Large Volume Injection of Polycyclic Aromatic Hydrocarbons

Size: px
Start display at page:

Download "Large Volume Injection of Polycyclic Aromatic Hydrocarbons"

Transcription

1 JSB is an authorised partner of Large Volume Injection of Polycyclic Aromatic Hydrocarbons Application Note - Environmental #113 Author Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Abstract Polycyclic Aromatic Hydrocarbons (PAHs) are formed from incomplete burning of carbon containing fuel. There are thousands of PAH compounds in the environment, and of those; there are several that have been established to be of concern for the environment. Extraction of PAH compounds involves a large amount of sample and solvent, and because of this, there is a lot of solvent waste. The use of Large Volume Injection (LVI) in conjunction with a Programmable Temperature Vaporizer (PTV) aids in eliminating some of this solvent waste, and reduces labor and shipping costs due to the ability to extract smaller volumes of sample without sacrificing sensitivity. This analysis will compare PAH compound response of a standard injection versus a large volume injection. Introduction: The most common practice for extraction of PAH compounds from a water matrix involves one liter of sample and more than 150mls of solvent. The sample goes through several liquid-liquid extraction and evaporation steps in order to achieve the final extract for sampling and analysis. USEPA Method 3511 proposes a micro-extraction technique that uses less than 5mls of sample and approximately 2mls of solvent. This technique is also much less time consuming. However, micro-extraction is not nearly as sensitive when analyzing for PAH compounds. In answer to this problem, laboratories can utilize Large Volume Injection in conjunction with a Programmable Temperature Vaporizer. When using this sampling technique, an analyst has the ability to inject several microliters of sample into the GC inlet, thus multiplying the amount of sample to be detected and analyzed by the MS. Consequently, sensitivity can be increased while extraction time and solvent use can be decreased. Discussion: As EPA detection limits get lower and lower, laboratories need to adapt to these expectations. Large Volume Injections enable laboratories to lower detection limits because the amount of sample introduced to the system is increased. Programmable Temperature Vaporization using solvent split mode is the usual technique for solvent elimination and pre-concentration of analytes. During vaporization, the analytes are transferred to the column for separation and analysis. When using LVI, the PTV inlet is set to solvent vent mode. The sample is introduced at a low temperature, and the solvent is eliminated using the purge flow to the split vent. The analytes are retained on the inlet liner while the solvent is vented. Next, the inlet is quickly heated in order transfer the analytes to the GC column in splitless mode. After the transfer, the inlet is set to purge any residual contaminants from the inlet.

2 ANNE JUREK Using LVI in conjunction with PTV enables laboratories to introduce more sample onto the GC column, therefore increasing sensitivity. This study will compare a 1µl injection of PAH compounds using a standard split/splitless injection to a 5µl injection of the same PAH compounds utilizing LVI and PTV. Experimental: The sampling system used for this analysis was the EST Analytical FLEX autosampler fitted with a 10µl liquid syringe. The Agilent 7890 GC and 5975 MS were used for separation and analysis. A Restek Rxi-5 Sil MS 30m x 250mm x 0.25µm column was mounted in the GC. The Agilent Split/Splitless inlet was used for the 1ul injections while the Titan PTV LVI was used for the 5ul injections. Refer to Tables 1 and 2 for the sampling and analysis parameters. Autosampler Flex (1µl) Flex (5µl) General Method Type Liquid Liquid Rinse Rinse Volume Rinse Fill Rate Rinse Cycles Rinse Dispense Rate 60% (6µl) 70% (7µl) 50% 50% % 100% Waste Depth 50% 50% Sample Sample Volume 10% (1µl) 50% (5µl) Rinse Volume 50% (5µl) 60% (6µl) Rinse Cycles 1 1 Pump Volume 50% (5µl) 60% (6µl) Pump Cycles 3 3 Air Volume Gap Air Fill Volume 10% (1µl) 10% (1µl) Single Injection Port Injection Rate Injection Volume Pre-Injection Delay 90% 90% 20% (2µl) 60% (6µl) 1 sec 1 sec Post Injection Delay 1 sec 1 sec Injection Start Output End End Rinse Rinse Volume Rinse Fill Rate Rinse Cycles 60% (6µl) 70% (7µl) 50% 50% 2 2 Rinse Dispense Rate 100% 100% Waste Depth 50% 50% Table 1: FLEX Autosampler Experimental Parameters

3 GC/MS Agilent 7890/5975 (1µl) Agilent 7890/5975 (5µl) Inlet Split/Splitless PTV Solvent Vent Inlet Temp. 280ºC 45ºC for 0.2 min, 200ºC/min to 125ºC for 0 min, 700ºC/min to 280ºC for 33.5min Inlet Head Pressure psi psi Split 20:1 NA Purge Flow to Split Vent NA 50ml/min at 1.5 min Vent Flow NA 100ml/min Vent pressure NA 0psi until 0.1min Cryo NA On at 50ºC Liner Column Oven Temp. Program Restek SKY liner, Splitless, Single Taper with Glass Wool, 4mm x 6.5 x 78.5 Rxi-5Sil MS 30m x 0.25mm I.D. x 0.25µm film thickness 45ºC hold for 4.0 min, ramp 10ºC/min to 320ºC hold for 2.0 min, 33.5 min run time TITAN XL SB Deactivated Liner with Glass Wool Rxi-5Sil MS 30m x 0.25mm I.D. x 0.25µm film thickness 45ºC hold for 4.0 min, ramp 10ºC/min to 320ºC hold for 2.0 min, 33.5 min run time Column Flow Rate 1.0ml/min. 1.0ml/min. Gas Helium Helium Total Flow 24ml/min 54.4ml/min. Source Temp. 230ºC 230ºC Quad Temp. 150ºC 150ºC MS Transfer Line Temp. 280ºC 280ºC Solvent Delay 5.0 min 5.0 min Acquisition Mode Scan Scan Scan Range m/z m/z Sampling Rate 3.12 scans/sec 3.12 scans/sec Table 2: GC/MS Experimental Parameters The PAH standards were acquired from Restek. Two different curves were run for the experiment. The 1µl injection curve standard range was from 0.5 ppm to 200ppm or 0.50 to 200ng on column. For the large volume, 5µl, injection the standard range was from 0.05ppm to 50ppm or 0.25 to 250ng on column. Next seven replicates of the low standard were run in order to determine MDLs. Finally, in order to ascertain the precision and accuracy of the injection techniques, seven replicates of the mid-point of each curve were run. Experimental results are listed in Table 3 and 4, while the chromatograms of the two injections are displayed in Figures 1 and 2.

4 1µl Injection, Standard Injection Compound Curve Ave. %RSD %Recovery MDL %RSD Curve RF 50ng 50ng Naphthalene Acenaphthalene Acenaphthene Fluorene Phenanthrene Anthracene Fluoranthene Pyrene Benz(a)anthracene Chrysene Benzo(b)fluoranthene Benzo(k)fluoranthene Benzo(a)pyrene Indeno(1,2,3-cd)pyrene Dibenz(a,h)anthracene Benzo(g,h,i)perylene Ave Table 3: Experimental Results Summary 1µl Injection 5µl Injection, Large Volume Injection Compound Curve Ave. %RSD %Recovery MDL %RSD Curve RF 50ng 50ng Naphthalene Acenaphthalene Acenaphthene Fluorene Phenanthrene Anthracene Fluoranthene Pyrene Benz(a)anthracene Chrysene Benzo(b)fluoranthene Benzo(k)fluoranthene Benzo(a)pyrene Indeno(1,2,3-cd)pyrene Dibenz(a,h)anthracene Benzo(g,h,i)perylene Ave Table 4: Experimental Results Summary 5µl Injection

5 Figure 1: 50ng on column, 1ul Injection Conclusions: Figure 1: 50ng on column, 5ul Injection The LVI technique, in conjunction with PTV, enabled the curve range to go down to 0.25ng on column without sacrificing compound response. The ability to inject more than one microliter of sample onto the GC column enhances the sensitivity of the system. When comparing the results of a standard injection to an LVI, the linearity, compound response, detection limits, and precision and accuracy were all comparable. Consequently, the use of LVI in conjunction with the FLEX Autosampling system is a viable option for PAH analysis. Furthermore, as detection limits become more and more stringent, the volume of the injection can increase in order to accommodate the new requirements. EST analytical and JSB shall not be liable for errors contained herein or for incidental or consequential damages in connection with this publication. Inform ation, descriptions, and specifications in this publication are subject to change without notice Headquarters JSB International Tramstraat CM Eindhoven T +31 (0) F +31 (0) With courtesy of Zoex Europe Tramstraat CM Eindhoven T +31 (0) F +31 (0) Sales and Service Netherlands Apolloweg 2B 8239 DA Lelystad T +31 (0) F +31 (0) Belgium Grensstraat 7 Box Diegem T +32 (0) F +32 (0) Germany Max-Planck-Strasse 4 D Kamp-Lintfort T +49 (0) F +49 (0) UK & Ireland Cedar Court, Grove Park Business Est. White Waltham, Maidenhead Berks, SL6 3LW T +44 (0) F +44 (0) info@go-jsb.com

Complete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges

Complete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges Complete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges Alexandria Pavkovich Jason Thomas Trent Sprenkle Outline Background EPA Method Requirements Background

More information

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System Application Note Authors Frank David Research Institute for Chromatography, Pres.

More information

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI) Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Brian Fitz and Bill Wilson Agilent Technologies, Inc. 285 Centerville Road Wilmington,

More information

Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut

Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut Application Note Environmental Author Ken Lynam Agilent Technologies, Inc. Abstract Specially designed self-tightening

More information

Application Note. Abstract. Authors. Environmental Analysis

Application Note. Abstract. Authors. Environmental Analysis High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID using the Agilent Low Thermal Mass (LTM II) System Application Note Environmental Analysis Authors Frank David and Karine Jacq Research

More information

Method-Specific Sample Preparation for EPH Analysis

Method-Specific Sample Preparation for EPH Analysis Method-Specific Sample Preparation for EPH Analysis Resprep SPE Cartridges for Complete Extractable Petroleum Hydrocarbon Fractionation Full separation between fractions. Guaranteed low background level.

More information

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Application Note Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Authors Kelly Beard and James McCurry Agilent Technologies, Inc. Abstract An Agilent

More information

Agilent Multimode Inlet for Gas Chromatography

Agilent Multimode Inlet for Gas Chromatography Agilent Multimode Inlet for Gas Chromatography Technical Note Agilent Multimode Inlet for the 7890A GC Designed to give you ease of use and maximum flexibility, the Agilent Multimode Inlet does everything

More information

CONSTITUTION OF COAL TAR PITCH AND ITS EFFECT ON PROPERTIES

CONSTITUTION OF COAL TAR PITCH AND ITS EFFECT ON PROPERTIES CONSTITUTION OF COAL TAR PITCH AND ITS EFFECT ON PROPERTIES H. K. Mayer 1 and I. C. Lewis 2 1 GrafTech International Ltd., 12900 Snow Road, Parma, OH 44130 2 Consultant Introduction Corresponding author

More information

Determination of PAH Compounds from Aqueous Samples Using a Non-Halogenated Extraction Solvent and Atlantic C18 Disks

Determination of PAH Compounds from Aqueous Samples Using a Non-Halogenated Extraction Solvent and Atlantic C18 Disks Determination of PAH Compounds from Aqueous Samples Using a Non-Halogenated Extraction Jim Fenster, Kevin Dinnean,, David Gallagher, Michael Ebitson, Horizon Technology, Inc., Salem, NH Introduction Polycyclic

More information

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI) Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Bill Wilson and Chin-Kai Meng Agilent Technologies, Inc. 2850 Centerville Road Wilmington,

More information

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS FLASH VAPORISATION INJECTION Split Splitless On-Column COOL INJECTION Large Volume Injection (LVI) On-Column On-Column-SVE (with solvent vapour

More information

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D759 Application Petrochemical Author James D. McCurry Agilent Technologies 285 Centerville

More information

Method Development for Capillary GC Systems. Slide 1

Method Development for Capillary GC Systems. Slide 1 Method Development for Capillary GC Systems Slide 1 AREAS TO OPTIMIZE Injector Carrier gas Column temperature Slide 2 COMMON INJECTOR MODES Vaporization Injection Modes Megabore Direct Split Splitless

More information

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel Application Note Energy and Chemicals Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel Applying the Energy Institute Method IP 8 with an Agilent J&W

More information

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application Petroleum Authors ChunXiao Wang Agilent Technologies (Shanghai) Co.,Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai

More information

Emissions Characteristics of Volatile and Semi-volatile Organic Compounds and Polynuclear Aromatic Hydrocarbons (PAHs) from Hot and Warm Mix Asphalts

Emissions Characteristics of Volatile and Semi-volatile Organic Compounds and Polynuclear Aromatic Hydrocarbons (PAHs) from Hot and Warm Mix Asphalts 1/10 Emissions Characteristics of Volatile and Semi-volatile Organic Compounds and Polynuclear Aromatic Hydrocarbons (PAHs) from Hot and Warm Mix Asphalts Frank Farshidi, University of California, Davis,

More information

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585 GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc. 850 Centerville

More information

New Integrated GC Guard Columns for Built-In Protection

New Integrated GC Guard Columns for Built-In Protection New Integrated GC Guard Columns for Built-In Protection GUARDIAN: Quick and Easy to Set-up With Guardian you get your column plus 5 or 10 meters of additional integrated deactivated tubing. However, it

More information

Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter

Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter Cd The Chrom Doctor Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands. In Part 1 of this series we focused

More information

State of Texas UST Monitoring visit us online at The latest TNRCC 1005/1006 UST methods used by Texas.

State of Texas UST Monitoring visit us online at  The latest TNRCC 1005/1006 UST methods used by Texas. At-a-Glance Product Information from Restek State of Texas UST Monitoring To help laboratories comply with and use these analytical procedures, Restek has been active in following the state guidance. Based

More information

TEKLAB, INC ANALYTICAL QC SUMMARY REPORT. CLIENT: Kelron Environmental Work Order: Project: CHMGP/ Page 1 of 6.

TEKLAB, INC ANALYTICAL QC SUMMARY REPORT. CLIENT: Kelron Environmental Work Order: Project: CHMGP/ Page 1 of 6. TEKLAB, INC CLIENT: Kelron Environmental Date: 27-Dec-6 Sample ID: MB-3498 SampType: MBLK SeqNo: 145961 Acenaphthene ND.1 Acenaphthylene ND.1 Anthracene ND.1 Benzo(a)anthracene ND.1 Benzo(a)pyrene ND.1

More information

Using the PSD for Backflushing on the Agilent 8890 GC System

Using the PSD for Backflushing on the Agilent 8890 GC System Application Note Petrochemicals Using the PSD for Backflushing on the Agilent 889 GC System Author Brian Fitz Agilent Technologies, Inc. Wilmington, DE, USA. Abstract An Agilent 889 series GC equipped

More information

GC Best Practices & Troubleshooting. Group/Presentation Title Agilent Restricted Month ##, 200X

GC Best Practices & Troubleshooting. Group/Presentation Title Agilent Restricted Month ##, 200X GC Best Practices & Troubleshooting Group/Presentation Title Agilent Restricted Month ##, 200X Troubleshooting Tips 1. Isolate the problem. (Blank Runs, Inject Un-retained Compound, Know what it is not)

More information

From Helium to Hydrogen: GC-MS Case Study on SVOCs in Water

From Helium to Hydrogen: GC-MS Case Study on SVOCs in Water From Helium to Hydrogen: GC-MS Case Study on SVOCs in Water Jessie Butler Alexander N. Semyonov December 13, 2012 Helium? It s gone! Not only at Disneyland Resort 2 Helium shortage grounds Mickey at Tokyo

More information

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector Detection of Sulfur Compounds in Natural Gas According to ASTM D554 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector Application Note Author Rebecca Veeneman Abstract Sulfur compounds in natural

More information

Fausto Munari e Andrea Cadoppi ThermoFisher - Italy

Fausto Munari e Andrea Cadoppi ThermoFisher - Italy The world leader in serving science Ultra Fast GC Determination of Total Hydrocarbons (C7-C40) and BTEX in Water and Soils through Direct Resistively Heated capillary columns and Robotic Autosampler. Fausto

More information

GC Inlets. An Overview. Simon Jones GC Applications Engineer

GC Inlets. An Overview. Simon Jones GC Applications Engineer GC Inlets An Overview Simon Jones GC Applications Engineer Types of Inlets Purged Packed Split / Splitless Cool On Column Programmable Temperature Vaporization Volatiles Interface Multi Mode Inlet Where

More information

The next level of True Blue Performance

The next level of True Blue Performance Restek Supplies & Accessories The next level of True Blue Performance Topaz GC inlet liners feature revolutionary technology and inertness to deliver the next level of: Deactivation Reproducibility Productivity

More information

Applicant: GUANGDONG XIN YU TECHNOLOGY INDUSTRIAL Date: Mar 21, 2018 CO.,LTD. LAI MEI INDUSTRIAL ZONE, CHENGHAI DISTRICT, SHAN TOU, GUANGDONG, CHINA

Applicant: GUANGDONG XIN YU TECHNOLOGY INDUSTRIAL Date: Mar 21, 2018 CO.,LTD. LAI MEI INDUSTRIAL ZONE, CHENGHAI DISTRICT, SHAN TOU, GUANGDONG, CHINA Applicant: GUANGDONG XIN YU TECHNOLOGY INDUSTRIAL Date: Mar, 08 CO.,LTD. LAI MEI INDUSTRIAL ZONE, CHENGHAI DISTRICT, SHAN TOU, GUANGDONG, CHINA Attn: MS LIN Sample Description: Thirteen () styles of submitted

More information

Technical Procedure for Gas Chromatography (GC-FID)

Technical Procedure for Gas Chromatography (GC-FID) Technical Procedure for Gas Chromatography (GC-FID) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph (GC- FID). 2.0 Scope This procedure applies to all

More information

Agilent 7693A Automated Liquid Sampler

Agilent 7693A Automated Liquid Sampler Agilent 7693A Automated Liquid Sampler Specifications Overview The Agilent 7693A is a state-of-the-art sample handling and injection system that provides the highest levels of precision and reliability

More information

Agilent Multimode Inlet

Agilent Multimode Inlet Agilent Multimode Inlet Large Volume Injection Tutorial Agilent Technologies Notices Agilent Technologies, Inc. 2009 No part of this manual may be reproduced in any form or by any means (including electronic

More information

EPA Method 625: Determination of Bases, Neutrals and Acids by Solid Phase Extraction and GC/MS Detection

EPA Method 625: Determination of Bases, Neutrals and Acids by Solid Phase Extraction and GC/MS Detection EPA Method 625: Determination of Bases, Neutrals and Acids by Solid Phase Extraction and GC/MS Detection UCT Part Numbers: EC82701M15-1000 mg 8270 sorbent/15 ml cartridge or EC82702M15-2000 mg 8270 sorbent/15

More information

DANI Transformer Oil Gas Analyzer

DANI Transformer Oil Gas Analyzer DANI Transformer Oil Gas Analyzer APPLICATION NOTE - AN169 Introduction Transformers are electrical devices used for energy transfer by electromagnetic induction between two or more circuits. Large oil

More information

Analysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID

Analysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID Analysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID Introduction P Gorst-Allman (LECO Africa Pty. Ltd) and B-J de Vos (NMISA). The analysis

More information

6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR

6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR 6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR A15836 This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use

More information

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS) Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph-mass spectrometer (GC-MS). 2.0 Scope

More information

Agilent 7693A Automated Liquid Sampler

Agilent 7693A Automated Liquid Sampler IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Agilent 7693A Automated Liquid Sampler

More information

SPECIFICATION. Bench top triple quadrupole Mass Spectrometry System coupled to a gas performance Chromatography (GC-MS/MS)

SPECIFICATION. Bench top triple quadrupole Mass Spectrometry System coupled to a gas performance Chromatography (GC-MS/MS) SPECIFICATION Bench top triple quadrupole Mass performance Chromatography (GC-MS/MS) 1. Scope These specifications describe the requirements for a bench top triple quadrupole mass spectrometry system coupled

More information

GC Method Compliance and Large Valve Oven Application

GC Method Compliance and Large Valve Oven Application GC Method Compliance and Large Valve Oven Application Solution Focused Kelly Beard June 20, 2016 Compliant Methods Compliant Methods What are our options? Analyzers and Analyzer Kits High Volume Well Known

More information

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract. Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers Application Note Abstract Versa With the rising prices of fossil fuels, more emphasis is being put on renewable resources

More information

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Application Note Petrochemicas High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Author James D. McCurry, Ph.D. Agilent Technologies, Inc. Abstract An Agilent

More information

Jaap de Zeeuw Restek Middelburg, The Netherlands. Copyrights: Restek Corporation

Jaap de Zeeuw Restek Middelburg, The Netherlands. Copyrights: Restek Corporation Liner Selection in GC Jaap de Zeeuw Restek Middelburg, The Netherlands What is a liner? A liner is an inert glass tube that is positioned inside the injection port of the GC The sample solution is introduced

More information

Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems

Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems Application Note Hydrocarbons Authors Michael Woodman Agilent Technologies, Inc. Chemical Analysis Solutions 2850

More information

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System Jim McCurry Roger Firor Agilent Technologies Wilmington, DE Page 1 Scope of the Analyzer Designed for ASTM D7798 For

More information

Characterization of Tiki Torch Fuels

Characterization of Tiki Torch Fuels 1940 N. Stark Road Midland, MI 48642 USA Phone: 855-IA-SOLVE (855-427-6583) Fax: (989) 486-9429 www.impactanalytical.com Customer: Philip Tyson Report Number*: R140075 Company: The Coconut Group Date Submitted:

More information

Analysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES

Analysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES Analysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES Application note Petrochemical Author Neli Drvodelic Agilent Technologies Melbourne, Australia Introduction

More information

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC Fast GC SPEEDY GC Slide 4 Optimization Goals Primary: Minimize time for the separation of a given number of peaks Secondary: Maximize the number of peaks separated by a given column Slide 5 HOW CAN THESE

More information

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC Technical Overview Trace gas analysis is a challenge in today s world. The ability to analyze lower component

More information

Liquid Injection Techniques in GC and GC-MS

Liquid Injection Techniques in GC and GC-MS Liquid Injection Techniques in GC and GC-MS Inge De Dobbeleer Regional Marketing Manager GC and GC-MS, EMEA Thermo Fisher Scientific, Breda/The Netherlands The world leader in serving science Content Hot

More information

Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides

Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides Application Note Food Testing & Agriculture Authors Limian Zhao and Chin-Kai Meng Agilent Technologies, Inc. 85 Centerville

More information

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis Page 1 Alternate Carrier Gas Considerations and Faster GC Analysis Faster GC Total Analytical Cycle Times A Variety of Approaches Pre-Run ALS Set-Up Chromatographic Run Post-Run Bake-Out Post-Run Cool-Down

More information

Method Detection Limits for EPA Method 8015 Diesel Range Organics using Fully Automated Extraction and Concentration

Method Detection Limits for EPA Method 8015 Diesel Range Organics using Fully Automated Extraction and Concentration Method Detection Limits for EPA Method 8015 Diesel Range Organics using Fully Automated Extraction and Concentration Michael Ebitson, David Gallagher, Horizon Technology, Inc. Key Words EPA Method 8015,

More information

State of Alaska UST Monitoring

State of Alaska UST Monitoring Contact Information At-a-Glance Product Information from Restek Alaska Department of Environmental Conservation 410 Willoughby Avenue Juneau, AK 99801-1795 Phone: (907) 465-5203 Fax: (907) 465 5218 www.state.ak.us/dec/deh/laboratories/ust.htm

More information

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector Application Note Hydrocarbon Processing Authors ChunXiao Wang Agilent Technologies

More information

Test Report. Final conclusion: PASS SHA18. Gohandmade Dsign International A/S HANNE Madevej Aabenraa DENMARK

Test Report. Final conclusion: PASS SHA18. Gohandmade Dsign International A/S HANNE Madevej Aabenraa DENMARK Gohandmade Dsign International A/S HANNE Madevej 11-13 6200 Aabenraa DENMARK Date of order: 14.06.18 Contact: Penny Guo Direct dial: +86(21) 60725688 x 5659 Email: penny.guo@hansecontrol.com Test Report

More information

PAH and Nitro-PAH emissions from GDI vehicles

PAH and Nitro-PAH emissions from GDI vehicles 19 th ETH-Conference on Combustion Generated Nanoparticles June 28 th July 1 st, 2015 ETH Zürich, Switzerland PAH and Nitro-PAH emissions from GDI vehicles Dr. Maria Muñoz Fernandez Coauthors: Dr. Norbert

More information

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis Introduction Petroleum & Petrochemical Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis By Katarina Oden, Barry Burger, and Amanda Rigdon Crude oil consists of thousands of different

More information

Online sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve

Online sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve Online sample cleanup on the Agilent 129 Infinity LC using a built in 2-position/6-port valve Analysis of Sudan red compounds in paprika powder Application Note Food Authors Angelika Gratzfeld-Huesgen

More information

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector Authors: The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector Introduction Specification of the hydrocarbon composition of LPG is required as traces

More information

Application. Gas Chromatography June 1995

Application. Gas Chromatography June 1995 Determining Oxygenates in Gasoline: ASTM Method D Application Gas Chromatography June 99 Authors Michael J. Szelewski Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-60 USA Matthew S. Klee

More information

5 PHASE GC COLUMNS DISCOVER NEW OPTIONS. Phenomenex I Web:

5 PHASE GC COLUMNS DISCOVER NEW OPTIONS. Phenomenex I Web: 5 PHASE GC COLUMNS DISCOVER NEW OPTIONS ZEBRON 5 PHASE GC COLUMNS Choosing the Correct 5 Phase Column Equivalent to USP Phase G27 Low bleed (MS Certified) especially suited to high sensitivity work using

More information

Supporting Information for. Diesel Particle Filter and Fuel Effects on Heavy-Duty Diesel Engine Emissions

Supporting Information for. Diesel Particle Filter and Fuel Effects on Heavy-Duty Diesel Engine Emissions Supporting Information for Diesel Particle Filter and Fuel Effects on Heavy-Duty Diesel Engine Emissions Matthew Ratcliff, A. John Dane, Aaron Williams, John Ireland, Jon Luecke, Robert L. McCormick, and

More information

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc.

More information

Agilent 6850 Series II Network GC System G2630A Performance Specifications

Agilent 6850 Series II Network GC System G2630A Performance Specifications IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Agilent 6850 Series II Network GC

More information

Set Your Sights on Superior Performance

Set Your Sights on Superior Performance Restek Sample Handling Set Your Sights on Superior Performance Restek PAL SPME Arrow Rugged stainless steel construction ensures longer lifetimes. Faster extraction means higher sample throughput. Better

More information

High-Temperature Simulated Distillation System Based on the 6890N GC Application

High-Temperature Simulated Distillation System Based on the 6890N GC Application High-Temperature Simulated Distillation System Based on the 6890N GC Application Petroleum Authors ChunXiao Wang Agilent Technologies (Shanghai) Co., Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai

More information

Techniques for Improving the Reproducibility of GC Analysis

Techniques for Improving the Reproducibility of GC Analysis Techniques for Improving the Reproducibility of GC Analysis Primary Areas of Concern Sample Auto-Injector Inlet Column Detector Sample Extract Handling and Care It is critical that the sample extract be

More information

Set Your Sights on Superior Performance

Set Your Sights on Superior Performance Restek Sample Handling Set Your Sights on Superior Performance Rugged stainless steel construction ensures longer lifetimes. Faster extraction means higher sample throughput. Better sensitivity allows

More information

Analysis of Petroleum Fractions by ASTM D2887

Analysis of Petroleum Fractions by ASTM D2887 Analysis of Petroleum Fractions by ASTM D2887 Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 2582 Key Words Simulated distillation, D2887, TRACE TR-SimDist Abstract ASTM

More information

GC Injector Liners. Bruker Varian Agilent (HP) Perkin Elmer Finnigan. Phone: Page 4

GC Injector Liners. Bruker Varian Agilent (HP) Perkin Elmer Finnigan.   Phone: Page 4 Bruker Varian Agilent (HP) Perkin Elmer Finnigan www.excelcomponent.com sales@excelcomponent.com Phone: 909-229-6363 Page 4 Deactivated Split/Splitless Injector We offer a full line of Split/Splitless

More information

Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS)

Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS) Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the pyrolysis-gas chromatograph-mass spectrometer

More information

TECHNICAL INFORMATION

TECHNICAL INFORMATION PRO-gran: TECHNICAL INFORMATION The ultimate in pitch performance PRO-gran: THE ULTIMATE PITCH INFILL PRO-gran has literally changed the game in high-performance infill for artificial sports pitches. A

More information

Agilent 9000 Intuvo GC Applications. Matthew Giardina, Ph.D. James McCurry, Ph.D. Rebecca Veeneman, Ph. D.

Agilent 9000 Intuvo GC Applications. Matthew Giardina, Ph.D. James McCurry, Ph.D. Rebecca Veeneman, Ph. D. Agilent 9000 Intuvo GC Applications Matthew Giardina, Ph.D. James McCurry, Ph.D. Rebecca Veeneman, Ph. D. Summary of Intuvo Applications by Industry Environmental EPA Method 8270 Semivolatile Organic Compounds

More information

Sulfur Detection at ppb Levels in Light Hydrocarbon Streams

Sulfur Detection at ppb Levels in Light Hydrocarbon Streams Sulfur Detection at ppb Levels in Light Hydrocarbon Streams Based on a New Super Permeable PLOT Column Agilent Select Low Sulfur Johan Kuipers Channel Training Specialist Oct 12 th, 2010 1 October 18,

More information

Quantification and Characterization of Sulfur in Low-Sulfur Reformulated Gasolines by GC-ICP-MS Application

Quantification and Characterization of Sulfur in Low-Sulfur Reformulated Gasolines by GC-ICP-MS Application Quantification and Characterization of Sulfur in Low-Sulfur Reformulated Gasolines by GC-ICP-MS Application Authors Steven M. Wilbur and Emmett Soffey Agilent Technologies 338 146th Place SE Bellevue,

More information

Application Note. Authors. Abstract. Energy & Chemicals

Application Note. Authors. Abstract. Energy & Chemicals Determination of Aromatic Content in Diesel Fuel According to ASTM D5186 Enhancing the Agilent 126 Infi nity Analytical SFC System with a Flame Ionization Detector Application Note Energy & Chemicals Authors

More information

Large Volume Injection (LVI) for GC:

Large Volume Injection (LVI) for GC: Large Volume Injection (LVI) for GC: Do s; Don ts; Why s; Won ts Robert G. Brown NEMC August 4-8, 2014 Washington, D.C www.lancasterlabsenv..com Some (Recent) History October/November, 2003 Paolo Magni

More information

Operation and Applications of Differential Flow Modulation

Operation and Applications of Differential Flow Modulation Operation and Applications of Differential Flow Modulation H2 Collection channel Column 1 H2 Column 2 FID Roger L Firor, Ph.D. Agilent Technologies Chemical Analysis Group Wilmington, DE USA Flow Modulator

More information

AppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS

AppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS AppNote 6/26 Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) Andreas Hoffmann GERSTEL GmbH & Co.KG, Eberhard-Gerstel-Platz 1,

More information

Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas

Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas Liliana Krotz, Walter Galotta, and Guido Giazzi Thermo Fisher Scientific, Milan, Italy Overview

More information

Setting up SilFlow for BackFlush in your GC

Setting up SilFlow for BackFlush in your GC Setting up SilFlow for BackFlush in your GC What is backflush and why use it? The BackFlush system eliminates the need to bake heavy sample fractions off the capillary column. Oils, tars and other semivolatile

More information

One-Minute Essential Oils Characterization by Gas Chromatography through Nanovolume Injection

One-Minute Essential Oils Characterization by Gas Chromatography through Nanovolume Injection Application Note: 10100 One-Minute Essential Oils Characterization by Gas Chromatography through Nanovolume Injection Thermo Fisher Scientific Inc., Milan, Italy Key Words 1-minute Essential Oils Analysis

More information

Chemical Laboratory - Kao., Test Report No. KE/205/320 Date : Mar 7, 205 Page 2 of 6 Summary of Test : (Detail test results on next page) Test Paramet

Chemical Laboratory - Kao., Test Report No. KE/205/320 Date : Mar 7, 205 Page 2 of 6 Summary of Test : (Detail test results on next page) Test Paramet Chemical Laboratory - Kao., Test Report No. KE/205/320 Date : Mar 7, 205 Page of 6 TAIXIANG RUBBER (SHEN ZHEN) CO., LTD. LISONG LANG INDUSTRIAL AREA GONG MING TOWN,SHENZHEN CITY,CHINA The following sample(s)

More information

Agilent and ASTM. Update on Recent Activities. Page 1

Agilent and ASTM. Update on Recent Activities. Page 1 Agilent and ASTM Update on Recent Activities Page 1 ASTM Committees for Refining, Fuels, Petroelum Products ASTM International D2 Petroleum products & lubricants D3 Gas fuels (natural gas) gas fuels D16

More information

Liquefied Gas Injector. Solution for the Sampling and Analysis of Liquefied Gases

Liquefied Gas Injector. Solution for the Sampling and Analysis of Liquefied Gases Liquefied Gas Injector Solution for the Sampling and Analysis of Liquefied Gases Safe and Representative Sampling of Liquefied Gases The analysis of impurities and contaminants in liquefied gases is an

More information

Rapid Determination of 54 Kinds of Pesticide Residues in Vegetables by PTV-GC-MS/MS

Rapid Determination of 54 Kinds of Pesticide Residues in Vegetables by PTV-GC-MS/MS PO-CON1517E Rapid Determination of 54 Kinds of Pesticide Residues in Vegetables by PTV-GC-MS/MS Pittcon 15 22-11P Shiheng Luo, Feifei Tian, Xizhi Wang, Guixiang Yang, Taohong Huang, Shin-ichi Kawano, Yuki

More information

AppNote 1/2010 KEYWORDS ABSTRACT. Biodiesel, Automation, ASTM D

AppNote 1/2010 KEYWORDS ABSTRACT. Biodiesel, Automation, ASTM D AppNote 1/2010 Full Automation of ASTM Method D6584-07 Standard Test Method for the Determination of Free and Total Glycerin in B-100 Biodiesel Methyl Esters by Gas Chromatography using a GERSTEL Dual

More information

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels James D. McCurry Senior Scientist Agilent Technologies Wilmington, DE USA Page 1 Application Summary There is a need to measure

More information

C2, C3, C4 Monomer Analysis

C2, C3, C4 Monomer Analysis C2, C3, C4 Monomer Analysis Malgorzata Sierocinska Agilent Technologies Waldbronn Page 1 Why Analyze Monomers? To Insure Consistent Production of High Quality Polymer Protect against food contamination

More information

Heavy Fuel Oil 6303 (2002)

Heavy Fuel Oil 6303 (2002) Origin Imperial Oil Ltd., Nova Scotia, Canada (2002) Synonyms: Bunker C, Land Bunker API Gravity 11.47 (calc) Equation for Predicting Evaporation %Ev = ( -0.16 + 0.013 T) t 1/2 Where: %Ev = weight percent

More information

ANALYSIS OF GASOLINE RANGE HYDROCARBONS ON BP1-PONA

ANALYSIS OF GASOLINE RANGE HYDROCARBONS ON BP1-PONA PET 01 - PETROLEUM ANALYSIS OF GASOLINE RANGE HYDROCARBONS ON BP1-PONA GASOLINE RANGE HYDROCARBONS Column Part No.: 054950 BP1, PONA 50 m x 0.15 mm ID Initial Temp.: 30 C, 5 min hold Rate 1: 2 C/min Temp.

More information

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors Model 8610C Gas Chromatograph Mounts up to Six Detectors and Five Injectors Ambient to 400 C Temperature Programmable Column Ovens Dimensions: 19 wide x 13.5 high x 14.5 deep Implement virtually any EPA

More information

PRODUCT SPECIFICATION. Product Specification Crystal 9000 Gas Chromatograph. Chromatec Crystal Laboratory Gas Chromatography System

PRODUCT SPECIFICATION. Product Specification Crystal 9000 Gas Chromatograph. Chromatec Crystal Laboratory Gas Chromatography System Chromatec Crystal 9000 Laboratory Gas Chromatography System GC Chromatec-Crystal 9000 Crystal 9000 is a leading gas chromatograph in Chromatec GC product line. Crystal 9000 has highest performances and

More information

Restek Corporation 110 Benner Circle, Bellefonte, Pa Barry L. Burger, Neil Johansen, Valerie Gamble, Donald Rhoades

Restek Corporation 110 Benner Circle, Bellefonte, Pa Barry L. Burger, Neil Johansen, Valerie Gamble, Donald Rhoades High Speed PONA Analysis For Detailed Hydrocarbon Analysis Extended (DHAX) Using Hydrogen Carrier Gas For The Determination of Individual Components In Spark Ignition Fuels Barry L. Burger, Neil Johansen,

More information

URB '-carbamoylbiphenyl-3-yl cyclohexylcarbamate. DEA Reference Material Collection. Form Chemical Formula Molecular Weight Melting Point ( o C)

URB '-carbamoylbiphenyl-3-yl cyclohexylcarbamate. DEA Reference Material Collection. Form Chemical Formula Molecular Weight Melting Point ( o C) O NH O NH O. GENERAL INFORMATION IUPAC Name: 3'-carbamoylbiphenyl-3-yl cyclohexylcarbamate CAS#: 56-08-6 Synonyms: Source: Appearance: UV max (nm): KDS-03 DEA Reference Material Collection White powder

More information

The Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over

The Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over The Agilent 12 Series high performance autosampler SL: Area precision, injection volume linearity, imum accessible volume, carry-over Technical Note 8 6 4 2.5 1 Introduction A main performance criterion

More information

DETECTORS Photo Ionization Detector - PID. Overview

DETECTORS Photo Ionization Detector - PID. Overview Overview DETECTORS The Photo Ionization Detector (PID) responds to all molecules whose ionization potential is below 10.6eV, including aromatics and molecules with carbon double bonds. The PID is nondestructive,

More information

Characterization and Source Identification of an Unknown Spilled Oil Using Fingerprinting Techniques by GC MS and GC FID

Characterization and Source Identification of an Unknown Spilled Oil Using Fingerprinting Techniques by GC MS and GC FID 15 LCGC VOLUME 1 NUMBER 1 OCTOBER 2 www.chromatographyonline.com Characterization and Source Identification of an Unknown Spilled Oil Using Fingerprinting Techniques by GC MS and GC FID This article describes

More information