Valving Schemes for the BioLogic Duo-Flow Chromatography System

Size: px
Start display at page:

Download "Valving Schemes for the BioLogic Duo-Flow Chromatography System"

Transcription

1 Valving Schemes for the iologic Duo-Flow Chromatography System

2 table of contents pplication asic ion Valve Operation using,, and Purge Positions position is for filling a loop with sample or for running buffer through a column; position is for injecting contents of the sample loop onto a column; Purge position is for priming the Duo-Flow workstation gradient pump and purging lines to waste without removing the column from the system. Multiple uffer Selection Gradient pump inlet valves may be used to select a different buffer, to add sanitation or storage buffer at the end of a run, or to load sample directly through the gradient pump. Reverse-Flow Chromatography for ffinity Purifications This is especially useful when purifying antibodies using Protein or other affinity columns. The protein is bound at the top of the column and nonbinding material is washed out. Flow to the column is then reversed and the bound proteins elute from the top of the column in very concentrated form, which helps prevent denaturation. -Column Switching Test columns side by side and compare their separation characteristics. dditional gradient pump inlet valves for multiple buffer selection may also be used with this application. 8-Column Switching Test up to 8 columns side by side and compare their separation characteristics. dditional gradient pump inlet valves for multiple buffer selection typically are used with this application. Mini utosampler with utomated Loop Fill and Rinse Use an auxiliary pump and a stream select valve as a - or 8-place mini autosampler. Combine a peristaltic pump, such as the Model EP- Econo TM pump or Econo gradient pump, with an inlet valve for automated filling and rinsing of a sample loop. Large-Volume ion onto a Low ack-pressure Column Use an auxiliary pump and a valve for direct injection of sample onto a column. This application uses a peristaltic pump such as the Model EP- Econo pump or Econo gradient pump to load large amounts of crude sample onto a column for a first-step separation. dditional pump inlet valves for multiple sample selection, rinsing, and buffer selection may also be used with this application. 8 Large-Volume Fraction Collection Use an VR9-8 stream select valve as a large volume fraction collector. 9 Tandem Chromatography Tandem chromatography is used to automatically collect selected fractions from one column and inject them onto a second column. Eluted fractions from the first column are collected into a DynaLoop TM or other sample loop. fter injection of the collected sample, it is eluted from the second column. Gradient pump inlet valves are used if different buffers are needed for equilibration and elution of each column. Valves Needed and x VR9-8* *Or x or SVT- x x Up to x SVT- or are optional for buffer selection and x VR9-8 Up to x SVT- or are optional for buffer selection and VR9-8* *Or or SVT- Up to x SVT-, or, or VR9-8 are optional for sample and buffer selection and VR9-8 x, VR9-8, SVT- Up to x are optional for buffer selection

3 pplication 0 Flow-Through Profiling sample is loaded onto one column and nonbinding components (flow-through proteins) pass through up to other columns. Each column may then be eluted individually. Gradient pump inlet valves are used if different buffers are needed for equilibration and elution of each column. Sequential inding and sample is loaded onto one column, nonbinding proteins are washed away, and then a fraction of eluted proteins is diverted to a second column. This column is washed, eluted, and fractions diverted to a third column, which in turn is washed and eluted. Gradient pump inlet valves are used if different buffers are needed for equilibration and elution of each column. Using the Econo Gradient Pump and a Splitter Valve to Collect Sets of Fractions The splitter valve is connected to and powered from the Econo gradient pump, which itself is controlled by the Duo-Flow system. The % Split is set in the Duo-Flow software. Valves Needed x Up to x are optional for buffer selection x Up to x are optional for buffer selection Splitter valve and an Econo gradient pump The flow exiting the column passes through the detectors and goes to the common port of the splitter valve. portion of the flow, defined by the user, is split and goes to a primary collector. The remaining portion of the flow goes to a secondary collector. The secondary collector runs in stand-alone mode, and is started manually. typical application for stream splitting is to automatically collect a small amount of each fraction for biological activity assays, thus eliminating post-run liquid transfer.

4 valve selection dvanced Chromatography pplications using Valving Introduction The iologic Duo-Flow chromatography system is a powerful and versatile biocompatible chromatography system for the purification of proteins, peptides, nucleic acids, and other small molecules. The modular nature of the Duo-Flow system makes it very easy to upgrade the gradient pump and to add peripheral devices such as detectors, auxiliary sample loading pumps, and fraction collectors. In addition, the unique versatility of the Duo-Flow system is enhanced by the availability of several different valves for automating both routine and more complex separation protocols. For the user, this means reliable operation and reproducible results, run after run, year after year. This guide describes typical valving applications. The applications range from use of a single sample injection valve to simply equilibrate a column, inject sample, and elute the column, to flow-through profiling using columns and valves. The Duo-Flow system software makes working with valves easy and intuitive. The software automatically senses each valve and displays it in a graphical format on the Manual screen. The Setup editor makes it simple to program any valving scheme by its valve aliasing feature and its ability to save an unlimited number of user-identified valving configurations. The Make-efore-reak dvantage To accommodate the wide range of pressure and flow rate of the Duo-Flow, the and VR9-8 valves use a patented make-before-break design that prevents pressure spikes when the valves rotate from one port to another. This is beneficial when using fragile columns or flow-sensitive detectors and prevents pump shutdowns due to transient over-pressure situations. Reliability Guaranteed Reliability is a vital prerequisite of any valve, especially when used with the wide range of buffers, salt solutions and solvents seen in today s laboratory. Thousands of io-rad chromatography valves are in daily use worldwide handling everything from.0 M salt solutions to 00% acetonitrile. ll valves are completely biocompatible with no metallic parts in contact with the liquid flow. Up to SV and VR Valves Can e Connected to the Duo-Flow System SVT- Valve Type Ports VR9-8 Low-pressure solenoid valve for stream selection Common,,,, SVT- Low-pressure solenoid valve for stream selection or flow diversion Common,, VR9-8 High-pressure make-before-break stream select valve Common, 8 High-pressure make-before-break injection valve (,, and Purge positions)

5 uffer uffer Column ion Purge C Column Column uffer uffer uffer uffer application in Position (Figure ) The column may be equilibrated or eluted. is loaded into the loop. If the loop is overfilled, the sample runs to waste. Never remove the syringe before the sample is injected onto the column. in Position (Figure ) The sample is injected onto the column. in Purge Position (Figure C) Flow from the gradient pump goes directly to waste, bypassing the column. asic ion Valve Operation using,, and Purge Positions position is for filling a loop with sample or for running buffer through a column; position is for injecting contents of the sample loop onto a column; Purge position is for priming the Duo-Flow workstation gradient pump and purging lines to waste without removing the column from the system.

6 C* 8 Example C 8 application Multiple uffer Selection Gradient pump inlet valves may be used to select a different buffer, to add sanitation or storage buffer at the end of a run, or to load sample directly through the gradient pump. ny valve may be attached to gradient pump inlets or. Two examples are shown. Use the Manual screen valve control to switch between valve ports and to prime each valve position using a syringe before running a method. uffer * common port uffer VR9-8 Water (Sanitizing Solution) Example ion VR9-8 C SVT- Column

7 Valve uffer uffer Column ion Valve uffer Valve uffer Valve Model 8 Model 8 Column application Step (Figure ) has first been loaded into the loop of valve in position. Valve moves to position and the sample passes to valve. With valve in position, sample is pumped onto the top of the column and nonbinding material is washed out of the column (top to bottom). Step (Figure ) Note that ports and of valve have been connected using a short piece of tubing. Valve moves to position for column washing and elution. The loop is now out of the flow path. With valve in position, the column is eluted by reverse flow (bottom to top). Reverse-Flow Chromatography for ffinity Purifications This is especially useful when purifying antibodies using Protein or other affinity columns. The protein is bound at the top of the column and nonbinding material is washed out. Flow to the column is then reversed and the bound proteins elute from the top of the column in very concentrated form, which helps prevent denaturation.

8 Valve uffer uffer application Flow Through Column columns are connected to valve as shown. With valve in position, flow from the gradient pump goes to column (Figure ). Flow Through Column With valve in position, flow from the gradient pump goes to column (Figure ). Using an for -Column Switching Test columns side by side and compare their separation characteristics. ion Valve ion Column Column Model 8 uffer Model 8 Valve uffer Valve Column Column

9 iologic uffer uffer uffer uffer Valve ion uffer (Sanitizing Solution) Column Model 8 uffer Valve Water Column application -Column Switching with uffer Selection dditional gradient pump inlet valves for multiple buffer selection may also be used with this application.

10 uffer uffer uffer uffer application Using VR9-8 Valves for 8-Column Switching Test up to 8 columns side by side and compare their separation characteristics. dditional gradient pump inlet valves for multiple buffer selection typically are used with this application. VR9-8 valves act as column inlet and outlet valves to determine which column is in line. valves act as gradient pump inlet valves for buffer selection. When valve is in position, sample is injected onto the selected column. uffer Valve For sample application Up to 8 different columns uffer ion C* C Model 8 (Sanitizing Solution) 8 8 uffer Column Inlet Valve VR9-8 s column switching valve to receive sample Column Outlet Valve VR9-8 s column switching valve to direct eluent to detectors and fraction collector * common port

11 Step Valve Loop Overfill to uffer uffer Valve VR9-8 C* * common port Econo Gradient Pump 8 Rinse Step Valve is in position for sample loading into the loop. Valve is an VR9-8 valve for sample and rinse selection. The Duo-Flow software automatically loads the selected sample into the loop. Step Valve moves to position and the sample is injected onto the column. Step When sample injection is complete, valve is automatically returned to position. The software can then automatically select a defined rinse position and rinse the loop prior to another sample injection. Model 8 Step Valve VR9-8 Step C Valve Econo Gradient Pump 8 Rinse Model 8 application uffer uffer Mini utosampler with utomated Loop Fill and Rinse Use an auxiliary pump and a stream select valve as a - or 8-place mini autosampler. Combine a peristaltic pump, such as the Model EP- Econo pump or Econo gradient pump, with an inlet valve for automated filling and rinsing of a sample loop.

12 Step 0 psi Valve uxiliary Pump Purge Low Pressure Column uffer uffer application Step Valve is in position and the column is equilibrated. The auxiliary pump is primed with sample. Step Valve moves to position and the sample is pumped directly onto the column by the auxiliary pump. The iologic Duo-Flow workstation gradient pump stops while the auxiliary pump is injecting sample. Step (not shown) When sample injection is complete, valve is automatically returned to position and the workstation gradient pump is used to elute sample from the column. Large-Volume ion onto a Low ack-pressure Column Use an auxiliary pump and a valve such as the SVT- for direct injection of sample onto a column. This application uses a peristaltic pump such as the Model EP- Econo pump or the Econo gradient pump to load large amounts of crude sample onto a column for a first-step separation. 0 psi back-pressure device may be placed after the workstation gradient pump to improve pump performance. dditional pump inlet valves for multiple sample selection, rinsing, and buffer selection may also be used with this application. SVT- C* C Valve Econo Gradient Pump Rinse SVT- Rinse Econo Gradient Pump Model 8 Step 0 psi Low Pressure Column Model 8 * common port uffer uffer

13 Valve uffer uffer Example Example Fraction Fraction Fraction * common port Valve VR9-8 s stream select valve Fraction Valve VR9-8 C Fraction ion Fraction 8 Fraction Fraction Fraction Fraction C* Fraction 8 Fraction Fraction Fraction Fraction Fraction 8 Fraction 0 Fraction Valve VR9-8 C Fraction 9 8 Fraction Fraction 8 Fraction Fraction application 8 Large-Volume Fraction Collection Use an VR9-8 stream select valve as a large-volume fraction collector. n VR9-8 valve can function as an 8-place collector in place of a regular fraction collector (Example ). n additional VR9-8 can be plumbed to the first one to provide -place capacity (Example ). Individual change valve steps are programmed in the method Protocol to collect the desired fractions.

14 application 9 Valve Column Tandem Chromatography uffer uffer Tandem chromatography is used to automatically collect selected fractions from one column and inject them onto a second column. Eluted fractions from the first column are collected into a DynaLoop or other sample loop. fter injection of the collected sample, it is eluted from the second column. Gradient pump inlet valves are used if different buffers are needed for equilibration and elution of each column. n example of this application is the purification of monoclonal antibodies, where column is a cation exchanger and column is ceramic hydroxyapatite. This automated procedure can be performed at neutral ph, as an alternative to affinity chromatography, which requires low-ph buffers to elute the antibodies. Step (not shown) The sample is loaded into the loop of valve and injected onto column. With the appropriate buffer selected from the gradient pump inlet valves, column is eluted. Valve directs the flow to the fraction collector for collection or diversion to waste. Step t a predetermined time or volume, valve moves to position and a portion of the column eluate is captured in the loop/dynaloop attached to valve. uffer uffer Step Model 8 Valve Valve VR9-8 C* 8 uffer (Sanitizing Solution) C Valve SVT- * common port DynaLoop Column step

15 Valve Purge Column uffer uffer uffer Valve uffer DynaLoop Column uffer (Sanitizing Solution) Valve VR9-8 C 8 Step Valve moves to Purge position to put valve and column into the flow path. Valves and change position to connect column to the detectors and fraction collector. With the appropriate buffer selected from the gradient pump inlet valves, the sample stored in the loop/dynaloop is injected onto column when valve moves to position (shown). Step (not shown) Valve reverts to position and column is eluted. Model 8 C Valve SVT- steps &

16 Purge Valve application 0 uffer uffer uffer Flow-Through Profiling using Valves uffer sample is loaded onto one column and nonbinding components (flow-through proteins) pass through up to other columns. Each column may then be eluted individually. Gradient pump inlet valves are used if different buffers are needed for equilibration and elution of each column. Step With valve in Purge position and valves and in position, the gradient pump may be primed with the buffers from the gradient pump inlet valves. T-unions are used to connect tubing. Port of valves and should be end-capped to prevent misdirected flow. To Equilibrate Columns Individually: Column Valve Valve Valve or Purge Purge uffer (Sanitizing Solution) uffer T-union uffer T-union ion Column Column UV Detector Column Model 8 Valve Valve step

17 Valve uffer uffer uffer uffer Column uffer uffer (Sanitizing Solution) uffer Valve Column Step To elute column, valve is in position. Valves and are both in position and are out of the flow path. The flow from column goes to the fraction collector. Column UV Detector Model 8 Valve step

18 Step To elute column, the gradient pump inlet valves both change to buffer. Valve changes to Purge position and valves and remain in position. The flow elutes column, bypassing columns and, and goes to the fraction collector. uffer uffer uffer uffer Purge step Column Valve uffer uffer (Sanitizing Solution) uffer Valve Column Purge Valve Valve Column uffer uffer uffer uffer Column uffer uffer Step To elute column, the gradient pump inlet valves change to buffer. (sanitizing solution) uffer Column Valve UV Detector Model 8 Valve remains in Purge position, valve is in position and valve remains in position. The flow elutes column and goes to the fraction collector. Column Valve step UV Detector Model 8

19 Valve uffer uffer uffer uffer Step ion Column uffer uffer (Sanitizing Solution) uffer Valve Column Step (not shown) Valve is placed initially in position. Valves and remain in position. ll three columns are equilibrated and the flow goes to waste via the fraction collector diverter valve. Step is loaded into the loop. Valve is placed in position and the sample is injected onto column. With valves and remaining in position, the unbound components of the sample pass through all three columns and then to the fraction collector (or to waste via the fraction collector diverter valve). The flow path is shown in blue. Column UV Detector Model 8 Valve step

20 Purge Valve application uffer uffer uffer uffer uffer Sequential inding and using Valves sample is loaded onto one column, nonbinding proteins are washed away, and then a fraction of eluted proteins is diverted to a second column. This column is washed, eluted, and fractions diverted to a third column, which in turn is washed and eluted. Gradient pump inlet valves are used if different buffers are needed for equilibration and elution of each column. Step With valve in Purge position and valves and in position, the gradient pump may be primed with the buffers from the gradient pump inlet valves. T-unions are used to connect tubing. Port of valves and should be end-capped to prevent misdirected flow. To Equilibrate Columns Individually: Column Valve Valve Valve or Purge Purge (Sanitizing Solution) uffer T-union uffer T-union ion Column Column UV Detector Column Model 8 Valve Valve step

21 Step (not shown) Valve is placed in position to load sample into the loop. Valve is in position and valve is in position. Columns and are now bypassed. Step Valves changes to position and sample is loaded onto column. uffer uffer uffer uffer ion Column step Valve uffer uffer (Sanitizing Solution) uffer Step Valve Column Valve Valve Column uffer uffer uffer uffer Column uffer uffer (Sanitizing Solution) Step Valve changes to position to wash column. The flow goes to the fraction collector or to waste via the fraction collector diverter valve. uffer Column Column Valve Valve UV Detector Model 8 step UV Detector Model 8

22 application Step Valve changes to position. The flow from column passes through column and then to the fraction collector. The flow path is shown. uffer uffer uffer uffer uffer (Sanitizing Solution) Step Valve changes to Purge position to bypass column. With valves and in position, the flow passes only through column and then to the fraction collector (or waste via the fraction collector, as shown). uffer uffer uffer Purge Column uffer Column UV Detector uffer Column Model 8 uffer Valve Valve Valve uffer (Sanitizing Solution) uffer uffer Column Column UV Detector Column Model 8 step Valve Valve Valve step

23 Step Valve changes to position. The flow passes through column and column. The flow path is shown. uffer uffer uffer uffer Purge Column step Valve uffer uffer (Sanitizing Solution) uffer Valve Column Purge Valve Valve Column uffer uffer uffer uffer Column uffer uffer (Sanitizing Solution) Step Valve changes to position. Column is now bypassed. uffer Column Valve UV Detector Model 8 Valve changes to position and column is now washed and eluted (shown). Column Valve step UV Detector Model 8

24 application ion Model 8 Using an Econo Gradient Pump and a Splitter Valve to Collect Sets of Fractions. The splitter valve is connected to and powered from the Econo gradient pump, which itself is controlled by the Duo-Flow system. The % Split is set in the Duo-Flow software. The flow exiting the column passes through the detectors and goes to the common port of the splitter valve. portion of the flow, defined by the user, is split and goes to a primary collector. The remaining portion of the flow goes to a secondary collector. The secondary collector runs in stand-alone mode, and is started manually. typical application for stream splitting is to automatically collect a small amount of each fraction for biological activity assays, thus eliminating post-run liquid transfer. Splitter Valve Column e.g., 9% % Econo Gradient Pump uffer uffer Model 8

25 For Users of the iologic HR System comparison ll the valving applications described for the iologic Duo-Flow system are also applicable to the iologic HR system. However, the user must consult the following valve port comparison guide as the HR system uses V- and V9-8 valves, not VR valves. These valves are not interchangeable between systems. oth systems can use valves. The Duo-Flow system uses SVT- valves whereas the HR system uses SV- valves. Comparison of Port Functions on and V- ion Valves Port Valve for the iologic Duo-Flow System Port V- Valve for the iologic HR System /overflow from sample loop Purge inject port to fill loop From gradient pump loop loop To column /overflow from sample loop From gradient pump inject port to fill loop loop loop Purge To column iologic Duo-Flow Valve Position Corresponding iologic HR V- Valve Position

26 Ordering Information Catalog # Description iologic Duo-Flow Valves 0-00 utomated ion Valve 0-00 Valve Rebuild Kit VR9-8 utomated Stream Select Valve 0-00 VR9-8 Valve Rebuild Kit 0-00 SVT- Diverter Valve 0-0 SVT- Valve Rebuild Kit 0-0 Select Valve, -port, -position solenoid random access valve, 0 psi ( bar) limit iologic Dynamic (Sliding Piston) Loops 0-0 DynaLoop Kit, includes a ml DynaLoop sliding piston loop, DynaLoop parts kit, instructions 0-0 DynaLoop 90 Kit, includes a 90 ml DynaLoop sliding piston loop, DynaLoop parts kit, instructions 0-00 DynaLoop Parts Kit, includes end-cap O-rings, sliding seal O-ring, filter, nut fittings, /8" ferrules, /-8 nuts and ferrules, /8" Teflon tubing, 0 ft 0-0 DynaLoop Sliding Piston Loop, ml sample loop replacement 0-0 DynaLoop 90 Sliding Piston Loop, 90 ml sample loop replacement 0-0 DynaLoop Sliding Seal Replacement iologic System ccessories 0-00 V- Manual ion Valve, -port, -position sample load valve,,000 psi (0 bar) limit 0-00 V- utomated ion Valve, -port, -position automatic sample load valve,,000 psi (0 bar) limit 0-00 V9-8 Stream Select Valve, 9-port, 8-position valve,,000 psi (0 bar) limit V9-8 utomated Stream Select Valve, 9-port, 8-position automatic valve,,000 psi (0 bar) limit 0-00 SV- Diverter Valve, -port, -position solenoid diversion valve, 0 psi ( bar) limit Model EP- Econo Pump -80 Model EP- Econo Pump, 0 V -8 Model EP- Econo Pump, 0 V -8 System Cable, connects Econo pump to iologic HR and Duo-Flow systems Econo Gradient Pump -900 Econo Gradient Pump, 00/0 V, includes tubing and fittings kit -900 Econo Gradient Pump, 0/0 V, includes tubing and fittings kit -90 Splitter Valve Kit, includes valve, tubing and adaptors Teflon is a trademark of E. I. dupont de Nemours and Company. io-rad Laboratories Life Science Group Web Site U.S. (800) IORD ustralia ustria (0) elgium 09-8 Canada (90) - China / Denmark 9 99 Finland 8 (0) France Germany Hong Kong India (9-) -00 Israel 0 9 Italy Japan Korea Latin merica Mexico -0 The Netherlands 08-0 New Zealand Norway Russia Singapore -98 Spain Sweden (0)8-00 Switzerland United Kingdom ulletin 8 US/EG Rev Sig 099

BioLogic DuoFlow Valving Guide

BioLogic DuoFlow Valving Guide Protein Purification iologic DuoFlow Valving Guide Valving Schemes for the iologic DuoFlow Chromatography System Protein Purification iologic DuoFlow Valving Guide Introduction The iologic DuoFlow Chromatography

More information

DynaLoop 25 and DynaLoop 90. Instruction Manual

DynaLoop 25 and DynaLoop 90. Instruction Manual 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 0 5 10 15 20 25 DynaLoop 25 and DynaLoop 90 Instruction Manual Catalog Numbers 750-0451 (25 ml) 750-0452 (90 ml) For technical support, call your

More information

Runwithit! The Mini-PROTEAN. 3 Mini Vertical Electrophoresis System is So Easy to Use, You Won t Want to Run With Anything Else.

Runwithit! The Mini-PROTEAN. 3 Mini Vertical Electrophoresis System is So Easy to Use, You Won t Want to Run With Anything Else. Runwithit! The Mini-PROTEAN 3 Mini Vertical Electrophoresis System is So Easy to Use, You Won t Want to Run With Anything Else. The User-Friendly 3 Mini-PROTEAN A Simple Approach to Gel Electrophoresis.

More information

Electrophoresis + Blotting + Imaging. V3 Western Workflow Visualize. Verify. Validate.

Electrophoresis + Blotting + Imaging. V3 Western Workflow Visualize. Verify. Validate. Electrophoresis + Blotting + Imaging V3 Western Workflow Visualize. Verify. Validate. Why choose the V3 Western Workflow from Bio-Rad? 1 2 Separate Proteins Visualize Separation Transfer Proteins Speed

More information

ÄKTA system Training Guide. Purifier 100 / 10. amersham bio science

ÄKTA system Training Guide. Purifier 100 / 10. amersham bio science ÄKTA system Training Guide Purifier 00 / 0 amersham bioscience amersham bioscience amersham bioscience amersham bio science ÄKTApurifier Buffers Injector Box Monitor Conductivity ph On-line Filter UV Detector

More information

UView Mini Transilluminator

UView Mini Transilluminator UView Mini Transilluminator Instruction Manual Catalog #166-0531 Hardware Instruction Manual The UView mini transilluminator is suitable for research use only. It must be used by specialized personnel

More information

Plus Dodeca Cell System

Plus Dodeca Cell System Expression Proteomics PROTEAN Plus Dodeca Cell System 12 High-Resolution, Reproducible Second-Dimension Gels in an Incredible 6 Hours PROTEAN Plus Dodeca Cell Second-Dimension Gels in only 6 Hours The

More information

Amplification: T100 Thermal Cycler. T100 Thermal Cycler

Amplification: T100 Thermal Cycler. T100 Thermal Cycler Amplification: T100 Thermal Cycler T100 Thermal Cycler The Smart PCR Choice The T100 Thermal Cycler s intuitive touch screen makes running PCR easier than ever before. The T100 Thermal Cycler s performance,

More information

HIGH PRESSURE SWITCHING VALVES

HIGH PRESSURE SWITCHING VALVES HIGH PRESSURE SWITCHING VALVES Rheodyne offers high pressure switching valves to simplify procedures and improve the speed, resolution, and sensitivity of HPLC analysis. The switching valves are available

More information

Cooling Module. Instruction Manual

Cooling Module. Instruction Manual Cooling Module Instruction Manual Catalog Numbers 170-3654, 170-3655, and 170-3688 For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-800-424-6723) Warranty Bio-Rad

More information

Chromatography. Liquid. HPLC Accessories SupelPRO Valves SUPELCO

Chromatography. Liquid. HPLC Accessories SupelPRO Valves SUPELCO 8 SupelPRO Valves Order:.800..00 Technical Service:.800.9.0 Web: www.sigma-aldrich.com/supelco SupelPRO Automated Fluidics Instruments Supelco s SupelPRO series are precision, electronically-controlled,

More information

ÄKTA system Training Guide

ÄKTA system Training Guide ÄKTA system Training Guide Purifier 00 / 0 amersham bioscience amersham bioscience amersham bioscience amersham bioscience ÄKTApurifier Buffers Injector Box Monitor Conductivity ph On-line Filter UV Detector

More information

Amplification: T100 Thermal Cycler. T100 Thermal Cycler

Amplification: T100 Thermal Cycler. T100 Thermal Cycler Amplification: T100 Thermal Cycler T100 Thermal Cycler The Smart PCR Choice The T100 thermal cycler s intuitive touch screen makes running PCR easier than ever before. The T100 thermal cycler s performance,

More information

S 1125 HPLC Pump System

S 1125 HPLC Pump System Routine S 170 EXPERT HPLC HPLC System SYSTEMS 500 The Sykam ROUTINE HPLC System S 500 is intended for all routine analysis. The system configuration is highly variable and several upgrade options make

More information

NI PS-15/16/17 Side Mount Brackets

NI PS-15/16/17 Side Mount Brackets INSTALLATION GUIDE NI PS-15/16/17 Side Mount Brackets This document provides the installation procedure for the NI PS-15/16/17 Side Mount Brackets. This accessory is used to mount NI PS-15/16/17 power

More information

SFC-FID ASTM System. SFC-4000 Series SFC. Performance Innovation Reliability

SFC-FID ASTM System. SFC-4000 Series SFC. Performance Innovation Reliability SFC-FID ASTM System SFC-4000 Series SFC Performance Innovation Reliability The JASCO SFC-FID system has been developed as a simple and validated turn-key solution for the measurement of the two ASTM standard

More information

Information Technology and Economic Development: An Introduction to the Research Issues

Information Technology and Economic Development: An Introduction to the Research Issues November 1998 Working Papers No. 153 Information Technology and Economic Development: An Introduction to the Research Issues TABLE 1 WORLD-WIDE INFORMATION TECHNOLOGY MARKET BREAKDOWN, 1995

More information

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve Technical Note This Technical Note describes the purpose, installation, and configuration and use of a diverter valve

More information

Thermo Scientific Dionex UltiMate 3000 Thermostatted Column Compartment

Thermo Scientific Dionex UltiMate 3000 Thermostatted Column Compartment chromatography Thermo Scientific Dionex UltiMate 3000 Thermostatted Column Compartment Product Specifications Thermo Scientific Dionex UltiMate 3000 products are UHPLC compatible by design, establishing

More information

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS MS Identification of Trace level Impurities from a Non-MS Compatible Mobile Phase Using ACQUITY UPLC System with 2D Technology by Heart-cutting and Online Sample Concentration Bronsky Gopinadh, Dilshad

More information

98 동방하이테크상사 T.(02) F.(02)

98 동방하이테크상사   T.(02) F.(02) 98 동방하이테크상사 http://www.allpump.co.kr T.(0)57-69 F.(0)57-69 00--7 troduction to Upchurch Scientific Valves Valves troduction Valves jection and Selection Valves for Pressures to,000 psi (69 bar) Switching

More information

Motor Valve MV-7/IMV-7

Motor Valve MV-7/IMV-7 Motor Valve MV-7/IMV-7 INSTRUCTIONS 1. Introduction Motor Valves MV-7/IMV-7 are a 3-position, 7-port motorized valves designed to he used as automatic sample injection valves or system valves. The valves

More information

Agilent G5632 Valve Kit Instructions

Agilent G5632 Valve Kit Instructions Agilent G5632 Valve Kit Instructions Agilent G5632 Valve Kit Instructions Contents Typical Applications 2 Alternating Column Regeneration (2 position/10 port valves only) 2 Dual column selection 3 Sample

More information

Bio-Rad Laboratories. Liquichek Hematology Controls

Bio-Rad Laboratories. Liquichek Hematology Controls Bio-Rad Laboratories H E M A T O L O G Y C O N T R O L S Liquichek Hematology Controls Bio-Rad Laboratories H E M A T O L O G Y C O N T R O L S Your Choice in Quality Control With more products and more

More information

Model 2110 Fraction Collector Instruction Manual

Model 2110 Fraction Collector Instruction Manual Model 2110 Fraction Collector Instruction Manual Catalog Numbers 731-8120 731-8122 Warranty Model Serial No. Date of Delivery Warranty Period Unless otherwise specified, instruments sold by Bio-Rad Laboratories

More information

Total credit to the non-financial sector (core debt), % of GDP Table F1.1

Total credit to the non-financial sector (core debt), % of GDP Table F1.1 Total credit to the non-financial sector (core debt), % of GDP Table F1.1 2012 2013 2014 2015 2016 Q2 16 Q3 16 Q4 16 Q1 17 Q2 17 Argentina 62.6 66.4 64.6 75.5 75.3 81.7 80.1 75.3 71.6 72.7 Australia 208.0

More information

Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas

Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas Liliana Krotz, Walter Galotta, and Guido Giazzi Thermo Fisher Scientific, Milan, Italy Overview

More information

Spring Driven Reels. Manufactured by Cavotec RMS

Spring Driven Reels. Manufactured by Cavotec RMS Spring Driven Reels Manufactured by Cavotec RMS Spring Driven Reels Who we are Cavotec is a multi-national group of companies serving the following industries: mining and tunnelling, ports and maritime,

More information

INSTALLATION GUIDE NI WSN-3291 Outdoor IP Enclosure For NI WSN-32xx Series Devices

INSTALLATION GUIDE NI WSN-3291 Outdoor IP Enclosure For NI WSN-32xx Series Devices INSTALLATION GUIDE NI WSN-3291 Outdoor IP Enclosure For NI WSN-32xx Series Devices This installation guide describes how to install and use the National Instruments WSN-3291 outdoor IP enclosure. The NI

More information

Infographics on Electromobility (January 2019)

Infographics on Electromobility (January 2019) Infographics on Electromobility (January 2019) Publisher: BMW Group Corporate Communications Electromobility Last Update: 04.01.2019 Contact: presse@bmw.de ELECTROMOBILITY IN GERMANY. SHARE IN NEW REGISTRATIONS

More information

VERITY Purification Systems Build the perfect system to fit your research

VERITY Purification Systems Build the perfect system to fit your research VERITY Purification Systems Build the perfect system to fit your research SPEC SHEET PURIFICATION VERITY PURIFICATION SYSTEMS WITH VERITY 3011 ISOCRATIC PUMP MAKE YOUR LAB LIFE EASIER From automated GPC

More information

AUTOMATED TORSION BENCH DRIVETWIST

AUTOMATED TORSION BENCH DRIVETWIST Force and torque measurement systems AUTOMATED TORSION BENCH DRIVETWIST ANDILOG Technologies info@andilog.com Phone : +33 (0) 820 888 202 Fax : +33 (0) 820 888 902 Automated torsion bench DRIVETWIST Automated

More information

BioDiesel Processes. Comprehensive Separation and Filtration Technologies for PROCESS SURVEY PILOT PROGRAM PROCESS SUPPORT TECHNICAL RECOMMENDATIONS

BioDiesel Processes. Comprehensive Separation and Filtration Technologies for PROCESS SURVEY PILOT PROGRAM PROCESS SUPPORT TECHNICAL RECOMMENDATIONS Comprehensive Separation and Filtration Technologies for BioDiesel Processes PROCESS SURVEY PILOT PROGRAM PROCESS SUPPORT TECHNICAL RECOMMENDATIONS TECHNOLOGY IMPLEMENTATION PROPOSALS Where to Apply Pall

More information

Allegro Bioprocessing Workstations. Allegro TMBioprocessing Workstations. USD 2793b. The system s support trays are made of a highthe

Allegro Bioprocessing Workstations. Allegro TMBioprocessing Workstations. USD 2793b. The system s support trays are made of a highthe USD 2793a USD 2793b Allegro TMBioprocessing Workstations Allegro Bioprocessing Workstations To support supportaawide widerange rangeofoffluid fluidmanagement management To applicationswith withallegro

More information

PrimeLine TM HPLC Pumps

PrimeLine TM HPLC Pumps PrimeLine TM HPLC Pumps Isocratic Model 501 Gradient System 500G PrimeLine TM Isocratic and Gradient HPLC Pumps Feature: Self-priming pump head requires no manual priming Patented Floating Pump Seal design

More information

Allegro 2D Standard Systems

Allegro 2D Standard Systems USD3190 Allegro 2D Standard Systems Faster Implementation of Validated Single-Use Systems The Allegro 2D systems are state-of-the art single-use systems designed and built by Pall Life Sciences in order

More information

UNIVERSAL TESTING MACHINES STENTOR II CC & ATLAS II CC

UNIVERSAL TESTING MACHINES STENTOR II CC & ATLAS II CC UNIVERSAL TESTING MACHINES STENTOR II CC & ATLAS II CC The product line STENTOR II CC and ATLAS II CC STENTOR II CC The single column testing machine up to 5kN - 3 models: 1kN, 2kN and 5kN - 2 sizes: 200mm

More information

GE Healthcare. ÄKTApilot. User Manual

GE Healthcare. ÄKTApilot. User Manual Pilot_UM.book Page 1 Tuesday, May 8, 2007 1:06 PM GE Healthcare ÄKTApilot User Manual Important user information All users must read this entire manual to fully understand the safe use of ÄKTApilot. WARNING!

More information

Xampler laboratory membrane cartridges

Xampler laboratory membrane cartridges GE Healthcare Data File 18-1170-21 AC Membrane separations Xampler laboratory membrane cartridges GE Healthcare Xampler ultrafiltration and microfiltration cartridges are designed for laboratory scale

More information

UNIVERSAL TESTING MACHINERY STENTOR II CC - ATLAS II CC - T-DRIVE CC

UNIVERSAL TESTING MACHINERY STENTOR II CC - ATLAS II CC - T-DRIVE CC UNIVERSAL TESTING MACHINERY STENTOR II CC - ATLAS II CC - T-DRIVE CC The product line STENTOR II CC and ATLAS II CC STENTOR II CC A single column testing machine for up to 5kN / 1,000 lbs - 3 models: 1kN,

More information

NEW HPLC - SYSTEM LINE

NEW HPLC - SYSTEM LINE NEW HPLC - SYSTEM LINE S c h a m b e c k S F D G m b H Drieschweg 13A D-53604 Bad Honnef/Germany Phone: +49 2224 9239-0 E-Mail: info@schambeck-sfd.com Internet: www.schambeck-sfd.com S 9425 HPLC solvent

More information

breakthrough versatility

breakthrough versatility ipro 5000 Series Total Nitrogen/Total Sulfur (TN/TS) Analyzer breakthrough versatility with exceptional accuracy and precision Petrochemicals Refineries Chemical Plants Commercial Testing Laboratories

More information

Agilent InfinityLab LC Purification Solutions SELECTION GUIDE

Agilent InfinityLab LC Purification Solutions SELECTION GUIDE gilent InfinityLab LC urification Solutions SELECTION GUIDE GILENT INFINITYLB LC URIFICTION SOLUTIONS EFFICIENT LC URIFICTION SOLUTIONS FOR NLYTICL TO RERTIVE SCLE gilent InfinityLab LC urification Solutions

More information

INSTALLATION GUIDE NI 9921/9922 Outdoor IP Enclosure For NI WLS-9000 Series Devices

INSTALLATION GUIDE NI 9921/9922 Outdoor IP Enclosure For NI WLS-9000 Series Devices INSTALLATION GUIDE NI 99/99 Outdoor IP Enclosure For NI WLS-9000 Series Devices This guide describes how to install and use the National Instruments 99 and National Instruments 99 outdoor IP enclosures.

More information

HPLC SYSTEM S 500 SERIES

HPLC SYSTEM S 500 SERIES HPLC SYSTEM S 500 SERIES The Sykam HPLC System S 500 Series is intended for all routine analysis as well as for the ambitious analyst. The system configuration is highly variable and several upgrade options

More information

testing to perfection MultiTest-i Technical Datasheet

testing to perfection MultiTest-i Technical Datasheet testing to perfection MultiTest-i Technical Datasheet MultiTest 0.5-i, 1-i & 2.5-i 0.5 kn 1 kn 2.5 kn Height Width Depth Vertical Daylight* MultiTest 0.5-i 1710 mm (67.3) 290 mm (11.4) 414 mm (16.3) 1359

More information

A Comparison of Boat Introduction and Direct Injection using the Thermo Scientific ipro 5000 Series Combustion Analyzer

A Comparison of Boat Introduction and Direct Injection using the Thermo Scientific ipro 5000 Series Combustion Analyzer Application Note: 42165 A Comparison of Boat Introduction and Direct Injection using the Thermo Scientific ipro 5000 Series Combustion Analyzer Kristian J. Hoffman, Angela Seipel, Application Specialists,

More information

High Pressure Valves. MX Series II Modules Manual Sample Injection Switching

High Pressure Valves. MX Series II Modules Manual Sample Injection Switching High Pressure Valves MX Series II Modules Manual Sample Injection Introduction to High Pressure Valves HIGH Pressure Valves High Pressure Valves Introduction Rheodyne valves fit virtually any flow control

More information

Lifting Instructions for 1336 PLUS and FORCE D Frame Drives

Lifting Instructions for 1336 PLUS and FORCE D Frame Drives Installation Instructions Lifting Instructions for 1336 PLUS and FORCE D Frame Drives This publication will guide you through the steps needed to properly lift and mount the following drives: 1336 PLUS

More information

Instructions for Use SeQuant CARS Continuous Auto-Regeneration System for SeQuant SAMS Suppressors

Instructions for Use SeQuant CARS Continuous Auto-Regeneration System for SeQuant SAMS Suppressors Instructions for Use SeQuant CARS Continuous Auto-Regeneration System for SeQuant SAMS Suppressors Instructions for Use SeQuant CARS Continuous Auto-Regeneration System CONTENTS General Information...

More information

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE AN HPLC THAT WORKS AS HARD AS YOU DO Confidence in your results comes with high performance, day-to-day reliability, and overall system robustness.

More information

AUSTRIA. Table 1. FDI flows in the host economy, by geographical origin. (Millions of US dollars)

AUSTRIA. Table 1. FDI flows in the host economy, by geographical origin. (Millions of US dollars) Table 1. FDI flows in the host economy, by geographical origin World 5 690 138 6 195 3 184 10 784 7 933 31 154 6 858 9 303 840 10 618 5 760 Developed economies 5 740 13 5 341 2 689 10 137 6 291 29 721

More information

Evaluating Battery Run-Down with the N6781A 2-Quadrant Source/Measure Unit and the 14585A Control and Analysis Software.

Evaluating Battery Run-Down with the N6781A 2-Quadrant Source/Measure Unit and the 14585A Control and Analysis Software. Evaluating Battery Run-Down with the N6781A 2-Quadrant Source/Measure Unit and the 14585A Control and Analysis Software Application Note Analyzing and optimizing battery run time requires running tests

More information

High Flow Cartridge Valves 2/2 Way Series NG16 NG100

High Flow Cartridge Valves 2/2 Way Series NG16 NG100 High Flow Cartridge Valves 2/2 Way Series NG16 NG100 Table of Contents Designation Symbol Page General Description and Operating Principle 3 Sectional View and Types of Cones and Sleeves 4 Specifications

More information

testing to perfection MultiTest-i Technical Datasheet

testing to perfection MultiTest-i Technical Datasheet testing to perfection MultiTest-i Technical Datasheet MultiTest 0.5-i, 1-i & 2.5-i 0.5kN 1kN 2.5kN Height Width Depth Vertical Daylight* MultiTest 0.5-i 1710mm (67.3) 290mm (11.4) 414mm (16.3) 1359mm (53.5)

More information

Determination of Sudan Dyes I IV in Curry Paste

Determination of Sudan Dyes I IV in Curry Paste Determination of Sudan Dyes I IV in Curry Paste Suparerk Tukkeeree and Jeffrey Rohrer 2 Thermo Fisher Scientific, Bangkok, Thailand; 2 Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 23 Key

More information

Agilent Valve Kit G4231A/B/C and G4232B/C/D

Agilent Valve Kit G4231A/B/C and G4232B/C/D Agilent Valve Kit G4231A/B/C and G4232B/C/D Instructions Agilent Valve Kit G4231A/B/C and G4232B/C/D - Instructions This Technical Note describes the installation of the Agilent G4231A/B/C and G4232B/C/D

More information

Fixed Flow Splitter *

Fixed Flow Splitter * Fixed Flow * *Patent Pending Analytical Scientific Instruments Fluid resistor technology eliminates tedious adjustments to capillary tubing to change split ratios. Easy to use interchangeable fluid resistors

More information

GlycoChrom TM HPLC Analyzer. Hardware Installation and System Test

GlycoChrom TM HPLC Analyzer. Hardware Installation and System Test GlycoChrom TM HPLC Analyzer Hardware Installation and System Test PIN 410-5004 For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-8 00-424-6723) ,, Table of Contents

More information

78 동방하이테크상사 T.(02) F.(02)

78 동방하이테크상사  T.(02) F.(02) 78 동방하이테크상사 http://www.allpump.co.kr T.(0)7-9 F.(0)7-9 00--7 Introduction to Introduction Rheodyne valves fit virtually any high-pressure flow control application. As you ll see from the table below and

More information

Agilent SD-1 Purification System ml/min

Agilent SD-1 Purification System ml/min Agilent SD-1 Purification System - 200 ml/min Introduction Installation of the Capillary Kit This technical note shows you how to install the Stainless Steel Capillary Kit (SD- 1 Purification System -

More information

LC20 CHROMATOGRAPHY ENCLOSURE OPERATOR S MANUAL Dionex Corporation

LC20 CHROMATOGRAPHY ENCLOSURE OPERATOR S MANUAL Dionex Corporation LC20 CHROMATOGRAPHY ENCLOSURE OPERATOR S MANUAL 1993 Dionex Corporation Document No. 034859 Revision 02 September 1993 1993 Dionex Corporation All rights reserved worldwide Printed in the United States

More information

CHROMATOGRAPHY. SYKAM S 6000 Valve Series

CHROMATOGRAPHY. SYKAM S 6000 Valve Series CHROMATOGRAPHY SYKAM S 6000 Valve Series Version: March 2011 SYKAM S 6000 Valve Series Rotor Seals Sykam offers a variety of different Rotor Seals materials to satisfy any possible application. All materials

More information

Setting up SilFlow for BackFlush in your GC

Setting up SilFlow for BackFlush in your GC Setting up SilFlow for BackFlush in your GC What is backflush and why use it? The BackFlush system eliminates the need to bake heavy sample fractions off the capillary column. Oils, tars and other semivolatile

More information

HELICOPTERS Service. Helicopter Maintenance Upgrades

HELICOPTERS Service. Helicopter Maintenance Upgrades HELICOPTERS Service Helicopter Maintenance Upgrades MISSION SUCCESS WITH AIRBUS HELICOPTERS UPGRADES Airbus Helicopters Customer Centers are at your service: providing local, on-the-spot solutions Airbus

More information

Drive solutions for the oil and gas industry.

Drive solutions for the oil and gas industry. Drive solutions for the oil and gas industry. maxon Heavy Duty drives: Robust design for extreme operating conditions. Founded in Switzerland. Available worldwide. maxon a strong global brand. maxon motor,

More information

Thermo Scientific Sorvall MTX and MX Plus Micro-Ultracentrifuge Series. powerful separation technology. quiet, compact design

Thermo Scientific Sorvall MTX and MX Plus Micro-Ultracentrifuge Series. powerful separation technology. quiet, compact design Thermo Scientific Sorvall MTX and MX Plus Micro-Ultracentrifuge Series powerful separation technology quiet, compact design Thermo Scientific Sorvall MTX and MX Plus Micro-Ultracentrifuge Series Our Sorvall

More information

I. Global wine markets,

I. Global wine markets, I. Global wine markets, 2007-09 Table 1: Summary of the world's wine markets, 2007-09 (annual average) Vine % of cropland Wine Wine Wine Wine Wine Per adult wine Pop- GDP GDP Area under vines produced

More information

Lighting up your way. Illuminated Vista Signs

Lighting up your way. Illuminated Vista Signs Lighting up your way Illuminated ista Signs Convex, double sided signs Flat light boxes Perfect for advertising Easy assembly Clean, streamline design Aluminum and black finishes Environmentally friendly

More information

Chromaflow columns. GE Healthcare. General column description. convenience saving of labor reproducibility contained packing scalability

Chromaflow columns. GE Healthcare. General column description. convenience saving of labor reproducibility contained packing scalability GE Healthcare Data File 18-1138-92 AE Process columns Chromaflow columns Chromaflow columns are a family of convenient to use, process-scale columns. A patented nozzle in the top and bottom of the column

More information

STI OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

STI OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY Extended range HPLC has been investigated for many years; however, progress was delayed because of

More information

Installation of the 2 position/10 port valve into the 1100 series thermostatted column compartment

Installation of the 2 position/10 port valve into the 1100 series thermostatted column compartment Agilent G1316A Thermostatted Column Compartment Technical Note Installation of the 2 position/10 port valve into the 1100 series thermostatted column compartment System Requirements 2 Performance Specifications

More information

manifolds Innovative solutions to increase quality and reduce cost You can see our thinking

manifolds Innovative solutions to increase quality and reduce cost You can see our thinking www.norgren.com manifolds Innovative solutions to increase quality and reduce cost You can see our thinking engineering advantage creating advantage with people, products, innovation and services A world

More information

Agilent 218 Purification Solution

Agilent 218 Purification Solution Agilent 218 Purification Solution System User Guide Purification Solution System User Guide Agilent Technologies Notices Agilent Technologies, Inc. 2012, 2013 No part of this manual may be reproduced in

More information

Rectangular photoelectric proximity switches

Rectangular photoelectric proximity switches Rectangular photoelectric proximity switches Features Diffuse sensors with operating distance of 300 mm or 800 mm Through-beam sensors with operating distance of 15 or 50 m Reflex sensors with operating

More information

2-Way Spool Type Cartridge Valve (CKE) ISO 7368 NB16 - NB50

2-Way Spool Type Cartridge Valve (CKE) ISO 7368 NB16 - NB50 2-Way Spool Type Cartridge Valve Series D 2-Way Spool Type Cartridge Valve (CKE) ISO 7368 N16 - N50 General Description Cartridge valves, also known as 2-way valves or logic valves, conform to ISO 7368

More information

Supported Liquid Extraction (SLE)

Supported Liquid Extraction (SLE) Supported Liquid Extraction Diatomaceous Earth Synthetic Supported Liquid Extraction (SLE) Choose Between Two Unique Sorbents www.phenomenex.com/sle Novum Synthetic SLE As a unique, synthetic SLE sorbent,

More information

OECD unemployment rate down to 6.5% in January 2016

OECD unemployment rate down to 6.5% in January 2016 OECD unemployment rate down to 6.% in January 2016 The OECD unemployment rate fell by 0.1 percentage point to 6.% in January 2016, 1.6 percentage points below the January 2013 peak. Across the OECD area,

More information

PREVIEW. Fieldbus System for Distributed I/O NEW PRODUCTS

PREVIEW. Fieldbus System for Distributed I/O NEW PRODUCTS NEW PRODUCTS PREVIEW Fieldbus System for Distributed I/O! Dual Port EtherNet/IP TM Gateway controls up to 8 Valve Manifolds and 8 Input blocks Controls up to 128 Inputs and 128 Outputs Supports QuickConnect

More information

TOP SMART intelligent position monitor:

TOP SMART intelligent position monitor: TOP SMART intelligent position monitor: Air Torque also manufacture the Top smart intrinsically safe compact valve position monitor with integrated diagnostics, self calibrating limit switches, integrated

More information

Combustion ion chromatography with a Dionex Integrion HPIC System

Combustion ion chromatography with a Dionex Integrion HPIC System TECHNICAL NOTE Combustion ion chromatography with a Dionex Integrion HPIC System Authors Daniel Khor,1 Jeff Rohrer,1 Adelon Austin,2 Mark Manahan2 1 Thermo Fisher Scientific, Sunnyvale, CA, USA; 2COSA

More information

HPLC System S 600 Series

HPLC System S 600 Series HPLC System S 600 Series The Sykam HPLC System S 600 Series is intended for the ambitious analyst. The system configuration is highly variable and several upgrade options make this system suitable for

More information

1. INTERNATIONAL OVERVIEW. 1.0 Area and population. population (1,000) area

1. INTERNATIONAL OVERVIEW. 1.0 Area and population. population (1,000) area 1.0 Area and population area population (1,000) km 2 2000 2010 2018 1 inhabitants per km 2 Belgium 30,530 10,251 10,920 11,443 375 Germany 357,380 82,212 81,777 82,952 232 Estonia 45,230 1,397 1,331 1,315

More information

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS Rev 1 3/6/2004 Selerity Technologies Inc. www.selerity.com SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS What is supercritical fluid chromatography? A chromatographic technique in

More information

Agilent Valve Kit G4231A/B and G4232A/B/C

Agilent Valve Kit G4231A/B and G4232A/B/C Agilent Valve Kit G4231A/B and G4232A/B/C Instructions Contents Typical Applications 2 Dual column selection (G4231A/B or G4232A/B/C) 2 Sample enrichment and sample cleanup (G4231A/B or G4232A/B/C) 2 Alternating

More information

Global Dialysis - Cost per Dialysis Session

Global Dialysis - Cost per Dialysis Session Cost per session for dialysis where a cost has been entered for a dialysis centre. All costs have been converted to US Dollars as at 19 May 2006. Note that the costs do not state what is covered by a dialysis

More information

QuickSplit TM Flow Splitters

QuickSplit TM Flow Splitters Quick TM Flow ters Quick TM Flow ter Features: Fluid resistor technology eliminates tedious adjustments to capillary tubing for split ratio optimization ratios are stable and reproducible, and not affected

More information

Keysight Technologies Battery Temperature Profiling While Charging and Discharging. Application Note

Keysight Technologies Battery Temperature Profiling While Charging and Discharging. Application Note Keysight Technologies Battery Temperature Profiling While Charging and Discharging Application Note 02 Keysight Battery Temperature Profiling While Charging and Discharging - Application Note Why is Temperature

More information

Packing Data Installation Instructions

Packing Data Installation Instructions Packing Data PK Installation Instructions This document describes how to install a Dataliner DL10 display in a panel or enclosure and connect power. European Union Directive Compliance wiring and safety

More information

OECD unemployment rate falls to 6.0% in March 2017

OECD unemployment rate falls to 6.0% in March 2017 OECD unemployment rate falls to 6.0% in March 017 The OECD unemployment rate for the population as a whole fell by 0.1 percentage point, to 6.0%, in March 017. Across the OECD area, 37.5 million people

More information

Totally Automated Method for the Determination of Sudan Dyes in Food via On-Line Filtration, SPE and HPLC Analysis

Totally Automated Method for the Determination of Sudan Dyes in Food via On-Line Filtration, SPE and HPLC Analysis Totally Automated Method for the Determination of Sudan Dyes in Food via On-Line Filtration, SPE and HPLC Analysis Joan M Stevens, Ph.D., Mark Crawford and Greg Robinson Single Presenter Name Abstract:

More information

Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D

Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Instructions Agilent Valve Kit G4231A/C and G4232C/D - Instructions This technical note describes the installation and application of the Agilent

More information

CSM-K. high capacity clean steam generator

CSM-K. high capacity clean steam generator CSM-K high capacity clean steam generator High capacity clean steam generator Spirax Sarco, the world s leading steam system specialist, has combined modern technology with package design expertise to

More information

FITTINGS, LOOPS, AND VALVE ACCESSORIES

FITTINGS, LOOPS, AND VALVE ACCESSORIES FITTINGS, LOOPS, AND VALVE ACCESSORIES Fittings Rheodyne offers a variety of fittings. They differ in material of construction, tubing size, and pressure rating. Materials/Biocompatibility. 316 stainless

More information

Performance Derby: MSCI Regions/Countries Earnings & Revenues Growth 2019E / 2018E / 2017A

Performance Derby: MSCI Regions/Countries Earnings & Revenues Growth 2019E / 2018E / 2017A Performance Derby: MSCI Regions/Countries Earnings & Revenues Growth 2019E / 2018E / 2017A February 28, 2018 Dr. Ed Yardeni 516-972-7683 eyardeni@yardeni.com Joe Abbott 732-497-5306 jabbott@yardeni.com

More information

[ CARE AND USE MANUAL ] WATERS ADVANCED PURIFICATION COLUMNS I. INTRODUCTION CONTENTS I. INTRODUCTION II. USING THE COLUMN III. CARE AND MAINTENANCE

[ CARE AND USE MANUAL ] WATERS ADVANCED PURIFICATION COLUMNS I. INTRODUCTION CONTENTS I. INTRODUCTION II. USING THE COLUMN III. CARE AND MAINTENANCE WATERS ADVANCED PURIFICATION COLUMNS I. INTRODUCTION Waters Advanced Purification (AP) series of glass columns are constructed of biocompatible glass and polymeric materials and can be easily used with

More information

Analysis of Petroleum Fractions by ASTM D2887

Analysis of Petroleum Fractions by ASTM D2887 Analysis of Petroleum Fractions by ASTM D2887 Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 2582 Key Words Simulated distillation, D2887, TRACE TR-SimDist Abstract ASTM

More information

Agilent 1200 Series Valves

Agilent 1200 Series Valves Agilent 1200 Series Valves User Manual Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or by any means (including electronic storage and

More information

Made UNDER high pressure

Made UNDER high pressure Made UNDER high pressure Made FOR high pressure THE CONTRINEX HIGH-PRESSURE-RESISTANT INDUCTIVE SENSOR SERIES Contrinex is a leading international high-tech sensor company with headquarters in Switzerland.

More information

BP Statistical Review of World Energy June 2017

BP Statistical Review of World Energy June 2017 BP Statistical Review of World Energy June 217 Primary energy 8 Consumption 8 Consumption by fuel 9 66 th edition Primary energy Consumption* Growth rate per annum Million tonnes oil equivalent 26 27 28

More information