Analysis of nitrogen isotope ratios in amino acids by GC/C/IRMS
|
|
- Lindsay Richard
- 6 years ago
- Views:
Transcription
1 Analysis of nitrogen isotope ratios in amino acids by GC/C/IRMS 27 March 2013 Yoshito Chikaraishi The initial version was released by YC in May The present version reflects feedbacks from Svenja Kruse (Univ. of British Columbia) and Shawn A. Steffan (Univ. of Wisconsin-Madison), Elizabeth Bosak (Univ. of Wisconsin-Madison) in March 2012.
2 Instrument Instruments in my lab: GC Agilent 6890N Injector Gerstel PTV Column Agilent HP Ultra-2 (50m, 0.32mm, 0.52µm) Flow rate (He) 1.4ml/min Oven program 40ºC (2.5 min) -15ºC/min (0 min) - 3ºC/min (0 min) - 6ºC/min - 220ºC (17.3min) Interface GC combustion III (oxidation at 1000ºC, reduction at 550ºC) IRMS Delta plus XP Standard mix of authentic amino acids: after HCl hydrolysis and derivatization, derivatives are stored at 20ºC, in DCM (Currently, I use new standard mix. See page 10 ). Compound Source Known Amino acid Pv/OiPr derivatives δ 15 N MW N(wt%) mg/10ml MW N(wt%) ng/µl ngn/µl nmol/µl L-alanine Ala CERKU glycine Gly CERKU L-valine Val BioGeos L-leucine Leu CERKU L-norleucine Ile BioGeos L-aspartic acid Asp BioGeos L-serine Ser BioGeos L-glutamic acid Glu BioGeos L-phenylalanine Phe BioGeos L-hydroxyproline Hyp BioGeos
3 Procedure for δ 15 N analysis 1. Set up GC/IRMS (e.g., connect GC and IRMS, focusing) 2. Check N 2 leaks for both back-flush and straight mode 3. Check stability by On/OFF of Ref. gas 4. Check reproducibility by measuring δ 15 N-known authentic AAs (amino acid standards), as Pv/OiPr derivatives 5. Make up a sequence file 6. Measure both authentic AAs and samples of interest 7. Check baseline separation on the chromatograms 8. Normalize the measured values (i.e., δ 15 N, vs ref.gas) to international scale (i.e., δ 15 N, vs Air)
4 1. Focusing Check these 9 boxes To find the best (optimum) position, apply the following steps: 1. After peak center, open auto focus. 2. Enter 50 and 500 for the step width in these two boxes 3. Click OK 4. After peak center, enter 25 and 250 in these two boxes, and click OK 5. After peak center, enter 12 and 120 in these two boxes, and click OK 6. After peak center, 6 and 60 in these two boxes, and click OK 7. After peak center, enter 3 and 30 in these two boxes, and click OK 8. After peak center, enter 2 and 20 in these two boxes, and click OK 9. After peak center, enter 1 and 10 in these to boxes, and click OK (twice)
5 3. Check stability by On/OFF of Ref. gas Generally, 1σ <
6 4. Check reproducibility by δ 15 N of known authentic AAs Set the isotope value of the ref. gas to 0 on the PC software, measure the authentic AA derivatives, and check the linearity of the relationship between the measured δ 15 N (, vs ref. gas) and known δ 15 N (, vs air) values Ref. N 2 Ref. N 2 m/z 28 Intensity (mv) Ala Gly Val Leu Nle Asp Ser Glu Phe Hyp Retention time (min) 25 If you do not get a good linear relationship, please check the following points: 1. the condition of the ox./red. furnace 2. the condition of GC column 3. dirt in inlet/liner δ 15 N, vs Air y = x (R 2 = 0.999) 4. deterioration of capillary connection δ 15 N, vs refs
7 5. Make up a sequence file For example, Line GC method Sample Num. Injection vol. (µl) Sample name IRMS Method 1 Ultra-2 AA 1 1 STD Pv/OiPr Ultra-2_AA_15N 2 Ultra-2 AA 2 1 Sample-A Ultra-2_AA_15N 3 Ultra-2 AA 3 1 Sample-B Ultra-2_AA_15N 4 Ultra-2 AA 4 1 Sample-C Ultra-2_AA_15N 5 Ultra-2 AA 5 1 Sample-D Ultra-2_AA_15N 6 Ultra-2 AA 1 1 STD Pv/OiPr Ultra-2_AA_15N 7 Ultra-2 AA 6 1 Sample-E Ultra-2_AA_15N 8 Ultra-2 AA 7 1 Sample-F Ultra-2_AA_15N 9 Ultra-2 AA 8 1 Sample-G Ultra-2_AA_15N 10 Ultra-2 AA 9 1 Sample-H Ultra-2_AA_15N 11 Ultra-2 AA 10 1 Sample-I Ultra-2_AA_15N 12 Ultra-2 AA 1 1 STD Pv/OiPr Ultra-2_AA_15N
8 8. Normalization of measured values to the international scale -1 When you set the isotope value of the ref. gas to 0 on the PC software, you have measured values ( vs ref. gas). Known δ 15 N Measured value = δ 15 N, vs ref. gas 1st 2nd 3rd 4th 5th Average 1σ Ala Gly Val Leu Nle Asp Ser Glu Phe Hyp δ 15 N, vs Air y = x (R 2 = 0.999) δ 15 N, vs refs You can normalize δ 15 N to the international scale (i.e., δ 15 N, vs Air) by using the equation: y = x (in this case). The slope 1.0 means that we need this linear normalization instead of an one-point normalization. Known δ 15 N δ 15 N, vs Air 1st 2nd 3rd 4th 5th Average 1σ Δ Ala Gly Val Leu Nle Asp Ser Glu Phe Hyp Average Std.dev σ = Precision Δ (= average known) = Trueness In this way, Accuracy (= Mean of Δ) should be Zero ± something.
9 8. Normalization of measured values to the international scale -2 With the equation: (δ 15 N, vs Air) = (measured δ 15 N, vs refs) (in this case), you can calculate normalized δ 15 N (, vs Air) for the samples of interest. Measured value = δ 15 N, vs ref. gas Sample-A Sample-B Sample-C Sample-D Sample-E Sample-F Sample-G Ala Gly Val Leu Ile Pro Ser too small Met too small too small too small too small too small Glu Phe δ 15 N, vs Air Sample-A Sample-B Sample-C Sample-D Sample-E Sample-F Sample-G Ala Gly Val Leu Ile Pro Ser 0.9 n.d Met n.d. 3.0 n.d. n.d. n.d. n.d Glu Phe
10 New standard mix We currently developed wide-range (from 25.9 to ) isotopic reference materials of amino acids, which are commercially available from Indiana University ( and SI science ( Compoud Source known δ 15 N Amino acid Pv/OiPr derivatives MW N(wt%) mg/10ml MW N(wt%) ng/µl ngn/µl nmol/µl L-alanine Ala Indiana Univ glycine Gly SI science L-valine Val BioGeos L-leucine Leu SI science L-norleucine Nle SI science L-aspartic acid Asp SI science L-methionine Met SI science L-glutamic acid Glu SI science L-phenylalanine Phe Indiana Univ L-hydroxyproline Hyp SI science Known δ 15 N δ 15 N (, vs Air) by GC/IRMS δ 15 N1st δ 15 N2nd δ 15 N3rd δ 15 N4th Average 1σ Δ Ala Gly Val Leu Nle Asp Ser Glu Phe Hyp Average Std.dev δ 15 N ( vs Air) determined by GC/IRMS y = R 2 = Known δ 15 N ( vs Air)
Molecular Weights and Residue Weights of Protected Amino Acids
Molecular Weights and Residue Weights of Protected Amino Acids Amino Acid Boc Derivative M.W. Residue Wt. in Protected Peptide hain Fmoc Derivative M.W. Residue Wt. in Protected Peptide hain Alanine Boc-Ala
More informationFast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC
Fast GC SPEEDY GC Slide 4 Optimization Goals Primary: Minimize time for the separation of a given number of peaks Secondary: Maximize the number of peaks separated by a given column Slide 5 HOW CAN THESE
More informationStability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas
Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas Liliana Krotz, Walter Galotta, and Guido Giazzi Thermo Fisher Scientific, Milan, Italy Overview
More informationSupporting information
Supporting information Table S1. Tensor parameters and shift of metal ion from starting position for the 10 subsets (from a total of 100) resulting in the largest number of peak pairings, using 2D and
More informationSetting up SilFlow for BackFlush in your GC
Setting up SilFlow for BackFlush in your GC What is backflush and why use it? The BackFlush system eliminates the need to bake heavy sample fractions off the capillary column. Oils, tars and other semivolatile
More informationMethod Development for Capillary GC Systems. Slide 1
Method Development for Capillary GC Systems Slide 1 AREAS TO OPTIMIZE Injector Carrier gas Column temperature Slide 2 COMMON INJECTOR MODES Vaporization Injection Modes Megabore Direct Split Splitless
More informationAgilent 7693A Automated Liquid Sampler
IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7693A Automated Liquid Sampler
More informationMeeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems
Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems Application Note Hydrocarbons Authors Michael Woodman Agilent Technologies, Inc. Chemical Analysis Solutions 2850
More informationAppNote 1/2010 KEYWORDS ABSTRACT. Biodiesel, Automation, ASTM D
AppNote 1/2010 Full Automation of ASTM Method D6584-07 Standard Test Method for the Determination of Free and Total Glycerin in B-100 Biodiesel Methyl Esters by Gas Chromatography using a GERSTEL Dual
More informationPage 1. Alternate Carrier Gas Considerations and Faster GC Analysis
Page 1 Alternate Carrier Gas Considerations and Faster GC Analysis Faster GC Total Analytical Cycle Times A Variety of Approaches Pre-Run ALS Set-Up Chromatographic Run Post-Run Bake-Out Post-Run Cool-Down
More informationGas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593
Application Note Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Authors Kelly Beard and James McCurry Agilent Technologies, Inc. Abstract An Agilent
More information[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS
MS Identification of Trace level Impurities from a Non-MS Compatible Mobile Phase Using ACQUITY UPLC System with 2D Technology by Heart-cutting and Online Sample Concentration Bronsky Gopinadh, Dilshad
More informationAnalysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID
Analysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID Introduction P Gorst-Allman (LECO Africa Pty. Ltd) and B-J de Vos (NMISA). The analysis
More informationTECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure
TECHNICAL REPORT Modifying Agilent 1100 HPLC Systems to Achieve UHPLC-like performance with HALO Fused-Core Columns Agilent 1100 HPLC system With a few modifications, an Agilent 1100 HPLC can produce UHPLC-like
More informationAlternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis
Introduction Petroleum & Petrochemical Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis By Katarina Oden, Barry Burger, and Amanda Rigdon Crude oil consists of thousands of different
More informationGLYCERIN PACKAGE 100mL $ Glycerin, Ibs/gallon, MONG Material Organic, Non-Glycerin, Ash, Specific Gravity, Moisture by Karl Fisher, Methanol
FUEL BIODIESEL SUGGESTED SAMPLE SIZE (MINIMUM) FEE GLYCERIN PACKAGE 100mL $350.00 Glycerin, Ibs/gallon, MONG Material Organic, Non-Glycerin, Ash, Specific Gravity, Moisture by Karl Fisher, Methanol ETHANOL
More informationGAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS
GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS FLASH VAPORISATION INJECTION Split Splitless On-Column COOL INJECTION Large Volume Injection (LVI) On-Column On-Column-SVE (with solvent vapour
More informationHigh Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph
Application Note Petrochemicas High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Author James D. McCurry, Ph.D. Agilent Technologies, Inc. Abstract An Agilent
More informationFast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application
Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application Petroleum Authors ChunXiao Wang Agilent Technologies (Shanghai) Co.,Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai
More informationHigh-Temperature Simulated Distillation System Based on the 6890N GC Application
High-Temperature Simulated Distillation System Based on the 6890N GC Application Petroleum Authors ChunXiao Wang Agilent Technologies (Shanghai) Co., Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai
More informationApplication Note. Abstract. Authors. Environmental Analysis
High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID using the Agilent Low Thermal Mass (LTM II) System Application Note Environmental Analysis Authors Frank David and Karine Jacq Research
More informationAgilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011
Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc.
More informationGC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585
GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc. 850 Centerville
More informationMET-Biodiesel Capillary GC Columns
MET-Biodiesel Capillary GC Columns Product Specifications Product Features & Benefits Chromatograms FAQs Related Products Updated: February 2, 2009 Product Specifications 2 Product Specifications What
More informationTechnical Procedure for Gas Chromatography (GC-FID)
Technical Procedure for Gas Chromatography (GC-FID) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph (GC- FID). 2.0 Scope This procedure applies to all
More informationAppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS
AppNote 6/26 Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) Andreas Hoffmann GERSTEL GmbH & Co.KG, Eberhard-Gerstel-Platz 1,
More informationURB '-carbamoylbiphenyl-3-yl cyclohexylcarbamate. DEA Reference Material Collection. Form Chemical Formula Molecular Weight Melting Point ( o C)
O NH O NH O. GENERAL INFORMATION IUPAC Name: 3'-carbamoylbiphenyl-3-yl cyclohexylcarbamate CAS#: 56-08-6 Synonyms: Source: Appearance: UV max (nm): KDS-03 DEA Reference Material Collection White powder
More informationAnalysis of Petroleum Fractions by ASTM D2887
Analysis of Petroleum Fractions by ASTM D2887 Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 2582 Key Words Simulated distillation, D2887, TRACE TR-SimDist Abstract ASTM
More informationHigh Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System
High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System Application Note Authors Frank David Research Institute for Chromatography, Pres.
More informationImpact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter
Cd The Chrom Doctor Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands. In Part 1 of this series we focused
More informationTechnical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS)
Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the pyrolysis-gas chromatograph-mass spectrometer
More informationApplication Note. Author. Introduction. Energy and Fuels
Analysis of Free and Total Glycerol in B-100 Biodiesel Methyl Esters Using Agilent Select Biodiesel for Glycerides Application Note Energy and Fuels Author John Oostdijk Agilent Technologies, Inc. Introduction
More informationASTM D 6730 Detailed Hydrocarbon Analysis
ASTM D 6730 Detailed Hydrocarbon Analysis Jaap de Zeeuw, Jan Pijpelink and Barry Burger Restek Corporation ASTM D 6730-01(2006)e1 Determination of Individual Components in Spark Ignition Engine Fuels as
More informationFausto Munari e Andrea Cadoppi ThermoFisher - Italy
The world leader in serving science Ultra Fast GC Determination of Total Hydrocarbons (C7-C40) and BTEX in Water and Soils through Direct Resistively Heated capillary columns and Robotic Autosampler. Fausto
More informationDetermination of Free and Total Glycerin in B100 Biodiesel
Page 1 of 5 Page 1 of 5 Return to Web Version Determination of Free and Total Glycerin in B100 Biodiesel By: Michael D. Buchanan, Katherine K. Stenerson, and Vicki Yearick, Reporter US Vol 27.1 techservice@sial.com
More informationDetermination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC
Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC Application Note Energy and Fuels Authors Detlef Wilhelm Anatox GmbH & Co. KG Fürstenwalde, Germany Udo Huber Agilent
More informationAgilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass. James D. McCurry, Ph.D. Senior Scientist
Agilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass James D. McCurry, Ph.D. Senior Scientist Agilent Solutions for the Analysis of Ethanol-Based Fuels Derived From Biomass Presented
More informationThe Only No Compromise
C196-E071 The Only No Compromise Best UHPLC for the Next Era Nexera is revealed as the next generation Shimadzu LC, achieving a new greater level of performance. Nexera hardware meets all analysis requirements:
More informationAgilent 6850 Series II Network GC System G2630A Performance Specifications
IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Agilent 6850 Series II Network GC
More informationHitachi High-Speed Amino Acid Analyzer
Hitachi High-Speed Amino Acid Analyzer Hitachi's Amino Acid Analyzer, Model L-8900, is the culmination of the technology and expertise earned from our extensive experience in manufacturing amino acid analyzers
More informationAgilent 7693A Automated Liquid Sampler
Agilent 7693A Automated Liquid Sampler Specifications Overview The Agilent 7693A is a state-of-the-art sample handling and injection system that provides the highest levels of precision and reliability
More informationDual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector
Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector Application Note Hydrocarbon Processing Authors ChunXiao Wang Agilent Technologies
More informationApplication Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany
Determination of Aromatic Hydrocarbons in Petroleum Middle Distillates with the Agilent Infinity Binary HPLC System with RID Detection According to IP9()/ASTM D9 Application Note Energy & Chemicals - Petrochemicals
More informationINSTALLATION MANUAL for AAA DIRECT
INSTALLATION MANUAL for AAA DIRECT AAA Direct Installation Manual Document No. 023741-04 1 of 11 INSTALLATION MANUAL for AAA DIRECT Dionex Corporation, 2004 Document No. 031481 Revision 04 17 November
More informationTechnical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)
Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph-mass spectrometer (GC-MS). 2.0 Scope
More information6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR
6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR A15836 This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use
More informationDetection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector
Detection of Sulfur Compounds in Natural Gas According to ASTM D554 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector Application Note Author Rebecca Veeneman Abstract Sulfur compounds in natural
More informationCustomer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard
Agilent Preventive Maintenance provides factory recommended service for your analytical systems to assure reliable operation and the accuracy of your results. Delivered by highly-trained and certified
More informationInstalling the 6850 Direct Valve Column Connector Kit
Use this kit to bypass the inlet and connect your 6850 gas or liquid sampling valve directly to a column. To use this kit, your 6850 GC must have a purged packed inlet. This kit contains: Installing the
More informationHPLC SYSTEM S 500 SERIES
HPLC SYSTEM S 500 SERIES The Sykam HPLC System S 500 Series is intended for all routine analysis as well as for the ambitious analyst. The system configuration is highly variable and several upgrade options
More informationPATROL UPLC Process Analysis System for Production
PATROL UPLC Process Analysis System for Production The PATROL UPLC Process Analysis System for Production is an integrated system solution designed and engineered to perform online and atline reaction
More informationGC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method
GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN1413:211 Method Application Note Author James D. McCurry, Ph.D. Agilent Technologies Abstract
More informationAgilent 1290 Infinity Autosampler
Agilent 1290 Infinity Autosampler Installation of the Large Volume Injection Kit General Information This note describes how to install the large volume injection kit into an Agilent 1290 Infinity Autosampler.
More informationFree and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584
Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN 14105 and ASTM D6584 Introduction With today s increasing concern for the environment and the depletion of fossil
More informationRefinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography
Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography Dr. Robert Lorenz Analytical Chemist Chevron Energy Technology Company Gulf Coast Conference Galveston, TX October
More informationDetermination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105
Application Note: 10215 Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105 Fausto Munari, Daniela Cavagnino, Andrea Cadoppi, Thermo Fisher Scientific, Milan,
More informationMethanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.
Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers Application Note Abstract Versa With the rising prices of fossil fuels, more emphasis is being put on renewable resources
More informationAgilent and ASTM. Update on Recent Activities. Page 1
Agilent and ASTM Update on Recent Activities Page 1 ASTM Committees for Refining, Fuels, Petroelum Products ASTM International D2 Petroleum products & lubricants D3 Gas fuels (natural gas) gas fuels D16
More informationUsing the PSD for Backflushing on the Agilent 8890 GC System
Application Note Petrochemicals Using the PSD for Backflushing on the Agilent 889 GC System Author Brian Fitz Agilent Technologies, Inc. Wilmington, DE, USA. Abstract An Agilent 889 series GC equipped
More informationAgilent Multimode Inlet for Gas Chromatography
Agilent Multimode Inlet for Gas Chromatography Technical Note Agilent Multimode Inlet for the 7890A GC Designed to give you ease of use and maximum flexibility, the Agilent Multimode Inlet does everything
More informationThe next level of True Blue Performance
Restek Supplies & Accessories The next level of True Blue Performance Topaz GC inlet liners feature revolutionary technology and inertness to deliver the next level of: Deactivation Reproducibility Productivity
More informationPractical Steps in GC Troubleshooting
Practical Steps in GC Troubleshooting Techniques, Tips, and Tricks Mark Sinnott Application Engineer GC Columns & Supplies Page 1 Everything was just fine and then this happened! How do I go about TROUBLESHOOTING?
More informationJaap de Zeeuw Restek Middelburg, The Netherlands. Copyrights: Restek Corporation
Liner Selection in GC Jaap de Zeeuw Restek Middelburg, The Netherlands What is a liner? A liner is an inert glass tube that is positioned inside the injection port of the GC The sample solution is introduced
More informationGC Analysis of PCBs in Fish Oil by Direct Injection Without Any Sample Cleanup. Philip L. Wylie, Ph.D. Sr. Applications Chemist Agilent Technologies
GC Analysis of PCBs in Fish Oil by Direct Injection Without Any Sample Cleanup Philip L. Wylie, Ph.D. Sr. Applications Chemist Agilent Technologies Pittsburgh Conference, 2008 Paper 1120-1 Tuesday, 8:30
More informationDetection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology
Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology Electronic Noses An electronic nose produces a recognizable response based
More informationVERITY Purification Systems Build the perfect system to fit your research
VERITY Purification Systems Build the perfect system to fit your research SPEC SHEET PURIFICATION VERITY PURIFICATION SYSTEMS WITH VERITY 3011 ISOCRATIC PUMP MAKE YOUR LAB LIFE EASIER From automated GPC
More informationGCMS-QP2010 Series. Consumables. Shimadzu Gas Chromatograph/Mass Spectrometer
GCMS-QP2010 Series Consumables Shimadzu Gas Chromatograph/Mass Spectrometer Diag. # 1) Injection Port - Septum Nut 2) Injection Port - Needle Guide 3) Septum 4) O-ring for Glass Insert 5) Glass Insert
More informationS 1125 HPLC Pump System
Routine S 170 EXPERT HPLC HPLC System SYSTEMS 500 The Sykam ROUTINE HPLC System S 500 is intended for all routine analysis. The system configuration is highly variable and several upgrade options make
More informationLarge Volume Injection of Polycyclic Aromatic Hydrocarbons
JSB is an authorised partner of Large Volume Injection of Polycyclic Aromatic Hydrocarbons Application Note - Environmental #113 Author Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Abstract
More informationAlliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE
Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE AN HPLC THAT WORKS AS HARD AS YOU DO Confidence in your results comes with high performance, day-to-day reliability, and overall system robustness.
More informationTechnical Overview. Introduction
Performance characteristics of the Agilent 129 Infinity Binary Pump More resolution and speed for conventional, superficially porous and sub-2 µm column packing material Technical Overview Introduction
More informationProof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut
Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut Application Note Environmental Author Ken Lynam Agilent Technologies, Inc. Abstract Specially designed self-tightening
More informationImpurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution
Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 129 Infinity II HDR-DAD Impurity Analyzer Solution Application ote Small Molecule Pharmaceuticals Author Sonja Schneider Agilent Technologies,
More informationAnalysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction
Analysis of Glycerin and Glycerides in Biodiesel (B1) Using ASTM D68 and EN11 Application HPI/Petrochemicals/Polymers Author James D. McCurry Agilent Technologies, Inc. 8 Centerville Road Wilmington, DE
More informationIncreased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel
Application Note Energy and Chemicals Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel Applying the Energy Institute Method IP 8 with an Agilent J&W
More informationPhenomenex Exclusive! Inlet Base Seals. Agilent 5000, 6000 and. Guaranteed fit for Series GCs. Easy Seals
Phenomenex Exclusive! Easy Seals Inlet Base Seals Guaranteed fit for Agilent 5000, 6000 and 7000 Series GCs The Easy Seals Gold Standard No Strained Hands No Troublesome Washers or Adaptors Continuing
More informationHPLC System S 600 Series
HPLC System S 600 Series The Sykam HPLC System S 600 Series is intended for the ambitious analyst. The system configuration is highly variable and several upgrade options make this system suitable for
More informationSupporting Information. for
Supporting Information for Metabolomic Assessments of Key Maize Resources: GC-MS and NMR Profiling of Grain from B73 Hybrids of the Nested Association Mapping (NAM) Founders and of Geographically Diverse
More informationFast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC
Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC Technical Overview Trace gas analysis is a challenge in today s world. The ability to analyze lower component
More informationTechniques for Making Your GC Analysis More Repeatable, Reproducible and Robust
Techniques for Making Your GC Analysis More Repeatable, Reproducible and Robust Mark Sinnott Application Engineer Columns and Consumables June 20, 2008 Primary Areas of Concern Sample Auto-Injector Inlet
More informationStudy of viscosity - temperature characteristics of rapeseed oil biodiesel and its blends
Study of viscosity - temperature characteristics of rapeseed oil biodiesel and its blends Li Kong 1, Xiu Chen 1, a, Xiaoling Chen 1, Lei Zhong 1, Yongbin Lai 2 and Guang Wu 2 1 School of Chemical Engineering,
More informationOptimized Method for Analysis of Commercial and Prepared Biodiesel using UltraPerformance Convergence Chromatography (UPC 2 )
Optimized Method for Analysis of Commercial and Prepared Biodiesel using UltraPerformance Convergence Chromatography (UPC 2 ) Mehdi Ashraf-Khorassani, 1 Giorgis Isaac, 2 and Larry T. Taylor 1 1 Department
More informationApplication Note. Authors. Abstract. Energy & Chemicals
Determination of Aromatic Content in Diesel Fuel According to ASTM D5186 Enhancing the Agilent 126 Infi nity Analytical SFC System with a Flame Ionization Detector Application Note Energy & Chemicals Authors
More informationASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column
ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column Cory S. Fix, Director of Application Development cory.fix@vgcchromatography.com Willie Steinecker,
More informationGC Method Compliance and Large Valve Oven Application
GC Method Compliance and Large Valve Oven Application Solution Focused Kelly Beard June 20, 2016 Compliant Methods Compliant Methods What are our options? Analyzers and Analyzer Kits High Volume Well Known
More informationAchieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)
Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Brian Fitz and Bill Wilson Agilent Technologies, Inc. 285 Centerville Road Wilmington,
More informationOxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels
Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels James D. McCurry Senior Scientist Agilent Technologies Wilmington, DE USA Page 1 Application Summary There is a need to measure
More informationComplete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges
Complete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges Alexandria Pavkovich Jason Thomas Trent Sprenkle Outline Background EPA Method Requirements Background
More informationMetathesis Catalysts
A Standard System of Characterization for Olefin Metathesis Catalysts Tobias Ritter, Andrew Hejl, Anna G. Wenzel, Timothy W. Funk, Robert H. Grubbs* Supporting Information Table of contents Table of contents...
More informationGC Inlets. An Overview. Simon Jones GC Applications Engineer
GC Inlets An Overview Simon Jones GC Applications Engineer Types of Inlets Purged Packed Split / Splitless Cool On Column Programmable Temperature Vaporization Volatiles Interface Multi Mode Inlet Where
More informationThe Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over
The Agilent 12 Series high performance autosampler SL: Area precision, injection volume linearity, imum accessible volume, carry-over Technical Note 8 6 4 2.5 1 Introduction A main performance criterion
More informationTechniques for Making Your GC Analysis More Repeatable, Reproducible and Robust
Techniques for Making Your GC Analysis More Repeatable, Reproducible and Robust Mark Sinnott Application Engineer Columns and Consumables June 20, 2008 Primary Areas of Concern Sample Auto-Injector Inlet
More informationAgilent 1290 Infinity II Ultra Low Dispersion Kit
Agilent 1290 Infinity II Ultra Low Dispersion Kit Technical Note 1290 Infinity II Ultra Low Dispersion Kit - Technical Note In this note we describe how to install the Agilent 1290 Infinity II Ultra Low
More informationFor more information about Agilent Technologies services please visit our web site using the following URL
Agilent Preventive Maintenance provides factory recommended service for your analytical systems to assure reliable operation and the accuracy of your results. Delivered by highly-trained and certified
More informationTechniques for Improving the Reproducibility of GC Analysis
Techniques for Improving the Reproducibility of GC Analysis Primary Areas of Concern Sample Auto-Injector Inlet Column Detector Sample Extract Handling and Care It is critical that the sample extract be
More informationChoose for Your Instrument
Choose for Your Instrument GC Supplies - Instrument Quick Pick Guide Agilent Technologies Bruker/Scion (Varian) PerkinElmer Shimadzu Thermo Scientific GC Supplies - Instrument Quick Pick Guide Choose for
More informationDETERMINATION OF N-BUTANOL AND ISOBUTANOL IN GASOLINE USING GAS CHROMATOGRAPHY (GC-FID)
DETERMINATION OF N-BUTANOL AND ISOBUTANOL IN GASOLINE USING GAS CHROMATOGRAPHY (GC-FID) Vladimir Honig, Jan Taborsky, Zdenek Linhart Czech University of Life Sciences Prague honig@af.czu.cz Abstract. The
More informationHPLC Model 210D. UV detector lamp switch. Rheodyne 7725i injection valve. Prime/purge valve. Column. compartment. thumbscrews. Storage.
1. Overview The front panel houses the UV detector lamp switch, Rheodyne injection valve, prime/purge valve, pressure gauge, and the storage compartment. In the storage compartment are the LED display
More informationAgilent Multimode Inlet
Agilent Multimode Inlet Large Volume Injection Tutorial Agilent Technologies Notices Agilent Technologies, Inc. 2009 No part of this manual may be reproduced in any form or by any means (including electronic
More informationSulfur Detection at ppb Levels in Light Hydrocarbon Streams
Sulfur Detection at ppb Levels in Light Hydrocarbon Streams Based on a New Super Permeable PLOT Column Agilent Select Low Sulfur Johan Kuipers Channel Training Specialist Oct 12 th, 2010 1 October 18,
More information