1 UNIVERSITY OF HOUSTON ENVIRONMENTAL HEALTH AND SAFETY POLICIES AND PROCEDURES SECTION: SUBJECT: ENVIRONMENTAL HEALTH AND SAFETY LABORATORY FUME HOODS I. PURPOSE These laboratory fume hood survey procedures detail Environmental Health and Safety s (EHS) efforts to determine if the airflow rate is adequate for safe use by laboratory personnel. A fume hood is generally used in a laboratory as a safety device to provide local control and removal of airborne contaminants, ventilation, containment, shielding, or other uses. The proper operation of fume hoods is vital to the safety of the laboratory user. EHS will inspect all fume hoods as part of the regularly scheduled laboratory safety audit program. These procedures apply to all fume hoods located on the main campus as well as the College of Pharmacy located in the Texas Medical Center. II. PROCEDURES The fume hoods should be operated at the proper working sash height or door opening as follows: Bench fume hood with horizontal sash - 14" working sash height Bench fume hood with vertical sliding doors 14" working gap between doors Walk-in fume hood with horizontal doors determined by the needs of the specific investigator Walk-in fume hood with vertical doors determined by the needs of the specific investigator Plant Operations (PO) personnel make all corrective adjustments of the fume hood flow rates. In general, PO is responsible for routine maintenance and repair of the fume hoods. EHS will immediately report all fume hoods found to be deficient to PO this will normally be required due to a low flow rate. Other needed repairs, such as malfunctioning low-flow alarms or inoperative sashes or doors, should also be called-in by laboratory personnel when the repairs are needed. EHS may be required to determine if fume hoods are safe for PO personnel to work where chemicals or radioactive materials are used. A database of all fume hoods is maintained by PO. A PO number uniquely identifies each fume hood. EHS will use the PO number to refer to fume hoods in any instance if the fume hood number is known. Inspection dates and flow rates found during EHS surveys will be maintained by the EHS. However PO will have access to this information and be permitted to add it to their database. UH EHS 1

2 The following equipment is needed to perform the fume hood survey: 1. Hot wire anemometer or similar instrument to measure fume hood flow rates 2. Tape measure 3. Calculator 4. Fume Hood Air Velocity form 5. Yellow verification of measurement label with sash height arrow and black marking pen 6. Notice of Fume Hood Deficiency form The following measurement techniques are to be used: 1. All readings will be taken with the hot wire anemometer or similar instument. The anemometer is used to measure inward linear air velocity in ft/min on the same plane as the fume hood sash or doors (Example: face velocity). 2. Equally spaced air velocity readings will be taken based on a square foot grid system, using the Fume Hood Air Velocity Measurements form, as follows: Bench fume hood with horizontal sash sash at working height (normally between 14 and 18 ) Bench or walk-in fume hood with sliding doors consult with investigator for normal working opening Walk-in fume hood with horizontal sashes - consult with investigator for normal working opening 3. The average face velocity will be calculated by averaging the air velocity readings. Criteria used for satisfactory face velocity The target range of desired face velocities for conventional bench type fume hoods will be between fpm. As stated previously, walk in fume hoods will be handled on an individual basis with the specific investigator involved. Some of the newer laboratory buildings have low flow fume hoods that operate between fpm. The accepted criteria for these fume hoods will be the manufactures lowest operating range as indicated on the hood sticker. For fume hoods that don t meet the minimum flow rates, the following steps will be taken: o Immediately call the Physical Plant Customer Service Center (FIXIT) at to report the deficiencies. Write the Work Order number on a Fume Hood Deficiency Notice. o Make a copy of the Fume Hood Deficiency Notice for the inspection records. UH EHS 2

3 o Post the Notice of Fume Hood Deficiency on the fume hood. o The laboratory contact will be advised that the fume hood has failed the inspection and should not be used until repairs are completed by PO personnel. For fume hoods that fall in the target range, the following steps will be taken: o Record date, average face velocity, and initials of inspector on the yellow verification of measurement sticker. o Post the sticker at the working height of the fume hood. For fume hoods that exceed the target range, the following steps will be taken: o Record date, average face velocity, and initials of inspector on the yellow verification of measurement sticker. o Post the sticker at the normal working height of the fume hood. o EHS will then forward this information to the Facilities Management Group and request that the fume hood be brought into the desired range. III. REFERENCES ANSI/ASHRAE , Method of Testing Performance of Laboratory Fume Hoods. (1995 or most recent addition) Prudent Practices in the Laboratory: Handling and Disposing of Chemicals National Research Council, National Academy Press, Washington, D.C. (1995 or most recent edition) UH EHS 3

4 University of Houston NOTICE OF FUMEHOOD DEFICIENCY Building: Room No: Fume Hood ID Number: As part of the laboratory safety inspection program, this fume hood was surveyed by Environmental Health and Safety Department (EHS) personnel, and was found to have the following problem(s): Low Flow Rate No Flow (< 25 ft/min) Flow Blocked by Equipment Inside Flow Blocked by Equipment Outside Sash Inoperative Other This fume hood has been called-in by EHS personnel for repairs. THIS FUME HOOD SHOULD NOT BE USED UNTIL THE REPAIRS ARE COMPLETED BY PLANT OPERATIONS PERSONNEL AND THIS NOTICE HAS BEEN REMOVED BY PLANT OPERATIONS OR ENVIRONMENTAL HEALTH AND SAFETY PERSONNEL. For the status of fume hood repairs, contact the Pant Operations Customer Service Center at (FIXIT) or online at and then click on status Inspected by: Date: Work Order Number: UH EHS 4

5 University of Houston FUME HOOD FACE VELOCITY MEASUREMENTS Building: Room Number: Date: Check if after repairs Date: Fume Hood I.D. Number: Type of Fume Hood: Bench hood with horizontal sash Bench hood with sliding doors Walk-in with horizontal sashes Walk-in with sliding doors Other: Air Velocity Measurements for Bench Hood with Horizontal Sash at Working Height Average Air Velocity: ft/min PASS or FAIL (Work Order Number: ) Repairs needed: Low air velocity Low flow alarm doesn t sound Other: Additional Air Velocity Measurements if Needed Average Air Velocity: ft/min Comments: UH EHS 5

Fume Hood Standard. The appropriate maintenance/engineering division on each campus or other business unit operating on UofT property shall:

Fume Hood Standard. The appropriate maintenance/engineering division on each campus or other business unit operating on UofT property shall: Fume Hood Standard A fume hood is an enclosed working chamber fitted with an exhaust ventilation system. It is one component of an exposure control system which is designed to contain, dilute and disperse

More information

Fume Hood Program. Program originated: 08/2008 Updated: 06/2016. Gordon Krueger Environmental Safety Specialist

Fume Hood Program. Program originated: 08/2008 Updated: 06/2016. Gordon Krueger Environmental Safety Specialist Fume Hood Program Program originated: 08/2008 Updated: 06/2016 Gordon Krueger Environmental Safety Specialist 1 TABLE OF CONTENTS I. INTRODUCTION 3 II. LABORATORY VENTILATION 3 A. Chemical Fume Hood Types

More information

---:...-- Environmental Health & Safety RADIATION PROCEDURES MANUAL. Procedure Cover Sheet. Procedure Number: EHS REV 2

---:...-- Environmental Health & Safety RADIATION PROCEDURES MANUAL. Procedure Cover Sheet. Procedure Number: EHS REV 2 Environmental Health & Safety RADIATION PROCEDURES MANUAL Procedure Cover Sheet Procedure Title: Fume Hoods Procedure Number: EHS-08-03-REV 2 Effective Date: 18 September, 2018 ---:...-- Approved By_-=--------::.--...l~---

More information

DEPAUL UNIVERSITY. Fume Hoods Manual. Environmental Health & Safety. April 2017

DEPAUL UNIVERSITY. Fume Hoods Manual. Environmental Health & Safety. April 2017 DEPAUL UNIVERSITY Fume Hoods Manual Environmental Health & Safety April 2017 1 TABLE OF CONTENTS SECTION PAGE NO. 1. INTRODUCTION 2 2. TYPES OF FUME HOODS 2 2.1 CONVENTIONAL FUME HOODS 2 2.2 HIGH PERFORMANCE

More information

January, 2014 Page 1 of 6

January, 2014 Page 1 of 6 Part 1 General 1.1 Summary.1 Unless otherwise indicated, follow the standards below when specifying chemical fume hood modifications or installations. These standards are not intended to restrict or replace

More information

3.5 Exposure Control Measures

3.5 Exposure Control Measures 3.5 Exposure Control Measures 3.5.1 Chemical Fume Hoods Chemical fume hoods are the most reliable and primary engineering control used in the laboratory to protect against the inhalation of hazardous vapors

More information

Design Standard - Fume Hoods & Fume Hood Exhausts

Design Standard - Fume Hoods & Fume Hood Exhausts Design Standard - Fume Hoods & Fume Hood Exhausts 1.0 Introduction The fume hood is the primary control device in most laboratories for protecting employees and students from exposure to hazardous chemicals.

More information

LABORATORY HOOD MANUAL

LABORATORY HOOD MANUAL LABORATORY HOOD MANUAL DRAKE UNIVERSITY ENVIRONMENTAL HEALTH & SAFETY November 2015 TABLE OF CONTENTS I. INTRODUCTION... 1 II. FUME HOOD SERVICE... 2 Fume Hood Service Provided by EH&S... 2 Fume Hood Service

More information

TABLE OF CONTENTS STANDARDS FOR THE DESIGN, CONSTRUCTION, MAINTENANCE AND USE OF LABORATORY FUME HOODS. Effective August 2010

TABLE OF CONTENTS STANDARDS FOR THE DESIGN, CONSTRUCTION, MAINTENANCE AND USE OF LABORATORY FUME HOODS. Effective August 2010 TABLE OF CONTENTS TABLE OF CONTENTS... 1 INTRODUCTION... 2 BACKGROUND... 2 FUME HOOD DESIGN AND CONSTRUCTION... 3 VARIABLE AIR VOLUME FUME HOODS... 3 RECIRCULATING HOODS... 3 SPECIAL LABORATORY FUME HOODS...

More information

UNR Ventilation Policy

UNR Ventilation Policy UNR Ventilation Policy UNR Ventilation Policy TABLE OF CONTENTS SECTION TITLE PAGE 1.0 Scope and Application 1 2.0 Responsibilities 2 2.1 EH&S 2 2.2 Facilities Services Department 3 2.3 End Users 4 3.0

More information

USER'S GUIDE TO FUME HOODS. Procedures for Increasing the Effectiveness of Laboratory Hoods. Revised December, Hood Basics

USER'S GUIDE TO FUME HOODS. Procedures for Increasing the Effectiveness of Laboratory Hoods. Revised December, Hood Basics USER'S GUIDE TO FUME HOODS Procedures for Increasing the Effectiveness of Laboratory Hoods Revised December, 1998 A laboratory fume hood system is designed to protect the operator from undesirable substances

More information

Laboratory Fume Hood. Operation & Maintenance Manual APR 2005

Laboratory Fume Hood. Operation & Maintenance Manual APR 2005 Laboratory Fume Hood Operation & Maintenance Manual APR 2005 TABLE OF CONTENTS I. Introduction p. 3 II. Fume Hood Services p. 4 III. Operation Recommendations p. 4 IV. General Fume Hood Maintenance p.

More information

UNIVERSITY OF ROCHESTER DESIGN STANDARDS JANUARY 2012 SECTION FUME HOODS AND LOCAL EXHAUST SYSTEMS

UNIVERSITY OF ROCHESTER DESIGN STANDARDS JANUARY 2012 SECTION FUME HOODS AND LOCAL EXHAUST SYSTEMS SECTION 15860 - FUME HOODS AND LOCAL EXHAUST SYSTEMS This standard describes fundamental good practices relating to the design, selection, installation, commissioning, and testing of local exhaust systems

More information

CHAPTER 40 PROFESSIONAL LICENSING AND FACILITY REGULATION. PART 13 Radiation Safety Requirements for Particle Accelerators

CHAPTER 40 PROFESSIONAL LICENSING AND FACILITY REGULATION. PART 13 Radiation Safety Requirements for Particle Accelerators 216-RICR-40-20-13 TITLE 216 DEPARTMENT OF HEALTH CHAPTER 40 PROFESSIONAL LICENSING AND FACILITY REGULATION SUBCHAPTER 20 - RADIATION PART 13 Radiation Safety Requirements for Particle Accelerators 13.1

More information

How to Use a Chemical Fume Hood Safely

How to Use a Chemical Fume Hood Safely Published on UC Davis Safety Services (https://safetyservices.ucdavis.edu) How to Use a Chemical Fume Hood Safely SafetyNet #: 35 A chemical fume hood is designed to protect the worker from volatile and

More information

Element D Services Heating Ventilating and Air Conditioning ODG010107

Element D Services Heating Ventilating and Air Conditioning ODG010107 Element D Services Heating Ventilating and Air Conditioning ODG010107 D304202 Laboratory Exhaust and Ventilation Part 1. General 1. This section supplements Design Guideline Element D3042 on air handling

More information

Fume hood construction details. VANGUARD FUME HOODS Listed by Intertek to UL 1805 standard page 1. Features

Fume hood construction details. VANGUARD FUME HOODS Listed by Intertek to UL 1805 standard page 1. Features Fume hood construction details 1 9 2 10 3 11 13 4 14 5 15 6 16 17 7 12 18 8 Features 1 Fluescent light fixture with interi sealed glass offering a safe and bright wking area (100 foot candles) 2 SEFA-8

More information

Fume hood construction details VANGUARD FUME HOODS. Listed by Intertek to UL 1805 standard. page 1. Features

Fume hood construction details VANGUARD FUME HOODS. Listed by Intertek to UL 1805 standard. page 1. Features Fume hood construction details 1 2 3 4 5 6 7 9 10 11 13 14 15 16 17 12 18 8 Features 1 Fluescent light fixture with interi sealed glass offering a safe and bright wking area (100 foot candles) 2 SEFA-8

More information

MOBILE FILTRATION FUME CABINET

MOBILE FILTRATION FUME CABINET MOBILE FILTRATION FUME CABINET MODELS INCLUDE: MFC1000 MFC1000E MFC1000ES WITH MONITORING SYSTEM Thermoline Scientific Equipment Pty. Ltd. T/A Thermoline Scientific ABN 80 000 859 129 10-12 Ross Place.

More information

Lab Safety & Energy Optimization. Introduction & Agenda

Lab Safety & Energy Optimization. Introduction & Agenda Lab Safety & Energy Optimization Determining Appropriate Airflow Specifications for VAV Fume Hoods Thomas C. Smith tcsmith@labhoodpro.com Introduction & Agenda Thomas C. Smith President, BSME (NCSU), MSEE

More information

Work Instructions: Inspecting Forklifts

Work Instructions: Inspecting Forklifts Work Instructions: Inspecting Forklifts Overview Purpose To explain the procedure for inspecting forklifts which is designed to: Ensure forklifts are working properly Comply with OSHA regulations Scope

More information

PART I RADIATION SAFETY REQUIREMENTS FOR PARTICLE ACCELERATORS

PART I RADIATION SAFETY REQUIREMENTS FOR PARTICLE ACCELERATORS PART I RADIATION SAFETY REQUIREMENTS FOR PARTICLE ACCELERATORS Sec. I.1 Purpose and Scope a. This Part establishes procedures for the registration [or licensing] and the use of particle accelerators. b.

More information

DIAGNOSTIC TESTING USING ASHRAE 110. Gerhard Knutson, Ph.D., CIH Knutson Ventilation, Inc

DIAGNOSTIC TESTING USING ASHRAE 110. Gerhard Knutson, Ph.D., CIH Knutson Ventilation, Inc DIAGNOSTIC TESTING USING ASHRAE 110 Gerhard Knutson, Ph.D., CIH Knutson Ventilation, Inc. 952-928-0195 ASHARE 110 A Method of Testing Three components Face velocity measurements Smoke visualization Tracer

More information

NOTE: Hard copies of this document may not be the current version. Refer to the IAmTheKey to verify the current version.

NOTE: Hard copies of this document may not be the current version. Refer to the IAmTheKey to verify the current version. 6 - FUME HOOD 1 MERCURY ACCESS Title: Fume Hood 1 Mercury Access Doc. No. 2016-MMTS-6 NOTE: This document will be reviewed at least annually to ensure its suitability. Revision History Rev. No. Change

More information

GUIDELINE FOR GLOVEBOXES THIRD EDITION AGS-G001 February 2007

GUIDELINE FOR GLOVEBOXES THIRD EDITION AGS-G001 February 2007 GUIDELINE FOR GLOVEBOXES THIRD EDITION AGS-G001 February 2007 AMERICAN GLOVEBOX SOCIETY STANDARDS DEVELOPMENT COMMITTEE All Rights Reserved Copyright 2007 No portion may be duplicated without written consent

More information

STATE UNIVERSITY CONSTRUCTION FUND

STATE UNIVERSITY CONSTRUCTION FUND DIRECTIVE 15H-8 Issue date: June 2016 1. General Design Criteria LABORATORY DESIGN a. Design a Pressure Independent Variable Volume system that satisfies the airflow demand of the fume hoods, minimum air

More information

INSTALLATION AND OPERATIONS MANUAL FUEL POLISHING MODULE (FPM)

INSTALLATION AND OPERATIONS MANUAL FUEL POLISHING MODULE (FPM) INSTALLATION AND OPERATIONS MANUAL FUEL POLISHING MODULE (FPM) Models 60M, 90M, 180M DIESEL ENGINES ONLY Don t Leave Shore Without It. IMPORTANT READ BEFORE INSTALLING THE INSTALLATION AND SETUP SHOULD

More information

Service Action Campaign Code: 20T8

Service Action Campaign Code: 20T8 IMPORTANT INFORMATION Please inform and provide a copy of this document to every person in your dealership with campaign-related responsibilities, including Service, Parts and Accounting personnel. By

More information

Element D Services Heating, Ventilating, and Air Conditioning

Element D Services Heating, Ventilating, and Air Conditioning PART 1 - GENERAL 1.01 OVERVIEW A. This section supplements Design Guideline Element D3042 on exhaust and ventilation with specific criteria for projects involving design of Laboratory Exhaust and Systems.

More information

Technical Standard for Radiological Air Cleaning and Ventilation Systems at Jefferson Lab

Technical Standard for Radiological Air Cleaning and Ventilation Systems at Jefferson Lab Technical Standard for Radiological Air Cleaning and Ventilation Systems at Jefferson Lab Thomas Jefferson National Accelerator Facility 12000 Jefferson Ave Newport News, VA 23606 January, 2007 Erik Abkemeier

More information

Model 8340 Intrinsically Safe VELOCICHECK Air Velocity Meter

Model 8340 Intrinsically Safe VELOCICHECK Air Velocity Meter Ventilation Testing/Balancing Model 8340 Intrinsically Safe VELOCICHECK Air Velocity Meter Operation and Service Manual 1980076, Revision E October 2002 Model 8340 Intrinsically Safe VELOCICHECK Air Velocity

More information

Element D Services Heating, Ventilating, and Air Conditioning

Element D Services Heating, Ventilating, and Air Conditioning Laboratory Exhaust and PART 1 - GENERAL 1.01 OVERVIEW A. This section supplements Design Guideline Element D3042 on exhaust and ventilation with additional criteria for projects involving design of laboratory

More information

SLIM LINE FUME HOOD MONITOR

SLIM LINE FUME HOOD MONITOR SLIM LINE FUME HOOD MONITOR Phoenix Controls FHM420 Fume Hood Monitor is sized to fit any fume hood and includes icons for simple and universal operation. The FHM420 slim line monitor works in conjunction

More information

Air Sentry High Performance Fume Hood Operation & Maintenance Manual

Air Sentry High Performance Fume Hood Operation & Maintenance Manual Air Sentry High Performance Fume Hood Operation & Maintenance Manual Revision 8 TABLE OF CONTENTS I. Introduction p. 3 II. High Performance Fume Hoods p. 4 III. Vortex Control System p. 5 IV. Uncrating

More information

Safety Recall Code: 46G4

Safety Recall Code: 46G4 Safety Recall Code: 46G4 Subject Brake Pedal Clip Release Date April 5, 2016 Affected Vehicles U.S.A. & CANADA: 2011-2016 Model Year Volkswagen Touareg Vehicles Check Campaigns/Actions screen in Elsa on

More information

BIOLOGICAL SAFETY CABINET

BIOLOGICAL SAFETY CABINET BIOLOGICAL SAFETY CABINET BK3-4 and BK3-6 SERIES Installation and Operations Manual BIOLOGICAL SAFETY CABINET Installation and Operations Manual Revision: 29-Aug-2018 Page 2 of 34 MODELS COVERED BK3-4

More information

CHIP & HOG TRUCK DRIVER EXAM

CHIP & HOG TRUCK DRIVER EXAM CHIP & HOG TRUCK DRIVER EXAM REVISED JULY 2016 Al-Pac Driver I.D. # : PLASTIC IS DRASTIC! First Name : Last Name : Address : City & Province : Postal Code : Phone Number - Main : Phone Number - Alternative

More information

U-Factor, Solar Heat Gain Coefficient, Visible Transmittance and Condensation Resistance Calculation Report

U-Factor, Solar Heat Gain Coefficient, Visible Transmittance and Condensation Resistance Calculation Report U-Factor, Solar Heat Gain Coefficient, Visible Transmittance and Condensation Resistance Calculation Report REPORT NO: NCTL-110-16086-01 SIMULATION DATE: 08/30/13 REPORT DATE: 08/30/13 Client: Product

More information

Laboratory Exhaust Systems

Laboratory Exhaust Systems Laboratory Exhaust Systems Vektor -HS, Vektor -MS and Vektor -CS with Variable Geometry Nozzle September 017 Vektor -HS, Vektor -MS and Vektor -CS Greenheck s Variable Geometry Nozzle exhaust systems allow

More information

HILLSBOROUGH COUNTY SHERIFF S OFFICE DAVID GEE, SHERIFF STANDARD OPERATING PROCEDURE

HILLSBOROUGH COUNTY SHERIFF S OFFICE DAVID GEE, SHERIFF STANDARD OPERATING PROCEDURE Date: 07/28/90 Revision: 11/15/05 Reviewed: 11/15/05 HILLSBOROUGH COUNTY SHERIFF S OFFICE DAVID GEE, SHERIFF STANDARD OPERATING PROCEDURE Number: DTN 917.06 Page: 1 of 7 SUBJECT: TESTING FOR BREATH ALCOHOL

More information

PAD-MOUNTED SWITCHGEAR INSPECTION & MAINTENANCE RECOMMENDATIONS TYPE PSI/II. 15kV 25kV. Qualified Persons

PAD-MOUNTED SWITCHGEAR INSPECTION & MAINTENANCE RECOMMENDATIONS TYPE PSI/II. 15kV 25kV. Qualified Persons Page 1 Qualified Persons... 1 Safety Information... 2 Procedures... 3 Maintenance... 3 Security Inspection (Exterior)... 3 Verify Securing Devices... 3 Maintaining the Exterior... 3 Replacing Labels...

More information

AUDIO-VISUAL EQUIPMENT

AUDIO-VISUAL EQUIPMENT 11 52 00 AUDIO-VISUAL EQUIPMENT 1. GENERAL A. Related sections: i. 00 00 13 Designing Learning Environments ii. 11 52 13 Projection Screens iii. 12 56 52 Audio-Visual Equipment iv. 72 41 00 General Audio-Visual

More information

Proposed Addendum ag to Standard , Ventilation for Acceptable Indoor Air Quality

Proposed Addendum ag to Standard , Ventilation for Acceptable Indoor Air Quality BSR/ASHRAE Addendum ag to ANSI/ASHRAE Standard 62.1-2016 Public Review Draft Proposed Addendum ag to Standard 62.1-2016, Ventilation for Acceptable Indoor Air Quality First Public Review (January 2019)

More information

Polypropylene Total Exhaust Fume Hoods

Polypropylene Total Exhaust Fume Hoods Polypropylene Total Exhaust Fume Hoods 24 36 48 Provides Comprehensive Operator Protection and Optimum Performance 60 72 96 PTEFH-48 shown on optional vented base cabinet Provides Personnel Protection

More information

Laboratory Fume Hood

Laboratory Fume Hood Attention! Important safety instructions. Read and understand the manual. Laboratory Fume Hood User and Installation Guide Mott Manufacturing Limited PO Box 1120 Brantford, Ontario N3T 5T3 Canada THIS

More information

FUME HOOD SELECTION GUIDE

FUME HOOD SELECTION GUIDE FUME HOOD SELECTION GUIDE FUME HOOD SELECTION GUIDE Laboratory fume hoods are one of the most vital pieces of safety equipment in a working laboratory. Fume hoods are designed to safely exhaust dangerous

More information

ZONE PRESENCE SENSOR FEATURES SPECIFICATIONS TABLE OF CONTENTS. VAV system using a Zone Presence Sensor.

ZONE PRESENCE SENSOR FEATURES SPECIFICATIONS TABLE OF CONTENTS. VAV system using a Zone Presence Sensor. ZONE PRESENCE SENSOR Phoenix Controls Zone Presence Sensor (ZPS ) detects the presence of personnel in front of a fume hood. When an operator is present, the ZPS signals the fume hood system into the standard,

More information

PRECAUTIONS To ensure personal safety and proper maintenance, please read this section. CAUTION

PRECAUTIONS To ensure personal safety and proper maintenance, please read this section. CAUTION Thank you very much for choosing an EIZO product. Please read this User s Manual carefully to familiarize you with safe and effective usage procedures. Please retain this manual for future reference. PRECAUTIONS

More information

RADIATION CONTROL - RADIATION SAFETY REQUIREMENTS FOR PARTICLE ACCELERATORS NOT USED IN THE HEALING ARTS

RADIATION CONTROL - RADIATION SAFETY REQUIREMENTS FOR PARTICLE ACCELERATORS NOT USED IN THE HEALING ARTS DEPARTMENT OF PUBLIC HEALTH AND ENVIRONMENT RADIATION CONTROL - RADIATION SAFETY REQUIREMENTS FOR PARTICLE ACCELERATORS NOT USED IN THE HEALING ARTS 6 CCR 1007-1 Part 09 [Editor s Notes follow the text

More information

BIOLOGICAL SAFETY CABINET

BIOLOGICAL SAFETY CABINET BIOLOGICAL SAFETY CABINET BK3 SERIES Installation and Operations Manual BIOLOGICAL SAFETY CABINET Installation and Operations Manual Revision: 17-May-16 Page 2 of 31 MODELS COVERED BK-3-4 Class II Type

More information

UniTrak Sensor Assembly Kit Horizontal, Vertical, and Combination Sashes

UniTrak Sensor Assembly Kit Horizontal, Vertical, and Combination Sashes Document No. 546-00449 UniTrak Sensor Assembly Kit Horizontal, Vertical, and Combination Sashes Product Description The UniTrak Sensor Assembly Kit contains the hardware to complete one fume hood sensor

More information

The IFSQN ISO Implementation Package Start Up Guide

The IFSQN ISO Implementation Package Start Up Guide Welcome to the IFSQN ISO 22000 Food Safety Management System Package Start Up Guide which will guide you through the contents of the package. The IFSQN ISO 22000 2018 Implementation Package includes: Food

More information

Extraction Arms. Capturing fumes at the source

Extraction Arms. Capturing fumes at the source Extraction Arms Capturing fumes at the source Mm-75 Mm-100 EA Ideal for Bench work and Smaller Working Areas The Miniman 75 mm is ideal for all types of smaller dust and fume applications. Totally free

More information

Fume cupboards and extraction devices

Fume cupboards and extraction devices Fume cupboards and extraction devices Energy efficiency, maximum ergonomics and a larger internal workspace make working with our new fume cupboards even safer and more convenient. A new design together

More information

Healthcare Solutions Room Pressure Monitors and Controls

Healthcare Solutions Room Pressure Monitors and Controls Healthcare Solutions Room Pressure Monitors and Controls TOUCHSCREEN ROOM PRESSURE MONITOR The Touchscreen Room Pressure Monitor (PMT) provides precise room pressure monitoring for critical spaces such

More information

Rotating Vane Anemometer

Rotating Vane Anemometer Ventilation Testing Rotating Vane Anemometer Alnor Model RVA801 AIRFLOW Model LCA301 Operation and Service Manual Copyright TSI Incorporated / 2007-2012 / All rights reserved. Address TSI Incorporated

More information

VEHICLE MAINTENANCE PROGRAM

VEHICLE MAINTENANCE PROGRAM VEHICLE MAINTENANCE PROGRAM Prepared by: SCOTT PARTRIDGE Transportation Manager Version: 5.05.14 Effective Date: 5 MAY 2014 The undersigned has read, understands and agrees to comply with the information

More information

BIOLOGICAL SAFETY CABINET

BIOLOGICAL SAFETY CABINET BIOLOGICAL SAFETY CABINET BK3 SERIES Installation and Operations Manual BIOLOGICAL SAFETY CABINET Installation and Operations Manual Revision: 08-Aug-16 Page 2 of 32 MODELS COVERED BK-3-4 Class II Type

More information

RAILWAY FREIGHT AND PASSENGER TRAIN BRAKE INSPECTION AND SAFETY RULES

RAILWAY FREIGHT AND PASSENGER TRAIN BRAKE INSPECTION AND SAFETY RULES RAILWAY FREIGHT AND PASSENGER TRAIN BRAKE INSPECTION AND SAFETY RULES TC O 0-95 Approved July 10, 2008 Effective October 1, 2008 Contents PART I GENERAL 3 1 Short Title 3 2 Scope 3 3 Definitions 3 4 Certification

More information

Passenger Train Brake Inspection and Safety Rules: Guidelines for British Columbia s Heritage Railways

Passenger Train Brake Inspection and Safety Rules: Guidelines for British Columbia s Heritage Railways Passenger Train Brake Inspection and Safety Rules: Guidelines for British Columbia s Heritage Railways Part I: General 1. SHORT TITLE 1.1 For ease of reference, these rules may be referred to as the "Train

More information

Rig Master Power by Mobile Thermo Systems Inc.

Rig Master Power by Mobile Thermo Systems Inc. RigMaster Power Dealer Warranty Policy The Limited Warranty This limited warranty applies to the RigMaster Auxiliary Power Unit (RigMaster APU) which consists of the following components: 1. The generator

More information

FLÄKTGROUP PM-MOTOR WITH INTEGRATED FC 106 FREQUENCY CONVERTER

FLÄKTGROUP PM-MOTOR WITH INTEGRATED FC 106 FREQUENCY CONVERTER FLÄKTGROUP PM-MOTOR WITH INTEGRATED FC 106 FREQUENCY CONVERTER INSTALLATION AND MAINTENANCE INSTRUCTIONS Risk of electric shock: Motor terminals may still be live if the impeller is rotating, even when

More information

Appendix C: Biosafety Cabinet Specifications C-1. Product Service

Appendix C: Biosafety Cabinet Specifications C-1. Product Service Appendix C: Biosafety Cabinet Specifications C-1 Product Service 1-800-522-7658 111 Appendix C: Biosafety Cabinet Specifications C-2 112 Product Service 1-800-522-7658 Appendix C: Biosafety Cabinet Specifications

More information

Maintenance Information

Maintenance Information (Dwg. MHP1539) 16600470 Edition 1 May 2007 Z Rail Aluminum and Steel Overhead Rail System Maintenance Information Save these Instructions The minimum maintenance required for a rail system requires inspection

More information

BS EN :2011 BS :2002+A1:2013 UKSRG Guidelines-4:2011 EN124:1994 HA104\09

BS EN :2011 BS :2002+A1:2013 UKSRG Guidelines-4:2011 EN124:1994 HA104\09 PendulumTest(s) carried out in accordance with: BS EN 13036-4:2011 BS 7976-2:2002+A1:2013 UKSRG Guidelines-4:2011 EN124:1994 HA104\09 On behalf of: Fibrelite Limited. Prepared by: Phil Nolan GET-A-GRIP!

More information

RECALL - ECM UPDATE FOR CARBON MONOXIDE EMISSIONS

RECALL - ECM UPDATE FOR CARBON MONOXIDE EMISSIONS RECALL - ECM UPDATE FOR CARBON MONOXIDE EMISSIONS # 14535: Product Emission - Carbon Monoxide Emissions - (Mar 3, 2015) Subject: 14535 - Carbon Monoxide Emissions Models: 2010 Chevrolet Equinox 2010 GMC

More information

Recall - Parking Brake Lever Assembly

Recall - Parking Brake Lever Assembly 1991 Buick Park Avenue V6-231 3.8L VIN L SFI Copyright 2013, ALLDATA 10.52 Page 1 Technical Service Bulletin # 92C25 Date: 911001 Recall - Parking Brake Lever Assembly Model Year: 1992 Bulletin No.: 92-C-2

More information

AUDIO-VISUAL EQUIPMENT

AUDIO-VISUAL EQUIPMENT 11 52 00 AUDIO-VISUAL EQUIPMENT 1. GENERAL A. Related sections: i. 00 00 13 Designing Learning Environments ii. 11 52 13 Projection Screens iii. 12 56 52 Audio-Visual Equipment iv. 72 41 00 General Audio-Visual

More information

Driving for Work. Managing Risks. Deirdre Sinnott Senior Inspector Work Related Vehicle Safety Program

Driving for Work. Managing Risks. Deirdre Sinnott Senior Inspector Work Related Vehicle Safety Program Driving for Work Managing Risks Deirdre Sinnott Senior Inspector Work Related Vehicle Safety Program Key Risk Management steps Do I have responsibilities? Business risk inventory Pay attention to Big Data

More information

Rotating Vane Anemometer

Rotating Vane Anemometer Ventilation Testing Rotating Vane Anemometer Alnor Model RVA801 AIRFLOW Model LCA301 Operation and Service Manual Copyright TSI Incorporated / 2007-2008 / All rights reserved. Address TSI Incorporated

More information

4.1 Functional diagram FM100 fume hood monitor... 4 Functional description FM Interfaces Terminal diagram FM100...

4.1 Functional diagram FM100 fume hood monitor... 4 Functional description FM Interfaces Terminal diagram FM100... 2.0 LabSystem Fume hood monitors Table of contents Section Title Page 1.1 Introduction..................................... 2 2.1 Functional description............................. 2 2.1.1 Actual value

More information

UWA Electrical Safety Procedure

UWA Electrical Safety Procedure UWA Electrical Safety Procedure Who it applies to The requirements of WA legislation are applicable to academic personnel (e.g. lecturers and researchers), students of various levels, personnel that support

More information

Fume Hood Operation & Maintenance Manual

Fume Hood Operation & Maintenance Manual Fume Hood Operation & Maintenance Manual Fume Hood Operation & Maintenance Manual TABLE OF CONTENTS Fume Hood Applications 2 Safety 3 Aerodynamic Principles & Safety 4 Fume Hood Cleaning 7 Basic Maintenance

More information

CUSTOMER - FREQUENTLY ASKED QUESTIONS

CUSTOMER - FREQUENTLY ASKED QUESTIONS CUSTOMER - FREQUENTLY ASKED QUESTIONS Version 1 EROAD ELD Do you allow yard moves and personal conveyance and how does this work? Yes, the EROAD ELD allows yard moves (ON YM) and personal conveyance (OFF

More information

Aspen 200 Powder Safety Workstation. Operation Manual

Aspen 200 Powder Safety Workstation. Operation Manual Aspen 200 Powder Safety Workstation Operation Manual Table of Contents Disclaimer...............................................1 Specifications............................................1 Getting Started

More information

Application. Tek-Air Systems, Inc. 41 Eagle Road Danbury, CT (203) FAX: (203) SALES FAX: (203)

Application. Tek-Air Systems, Inc. 41 Eagle Road Danbury, CT (203) FAX: (203) SALES FAX: (203) TEK-AIR TECHNICAL PRODUCT DATA SHEET FVC2600 SASH SENSING VAV FUME HOOD SYSTEM MODEL 2600: Variable volume control based on sash position utilizes microprocessor based electronics and linear air control

More information

The Participant will select one of the two Approval Categories (below) and submit the appropriate documentation.

The Participant will select one of the two Approval Categories (below) and submit the appropriate documentation. FORM WCCL-PC3 WCCL TEST STAND APPROVAL PROCEDURE AND CHECKLIST AHRI CERTIFICATION PROGRAM FOR WATER-COOLED WATER CHILLING AND HEAT PUMP WATER-HEATING (WCCL) PACKAGES WCCL Participant: Date of Inspection:

More information

U-Factor, Solar Heat Gain Coefficient, Visible Transmittance and Condensation Resistance Calculation Report

U-Factor, Solar Heat Gain Coefficient, Visible Transmittance and Condensation Resistance Calculation Report U-Factor, Solar Heat Gain Coefficient, Visible Transmittance and Condensation Resistance Calculation Report REPORT NO: NCTL-110-14975-01 SIMULATION DATE: 03/25/13 REPORT DATE: 03/25/13 Client: Product

More information

V05 General Purpose Bench Fume Hood

V05 General Purpose Bench Fume Hood V05 General Purpose Bench Fume Hood with Vertical Rising ash Accessories Included: 4 120 VAC 20 amp GFCI protected duplex receptacles 1 LED light fixture with illumination and color controller 1 Electronic

More information

Ventilated Balance Enclosures

Ventilated Balance Enclosures Ventilated Balance Enclosures by PLAS n LABS Meets and exceeds ANSI/ASHRAE 110:2016 standard for containment & performance PLAS n LABS, INC. www.plas-labs.com PLI@PLAS-LABS.com 517-372-7177 Ventilated

More information

Internal Audit Report. Fuel Consumption Oversight and Coordination TxDOT Internal Audit Division

Internal Audit Report. Fuel Consumption Oversight and Coordination TxDOT Internal Audit Division Internal Audit Report Fuel Consumption Oversight and Coordination TxDOT Internal Audit Division Objective To determine if a process exists to ensure retail fuel consumption is appropriately managed and

More information

Thermo Scientific Hamilton Horizon Fume Hoods. Operation, Maintenance and Installation Instructions

Thermo Scientific Hamilton Horizon Fume Hoods. Operation, Maintenance and Installation Instructions Thermo Scientific Hamilton Horizon Fume Hoods Operation, Maintenance and Installation Instructions Thermo Scientific Horizon Fume Hood Table of Contents Introduction.............................................................

More information

ANNEXES. to the. Proposal for a Regulation of the European Parliament and of the Council

ANNEXES. to the. Proposal for a Regulation of the European Parliament and of the Council EUROPEAN COMMISSION Brussels, 17.5.2018 COM(2018) 296 final ANNEXES 1 to 8 ANNEXES to the Proposal for a Regulation of the European Parliament and of the Council on the labelling of tyres with respect

More information

SAMPLING SAMPLING Model: Product Data Sheet Pico Vibe. 4mm Vibration Motor - 7mm Type

SAMPLING SAMPLING Model: Product Data Sheet Pico Vibe. 4mm Vibration Motor - 7mm Type - Target Specification for Prototype Samples - Product Data Sheet Pico Vibe mm Vibration Motor - 7mm Type mm Vibration Motor - 7mm Type Shown on 6mm Isometric Grid Model: Ordering Information Key Features

More information

Urgent Field Safety Notice

Urgent Field Safety Notice December X, 2017 Urgent Field Safety Notice Attention: Directors of Respiratory Care, Inpatient Critical Care Units, and Risk Management Dear Valued Customer: The purpose of this letter is to advise you

More information

Car Park Label Scheme for Persons with Physical Disabilities

Car Park Label Scheme for Persons with Physical Disabilities Car Park Label Scheme for Persons with Physical Disabilities Applying for a Class 1 or Class 2 label 1. If I am a motorcyclist, can I apply for a Label? The Car Park Label Scheme is only applicable to

More information

A Study of Factors Affecting Fume Hood Energy Consumption

A Study of Factors Affecting Fume Hood Energy Consumption A Study of Factors Affecting Fume Hood Energy Consumption Joseph S Marian, Jr. American Auto-Matrix, One Technology Lane, Export, Pennsylvania 15632, USA Contents Contents... 2 1. Introduction... 4 1.1

More information

EXT HEPA/CARBON FUME HOOD

EXT HEPA/CARBON FUME HOOD EXT HEPA/CARBON FUME HOOD POLYPROPYLENE Installation and Operations Manual EXT HEPA/CARBON FUME HOOD Installation and Operations Manual Revision: 28-Jun-18 Page 2 of 12 EXT HEPA/CARBON FUME HOOD Installation

More information

South Tacoma Groundwater Protection District Spill Prevention and Response Plan

South Tacoma Groundwater Protection District Spill Prevention and Response Plan Name of Business Address Facility Phone ( ) - Types of Work or Hazardous Substances Used This spill plan is designed to handle the requirements for this system and associated hazardous substances. Update

More information

ONSITE WASTEWATER TREATMENT SYSTEM Property Transfer Certification Instructions and Inspection Report

ONSITE WASTEWATER TREATMENT SYSTEM Property Transfer Certification Instructions and Inspection Report ONSITE WASTEWATER TREATMENT SYSTEM Property Transfer Certification Instructions and Inspection Report These instructions are for homeowners and inspectors completing the onsite wastewater treatment system

More information

Design Requirements The system shall be designed, manufactured, tested and installed in compliance with the current versions or issuance of:

Design Requirements The system shall be designed, manufactured, tested and installed in compliance with the current versions or issuance of: SECTION 15877 FUME HOODS 1. General 1.1. Summary These specifications provide a performance specification for a chemical laboratory Fume Hood (herein referred to as FH). The intent of the specification

More information

13142B Loss of Battery Charge Replace Generator Control Module

13142B Loss of Battery Charge Replace Generator Control Module 13142B Loss of Battery Charge Replace Generator Control Module 2012-2013 Buick LaCrosse, Regal 2013 Chevrolet Malibu Eco Equipped with eassist The Part Inspection Procedure in this bulletin has been revised.

More information

FVC2100/2200 TEK-AIR TECHNICAL PRODUCT DATA SHEET FUME HOOD FACE VELOCITY MONITOR AND CONTROLLER. Application. General Description

FVC2100/2200 TEK-AIR TECHNICAL PRODUCT DATA SHEET FUME HOOD FACE VELOCITY MONITOR AND CONTROLLER. Application. General Description TEK-AIR TECHNICAL PRODUCT DATA SHEET FVC2100/2200 FUME HOOD FACE VELOCITY MONITOR AND CONTROLLER MODEL 2100: Constant face velocity, variable volume MODEL 2200: Constant face velocity, variable volume

More information

*** DRAFT *** Lubricant Service Designations for Automotive Manual Transmissions, Manual Transaxles, and Axles

*** DRAFT *** Lubricant Service Designations for Automotive Manual Transmissions, Manual Transaxles, and Axles *** DRAFT *** Lubricant Service Designations for Automotive Manual Transmissions, Manual Transaxles, and Axles API PUBLICATION 1560 EIGHTH EDITION, 2011 American Petroleum Institute 1220 L Street, Northwest

More information

SafeFAST Top. Class II Microbiological Safety Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE.

SafeFAST Top. Class II Microbiological Safety Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE. Class II Microbiological Safety Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE. Class II Microbiological Safety Cabinets SAFETY CABINETS WITH AUTOMATIC REGULATION AND MICROPROCESSOR BASED MONITORING

More information

Recordkeeping Requirements of the Federal Motor Carrier Safety Regulations

Recordkeeping Requirements of the Federal Motor Carrier Safety Regulations Recordkeeping Requirements of the Federal Motor Carrier Safety Regulations The following table summarizes the recordkeeping requirements of the Federal Motor Carrier Safety Regulations under 49 CFR Parts

More information

UNITED STATES DEPARTMENT OF LABOR MINE SAFETY AND HEALTH ADMINISTRATION Metal and Nonmetal Mine Safety and Health REPORT OF INVESTIGATION

UNITED STATES DEPARTMENT OF LABOR MINE SAFETY AND HEALTH ADMINISTRATION Metal and Nonmetal Mine Safety and Health REPORT OF INVESTIGATION MAI-2009-15 UNITED STATES DEPARTMENT OF LABOR MINE SAFETY AND HEALTH ADMINISTRATION Metal and Nonmetal Mine Safety and Health REPORT OF INVESTIGATION Surface Nonmetal Mine (Construction Sand and Gravel)

More information

National Animal Supplement Council

National Animal Supplement Council NASC established a Preferred Supplier Program for companies who wish to provide products and services to NASC member companies, including raw materials (ingredients), contract manufacturing (dosage form

More information

Operation of Fork Lift Trucks

Operation of Fork Lift Trucks Operation of Fork Lift Trucks 1. PURPOSE The purpose of this Procedure is to define the rules for the safe operation of Forklift trucks within the University. 2. SCOPE These rules apply to all rider-type

More information

Installation instruction MicaFlex fume hood control. Fume hood control. Innehållsförteckning. Mi-212gb_

Installation instruction MicaFlex fume hood control. Fume hood control. Innehållsförteckning. Mi-212gb_ MicaFlex Flow 0.5 m/s ormal ormal Caution Alarm TEST ORMAL HIGH LOW FLOW FLOW FLOW RESET t s PGM ESC Keypad functions in programming mode Operator Monitor FHM catrone Installation instruction MicaFlex

More information