Performance Characteristics of the Agilent 1260 Infinity II Bio-Inert LC
|
|
- Michael May
- 6 years ago
- Views:
Transcription
1 Performance haracteristics of the gilent 16 Infinity II io-inert L Technical Overview uthors Sonja Schneider and Uschi Schweikert gilent Technologies, Inc. Waldbronn, Germany Introduction The gilent 16 Infinity II io-inert L is the next generation of gilent io-inert L, specially designed for conditions used in biochromatography (high salt concentrations such as M Nal, up to 8 M urea, and high and low ph solvents such as. M NaOH or. M Hl) by working with a completely inert sample flowpath. ll capillaries and fittings throughout the multisampler, multicolumn thermostat, and detectors, are metal free, and biomolecules come in contact only with ceramics or PEEK. esign elements in the gilent 16 Infinity II io-inert Multisampler and gilent 16 Infinity II io-inert Multicolumn Thermostat (MT) provide the user with the highest flexibility. ased on the proven technology of the gilent 16 Infinity Series liquid chromatography platform, the 16 Infinity II io-inert L has equivalent performance specifications as the standard 16 Infinity io-inert Quaternary L 1. The 16 Infinity II io-inert L is compatible with standard methods, with a maximum pressure of 6 bar.
2 The 16 Infinity II io-inert L offers: Flow rate range up to 1 ml/min Maximum pressure of 6 bar for UHPL. Higher pressures allow support of either higher flow rates (more speed) or longer columns (more resolution). The gilent 16 Infinity II io-inert Multisampler is based on a proven flow-through design using a ceramic needle. Further elements are the sample hotel for various sample containers that hold up to eight vial or microtiter plates, the robotics and needle handling routine for shortest injection cycle times, and the multiwash function for lowest carryover. n advanced column capacity for up to six columns, with individual precolumn solvent heating in a single MT. The door provides superior usability with flexible flap positions at 9 or 18, and it can be removed completely for highest accessibility. The MT facilitates precise column thermostatting for maximum application flexibility. These performances are possible by Peltier cooling and heating with two independent temperature zones from 1 degrees below ambient (minimum 4 ) to 8. In addition, the MT can be equipped with a column tag reading option for automatic and trackable column management. gilent 16 Infinity II L iode rray etector () WR with bio inert standard flow cell Standard-bore UHPL and conventional applications run on the same system configuration. Revolutionary gilent InfinityLab Quick onnect UHPL column fittings for dead-volume-free fluidic connections that enable efficient, fast, and convenient column exchange. In addition, easy-to install precolumn gilent InfinityLab Quick-onnect heat exchangers are available for precolumn solvent heating (Figure 1). Experimental The 16 Infinity II io-inert L that was tested comprised: gilent 16 Infinity II io-inert Pump (G64) gilent 16 Infinity II io-inert Multisampler (G668) with sample cooler (Option #1) gilent 16 Infinity II Multicolumn Thermostat (G7116) with bio-inert heat exchanger (Option #19) gilent 16 Infinity II iode rray etector WR (G711) with bio inert flow cell (Option #8) Figure 1. gilent InfinityLab Quick onnect fittings and gilent InfinityLab Quick-onnect heat exchangers.
3 Pump performance retention time (RT) precision The most important parameter influencing retention time (RT) precision is pump performance. RT precision was tested with different gradient and isocratic conditions using 4.6 and.1 mm id columns. The relative standard deviation (RS) of RTs for conventional gradient runs was <.4 %RS, except for the first peak for n = 7 (Figure.) For fast gradients, with a run time of approximately 1 minute, the RS for RTs was <.1 %RS (Figure 3) for n = Peak I RS RT (%) 1. Uracil.113. Phenol Methyl paraben Ethyl paraben.8. Propyl paraben n,n-iethyl-m-toluamide utyl paraben. 8. Toluene Heptyl paraben Sample from Sigma-ldrich: Reversed Phase Test Mix, Order No.: U 1 1 ml (uracil, phenol, n,n-diethyl-m-toluamide, toluene) HPL Gradient System iagnostic Mix, Order No.: ml (phenol, methyl paraben, ethyl paraben, propyl paraben, butyl paraben, heptyl paraben, uracil) Sample preparation ilute each sample to ml with water/acetonitrile 1:1 Mix the two diluted samples 1:1 olumn gilent ZORX S 18, 4.6 mm 1 mm, µm Mobile phase ) Water, ) cetonitrile Gradient minutes %, 1 minutes 9 % Flow rate 1 ml/min Stop time 1 minutes Post time minutes Injection volume µl olumn temperature 3 4/4 nm; Ref 4/1 nm io-inert standard flow cell, 1 mm Peak width <. minutes (1 Hz) Figure. Precision of RTs for conventional gradient runs Peak I RS RT (%) 1. cetophenone.89. Propiophenone utyrophenone Valerophenone.. Hexanophenone Heptanophenone Octanophenone.9 8. enzophenone cetanilide Sample: RRL heckout sample (p/n ) cetophenone, propiophenone, butyrophenone, valerophenone, hexanophenone, heptanophenone, octanophenone, benzophenone, acetanilide olumn gilent Poroshell 1 E 18,.1 mm mm,.7 µm Mobile phase ) Water, ) cetonitrile Gradient minutes 3 % 1 minute 9 % Flow rate ml/min Stop time 1. minutes Post time 1 minute Injection volume 1 µl olumn temperature 4/1 nm; Ref 36/1 nm io-inert standard flow cell, 1 mm Peak width <.6 minutes (4 Hz) Figure 3. Precision of RTs for fast gradient runs. 3
4 Figure 4 shows conventional isocratic conditions with an RT precision of <.6 %RS for n = 7. omparison of an gilent 16 Infinity io-inert Quaternary L and an gilent 16 Infinity II io inert L step gradient Tracer experiments are frequently used to verify the solvent mixing ripple at different gradient mixtures to evaluate pump performance. Figure shows a step gradient from to 1 % in 1 % steps with caffeine as the tracer compound. n overlay of step gradients generated with the 16 Infinity io-inert Quaternary L and the 16 Infinity II io-inert L is shown. The performance of both systems is highly comparable regarding the mixing properties of the pump. 1,7 1,6 1, 1,4 1,3 1, 1,1 1, Peak I RS RT (%) RS rea (%) 1. imethylphtalate iethylphtalate iphenyl o-terphenyl Isocratic sample (p/n ) olumn gilent ZORX S 18, mm, 1.8 µm Mobile phase ) Water, ) cetonitrile Isocratic 7 % Flow rate 1. ml/min Stop time 9 minutes Injection volume µl, draw speed µl/min olumn temperature 4 4/4 nm; Ref 38/1 nm io-inert standard flow cell, 1 mm Peak width <. minutes, (1 Hz) Figure 4. RT precision for conventional isocratic runs gilent 16 Infinty io-inert Quaternary L, 149 bar gilent 16 Inffinity II io-inert L, 13 bar olumn Restriction capillary Mobile phase ) Water + % isopropanol ) Water + % isopropanol + 1 mg/l caffeine Step gradient From to 1% in 1% steps Flow rate 1 ml/min Stop time minutes olumn temperature 36 73/4 nm; Ref 38/1 nm; peak width: <.13 minutes, ( Hz) io-inert standard flow cell, 1 mm Figure. Overlay of step gradients. gilent 16 Infinity io-inert Quaternary L and an gilent 16 Infinity II io-inert L. 4
5 Low backpressure application protein separation by size exclusion chromatography (SE) at 1 bar Low backpressure applications, commonly used in protein analysis, were tested with the 16 Infinity II io inert L to prove pressure stability and precision of RT with the 6 bar system. High precision of RTs was obtained (RS of <.4 %, n = ). Figure 6 shows a separation of bovine serum albumin (S) from its dimer using an SE column with a low backpressure, resulting in a total system pressure of 1 bar. Injector performance area precision Precise injection is mandatory for good quantitative results in liquid chromatography. The 16 Infinity II io-inert Multisampler can inject precisely over an injection range of. to 1 µl. Figure 7 shows an example chromatogram for an injection volume of 1 µl. The RS was <.37 % for n = 7. The RS for an injection volume of. µl was <.4 % for n = S imer 13 ka S 66 ka olumn SE, 1 3 mm, 13 µm Mobile phase ) Phosphate buffered saline, ph = 7. Isocratic 1 % Flow rate. ml/min Stop time 6 minutes olumn temperature Room temperature 8/4 nm; Ref 36/1 nm; peak width: <.1 minutes, (. Hz) io-inert standard flow cell, 1 mm RS of RT in % S.43 S imer.16 Figure 6. Low backpressure application: separation of bovine serum albumin (S) at 1 bar Peak I RS RT (%) RS rea (%) 1. cetophenone Propiophenone utyrophenone Valerophenone Hexanophenone Heptanophenone Octanophenone enzophenone cetanilide Sample: RRL heckout sample (p/n ) cetophenone, propiophenone, butyrophenone, valerophenone, hexanophenone, heptanophenone, octanophenone, benzophenone, acetanilide olumn gilent Poroshell 1 E 18, 3 mm,.7 µm Mobile phase ) Water, ) cetonitrile Gradient minutes %, 8 minutes 8 % Flow rate 1. ml/min Stop time 8 minutes Post time 4 minutes Injection volume 1 µl olumn temperature 3 4/1 nm; Ref 4/1 nm io-inert standard flow cell, 1 mm Peak width <. minutes (1 Hz) Figure 7. rea precision for conventional gradient runs for 1 and. µl.
6 Figure 8 shows an example chromatogram for an injection volume of µl. The RS is <.14 % for n = 7. The injector settings are important for optimum precision of areas. If the highest precision is needed, the draw speed of the injector should be set to lower values, especially if large volumes or highly viscous samples are injected. It is important to avoid solvent evaporation out of the sample vials, and decomposition problems using a cooled autosampler. Injector performance carryover For the injection, the draw speed was set to µl/min, and an exterior needle wash for 1 seconds was used (Figure 9). No carryover was detected for the conditions used. fter a 1,-ng sample injection, unadulterated solvent was injected. Recommendations for carryover and cleaning procedures Flush port wash solvent must always be installed and used. The solvent chosen should be able to dissolve the sample compounds. It is highly recommended to reconnect the capillary connections from time to time to prevent cavities, which can lead to enhanced carryover. 1,7 1,6 1, 1,4 1,3 1, 1,1 1, Peak I RS RT (%) RS rea (%) 1. imethylphtalate iethylphtalate iphenyl o-terphenyl Isocratic sample (p/n ) olumn gilent ZORX S 18, mm, 1.8 µm Mobile phase ) Water, ) cetonitrile Isocratic 7 % Flow rate 1. ml/min Stop time 9 minutes Injection volume µl, draw speed µl/min olumn temperature 4 4/4 nm; Ref 38/1 nm io-inert standard flow cell, 1 mm Peak width <. minutes, (1 Hz) Figure 8. rea precision for isocratic runs with µl injection volume. 1, ,8 1. 1,7 1, ng 1,6 hlorhexidine 1. 1,.8 1,4.6 1,3.4 No carryover detected 1, 1,1. 1, Sample: hlorhexidine 1 mg/ml dissolved in.1 % TF in H O dd olumn gilent Poroshell 1 E 18, 4.6 mm mm Mobile phase ).1 % TF in in H O dd ).1 % TF in cetonitrile Isocratic 4 % Flow rate.6 ml/min Stop time. minutes Injection volume 1 µl, draw speed µl/min 1-second needle wash (exterior) olumn temperature 4/4 nm; Ref 36/1 nm io-inert standard flow cell, 1 mm Peak width <.1 minutes, ( Hz) Figure 9. No carryover was detected after injection of 1, ng chlorhexidine. 6
7 Injector performance injection volume linearity Injection volume linearity was tested using caffeine standards. ll injection volumes contained ng of caffeine. s a result, the injection volume was varied, but the injected amount always remained the same (Figure 1). The peak heights and areas should be the same for all injection volumes. The experiments showed that all areas were within 1.4 % RS over the complete injection volume range of.78 to 1 µl for n =. Performance of the gilent 16 Infinity II WR The 16 Infinity II Wide Wavelength Range (WR) with the bio inert standard flow cell is recommended for all common applications. etector performance linearity Linearity was tested using caffeine standards from 1. to 1,6 ng injected with n =. Good linearity was obtained at this concentration range. The coefficient of correlation was The response factors were within the % error range over an absorbance range of 1.7 to,36 (Figure 11). ll areas were within 1.4 %RS.78 µl 1.6 µl 3.13 µl 6. µl 1. µl µl µl 1 µl Sample: affeine 1 mg/ml, 7 times 1: diluted olumn gilent ZORX Eclipse Plus 18, mm, 1.8 µm Mobile phase ) Water, ) cetonitrile Isocratic 3 % Flow rate.8 ml/min Stop time. minutes Injection volume.78 to 1 µl, draw speed µl/min olumn temperature 4/4 nm; Ref 38/8 nm io-inert standard flow cell, 1 mm Peak width <.1 minutes, ( Hz) Figure 1. Injection volume linearity from.78 up to 1 µl; injected amount was always the same. Response factor (mount/rea) 4.E-1 3.E-1 Height = E-1.E-1.E-1 1.E-1 gilent 16 linearity correlation.9999 Height =, ,6 mount (ng injected) % Range verage response factor =.83 Sample: Enterprise Edition affeine Extended Standards Kit olumn gilent Poroshell 1 E 18, 3. mm, 1.9 µm Mobile phase ) Water, ) cetonitrile Isocratic 1 % Flow rate.8 ml/min Stop time. minutes Injection volume 1 µl, µl (1,6 ng) olumn temperature 3 73/1 nm; Ref 38/1 nm Universal bio-inert standard flow cell, 1 mm Peak width <.1 minutes, ( Hz) Figure 11. Linearity of the gilent 16 Infinity II WR with bio-inert flow cell. 7
8 Multicolumn thermostat comparison of standard and bio inert heat exchangers The column temperature regulation using the bio-inert heat exchanger was compared to the standard (stainless steel) heat exchanger in the MT. Samples were sensitive to temperature changes regarding RT (sulfa drugs: sulfadiazine, sulfathiazole, sulfamerazine, and sulfamethazine). With the change of solvent temperature, component (sulfathiazole) changed position within the order of eluted components during the isocratic run. t 1, component eluted at the third position, whereas at 3 and above it eluted at the second position using the standard heat exchanger (Figure 1) Peak I. Sulfadiazine. Sulfathiazole. Sulfamerazine. Sulfamethazine olumn gilent InfinityLab, E 1, mm,.7 µm Mobile phase ) Water +.1 % TF ) cetonitrile +.1 % TF Flow rate 1 ml/min Gradient minutes, 1 % 1 minutes, % 1 minutes, 1 % Stop time 1.1 minutes Post time minutes olumn temperature 6,, 4, 3,, 1 4/4 nm; Ref 38/1 nm bio-inert standard flow cell, 1 mm Peak width <.1 minutes, ( Hz) Figure 1. Thermosensitive sulfa drugs, with column temperature regulated through the standard heat exchanger. 8
9 Using the bio-inert heat exchanger, no major differences have been seen regarding RT changes (Figure 13). Table 1 shows the relative RT of sulfathiazole (component ) referred to sulfamerazine (component ) with n = 3. oth heat exchangers are highly comparable Peak I. Sulfadiazine. Sulfathiazole. Sulfamerazine. Sulfamethazine olumn gilent InfinityLab, E 1, mm,.7 µm Mobile phase ) Water +.1 % TF ) cetonitrile +.1 % TF Flow rate 1 ml/min Gradient minutes, 1 % 1 minutes, % 1 minutes, 1 % Stop time 1.1 minutes Post time minutes olumn temperature 6,, 4, 3,, 1 4/4 nm; Ref 38/1 nm bio-inert standard flow cell, 1 mm Peak width <.1 minutes, ( Hz) Figure 13. Thermosensitive sulfa drugs, with column temperature regulated through the bio-inert heat exchanger. Table 1. omparison of standard and bio-inert heat exchanger. Relative RT of sulfathiazole (component ) referred to sulfamerazine (component ) Standard heat exchanger io-inert heat exchanger
10 onclusions The performance of the gilent 16 Infinity II io-inert L meets the requirements of modern analytical liquid chromatography. The performance is highly comparable with the gilent 16 Infinity io-inert Quaternary L. The 16 Infinity II io-inert L is well suited for 4.6, 3., and.1 mm id columns, and can be used for conventional HPL and UHPL on columns packed with 1.8 µm particles. Precision of RTs for conventional L is typically <.4 % RS. The precision for peak areas is typically <.14 % for injection volumes > µl. No carryover was detected after 1, ng chlorhexidine injection. References 1. Schneider, S. Performance characteristics of the 16 Infinity io-inert Quaternary L System, gilent Technologies Technical Overview, publication number EN, 11.. gilent 16 Infinity io-inert Quaternary L, Features, Technical etails, pplications, and Specifications, gilent Technologies ata Sheet, publication number EN, 1. The gilent 16 Infinity II WR with a io-inert standard flow cell is recommended for all common applications. It is especially well suited for bio analytical applications, for example, protein analysis using ion exchange chromatography (IEX) or size exclusion/gel filtration (SE). Low pressure applications around 1 bar, commonly used in protein analysis and analytical scale preparative L, showed high RT precision. This information is subject to change without notice. gilent Technologies, Inc., 17 Published in the US, pril 1, EN
The Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over
The Agilent 12 Series high performance autosampler SL: Area precision, injection volume linearity, imum accessible volume, carry-over Technical Note 8 6 4 2.5 1 Introduction A main performance criterion
More informationTechnical Overview. Introduction
Performance characteristics of the Agilent 129 Infinity Binary Pump More resolution and speed for conventional, superficially porous and sub-2 µm column packing material Technical Overview Introduction
More informationAgilent InfinityLab LC Purification Solutions SELECTION GUIDE
gilent InfinityLab LC urification Solutions SELECTION GUIDE GILENT INFINITYLB LC URIFICTION SOLUTIONS EFFICIENT LC URIFICTION SOLUTIONS FOR NLYTICL TO RERTIVE SCLE gilent InfinityLab LC urification Solutions
More informationImpurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution
Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 129 Infinity II HDR-DAD Impurity Analyzer Solution Application ote Small Molecule Pharmaceuticals Author Sonja Schneider Agilent Technologies,
More informationTECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure
TECHNICAL REPORT Modifying Agilent 1100 HPLC Systems to Achieve UHPLC-like performance with HALO Fused-Core Columns Agilent 1100 HPLC system With a few modifications, an Agilent 1100 HPLC can produce UHPLC-like
More informationAgilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve
Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve Technical Note This Technical Note describes the purpose, installation, and configuration and use of a diverter valve
More informationApplication Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany
Determination of Aromatic Hydrocarbons in Petroleum Middle Distillates with the Agilent Infinity Binary HPLC System with RID Detection According to IP9()/ASTM D9 Application Note Energy & Chemicals - Petrochemicals
More informationBoosting Trace Detection Performance with the Vanquish Diode Array Detector and High-Sensitivity LightPipe Flow Cell
Boosting Trace Detection Performance with the Vanquish Diode Array Detector and High-Sensitivity LightPipe Flow Cell Alexandra Manka and Holger Franz Thermo Fisher Scientific, Germering, Germany Technical
More informationS 1125 HPLC Pump System
Routine S 170 EXPERT HPLC HPLC System SYSTEMS 500 The Sykam ROUTINE HPLC System S 500 is intended for all routine analysis. The system configuration is highly variable and several upgrade options make
More informationOnline sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve
Online sample cleanup on the Agilent 129 Infinity LC using a built in 2-position/6-port valve Analysis of Sudan red compounds in paprika powder Application Note Food Authors Angelika Gratzfeld-Huesgen
More informationAgilent 1290 Infinity II Ultra Low Dispersion Kit
Agilent 1290 Infinity II Ultra Low Dispersion Kit Technical Note 1290 Infinity II Ultra Low Dispersion Kit - Technical Note In this note we describe how to install the Agilent 1290 Infinity II Ultra Low
More informationAgilent G5632 Valve Kit Instructions
Agilent G5632 Valve Kit Instructions Agilent G5632 Valve Kit Instructions Contents Typical Applications 2 Alternating Column Regeneration (2 position/10 port valves only) 2 Dual column selection 3 Sample
More informationDetection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector
Detection of Sulfur Compounds in Natural Gas According to ASTM D554 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector Application Note Author Rebecca Veeneman Abstract Sulfur compounds in natural
More informationDetermination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC
Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC Application Note Energy and Fuels Authors Detlef Wilhelm Anatox GmbH & Co. KG Fürstenwalde, Germany Udo Huber Agilent
More informationGas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593
Application Note Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Authors Kelly Beard and James McCurry Agilent Technologies, Inc. Abstract An Agilent
More informationHPLC SYSTEM S 500 SERIES
HPLC SYSTEM S 500 SERIES The Sykam HPLC System S 500 Series is intended for all routine analysis as well as for the ambitious analyst. The system configuration is highly variable and several upgrade options
More informationHPLC System S 600 Series
HPLC System S 600 Series The Sykam HPLC System S 600 Series is intended for the ambitious analyst. The system configuration is highly variable and several upgrade options make this system suitable for
More informationApplication Note. Authors. Abstract. Energy & Chemicals
Determination of Aromatic Content in Diesel Fuel According to ASTM D5186 Enhancing the Agilent 126 Infi nity Analytical SFC System with a Flame Ionization Detector Application Note Energy & Chemicals Authors
More informationNEW HPLC - SYSTEM LINE
NEW HPLC - SYSTEM LINE S c h a m b e c k S F D G m b H Drieschweg 13A D-53604 Bad Honnef/Germany Phone: +49 2224 9239-0 E-Mail: info@schambeck-sfd.com Internet: www.schambeck-sfd.com S 9425 HPLC solvent
More informationHigh Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell
High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell Application Note Consumer Products Authors Gerd Vanhoenacker,
More informationACQUITY UPLC I-Class System with 2D Technology
ACQUITY UPLC I-Class System with 2D Technology The Waters ACQUITY UPLC I-Class System with 2D Technology allows chemists to increase sensitivity and selectivity, eliminate unwanted interferences, characterize
More informationDetermination of Sudan Dyes I IV in Curry Paste
Determination of Sudan Dyes I IV in Curry Paste Suparerk Tukkeeree and Jeffrey Rohrer 2 Thermo Fisher Scientific, Bangkok, Thailand; 2 Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 23 Key
More informationC196-E081C. Ultra High Performance Liquid Chromatograph Nexera X2. Specifications
C196-E081C Ultra High Performance Liquid Chromatograph Nexera X2 Specifications Specifications System Controllers CBM-20A / 20Alite CBM-20A (228-45012-XX) CBM-20Alite (228-45011-XX) Connectable units Solvent
More information[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS
MS Identification of Trace level Impurities from a Non-MS Compatible Mobile Phase Using ACQUITY UPLC System with 2D Technology by Heart-cutting and Online Sample Concentration Bronsky Gopinadh, Dilshad
More informationGAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS
GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS FLASH VAPORISATION INJECTION Split Splitless On-Column COOL INJECTION Large Volume Injection (LVI) On-Column On-Column-SVE (with solvent vapour
More informationPATROL UPLC Process Analysis System for Production
PATROL UPLC Process Analysis System for Production The PATROL UPLC Process Analysis System for Production is an integrated system solution designed and engineered to perform online and atline reaction
More informationLiquid Chromatography
SCINTOMICS Liquid Chromatography Always the right system for your needs. SCINTOMICS: Always the right system for your needs Advanced Research Daily QC Routine and beyond High Performance analytical/semi-preparative
More informationHigh-throughput protein aggregate analysis of monoclonal antibodies using a novel dual-channel UHPLC instrument
APPLICATION NOTE 72598 High-throughput protein aggregate analysis of monoclonal antibodies using a novel dual-channel UHPLC instrument Authors Nicola McGillicuddy, 1 Amy Farrell, 1 Sara Carillo, 1 Martin
More informationHIGH PRESSURE SWITCHING VALVES
HIGH PRESSURE SWITCHING VALVES Rheodyne offers high pressure switching valves to simplify procedures and improve the speed, resolution, and sensitivity of HPLC analysis. The switching valves are available
More informationAgilent InfinityLab Quick Change Valve G4237A
Agilent InfinityLab Quick Change Valve G4237A Instructions Agilent G4237A Valve Kit Instructions This technical note describes the installation and application of the Agilent InfinityLab Quick Change 4
More informationAgilent 1290 Infinity Autosampler
Agilent 1290 Infinity Autosampler Installation of the Large Volume Injection Kit General Information This note describes how to install the large volume injection kit into an Agilent 1290 Infinity Autosampler.
More informationAgilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011
Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc.
More informationThermo Scientific Dionex UltiMate 3000 Thermostatted Column Compartment
chromatography Thermo Scientific Dionex UltiMate 3000 Thermostatted Column Compartment Product Specifications Thermo Scientific Dionex UltiMate 3000 products are UHPLC compatible by design, establishing
More informationHigh Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph
Application Note Petrochemicas High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Author James D. McCurry, Ph.D. Agilent Technologies, Inc. Abstract An Agilent
More informationAgilent InfinityLab Quick Change Valves G4231A/C and G4232C/D
Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Instructions Agilent Valve Kit G4231A/C and G4232C/D - Instructions This technical note describes the installation and application of the Agilent
More informationSimplify your search for. You will find both PEEK-lined stainless steel hardware as well as standard ss column hardware at the below stockingstations.
Simplify your search for COLUMN HARDWARE Discover how to maximize results with Isolation Technologies column hardware and accessories that are specifically engineered to meet critical UHPLC, HPLC, and
More informationAlliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE
Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE AN HPLC THAT WORKS AS HARD AS YOU DO Confidence in your results comes with high performance, day-to-day reliability, and overall system robustness.
More informationThermo Scientific Dionex UltiMate 3000 Pumps
Chromatography Thermo Scientific Dionex UltiMate 3000 Pumps The Benchmark in LC Solvent Delivery Product Specifications Thermo Scientific Dionex UltiMate 3000 products are UHPLC compatible by design, establishing
More informationSTABILITY AND RELIABILITY: NEW APPROACHES IN PREPARATIVE HPLC COLUMN DESIGN
STABILITY AND RELIABILITY: NEW APPROACHES IN PREPARATIVE HPLC COLUMN DESIGN Fang Xia, Jie Y. Cavanaugh, Darcy Shave, Gary Izzo, Michael Savaria, Thomas Grady, Markus Wanninger, Donald Ziniti, Brad Francis,
More informationFast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC
Fast GC SPEEDY GC Slide 4 Optimization Goals Primary: Minimize time for the separation of a given number of peaks Secondary: Maximize the number of peaks separated by a given column Slide 5 HOW CAN THESE
More informationIncreased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel
Application Note Energy and Chemicals Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel Applying the Energy Institute Method IP 8 with an Agilent J&W
More informationVERITY Purification Systems Build the perfect system to fit your research
VERITY Purification Systems Build the perfect system to fit your research SPEC SHEET PURIFICATION VERITY PURIFICATION SYSTEMS WITH VERITY 3011 ISOCRATIC PUMP MAKE YOUR LAB LIFE EASIER From automated GPC
More informationAgilent Multimode Inlet for Gas Chromatography
Agilent Multimode Inlet for Gas Chromatography Technical Note Agilent Multimode Inlet for the 7890A GC Designed to give you ease of use and maximum flexibility, the Agilent Multimode Inlet does everything
More informationACQUITY UPLC I-Class System Specifications
ACQUITY UPLC I-Class System s Revision A Copyright Waters Corporation 2011 All rights reserved Copyright notice 2011 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND IN IRELAND. ALL RIGHTS
More informationAgilent G4239C Valve Kit - Instructions
Agilent G4239C Valve Kit - Instructions Agilent G4239C Valve Kit - Instructions Technical Information about the Agilent Valve Kit G4239C. Contents Typical Applications 2 Multicolumn Selection 2 Method
More informationMeeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems
Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems Application Note Hydrocarbons Authors Michael Woodman Agilent Technologies, Inc. Chemical Analysis Solutions 2850
More informationYour Essential Resource for INFINITYLAB LC SUPPLIES
Your Essential Resource for INFINITYLAB LC SUPPLIES ACROSS THE LAB, AROUND THE WORLD, WITH YOU EVERY STEP OF THE WAY. Real stories from the lab. WHO WE ARE. WHAT WE DO. Whatever your lab needs, Agilent
More informationCombustion ion chromatography with a Dionex Integrion HPIC System
TECHNICAL NOTE Combustion ion chromatography with a Dionex Integrion HPIC System Authors Daniel Khor,1 Jeff Rohrer,1 Adelon Austin,2 Mark Manahan2 1 Thermo Fisher Scientific, Sunnyvale, CA, USA; 2COSA
More informationAgilent 1260 Infinity Manual Injector
Agilent 1260 Infinity Manual Injector User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2006-2011, 2012 No part of this manual may be reproduced in any form or by any means (including
More informationAgilent 1200 Series High Performance Autosamplers & Micro Well Plate Autosampler
Agilent 1200 Series High Performance Autosamplers & Micro Well Plate Autosampler User Manual 1200 Series Autosamplers User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2006, 2008 No part
More informationAgilent Valve Kit G4231A/B and G4232A/B/C
Agilent Valve Kit G4231A/B and G4232A/B/C Instructions Contents Typical Applications 2 Dual column selection (G4231A/B or G4232A/B/C) 2 Sample enrichment and sample cleanup (G4231A/B or G4232A/B/C) 2 Alternating
More informationAgilent 1260 Infinity Bio-inert High Performance Autosampler
Agilent 1260 Infinity Bio-inert High Performance Autosampler User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2011-2012 No part of this manual may be reproduced in any form or by any
More informationSetting up SilFlow for BackFlush in your GC
Setting up SilFlow for BackFlush in your GC What is backflush and why use it? The BackFlush system eliminates the need to bake heavy sample fractions off the capillary column. Oils, tars and other semivolatile
More informationUsing a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application
Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D759 Application Petrochemical Author James D. McCurry Agilent Technologies 285 Centerville
More informationChromatography. Liquid. HPLC Accessories SupelPRO Valves SUPELCO
8 SupelPRO Valves Order:.800..00 Technical Service:.800.9.0 Web: www.sigma-aldrich.com/supelco SupelPRO Automated Fluidics Instruments Supelco s SupelPRO series are precision, electronically-controlled,
More informationAgilent 1290 Infinity Multisampler
Agilent 1290 Infinity Multisampler User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2014 No part of this manual may be reproduced in any form or by any means (including electronic storage
More informationFast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application
Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application Petroleum Authors ChunXiao Wang Agilent Technologies (Shanghai) Co.,Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai
More informationAgilent 7693A Automated Liquid Sampler
Agilent 7693A Automated Liquid Sampler Specifications Overview The Agilent 7693A is a state-of-the-art sample handling and injection system that provides the highest levels of precision and reliability
More informationAgilent Online-SPE Direct Inject Kit G4744A
Agilent Online-SPE Direct Inject Kit G4744A Installation Guide Agilent Online-SPE Direct Inject Kit G4744A - Installation Guide Contents Application 2 Delivery Checklist 4 Online SPE Direct Inject Kit
More informationAgilent InfinityLab LC Series Multi-wash Upgrade Kit
Agilent InfinityLab LC Series Multi-wash Upgrade Kit General Information Installation Note The multi-wash upgrade kit is designed for low carry over and can be installed in any Agilent InfinityLab LC Series
More informationApplication Note. Abstract. Authors. Environmental Analysis
High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID using the Agilent Low Thermal Mass (LTM II) System Application Note Environmental Analysis Authors Frank David and Karine Jacq Research
More informationmono-layer ACR High purity Reinforced 1-10 Excellent Fair mono-layer DD High purity Reinforced Excellent Excellent
Extended ph range from 1-10 provide longer lifetimes with harsh mobile phases Novel bonding reduces residual silanols which cause peak tailing Polymer coated capsule type silica Excellent selectivity for
More informationMET-Biodiesel Capillary GC Columns
MET-Biodiesel Capillary GC Columns Product Specifications Product Features & Benefits Chromatograms FAQs Related Products Updated: February 2, 2009 Product Specifications 2 Product Specifications What
More informationAgilent InfinityLab LC Series 1260 Infinity II Fraction Collectors
Agilent InfinityLab LC Series 1260 Infinity II Fraction Collectors Inlet / Waste Tubings Replacement This technical note describes the installation of the inlet/waste tubing kit to a 1260 Infinity II Fraction
More informationHigh-Temperature Simulated Distillation System Based on the 6890N GC Application
High-Temperature Simulated Distillation System Based on the 6890N GC Application Petroleum Authors ChunXiao Wang Agilent Technologies (Shanghai) Co., Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai
More informationDual-Needle Infinity II Upgrade Kit
Dual-Needle Infinity II Upgrade Kit Installation Note General Information The Dual-needle Upgrade Kit is designed for very fast injection for high throughput and unmatched flexibility to obtain the best
More informationHigh Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System
High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System Application Note Authors Frank David Research Institute for Chromatography, Pres.
More informationAgilent 1200 Series Manual Injector
Agilent 1200 Series Manual Injector User Manual 1200 Series MI User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2006, 2008 No part of this manual may be reproduced in any form or by
More informationDual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector
Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector Application Note Hydrocarbon Processing Authors ChunXiao Wang Agilent Technologies
More informationTechnical Procedure for Gas Chromatography (GC-FID)
Technical Procedure for Gas Chromatography (GC-FID) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph (GC- FID). 2.0 Scope This procedure applies to all
More informationMethod Development for Capillary GC Systems. Slide 1
Method Development for Capillary GC Systems Slide 1 AREAS TO OPTIMIZE Injector Carrier gas Column temperature Slide 2 COMMON INJECTOR MODES Vaporization Injection Modes Megabore Direct Split Splitless
More informationFausto Munari e Andrea Cadoppi ThermoFisher - Italy
The world leader in serving science Ultra Fast GC Determination of Total Hydrocarbons (C7-C40) and BTEX in Water and Soils through Direct Resistively Heated capillary columns and Robotic Autosampler. Fausto
More informationApplication Note. Author. Introduction. Energy and Fuels
Analysis of Free and Total Glycerol in B-100 Biodiesel Methyl Esters Using Agilent Select Biodiesel for Glycerides Application Note Energy and Fuels Author John Oostdijk Agilent Technologies, Inc. Introduction
More informationGC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585
GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc. 850 Centerville
More informationAppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS
AppNote 6/26 Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) Andreas Hoffmann GERSTEL GmbH & Co.KG, Eberhard-Gerstel-Platz 1,
More informationOptimized Method for Analysis of Commercial and Prepared Biodiesel using UltraPerformance Convergence Chromatography (UPC 2 )
Optimized Method for Analysis of Commercial and Prepared Biodiesel using UltraPerformance Convergence Chromatography (UPC 2 ) Mehdi Ashraf-Khorassani, 1 Giorgis Isaac, 2 and Larry T. Taylor 1 1 Department
More informationAgilent InfinityLab Quick Change Valve G4235A Instructions
Agilent InfinityLab Quick Change Valve G4235A Instructions Contents Typical Application 2 Delivery Checklist 3 Specifications 4 Installing the Valve Heads 5 Agilent Technologies Typical Application This
More informationSTI OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY
OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY Extended range HPLC has been investigated for many years; however, progress was delayed because of
More informationDetection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology
Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology Electronic Noses An electronic nose produces a recognizable response based
More informationHitachi High-Speed Amino Acid Analyzer
Hitachi High-Speed Amino Acid Analyzer Hitachi's Amino Acid Analyzer, Model L-8900, is the culmination of the technology and expertise earned from our extensive experience in manufacturing amino acid analyzers
More informationSpark Holland introduces ALIAS for ÄKTA pure All-round, state-of-the-art
BETTER SAMPLE CARE Spark Holland introduces ALIAS for ÄKTA pure All-round, state-of-the-art autosampler for ÄKTA pure Spark Holland introduces ALIAS for ÄKTA pure All-round, state-of-the-art autosampler
More informationÄKTA system Training Guide. Purifier 100 / 10. amersham bio science
ÄKTA system Training Guide Purifier 00 / 0 amersham bioscience amersham bioscience amersham bioscience amersham bio science ÄKTApurifier Buffers Injector Box Monitor Conductivity ph On-line Filter UV Detector
More informationAchieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)
Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Brian Fitz and Bill Wilson Agilent Technologies, Inc. 285 Centerville Road Wilmington,
More informationPhase Distribution of Ethanol, and Water in Ethyl Esters at K and K
Phase Distribution of Ethanol, and Water in Ethyl Esters at 298.15 K and 333.15 K Luis A. Follegatti Romero, F. R. M. Batista, M. Lanza, E.A.C. Batista, and Antonio J.A. Meirelles a ExTrAE Laboratory of
More informationAgilent 1100 Series Capillary LC System. System Manual
Agilent 1100 Series Capillary LC System System Manual s1 Notices Agilent Technologies, Inc. 2002 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval
More informationAnalyte Focusing in Elevated Temperature HPLC
Analyte Focusing in Elevated Temperature HPLC Jody Clark Brian Jones, Stephanie J. Marin, Dale Felix Selerity Technologies, Inc. Salt Lake City, UT 84104 www.selerity.com Introduction HPLC at extreme temperatures
More informationPRODUCT MANUAL. CARBONATE REMOVAL DEVICE 200 CRD 200 (4 mm) (P/N ) CRD 200 (2 mm) (P/N )
CRD 200 Product Manual Page 1 of 18 PRODUCT MANUAL FOR CARBONATE REMOVAL DEVICE 200 CRD 200 (4 mm) (P/N 062983) CRD 200 (2 mm) (P/N 062986) Dionex Corporation, 2006 Document No. 065068 Revision 04 November
More informationAgilent 7693A Automated Liquid Sampler
IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7693A Automated Liquid Sampler
More informationUsing the PSD for Backflushing on the Agilent 8890 GC System
Application Note Petrochemicals Using the PSD for Backflushing on the Agilent 889 GC System Author Brian Fitz Agilent Technologies, Inc. Wilmington, DE, USA. Abstract An Agilent 889 series GC equipped
More informationDetermination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105
Application Note: 10215 Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105 Fausto Munari, Daniela Cavagnino, Andrea Cadoppi, Thermo Fisher Scientific, Milan,
More informationAgilent 1260 Infinity Standard Autosampler
Agilent 1260 Infinity Standard Autosampler User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2007-2012, 2013 No part of this manual may be reproduced in any form or by any means (including
More informationPractical Steps in GC Troubleshooting
Practical Steps in GC Troubleshooting Techniques, Tips, and Tricks Mark Sinnott Application Engineer GC Columns & Supplies Page 1 Everything was just fine and then this happened! How do I go about TROUBLESHOOTING?
More informationAchieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)
Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Bill Wilson and Chin-Kai Meng Agilent Technologies, Inc. 2850 Centerville Road Wilmington,
More informationSystem Manual. Agilent 1260 Infinity Capillary LC System
System Manual Agilent 1260 Infinity Capillary LC System Notices Agilent Technologies, Inc. 2010, 2012 No part of this manual may be reproduced in any form or by any means (including electronic storage
More informationASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column
ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column Cory S. Fix, Director of Application Development cory.fix@vgcchromatography.com Willie Steinecker,
More informationDionex CRD Carbonate Removal Device
Thermo Scientific Dionex CRD Carbonate Removal Device Product Manual P/N: June 2013 Part of Thermo Fisher Scientific Product Manual for Dionex CRD Carbonate Removal Device 180 CRD 180, 0.4 mm (P/N 079960)
More informationAnalysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES
Analysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES Application note Petrochemical Author Neli Drvodelic Agilent Technologies Melbourne, Australia Introduction
More informationAnalysis of Petroleum Fractions by ASTM D2887
Analysis of Petroleum Fractions by ASTM D2887 Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 2582 Key Words Simulated distillation, D2887, TRACE TR-SimDist Abstract ASTM
More informationNeueste LC Entwicklungen
Neueste LC Entwicklungen 2D-LC, High Dynamic Range Detektion, quaternäre Pumpe mit 1200 bar Sebastian Krahe Produktspezialist LC Agilent Technologies, Deutschland/Österreich 2. Oktober 2012 Hamburg 1 Agilent
More informationStability Axial Compression Technology Application Ordering Information
42 YMC-Actus YMC-Actus 43 YMC-Actus Contents Stability... 44 Axial Compression Technology... 45 Application...46-47 Ordering Information...48-49 Introduction Fast semi-preparative chromatography Semi-preparative
More information