Installation Instructions Spector compact Conductivity Transmitter System LRGT 16-1

Size: px
Start display at page:

Download "Installation Instructions Spector compact Conductivity Transmitter System LRGT 16-1"

Transcription

1 Installation Instructions Spector compact Conductivity Transmitter System LRGT 16-1

2 IMPORTANT NOTES...4 SAFETY NOTES...4 Danger...4 EXPLANATORY NOTES...4 Scope of supply...4 System description...4 Function...5 INSTALLATION...5 Measuring electrode...5 Reduction of length of installation and measuring length (control range)...6 ALIGNING, REMOVING AND MOUNTING OF TERMINAL BOX...8 Align terminal box...8 Remove terminal box...8 Install terminal box...8 WIRING...9 Connecting conductivity transmitter...9 Wiring diagram...10 COMPONENT PARTS...16 COMMISSIONING PROCEDURE...12 Check wiring...12 Set code switch...12 Factory setting...12 Adjust measuring length (control range) and actual value output...12 Check temperature coefficient Tk setting...12 Adjust cell constant...13 Functional test...13 LED display and fault indication

3 TROUBLESHOOTING...15 Fault finding list...15 Changing electronic insert...16 TECHNICAL DATA...16 DIMENSIONS OF LRGT APPLICATION AS CONDUCTIVITY LIMITER AND CONTINUOUS BLOWDOWN CONTROLLER...20 INDUSTRIAL CONTROLLER KS Component parts / front view...20 Designs...20 Wiring...21 Operating structure...22 Configuration level...22 Parameter level...25 Operating level

4 Important Notes Safety Note Use conductivity transmitter system LRGT 16-1 only for monitoring electrical conductivity in liquids. The equipment has to be supplied with 24 V DC from a safety power supply unit (PSU) in accordance with DIN VDE 0106 (safe supply isolation). The PSU must be equipped with an overcurrent protective device according to DIN EN /VDE The transmitter system LRGT 16-1 must only be installed by qualified staff. Qualified staff are those persons who - through adequate training in electrical engineering, the use and application of safety equipment in accordance with regulations concerning electrical safety systems, and first aid & accident prevention have achieved a recognised level of competence appropriate to the installation and commissioning of this critical safety device. Danger When loosening the conductivity transmitter system steam or hot water might escape. This presents the danger of scalding. It is therefore essential not to dismantle the conductivity transmitter system unless the boiler pressure is verified to be zero. Explanatory Notes Scope of Supply LRGT Control electrode with joint ring D 33 x 39 DIN E8P 1 Terminal box with conductivity transmitter 1 Installation manual System Description The LRGT 16-1 is a compact-design conductivity transmitter system consisting of a conductivity electrode, a temperature sensor for detecting the fluid temperature and an integrated conductivity transmitter. The equipment works according to the conductivity measurement two-electrode control principle and provides a conductivity-dependent and temperature compensated measuring current of 4-20 ma. The conductivity transmitter system is designed to measure the electrical conductivity in conductive fluids. If the LRGT 16-1 is connected to the industrial controller type KS 90 the equipment can be used as conductivity limiter and/or continuous blowdown controller in steam boilers, (pressurised) hot-water boilers and condensate systems. Type Approvals TÜV WÜL EG BAF-MUC

5 Function A measuring current of variable frequency flows through the fluid and creates a potential gradient between the measuring electrode and the measuring tube. The potential gradient is evaluated as measuring voltage U U. In order to compensate for polarisation, which would reduce the measuring current, the measuring frequency is adapted accordingly. The resulting measuring current is sensed and transmitted as measuring voltage U I. The electrical conductivity changes as a function of temperature. The resistance thermometer integrated in the electrode measures the fluid temperature and establishes a reference temperature on which the measured values will be based. The electrical conductivity is calculated from the measuring voltages U U and U I and - depending on the adjusted temperature coefficient Tk - is based linearly on the reference temperature 25 C. After converting it into a conductivity-proportional current signal, a current of 4-20 ma is provided for external processing. The cables leading to the measuring electrode, the measuring tube and the resistance thermometer are monitored for interruption and short circuits. In addition, the electronic insert is protected against excessive temperatures in the terminal box. In case of a malfunction the LEDs are permanently illuminated or flashing and the current signal is set to 0 or 0.5 ma. The transmitter is parameterised via code switch. The code switch can also be used for adaptation of the cell constant and triggering a functional test. The electrical conductivity is measured in µs/cm. In some countries use is also made of ppm (parts per million). Conversion: 1 µs/cm = 0.5 ppm. Installation The conductivity transmitter system is designed such that it can be screwed directly into boiler standpipes or suitably tapped flanges. The electrode must be installed such that it is always submerged. The installation in protection tubes is not permitted according to WÜ 100. Ensure a minimum distance of approx. 20 mm between the lower part of the measuring tube and the boiler wall, the smoke tubes, any other metal parts and the low water level (LW). Installation of the Measuring Electrode 1. Check seating surfaces of threads or flange provided on vessel or boiler standpipe. 2. Place ring joint onto the seating surface of the transmitter. Note: Use only ring joint (of stainless steel E8P) D 33 x 39 to DIN 7603 supplied with the electrode. 3. Apply a light smear of silicone grease (e. g. DOW Corning 111 Compound) to electrode thread. 4. Screw conductivity transmitter electrode into threads or flange provided on vessel or boiler standpipe and tighten with a 41 mm open-end spanner. The torque required when cold is 150 Nm. The tapped hole must conform to DIN 3852 part 2, 1" BSP (DIN ISO 228) form X or Y, standard version. The subsequent hole must have a diameter of > 24 mm. 5

6 Reducing length of installation and measuring length (control range) The measuring tube and the electrode can be cut to suit the individual requirements of the installation. Length of installation A [mm] Measuring length L [mm] Possible reduction of length R [mm] 1. Carefully determine the required reduction of length - measured from the bottom end of the measuring tube - and mark this on the measuring tube. 2. Only slacken locking grub screw using a screwdriver size 1.3 for socket-head cap screws. 3. Unscrew measuring tube and cut it to the length marked. 4. Unscrew spacer disc. 5. Mark the reduction of length - measured from the extreme end of the tip - on the electrode tip. 6. Cut electrode tip to length. Cut back mm of the insulation. 7. Screw back the spacer disc onto the electrode tip and fasten fingertight with a 12 mm open end spanner. 8. Screw back measuring tube with a torque of 50 Nm and tighten locking grub screw. Attention! The seating surfaces and threads on the vessel and mounting flanges must be accurately machined.! Do not bend the electrode tip when cutting. Do not subject electrode to physical shocks.! Do not lag the electrode body.! The conductivity transmitter system can be installed horizontally or with a vertical inclination. The measuring surface must be permanently submerged. Tools! 41 mm open-end spanner! 12 mm open-end spanner! Screwdriver size 1.3 for socket-head cap screws! Hacksaw! Flat file, cut 2 6

7 Reducing length of installation and measuring length (control range) 7

8 Aligning, removing and mounting the terminal box Align terminal box 1. Undo cover screws (cross-recess head screws M4) and remove one terminal cover. The arrow on the name plate points towards this cover. 2. Loosen fixing nut using 18 (19) mm open-end spanner and turn terminal box into desired direction (cable outlet). 3. Re-tighten fixing nut with a torque of 30 Nm. 4. Re-install cover, tighten cover screws, ensuring correct seat of the cover gasket. If required you can also remove the terminal box from the electrode and re-install it afterwards. Remove terminal box 1. Loosen cover screws (cross-recess head screws M4) and remove one terminal cover. The arrow on the name plate points towards this cover. 2. Undo fixing screws for cover plate and remove plate. 3. Pull electrode cables from terminal lugs (electronic insert, functional earthing). 4. Loosen fixing nut with 18 (19) mm open-end spanner and remove it. 5. Pull electrode cables through fixing nut. 6. Remove terminal box. Pull all electrode cables through hole for fixing screw. Install terminal box 1. Pull all electrode cables through hole for fixing screw in terminal box. 2. Place terminal box onto electrode and turn it into the desired direction, taking the cable outlet into consideration. Ensure correct seat of the gasket between the electrode and the terminal box. 3. Pull all electrode cables through fixing nut. Screw nut on fixing bolt and tighten with a torque of 30 Nm. 4. Connect electrode cables according to wiring diagram to terminal lugs (electronic insert, functional earthing). 5. Screw back cover plate. Install terminal cover and tighten cover screws, ensuring correct seat of cover gasket. 8

9 Wiring Connect conductivity transmitter The equipment has to be supplied with 24 V DC from a safety power supply unit (PSU) in accordance with DIN VDE 0106 (safe supply isolation). The PSU must be equipped with an overcurrent protective device according to DIN EN /VDE The following steps 1-5 do not apply for equipment fitted with a connector. For wiring use flexible screened control cable, min. conductor size 0.75 mm². 1. Remove terminal strip and pull connecting cable through cable gland. 2. Strip off approx. 40 mm of cable insulation coating and remove approx. 5 mm of conductor end insulation. 3. Carry out wiring in accordance with wiring diagram. 4. Plug in terminal strip. 5. Seal cable entry by tightening cable glands. Use sealing plug to seal off cable glands that are not used. 6. Install terminal cover and tighten cover screws, ensuring correct seat of cover gasket. Tools! Screwdriver size 1 for cross-recess head screws! Screwdriver size 2.5 for slotted screws, completely insulated to VDE 0680! 18 (19) mm open-end spanner 9

10 Wiring diagram Version with terminal strip: Connection of terminal lugs 1-6 (electrode cables) 1, 2 Measuring electrode 3, 4 Measuring tube (mass) 5, 6 Resistance thermometer 10

11 Functional elements Representation without cover plate 11

12 Commissioning Procedure Check wiring 1. Make sure that the conductivity transmitter system is wired in accordance with the wiring diagram. 2. Check that the supply voltage corresponds to the name plate rating. Set code switch The transmitter is parameterised by means of a ten-pole code switch. This code switch can also be used to adjust the cell constant and trigger off the functional test. The individual switch positions can be changed by using e. g. a ball pen. Factory setting S1 S2 S3 S4 S5 S6 S7 S8 S9 S10 ON ON ON OFF ON ON OFF OFF OFF OFF In the following table the factory setting is highlighted in grey: Adjust measuring length (control range) and actual value output Code switch Control range Current output ma = µs/cm [µs/cm at 25 C] 4 ma correspond to 20 ma correspond to OFF OFF OFF ON OFF OFF OFF ON OFF ON ON OFF OFF OFF ON ON OFF ON OFF ON ON ON ON ON During the commissioning procedure (when the supply voltage is applied) the first output will be 4 ma, then the output current increases until it reaches the actual value. Check temperature coefficient Tk setting For a linear compensation of the measured conductivity value based at 25 C the temperature coefficient Tk is set at our works to 2.1 % per C. Once the service temperature is reached you can check this setting by carrying out a comparison measurement - e. g. as part of the commissioning procedure. If there is a deviation between the indicated conductivity and the measured comparison value you can correct the measurement result of the transmitter by adjusting a lower or higher temperature coefficient. Continue modifying the Tk setting until the indicated conductivity value and measured comparison value agree. Code switch Temperature coefficient Tk [%/ C] OFF OFF OFF OFF 0 (no compensation) ON OFF OFF OFF 1.6 OFF ON OFF OFF 1.7 ON ON OFF OFF

13 Code switch Temperature coefficient Tk [%/ C] OFF OFF ON OFF 1.9 ON OFF ON OFF 2.0 OFF ON ON OFF 2.1 ON ON ON OFF 2.2 OFF OFF OFF ON 2.3 ON OFF OFF ON 2.4 OFF ON OFF ON 2.5 ON ON OFF ON 2.6 OFF OFF ON ON 2.7 ON OFF ON ON 2.8 OFF ON ON ON 2.9 ON ON ON ON 3.0 Adaptation of the cell constant The cell constant adjusted at our factory is a characteristic geometric quantity of the equipment and influences the calculation of conductivity. However, during operation this constant can drift, e. g. due to dirt accumulated on the measuring electrode. If the indicated conductivity value deviates from the measured comparative value you can correct the measurement result of the transmitter by re-adjusting the cell constant. Depending on the deviation set code switches 8 or 9 to ON, the cell constant will then change continuously. Observe the conductivity indication and when the indicated value agrees with the comparison measurement set the switch back to OFF. Should an adjustment not be possible remove the transmitter and clean the measuring surface. To restore the factory setting of the cell constant - e. g. after cleaning the measuring surface - set code switches 8 and 9 simultaneously to ON. After approx. 1 sec. switch back to OFF. The factory setting is restored. Code switch Deviation of the indicated 8 9 Function conductivity non OFF OFF No change Indicated value < Measured comparative ON OFF Cell constant increases value Indicated value > Measured comparative OFF ON Cell constant decreases value ON ON Restores factory setting Functional Test Use code switch 10 for performing a functional test. A value is simulated which exceeds the final control range value and a current output of 20 ma is provided. Code switch 10 OFF ON Functional test Normal operation Simulation Final control range value exceeded When the functional test is finished set code switch back to OFF. 13

14 LED Display and Fault Indications A green and a red LED for status and fault indication are provided in the centre of the electronic insert. Normal operation green LED red LED Current output [ma] Conductivity 0 to +10 % of control range lit proportional to measured value Conductivity 90 to 100 % of control range lit proportional to measured value Conductivity 10 to 90 % of control range lit lit proportional to measured value Fault green LED red LED Current output [ma] Value below 0 % setting, flashes 4 ma e. g. electrode exposed Connections to measuring tube/electrode interrupted flashes 0 ma Value above 100 % setting, e. g. control range too small flashes 0 ma Connections to measuring tube/electrode shortcircuited Temperature in terminal box exceeds 85 C flashes 0 ma flashes flashes 0 ma Connection to resistance thermometer interrupted or short-circuited flash alternately 0.5 ma 14

15 Troubleshooting Fault Finding List Equipment does not work Fault: Remedy: Fault: Remedy: Supply voltage failure - no function. Check power supply and all electrical connections. Electronic insert defective. Replace electronic insert. Equipment signals malfunction LED display: Red LED flashing. Current output: 4 ma Fault: Value below 0 % setting, e. g. electrode exposed. Remedy: Check water level and installation. LED display: Red LED flashing. Current output: 0 ma Fault: Connection to measuring tube/electrode interrupted. Remedy: Check electrode cables (electronic insert, terminal lugs 1-4). Replace equipment if necessary. LED display: Green LED flashing. Current output: 0 ma Fault: Value above 100 % setting, e. g. control range too small. Remedy: Adjust larger control range. LED display: Red LED flashing. Current output: 0 ma Fault: Connections to measuring tube/electrode short-circuited. Remedy: Check electrode cables (electronic insert, terminal lugs 1-4). Replace equipment if necessary. LED display: Red and green LEDs flash alternately. Current output: 0 ma Fault: Temperature in terminal box exceeds 85 C. Remedy: Check ambient temperature. Make sure it does not exceed 70 C. LED display: Red and green LEDs flash alternately. Current output: 0.5 ma Fault: Electrical connections to resistance thermometer interrupted or short-circuited or thermometer defective. Remedy: Check thermometer connections (electronic insert, terminal lugs 5 + 6). Replace equipment if necessary. 15

16 Inaccurate readings Fault: Remedy: Indicated conductivity higher than measured comparison value. During commissioning: Reduce temperature coefficient Tk. During operation: Reduce cell constant. Fault: Remedy: Fault: Remedy: Indicated conductivity lower than measured comparison value. During commissioning: Increase temperature coefficient Tk. During operation: Increase cell constant. The measurement result can no longer be adjusted by changing the cell constant. Remove equipment and clean measuring surface. Exchanging electronic insert 1. Undo lid screws and remove both housing lids. 2. Unscrew fixing screws of cover plate and remove the plate. 3. Pull electrode wires from terminal lugs on circuit board. Plug out the lower terminal strip. 4. Undo fixing screws of electronic insert and remove insert. The electronic insert is available as spare part (type LRV 1-40). When ordering please state the version (VS) and material number indicated on the name plate. 5. Install the new electronic insert in reverse order. Tools! Screwdriver size 1 for cross-recess head screws.! Screwdriver size 2.5 for slotted screws, completely insulated to VDE 0680 Technical data Type code Type L R G T Conductivity measurement Closed-loop control Sensor Transmitter Type of measurement (conductimetry = 1) Pressure rating, PN 40 = 6 Design series Service pressure 32 bar at 238 C 16

17 Connection 1" BSP to DIN ISO 228 Materials Screwed housing: Measuring tube/electrode/screw: Electrode rod insulation: PTFE Spacer disc: PEEK Terminal box: aluminium Length of installation / Measuring length / Reduction of length Length of installation A [mm] Measuring length L [mm] Possible reduction of length R [mm] Temperature sensor Resistance thermometer Pt 1000 Cell constant C 0.2 cm -1 Wiring 5pole screw-type terminal strip, removable, conductor size 1.5 mm² or via 5pole connector Cable entry Cable gland with integral cable clamp, 1 x PG 9 (M16), 1 x PG 16 (M20) Protection IP 65 to DIN Max. admissible ambient temperature 70 C Temperature for storage and transport 40 to + 80 C Supply voltage 24 V DC (18-36 V DC) 17

18 Power consumption 3 W Fuse Electronic thermal fuse T max 85 C, hysteresis 2 C Conductivity control range / Actual value output Control range*) [µs/cm at 25 C] 0.5 Current output ma = µs/cm 4 ma corresponds to 20 ma corresponds to Adjustable via code switch. Max. load for actual value output 750 Ω. *) Conversion µs/cm into ppm (parts per million) 1 µs/cm = 0.5 ppm Measurement cycle 1 sec. Accuracy 5 %, referred to measured conductivity without temperature compensation Temperature compensation linear, Tk adjustable via code switch, 0 %/ C, %/ C in steps of 0.1 %/ C Time constant T (measured acc. to DIN 3440) Temperature: 9 sec. Conductivity: 14 sec. 18

19 Dimensions of Conductivity Transmitter System LRGT

20 Application as conductivity transmitter and continuous blowdown controller Use conductivity transmitter system for conductivity limiting (TDS and continuous blowdown control) in steam boilers only in conjunction with industrial controller KS 90. The following describes the configuration and adjustment of the controller KS 90 when used for conductivity (TDS) limiting and continuous boiler blowdown control. Note: The installation manual for the industrial controller KS 90 supplied with the equipment must also be taken into consideration! Industrial Controller KS 90 Functional elements / Front view Yellow LED C lit: Yellow LED H lit: Green LED AL 1 lit: Continuous blowdown valve opens Continuous blowdown valve closes Max. limit exceeded Versions The following versions can be used: Supply voltage 230/115 V AC 24 V AC or

21 Wiring Keep mains cables separate from signal and measuring cables. We recommend twisted and screened measuring cables (screen connected to earth). Connected final elements must be equipped with protective circuits to manufacturer specification. This avoids voltage peaks which can cause trouble to the controller. The instruments must be protected by an individual or common fuse for a max. power consumption of 10 VA per unit (standard fuse ratings, min. 1.0 A). Signal and measurement circuits may carry max. 50 V r.m.s. against ground, mains circuits may carry max. 250 V r.m.s. between terminals. Output 1: Output 2: Output 3: Continuous blowdown valve closes, relay contact closed Continuous blowdown valve opens, relay contact closed Max. limit exceeded, relay contact open 21

22 Operating Structure After switching on mains (1), the controller is initialised (Test, Con1, Con2), then turns into the #OPERATING LEVEL 1 and controls the process. The controller is matched to the control task at the # PARAMETER LEVEL 2 and to the process at the #CONFIGURATION LEVEL 3. The level transfer is done with the key. To exit the operating level, S.I.L. switch LOCK must be open (factory setting). Exit from the parameter level is via timeout (30 s), too. After keying completely through the configuration level, transfer is into operating level (2 configuration not changed) or into initialisation (3 configuration has been changed). Configuration Level Adjustment: Type of input: 4-20 ma, Inl /InH adjustable Controller function: Three-position stepping controller Function of the outputs: Relay 1: valve CLOSED Relay 2: valve OPEN Relay 3: Alarm 1 22

23 Controller Functions Adjustment: Alarm 1: Sensor alarm + limit contact Absolute limit contact Adjustment: Sensor alarm reaction as X < W, i. e. valve closes 23

24 Adjustment:

25 Parameter Level No. Parameter name Symbol Adjustment range 01 Set-point SP 2 Between lower and upper set point limit ( SPL...SPH ) 02 Segment time P t min, adjustment Limit contact 1 low L C L 1 Lower set-point limit SPL Limit contact 1 high L C H Alarm switch. differential for LC 1 S d A , adjustment 1 06 Operation locking L o c 0 07 Lower set-point limit S P L Start of control range ( InL )... Upper set-point limit ( SPH 1 ) 08 Upper set-point limit S P H Lower set-point limit ( SPL + 1 )... End of control range ( InH ) 09 Proportional band P b %, adjustment 10 % 10 Integral time t i 0 11 Derivative time t d 0 12 Actuator response time t t depend. on actuator, e. g. 120 s 13 Min. step time t t p s, adjustment 2.0 s 14 Trigger point separation S H %, adjustment e. g. 10 % 15 Filter time constant t F s, adjustment e. g. 1.0 s 16 Decimal point d P 0 / 1 / 2 (0 = no decimal point) 17 Start of control range I n L ( InH 1 ) 18 End of control range I n H ( InL +1 )

26 The displayed parameter values of SP 2, LCL, LCH, SdA 1, SPL, SPH, InL and InH depend on the decimal point. Specifications in % refer to the span I n L... I n H. First calibrate start and end of the control range (InL, InH) and then the upper and lower set-point limits (SPL, SPH). Conductivity Controller transmitter Control range* [µs/cm at 25 C] I n L S P L I n H S P H If the equipment is used for conductivity limiting purposes set the limit contact low (LCL1) to the predefined max. electrical conductivity value. When this limit is exceeded the green LED for Alarm 1 lights up and relay output 3 opens the burner protection circuit. The adjusted limit value can be retrieved at any time on the parameter level. Relay output 3 does not interlock automatically. The interlocking function must hence be implemented in the follow-up circuit (burner protection circuit), which has to comply with the requirements of DIN VDE Operating Level This level is for process control. Actual (process) value X and set-point W are displayed in manual and automatic modes. Manipulated variable Y is displayed by means of selector key. Manual and automatic modes are switched with Hand/automatic key. 26

27 The set-point in manual mode or set-point and/or manipulated variable in automatic mode can be changed by means of arrow keys. The change is effective after 2 s or by pressing selector key briefly. For manual operation of the continuous blowdown valve first switch to manual operation using the selector key and then switch to manipulated variable Y. Now it is possible to use the arrow keys for manual adjustments. If the measuring current coming from the conductivity transmitter falls below 2 ma or the line has the wrong polarity the following error message will appear: With this error message the continuous blowdown valve will be closed. 27

28 /602c 28

Compact System NRGS 11-2 NRGS Original Installation Instructions English

Compact System NRGS 11-2 NRGS Original Installation Instructions English Compact System NRGS 11-2 NRGS 16-2 EN English Original Installation Instructions 810366-05 1 Contents Important Notes Page Usage for the intended purpose...4 Safety note...4 LV (Low Voltage) Directive

More information

Level Electrode NRG NRG NRG NRG

Level Electrode NRG NRG NRG NRG Level Electrode NRG 16-11 NRG 17-11 NRG 19-11 NRG 111-11 EN English Original Installation Instructions 818578-05 1 Contents Important Notes Page Usage for the intended purpose...4 Safety note...4 Danger...4

More information

Conductivity Switch LRGS Original Installation Instructions English

Conductivity Switch LRGS Original Installation Instructions English Conductivity Switch LRGS 15-1 EN English Original Installation Instructions 818758-02 1 Contents Important Notes Page Usage for the intended purpose...4 Safety note...4 Danger...4 LV (Low Voltage) Directive

More information

NRG 16-12, NRG 17-12, NRG 19-12

NRG 16-12, NRG 17-12, NRG 19-12 GESTRA GESTRA Steam Systems NRG 16-12 NRG 17-12 NRG 19-12 Installation Instructions 818626-00 Level Electrode NRG 16-12, NRG 17-12, NRG 19-12 Contents Important Notes Page Usage for the intended purpose...4

More information

BA 46 BA 46-ASME BA 47 BA 47-ASME BAE BAE ASME BAE BAE ASME

BA 46 BA 46-ASME BA 47 BA 47-ASME BAE BAE ASME BAE BAE ASME Continuous Blowdown Valve BA 46 BA 46-ASME BA 47 BA 47-ASME BAE 46... BAE 46...-ASME BAE 47... BAE 47...-ASME BA 46 / BA 46-ASME, PN 40/CL 150/300, DN 15-DN 50 BA 47 / BA 47-ASME, PN 63/CL 600, DN 25,

More information

Operating Instructions. Angle Seat Control Valve. Type 7020

Operating Instructions. Angle Seat Control Valve. Type 7020 Operating Instructions Angle Seat Control Valve Type 7020 With: Digital Positioner Type 8048 Electro-pneumatic Positioner Type 8047 Pneumatic Positioner Type 8047 Version: 02/2006 Manual-7020e.doc Art.-No:

More information

Operating instructions Flow monitors SI6600 SI6700 SI /02 01/2011

Operating instructions Flow monitors SI6600 SI6700 SI /02 01/2011 Operating instructions Flow monitors SI6600 SI6700 SI6800 UK 704680/02 01/2011 Contents 1 Preliminary note 3 1.1 Explanation of symbols 3 2 Safety instructions 3 3 Functions and features 4 3.1 Application

More information

Type 3761 Pneumatic or Electropneumatic Positioner for Rotary Actuators. Fig. 1 Type 3761 Positioner. Mounting and Operating Instructions EB 8386 EN

Type 3761 Pneumatic or Electropneumatic Positioner for Rotary Actuators. Fig. 1 Type 3761 Positioner. Mounting and Operating Instructions EB 8386 EN Type 3761 Pneumatic or Electropneumatic Positioner for Rotary Actuators Fig. 1 Type 3761 Positioner Mounting and Operating Instructions EB 8386 EN Edition July 2007 Contents Contents Page 1 Design and

More information

ecotrans Lf 01/02 Microprocessor Transmitter / Switching Device for Conductivity

ecotrans Lf 01/02 Microprocessor Transmitter / Switching Device for Conductivity Data Sheet 20.27 Page / ecotrans Lf 0/02 Microprocessor Transmitter / Switching Device for Conductivity Type 2027 Housing for DIN rail mounting (5 x 7.5 mm to EN 0 75 A.) Brief description The JUMO ecotrans

More information

ΑΥΤΟΜΑΤΗ ΣΤΡΑΤΣΩΝΑ ΕΛΕΓΚΤΕΣ

ΑΥΤΟΜΑΤΗ ΣΤΡΑΤΣΩΝΑ ΕΛΕΓΚΤΕΣ ΑΥΤΟΜΑΤΗ ΣΤΡΑΤΣΩΝΑ ΕΛΕΓΚΤΕΣ TI-P403-52 / 4033xxB 1/5 AB Issue 3 and Blowdown Controllers Compact unit for TDS control on large or small boilers. Wall or panel mounted versions. 4 digit LED display (ppm

More information

INDUCTIVE CONDUCTIVITY SENSOR. Instruction Manual. Bürkert 2001 Subject to technical change without notice

INDUCTIVE CONDUCTIVITY SENSOR. Instruction Manual. Bürkert 2001 Subject to technical change without notice INDUCTIVE CONDUCTIVITY SENSOR Instruction Manual Bürkert 00 Subject to technical change without notice INTRODUCTION Table of Contents. INTRODUCTION.... Symbols used.... General safety instructions....

More information

Operating Instructions. Pneumatic Control Valve Low Temperature. Type Series GS3

Operating Instructions. Pneumatic Control Valve Low Temperature. Type Series GS3 Operating Instructions Pneumatic Control Valve Low Temperature Type 8026 Series GS3 With: Digital Positioner Type 8048 Electro-pneumatic Positioner Type 8047 Pneumatic Positioner Type 8047 Version: 03/2006

More information

Portable pressure calibrator, Wally Box III Model CPH7600

Portable pressure calibrator, Wally Box III Model CPH7600 Calibration technology Portable pressure calibrator, Wally Box III Model CPH7600 WIKA data sheet CT 17.01 Applications Calibration service companies and service industry Measurement and control laboratories

More information

Mounting and Operating Instructions EB 5868/5869 EN

Mounting and Operating Instructions EB 5868/5869 EN Electric Control Valves Types 3213/5857, 3213/5824, Types 3214/5824, 3214/3374, 3214/3274 with safety function: Types 3213/5825, 3214/5825, 3214/3374, 3214/3274 Pneumatic Control Valves Types 3213/2780-1,

More information

UNA 13 UNA 15 UNA 15 Stainless Steel Design. Installation Instructions Steam Traps UNA 13, UNA 15

UNA 13 UNA 15 UNA 15 Stainless Steel Design. Installation Instructions Steam Traps UNA 13, UNA 15 UNA 13 UNA 15 UNA 15 Stainless Steel Design Installation Instructions 810604-01 Steam Traps UNA 13, UNA 15 1 Contents Important Notes Page Use... 7 Safety note... 7 Danger... 7 Explanatory Notes Scope

More information

JUMO ecotrans Lf 01/02

JUMO ecotrans Lf 01/02 Data Sheet 07 Page /7 JUMO ecotrans Lf 0/0 Microprocessor Transmitter/Switching Device for Conductivity Housing for DIN rail mounting (5 mm 7.5 mm to EN 075 A.) Brief description The JUMO ecotrans Lf 0/0

More information

User Manual. Solar Charge Controller 3KW

User Manual. Solar Charge Controller 3KW User Manual Solar Charge Controller 3KW 1 CONTENTS 1 ABOUT THIS MANUAL... 3 1.1 Purpose... 3 1.2 Scope... 3 1.3 SAFETY INSTRUCTIONS... 3 2 INTRODUCTION... 4 2.1 Features... 4 2.2 Product Overview... 5

More information

Electrical linear actuator with safety function MC100FSR/24

Electrical linear actuator with safety function MC100FSR/24 Issue 2013-05-24 Operating manual 3.10-TA.210-20-EN Electrical linear actuator with safety function Application Linear actuator with automatic coupling and safety function for fine stroke adjustment of

More information

Signet Pressure Transmitter

Signet Pressure Transmitter Signet 80 Pressure 80.090 80.090 Rev. J 0/ English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents. Installation.

More information

Operating instructions Flow monitors SI6000 SI6100 SI /00 06/2011

Operating instructions Flow monitors SI6000 SI6100 SI /00 06/2011 Operating instructions Flow monitors SI6000 SI6100 SI6200 706053/00 06/2011 Contents 1 Safety instructions... 3 2 Functions and features... 4 2.1 Application area... 4 2.2 Operating principle flow monitoring...

More information

Mounting and Operating instructions

Mounting and Operating instructions Mounting and Operating instructions 2-point water level controllers DLR1/DHR1 with LW alarm (DLR1) with HW alarm (DHR1) Low water level limiter LMC1 In combination with level control electrodes EL041,

More information

Differential Pressure Transmitter

Differential Pressure Transmitter Specifications/Instructions Differential Pressure Transmitter General Model PY9000D is a differential pressure transmitter that uses a ceramic cantilever sensor. Deflection of the ceramic cantilever caused

More information

Operating instructions Flow monitors SI5000 SI / / 2007

Operating instructions Flow monitors SI5000 SI / / 2007 Operating instructions Flow monitors SI5000 SI5001 704056 / 01 06 / 2007 Contents 1 Safety instructions... 3 2 Functions and features... 4 2.1 Application area... 4 2.2 Operating principle flow monitoring...

More information

Condensate Drain Valve AK 45. Original Installation Instructions English

Condensate Drain Valve AK 45. Original Installation Instructions English Condensate Drain Valve AK 45 EN English Original Installation Instructions 810467-03 1 Contents Important Notes Page Usage for the intended purpose...4 Safety note...4 Danger...4 Attention...4 PED (Pressure

More information

Pneumatic or Electropneumatic Positioner for Rotary Actuators Type Fig. 1 Type 3761 Positioner. Mounting and Operating Instructions EB 8386 EN

Pneumatic or Electropneumatic Positioner for Rotary Actuators Type Fig. 1 Type 3761 Positioner. Mounting and Operating Instructions EB 8386 EN Pneumatic or Electropneumatic Positioner for Rotary Actuators Type 3761 Fig. 1 Type 3761 Positioner Mounting and Operating Instructions EB 8386 EN Edition June 2004 Contents Contents Page 1 Design and

More information

ST Wiring diagram. Product description. Temperature controller. Order number

ST Wiring diagram. Product description. Temperature controller. Order number ST7-31.3 Temperature controller Order number 9154.12 Wiring diagram Product description The switching exits of the thermostatic controller can be programmed as -two-point controller with alarm -three-point

More information

NRG Installation Instructions Level Probe Type NRG A Siebe Group Product 1

NRG Installation Instructions Level Probe Type NRG A Siebe Group Product 1 NRG 211-1 Installation Instructions 810425-00 Level Probe Type NRG 211-1 A Siebe Group Product 1 Dimensions ~80 ~118 97 478 334.5 Fig. 1 43.5 2 Parts Drawings G H I J K F E L D C M N B O A Fig. 2 3 Parts

More information

Operating and Installation Instructions Diaphragm Vacuum Pumps and Compressors

Operating and Installation Instructions Diaphragm Vacuum Pumps and Compressors Operating and Installation Instructions Diaphragm Vacuum Pumps and Compressors Type range: UN813.3ANI UN813.4ANI UN813.3ANDCB UN813.4ANDCB UN813.5ANI Fig. 1: UN813.3ANI Fig. 2: UN813.4ANI You have selected

More information

Electronic pressure switch. Type HEDE 10. Contents. Features. RE Edition: Replaces:

Electronic pressure switch. Type HEDE 10. Contents. Features. RE Edition: Replaces: Electronic pressure switch Type HEDE 10 RE 30277 Edition: 2017-02 Replaces: 2016-06 Component series 3X Maximum operating pressure 600 bar Features Suitable for measuring pressures in hydraulic systems

More information

Insertion magnetic inductive flowmeter

Insertion magnetic inductive flowmeter Insertion magnetic inductive flowmeter Sensor without moving parts Flowmeter with On/Off control Application related calibration by Teach-In function Clean in place (CIP) FDA-compliant materials Type 8041

More information

Electropneumatic Converters i/p Converters Type 6111 Mounting and Operating Instructions EB 6111 EN

Electropneumatic Converters i/p Converters Type 6111 Mounting and Operating Instructions EB 6111 EN Electropneumatic Converters i/p Converters Type 6111 Fig. 1 Type 6111 in standard version Fig. Type 6111 mounted on a supply air manifold Fig. 3 Type 6111 in field enclosure Mounting and Operating Instructions

More information

Electrical linear actuator with safety function MC100FSR/230

Electrical linear actuator with safety function MC100FSR/230 Issue 2013-05-24 Operating manual 3.10-TA.210-10-EN Electrical linear actuator with safety function Application Linear actuator with automatic coupling and safety function for fine stroke adjustment of

More information

User Manual Solar Charge Controller 3KW

User Manual Solar Charge Controller 3KW User Manual Solar Charge Controller 3KW Version: 1.3 CONTENTS 1 ABOUT THIS MANUAL... 1 1.1 Purpose... 1 1.2 Scope... 1 1.3 SAFETY INSTRUCTIONS... 1 2 INTRODUCTION... 2 2.1 Features... 2 2.2 Product Overview...

More information

VALVE CONTROLLERS Controllers for Dust Extr 2010 / 2011 action Technology

VALVE CONTROLLERS Controllers for Dust Extr 2010 / 2011 action Technology VALVE CONTROLLERS Controllers for Dust Extraction 2010 Technology / 2011 Valve controllers for all cases HESCH has the skills and technology to tackle any control task for dedusting of filter and dust

More information

Technical Data Electric double-motor-actuators

Technical Data Electric double-motor-actuators M768 Electric double-motor-actuators Contents Page General data Overview, application, design and mode of operation, mounting position Technical details General Motors Switching and signaling unit Electrical

More information

TLR2 RESOL TLR2 * * Mounting Connection Handling Fault localization. Manual

TLR2 RESOL TLR2 * * Mounting Connection Handling Fault localization. Manual RESOL TLR2 Mounting Connection Handling Fault localization *48001550* 48001550 TLR2 Thanks for buying a RESOL product. Read this manual carefully to get the best perfomance from this unit. B Manual www.resol.de

More information

Signet Pressure Transmitter

Signet Pressure Transmitter Signet 8450 Pressure English 8450.090 8450.090 Rev. G /06 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents

More information

Supplementary instructions. Air cooling - FIBERTRAC 31 with Conduit tube connection. Active air cooling system for radiation-based sensors

Supplementary instructions. Air cooling - FIBERTRAC 31 with Conduit tube connection. Active air cooling system for radiation-based sensors Supplementary instructions Air cooling - FIBERTRAC 31 with Conduit tube connection Active air cooling system for radiation-based Document ID: 52847 Contents Contents 1 Product description 1.1 Configuration...

More information

WP37 & HT37 User Manual

WP37 & HT37 User Manual WP37 & HT37 User Manual Origio WP37 & HT37 Warming Plates & Heated Trolley Warming Plates WP37 300 WP37 500 Heated Trolley HT37 Origio WP37 Component Description Heated work surface Model WP37 300 Power

More information

& HIGH CURRENT DC POWER SUPPLIES INSTRUCTION MANUAL

& HIGH CURRENT DC POWER SUPPLIES INSTRUCTION MANUAL 72-6850 & 72-6852 HIGH CURRENT DC POWER SUPPLIES INSTRUCTION MANUAL Table of Contents Introduction 2 Specification 2 Safety 4 EMC 5 Installation 6 Connections 6 Operation 7 Maintenance and Repair 8 www.tenma.com

More information

ECON ELECTRIC ACTUATOR Fig ELA60 Ex

ECON ELECTRIC ACTUATOR Fig ELA60 Ex ECON ELECTRIC ACTUATOR Fig. 7907, type ELA60 Fig. 7907 ELA60 IP67 (optionally IP68) Fig. 7907 ELA60 Ex EX ǁ 2G Ex d IIB T4 Gb Small & Compact quarter turn actuator Mechanical position indicator High output

More information

Thermal flow sensor TFS-35

Thermal flow sensor TFS-35 průmyslová elektronika Thermal flow sensor TFS-35 For sensing limit flow rates of fluid media and for monitoring their temperature. These sensors are intended for installation in pipes, in which the actual

More information

Inline Flow meter for compressed air and gases DN15 (1/2 ) - DN80 (3 )

Inline Flow meter for compressed air and gases DN15 (1/2 ) - DN80 (3 ) EE771/EE772 Inline Flow meter for compressed air and gases DN15 (1/2 ) - DN80 (3 ) The inline flow meter EE771/EE772, based on the measurement principle of thermal mass flow, is ideally suited for the

More information

Electronic Ignition Module

Electronic Ignition Module 7 607 Electronic Ignition Module TQO31A27 Electronic ignition module for use with oil or gas burners operating with singleor double-pole ignition. The TQO31A27 and this Data Sheet are intended for use

More information

PRS 9. Installation Instructions Cycling Timer PRS 9. A Siebe Group Product 1

PRS 9. Installation Instructions Cycling Timer PRS 9. A Siebe Group Product 1 PRS 9 Installation Instructions 810407-00 Cycling Timer PRS 9 A Siebe Group Product 1 Dimensions / Parts Drawings 128.5 169 30.01 (6TE) Fig. 1 A B C D E I H G J F MAX 70 C Fig. 2 MAX 95 % 2 Key A B C D

More information

Supplementary instructions. Air cooling - FIBERTRAC 31. Active air cooling system for radiation-based sensors. Document ID: 50339

Supplementary instructions. Air cooling - FIBERTRAC 31. Active air cooling system for radiation-based sensors. Document ID: 50339 Supplementary instructions Air cooling - FIBERTRAC 31 Active air cooling system for radiation-based sensors Document ID: 50339 Contents Contents 1 Product description 1.1 Configuration... 3 2 Mounting

More information

OD DIGITAL DISSOLVED OXYGEN CONTROLLER POLAROGRAPHIC CELL

OD DIGITAL DISSOLVED OXYGEN CONTROLLER POLAROGRAPHIC CELL UNI EN ISO 9001:2000 CERT. N. 9115 BCEL INSTRUCTION MANUAL OD 265.2 DIGITAL DISSOLVED OXYGEN CONTROLLER POLAROGRAPHIC CELL Rev. B - 03/07 B&C Electronics Srl - Via per Villanova 3-20040 Carnate (Mi) -

More information

Proportional directional valve, pilot-operated, with integrated electronics (OBE)

Proportional directional valve, pilot-operated, with integrated electronics (OBE) Proportional directional valve, pilot-operated, with integrated electronics (OBE) Type WFCE RE 943 Edition: 7- Replaces: 6- Size 6 5 Component series X Maximum operating pressure 4 bar Maximum flow 5 l/min

More information

ST48-WHUV.102. Wiring diagram. Product description. PID controller. Order number

ST48-WHUV.102. Wiring diagram. Product description. PID controller. Order number ST48-WHUV.12 PID controller Order number 935.15 Wiring diagram Product description This micro-processed controller serves for temperature control at high measuring accuracy. Beside resistance sensors and

More information

INSTRUCTION MANUAL BR 25 FOR RAILWAY APPLICATION 5COR BR25AT REV00

INSTRUCTION MANUAL BR 25 FOR RAILWAY APPLICATION 5COR BR25AT REV00 INSTRUCTION MANUAL BR 25 FOR RAILWAY APPLICATION 5COR469100 BR25AT REV00 I CONTENT: 1 SAFETY 1.1 Safety instructions 1 1.2 Specified applications 1 1.3 Safety notes on the equipment operation 1 2 BR 25

More information

Model 556 Converter Installation and Maintenance Instructions

Model 556 Converter Installation and Maintenance Instructions IM-P403-98 EMM Issue 5 Model 556 Converter Installation and Maintenance Instructions 1. Using these Installation and Maintenance Instructions 2. Returns and disposal 3. Intended use 4. Safety 5. Description

More information

Operating instructions Flow monitors SI5002 SI5003 SI / / 2012

Operating instructions Flow monitors SI5002 SI5003 SI / / 2012 Operating instructions Flow monitors SI5002 SI5003 SI0521 704318 / 03 08 / 2012 Contents 1 Safety instructions... 3 2 Functions and features... 4 2.1 Application area... 4 2.2 Operating principle flow

More information

Operating Instructions (from serial number 3000)

Operating Instructions (from serial number 3000) Conductivity meter Leitfähigkeits-Messgerät Conductivimètre Operating Instructions (from serial number 3000) Bedienungsanleitung (ab Serien-Nummer 3000) Manuel d utilisation (à partir du numéro de série

More information

4/3 proportional directional control valve, without position control, with on-board electronics (OBE)

4/3 proportional directional control valve, without position control, with on-board electronics (OBE) 4/3 proportional directional control valve, without position control, with on-board electronics (OBE) Type 4WRBAE..E.. /..W.. Nominal size (NG) 6, 10 Unit series 2X Maximum working pressure P, A, B 315

More information

Installation and Operational Instructions for EAS - HTL housed overload clutch Sizes 01 3 Type 490._24.0

Installation and Operational Instructions for EAS - HTL housed overload clutch Sizes 01 3 Type 490._24.0 Please read these Operational Instructions carefully and follow them accordingly! Ignoring these Instructions may lead to malfunctions or to clutch failure, resulting in damage to other parts. Contents:

More information

Frost protection for pipes. 16 Technical handbook. Thermostat with line or ambient temperature sensor

Frost protection for pipes. 16 Technical handbook. Thermostat with line or ambient temperature sensor Frozen pipes can be a costly problem. When pipes are exposed to subzero temperatures they can burst, leading to considerable damage and disruption. The Raychem frost protection system provides an efficient

More information

ZB0050 / ZB0051 ZB0070 / ZB0071

ZB0050 / ZB0051 ZB0070 / ZB0071 Operating instructions Safety Rope Emergency Stop Switches UK ZB0050 / ZB0051 ZB0070 / ZB0071 7390877 / 02 08/2013 Contents 1 Safety instructions...3 2 Installation / set-up...4 2.1 Applications...4 2.2

More information

Series 7000 Torque Sensor for PTO-shafts

Series 7000 Torque Sensor for PTO-shafts Properties PTO (Power Take-Off) shaft with integrated torque and angle measurement Non-contact measurement system, high robustness Special for PTO shafts 1 ¾ und 1 3/8 Plug & Play solution, no additional

More information

SBM TECHNOLOGY PET Blow Valve Block

SBM TECHNOLOGY PET Blow Valve Block Advanced high pressure seal technology Robust, compact & lightweight design Proven & reliable technology Maintenance & service friendly High flow rate exhaust function Minimal dead space volume (< 35ml)

More information

Mounting and Operating Instructions EB 8039 EN. Type 3351 Pneumatic On/off Valve. Type 3351 Pneumatic On/off Valve. Type 3351 Pneumatic On/off Valve

Mounting and Operating Instructions EB 8039 EN. Type 3351 Pneumatic On/off Valve. Type 3351 Pneumatic On/off Valve. Type 3351 Pneumatic On/off Valve Type 3351 Pneumatic On/off Valve Type 3351 Pneumatic On/off Valve Type 3351 Pneumatic On/off Valve Version with handwheel Mounting and Operating Instructions EB 8039 EN Edition May 2016 Definition of signal

More information

Mounting and Operating Instructions EB 3913 EN. Electropneumatic Converters i/p Converter Type

Mounting and Operating Instructions EB 3913 EN. Electropneumatic Converters i/p Converter Type Electropneumatic Converters i/p Converter Type 3913-0001 Fig. 1 Type 3913-0001 Electropneumatic Converter with pressure gauge and bracket Mounting and Operating Instructions EB 3913 EN Edition August 2011

More information

Series 430 Type 3430 Pneumatic Indicating Controller for Temperature with Capillary Sensor Type 3432 Controller Station Type 3436 Transmitter Module

Series 430 Type 3430 Pneumatic Indicating Controller for Temperature with Capillary Sensor Type 3432 Controller Station Type 3436 Transmitter Module Series 430 Type 3430 Pneumatic Indicating Controller for Temperature with Capillary Sensor Type 3432 Controller Station Type 3436 Transmitter Module Application Temperature controller for process engineering

More information

Screw-in resistance thermometers with plug connector

Screw-in resistance thermometers with plug connector Data Sheet 90.2004 Page 1/8 s with plug connector for temperatures from -50 to +300 C highly resistant to shock and vibration connector locked in for reliable contact, IP65 or IP69K protection available

More information

Operating instructions

Operating instructions 1 2 3 + - 4 ¼ 20 ma 4 12 ¼ 36 V DC E12 - + 5 6 7 8 Level and Pressure Operating instructions VEGABAR 20 DISPLAY VEGADIS 10 0-20 bar p Safety information Safety information The described module must only

More information

Pressure chlorine changeover unit C 7520

Pressure chlorine changeover unit C 7520 BW 2 24 04 / 1 Content 1. Scope of delivery 2. Device description 3. Installation 4. Operation 5. Shutdown 6. Maintenance 7. Troubleshooting 1 Scope of delivery The chlorine gas changeover unit C 7520

More information

Operating instructions Binary level sensor LMT /00 06/2011

Operating instructions Binary level sensor LMT /00 06/2011 Operating instructions Binary level sensor LMT UK 706052/00 06/2011 Contents 1 Preliminary note...3 1.1 Symbols used...3 2 Safety instructions...3 3 Functions and features...4 3.1 Application area...4

More information

Angle seat valve with diaphragm actuator VZXA-...-M

Angle seat valve with diaphragm actuator VZXA-...-M Angle seat valve with diaphragm actuator VZXA-...-M Instructions Operating (Translation of the original instructions) Festo AG & Co. KG Ruiter Straße 82 73734 Esslingen Germany +49 711 347-0 www.festo.com

More information

Pressure transmitter for general purpose Type MBS 1700 and MBS 1750

Pressure transmitter for general purpose Type MBS 1700 and MBS 1750 MAKING MODERN LIVING POSSIBLE Data sheet Pressure transmitter for general purpose Type MBS 1700 and MBS 1750 The compact pressure transmitters MBS 1700 and MBS 1750 are designed for use as a general purpose

More information

Series 7000 Torque Sensor for PTO-shafts

Series 7000 Torque Sensor for PTO-shafts Properties PTO (Power Take-Off) shaft with integrated torque and angle measurement Non-contact measurement system, high robustness Special for PTO shafts 1 ¾ und 1 3/8 Plug & Play solution, no additional

More information

Solid State Electronic Pressure Switch with Integral LED Display Model PSD-10

Solid State Electronic Pressure Switch with Integral LED Display Model PSD-10 Solid State Electronic Pressure Switch with Integral LED Display Model PSD-10 Electronic Pressure Measurement WIKA Datasheet PSD-10 Applications Hydraulics and pneumatics Filter monitoring Pump control

More information

PURE SINE WAVE DC TO AC POWER INVERTER

PURE SINE WAVE DC TO AC POWER INVERTER PURE SINE WAVE DC TO AC POWER INVERTER 60S-12A / 60S-24A 60S-12E / 60S-24E 100S-12A / 100S-24A 100S-12E / 100S-24E 150S-12A / 150S-24A 150S-12E / 150S-24E Instruction manual SINE WAVE INVERTER Please read

More information

ETP80. Electronic Transmitter for Pressure, Differential Pressure and Level

ETP80. Electronic Transmitter for Pressure, Differential Pressure and Level ETP80 Electronic Transmitter for Pressure, Differential Pressure and Level The ETP80 is a series of pressure transmitters with a wide range of s to suit most applications. The transmitters can measure

More information

Mounting and operating instructions EB 5801 EN. Electric Actuators Type 5801 (Rotary Actuator) Type 5802 (Linear Actuator)

Mounting and operating instructions EB 5801 EN. Electric Actuators Type 5801 (Rotary Actuator) Type 5802 (Linear Actuator) Electric Actuators Type 5801 (Rotary Actuator) Type 5802 (Linear Actuator) Linear Actuator with Type 3260 Control Valve Rotary actuator with lever system Linear actuator with Type 3321 (V2001) Control

More information

4/4 directional control valves, direct operated, with electrical position feedback and integrated electronics (OBE)

4/4 directional control valves, direct operated, with electrical position feedback and integrated electronics (OBE) 4/4 directional control valves, direct operated, with electrical position feedback and integrated electronics (OBE) Type 4WRPEH RE 29121 Edition: 2014-01 Size 6 Component series 3X Maximum operating pressure

More information

Overview of types. Technical data

Overview of types. Technical data Technical data sheet 6..S+NVFT24-MFT-E Globe valves, 2-way, with flange PN 16, actuator with emergency control function for closed hot water systems up to 150 C in district heating applications approved

More information

Mounting and Operating Instructions EB 8222 EN. Type 3310/AT and Type 3310/3278 Pneumatic Control Valves. Type 3310 Segmented Ball Valve

Mounting and Operating Instructions EB 8222 EN. Type 3310/AT and Type 3310/3278 Pneumatic Control Valves. Type 3310 Segmented Ball Valve Type 3310/AT and Type 3310/3278 Pneumatic Control Valves Type 3310 Segmented Ball Valve Fig. 1 Type 3310/3278 with positioner Fig. 2 Type 3310/AT Mounting and Operating Instructions EB 8222 EN Edition

More information

Level Measuring Sensor for Diesel Fuel Tanks

Level Measuring Sensor for Diesel Fuel Tanks Installation and Operating Manual Level Measuring Sensor for Diesel Fuel Tanks Executions with switchable output voltage signal: 0-2.3 V and industrial standard signal 0-10 V: Tank Sensor Diesel 010-250

More information

Process controller. for pneumatic and electric actuators. Process controller V. Process controller

Process controller. for pneumatic and electric actuators. Process controller V. Process controller for pneumatic and electric actuators Process controller 110-240V Process controller dtron 316 in Rittal-cabinet with current loop sensor input for controlling pneumatic and electric actuators with a position

More information

Mounting and Operating Instructions EB 8111/8112 EN. Series V2001 Valves Type 3321 Globe Valve

Mounting and Operating Instructions EB 8111/8112 EN. Series V2001 Valves Type 3321 Globe Valve Series V2001 Valves Type 3321 Globe Valve Type 3321 Globe Valve with rod-type yoke and Type 3372 Electropneumatic Actuator (350 cm²) Mounting and Operating Instructions EB 8111/8112 EN Edition June 2013

More information

Installation and Operating Instructions

Installation and Operating Instructions Plastic encased safety interlocking switch Type PSEN me4 Intended use The plastic encased safety switches PSEN me4 with separate actuator are suitable for the mounting on safety guards. They are designated

More information

Angle seat valve with piston actuator VZXA-...-K

Angle seat valve with piston actuator VZXA-...-K Angle seat valve with piston actuator VZXA-...-K Festo AG & Co. KG Postfach 73726 Esslingen Germany +49 711 347-0 www.festo.com 3 Further information Accessories www.festo.com/catalogue Spare parts www.festo.com/spareparts

More information

Type Operating Instructions. Bedienungsanleitung Manuel utilisateur

Type Operating Instructions. Bedienungsanleitung Manuel utilisateur Water flow rate transmitter Wasser-Durchfluss-Transmitter Transmetteur de débit d eau Operating Instructions Bedienungsanleitung Manuel utilisateur We reserve the right to make technical changes without

More information

4/2 servo solenoid valves with on-board electronics (OBE), positive overlap and position feedback

4/2 servo solenoid valves with on-board electronics (OBE), positive overlap and position feedback Electric Drives and Controls Hydraulics Linear Motion and Assembly Technologies Pneumatics Service 4/2 servo solenoid valves with on-board electronics (OBE), positive overlap and position feedback RA 29024/01.05

More information

INSTRUCTION MANUAL RELAY FOR ON LOAD TAP CHANGER APPLICATIONS OR-25 5COR OR25 REV03

INSTRUCTION MANUAL RELAY FOR ON LOAD TAP CHANGER APPLICATIONS OR-25 5COR OR25 REV03 INSTRUCTION MANUAL RELAY FOR ON LOAD TAP CHANGER APPLICATIONS OR-25 5COR469000 OR25 REV03 I II CONTENT: 1 SAFETY 1.1 Safety instructions 2 1.2 Specified applications 2 1.3 Safety notes on the equipment

More information

LTX RF LEVEL SENSOR. Instruction Manual

LTX RF LEVEL SENSOR. Instruction Manual LTX RF LEVEL SENSOR Instruction Manual FOR MODELS LTX01, LTX02, LTX05 Intempco Document No: LTX - M01 Rev. 1 Issue Date: April 2005 LTX01 RF LEVEL SENSOR USER MANUAL Software Rev : Rev. Date : June 2004

More information

Thermal flow sensor TFS-35

Thermal flow sensor TFS-35 průmyslová elektronika Thermal flow sensor TFS-35 For flow monitoring rates of liquids media and for monitoring of their temperature These sensors are intended for installation in pipes, in which the actual

More information

Modular resistance thermometer - Quicktemp TP60/TW39...T500 -

Modular resistance thermometer - Quicktemp TP60/TW39...T500 - Das modulare MODULAR von Hengesbach QUICK-CALIBRATE INSIDE FEATURES 1 BASIC DEVICE - PROCESS ADAPTATION USING PROTECTIVE SLEEVE DESIGNS QUICKTEMP MODULAR PROCESS CONNECTION SYSTEM FOR DEVICE REPLACEMENT

More information

4/3 directional control valve, pilot operated, with electric position feedback and integrated electronics (OBE)

4/3 directional control valve, pilot operated, with electric position feedback and integrated electronics (OBE) 4/ directional control valve, pilot operated, with electric position feedback and integrated electronics (OBE) RE 977/.1 Replaces: 1.9 1/16 Type 4WRVE 1...7, symbols V, V1 Sizes 1, 16, 5, 7 Component series

More information

LKS 160 FOR AIR DAMPERS ON MULTI-STAGE AND MODULATING BURNERS

LKS 160 FOR AIR DAMPERS ON MULTI-STAGE AND MODULATING BURNERS ACTUATOR LKS 160 FOR AIR DAMPERS ON MULTI-STAGE AND MODULATING BURNERS PRODUCT HANDBOOK APPLICATION The LKS 160 air damper actuator is designed to be fitted to multi-stage or modulating oil and gas burners

More information

Typical Feed Water Heater Isolation System.

Typical Feed Water Heater Isolation System. Type AVS 4/5 Typical Feed Water Heater Isolation System. Features and Benefits These valves are designed to bypass high pressure feedwater around a group of heaters in the event of high water level in

More information

Signet Temperature Transmitter

Signet Temperature Transmitter Signet 80 Temperature English 80.090 80.090 Rev. G /0 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.

More information

Proportional cartridge throttle valve, with on-board electronics (OBE) and inductive position transducer, pilot operated

Proportional cartridge throttle valve, with on-board electronics (OBE) and inductive position transducer, pilot operated Proportional cartridge throttle valve, with on-board electronics (OBE) and inductive position transducer, pilot operated RE 29216/09.05 1/18 Type FESE Nominal size 16, 25, 32, 40, 50 Unit series 1 Maximum

More information

Svendborg Conveyor Brake Maintenance

Svendborg Conveyor Brake Maintenance CLIENT LOGO OPF MAINTENANCE WORK INSTRUCTION CLIENT DOC NUMBER Scope PREPARATION AND INFORMATION The Svendborg Brake is designed to transfer a braking torque from the calliper to a brake disc in order

More information

Product Information ECN 425 EQN 437. Absolute Rotary Encoders with Hollow Shaft and Expanding Ring Coupling for Safety-Related Applications

Product Information ECN 425 EQN 437. Absolute Rotary Encoders with Hollow Shaft and Expanding Ring Coupling for Safety-Related Applications Product Information ECN 425 EQN 437 Absolute Rotary Encoders with Hollow Shaft and Expanding Ring Coupling for Safety-Related Applications 4/2014 ECN 425, EQN 437 Rotary encoders for absolute position

More information

Digital ph transmitter

Digital ph transmitter Digital ph transmitter Compact, remote versions for DN 15 to 200 Fully programmable ph transmitter functions for all kinds of ph-measurement tasks Large range of process connections with various fittings

More information

Level Sensor with G1/2" Thread Hygienic

Level Sensor with G1/2 Thread Hygienic SENSORS FOR FOOD AND BIOPHARMA. Product Information NVS-141, -143, -146, -161, -163, -166, -181, -183, -186 FOOD Level Sensor with Thread Hygienic Application/intended use Limit detection of aqueous and

More information

Operating manual UPS - System

Operating manual UPS - System Operating manual UPS - System POWERMASTER M MIL 1000VA 7Min. BAX 3330 E UPS-Division Issued 15. August 2006 JOVYATLAS JOVYATLAS Elektrische Umformtechnik GmbH Groninger Straße 29-37 D-26789 Leer/Ostfriesland

More information

Flow - magnetic inductive - probe form

Flow - magnetic inductive - probe form GHM Messtechnik GmbH Hans-Sachs-Str. 6 938 Regenstauf Germany Fon +9 (0) 90-9383 - 0 Fax - 33 www.ghm-messtechnik.de info@ghm-messtechnik.de Flow - magnetic inductive - probe form Characteristics System

More information

Angle seat valve with piston actuator VZXA-...-K

Angle seat valve with piston actuator VZXA-...-K Angle seat valve with piston actuator VZXA-...-K Instructions Operating (Translation of the original instructions) Festo AG & Co. KG Ruiter Straße 82 73734 Esslingen Germany +49 711 347-0 www.festo.com

More information

Level Sensor with M12 Thread Hygienic

Level Sensor with M12 Thread Hygienic SENSORS FOR FOOD AND BIOPHARMA. Product Information NVS-041, -043, -046, -061, -063, -066, -081, -083, -086 FOOD Level Sensor with M12 Thread Hygienic Application/intended use Limit detection of aqueous

More information