2-D GC/MS and the Deans Switch: Theory and Applications. Fred Feyerherm, Agilent Technologies Inc. Houston, Tx
|
|
- Coleen Gibbs
- 5 years ago
- Views:
Transcription
1 2-D GC/MS and the Deans Switch: Theory and Applications Fred Feyerherm, Agilent Technologies Inc. Houston, Tx
2 Introduction Simplified 2-D GC can be used to accomplish the separation of specific analytes from very complex matrices. In complex sample matrices, there are often too many overlapping compounds to allow resolution of the compound(s) of interest, even with the highest resolution columns available. On any single column, the analytes co-elute with several other compounds, making their analysis difficult or impossible. This approach can solve difficult analytical problems with a relatively low cost, easy to use system that is very robust and easy to maintain.
3 Multidimensional (2-D) GC Very old (>25 yrs) but powerful separation technique. Based on cutting peak(s) from one GC column onto another with stationary phase of different selectivity (polarity). Compounds that co-elute with analyte on the first column separate from the analyte(s) of interest on the second column of different polarity. THC in Hair NIDA drugs in Oral Fluids THC in Blood Especially useful at ultra-trace concentrations (<1 ng/ml)
4 Why 2-D GC now? What s Changed? 6890 Simplified 2-D GC systems are much easier to use Micro-fluidics Deans Switch Column connections are easier, zero dead volume, inert, and reliable. Setting gas flows done with EPC and Flow Calculator. Retention time drift greatly reduced with modern oven and EPC. Inertness problems with switch hardware eliminated with surface coatings. RT control is very tight and the new switch is very quick.
5 Simplified 2-Dimensional GC 7683 Autosampler Cut FID1 FID2 Column 1 Column N GC Deans Switch
6 Deans Switch System: Used to heart cut individual peaks from DB-17ms to DB-1ms column FID T2 S/S Inlet restrictor 0.1 mm UDFS T1 purge restrictor solenoid valve PCM DB-17ms 15m x 0.25 x 0.25 um MS T3 DB-1ms 15m x 0.25 x 0.25 um Oven Wall
7 Deans Switch System Used to heart cut peak from DB-17MS to DB-1MS FID restrictor 0.3 m x.1 mm FS solenoid valve Injector PCM DB-17MS 15m x.25 x 0.25 um MSD DB-1MS 15m x.25 x 0.25 um
8 Deans Switch System Switch off, DB-17MS effluent goes to monitor FID (no cut) FID restrictor solenoid valve (off) Injector PCM 27 psi 20 psi DB-17MS MSD DB-1MS
9 Deans Switch System Switch on, DB-17MS effluent is cut to DB-1MS FID restrictor solenoid valve (on) Injector PCM 27 psi 20 psi DB-17MS MSD DB-1MS
10 Deans Switch System (after the cut) Switch off, DB-17MS effluent goes to monitor FID FID restrictor solenoid valve (off) Injector PCM 27 psi 20 psi DB-17MS MSD DB-1MS
11 EPC Backflush to save time and clean out first column! Program Inlet pressure down to 0.5 psi after the last cut, then after the last analyte elutes from column 2, program the PCM to 60 psi so heavies are backflushed out split vent. FID restrictor solenoid valve (off) S/S Inlet PCM 27 psi 0.5 psi 20 psi 60 psi DB-17ms MS DB-1ms
12 New Micro-fluidics Deans Switch
13 Deans Switch column connections
14 Simple Run Table Entry Cuts Sample onto Column 2
15 Method Developers Calculator to Tools correctly set flows and restrictor size
16 Cryo Trap Re-focuses Sample on Column 2 Sharpens peak very important for lower concentrations Adds ability to perform 2nd oven ramp FID restrictor solenoid valve (on) Injector AUX 4 27 psi 20 psi DB-17MS MSD Cryo DB-1MS
17 First Application Example 0.1 pg/mg THC in Hair DB-35ms DB-5ms 2002 at Northwest Toxicology in Salt Lake City Using Finnigan TSQ 7000 Triple Quadrupole MS Same samples analyzed with Deans switch
18 Low Level Detection Required SAMSHA guidelines set confirmatory cutoff at 0.05 pg/mg 20 mg sample size used washed, cut up, digested, extracted derivitized with PFPA/PFPOH 50 ul final extract volume (in toluene) THC-COOH concentration in final extract = 20 fg/ul
19 GC/MS Conditions Agilent 6890N N Negative chemical ionization (NCI) mode -Autotuned Source temperature 160C SIM ions (5 ions): THC-COOH 620, 492, 472 d3-thc-cooh 623, 495 Ammonia reagent (buffer) gas (5X increase in sensitivity vs methane)
20 Ideal Conditions (2 ul unextracted Abundance standard) 1200 Ion ( to ): THC-03.D Ion ( to ): THC-03.D 8.45 Ion ( to ): THC-03.D pg/mg (unextracted) Time-->
21 Abundance Hair Matrix Effect: 0.1 pg/mg Ion ( to ): THC-05.D Ion ( to ): THC-05.D Ion ( to ): THC-05.D Time-->
22 Deans Switch System (THC in Hair) Heartcut THC peak from DB-35MS to DB-5MS FID restrictor 4.2m x.1 mm FS tubing UNIS Inlet solenoid valve PCM DB-35MS 40m x.18 mm x 0.18 um MSD DB-5MS 40m x.18 mm x 0.18 um
23 Deans Switch System Switch off, DB-35MS effluent goes to monitor FID (no cut) FID restrictor solenoid valve (off) UNIS Inlet PCM 80 psi 50 psi DB-35MS MSD DB-5MS
24 Deans Switch System Switch on, DB-35MS effluent is cut to DB-5MS FID restrictor solenoid valve (on) UNIS Inlet PCM 80 psi 50 psi DB-35MS MSD DB-5MS
25 Results: 0.1 pg/mg THC-COOH in Hair Abundance 1100 Ion ( to ): THC-032.D m/e pg/mg THC-COOH in hair Time-->
26 Second Application Example THC in Oral Fluid TMS derivative DB-17ms DB1ms Both are 15m x 0.25 x 0.25
27 Deans Switch System Used to heart cut peak from DB-17MS to DB-1MS FID restrictor 0.3 m x.1 mm FS solenoid valve Injector PCM DB-17MS 15m x.25 x 0.25 um MSD DB-1MS 15m x.25 x 0.25 um
28 Method Developers Calculator to Tools correctly set flows and restrictor size
29 THC parent drug in Oral Fluid Abundance 1400 Ion ( to ): THC03.D\data.ms Ion ( to ): THC03.D\data.ms Ion ( to ): THC03.D\data.ms ng/ml Time-->
30 Third Application Example THC parent and THCA in Blood Arizona DPS TMS derivative RTX-200 DB17ms
31 Example of Inadequate Resolution DB-17ms DB-1ms Abundance Ion ( to ): 02-A.D Ion ( to ): 02-A.D Ion ( to ): 02-A.D Time-->
32 Deans Calculator
33 FID channel of the Deans Switch with no cuts to the secondary column Abundance Signal: 05B.D\FID1A.CH Time-->
34 Cut Times determined from FID
35 Abundance Signal: 05C.D\FID1A.CH THC Cut THCC Cut Abundance Time--> Ion ( to ): 05C.D Ion ( to ): 05C.D THCC-TMS Abundance 1000 Time--> TIC: 08g.D THC- TMS Co-eluting peaks from primary column THCC-TMS THC-TMS Time-->
36 0.5 ng/ml THC-TMS in Blood Abundance Ion ( to ): 011SP011.D Ion ( to ): 011SP011.D Ion ( to ): 011SP011.D THC-TMS, 0.5 ng/ml Ion Time-->
37 Summary Developments in GC hardware in recent years have made Deans switch systems easier to use and maintain. 2D GC is a powerful tool that can be used by itself or combined with other selective techniques to solve difficult separation problems. 2D GCMS (GC/GC/MS) can be used instead of GC/MS/MS in many applications.
Simple Heart Cutting with Deans. Capillary Flow Technolgy. Simon Jones Applications Engineer
Simple Heart Cutting with Deans Switch and Backflushing with Capillary Flow Technolgy Simon Jones Applications Engineer GC or GC/MS Analysis in Complex Matrices In complex sample matrices, there are often
More informationUsing a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application
Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D759 Application Petrochemical Author James D. McCurry Agilent Technologies 285 Centerville
More informationC2, C3, C4 Monomer Analysis
C2, C3, C4 Monomer Analysis Malgorzata Sierocinska Agilent Technologies Waldbronn Page 1 Why Analyze Monomers? To Insure Consistent Production of High Quality Polymer Protect against food contamination
More informationOxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels
Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels James D. McCurry Senior Scientist Agilent Technologies Wilmington, DE USA Page 1 Application Summary There is a need to measure
More informationUsing the PSD for Backflushing on the Agilent 8890 GC System
Application Note Petrochemicals Using the PSD for Backflushing on the Agilent 889 GC System Author Brian Fitz Agilent Technologies, Inc. Wilmington, DE, USA. Abstract An Agilent 889 series GC equipped
More informationGC Method Compliance and Large Valve Oven Application
GC Method Compliance and Large Valve Oven Application Solution Focused Kelly Beard June 20, 2016 Compliant Methods Compliant Methods What are our options? Analyzers and Analyzer Kits High Volume Well Known
More informationAgilent and ASTM. Update on Recent Activities. Page 1
Agilent and ASTM Update on Recent Activities Page 1 ASTM Committees for Refining, Fuels, Petroelum Products ASTM International D2 Petroleum products & lubricants D3 Gas fuels (natural gas) gas fuels D16
More informationGas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593
Application Note Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Authors Kelly Beard and James McCurry Agilent Technologies, Inc. Abstract An Agilent
More informationAgilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass. James D. McCurry, Ph.D. Senior Scientist
Agilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass James D. McCurry, Ph.D. Senior Scientist Agilent Solutions for the Analysis of Ethanol-Based Fuels Derived From Biomass Presented
More informationFast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC
Fast GC SPEEDY GC Slide 4 Optimization Goals Primary: Minimize time for the separation of a given number of peaks Secondary: Maximize the number of peaks separated by a given column Slide 5 HOW CAN THESE
More informationAgilent G2855A Deans Switching System
Agilent G2855A Deans Switching System Installation and Operation Agilent Technologies Notices Agilent Technologies, Inc. 2003 No part of this manual may be reproduced in any form or by any means (including
More informationGC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585
GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc. 850 Centerville
More informationTechnical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)
Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph-mass spectrometer (GC-MS). 2.0 Scope
More informationPage 1. Alternate Carrier Gas Considerations and Faster GC Analysis
Page 1 Alternate Carrier Gas Considerations and Faster GC Analysis Faster GC Total Analytical Cycle Times A Variety of Approaches Pre-Run ALS Set-Up Chromatographic Run Post-Run Bake-Out Post-Run Cool-Down
More informationOperation and Applications of Differential Flow Modulation
Operation and Applications of Differential Flow Modulation H2 Collection channel Column 1 H2 Column 2 FID Roger L Firor, Ph.D. Agilent Technologies Chemical Analysis Group Wilmington, DE USA Flow Modulator
More informationTechnical Procedure for Gas Chromatography (GC-FID)
Technical Procedure for Gas Chromatography (GC-FID) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph (GC- FID). 2.0 Scope This procedure applies to all
More informationGC Analysis of PCBs in Fish Oil by Direct Injection Without Any Sample Cleanup. Philip L. Wylie, Ph.D. Sr. Applications Chemist Agilent Technologies
GC Analysis of PCBs in Fish Oil by Direct Injection Without Any Sample Cleanup Philip L. Wylie, Ph.D. Sr. Applications Chemist Agilent Technologies Pittsburgh Conference, 2008 Paper 1120-1 Tuesday, 8:30
More informationIncreased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel
Application Note Energy and Chemicals Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel Applying the Energy Institute Method IP 8 with an Agilent J&W
More informationTypical Gas Chromatographic System
Page 72 Air Hydrogen Carrier Gas Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector Electrometer Cylinders or Generators Flow Controller Column Recorder/
More informationOptimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides
Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides Application Note Food Testing & Agriculture Authors Limian Zhao and Chin-Kai Meng Agilent Technologies, Inc. 85 Centerville
More informationGC Inlets. An Overview. Simon Jones GC Applications Engineer
GC Inlets An Overview Simon Jones GC Applications Engineer Types of Inlets Purged Packed Split / Splitless Cool On Column Programmable Temperature Vaporization Volatiles Interface Multi Mode Inlet Where
More informationSetting up SilFlow for BackFlush in your GC
Setting up SilFlow for BackFlush in your GC What is backflush and why use it? The BackFlush system eliminates the need to bake heavy sample fractions off the capillary column. Oils, tars and other semivolatile
More informationGAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS
GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS FLASH VAPORISATION INJECTION Split Splitless On-Column COOL INJECTION Large Volume Injection (LVI) On-Column On-Column-SVE (with solvent vapour
More informationAchieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)
Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Bill Wilson and Chin-Kai Meng Agilent Technologies, Inc. 2850 Centerville Road Wilmington,
More informationPRODUCT SPECIFICATION. Product Specification Crystal 9000 Gas Chromatograph. Chromatec Crystal Laboratory Gas Chromatography System
Chromatec Crystal 9000 Laboratory Gas Chromatography System GC Chromatec-Crystal 9000 Crystal 9000 is a leading gas chromatograph in Chromatec GC product line. Crystal 9000 has highest performances and
More informationTechnical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS)
Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the pyrolysis-gas chromatograph-mass spectrometer
More informationMethod Development for Capillary GC Systems. Slide 1
Method Development for Capillary GC Systems Slide 1 AREAS TO OPTIMIZE Injector Carrier gas Column temperature Slide 2 COMMON INJECTOR MODES Vaporization Injection Modes Megabore Direct Split Splitless
More informationHIGH PRESSURE SWITCHING VALVES
HIGH PRESSURE SWITCHING VALVES Rheodyne offers high pressure switching valves to simplify procedures and improve the speed, resolution, and sensitivity of HPLC analysis. The switching valves are available
More informationOnline sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve
Online sample cleanup on the Agilent 129 Infinity LC using a built in 2-position/6-port valve Analysis of Sudan red compounds in paprika powder Application Note Food Authors Angelika Gratzfeld-Huesgen
More informationAnalysis and. Separation of Oxygenates in Hydrocarbon Matrices. Simon Jones Application Engineer Folsom, CA
Analysis and Chromatographic Separation of Oxygenates in Hydrocarbon Matrices Simon Jones Application Engineer Folsom, CA August 20, 2009 Agenda WCOT vs. PLOT columns OxyPlot A Unique stationary phase
More informationImpact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter
Cd The Chrom Doctor Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands. In Part 1 of this series we focused
More informationFausto Munari e Andrea Cadoppi ThermoFisher - Italy
The world leader in serving science Ultra Fast GC Determination of Total Hydrocarbons (C7-C40) and BTEX in Water and Soils through Direct Resistively Heated capillary columns and Robotic Autosampler. Fausto
More informationAgilent G5632 Valve Kit Instructions
Agilent G5632 Valve Kit Instructions Agilent G5632 Valve Kit Instructions Contents Typical Applications 2 Alternating Column Regeneration (2 position/10 port valves only) 2 Dual column selection 3 Sample
More informationAgilent s GC Analyzer Solutions: Introduction to the Controlled Substances Analyzer
Agilent s GC Analyzer Solutions: Introduction to the Controlled Substances Analyzer Get your laboratory on the Analytical FAST TRACK December 2013 1 Today s Challenges Expecting More from Less Business
More informationAgilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve
Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve Technical Note This Technical Note describes the purpose, installation, and configuration and use of a diverter valve
More informationFree and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584
Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN 14105 and ASTM D6584 Introduction With today s increasing concern for the environment and the depletion of fossil
More informationRefinery Gas. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.
Refinery Gas Analysis by Gas Chromatography Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Refinery Gas Analysis Reliability Placing refinery gas analyzers in the field for over
More informationModel 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors
Model 8610C Gas Chromatograph Mounts up to Six Detectors and Five Injectors Ambient to 400 C Temperature Programmable Column Ovens Dimensions: 19 wide x 13.5 high x 14.5 deep Implement virtually any EPA
More informationHigh Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph
Application Note Petrochemicas High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Author James D. McCurry, Ph.D. Agilent Technologies, Inc. Abstract An Agilent
More informationGC Best Practices & Troubleshooting. Group/Presentation Title Agilent Restricted Month ##, 200X
GC Best Practices & Troubleshooting Group/Presentation Title Agilent Restricted Month ##, 200X Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what it is not)
More informationAgilent Valve Kit G4231A/B and G4232A/B/C
Agilent Valve Kit G4231A/B and G4232A/B/C Instructions Contents Typical Applications 2 Dual column selection (G4231A/B or G4232A/B/C) 2 Sample enrichment and sample cleanup (G4231A/B or G4232A/B/C) 2 Alternating
More informationAgilent 6850 Series II Network GC System G2630A Performance Specifications
IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Agilent 6850 Series II Network GC
More informationApplication Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER
Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D7423-09 using AC OXYTRACER Fast Analysis in
More informationOptimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE
Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System Jim McCurry Roger Firor Agilent Technologies Wilmington, DE Page 1 Scope of the Analyzer Designed for ASTM D7798 For
More informationPractical Steps in GC Troubleshooting
Practical Steps in GC Troubleshooting Techniques, Tips, and Tricks Mark Sinnott Application Engineer GC Columns & Supplies Page 1 Everything was just fine and then this happened! How do I go about TROUBLESHOOTING?
More informationLarge Volume Injection of Polycyclic Aromatic Hydrocarbons
JSB is an authorised partner of Large Volume Injection of Polycyclic Aromatic Hydrocarbons Application Note - Environmental #113 Author Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Abstract
More information260 Solvent Vapor Exit Accessory
Theory of operation The Solvent Vapor Exit (SVE) is a GC accessory for performing large volume injections with a Cool On-Column inlet (COC). At the start of the run, the SVE solenoid valve is open as the
More informationDETECTORS Photo Ionization Detector - PID. Overview
Overview DETECTORS The Photo Ionization Detector (PID) responds to all molecules whose ionization potential is below 10.6eV, including aromatics and molecules with carbon double bonds. The PID is nondestructive,
More informationASTM D for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software
ASTM D5501-04 for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software Larry Meeker and Alice J. Di Gioia Waters Corporation Houston Field Laboratory 5909 West Loop, South
More informationDetection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector
Detection of Sulfur Compounds in Natural Gas According to ASTM D554 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector Application Note Author Rebecca Veeneman Abstract Sulfur compounds in natural
More informationASTM D 6730 Detailed Hydrocarbon Analysis
ASTM D 6730 Detailed Hydrocarbon Analysis Jaap de Zeeuw, Jan Pijpelink and Barry Burger Restek Corporation ASTM D 6730-01(2006)e1 Determination of Individual Components in Spark Ignition Engine Fuels as
More informationAgilent InfinityLab Quick Change Valves G4231A/C and G4232C/D
Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Instructions Agilent Valve Kit G4231A/C and G4232C/D - Instructions This technical note describes the installation and application of the Agilent
More informationAgilent Valve Kit G4231A/B/C and G4232B/C/D
Agilent Valve Kit G4231A/B/C and G4232B/C/D Instructions Agilent Valve Kit G4231A/B/C and G4232B/C/D - Instructions This Technical Note describes the installation of the Agilent G4231A/B/C and G4232B/C/D
More informationAchieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)
Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Brian Fitz and Bill Wilson Agilent Technologies, Inc. 285 Centerville Road Wilmington,
More informationSPECIFICATION. Bench top triple quadrupole Mass Spectrometry System coupled to a gas performance Chromatography (GC-MS/MS)
SPECIFICATION Bench top triple quadrupole Mass performance Chromatography (GC-MS/MS) 1. Scope These specifications describe the requirements for a bench top triple quadrupole mass spectrometry system coupled
More informationFast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC
Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC Technical Overview Trace gas analysis is a challenge in today s world. The ability to analyze lower component
More informationUltra Clean Gas Filters Reference Guide ULTRA CLEAN GAS FILTERS
ULTRA CLEAN GAS FILTERS ULTRA CLEAN GAS FILTERS Table of Contents Ultra Clean Gas Filter s for GC and GC/MS...3 Triple Filter for MS/ECD/NPD... 3 Complete Triple Filter Bundle for FID...3 Complete Triple
More informationCustomer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard
7890 GC Standard Agilent Preventive Maintenance provides factory recommended service for your analytical systems to assure reliable operation and the accuracy of your results. Delivered by highly-trained
More informationBeverage Grade Carbon Dioxide
Analysis by Gas Chromatography Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION The Challenge Carbon dioxide, used in the production of carbonated soft drinks and other beverages,
More information[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS
MS Identification of Trace level Impurities from a Non-MS Compatible Mobile Phase Using ACQUITY UPLC System with 2D Technology by Heart-cutting and Online Sample Concentration Bronsky Gopinadh, Dilshad
More informationIFPAC 2003 Dr. Berthold Andres
IFPAC 2003 Dr. Berthold Andres ABB Automation Products Germany Microelectromechanical Systems for Process Analytics Copyright 2002 ABB. All rights reserved. - Process Analyzer and Instrumentation Water
More informationAgilent Multimode Inlet for Gas Chromatography
Agilent Multimode Inlet for Gas Chromatography Technical Note Agilent Multimode Inlet for the 7890A GC Designed to give you ease of use and maximum flexibility, the Agilent Multimode Inlet does everything
More informationDetection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology
Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology Electronic Noses An electronic nose produces a recognizable response based
More informationSELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS
Rev 1 3/6/2004 Selerity Technologies Inc. www.selerity.com SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS What is supercritical fluid chromatography? A chromatographic technique in
More informationGC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method
GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN1413:211 Method Application Note Author James D. McCurry, Ph.D. Agilent Technologies Abstract
More informationPerforming ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software
Install a capillary column in the oven of the SRI GC. The ASTM method suggests a 12 meter.32mm id narrow-bore column coupled with a 2.5 meter guard column but permits the use of any column which exhibits
More informationASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column
ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column Cory S. Fix, Director of Application Development cory.fix@vgcchromatography.com Willie Steinecker,
More informationSTI OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY
OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY Extended range HPLC has been investigated for many years; however, progress was delayed because of
More informationRefinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography
Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography Dr. Robert Lorenz Analytical Chemist Chevron Energy Technology Company Gulf Coast Conference Galveston, TX October
More informationApplication. Gas Chromatography June 1995
Determining Oxygenates in Gasoline: ASTM Method D Application Gas Chromatography June 99 Authors Michael J. Szelewski Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-60 USA Matthew S. Klee
More informationJaap de Zeeuw Restek Middelburg, The Netherlands. Copyrights: Restek Corporation
Liner Selection in GC Jaap de Zeeuw Restek Middelburg, The Netherlands What is a liner? A liner is an inert glass tube that is positioned inside the injection port of the GC The sample solution is introduced
More informationTroubleshooting Tips & Tricks for your GC Analyzer & CFT Application
Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application 7890A/7890B GC Overview October 29, 2014 1 Definitions Carrier Gas Pressurized gas used to transport the sample through the system.
More informationLiquefied Gas Injector. Solution for the Sampling and Analysis of Liquefied Gases
Liquefied Gas Injector Solution for the Sampling and Analysis of Liquefied Gases Safe and Representative Sampling of Liquefied Gases The analysis of impurities and contaminants in liquefied gases is an
More informationAgilent Solutions for the Petrochemical and Oleochemical Industries
e-seminar Series - Agilent Solutions for the Petrochemical and Oleochemical Industries Date: 12 June 2009 Time : 10am Singapore Time For audio, please dial one of the phone numbers listed: International
More informationCustomer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard
Agilent Preventive Maintenance provides factory recommended service for your analytical systems to assure reliable operation and the accuracy of your results. Delivered by highly-trained and certified
More informationProof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut
Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut Application Note Environmental Author Ken Lynam Agilent Technologies, Inc. Abstract Specially designed self-tightening
More informationSame Column and Gas Type Try Different/Faster Velocities. Same Column, Switch He to H2 Carrier Then Try Faster Velocities
: What Are You Willing To Do? Daron Decker Chromatography Technical Specialist Page 1 Agilent Restricted 3 Options Same Column and Gas Type Try Different/Faster Velocities Same Column, Switch He to H2
More informationURB '-carbamoylbiphenyl-3-yl cyclohexylcarbamate. DEA Reference Material Collection. Form Chemical Formula Molecular Weight Melting Point ( o C)
O NH O NH O. GENERAL INFORMATION IUPAC Name: 3'-carbamoylbiphenyl-3-yl cyclohexylcarbamate CAS#: 56-08-6 Synonyms: Source: Appearance: UV max (nm): KDS-03 DEA Reference Material Collection White powder
More informationAUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL
AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL Challenges of Water Monitoring Volatile Organic Compounds (VOCs) can have negative health impacts even at ppb levels VOC concentrations can
More informationGC Injector Liners. Bruker Varian Agilent (HP) Perkin Elmer Finnigan. Phone: Page 4
Bruker Varian Agilent (HP) Perkin Elmer Finnigan www.excelcomponent.com sales@excelcomponent.com Phone: 909-229-6363 Page 4 Deactivated Split/Splitless Injector We offer a full line of Split/Splitless
More informationSulfur Detection at ppb Levels in Light Hydrocarbon Streams
Sulfur Detection at ppb Levels in Light Hydrocarbon Streams Based on a New Super Permeable PLOT Column Agilent Select Low Sulfur Johan Kuipers Channel Training Specialist Oct 12 th, 2010 1 October 18,
More informationSimulated Distillation Analyzers, Software, Standards, Consumables, Training
Simulated Distillation Analyzers, Software, Standards, Consumables, Training www.separationsystems.com Offering the Fullest Range of Optimized Solutions Simulated distillation (SimDis) has been used to
More informationInstalling a Purged Packed Inlet
Installing a Purged Packed Inlet Agilent 6850 Gas Chromatograph This kit contains: Accessory G2651A Description Purged packed column inlet system 1 Top block insulation 1 Bottom block insulation 1 Insulation
More informationApplication Note. Authors. Abstract. Energy & Chemicals
Determination of Aromatic Content in Diesel Fuel According to ASTM D5186 Enhancing the Agilent 126 Infi nity Analytical SFC System with a Flame Ionization Detector Application Note Energy & Chemicals Authors
More informationForensic Blood Alcohol Determination with the Intuvo 9000 GC
Forensic Blood Alcohol Determination with the Intuvo 9000 GC Rebecca Veeneman, Ph.D Applications Chemist The GC Usability Gap Usability lags features and performance Over the years, GC features and performance
More informationperformance productivity reliability
Solutions that meet your demands for: performance productivity reliability Excellent Choices for Global Hydrocarbon Processing Applications Crude Oil & Natural Gas Natural Gas > Return to Table of Contents
More informationTHC Heated FID 3-300A
Rack Mount/Table Top TVOC Analyzer THC Heated FID 3-300A Space saving 19"/3PU space saving rack mount and table top heated emission analyzer for the continuous determination of the mass concentration of
More informationEvaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS
Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS Application Environmental Authors Wei Luan and Chuanhong Tu Agilent Technologies (Shanghai) Co., Ltd. 412 Ying
More informationChromatography. Liquid. HPLC Accessories SupelPRO Valves SUPELCO
8 SupelPRO Valves Order:.800..00 Technical Service:.800.9.0 Web: www.sigma-aldrich.com/supelco SupelPRO Automated Fluidics Instruments Supelco s SupelPRO series are precision, electronically-controlled,
More informationThe Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector
Authors: The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector Introduction Specification of the hydrocarbon composition of LPG is required as traces
More informationDetermination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105
Application Note: 10215 Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105 Fausto Munari, Daniela Cavagnino, Andrea Cadoppi, Thermo Fisher Scientific, Milan,
More informationFrom Helium to Hydrogen: GC-MS Case Study on SVOCs in Water
From Helium to Hydrogen: GC-MS Case Study on SVOCs in Water Jessie Butler Alexander N. Semyonov December 13, 2012 Helium? It s gone! Not only at Disneyland Resort 2 Helium shortage grounds Mickey at Tokyo
More informationHigh Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System
High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System Application Note Authors Frank David Research Institute for Chromatography, Pres.
More informationAgilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011
Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc.
More informationAccuracy Starts Here. Agilent Captiva Filtration Selection Guide
Accuracy Starts Here Agilent Captiva Filtration Selection Guide TAbLe of CoNTeNTS Tips & Tools Learn the core concepts surrounding Solid Phase Extraction and best practices for Sample Prep. View the video
More informationApplication Note. Author. Introduction. Energy and Fuels
Analysis of Free and Total Glycerol in B-100 Biodiesel Methyl Esters Using Agilent Select Biodiesel for Glycerides Application Note Energy and Fuels Author John Oostdijk Agilent Technologies, Inc. Introduction
More informationTechnical Overview. Introduction
Performance characteristics of the Agilent 129 Infinity Binary Pump More resolution and speed for conventional, superficially porous and sub-2 µm column packing material Technical Overview Introduction
More informationApplication Note. Abstract. Authors. Environmental Analysis
High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID using the Agilent Low Thermal Mass (LTM II) System Application Note Environmental Analysis Authors Frank David and Karine Jacq Research
More informationEmerging Techniques; Hyphenated Technologies for Polymer and Biofuel Application
Emerging Techniques; Hyphenated Technologies for Polymer and Biofuel Application Scott Fraser (PhD) National Product Specialist Thermal & Elemental Analysis 1 2009 2010 PerkinElmer Hyphenation Possibilities
More informationInstalling a Purged Packed (PP) Inlet
Installing a Purged Packed (PP) Inlet Agilent 6850 Gas Chromatograph Accessory G2651A These instructions are divided into two parts: Part 1 to prepare the Electronic Pressure Control ("EPC") module for
More information