authorized Consumption Billed Authorized

More information

Testing CNG. Retail Motor Fuel Dispensers

Testing CNG. Retail Motor Fuel Dispensers The s of Testing CNG Retail Motor Fuel Dispensers 1 Certain commercial entities, equipment, or materials may be identified in this presentation in order to describe a procedure or concept adequately. Such

More information

TSI PTI-Prototype for PN- Periodic Technical Inspection

TSI PTI-Prototype for PN- Periodic Technical Inspection TSI PTI-Prototype for PN- Periodic Technical Inspection Jürgen Spielvogel Senior Global Product Manager Particle Instruments 22 nd ETH-Conference on Combustion Generated Nanoparticles 18 21 June 2018 PN

More information

ANALYTICAL STUDY ON THE UNCERTAINTY OF LOAD CELLS CALIBRATED WITH DEADWEIGHT-FORCE-STANDARD MACHINE AND FORCE-COMPARATOR MACHINE

ANALYTICAL STUDY ON THE UNCERTAINTY OF LOAD CELLS CALIBRATED WITH DEADWEIGHT-FORCE-STANDARD MACHINE AND FORCE-COMPARATOR MACHINE ANALYTICAL STUDY ON THE UNCERTAINTY OF LOAD CELLS CALIBRATED WITH DEADWEIGHT-FORCE-STANDARD MACHINE AND FORCE-COMPARATOR MACHINE Sumet Heamawatanachai 1,*, Kittipong Chaemthet 1, Noppadon Sumyong 1, and

More information

Title: Terms and Conditions for the Approval of Coriolis Liquid Meters. Effective Date: Page: 1 of 9 Revision: 1

Title: Terms and Conditions for the Approval of Coriolis Liquid Meters. Effective Date: Page: 1 of 9 Revision: 1 Effective Date: 2012-12-01 Page: 1 of 9 Revision: 1 1.0 Application These terms and conditions apply to Coriolis metering assemblies and to metering installations used in trade 1. 2.0 Definitions In this

More information

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF TORQUE METROLOGY (Torque Generating Device Calibration)

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF TORQUE METROLOGY (Torque Generating Device Calibration) CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF TORQUE METROLOGY (Torque Generating Device Calibration) Approved By: Chief Executive Officer: Ron Josias Executive - Accreditation: Mpho Phaloane Author:

More information

LS Rotary Piston flow meter

LS Rotary Piston flow meter LS Rotary Piston flow meter Model LS rotary piston flow meter (hereafter for short called flow meter) is a kind of volumetric flow-measuring instrument, mainly used to measure volume flow of liquid. Its

More information

Study of Fuel Economy Standard and Testing Procedure for Motor Vehicles in Thailand

Study of Fuel Economy Standard and Testing Procedure for Motor Vehicles in Thailand Study of Fuel Economy Standard and Testing Procedure for Motor Vehicles in Thailand MR.WORAWUTH KOVONGPANICH TESTING MANAGER THAILAND AUTOMOTIVE INSTITUTE June 20 th, 2014 Overview Background Terminology

More information

TRACEABILITY OF THE CALIBRATION OF TEST CAR FOR ROLL BRAKE TESTER Aimo Pusa Raute-MIKES Finland, by Raute Precision Oy, Force and Mass Laboratory

TRACEABILITY OF THE CALIBRATION OF TEST CAR FOR ROLL BRAKE TESTER Aimo Pusa Raute-MIKES Finland, by Raute Precision Oy, Force and Mass Laboratory TRACEABILITY OF THE CALIBRATION OF TEST CAR FOR ROLL BRAKE TESTER Aimo Pusa Raute-MIKES Finland, by Raute Precision Oy, Force and Mass Laboratory Abstract The calibration of the roll brake tester for trucks

More information

Verifying the accuracy of involute gear measuring machines R.C. Frazer and J. Hu Design Unit, Stephenson Building, University ofnewcastle upon Tyne,

Verifying the accuracy of involute gear measuring machines R.C. Frazer and J. Hu Design Unit, Stephenson Building, University ofnewcastle upon Tyne, Verifying the accuracy of involute gear measuring machines R.C. Frazer and J. Hu Design Unit, Stephenson Building, University ofnewcastle upon Tyne, Abstract This paper describes the most common methods

More information

An Analysis of Alternative Fuels for Automotive Engines. Joey Dille

An Analysis of Alternative Fuels for Automotive Engines. Joey Dille An Analysis of Alternative Fuels for Automotive Engines Joey Dille 1 Problem Hundreds of millions of people use the car as their preferred method of transportation, but cars pollute the air and contribute

More information

INNOVATIVE RADIOTRACER TECHNIQUES APPLIED TO THE DEVELOPMENT OF ENGINES AND LUBRICANTS

INNOVATIVE RADIOTRACER TECHNIQUES APPLIED TO THE DEVELOPMENT OF ENGINES AND LUBRICANTS INNOVATIVE RADIOTRACER TECHNIQUES APPLIED TO THE DEVELOPMENT OF ENGINES AND LUBRICANTS Sytec M2A, 12th-13th October, 2015 COMPANY PROFILE DSi founded in 2000 18 employees including 9 highly qualified engineers

More information

National comparison on verification of fuel dispensers

National comparison on verification of fuel dispensers 16 th International Congress of Metrology, 06011 (2013) DOI: 10.1051/ metrology/201306011 C Owned by the authors, published by EDP Sciences, 2013 National comparison on verification of fuel dispensers

More information

Gauges, Sight Glasses and Vacuum Breakers

Gauges, Sight Glasses and Vacuum Breakers Gauges, Sight Glasses and Vacuum Breakers Gauges, Sight Glasses and Vacuum Breakers Gauges Pressure gauges Pressure gauges should be installed in at least the following situations: Upstream of a pressure

More information

SOOT SENSOR FOR EMISSION ONBOARD CONTROL SYSTEMS

SOOT SENSOR FOR EMISSION ONBOARD CONTROL SYSTEMS SOOT SENSOR FOR EMISSION ONBOARD CONTROL SYSTEMS Victor GHEORGHIU Prof. PhD ME HAW Hamburg University of Applied Sciences, Germany Return CONTENT Emissions Legislation Characteristic of PM-Measuring with

More information

COMPRESSIBLE FLOW ANALYSIS IN A CLUTCH PISTON CHAMBER

COMPRESSIBLE FLOW ANALYSIS IN A CLUTCH PISTON CHAMBER COMPRESSIBLE FLOW ANALYSIS IN A CLUTCH PISTON CHAMBER Masaru SHIMADA*, Hideharu YAMAMOTO* * Hardware System Development Department, R&D Division JATCO Ltd 7-1, Imaizumi, Fuji City, Shizuoka, 417-8585 Japan

More information

EU Projekt HySYS Fuel Cell Hybrid Vehicle System Component Development

EU Projekt HySYS Fuel Cell Hybrid Vehicle System Component Development EU Projekt HySYS Fuel Cell Hybrid Vehicle System Component Development Dr. Jörg Wind, Daimler AG ECPE - HOPE Symposium Automotive Power Electronics 7-8 October 2008, Sindelfingen FC Hybrid Vehicle System

More information

UNDERSTANDING CMM SPECIFICATIONS AND CALIBRATION METHODS

UNDERSTANDING CMM SPECIFICATIONS AND CALIBRATION METHODS UNDERSTANDING CMM SPECIFICATIONS AND CALIBRATION METHODS EDUCATION A TECHNICAL PRESENTATION FROM THE LEADING MANUFACTURER OF METROLOGY INSTRUMENTS Bulletin No. 2183 EDU-14003A-H Mitutoyo Institute of Metrology

More information

Workshop on Automotive Stack Design Options, Platform Concept, and Cost Targets

Workshop on Automotive Stack Design Options, Platform Concept, and Cost Targets AUTOSTACK Workshop Feb8 th 2011, Grenoble Workshop on Automotive Stack Design Options, Platform Concept, and Cost Targets F. Finsterwalder Daimler AG Content 1. Introduction 2. Summary of system requirements

More information

4. With a neat sketch explain in detail about the different types of fuel injection system used in SI engines. (May 2016)

4. With a neat sketch explain in detail about the different types of fuel injection system used in SI engines. (May 2016) SYED AMMAL ENGINEERING COLLEGE (Approved by the AICTE, New Delhi, Govt. of Tamilnadu and Affiliated to Anna University, Chennai) Established in 1998 - An ISO 9001:2000 Certified Institution Dr. E.M.Abdullah

More information

Pressure balance Differential pressure version Model CPB5600DP

Pressure balance Differential pressure version Model CPB5600DP Calibration technology Pressure balance Differential version Model CPB5600DP WIKA data sheet CT 31.56 for further approvals see page 7 Applications Primary standard for the calibration of differential

More information

Datasheet Rev.1 Feb 2014 High precision flowmeter

Datasheet Rev.1 Feb 2014 High precision flowmeter Datasheet Rev.1 Feb 2014 High precision flowmeter DeltaTMHP Warranty of high accuracy measurement Compliance with the requirements while limiting the measurement uncertainties to their minimum (ISO5167,

More information

Exercise 4-1. Flowmeters EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Rotameters. How do rotameter tubes work?

Exercise 4-1. Flowmeters EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Rotameters. How do rotameter tubes work? Exercise 4-1 Flowmeters EXERCISE OBJECTIVE Learn the basics of differential pressure flowmeters via the use of a Venturi tube and learn how to safely connect (and disconnect) a differential pressure flowmeter

More information

Numerical Simulation and Performance Analysis of Rotary Vane Compressors for Automobile Air Conditioner

Numerical Simulation and Performance Analysis of Rotary Vane Compressors for Automobile Air Conditioner Purdue University Purdue e-pubs International Compressor Engineering Conference School of Mechanical Engineering 24 Numerical Simulation and Performance Analysis of Rotary Vane Compressors for Automobile

More information

HTM Series Helical Turbine Meters

HTM Series Helical Turbine Meters HTM Series Helical Turbine Meters The HTM series of turbine meters is designed for high accuracy volume measurement of crude oils and other liquid hydrocarbons. M&T helical turbine meters are preferred

More information

Series 8000 Vortex Flow Meter

Series 8000 Vortex Flow Meter VFM Series 8000 Vortex Flow Meter Operating Principle The operation of the VFM 8000 is based on the Karman Vortex Street. Vortices are formed as the fluid flows around the shedder body which are alternately

More information

Report on Recent Activities: CMS/ITRI, Chinese Taipei (Taiwan)

Report on Recent Activities: CMS/ITRI, Chinese Taipei (Taiwan) Report on Recent Activities: CMS/ITRI, Chinese Taipei (Taiwan) APMP 2015 TCFF Meeting Chun-Min Su, Ph.D. CMS/ITRI 2015/11/2 @ Beijing, China 1 Content Improvements to the Standard Facilities International

More information

White Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information

White Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Abstract High Temperature Simulated Distillation (High Temp SIMDIS) is one of the most frequently used techniques to determine

More information

PETROL. Double Case PD Flowmeters. Trust in us Our Quality Counts. Technical Bulletin. PETROL Instruments S.r.l.

PETROL. Double Case PD Flowmeters. Trust in us Our Quality Counts. Technical Bulletin. PETROL Instruments S.r.l. Double Case PD Flowmeters Technical Bulletin The trusted specialist manufacturer of high precision instruments that relies upon the accurate measurement of fluid volumes Trust in us Our Quality Counts

More information

Test bed for hydrostatic bearing gap

Test bed for hydrostatic bearing gap Fluid Power 2013 Instructions for authors Test bed for hydrostatic bearing gap Subtitle Georg Mörwald Institute of Production Engineering, TU Graz Jörg Edler Institute of Production Engineering, TU Graz

More information

2.6. Air Flow Control Valve Type PRD

2.6. Air Flow Control Valve Type PRD 2.6. Air Flow Control Valve Type PRD Page 1/10 Air Flow Control Valve, Type PRD Page 2/10 Air Flow Control Valve Type PRD Description and Design The PRD remains the air valve of choice for critical heating,

More information

Certificate of Approval NMI 12/1/8

Certificate of Approval NMI 12/1/8 Bradfield Road, West Lindfield NSW 2070 Certificate of Approval Issued by the Chief Metrologist under Regulation 60 of the National Measurement Regulations 1999 This is to certify that an approval for

More information

Hybrid Nanopositioning Systems with Piezo Actuators

Hybrid Nanopositioning Systems with Piezo Actuators Hybrid Nanopositioning Systems with Piezo Actuators Long Travel Ranges, Heavy Loads, and Exact Positioning Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page

More information

Introduction of measurement technics regarding mass emissions and real time fuel consumption using direct exhaust gas flow meter

Introduction of measurement technics regarding mass emissions and real time fuel consumption using direct exhaust gas flow meter THAI Automotive Summit 2015 Introduction of measurement technics regarding mass emissions and real time fuel consumption using direct exhaust gas flow meter Masanobu Akita HORIBA, Ltd. 2015 HORIBA, Ltd.

More information

Notes on Soot Measurement of Diesel Engines Wolfgang Schindler Wolfgang Singer

Notes on Soot Measurement of Diesel Engines Wolfgang Schindler Wolfgang Singer Notes on Soot Measurement of Diesel Engines Wolfgang Schindler Wolfgang Singer AVL List GmbH, Graz, Austria Why bother about Soot Measurement? HC Oil cond. HC Fuel cond. Soot C Water Metal & Others Gravimetric

More information

Electro-Proportional Terms and Definitions

Electro-Proportional Terms and Definitions Electro-Proportional Terms and Definitions Valve Deadband The span of operation where there is no flow or pressure output for some specified range of command Hydraulic Valve Gain The characteristic relating

More information

ETHANOL BLEND FUEL PERFORMANCE ON EVAPORATIVE EMISSION OF MOTORCYCLE IN THAILAND

ETHANOL BLEND FUEL PERFORMANCE ON EVAPORATIVE EMISSION OF MOTORCYCLE IN THAILAND ETHANOL BLEND FUEL PERFORMANCE ON EVAPORATIVE EMISSION OF MOTORCYCLE IN THAILAND Thummarat Thummadetsak Padol Sukajit Somchai Siangsanorh PTT Research & Technology Institute PTT Public Company Limited

More information

Transferpette. Testing Instructions (SOP) 1. Introduction. October 2013

Transferpette. Testing Instructions (SOP) 1. Introduction. October 2013 Transferpette Testing Instructions (SOP) October 2013 1. Introduction The standards DIN EN ISO 8655 describe both the design and the testing of the piston operated pipette Transferpette. The following

More information

The purpose of this rule is to limit VOC emissions from the transfer of organic liquids.

The purpose of this rule is to limit VOC emissions from the transfer of organic liquids. RULE 4624 TRANSFER OF ORGANIC LIQUID (Adopted April 11, 1991; Amended September 19, 1991; Amended May 21, 1992; Amended December 17, 1992; Amended December 20, 2007) 1.0 Purpose The purpose of this rule

More information