NANOpure Infinity Base Unit

Size: px
Start display at page:

Download "NANOpure Infinity Base Unit"

Transcription

1 BARNSTEAD THERMOLYNE CORPORATION NANOpure Infinity Base Unit OPERATING MANUAL AND PARTS LIST Series 896 Model # Voltage D D D LT896X1 5/11/98 Serial Number 1

2 Table of Contents Safety Information... 4 Alert Signals... 4 Warnings... 4 Introduction... 6 Specifications... 7 Dimensions and Clearance Requirements... 7 Feedwater Requirements... 7 Product Water... 7 Electrical Requirements... 8 Environmental Conditions... 8 Declaration of Conformity... 8 Unpackaging and Installation... 9 Unpackaging... 9 Choosing a Site... 9 Tubing Adapter Installation Remote Dispenser Housing Installing Remote Dispenser Housing on Bench Mounted Units Installing Remote Dispenser Housing on Wall Mounted Units Installing Remote Dispenser Housing in Remote Locations Or Accessories Bench Mounting Wall Mounting Water Connections Feed Water Connection Controls Main Power Switch Control Panel Switches Initial Operation Initial Sanitization Cartridge Installation and Rinse Up Normal Operation Remote Dispenser Operation Use of Standby Mode Selecting Set Point Resetting Sanitization Timer Reading Purity of Product Water and Feed Water With Optional Inlet Cell Accessory Installing Optional Float or Pressure Switch Installing Optional Inlet Cell Optional N.I.S.T. Calibration Module Maintenance and Servicing System Sanitization and Cartridge Replacement General Cleaning Instructions Micron Filter Replacement Fuse Replacement

3 TABLE OF CONTENTS Shutdown Problem Solving Replacement Parts Ordering Procedures Two Year Limited Warranty

4 Safety Information Alert Signals Warning Warnings alert you to a possibility of personal injury. Caution Cautions alert you to a possibility of damage to equipment. Note Notes alert you to pertinent facts and conditions. Your Barnstead Thermolyne NANOpure Infinity Base Unit has been designed with function, reliability, and safety in mind. It is your responsibility to install it in conformance with local electrical codes. This manual contains important safety information. You must carefully read and understand contents of this manual prior to use of this equipment. For safe operation, please pay attention to alert indicators throughout manual. Water purification technology employs one or more of following: chemicals, electrical devices, mercury vapor lamps, steam and heated vessels. Care should be taken when installing, operating or servicing Barnstead products. The specific safety notes pertinent to this Barnstead product are listed below. Warnings To avoid electrical shock, always: 1. Use a properly grounded electrical outlet of correct voltage and current handling capacity. 2. Do not locate NANOpure Infinity Base Unit directly over equipment that requires electrical service. Routine maintenance of this unit may involve water spillage and subsequent electrical shock hazard if improperly located. 3. Replace fuses with those of same type and rating. 4. Disconnect from power supply prior to maintenance and servicing. 4

5 SAFETY INFORMATION To avoid personal injury: 1. Do not use in presence of flammable or combustible materials; fire or explosion may result. This device contains components which may ignite such materials. 2. This device is to be used with water feeds only. Sanitizing/cleaning agents must be used in compliance with instructions in this manual. Failure to comply with above could result in explosion and personal injury. 3. Avoid splashing disinfecting solutions on clothing or skin. 4. Ensure all piping connections are tight to avoid chemical leakage. 5. Ensure adequate ventilation. 6. Carefully follow manufacturer s safety instructions on labels of chemical containers and material safety data sheets. 7. Depressurize system prior to disengaging cartridge hold-down bracket. 8. Refer servicing to qualified personnel. 9. Avoid contact of strong oxidizing agents, such as nitric acid, with ion exchange cartridges. An explosion may result. 5

6 Introduction Control Panel Remote Dispenser Remote Dispenser Housing Congratulations on your purchase of a Barnstead Thermolyne NANOpure Infinity Base Unit. The unit is a water purification system designed to provide high resistivity, reagent grade water that exceeds ASTM Type I, CAP and NCCLS Type I standards. It uses a four-stage deionization process combined with a 0.2 micron filter to polish suitable feed water (potable tap, distilled, deionized, or reverse osmosis) to produce water with a resistivity of up to 18.3 megohm-cm. Water resistivity is continuously monitored by a resistivity cell and displayed on a digital display. The NANOpure Infinity Base Unit can be easily upgraded to incorporate a UV (ultraviolet) system for organic determinant analysis, a UF (ultrafilter) system for such demanding applications as cell and tissue culture, in vitro fertilization and Monoclonal antibody production or a UV/UF (ultraviolet and ultrafilter) system for biological research requiring water free of virtually all contamination. The electronics can be verified and calibrated utilizing N.I.S.T. Traceable Calibration module. See accessory ordering information (page 41). Please read instructions carefully to ensure that you receive maximum benefit from NANOpure Infinity Base Unit. Also, be sure to fill out and return enclosed warranty registration card. We would like to receive information requested, and it will help us assure you of proper warranty coverage. General Usage Do not use this product for anything or than its intended usage. Figure 1 NANOpure Infinity Base Unit Front 6

7 Specifications Dimensions and Clearance Requirements Dimensions (include remote dispenser housing mounted to unit ) Wall mounted models Width 20 (50.8 cm) Depth 17-1/2 (44.5 cm) Height 34-7/8 (88.6 cm). Bench mounted models Width 25 (63.5 cm) Depth 17-1/2 (44.5 cm) Height 22-5/8 (57.5 cm) Dimensions (remote dispenser housing mounted away from unit) Wall and Bench mounted models Width 20 (50.8 cm) Depth 17-1/2 (44.5 cm) Height 22-5/8 (57.5 cm) Clearances Sides - 6 (15 cm) minimum for servicing. Above - 3 (7.6 cm) minimum for removal of outer case. Front - 20 (50.8 cm) minimum for opening front door. Cartridge replacement is easily accomplished by opening front door. Feed Water Requirements Types 1 Tap (Potable), RO, DI, distilled. TOC Less than 1.0 ppm. Turbidity 1.0 N.T.U. maximum. Pressure Range Gravity feed to 100 psig (7kg/cm 2 ) maximum. Temperature Range 4.4 C C ( F) Product Water Water Quality Resistivity TOC Flow Rate ASTM, CAP and NCCLS Type I Less than 3.0 ppb 1.5 lpm maximum at minimum inlet feed water pressure of 30 PSIG at 60 HZ and with a new final filter. 1 Suitability as qualified by laboratory analysis. 7

8 SPECIFICATIONS Electrical Requirements The NANOpure Infinity Base Unit is equipped with a power cord to be plugged into an electrical outlet of appropriate voltage. Voltage and Frequency (Nominal) Model D VAC, Hz VAC, Hz, 1 phase Model D VAC, Hz VAC, Hz, 1 phase Model D VAC, Hz VAC, Hz, 1 phase Environmental Conditions Operating: 4 C - 49 C; 20% to 80% relative humidity, non-condensing. Installation Category II (over-voltage) in accordance with IEC 664. Pollution Degree 2 in accordance with IEC 664. Altitude limit: 3,500 meters. Storage: -25 C to 65 C; 10% to 85% relative humidity. Declaration of Conformity (-33 models only) Barnstead Thermolyne hereby declares under its sole responsibility that this product conforms with technical requirements of following standards: EMC: EN Generic Emission Standard; EN Generic Immunity Standard; Safety: IEC Safety requirements for electrical equipment for measurement, control, and laboratory use; Part I: General Requirements per provisions of Electromagnetic Compatability Directive 89/336/EEC, as amended by 92/31/EEC and 93/68/EEC, and per provisions of Low Voltage Directive 73/23/ EEC, as amended by 93/68/EEC. The authorized representative located within European Community is: European Manager Barnstead Thermolyne Saarbrückener Str. 248 D Braunschweig Germany Copies of Declaration of Conformity are available upon request. 8

9 Unpackaging and Installation Caution Wall composition, condition and construction, as well as fastener type, must be considered when mounting this unit. The mounting surface and fasteners selected must be capable of supporting a minimum of 275 lbs. Inadequate support and/ or fasteners may result in damage to mounting surface and/or equipment. If you are unsure of mounting surface composition, condition and construction or correct fasteners, consult your building maintenance group or contractor. Warning Do not locate NANOpure Infinity Base Unit directly over equipment that requires electrical service. Routine maintenance of this unit may involve water spillage and subsequent electrical shock hazard if improperly located. Do not use in presence of flammable or combustible materials; fire or explosion may result. The device contains components which may ignite such materials. Unpackaging 1. Remove unit from its shipping container. Remove all contents carefully. Ensure that feed tubing, sanitization cartridge, remote dispenser housing and its two brackets, wall bracket, accessory parts bag and power cord are removed from packaging materials before discarding. Put NANOpure Infinity Base Unit on a bench. Choosing a Site The NANOpure Infinity Base Unit system features a pivoting control panel display and a dispenser which allow system to be mounted almost anywhere within laboratory. Use wall bracket for wall mounted systems as a template to drill mounting holes. (The NANOpure Infinity Base Unit does not include screws and fasteners for mounting.) Allow a minimum of 6 inches (15 cm) clearance on all sides of unit for servicing, 3 inches (7.6 cm) on top for outer case removal and 20 inches (50.8 cm) in front for opening door. Note The outlet of a gravity feed storage reservoir must be above or at same level as inlet of NANOpure Infinity Base Unit. 9

10 INSTALLATION Tubing Adapter Installation The following instructions will apply when you need to attach a piece of tubing to your NANOpure Infinity Base Unit during installation, unless orwise noted in installation instructions. To make tubing connections: 1. Completely disassemble fitting. Refer to Figure 2 to familiarize yourself with names of component parts. 2. Make sure tubing is cut off reasonably square and that no plastic burrs or ridges are present. 3. Place grab ring and backup ring in hex nut in order and orientation shown in Figure 2. Thread nut into adapter. DO NOT use o-ring at this time. 4. Push tubing through nut until it bottoms out in adapter. 5. Remove adapter nut and tubing. Place o-ring over tubing. Be careful not topush backup ring or grab ring furr back on tubing when installing o-ring. 6. Install hex nut on adapter and hand tighten. Figure 2 Typical Polypropylene Tubing Adapter Installation 10

11 INSTALLATION Remote Dispenser Housing The Infinity unit comes complete with a remote dispenser allowing you to deliver water up to eight feet (2.4 meters) away from unit. Install remote dispenser housing on bottom right of unit if wall mounting or top right side of unit if bench mounting. The dispenser can also be remotely installed up to six feet (1.8 meters) from right hand side of unit. An accessory parts bag shipped with unit contains seven hole plugs, six screws (three flat head, three pan head) and one nut. You will also need to find two small brackets packed separately in Infinity box. Installing Remote Dispenser Housing on Bench Mounted Units 1. Remove locking screw located on right hand side of Infinity unit between rails. 2. Line up two slots on left side of remote dispenser housing with top of two rails on Infinity unit. Slide remote dispenser housing down onto rails. 3. Replace locking screw to secure remote dispenser housing. 11

12 INSTALLATION Installing Remote Dispenser Housing on Wall Mounted Units If you are installing remote dispenser housing on a wall mounted unit, remove two rails and locking screw on right side of Infinity unit. Replace screws with hole plugs provided. Retain two rails for possible future use. Prior to installing NANOpure Infinity Base Unit on a wall, lay unit on its back to attach remote dispenser housing. Bottom View of NANOpure unit Remote Dispenser Housing Short Leg Bracket L-shaped Bracket B 1. Remove two screws from back of remote dispenser housing. Keep m. You will use m later. 2. Place L-shaped bracket to left and back sides of remote dispenser housing. Replace two screws (1/2 x 8/32 pan head) removed in step 1 into back of housing to attach bracket (See Figure 3, A). 3. Place one flat head screw (1/2 x 8/32 ) into left side bottom of L-shaped bracket. Secure it with nut supplied. (See Figure 3, B) 4. Attach remaining bracket with two pan head screws(1/2 x 8/32 ) to bottom right side of Infinity unit with short legs pointing to right. (See Figure 3, C). 5. Locate remote dispenser housing/ L-shaped bracket assembled in step 2 and two remaining flat head screws. Line up two holes in bracket mounted on bottom of Infinity unit with two holes in remote dispenser housing/l-shaped bracket. Secure with two flat head screws. (See Figure 3, D). D C B Figure 3 Remote Dispenser Housing Installation 12 A 6. Screw in remaining pan head screw through remote dispenser housing with bracket in hole on far right of unit, where y line up.

13 INSTALLATION Remote Dispenser Housing lying on its top B L-shaped Bracket Short Leg Bracket A Installing Remote Dispenser Housing in Remote Locations If you are installing remote dispenser housing in a remote location away from Infinity unit, remove two rails and locking screw on right side of unit. Replace screws with hole plugs provided. Retain two rails for possible future use. The remote dispenser housing can be installed up to six feet (1.8 meters) away from unit. 1. Remove two screws from back of remote dispenser housing. Keep m. You will use m later. 2. Lay remote dispenser housing on its top. Place L-shaped bracket to right and back sides of remote dispenser housing. Replace two screws (1/2 x 8/32 pan head) removed in step 1 into back of housing to attach bracket (See Figure 4, A). 3. Place one flat head screw (1/2 x 8/32 ) into left side bottom of L-shaped bracket. Secure it with nut supplied. (See Figure 4, B). 4. Locate short leg bracket and two flat head screws (1/2 x 8/32 ). Line up two holes on short legs with two holes on remote dispenser housing. The long leg should point up. Screw in two flat head screws to attach bracket. (See Figure 4, C). Figure 4 Remote Dispenser Housing Installation in Remote Locations C B 5. Put two screws (customer supplied) in wall 3-1/2 inches apart on a straight horizontal line. Do not put screws all of way into wall. The space between bottom of head on screw and wall should be 1/8-inch. 13

14 INSTALLATION 6. Line up key hole slots on remote dispenser housing with two screws in wall. Slide remote dispenser housing down onto screws. 7. Put two more screws (customer supplied) into two remaining holes in remote dispenser housing to secure remote dispenser housing bracket to wall. 8. Install two hole plugs into two holes in side of bracket. Lock dispenser into its housing by placing tubing end of dispenser into U-shaped groove in housing and pushing it toward back of unit. You know it is locked when you hear a click. After remote dispenser is fully inserted into pivoting holder you must rotate down. To remove dispenser from its housing, lift/ rotate up and pull dispenser straight out. Or accessories Optional accessories for this unit include an inlet cell, float switch, low pressure switch and N.I.S.T. Traceable Calibration Module. Installation procedures for se accessories are explained in later sections of this manual. It is best to install se accessories prior to mounting NANOpure Infinity Base Unit. Bench Mounting 1. Place NANOpure Infinity Base Unit on a bench top that is accessible to water and electricity and that is convenient to your work area, noting clearance requirements. 14

15 INSTALLATION Warning Do not locate NANOpure Infinity Base Unit directly over equipment that requires electrical service. Routine maintenance of this unit may involve water spillage and subsequent electrical shock hazard if improperly located. Do not use in presence of flammable or combustible materials; fire or explosion may result. The device contains components which may ignite such materials. Note The outlet of a gravity feed storage reservoir must be above or at same level as inlet of NANOpure Infinity Base Unit. Caution Wall composition, condition and construction, as well as fastener type, must be considered when mounting this unit. The mounting surface and fasteners selected must be capable of supporting a minimum of 275 lbs. Inadequate support and/or fasteners may result in damage to mounting surface and/or equipment. If you are unsure of mounting surface composition, condition and construction or correct fasteners, consult your building maintenance group or contractor. Wall Mounting Install NANOpure Infinity on a wall in a convenient location that is accessible to water and electricity. 1. Locate wall bracket packed separately from unit. 2. Using wall bracket as a template, locate and drill mounting holes in wall. A minimum of four (customer-supplied) fasteners will be required two on top and two on bottom. 3. Attach wall bracket to wall using customer-supplied fasteners. 4. Remove locking screws on each side of wall bracket. 5. Pull two locking slides on each side of wall bracket out as far as y will go. 6. Hang Infinity unit on wall bracket by sliding mounting pins into wall bracket slots. 7. Push locking slides on each side of wall bracket in as far as y will go. 8. Replace locking screws. Note Prior to installing NANOpure Infinity Base Unit on wall, ensure that remote dispenser housing and/or or accessories such as Inlet Cell are installed. See appropriate section of this manual for installation of accessories. 15

16 INSTALLATION Water Connections Feed Water Connection 1. Locate length of 3/8 O.D. tubing provided with a quick disconnect insert on one end and a 3/8 O.D. X 1/4 NPT tubingadapter on or. 2. Install tubing adapter onto your incoming water line. We recommend a customer supplied shut off valve be installed in your feed water line. Do not connect feed water to your NANOpure Infinity Base Unit. You will connect this during initial sanitization procedure. Pump Interlock Plug Main Power Switch Feed Water Inlet Connection Figure 5 Feed Water Connection (Left Side with Front of Unit Facing You) 16

17 Controls Familiarize yourself with controls prior to proceeding. More detailed information is available on pages Main Power Switch The main power switch on NANOpure Infinity Base Unit is located on lower left side of unit (as you face front of unit), directly above power cord receptacle. Control Panel The NANOpure Infinity Base Unit is controlled through a panel which incorporates six switches to control its functions: START/STOP, STANDBY, MODE, DOWN, UP and ENTER. This control panel utilizes a digital display to show system information and purity of product water. It will also show purity of feed water when an optional inlet cell is utilized. Two hidden LEDs illuminate to inform you it is time to sanitize unit (SAN) or replace UV lamp (UV) in UV models. The SAN light illuminates after approximately six months of use. The UV light illuminates after approximately three months of use in models which utilize a UV lamp. The control panel can be adjusted to angle you need for viewing digital display. To adjust angle of control panel: START/STOP STANDBY MODE DOWN UP ENTER SAN UV 1. Open front door. Inside door are two rectangular openings which show two screws held in place by two wing nuts, one in each opening. Figure 6 NANOpure Infinity Base Unit Controls 17

18 CONTROLS 2. Turn wing nuts to left to unlock and loosen m. The control panel will now pivot top to bottom in its opening. Reposition control panel to your desired viewing position. 3. Lock control panel in place by turning wing nuts to right until y are hand-tightened. Switches When main power switch (on lower left side of unit) is on, six switches on control panel function as follows: START/STOP allows you to turn unit on or off. STANDBY allows you to put unit into standby, recirculating water for 10 minutes every hour. When in standby, display shows Standby Mode during periods of inactivity. During 10- minute recirculation, display reads Recirculation. MODE allows you to change from purity reading to maintenance prompts and activities. DOWN and UP allow you to scroll between different commands and values. ENTER allows you to activate a certain mode. 18

19 Warning Avoid splashing disinfecting solutions on clothing or skin. Ensure all piping connections are tight to avoid leakage. Ensure adequate ventilation. Carefully follow manufacturer s safety instructions and material safety data sheets. This device is to be used with water feeds only. Sanitizing/cleaning agents must be used in compliance with instructions in this manual. Failure to comply with above could result in explosion and personal injury. Use a properly grounded electrical outlet of correct voltage and current handling capacity. Note The top cartridge nipple is one with right-angle turn and one flange. The bottom cartridge nipple extends straight out from cartridge. The flange on top cartridge nipple should be able to slide down keyway wall behind sheet metal. Caution Make sure cartridge hold-down bracket is in place. Initial Operation Initial sanitization Your NANOpure Infinity Base Unit has been shipped with a sanitization cartridge (Catalog No. D50258) that you will install. Three empty cartridges have been installed in unit, allowing you to sanitize your NANOpure Infinity Base Unit prior to using it for first time. To sanitize your NANOpure Infinity Base Unit: 1. Open front door. Disengage cartridge hold-down bracket by pulling it out and up. Remove D50258 sanitization cartridge from its bag. Wet cartridge o-rings with water. Install D50258 sanitization cartridge in empty cartridge position (#4). Press top cartridge nipple into upper socket until it bottoms out. Then, lower cartridge inserting bottom cartridge nipple into lower socket until it is firmly seated. Replace cartridge hold-down bracket. Close front door. 2. Attach feed water line (from step 2 Feed Water Connection, see Figure 5) to unit by snapping quick disconnect coupling into quick disconnect body in lower left side of unit. 3. Plug power cord into unit s power entry module and plug into a live outlet. Place main power switch to I (on) position. Place a suitable container under remote dispenser. Depress trigger to open remote dispenser. The remote dispenser can be locked by depressing button on its right side while depressing trigger. 19

20 INITIAL OPERATION Note To disrupt flow of water to NANOpure Infinity Base Unit, press stainless steel thumb pad on quick disconnect fitting. The insert with valve is easily removed. To resupply Infinity unit with water, Attach feed water line (from step 2 Feed Water Connection) to unit by snapping quick disconnect coupling into quick disconnect body in lower left back of unit. Note Product purity during sanitization will be very low. Do not pay close attention to se readings or below set point indication during sanitization procedure. Note The cartridges will still contain water when removed. Place cartridges upside down in a sink, bucket or or waterproof container to drain m after removal. 4. Push START/STOP switch to start. 5. Run water to drain for 30 seconds, close dispenser by depressing handle. Depressing handle will unlock button and close dispenser. 6. Recirculate water for 30 minutes. 7. After 30 minutes, reset sanitization timer according to instructions on page 27 and n press START/STOP switch to turn unit off. Remove feed water line. Turn main power switch to O position. 8. Open remote dispenser, lock it open and point it down a drain or into a suitable container. Depressurize unit until you are no longer getting water out of remote dispenser. 9. Open front door. Disengage cartridge hold-down bracket. Remove all of cartridges by pulling first up and n out. Drain water from three empty cartridges and keep m for future use. Throw away sanitization cartridge. 10. Proceed to Cartridge Installation and Rinse Up. Warning Depressurize system prior to disengaging cartridge hold-down bracket. 20

21 INITIAL OPERATION Note The NANOpure Infinity Base Unit cartridges must be installed in proper order. If you have just sanitized your NANOpure Infinity Base Unit, retain empty cartridges shipped with your unit for use during future sanitization procedures. Discard used D50258 sanitization cartridge. The top cartridge nipple is one with right-angle turn and one flange. The bottom cartridge nipple extends straight out from cartridge. The flange on top cartridge nipple should be able to slide down keyway wall behind sheet metal. Cartridge position by number Figure 7 Proper Cartridge Order Cartridge Installation and Rinse Up 1. Open front door. 2. The cartridges will be installed according to charts below. Position 1 is farst left, position 2 next one to right, position 3 next one to right and position 4 farst right. (See Figure 7). Proceed to step 3. Type 1 Cartridge Kit Tap/DI Feed #D50256 Position Type Catalog# 1 Pretreatment D Ultrapure D Ultrapure D Ultrapure D50253 ORGANICfree Cartridge Kit Tap/DI Feed #D50254 Position Type Catalog# 1 Pretreatment D Ultrapure D Ultrapure D ORGANICfree D50252 Type 1 Cartridge Kit R/O or Distilled Feed #D50257 Position Type Catalog# 1 Pretreatment D Ultrapure D Ultrapure D Ultrapure D50253 ORGANICfree Cartridge Kit R/O or Distilled Feed #D50255 Position Type Catalog# 1 Pretreatment D Ultrapure D Ultrapure D ORGANICfree D

22 INITIAL OPERATION 3. Remove protective bag and wet o-rings with water on both cartridge nipples. Note Ensure both o-rings of cartridge are down inside connector before moving onto next cartridge installation. Caution Make sure cartridge hold-down bracket is in place. Caution Do not allow NANOpure Infinity Base Unit to operate unless water is available to unit. 4. Install appropriate cartridge in cartridge position #1 by pressing top cartridge nipple into upper socket until it bottoms out. 5. Lower cartridge and insert bottom cartridge nipple into lower socket until it is firmly seated. 6. Repeat steps 3-5 with remaining cartridges in your kit, placing m in next three positions from left to right according to Cartridge Kits charts on page 21. Do not install 0.2 micron filter and bell assembly at this time. 7. Replace cartridge hold-down bracket. 8. Close front door. 9. Reattach feed water line to unit by snapping quick disconnect coupling into quick disconnect body in lower left back of unit. 10. Turn unit on by turning main power switch to I (on) position. 11. Point outlet of remote dispenser down a drain or into a suitable container. Depress draw-off trigger and lock it open. 12. Press START/STOP switch to start. 22

23 INITIAL OPERATION Note For more demanding applications where low TOC water is required, a rinse of liters through cartridges and filter may be necessary. Caution Do not overtighten 0.2 micron filter assembly onto Luer fitting or use excessive force in seating it. The filter and/or Luer fitting can be damaged by overtightening or excessive force. 13. Run product water to drain for 5 minutes through remote dispenser. Close remote dispenser and let water recirculate until resistivity reaches your desired purity level. (See note on this page). 14. Remove a new 0.2 micron filter and bell assembly from its bag and insert it into Luer fitting on remote dispenser. Gently turn it clockwise until it is fully seated in Luer fitting. 15. Remove protective cap from filter bell. Open remote dispenser and flush 1-2 liters of water through 0.2 micron filter. The product water is now ready for use. (See note on this page). 23

24 Normal Operation Note On initial start-up, purity meter may display "Prod.... MΩ-cm". This is caused by air in cell and should be replaced by a resistivity reading almost immediately. If dots aren t replaced by a value after one minute, refer to Problem Solving section for this manual. 1. Turn system power on by depressing main power switch to I position. 2. Press START/STOP switch on control panel. 3. The system greeting lights indicate which module, if any, is installed on base unit. This greeting is one of following: NANOPURE, NANOPURE UV, NANOPURE UF, or NANOPURE UV/UF. Your base unit can be upgraded to include any of three modules: UV module (Catalog No. D8974), UF module (D8984) or UV/UF module (D8994). While system greeting is displayed for approximately five seconds, unit checks for presence of an inlet cell. 4. The system electronics will check its calibration. (All units have been factory calibrated.) If calibration is OK display will show Calibration OK. After this it will show Prod.... MΩ-cm. Then display will read Prod. 10 (±0.2) MΩcm. This is a reading of calibration reference value. 5. The display will n begin reading resistivity of product water. Initially it will read Prod.... MΩ-cm until air is bled from system. 6. Allow water s resistivity to rise to desired purity before drawing off water. 7. The system should be left on or in standby during work day. (See Use of Standby mode.) 24

25 NORMAL OPERATION Remote Dispenser Operation Lock dispenser into its housing by placing tubing end of dispenser into U-shaped groove in housing and pushing it toward back of unit until it stops and n pull down. You know it is locked when you hear a click. To operate: 1. Remove protective cap from filter bell. 2. Depress draw-off trigger. Note Do not turn off NANOpure Infinity Base Unit during non-work hours. Doing so will allow bacterial growth and or contamination of water in system. As a result, system will require a lengthy rinse-up period at beginning of work day to achieve high-quality product water. We recommend using Standby Mode. 3. When draw off is complete, release draw-off trigger and replace protective cap on filter bell. 4. For unattended operation, depress button on right side of dispenser at same time as trigger is depressed. 5. To unlock draw-off trigger and stop dispensing water, press draw-off trigger. To remove dispenser from its housing, pull it up and out. See page 14 for additional information. Use of Standby Mode At end of work day, press Standby switch on front of NANOpure Infinity Base Unit to place unit in Standby mode for night. Press STANDBY switch and display will read Standby Mode for 50 minutes and n go into a 10-minute recirculation mode. While pump is energized, display will read Recirculation. 25

26 NORMAL OPERATION Selecting Set Point The NANOpure Infinity Base Unit electronics include a user programmable set point which alerts you when water quality falls below programmed set point. The set point is user selectable from 0-18 MΩ-cm. The display will alternate actual resistivity measurements with a Below Setpoint message when measurement is below set point. The set point indication is set at factory at 10 MΩ-cm. You must start this procedure from off (stop) position. 1. Press and hold ENTER switch. 2. Press START/STOP switch. 3. Release switches. 4. The display will read Setpt. 10 MΩ-cm. 5. To adjust value, press UP or DOWN arrow until your desired value is displayed and press ENTER. If resistivity of product water falls below set point value, display will read resistivity value for approximately seven seconds and a Below Setpoint display for approximately three seconds. Once resistivity value rises above set point, unit will only read resistivity. 26

27 NORMAL OPERATION Resetting Sanitization Timer 1. While NANOpure Infinity is operating, press MODE switch. 2. The display will read Reset SAN Timer 3. Press ENTER switch to reset Sanitization Timer. 4. The display will read Enter to Reset 5. Press ENTER switch again and timer will be reset. This will reset sanitization timer for six months. Note Press ENTER button at any step and unit will go to resistivity of product water. Pressing UP and DOWN switches allows scrolling between various purity of product water and feed water displays. Reading Purity of Product Water and Feed Water With Optional Inlet Cell Accessory 1. Turn system power on by depressing main power switch to I position. 2. Press START/STOP switch on control panel. 3. The display will show type of system for approximately five seconds. 4. The system electronics will check its calibration. All units have been factory calibrated. If calibration is OK display will show Calibration OK. After this it will show Prod.... MΩ-cm. Then display will read PROD. 10 (±0.2) MΩ-cm. This is a reading of calibration reference value. 27

28 NORMAL OPERATION 5. The display will n begin reading resistivity of product water. Initially it will read Prod.... MΩ-cm until air is bled from system. 6. Press DOWN switch to: Display conductivity of product water Display Total Dissolved Solids of product water Display temperature of product water 7. If no inlet cell is used, reading will return to resistivity of product water. If an inlet cell is attached, it will read following: Display resistivity of inlet water Display conductivity of inlet water Display Total Dissolved Solids of inlet water Display temperature of inlet water Display resistivity of product water 28

29 Installing Float or Pressure Switch Warning Disconnect from power supply prior to maintenance and servicing. Accessories D8964 (float switch) and D2706 (pressure switch) are designed to protect NANOpure Infinity Base Unit pump by alerting NANOpure Infinity Base Unit of an inadequate feed water condition so that pump can be shut down. If an inadequate feed water condition exists and NANOpure Infinity Base Unit pump is shut down due to this condition, display will read Check Inlet. Use following instructions for installation. 1. Disconnect unit from electrical power. 2. If using D8964 float switch, follow installation instructions included with float switch for installation into a tank. 3. If using D2706 low pressure switch, install PVC tee (supplied with D2706) in incoming water line. Screw switch into top of tee, n connect inlet tubing to NANOpure Infinity Base Unit with remaining opening. 4. Route cable from float or low pressure switch to top left of NANOpure Infinity Base Unit. 5. Remove jumper plug from back, left, upper portion of unit and save for future use. 6. Plug cable into jumper plug outlet. 7. Reconnect electrical power. 29

30 Installing Inlet Cell The NANOpure Infinity Base Unit system offers, as an optional feature, ability to monitor resistivity of NANOpure Infinity Base Unit inlet water. To install inlet cell: 1. Disconnect inlet water and electrical service to unit. Open Remote Dispenser to depressurizeunit. 2. Open front door. Inlet Cell Cable Inlet Cell Inlet Cell Well 3. Remove screws holding Infinity unit s case in place. As unit faces you, re are three screws in left side near back, three in right side near back and five in front, two on each side and one above cartidge hold-down bracket. Remove jumper plug, Float Switch or Pressure Switch Cable from back, left, upper portion of unit. 4. Remove remote dispenser from its housing. 5. Lift up to remove outer case. 6. Remove electric section cover directly behind cartridges. 7. The inlet cell well is located in lower front left of unit. (see Figure 8). Figure 8 Inlet Cell Installation and Location 30

31 INSTALLING THE INLET CELL 8. Remove 1/2-inch plug located in front of cell well. 9. Install inlet cell by turning clear plastic bushing clockwise. Do not turn black snap cap. Hand-tighten bushing. Do not use a wrench. Overtightening may damage seal. 10. Thread inlet cell cable through large hole in lower left of unit and back into unit through anor large hole in upper left of same side of unit. 11. Connect white plastic cell connector to position P3 on resistivity circuit board (PC896X2) located on inside back wall of unit. The flat part of connector should face upward. 12. Replace electric section cover. Outlet Cell Connection Position P2 Inlet Cell Connection Position P3 Inlet Cell Cable 13. Replace outer case. Replace screws you removed to take off case. Reattach jumper plug, Float Switch or Pressure Switch you removed earlier. 14. Reconnect feed water. Figure 9 Inlet Cell Cable Connection to Circuit Board 31

32 N.I.S.T. Calibration Module Performing an Electronic Calibration Using Optional N.I.S.T. Calibration Module If you purchased optional N.I.S.T. calibration module (Catalog No. E896X5) you can perform a calibration of NANOpure Infinity purity sensing electronics traceable to N.I.S.T. standards. 1. Disconnect electrical service to unit. Open remote dispenser to depressurize unit. 2. Open front door. 3. Remove screws that hold Infinity unit s outer case in place. As you face unit, re are three screws in left side near back, three in right side near back and five in front, two on each side and one above cartridge hold-down bracket. Remove jumper plug or float/pressure switch from back, left, upper portion of unit. 4. Remove remote dispenser from its housing. 5. Lift and remove outer case. 6. Remove electric section cover directly behind cartridges. 7. Locate outlet cell connection on circuit board (see figure 9) and remove outlet cell connection. 32

33 N.I.S.T. CALIBRATION MODULE 8. Connect N.I.S.T. calibration module to connection point (J2), To Outlet Cell on calibration board (PC896X3A). 9. Replace electric section cover removed in step Plug unit into a live outlet and turn unit on by depressing main power switch. 11. Press and hold DOWN switch on controls. 12. Press START/STOP switch. 13. Release switches. 14. The display will show Calibrate?, press ENTER. Note The calibration module can be returned to Barnstead Thermolyne for recertification. Please contact Barnstead Thermolyne for instructions. 15. Display will read Calibrating. When calibration is complete display will show Calibration Done. 16. Turn main power switch to O (off) position. Unplug unit. 17. Remove calibration module and reconnect cell. 18. Replace outer case. Replace screws you removed earlier. Reattach jumper plug or float/pressure switch you removed earlier. 33

34 Maintenance and Servicing Note The Sanitization Timer will illuminate after six months, reminding you to sanitize unit. Complete system sanitization procedure and reset sanitization lamp/ indicator by going through Reset Sanitization Timer procedure on page 27. Warning Disconnect from power supply prior to maintenance and servicing. Note The cartridges will still contain a cartridge bed of water when removed. Therefore, you will want to have a sink, bucket or or waterproof container available to place m in after removal. Warning Depressurize system completely prior to lifting cartridge hold-down bracket and removing cartridges. System Sanitization and Cartridge Replacement The frequency with which you will need to clean your unit and replace your cartridges is dependent on your feed water s characteristics, your purity requirements and your usage. Sanitize your NANOpure Infinity Base Unit and replace cartridges when product water purity drops below acceptable levels of resistivity, when organic levels become too high, or if a new 0.2 micron filter clogs rapidly after installation even though cartridges were thoroughly rinsed before 0.2 micron filter was installed. To sanitize NANOpure Infinity Base Unit, purification cartridges must be replaced with a sanitization cartridge and three empty cartridges supplied with your unit. The simple-touse sanitization cartridge is available from BarnsteadlThermolyne (Catalog Number D50258). This is used in addition to empty cartridges included with this unit to effect a complete sanitization. 1. Disconnect electrical service to unit. 2. Remove feed water line by depressing stainless steel thumb pad. 3. Depressurize system by opening remote dispenser. 4. Open front door. 5. Be aware that exhausted cartridge will contain a cartridge bed of excess water, you will want to have a container available to place it in after removal Disengage cartridge hold-down bracket.

35 MAINTENANCE AND SERVICING Note The flange on top cartridge nipple should be able to slide down keyway wall behind sheet metal. 7. Remove exhausted cartridge in right-hand (#4) position by pulling it first up and n out. 8. Remove a D50258 sanitization cartridge from its packaging. Press top cartridge nipple of D50258 sanitization cartridge into upper socket in right-hand (4) position until it bottoms out. 9. Lower cartridge and insert bottom cartridge nipple into lower socket until it is firmly seated. 10. Install three empty cartridges supplied with your NANOpure Infinity Base Unit in positions 1, 2 & Reposition cartridge hold-down bracket. 12. Close front door. Remove 0.2 micron filter and bell assembly. 13. Sanitize, install new cartridges and rinse according to instructions for Initial Sanitization and Cartridge Installation in Initial Operation section. General Cleaning Instructions Disconnect electrical service to unit. Wipe exterior surfaces with lightly dampened cloth containing mild soap solution. 35

36 MAINTENANCE AND SERVICING Caution Do not overtighten 0.2 micron filter assembly onto Luer fitting or use excessive force in seating it. The filter and/or Luer fitting can be damaged by overtightening or excessive force. Note If a newly installed 0.2 micron filter clogs rapidly after installation, NANOpure Infinity Base Unit may need to be sanitized to remove bacterial contaminants. See System Sanitization. Warning Replace fuses with those of same type and rating. 0.2 Micron Filter Replacement Replace 0.2 micron final filter whenever any of following conditions occur: every 30 days, product water flow rate is reduced or bacteria break through. The 0.2 micron filter is shipped assembled with a bell. To replace 0.2 micron filter assembly: 1. Remove old 0.2 micron filter assembly by turning it counterclockwise until it is free from Luer fitting. 2. Remove new 0.2 micron filter assembly from its bag and insert it into Luer fitting. Gently turn it clockwise until it is fully seated in Luer fitting. 3. Rinse 1-2 liters of water through filter to drain prior to using product water. Fuse Replacement 1. Turn off NANOpure Infinity Base Unit and disconnect it from power supply by removing power cord directly below main power switch. Locate fuse drawer above power cord receptacle. 2. Pull out fuse drawer located in power entry module. 3. Remove old fuses and replace with fuses of same type and rating. (See Parts Listing.) 4. Replace fuse drawer. 36

37 MAINTENANCE AND SERVICING 5. Reattach power cord and reconnect unit to power supply. 6. Operate normally. Shutdown If NANOpure Infinity Base Unit is to be shut down for an extended period of time, unit should be completely drained and cartridges removed to prevent growth of bacteria. If system has remained inactive and full of water for more than 96 hours, unit should be drained, sanitized and new cartridges installed prior to use. 37

38 38 roblem P s Cause ossible P s Solution completely Infinity NANOpure operating, not (pump inactive active). not panel control NANOpure to power electrical No Infinity. blown. fuse Main cord power Infinity NANOpure Ensure source power live a to connected is electrical into plugged completely and module entry power as well as outlet in switch on/off Power unit. on position. (on) I in indicated as fuse main Replace section Replacement" "Fuse 36). (page flow product no or Reduced Dispenser. Remote from clogged. filter micron.2 0 e th in indicated as filter final Replace Replacement" Filter Micron "0.2 36). (page manual of section Inlet" "Check reads Display run. not does pump and reservoir), (in protector Pump or switch pressure line water feed to connected not plug jumper interlock. pump or pressure water feed Insufficient NANOpure satisfy to volume Unit. Base Infinity jumper protector, pump Connect to cord switch pressure or plug left upper on located receptacle NANOpure of section hand Infinity. connect off, shut line water feed Open Infinity NANOpure to line water feed sufficient ensure and Unit Base exceed must pressure (line pressure kg/cm (0.35 psig 5 2 t sufficien or ), water feed in level water reservoir). 1". "Error reads isplay D. error communication PC board failure. System electronics (turn Infinity NANOpure down Power unit. restart and off) printed meter resistivity Replace board. circuit 2". "Error reads isplay D. system in Air connection. cell Bad through water more Run air purge to Infinity NANOpure unit. from it as cell of connection Check board. circuit to attaches 3". "Error reads isplay D t rese has microprocessor Meter operation. during failure. System electronics (turn Infinity NANOpure down Power unit. restart and off) printed meter resistivity Replace board. circuit 4". "Error reads isplay D. calibration of out eter M g utilizin procedure Perform calibration calibration N.I.S.T. optional module. Problem Solving

39 39 roblem P e Caus ossible P s Solution - M... "Prod. reads Display seconds. 10 than more for cm" cell. in Air connection. cell Bad your started just you If one wait Infinity, NANOpure itself. corrects it if see to minute cell of connection Check circuit to attaches it as board. rinse not will water Recirculated level. purity desired to up cartridges. Exhausted order. of out Cartridges as cartridges all Replace "Cartridge in indicated 35) (page section Replacement" manual. this of in cartridges Install in indicated as order proper manual. in located chart life. cartridge hort S e ar used being Cartridges date. expiration beyond water. feed quality Poor date. expiration Check capacity lose to begin Cartridges years two stored being after manufacture. of date from with cartridges Replace ones. unexpired is ROpure Barnstead a If that check source, water feed functioning is membrane properly. feed is Still Barnstead a If that ensure source, water to temperature distillate not does Infinity NANOpure 49 C (120 C). exceed water feed is water tap If of quality check source, quality cases some In water. with change will water of source Changing seasons. well or water, well to water (city in result will water) city to water change. quality water a central from a is water feed If verify system, purification water proper and quality water system. of functioning PROBLEM SOLVING

40 Replacement Parts Recommended Spares Consumables Consumable parts are those required to support day-to-day operation of this equipment. Barnstead Thermolyne establishes two types of consumables; those items that must periodically be replaced to maintain performance (filters, resin cartridges, etc.) and or items of limited life (indicator lights, fuses, etc.) that you can expect to replace on a more or less random basis. Where practical, Barnstead Thermolyne recommends frequency of replacement, or provides information on life expectancy from which you may calculate a replacement interval compatible with your usage pattern. The replacement of consumable parts is discussed in Maintenance and Servicing section to assist you in accomplishing your own service. Consumables may be ordered separately and in some cases, as an expendables kit. Check with your Barnstead Thermolyne representative for additional information on expendables kit. Recommended Description Catalog No. Quantity Type 1 Cartridge Kit Tap/DI Feed D ORGANICfree Cartridge Kit Tap/DI Feed D Type 1 Cartridge Kit R/O or Distilled Feed D ORGANICfree Cartridge Kit R/O or Distilled Feed D Remote Dispenser Final Filter FL703X2 2 Mains Fuse, 120 Volt - Type T Time Lag, 3.0 Amp, 250 Volt Mains Fuse, 230 Volt - Type T Time Lag, 2.0 Amp, 250 Volt Teflon Tape, Roll Main Transformer Fuse, 120 Volt - Type T Time-Lag, 1.0 Amp, 250 Volt Main Transformer Fuse, 230 Volt - Type T Time-Lag, 0.5 Amp, 250 Volt FZX35 1 Sanitization Cartridge D Registered trademark of Dupont 40

41 REPLACEMENT PARTS General Maintenance Parts General maintenance parts are defined as laboratory level repair parts which do not require great expertise or special tools for installation. Barnstead Thermolyne recommends that you stock general maintenance parts as an aid to ensuring continued operation of this equipment. Description Catalog No. Recommended Quantity Quick Disconnect Inlet Valve Body 1/4 NPT CUX11 1 Insert 3/8 tube CUX9 1 Check Valve PMX136 1 Safety Stock For critical applications where performance with minimum downtime is required, Barnstead Thermolyne recommends that you maintain a local stock of those parts listed in General Maintenance Parts and Safety Stock sections. Description Catalog No. Recommended Quantity Display ME896X2 1 Main PC Board PC896X5 1 Recirculation Pump & Motor, 100 & 120 Volt PU896X1 1 Recirculation Pump & Motor, 240 Volt PU896X2 1 Pressure Regulator Resistivity Cell E896X4 1 Power Supply PC Board PC759X4A 1 Calibration PC Board PC896X3A 1 Resistivity Meter PC Board PC896X2 1 Optional Accessory Ordering Information Description Catalog No. UV Module D8974 UF Module, 10,000 MWCO D8984 Inlet Cell Bushing BR550X3 Cell O Ring GSX29 UV/UF Module D8994 N.I.S.T. Calibration Module E896X5 Inlet Cell E896X3 Low-Pressure Switch D2706 Float Switch D

NANOpure Infinity UV Module Upgrade

NANOpure Infinity UV Module Upgrade BARNSTEAD THERMOLYNE CORPORATION NANOpure Infinity UV Module Upgrade INSTALLATION MANUAL D8974 NANOpure Infinity UV Module Upgrade Kit LT897x5 7/97 1 2 Table of Contents Safety Information... 4 Alert Signals...

More information

BARNSTEAD THERMOLYNE CORPORATION. EASYpure LF OPERATION MANUAL AND PARTS LIST. Series 1052 LT1052X2 9/8/97

BARNSTEAD THERMOLYNE CORPORATION. EASYpure LF OPERATION MANUAL AND PARTS LIST. Series 1052 LT1052X2 9/8/97 BARNSTEAD THERMOLYNE CORPORATION EASYpure LF OPERATION MANUAL AND PARTS LIST Series 1052 LT1052X2 9/8/97 1 IMPORTANT INFORMATION This manual contains important operating and safety information. You must

More information

NANOpure Analytical Deionization System

NANOpure Analytical Deionization System BARNSTEAD THERMOLYNE CORPORATION NANOpure Analytical Deionization System OPERATION MANUAL AND PARTS LIST SERIES 851 ANALYTICAL GRADE SYSTEMS Model # Voltage Mount D4741 120 Wall D4742 240 Wall D4743 100

More information

EASYpure II RF. OPERATION MANUAL AND PARTS LIST Series Model D Volts LT1305X1 8/22/05

EASYpure II RF. OPERATION MANUAL AND PARTS LIST Series Model D Volts LT1305X1 8/22/05 EASYpure II RF OPERATION MANUAL AND PARTS LIST Series 1305 Model D7031 100-240 Volts LT1305X1 8/22/05 Table of Contents Safety Information...4 Alert Signals...4 Warnings...4 Introduction...6 General Specifications...7

More information

EASYpure II UV. OPERATION MANUAL AND PARTS LIST Series Model D Volts LT1302X1 8/22/05

EASYpure II UV. OPERATION MANUAL AND PARTS LIST Series Model D Volts LT1302X1 8/22/05 EASYpure II UV OPERATION MANUAL AND PARTS LIST Series 1302 Model D7401 100-240 Volts LT1302X1 8/22/05 Table of Contents Safety Information...4 Alert Signals...4 Warnings...4 Introduction...6 Product Overview...6

More information

BARNSTEAD THERMOLYNE ROTO MIX TYPE

BARNSTEAD THERMOLYNE ROTO MIX TYPE BARNSTEAD THERMOLYNE ROTO MIX TYPE 48200 OPERATION MANUAL AND PARTS LIST SERIES 1097 60 RPM 100 RPM Models Models M48225 M48215 M48220-33 M48210-33 M48224 LT1097X1 4/27/98 1 Table of Contents Safety Information...3

More information

B-pure Pressure Cartridge System

B-pure Pressure Cartridge System BARNSTEAD THERMOLYNE CORPORATION B-pure Pressure Cartridge System OPERATION MANUAL AND PARTS LIST SERIES 583 Model # Voltage D4505 D4511 D4521 120 D4522-33 230 D4523 100 D4524 120 D4525 240 D5831 120 D5832

More information

BARNSTEAD THERMOLYNE CORPORATION MEL-TEMP 3.0 OPERATING INSTRUCTIONS. Model No LT1186X4 5/1/00

BARNSTEAD THERMOLYNE CORPORATION MEL-TEMP 3.0 OPERATING INSTRUCTIONS. Model No LT1186X4 5/1/00 BARNSTEAD THERMOLYNE CORPORATION MEL-TEMP 3.0 OPERATING INSTRUCTIONS Model No. 1401 1402 LT1186X4 5/1/00 Table of Contents Safety Information...3 Alert Signals...3 Intended Use...3 General Usage...3 Warnings...3

More information

Type D0800. Bantam Cartridge Holder OPERATION MANUAL AND PARTS LIST SERIES 494. Mo de l # Vo ltag e D D /12/08

Type D0800. Bantam Cartridge Holder OPERATION MANUAL AND PARTS LIST SERIES 494. Mo de l # Vo ltag e D D /12/08 Type D0800 Bantam Cartridge Holder OPERATION MANUAL AND PARTS LIST SERIES 494 Mo de l # Vo ltag e D0800 120 D0805 240 60119 3/12/08 Table of Contents Safety Information...3 Alert Signals...3 Warnings...3

More information

Type Maxi Mix III Rotary Shaker OPERATION MANUAL AND PARTS LIST SERIES 1046 BARNSTEAD THERMOLYNE

Type Maxi Mix III Rotary Shaker OPERATION MANUAL AND PARTS LIST SERIES 1046 BARNSTEAD THERMOLYNE BARNSTEAD THERMOLYNE Type 65800 Maxi Mix III Rotary Shaker OPERATION MANUAL AND PARTS LIST SERIES 1046 LT1046X1 2/20/98 Model # Voltage M65820-26 230 M65820-26-60 230 M65824 100 M65825 120 M65920 230 Table

More information

Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by:

Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by: Storage System Ultrasound Probe Cabinet Operator's Manual Manufactured by: CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA 319.248.6757 / 800.445.6741 WWW.CIVCO.COM Copyright 2018 All

More information

MicroCoat. System Operating Manual MC2000 Series. MC785, MC785-WF Spray Valves. US: UK: Mexico:

MicroCoat. System Operating Manual MC2000 Series. MC785, MC785-WF Spray Valves. US: UK: Mexico: MicroCoat System Operating Manual MC2 Series MC785, MC785-WF Spray Valves A NORDSON COMPANY US: 8-498-8865 UK: 8 585733 Mexico: 1-8-556-3484 Introduction The MicroCoat System provides precise lubrication

More information

LABORATORY ZERO AIR GENERATOR MODEL N-GC1500 USER S MANUAL

LABORATORY ZERO AIR GENERATOR MODEL N-GC1500 USER S MANUAL LABORATORY ZERO AIR GENERATOR MODEL N-GC1500 USER S MANUAL Content 1. Introduction...2 2. Important safety instruction...3 3. System component...4 4. Engineering design overview...5 5. Installation...6

More information

VSP SERIES Installation, Operation and Maintenance Manual

VSP SERIES Installation, Operation and Maintenance Manual VSP SERIES Installation, Operation and Maintenance Manual VARIABLE SPEED PERISTALTIC METERING PUMPS VSP-12 and VSP-20 Models READ ALL WARNINGS CAREFULLY BEFORE INSTALLING PUMP PUMP DATA/SPECIFICATIONS

More information

RediCoat by Nordson. Hopper and VBF Dolly Systems. Customer Product Manual Part A Issued 10/07

RediCoat by Nordson. Hopper and VBF Dolly Systems. Customer Product Manual Part A Issued 10/07 RediCoat by Nordson Hopper and VBF Dolly Systems Customer Product Manual Part 1082648A Issued 10/07 This document is subject to change without notice. Nordson Corporation Amherst, Ohio USA 2 Table of Contents

More information

Safety Information...3 Alert Signals...3 Warnings...3 Introduction...4 Principles of Operation...4 Declaration of Conformity...

Safety Information...3 Alert Signals...3 Warnings...3 Introduction...4 Principles of Operation...4 Declaration of Conformity... 9/9/10 Table of Contents Safety Information...3 Alert Signals...3 Warnings...3 Introduction...4 Principles of Operation...4 Declaration of Conformity...4 Technical Specifications...5 Installation...5 Operation...6

More information

Product Information. Kugelrohr Short-Path Distillation Apparatus. Operation Instructions Z V Apparatus Z V Apparatus

Product Information. Kugelrohr Short-Path Distillation Apparatus. Operation Instructions Z V Apparatus Z V Apparatus Product Information Kugelrohr Short-Path Distillation Apparatus Operation Instructions Z401137 115V Apparatus Z401145 230V Apparatus Table of Contents Safety Information Intended Use 3 General Usage 3

More information

MODEL ELC-12/40-CVM-D BATTERY CHARGER

MODEL ELC-12/40-CVM-D BATTERY CHARGER NATIONAL RAILWAY SUPPLY MODEL ELC-12/40-CVM-D BATTERY CHARGER Installing, Operating and Service Instructions for the ELC-12/40-CVM-D Solid State Charger PLEASE SAVE THESE IMPORTANT SAFETY AND OPERATING

More information

CHROME ALLEY CAT L.E.D. FUEL & BATTERY GAUGE 7381

CHROME ALLEY CAT L.E.D. FUEL & BATTERY GAUGE 7381 CHROME ALLEY CAT L.E.D. FUEL & BATTERY GAUGE 7381 Thank You For Choosing Küryakyn! Protect yourself and others from potential injury and property damage or loss. Pay close attention to all instructions,

More information

MECHANICAL FUEL METER. FM-100 and FM-200 Series Owner s Manual

MECHANICAL FUEL METER. FM-100 and FM-200 Series Owner s Manual MECHANICAL FUEL METER FM-100 and FM-200 Series Owner s Manual To the owner Congratulations on receiving your GPI Mechanical Fuel Meter. We at GPI are pleased to provide you with a fuel meter designed to

More information

Model and Series 115 VAC INDUSTRIAL DIAPHRAGM PUMPS. PumpAgents.com - buy pumps and parts online INDUSTRIAL DIAPHRAGM PUMPS

Model and Series 115 VAC INDUSTRIAL DIAPHRAGM PUMPS. PumpAgents.com - buy pumps and parts online INDUSTRIAL DIAPHRAGM PUMPS Model 31801 and 31800 Series 115 VAC INDUSTRIAL DIAPHRAGM PUMPS INDUSTRIAL DIAPHRAGM PUMPS FEATURES Run Dry Ability Self-Priming Thermal Overload Protected Motor Easy Installation Low Amp Draw Compact

More information

PURE WATER. Installation, Operation and Maintenance Manual. 3-Stage Under Counter Filtration with UV Model PWSYS-UV-3STG. Important.

PURE WATER. Installation, Operation and Maintenance Manual. 3-Stage Under Counter Filtration with UV Model PWSYS-UV-3STG. Important. Installation, Operation and Maintenance Manual 3-Stage Under Counter Filtration with UV Model PWSYS-UV-3STG IOM-WQ-PWSYS-UV-3STG PURE WATER Important Please read the entire manual before proceeding with

More information

DR Series Radial Diaphragm Valves

DR Series Radial Diaphragm Valves DR Series Radial Valves Service Instructions Valves with Plastic Actuators (1/2 through 1 inch) Valves with Aluminum Actuators (1/2 through 2 inch) Manual model Pneumatic model Manual model Pneumatic model

More information

Model 2310 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

Model 2310 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority VX-2 Model 2310 Series Clearing The Airway Is Our #1 Priority Operating Instructions & Maintenance Manual, INC. 11064 Randall Street Sun Valley, CA 91352 USA www.sscor.com Email: marketing@sscor.com techsupport@sscor.com

More information

R & D SPECIALTIES ROTROL I USER'S MANUAL

R & D SPECIALTIES ROTROL I USER'S MANUAL R & D SPECIALTIES ROTROL I USER'S MANUAL TABLE OF CONTENTS INTRODUCTION...2 SPECIFICATIONS...2 CONTROLS AND INDICATORS...3 TIME DELAYS...4 INSTALLATION...5 SYSTEM OPERATION...9 TROUBLESHOOTING...13 OPTIONAL

More information

WARNING: FAILURE TO FOLLOW THESE RULES MAY RESULT IN SERIOUS PERSONAL INJURY CAUTION: INSTALLATION LOCATION:

WARNING: FAILURE TO FOLLOW THESE RULES MAY RESULT IN SERIOUS PERSONAL INJURY CAUTION: INSTALLATION LOCATION: 2 Please read the safety and installation instructions carefully to help ensure a correct and SAFE installation of your Second Wind Ultraviolet Germicidal Photo-catalytic Air Purifier. WARNING: FAILURE

More information

Centrifuge Operator / Service Manual

Centrifuge Operator / Service Manual 3000 Centrifuge Centrifuge Operator / Service Manual cat.# 26230 & 26231 The Q-sep 3000 centrifuge complies with all requirements of UL standard 3101 20, Can/CSA C22.2 No. 1010.1, and Can/CSA C22.2 No.

More information

NILFISK BA 500 Service Manual

NILFISK BA 500 Service Manual NILFISK BA 500 Service Manual Model 66324400 12/94 Form Number 043023 TABLE OF CONTENTS Batteries...21 Brush Drive Belt Adjustment Or Replacement...7 Brush Drive Motor - Carbon brush Inspection... 8 Brush

More information

Troubleshooting Guide: 355 Lights (24V)

Troubleshooting Guide: 355 Lights (24V) Troubleshooting Guide: 355 Lights (24V) Contents Description Refer To: Troubleshooting - Troubleshooting Chart Adjustments / Repair Procedures Bulb Replacing the Bulb Fuse(s) Replacing the Fuse (Ceiling)

More information

VWR Zero Air Generator

VWR Zero Air Generator Technical Information Installation, Operation, and Maintenance Manual VWR 89237-564 Zero Air Generator Figure 1 - VWR Zero Air Generator These instructions must be thoroughly read and understood before

More information

Battery Backup Sump Pump System Instruction Manual

Battery Backup Sump Pump System Instruction Manual Battery Backup Sump Pump System Instruction Manual Push button second to test or reset alarm. Push 5 seconds to silence alarm for 24 hours. Warning alarms The fluid in the battery is low. Add distilled

More information

AD-830. Parts Manual thru 1997

AD-830. Parts Manual thru 1997 AD-830 Parts Manual 1995 thru 1997 American Dryer Corporation 88 Currant Road Fall River, MA 02720-4781 Telephone: (508) 678-9010 / Fax: (508) 678-9447 E-mail address: service@amdry.com 072696 092497DX/abe

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS THANK YOU FOR CHOOSING KURYAKYN! Protect yourself and others from possible injury and property damage or loss. Pay close attention to all instructions, warnings, cautions, and notices regarding the installation,

More information

Troubleshooting Guide: 355 Lights (12V)

Troubleshooting Guide: 355 Lights (12V) Troubleshooting Guide: 355 Lights (12V) Contents Description Refer To: Troubleshooting - Troubleshooting Chart Adjustments / Repair Procedures Bulb Replacing the Bulb Fuse(s) Replacing the Fuse (Ceiling)

More information

M OTO R. The electric motor. DC - direct current voltage.

M OTO R. The electric motor. DC - direct current voltage. TMD-5 Lubricator Automatic Cyclic Before installing or operating this lubricator, please read this manual carefully. Failure to follow instructions can result in damage to the product and/or serious bodily

More information

FlexChem Model 404 Rotating Oven

FlexChem Model 404 Rotating Oven www.scigene.com FlexChem Model 404 Rotating Oven USER MANUAL Cat. #1052-20-1, 1052-20-2, 1052-22-1 & 1052-22-2 FOR RESEARCH USE ONLY Not for Use in Diagnostic Procedures SciGene 306 Potrero Ave, Sunnyvale,

More information

5000TOC Sensor Service Manual

5000TOC Sensor Service Manual Part No. 84449 5000TOC Sensor Service Manual This document contains proprietary information, which is protected by copyright. All rights are reserved. No part of this document may be photocopied (other

More information

Installation Operation Parts

Installation Operation Parts OWNER S MANUAL BATTERY BACKUP SUMP Installation Operation Parts For further operating, installation or maintenance assistance, Call 98-8-05 PRINTED IN U.S.A. M-8 (/9) RULES FOR SAFE INSTALLATION AND OPERATION

More information

ValveMate 7000 Controller Operating Manual

ValveMate 7000 Controller Operating Manual A NORDSON COMPANY ValveMate 7000 Controller Operating Manual Steady Test Clear Time Set Fast Slow POWER RUN SETUP CYCLE A NORDSON COMPANY VALVEMATE 7000 Run Setup Purge Program Fast Slow Time Set Pressure

More information

HD 7700 Setup & Operator Manual

HD 7700 Setup & Operator Manual HD 7700 Setup & Operator Manual Issue 1 December, 01 Performance Design Inc. The Heavy Duty Ultima (HD 7700) electric punch has been designed to punch most any job that may pass through your bindery or

More information

S25500 MAXI STIRRER OPERATION MANUAL AND PARTS LIST SERIES 1352

S25500 MAXI STIRRER OPERATION MANUAL AND PARTS LIST SERIES 1352 S25500 MAXI STIRRER OPERATION MANUAL AND PARTS LIST SERIES 152 Model S1505 Series 152 120 Volt 50/60 HZ Model S1500-- Series 152 220-240 Volt 50/60 HZ Model S2555 Series 152 120 Volt 50/60 HZ Model S2550-

More information

OPERATION MANUAL MODELS TWE-250 TWE-321 TWE-375 TRU WELD EQUIPMENT COMPANY 6400 N. HONEYTOWN ROAD SMITHVILLE, OHIO (330)

OPERATION MANUAL MODELS TWE-250 TWE-321 TWE-375 TRU WELD EQUIPMENT COMPANY 6400 N. HONEYTOWN ROAD SMITHVILLE, OHIO (330) OPERATION MANUAL MODELS TWE-250 TWE-321 TWE-375 TRU WELD EQUIPMENT COMPANY 6400 N. HONEYTOWN ROAD SMITHVILLE, OHIO 44677 (330) 669 2773 CONTENTS Section Description Pages 1 Introduction 3 2 External Features

More information

Operating Manual for. Dispenser. Read this manual thoroughly before operating the Labmax Bottle Top Dispenser!

Operating Manual for. Dispenser. Read this manual thoroughly before operating the Labmax Bottle Top Dispenser! Operating Manual for Dispenser Read this manual thoroughly before operating the Labmax Bottle Top Dispenser! General Safety Precautions When using the Labmax please observe the following safety precautions

More information

BT403. A Geno Technology, Inc. (USA) brand name. BT-300 Power Supply. Cat. No. BT

BT403. A Geno Technology, Inc. (USA) brand name. BT-300 Power Supply. Cat. No. BT BT403 A Geno Technology, Inc. (USA) brand name BT-300 Power Supply Cat. No. BT403 1-800-628-7730 1-314-991-6034 info@btlabsystems.com WARNING... 3 SAFETY INFORMATION... 3 ENVIRONMENTAL CONDITIONS... 4

More information

MICROPROCESSOR CONTROL WATER SUPPLY SYSTEM (MPR-01AW)

MICROPROCESSOR CONTROL WATER SUPPLY SYSTEM (MPR-01AW) MICROPROCESSOR CONTROL WATER SUPPLY SYSTEM (MPR-01AW) Installation, Operation, and Maintenance Manual REV. "B" REV. "C" PLEASE READ ENTIRE DOCUMENT BEFORE PROCEEDING WITH INSTALLATION Energy Saver has

More information

OPERATIONS MANUAL. Mini Centrifuge Model MCF Certified

OPERATIONS MANUAL. Mini Centrifuge Model MCF Certified OPERATIONS MANUAL Mini Centrifuge Model MCF-2360 Certified Contents 1. Safety 2 2. Introduction 7 3. Package Contents 7 4. Specifications 8 5. Features 8 6. Parts of the Mini Centrifuge 9 7. Installation

More information

Operators Manual. Recirculating Chiller /06/08

Operators Manual. Recirculating Chiller /06/08 Operators Manual Recirculating Chiller 110-197 11/06/08 Table of Contents Section 1. General Information 1.1 Warranty 1.2 Unpacking 1.3 Package Contents 1.4 Description of the Recirculating Chiller 1.5

More information

MOTOR. The electric motor. DC - direct current voltage.

MOTOR. The electric motor. DC - direct current voltage. TMD-5 Lubricator Automatic Cyclic Before installing or operating this lubricator, please read this manual carefully. Failure to follow instructions can result in damage to the product and/or serious bodily

More information

THIS PRODUCT IS FOR PROFESSIONAL LABORATORY USE ONLY USER'S MANUAL. WELLS ENGINE UNIT 230 VOLT Product No. U905, U906, U907, U908

THIS PRODUCT IS FOR PROFESSIONAL LABORATORY USE ONLY USER'S MANUAL. WELLS ENGINE UNIT 230 VOLT Product No. U905, U906, U907, U908 DENTAL, INC. TECHNICAL BULLETIN U807-022510 5860 FLYNN CREEK ROAD READ ALL INSTRUCTIONS P.O. BOX 106 BEFORE PROCEEDING COMPTCHE, CALIFORNIA, U.S.A. 95427 SAVE THIS FOR FUTURE REFERENCE THIS PRODUCT IS

More information

Installation Guide. Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A

Installation Guide. Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A Installation Guide Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A Agilent Technologies 2001 All Rights Reserved. Reproduction, adaptation, or translation without permission is prohibited,

More information

Instruction Manual. Vortex Mixers

Instruction Manual. Vortex Mixers Instruction Manual Vortex Mixers TABLE OF CONTENTS Package Contents............... 1 Warranty............... 1 Installation............... 2 Maintenance & Servicing............... 2 Environmental Conditions...............

More information

INSTRUCTIONS. SILEX demineralizers. Type 1 B and 2 B. Our World is Water. D1A-30Bver4. Instructions. SILHORKO-EUROWATER A/S Phone

INSTRUCTIONS. SILEX demineralizers. Type 1 B and 2 B. Our World is Water. D1A-30Bver4. Instructions. SILHORKO-EUROWATER A/S Phone Instructions D1A-30Bver4 INSTRUCTIONS SILEX demineralizers Type 1 B and 2 B SILHORKO-EUROWATER A/S Phone + 45 86 57 12 22 DK-8660 Skanderborg info@eurowater.com Denmark www.eurowater.com Our World is Water

More information

DA Series Diaphragm Valve

DA Series Diaphragm Valve DA Series Diaphragm Valve Service Instructions Pneumatically Actuated Valve Toggle Valve Valve Valve Valves are shown with tube butt weld ends. These instructions also apply to DA series valves with any

More information

M-3025CB-AV Fuel Pump

M-3025CB-AV Fuel Pump SAVE THESE INSTRUCTIONS M-3025CB-AV Fuel Pump Owner s Manual TABLE OF CONTENTS General Information... 2 Safety Instructions... 2 Installation... 3 Operation... 4 Maintenance... 4 Repair... 5 Troubleshooting...

More information

Models & Inch & 4 Inch Digital Inline Pressure Control Valves SM64504

Models & Inch & 4 Inch Digital Inline Pressure Control Valves SM64504 SM64504 September 2008 Aerospace Group Conveyance Systems Division Carter Brand Ground Fueling Equipment Applicable addition manuals: None Maintenance & Repair Manual 3 Inch & 4 Inch Digital Inline Pressure

More information

MicroCoat System Operating Manual MC4000 Series MC785M, MC785M-WF Spray Valves

MicroCoat System Operating Manual MC4000 Series MC785M, MC785M-WF Spray Valves MicroCoat System Operating Manual MC Series MC785M, MC785M-WF Spray Valves A NORDSON COMPANY Introduction The MicroCoat System provides precise lubrication control for metal stamping operations. The MC

More information

System 3 ZB1PS Device Chassis. Operator s Manual

System 3 ZB1PS Device Chassis. Operator s Manual System 3 ZB1PS Device Chassis Operator s Manual System 3 ZB1PS Operator's Manual Copyright 2007 Tucker-Davis Technologies, Inc. (TDT). All rights reserved. No part of this manual may be reproduced or transmitted

More information

Model 3051 Sensor Module Replacement

Model 3051 Sensor Module Replacement Model 3051 Sensor Module Replacement 00809-0400-4001 English Rev. BA SAFETY MESSAGES Procedures and instructions in this manual may require special precautions to ensure the safety of the personnel performing

More information

DWYER INSTRUMENTS, INC. Phone: 219/ P.O. BOX 373 MICHIGAN CITY, IN 46361, U.S.A. Fax: 219/

DWYER INSTRUMENTS, INC. Phone: 219/ P.O. BOX 373 MICHIGAN CITY, IN 46361, U.S.A. Fax: 219/ Series 43000 Capsu-Photohelic Pressure Switch/Gage Specifications - Installation and Operating Instructions Bulletin B-34 Ø4-3/4 [120.65] 3-7/8 SQ [98.43] 3/4 CONDUIT 4-3/8 [111.13] HOUSING REMOVAL 3-1/16

More information

MODEL ELC-12/60-D BATTERY CHARGER

MODEL ELC-12/60-D BATTERY CHARGER *32198* NATIONAL RAILWAY SUPPLY Installing, Operating and Service Instructions for the 12/60 Solid State Charger MODEL ELC-12/60-D BATTERY CHARGER PLEASE SAVE THESE IMPORTANT SAFETY AND OPERATING INSTRUCTIONS

More information

250L Cartridge Dual Seal

250L Cartridge Dual Seal INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS 250L Cartridge Dual Seal Installation, Operation and Maintenance Instructions TABLE OF CONTENTS 1.0 Cautions...2 2.0 Transport and Storage...2 3.0 Description...2

More information

MODEL CRSY102 CONDENSATE RECIRCULATING SYSTEM OPERATIONS MANUAL

MODEL CRSY102 CONDENSATE RECIRCULATING SYSTEM OPERATIONS MANUAL MODEL CRSY102 CONDENSATE RECIRCULATING SYSTEM OPERATIONS MANUAL PO Box 715 Marietta, OH 45750 800-648-3042 740-373-6809 Fax 740-374-3760 www.caronproducts.com service@caronproducts.com Dear Valued Customer:

More information

BlueSwan Pipette Controller

BlueSwan Pipette Controller BlueSwan Pipette Controller BSST-0100 Operation Manual Ver. 1.0 Copyright 2016 Blue-Ray Biotech Corp. All Rights Reserved. Table of Contents 1. Safety Precautions... 1 2. General Description... 2 2.1

More information

INSTALLATION ZOMBIE L.E.D. FUEL & BATTERY GAUGE. THANK YOU FOR CHOOSING KϋRYAKYN!

INSTALLATION ZOMBIE L.E.D. FUEL & BATTERY GAUGE. THANK YOU FOR CHOOSING KϋRYAKYN! THANK YOU FOR CHOOSING KϋRYAKYN! PROTECT YOURSELF AND OTHERS FROM POSSIBLE INJURY AND PROPERTY DAMAGE OR LOSS. PAY CLOSE ATTENTION TO ALL INSTRUCTIONS, WARNINGS, CAUTIONS, AND NOTICES REGARDING THE, USE,

More information

MCC Automatic Start-Stop Microcomputer Charger Control

MCC Automatic Start-Stop Microcomputer Charger Control MCC Automatic Start-Stop Microcomputer Charger Control Installation & Operation Instructions Warning: Read ALL the instructions before starting. If anything is not clear, call your dealer or contact Arrgh!!

More information

X4 Installation and Operation Manual - POWER FLAME INCORPORATED

X4 Installation and Operation Manual - POWER FLAME INCORPORATED 7.13.2 Set the burner s combustion air inlet damper to the approximate setting as shown in this manual for the desired firing rate. Also, verify that the correct main orifice is installed in the main orifice

More information

PRODUCT SERVICE MANUAL FOR BK12DHZ PUMPS

PRODUCT SERVICE MANUAL FOR BK12DHZ PUMPS PRODUCT SERVICE MANUAL FOR BK12DHZ PUMPS WARNING This manual, and the GENERAL INSTRUCTION MANUAL SRM00046, should be read thoroughly prior to pump installation, operation or maintenance. Manual No. SRM00095

More information

INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS.

INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS. INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS. Before installing and/or using this product, please check for any visual damage of the enclosure, power

More information

Advantage-D. Operating Instructions and Maintenance Manual. Central Vacuum Systems (Expandable/Modular Models) (Ver.

Advantage-D. Operating Instructions and Maintenance Manual. Central Vacuum Systems (Expandable/Modular Models) (Ver. Advantage-D Series 3 Central Vacuum Systems (Expandable/Modular Models) (Ver. 8/05) Operating Instructions and Maintenance Manual DESCRIPTION The Becker Advantage-D and Advantage-L central vacuum systems

More information

USER S MANUAL TUBE SEALER. Model SE250. No. CAT.SE25010Ce

USER S MANUAL TUBE SEALER. Model SE250. No. CAT.SE25010Ce USER S MANUAL TUBE SEALER Model SE250 No. CAT.SE25010Ce Centron Technologies Corporation 319-25 Sadang-4-dong, Dongjak-ku Seoul, Korea 156-823 Tel. +82-2.522.7807 Fax +82-2.522.7806 Important Note 1. Safety

More information

Type Maxi Mix III Rotary Shaker

Type Maxi Mix III Rotary Shaker Type 65800 Maxi Mix III Rotary Shaker OPERATION MANUAL AND PARTS LIST SERIES 1046 LT1046X1 3/19/08 Model # Voltage M65820-26 230 M65820-26-60 230 M65824 100 M65825 120 M65920 230 Table of Contents Safety

More information

TS1251 PRESSURE DISPENSER USER'S MANUAL

TS1251 PRESSURE DISPENSER USER'S MANUAL TS1251 PRESSURE DISPENSER USER'S MANUAL TABLE OF CONTENTS SECTION DESCRIPTION PAGE NUMBER 1.0 CAUTIONS AND WARNINGS... 3 2.0 INTRODUCTION... 4 3.0 DESCRIPTION... 4 & 5 4.0 SET UP AND INSTALLATION... 6

More information

Model Fixed Rack-Mount Kit

Model Fixed Rack-Mount Kit Model 4299-7 Fixed Rack-Mount Kit Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139 1-800-935-5595 tek.com/keithley Assembly and Mounting Instructions Introduction This document contains the

More information

PRODUCT INFORMATION BULLETIN #3365 DIGITAL MOTOR CONTROL PLATTER SYSTEMS For Serial Number and After

PRODUCT INFORMATION BULLETIN #3365 DIGITAL MOTOR CONTROL PLATTER SYSTEMS For Serial Number and After PRODUCT INFORMATION BULLETIN #3365 DIGITAL MOTOR CONTROL PLATTER SYSTEMS For Serial Number 28640996 and After Record Platter System Identification Numbers Here: Model # Serial # Table of Contents Program

More information

LM-200 MECHANICAL LUBE METER Owner s Manual

LM-200 MECHANICAL LUBE METER Owner s Manual LM-200 MECHANICAL LUBE METER Owner s Manual To the owner Congratulations on receiving your GPI Mechanical Lube Meter. We at GPI are pleased to provide you with a Lube Meter designed to give you maximum

More information

DCell Suction. Model DM Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

DCell Suction. Model DM Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority DCell Suction Model DM10-001 Clearing The Airway Is Our #1 Priority Operating Instructions & Maintenance Manual, INC. 11064 Randall Street Sun Valley, CA 91352 USA www.sscor.com Email: marketing@sscor.com

More information

S25500 MAXI STIRRER OPERATION MANUAL AND PARTS LIST SERIES 849

S25500 MAXI STIRRER OPERATION MANUAL AND PARTS LIST SERIES 849 BARNSTEAD THERMOLYNE S25500 MAXI STIRRER OPERATION MANUAL AND PARTS LIST SERIES 849 Model S25535 Series 849 120 Volt 50/60 HZ Model S25530 Series 849 240 Volt 50/60 HZ Model S25530-26 Series 849 220-240

More information

SOS SERIES SOS1 SOS2. Spares On Site Battery Cabinet Installation Guide rEV3

SOS SERIES SOS1 SOS2. Spares On Site Battery Cabinet Installation Guide rEV3 Atlantic Battery Systems 1065 Market Street Paterson, NJ 07513 Phone: (800) 875-0073 Fax: (973) 523-2344 sales@atbatsys.com www.atbatsys.com SOS1 SOS2 SOS SERIES Spares On Site Battery Cabinet Installation

More information

Automatic Burnout Furnaces 115 to 230-volt Models OPERATOR S MANUAL

Automatic Burnout Furnaces 115 to 230-volt Models OPERATOR S MANUAL Automatic Burnout Furnaces 115 to 230-volt Models OPERATOR S MANUAL TABLE OF CONTENTS Introduction...3 Warranty...3 On-Line Warranty Registration...3 Safety Instructions........................................................3

More information

Instruction Sheet. Subject: Installations Instructions. Air Compressor Kits. Importan

Instruction Sheet. Subject: Installations Instructions. Air Compressor Kits. Importan MULTIPLEX 2100 FUTURE, SELLERSBURG, IN 47172 800-367-4233 WWW.MANITOWOCBEVERAGE.COM/US Instruction Sheet Subject: Installations Instructions Importan Read the following warnings before beginning an installation.

More information

Installation Instructions

Installation Instructions Installation Instructions for 15912 to 15916 Electric Fuel Pumps & Fuel Pressure Regulators Installation Instructions WARNING! These instructions must be read and fully understood before beginning the

More information

STATIM CASSETTE AUTOCLAVE TROUBLESHOOTING GUIDE

STATIM CASSETTE AUTOCLAVE TROUBLESHOOTING GUIDE STATIM CASSETTE AUTOCLAVE TROUBLESHOOTING GUIDE Version 2.1 Original Statim Error Codes The Statim Sterilizer should never leak steam or have water dripping from under the machine. If either of these conditions

More information

DELTA O-RING CARTRIDGE SEAL ASSEMBLY AND INSTALLATION INSTRUCTIONS INTRODUCTION:

DELTA O-RING CARTRIDGE SEAL ASSEMBLY AND INSTALLATION INSTRUCTIONS INTRODUCTION: DELTA O-RING CARTRIDGE SEAL ASSEMBLY AND INSTALLATION INSTRUCTIONS INTRODUCTION: These instructions are provided to familiarize the user with the seal and its use. The instructions must be read carefully

More information

CLEAN ROOM DEVICES, LLC "WHERE TUBING AND FITTINGS COME TOGETHER"

CLEAN ROOM DEVICES, LLC WHERE TUBING AND FITTINGS COME TOGETHER CLEAN ROOM DEVICES, LLC "WHERE TUBING AND FITTINGS COME TOGETHER" CRD600AF Automatic Fitting Inserter With Auto Feed OPERATIONS MANUAL (Shown with optional alcohol dispenser) 1 VERSION 1.1 LAST EDITED

More information

3 Inch & 4 Inch Digital Bypass Pressure Control Valves

3 Inch & 4 Inch Digital Bypass Pressure Control Valves SM64505 September 2008 Aerospace Group Conveyance Systems Division Carter Brand Ground Fueling Applicable addition manuals: None Maintenance & Repair Manual 3 Inch & 4 Inch Digital Bypass Pressure Control

More information

DIGITAL BATTERY TORQUE WRENCH (BC-RAD SELECT) USER GUIDE

DIGITAL BATTERY TORQUE WRENCH (BC-RAD SELECT) USER GUIDE DIGITAL BATTERY TORQUE WRENCH (BC-RAD SELECT) USER GUIDE W.CHRISTIE (INDUSTRIAL) LTD CHRISTIE HOUSE, MEADOWBANK ROAD, ROTHERHAM, SOUTH YORKSHIRE, S61 2NF, UK T: +44(0)1709 550088 F: +44(0)1709 550030 E:

More information

User s Manual. Automatic Switch-Mode Battery Charger

User s Manual. Automatic Switch-Mode Battery Charger User s Manual Automatic Switch-Mode Battery Charger IMPORTANT Read, understand, and follow these safety rules and operating instructions before using this battery charger. Only authorized and trained service

More information

SY-SERIES S04DA75 THRU S06DA180 & S04SR75 THRU S06SR180 DOUBLE ACTING & SPRING RETURN SCOTCH YOKE ACTUATORS

SY-SERIES S04DA75 THRU S06DA180 & S04SR75 THRU S06SR180 DOUBLE ACTING & SPRING RETURN SCOTCH YOKE ACTUATORS SY-SERIES S04DA75 THRU S06DA180 & S04SR75 THRU S06SR180 DOUBLE ACTING & SPRING RETURN SCOTCH YOKE ACTUATORS INTRODUCTION A-T Controls SY scotch yoke actuators have been designed and engineered to provide

More information

6 Gauge Box Set IS0333

6 Gauge Box Set IS0333 Caution 6 Gauge Box Set IS0 Rev. B ecr 882 9/202 Disconnect the battery during installation. Tighten nuts on the back clamp only slightly more than you can tighten with your fingers. Six inch-pounds of

More information

40041 Heavy Duty ADA System with Booster Bracket for JK Heavy Duty ADA System with Booster Bracket for 2012 to Current JK

40041 Heavy Duty ADA System with Booster Bracket for JK Heavy Duty ADA System with Booster Bracket for 2012 to Current JK 40041 Heavy Duty ADA System with Booster Bracket for 2007-2011 JK 40044 Heavy Duty ADA System with Booster Bracket for 2012 to Current JK 40049 Heavy Duty ADA System Universal for all vehicles (booster

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Ultraviolet Single and Dual Germicidal Lamps 60 Hz Part Numbers: P103-UVLTT1LP1020A01, P103-UVLTT1LP3020A01, P103-UVLTT2LP1020A01, P103-UVLTT2LP3020A01 SAFETY CONSIDERATIONS Installation

More information

INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS.

INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS. INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS. Before installing and/or using this product, please check for any visual damage of the enclosure, power

More information

MR 5-30 or MR 5-30N Series Fuel Meter Owner s Manual

MR 5-30 or MR 5-30N Series Fuel Meter Owner s Manual SAVE THESE INSTRUCTIONS MR 5-30 or MR 5-30N Series Fuel Meter Owner s Manual TABLE OF CONTENTS General Information...2 Installation...3 Operation...3 Calibration...4 Maintenance...5 Troubleshooting...6

More information

MODEL NUMBER: MEDIUM DUTY ONBOARD AIR SYSTEM

MODEL NUMBER: MEDIUM DUTY ONBOARD AIR SYSTEM MODEL NUMBER: 10003 MEDIUM DUTY ONBOARD AIR SYSTEM IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this manual before installing and using

More information

portable ac/dc suction pump USER S MANUAL ITEM # 18605

portable ac/dc suction pump USER S MANUAL ITEM # 18605 portable ac/dc suction pump USER S MANUAL ITEM # 18605 index 1 1. Safety Information...2-4 2. AC Power Source...3 3. Electrical...3-4 4. Introduction...4 5. Illustration of Important Parts...5 6. Set-Up...6

More information

Pump Manual. Model Numbers: SP240VDFS57LPM SP240VDFS80LPM. Product Names: Digital Refueling Station - 57 L/min Digital Refueling Station - 80 L/min

Pump Manual. Model Numbers: SP240VDFS57LPM SP240VDFS80LPM. Product Names: Digital Refueling Station - 57 L/min Digital Refueling Station - 80 L/min Pump Manual Model Numbers: SP240VDFS57LPM SP240VDFS80LPM Product Names: Digital Refueling Station - 57 L/min Digital Refueling Station - 80 L/min Copyright: Scintex Fuel Depot 2012[Type text] Page 1 WARNING:

More information

Car Battery Charger Instructions for Use

Car Battery Charger Instructions for Use BATTERY CHARGER 12Volt 4Amp FOR INDOOR USE ONLY Power Details: Input: 230-240Vac; 50Hz; 52W Output: 12V DC; 2.8A Maximum Charge Rate: 4A RMS Read these instructions before operating this car battery charger

More information

NANOPAC-300 & 500 Power Supply. Instruction manual NANOPAC-300 & NANOPAC-500

NANOPAC-300 & 500 Power Supply. Instruction manual NANOPAC-300 & NANOPAC-500 NANOPAC-300 & 500 Power Supply Instruction manual NANOPAC-300 & NANOPAC-500 Version 01C Feb 5th, 2014 1 Packing list NANOPAC-300 or 500-1x NANOPAC-300 Power Supply or NANOPAC-500-1x Power Cord - 1x Instruction

More information

Declaration of Conformity

Declaration of Conformity Declaration of Conformity We, Manufacturer: Spartanics Ltd. 3605 Edison Place Rolling Meadows, Illinois 60008 Phone: 847-394-5700 Fax: 847-394-0409 USA ENGLISH declare under our sole responsibility that

More information

5 Gauge Box Set IS0342

5 Gauge Box Set IS0342 Caution 5 Gauge Box Set IS0342 Rev. B ecr 8832 9/202 Disconnect the battery during installation. Tighten nuts on the back clamp only slightly more than you can tighten with your fingers. Six inch-pounds

More information