1 Valmet Conductivity Measurement Series 3100 Installation & operating manual K15344 V1.1 EN

2

3 Table of contents 1. About the device 5 3. Operating Description...5 Sensors...5 Temperature compensation...5 Applications...6 Warranty Keypad...14 Measurement display...14 Service display...14 Menu description...15 Menu structure Installation 7 4. Maintenance & troubleshooting Installation notes...7 Dimensions...7 Mounting the TCU...8 Electric connections...9 Connecting the sensor...11 Mounting the sensor Cleaning the sensor...18 Removing & mounting the sluice sensor...19 Troubleshooting...20 Spare parts...21 Test equipment...21 Sensor connection diagram...22 Data sheets Declaration of conformity Conductivity TCU series Installation & operating manual K15344 V1.1 EN

4 Recycling and disposal When sorted by material, nearly all parts of the device can be recycled. A materials list is delivered with the device. Upon request, the manufacturer will provide more detailed instructions for recycling and disposal. Used devices may also be returned to the manufacturer for recycling and disposal against a separate fee. The equipment contains no batteries. 4

5 1. About the device 1.1. Description The series 3100 Conductivity TCU is designed for high and low conductivity measurement in combination with Valmet 4-electrode conductivity sensors series 4000 and It is designed for wall mounting or panel mounting in industrial plant environments. The TCU has four individually selectable set-up modes, two for UPW/WFI (low conductivity) and two for high conductivities. Each of the four set-up modes has access to individual setting of temperature compensation, output range, alarm, and calibration. The set-up modes are selectable remotely or from the keypad. The TCU has two 4 20 ma current outputs, a display, and a keypad. The line-powered version has two relays for alarms. Optionally the HART protocol for field device transmitters can be implemented. For the full specifications, please refer to HART Field Device Specification for series TCU versions, type no. (for HART communication: add "H" to type number. Wall mounted Panel mounted Loop powered Line powered Sensors All sensors are based on the 4-electrode principle according to which the current and the voltage are measured across separate electrodes. That eliminates the errors from the polarization effect that is generally a problem at high conductivities. The value of the cell constant is stored in the sensor. This makes all sensors interchangeable without sacrificing accuracy and without the need for recalibration or re-entering the cell-constant when exchanging the sensor. A large selection of sensor types is available meeting the needs for almost any application Temperature compensation In an electrolyte, all ions present will contribute to the conductivity. Different ions will contribute with different amounts and the amounts will vary with temperature. When the conductivity is used for the calculation of a chemical concentration, it is necessary to calculate what the conductivity would be at a fixed temperature (the conductivity is temperature compensated). This conductivity is calculated from measured conductivity and temperature. As different ions have different temperature dependences, the temperature compensation is to be selected to fit the actual application. The temperature is measured with a Pt-1000 resistor in the sensor. Ultrapure water (UPW) contains a very small amount of impurities, and the resulting conductivity arises from the impurity ions as well as from the autoprotolysis of the water. Strong or weak acids or bases (present as impurity ions in very small amounts) affect the autoprotolysis differently, and thus contribute differently to the conductivity. The conductivity from the impurity and from the autoprotolysis changes differently with temperature, and the temperature compensation algorithms have to compensate accordingly. The UPW-conductivity set-up includes temperature compensation algorithms for ultrapure water with the impurities found in common use. Conductivity TCU series Installation & operating manual K15344 V1.1 EN

6 1.4. Applications The series 3100 Conductivity TCU is designed for ultrapure water applications in power plants and in pharmaceutical industries, for cleaning in place (CIP) applications in breweries and diaries and for general use as a conductivity meter. UPW conductivity set up, temperature compensations. Reference temperature 25 C. OFF UPW-Neutral UPW-Acidic UPW-Alkaline UPW-Ammonia STD Salt IEC 746-BII C C C C C C UPW conductivity set up, temperature compensations. Reference temperature 25 C. OFF UPW-Neutral UPW-Acidic UPW-Alkaline UPW-Ammonia STD Salt IEC 746-BII C C C C C C Conductivity TCU series Installation & operating manual K15344 V1.1 EN

7 2. Installation 2.1. Installation notes When planning the installation of the series 3000 conductivity TCU, a number or issues have to be considered: Whether mounted on a wall, in a panel or onto a piping or cable tray, the place should be free of vibrations and mechanically shocks should not occur. Protect the instrument from process electrolyte, and from downpour. If necessary use a Protection Shield. Protect the instrument from excessively high or low temperatures and from direct sunlight. Be aware of the sensors requirements for liquor volume around the sensor tip. Mount the sensor at a place where it is protected from mechanical impact, where it can be accessed and extracted from the process liquor for service jobs, and where the process liquor is properly grounded via the process piping. Minimize the length of sensor cable as much as possible. Protect the sensor cable from mechanical impact, excessive high or low temperatures and process electrolyte. Look up for ground loop currents, which can cause a fluctuating or faulty reading. There should be no potential between the mounting place of the TCU case and the sensor. If mounting the TCU on a cable tray, the cable tree should be electrically connected to the process piping. Also, look up for a potential difference between the protective ground and the process piping and/or the mounting place of the TCU. A special situation arises where the process piping is made by isolating material e.g. PTFE lined steel piping, PVC or PP piping, giving no ground connection of the process electrolyte. If the sensor is of the type where no metal structure is in electrical contact with the process electrolyte giving a grounding of the process electrolyte that way, it is necessary to establish a grounding electrode in the process electrolyte, which must be connected to the metal structure of the sensor or to the housing of the TCU Dimensions Ø5.2 5 esc OK 26.5 Fig. 1. Installation dimensions, in millimeters. Conductivity TCU series Installation & operating manual K15344 V1.1 EN

8 2.3. Mounting the TCU Wall mounting: to access the sensor connector, open the lid and dismount the corner screws (4 pcs); see Fig. 2. Panel mounting: make a cut-out to the panel, size mm (5.34" 5.34"); Fig. 3. Pipe mounting: the bottom part of TCU can be turned 90 for mounting on a horizontal tube; see Fig Available pipe mounting kits: Type no. 4902a 4902b 4902c 4902d 4902e For pipe dimension mm mm mm mm mm Fig. 3. Panel mounting. Fig. 2. Wall mounting. Fig. 4. Pipe mounting. Conductivity TCU series Installation & operating manual K15344 V1.1 EN

9 2.4. Electric connections J2 J3 ma1 A Com B ma Sensor Power supply No V DC ma1 ma2 +A 0 +B Outputs Control J1 Power supply No V DC + - Optional S A Off On Off On S B Off Off On On Set-up 1 Set-up 2 Set-up 3 Set-up 4 Power supply No V DC + - SA SB Input load 4.4K Optional Fig. 5. Loop powered version. Switches SA and SB control the selection of set-up as indicated. Conductivity TCU series Installation & operating manual K15344 V1.1 EN

10 Fuse: RM5, 1 A slowblow Alarms Output Control U7 J2 J4 2 1 ma1+ Comm ma2+ Comm 1 2 +A 0 +B J2 J Sensor Alarm contacts shown in unpowered mode Supply voltage input V AC L N = = = J1 L N 230V ma1 ma2 +A +B Alarms Outputs Control Live Neutral Ground (housing) J5 ma1 & ma2: active outputs. Max. output voltage: 16V, 20mA Power supply V DC + - SA SB Input load 4.4K Optional The equipment is constructed for permanent installation and must be grounded for safety reasons. A power switch must be available close to the equipment. S A Off On Off On S B Off Off On On Set-up 1 Set-up 2 Set-up 3 Set-up 4 Fig. 6. Line powered version. Switches SA and SB control the selection of set-up as indicated. Conductivity TCU series Installation & operating manual K15344 V1.1 EN

11 2.5. Connecting the sensor 1. First release the union (Fig. 7 A) and separate the units (B). 2. Screw in the cable joint (C). 3. Reassemble the units and tighten them together until the packing is tight. 4. To disconnect the sensor plug (D), release the union and separate the units before screwing out the cable screw joint. NOTE: The cable gland must be tightened carefully to fulfill EMC requirements. U7 J3 J2 J Sensor 2 1 J1 L N 230V ma1 ma2 +A 0 +B Alarms Outputs Control J5 C B A D Fig. 7. Connecting the sensor. Conductivity TCU series Installation & operating manual K15344 V1.1 EN

12 2.6. Mounting the sensor The sensor should be mounted at a place where there will be no occurrence of air pockets or bubbles. The best installation is obtained across a pump (a) or in an upstream pipe (b), (e); Fig.. If the sensor is mounted in a downstream pipe the mounting point shall be moved away from locations where an air funnel may be created (c), (d), otherwise it may be necessary to insert a flow damper just after the measuring point (f), (g). If a continuous TCU is wanted even in case the pipe runs empty, an installation like (h) can be used. The sensor is mounted at the under side of the horizontal pipe in a cone, to let air bubbles pass without disturbing the TCU. Sensor (i) is used to for concentration control in a loop. The dosing is injected before the sensor, preferably with a pump between the dosage and the sensor to provide for mixing before the TCU. The sensor is mounted on the pressure side of pump to minimize influence from air bubbles. It is recommended to install valves so that the sensor can be removed for service without a shutdown of the process. c b a d Return g f h e i Fig. 8. Preferred sensor mounting locations. Conductivity TCU series Installation & operating manual K15344 V1.1 EN

13 Incorrect mounting Fig. 9 shows incorrect mounting locations: Sensor (a) is mounted in a tank where it is intended to control dosing via a pump. If the agitation in the tank stops, the control loop is opened, and the dosage control will fail. Control loop will also fail if the liquor level falls below the sensor measuring head. Sensor (b) is mounted in a horizontal pipe. If the flow is too low or in case of no flow, air bubbles may occur or the pipe may not be filled. Consequently, the sensor will be measuring partially in air, giving a faulty reading. Sensor (c) is installed where bubbles may occur due to local boiling caused by the vacuum created by the liquor in the long downstream pipe. Sensor (d) is installed in a downstream pipe close to an open ending. The pipe will never be filled and the sensor will give a faulty reading. Sensor (e) is installed in the suction side of a pump where the pressure is low while the pump is running. Consequently, air bubbles may be formed causing a faulty reading. a b Return e c d Fig. 9. Wrong mounting. Conductivity TCU series Installation & operating manual K15344 V1.1 EN

14 3. Operating 3.1. Keypad OK: Switching from measurement display to menu display; accepting a menu selection or value. Esc: Returning from menu or submenu. Up / down arrows: Navigating up and down in the menu. Changing value when entering text or numerical values. Left / right arrows: Navigating in a menu line to a position for modification. Valmet: Selects the measurement display or service display Measurement display Line 1: Active set-up and Set-up type Line 2: Temperature compensation Line 3: Alarm status Line 4: Process value Line 5: - - Line 8: Liquor temperature Fig. 2. Measurement display Service display Line 1: TCU serial number and active set-up Line 2: Software version Line 3: Process value Line 4: Measured conductivity, uncompensated Line 5: Pt1000 value and liquor temperature Line 6: Cell constant and cell constant mode Line 7: Current channel value and range Line 8: Voltage channel value and range Fig. 1. Keypad. Fig. 3. Service display. Conductivity TCU series Installation & operating manual K15344 V1.1 EN

15 3.4. Menu description Menu display is shown in Fig. 4 below. Each menu is described in this section. Fig. 4. Menu display. System menu Selections in System menu are valid for all set-ups. Set-up select: Selects between Set-up 1, Set-up 2, Set-up 3, Set-up 4 and External. Each set-up holds an individual set of values and selections. Set-up 1 and 2 are assigned to UPW conductivity whereas set-up 3 and 4 are assigned to standard conductivity. External set-up selection is controlled from the digital inputs (see section "Electric connection"). Set-up selection can also be controlled with HART. Temperature adjust: Correct the reading of measured liquor temperature by pressing <UP> an <DOWN> keys. Adjustment range ±1 C. Contrast adjust: Correct the display contrast by pressing <UP> and <DOWN> keys. Language: Select TCU language: English, Swedish, Finnish, Spanish, or Portuguese. Conductivity unit: Select between S/cm and S/m. Temperature unit: Select between C and F. Backlight: Selects between Off and On. When set Off, the backlight will be on for 20 seconds after each key pressed and for 30 seconds after power-on. A loop-powered TCU has no backlight. Run mode: Selects between 1 second update or Fast update. Controls how often a measurement is made. In Fast update the TCU samples the sensor as fast as possible. The sampling rate will be increased from 3 to 4 samples per second when Manual cell constant is used. HART mode: Selects between Read/write, Read only, and Off. HART functionality is optional and is only implemented in TCU with type id ending with letter H. HART version: Selects between HART version 6 and HART version 5. This affects the response from command 0 and 15. Poll address: Used from HART. Initial set to 0. Passwords set: Sets the passwords for each submenu in the main menu: System, Adjustments, Temperature Comp., ma Output and Alarm. Each password is a 4-digit number. A password of 0000 gives free access. This is factory default. If the password is forgotten press <LEFT> and <UP> at the same time. This will clear the password. If password is used on one or more menu points the System menu shall also be protected to protect the password. Damping settings: Allows the user to filter the incoming measurement signal. The damping value (seconds) sets the filtering period; if damping value is set to zero, "Damping disabled" is active. Conductivity TCU series Installation & operating manual K15344 V1.1 EN

16 Adjustment menu Selections in adjustment menu are valid only for actual set-up. Cell constant mode: Selects between Automatic and Manual mode. In Automatic mode the cell constant is read from the sensor whereas in Manual mode a fixed value is used. When the mode is changed from Automatic to Manual the Automatic value is transferred to the Manual value. Manual cell constant can be used to reduce the sampling period in Fast mode or to make an individual adjustment. However, the automatic setting of cell constant is lost when the sensor is exchanged. Cell constant value: Here a manual cell constant value can be entered. Most sensors have a value between 0.17 to 0.24 depending on sensor type and the individual sensor. Conductivity reading is proportional to the cell constant value. Calc cc to cond.: Here you can type in a known conductivity, the TCU will then calculate the corresponding cell constant. This is useful for calibrating the sensor in a fixed solution. The calculated cell constant is used as default if cell constant mode is set to Manual. Temperature Comp. Selections in Temperature Comp. menu are valid only for actual set-up. Select the temperature compensation. See section 1.4. Additionally it is possible to restore all values for actual set-up to factory defaults. ma Output Selections in ma output menu are valid only for actual set-up. Output select: Alternatives are Off, Process value, PV 0-x-infinite, PV without TC, Temperature. Low / High limit: Set the 4 ma value and the 20 ma value for the selected variable. For UPW set-up conductivity in µs, for standard set-up in ms. For selection Off the high limit sets a fixed output current with 0 as 4 ma and 100 as 20 ma. Low limit is not used. PV 0-x-infinite compresses the output current as shown in Fig. 5. Low limit is not used. Alarm Not implemented in loop-powered versions. Selections in Alarm menu are valid only for actual set-up. Alarm select: Alternatives are Off, Process value, PV without TC, USP<645> (only UPW-type set-up), Temperature. Alarm type: Maximum NO, Maximum NC, Minimum NO or Minimum NC. NO: Relay contact open during non-alarm, closed during alarm. NC: Relay contact closed during nonalarm, open during alarm. Relay coil is powered when contact is closed. Hysteresis: 1% of value, 0.5 for temperature. Alarm limit: For UPW and Standard set-up: Conductivity in ms. For USP<645> the limit indicates the percentage of the USP<645>limit, to allow setting up a safety margin. Alarm display flash: Valid for all set-up modes with line-powered TCU. Selects between Off and On If set to On an alarm condition will make the backlight flash. Autoclean Not implemented in loop-powered versions. Selections in Autoclean menu are valid for all set-up modes. The Autoclean function controls the wash sequence of Autoclean Sensor type Alarm relay is used for activating the flushing, The normal setting of Alarm 1 is ignored. ma output 1 is frozen during the flushing and the holding time. Fig. 5. Output current compression. Conductivity TCU series Installation & operating manual K15344 V1.1 EN

17 3.5. Menu structure System Set-up select Temperature adjust Contrast adjust Backlight Adjustments Cell constant mode Cell constant value Calc cc to cond. Temperature comp. ma output Output1 select Low limit 1 High limit 2 Output2 select Low limit 2 Alarms Alarm1 select Alarm1 type Alarm1 limit Alarm2 select Alarm2 type Autoclean Runmode High limit 2 Alarm2 limit Language Conductivity unit Alarm display flash Temperature unit HART mode Poll address Passwords set Damping settings Fig. 6. Menu structure. Conductivity TCU series Installation & operating manual K15344 V1.1 EN

18 4. Maintenance & troubleshooting 4.1. Cleaning the sensor Cleaning may be necessary if deposits have built up to an extent that affects the cell constant of the electrode part. When cleaning the sensor do not use hard material to scrape or grind. Use a soft brush or cloth and a suitable cleaning remedy depending on the nature of the deposits. Be careful not to damage the electrodes as this might affect the accuracy of the sensor. Chemical cleaning can be applied by using any reagent that will not affect the sensor material. Sulphamic acid (NH 2 SO 3 H) is recommended. Use 250 g in 1 litre of water. Heat the water to ease dissolving. The solution can be used cold or warm. Commercially available cleaning chemicals may contain corrosion inhibitors, which can create an electrical isolating layer on the electrodes. This will result in a high scaling signal after cleaning and a degraded performance of the sensor. List of usable cleaning acids for sensors. Insert for max. 2h at 30 C~ 86 F Acetic acid CH 3 COOH Citric acid C 3 H 4 (OH) (COOH) 3 Formic acid HCOOH Lactic acid C 3 H 4 (OH) COOH Nitric acid HNO 3 Oxalic acid (COOH) 2, max. 40% w/w Phosphoric acid H 3 PO 4 Sulphamic acid NH 2 SO 3 H Sulphuric acid H 2 SO 4, max. 30% w/w Tartaric acid C 2 H 2 (OH) 2 (COOH) 2 WARNING: Do not use Hydrochloric acid (HCl) or Hydrofluoric acid (HF)! Conductivity TCU series Installation & operating manual K15344 V1.1 EN

19 4.2. Removing & mounting the sluice sensor Removing (Fig. 1) Maximum pressure when removing: 5 bar. 1. Unscrew the nut while holding back the sensor to avoid a violent push-out. 2. Let the sensor move out under control until the stop. 3. Close the ball valve. 4. Unscrew the sensor from the sluice chamber. 1. Mounting (Fig. 2) 1. Check the O-ring for damage before remounting the sensor. 2. Open the ball valve. 3. Insert the sensor and tighten the 1" nut. 1. O-ring L 3. Sensor type L(mm) Fig. 2. Mounting the sluice sensor. 4. Fig. 1. Removing the sluice sensor. Conductivity TCU series Installation & operating manual K15344 V1.1 EN

20 4.3. Troubleshooting Locate the problem to the monitor, the sensor, sensor installation or to the electrical connections. If possible shift to another sensor or use a test probe to check the TCU. If the temperature reading is very wrong: Check Pt1000 in sensor for short circuit or interruption The check can be done by resistance measurement from the sensor connector. Temperature reading partly wrong: This error may arise from a poor thermal contact of the Pt1000 element to the sensor structure. The thermal contact can be checked by measuring the response time for a temperature change. Another source of error may be the installation. The steel part of the sensor with the reduced diameter is to be fully inserted in the process liquor flow. Conductivity/process value far out of range: Check the sensor cable for damage and check the sensor cable connection in the monitor for broken wires. Is the sensor tip surrounded with process liquor Check the measuring section of the sensor for scaling deposits and for electrode damages. No reading in display: Check line voltage, fuses and check that the connectors are in place. Setting of the temperature compensation Setting of analog output and accordance to setting of analog input of DCS Make a record of set-up info (Appendix A) If help is required from factory please make a record of the measurement display and the service display. This can be done with photos. Use backlight if possible or illuminate the display from up or down in a 45 angle to avoid reflections (Fig. 3). Do not use flash. Fig. 3. Taking a photo of TCU displays. Lamp Camera Conductivity TCU series Installation & operating manual K15344 V1.1 EN

21 4.4. Spare parts The TCU contains no wearing components. Since the sensor is the most exposed part, having spare sensors is recommended. Spare packings are recommended for the sensors: For type 4221 and 4226 sensors: Packing ring (4 pcs for each sensor) O-ring (3 pcs for each sensor) O-ring 44.2 x 3.0 for sluice sensor For type sensors: PTFE gasket with O-ring For type sensors: Gasket 4.5. Test equipment Test probe, type Pos. 1: Simulated temperature 90.1 C Pos. 2: Simulated temperature 25 C Pos. 1/2 Fig. 4. Test probe. Fig. 5. Packing system, 4216 sensors. Fig. 6. Packing system, 4223 and 4226 sensors. Conductivity TCU series Installation & operating manual K15344 V1.1 EN

22 4.6. Sensor connection diagram Test probe Red/Blue Pink/gray Green Yellow Gray Pink Blue Red Black White Brown Common Common sense Pt1000 sense Pt1000 Rcc sense Rcc I2 U2 Ground U1 I for 26 C temp. Reading 2k for CC = 0.2 Rt Rcc Rx Cable Rx 10 Ohm 100 Ohm 1 kohm 10 kohm 100 kohm 1 MOhm Simulated conductivity 20 ms/cm 2 ms/cm 200 µs/cm 20 µs/cm 2 µs/cm 0.2 µs/cm I1 U1 GND U2 I2 GND I1 U1 U2 I2 Sensor types 4221, 4223, 4338, 4339 Sensor types 4141, 4147, 4151, 4181 I1 U1 GND U2 I2 Rcc Rcc sense Common Common sense Pt1000 Pt1000 sense Sensor types 4224, 4226, 4336, 4337 Fig. 7. Sensor connection diagram and test probe. Rcc Pt 1000 Conductivity TCU series Installation & operating manual K15344 V1.1 EN

23 Valmet Conductivity Measurement 3100 Measurements for demanding applications The Valmet Conductivity instruments measure high and low conductivity using the 4-electrode principle. Designed for conductivity control in industries such as pulp, power, pharmaceutical, food, brewery, sugar and chemical. Four individually selectable set-up modes are available, two for very low conductivity and two for normal/high conductivity. No maintenance Long life 4-electrode measuring principle 13 preprogrammed temperature compensations Wide sensor-assortment for all applications Low installation costs Wide selection of process mounting accessories The Valmet Conductivity instruments are designed for very high precision and for a long and maintenance-free life. Rugged Accurate Reliable

24 Specifications Low conductivity set-up: Range 0 μs/cm to 2 S/cm (0 μs/m to 200 S/m). Temperature compensation: O, ultrapure water (neutral, acid, alkaline, ammonia), STD salt. High conductivity set-up: Range 20 μs/cm to 2 S/cm (2 ms/m to 200 S/m) Temperature compensation: O, STD salt, weak lye, weak acids Output: Two current outputs 4 20 ma for conductivity or temperature. Temperature measurement: From 40 to +250 C Display: LCD Sensor connection: Series 4000 sensors with 11-pole strip connector and MF20 cable adaptor Local or remote measuring range selection: Two low and two high conductivity set-up modes Control voltage: V DC HART communication as a standard Enclosure Material: Cast aluminium, PE coated Protection: IP67 Dimensions: 144 x 144 x 107 mm Mounting: Panel, wall or pipe Ambient temp.: From -10 to +60 C Storage temp.: From 20 to +70 C Weight: 2.3 kg Sensors for serie 3100 Series 4200 Industrial applications Series 4100 Sanitary applications Dimensions in mm Mounting options Wall Pipe Panel Valmet Corporation, BR81430_EN_01 06/2015. Wall mounting Type 3107H Type 3117H Panel mounting Type 3108H Type 3118H Power supply Loop powered Line powered V DC V AC, on ma1 6 VA Alarm No alarm Two relays 250 V AC, 8 A Display Non- Illuminated illuminated For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation. Pipe mounting kits Type For pipe No. dimension 4902 a mm b mm c mm d mm e mm

25 Sanitary Sensor 4111 Sanitary closed type 4-electrode conductivity sensor with internal electrodes. Used with conductivity TCU series 3100 and Mounting Mounting with 2 ISO 2852 clamp (not included) in a short Tee with a welding nipple. No air bubbles or sedimentation at the installation point. Do not mount on the top side of a horizontal pipe. Position the flow opening parallel to the direction of the flow. Technical data Materials PVDF, EPDM, silicone rubber and steel W (AISI 316L). Flange and cable connection: steel W (AISI 316L) Pressure 10 bar at 130 C Cell constant Reproduceability ±3 % Linearity 1.0 ± 0.5 % (electrically adjusted) From to 20 μs/cm ± 3.8 % per decade Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable Connection Weight PVC 16 x 0.25 mm 2 screened standard length 5 m max. temp. 70 C 11-pole D-connector with MF20 cable adapter 1.0 kg ø64 ø ø ±2 Valmet Corporation, D08670 V1.0 EN 11/2016 Dimensions in mm For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation Max. 50 Min. 34

26 Sanitary Sensor 4113 Sanitary open type 4-electrode conductivity sensor with external electrodes. Surface roughness of all wet parts Ra < 0.80 μm. Used with TCU series 3100 and Technical data Materials PTFE, silicone rubber and steel W (AISI 316L) Flange and cable connection: W (AISI 316L) Pressure 12 bar at 150 C 1.0 ± 0.5 % (electrically adjusted). Cell constant Calibrated in Ø63 mm SS pipe. Ø51 mm pipe gives an error of +6 %. Reproduceability ±2 % Linearity From to 20 μs/cm ± 2 % per decade Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2 screened standard length 5 m max. temp. 70 C Connection 11-pole strip connector with MF20 cable adapter Weight 1.1 kg R=30 25 ø64 For ISO clamp 2 ø Valmet Corporation, D07713 V1.1 EN 06/2016 Mounting Mounting with a 2 ISO 2852 clamp (not included). Calibrated in a measuring pipe: outside Ø63 mm, inside Ø60 mm. During factory calibration electrodes are directed towards the pipe wall as shown. If the original factory calibration will be used, mount the sensor in the same way; if mounted to a larger pipe or with a different electrode direction, recalibration is required. On a horizontal pipe, mount to the side of the pipe and make sure no air bubbles occur at the installation point. Hygienic mounting Mounting in an EHEDG approved process connection. Inside diameter of pipes at installation position must be 3 or more. For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation ± ø51

27 Sanitary Sensor 4113s Pharmaceutical sanitary open type for ultrapure injection water 4-electrode conductivity sensor with external electrodes. Surface roughness of all wet parts Ra < 0.80 μm. Used with TCU series 3100 and Technical data PTFE, silicone rubber and steel W Materials (AISI 316L) Pressure 12 bar at 150 C 1.0 ± 0.5 % (electrically adjusted) Calibrated in a Ø51 mm SS pipe. Cell constant Ø63 mm SS pipe gives an error of ~ -6 %. Ø63 mm plastic pipe gives an error of ~ %. Reproduceability ±2 % Linearity From to 0.2 μs/cm ± 2 % per decade. Below 5 μs/cm μs/cm. Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2 standard length 3 m max. temp. 70 C Can optionally be mounted with 5 m cable; measuring below 0.5 μs/ cm with this cable length is not recommended. Connection 11-pole strip connector with MF20 cable adapter Weight 0.6 kg 24.3 ø64 For ISO clamp 2 ø Typical mounting ø51 dimensions in millimeters Mounting Mounting with a 2 ISO 2852 clamp (not included). Minimum distance to surroundings: 24.3 mm. When mounting on a horizontal pipe, mount to the side of the pipe and make sure no air bubbles occur at the installation point. Construct the piping in a self-draining manner. 4113s s 8170

28 Hygienic mounting Mounting in an EHEDG approved process connection. Inside diameter of pipes at installation position must be 3 or more. Mounting with type 8151 stainless steel nipple in a 2 plastic Tee Warming! Direct mounting to a plastic pipe gives errors of -25% to 30 % 2 BSP 8151 O-ring x 2.62 included in 8151 ø63 plastic Tee Mounting in a Ø2 (51) Tee with reducers Warning! Do not mount directly to a pipe smaller than 2 (51 mm) Tee 2 (51) 2 (51) 1.5 (ø38) Mounting with 8155 stainless nipple in a 63.5mm stainless Tee Warning! Direct mounting to pipes wider than 2 (51 mm) gives errors ø63.5 Tee Valmet Corporation, D07726 V1.1 EN 06/2016 dimensions in millimeters For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation.

29 Sanitary Sensor 4115 Sanitary open type sensor 4-electrode conductivity sensor with external electrodes. Used with TCU series 3100 and Typical mounting Mounting with 2 ISO 2852 clamp (not included). Calibrated in a diam. 180 mm vessel; if the sensor will be mounted in a pipe smaller than 100 mm in diameter, recalibration is required. No air bubbles or sedimentation at the installation point. Do not mount on the top side of a horizontal pipe. Technical data Length 0.25 m or extended to 0.5 / 1.0 / 1.5 / 2.0 / 2.5 m Materials, sensor PTFE, silicone rubber and steel W (AISI 316L) Materials, flange & cable connection W (AISI 316L) Pressure max. 12 bar at 150 C Cell constant Reproduceability ± 2% Linearity 1.0 ± 0.5% (electrically adjusted) to 20 μs/cm ±2% per decade Temp. sensor Pt 1000 (IEC 751 class A) 25 R=30 L = 220 diam For ISO clamp 2 diam Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2 screened dimensions in millimeters standard length 5 m max. temp. 70 C ± 2 Connection 11-pole strip connector with MF20 cable adapter Valmet Corporation, D07438 V1.0 EN 04/2015 Weight 1.2 kg For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation diam. 51

30 Sanitary Sensor A sanitary open type 4-electrode conductivity sensor with external electrodes. Surface roughness of all wet parts Ra < 0.80 μm. Used with TCU series 3100 and Mounting Mounting with a 2 ½ ISO 2852 clamp (not included). Typical mounting in a short Tee. Keep the minimum distance to surroundings! During factory calibration the electrodes are directed towards the wall (as shown). If the original factory calibration will be used, mount the sensor as shown in the drawings. On a horizontal pipe, mount to the side of the pipe and make sure no air bubbles occur at the installation point. Construct the piping in self-draining manner. Hygienic mounting Mounting in an EHEDG approved process connection. Inside diameter of pipes at installation position must be 3 or more. Technical data dimensions in inches (mm) 3.05 (77.5) 0.96 (ø24.5) 20 R:0.26 (6.5) For ISO clamp 2½ Materials PTFE, silicone rubber and steel W (AISI 316L) Pressure 12 bar at 150 C 1.0 ± 0.5 % (electrically adjusted) 6.42 (163) 0.31 (8) 2.72 (69) 1 (25) Cell constant Calibrated in a Ø3 (Ø76 mm) pipe. Ø2.5 (Ø63 mm) pipe gives an error of +3 %. Reproduceability ±2 % Linearity From to 20 μs/cm ± 2 % per decade Temp. sensor Pt 1000 (IEC 751 class A) Typical mounting 0.63 (16) 3 (76) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Valmet Corporation, D07727 V1.1 EN 06/2016 Sensor cable Connection Weight PVC 12 x 0.25 mm 2 screened standard length 5 m max. temp. 70 C 11-pole strip connector with MF20 cable adapter 1.1 kg For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation. min (63)

31 Sanitary Sensor 4124s 3-A sanitary open type for ultra pure water 4-electrode conductivity sensor with external electrodes. Surface roughness of all wet parts Ra < 0.80 μm. Used with TCU series 3100 and Mounting Mounting with a 2 ½ ISO 2852 clamp (not included). Minimum distance to the surroundings is 30 mm. On a horizontal pipe, mount to the side of the pipe; make sure no air bubbles occur at installation point. Construct the piping in a self-draining manner. Hygienic mounting Mounting in an EHEDG approved process connection. Inside diameter of pipes at installation position must be 3 or more. Technical data PTFE, silicone rubber and steel W Materials (AISI 316L) Pressure 12 bar at 150 C Calibrated in a Ø63 mm SS pipe. Ø76 mm SS pipe gives an error of Cell constant ~ -3 % (77.5) 0.96 (24.5) 20 R:6.5 For ISO clamp 2 ½ 1.18 (30) Ø63 mm plastic pipe gives an error of ~ %. Reproduceability ±2 % Linearity From to 0.2 μs/cm ± 2 % per decade. Below 5 μs/cm μs/cm (163) 0.31 (8) 1 (25) 2.72 (69) Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2 standard length 3 m max. temp. 70 C Valmet Corporation, D07712 V1.1 EN 06/2016 Connection Weight Sensor can optionally be mounted with 5 m of cable; measuring below 0.5 μs/cm with this cable length is not recommended. 11-pole strip connector with MF20 cable adapter 1.0 kg For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation.

32 Sanitary Sensor 4131 Sanitary closed type 4-electrode conductivity sensor with internal electrodes. Universal flange for SMS38, DIN DN32 and 40 union. Used with TCU series 3100 and Mounting Sensor mounting in a short Tee with a welding male part. Can be mounted on vertical or horizontal pipes in all directions; on a horizontal pipe mounting to the side or under the pipe. Flow opening must be parallel to the flow direction. An arrow on the sensor indicates the position of the temperature sensor. Mount in a place where no air pockets or bubbles occur. Valmet Corporation, D07728 V1.0 EN 04/2015 Technical data Materials, sensor Flange & cable connection Temperature sensor Pressure Cell constant Reproduceability ±2 % PVDF, EPDM, silicone rubber and steel W (AISI 316L) Steel W (AISI 316L) Coated with PFA max. 10 bar at 130 C (150 C < 1h) 1.0 ± 0.5 % (electrically adjusted) Linearity From to 1 μs/cm ± 2 % per decade Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable Connection Weight Accessories PVC 12 x 0.25 mm 2 screened standard length 5 m max. temp. 70 C 11-pole strip connector with MF20 cable adapter 1.0 kg SMS38 DN32 DN40 Union nut Nipple with gasket Nipple height (h) 22 mm 16.5 mm 16.5 mm Nipple diameter Tee branch length (b) 4 mm 8 mm 8 mm For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation. dimensions in mm Typical mounting ø50 ø24 35 min. 38 h b SMS38

33 Sanitary Sensor 4135 Sanitary open type 4-electrode conductivity sensor with external electrodes. Used with conductivity TCU series 3100 and Universal flange for SMS38, DIN DN32 and 40 union. Mounting Mounting requires a welding nipple. Sensor can be mounted in any direction. Keep the minimum distance (R = 37 mm) to the surroundings! Mount the sensor in a place where no air pockets or bubbles occur. Technical data Length Materials L = 220 mm or to be specified, max. 2.5 m Pressure 12 bar at 150 C Cell constant Reproduceability ±2 % Linearity PTFE, silicone rubber and steel W (AISI 316L). 1.0 ± 0.5 % (electrically adjusted) Calibrated in an open vessel From to 20 μs/cm ± 2 % per decade Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable Connection Weight PVC 16 x 0.25 mm 2 screened standard length 5 m max. temp. 70 C 11-pole strip connector with MF20 cable adapter 0.9 kg Typical mounting 140 L ø ø R=37 Dimensions in mm Valmet Corporation, D08672 V1.0 EN 11/2016 Accessories SMS38 DN32 DN40 Union nut Nipple with gasket Nipple height (h) 22 mm 16.5 mm 16.5 mm Nipple diameter For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation. R=37

34 Sanitary Sensor 4141 Sanitary closed type 4-electrode conductivity sensor with internal electrodes. Suitable for acetone applications. Used with TCU series 3100 and Mounting Sensor mounting with a 2 ISO 2852 clamp (not included). When mounting on a horizontal pipe, mount on the side and in a place where no air pockets or bubbles occur. The flow opening should be parallel to the direction of the flow. Construct the piping in a self-draining manner. Technical data Materials PVDF, EPDM, silicone rubber and steel W (AISI 316L) Temperature sensor Pressure Coated with PFA max. 10 bar at 130 C (150 C < 1h) dimensions in mm Cell constant Reproduceability ± 2 % 1.0 ± 0.5 % (electrically adjusted) Linearity From to 1 μs/cm ± 2 % per decade; below 5 μs/cm μs/cm Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. ø64 ø ø Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2 screened Standard length 5 m Sensor can optionally be mounted with 3 m or 15 m of cable. Typical mounting 4141 Measuring below 0.2 μs/cm with 3 m cable is not recommended. Do not measure below 5 μs/cm with 15 m cable. max. temp. 70 C ø63 Valmet Corporation, D07736 V1.0 EN 04/2015 Connection Weight 11-pole strip connector with MF20 cable adapter 1.0 kg For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation

35 Sanitary Sensor 4141 PES Sanitary closed type non-standard item 4-electrode conductivity sensor with internal electrodes. Used with TCU series 3100 and Mounting Sensor mounting with a 2 ISO 2852 clamp (not included) to a short Tee with a welding nipple. The sensor can be mounted in pipes of all directions; do not mount on the upper side of a horizontal pipe. Mount in a place where no air bubbles or sedimentation occur. The flow opening should be parallel to the direction of the flow. An arrow on the sensor indicates the position of the temperature sensor. Technical data Materials PES, EPDM, silicone rubber and steel W (AISI 316L). Flange and cable connection: steel W (AISI 316L) Pressure max. 12 bar at 150 C Cell constant Reproduceability ± 2 % 1.0 ± 0.5 % (electrically adjusted) Linearity From to 1 μs/cm ± 2 % per decade Temp. sensor Pt 1000 (IEC 751 class A) ø64 ø ø dimensions in mm Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2 screened Standard length 5 m max. temp. 70 C Typical mounting Connection 11-pole strip connector with MF20 cable adapter Weight 1.0 kg ø Valmet Corporation, D07737 V1.0 EN 04/2015 For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation ø51 ø38

36 Sanitary Sensor 4147 Sanitary flow type 4-electrode conductivity sensor with internal electrodes (resistant to acetone). Used with TCU series 3100 and Mounting Sensor mounting with a 2 ISO 2852 clamp in a miniature flow chamber (not included). Mount the flow chamber vertically. Flange equipped with a hole for a 2 mm guide pin, to secure the correct flow direction (see figures). Mount in a place where no air pockets or bubbles occur. Technical data Materials Temperature sensor Pressure Cell constant Reproduceability ± 2 % Linearity PVDF, EPDM, silicone rubber and steel W (AISI 316L), sensor top W (AISI 316L). Coated with PFA max. 10 bar at 130 C (150 C < 1h) 1 ± 0.5 % (electrically adjusted); calibrated in one of the flow chambers. From to 100 μs/cm ± 3 % per decade Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2 Connection Weight Standard length 5 m max. temp. 70 C 11-pole strip connector with MF20 cable adapter 1.0 kg dimensions in mm ø23.5 ø64 ø24 Typical mounting clamp Measuring chamber volume 5.5 cm 2 Valmet Corporation, D07730 V1.0 EN 04/2015 Miniature flow chamber Type 8190 Type 8192 Accessories ld 6 mm ld 10 mm Clamp rings 8172: 2 ISO 8162: Mini Mini nipple & gasket 8160: Ø9.5 Extra O-ring Type , x 1.78 EPDM For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation R=0.8±0.1

37 Sanitary Sensor 4151 Sanitary closed type 4-electrode conductivity sensor with internal electrodes and VARIVENT flange (clamp not included). Used with TCU series 3100 and Mounting Sensor mounting in VARIVENT DN50 (Ø68) inline access unit. Can be mounted on vertical or horizontal pipes in all directions; on a horizontal pipe mounting to the side or under the pipe. Flow opening must be parallel to the flow direction. Mount in a place where no air pockets or bubbles occur. Valmet Corporation, D07729 V1.0 EN 04/2015 Technical data Materials Temperature sensor Pressure Cell constant Reproduceability ±2 % PVDF, EPDM silicone rubber and steel W (AISI 316L) Coated with PFA max. 10 bar at 130 C (150 C < 1h) 1.0 ± 0.5 % (electrically adjusted) Linearity From to 1 μs/cm ± 2 % per decade. Below 5 μs/cm μs/cm Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable Connection Weight PVC 12 x 0.25 mm 2 screened Standard length 5 m. Sensor can optionally be mounted with 3 or 15 m of cable. Measuring below 0.2 μs/cm with 3 m of cable is not recommended. Do not measure below 5 μs/cm with 15 m of cable. max. temp. 70 C 11-pole strip connector with MF20 cable adapter 1.5 kg For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation. dimensions in mm Typical mounting ø84 ø68 ø24 min. 50 min. 35 (R=6.5)

38 Sanitary Sensor 4154 Sanitary open type 4-electrode conductivity sensor with external electrodes. Surface roughness of all wet parts Ra < 0.80 μm. Used with TCU series 3100 and Mounting Mounting in a VARIVENT flange with a VARIVENT DN50 clamp (Ø68, not included). Minimum distance to surroundings: 30 mm. On a horizontal pipe, mount to the side of the pipe and make sure no air bubbles occur at the installation point. Design the piping in self-draining manner. Valmet Corporation, D08310 V1.0 EN 06/2016 Hygienic mounting Mounting in an EHEDG approved process connection. Inside diameter of pipes at installation position must be 3 or more. Technical data Materials Pressure 12 bar at 150 C Cell constant Reproduceability ±2 % Sensor: PTFE, silicone rubber and steel W (AISI 316L) 1.0 ± 0.5 % (electrically adjusted). Calibrated in a VARIVENT DN50 in-line access unit. Linearity From to 20 μs/cm ± 2 % per decade Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable Connection Weight PVC 12 x 0.25 mm 2, screened standard length 3 m max. temp. 70 C 11-pole strip connector with MF20 cable adapter 1.5 kg For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation. dimensions in mm ø84 ø68 25 min. 75 (R=6.5)

39 Sanitary Sensor 4154s Pharmaceutical sanitary open type 4-electrode conductivity sensor with external electrodes. Surface roughness of all wet parts Ra < 0.80 μm. Used with TCU series 3100 and Mounting Mounting in a VARIVENT flange with a VARI- VENT DN50 clamp (Ø68, not included). Minimum distance to surroundings: 30 mm. On a horizontal pipe, mount to the side of the pipe and make sure no air bubbles occur at the installation point. Design the piping in self-draining manner. Hygienic mounting Mounting in an EHEDG approved process connection. Inside diameter of pipes at installation position must be 3 or more. Technical data Materials Sensor: PTFE, silicone rubber and steel W (AISI 316L) ø68 30 Pressure 12 bar at 150 C ø84 Cell constant 1.0 ± 0.5 % (electrically adjusted). Calibrated in a VARIVENT DN50 inline access unit. 25 Reproduceability ±2 % Linearity From to 0.2 μs/cm ± 2 % per decade; below 5 μs/cm μs/cm Temp. sensor Pt 1000 (IEC 751 class A) dimensions in mm 25 min. 75 Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2, screened standard length 3 m Valmet Corporation, D08311 V1.0 EN 06/2016 Connection Weight option: 5 m cable; measuring below 0.5 μs/cm with this cable length is not recommended max. temp. 70 C 11-pole strip connector with MF20 cable adapter 1.5 kg For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation. min. 60 (R=6.5)

40 Sanitary Sensor 4181 Sanitary closed type 4-electrode conductivity sensor with internal electrodes. Used with TCU series 3100 and Mounting Sensor mounting with a 1.5 ISO 2852 clamp (not included). When mounting on a horizontal pipe, mount to the side of the pipe and in a place where no air bubbles occur. The flow opening should be parallel to the direction of the flow. Construct the piping in a self-draining manner. Valmet Corporation, D07738 V1.0 EN 04/2015 Technical data Materials PVDF, EPDM, silicone rubber and steel W (AISI 316L). Temperature sensor coated with PFA Pressure max. 10 bar at 130 C (150 C < 1h) Cell constant Reproduceability ± 2 % Linearity 1.0 ± 0.5 % (electrically adjusted) Calibrated in SS pipe. In SS pipe from 1 μs/cm to μs/cm ± 2 % per decade; from μs/cm to μs/cm error + 5 % per decade Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable Connection Weight PVC 12 x 0.25 mm 2 screened Standard length 5 m. Sensor can optionally be mounted with 3 m or 15 m of cable. Measuring below 0.2 μs/cm with 3 m cable is not recommended. Do not measure below 5 μs/cm with 15 m cable. max. temp. 70 C 11-pole strip connector with MF20 cable adapter 1.0 kg For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation. dimensions in mm ø38 ø50.5 ø24 UP 2 30 ø ø ø25 Typical mounting ø44 ø50

41 Sanitary Sensor 4183 Sanitary open type for tank mounting 4-electrode conductivity sensor with external electrodes. Surface roughness of all wet parts Ra < 0.80 μm. Used with TCU series 3100 and Mounting Mounting with a 1 ½ ISO 2852 clamp (not included). Mount in a place where no air bubbles or sedimentation occur. Calibrated in a measuring pipe with inside Ø180 mm. If mounted to a pipe smaller than Ø100 mm, recalibration is required. Valmet Corporation, D07739 V1.1 EN 06/2016 Hygienic mounting Mounting in an EHEDG approved process connection. Inside diameter of pipes at installation position must be 3 or more. Technical data Materials PTFE and steel W (AISI 316L). Flange and cable connection: steel W (AISI 316L). Inside pressure gasket: SIL Pressure max. 12 bar at 150 C Cell constant 1.0 ± 0.7 % (electrically adjusted) Calibrated in a 180 mm pipe. Reproduceability 2 % Linearity From to 20 μs/cm ± 2 % per decade Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2 screened Standard length 5 m max. temp. 70 C Connection 11-pole strip connector with MF20 cable adapter Weight 1.0 kg For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation. R= ø Typical mounting NA connect 38 mounting Standard ISO 1 ½ clamp mounting For ISO clamp 1 ½ ø24 NAV ½ ISO L=85 37 dimensions in mm

42 Sanitary Sensor 4183s Pharmaceutical sanitary open type for ultra-pure injection water 4-electrode conductivity sensor with external electrodes. Surface roughness of all wet parts Ra < 0.80 μm. Used with TCU series 3100 and Mounting Mounting with a 1 ½ ISO 2852 clamp (not included). Minimum distance to surroundings is 24.3 mm, On a horizontal pipe, mount to the side of the pipe and make sure no air bubbles occur at the installation point. Construct the piping in self-draining manner. Hygienic mounting Mounting in an EHEDG approved process connection. Inside diameter of pipes at installation position must be 3 or more. Technical data Materials PTFE, silicone rubber and steel W (AISI 316L) Pressure max. 12 bar at 150 C Cell constant Reproduceability ± 2 % 1.0 ± 0.5 % (electrically adjusted) Calibrated in a Ø51 mm SS pipe. Ø63 mm SS pipe gives an error of approx. -6 %. Ø63 mm plastic pipe gives an error of approx % Linearity From to 0.2 μs/cm ± 2 % per decade Below 5 μs/cm μs/cm Temp. sensor Pt 1000 (IEC 751 class A) 24.3 ø50.5 For ISO clamp 1½ ø24 Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2 screened Standard length 3 m. Valmet Corporation, D08070 V1.1 EN 06/2016 Connection Weight Option: mounting with 5 m cable; measuring below 0.5 μs/cm with this cable length is not recommended. max. temp. 70 C 11-pole strip connector with MF20 cable adapter 0.6 kg For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation. dimensions in mm

43 Sanitary Sensor 4185 Sanitary open type for tank mounting 4-electrode conductivity sensor with external electrodes. Wet steel parts of sensor polished to Ra<0.8 μm, PTFE part to Ra<1.2 μm. Used with TCU series 3100 and Mounting Sensor mounting with a 1 ½ ISO 2852 clamp (not included). Mount in a place where no air bubbles or sedimentation occur. The sensor has been calibrated in a measuring pipe with inside Ø180 mm. If the sensor will be mounted to a pipe smaller than Ø100 mm, recalibration is required. Valmet Corporation, D07740 V1.0 EN 04/2015 Technical data Materials PTFE and steel W (AISI 316L). Flange and cable connection: steel W (AISI 316L). Inside pressure gasket: SIL Pressure max. 12 bar at 150 C Cell constant Reproduceability ± 2 % Linearity 1.0 ± 0.7 % (electrically adjusted) Calibrated in a 180 mm pipe. From to 20 μs/cm ± 2 % per decade Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable Connection Weight PVC 12 x 0.25 mm 2 screened Standard length 5 m max. temp. 70 C 11-pole strip connector with MF20 cable adapter 1.2 kg For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation. dimensions in mm R=37 35 Typical mounting NA connect 38 mounting Standard ISO 1 ½ clamp mounting NAV 1-1 ø ½ ISO 10 4 For ISO clamp 1 ½ ø L=150 37

44 Sanitary Sensors 4193 & 4194 Sanitary open types 4-electrode conductivity sensors with external electrodes, for use with BSP 1 ¼ Ingold nipple. Used with TCU series 3100 and Mounting Sensor mounting with a 1 ¼ nut (included). Mount in a place where no air bubbles or sedimentation occur. Do not mount on the upper side of a horizontal pipe. If the sensor will be mounted to a small diamter pipe, recalibration is required. Technical data Type 4193 Type 4194 Materials L1 = 32 mm, L2 = 140 mm L1 = 42 mm, L2 = 140 mm PTFE, silicone rubber, Viton O-ring and steel W (AISI 316L). Flange and cable connection: steel W (AISI 316L). Pressure max. 12 bar at 150 C Cell constant Reproduceability ± 2 % Linearity 1.0 ± 0.5 % (electrically adjusted) Calibrated in an open vessel. From to 20 μs/cm ± 2 % per decade Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable Connection Weight PVC 12 x 0.25 mm 2 screened Standard length 5 m max. temp. 70 C 11-pole strip connector with MF20 cable adapter 1.5 kg dimensions in mm L2 33 Typical mounting 10 L ø36 ø18.5 ø R=30 Valmet Corporation, D07743 V1.0 EN 04/2015 Accessories 8977 Nipple BSP 1 ¼ For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation R=30

45 Industrial Sensor 4214 Industrial open compact type sensor 4-electrode conductivity sensor with external electrodes. Used with TCU series 3100 and Technical data Length Materials, sensor 83 mm PTFE, silicone rubber and steel W (AISI 316L) Pressure max. 12 bar at 150 C Cell constant Reproduceability ± 2 % 1.0 ± 0.5% (electrically adjusted); calibrated in a diam x 1.6 mm pipe. Linearity From to 20 μs/cm ± 2 % per decade. Below 5 μs/cm μs/cm Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2 standard length 5 m max. temp. 70 C Connection Weight 11-pole strip connector with MF20 cable adapter 1.0 kg diam diam

46 Accessories Slip-on flanges types 826x, 846x DIN Type SMS D Type DN D SMS slip-on flange D Typical mounting DIN 11851, DN ID Steel pipes ID49: Meas. error approx. +9% ID66: Meas. error approx. +3% ID73: Ref. diameter (0%) Use thread gasket D Plastic pipes D63: Meas. error approx. -5% D75: Meas. error approx. -3% D90: Ref. diameter (0%) Guide pin 8901 Tube must be placed on chamfered inside Packing on TOP SIDE of union nut! diam. 84 Valmet Corporation, D07439 V1.0 EN 04/2015 For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation.

47 Industrial Sensor 4215 Industrial open compact type sensor 4-electrode conductivity sensor with external electrodes. Used with TCU series 3100 and Technical data Length 0.25 m or extended to 0.5 / 1.0 / 1.5 / 2.0 / 2.5 m Materials, sensor PTFE, silicone rubber and steel W (AISI 316L) Pressure max. 12 bar at 150 C Cell constant 1.0 ± 0.5% (electrically adjusted); calibrated in a diam x 1.6 mm pipe. Reproduceability ± 2 % Linearity From to 20 μs/cm ± 2 % per decade. Below 5 μs/cm μs/cm Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2 Connection standard length 5 m max. temp. 70 C 11-pole strip connector with MF20 cable adapter diam diam Weight 1.2 kg Typical mounting ID Steel pipes ID49: Meas. error approx. +9% ID66: Meas. error approx. +3% ID73: Ref. diameter (0%) Strap to lock adjustable immersion depth L=0.5m Valmet Corporation, D07440 V1.0 EN 04/2015 For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation.

48 Industrial Sensor 4216 Open type sensor for sluice chamber 4-electrode conductivity sensor with external electrodes. Used with TCU series 3100 and Technical data Materials Sensor: PTFE, silicone rubber and W (AISI 316L) steel Sensor packing: O-ring EPDM NV41 Sluice valve packing: PTFE Pressure max. 12 bar at 150 C; during withdrawal max. 5 bar Cell constant 1.0 ± 0.5 % (electrically adjusted) Reproduceability ± 2 % Linearity From to 20 μs/cm ±2% per decade NV55 BSP 1.5 NPT Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. 302 Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2 standard length 5 m max. temp. 70 C Connection 11-pole strip connector with MF20 cable adapter ø29 Weight Sensor 2.9 kg Sluice chamber 3.0 kg dimensions in millimeters 46 R=37 67 Sluice chamber 8299 Typical mounting 1.5 ball valve for mounting of sensor and Valmet Corporation, D07710 V1.0 EN 04/2015 BSP 1 ½ max. 75 ø48 For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation. Dismount outer joint of ball valve before welding.

49 Industrial Sensor 4221 Durable industrial sensor 4-electrode conductivity sensor with internal electrodes. Used with TCU series 3100 and Technical data Length Materials 125 mm PTFE and steel W (AISI 316L) Pressure 25 bar at 210 C Cell constant Reproduceability Linearity 1.0 ± 0.5% (electrically adjusted) ± 1.5%, min. ± 8 ns/cm to 0.2 μs/cm ±1% per decade + 8 ns/cm Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 20 sec. Protection IP65 splash-proof (DIN ) Sensor cable Connection Weight PVC 12 x 0.25 mm 2 screened standard length 3 m, 5 m ~ 1 μs/ cm, 15 m ~ 5 μs/cm max. temp. 70 C 11-pole strip connector with MF20 cable adapter 0.7 kg 29.4 L= Ø3 Ø30 Ø25 75 Ø25 dimensions in millimeters Standard 11-pole strip

50 Typical mounting 70 Up Without guide pin Ø6.5 Weld tee Ø Flow assembly Type 71 Ø6 Type 72 Ø10 Guide pin 8933 Ø Ø Ø44.5** 8901 Ø Up 8935 Weld tee 8901 Guide pin 8936 Ø Ø54** 8938 Ø60.3 Without guide pin ¼ BSP 8903 Without guide pin Guide pin 1 ¼ Tee 8911 Ø33.6 min. 80** Ø44 min. 80** Up Valmet Corporation, D07552 V1.0 EN 04/2015 For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation.

51 Industrial Sensor 4221PE Industrial closed sensor for small pipes 4-electrode conductivity sensor with internal electrodes. Used with TCU series 3100 and Technical data Length Materials 125 mm PTFE and steel W (AISI 316L) Pressure 20 bar at 80 C Cell constant Reproduceability Linearity 1.0 ± 0.5% (electrically adjusted) ± 1.5%, min. ± 8 ns/cm to 0.2 μs/cm ±1% per decade. Below 5 μs/cm + 8 ns/cm Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 20 sec. Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.22 mm 2 type screened Connection Weight standard length 3 m max. temp. 70 C 11-pole strip connector with MF20 cable adapter 0.7 kg 29.4 L= Ø3 Ø30 Ø25 75 Ø25 dimensions in millimeters Standard 11-pole strip

52 Typical mounting 70 Up Without guide pin Ø6.5 Weld tee Ø Flow assembly Type 71 Ø6 Type 72 Ø10 Guide pin 8933 Ø Ø Ø44.5** 8901 Ø Up 8935 Weld tee 8901 Guide pin 8936 Ø Ø54** 8938 Ø60.3 Without guide pin ¼ BSP 8903 Without guide pin Guide pin 1 ¼ Tee 8911 Ø33.6 min. 80** Ø44 min. 80** Up Valmet Corporation, D07575 V1.0 EN 04/2015 For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation.

53 Industrial Sensor 4222 Durable industrial sensor 4-electrode conductivity sensor with internal electrodes. Used with TCU series 3100 and Technical data Length Materials, sensor 0.25 m or extended to required length, max. 2.5 m PTFE and steel W (AISI 316L) Pressure 25 bar at 210 C Cell constant Reproduceability Linearity 1.0 ± 0.5% (electrically adjusted) ± 1.5%, min. ± 8 ns/cm to 0.2 μs/cm ±1% per decade + 8 ns/cm Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 20 sec. Protection IP65 splash-proof (DIN ) Sensor cable Connection PVC 12 x 0.25 mm 2 screened standard length 3 m, 5 m ~ 1 μs/cm, 15 m ~ 5 μs/cm max. temp. 70 C 11-pole strip connector with MF20 cable adapter Weight min. 1.0 kg (0.25 m) 29.4 L = 0.25m m 136 Ø3 Ø30 Ø25 75 Ø25 dimensions in millimeters Standard 11-pole strip

54 Typical mounting Up Without guide pin ¼ BSP Weld tee 1 ¼ Tee Guide pin 8933 Ø Ø Ø44.5** 8901 Up Up 8935 Weld tee 8901 Guide pin 8936 Ø Ø54** 8938 Ø60.3 Valmet Corporation, D07560 V1.0 EN 04/2015 For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation.

55 Industrial Sensor 4223 Durable industrial sensor for sluice chamber 4-electrode conductivity sensor with internal electrodes. Used with TCU series 3100 and Technical data Materials Pressure Cell constant Reproduceability ± 1.5% Linearity sensor: PTFE and steel W (AISI 316L) packings: PTFE max. 15 bar at 150 C; during withdrawal max. 5 bar 1.0 ± 0.5% (electrically adjusted) to 0.2 μs/cm ±1% per decade + 8 ns/cm Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 20 sec. Protection IP65 splash-proof (DIN ) Sensor cable Connection Weight PVC 12 x 0.25 mm 2 screened standard length 3 m; cable extension 5 m ~ 1 μs/cm, 15 m ~ 5 μs/cm max. temp. 70 C 11-pole strip connector with MF20 cable adapter sensor 3.0 kg sluice chamber 3.5 kg dimensions in millimeters NV41 NV55 Ø29.6 Ø

56 Typical mounting - Dismount the outer joint of ball valve before welding! min Ø for D= for D= for D=105 D sluice valve chamber with 1.5 ball valve weld connection Ø48 1½ BSP for D= for D= for D= ½ Valmet Corporation, D07576 V1.0 EN 04/2015 For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation.

57 Industrial Sensor 4224 Durable industrial sensor 4-electrode conductivity sensor with external electrodes. Used with TCU series 3100 and Technical data Length 195 mm Materials Sensor of PTFE and steel W (AISI 316L) Pressure 25 bar at 210 C Cell constant 1.0 ± 0.5% (electrically adjusted) Reproduceability ± 1.5% Linearity to 20 μs/cm ±1% per decade Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 25 sec. Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2 screened standard length 5 m max. temp. 70 C Connection 11-pole strip connector with MF20 cable adapter Weight 1.1 kg dimensions in millimeters diam diam. 30 min diam. 25 L=

58 Typical mounting Alternative 1 Alternative DN Alternative 3 Alternative 4 Cable bend Use thread gasket Tube must be placed on chamfered inside Packing on TOP SIDE of union nut! Valmet Corporation, D07441 V1.2 EN 01/2016 For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation.

59 180 Industrial Sensor 4226 Durable industrial sensor for sluice chamber 4-electrode conductivity sensor with external electrodes. Used with TCU series 3100 and Technical data Materials Sensor: PTFE and steel W (AISI 316L) Sensor & sluice valve packings: PTFE Pressure Max. 15 bar at 150 C. During withdrawal max. 5 bar Cell constant 1.0 ± 0.5% (electrically adjusted) Reproduceability ± 1.5% Linearity to 20 μs/cm ±2% per decade Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 25 sec. Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2 screened Connection standard length 5 m max. temp. 70 C 11-pole strip connector with MF20 cable adapter NV41 NV55 ø24 ø29.6 min Type 270 ø48 Weight Sensor 3.0 kg Sluice chamber 3.5 kg Max. pressure / temperature dimensions in millimeters - at 15 bar 150 C - at 5 bar 200 C - at 216 lb 300 F - at 70 lb 390 F 270 sluice chamber with 1.5 ball valve Typical installation Type 270 Type 270

60 269 sluice valve with 1.5 ball valve Flanged sluice chambers Type b ø48 D d d1 L2 d4 K Typical installation L2 min Max. pressure during withdrawal 5 bar dimensions in millimeters Non-standard item Sluice chamber L2 D b k d1 d2 d4 Type 271 DIN DN40, PN Type 272 ANSI 1.5, 300 lb Type 273 DIN DN50, PN Type 274 ANSI 2, 150 lb Valmet Corporation, D07725 V1.0 EN 04/2015 For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation.

61 180 Industrial Sensor 4226 Durable industrial sensor for sluice chamber 4-electrode conductivity sensor with external electrodes. Used with TCU series 3100 and Technical data Materials Sensor: PTFE and steel W (AISI 316L) Sensor & sluice valve packings: PTFE Pressure Max. 15 bar at 150 C. During withdrawal max. 5 bar Cell constant 1.0 ± 0.5% (electrically adjusted) Reproduceability ± 1.5% Linearity to 20 μs/cm ±2% per decade Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 25 sec. Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2 screened Connection standard length 5 m max. temp. 70 C 11-pole strip connector with MF20 cable adapter NV41 NV55 ø24 ø29.6 min Weight Type 270 ø48 Sensor 3.0 kg Sluice chamber 3.5 kg Max. pressure / temperature dimensions in millimeters - at 15 bar 150 C - at 5 bar 200 C - at 216 lb 300 F - at 70 lb 390 F 270 sluice chamber with 1.5 ball valve Typical installation Type 270 Type 270

62 269 sluice valve with 1.5 ball valve Flanged sluice chambers Type b ø48 D d d1 L2 d4 K Typical installation L2 min Max. pressure during withdrawal 5 bar dimensions in millimeters Non-standard item Sluice chamber L2 D b k d1 d2 d4 Type 271 DIN DN40, PN Type 272 ANSI 1.5, 300 lb Type 273 DIN DN50, PN Type 274 ANSI 2, 150 lb Valmet Corporation, D07725 V1.0 EN 04/2015 For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation.

63 Industrial Sensor 4241 Closed compact type sensor for small pipes 4-electrode conductivity sensor with internal electrodes. Used with TCU series 3100 and Mounting Flow opening must be parallel to the flow direction. An arrow on sensoe indicates the position of the temperature sensor. Technical data Materials, sensor PVDF, EPDM, silicone rubber and steel W (AISI 316L). Flange & cable connection Steel W (AISI 316L) Temperature sensor Coated with PFA Pressure 10 bar at 130 C (150 C < 1 h) Cell constant 1.0 ± 0.5% (electrically adjusted) Reproduceability ± 2 % Linearity to 1 μs/cm ±2% per decade Below 5 μs/cm μs/cm Temp. sensor Pt 1000 (IEC 751 class A) Time delay T 90, 9 sec. Protection IP65 splash-proof (DIN ) Sensor cable PVC 12 x 0.25 mm 2 screened Standard length 5 m. Sensor can optionally be mounted with 3 or 15 m of cable. Measuring below 0.2 μs/cm with 3 m of cable is not recommended. Do not measure below 5 μs/cm with 15 m of cable. ø ø ø25 max. temp. 70 C Connection 11-pole strip connector with MF20 cable adapter dimensions in millimeters Weight 1.0 kg

64 40 DIN 11851, DN Od10 Guide pin Guide pin Guide pin PVC Tee NV40/ Guide pin ¼ BSP ¼ Tee dimensions in millimeters Valmet Corporation, D07731 V1.0 EN 04/2015 For more information, contact your local Valmet office. Specifications in this document are subject to change without notice. Product names in this publication are all trademarks of Valmet Corporation.

65 EN FR FI ES SV PT DE IT NL PL RU EU DECLARATION OF CONFORMITY VAATIMUSTENMUKAISUUSVAKUUTUS KONFORMITETSDEKLARATION KONFORMITÄTSERKLÄRUNG CONFORMITEITSVERKLARING ДЕКЛАРАЦИЯ СООТВЕТСТВИЯ DECLARATION DE CONFORMITE DECLARACION DE CONFORMIDAD DECLARAÇÃO DO CONFORMIDADE DICHIARAZIONE DI CONFORMITA DEKLARACJA ZGODNOŚCI Manufacturer Valmistaja Fabrikant Hersteller Fabrikant Производитель Fabricant Fabricante Fabricante Fabriccante Producent Device Laite Enheten Gerät Apparaat Устройство Artifice Dispositivo Dispositivo Dispositivo Urządzenie Valmet Automation Oy Kehräämöntie 3 / P.O.Box Kajaani, FINLAND Valmet Conductivity Measurements Series 2822 Conductivity Measurements Series 3000 Portable Conductivity Measurements Series 3100 Conductivity Measurements Series 3200 Dual Conductivity Measurements Series 3300 Concentration Measurements Series 3500 Causticizing Measurements with corresponding sensors This product is in conformity with the requirements of the following directives, and with standards and national legislation implementing these directives: Tuote täyttää seuraavien säännösten asettamat vaatimukset sekä nämä voimaansaattavat kansalliset päätökset: Denna produkt uppfyller kraven i följande direktiv och standarder samt nationell lagstiftning som inför dessa direktiv: Dieses Produkt erfüllt die folgenden Richtlinien und Normen und die nationalen Gesetze, die die Anwendung der obigen Richtlinien durchsetzen: Dit product is in overeenstemming met de eisen van de volgende richtlijnen en met normen en nationale wetgeving ter uitvoering van deze richtlijnen: Данное изделие соответствует требованиям следующих директив, а также стандартам и национальным законодательным актам, вводящим в действие данные директивы: Le produit est conforme aux directives et normes suivantes, et les dispositions nationales portant mise en œuvre desdites directives: Este producto es conforme a los requisitos de las siguientes directivas y normas y legislación nacional de aplicación de estas directivas: Este produto obedece aos requisitos das directivas e normas abaixo mencionadas e legislação nacional que transpõe as referidas directivas para o direito interno: Questo prodotto è conforme ai requisiti delle seguenti direttive, norme e standard e legislazione nazionale implementata sulla base di queste direttive: Ten produkt jest zgodny z wymaganiami następujących dyrektyw oraz ze standardami i przepisami krajowymi dotyczącymi wdrożenia tych dyrektyw: Loop-powered models: Line-powered models: 2014/30/EU 2002/95/EC 2012/19/EU 2014/30/EU 2014/35/EC 2002/95/EC 2012/19/EU (EMC Directive) (RoHS Directive) (WEEE Directive) (EMC Directive) (Low Voltage Directive) (RoHS Directive) (WEEE Directive) Juha Koistinen Vice President Valmet Automation Oy Control & Measurement Systems Signature Allekirjoitus Underskrift Unterschrift Handtekening Подпись Signature Firma Firma Assinatura Podpis Clarification of signature, position Nimenselvennys, asema Namnförtydligande, befattning Unterschrift in Maschinenschrift, Position Verduidelijking van handtekening, positie Расшифровка подписи, должность Nom en toutes letters, fonction Nombre y cargo Nome en função do assinante Firma in stampatello, mansione Imię i nazwisko, stanowisko

Valmet Concentration Measurement Series 3300

Valmet Concentration Measurement Series 3300 Valmet Concentration Measurement Series 3300 Installation & operating manual K15484 V1.1 EN Table of contents 1. About the device 5 4. System testing & adjustment 18 1.1. 1.2. 1.3. 1.4. 2. 2.1. 2.2. 2.3.

More information

Valmet Dual Conductivity Measurement Series 3200

Valmet Dual Conductivity Measurement Series 3200 Valmet Dual Conductivity Measurement Series 3200 Installation & operating manual K14278 V1.5 EN Table of contents 1. About the device 5 3. Software structure 15 1.1. 1.2. 1.3. 1.4. Introduction...5 Conductivity

More information

Valmet Cooking Liquor Measurement Series 3400

Valmet Cooking Liquor Measurement Series 3400 Valmet Cooking Liquor Measurement Series 3400 Installation & operating manual K14927 V1.1 EN Table of contents Warnings & safety information 4. System testing & adjustment 18 1. 1.1. 1.2. 1.3. 1.4. 1.5.

More information

Metso Conductivity for sensor series 9400 and 9500

Metso Conductivity for sensor series 9400 and 9500 Metso Conductivity Modules 2800 for sensor series 9400 and 9500 Metso Conductivity Modules 2800 Plug-in modules easy service Exchangeable scales with concentration reading Remotely controlled range switch

More information

JUMO tecline CR-4P Conductive 4-Electrode Conductivity Sensor

JUMO tecline CR-4P Conductive 4-Electrode Conductivity Sensor Page 1/6 JUMO tecine CR-4P Conductive 4-Electrode Conductivity Brief description The JUMO tecine CR-4P fills the gap between conductive conductivity measurement with two-electrode conductivity sensors,

More information

Inductive conductivity meter

Inductive conductivity meter Inductive conductivity meter Type 8228 can be combined with... Configurable outputs: up to 2 transistor and up to 2 analog 4... 20 ma outputs Removable backlighted display Simulation of process values

More information

Inductive conductivity meter

Inductive conductivity meter Inductive conductivity meter CIP version Type 8228 can be combined with... Standard version Perfect for concentrated liquids and wide conductivity range Pre-parameterized versions available for direct

More information

JUMO tecline Ci-S. Inductive conductivity and temperature sensor for general process engineering. Installation Instructions T94Z001K000

JUMO tecline Ci-S. Inductive conductivity and temperature sensor for general process engineering. Installation Instructions T94Z001K000 JUMO tecline Ci-S Inductive conductivity and temperature sensor for general process engineering Installation Instructions 20294200T94Z001K000 V1.00/EN/00582016 Contents Contents 1 Safety information................................................

More information

Digital ph transmitter

Digital ph transmitter Digital ph transmitter Compact, remote versions for DN 15 to 200 Fully programmable ph transmitter functions for all kinds of ph-measurement tasks Large range of process connections with various fittings

More information

Conductivity meter. PLC Solenoid valve. Type Type On/Off Diaphragm valve. Valve islands

Conductivity meter. PLC Solenoid valve. Type Type On/Off Diaphragm valve. Valve islands Conductivity meter Configurable outputs: two transistor and single or dual analog 4... 20 ma Removable backlighted display Universal process connection Type can be combined with... Three cell constants

More information

Magnetic-Inductive Flow Meter FMI

Magnetic-Inductive Flow Meter FMI SENSORS FOR FOOD AND BIO. Product information FMI-C, FMI-R FOOD Magnetic-Inductive Flow Meter FMI Application/Specified usage Magnetic-inductive flowmeter for the measurement of flow rate and volume in

More information

Continuous level sensor NSK

Continuous level sensor NSK SENSORS FOR FOOD AND BIOPHARMA. Product Information NSK-157, -357, -358 FOOD Continuous level sensor NSK Application Continuous level monitoring in metallic vessels up to 3 m in height Ideal for highly

More information

Digital ph transmitter

Digital ph transmitter Digital ph transmitter Compact, remote versions for DN 15 to 200 Fully programmable ph transmitter functions for all kinds of ph-measurement tasks Large range of process connections with various fittings

More information

SITRANS P measuring instruments for pressure

SITRANS P measuring instruments for pressure Overview Application The pressure transmitter is designed for the special requirements of the food, pharmaceutical and biotechnology industries. The use of high-grade materials guarantees compliance with

More information

SE 655 N-GEFFT0AM Digital Toroidal Conductivity Sensor

SE 655 N-GEFFT0AM Digital Toroidal Conductivity Sensor User Manual English SE 655 N-GEFFT0AM Digital Toroidal Conductivity Sensor Memosens Protocol Latest Product Information: www.knick.de 2 Safety Instructions WARNING Failure to observe this warning may result

More information

INDUCTIVE CONDUCTIVITY SENSOR. Instruction Manual. Bürkert 2001 Subject to technical change without notice

INDUCTIVE CONDUCTIVITY SENSOR. Instruction Manual. Bürkert 2001 Subject to technical change without notice INDUCTIVE CONDUCTIVITY SENSOR Instruction Manual Bürkert 00 Subject to technical change without notice INTRODUCTION Table of Contents. INTRODUCTION.... Symbols used.... General safety instructions....

More information

JUMO ecoline Ci Inductive conductivity and temperature sensor for general water engineering

JUMO ecoline Ci Inductive conductivity and temperature sensor for general water engineering Data Sheet 22943 Page 1/9 JUMO ecoline Ci Inductive conductivity and temperature sensor for general water engineering Brief description The sensor detects the electrolytic conductivity of a process liquid.

More information

Indumax P CLS50. Technical Information. Highly resistant, inductive conductivity sensor for standard, hazardous and high-temperature applications

Indumax P CLS50. Technical Information. Highly resistant, inductive conductivity sensor for standard, hazardous and high-temperature applications Technical Information Indumax P CLS50 Highly resistant, inductive conductivity sensor for standard, hazardous and high-temperature applications Application CLS50 conductivity sensors are especially suitable

More information

Magnetic-Inductive flow Meter fmi

Magnetic-Inductive flow Meter fmi 1 Product information fmi-c, fmi-r f o o d P H A R M A Magnetic-Inductive flow Meter fmi Application / Specified Usage Certification Magnetic-inductive flowmeter for the measurement of flow rate and volume

More information

Continuous level sensor NSK

Continuous level sensor NSK SENSORS FOR FOOD AND BIOPHARMA. Product Information NSK-187, -387, -388 FOOD Continuous level sensor NSK Application Continuous level monitoring in metallic vessels up to 2.5 m in height Ideal for highly

More information

JUMO ecoline Ci Inductive conductivity and temperature sensor for general water engineering

JUMO ecoline Ci Inductive conductivity and temperature sensor for general water engineering Data Sheet 22943 Page 1/9 JUMO ecoline Ci Inductive conductivity and temperature sensor for general water engineering Brief description The sensor detects the electrolytic conductivity of a process liquid.

More information

Conductivity Measuring Sensor InduMax H CLS 52

Conductivity Measuring Sensor InduMax H CLS 52 Technical Information TI 167C/24/ae Conductivity Measuring Sensor InduMax H CLS 52 Electrodeless conductivity sensor with quick temperature measurement and hygienic design The CLS 52 sensors are intended

More information

Octave Water Meter Installation Manual

Octave Water Meter Installation Manual Octave Water Meter Installation Manual 1.0 Notes For proper flow measurements, the OCTAVE s measuring tube should be completely full at all times. Non-wetted sensors show loss of signal. Though this will

More information

Electromagnetic flowmeters Series FLOMID Electromagnetic flowmeter for conductive liquids

Electromagnetic flowmeters Series FLOMID Electromagnetic flowmeter for conductive liquids Electromagnetic flowmeters Series FLOMID Electromagnetic flowmeter for conductive liquids Flow rate measurement is independent of density, temperature, viscosity and pressure Pulsed coil excitation to

More information

Compact System NRGS 11-2 NRGS Original Installation Instructions English

Compact System NRGS 11-2 NRGS Original Installation Instructions English Compact System NRGS 11-2 NRGS 16-2 EN English Original Installation Instructions 810366-05 1 Contents Important Notes Page Usage for the intended purpose...4 Safety note...4 LV (Low Voltage) Directive

More information

Miniature resistance thermometer For sanitary applications Model TR21-A, with flange connection

Miniature resistance thermometer For sanitary applications Model TR21-A, with flange connection Electrical temperature measurement Miniature resistance thermometer For sanitary applications Model TR21-A, with flange connection WIKA data sheet TE 60.26 Applications Sanitary applications Food and beverage

More information

Level Sensor with G1/2" Thread Hygienic

Level Sensor with G1/2 Thread Hygienic SENSORS FOR FOOD AND BIOPHARMA. Product Information NVS-141, -143, -146, -161, -163, -166, -181, -183, -186 FOOD Level Sensor with Thread Hygienic Application/intended use Limit detection of aqueous and

More information

Level Sensor with M12 Thread Hygienic

Level Sensor with M12 Thread Hygienic SENSORS FOR FOOD AND BIOPHARMA. Product Information NVS-041, -043, -046, -061, -063, -066, -081, -083, -086 FOOD Level Sensor with M12 Thread Hygienic Application/intended use Limit detection of aqueous

More information

JUMO tecline CR Electrolytic Conductivity Cells in stainless steel or titanium

JUMO tecline CR Electrolytic Conductivity Cells in stainless steel or titanium Page 1/10 JUMO tecline CR Electrolytic Conductivity Cells in stainless steel or titanium 202924 Series (formerly 2 EL6... Series) 2-electrode system for ranges from 0.05 µs/cm to 1 ms/cm large variety

More information

Magnetic-Inductive Flow Meter FMQ

Magnetic-Inductive Flow Meter FMQ SENSORS FOR FOOD AND BIO. Product information FMQ FOOD Magnetic-Inductive Flow Meter FMQ Application/Specified usage Magnetic-inductive flowmeter for the measurement of flow rate and volume in food and

More information

Installation Instructions Spector compact Conductivity Transmitter System LRGT 16-1

Installation Instructions Spector compact Conductivity Transmitter System LRGT 16-1 Installation Instructions 810701-00 Spector compact Conductivity Transmitter System LRGT 16-1 IMPORTANT NOTES...4 SAFETY NOTES...4 Danger...4 EXPLANATORY NOTES...4 Scope of supply...4 System description...4

More information

Magnetic Inductive Flow

Magnetic Inductive Flow GHM Messtechnik GmbH Hans-Sachs-Str. 26 93128 Regenstauf Germany Fon +49 (0) 9402-9383 - 0 Fax +49 (0) 9402-9383 - 33 www.ghm-messtechnik.de info@ghm-messtechnik.de Magnetic Inductive Flow Characteristics

More information

Signet Conductivity/Resistivity Electrodes

Signet Conductivity/Resistivity Electrodes Signet 2818-2823 Conductivity/Resistivity Electrodes 2818 2819 2820 2821 2822 2823 Sanitary Features Standard process connections - Polypro - SS on 10 and 20 cell - Tri-clamp 1-1½ in., 2 in. - Opt. ½ in.

More information

CombiTemp Temperature Measuring System

CombiTemp Temperature Measuring System CombiTemp Temperature Measuring ystem Flexible building block concept All wetted parts in acid-proof, stainless steel ome hygienic connections conform to -A standards, FDA demands and EHEDG guidelines

More information

ETP80. Electronic Transmitter for Pressure, Differential Pressure and Level

ETP80. Electronic Transmitter for Pressure, Differential Pressure and Level ETP80 Electronic Transmitter for Pressure, Differential Pressure and Level The ETP80 is a series of pressure transmitters with a wide range of s to suit most applications. The transmitters can measure

More information

Insertion magnetic inductive flowmeter

Insertion magnetic inductive flowmeter Insertion magnetic inductive flowmeter Sensor without moving parts Flowmeter with On/Off control Application related calibration by Teach-In function Clean in place (CIP) FDA-compliant materials Type 8041

More information

Electromagnetic Flow Monitor magphant

Electromagnetic Flow Monitor magphant Technical Information TI 036D/06/en No. 50078457 Electromagnetic Flow Monitor magphant Monitoring and measurement Flow monitoring with selectable limit values (relay output) Flow measurement via 4...20

More information

Pressure Transmitters for Sanitary Applications Model SA-11

Pressure Transmitters for Sanitary Applications Model SA-11 Electronic Pressure Measurement Pressure Transmitters for Sanitary Applications Model SA-11 WIKA Data Sheet PE 81.80 Applications Food and beverage industry Pharmaceutical industry Biotechnology industry

More information

SITRANS T measuring instruments for temperature

SITRANS T measuring instruments for temperature for installation in pipelines and tanks Overview The resistance thermometer is designed for installation in tanks and pipelines as well as for the measurement of temperature with hygiene requirements.

More information

SATRON VTe pressure transmitter

SATRON VTe pressure transmitter SATRON VTe pressure transmitter belongs to the series V-transmitters which will have both analog and smart properties. SATRON VTe is used for 0-4 kpa...0-00 MPa ranges. The transmitter communicates in

More information

SOLENOID DRIVEN DOSING PUMPS MX series

SOLENOID DRIVEN DOSING PUMPS MX series SOLENOID DRIVEN DOSING PUMPS MX series www.fwtsystems.it INTRODUCING MX SERIES FWT is pleased to introduce MX solenoid driven dosing pumps, a new series which will enable to widen the product range covering

More information

Chapter Page. 1 Introduction Description 1

Chapter Page. 1 Introduction Description 1 Chapter Page 1 Introduction 1 1.1 Description 1 Table of Contents 2 Calibration 3 2.1 Wet Calibration 3 2.2 Calibration Procedure 4 2.3 Installing Optional Display Board 5 2.4 Display Board Range Adjustment

More information

Operating instructions Binary level sensor LMT /00 06/2011

Operating instructions Binary level sensor LMT /00 06/2011 Operating instructions Binary level sensor LMT UK 706052/00 06/2011 Contents 1 Preliminary note...3 1.1 Symbols used...3 2 Safety instructions...3 3 Functions and features...4 3.1 Application area...4

More information

JUMO DELOS SI. Precision pressure transmitter with switching contacts and display. Application. Brief description. Key features.

JUMO DELOS SI. Precision pressure transmitter with switching contacts and display. Application. Brief description. Key features. Data sheet 405052 Page 1/10 JUMO DELOS SI Precision pressure transmitter with switching contacts and display Application Food & pharma CIP/SIP systems Machine and system construction Air conditioning and

More information

Compact Pressure Transmitter DAN-HH

Compact Pressure Transmitter DAN-HH 1 Product Information DAN-HH Compact Pressure Transmitter DAN-HH Application / Specified Usage Authorizations Pressure measurement in pipes and vessels High Temperature applications up to 150 C permanent

More information

FLS F6.60 MAGMETER FLOW SENSOR INSERTION FLOW SENSORS APPLICATIONS MAIN FEATURES

FLS F6.60 MAGMETER FLOW SENSOR INSERTION FLOW SENSORS APPLICATIONS MAIN FEATURES FLS F6.60 MAGMETER FLOW SENSOR The new F6.60 and F6.63 are flow meters without moving mechanical parts which can be applied for the measurement of dirty liquids so long as they are conductive and homogeneous.

More information

Electromagnetic flowmeters Series FLOMAT Insertion electromagnetic flowmeter for conductive liquids

Electromagnetic flowmeters Series FLOMAT Insertion electromagnetic flowmeter for conductive liquids Electromagnetic flowmeters Series FLOMAT Insertion electromagnetic flowmeter for conductive liquids For use in large diameter pipes as an economical solution for flow measurement Flow rate measurement

More information

HUMIDITY / TEMPERATURE TRANSMITTER

HUMIDITY / TEMPERATURE TRANSMITTER HUMIDITY / TEMPERATURE TRANSMITTER SEM161, SEM16 CONFIGURATION USING USB PORT POWERED CONFIGURATOR HUMIDITY, TEMPERATURE, DEW POINT MEASUREMENT INTERCHANGEABLE DIGITAL SENSOR HIGH ACCURACY AND STABILITY

More information

Signet Conductivity/Resistivity Electrodes

Signet Conductivity/Resistivity Electrodes Signet 2818-2823 Conductivity/Resistivity Electrodes 2818 2819 2820 2821 2822 2823 Sanitary Features Standard process connections - Polypro - SS on 10 and 20 cell - Tri-clamp 1-1½ in., 2 - Opt. ½ in. NPT

More information

Appendix. Clearance. Skewed profiles. Application Schematics

Appendix. Clearance. Skewed profiles. Application Schematics Appendix Application Schematics The purpose of the application schematics is to show different applications and the best sensor location for a particular application. Clearance Because the sensor will

More information

ABB Instrumentation. 50T Series Transmitters Model 54G/A gauge/absolute pressure transmitter for sanitary use. Specification sheet

ABB Instrumentation. 50T Series Transmitters Model 54G/A gauge/absolute pressure transmitter for sanitary use. Specification sheet Specification sheet 50T Series Transmitters Model 54G/A gauge/absolute pressure transmitter for sanitary use Ranges : - 100 to 4000 kpa - 1 to 40 bar - 14.5 to 600 psi Base accuracy : 0.2%(BFSL) Piezoresistive

More information

Pressure transmitter COMPACT ECOnomic

Pressure transmitter COMPACT ECOnomic Pressure transmitter COMPACT ECOnomic for food/pharmaceutical/biotechnology, Type series CA1110 Features Digitale pressure transmitter Hygienic design according to EHEDG Case and wetted parts of stainless

More information

Pressure transmitter for industrial applications Type MBS 4510

Pressure transmitter for industrial applications Type MBS 4510 MAKING MODERN LIVING POSSIBLE Data sheet Pressure transmitter for industrial applications Type MBS 4510 The high accuracy flush diaphragm pressure transmitter MBS 4510 is designed for use in non-uniform,

More information

XMP ci. Process Pressure Transmitter with HART -communication. Ceramic Sensor. accuracy according to IEC 60770: 0.1 % FSO.

XMP ci. Process Pressure Transmitter with HART -communication. Ceramic Sensor. accuracy according to IEC 60770: 0.1 % FSO. Process Pressure Transmitter with HART -communication Ceramic Sensor accuracy according to IEC 60770: 0.1 % FSO Nominal pressure from 0... 160 mbar up to 0... 20 bar Output signals 2-wire:... 20 ma others

More information

Pressure transmitters for industrial applications Type MBS 4510

Pressure transmitters for industrial applications Type MBS 4510 Pressure transmitters for industrial applications Type MBS 4510 Technical brochure Features Designed for use in severe industrial environments Enslosure and wetted parts of acid-resistant stainless steel

More information

ecotrans Lf 01/02 Microprocessor Transmitter / Switching Device for Conductivity

ecotrans Lf 01/02 Microprocessor Transmitter / Switching Device for Conductivity Data Sheet 20.27 Page / ecotrans Lf 0/02 Microprocessor Transmitter / Switching Device for Conductivity Type 2027 Housing for DIN rail mounting (5 x 7.5 mm to EN 0 75 A.) Brief description The JUMO ecotrans

More information

Dusty / Dusty Ex. Low-Cost Broken Bag Detection. Operating Instructions. SWR engineering Messtechnik GmbH

Dusty / Dusty Ex. Low-Cost Broken Bag Detection. Operating Instructions. SWR engineering Messtechnik GmbH EN Dusty / Dusty Ex Operating Instructions Low-Cost Broken Bag Detection SWR engineering Messtechnik GmbH CONTENTS Page 1. Introduction.............................................................. 3 1.1

More information

Flow measurement FlexFlow PF20H Hygienic flow sensor

Flow measurement FlexFlow PF20H Hygienic flow sensor Product highlights Parallel measurement of flow and temperature Flow measurement independent of the mounting position Large measuring range up to 400 cm/s Measurement at high media temperatures up to 125

More information

Operating instructions Flow monitors SI6600 SI6700 SI /02 01/2011

Operating instructions Flow monitors SI6600 SI6700 SI /02 01/2011 Operating instructions Flow monitors SI6600 SI6700 SI6800 UK 704680/02 01/2011 Contents 1 Preliminary note 3 1.1 Explanation of symbols 3 2 Safety instructions 3 3 Functions and features 4 3.1 Application

More information

Digital Pressure Gauge / Pressure Switch

Digital Pressure Gauge / Pressure Switch SENSORS FOR FOOD AND BIO. Product Information MAN-90-BAT, MAN-90P-BAT FOOD Digital Pressure Gauge / Pressure Switch Application Hygienic pressure measurement Overpressure save at highest accuracy Largest

More information

Rosemount 8750WA Magnetic Flowmeter System For Water and Wastewater Industries

Rosemount 8750WA Magnetic Flowmeter System For Water and Wastewater Industries Product Data Sheet January 214 813-1-475, Rev FA Rosemount 875WA Magnetic Flowmeter System For Water and Wastewater Industries THE 875WA MAGNETIC FLOWMETER Rosemount reliability in a customized offering

More information

Pressure chlorine changeover unit C 7520

Pressure chlorine changeover unit C 7520 BW 2 24 04 / 1 Content 1. Scope of delivery 2. Device description 3. Installation 4. Operation 5. Shutdown 6. Maintenance 7. Troubleshooting 1 Scope of delivery The chlorine gas changeover unit C 7520

More information

Metal tube variable area flowmeter for gases, liquids and steam

Metal tube variable area flowmeter for gases, liquids and steam Metal tube variable area flowmeter for gases, liquids and steam Metallic or plastic tube with a robust construction Indication by means of magnetic coupling Scales calibrated in l/h, m 3 /h, kg/h, t/h,

More information

Datasheet PDCSY-MW-CHM. Technical Overview. Features. Product warranty and total quality commitment. General Information.

Datasheet PDCSY-MW-CHM. Technical Overview. Features. Product warranty and total quality commitment. General Information. Datasheet Compact Heat Meters Technical Overview Heat energy is calculated by using a matched pair of high accuracy sensors to measure the difference between the forward and flow temperatures. The amount

More information

Magnetic-inductive flow meter EPS / UMF2. Technical data sheet

Magnetic-inductive flow meter EPS / UMF2. Technical data sheet Magnetic-inductive flow meter EPS / UMF2 Technical data sheet high accuracy: 0.3 % of actual maintenance-free no pressure drop numerous lining materials numerous electrode materials low-cost grounding

More information

Analytical Instruments Two-Electrode Conductivity Sensors

Analytical Instruments Two-Electrode Conductivity Sensors Data Sheet Analytical Instruments Two-Electrode Conductivity Sensors Model TB2 Superior accuracy in low conductivity measurement ranges. Measurement ranges from 0 to 199.9 µs/cm to 0 to 19.99 ms/cm. Resolution

More information

Technical specification

Technical specification 15. Nov. 2010 Page 1 (5) Metso Rotating Sensor part - Rotating shear force based transmitter for pulp consistency measurement in pulp and paper industry. - Sensing element selection based on pulp type

More information

CD34CNFLFxxx. Capacitive sensor, Foreground Suppression. Main features. Main functions. Description

CD34CNFLFxxx. Capacitive sensor, Foreground Suppression. Main features. Main functions. Description Capacitive sensor, Foreground Suppression Main features Compact housing Supply voltage: 10 to 30 VDC Output: 100 ma, NPN or PNP preset Make or break switching function LED indication for output and power

More information

Level limit switch Series RF 3000 Selection list

Level limit switch Series RF 3000 Selection list Table of content Page Overview P2 Applications P4 ------------------------------------------------------------------------------------------------------------- RF 3100 Standard version P6 -------------------------------------------------------------------------------------------------------------

More information

JUMO dtrans p20. Process pressure transmitter. Type Brief description. Key features. Block diagram

JUMO dtrans p20. Process pressure transmitter. Type Brief description. Key features. Block diagram Data Sheet 403025 Page 1/10 JUMO dtrans p20 Process pressure transmitter Brief description The JUMO dtrans p20 pressure transmitter with HART interface combines maximum precision with easy operation. It

More information

Signet 2551 Magmeter Flow Sensor

Signet 2551 Magmeter Flow Sensor Signet 2551 Magmeter Flow Sensor Available in a variety of wetted materials and ideal for pipe sizes up to DN900 (36 in.) Features Test certificate included for -X0, -X1 Patented Magmeter technology No

More information

Full Profile Insertion

Full Profile Insertion FPI Mag Meter Full Profile Insertion High Performance Electromagnetic Flow Meter Municipal Market Product Presentation Manufactured exclusively by McCrometer, Inc. Hemet, CA USA 55+ Years of Experience

More information

+ = Angle-seat Valve System for continuous control. Type 8802/2300 electro-pneumatically operated angle-seat valve

+ = Angle-seat Valve System for continuous control. Type 8802/2300 electro-pneumatically operated angle-seat valve Angle-seat Valve System for continuous control Type 8802/2300 electro-pneumatically operated angle-seat valve + = TWO YEAR WARRANTY on this valve system Burkert 8802 continuous control valve + = 2300 series

More information

Paddle-wheel flow controller for On/Off control

Paddle-wheel flow controller for On/Off control Paddle-wheel flow controller for On/Off control Type 8032 can be combined with... Indication, monitoring, transmitting and On/Off control in one device Programmable outputs (transistor or relay) Automatic-calibration:

More information

Pressure measurement PP20H Fully welded hygienic pressure sensor with flush membrane

Pressure measurement PP20H Fully welded hygienic pressure sensor with flush membrane Product highlights All market typical hygienic connections available EHEDG & 3-A certified versions available Condensate-proof measuring cell IO-Link available Space-saving installations from DN 25 on

More information

Positive displacement batch controller

Positive displacement batch controller 8075 Batch controller Positive displacement batch controller Type 8075 can be combined with... Compact version for DN15 to DN100 Dosing On site calibration by Teach-In Check of input/output signals Total

More information

Differential Pressure Sensors

Differential Pressure Sensors 1 910 1910P01 Differential Pressure Sensors for air or nonaggressive gases QBM66... Pressure-linear characteristic with selectable pressure measuring range Operating voltage AC 24 V or DC 13.5...33 V Output

More information

OD OD DIGITAL DISSOLVED OXYGEN CONTROLLERS

OD OD DIGITAL DISSOLVED OXYGEN CONTROLLERS UNI EN ISO 9001:2000 CERT. N. 9115 BCEL OPERATOR S MANUAL OD 545.2 - OD 565.2 DIGITAL DISSOLVED OXYGEN CONTROLLERS Rev. A Valid from S/N 28421 Input: Scales: Power supply: from polarographic cells 0/19.99

More information

... PureLine PQ. Technical Specification. PureLine PQ (Disinfection) Models PQ PQ UV Chamber. Cabinet

... PureLine PQ. Technical Specification. PureLine PQ (Disinfection) Models PQ PQ UV Chamber. Cabinet PureLine PQ PureLine PQ (Disinfection) Models PQ0200 - PQ 1100... Technical Specification UV Chamber Material: Internal finish: Stainless steel 316L

More information

Pressure transducer DMU 01K Compact version

Pressure transducer DMU 01K Compact version Pressure transducer DMU 01K Compact version Dimensions (in mm) Wiring diagram 8 Application Special OEM unit. For electronic pressure measurement in industrial or HVAC applications (such as hydraulic,

More information

SATRON VDt Differential Pressure Transmitter

SATRON VDt Differential Pressure Transmitter SATRON VDt differential pressure transmitter belongs to V-transmitter family. The series V transmitters have both analog and smart properties. SATRON VDt is used for -,1kPa...-15 MPa ranges. It is a 2-wire

More information

Compact Heat Meters. Features. -A Pulsed output. -B M-Bus output. Accessories. UK Sales Tel: International Tel:

Compact Heat Meters. Features. -A Pulsed output. -B M-Bus output. Accessories. UK Sales Tel: International Tel: Compact Heat Meters Features Compact design Simple operation Pulsed output Measures heating or cooling Specification Product Codes Water Meter Temp. range 10 to 90 C Nominal pressure 16bar Installation

More information

LTX RF LEVEL SENSOR. Instruction Manual

LTX RF LEVEL SENSOR. Instruction Manual LTX RF LEVEL SENSOR Instruction Manual FOR MODELS LTX01, LTX02, LTX05 Intempco Document No: LTX - M01 Rev. 1 Issue Date: April 2005 LTX01 RF LEVEL SENSOR USER MANUAL Software Rev : Rev. Date : June 2004

More information

ELADOS EMP III Diaphragm Pump type

ELADOS EMP III Diaphragm Pump type ELADOS EMP III Diaphragm Pump type 00800-01200 Mechanical stroke adjustment Stroke frequency regulation Pulse control Pulse multiplication and division Pulse memory Standard signal control Empty signal

More information

JUMO ecotrans Lf 01/02

JUMO ecotrans Lf 01/02 Data Sheet 07 Page /7 JUMO ecotrans Lf 0/0 Microprocessor Transmitter/Switching Device for Conductivity Housing for DIN rail mounting (5 mm 7.5 mm to EN 075 A.) Brief description The JUMO ecotrans Lf 0/0

More information

Oscillating piston flowmeters Series COVOL Positive displacement flowmeter for liquids

Oscillating piston flowmeters Series COVOL Positive displacement flowmeter for liquids Oscillating piston flowmeters Series COVOL Positive displacement flowmeter for liquids Suitable for all kind of liquids Viscosity up to 120000 mpa s Excellent performance with changing process conditions

More information

HBLT-Wire LEVEL SENSOR For analogue measurements of NH3 and HFC in refrigeration systems

HBLT-Wire LEVEL SENSOR For analogue measurements of NH3 and HFC in refrigeration systems Instruction manual HBLT-Wire LEVEL SENSOR For analogue measurements of NH3 and HFC in refrigeration systems Instruction manual HBLT-wire Levelsensor (006-UK) 1 / 11 Table of contents Safety Instructions...

More information

Pipe surface resistance thermometer, for clamping Model TR57-M, miniature design

Pipe surface resistance thermometer, for clamping Model TR57-M, miniature design Electrical temperature measurement Pipe surface resistance thermometer, for clamping Model TR57-M, miniature design WIKA data sheet TE 60.57 Applications Sanitary applications Food and beverage industry

More information

General Purpose Conductivity Sensors ENDURANCE. Model 400 and 400VP Series. Product Data Sheet PDS /rev.G February 2006

General Purpose Conductivity Sensors ENDURANCE. Model 400 and 400VP Series. Product Data Sheet PDS /rev.G February 2006 Product Data Sheet PDS 71-400/rev.G February 2006 Model 400 and 400VP Series ENDURANCE General Purpose Conductivity Sensors INITIAL CALIBRATION NOT REQUIRED. Pre-determined cell constant ensures out-of-the

More information

Instruction Manual Portamess 911 Cond

Instruction Manual Portamess 911 Cond Instruction Manual Portamess 911 Cond ProMinent Dosiertechnik GmbH D-69123 Heidelberg Germany ProMinent Dosiertechnik GmbH Im Schuhmachergewann 5 11 D-69123 Heidelberg Postfach 10 17 60 D-69007 Heidelberg

More information

유량계전문제조업체. Electro Magnetic FLOWMETER

유량계전문제조업체. Electro Magnetic FLOWMETER 유량계전문제조업체 w w w. f c k o r e a i n s. c o m Electro Magnetic FLOWMETER ELECTRO-MAGNETIC FLOW METER Contents Principle of Operation 04 Grounding Method 05~07 Piping Installation 08 CN-400 CNT-2 Flanged

More information

Diaphragm monitoring system with clamp connection For sanitary applications Model DMS-FP

Diaphragm monitoring system with clamp connection For sanitary applications Model DMS-FP Pressure Diaphragm monitoring system with clamp connection For sanitary applications Model DMS-FP WIKA data sheet DS 95.20 Applications Hygienic pressure measurement for the pharmaceutical industry and

More information

Reliable fluid detection with non-invasive ultrasonic measurement

Reliable fluid detection with non-invasive ultrasonic measurement Technical specification PIOX S502ID Reliable fluid detection with non-invasive ultrasonic measurement Permanently installed ultrasonic measuring system for the non-invasive detection of one of 2...5 fluids

More information

Differential pressure sensor

Differential pressure sensor s 1 916 QBM3x20-.. QBM3x20-..D Differential pressure sensor for air and non-aggressive gases QBM3020-.. QBM3120-.. High degree of measuring accuracy Adjustable characteristic curve (pressure-linear or

More information

FLOWX3 F3.60M.S Insertion Magmeter

FLOWX3 F3.60M.S Insertion Magmeter FLOWX3 F3.60M.S Insertion Magmeter The FLOWX3 F3.60M.S Insertion Magmeters are suitable to measure flowrate in both metal and thermoplastic pipelines. No moving mechanical parts and the high quality materials

More information

SANITARY PRESSURE GAUGES

SANITARY PRESSURE GAUGES 117 SANITARY PRESSURE GAUGES Type 2030 Digital Sanitary Gauge...119 Type 1032, 2 1 /2, 3 1 /2 and 4 1 /2 Gauge...120 Type 1036 3 1 /2 Gauge w/type 1037 Sanitary Fitting...121 Type 1032, 2 Fractional Gauge...122

More information

Signet Pressure Transmitter

Signet Pressure Transmitter Signet 80 Pressure 80.090 80.090 Rev. J 0/ English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents. Installation.

More information

Magnetic float switch For vertical installation Model FLS (models with Ex approval: 60, AL-ADF)

Magnetic float switch For vertical installation Model FLS (models with Ex approval: 60, AL-ADF) Level measurement Magnetic float switch For vertical installation Model FLS (models with Ex approval: 60, AL-ADF) TC Fluid Control WIKA data sheet LM LM 3001 30.01 for further approvals see page 3 Applications

More information

Mounting and Operating Instructions EB 3913 EN. Electropneumatic Converters i/p Converter Type

Mounting and Operating Instructions EB 3913 EN. Electropneumatic Converters i/p Converter Type Electropneumatic Converters i/p Converter Type 3913-0001 Fig. 1 Type 3913-0001 Electropneumatic Converter with pressure gauge and bracket Mounting and Operating Instructions EB 3913 EN Edition August 2011

More information

DMP 331P DMP 331P. Industrial Pressure Transmitter. Pressure Ports And Process Connections With Flush Welded Stainless Steel Diaphragm

DMP 331P DMP 331P. Industrial Pressure Transmitter. Pressure Ports And Process Connections With Flush Welded Stainless Steel Diaphragm DMP P Industrial Pressure Transmitter Pressure Ports And Process Connections With Flush Welded Stainless Steel Diaphragm accuracy according to IEC 60770: standard: 0.5 % FSO option: 0.5 % FSO Industrial

More information