2-D GC/MS and the Deans Switch: Theory and Applications. Fred Feyerherm, Agilent Technologies Inc. Houston, Tx

Size: px
Start display at page:

Download "2-D GC/MS and the Deans Switch: Theory and Applications. Fred Feyerherm, Agilent Technologies Inc. Houston, Tx"

Transcription

1 2-D GC/MS and the Deans Switch: Theory and Applications Fred Feyerherm, Agilent Technologies Inc. Houston, Tx

2 Introduction Simplified 2-D GC can be used to accomplish the separation of specific analytes from very complex matrices. In complex sample matrices, there are often too many overlapping compounds to allow resolution of the compound(s) of interest, even with the highest resolution columns available. On any single column, the analytes co-elute with several other compounds, making their analysis difficult or impossible. This approach can solve difficult analytical problems with a relatively low cost, easy to use system that is very robust and easy to maintain.

3 Multidimensional (2-D) GC Very old (>25 yrs) but powerful separation technique. Based on cutting peak(s) from one GC column onto another with stationary phase of different selectivity (polarity). Compounds that co-elute with analyte on the first column separate from the analyte(s) of interest on the second column of different polarity. THC in Hair NIDA drugs in Oral Fluids THC in Blood Especially useful at ultra-trace concentrations (<1 ng/ml)

4 Why 2-D GC now? What s Changed? 6890 Simplified 2-D GC systems are much easier to use Micro-fluidics Deans Switch Column connections are easier, zero dead volume, inert, and reliable. Setting gas flows done with EPC and Flow Calculator. Retention time drift greatly reduced with modern oven and EPC. Inertness problems with switch hardware eliminated with surface coatings. RT control is very tight and the new switch is very quick.

5 Simplified 2-Dimensional GC 7683 Autosampler Cut FID1 FID2 Column 1 Column N GC Deans Switch

6 Deans Switch System: Used to heart cut individual peaks from DB-17ms to DB-1ms column FID T2 S/S Inlet restrictor 0.1 mm UDFS T1 purge restrictor solenoid valve PCM DB-17ms 15m x 0.25 x 0.25 um MS T3 DB-1ms 15m x 0.25 x 0.25 um Oven Wall

7 Deans Switch System Used to heart cut peak from DB-17MS to DB-1MS FID restrictor 0.3 m x.1 mm FS solenoid valve Injector PCM DB-17MS 15m x.25 x 0.25 um MSD DB-1MS 15m x.25 x 0.25 um

8 Deans Switch System Switch off, DB-17MS effluent goes to monitor FID (no cut) FID restrictor solenoid valve (off) Injector PCM 27 psi 20 psi DB-17MS MSD DB-1MS

9 Deans Switch System Switch on, DB-17MS effluent is cut to DB-1MS FID restrictor solenoid valve (on) Injector PCM 27 psi 20 psi DB-17MS MSD DB-1MS

10 Deans Switch System (after the cut) Switch off, DB-17MS effluent goes to monitor FID FID restrictor solenoid valve (off) Injector PCM 27 psi 20 psi DB-17MS MSD DB-1MS

11 EPC Backflush to save time and clean out first column! Program Inlet pressure down to 0.5 psi after the last cut, then after the last analyte elutes from column 2, program the PCM to 60 psi so heavies are backflushed out split vent. FID restrictor solenoid valve (off) S/S Inlet PCM 27 psi 0.5 psi 20 psi 60 psi DB-17ms MS DB-1ms

12 New Micro-fluidics Deans Switch

13 Deans Switch column connections

14 Simple Run Table Entry Cuts Sample onto Column 2

15 Method Developers Calculator to Tools correctly set flows and restrictor size

16 Cryo Trap Re-focuses Sample on Column 2 Sharpens peak very important for lower concentrations Adds ability to perform 2nd oven ramp FID restrictor solenoid valve (on) Injector AUX 4 27 psi 20 psi DB-17MS MSD Cryo DB-1MS

17 First Application Example 0.1 pg/mg THC in Hair DB-35ms DB-5ms 2002 at Northwest Toxicology in Salt Lake City Using Finnigan TSQ 7000 Triple Quadrupole MS Same samples analyzed with Deans switch

18 Low Level Detection Required SAMSHA guidelines set confirmatory cutoff at 0.05 pg/mg 20 mg sample size used washed, cut up, digested, extracted derivitized with PFPA/PFPOH 50 ul final extract volume (in toluene) THC-COOH concentration in final extract = 20 fg/ul

19 GC/MS Conditions Agilent 6890N N Negative chemical ionization (NCI) mode -Autotuned Source temperature 160C SIM ions (5 ions): THC-COOH 620, 492, 472 d3-thc-cooh 623, 495 Ammonia reagent (buffer) gas (5X increase in sensitivity vs methane)

20 Ideal Conditions (2 ul unextracted Abundance standard) 1200 Ion ( to ): THC-03.D Ion ( to ): THC-03.D 8.45 Ion ( to ): THC-03.D pg/mg (unextracted) Time-->

21 Abundance Hair Matrix Effect: 0.1 pg/mg Ion ( to ): THC-05.D Ion ( to ): THC-05.D Ion ( to ): THC-05.D Time-->

22 Deans Switch System (THC in Hair) Heartcut THC peak from DB-35MS to DB-5MS FID restrictor 4.2m x.1 mm FS tubing UNIS Inlet solenoid valve PCM DB-35MS 40m x.18 mm x 0.18 um MSD DB-5MS 40m x.18 mm x 0.18 um

23 Deans Switch System Switch off, DB-35MS effluent goes to monitor FID (no cut) FID restrictor solenoid valve (off) UNIS Inlet PCM 80 psi 50 psi DB-35MS MSD DB-5MS

24 Deans Switch System Switch on, DB-35MS effluent is cut to DB-5MS FID restrictor solenoid valve (on) UNIS Inlet PCM 80 psi 50 psi DB-35MS MSD DB-5MS

25 Results: 0.1 pg/mg THC-COOH in Hair Abundance 1100 Ion ( to ): THC-032.D m/e pg/mg THC-COOH in hair Time-->

26 Second Application Example THC in Oral Fluid TMS derivative DB-17ms DB1ms Both are 15m x 0.25 x 0.25

27 Deans Switch System Used to heart cut peak from DB-17MS to DB-1MS FID restrictor 0.3 m x.1 mm FS solenoid valve Injector PCM DB-17MS 15m x.25 x 0.25 um MSD DB-1MS 15m x.25 x 0.25 um

28 Method Developers Calculator to Tools correctly set flows and restrictor size

29 THC parent drug in Oral Fluid Abundance 1400 Ion ( to ): THC03.D\data.ms Ion ( to ): THC03.D\data.ms Ion ( to ): THC03.D\data.ms ng/ml Time-->

30 Third Application Example THC parent and THCA in Blood Arizona DPS TMS derivative RTX-200 DB17ms

31 Example of Inadequate Resolution DB-17ms DB-1ms Abundance Ion ( to ): 02-A.D Ion ( to ): 02-A.D Ion ( to ): 02-A.D Time-->

32 Deans Calculator

33 FID channel of the Deans Switch with no cuts to the secondary column Abundance Signal: 05B.D\FID1A.CH Time-->

34 Cut Times determined from FID

35 Abundance Signal: 05C.D\FID1A.CH THC Cut THCC Cut Abundance Time--> Ion ( to ): 05C.D Ion ( to ): 05C.D THCC-TMS Abundance 1000 Time--> TIC: 08g.D THC- TMS Co-eluting peaks from primary column THCC-TMS THC-TMS Time-->

36 0.5 ng/ml THC-TMS in Blood Abundance Ion ( to ): 011SP011.D Ion ( to ): 011SP011.D Ion ( to ): 011SP011.D THC-TMS, 0.5 ng/ml Ion Time-->

37 Summary Developments in GC hardware in recent years have made Deans switch systems easier to use and maintain. 2D GC is a powerful tool that can be used by itself or combined with other selective techniques to solve difficult separation problems. 2D GCMS (GC/GC/MS) can be used instead of GC/MS/MS in many applications.

Simple Heart Cutting with Deans. Capillary Flow Technolgy. Simon Jones Applications Engineer

Simple Heart Cutting with Deans. Capillary Flow Technolgy. Simon Jones Applications Engineer Simple Heart Cutting with Deans Switch and Backflushing with Capillary Flow Technolgy Simon Jones Applications Engineer GC or GC/MS Analysis in Complex Matrices In complex sample matrices, there are often

More information

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D759 Application Petrochemical Author James D. McCurry Agilent Technologies 285 Centerville

More information

C2, C3, C4 Monomer Analysis

C2, C3, C4 Monomer Analysis C2, C3, C4 Monomer Analysis Malgorzata Sierocinska Agilent Technologies Waldbronn Page 1 Why Analyze Monomers? To Insure Consistent Production of High Quality Polymer Protect against food contamination

More information

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels James D. McCurry Senior Scientist Agilent Technologies Wilmington, DE USA Page 1 Application Summary There is a need to measure

More information

Using the PSD for Backflushing on the Agilent 8890 GC System

Using the PSD for Backflushing on the Agilent 8890 GC System Application Note Petrochemicals Using the PSD for Backflushing on the Agilent 889 GC System Author Brian Fitz Agilent Technologies, Inc. Wilmington, DE, USA. Abstract An Agilent 889 series GC equipped

More information

GC Method Compliance and Large Valve Oven Application

GC Method Compliance and Large Valve Oven Application GC Method Compliance and Large Valve Oven Application Solution Focused Kelly Beard June 20, 2016 Compliant Methods Compliant Methods What are our options? Analyzers and Analyzer Kits High Volume Well Known

More information

Agilent and ASTM. Update on Recent Activities. Page 1

Agilent and ASTM. Update on Recent Activities. Page 1 Agilent and ASTM Update on Recent Activities Page 1 ASTM Committees for Refining, Fuels, Petroelum Products ASTM International D2 Petroleum products & lubricants D3 Gas fuels (natural gas) gas fuels D16

More information

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Application Note Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Authors Kelly Beard and James McCurry Agilent Technologies, Inc. Abstract An Agilent

More information

Agilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass. James D. McCurry, Ph.D. Senior Scientist

Agilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass. James D. McCurry, Ph.D. Senior Scientist Agilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass James D. McCurry, Ph.D. Senior Scientist Agilent Solutions for the Analysis of Ethanol-Based Fuels Derived From Biomass Presented

More information

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC Fast GC SPEEDY GC Slide 4 Optimization Goals Primary: Minimize time for the separation of a given number of peaks Secondary: Maximize the number of peaks separated by a given column Slide 5 HOW CAN THESE

More information

Agilent G2855A Deans Switching System

Agilent G2855A Deans Switching System Agilent G2855A Deans Switching System Installation and Operation Agilent Technologies Notices Agilent Technologies, Inc. 2003 No part of this manual may be reproduced in any form or by any means (including

More information

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585 GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc. 850 Centerville

More information

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS) Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph-mass spectrometer (GC-MS). 2.0 Scope

More information

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis Page 1 Alternate Carrier Gas Considerations and Faster GC Analysis Faster GC Total Analytical Cycle Times A Variety of Approaches Pre-Run ALS Set-Up Chromatographic Run Post-Run Bake-Out Post-Run Cool-Down

More information

Operation and Applications of Differential Flow Modulation

Operation and Applications of Differential Flow Modulation Operation and Applications of Differential Flow Modulation H2 Collection channel Column 1 H2 Column 2 FID Roger L Firor, Ph.D. Agilent Technologies Chemical Analysis Group Wilmington, DE USA Flow Modulator

More information

Technical Procedure for Gas Chromatography (GC-FID)

Technical Procedure for Gas Chromatography (GC-FID) Technical Procedure for Gas Chromatography (GC-FID) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph (GC- FID). 2.0 Scope This procedure applies to all

More information

GC Analysis of PCBs in Fish Oil by Direct Injection Without Any Sample Cleanup. Philip L. Wylie, Ph.D. Sr. Applications Chemist Agilent Technologies

GC Analysis of PCBs in Fish Oil by Direct Injection Without Any Sample Cleanup. Philip L. Wylie, Ph.D. Sr. Applications Chemist Agilent Technologies GC Analysis of PCBs in Fish Oil by Direct Injection Without Any Sample Cleanup Philip L. Wylie, Ph.D. Sr. Applications Chemist Agilent Technologies Pittsburgh Conference, 2008 Paper 1120-1 Tuesday, 8:30

More information

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel Application Note Energy and Chemicals Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel Applying the Energy Institute Method IP 8 with an Agilent J&W

More information

Typical Gas Chromatographic System

Typical Gas Chromatographic System Page 72 Air Hydrogen Carrier Gas Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Regulators Injection Port Detector Electrometer Cylinders or Generators Flow Controller Column Recorder/

More information

Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides

Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides Application Note Food Testing & Agriculture Authors Limian Zhao and Chin-Kai Meng Agilent Technologies, Inc. 85 Centerville

More information

GC Inlets. An Overview. Simon Jones GC Applications Engineer

GC Inlets. An Overview. Simon Jones GC Applications Engineer GC Inlets An Overview Simon Jones GC Applications Engineer Types of Inlets Purged Packed Split / Splitless Cool On Column Programmable Temperature Vaporization Volatiles Interface Multi Mode Inlet Where

More information

Setting up SilFlow for BackFlush in your GC

Setting up SilFlow for BackFlush in your GC Setting up SilFlow for BackFlush in your GC What is backflush and why use it? The BackFlush system eliminates the need to bake heavy sample fractions off the capillary column. Oils, tars and other semivolatile

More information

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS FLASH VAPORISATION INJECTION Split Splitless On-Column COOL INJECTION Large Volume Injection (LVI) On-Column On-Column-SVE (with solvent vapour

More information

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI) Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Bill Wilson and Chin-Kai Meng Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

PRODUCT SPECIFICATION. Product Specification Crystal 9000 Gas Chromatograph. Chromatec Crystal Laboratory Gas Chromatography System

PRODUCT SPECIFICATION. Product Specification Crystal 9000 Gas Chromatograph. Chromatec Crystal Laboratory Gas Chromatography System Chromatec Crystal 9000 Laboratory Gas Chromatography System GC Chromatec-Crystal 9000 Crystal 9000 is a leading gas chromatograph in Chromatec GC product line. Crystal 9000 has highest performances and

More information

Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS)

Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS) Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the pyrolysis-gas chromatograph-mass spectrometer

More information

Method Development for Capillary GC Systems. Slide 1

Method Development for Capillary GC Systems. Slide 1 Method Development for Capillary GC Systems Slide 1 AREAS TO OPTIMIZE Injector Carrier gas Column temperature Slide 2 COMMON INJECTOR MODES Vaporization Injection Modes Megabore Direct Split Splitless

More information

HIGH PRESSURE SWITCHING VALVES

HIGH PRESSURE SWITCHING VALVES HIGH PRESSURE SWITCHING VALVES Rheodyne offers high pressure switching valves to simplify procedures and improve the speed, resolution, and sensitivity of HPLC analysis. The switching valves are available

More information

Online sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve

Online sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve Online sample cleanup on the Agilent 129 Infinity LC using a built in 2-position/6-port valve Analysis of Sudan red compounds in paprika powder Application Note Food Authors Angelika Gratzfeld-Huesgen

More information

Analysis and. Separation of Oxygenates in Hydrocarbon Matrices. Simon Jones Application Engineer Folsom, CA

Analysis and. Separation of Oxygenates in Hydrocarbon Matrices. Simon Jones Application Engineer Folsom, CA Analysis and Chromatographic Separation of Oxygenates in Hydrocarbon Matrices Simon Jones Application Engineer Folsom, CA August 20, 2009 Agenda WCOT vs. PLOT columns OxyPlot A Unique stationary phase

More information

Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter

Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter Cd The Chrom Doctor Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands. In Part 1 of this series we focused

More information

Fausto Munari e Andrea Cadoppi ThermoFisher - Italy

Fausto Munari e Andrea Cadoppi ThermoFisher - Italy The world leader in serving science Ultra Fast GC Determination of Total Hydrocarbons (C7-C40) and BTEX in Water and Soils through Direct Resistively Heated capillary columns and Robotic Autosampler. Fausto

More information

Agilent G5632 Valve Kit Instructions

Agilent G5632 Valve Kit Instructions Agilent G5632 Valve Kit Instructions Agilent G5632 Valve Kit Instructions Contents Typical Applications 2 Alternating Column Regeneration (2 position/10 port valves only) 2 Dual column selection 3 Sample

More information

Agilent s GC Analyzer Solutions: Introduction to the Controlled Substances Analyzer

Agilent s GC Analyzer Solutions: Introduction to the Controlled Substances Analyzer Agilent s GC Analyzer Solutions: Introduction to the Controlled Substances Analyzer Get your laboratory on the Analytical FAST TRACK December 2013 1 Today s Challenges Expecting More from Less Business

More information

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve Technical Note This Technical Note describes the purpose, installation, and configuration and use of a diverter valve

More information

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584 Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN 14105 and ASTM D6584 Introduction With today s increasing concern for the environment and the depletion of fossil

More information

Refinery Gas. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Refinery Gas. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results. Refinery Gas Analysis by Gas Chromatography Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Refinery Gas Analysis Reliability Placing refinery gas analyzers in the field for over

More information

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors Model 8610C Gas Chromatograph Mounts up to Six Detectors and Five Injectors Ambient to 400 C Temperature Programmable Column Ovens Dimensions: 19 wide x 13.5 high x 14.5 deep Implement virtually any EPA

More information

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Application Note Petrochemicas High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Author James D. McCurry, Ph.D. Agilent Technologies, Inc. Abstract An Agilent

More information

GC Best Practices & Troubleshooting. Group/Presentation Title Agilent Restricted Month ##, 200X

GC Best Practices & Troubleshooting. Group/Presentation Title Agilent Restricted Month ##, 200X GC Best Practices & Troubleshooting Group/Presentation Title Agilent Restricted Month ##, 200X Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what it is not)

More information

Agilent Valve Kit G4231A/B and G4232A/B/C

Agilent Valve Kit G4231A/B and G4232A/B/C Agilent Valve Kit G4231A/B and G4232A/B/C Instructions Contents Typical Applications 2 Dual column selection (G4231A/B or G4232A/B/C) 2 Sample enrichment and sample cleanup (G4231A/B or G4232A/B/C) 2 Alternating

More information

Agilent 6850 Series II Network GC System G2630A Performance Specifications

Agilent 6850 Series II Network GC System G2630A Performance Specifications IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Agilent 6850 Series II Network GC

More information

Application Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER

Application Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D7423-09 using AC OXYTRACER Fast Analysis in

More information

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System Jim McCurry Roger Firor Agilent Technologies Wilmington, DE Page 1 Scope of the Analyzer Designed for ASTM D7798 For

More information

Practical Steps in GC Troubleshooting

Practical Steps in GC Troubleshooting Practical Steps in GC Troubleshooting Techniques, Tips, and Tricks Mark Sinnott Application Engineer GC Columns & Supplies Page 1 Everything was just fine and then this happened! How do I go about TROUBLESHOOTING?

More information

Large Volume Injection of Polycyclic Aromatic Hydrocarbons

Large Volume Injection of Polycyclic Aromatic Hydrocarbons JSB is an authorised partner of Large Volume Injection of Polycyclic Aromatic Hydrocarbons Application Note - Environmental #113 Author Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Abstract

More information

260 Solvent Vapor Exit Accessory

260 Solvent Vapor Exit Accessory Theory of operation The Solvent Vapor Exit (SVE) is a GC accessory for performing large volume injections with a Cool On-Column inlet (COC). At the start of the run, the SVE solenoid valve is open as the

More information

DETECTORS Photo Ionization Detector - PID. Overview

DETECTORS Photo Ionization Detector - PID. Overview Overview DETECTORS The Photo Ionization Detector (PID) responds to all molecules whose ionization potential is below 10.6eV, including aromatics and molecules with carbon double bonds. The PID is nondestructive,

More information

ASTM D for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software

ASTM D for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software ASTM D5501-04 for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software Larry Meeker and Alice J. Di Gioia Waters Corporation Houston Field Laboratory 5909 West Loop, South

More information

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector Detection of Sulfur Compounds in Natural Gas According to ASTM D554 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector Application Note Author Rebecca Veeneman Abstract Sulfur compounds in natural

More information

ASTM D 6730 Detailed Hydrocarbon Analysis

ASTM D 6730 Detailed Hydrocarbon Analysis ASTM D 6730 Detailed Hydrocarbon Analysis Jaap de Zeeuw, Jan Pijpelink and Barry Burger Restek Corporation ASTM D 6730-01(2006)e1 Determination of Individual Components in Spark Ignition Engine Fuels as

More information

Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D

Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Instructions Agilent Valve Kit G4231A/C and G4232C/D - Instructions This technical note describes the installation and application of the Agilent

More information

Agilent Valve Kit G4231A/B/C and G4232B/C/D

Agilent Valve Kit G4231A/B/C and G4232B/C/D Agilent Valve Kit G4231A/B/C and G4232B/C/D Instructions Agilent Valve Kit G4231A/B/C and G4232B/C/D - Instructions This Technical Note describes the installation of the Agilent G4231A/B/C and G4232B/C/D

More information

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI) Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Brian Fitz and Bill Wilson Agilent Technologies, Inc. 285 Centerville Road Wilmington,

More information

SPECIFICATION. Bench top triple quadrupole Mass Spectrometry System coupled to a gas performance Chromatography (GC-MS/MS)

SPECIFICATION. Bench top triple quadrupole Mass Spectrometry System coupled to a gas performance Chromatography (GC-MS/MS) SPECIFICATION Bench top triple quadrupole Mass performance Chromatography (GC-MS/MS) 1. Scope These specifications describe the requirements for a bench top triple quadrupole mass spectrometry system coupled

More information

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC Technical Overview Trace gas analysis is a challenge in today s world. The ability to analyze lower component

More information

Ultra Clean Gas Filters Reference Guide ULTRA CLEAN GAS FILTERS

Ultra Clean Gas Filters Reference Guide ULTRA CLEAN GAS FILTERS ULTRA CLEAN GAS FILTERS ULTRA CLEAN GAS FILTERS Table of Contents Ultra Clean Gas Filter s for GC and GC/MS...3 Triple Filter for MS/ECD/NPD... 3 Complete Triple Filter Bundle for FID...3 Complete Triple

More information

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard 7890 GC Standard Agilent Preventive Maintenance provides factory recommended service for your analytical systems to assure reliable operation and the accuracy of your results. Delivered by highly-trained

More information

Beverage Grade Carbon Dioxide

Beverage Grade Carbon Dioxide Analysis by Gas Chromatography Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION The Challenge Carbon dioxide, used in the production of carbonated soft drinks and other beverages,

More information

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS MS Identification of Trace level Impurities from a Non-MS Compatible Mobile Phase Using ACQUITY UPLC System with 2D Technology by Heart-cutting and Online Sample Concentration Bronsky Gopinadh, Dilshad

More information

IFPAC 2003 Dr. Berthold Andres

IFPAC 2003 Dr. Berthold Andres IFPAC 2003 Dr. Berthold Andres ABB Automation Products Germany Microelectromechanical Systems for Process Analytics Copyright 2002 ABB. All rights reserved. - Process Analyzer and Instrumentation Water

More information

Agilent Multimode Inlet for Gas Chromatography

Agilent Multimode Inlet for Gas Chromatography Agilent Multimode Inlet for Gas Chromatography Technical Note Agilent Multimode Inlet for the 7890A GC Designed to give you ease of use and maximum flexibility, the Agilent Multimode Inlet does everything

More information

Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology

Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology Electronic Noses An electronic nose produces a recognizable response based

More information

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS Rev 1 3/6/2004 Selerity Technologies Inc. www.selerity.com SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS What is supercritical fluid chromatography? A chromatographic technique in

More information

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN1413:211 Method Application Note Author James D. McCurry, Ph.D. Agilent Technologies Abstract

More information

Performing ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software

Performing ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software Install a capillary column in the oven of the SRI GC. The ASTM method suggests a 12 meter.32mm id narrow-bore column coupled with a 2.5 meter guard column but permits the use of any column which exhibits

More information

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column Cory S. Fix, Director of Application Development cory.fix@vgcchromatography.com Willie Steinecker,

More information

STI OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

STI OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY Extended range HPLC has been investigated for many years; however, progress was delayed because of

More information

Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography

Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography Dr. Robert Lorenz Analytical Chemist Chevron Energy Technology Company Gulf Coast Conference Galveston, TX October

More information

Application. Gas Chromatography June 1995

Application. Gas Chromatography June 1995 Determining Oxygenates in Gasoline: ASTM Method D Application Gas Chromatography June 99 Authors Michael J. Szelewski Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-60 USA Matthew S. Klee

More information

Jaap de Zeeuw Restek Middelburg, The Netherlands. Copyrights: Restek Corporation

Jaap de Zeeuw Restek Middelburg, The Netherlands. Copyrights: Restek Corporation Liner Selection in GC Jaap de Zeeuw Restek Middelburg, The Netherlands What is a liner? A liner is an inert glass tube that is positioned inside the injection port of the GC The sample solution is introduced

More information

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application 7890A/7890B GC Overview October 29, 2014 1 Definitions Carrier Gas Pressurized gas used to transport the sample through the system.

More information

Liquefied Gas Injector. Solution for the Sampling and Analysis of Liquefied Gases

Liquefied Gas Injector. Solution for the Sampling and Analysis of Liquefied Gases Liquefied Gas Injector Solution for the Sampling and Analysis of Liquefied Gases Safe and Representative Sampling of Liquefied Gases The analysis of impurities and contaminants in liquefied gases is an

More information

Agilent Solutions for the Petrochemical and Oleochemical Industries

Agilent Solutions for the Petrochemical and Oleochemical Industries e-seminar Series - Agilent Solutions for the Petrochemical and Oleochemical Industries Date: 12 June 2009 Time : 10am Singapore Time For audio, please dial one of the phone numbers listed: International

More information

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard Agilent Preventive Maintenance provides factory recommended service for your analytical systems to assure reliable operation and the accuracy of your results. Delivered by highly-trained and certified

More information

Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut

Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut Application Note Environmental Author Ken Lynam Agilent Technologies, Inc. Abstract Specially designed self-tightening

More information

Same Column and Gas Type Try Different/Faster Velocities. Same Column, Switch He to H2 Carrier Then Try Faster Velocities

Same Column and Gas Type Try Different/Faster Velocities. Same Column, Switch He to H2 Carrier Then Try Faster Velocities : What Are You Willing To Do? Daron Decker Chromatography Technical Specialist Page 1 Agilent Restricted 3 Options Same Column and Gas Type Try Different/Faster Velocities Same Column, Switch He to H2

More information

URB '-carbamoylbiphenyl-3-yl cyclohexylcarbamate. DEA Reference Material Collection. Form Chemical Formula Molecular Weight Melting Point ( o C)

URB '-carbamoylbiphenyl-3-yl cyclohexylcarbamate. DEA Reference Material Collection. Form Chemical Formula Molecular Weight Melting Point ( o C) O NH O NH O. GENERAL INFORMATION IUPAC Name: 3'-carbamoylbiphenyl-3-yl cyclohexylcarbamate CAS#: 56-08-6 Synonyms: Source: Appearance: UV max (nm): KDS-03 DEA Reference Material Collection White powder

More information

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL Challenges of Water Monitoring Volatile Organic Compounds (VOCs) can have negative health impacts even at ppb levels VOC concentrations can

More information

GC Injector Liners. Bruker Varian Agilent (HP) Perkin Elmer Finnigan. Phone: Page 4

GC Injector Liners. Bruker Varian Agilent (HP) Perkin Elmer Finnigan.   Phone: Page 4 Bruker Varian Agilent (HP) Perkin Elmer Finnigan www.excelcomponent.com sales@excelcomponent.com Phone: 909-229-6363 Page 4 Deactivated Split/Splitless Injector We offer a full line of Split/Splitless

More information

Sulfur Detection at ppb Levels in Light Hydrocarbon Streams

Sulfur Detection at ppb Levels in Light Hydrocarbon Streams Sulfur Detection at ppb Levels in Light Hydrocarbon Streams Based on a New Super Permeable PLOT Column Agilent Select Low Sulfur Johan Kuipers Channel Training Specialist Oct 12 th, 2010 1 October 18,

More information

Simulated Distillation Analyzers, Software, Standards, Consumables, Training

Simulated Distillation Analyzers, Software, Standards, Consumables, Training Simulated Distillation Analyzers, Software, Standards, Consumables, Training www.separationsystems.com Offering the Fullest Range of Optimized Solutions Simulated distillation (SimDis) has been used to

More information

Installing a Purged Packed Inlet

Installing a Purged Packed Inlet Installing a Purged Packed Inlet Agilent 6850 Gas Chromatograph This kit contains: Accessory G2651A Description Purged packed column inlet system 1 Top block insulation 1 Bottom block insulation 1 Insulation

More information

Application Note. Authors. Abstract. Energy & Chemicals

Application Note. Authors. Abstract. Energy & Chemicals Determination of Aromatic Content in Diesel Fuel According to ASTM D5186 Enhancing the Agilent 126 Infi nity Analytical SFC System with a Flame Ionization Detector Application Note Energy & Chemicals Authors

More information

Forensic Blood Alcohol Determination with the Intuvo 9000 GC

Forensic Blood Alcohol Determination with the Intuvo 9000 GC Forensic Blood Alcohol Determination with the Intuvo 9000 GC Rebecca Veeneman, Ph.D Applications Chemist The GC Usability Gap Usability lags features and performance Over the years, GC features and performance

More information

performance productivity reliability

performance productivity reliability Solutions that meet your demands for: performance productivity reliability Excellent Choices for Global Hydrocarbon Processing Applications Crude Oil & Natural Gas Natural Gas > Return to Table of Contents

More information

THC Heated FID 3-300A

THC Heated FID 3-300A Rack Mount/Table Top TVOC Analyzer THC Heated FID 3-300A Space saving 19"/3PU space saving rack mount and table top heated emission analyzer for the continuous determination of the mass concentration of

More information

Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS

Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS Application Environmental Authors Wei Luan and Chuanhong Tu Agilent Technologies (Shanghai) Co., Ltd. 412 Ying

More information

Chromatography. Liquid. HPLC Accessories SupelPRO Valves SUPELCO

Chromatography. Liquid. HPLC Accessories SupelPRO Valves SUPELCO 8 SupelPRO Valves Order:.800..00 Technical Service:.800.9.0 Web: www.sigma-aldrich.com/supelco SupelPRO Automated Fluidics Instruments Supelco s SupelPRO series are precision, electronically-controlled,

More information

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector Authors: The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector Introduction Specification of the hydrocarbon composition of LPG is required as traces

More information

Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105

Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105 Application Note: 10215 Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105 Fausto Munari, Daniela Cavagnino, Andrea Cadoppi, Thermo Fisher Scientific, Milan,

More information

From Helium to Hydrogen: GC-MS Case Study on SVOCs in Water

From Helium to Hydrogen: GC-MS Case Study on SVOCs in Water From Helium to Hydrogen: GC-MS Case Study on SVOCs in Water Jessie Butler Alexander N. Semyonov December 13, 2012 Helium? It s gone! Not only at Disneyland Resort 2 Helium shortage grounds Mickey at Tokyo

More information

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System Application Note Authors Frank David Research Institute for Chromatography, Pres.

More information

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc.

More information

Accuracy Starts Here. Agilent Captiva Filtration Selection Guide

Accuracy Starts Here. Agilent Captiva Filtration Selection Guide Accuracy Starts Here Agilent Captiva Filtration Selection Guide TAbLe of CoNTeNTS Tips & Tools Learn the core concepts surrounding Solid Phase Extraction and best practices for Sample Prep. View the video

More information

Application Note. Author. Introduction. Energy and Fuels

Application Note. Author. Introduction. Energy and Fuels Analysis of Free and Total Glycerol in B-100 Biodiesel Methyl Esters Using Agilent Select Biodiesel for Glycerides Application Note Energy and Fuels Author John Oostdijk Agilent Technologies, Inc. Introduction

More information

Technical Overview. Introduction

Technical Overview. Introduction Performance characteristics of the Agilent 129 Infinity Binary Pump More resolution and speed for conventional, superficially porous and sub-2 µm column packing material Technical Overview Introduction

More information

Application Note. Abstract. Authors. Environmental Analysis

Application Note. Abstract. Authors. Environmental Analysis High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID using the Agilent Low Thermal Mass (LTM II) System Application Note Environmental Analysis Authors Frank David and Karine Jacq Research

More information

Emerging Techniques; Hyphenated Technologies for Polymer and Biofuel Application

Emerging Techniques; Hyphenated Technologies for Polymer and Biofuel Application Emerging Techniques; Hyphenated Technologies for Polymer and Biofuel Application Scott Fraser (PhD) National Product Specialist Thermal & Elemental Analysis 1 2009 2010 PerkinElmer Hyphenation Possibilities

More information

Installing a Purged Packed (PP) Inlet

Installing a Purged Packed (PP) Inlet Installing a Purged Packed (PP) Inlet Agilent 6850 Gas Chromatograph Accessory G2651A These instructions are divided into two parts: Part 1 to prepare the Electronic Pressure Control ("EPC") module for

More information