1 Determination of Aromatic Hydrocarbons in Petroleum Middle Distillates with the Agilent Infinity Binary HPLC System with RID Detection According to IP9()/ASTM D9 Application Note Energy & Chemicals - Petrochemicals Author Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany Abstract This Application Note shows the determination of aromatic hydrocarbons under normal phase conditions with the Agilent Infi nity Binary LC system with RID detection in compliance with the IP9()/ASTM D9 methods. Nagesh Babu and Sumitra Sankar Gulf Bio Analytical, United Arab Emirates, Dubai
2 Introduction The IP9()/ASTM D9 methods cover the determination of mono-aromatic (MAH), di-aromatic (DAH), and poly-aromatic hydrocarbon (TRI+) contaminants in petroleum middle distillates and diesel fuels boiling in the range from C to C, for example gas oil. The aromatic hydrocarbon content of motor diesel fuel is a factor that can affect exhaust emissions and fuel combustion characteristics, as measured by the cetane number. Using this instrumental setup, it is possible to determine mono and di-aromatic hydrocarbons according to IP/ASTM D79. It is also possible to determine the total aromatic hydrocarbons and total saturates according to ASTM D79. This Application Note shows the possibility to determine aromatic hydrocarbons under normal phase conditions with RID detection by using the Agilent Infinity binary LC system while complying with the IP9()/ ASTM D9 methods. Experimental Instruments Agilent Infi nity LC System Agilent Infinity Binary Pump (GB) with Agilent Infinity Standard Degasser (GA) Agilent Infinity Standard Autosampler (G9B) with Agilent 9 Infinity Thermostat (GB) Agilent Infinity Thermostatted Column Compartment (GA) with Agilent Infinity Series Quick-Change -postition/-port valve (G A). Agilent Infinity Refractive Index Detector (GA) Software Agilent OpenLAB CDS ChemStation Edition for LC & LC/MS Systems, Rev. C.. Columns Agilent ZORBAX NH,. mm, µm (p/n ) Agilent ZORBAX NH, Guard Cartridges,.. mm, µm (p/n 89-98) Agilent ZORBAX High Performance Guard Fittings Kit (p/n ) Reagents Mobile Phase: Heptane, HPLC Grade Standards System Performance Standard (SPS) - ASTM-P--PAK: cyclohexane ( mg/ml), o-xylene ( mg/ml, MAH), dibenzothiophene ( mg/ml, DAH), 9-methylanthracene ( mg/ml, TRI+). Calibration Standard: ASTM-P--SET Std. Std. Std. Cyclohexane o-xylene (MAH).. Debenzothiophene (DAH).. 9-Methylantracene (TRI+).. Std.
3 LC method RID Peak width >. minutes ( second response time) (.8 Hz) Temperature. C Column compartment Initial valve position Port & Temperature. C Change valve position 7.8 minutes: Port & Sampler Injection mode Standard injection Injection volume. μl Needle wash in vial with heptane Binary pump Flow. ml/min Stop time. min Solvent A Heptane, isocratic Valve operation (Figure ) In the OFF position, the sample is injected into the column through the autosampler. Once the DAH is eluted from the column, the valve is switched to the ON position, which reverses the flow in the column, backflushing the TRI+ aromatics to the RID detector. The time B (seconds) for switching to backflush is calculated based on an injection of the System Performance Standard (SPS) without valve switching using the following equation: B = t A +. (t B t A ) A Pump Autosampler B Pump Autosampler NH-column NH-column Detector Where: t A is the retention time of dibenzothiophene (DAH), in seconds. t B is the retention time of 9-methylanthracene (TRI+), in seconds. Detector Figure. Valve configuration for the IP 9 method. A) OFF Position & is the loading position. B) After switching to ON position & the compounds are eluted and analyzed in backflush.
4 Sample preparation Place between.9 g to. g of sample, to the nearest. g, into a -ml volumetric flask and fill to the mark with n-heptane. Shake thoroughly to mix, and filter the sample to remove the particulate matter. RID, Refractive index signal Non-aromatics. Mono-aromatics.97 Di-aromatics.8 Results and Discussion With the final method, all four calibration standards were measured (Figure ). The measurement showed a good separation of the inherent compounds, a clear separation of the aromatic compounds from the aliphatic cyclohexane, and a separation of the poly nuclear aromatic hydrocarbons to a much higher retention time. This is necessary for their determination in real samples. For the non-aromatic as well as aromatic compounds (MAH, DAH, TRI+), individual calibration curves were created. They ranged from to mg/ml for cyclohexane,. to mg/ml for MAH and DAH, and. to mg/ml for TRI+ aromatics (Figure ). All calibration curves for the aromatic hydrocarbons showed linearity coefficients better than To determine the limit-of-detection (LOD), the signals and the noise were measured for a / dilution of Standard. The typical LOD for aromatic compounds is µg/ml (calculated for a signal-tonoise ratio of ). Statistical evaluation of retention time and area for the aromatic compounds by multiple injection of calibration standard level showed RSD values typically below. % and. %, respectively (Table ). nriu tri+-aromatics Time (min) Figure. Overlay of IP9 Calibration standard (ASTM-P--SET), cyclohexane (non-aromatics), o-xylene (MAH), dibenzothiophene (DAH), 9-methylanthracene (TRI+). A Non-aromatics, RID A = *Amt 97.7 Correlation:.997,, C Di-aromatics, RID A =.*Amt Correlation:.99997,,, Figure. Calibration curves for the four compounds tested. B Mono-aromatics, RID A =.878*Amt +. Correlation:.99987,,, 8 D Tri+-aromatics, RID A =.88*Amt Correlation:.9998,,, Table. Statistical evaluation of retention time and area for the aromatic compounds by multiple injection (n = ) of calibration standard level. Mono-aromatics Di-aromatics Tri+-aromatics RT (min).7.. RT RSD (%)... RSD (%)... LOD (µg/ml).7
5 Figure displays a RI chromatogram of a real sample. This sample shows approximately. mg/ml of tri+aromatic compounds. Conclusion This Application Note demonstrates the use of the Agilent Infinity Binary LC in combination with the Agilent Infinity RI Detector for the determination mono-aromatic (MAH), di-aromatic (DAH), and tri+-aromatic (TRI+) contaminants in petroleum middle distillates and diesel fuels according to IP 9()/ASTM D9. The calibration of the RID showed excellent linearity for MAH, DAH, and TRI+ aromatics. The retention time RSD is typically below. % and the area RSD is typically below. %. nriu RID, Refractive index signal Non-aromatics Mono-aromatics Di-aromatics 8 Time (min) Figure. Real diesel oil sample showing aromatic compounds. Tri-aromatics References. IP9()/ASTM D9, Standard Test Method for Determination of Aromatic Hydrocarbon Types in Middle Distillates High Performance Liquid Chromatography Method with Refractive Index Detection.. D9.htm. Agilent Application Note, Publication number 99-7EN. Agilent Application note, Publication number 99-7EN
6 This information is subject to change without notice. Agilent Technologies, Inc., Published in the USA, December, 99-7EN
Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems
Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems Application Note Hydrocarbons Authors Michael Woodman Agilent Technologies, Inc. Chemical Analysis Solutions 2850
More informationDetermination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC
Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC Application Note Energy and Fuels Authors Detlef Wilhelm Anatox GmbH & Co. KG Fürstenwalde, Germany Udo Huber Agilent
More informationMeeting the Requirements of ASTM D ( IP548/06 ) with Agilent 1200 LC Systems
Meeting the Requirements of ASTM D 6591-06 ( IP548/06 ) with Agilent 1200 LC Systems M.Woodman, Chemical Analysis Solutions Unit, Wilmington, Delaware, USA M.Sierocinska, European Field Support Centre,
More informationApplication Note. Authors. Abstract. Energy & Chemicals
Determination of Aromatic Content in Diesel Fuel According to ASTM D5186 Enhancing the Agilent 126 Infi nity Analytical SFC System with a Flame Ionization Detector Application Note Energy & Chemicals Authors
More informationImpurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution
Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 129 Infinity II HDR-DAD Impurity Analyzer Solution Application ote Small Molecule Pharmaceuticals Author Sonja Schneider Agilent Technologies,
More informationGas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593
Application Note Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Authors Kelly Beard and James McCurry Agilent Technologies, Inc. Abstract An Agilent
More informationEvaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS
Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS Application Environmental Authors Wei Luan and Chuanhong Tu Agilent Technologies (Shanghai) Co., Ltd. 412 Ying
More informationAgilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011
Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc.
More informationHigh Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell
High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell Application Note Consumer Products Authors Gerd Vanhoenacker,
More informationHigh Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph
Application Note Petrochemicas High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Author James D. McCurry, Ph.D. Agilent Technologies, Inc. Abstract An Agilent
More informationThe Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over
The Agilent 12 Series high performance autosampler SL: Area precision, injection volume linearity, imum accessible volume, carry-over Technical Note 8 6 4 2.5 1 Introduction A main performance criterion
More informationOnline sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve
Online sample cleanup on the Agilent 129 Infinity LC using a built in 2-position/6-port valve Analysis of Sudan red compounds in paprika powder Application Note Food Authors Angelika Gratzfeld-Huesgen
More informationDetection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector
Detection of Sulfur Compounds in Natural Gas According to ASTM D554 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector Application Note Author Rebecca Veeneman Abstract Sulfur compounds in natural
More informationUsing the PSD for Backflushing on the Agilent 8890 GC System
Application Note Petrochemicals Using the PSD for Backflushing on the Agilent 889 GC System Author Brian Fitz Agilent Technologies, Inc. Wilmington, DE, USA. Abstract An Agilent 889 series GC equipped
More informationAgilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve
Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve Technical Note This Technical Note describes the purpose, installation, and configuration and use of a diverter valve
More informationAgilent 1290 Infinity Autosampler
Agilent 1290 Infinity Autosampler Installation of the Large Volume Injection Kit General Information This note describes how to install the large volume injection kit into an Agilent 1290 Infinity Autosampler.
More informationAnalysis of Petroleum Fractions by ASTM D2887
Analysis of Petroleum Fractions by ASTM D2887 Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 2582 Key Words Simulated distillation, D2887, TRACE TR-SimDist Abstract ASTM
More informationDual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector
Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector Application Note Hydrocarbon Processing Authors ChunXiao Wang Agilent Technologies
More informationFast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application
Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application Petroleum Authors ChunXiao Wang Agilent Technologies (Shanghai) Co.,Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai
More informationApplication Note. Abstract. Authors. Environmental Analysis
High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID using the Agilent Low Thermal Mass (LTM II) System Application Note Environmental Analysis Authors Frank David and Karine Jacq Research
More informationHigh Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System
High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System Application Note Authors Frank David Research Institute for Chromatography, Pres.
More informationApplication Note. Author. Introduction. Energy and Fuels
Analysis of Free and Total Glycerol in B-100 Biodiesel Methyl Esters Using Agilent Select Biodiesel for Glycerides Application Note Energy and Fuels Author John Oostdijk Agilent Technologies, Inc. Introduction
More informationUsing a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application
Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D759 Application Petrochemical Author James D. McCurry Agilent Technologies 285 Centerville
More informationDetermination of Sudan Dyes I IV in Curry Paste
Determination of Sudan Dyes I IV in Curry Paste Suparerk Tukkeeree and Jeffrey Rohrer 2 Thermo Fisher Scientific, Bangkok, Thailand; 2 Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 23 Key
More informationAppNote 1/2010 KEYWORDS ABSTRACT. Biodiesel, Automation, ASTM D
AppNote 1/2010 Full Automation of ASTM Method D6584-07 Standard Test Method for the Determination of Free and Total Glycerin in B-100 Biodiesel Methyl Esters by Gas Chromatography using a GERSTEL Dual
More informationHigh-Temperature Simulated Distillation System Based on the 6890N GC Application
High-Temperature Simulated Distillation System Based on the 6890N GC Application Petroleum Authors ChunXiao Wang Agilent Technologies (Shanghai) Co., Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai
More informationDetermination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105
Application Note: 10215 Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105 Fausto Munari, Daniela Cavagnino, Andrea Cadoppi, Thermo Fisher Scientific, Milan,
More informationS 1125 HPLC Pump System
Routine S 170 EXPERT HPLC HPLC System SYSTEMS 500 The Sykam ROUTINE HPLC System S 500 is intended for all routine analysis. The system configuration is highly variable and several upgrade options make
More informationGC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585
GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc. 850 Centerville
More informationTechnical Overview. Introduction
Performance characteristics of the Agilent 129 Infinity Binary Pump More resolution and speed for conventional, superficially porous and sub-2 µm column packing material Technical Overview Introduction
More informationAnalysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction
Analysis of Glycerin and Glycerides in Biodiesel (B1) Using ASTM D68 and EN11 Application HPI/Petrochemicals/Polymers Author James D. McCurry Agilent Technologies, Inc. 8 Centerville Road Wilmington, DE
More informationIncreased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel
Application Note Energy and Chemicals Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel Applying the Energy Institute Method IP 8 with an Agilent J&W
More informationFree and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584
Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN 14105 and ASTM D6584 Introduction With today s increasing concern for the environment and the depletion of fossil
More informationTechnical Procedure for Gas Chromatography (GC-FID)
Technical Procedure for Gas Chromatography (GC-FID) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph (GC- FID). 2.0 Scope This procedure applies to all
More informationFast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC
Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC Technical Overview Trace gas analysis is a challenge in today s world. The ability to analyze lower component
More informationAgilent Online-SPE Direct Inject Kit G4744A
Agilent Online-SPE Direct Inject Kit G4744A Installation Guide Agilent Online-SPE Direct Inject Kit G4744A - Installation Guide Contents Application 2 Delivery Checklist 4 Online SPE Direct Inject Kit
More informationBoosting Trace Detection Performance with the Vanquish Diode Array Detector and High-Sensitivity LightPipe Flow Cell
Boosting Trace Detection Performance with the Vanquish Diode Array Detector and High-Sensitivity LightPipe Flow Cell Alexandra Manka and Holger Franz Thermo Fisher Scientific, Germering, Germany Technical
More informationAgilent InfinityLab LC Series Multi-wash Upgrade Kit
Agilent InfinityLab LC Series Multi-wash Upgrade Kit General Information Installation Note The multi-wash upgrade kit is designed for low carry over and can be installed in any Agilent InfinityLab LC Series
More informationPerformance Characteristics of the Agilent 1260 Infinity II Bio-Inert LC
Performance haracteristics of the gilent 16 Infinity II io-inert L Technical Overview uthors Sonja Schneider and Uschi Schweikert gilent Technologies, Inc. Waldbronn, Germany Introduction The gilent 16
More informationHPLC System S 600 Series
HPLC System S 600 Series The Sykam HPLC System S 600 Series is intended for the ambitious analyst. The system configuration is highly variable and several upgrade options make this system suitable for
More informationAgilent G5632 Valve Kit Instructions
Agilent G5632 Valve Kit Instructions Agilent G5632 Valve Kit Instructions Contents Typical Applications 2 Alternating Column Regeneration (2 position/10 port valves only) 2 Dual column selection 3 Sample
More informationGAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS
GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS FLASH VAPORISATION INJECTION Split Splitless On-Column COOL INJECTION Large Volume Injection (LVI) On-Column On-Column-SVE (with solvent vapour
More informationNEW HPLC - SYSTEM LINE
NEW HPLC - SYSTEM LINE S c h a m b e c k S F D G m b H Drieschweg 13A D-53604 Bad Honnef/Germany Phone: +49 2224 9239-0 E-Mail: info@schambeck-sfd.com Internet: www.schambeck-sfd.com S 9425 HPLC solvent
More informationAppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS
AppNote 6/26 Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) Andreas Hoffmann GERSTEL GmbH & Co.KG, Eberhard-Gerstel-Platz 1,
More informationHPLC SYSTEM S 500 SERIES
HPLC SYSTEM S 500 SERIES The Sykam HPLC System S 500 Series is intended for all routine analysis as well as for the ambitious analyst. The system configuration is highly variable and several upgrade options
More informationMethanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.
Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers Application Note Abstract Versa With the rising prices of fossil fuels, more emphasis is being put on renewable resources
More informationHigh-throughput protein aggregate analysis of monoclonal antibodies using a novel dual-channel UHPLC instrument
APPLICATION NOTE 72598 High-throughput protein aggregate analysis of monoclonal antibodies using a novel dual-channel UHPLC instrument Authors Nicola McGillicuddy, 1 Amy Farrell, 1 Sara Carillo, 1 Martin
More informationTECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure
TECHNICAL REPORT Modifying Agilent 1100 HPLC Systems to Achieve UHPLC-like performance with HALO Fused-Core Columns Agilent 1100 HPLC system With a few modifications, an Agilent 1100 HPLC can produce UHPLC-like
More informationAgilent InfinityLab LC Purification Solutions SELECTION GUIDE
gilent InfinityLab LC urification Solutions SELECTION GUIDE GILENT INFINITYLB LC URIFICTION SOLUTIONS EFFICIENT LC URIFICTION SOLUTIONS FOR NLYTICL TO RERTIVE SCLE gilent InfinityLab LC urification Solutions
More informationAlternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis
Introduction Petroleum & Petrochemical Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis By Katarina Oden, Barry Burger, and Amanda Rigdon Crude oil consists of thousands of different
More informationSimultaneous Determination of Fatty Acid Methyl Esters Contents in the Biodiesel by HPLC-DAD Method
2016 International Conference on Applied Mechanics, Mechanical and Materials Engineering (AMMME 2016) ISBN: 978-1-60595-409-7 Simultaneous Determination of Fatty Acid Methyl Esters Contents in the Biodiesel
More information[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS
MS Identification of Trace level Impurities from a Non-MS Compatible Mobile Phase Using ACQUITY UPLC System with 2D Technology by Heart-cutting and Online Sample Concentration Bronsky Gopinadh, Dilshad
More informationGC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method
GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN1413:211 Method Application Note Author James D. McCurry, Ph.D. Agilent Technologies Abstract
More informationDual-Needle Infinity II Upgrade Kit
Dual-Needle Infinity II Upgrade Kit Installation Note General Information The Dual-needle Upgrade Kit is designed for very fast injection for high throughput and unmatched flexibility to obtain the best
More informationAchieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)
Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Brian Fitz and Bill Wilson Agilent Technologies, Inc. 285 Centerville Road Wilmington,
More informationApplication. Gas Chromatography June 1995
Determining Oxygenates in Gasoline: ASTM Method D Application Gas Chromatography June 99 Authors Michael J. Szelewski Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-60 USA Matthew S. Klee
More informationOptimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE
Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System Jim McCurry Roger Firor Agilent Technologies Wilmington, DE Page 1 Scope of the Analyzer Designed for ASTM D7798 For
More informationAlliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE
Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE AN HPLC THAT WORKS AS HARD AS YOU DO Confidence in your results comes with high performance, day-to-day reliability, and overall system robustness.
More informationAgilent G4236A 2D-LC Valve Installation Guide
Agilent G4236A 2D-LC Valve Installation Guide Technical Note Contents Delivery Checklist 2 Installing Valve Heads 3 Removing the Transportation Lock and the Valve Dummy 3 Installing the Valve Head and
More informationVERITY Purification Systems Build the perfect system to fit your research
VERITY Purification Systems Build the perfect system to fit your research SPEC SHEET PURIFICATION VERITY PURIFICATION SYSTEMS WITH VERITY 3011 ISOCRATIC PUMP MAKE YOUR LAB LIFE EASIER From automated GPC
More informationComplete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges
Complete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges Alexandria Pavkovich Jason Thomas Trent Sprenkle Outline Background EPA Method Requirements Background
More informationPhase Distribution of Ethanol, and Water in Ethyl Esters at K and K
Phase Distribution of Ethanol, and Water in Ethyl Esters at 298.15 K and 333.15 K Luis A. Follegatti Romero, F. R. M. Batista, M. Lanza, E.A.C. Batista, and Antonio J.A. Meirelles a ExTrAE Laboratory of
More informationAgilent 1290 Infinity II Ultra Low Dispersion Kit
Agilent 1290 Infinity II Ultra Low Dispersion Kit Technical Note 1290 Infinity II Ultra Low Dispersion Kit - Technical Note In this note we describe how to install the Agilent 1290 Infinity II Ultra Low
More informationSimulated Distillation Analyzers, Software, Standards, Consumables, Training
Simulated Distillation Analyzers, Software, Standards, Consumables, Training www.separationsystems.com Offering the Fullest Range of Optimized Solutions Simulated distillation (SimDis) has been used to
More informationGC-101. Testing of Biodiesel Using Gas Chromatography. by Bob Armantrout. Presented at: Local Biodiesel: A Biodiesel Coop Conference
GC-101 Testing of Biodiesel Using Gas Chromatography by Bob Armantrout (Image Research Machines plc) Presented at: Local Biodiesel: A Biodiesel Coop Conference Strategies For Community-based Biodiesel
More informationDetermination of Free and Total Glycerin in B100 Biodiesel
Page 1 of 5 Page 1 of 5 Return to Web Version Determination of Free and Total Glycerin in B100 Biodiesel By: Michael D. Buchanan, Katherine K. Stenerson, and Vicki Yearick, Reporter US Vol 27.1 techservice@sial.com
More informationElvaX Sulfur in Oil analyzer
ElvaX Sulfur in Oil analyzer Introduction ElvaX Sulfur in Oil was designed especially for analysis of petrochemicals for sulfur content in wide concentration range from ppm to 5%. Many international standards
More informationAgilent Valve Kit G4231A/B and G4232A/B/C
Agilent Valve Kit G4231A/B and G4232A/B/C Instructions Contents Typical Applications 2 Dual column selection (G4231A/B or G4232A/B/C) 2 Sample enrichment and sample cleanup (G4231A/B or G4232A/B/C) 2 Alternating
More informationAchieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)
Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Bill Wilson and Chin-Kai Meng Agilent Technologies, Inc. 2850 Centerville Road Wilmington,
More informationAgilent Valve Kit G4231A/B/C and G4232B/C/D
Agilent Valve Kit G4231A/B/C and G4232B/C/D Instructions Agilent Valve Kit G4231A/B/C and G4232B/C/D - Instructions This Technical Note describes the installation of the Agilent G4231A/B/C and G4232B/C/D
More informationPerforming ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software
Install a capillary column in the oven of the SRI GC. The ASTM method suggests a 12 meter.32mm id narrow-bore column coupled with a 2.5 meter guard column but permits the use of any column which exhibits
More informationMET-Biodiesel Capillary GC Columns
MET-Biodiesel Capillary GC Columns Product Specifications Product Features & Benefits Chromatograms FAQs Related Products Updated: February 2, 2009 Product Specifications 2 Product Specifications What
More informationInstallation of the 2 position/10 port valve into the 1100 series thermostatted column compartment
Agilent G1316A Thermostatted Column Compartment Technical Note Installation of the 2 position/10 port valve into the 1100 series thermostatted column compartment System Requirements 2 Performance Specifications
More informationTechnical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)
Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph-mass spectrometer (GC-MS). 2.0 Scope
More informationThe Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector
Authors: The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector Introduction Specification of the hydrocarbon composition of LPG is required as traces
More informationNeueste LC Entwicklungen
Neueste LC Entwicklungen 2D-LC, High Dynamic Range Detektion, quaternäre Pumpe mit 1200 bar Sebastian Krahe Produktspezialist LC Agilent Technologies, Deutschland/Österreich 2. Oktober 2012 Hamburg 1 Agilent
More informationASTM D for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software
ASTM D5501-04 for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software Larry Meeker and Alice J. Di Gioia Waters Corporation Houston Field Laboratory 5909 West Loop, South
More informationA Single Method for the Direct Determination of Total Glycerols in All Biodiesels Using Liquid Chromatography and Charged Aerosol Detection
A Single Method for the Direct Deteration of Total Glycerols in All Biodiesels Using Liquid Chromatography and Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth, Christopher Crafts, Thermo
More informationAgilent InfinityLab Quick Change Valves G4231A/C and G4232C/D
Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Instructions Agilent Valve Kit G4231A/C and G4232C/D - Instructions This technical note describes the installation and application of the Agilent
More informationAgilent SD-1 Purification System ml/min
Agilent SD-1 Purification System - 200 ml/min Introduction Installation of the Capillary Kit This technical note shows you how to install the Stainless Steel Capillary Kit (SD- 1 Purification System -
More informationAgilent 218 Purification Solution
Agilent 218 Purification Solution System User Guide Purification Solution System User Guide Agilent Technologies Notices Agilent Technologies, Inc. 2012, 2013 No part of this manual may be reproduced in
More informationASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column
ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column Cory S. Fix, Director of Application Development cory.fix@vgcchromatography.com Willie Steinecker,
More informationGlycoChrom TM HPLC Analyzer. Hardware Installation and System Test
GlycoChrom TM HPLC Analyzer Hardware Installation and System Test PIN 410-5004 For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-8 00-424-6723) ,, Table of Contents
More informationPATROL UPLC Process Analysis System for Production
PATROL UPLC Process Analysis System for Production The PATROL UPLC Process Analysis System for Production is an integrated system solution designed and engineered to perform online and atline reaction
More informationSetting up SilFlow for BackFlush in your GC
Setting up SilFlow for BackFlush in your GC What is backflush and why use it? The BackFlush system eliminates the need to bake heavy sample fractions off the capillary column. Oils, tars and other semivolatile
More informationAgilent 7693A Automated Liquid Sampler
Agilent 7693A Automated Liquid Sampler Specifications Overview The Agilent 7693A is a state-of-the-art sample handling and injection system that provides the highest levels of precision and reliability
More informationDual Plasma Sulfur and Nitrogen Chemiluminescence Detectors. Unsurpassed Stability, Selectivity, and Sensitivity for your GC Analysis
Dual Plasma Sulfur and Nitrogen Chemiluminescence Detectors Unsurpassed Stability, Selectivity, and Sensitivity for your GC Analysis Now with New Dual Plasma Technology Overview The Agilent Technologies
More informationAnalysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES
Analysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES Application note Petrochemical Author Neli Drvodelic Agilent Technologies Melbourne, Australia Introduction
More informationRefinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography
Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography Dr. Robert Lorenz Analytical Chemist Chevron Energy Technology Company Gulf Coast Conference Galveston, TX October
More informationApplication Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER
Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D7423-09 using AC OXYTRACER Fast Analysis in
More informationCONSTITUTION OF COAL TAR PITCH AND ITS EFFECT ON PROPERTIES
CONSTITUTION OF COAL TAR PITCH AND ITS EFFECT ON PROPERTIES H. K. Mayer 1 and I. C. Lewis 2 1 GrafTech International Ltd., 12900 Snow Road, Parma, OH 44130 2 Consultant Introduction Corresponding author
More informationa review of analytical methods for the quantification of aromatics in diesel fuels
a review of analytical methods for the quantification of aromatics in diesel fuels Prepared for the Automotive Emissions Management Group by the Special Task Force on Diesel Fuel Emissions (AE/STF-7) P.
More informationWhite Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information
Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Abstract High Temperature Simulated Distillation (High Temp SIMDIS) is one of the most frequently used techniques to determine
More informationStability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas
Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas Liliana Krotz, Walter Galotta, and Guido Giazzi Thermo Fisher Scientific, Milan, Italy Overview
More informationAgilent 1200 Series Valves
Agilent 1200 Series Valves User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or by any means (including electronic storage and
More informationAgilent Multimode Inlet for Gas Chromatography
Agilent Multimode Inlet for Gas Chromatography Technical Note Agilent Multimode Inlet for the 7890A GC Designed to give you ease of use and maximum flexibility, the Agilent Multimode Inlet does everything
More informationeconomics and performance Spark Holland introduces OPTIMAS The optimum balance of BETTER SAMPLE CARE
BETTER SAMPLE CARE Spark Holland introduces OPTIMAS The optimum balance of economics and performance Spark Holland introduces OPTIMAS The optimum balance of economics and performance OPTIMAS The optimum
More informationAnalysis of Per/Polyfluoroalkyl Substances in Water Using an Agilent 6470 Triple Quadrupole LC/MS
Analysis of Per/Polyfluoroalkyl Substances in Water Using an Agilent 647 Triple Quadrupole LC/MS Application Note Authors Kathy Hunt and Ralph Hindle Vogon Laboratory Services Ltd., Cochrane, AB Canada
More informationAPPLICATION OF SOLID PHASE MICROEXTRACTION (SPME) IN PROFILING HYDROCARBONS IN OIL SPILL CASES
APPLICATION OF SOLID PHASE MICROEXTRACTION (SPME) IN PROFILING HYDROCARBONS IN OIL SPILL CASES Zuraidah Abdullah Munir*, Nor ashikin Saim, Nurul Huda Mamat Ghani Department of Chemistry, Faculty of Applied
More informationOptimized Supercritical Fluid Chromatographic Instrumentation for the Analysis of Petroleum Fractions
Optimized Supercritical Fluid Chromatographic Instrumentation for the Analysis of Petroleum Fractions Bruce E. Richter*, Brian A. Jones, and Nathan L Porter Sensar/Larson-Davis, 1652 W. 820 N., Provo,
More information