1 Meeting the Requirements of ASTM D ( IP548/06 ) with Agilent 1200 LC Systems M.Woodman, Chemical Analysis Solutions Unit, Wilmington, Delaware, USA M.Sierocinska, European Field Support Centre, Waldbronn, Germany
2 Agenda What is the source and application for ASTM D (and IP548/06)? Description of Recent ASTM D Method Updates New D / IP548/06 Method Requirements Alternate methods for biodiesel blends Recommendations for the best configuration Results from locally collected diesel samples Summary Page 2
3 Usage and Source of ASTM/IP/EN methods used ASTM D and IP548/06 measure aromatic hydrocarbon groups in petroleum distillates having specified boiling point ranges of o C. These methods are mainly applied to fuel analysis, particularly diesel motor fuel. Kerosene based fuels for aviation are covered by another related method ASTM ( is the source of ASTM methods. The Energy Institute ( is the source of the IP (sometimes referenced as EI) methods IP (or EI) and EN methods (European Committee for Standardization ) often have parallel ASTM methods The method revision designation is very important Page 3
4 Related IP, EN and ASTM Methods IP Method and Revision Method Overview Special Parameters ASTM Method? Comments IP391/ o C diesel fuel petro/bio blends up to B-5 no backflush amino and/or cyano column no equivalent available at this time same as method EN12916:2006 MAH, DAH, Tri+AH reported IP548/ o C diesel fuel backflush required D MAH, DAH, Tri+AH reported amino and/or cyano column FAME interferes with result IP436/ o C aviation fuel, kerosene no backflush amino and/or cyano column D MAH and DAH reported not for samples with Tri+AH Page 4
5 General Conditions and Schematic of ASTM D 6591 Mobile phase: n-heptane at 1ml/min flow rate Column: Amino-propyl and/or Amino-Cyano-propyl bonded phases. Amino-propyl 4.6x250mm 5um is commonly used Column Temperature: approx o C Detection: Refractive Index Special Feature: 6 port valve for backflushing the column to elute tri+ aromatics as a single peak Page 5
6 General Schematic of ASTM D Figure 1, Agilent publication E Page 6
7 Recent HGA Method Updates: ASTM D Terms redefined more specific description of polyaromatics adopted Designate a new IP method correlation IP (used with ASTM D ) replaced by IP548/06 Exclusion or warning regarding Bio/Petrodiesel Blends the method recognizes difficulties with biodiesel fuel blends and refers the analyst to IP391/07 (EN12916:2006) for samples containing biodiesel (FAME) components due to inclusion of FAME components in the tri+ah backflush peak, erroneously high tri+ aromatic results are reported if D is used with biodiesel blends unusually high tri+ah results with this method should be investigated to determine if unsuspected FAME is present due to fuel contamination or mislabeling Page 7
8 General Conditions and Schematic of D System Performance Standard 1. cyclohexane 2. o-xylene (dimethylbenzene) 3. dibenzothiophene 4. 9-methyl anthracene ASTM D System Performance Std., Agilent p/n Mobile phase: n-heptane at 1ml/min flow rate Column: ZORBAX NH2 4.6x250mm, 5um Column Temperature: 20 o C Detection: Refractive Index, 35 o C Page 8
9 General Conditions and Schematic of D Calibration Standards 1. cyclohexane 2. o-xylene (1,2-dimethylbenzene) 3. 1-methyl-naphthalene 4. phenanthrene with backflush at ~9 min ASTM D Calibration Standard A-D, Agilent kit p/n Mobile phase: n-heptane at 1ml/min flow rate Column: ZORBAX NH2 4.6x250mm, 5um Column Temperature: 20 o C Detection: Refractive Index, 35 o C Page 9
10 Analysis of Petrodiesel, overlay of 3 injections 1. saturates 2. mono-aromatics 3. di-aromatics 4. tri+ aromatics t= <30 minutes Ultra Low Sulfur Diesel (ULSD), retail pump Page 10
11 Analysis of Petrodiesel, overlay of 3 brands 1. saturates 2. mono-aromatics 3. di-aromatics 4. tri+ aromatics t= <30 minutes Ultra Low Sulfur Diesel (ULSD), retail pumps a winter blend, old retained sample summer blends Page 11
12 ASTM D Separation Criteria The official method clearly describes the appearance and statistical evaluation of method conditions for the new revision Resolution, selectivity, precision and calibration linearity are clearly stated In the following slides we review the performance of the Agilent approach to this method. indicates that Agilent meets requirements
13 ASTM D Criteria Column Selection Column Descriptions: 4.6 x 250 mm, stainless steel column packed with 5um ZORBAX NH2 (p/n ) SPS 1. cyclohexane 2. o-xylene (1,2-dimethylbenzene) 3. dibenzothiophene 4. 9-methyl anthracene ml/min heptane 20C column temp. 35C RID temp. 20ul SPS no backflush. 6.4 Column System Any stainless steel HPLC column(s) packed with an approved amino-bonded (or polar amino/cyano-bonded) silica stationary phase is suitable
14 SPS Separation and Resolution Criteria SPS 1. cyclohexane 2. o-xylene (1,2-dimethylbenzene) 3. dibenzothiophene 4. 9-methyl anthracene ml/min heptane 20C column temp. 35C RID temp. 20ul SPS backflush 9.4 min provided it meets the resolution requirements laid down in Ensure that baseline separation is obtained between all components of the SPS. [without backflush]. 9.6 Calculate the backflush time, B, in seconds, using the following equation: B = ta + 0.4(tB ta)
15 Calculation of Resolution (Cyclohexane to 1,2 dimethylbenzene (o-xylene)) Ensure that the resolution between cyclohexane and 1,2 dimethylbenzene is not less than Column Resolution -- Calculate the resolution, R, between cyclohexane and 1,2 dimethylbenzene using the following equation. 2(t 2 -t 1 ) (difference in retention time) R = (y 1 +y 2 ) (averaging of peak widths). Name R. Time width (hh) Rsln. 1. cyclohexane ,2-dimethylbenzene) dibenzothiophene methyl anthracene
16 Linearity (section 9.4: R = >0.999, Intercept <0.01g/100ml) RID peak area vs. Std.Conc., o-xylene y = x R 2 = RID Peak Area Series1 Linear (Series1) Conc. (g/100 ml). correlation coefficients all >0.999 (fits are nominally 1), linear fit of the calibration points Intercepts all =<0.005g/100ml
17 Reproducibility for Retention and Area, Cal. A-D Calibrant A Analyte R.T. Avg., n=3 R.T. Stdev R.T.RSD% Area Avg., n=3 Area Stdev Area.RSD% xylene % 3.29E % 1-Me-naphthalene % 5.76E % phenanthrene % 7.12E % Calibrant B Analyte R.T. Avg., n=3 R.T. Stdev R.T.RSD% Area Avg., n=3 Area Stdev Area.RSD% xylene % 8.33E % 1-Me-naphthalene % 1.46E % phenanthrene % 3.55E % Calibrant C Analyte R.T. Avg., n=3 R.T. Stdev R.T.RSD% Area Avg., n=3 Area Stdev Area.RSD% xylene % 2.06E % 1-Me-naphthalene % 3.66E % phenanthrene % 8.87E % Calibrant D Analyte R.T. Avg., n=3 R.T. Stdev R.T.RSD% Area Avg., n=3 Area Stdev Area.RSD% xylene % 4.03E % 1-Me-naphthalene % 2.96E % phenanthrene % 1.76E % Average RSD% All Runs 0.028% 0.555%
18 Recommendations for the best configuration Agilent pump, minimum isocratic G1311A Agilent sampler, minimum G1329A Agilent thermostatted column compartment (TCC) with 6 port 2 position switching/backflush valve Agilent RID G1362A Agilent ChemStation B or EZChrom software, or third party compatible control/data software Agilent ZORBAX NH2 4.6x250mm 5um column equilibrated to heptane mobile phase Agilent calibrants for system performance and calibration HPLC grade n-heptane, 2-propanol Page 18
19 What to order, for consumables, ASTM D ASTM D (IP548/06) System Performance Standard (SPS) Agilent p/n ASTM D (IP548/06) System Calibration Solutions A-D Agilent p/n Column(s): ZORBAX NH2 4.6x250mm, 5um, Agilent p/n guard cartridge (pk/4) 4.6x12.5mm, 5um, Agilent p/n guard hardware (holder) kit, 4.6mm, Agilent p/n Normal phase pump seals, (pk/2), Agilent p/n Mobile phase: n-heptane, HPLC grade, minimum 4 Liter or 2.5 Liter bottles, 2-propanol, HPLC grade, 1 Liter (installation, maintenance) Page 19
20 Performance with Selected Samples Samples were collected from retail and commercial diesel and biodiesel distribution centers around the Chicago area One older sample was collected in the cold winter months, all others in recent warm summer months. Fuel blends vary seasonally in this area, with change occurring around early September and April Sample preparation: Accurately weigh ~1g fuel and dilute to 10ml with n-heptane. Mix thoroughly and inject 10ul. Calculations: Determine sample concentrations from calibration curves using 1,2 dimethylbenzene for MAH, 1-methyl-naphthalene for DAH and phenanthrene for Tri+AH. Correct for sample weight and dilution variables. Conditions: as described in previous slides Page 20
21 Analysis of Petrodiesel, overlay of 3 brands 1. saturates 2. mono-aromatics 3. di-aromatics 4. tri+ aromatics t= <30 minutes Ultra Low Sulfur Diesel (ULSD), retail pumps a winter blend, old retained sample summer blends Page 21
22 Analysis of Petrodiesel, n=3, Vendor 1 Group Avg., n=3 RSD% MAH g/100ml 0.68% DAH 4.74 g/100ml 2.77% Tri+AH 0.70 g/100ml 6.63% Vendor 1, Retail pump, summer MAH DAH Tri+AH Mobil Page 22
23 Analysis of Petrodiesel, n=3, Vendor 2 Group Avg., n=3 RSD% MAH g/100ml 0.12% DAH 5.10 g/100ml 0.60% Tri+AH 0.77 g/100ml 1.52% Vendor 2, Retail pump, summer MAH DAH Tri+AH Marathon Page 23
24 Analysis of BioPetrodiesel with ASTM D Amount units Group g/100ml MAH 4.12 g/100ml DAH 3.59 g/100ml Tri+AH Vendor 3, consumer retail pump, summer, dispensing pump labeled diesel erroneous high Tri+AH due to FAME in sample Citgo Page 24
25 Analysis of BioPetrodiesel with ASTM D Amount units Group g/100ml MAH 5.12 g/100ml DAH 4.12 g/100ml Tri+AH Vendor 4, commercial biodiesel blend, dispensing pump labeled diesel erroneous high Tri+AH due to FAME in sample RR Page 25
26 Analysis of Biodiesel B-11 with Method IP391/07 Group Avg., n=3 RSD% MAH g/100ml 5.06% DAH 5.15 g/100ml 3.38% Tri+AH 0.72 g/100ml 4.58% Vendor 4, ULSD-based biodiesel blend retail auto/light truck Results are n=3 by IP391/07 method approved for diesel and biodiesel blends Tri+AH region FAME resolved RR Page 26
27 ASTM D Observations The method retention time stability is very sensitive to column activity, as noted in the official method. This is typical for many normal phase separation methods Moisture in solvents or samples is problematic and affects column activity Variable retention is primarily due to moisture and temperature, though good system maintenance is essential Use of normal phase seals in the pump is very important Page 27
28 Summary ASTM D Status Specialized all-agilent column arrangement meets SPS resolution requirements for cyclohexane and 1,2-dimethylbenzene (o-xylene) Adding the 6-port valve to the system configuration is required, however it is possible to easily switch between current IP391/07 (no backflush, for biodiesel blends) and ASTM D 6591 (with backflush, no FAME content) Page 28
29 References 1. IP391/07 / EN12916:2006, Energy Institute, UK 2. Agilent application note E 3. ASTM Methods D , D Page 29
30 Appendix and supplemental slides may be used depending on audience and need. Page 30
31 Acknowledgements Cynthia guidance, resources and encouragement Maggie continued technical support and for persistence in keeping this development on the road map Melanie/Georgia/Phil standards packaging and work with the Ultra Scientific team Andre creating the opportunity and motivation Ultra Scientific too many to name, once again committed to a successful development with Agilent Page 31
Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems
Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems Application Note Hydrocarbons Authors Michael Woodman Agilent Technologies, Inc. Chemical Analysis Solutions 2850
More informationApplication Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany
Determination of Aromatic Hydrocarbons in Petroleum Middle Distillates with the Agilent Infinity Binary HPLC System with RID Detection According to IP9()/ASTM D9 Application Note Energy & Chemicals - Petrochemicals
More informationDetermination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC
Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC Application Note Energy and Fuels Authors Detlef Wilhelm Anatox GmbH & Co. KG Fürstenwalde, Germany Udo Huber Agilent
More informationApplication Note. Authors. Abstract. Energy & Chemicals
Determination of Aromatic Content in Diesel Fuel According to ASTM D5186 Enhancing the Agilent 126 Infi nity Analytical SFC System with a Flame Ionization Detector Application Note Energy & Chemicals Authors
More informationGas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593
Application Note Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593 Authors Kelly Beard and James McCurry Agilent Technologies, Inc. Abstract An Agilent
More informationAgilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011
Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc.
More informationGC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585
GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585 Application Note Fuels Author James D. McCurry, Ph.D. Agilent Technologies, Inc. 850 Centerville
More informationOptimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE
Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System Jim McCurry Roger Firor Agilent Technologies Wilmington, DE Page 1 Scope of the Analyzer Designed for ASTM D7798 For
More informationEvaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS
Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS Application Environmental Authors Wei Luan and Chuanhong Tu Agilent Technologies (Shanghai) Co., Ltd. 412 Ying
More informationOxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels
Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels James D. McCurry Senior Scientist Agilent Technologies Wilmington, DE USA Page 1 Application Summary There is a need to measure
More informationHigh Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph
Application Note Petrochemicas High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph Author James D. McCurry, Ph.D. Agilent Technologies, Inc. Abstract An Agilent
More informationAgilent and ASTM. Update on Recent Activities. Page 1
Agilent and ASTM Update on Recent Activities Page 1 ASTM Committees for Refining, Fuels, Petroelum Products ASTM International D2 Petroleum products & lubricants D3 Gas fuels (natural gas) gas fuels D16
More informationApplication Note. Abstract. Authors. Environmental Analysis
High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID using the Agilent Low Thermal Mass (LTM II) System Application Note Environmental Analysis Authors Frank David and Karine Jacq Research
More informationUsing the PSD for Backflushing on the Agilent 8890 GC System
Application Note Petrochemicals Using the PSD for Backflushing on the Agilent 889 GC System Author Brian Fitz Agilent Technologies, Inc. Wilmington, DE, USA. Abstract An Agilent 889 series GC equipped
More informationMethanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.
Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers Application Note Abstract Versa With the rising prices of fossil fuels, more emphasis is being put on renewable resources
More informationDetection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector
Detection of Sulfur Compounds in Natural Gas According to ASTM D554 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector Application Note Author Rebecca Veeneman Abstract Sulfur compounds in natural
More informationHigh Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell
High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell Application Note Consumer Products Authors Gerd Vanhoenacker,
More informationApplication Note. Author. Introduction. Energy and Fuels
Analysis of Free and Total Glycerol in B-100 Biodiesel Methyl Esters Using Agilent Select Biodiesel for Glycerides Application Note Energy and Fuels Author John Oostdijk Agilent Technologies, Inc. Introduction
More informationUsing a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application
Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D759 Application Petrochemical Author James D. McCurry Agilent Technologies 285 Centerville
More informationApplication. Gas Chromatography June 1995
Determining Oxygenates in Gasoline: ASTM Method D Application Gas Chromatography June 99 Authors Michael J. Szelewski Agilent Technologies, Inc. 0 Centerville Road Wilmington, DE 90-60 USA Matthew S. Klee
More informationAgilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass. James D. McCurry, Ph.D. Senior Scientist
Agilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass James D. McCurry, Ph.D. Senior Scientist Agilent Solutions for the Analysis of Ethanol-Based Fuels Derived From Biomass Presented
More informationAgilent G5632 Valve Kit Instructions
Agilent G5632 Valve Kit Instructions Agilent G5632 Valve Kit Instructions Contents Typical Applications 2 Alternating Column Regeneration (2 position/10 port valves only) 2 Dual column selection 3 Sample
More informationIncreased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel
Application Note Energy and Chemicals Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel Applying the Energy Institute Method IP 8 with an Agilent J&W
More informationGroup-Type Analysis (PiPNA) in Diesel and Jet Fuel by Flow Modulated GCxGC FID.
Group-Type Analysis (PiPNA) in Diesel and Jet Fuel by Flow Modulated GCxGC FID. Dedicated PiPNA + FAME For (Bio-)Diesel and Jet Fuels Robust System, Easy to use No Cryogenic coolant Required Keywords:
More informationAnalysis of Petroleum Fractions by ASTM D2887
Analysis of Petroleum Fractions by ASTM D2887 Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 2582 Key Words Simulated distillation, D2887, TRACE TR-SimDist Abstract ASTM
More informationImpurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution
Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 129 Infinity II HDR-DAD Impurity Analyzer Solution Application ote Small Molecule Pharmaceuticals Author Sonja Schneider Agilent Technologies,
More informationAnalysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction
Analysis of Glycerin and Glycerides in Biodiesel (B1) Using ASTM D68 and EN11 Application HPI/Petrochemicals/Polymers Author James D. McCurry Agilent Technologies, Inc. 8 Centerville Road Wilmington, DE
More informationHigh Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System
High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System Application Note Authors Frank David Research Institute for Chromatography, Pres.
More informationApplication Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER
Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D7423-09 using AC OXYTRACER Fast Analysis in
More informationGC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method
GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN1413:211 Method Application Note Author James D. McCurry, Ph.D. Agilent Technologies Abstract
More informationDetermination of Free and Total Glycerin in B100 Biodiesel
Page 1 of 5 Page 1 of 5 Return to Web Version Determination of Free and Total Glycerin in B100 Biodiesel By: Michael D. Buchanan, Katherine K. Stenerson, and Vicki Yearick, Reporter US Vol 27.1 techservice@sial.com
More informationAlternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis
Introduction Petroleum & Petrochemical Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis By Katarina Oden, Barry Burger, and Amanda Rigdon Crude oil consists of thousands of different
More informationComplete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges
Complete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges Alexandria Pavkovich Jason Thomas Trent Sprenkle Outline Background EPA Method Requirements Background
More informationSimulated Distillation Analyzers, Software, Standards, Consumables, Training
Simulated Distillation Analyzers, Software, Standards, Consumables, Training www.separationsystems.com Offering the Fullest Range of Optimized Solutions Simulated distillation (SimDis) has been used to
More informationAgilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve
Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve Technical Note This Technical Note describes the purpose, installation, and configuration and use of a diverter valve
More informationFree and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584
Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN 14105 and ASTM D6584 Introduction With today s increasing concern for the environment and the depletion of fossil
More informationAgilent Solutions for the Petrochemical and Oleochemical Industries
e-seminar Series - Agilent Solutions for the Petrochemical and Oleochemical Industries Date: 12 June 2009 Time : 10am Singapore Time For audio, please dial one of the phone numbers listed: International
More informationASTM D 6730 Detailed Hydrocarbon Analysis
ASTM D 6730 Detailed Hydrocarbon Analysis Jaap de Zeeuw, Jan Pijpelink and Barry Burger Restek Corporation ASTM D 6730-01(2006)e1 Determination of Individual Components in Spark Ignition Engine Fuels as
More informationEmissions from Heavy-Duty Diesel Engine with EGR using Oil Sands Derived Fuels
Emissions from Heavy-Duty Diesel Engine with EGR using Oil Sands Derived Fuels W. Stuart Neill National Research Council Canada Ottawa, Ontario, Canada 9 th DEER Conference, Newport, Rhode Island August
More informationWhite Paper.
The Advantage of Real Atmospheric Distillation Complying with the ASTM D7345 Test Method in the Distillation Process Introduction / Background In the past, refiners enjoyed a constant supply of the same
More informationTECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure
TECHNICAL REPORT Modifying Agilent 1100 HPLC Systems to Achieve UHPLC-like performance with HALO Fused-Core Columns Agilent 1100 HPLC system With a few modifications, an Agilent 1100 HPLC can produce UHPLC-like
More informationHigh-Temperature Simulated Distillation System Based on the 6890N GC Application
High-Temperature Simulated Distillation System Based on the 6890N GC Application Petroleum Authors ChunXiao Wang Agilent Technologies (Shanghai) Co., Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai
More informationModule8:Engine Fuels and Their Effects on Emissions Lecture 36:Hydrocarbon Fuels and Quality Requirements FUELS AND EFFECTS ON ENGINE EMISSIONS
FUELS AND EFFECTS ON ENGINE EMISSIONS The Lecture Contains: Transport Fuels and Quality Requirements Fuel Hydrocarbons and Other Components Paraffins Cycloparaffins Olefins Aromatics Alcohols and Ethers
More informationRestek Corporation 110 Benner Circle, Bellefonte, Pa Barry L. Burger, Neil Johansen, Valerie Gamble, Donald Rhoades
High Speed PONA Analysis For Detailed Hydrocarbon Analysis Extended (DHAX) Using Hydrogen Carrier Gas For The Determination of Individual Components In Spark Ignition Fuels Barry L. Burger, Neil Johansen,
More informationFast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC
Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC Technical Overview Trace gas analysis is a challenge in today s world. The ability to analyze lower component
More informationRefinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography
Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography Dr. Robert Lorenz Analytical Chemist Chevron Energy Technology Company Gulf Coast Conference Galveston, TX October
More informationMethod Development for Capillary GC Systems. Slide 1
Method Development for Capillary GC Systems Slide 1 AREAS TO OPTIMIZE Injector Carrier gas Column temperature Slide 2 COMMON INJECTOR MODES Vaporization Injection Modes Megabore Direct Split Splitless
More informationAgilent InfinityLab Quick Change Valves G4231A/C and G4232C/D
Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Instructions Agilent Valve Kit G4231A/C and G4232C/D - Instructions This technical note describes the installation and application of the Agilent
More informationSELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS
Rev 1 3/6/2004 Selerity Technologies Inc. www.selerity.com SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS What is supercritical fluid chromatography? A chromatographic technique in
More informationDetermination of Sudan Dyes I IV in Curry Paste
Determination of Sudan Dyes I IV in Curry Paste Suparerk Tukkeeree and Jeffrey Rohrer 2 Thermo Fisher Scientific, Bangkok, Thailand; 2 Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 23 Key
More informationTennessee Department of Agriculture
Tennessee Department of Agriculture Biodiesel Quality Program Education, Communication, Cooperation, & Regulation Presented by Randy Jennings Tennessee Department of Agriculture Regulatory Services February
More information[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS
MS Identification of Trace level Impurities from a Non-MS Compatible Mobile Phase Using ACQUITY UPLC System with 2D Technology by Heart-cutting and Online Sample Concentration Bronsky Gopinadh, Dilshad
More informationAlliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE
Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE AN HPLC THAT WORKS AS HARD AS YOU DO Confidence in your results comes with high performance, day-to-day reliability, and overall system robustness.
More informationDetermination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105
Application Note: 10215 Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105 Fausto Munari, Daniela Cavagnino, Andrea Cadoppi, Thermo Fisher Scientific, Milan,
More informationGAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS
GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS FLASH VAPORISATION INJECTION Split Splitless On-Column COOL INJECTION Large Volume Injection (LVI) On-Column On-Column-SVE (with solvent vapour
More informationAgilent 1290 Infinity II Ultra Low Dispersion Kit
Agilent 1290 Infinity II Ultra Low Dispersion Kit Technical Note 1290 Infinity II Ultra Low Dispersion Kit - Technical Note In this note we describe how to install the Agilent 1290 Infinity II Ultra Low
More informationAgilent Valve Kit G4231A/B/C and G4232B/C/D
Agilent Valve Kit G4231A/B/C and G4232B/C/D Instructions Agilent Valve Kit G4231A/B/C and G4232B/C/D - Instructions This Technical Note describes the installation of the Agilent G4231A/B/C and G4232B/C/D
More informationWhite Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information
Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Abstract High Temperature Simulated Distillation (High Temp SIMDIS) is one of the most frequently used techniques to determine
More informationAgilent Valve Kit G4231A/B and G4232A/B/C
Agilent Valve Kit G4231A/B and G4232A/B/C Instructions Contents Typical Applications 2 Dual column selection (G4231A/B or G4232A/B/C) 2 Sample enrichment and sample cleanup (G4231A/B or G4232A/B/C) 2 Alternating
More informationVoting Draft Standard
page 1 of 7 Voting Draft Standard EL-V1M4 Sections 1.7.1 and 1.7.2 March 2013 Description This proposed standard is a modification of EL-V1M4-2009-Rev1.1. The proposed changes are shown through tracking.
More informationGC Method Compliance and Large Valve Oven Application
GC Method Compliance and Large Valve Oven Application Solution Focused Kelly Beard June 20, 2016 Compliant Methods Compliant Methods What are our options? Analyzers and Analyzer Kits High Volume Well Known
More informationLarge Volume Injection of Polycyclic Aromatic Hydrocarbons
JSB is an authorised partner of Large Volume Injection of Polycyclic Aromatic Hydrocarbons Application Note - Environmental #113 Author Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Abstract
More informationSUPPRESSION IN ANALYSIS OF ARTIFICIAL SWEETENERS WITH
APPROACHES TO DECREASE ION SUPPRESSION IN ANALYSIS OF ARTIFICIAL SWEETENERS WITH UPLC-TQ MS Noora Perkola Finnish Environment Institute 3rd Nordic MS Symposium September 12 2012, Jurmala, Latvia ARTIFICIAL
More informationDual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector
Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector Application Note Hydrocarbon Processing Authors ChunXiao Wang Agilent Technologies
More informationThe Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over
The Agilent 12 Series high performance autosampler SL: Area precision, injection volume linearity, imum accessible volume, carry-over Technical Note 8 6 4 2.5 1 Introduction A main performance criterion
More informationFast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC
Fast GC SPEEDY GC Slide 4 Optimization Goals Primary: Minimize time for the separation of a given number of peaks Secondary: Maximize the number of peaks separated by a given column Slide 5 HOW CAN THESE
More informationAppNote 1/2010 KEYWORDS ABSTRACT. Biodiesel, Automation, ASTM D
AppNote 1/2010 Full Automation of ASTM Method D6584-07 Standard Test Method for the Determination of Free and Total Glycerin in B-100 Biodiesel Methyl Esters by Gas Chromatography using a GERSTEL Dual
More informationWET GRIP TEST METHOD IMPROVEMENT for Passenger Car Tyres (C1) GRBP 68 th session
Transmitted by the expert from ETRTO Informal document GRB-68-15 (68 th GRB, 12-14 September 2018, agenda item 6) WET GRIP TEST METHOD IMPROVEMENT for Passenger Car Tyres (C1) Overview of Tyre Industry
More informationC2, C3, C4 Monomer Analysis
C2, C3, C4 Monomer Analysis Malgorzata Sierocinska Agilent Technologies Waldbronn Page 1 Why Analyze Monomers? To Insure Consistent Production of High Quality Polymer Protect against food contamination
More informationa review of analytical methods for the quantification of aromatics in diesel fuels
a review of analytical methods for the quantification of aromatics in diesel fuels Prepared for the Automotive Emissions Management Group by the Special Task Force on Diesel Fuel Emissions (AE/STF-7) P.
More informationPerforming ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software
Install a capillary column in the oven of the SRI GC. The ASTM method suggests a 12 meter.32mm id narrow-bore column coupled with a 2.5 meter guard column but permits the use of any column which exhibits
More informationTechnical Procedure for Gas Chromatography (GC-FID)
Technical Procedure for Gas Chromatography (GC-FID) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph (GC- FID). 2.0 Scope This procedure applies to all
More informationOn-Line Process Analyzers: Potential Uses and Applications
On-Line Process Analyzers: Potential Uses and Applications INTRODUCTION The purpose of this report is to provide ideas for application of Precision Scientific process analyzers in petroleum refineries.
More informationOnline sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve
Online sample cleanup on the Agilent 129 Infinity LC using a built in 2-position/6-port valve Analysis of Sudan red compounds in paprika powder Application Note Food Authors Angelika Gratzfeld-Huesgen
More informationZürich Testing on Fuel Effects and Future Work Programme
Zürich Testing on Fuel Effects and 2016-2017 Future Work Programme Benjamin Brem 1,2, Lukas Durdina 1,2 and Jing Wang 1,2 1 Empa 2 ETH Zürich FORUM on Aviation and Emissions Workshop Amsterdam 15.04.2016
More informationRefill Capacities and Recommendations
Page 1 of 13 Shutdown SIS Previous Screen Product: TRUCK ENGINE Model: C7 TRUCK ENGINE WAX Configuration: C7 On-highway Engine WAX00001-UP Operation and Maintenance Manual C7 On-highway Engine Media Number
More informationSulfur Detection at ppb Levels in Light Hydrocarbon Streams
Sulfur Detection at ppb Levels in Light Hydrocarbon Streams Based on a New Super Permeable PLOT Column Agilent Select Low Sulfur Johan Kuipers Channel Training Specialist Oct 12 th, 2010 1 October 18,
More information3.0 FUEL INJECTION AND INDUCTION SYSTEM CLEANER CONCENTRATE
Fig. 1: Identifying BMW Group Fuel System Cleaner Plus Bottle BMW Group Fuel System Cleaner Plus is the only BMW approved in tank additive. Using non approved fluids or tools can lead to premature component
More informationFast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application
Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application Petroleum Authors ChunXiao Wang Agilent Technologies (Shanghai) Co.,Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai
More informationASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column
ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column Cory S. Fix, Director of Application Development cory.fix@vgcchromatography.com Willie Steinecker,
More informationAnalysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES
Analysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES Application note Petrochemical Author Neli Drvodelic Agilent Technologies Melbourne, Australia Introduction
More informationAppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS
AppNote 6/26 Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) Andreas Hoffmann GERSTEL GmbH & Co.KG, Eberhard-Gerstel-Platz 1,
More informationMET-Biodiesel Capillary GC Columns
MET-Biodiesel Capillary GC Columns Product Specifications Product Features & Benefits Chromatograms FAQs Related Products Updated: February 2, 2009 Product Specifications 2 Product Specifications What
More informationDetermination of chlorine/chlorides using MWD- XRF
Determination of chlorine/chlorides using MWD- XRF Ranzy Morgan Director, Sales & Marketing Petroleum Analyzers Scott Blumenshine Senior Chemist Technical Support Flint Hills Resources XOS, Inc. East Greenbush,
More informationAnalysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID
Analysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID Introduction P Gorst-Allman (LECO Africa Pty. Ltd) and B-J de Vos (NMISA). The analysis
More informationISBN SANS 342:2006 Edition 4 SOUTH AFRICAN NATIONAL STANDARD Automotive diesel fuel Published by Standards South Africa 1 dr lategan roa
ISBN 0-626-18752-4 SOUTH AFRICAN NATIONAL STANDARD Automotive diesel fuel Published by Standards South Africa 1 dr lategan road groenkloof private bag x191 pretoria 0001 tel: 012 428 7911 fax: 012 344
More informationGC Inlets. An Overview. Simon Jones GC Applications Engineer
GC Inlets An Overview Simon Jones GC Applications Engineer Types of Inlets Purged Packed Split / Splitless Cool On Column Programmable Temperature Vaporization Volatiles Interface Multi Mode Inlet Where
More informationDxC 700 AU Job Aid Booklet
DxC 700 AU Job Aid Booklet For Training Purposes Only These job aids are shortened versions of procedures found in the references below. Information in the job aid is correct as of the date published.
More informationInstallation of the 2 position/10 port valve into the 1100 series thermostatted column compartment
Agilent G1316A Thermostatted Column Compartment Technical Note Installation of the 2 position/10 port valve into the 1100 series thermostatted column compartment System Requirements 2 Performance Specifications
More informationElvaX Sulfur in Oil analyzer
ElvaX Sulfur in Oil analyzer Introduction ElvaX Sulfur in Oil was designed especially for analysis of petrochemicals for sulfur content in wide concentration range from ppm to 5%. Many international standards
More informationDetection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology
Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology Electronic Noses An electronic nose produces a recognizable response based
More informationSynthetic Fuel Formulation from Natural Gas via GTL: A Synopsis and the Path Forward
Synthetic Fuel Formulation from Natural Gas via GTL: A Synopsis and the Path Forward Elfatih Elmalik 1,2, Iqbal Mujtaba 1, Nimir Elbashir 2 1 University of Bradford, UK 2 Texas A&M University at Qatar
More informationComparison of IG Testing Standard. IGMA Summer Meeting August 8, 2004
Comparison of IG Testing Standard IGMA Summer Meeting August 8, 2004 1 The Task Compare the new Harmonized standards to: Old ASTM stds. Canadian stds. CEN stds. (Europe) ISO std. (proposed) 2 The Standards
More informationOperation and Applications of Differential Flow Modulation
Operation and Applications of Differential Flow Modulation H2 Collection channel Column 1 H2 Column 2 FID Roger L Firor, Ph.D. Agilent Technologies Chemical Analysis Group Wilmington, DE USA Flow Modulator
More informationAchieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)
Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Bill Wilson and Chin-Kai Meng Agilent Technologies, Inc. 2850 Centerville Road Wilmington,
More informationTechnical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)
Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph-mass spectrometer (GC-MS). 2.0 Scope
More informationS 1125 HPLC Pump System
Routine S 170 EXPERT HPLC HPLC System SYSTEMS 500 The Sykam ROUTINE HPLC System S 500 is intended for all routine analysis. The system configuration is highly variable and several upgrade options make
More informationUltra Low Sulfur Diesel Fuel For Use In Locomotive Marine
We have made it easy for you to find a PDF Ebooks without any digging. And by having access to our ebooks online or by storing it on your computer, you have convenient answers with ultra low sulfur diesel
More informationQuality-by-Design in Method Development. Mijo Stanic General Manager and Technical Director
Quality-by-Design in Method Development Mijo Stanic General Manager and Technical Director Agenda Chromicent who we are Traditional vs. Systematic strategies for method development QbD a new approach in
More informationPage 1. Alternate Carrier Gas Considerations and Faster GC Analysis
Page 1 Alternate Carrier Gas Considerations and Faster GC Analysis Faster GC Total Analytical Cycle Times A Variety of Approaches Pre-Run ALS Set-Up Chromatographic Run Post-Run Bake-Out Post-Run Cool-Down
More information