1 Determination of Aromatic Hydrocarbons in Petroleum Middle Distillates with the Agilent Infinity Binary HPLC System with RID Detection According to IP9()/ASTM D9 Application Note Energy & Chemicals - Petrochemicals Author Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany Abstract This Application Note shows the determination of aromatic hydrocarbons under normal phase conditions with the Agilent Infi nity Binary LC system with RID detection in compliance with the IP9()/ASTM D9 methods. Nagesh Babu and Sumitra Sankar Gulf Bio Analytical, United Arab Emirates, Dubai

2 Introduction The IP9()/ASTM D9 methods cover the determination of mono-aromatic (MAH), di-aromatic (DAH), and poly-aromatic hydrocarbon (TRI+) contaminants in petroleum middle distillates and diesel fuels boiling in the range from C to C, for example gas oil. The aromatic hydrocarbon content of motor diesel fuel is a factor that can affect exhaust emissions and fuel combustion characteristics, as measured by the cetane number. Using this instrumental setup, it is possible to determine mono and di-aromatic hydrocarbons according to IP/ASTM D79. It is also possible to determine the total aromatic hydrocarbons and total saturates according to ASTM D79. This Application Note shows the possibility to determine aromatic hydrocarbons under normal phase conditions with RID detection by using the Agilent Infinity binary LC system while complying with the IP9()/ ASTM D9 methods. Experimental Instruments Agilent Infi nity LC System Agilent Infinity Binary Pump (GB) with Agilent Infinity Standard Degasser (GA) Agilent Infinity Standard Autosampler (G9B) with Agilent 9 Infinity Thermostat (GB) Agilent Infinity Thermostatted Column Compartment (GA) with Agilent Infinity Series Quick-Change -postition/-port valve (G A). Agilent Infinity Refractive Index Detector (GA) Software Agilent OpenLAB CDS ChemStation Edition for LC & LC/MS Systems, Rev. C.. Columns Agilent ZORBAX NH,. mm, µm (p/n ) Agilent ZORBAX NH, Guard Cartridges,.. mm, µm (p/n 89-98) Agilent ZORBAX High Performance Guard Fittings Kit (p/n ) Reagents Mobile Phase: Heptane, HPLC Grade Standards System Performance Standard (SPS) - ASTM-P--PAK: cyclohexane ( mg/ml), o-xylene ( mg/ml, MAH), dibenzothiophene ( mg/ml, DAH), 9-methylanthracene ( mg/ml, TRI+). Calibration Standard: ASTM-P--SET Std. Std. Std. Cyclohexane o-xylene (MAH).. Debenzothiophene (DAH).. 9-Methylantracene (TRI+).. Std.

3 LC method RID Peak width >. minutes ( second response time) (.8 Hz) Temperature. C Column compartment Initial valve position Port & Temperature. C Change valve position 7.8 minutes: Port & Sampler Injection mode Standard injection Injection volume. μl Needle wash in vial with heptane Binary pump Flow. ml/min Stop time. min Solvent A Heptane, isocratic Valve operation (Figure ) In the OFF position, the sample is injected into the column through the autosampler. Once the DAH is eluted from the column, the valve is switched to the ON position, which reverses the flow in the column, backflushing the TRI+ aromatics to the RID detector. The time B (seconds) for switching to backflush is calculated based on an injection of the System Performance Standard (SPS) without valve switching using the following equation: B = t A +. (t B t A ) A Pump Autosampler B Pump Autosampler NH-column NH-column Detector Where: t A is the retention time of dibenzothiophene (DAH), in seconds. t B is the retention time of 9-methylanthracene (TRI+), in seconds. Detector Figure. Valve configuration for the IP 9 method. A) OFF Position & is the loading position. B) After switching to ON position & the compounds are eluted and analyzed in backflush.

4 Sample preparation Place between.9 g to. g of sample, to the nearest. g, into a -ml volumetric flask and fill to the mark with n-heptane. Shake thoroughly to mix, and filter the sample to remove the particulate matter. RID, Refractive index signal Non-aromatics. Mono-aromatics.97 Di-aromatics.8 Results and Discussion With the final method, all four calibration standards were measured (Figure ). The measurement showed a good separation of the inherent compounds, a clear separation of the aromatic compounds from the aliphatic cyclohexane, and a separation of the poly nuclear aromatic hydrocarbons to a much higher retention time. This is necessary for their determination in real samples. For the non-aromatic as well as aromatic compounds (MAH, DAH, TRI+), individual calibration curves were created. They ranged from to mg/ml for cyclohexane,. to mg/ml for MAH and DAH, and. to mg/ml for TRI+ aromatics (Figure ). All calibration curves for the aromatic hydrocarbons showed linearity coefficients better than To determine the limit-of-detection (LOD), the signals and the noise were measured for a / dilution of Standard. The typical LOD for aromatic compounds is µg/ml (calculated for a signal-tonoise ratio of ). Statistical evaluation of retention time and area for the aromatic compounds by multiple injection of calibration standard level showed RSD values typically below. % and. %, respectively (Table ). nriu tri+-aromatics Time (min) Figure. Overlay of IP9 Calibration standard (ASTM-P--SET), cyclohexane (non-aromatics), o-xylene (MAH), dibenzothiophene (DAH), 9-methylanthracene (TRI+). A Non-aromatics, RID A = *Amt 97.7 Correlation:.997,, C Di-aromatics, RID A =.*Amt Correlation:.99997,,, Figure. Calibration curves for the four compounds tested. B Mono-aromatics, RID A =.878*Amt +. Correlation:.99987,,, 8 D Tri+-aromatics, RID A =.88*Amt Correlation:.9998,,, Table. Statistical evaluation of retention time and area for the aromatic compounds by multiple injection (n = ) of calibration standard level. Mono-aromatics Di-aromatics Tri+-aromatics RT (min).7.. RT RSD (%)... RSD (%)... LOD (µg/ml).7

5 Figure displays a RI chromatogram of a real sample. This sample shows approximately. mg/ml of tri+aromatic compounds. Conclusion This Application Note demonstrates the use of the Agilent Infinity Binary LC in combination with the Agilent Infinity RI Detector for the determination mono-aromatic (MAH), di-aromatic (DAH), and tri+-aromatic (TRI+) contaminants in petroleum middle distillates and diesel fuels according to IP 9()/ASTM D9. The calibration of the RID showed excellent linearity for MAH, DAH, and TRI+ aromatics. The retention time RSD is typically below. % and the area RSD is typically below. %. nriu RID, Refractive index signal Non-aromatics Mono-aromatics Di-aromatics 8 Time (min) Figure. Real diesel oil sample showing aromatic compounds. Tri-aromatics References. IP9()/ASTM D9, Standard Test Method for Determination of Aromatic Hydrocarbon Types in Middle Distillates High Performance Liquid Chromatography Method with Refractive Index Detection.. D9.htm. Agilent Application Note, Publication number 99-7EN. Agilent Application note, Publication number 99-7EN

6 This information is subject to change without notice. Agilent Technologies, Inc., Published in the USA, December, 99-7EN

Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems

Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems Application Note Hydrocarbons Authors Michael Woodman Agilent Technologies, Inc. Chemical Analysis Solutions 2850

More information

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC Application Note Energy and Fuels Authors Detlef Wilhelm Anatox GmbH & Co. KG Fürstenwalde, Germany Udo Huber Agilent

More information

Meeting the Requirements of ASTM D ( IP548/06 ) with Agilent 1200 LC Systems

Meeting the Requirements of ASTM D ( IP548/06 ) with Agilent 1200 LC Systems Meeting the Requirements of ASTM D 6591-06 ( IP548/06 ) with Agilent 1200 LC Systems M.Woodman, Chemical Analysis Solutions Unit, Wilmington, Delaware, USA M.Sierocinska, European Field Support Centre,

More information

Application Note. Authors. Abstract. Energy & Chemicals

Application Note. Authors. Abstract. Energy & Chemicals Determination of Aromatic Content in Diesel Fuel According to ASTM D5186 Enhancing the Agilent 126 Infi nity Analytical SFC System with a Flame Ionization Detector Application Note Energy & Chemicals Authors

More information

Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution

Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 129 Infinity II HDR-DAD Impurity Analyzer Solution Application ote Small Molecule Pharmaceuticals Author Sonja Schneider Agilent Technologies,

More information

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Application Note Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Authors Kelly Beard and James McCurry Agilent Technologies, Inc. Abstract An Agilent

More information

Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS

Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS Application Environmental Authors Wei Luan and Chuanhong Tu Agilent Technologies (Shanghai) Co., Ltd. 412 Ying

More information

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc.

More information

High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell

High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell Application Note Consumer Products Authors Gerd Vanhoenacker,

More information

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Application Note Petrochemicas High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Author James D. McCurry, Ph.D. Agilent Technologies, Inc. Abstract An Agilent

More information

The Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over

The Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over The Agilent 12 Series high performance autosampler SL: Area precision, injection volume linearity, imum accessible volume, carry-over Technical Note 8 6 4 2.5 1 Introduction A main performance criterion

More information

Online sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve

Online sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve Online sample cleanup on the Agilent 129 Infinity LC using a built in 2-position/6-port valve Analysis of Sudan red compounds in paprika powder Application Note Food Authors Angelika Gratzfeld-Huesgen

More information

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector Detection of Sulfur Compounds in Natural Gas According to ASTM D554 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector Application Note Author Rebecca Veeneman Abstract Sulfur compounds in natural

More information

Using the PSD for Backflushing on the Agilent 8890 GC System

Using the PSD for Backflushing on the Agilent 8890 GC System Application Note Petrochemicals Using the PSD for Backflushing on the Agilent 889 GC System Author Brian Fitz Agilent Technologies, Inc. Wilmington, DE, USA. Abstract An Agilent 889 series GC equipped

More information

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve Technical Note This Technical Note describes the purpose, installation, and configuration and use of a diverter valve

More information

Agilent 1290 Infinity Autosampler

Agilent 1290 Infinity Autosampler Agilent 1290 Infinity Autosampler Installation of the Large Volume Injection Kit General Information This note describes how to install the large volume injection kit into an Agilent 1290 Infinity Autosampler.

More information

Analysis of Petroleum Fractions by ASTM D2887

Analysis of Petroleum Fractions by ASTM D2887 Analysis of Petroleum Fractions by ASTM D2887 Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 2582 Key Words Simulated distillation, D2887, TRACE TR-SimDist Abstract ASTM

More information

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector Application Note Hydrocarbon Processing Authors ChunXiao Wang Agilent Technologies

More information

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application Petroleum Authors ChunXiao Wang Agilent Technologies (Shanghai) Co.,Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai

More information

Application Note. Abstract. Authors. Environmental Analysis

Application Note. Abstract. Authors. Environmental Analysis High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID using the Agilent Low Thermal Mass (LTM II) System Application Note Environmental Analysis Authors Frank David and Karine Jacq Research

More information

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System Application Note Authors Frank David Research Institute for Chromatography, Pres.

More information

Application Note. Author. Introduction. Energy and Fuels

Application Note. Author. Introduction. Energy and Fuels Analysis of Free and Total Glycerol in B-100 Biodiesel Methyl Esters Using Agilent Select Biodiesel for Glycerides Application Note Energy and Fuels Author John Oostdijk Agilent Technologies, Inc. Introduction

More information

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D759 Application Petrochemical Author James D. McCurry Agilent Technologies 285 Centerville

More information

Determination of Sudan Dyes I IV in Curry Paste

Determination of Sudan Dyes I IV in Curry Paste Determination of Sudan Dyes I IV in Curry Paste Suparerk Tukkeeree and Jeffrey Rohrer 2 Thermo Fisher Scientific, Bangkok, Thailand; 2 Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 23 Key

More information

AppNote 1/2010 KEYWORDS ABSTRACT. Biodiesel, Automation, ASTM D

AppNote 1/2010 KEYWORDS ABSTRACT. Biodiesel, Automation, ASTM D AppNote 1/2010 Full Automation of ASTM Method D6584-07 Standard Test Method for the Determination of Free and Total Glycerin in B-100 Biodiesel Methyl Esters by Gas Chromatography using a GERSTEL Dual

More information

High-Temperature Simulated Distillation System Based on the 6890N GC Application

High-Temperature Simulated Distillation System Based on the 6890N GC Application High-Temperature Simulated Distillation System Based on the 6890N GC Application Petroleum Authors ChunXiao Wang Agilent Technologies (Shanghai) Co., Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai

More information

Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105

Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105 Application Note: 10215 Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105 Fausto Munari, Daniela Cavagnino, Andrea Cadoppi, Thermo Fisher Scientific, Milan,

More information

S 1125 HPLC Pump System

S 1125 HPLC Pump System Routine S 170 EXPERT HPLC HPLC System SYSTEMS 500 The Sykam ROUTINE HPLC System S 500 is intended for all routine analysis. The system configuration is highly variable and several upgrade options make

More information

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585 GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc. 850 Centerville

More information

Technical Overview. Introduction

Technical Overview. Introduction Performance characteristics of the Agilent 129 Infinity Binary Pump More resolution and speed for conventional, superficially porous and sub-2 µm column packing material Technical Overview Introduction

More information

Analysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction

Analysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction Analysis of Glycerin and Glycerides in Biodiesel (B1) Using ASTM D68 and EN11 Application HPI/Petrochemicals/Polymers Author James D. McCurry Agilent Technologies, Inc. 8 Centerville Road Wilmington, DE

More information

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel Application Note Energy and Chemicals Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel Applying the Energy Institute Method IP 8 with an Agilent J&W

More information

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584 Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN 14105 and ASTM D6584 Introduction With today s increasing concern for the environment and the depletion of fossil

More information

Technical Procedure for Gas Chromatography (GC-FID)

Technical Procedure for Gas Chromatography (GC-FID) Technical Procedure for Gas Chromatography (GC-FID) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph (GC- FID). 2.0 Scope This procedure applies to all

More information

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC Technical Overview Trace gas analysis is a challenge in today s world. The ability to analyze lower component

More information

Agilent Online-SPE Direct Inject Kit G4744A

Agilent Online-SPE Direct Inject Kit G4744A Agilent Online-SPE Direct Inject Kit G4744A Installation Guide Agilent Online-SPE Direct Inject Kit G4744A - Installation Guide Contents Application 2 Delivery Checklist 4 Online SPE Direct Inject Kit

More information

Boosting Trace Detection Performance with the Vanquish Diode Array Detector and High-Sensitivity LightPipe Flow Cell

Boosting Trace Detection Performance with the Vanquish Diode Array Detector and High-Sensitivity LightPipe Flow Cell Boosting Trace Detection Performance with the Vanquish Diode Array Detector and High-Sensitivity LightPipe Flow Cell Alexandra Manka and Holger Franz Thermo Fisher Scientific, Germering, Germany Technical

More information

Agilent InfinityLab LC Series Multi-wash Upgrade Kit

Agilent InfinityLab LC Series Multi-wash Upgrade Kit Agilent InfinityLab LC Series Multi-wash Upgrade Kit General Information Installation Note The multi-wash upgrade kit is designed for low carry over and can be installed in any Agilent InfinityLab LC Series

More information

Performance Characteristics of the Agilent 1260 Infinity II Bio-Inert LC

Performance Characteristics of the Agilent 1260 Infinity II Bio-Inert LC Performance haracteristics of the gilent 16 Infinity II io-inert L Technical Overview uthors Sonja Schneider and Uschi Schweikert gilent Technologies, Inc. Waldbronn, Germany Introduction The gilent 16

More information

HPLC System S 600 Series

HPLC System S 600 Series HPLC System S 600 Series The Sykam HPLC System S 600 Series is intended for the ambitious analyst. The system configuration is highly variable and several upgrade options make this system suitable for

More information

Agilent G5632 Valve Kit Instructions

Agilent G5632 Valve Kit Instructions Agilent G5632 Valve Kit Instructions Agilent G5632 Valve Kit Instructions Contents Typical Applications 2 Alternating Column Regeneration (2 position/10 port valves only) 2 Dual column selection 3 Sample

More information

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS FLASH VAPORISATION INJECTION Split Splitless On-Column COOL INJECTION Large Volume Injection (LVI) On-Column On-Column-SVE (with solvent vapour

More information

NEW HPLC - SYSTEM LINE

NEW HPLC - SYSTEM LINE NEW HPLC - SYSTEM LINE S c h a m b e c k S F D G m b H Drieschweg 13A D-53604 Bad Honnef/Germany Phone: +49 2224 9239-0 E-Mail: info@schambeck-sfd.com Internet: www.schambeck-sfd.com S 9425 HPLC solvent

More information

AppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS

AppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS AppNote 6/26 Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) Andreas Hoffmann GERSTEL GmbH & Co.KG, Eberhard-Gerstel-Platz 1,

More information

HPLC SYSTEM S 500 SERIES

HPLC SYSTEM S 500 SERIES HPLC SYSTEM S 500 SERIES The Sykam HPLC System S 500 Series is intended for all routine analysis as well as for the ambitious analyst. The system configuration is highly variable and several upgrade options

More information

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract. Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers Application Note Abstract Versa With the rising prices of fossil fuels, more emphasis is being put on renewable resources

More information

High-throughput protein aggregate analysis of monoclonal antibodies using a novel dual-channel UHPLC instrument

High-throughput protein aggregate analysis of monoclonal antibodies using a novel dual-channel UHPLC instrument APPLICATION NOTE 72598 High-throughput protein aggregate analysis of monoclonal antibodies using a novel dual-channel UHPLC instrument Authors Nicola McGillicuddy, 1 Amy Farrell, 1 Sara Carillo, 1 Martin

More information

TECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure

TECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure TECHNICAL REPORT Modifying Agilent 1100 HPLC Systems to Achieve UHPLC-like performance with HALO Fused-Core Columns Agilent 1100 HPLC system With a few modifications, an Agilent 1100 HPLC can produce UHPLC-like

More information

Agilent InfinityLab LC Purification Solutions SELECTION GUIDE

Agilent InfinityLab LC Purification Solutions SELECTION GUIDE gilent InfinityLab LC urification Solutions SELECTION GUIDE GILENT INFINITYLB LC URIFICTION SOLUTIONS EFFICIENT LC URIFICTION SOLUTIONS FOR NLYTICL TO RERTIVE SCLE gilent InfinityLab LC urification Solutions

More information

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis Introduction Petroleum & Petrochemical Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis By Katarina Oden, Barry Burger, and Amanda Rigdon Crude oil consists of thousands of different

More information

Simultaneous Determination of Fatty Acid Methyl Esters Contents in the Biodiesel by HPLC-DAD Method

Simultaneous Determination of Fatty Acid Methyl Esters Contents in the Biodiesel by HPLC-DAD Method 2016 International Conference on Applied Mechanics, Mechanical and Materials Engineering (AMMME 2016) ISBN: 978-1-60595-409-7 Simultaneous Determination of Fatty Acid Methyl Esters Contents in the Biodiesel

More information

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS MS Identification of Trace level Impurities from a Non-MS Compatible Mobile Phase Using ACQUITY UPLC System with 2D Technology by Heart-cutting and Online Sample Concentration Bronsky Gopinadh, Dilshad

More information

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN1413:211 Method Application Note Author James D. McCurry, Ph.D. Agilent Technologies Abstract

More information

Dual-Needle Infinity II Upgrade Kit

Dual-Needle Infinity II Upgrade Kit Dual-Needle Infinity II Upgrade Kit Installation Note General Information The Dual-needle Upgrade Kit is designed for very fast injection for high throughput and unmatched flexibility to obtain the best

More information

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI) Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Brian Fitz and Bill Wilson Agilent Technologies, Inc. 285 Centerville Road Wilmington,

More information

Application. Gas Chromatography June 1995

Application. Gas Chromatography June 1995 Determining Oxygenates in Gasoline: ASTM Method D Application Gas Chromatography June 99 Authors Michael J. Szelewski Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-60 USA Matthew S. Klee

More information

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System Jim McCurry Roger Firor Agilent Technologies Wilmington, DE Page 1 Scope of the Analyzer Designed for ASTM D7798 For

More information

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE AN HPLC THAT WORKS AS HARD AS YOU DO Confidence in your results comes with high performance, day-to-day reliability, and overall system robustness.

More information

Agilent G4236A 2D-LC Valve Installation Guide

Agilent G4236A 2D-LC Valve Installation Guide Agilent G4236A 2D-LC Valve Installation Guide Technical Note Contents Delivery Checklist 2 Installing Valve Heads 3 Removing the Transportation Lock and the Valve Dummy 3 Installing the Valve Head and

More information

VERITY Purification Systems Build the perfect system to fit your research

VERITY Purification Systems Build the perfect system to fit your research VERITY Purification Systems Build the perfect system to fit your research SPEC SHEET PURIFICATION VERITY PURIFICATION SYSTEMS WITH VERITY 3011 ISOCRATIC PUMP MAKE YOUR LAB LIFE EASIER From automated GPC

More information

Complete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges

Complete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges Complete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges Alexandria Pavkovich Jason Thomas Trent Sprenkle Outline Background EPA Method Requirements Background

More information

Phase Distribution of Ethanol, and Water in Ethyl Esters at K and K

Phase Distribution of Ethanol, and Water in Ethyl Esters at K and K Phase Distribution of Ethanol, and Water in Ethyl Esters at 298.15 K and 333.15 K Luis A. Follegatti Romero, F. R. M. Batista, M. Lanza, E.A.C. Batista, and Antonio J.A. Meirelles a ExTrAE Laboratory of

More information

Agilent 1290 Infinity II Ultra Low Dispersion Kit

Agilent 1290 Infinity II Ultra Low Dispersion Kit Agilent 1290 Infinity II Ultra Low Dispersion Kit Technical Note 1290 Infinity II Ultra Low Dispersion Kit - Technical Note In this note we describe how to install the Agilent 1290 Infinity II Ultra Low

More information

Simulated Distillation Analyzers, Software, Standards, Consumables, Training

Simulated Distillation Analyzers, Software, Standards, Consumables, Training Simulated Distillation Analyzers, Software, Standards, Consumables, Training www.separationsystems.com Offering the Fullest Range of Optimized Solutions Simulated distillation (SimDis) has been used to

More information

GC-101. Testing of Biodiesel Using Gas Chromatography. by Bob Armantrout. Presented at: Local Biodiesel: A Biodiesel Coop Conference

GC-101. Testing of Biodiesel Using Gas Chromatography. by Bob Armantrout. Presented at: Local Biodiesel: A Biodiesel Coop Conference GC-101 Testing of Biodiesel Using Gas Chromatography by Bob Armantrout (Image Research Machines plc) Presented at: Local Biodiesel: A Biodiesel Coop Conference Strategies For Community-based Biodiesel

More information

Determination of Free and Total Glycerin in B100 Biodiesel

Determination of Free and Total Glycerin in B100 Biodiesel Page 1 of 5 Page 1 of 5 Return to Web Version Determination of Free and Total Glycerin in B100 Biodiesel By: Michael D. Buchanan, Katherine K. Stenerson, and Vicki Yearick, Reporter US Vol 27.1 techservice@sial.com

More information

ElvaX Sulfur in Oil analyzer

ElvaX Sulfur in Oil analyzer ElvaX Sulfur in Oil analyzer Introduction ElvaX Sulfur in Oil was designed especially for analysis of petrochemicals for sulfur content in wide concentration range from ppm to 5%. Many international standards

More information

Agilent Valve Kit G4231A/B and G4232A/B/C

Agilent Valve Kit G4231A/B and G4232A/B/C Agilent Valve Kit G4231A/B and G4232A/B/C Instructions Contents Typical Applications 2 Dual column selection (G4231A/B or G4232A/B/C) 2 Sample enrichment and sample cleanup (G4231A/B or G4232A/B/C) 2 Alternating

More information

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI) Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Bill Wilson and Chin-Kai Meng Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

Agilent Valve Kit G4231A/B/C and G4232B/C/D

Agilent Valve Kit G4231A/B/C and G4232B/C/D Agilent Valve Kit G4231A/B/C and G4232B/C/D Instructions Agilent Valve Kit G4231A/B/C and G4232B/C/D - Instructions This Technical Note describes the installation of the Agilent G4231A/B/C and G4232B/C/D

More information

Performing ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software

Performing ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software Install a capillary column in the oven of the SRI GC. The ASTM method suggests a 12 meter.32mm id narrow-bore column coupled with a 2.5 meter guard column but permits the use of any column which exhibits

More information

MET-Biodiesel Capillary GC Columns

MET-Biodiesel Capillary GC Columns MET-Biodiesel Capillary GC Columns Product Specifications Product Features & Benefits Chromatograms FAQs Related Products Updated: February 2, 2009 Product Specifications 2 Product Specifications What

More information

Installation of the 2 position/10 port valve into the 1100 series thermostatted column compartment

Installation of the 2 position/10 port valve into the 1100 series thermostatted column compartment Agilent G1316A Thermostatted Column Compartment Technical Note Installation of the 2 position/10 port valve into the 1100 series thermostatted column compartment System Requirements 2 Performance Specifications

More information

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS) Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph-mass spectrometer (GC-MS). 2.0 Scope

More information

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector Authors: The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector Introduction Specification of the hydrocarbon composition of LPG is required as traces

More information

Neueste LC Entwicklungen

Neueste LC Entwicklungen Neueste LC Entwicklungen 2D-LC, High Dynamic Range Detektion, quaternäre Pumpe mit 1200 bar Sebastian Krahe Produktspezialist LC Agilent Technologies, Deutschland/Österreich 2. Oktober 2012 Hamburg 1 Agilent

More information

ASTM D for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software

ASTM D for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software ASTM D5501-04 for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software Larry Meeker and Alice J. Di Gioia Waters Corporation Houston Field Laboratory 5909 West Loop, South

More information

A Single Method for the Direct Determination of Total Glycerols in All Biodiesels Using Liquid Chromatography and Charged Aerosol Detection

A Single Method for the Direct Determination of Total Glycerols in All Biodiesels Using Liquid Chromatography and Charged Aerosol Detection A Single Method for the Direct Deteration of Total Glycerols in All Biodiesels Using Liquid Chromatography and Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth, Christopher Crafts, Thermo

More information

Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D

Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Instructions Agilent Valve Kit G4231A/C and G4232C/D - Instructions This technical note describes the installation and application of the Agilent

More information

Agilent SD-1 Purification System ml/min

Agilent SD-1 Purification System ml/min Agilent SD-1 Purification System - 200 ml/min Introduction Installation of the Capillary Kit This technical note shows you how to install the Stainless Steel Capillary Kit (SD- 1 Purification System -

More information

Agilent 218 Purification Solution

Agilent 218 Purification Solution Agilent 218 Purification Solution System User Guide Purification Solution System User Guide Agilent Technologies Notices Agilent Technologies, Inc. 2012, 2013 No part of this manual may be reproduced in

More information

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column Cory S. Fix, Director of Application Development cory.fix@vgcchromatography.com Willie Steinecker,

More information

GlycoChrom TM HPLC Analyzer. Hardware Installation and System Test

GlycoChrom TM HPLC Analyzer. Hardware Installation and System Test GlycoChrom TM HPLC Analyzer Hardware Installation and System Test PIN 410-5004 For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-8 00-424-6723) ,, Table of Contents

More information

PATROL UPLC Process Analysis System for Production

PATROL UPLC Process Analysis System for Production PATROL UPLC Process Analysis System for Production The PATROL UPLC Process Analysis System for Production is an integrated system solution designed and engineered to perform online and atline reaction

More information

Setting up SilFlow for BackFlush in your GC

Setting up SilFlow for BackFlush in your GC Setting up SilFlow for BackFlush in your GC What is backflush and why use it? The BackFlush system eliminates the need to bake heavy sample fractions off the capillary column. Oils, tars and other semivolatile

More information

Agilent 7693A Automated Liquid Sampler

Agilent 7693A Automated Liquid Sampler Agilent 7693A Automated Liquid Sampler Specifications Overview The Agilent 7693A is a state-of-the-art sample handling and injection system that provides the highest levels of precision and reliability

More information

Dual Plasma Sulfur and Nitrogen Chemiluminescence Detectors. Unsurpassed Stability, Selectivity, and Sensitivity for your GC Analysis

Dual Plasma Sulfur and Nitrogen Chemiluminescence Detectors. Unsurpassed Stability, Selectivity, and Sensitivity for your GC Analysis Dual Plasma Sulfur and Nitrogen Chemiluminescence Detectors Unsurpassed Stability, Selectivity, and Sensitivity for your GC Analysis Now with New Dual Plasma Technology Overview The Agilent Technologies

More information

Analysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES

Analysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES Analysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES Application note Petrochemical Author Neli Drvodelic Agilent Technologies Melbourne, Australia Introduction

More information

Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography

Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography Dr. Robert Lorenz Analytical Chemist Chevron Energy Technology Company Gulf Coast Conference Galveston, TX October

More information

Application Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER

Application Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D7423-09 using AC OXYTRACER Fast Analysis in

More information

CONSTITUTION OF COAL TAR PITCH AND ITS EFFECT ON PROPERTIES

CONSTITUTION OF COAL TAR PITCH AND ITS EFFECT ON PROPERTIES CONSTITUTION OF COAL TAR PITCH AND ITS EFFECT ON PROPERTIES H. K. Mayer 1 and I. C. Lewis 2 1 GrafTech International Ltd., 12900 Snow Road, Parma, OH 44130 2 Consultant Introduction Corresponding author

More information

a review of analytical methods for the quantification of aromatics in diesel fuels

a review of analytical methods for the quantification of aromatics in diesel fuels a review of analytical methods for the quantification of aromatics in diesel fuels Prepared for the Automotive Emissions Management Group by the Special Task Force on Diesel Fuel Emissions (AE/STF-7) P.

More information

White Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information

White Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Abstract High Temperature Simulated Distillation (High Temp SIMDIS) is one of the most frequently used techniques to determine

More information

Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas

Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas Liliana Krotz, Walter Galotta, and Guido Giazzi Thermo Fisher Scientific, Milan, Italy Overview

More information

Agilent 1200 Series Valves

Agilent 1200 Series Valves Agilent 1200 Series Valves User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or by any means (including electronic storage and

More information

Agilent Multimode Inlet for Gas Chromatography

Agilent Multimode Inlet for Gas Chromatography Agilent Multimode Inlet for Gas Chromatography Technical Note Agilent Multimode Inlet for the 7890A GC Designed to give you ease of use and maximum flexibility, the Agilent Multimode Inlet does everything

More information

economics and performance Spark Holland introduces OPTIMAS The optimum balance of BETTER SAMPLE CARE

economics and performance Spark Holland introduces OPTIMAS The optimum balance of BETTER SAMPLE CARE BETTER SAMPLE CARE Spark Holland introduces OPTIMAS The optimum balance of economics and performance Spark Holland introduces OPTIMAS The optimum balance of economics and performance OPTIMAS The optimum

More information

Analysis of Per/Polyfluoroalkyl Substances in Water Using an Agilent 6470 Triple Quadrupole LC/MS

Analysis of Per/Polyfluoroalkyl Substances in Water Using an Agilent 6470 Triple Quadrupole LC/MS Analysis of Per/Polyfluoroalkyl Substances in Water Using an Agilent 647 Triple Quadrupole LC/MS Application Note Authors Kathy Hunt and Ralph Hindle Vogon Laboratory Services Ltd., Cochrane, AB Canada

More information

APPLICATION OF SOLID PHASE MICROEXTRACTION (SPME) IN PROFILING HYDROCARBONS IN OIL SPILL CASES

APPLICATION OF SOLID PHASE MICROEXTRACTION (SPME) IN PROFILING HYDROCARBONS IN OIL SPILL CASES APPLICATION OF SOLID PHASE MICROEXTRACTION (SPME) IN PROFILING HYDROCARBONS IN OIL SPILL CASES Zuraidah Abdullah Munir*, Nor ashikin Saim, Nurul Huda Mamat Ghani Department of Chemistry, Faculty of Applied

More information

Optimized Supercritical Fluid Chromatographic Instrumentation for the Analysis of Petroleum Fractions

Optimized Supercritical Fluid Chromatographic Instrumentation for the Analysis of Petroleum Fractions Optimized Supercritical Fluid Chromatographic Instrumentation for the Analysis of Petroleum Fractions Bruce E. Richter*, Brian A. Jones, and Nathan L Porter Sensar/Larson-Davis, 1652 W. 820 N., Provo,

More information