HIGH PRESSURE SWITCHING VALVES
|
|
- Gwenda James
- 6 years ago
- Views:
Transcription
1 HIGH PRESSURE SWITCHING VALVES Rheodyne offers high pressure switching valves to simplify procedures and improve the speed, resolution, and sensitivity of HPLC analysis. The switching valves are available in stainless steel, titanium, and PEEK. The stainless steel and PEEK valves are available with 1/16- and 1/8-inch ports. There are two major areas of application: column switching and column selection. Column Selection These valves substitute one column for another without the need to manually disconnect the plumbing. This makes it easy to dedicate a separate column to each analysis. Dedicated columns eliminate equilibration delays, reduce interferences, and prolong column life. The valve is simply turned to select the column desired for a particular analysis. The columns switched off-line are automatically sealed at both ends. The various models each have unique column selection capabilities, discussed on pages Column Switching These valves re-route mobile phase during the chromatographic run, with no change in the mode of separation (such as backflushing and fraction cutting), or with changes, using sequential separation with different columns and/or mobile phases (such as trace enrichment, sample clean-up, and coupled-column analysis of complex samples). The Model 7000 (and its titanium and PEEK versions) is the most generally useful valve for column switching. But other models are better in certain cases. Examples are illustrated on page 20. Table V. Specifications of Switching Valves Model 7000,7030,7040 (SS & Titanium) Passage Diameter.6 mm 7060 (SS).4 mm 7000L,7030L,7040L, 7060L(SS & Titanium) 1.0 mm 9010,9030,9060 (PK).4 mm 3000,3030,3060 (PK) 1.0 mm , , (SS) 1.0 mm Factory Set Pressure 3000 psi (207 bar) 3000psi (207 bar) 4000 psi (276 bar) SS = stainless steel PK = PEEK *Titanium valves have a maximum temperature of 50 C, due to the Tefzel rotor seal. Many Rheodyne switching valves can be easily converted from one model to another. Model 7040 is a 7000 with an external loop across ports 1 and 5. The Model 7030 is a 7000 with a different rotor seal. Most models can be field-retrofitted with a pneumatic actuator. Valves with factoryinstalled actuators are designated by the letter "P", e.g. Model 7000P. Many models are available in a standardbore and a large-bore "L-version". The large-bore and/or the 1/8-inch valves can be used when the pressure drop must be limited. See the graph on page 10. Max. Field Set Pressure 7000 psi (483 bar) 7000 psi (483 bar) 4000 psi (276 bar) Maximum Temp. *
2 Six-position Column Selection Valves Tandem Rheodyne offers manually operated selector valves of different sizes and materials. The models that accept 1/16-inch fittings are: 7060 (stainless steel), (titanium), and 9060 (PEEK). Those that accept 1/8-inch fittings are: 3060 (PEEK) and (stainless steel). You can select among six columns for different analytical methods on the same chromatograph. This has several advantages over manually changing columns: selection is immediate, fittings are not subjected to the wear of repeated tightening, and the valve seals the off-line columns at both ends and keeps them on standby for future use. The common (center) port of this valve connects to the injector. Turning the valve handle directs flow into one of up to six columns connected to the six peripheral ports. A second valve is placed at the column outlets to select the operating column effluent and direct it to the detector. It is useful to use the sixth port for a bypass/ flush-out tube. This permits rapid mobile phase chageover, without exposing any column to a mobile phase other than the one with which it is routinely used. This scheme is shown above. Note that two separtate valves are required. The valves accepting 1/16-inch fittings have internal passages of 0.4-mm I.D., with a total internal volume less than 2 µl. This has an insignificant effect on resolution, as discussed on page 19. The valves accepting 1/8-inch fittings and the large-bore "L-versions" are used in applications where excessive pressure drop must be avoided. See page 15 for specifications and page 18 for pressure drop information. These valves can be used for mobile phase selection by connecting the common (central) port to a pump inlet. See page 23 for economical alternatives using Type 50 Teflon valves. The valves have a spring-loaded detent mechanism which allows continuous rotation in either direction and ensures that the shaft "falls into" each position at the precise 60 spacing position Valve 7060P 6-position valve with 7060L 7060LP 6-position Large Bore Valve 6-position Large Bore Valve with Titanium Valve 9060 PEEK Valve 9060P PEEK Valve with 3060 PEEK (1/8-inch) Valve 3060P PEEK Valve with Stainless Steel (1/8-inch) Valve P Stainless Steel Valve with The Tandem consist of two six-position valves mounted on a common pneumatic actuator. The valve rotors are geared together so that they turn simultaneously. This product allows the connection of up to six columns, or five columns and a bypass/flush-out tube which allows solvent changeover to be made without flow through any column. The actuator is controlled by the Model 7164 Air Control Valve (supplied), but can also be controlled by a programmer via a solenoid air valve (not supplied, see page 22). The column in use is indicated by a pointer on the actuator shaft. A six-position electrical switch is mounted on the shaft, permitting remote indication of valve position. The tandem actuator requires compressed gas: minimum 80 psi (6 bar), maximum 100 psi (7 bar) Tandem 7066L Large Bore Tandem Titanium Tandem 9066 PEEK Tandem 3066 PEEK (1/8-inch) Tandem Stainless Steel (1/8-inch) Tandem
3 Column Switching Valves Two-position six-port switching valves are available in different sizes and materials. The models that accept 1/16-inch fittings are: 7000 (stainless steel), (titanium), and 9010 (PEEK). The models that accept 1/8-inch fittings are 3000 (PEEK) and (stainless steel). Turning the handle back and forth 60 switches the flow paths from the solid lines to the dashed lines illustrated above. These valves have many column switching uses, where the flow is redirected among columns during the chromatographic run. Applications include trace enrichment, Sample clean-up, and coupled-column analysis (see the diagrams on page 20). These valves are also useful for selecting between two columns, as illustrated above and in Fig. 8. Compare them with the three-way valves on page 18, in which each end of the off-line column is independently sealed, instead of connected together head-to-tail. If switching is done before all pressure has bled from the column, independent seals produce less shock to the column. Model 9010 PEEK injector (see page 7) is listed here because it can be converted to a switching valve functionally identical to Model 7000 by removing the loop Switching Valve 7000P Switching Valve with 7000L 7000LP Large Bore Valve Large Bore Valve with Titanium Valve L Large Bore Titanium Valve 9010 PEEK Injector 3000 PEEK Valve (1/8-inch) Stainless Steel Valve (1/8-inch) from pump Sample Injector to detector Column Selection Valve Figure 7. In this photograph the Model 7000 Switching Valve is connected to select between two columns, one 10 cm and the other 15 cm. Turning the valve handle selects one of the columns for the injector-to-detector flow path and places the other column off-line, sealing it and keeping it in standby condition. A Model 7125 Sample Injector is shown connected to the selection valve.
4 Three-way Valves Four-way Valves Two-position six-port valves with a double three-way switching pattern are available in different sizes and materials. The models that accept 1/16-inch fittings are: 7030 (stainless steel), (titanium), and 9030 (PEEK). The models that accept 1/8-inch are: 3030 (PEEK) and (stainless steel). Turning the handle back and forth 60 switches the flow paths from the solid lines to the dashed lines illustrated in the figure above. You can perform the column switching schemes shown in Figs. 11f and 11l way Valve 7030P 3-way Valve with 7030L 7030LP Large Bore 3-way Valve Large Bore 3-way Valve with way Titanium Valve way PEEK Valve 9030P 3-way PEEK Valve with way PEEK Valve (1/8-inch) way Stainless Steel Valve (1/8-inch) Model 7040 is a six-port two position valve with an external loop, which exchanges the flow pattern from ports (2 3 and 4 6) to (2 6 and 4 3). See Figs. 11c, 11j, and 11k. Model 9010 can be converted to a 4-way valve by changing the loop to ports 1 and 5. Model 3000 can be converted by adding an external loop to ports 1 and way Valve 7040P 4-way/Pneumatic 7040L Large Bore 4-way Valve 7040LP Large Bore 4-way/Pneumatic way Titanium Valve Pressure Drop (psi) Pressure Drop (Atm) Flow Rate (ml/min) Figure 8. Pressure drop vs. flow rate for Model 7000 and Model 7000L (large-bore) valves; water at 20 C. Experimental measurements: The flow channel is one stator inlet port, one rotor seal groove, one stator outlet port, and two connectiong tubes. Solid squares = (1-mm 7000L valve) + (two 1-mm x 5-cm tubes). Open squares = (0.6-mm 7000 valve) + (two 1-mm x 5-cm tubes). Cross mark = (0.6-mm 7000 valve) + (two 0.5-mm x 5-cm tubes). Solid lines are theoretical values for 10-cm long tubes of 1 mm and 0.5 mm I.D.
5 Effect of Valves and Tubing on Resolution Rheodyne column selecting and column switching valves cause no observable loss of resolution with analytical columns when small-bore connecting tubing is used. Since dispersion caused by tubing is proportional to the fourth power of diameter, large-bore tubing should be avoided. A size 0.25 mm (0.010 inch) is recommended. The chromatograms in Fig. 10, made using a typical analytical chromatograph, show these effects. Fig. A is the control (injector column detector); no valves are in the system. In Figs. B, C, and D, two Model 7060 column selection valves were placed side by side (injector valve #1 column valve#2 detector), as illustrated above. The injector and detector were connected to these valves by the same tubing used in the control. The extra tubing pieces required to connect the valves to the column were a 10-cm length for valve#1 column and a 35-cm length for column valve#2. The diameters of these tubes are indicated in the figure. column only valve with.007 inch valve with.012 inch valve with.020 inch Figure 9. This shows the loss of resolutuon caused by the addition of two Model 7060 column selection valves when using connection tubes of three different inside diameters. Conditions for all cases: 4.6-mm x 12.5-cm column, 5 µm C-18 packing, 50% acetonitrile in water, 2 ml/min, 21 C, 5 µl sample partial-filled into a Model 7125 injector, 10-cm x 0.18-mm (.007-inch) bore injector outlet tube (to column or valve), 20-cm x 0.18-mm bore detector inlet tube (from column or valve), low-dispersion 1-cm path UV detector cell, 0.2-sec detector time constant. See text for details.
6 20a. Model 7000 Two-column selection A single two-positon valve can select either column A or column B. The off-line column is sealed by connecting head to tail. Switching should be done when the flow is off and the pressure is at zero. 20b. Model 7060 Five-column selection Two six-postion valves, one at each end of the columns, select between 5 columns and bypass/flush-out line. Offline columns are independently sealed. Two valves required. 20c. Model 7040 Column backflushing Flow direction can be reversed using a four-way valve. In this diagram, peaks are analyzed in the normal way, then the valve is switched and highly retained peaks are backflushed to the detector. Compare 20d. 20d. Model 7000 Pre-column backflushing Sample is injected into series-connected pre-column and analytical column. After valve switch, analytical column flow continues in normal direction. Only the precolumn is backflushed, eluting highly retained peaks directly to detector. 20e. Model 7000 Sample clean-up (stripping) Protects analytical column from proteins by irreversibly sorbing then onto precolumn. Analytes are also initially sorbed, but, after valve switch, a solvent change or gradient backflushes analytes into analytical column. Compare 20f, 20h. 20f. Model 7030 Sample clean-up (bypass) Protects analytical column from proteins by bypassing analytical column. Analytes are sorbed from buffer onto reversedphase pre-column. After valve switch a solvent change or gradient elutes analytes into analytical column. 20g. Model 7000 Sample clean-up (bypass) Identical to 20f, but using Model Here analytical column is isolated by connectiing head to tail. In 20f ends are independently isolated. This shows subtle differences between using different models. 20h. Model 7000 Sample clean-up (stripping) Unretained analytes pass through precolumn to analytical column where they are separated. Highly-retained undesired compounds are washed off the precolumn after valve switch. Two pumps required. Compare to 20e, 20f. 20i. Model 7000 Sample enrichment Large volumes of sample can be injected into the pre-column, where trace components are concentrated. After valve switch, sample is backflushed off pre-column into analytical column by the second pump. 20j. Model 7040 Alternating enrichment Similar to 20i, but two pre-columns save time. One is backflushed into analytical column while other reequilibrates and concentrates sample. The two precolumns alternate in their duty cycle. 20k. Model 7040 Alternating regeneration Two four-way valves can interchange two components between two flow streams. Here column A is regenerating while column B is analyzing. The columns then interchange by switching the two valves. 20l. Model 7030 Stop-flow scanning A three-way valve seals the column inlet, abruptly stopping the flow in the detector so that the peak can be scanned. The pumped mobile phase is diverted to drain line, so the pump need not be stopped.
78 동방하이테크상사 T.(02) F.(02)
78 동방하이테크상사 http://www.allpump.co.kr T.(0)7-9 F.(0)7-9 00--7 Introduction to Introduction Rheodyne valves fit virtually any high-pressure flow control application. As you ll see from the table below and
More informationChromatography. Liquid. HPLC Accessories SupelPRO Valves SUPELCO
8 SupelPRO Valves Order:.800..00 Technical Service:.800.9.0 Web: www.sigma-aldrich.com/supelco SupelPRO Automated Fluidics Instruments Supelco s SupelPRO series are precision, electronically-controlled,
More informationAgilent G5632 Valve Kit Instructions
Agilent G5632 Valve Kit Instructions Agilent G5632 Valve Kit Instructions Contents Typical Applications 2 Alternating Column Regeneration (2 position/10 port valves only) 2 Dual column selection 3 Sample
More informationAgilent InfinityLab Quick Change Valves G4231A/C and G4232C/D
Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Instructions Agilent Valve Kit G4231A/C and G4232C/D - Instructions This technical note describes the installation and application of the Agilent
More informationAgilent Valve Kit G4231A/B/C and G4232B/C/D
Agilent Valve Kit G4231A/B/C and G4232B/C/D Instructions Agilent Valve Kit G4231A/B/C and G4232B/C/D - Instructions This Technical Note describes the installation of the Agilent G4231A/B/C and G4232B/C/D
More informationAgilent Valve Kit G4231A/B and G4232A/B/C
Agilent Valve Kit G4231A/B and G4232A/B/C Instructions Contents Typical Applications 2 Dual column selection (G4231A/B or G4232A/B/C) 2 Sample enrichment and sample cleanup (G4231A/B or G4232A/B/C) 2 Alternating
More informationHigh Pressure Valves. MX Series II Modules Manual Sample Injection Switching
High Pressure Valves MX Series II Modules Manual Sample Injection Introduction to High Pressure Valves HIGH Pressure Valves High Pressure Valves Introduction Rheodyne valves fit virtually any flow control
More informationTECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure
TECHNICAL REPORT Modifying Agilent 1100 HPLC Systems to Achieve UHPLC-like performance with HALO Fused-Core Columns Agilent 1100 HPLC system With a few modifications, an Agilent 1100 HPLC can produce UHPLC-like
More informationFITTINGS, LOOPS, AND VALVE ACCESSORIES
FITTINGS, LOOPS, AND VALVE ACCESSORIES Fittings Rheodyne offers a variety of fittings. They differ in material of construction, tubing size, and pressure rating. Materials/Biocompatibility. 316 stainless
More information[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS
MS Identification of Trace level Impurities from a Non-MS Compatible Mobile Phase Using ACQUITY UPLC System with 2D Technology by Heart-cutting and Online Sample Concentration Bronsky Gopinadh, Dilshad
More informationHigh Pressure Valves. MX Series II Modules Manual Sample Injection Switching
High Pressure Valves MX Series II Modules Manual Sample Injection Switching Introduction to High Pressure Valves HIGH PRESSURE VALVES High Pressure Valves Introduction Rheodyne valves fit virtually any
More informationS 1125 HPLC Pump System
Routine S 170 EXPERT HPLC HPLC System SYSTEMS 500 The Sykam ROUTINE HPLC System S 500 is intended for all routine analysis. The system configuration is highly variable and several upgrade options make
More informationÄKTA system Training Guide. Purifier 100 / 10. amersham bio science
ÄKTA system Training Guide Purifier 00 / 0 amersham bioscience amersham bioscience amersham bioscience amersham bio science ÄKTApurifier Buffers Injector Box Monitor Conductivity ph On-line Filter UV Detector
More informationValving Schemes for the BioLogic Duo-Flow Chromatography System
Valving Schemes for the iologic Duo-Flow Chromatography System table of contents pplication asic ion Valve Operation using,, and Purge Positions position is for filling a loop with sample or for running
More informationOnline sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve
Online sample cleanup on the Agilent 129 Infinity LC using a built in 2-position/6-port valve Analysis of Sudan red compounds in paprika powder Application Note Food Authors Angelika Gratzfeld-Huesgen
More informationNEW HPLC - SYSTEM LINE
NEW HPLC - SYSTEM LINE S c h a m b e c k S F D G m b H Drieschweg 13A D-53604 Bad Honnef/Germany Phone: +49 2224 9239-0 E-Mail: info@schambeck-sfd.com Internet: www.schambeck-sfd.com S 9425 HPLC solvent
More informationAgilent 1200 Series Valves
Agilent 1200 Series Valves User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or by any means (including electronic storage and
More informationQuickSplit TM Flow Splitters
Quick TM Flow ters Quick TM Flow ter Features: Fluid resistor technology eliminates tedious adjustments to capillary tubing for split ratio optimization ratios are stable and reproducible, and not affected
More informationInstallation of the 2 position/10 port valve into the 1100 series thermostatted column compartment
Agilent G1316A Thermostatted Column Compartment Technical Note Installation of the 2 position/10 port valve into the 1100 series thermostatted column compartment System Requirements 2 Performance Specifications
More informationVALVES NEW MANUAL INJECTOR RHEBUILD KITS SAMPLE LOOPS MICRO VALVES PAGE 132 PAGE 136 PAGE 139 PAGE
VALVES NEW MANUAL INJECTOR PAGE RHEBUILD KITS PAGE 6 SAMPLE LOOPS PAGE 9 MICRO VALVES PAGE 6 SEE STARRED PRODUCTS FOR TOP SELLERS! Valve Overview VALVES 5 Rotary Shear Valves Rheodyne Rotary Shear Valves
More informationApplication Note. General Use Inlet Solvent Filters. Inlet Solvent Filters. Filters and Degassers. New!
Inlet Solvent Filters Filters and Degassers General Use Inlet Solvent Filters Large Surface Areas disposable µm, 0 µm, and 0 µm Pore Sizes Available General Use and Prep Filters for Higher Flow Applications
More informationImpurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution
Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 129 Infinity II HDR-DAD Impurity Analyzer Solution Application ote Small Molecule Pharmaceuticals Author Sonja Schneider Agilent Technologies,
More informationVERITY Purification Systems Build the perfect system to fit your research
VERITY Purification Systems Build the perfect system to fit your research SPEC SHEET PURIFICATION VERITY PURIFICATION SYSTEMS WITH VERITY 3011 ISOCRATIC PUMP MAKE YOUR LAB LIFE EASIER From automated GPC
More informationGAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS
GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS FLASH VAPORISATION INJECTION Split Splitless On-Column COOL INJECTION Large Volume Injection (LVI) On-Column On-Column-SVE (with solvent vapour
More informationAgilent G4239C Valve Kit - Instructions
Agilent G4239C Valve Kit - Instructions Agilent G4239C Valve Kit - Instructions Technical Information about the Agilent Valve Kit G4239C. Contents Typical Applications 2 Multicolumn Selection 2 Method
More informationSetting up SilFlow for BackFlush in your GC
Setting up SilFlow for BackFlush in your GC What is backflush and why use it? The BackFlush system eliminates the need to bake heavy sample fractions off the capillary column. Oils, tars and other semivolatile
More informationInstrument Accessories & Parts, Tubing & Connections
Instrument Accessories & Parts, Tubing & Connections ASI Piston Seals for Agilent HPLC Systems Over time the piston seal in every HPLC system becomes worn and must be replaced. Our piston seals are constructed
More informationÄKTA system Training Guide
ÄKTA system Training Guide Purifier 00 / 0 amersham bioscience amersham bioscience amersham bioscience amersham bioscience ÄKTApurifier Buffers Injector Box Monitor Conductivity ph On-line Filter UV Detector
More information98 동방하이테크상사 T.(02) F.(02)
98 동방하이테크상사 http://www.allpump.co.kr T.(0)57-69 F.(0)57-69 00--7 troduction to Upchurch Scientific Valves Valves troduction Valves jection and Selection Valves for Pressures to,000 psi (69 bar) Switching
More informationGas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593
Application Note Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Authors Kelly Beard and James McCurry Agilent Technologies, Inc. Abstract An Agilent
More informationHPLC SYSTEM S 500 SERIES
HPLC SYSTEM S 500 SERIES The Sykam HPLC System S 500 Series is intended for all routine analysis as well as for the ambitious analyst. The system configuration is highly variable and several upgrade options
More informationFixed Flow Splitter *
Fixed Flow * *Patent Pending Analytical Scientific Instruments Fluid resistor technology eliminates tedious adjustments to capillary tubing to change split ratios. Easy to use interchangeable fluid resistors
More informationAgilent InfinityLab Quick Change Valve G4237A
Agilent InfinityLab Quick Change Valve G4237A Instructions Agilent G4237A Valve Kit Instructions This technical note describes the installation and application of the Agilent InfinityLab Quick Change 4
More informationHPLC Model 210D. UV detector lamp switch. Rheodyne 7725i injection valve. Prime/purge valve. Column. compartment. thumbscrews. Storage.
1. Overview The front panel houses the UV detector lamp switch, Rheodyne injection valve, prime/purge valve, pressure gauge, and the storage compartment. In the storage compartment are the LED display
More informationACQUITY UPLC I-Class System with 2D Technology
ACQUITY UPLC I-Class System with 2D Technology The Waters ACQUITY UPLC I-Class System with 2D Technology allows chemists to increase sensitivity and selectivity, eliminate unwanted interferences, characterize
More informationseal, needle seat, needle assembly, seat cap kit Includes: rotor seal, needle assembly, needle seat 1100 G kit 25271
Restek Replacement Parts for Agilent HPLC Systems Meet or exceed OEM performance. Agilent part # qty. cat.# price Preventive Maintenance Kit Includes: rotor seal, needle seat, needle assembly, seat cap
More informationPRODUCT MANUAL. CARBONATE REMOVAL DEVICE 200 CRD 200 (4 mm) (P/N ) CRD 200 (2 mm) (P/N )
CRD 200 Product Manual Page 1 of 18 PRODUCT MANUAL FOR CARBONATE REMOVAL DEVICE 200 CRD 200 (4 mm) (P/N 062983) CRD 200 (2 mm) (P/N 062986) Dionex Corporation, 2006 Document No. 065068 Revision 04 November
More informationChrom Tech your chromatography specialists. IDEX Injection Valves Manual. IDEX Injection Valves Actuated. 1/4-28 Flat-Bottom Injection Port Adapter
HPLC ACCESSORIES IDEX Injection Valves IDEX Injection Valves Manual IDEX MANUAL INJECTION VALVES DESCRIPTION POS. SENSING SWITCH MAX PSI FITTING HIGH PRESSURE, ANALYTICAL (accepts a 22 gauge needle) FITS
More informationHPLC System S 600 Series
HPLC System S 600 Series The Sykam HPLC System S 600 Series is intended for the ambitious analyst. The system configuration is highly variable and several upgrade options make this system suitable for
More informationThe Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over
The Agilent 12 Series high performance autosampler SL: Area precision, injection volume linearity, imum accessible volume, carry-over Technical Note 8 6 4 2.5 1 Introduction A main performance criterion
More informationUsing a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application
Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D759 Application Petrochemical Author James D. McCurry Agilent Technologies 285 Centerville
More informationAgilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve
Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve Technical Note This Technical Note describes the purpose, installation, and configuration and use of a diverter valve
More informationAgilent G4236A 2D-LC Valve Installation Guide
Agilent G4236A 2D-LC Valve Installation Guide Technical Note Contents Delivery Checklist 2 Installing Valve Heads 3 Removing the Transportation Lock and the Valve Dummy 3 Installing the Valve Head and
More informationFilters & Frits. IDEX Health & Science introduces Upchurch Scientific precolumn and inline very high pressure (VHP) filters on page 159
Using a scale to count Upchurch Scientific frits in Oak Harbor, WA. IDEX Health & Science introduces Upchurch Scientific precolumn and inline very high pressure (VHP) filters on page 159 156 Inlet Solvent
More information1110 Valves. Introduction
Introduction Specific part numbers are not given in this portion of the service manual. For all replacement part numbers, refer to the IPB portion of this document. This document does not provide instruction
More informationBioLogic DuoFlow Valving Guide
Protein Purification iologic DuoFlow Valving Guide Valving Schemes for the iologic DuoFlow Chromatography System Protein Purification iologic DuoFlow Valving Guide Introduction The iologic DuoFlow Chromatography
More informationAgilent InfinityLab LC Purification Solutions SELECTION GUIDE
gilent InfinityLab LC urification Solutions SELECTION GUIDE GILENT INFINITYLB LC URIFICTION SOLUTIONS EFFICIENT LC URIFICTION SOLUTIONS FOR NLYTICL TO RERTIVE SCLE gilent InfinityLab LC urification Solutions
More informationAdjustable Flow Splitter *
Adjustable Flow Splitter * *Patent Pending Analytical Scientific Instruments Adjustable metering valve enables precise direct control over split ratios. Fluid resisr technology eliminates tedious adjustments
More informationPage 1. Alternate Carrier Gas Considerations and Faster GC Analysis
Page 1 Alternate Carrier Gas Considerations and Faster GC Analysis Faster GC Total Analytical Cycle Times A Variety of Approaches Pre-Run ALS Set-Up Chromatographic Run Post-Run Bake-Out Post-Run Cool-Down
More informationIDEX Health & Science introduces Upchurch Scientific precolumn and inline very high pressure (VHP) filters on page 159
Filters & Frits Using a scale to count Upchurch Scientific frits in Oak Harbor, WA. IDEX Health & Science introduces Upchurch Scientific precolumn and inline very high pressure (VHP) filters on page 159
More informationTechnical Overview. Introduction
Performance characteristics of the Agilent 129 Infinity Binary Pump More resolution and speed for conventional, superficially porous and sub-2 µm column packing material Technical Overview Introduction
More informationMotor Valve MV-7/IMV-7
Motor Valve MV-7/IMV-7 INSTRUCTIONS 1. Introduction Motor Valves MV-7/IMV-7 are a 3-position, 7-port motorized valves designed to he used as automatic sample injection valves or system valves. The valves
More informationHyperShear TM HPLC and UHPLC Mixers
HyperShear TM HPLC and UHPLC s Reduce baseline noise, increase sensitivity and improve gradient accuracy Decrease mixing and delay volume without sacrificing mixing efficiency Increase reaction efficiency
More informationTechnical Offer. Antek Model 6200 Analyzer Determine Total Sulfur and Total Nitrogen in a single analyzer
Technical Offer Antek Model 6200 Analyzer 2015.4 Determine Total Sulfur and Total Nitrogen in a single analyzer Meet sulfur regulations and control your process with continuous sulfur and/or nitrogen measurements
More informationRefinery Gas. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.
Refinery Gas Analysis by Gas Chromatography Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Refinery Gas Analysis Reliability Placing refinery gas analyzers in the field for over
More informationDetection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector
Detection of Sulfur Compounds in Natural Gas According to ASTM D554 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector Application Note Author Rebecca Veeneman Abstract Sulfur compounds in natural
More informationAgilent Online-SPE Direct Inject Kit G4744A
Agilent Online-SPE Direct Inject Kit G4744A Installation Guide Agilent Online-SPE Direct Inject Kit G4744A - Installation Guide Contents Application 2 Delivery Checklist 4 Online SPE Direct Inject Kit
More informationLC20 CHROMATOGRAPHY ENCLOSURE OPERATOR S MANUAL Dionex Corporation
LC20 CHROMATOGRAPHY ENCLOSURE OPERATOR S MANUAL 1993 Dionex Corporation Document No. 034859 Revision 02 September 1993 1993 Dionex Corporation All rights reserved worldwide Printed in the United States
More informationHyperShear TM HPLC Mixers
HyperShear TM HPLC s HyperShear TM HPLC Features: Static or Dynamic s with highly efficient vortex sheer mixing Reduce baseline noise, increase sensitivity, and improve gradient accuracy Mixing volume
More informationHyperShear TM HPLC Mixers
HyperShear TM HPLC s HyperShear TM HPLC Features: Static or Dynamic s with highly efficient vortex shear mixing Reduce baseline noise, increase sensitivity, and improve gradient accuracy Mixing volume
More informationApplication. Gas Chromatography June 1995
Determining Oxygenates in Gasoline: ASTM Method D Application Gas Chromatography June 99 Authors Michael J. Szelewski Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-60 USA Matthew S. Klee
More informationAlliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE
Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE AN HPLC THAT WORKS AS HARD AS YOU DO Confidence in your results comes with high performance, day-to-day reliability, and overall system robustness.
More information2-D GC/MS and the Deans Switch: Theory and Applications. Fred Feyerherm, Agilent Technologies Inc. Houston, Tx
2-D GC/MS and the Deans Switch: Theory and Applications Fred Feyerherm, Agilent Technologies Inc. Houston, Tx Introduction Simplified 2-D GC can be used to accomplish the separation of specific analytes
More informationDionex CRD Carbonate Removal Device
Thermo Scientific Dionex CRD Carbonate Removal Device Product Manual P/N: June 2013 Part of Thermo Fisher Scientific Product Manual for Dionex CRD Carbonate Removal Device 180 CRD 180, 0.4 mm (P/N 079960)
More informationAgilent 1200 Series Manual Injector
Agilent 1200 Series Manual Injector User Manual 1200 Series MI User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2006, 2008 No part of this manual may be reproduced in any form or by
More informationDetermination of Sudan Dyes I IV in Curry Paste
Determination of Sudan Dyes I IV in Curry Paste Suparerk Tukkeeree and Jeffrey Rohrer 2 Thermo Fisher Scientific, Bangkok, Thailand; 2 Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 23 Key
More informationFittings SECTION 1 OF 4 SECTIONS ON FITTINGS
Fittings SECTION 1 OF 4 SECTIONS ON FITTINGS Deburring an Upchurch Scientific fitting in Oak Harbor, WA. The Very High Pressure (VHP) reusable fitting, VHP-320, is one of the exciting new products in the
More informationGlycoChrom TM HPLC Analyzer. Hardware Installation and System Test
GlycoChrom TM HPLC Analyzer Hardware Installation and System Test PIN 410-5004 For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-8 00-424-6723) ,, Table of Contents
More informationCHROMATOGRAPHY. SYKAM S 6000 Valve Series
CHROMATOGRAPHY SYKAM S 6000 Valve Series Version: March 2011 SYKAM S 6000 Valve Series Rotor Seals Sykam offers a variety of different Rotor Seals materials to satisfy any possible application. All materials
More informationThermo Scientific Dionex UltiMate 3000 Thermostatted Column Compartment
chromatography Thermo Scientific Dionex UltiMate 3000 Thermostatted Column Compartment Product Specifications Thermo Scientific Dionex UltiMate 3000 products are UHPLC compatible by design, establishing
More informationAgilent Dynamax Analytical HPLC Column Systems 4.6 mm id Data Sheet
Agilent Dynamax Analytical HPLC Column Systems 4.6 mm id Data Sheet 1. Description The Agilent Dynamax modular HPLC column system uses a unique technology, Dynamic, to maintain the structure of the packed
More informationPATROL UPLC Process Analysis System for Production
PATROL UPLC Process Analysis System for Production The PATROL UPLC Process Analysis System for Production is an integrated system solution designed and engineered to perform online and atline reaction
More informationHigh-throughput protein aggregate analysis of monoclonal antibodies using a novel dual-channel UHPLC instrument
APPLICATION NOTE 72598 High-throughput protein aggregate analysis of monoclonal antibodies using a novel dual-channel UHPLC instrument Authors Nicola McGillicuddy, 1 Amy Farrell, 1 Sara Carillo, 1 Martin
More information[ CARE AND USE MANUAL ] WATERS ADVANCED PURIFICATION COLUMNS I. INTRODUCTION CONTENTS I. INTRODUCTION II. USING THE COLUMN III. CARE AND MAINTENANCE
WATERS ADVANCED PURIFICATION COLUMNS I. INTRODUCTION Waters Advanced Purification (AP) series of glass columns are constructed of biocompatible glass and polymeric materials and can be easily used with
More informationAchieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)
Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Bill Wilson and Chin-Kai Meng Agilent Technologies, Inc. 2850 Centerville Road Wilmington,
More informationUpchurch High Pressure Fittings & Connectors
.010 thru-hole P-742 PEEK ZDV Union. 9 2.050 thru-hole P-760 PEEK ZDV Union P-704 PEEK ZDV Union. 9 2. 9 2 Fingertight Fittings ONE-PIECE FINGERTIGHT FITTINGS INTRODUCTION For Coned Ports For convenience
More informationHigh Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph
Application Note Petrochemicas High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Author James D. McCurry, Ph.D. Agilent Technologies, Inc. Abstract An Agilent
More informationReciPro Liquid Metering Pumps. Reproducible flow rates High pressure capabilities Corrosion resistant Remote control options
ReciPro Liquid Metering Pumps Reproducible flow rates High pressure capabilities Corrosion resistant Remote control options ReciPro HP Series Eldex metering pumps can be used wherever precise flow rate
More informationValve Box and Valve Accessories G4315A, G4316A, G4317A
Valve Box and Valve Accessories G4315A, G4316A, G4317A Installation Guide The G4315A, G4316A, and G4317A accessories consist of separate kits depending on the custom configuration ordered. These kits may
More informationHPLC Tubing and Fittings Hints & Tips for the Chromatographer
HPLC Tubing and Fittings Hints & Tips for the Chromatographer By William S. Letter (Chiralizer Services, L.L.C.) INTRODUCTION: Setting up a high pressure liquid chromatography (HPLC) system to run trouble-free
More informationCombustion ion chromatography with a Dionex Integrion HPIC System
TECHNICAL NOTE Combustion ion chromatography with a Dionex Integrion HPIC System Authors Daniel Khor,1 Jeff Rohrer,1 Adelon Austin,2 Mark Manahan2 1 Thermo Fisher Scientific, Sunnyvale, CA, USA; 2COSA
More informationUsing the PSD for Backflushing on the Agilent 8890 GC System
Application Note Petrochemicals Using the PSD for Backflushing on the Agilent 889 GC System Author Brian Fitz Agilent Technologies, Inc. Wilmington, DE, USA. Abstract An Agilent 889 series GC equipped
More informationApplication Note. Authors. Abstract. Energy & Chemicals
Determination of Aromatic Content in Diesel Fuel According to ASTM D5186 Enhancing the Agilent 126 Infi nity Analytical SFC System with a Flame Ionization Detector Application Note Energy & Chemicals Authors
More informationOne-Piece UPchurch Fingertight Fittings 1,2. ChromTRAC. TOP SELLER see starred products in price block. Fingertight Fittings.
Fingertight Fittings 4 One-Piece Fingertight Fittings Upchurch Scientific s original One-Piece Fingertight Fittings have 0-3 threads and are designed to be used with /6 OD tubing, except the M-645 (6-40,
More informationBoosting Trace Detection Performance with the Vanquish Diode Array Detector and High-Sensitivity LightPipe Flow Cell
Boosting Trace Detection Performance with the Vanquish Diode Array Detector and High-Sensitivity LightPipe Flow Cell Alexandra Manka and Holger Franz Thermo Fisher Scientific, Germering, Germany Technical
More informationAgilent 1290 Infinity Autosampler
Agilent 1290 Infinity Autosampler Installation of the Large Volume Injection Kit General Information This note describes how to install the large volume injection kit into an Agilent 1290 Infinity Autosampler.
More informationFast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC
Fast GC SPEEDY GC Slide 4 Optimization Goals Primary: Minimize time for the separation of a given number of peaks Secondary: Maximize the number of peaks separated by a given column Slide 5 HOW CAN THESE
More informationLiquid Chromatography
SCINTOMICS Liquid Chromatography Always the right system for your needs. SCINTOMICS: Always the right system for your needs Advanced Research Daily QC Routine and beyond High Performance analytical/semi-preparative
More informationHigh Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell
High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell Application Note Consumer Products Authors Gerd Vanhoenacker,
More informationHyperShear TM HPLC and UHPLC Mixers
HyperShear TM HPLC and UHPLC s HyperShear TM HPLC and UHPLC Features: Micro Flow, Low Flow, Analytical Flow, High Flow and Combo Series Reduce baseline noise, increase sensitivity and improve gradient
More informationOptimized Method for Analysis of Commercial and Prepared Biodiesel using UltraPerformance Convergence Chromatography (UPC 2 )
Optimized Method for Analysis of Commercial and Prepared Biodiesel using UltraPerformance Convergence Chromatography (UPC 2 ) Mehdi Ashraf-Khorassani, 1 Giorgis Isaac, 2 and Larry T. Taylor 1 1 Department
More informationAnalysis of Petroleum Fractions by ASTM D2887
Analysis of Petroleum Fractions by ASTM D2887 Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 2582 Key Words Simulated distillation, D2887, TRACE TR-SimDist Abstract ASTM
More informationCTC Analytics Service Note June 2015 PAL-xt: Material change for Rotor Seal of C2V-3006-CTC
To: Dealer/OEM orld-ide, PAL Systems Platform PAL and PAL-xt Systems From: Toni Eberwein Date: June 19, 2015 Subject: change for rotor seal C2-30R (for valve C2V-300D-CTC) CTC Alytics Service Note June
More informationPrimeLine TM HPLC Pumps
PrimeLine TM HPLC Pumps Isocratic Model 501 Gradient System 500G PrimeLine TM Isocratic and Gradient HPLC Pumps Feature: Self-priming pump head requires no manual priming Patented Floating Pump Seal design
More informationMaintaining Your Agilent 1100 Series HPLC System
Operate your lab at peak performance Troubleshoot difficult problems Select the right Agilent supplies For detailed information about supplies from Agilent, please contact your local Agilent sales representative,
More informationHPLC Essentials Genuine Restek Replacement Parts for HPLC Systems
HPLC Essentials Genuine Replacement Parts for HPLC Systems 20 YEARS Turning Visions into Reality 800-356-1688 814-353-1300 2 Becky Wittrig HPLC Products Marketing Manager 4+ years of service! did you know?
More informationSimplify your search for. You will find both PEEK-lined stainless steel hardware as well as standard ss column hardware at the below stockingstations.
Simplify your search for COLUMN HARDWARE Discover how to maximize results with Isolation Technologies column hardware and accessories that are specifically engineered to meet critical UHPLC, HPLC, and
More information819 Injection Module User s Guide
819 Injection Module User s Guide LT2533/ 2004 Gilson, Inc. All rights reserved August 2004 Table of Contents Declaration of Conformity 1 Introduction 2 Installation 3 Operation 4 Maintenance Unpacking...
More informationLC30 CHROMATOGRAPHY OVEN OPERATOR S MANUAL Dionex Corporation
LC30 CHROMATOGRAPHY OVEN OPERATOR S MANUAL 1995 Dionex Corporation Document No. 034860 Revision 03 January 1995 1995 Dionex Corporation All rights reserved worldwide. Printed in the United States of America.
More informationPipetted Liquid Sampling. Automated sample preparation platform for online LC-MS/MS bioanalysis
Pipetted Liquid Sampling Automated sample preparation platform for online LC-MS/MS bioanalysis Integrated Sample Processing SCAP PLS system components mounted on a robotic platform Integrated Sample Processing
More information