Technical Service Manual. MaxVenturi Oxygen Delivery Device Rev. H

Size: px
Start display at page:

Download "Technical Service Manual. MaxVenturi Oxygen Delivery Device Rev. H"

Transcription

1 2305 South 1070 West Salt Lake City, Utah (801) Fax (801) Toll Free USA maxtec Technical Service Manual MaxVenturi Oxygen Delivery Device R211M01 Rev. H

2 CONTENTS DESCRIPTION PAGE SECTION 1: 1. Introduction... 3 SECTION 2: 2. Diagnostics Error Codes Flow Diagnostic Procedure Background Equipment Needed Configuration Test Procedure Oxygen Flow Diagnostic Procedure MaxVenturi Leak Test Procedure... 8 SECTION 3 3. Troubleshooting... 9 SECTION 4 4. Replacement Procedures Front Cover/Circuit Board Replacement Flow Meter Cleaning Procedure Cable Replacement Procedure Valve Cartridge Replacement Procedure Orifice Replacement Procedure Battery Replacement Procedure Conic Replacement Procedure O-Ring Maintenance Procedure SECTION 5 5. Spare and replacement parts REV. H 2

3 INTRODUCTION 1. INTRODUCTION NOTICE Repair of this equipment must be performed by a qualified service technician experienced in repair of portable handheld medical equipment. The MaxVenturi series of products has been designed for maximum reliability, stable performance and low maintenance. Send Maxtec products/material in need of factory repair to: Maxtec Customer Service Line: HOW TO USE THIS MANUAL Maxtec Customer Service Department 2305 South 1070 West Salt Lake City, Utah (Include RMA number) This manual provides service technicians with information needed to maintain and service the MaxVenturi series of products. The manual is divided into several sections. The DIAGNOSTICS section describes self-test and service diagnostics for checking system functions. The TROUBLESHOOTING section provides a guide to assist service technicians in locating the source of problems. The REPLACEMENT PROCEDURES section contains instruction for removal and replacement of assemblies that are considered field-replaceable. GENERAL TROUBLESHOOTING GUIDELINES Troubleshooting the MaxVenturi series of products should always begin by assessing the problem with the user who experienced the trouble. This may eliminate unnecessary troubleshooting steps. Once a general problem has been identified, refer to the trouble shooting guidelines in Section 3 to determine the proper corrective action. After a component has been replaced, verify that the unit is operating properly by running the appropriate diagnostic procedure. CAUTION The MaxVenturi contains electronic components that are susceptible to damage by electrostatic discharge. When disassembling the device, work at a static controlled workstation; wear a static control wrist strap to discharge accumulated static charges from you and any tool you are using. Handle the circuit board by the case. Use antistatic containers for transporting circuit boards. Note: Some illustrations and pictures may be for representation only. However, the repair operations remain the same as set forth in this service manual. REV. H 3

4 TROUBLESHOOTING 2.0 DIAGNOSTICS 2.1 ERROR CODES The MaxVenturi series of products contains internal programming designed to indicate common errors that occur during the use and maintenance of the device. The analyzer has a self-test feature built into the software to detect faulty calibrations, oxygen sensor failures, and low battery voltage. When errors occur in the function of the oxygen analyzer assembly portion of the device, error codes are displayed to signify to the end user or service technician the type of error and corrective action(s) needed to restore proper function of the device. Listed below are the types of error codes displayed, the types of error they may signify, and the corrective action(s) needed to restore proper function. E02: No signal from sensor Possible Errors: No sensor attached. No signal is being sent through the cable to the analyzer. Corrective action: Disconnect and reconnect the sensor. Unit should perform an auto calibration and should read 20.9%. Should the unit fail to calibrate properly, see troubleshooting section of this manual. E03: No valid calibration data available Possible Errors: Calibration is occurring when oxygen levels are not at room air or 100% oxygen. Unit is not in thermal equilibrium. Corrective action: Make sure the unit has reached thermal equilibrium. Perform calibration as described in user manual. E04: Battery below minimum operating voltage Possible Errors: Battery power is low. Corrective action: Replace the batteries as described in replacement procedures section of this manual. CAL Err St: O2 Sensor reading is not stable Corrective action: Wait for displayed oxygen reading to stabilize when calibrating the device at 100% oxygen. Wait for unit to reach thermal equilibrium. (Note this can take up to one half hour if the device is stored in temperatures outside its operating temperature range.) CAL Err lo: Sensor voltage too low Corrective action: Repeat calibration routine as described in user manual. If unit repeats this error more than three times, refer to troubleshooting section of this manual. REV. H 4

5 TROUBLESHOOTING CAL Err hi: Sensor voltage too high Corrective action: Repeat calibration routine as described in user manual. If unit repeats this error more than three times, refer to troubleshooting section of this manual. CAL Err Bat: Battery voltage is too low to recalibrate. Corrective Actions: Replace batteries (See replacement procedures section 4.6.) Repeat calibration routine to insure proper function. 2.2 FLOW DIAGNOSTIC PROCEDURE Background The MaxVenturi series of products uses the venturi effect to entrain room air by using a stream of oxygen flowing perpendicular to the entrainment port. The output of mixed oxygen and air can be affected by various factors the velocity of the oxygen exiting the orifice (affected by pressure and orifice size), the shape and position of the diffuser, the length and diameter of tubing attached to the exit port of the device, the type of user interface, and the presence and type of inlet filter. A test system was developed by Maxtec to measure the flow characteristics of the MaxVenturi device and to detect product failures or defects. Note: The flow diagnostic procedure must be performed with this test system setup to ensure correct diagnostic data Equipment Needed 50 psi supply source of medical grade oxygen TSI 4000 series flow meter Model 4040 E see with inlet filter (Filter 22mm ISO Taper) PC used to fix settings of the TSI 4000 series flow meter. 22mm dia. corrugated tubing 22 length. Flow Inspection Tool Maxtec Part #: MT130 (0-60 LPM MaxVenturi) Flow Inspection Tool Maxtec Part #: MT (0-120 LPM High Flow MaxVenturi) Inlet filter Maxtec Part #: RP34P Configuration TSI Configuration Using the supplied software and PC modify the settings of the TSI4040 flow meter to measure Air at STP. Ensure that this device is maintained at this setting for the duration of use Test Bench Configuration On the outlet port of the TSI 4040 flow meter, attach the MT130 for the 0-60 LPM MaxVenturi or the MT for the LPM High Flow MaxVenturi. Attach the 22mm diameter corrugate tubing onto the inlet port of the TSI4040 flow meter. It is important to use the inlet filter (Filter 22mm ISO Taper) supplied with the TSI4040 to ensure accurate flow measurements. REV. H 5

6 TROUBLESHOOTING Test Procedure 1. Insert the disposable inlet filter (RP34P02) onto the MaxVenturi to be tested. Check that the disposable filter is not excessively dirty or moist. Replace filter when it is visibly contaminated with dirt or moisture. 2. Mount the MaxVenturi to a pole or other device. Attach the oxygen supply line to the oxygen inlet fitting on the back of the MaxVenturi. 3. Turn oxygen % control valve completely off. 4. Open on/off valve on the back of the MaxVenturi. Turn on main oxygen supply valve (from the O 2 tank). 5. Slowly open the flow control valve. Adjust the flow control valve until the MaxVenturi flow meter reaches 20 LPM. Be sure to read the flow meter with your eyes vertically level and to the center of the ball. 6. Turn on the TSI flow meter if it is not already on. Connect the 22mm hose to the outlet of the MaxVenturi. 7. Read the flow on the TSI flow meter. Check Venturi Testing Chart (see below) to verify that the flow reading falls within acceptable range. 8. Adjust the flow control valve until the flow meter reaches 35 LPM (or 60 LPM for the High Flow MaxVenturi) and verify the TSI flow meter is reading within the acceptable range. It is important that the O 2 pressure remain at 50 psi. If the pressure drops due to the increased flow adjust the pressure back up to 50 psi. 9. Continue adjusting the flow control valve to the other flow settings and verify the TSI flow meter is reading with the acceptable range. 10. Failure occurs when one or more flow readings fall outside the acceptable range. Note: When failure occurs perform the Oxygen Flow Diagnostic Procedure found in Section and then proceed to the Troubleshooting section of this document LPM MaxVenturi Total Flow Testing Chart Flow Setting Acceptable Range 20 LPM 15.0 LPM 20.0 LPM 35 LPM 28.9 LPM 35.0 LPM 50 LPM 39.1 LPM 47.0 LPM LPM High Flow MaxVenturi Total Flow Testing Chart Flow Setting Acceptable Range 20 LPM 13.3 LPM 16.2 LPM 60 LPM 43.9 LPM 53.4 LPM 80 LPM 61.6 LPM 75.3 LPM 120 LPM 99.7 LPM LPM REV. H 6

7 TROUBLESHOOTING Oxygen Flow Diagnostic Procedure The Oxygen Flow diagnostic procedure should be performed anytime the MaxVenturi device has a failure at one or more of the flow rates described in section Insert the disposable inlet filter (RP34P02) onto the MaxVenturi being tested. Check that the disposable filter is not excessively dirty or moist. Replace filter when it is visibly contaminated with dirt or moisture. 2. Mount the MaxVenturi to a pole or other device. Attach the oxygen supply line to the oxygen inlet fitting on the back of the MaxVenturi. 3. Turn oxygen % control valve completely off. 4. Open on/off valve on the back of the MaxVenturi. Turn on main oxygen supply valve (from the O2 tank). 5. Plug the inlet of the disposable filter. 6. Slowly open the flow control valve. Adjust the flow control valve until the MaxVenturi flow meter reaches 20 LPM. Be sure to read the flow meter with your eyes vertically level and to the center of the ball. 7. Turn on the TSI flow meter if it is not already on. Connect the 22mm hose to the outlet of the MaxVenturi. 8. Read the flow on the TSI flow meter. Check Oxygen Flow Testing Chart (see below) to verify that the flow reading falls within acceptable range. 9. Adjust the flow control valve until the flow meter reaches 35 LPM (or 60 LPM for the High Flow MaxVenturi) and verify the TSI flow meter is reading within the acceptable range. It is important that the O 2 pressure remain at 50 psi. If the pressure drops due to the increased flow adjust the pressure back up to 50 psi. 10. Continue adjusting the flow control valve to the other flow settings and verify the TSI flow meter is reading with the acceptable range. 11. Failure occurs when any flow reading falls outside the acceptable range. *Proceed to Troubleshooting Section for further diagnosis LPM MaxVenturi Oxygen Flow Testing Chart Flow Setting Acceptable Range 20 LPM 1.95 LPM 2.36 LPM 35 LPM 3.59 LPM 4.34 LPM 50 LPM 5.29 LPM 6.38 LPM LPM High Flow MaxVenturi Oxygen Flow Testing Chart Flow Setting Acceptable Range 20 LPM 4.47 LPM 4.70 LPM 60 LPM LPM LPM 80 LPM LPM LPM 120 LPM LPM LPM REV. H 7

8 TROUBLESHOOTING MaxVenturi Leak Test Procedure 1. Attach oxygen supply line to the oxygen inlet fitting on the back of the MaxVenturi being tested. 2. Close all flow valves. Plug the sensor port and the outlet port with plugs or caps (A new device is shipped with red caps provided). Insert a barbed adapter into the air inlet. Place a tube on the barbed adapter. Place the end of the tube into a container of water. 3. Open all the valves supplying oxygen to the MaxVenturi. Watch for bubbles in the container of water. The maximum number of allowable bubbles is 1 bubble per minute. 4. If the unit leaks bubbles into the container of water, loosen the oxygen valve, then retighten, and repeat leak test. Also check flow valve for leaks in the same manner. If any valve continues to leak, remove the valve and replace the o-ring (RP56P02-901) on the tip of the valve stem if necessary. 5. If the unit continues to leak, return unit to Maxtec Service department. REV. H 8

9 TROUBLESHOOTING 3.0 TROUBLESHOOTING This section contains information to assist the Service Technician in determining possible faults affecting the performance of the MaxVenturi device. The MaxVenturi device is comprised of several subassemblies. For sake of troubleshooting; the MaxVenturi analyzer is the portion of the device that contains the printed circuit board, the plastic cover, the overlay, and any other parts permanently attached to those components. This subassembly is only replaceable and cannot be repaired. Another subassembly of the MaxVenturi is the sensor. This subassembly is not repairable in the field, such that if failures occur with the sensor, it will need to be replaced. The following section is a list of problems that may occur and possible causes of that failure. PROBLEM: Flow does not reach 55 LPM (120 LPM for the High Flow MaxVenturi) even with the flow valve wide open. POSSIBLE CAUSE(S): The inlet pressure of the oxygen supply may be too low. Check the pressure of the gas entering the device. Make sure the inlet pressure falls between psi. (Consult user manual for flow ranges at pressures outside of this range). If the inlet pressure is within range, remove the device from its fixture and tilt the unit back and forth to see if the ball-float rolls without impedance. If it sticks or hesitates, see repair procedure 4.2. If the inlet pressure is within range, and the ball-float rolls without impedance, the MaxVenturi orifice may be plugged or defective. To check orifice and/or repair it see repair procedure 4.5. PROBLEM: The oxygen level on the display does not reach 100%, even with the oxygen control knob fully open. POSSIBLE CAUSE(S): The device needs to be calibrated. If the unit is being used with oxygen concentrations close to 100%, it is best to calibrate the device at 100%. Refer to the calibration section of the user manual to follow the 100% oxygen calibration routine. If the oxygen level on the display does not reach 95%, check that the correct needle valve is in the O2% Control Valve. See repair section 4.4. PROBLEM: The oxygen level displayed does not go as low as the level indicated in the specifications, even with the oxygen control knob completely closed. POSSIBLE CAUSE(S): The device needs to be calibrated. Refer to calibration section of the user manual. Ensure that the humidifier, patient circuit, and patient interface being used are on the approved list of disposables. Ensure that the room air inlet filter is not soiled or wet. Replace the filter if needed. Make sure that the nasal canula or other tubing is not occluded. PROBLEM: Unit fails to energize. POSSIBLE CAUSE(S): The batteries are exhausted. Replace batteries following the repair procedure found in Section 4.6. The batteries are oriented incorrectly. Reinstall the batteries in the correct orientation. The batteries are not contacting the terminals. Check for loose connection. The analyzer screen is damaged. See replacement procedure 4.1. REV. H 9

10 TROUBLESHOOTING PROBLEM: The analyzer gives an E02 code when a sensor is connected to the cable. POSSIBLE CAUSES: The sensor has expired. Check that voltage from sensor is present. Testing the sensor on another unit may indicate if the sensor is broken or if the analyzer is defective. The sensor cable is broken. Disconnect, and then reconnect the sensor. If the error persists, check for a short in the sensor cable. The cable can be broken when excessive force is used to pull the sensor out of its port by pulling on the cord instead of the sensor. To replace sensor cable see replacement procedure 4.3. PROBLEM: The MaxVenturi unit does not pass the Flow Diagnostic Procedure at one or more flow levels AND the MaxVenturi unit does pass the Oxygen Flow Diagnostic Procedure at one or more flow levels. POSSIBLE CAUSES: Note - possible causes are listed from most to least likely. Proceed to section 4.5 to replace the orifice. (Section 4.5 applies only to new models). The diameter of the orifice is incorrect due to damage, manufacturing defect, or the wrong orifice for the version of MaxVenturi is used. Check the diameter of the orifice using pin gauges. The diameter of the 0-60 LPM MaxVenturi orifice should be.020 and not less.019 and not more than.021. The diameter of the LPM High Flow MaxVenturi orifice should be.046 and not less.045 and not more than.047 Use a smaller pin gauge first and increase the size of the pin gauge until the pin just barely slides through the hole and the next largest pin gauge does not fit. If the size of the hole is too small or too large follow the replacement procedure in section 4.5. (Section 4.5 applies only to new models). The flow meter has a leak around its seals. This can be caused by a missing or pinched o-ring. Visually check to make sure the o-ring has a seal around all its edges. If the o-rings appear to be correctly positioned, connect an oxygen source to the unit and use snoop or soapy water around the perimeter of the flow meter to check for leaks. The inner diameter of the diffuser tube is incorrect due to damage or manufacturing defect. Check the inner diameter of the tube with pin gauges. The MaxVenturi smallest tube diameter should be.252 ±.005. Start with smaller pin gauges that fit and proceed until the pin just barely fits in the hole and the next larger pin gauge does not fit. Also check the if the diffuser tube has moved or is loose. The air inlet filter is occluded, dirty, damaged, or wet. Visually inspect it for occlusion, dirt, damage or moisture. Replace if necessary. PROBLEM: The MaxVenturi unit does not pass the Flow Diagnostic at one or more flow levels AND The MaxVenturi unit does not pass the Oxygen Flow Diagnostic Procedure at one or more flow levels. POSSIBLE CAUSES: Note Possible causes are listed from most to least likely. A foreign object has partially or fully occluded the opening of the orifice. It is important that any time the orifice is suspect that it be checked for foreign material. This procedure is explained in the orifice repair section 4.5. (Section 4.5 applies only to new models). If a particle is found in the hole remove it by inserting a small pin gauge into the hole and then blowing out the cavity with air. Be sure the particle is removed before reinstalling the flow control valve. The orifice has a manufacturing defect. Replace the orifice as described in section 4.5. (Section 4.5 applies only to new models). The flow meter has a manufacturing defect. This can best be detected by replacing the flow meter with a flow meter that is known to be correct and passes the Flow Diagnostic Procedure. Replacing the flow meter with a good one and performing the Flow Diagnostic Procedure again on the unit will either determine the flow meter is defective or rule it out. PROBLEM: Air does not come out the outlet of the MaxVenturi when the on/off valve is open, the flow control valve is partially open, and an oxygen source is connected. REV. H 10

11 TROUBLESHOOTING POSSIBLE CAUSES: The oxygen source is not on. Check the valve on the cylinder or any other supply valves. The oxygen source is depleted. Check the pressure of the gas on the cylinder. The connections between the oxygen cylinder and the MaxVenturi have a leak. Check for leak and tighten connections if necessary. PROBLEM: Gas flows from the MaxVenturi, when the on/off valve is open but the flow control valve is closed as well as the O2% control knob. POSSIBLE CAUSES: An o-ring is missing from either of the two control valves. Remove each of them and check that there are two o- rings on each of them. Sometimes the smaller o-ring will come off of the valve and be lodged inside the opening. Remove it from the opening with a pair of tweezers and reapply it. When the valves are reinserted, be sure to check that the o-ring is seated properly on the groove. An o-ring is damaged on either of the two control valves. Remove each of them and check for damage. A needle valve is broken. Check that the control valves needles are not broken or that the end of the needle valve has come loose. PROBLEM: The ball in the flow meter is jumpy. POSSIBLE CAUSE: A particle is on the ball in the flow meter or in the tube of the flow meter. To clean the flow meter, see Section 4.2. PROBLEM: The ball in the flow meter suddenly drops and won t return to higher flow levels. POSSIBLE CAUSE: The orifice has become occluded with a foreign object. See the procedure to check for debris in Section 4.5. PROBLEM: One or both of the buttons on the MaxVenturi have to be pushed really hard to work. POSSIBLE CAUSE: The foam sticker under the overlay is missing or has become dislodged. Circuit board buttons have become warn out. Circuit board has broken out of cover. To repair, replace the front cover/circuit board as described in section 4.1. PROBLEM: The battery symbol appears on the LCD screen. POSSIBLE CAUSE: The batteries are low in power. Replace the batteries see section 4.6. Comment [MS1]: PROBLEM: The display reads CAL Err lo and the user has tried to calibrate the sensor 3 times. REV. H 11

12 TROUBLESHOOTING POSSIBLE CAUSE: The Max250 Sensor is defective or dying. Replace the sensor with a new sensor if available. Calibrate the unit. This will determine if it is the sensor or the circuit board (analyzer). If no sensor is available, then do the following to determine if the sensor is bad. 1. Disconnect the sensor from the cable of the MaxVenturi. 2. Use a multi-meter to measure the mv output of the sensor. Contact one probe of the multi-meter to the center post of the connector and the second probe to the outside edge of the connector. The mv reading should be approximately mv at standard pressure (Sea level). 3. To determine what the mv output of the user s sensor is at standard pressure, use the following equation where P(u) is the barometric pressure at the location of the sensor in mbar, V(s) is the output voltage of the sensor in mv, and V(c) is the corrected output voltage in mv. V(c) = V(s) X 1013 mbar / P(u) 4. V(c) should fall within the mv range. If V(c) falls below that range then the sensor needs to be replaced. Contact Maxtec Customer Service for a replacement sensor. If the V(c) falls within the correct range, reconnect the sensor and attempt to calibrate the unit again. Be sure to follow the correct procedure as outlined in the user manual for the MaxVenturi. If the problem persists contact customer service for a replacement MaxVenturi Analyzer Assembly and follow Section 4.1 to replace the assembly. PROBLEM: The display reads CAL Err hi and the user has tried to calibrate the sensor 3 times at 100% oxygen. POSSIBLE CAUSE: The Max250 Sensor is defective. Replace the sensor with a new sensor. Calibrate the unit at 100% oxygen following the user manuals instructions. This is done by completely blocking off the inlet of the MaxVenturi and leaving the sensor in its port. Turn the Flow control valve on and wait for the gas to equilibrate. Push the CAL button and hold it for 3 seconds or until the display reads CAL. If the unit calibrates then it is likely that the first sensor is defective. Contact customer service for a replacement Max250 sensor. If the unit fails to calibrate it is likely that the MaxVenturi Analyzer Assembly is damaged or defective. If the problem persists contact customer service for a replacement MaxVenturi Analyzer Assembly and follow Section 4.1 to replace the assembly. *If a replacement sensor is not readily available then the following procedure can be used to determine if the sensor is defective. 1. Disconnect the sensor from the cable of the MaxVenturi, leave the sensor in the sensor port. 2. Connect a 100% oxygen supply to the oxygen inlet of the MaxVenturi. 3. Plug the air inlet port, open the on/off valve of the unit and turn up the flow control valve. 4. Use a multi-meter to measure the mv output of the sensor. Contact one probe of the multi-meter to the center post of the connector and the second probe to the outside edge of the connector. The mv reading should be approximately mv at standard pressure (See level). 5. To determine what the mv output of the user s sensor is at standard pressure, use the following equation where P(u) is the barometric pressure at the location of the sensor in mbar, V(s) is the output voltage of the sensor in mv, and V(c) is the corrected output voltage in mv. V(c) = V(s) X 1013 mbar / P(u) 6. V(c) should fall within the mv range. If V(c) falls below that range then the sensor needs to be replaced. Contact Maxtec Customer Service for a replacement sensor. REV. H 12

13 4.0 REPLACEMENT PROCEDURES The following section contains repair procedures that can be performed in the field with the appropriate replacement parts. Any repair procedure not included in this section should be performed by the manufacturer. Only repairs shown here should be performed in the field, by an authorized repair technician. All other repairs should be performed by the manufacturer. *VERY IMPORTANT NOTE: Check to see if you have an old or new MaxVenturi manifold. This will determine whether you also need revision D of this document for an old manifold (Rev H). If there is a 2 scribed on the back of the manifold below the pole clamp then you have a new manifold (Rev J), if no number is present then you have an old manifold (Rev H). If the unit being repaired is a Revision H, the orifice repair procedure described in section 4.6 is not field repairable and should be returned to Maxtec Service Department for Repair or an authorized service facility. To become an authorized service facility, contact Maxtec Technical Service Department at FRONT COVER/CIRCUIT BOARD REPLACEMENT For front cover/circuit board replacement use: Maxtec part # R211P13 (MaxVenturi). Follow the steps provided below to replace the front cover assembly. For replacement parts contact the Maxtec Service Department at MAXTEC or Step 1. Remove the MaxVenturi Analyzer of the MaxVenturi from the main body by removing the four screws on the back. Use a #2 Phillips screwdriver (a screwdriver is provided with each unit). Step 2. Detach the two electrical connectors from the back of the printed circuit board (PCB). Push in the center of each connector to open the latch of each connector. REV. H 13

14 Step 3. Remove the replacement MaxVenturi Analyzer assembly from its packaging. Ensure proper care is taken to prevent electrostatic discharge which will damage the printed circuit board. Step 4. Attach the battery cable (larger connector) to the appropriate connector (larger connector) on the PCB of the replacement MaxVenturi Analyzer. Step 5. Attach the sensor cable (smaller connector) to the appropriate connector (small connector) on the PCB on the replacement MaxVenturi Analyzer. REV. H 14

15 Step 6. Align the front cover in the upright position on the front of the unit and install the 4 screws through the back to secure the replacement MaxVenturi Analyzer. Ensure that the wires do not get pinched between the front cover and the main body of the device. Step 7. Check that unit powers up and reads correctly. REV. H 15

16 4.2 FLOW METER CLEANING PROCEDURE For replacement parts contact the Maxtec Service Department at MAXTEC or Step 1. Using a 5/32 Allen (hex) wrench, remove the lowest stainless steel set screw holding the flow meter to the main body of the device. Step 2. Using a ball-end of a 5/32 Allen (hex) wrench, remove the top stainless steel set screw holding the flow meter. (The wrench must be held at angle. The screw will not come out of the hole, but the flow meter will come off the front.) Step 3. Two o-rings sit in grooves on the front of the unit. It is suggested that these O-rings be replaced during maintenance. Contact Maxtec Service Department for replacement o-rings. REV. H 16

17 Step 4. Using a 6 mm Allen (hex) wrench, remove the metal plug from the flow meter by twisting counterclockwise. Step 5. Use a pair of tweezers or needle nose pliers to remove the plastic plug inside the flow meter. Step 6. Tilt the flow meter to remove the ball. REV. H 17

18 Step 7. Remove foreign particles in the flow meter using a long cotton applicator. If particles remain, use of rubbing alcohol may be required. Blow out any debris with an air hose if necessary. Step 8. Insert the ball back into the hole. Reinsert the plastic plug inside the flow meter. Reattach the metal plug of the flow meter. Step 9. Check to make sure the ball rolls freely back and forth in the flow meter. REV. H 18

19 Step 10. Place two new o-rings in the manifold grooves on the front of the unit. Step 11. Carefully line up the body of the device over the flow meter so the grooves for the o-ring line up with the o- rings on the flow meter. Step 12. Fasten the flow meter in place with the two stainless steel screws removed in steps 1 and 2. It works best to start both screws before tightening both of them. REV. H 19

20 4.3 CABLE REPLACEMENT PROCEDURE The following procedure should be followed when the cable to the sensor is defective. Please read the procedure completely before beginning. The cable replacement kit part number is R211P For replacement parts contact the Maxtec Service Department at MAXTEC or Step 1. Remove the MaxVenturi Analyzer Assembly from the front of the unit by removing the (4) mounting screws in the back. Step 2. Disconnect the two wire connections by pressing in the center of the connecter and pulling lightly on the wires. REV. H 20

21 Step 3. Remove the batteries from the unit to allow for more room to work. (Battery holder may be different than what is shown) Step 4. Push on the cable from the back to break any glue that is sticking to the cable and the rubber strain relief. Step 5. Cut the wire between the zip tie and the rubber strain relief. REV. H 21

22 Step 6. Remove the connector from the end of the red and black wires by inserting a small pin into the holes and depressing the lever of each pin. Step 7. Carefully pull the wire out from the rubber strain relief through the back of the device. Step 8. Remove the replacement cable from its packaging and insert the stripped end of the wires into the rubber strain relief in the back of the unit. Pull on the cable till approximately one inch of the shielded cable protrudes into the main cavity of the unit. REV. H 22

23 Step 9. Wrap the end of the cable with a zip tie and tighten the zip tie. Step 10. Clip off the excess length of the zip tie and pull on the cable till the zip tie is against the rubber grommet inside the main cavity. Step 11. Apply a few drops of glue on the zip tie next to the cable. REV. H 23

24 Step 12. Attach two clips using wire crimpers to the red and black wires. (Clips may be pre-attached by service department at Maxtec upon request.) Step 13. Insert the clips into the connector removed in step 6. The red wire should be inserted into the hole with the arrow on the connector. (See photo.) Step 14. Place the batteries back into the unit. Reconnect the power cable connector and the sensor cable connector to the back of the PCB. REV. H 24

25 Step 15. Connect a sensor to the sensor cable. Turn the unit on and verify that the unit powers up and that a reading shows up on the LCD screen. Step 16. Align the MaxVenturi Analyzer Assembly to the edges of the main body and reattach it using the 4 screws removed in step 1. REV. H 25

26 4.4 VALVE CARTRIDGE REPLACEMENT PROCEDURE The following procedure should be performed when a unit has a broken or defective flow control knob/cartridge or a broken O2% control knob/cartridge. Note: Replacement valve cartridges will not appear or assemble the same as previous versions of the knob, but will function the same. For replacement parts contact the Maxtec Service Department at MAXTEC or Step 1a. If you have the old style knob as seen below, pull the knob off. OR OR Step 1b. If you have new style knob as seen below, first remove the cap on the knob and then remove the knob by loosening the retaining nut with a 7mm nut driver while holding the knob. Pull knob to remove. REV. H 26

27 Step 2. Using a ½ open ended wrench, loosen and remove the valve cartridge. Be careful not to scratch the venture body. Step 3. O 2 control cartridge (R211P for 0-60 LPM MaxVenturi, or R211P for LPM MaxVenturi) and flow control cartridge (R211P30-001) are different. Check to make sure the replacement cartridge matches the cartridge being replaced. REV. H 27

28 Step 4. Before installing cartridge remove the knob as noted in step1b. Use wrench to tighten till the fitting is just barely snug. (Do not Over tighten the nut.) O-ring damage caused by over tightening can be detected during a leak test. Step 5. Reinstall the black knob by placing the knob over the valve stem and tightening the retaining nut with a 7mm nut driver while holding the knob. Step 6. Install the cap on the new knob. Be sure to align the wide gap in the cap with the two protruding ribs inside the cap. REV. H 28

29 Step 7. Verify that the new valve cartridge opens and closes properly. REV. H 29

30 4.5 ORIFICE REPLACEMENT PROCEDURE This test should be performed when a leak was detected using step 4.5 or when a close inspection of the orifice indicates that it is faulty or damaged. This replacement procedure is for Revision J of the Manifold and later. See earlier note under 4.0 section heading. For replacement parts contact the Maxtec Service Department at MAXTEC or Step 1. Remove the flow control valve of the MaxVenturi as seen in Section 4.4 steps 1-2. Step 2. IMPORTANT: In order to determine the exact cause of the part failure hold the MaxVenturi so that light is entering the inlet port of the venturi and peer down the orifice opening through the opening where the flow control valve was removed. The opening of the orifice should be circular and devoid of debris. If the opening is circular and free of debris proceed with the rest of this procedure. If there is debris, refer to the Troubleshooting section to see how to remove debris. Then reattach the valves. Step 3. Place the orifice punch out tool down the outlet of the MaxVenturi. Step 5. Use an arbor press to push out the orifice. Apply increasing amounts of force gradually until the part is removed. REV. H 30

31 Step 6. Remove the orifice from the valve cartridge opening by tipping the unit. Use tweezers if necessary. Step 7a. Insert the small o-ring (RP56P02-901) onto the outside of the new orifice. Step 8. Insert the orifice onto the orifice positioning tool (MT152). REV. H 31

32 Step 9. Thread the positioning tool(mt152) (with the orifice) completely into the opening for the valve cartridge of the MaxVenturi. Step 10. Place the device on its end and press the orifice into place using an arbor press. The orifice is seated when the positioning tool knob is fully depressed as shown. Step 11. Remove the orifice insertion tool by twisting the tool counter-clockwise. REV. H 32

33 Step 12. Reinsert the valve cartridge and reinstall the black knob and cap as seen in section 4.4. Step 13. Test the function of the device according to section and perform a leak test according to REV. H 33

34 4.6 BATTERY REPLACEMENT PROCEDURE Step 1. Remove the MaxVenturi Analyzer Assembly from the front of the unit by removing the (4) mounting screws in the back. Step 2. Take out the battery holder. Remove the dead batteries from the holder. A screwdriver or other tool may be used to pry the batteries out. Step 3. Insert new batteries. Ensure that they are oriented correctly following the diagram on the battery holder. REV. H 34

35 Step 4. Reinsert the battery holder on the Velcro strip. Step 5. Connect a sensor to the sensor cable. Turn the unit on and verify that the unit powers up and that a reading shows up on the LCD screen. Verify that the battery symbol is no longer displayed. Step 6. Align the MaxVenturi Analyzer Assembly to the edges of the main body and reattach it using the 4 screws removed in step 1. REV. H 35

36 4.7 CONIC REPLACEMENT PROCEDURE For replacement parts contact the Maxtec Service Department at MAXTEC or Step 1. First, squeeze the base of the conic with large pliers to break glue. Then, Lightly strike the male or female conic with a plastic mallet. Strike on opposite sides of the part until the part comes off. Scrape off excess glue inside the main body where the part was glued in. Step 2. Carefully verify that all glue is removed before proceeding. Step 3. Apply a small amount of glue (Loctite #414) around the edge of the opening for the fitting. Step 4. Position a new fitting in the opening of the hole and insert the fitting into the hole. Strike the part gently with a plastic mallet to ensure the part snaps into place. Step 5. Wipe off any excess glue and allow the glue to dry. REV. H 36

37 4.8 O-RING MAINTENANCE PROCEDURE For replacement parts contact the Maxtec Service Department at MAXTEC or Step 1. Remove the black knob from the flow valve cartridge. Step 2. Use a ½ open ended wrench to remove the valve cartridge. Step 3. Remove the two o-rings from the valve cartridge. Step 4. Replace the two o-rings with new o-rings. O-rings can be purchased from Maxtec. Replacement part numbers are RP56P (smaller o-ring), and RP56P (larger o-ring). REV. H 37

38 Step 5. Reinsert the flow valve cartridge into the manifold using the ½ open ended wrench. Step 6. Repeat the preceding steps to replace the o-rings of the O2% valve cartridge Step 7. Be sure not to switch the two control knobs. See the image below for differentiation. Please see section 5.0 (Spare and Replacement Parts) for additional images of valves: R211P R211P R211P REV. H 38

39 5.0 SPARE AND REPLACEMENT PARTS For replacement parts contact the Maxtec Service Department at MAXTEC or Part Number Description Photo R110P10 Diverter R125P Sensor MAX 250E R211P04 Manifold R211P LPM MaxVenturi Orifice (new style) R211P LPM High Flow MaxVenturi Orifice REV. H 39

40 Part Number Description Photo R211P06 MaxVenturi Diffuser R211P07 Female Conic R211P08 Male Conic R211P Chamber filter R211P Porous oxygen filter R100P Pole Mount Clamp R100P38 Hand Screw for Pole Mount Clamp REV. H 40

41 Part Number Description Photo R100P39 Locking Pole Clamp R100P45 Key For Locking Pole Clamp R211P11 Battery Clip Sub-Assembly R211P Sensor cable Replacement Kit Contains replacement cable, crimp pins, 2 pin connector, zip-tie, and 3g tube of Loctite 414 R211P25 Male Conic Upgrade Kit Contains Male conic and 3 gram tube of Loctite 414 R211P22 NIST Fitting REV. H 41

42 Part Number Description Photo R211P15 On/Off Valve Assembly R211P13 (MaxVenturi) Analyzer Assembly RP06P29 #4-40 screws RP06P30 #10-32 Socket Head Cap Screws RP06P31 #10-32 Pan-head screws R211P Flow Control Valve Replacement Kit (For both the 0-60 and MaxVenturi) REV. H 42

43 Part Number Description Photo R211P O2% Valve Replacement Kit (0-60 LPM MaxVenturi only) R211P O2% Valve Replacement Kit (0-120 LPM High Flow MaxVenturi only) REV. H 43

44 Part Number Description Photo RP31P05 Cable Strain Relief RP34P LPM MaxVenturi Flowmeter RP34P08 RP56P RP56P LPM High Flow MaxVenturi Flowmeter O-rings for flow meter QTY: 2 Orifice o-ring QTY: 1 RP57P48 Metal pin clips QTY: 2 RP57P49 2-wire male connector RP61P01 Zip Tie REV. H 44

45 Part Number Description Photo R211P Female Conic Upgrade Kit Contains Female conic and 3 gram tube of Loctite 414 RP56P Valve Cartridge O-rings (smaller) RP56P Valve Cartridge O-ring (larger) REV. H 45

FOR FUTURE REFERENCE SERIES 93HPS

FOR FUTURE REFERENCE SERIES 93HPS Hypro Series 93HPS Hydraulically Driven Wetseal Multistage Pumps Repair Manual KEEP FOR FUTURE REFERENCE Form L-1578R Rev. A SERIES 93HPS Hydraulically Driven Stainless Steel Multistage Centrifugal Pumps

More information

HYDRAULICS. TX420 & & lower. Hydraulic Tandem Pump Removal. 4. Remove the LH side panel (Fig. 0388).

HYDRAULICS. TX420 & & lower. Hydraulic Tandem Pump Removal. 4. Remove the LH side panel (Fig. 0388). TX420 & 425 240000299 & lower 4. Remove the LH side panel (Fig. 0388). Hydraulic Tandem Pump Removal Note: Cleanliness is a key factor in a successful repair of any hydraulic system. Thoroughly clean all

More information

Installation Manual TWM Performance Short Shifter Cobalt SS/SC, SS/TC, HHR SS, Ion Redline and Saab 9-3

Installation Manual TWM Performance Short Shifter Cobalt SS/SC, SS/TC, HHR SS, Ion Redline and Saab 9-3 Page 1 Installation Manual TWM Performance Short Shifter Cobalt SS/SC, SS/TC, HHR SS, Ion Redline and Saab 9-3 Please Note: It is preferable to park on a flat surface, as you will have to engage and disengage

More information

Robotic Cleaner Troubleshooting Guide. Copyright 2010 Hayward Industries Inc.

Robotic Cleaner Troubleshooting Guide. Copyright 2010 Hayward Industries Inc. Robotic Cleaner Troubleshooting Guide Copyright 2010 Hayward Industries Inc. Table of Contents TigerShark & TigerShark QC Important safety instructions Page 1 Service Tools Page 2 Filter Removal/Cleaning

More information

REPAIR PROCEDURES MANUAL

REPAIR PROCEDURES MANUAL REPAIR PROCEDURES MANUAL PVX Series Vane Pumps A Design Series Step-by-Step Guide to Troubleshooting and Repairing PVX Series Vane Pumps Introduction Thank you for choosing Continental Hydraulics PVX Vane

More information

Open Center Compact Valve Custom Installation Guide Rev A

Open Center Compact Valve Custom Installation Guide Rev A 200-0762-01 Open Center Compact Valve Custom Installation Guide 602-0575-01 Rev A 2014-12 Overview This guide provides information for completing a custom AutoSteer valve installation on wheeled farm vehicles

More information

NOTE: Visit our website at for video repair procedures, under the Tools section.

NOTE: Visit our website at   for video repair procedures, under the Tools section. Repair Instructions Hypro Repair Tools: Tool Box No. 3010-0168 1/4" Allen Wrench No. 3020-0008 Support Bars (2) No. 3010-0064 Port Brush No. 3010-0066 1/16" Allen Wrench No. 3020-0009 Brush Holder No.

More information

CP2000 Ballast - Replacing Thyristor Instruction Sheet

CP2000 Ballast - Replacing Thyristor Instruction Sheet CP2000 Ballast - Replacing Thyristor Instruction Sheet INTRODUCTION These instructions are for the CP2000 Ballast and explain how to: replace the existing 80 Amp Thyristor with a new 100 Amp Thyristor

More information

HIGH PRESSURE CONTROL VALVE PISTON BALANCED

HIGH PRESSURE CONTROL VALVE PISTON BALANCED PISTON BALANCED All Rights Reserved. All contents of this publication including illustrations are believed to be reliable. And while efforts have been made to ensure their accuracy, they are not to be

More information

Service Manual Air Tech Second Stage

Service Manual Air Tech Second Stage Service Manual Air Tech Second Stage Copyright 2002, Cressi-sub Revised 3/2002 2 Air Tech Second Stage Service Manual Contents BEFORE STARTING... 3 DISASSEMBLY... 3 PARTS CLEANING AND LUBRICATION... 9

More information

Auto Sentry-eXP Maintenance. Revised 12/21/07

Auto Sentry-eXP Maintenance. Revised 12/21/07 Auto Sentry-eXP Maintenance Revised 12/21/07 Maintenance Procedures for Auto Sentry exp Bill Dispenser Credit Card Reader Bill Acceptor Bill Dispenser Maintenance Bill Dispenser Problem / Cause Bill Dispenser

More information

ELECTRICAL SYSTEM UPGRADE

ELECTRICAL SYSTEM UPGRADE NEW CONTROLLER & ELECTRICAL SYSTEM UPGRADE FOR DAIRY TECH, INCORPORATED 10, 30 & 60G PASTEURIZERS Parts to Include 2 Wire ties (Nuts) 2 sticky wire mount pads Large Rubber Grommet (for bottom of electric

More information

Part# JL AIR IT UP 4 Tire On Board Air Delivery System. (Requires External Air Source)

Part# JL AIR IT UP 4 Tire On Board Air Delivery System. (Requires External Air Source) Part# 18-1819 JL AIR IT UP 4 Tire On Board Air Delivery System (Requires External Air Source) The most up-to-date instructions always visit www.updownair.com www.updownair.com 833-226-4863 I M P O R T

More information

Throttling Pendulum Gate O-ring Replacement Procedure Rev. A, May 2012

Throttling Pendulum Gate O-ring Replacement Procedure Rev. A, May 2012 Throttling Pendulum Gate O-ring Replacement Procedure Rev. A, May 2012 This procedure covers the replacement of the gate O-ring of the Nor-Cal Pendulum Gate Valve, with the gate valve housing installed

More information

Maintenance Information

Maintenance Information 16573370 Edition 2 February 2014 Air Grinder 99V Series Maintenance Information Save These Instructions Product Safety Information WARNING Failure to observe the following warnings, and to avoid these

More information

Marsh Shipping Supply Co. LLC. Marsh TD2100 Electric Taper Technical Manual

Marsh Shipping Supply Co. LLC. Marsh TD2100 Electric Taper Technical Manual Marsh Shipping Supply Co. LLC Marsh TD2100 Electric Taper Technical Manual 2 A wall-socket must be close to the product and readily accessible. The overall system is protected against overload by the branch

More information

KAM IAS ISOKINETIC AUTOMATIC SAMPLER. User Manual IASMANUAL-0513 KAM CONTROLS, INC Ann Arbor Drive Houston, Texas USA

KAM IAS ISOKINETIC AUTOMATIC SAMPLER. User Manual IASMANUAL-0513 KAM CONTROLS, INC Ann Arbor Drive Houston, Texas USA An ISO 900 certified company TEL + 73 784-0000 FAX + 73 784-000 Email Sales@Kam.com KAM IAS ISOKINETIC AUTOMATIC SAMPLER PER API 8.2, ASTM D477 AND ISO 37 User Manual IASMANUAL-053 3939 Ann Arbor Drive

More information

P-600. Technical Manual. Troubleshooting Repairs Replacements

P-600. Technical Manual. Troubleshooting Repairs Replacements P-600 Technical Manual Troubleshooting Repairs Replacements Table of Contents P-600 Lift Symptoms and Problems Finding the Problem Before Getting Inside 3 Pneumatic Systems 4 Electrical Systems 5 Mechanical

More information

METERING VALVE 2" STEM GUIDED

METERING VALVE 2 STEM GUIDED 2" STEM GUIDED All Rights Reserved. All contents of this publication including illustrations are believed to be reliable. And while efforts have been made to ensure their accuracy, they are not to be construed

More information

5 Removal and replacement

5 Removal and replacement 5 Removal and replacement This chapter describes the removal and replacement of field-replaceable units (FRUs) only. Removal and replacement strategy User-replaceable parts Covers Internal assemblies ENWW

More information

Page 1 of 26 Oteco Inc. Houston, Texas

Page 1 of 26 Oteco Inc. Houston, Texas Page 1 of 26 Page 2 of 26 1.0 OVERVIEW CONTENTS 2.0 INSTALLATION GUIDELINES 2.1 Preferred Valve Orientation 2.2 Pressure Rating and Orientation of Discharge line 2.3 Reaction Forces and Anchoring of Reset

More information

TK08 TUNING KIT TECHNICAL MANUAL FOR 16/26/27/28 SERIES SHOCKS Revised 6/25/14 Tech Line: (952) Fax (952)

TK08 TUNING KIT TECHNICAL MANUAL FOR 16/26/27/28 SERIES SHOCKS Revised 6/25/14 Tech Line: (952) Fax (952) Tech Line: (952) 985-5675 Fax (952) 985-5679 21730 Hanover Ave. Lakeville, MN 55044 www.qa1.net www.facebook.com/qa1motorsports TK08 TUNING KIT TECHNICAL MANUAL CONTENTS UNDER PRESSURE! USE EXTREME CAUTION

More information

GPS AutoSteer System Installation Manual

GPS AutoSteer System Installation Manual GPS AutoSteer System Installation Manual Supported Vehicles Agco Gleaner Combines R65 R66 R75 R76 PN: 602-0288-01-A LEGAL DISCLAIMER Note: Read and follow ALL instructions in this manual carefully before

More information

Service Manual Air Plus Second Stage

Service Manual Air Plus Second Stage Service Manual Air Plus Second Stage Includes XS Series Second Stage Copyright 2002, Cressi-sub Revised 3/2002 2 Air Plus Second Stage Service Manual Contents BEFORE STARTING... 3 DISASSEMBLY... 3 PARTS

More information

LIPPERTCOMPONENTS, INC.

LIPPERTCOMPONENTS, INC. LIPPERTCOMPONENTS, INC. SCHWINTEK INWALL SLIDEOUT SYSTEM OPERATION AND SERVICE MANUAL Contents I. Controls 1-1 System components 1 1-1A versions C1 & C2 2 1-2 Motor wiring harness connections 3 1-3 Extend

More information

MAINTENANCE MANUAL FOR THERMOSTATIC TEMPERATURE REGULATING VALVE TRAC STYLE P

MAINTENANCE MANUAL FOR THERMOSTATIC TEMPERATURE REGULATING VALVE TRAC STYLE P MANUAL NUMBER P-EFS-1 MAINTENANCE MANUAL FOR THERMOSTATIC TEMPERATURE REGULATING VALVE TRAC STYLE P TRAC Regulator Company Inc. 160 South Terrace Avenue Mount Vernon, New York USA 10550-2408 Phone: (914)

More information

Primus Wind Power. AIR Circuit Replacement Instructions

Primus Wind Power. AIR Circuit Replacement Instructions Primus Wind Power AIR Circuit Replacement Instructions Items included with this kit: (see Exploded View p. 3) ITEM ITEM NAME QUANTITY 1 SCREW - SOCKET HEAD - 10-24 X 1-1/2" 4 includes 1 spare 2 O-RING

More information

LAWN SPRINKLER, IRRIGATION PUMP

LAWN SPRINKLER, IRRIGATION PUMP LAWN SPRINKLER, IRRIGATION PUMP MODEL #, SP0P, SP5P, SP20P, EL0P, EL5P, EL20P SAFETY INFORMATION Please read and understand this entire manual before attempting to assemble, operate or install the product.

More information

Discount-Equipment.com

Discount-Equipment.com REQUIRED TOOLS LS Series Remix Shaft Installation Instructions /8", /6", /2" Allen Wrenches Snap Ring Pliers (Light Duty) /" Combination Wrench Loctite #22 Blue /" Socket w/ /8" Ratchet Electric Drill

More information

TECHNICAL SERVICE BULLETIN: TSB

TECHNICAL SERVICE BULLETIN: TSB TECHNICAL SERVICE BULLETIN: TSB2015-007 To: From: Global Service Organization, ichem VELOCITY and iricell Distributors Iris Diagnostics Service and Support Date: 10 June 2015 Subject: STRIP PROVIDER MODULE

More information

Periodic Maintenance 7-1

Periodic Maintenance 7-1 Periodic Maintenance This section provides periodic maintenance guidelines for keeping the InkCenter Refill System in optimal operational condition and detailed descriptions of maintenance procedures.

More information

TK08 TUNING KIT TECHNICAL MANUAL FOR 16/26/27/28 SERIES SHOCKS Revised 6/17/14

TK08 TUNING KIT TECHNICAL MANUAL FOR 16/26/27/28 SERIES SHOCKS Revised 6/17/14 Tech Line: (952) 985-5675 Fax (952) 985-5679 21730 Hanover Ave. Lakeville, MN 55044 www.qa1.net www.facebook.com/qa1motorsports TK08 TUNING KIT TECHNICAL MANUAL CONTENTS UNDER PRESSURE! USE EXTREME CAUTION

More information

Model 2310 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

Model 2310 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority VX-2 Model 2310 Series Clearing The Airway Is Our #1 Priority Operating Instructions & Maintenance Manual, INC. 11064 Randall Street Sun Valley, CA 91352 USA www.sscor.com Email: marketing@sscor.com techsupport@sscor.com

More information

Maintenance Information

Maintenance Information 04581245 Edition 2 May 2014 Air Grinder, Die Grinder and Sander Series G2 (Angle) Maintenance Information Save These Instructions Product Safety Information WARNING Failure to observe the following warnings,

More information

GH-BETTIS SERVICE INSTRUCTIONS DISASSEMBLY & REASSEMBLY FOR MODELS HD521-M4, HD721-M4 AND HD731-M4 DOUBLE ACTING SERIES PNEUMATIC ACTUATORS

GH-BETTIS SERVICE INSTRUCTIONS DISASSEMBLY & REASSEMBLY FOR MODELS HD521-M4, HD721-M4 AND HD731-M4 DOUBLE ACTING SERIES PNEUMATIC ACTUATORS GH-BETTIS SERVICE INSTRUCTIONS DISASSEMBLY & REASSEMBLY FOR MODELS HD521-M4, HD721-M4 AND HD731-M4 DOUBLE ACTING SERIES PNEUMATIC ACTUATORS WITH HYDRAULIC CONTROL PACKAGE PART NUMBER: SE-023 REVISION:

More information

DAH Left Hand Hose Inlet Kit Instruction Sheet For Use with 700 SERIES DAH Dilution at Hand Blend Centers

DAH Left Hand Hose Inlet Kit Instruction Sheet For Use with 700 SERIES DAH Dilution at Hand Blend Centers DAH Left Hand Hose Inlet Kit 98-54-1 Instruction Sheet For Use with 700 SERIES DAH Dilution at Hand Blend Centers BASIC KIT FUNCTION: The left hand hose inlet kit has two main functions. The kit can be

More information

CRD610 Automatic Fitting Inserter

CRD610 Automatic Fitting Inserter CRD610 Automatic Fitting Inserter OPERATIONS MANUAL VERSION 1.2 LAST EDITED 12.12.2018 cleanroomdevices.com 1 Table of Contents Title Page. 1 Table of Contents...2 1.0 General Product & Safety Information....3

More information

GPS AutoSteer System Installation Manual

GPS AutoSteer System Installation Manual GPS AutoSteer System Installation Manual Supported Vehicles Case IH Combines 7010 7120 8010 8120 AFX 8010 9120 PN: 602-0283-01-A LEGAL DISCLAIMER Note: Read and follow ALL instructions in this manual carefully

More information

Installation Instructions

Installation Instructions Instructions Aerada 900 Series Futura Faucet With Battery (BIR) Infrared Control S53-284 4" Centerset/Centershank S53-289 Centerset/Centershank w/plate Table of Contents Pre- Information............2 900

More information

GPS AutoSteer System Installation Manual

GPS AutoSteer System Installation Manual GPS AutoSteer System Installation Manual Supported Vehicles Case IH Vehicles Case 2577 Combines Case 2588 Combines Accuguide Ready PN: 602-0233-01-A LEGAL DISCLAIMER Note: Read and follow ALL instructions

More information

EZ LITE CRUISER. Service & Maintenance Manual

EZ LITE CRUISER. Service & Maintenance Manual EZ LITE CRUISER Service & Maintenance Manual Table Of Contents Introduction to the EZ Lite Cruiser 3 Identification of Components 4 Controller System Component Diagram 6 Controller System I/O Ports Detail

More information

Remove the 3-11mm nuts holding mirror on. Don t drop the nuts!

Remove the 3-11mm nuts holding mirror on. Don t drop the nuts! 2005-2012 Ford Mustang Puddle Lamp Kit Parts List: Quantity: Tool List: LED Lamps 2 Flat head screwdriver Seals 2 Ratchet & Socket set OR Nuts 2 Adjustable Wrench Wiring harness 1 Drill & 11/16 th bit

More information

INSTALLATION INSTRUCTIONS South Highway 11 Westminster, SC Toll Free (888) (864) FAX (864)

INSTALLATION INSTRUCTIONS South Highway 11 Westminster, SC Toll Free (888) (864) FAX (864) These instructions apply to the servicing of the Lift Technologies MaxiMizer Integral Sideshifters Cylinder Head. WARNING! Unless the steps in the following Installation Instructions are properly followed

More information

Southwest Windpower Instruction Sheet AIR-X Circuit Replacement Kit

Southwest Windpower Instruction Sheet AIR-X Circuit Replacement Kit Southwest Windpower Instruction Sheet AIR-X Circuit Replacement Kit Tools Required 5 / 32 Hex key 5 / 16 Hex key 7 / 64 Hex key Standard screwdriver Pair of external snap ring pliers Rubber mallet Hammer

More information

Safe-T-element Installation Instructions

Safe-T-element Installation Instructions Safe-T-element Installation Instructions For: PTI STEZA (2x2 Burner Configuration) & PTI STEZB (3x1 Burner Configuration) Revision K (May. 3 2012) TABLE OF CONTENTS 1. PREPARATION... 3 1.1 General Safety

More information

200 Shadylane Drive Philipsburg, PA Phone: (814) Fax: (814) Service Manual. Model 642E Centrifuge

200 Shadylane Drive Philipsburg, PA Phone: (814) Fax: (814) Service Manual. Model 642E Centrifuge 200 Shadylane Drive Philipsburg, PA 16866 Phone: (814) 342-6205 Fax: (814) 342-4510 www.druckerdiagnostics.com Service Manual Model 642E Centrifuge MODEL 642E SERVICE MANUAL REV: Original Issue 1 CONTENTS

More information

* * APPLICABLE MODELS: 2014 > Mazda 3

* * APPLICABLE MODELS: 2014 > Mazda 3 PART NUMBER: 0000 8C L48 (DIO) / 0000 89 L84 (PIO) GENUINE ACCESSORIES INSTALLATION INSTRUCTIONS Rev. AAA *550-0700-000* APPLICABLE MODELS: 2014 > Mazda 3 REQUIRED COMPONENTS: ITEM QTY DESCRIPTION Usage

More information

VECTRIX VX-2 SERVICE MANUAL. Version 1.0/May 2011 VECTRIX, LLC

VECTRIX VX-2 SERVICE MANUAL.   Version 1.0/May 2011 VECTRIX, LLC www.vectrix.com CONTENTS SECTION A: Tools 1 Tools Needed SECTION B: Mechanical Parts 1 Front Fairing 2 Front Console Cover 3 Speedometer Cover 4 Front Vertical Panel Cover-Lower 5 Front Vertical Panel

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS REAR DISC CONVERSION KIT A126-2 1988-98 C1500 2WD 10" REAR DRUM Thank you for choosing STAINLESS STEEL BRAKES CORPORATION for your braking needs. Pleases take the time to read

More information

Section 7 - Troubleshooting Guide

Section 7 - Troubleshooting Guide Section 7 - Troubleshooting Guide Section 7 - Troubleshooting Guide IMPORTANT While this troubleshooting guide provides information to aid in troubleshooting problems with the range, it does not contain

More information

Checker Unload Checkstand Installation Instructions

Checker Unload Checkstand Installation Instructions Front-End Checkouts Commercial Refrigeration Equipment Display Merchandisers Store Fixtures Checker Unload Checkstand Installation Instructions Attention Store Manager: Please find enclosed in this packet

More information

* * APPLICABLE MODELS: 2014 > Mazda 6

* * APPLICABLE MODELS: 2014 > Mazda 6 PART NUMBER: 0000 8C H02(DIO) / 0000 89 H18(PIO) GENUINE ACCESSORIES INSTALLATION INSTRUCTIONS Rev. AAA *550-0694-000* APPLICABLE MODELS: 2014 > Mazda 6 REQUIRED COMPONENTS: ITEM QTY DESCRIPTION Usage

More information

Tooling Assistance Center

Tooling Assistance Center Safeguards are designed into this application equipment to protect operators and maintenance personnel from most hazards during equipment operation. However, certain safety precautions must be taken by

More information

FlexJet - Flex Cable Replacement

FlexJet - Flex Cable Replacement P/N: 109515R0 14140 NE 200th St. Woodinville, WA. 98072 PH: (425) 398-8282 FX: (425) 398-8383 FlexJet - Flex Cable Replacement Notices: Warning! Ensure that all AC power cables are removed from the printer

More information

TBI /2012 TRAUMATIC BRAIN INJURY DEVICE

TBI /2012 TRAUMATIC BRAIN INJURY DEVICE USER MANUAL TBI 0310 6/2012 TRAUMATIC BRAIN INJURY DEVICE Page 1 of 26 Setting up the TBI 0310 Head Impactor The TBI 0310 Head Impactor when fully assembled has the following components: 1. Control box

More information

Sachs shock manual. ( ) 2 & 4 Stroke RR Enduro. ( ) RS Dual Sport

Sachs shock manual. ( ) 2 & 4 Stroke RR Enduro. ( ) RS Dual Sport Sachs shock manual (2013 2015) 2 & 4 Stroke RR Enduro (2014-2015) RS Dual Sport 1 Introduction The procedures in this manual must take place in a clean environment using professional tools and some specific,

More information

Operation and Maintenance Manual

Operation and Maintenance Manual BM / BMA Hydrometers ½ Operation and Maintenance Manual i This manual is intended for use by the users of this equipment. The information contained herein is the property of the Arad Ltd. Dalia and may

More information

Overload Warning System DS 50

Overload Warning System DS 50 Overload Warning System DS 50 Grove - FMTV Trouble Shooting Manual Level 1 - Basics - 1 Table of contents Troubleshooting Manual PAT Overload Warning system DS 50 Level 1 -Basics- Item Content Page 1.

More information

Installation instructions for Alpha Racing Quickshifter/Blipper for a pre-2014 S1000RR

Installation instructions for Alpha Racing Quickshifter/Blipper for a pre-2014 S1000RR Skill Level: Intermediate Installation instructions for Alpha Racing Quickshifter/Blipper for a pre-2014 S1000RR Tools Required: Alen keys/socket/drivers Torx keys/sockets/drivers Metric box wrenches and

More information

* * APPLICABLE MODELS: 2017 > CX-5

* * APPLICABLE MODELS: 2017 > CX-5 PART NUMBER: 0000 8C R06(DIO) / 0000 89 R28(PIO) GENUINE ACCESSORIES INSTALLATION INSTRUCTIONS Rev. AAA *550-0681-000* APPLICABLE MODELS: 2017 > CX-5 REQUIRED COMPONENTS: ITEM QTY DESCRIPTION Usage Chart

More information

Pressure Transducer Field Installation Instructions for the LI-6252 CO 2 and LI-6262 CO 2 /H 2 O Analyzers

Pressure Transducer Field Installation Instructions for the LI-6252 CO 2 and LI-6262 CO 2 /H 2 O Analyzers 6262-05 6262-03 Pressure Transducer Field Installation Instructions for the LI-6252 CO 2 and LI-6262 CO 2 /H 2 O Analyzers Please read these instructions before beginning installation. Please note that

More information

Installation Instructions COMPETITION/PLUS SHIFTER Ford Mustang MT82 6-Speed Manual Transmission Catalog#

Installation Instructions COMPETITION/PLUS SHIFTER Ford Mustang MT82 6-Speed Manual Transmission Catalog# Installation Instructions COMPETITION/PLUS SHIFTER 2015-2017 Ford Mustang MT82 6-Speed Manual Transmission Catalog# 3916037 Rev. 00 WORK SAFELY! For maximum safety, perform this installation on a clean,

More information

OWNER S MANUAL. ROTARY SURFACE CLEANER Models 105C, 105F, 105CW, & 105FW. Revision 2.01

OWNER S MANUAL. ROTARY SURFACE CLEANER Models 105C, 105F, 105CW, & 105FW. Revision 2.01 OWNER S MANUAL ROTARY SURFACE CLEANER Models 105C, 105F, 105CW, & 105FW Revision 2.01 ROTARY SURFACE CLEANER WARNING HIGH PRESSURE CAN CAUSE SERIOUS INJURY, MAXIMUM WORKING PRESSURE IS 4000 P.S.I. Any

More information

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL INSTALLATION, OPERATION, AND MAINTENANCE MANUAL MODEL 4D-200 REDUCED PRESSURE PRINCIPLE (RPZ) & MODEL 4D-700 REDUCED PRESSURE DETECTOR ASSEMBLY (RPDA) BACKFLOW PREVENTERS 2 ½ 10 Conbraco Industries Inc.

More information

Rollstar Shade Installation Instructions

Rollstar Shade Installation Instructions Rollstar Shade Installation Instructions All Lifting Systems Inside or Outside Mount Thank you for purchasing your new Rollstar shade. It has been custom-made from the highest quality materials to the

More information

P44 Stepper. User Manual

P44 Stepper. User Manual P44 Stepper User Manual Table of Contents Introduction 1 Safety Warning 2 Overview of Parts 2 Attaching to Chair 3 Manoeuvring Around 4 Setting Pedal Stops 4 Getting On 5 Setting the Resistance (models

More information

Maintenance Information

Maintenance Information 16572679 Edition 2 May 2014 Air Drill QP Series Maintenance Information Save These Instructions Product Safety Information WARNING Failure to observe the following warnings, and to avoid these potentially

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS REAR DISC BRAKE CONVERSION KIT A126-1 1973-87 CHEVROLET 1/2 TON 2WD Thank you for choosing STAINLESS STEEL BRAKES CORPORATION for your braking needs. Pleases take the time to

More information

INSTRUCTIONS Refillable, reusable. Extra versatile. Pressurized by free air

INSTRUCTIONS Refillable, reusable. Extra versatile. Pressurized by free air SURE SHOT SPRAYERS INSTRUCTIONS Refillable, reusable. Extra versatile. Pressurized by free air TOP LOAD WITH LIGHT LIQUIDS PRESSURIZE WITH AIR CHUCK SPRAY ANYWHERE MODEL A 32 OZ. DIRECTIONS FOR OPERATING

More information

Wheel Angle Sensor Kit Installation

Wheel Angle Sensor Kit Installation Wheel Angle Sensor Kit Installation Item Component Part Number Qty 1. WAS Bracket Kit 200-0247-02 1 2. WAS Assembly Kit 200-0468-01 1 3. Instruction Guide 602-0401-01 1 602-0401-01-A Overview Always shut

More information

WEBER CARBURETOR TROUBLESHOOTING GUIDE

WEBER CARBURETOR TROUBLESHOOTING GUIDE This guide is to help pinpoint problems by diagnosing engine symptoms associated with specific vehicle operating conditions. The chart will guide you step by step to help correct these problems. For successful

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS REAR DISC BRAKE CONVERSION KITS A112, A112-1 & A112-93 1979-93 FORD MUSTANG with 7.5" & 8.8" AXLES Thank you for choosing STAINLESS STEEL BRAKES CORPORATION for your braking needs.

More information

HAYWARD FLOW CONTROL TBH SERIES TRUE UNION BALL VALVE INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS

HAYWARD FLOW CONTROL TBH SERIES TRUE UNION BALL VALVE INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS HAYWARD FLOW CONTROL TBH SERIES TRUE UNION BALL VALVE INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS Pg. 1 of 10 PLEASE READ THE FOLLOWING INFORMATION PRIOR TO INSTALLING AND USING HAYWARD TBH SERIES

More information

Torqueflite Manual/Automatic Valve Body

Torqueflite Manual/Automatic Valve Body TCI 122400 Torqueflite Manual/Automatic Valve Body This valve body can be installed in a few hours by carefully following directions. Read all instructions first to familiarize yourself with the parts

More information

OPERATION AND MAINTENANCE MANUAL

OPERATION AND MAINTENANCE MANUAL WREN IBT SERIES HYDRAULIC TORQUE WRENCHES IBT SQUARE DRIVE SERIES OPERATION AND MAINTENANCE MANUAL FOR WREN Products: POINT 75, 1IBT, 3IBT, 5IBT, 8IBT, 10IBT, 20IBT, 25IBT, 35IBT, 50IBT SQUARE DRIVE HYDRAULIC

More information

Desktop 5.5 Z Axis Retrofit

Desktop 5.5 Z Axis Retrofit Page 1 Kit parts Desktop 5.5 Z Axis Retrofit Carriage plate with stop bolt and Z proximity switch installed Zip ties Spare bolts Spindle mounting plate with stop bolt, spring mount, and rail Z proximity

More information

Firehawk Second Stage Regulator Fire Service

Firehawk Second Stage Regulator Fire Service Firehawk Second Stage Regulator Fire Service MAINTENANCE AND REPAIR TAL 1701 (L) Rev. 2 MSA 2017 Prnt. Spec. 10000005389(I) Mat. 10147454 Doc. 10147454 TAL 1701 (L) Rev. 2-10147454 2 NON-CBRN FIREHAWK

More information

SERVICE PROCEDURES FOR CLUTCH HYDRAULIC UNITS

SERVICE PROCEDURES FOR CLUTCH HYDRAULIC UNITS SERVICE PROCEDURES FOR CLUTCH HYDRAULIC UNITS SAFETY PROCEDURES Always follow the vehicle manufacturer's recommended safety procedures in your Shop and Owners Manual. REQUIRED TOOLS Flat blade screwdriver,

More information

Maintenance Information

Maintenance Information 80234313 Edition 1 June 2006 Air Grinder, Die Grinder, Sander and Belt Sander Series G1 (Angle) Maintenance Information Save These Instructions WARNING Always wear eye protection when operating or performing

More information

INSTALLATION INSTRUCTIONS FOR THE TRUCK MOUNTED VIPER ADDITIVE INJECTION SYSTEM GTP-8776C

INSTALLATION INSTRUCTIONS FOR THE TRUCK MOUNTED VIPER ADDITIVE INJECTION SYSTEM GTP-8776C INSTALLATION INSTRUCTIONS FOR THE TRUCK MOUNTED VIPER ADDITIVE INJECTION SYSTEM GTP-8776C This additive injection system was designed to be used with five gallon jug of additive. The system is supplied

More information

FOR SIT-STAND WORKSTATION

FOR SIT-STAND WORKSTATION INSTALLATION MANUAL FOR SIT-STAND WORKSTATION Weight Capacity: 6.5-24.5 lbs. 6017180 Rev. B Contents Tools Required / Supplied Part Kits / Warnings/Disclaimers...2 Base Installation Clamp Mount Base Location...3

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS MODEL: GRAFTON RH-510 1 REV.A Restoration Hardware Balance Pressure Tub /Shower Set Specification Diagram Ensure that the stop ring () is correctly installed, prior to finished

More information

Wheel Angle Sensor Kit Installation

Wheel Angle Sensor Kit Installation Wheel Angle Sensor Kit Installation Item Component Part Number Qty 1. Bracket Kit, WAS, Titan 200-0769-01 1 2. WAS Assembly 200-0468-01 1 3. Instruction Guide 602-0576-01 1 602-0576-01 Rev A Overview Always

More information

Installation Manual TWM Performance Short Shifter Subaru STi 2008+

Installation Manual TWM Performance Short Shifter Subaru STi 2008+ - 1 - Installation Manual TWM Performance Short Shifter Subaru STi 2008+ Please Note: It is preferable to park on a flat surface, as you will have to engage and disengage the hand brake and shift from

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS FUEL PUMP HANGER MITSUBISHI LANCER EVOLUTION X Support: info@radiumauto.com Document# 19-0116 WARNING: DO NOT EXPOSE WORK AREA TO ANY SPARKS OR FIRE. DO NOT SMOKE WHILE WORKING

More information

TOYOTA HIGHLANDER RUNNING BOARD HIGHLANDER HV Preparation

TOYOTA HIGHLANDER RUNNING BOARD HIGHLANDER HV Preparation Preparation Part Number: PT738-48080 Kit Contents Item # Quantity Reqd. Description 1 1 Driver Side Running Board 2 1 Passenger Side Running Board 3 4 /Middle Mount Bracket 4 2 Rear Mount Bracket 5 2 Rear

More information

Maintenance Information

Maintenance Information 80234313 Edition 2 May 2014 Air Grinder, Die Grinder, Sander and Belt Sander Series G1 (Angle) Maintenance Information Save These Instructions Product Safety Information WARNING Failure to observe the

More information

OXYGEN GENERATING SYSTEMS INTL. Oxygen System Start-up, Overview and Maintenance Guide

OXYGEN GENERATING SYSTEMS INTL. Oxygen System Start-up, Overview and Maintenance Guide OXYGEN GENERATING SYSTEMS INTL. Oxygen System Start-up, Overview and Maintenance Guide OXYGEN GENERATOR SYSTEM START-UP STEP 1 Walk around the skid and inspect for damage. Then check that all electrical

More information

SERVICE GUIDE For WARN PULLZALL 24Vdc P/N &

SERVICE GUIDE For WARN PULLZALL 24Vdc P/N & SERVICE GUIDE For WARN PULLZALL 24Vdc P/N 885005 & 885006 REPAIR / REPLACEMENT INSTRUCTIONS TROUBLE SHOOTING GUIDE 987606A1.doc Page 1 of 46 WARNING This guide identifies potential hazards and has important

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS FUEL PUMP HANGER FORD MUSTANG: 2011+ S197 AND S550 Support: info@radiumauto.com Document# 19-0161 WARNING: DO NOT EXPOSE WORK AREA TO ANY SPARKS OR FIRE. DO NOT SMOKE WHILE WORKING

More information

Table of Contents. Timer Identification Timer ID BLU-U Features: 1K 6K BLU-U Features 1K 6K

Table of Contents. Timer Identification Timer ID BLU-U Features: 1K 6K BLU-U Features 1K 6K DUSA Pharmaceuticals, Inc. Table of Contents Go to Chart # Timer Identification Timer ID BLU-U Features: 1K 6K BLU-U Features 1K 6K BLU-U Features: 10K BLU-U Features 10K BLU-U Symptom Fans Running, Timer

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS REAR DISC BRAKE CONVERSION KIT A126-3 1988-98 CHEVY K1500 4WD 10" DRUMS Thank you for choosing STAINLESS STEEL BRAKES CORPORATION for your braking needs. Pleases take the time

More information

IBT Series Square Drive Torque Wrenches

IBT Series Square Drive Torque Wrenches IBT Series Square Drive Torque Wrenches Operation and Maintenance Manual Model.75, 1, 3, 5, 8, 10, 20, 25, 35, 50 http://www.torsionx.com Use the IBT Series Square Drive Torque Wrenches Model.75, 1, 3,

More information

UPC5000/UPC5010. Operation & Maintenance Manual. Portable & Rack-mountable Pneumatic Pressure Calibration Console

UPC5000/UPC5010. Operation & Maintenance Manual. Portable & Rack-mountable Pneumatic Pressure Calibration Console UPC5000/UPC5010 Portable & Rack-mountable Pneumatic Pressure Calibration Console Operation & Maintenance Manual CONDEC Sales Phone No. (888) 295-8475 CONDEC Web Site: www.4condec.com 63256 Contents About

More information

AV INSTALLATION, OPERATION & MAINTENANCE MANUAL

AV INSTALLATION, OPERATION & MAINTENANCE MANUAL Rack & Pinion Pneumatic Actuators AV INSTALLATION, OPERATION & MAINTENANCE MANUAL AVCS PO Box 68172 Minneapolis, MN 55418 TABLE OF CONTENTS CHAPTER 1: PRODUCT DESCRIPTION 3 CHAPTER 2: TECHNICAL FEATURES

More information

Installation For Service Only

Installation For Service Only For Service Only IMPORTANT! Aerada 1100 Series Low Arc Faucet S53-302 Battery Infrared Metering Faucet (Center Shank with 4" Trimplate) THIS SIDE UP! Packing List Read this entire installation manual to

More information

MODEL MA4210 Installation and Operation Manual Important:

MODEL MA4210 Installation and Operation Manual Important: MODEL MA4210 Installation and Operation Manual Important: This manual contains specific cautionary statements relative to worker safety. Read this manual thoroughly and follow as directed. It is impossible

More information

Service Manual. Balanced Piston First Stage

Service Manual. Balanced Piston First Stage Service Manual Balanced Piston First Stage Copyright 2002, Cressi-sub Revised 3/2002 2 Balanced Piston First Stage Service Manual Contents BEFORE STARTING... 3 DISASSEMBLY... 3 PARTS CLEANING AND LUBRICATION...

More information

If technical support is required, please contact Advent Technical Support at

If technical support is required, please contact Advent Technical Support at Document 128-9015A Created 12/12/11 Kit Contents: Item # Qty. Component Description 1 2 Headrest Assembly 2 2 Cables # 3 3 1 Power Cord # 9 4 1 FM Antenna 5 1 Control Box 6 2 IR Headphones 7 2 Remote Control

More information

Operation and Maintenance Manual http://www.torsionx.eu Use the MaxDrv Series Square Drive Torque Wrench Model.75, 1, 3, 5, 8, 10, 20, 25, 35, 50 to install and remove threaded fasteners requiring precise

More information