Testing Power Supplies for Medical Electronics Applications
|
|
- Audrey Harmon
- 5 years ago
- Views:
Transcription
1 white pƒper Testing Power Supplies for Medical Electronics Applications g Introduction The international stard IEC specifies testing performance for electrical devices used in medical applications. This stard has undergone some significant changes in the last number of years, with the recent adoption of the 3rd edition of the stard in In fact, IEC recently announced the planned introduction of the 4th edition of the IEC60601 stard, slated for adoption in 2018, making it a design consideration for the next generation of medical equipment. The medical electronics end-product is the applied part that must be evaluated for overall performance to meet stard s requirements. The power supply is considered a component of a medical device. Although a power supply cannot be certified, it can be designed tested to meet the necessary levels of performance according to the applied part s intended applications. Testing power supplies to this medical stard can help manufacturers optimize their medical designs to achieve the highest performance safety levels thereby enable first-pass success during agency approvals. This paper describes how Excelsys Technologies Xsolo power supplies have been designed tested to meet both the B BF type requirements per the IEC stard. g Safety Classifications for Medical Applied Parts In July 2014, the IEC updated the medical stard IEC as related to Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety Essential Performance Collateral Stard: Electromagnetic Disturbances Requirements Tests. The stard cites three classifications for meeting these requirements: CF, BF, B. The requirements for leakage current for each of these classifications, taken from the IEC specifications, are shown in Tables 1 2 on the following page. The CF, or Cardiac Floating, is the most deming performance level. Applied parts falling under this rating have application in products including, but not limited to, ventricular assist devices (VADs) or dialysis machines. In these applications, electrical connection can be made directly to the patient s heart during the operation of the device. Type BF or Body Floating is less stringent than CF is intended for applied parts which have conductive longterm contact with a patient. These products can include incubators, patient heating cooling equipment, ultrasound monitoring, cardiac monitoring, long term diagnostic equipment blood pressure monitoring to name a few. The least stringent requirement is for the type B, or Body, classification is reserved for applied parts that are not conductive can be immediately released from the patient. This includes, but is not limited to, LED lighting, medical lasers, medical imaging, hospital beds phototherapy equipment. Another difference between Type B, BF CF, is that type B applied parts may be 1
2 Table 1. Allowable values of PATIENT LEAKAGE CURRENTS PATIENT AUXILIARY CURRENTS (in µa) under NORMAL CONDITION SINGLE FAULT CONDITION Type b Type bf Type cf applied part applied part applied part Measuring Current Description Reference circuit nc sfc nc sfc nc sfc PATIENT Figure 19 dc AUXILIARY CURRENT ac PATIENT From PATIENT (a) dc LEAKAGE CONNECTION CURRENT to earth ac Caused by (c) Figure 17 dc an external voltage on ac a SIP/SOP Total PATIENT With the same types (a) dc LEAKAGE of APPLIED PART (h) Figure 20 ac 500 1, , CURRENT a connected together Caused by (c) Figure 17 dc an external voltage on (h) Figure 20 ac 500 1, , a SIP/SOP Table 1: Leakage current requirements according to specification: IEC , Table 3 - Reference: (a). Key: = NORMAL CONDITION; = SINGLE FAULT CONDITION Table 2. Allowable values of PATIENT LEAKAGE CURRENTS (in µa) under the special test conditions identified in Measuring Type b Type bf Type cf Current Description a Reference circuit applied part applied part applied part PATIENT Caused by an external voltage (b) Not applicable 5, LEAKAGE on the PATIENT CURRENT CONNECTION of an F f-type APPLIED PART Caused by an (d) Figure c external voltage on a metal ACCESSIBLE part not PROTECTIVELY earthed Total PATIENT Caused by an external voltage (b) Not applicable 5, LEAKAGE on the PATIENT (h) Figure 20 CURRENT b CONNECTION of an F f-type APPLIED PART Caused by an (d) Figure 18 1,000 1,000 c external voltage on a metal ACCESSIBLE (h) Figure 20 part not PROTECTIVELY earthed Table 2: Leakage current requirements according to specification: IEC Table 4 - Reference: (b). a b c The condition referred to in Table IV of the seond edition as MAINS VOLTAGE ON APPLIED PART, treated in that edition as a SINGLE FAULT CONDITION, is treated in this edition as a special test condition. The test with MAXIMUM MAINS VOLTAGE on a non-protectively EARTHED ACCESSIBLE PART is also a special test condition, but the allowable values are the same as for SINGLE FAULT CONDITION. See also the rationales for (d). Total PATIENT LEAKAGE CURRENT values are only applicable to equipment having multiple APPLIED PARTS. See (h). the individual APPLIED PARTS shall comply with the PATIENT LEAKAGE CURRENT values. This condition is not tested with TYPE CF APPLIED PARTS because it is covered by the test with MAXIMUM MAINS VOLTAGE on the APPLIED PART. See also the rationale for (D). 2
3 connected to earth ground. Type BF CF are not connected to earth ground this is why the specification refers to them as floating. g Testing Power Supplies to the Medical Stard Excelsys Technologies developed its XS series of power supplies for medical applications. Testing the power supplies to the stard was undertaken using both its basic low-leakage models, the XS N-000 XS N-004, respectively. The high efficiency XS1000 Power Supply delivers up to 1008W in an enclosed, fan cooled chassis. Nominal output voltages are 24V 48V with wide adjustment ranges user defined set points. Xsolo carries dual safety certification, EN nd Edition for Industrial Applications EN nd 3rd Edition for Medical Applications, meeting the stringent creepage clearance requirements, 4KV ac isolation <300uA leakage current. Boasting up to 92% efficiency, the Figure 1. Excelsys model XS1000 has been design tested for medical applications. XS1000 is ideal for use in acoustically-sensitive medical applications, harsh industrial environments, laboratory equipment HI-Rel/MIL-COTS applications. Excelsys used the isolation voltage ratings of each power supply to determine which one would be suitable for a particular application type (Table 3). Power supplies vary widely in terms quality reliability. There are design differences, depending on the supplier, many power supplies are not meant for medical applications. Type Input to Output Isolation Input to Ground Isolation Output to Ground Isolation B rated 4000V ac 1500V ac 500V ac BF/FC rated 4000V ac 1500V ac 1500V ac Table 3. Isolation voltages necessary for B, BF, or CF rated power supplies. g Testing Results The Excelsys testing procedures were performed in accordance with the IEC requirements using an ac source providing 264V ac at 63 Hz, which demonstrates worse-case conditions. The power supplies were tested to Tables specified by the IEC s specifications, as shown in Figures 2 3 on the following pages. The resulting data from Excelsys is shown in Tables 4 5 along with the requirements in the specification showing the limits for ac testing. These results indicate that the Xsolo products would be well-suited to either type BF or B applied applications. 3
4 Test 1: Patient Connection to Earth Result: PASS For the first test, Excelsys used the following power supply settings: Vin = 264V rms (maximum rated input voltage), Frequency = 63Hz (maximum rated input frequency) The test method was to follow the instructions shown in Figure 2 (based on IEC Fig 15.) Key Measure (with S 7 closed, if CLASS I ME EQUIPMENT) under all possible combinations of positions of S 1, S 5, S 10, S 13. S 1 open is SINGLE FAULT CONDITION. CLASS I ME EQUIPMENT only: Measure with S 7 open (SINGLE FAULT CONDITION) with S 1 closed under all possible combinations of S 5, S 10, S 13. For CLASS II ME EQUIPMENT, the PROTECTIVE EARTH CONNECTION S 7 are not used. Figure 2. (From IEC ). Measure circuit for patient leakage current from patient connection to earth. P 1 Excelsys measured the following parameters shown in Table 4: Measuring Test Condition XS N-000 XS N-004 Circuit (µa) (µa) S7 Open (Neutral) Vo- Vo+ Vo- Vo S5 Closed Reversed Input Polarity: S1 Open (Live) Table 4. Excelsys power supply test against IEC Table 15. Result: Pass. 4
5 Test 2: Patient Connection of an F-Type Applied Part Result: Pass For the second test, Excelsys used the following power supply settings: Vin = 264V rms (maximum rated input voltage), Frequency = 63Hz (maximum rated input frequency) Figure 3. (From IEC ). Measure circuit patient leakage current caused by an external voltage on the patient connection of an F-type applied part. The test method was to follow the instructions shown in Figure 3 (based on IEC Fig 16.) For legends, see Table 5. Key Measure (with S 7 closed, if CLASS I ME EQUIPMENT) with S1, closed under all possible combinations of positions of S 5, S 9, S 10, S 13. For CLASS II ME EQUIPMENT, the PROTECTIVE EARTH CONNECTION S 5 are not used. Excelsys measured the following parameters shown in Table 5: Measuring Test Condition XS N-000 XS N-004 Circuit (µa) (µa) S1 Open (Neutral) Vo- Vo+ Vo- Vo S5 Closed Reversed Input Polarity: S1 Open (Live) Table 5. Excelsys power supply test against IEC Table 16: Result: Pass. 5
6 g Summary Excelsys Technologies Xsolo power supplies have been designed tested to demonstrate that they are suitable for use in products intended for medical applications requiring compliance with either the B or BF classification. Moreover, It is not uncommon for applied products requiring type CF performance levels to need additional isolation methods such as isolated power converters in the system or medically-rated high-isolation transformers in the overall system design. Excelsys offers low leakage power supply versions, which enable incorporation of multiple power supplies into various medical-rated systems depending on the applications. Excelsys has a global applications team available to support assist customers with their system applications. References: Excelsys Technologies Ltd Excelsys Technologies Ltd 6
Why Electric Safety Standards
Paul Haggarty Why Electric Safety Standards To protect patients and caregivers from electrical shock To prevent electrical interference between instrument sub-systems/circuits To prevent electrical interference
More informationCHAPTER 2: ELECTRICAL SAFETY
CHAPTER 2: ELECTRICAL SAFETY 2.1 Physiological Effect of Electricity 2.2 Macroshock and Microshock Macroshock Microshock 2.3 Approaches toward Protection against Shock 2.4 Medical Safety Test Electrical
More informationTHE HEARTBEAT B E H I N D T H E HEARTBEAT MEDICAL POWER BROCHURE
THE HEARTBEAT B E H I N D T H E HEARTBEAT MEDICAL BROCHURE A B O U T EOS P O W E R EOS Power has been designing standard, modified and custom medical power solutions for the global medical market for
More informationSPECIFIC TECHNICAL CRITERIA
Issue Date: 2007-01-23 Page 4 of 57 Report Reference # E116994-A34-UL-1 SPECIFIC TECHNICAL CRITERIA TEST REPORT UL 60601-1 Medical Electrical Equipment Part 1: General requirements for safety Report Reference
More informationFOCUS ON MEDICAL POWER SUPPLIES. designing and producing a wide range of medical power supplies since 1986
FOCUS ON MEDICAL POWER SUPPLIES designing and producing a wide range of medical power supplies since 1986 ELECTRONICA 2016 special PSUs not available on the market. Multi-voltage output systems Limited
More information020: 2013 CEB SPECIFICATION MINIATURE CIRCUIT BREAKER (MCB)
020: 2013 CEB SPECIFICATION MINIATURE CIRCUIT BREAKER (MCB) CEYLON ELECTRICITY BOARD SRI LANKA Telephone: +94 11 232 0953 Fax: +94 11 232 3935 CONTENTS Page 1.0 Scope 3 2.0 System Parameters 3 3.0 Service
More informationOutdoor Distribution (15 kv through 25 kv) S&C Fault Tamer Fuse Limiter
Outdoor Distribution (5 kv through 25 kv) S&C Fault Tamer Fuse Limiter Introducing S&C s new generation of pole-top transformer protection... Application Although the S&C Fault Tamer Fuse Limiter handles
More informationElectrical Safety in Medical Equipment. D. J. McMahon rev cewood
Electrical Safety in Medical Equipment D. J. McMahon 141008 rev cewood 2017-10-23 Key Points Electrical Safety in Medical Equipment: - know the significance of NFPA-99 and the annual editions - define
More informationDate received into service; / /.
Innovating Together RIGEL SafeTest 50 electrical safety analyzer Copyright 2016 SEAWARD GROUP Last Update: April 20th 2016 Instruction Manual 410A570 Revision 1.2 Limited Warranty & Limitation of Liability
More informationTEST REPORT IEC Part 1: General requirements for basic safety and essential performance
jhćmtc Test Report issued under the responsibility of: TEST REPORT IEC 60601-1 Part 1: General requirements for basic safety and essential performance Report Number.... : T223-0128/16 Date of issue...
More informationCOVER PAGE FOR TEST REPORT
Issue Date: 2007-05-22 Page 1 of 2 Report Reference # E116994-A35-CB-1 Amendment 1 2008-04-24 COVER PAGE FOR TEST REPORT Test Item Description: DC Power Supplies Model/Type Reference: Class I Models: GNT2XXX-YYY
More informationElectrical Testing Company Pty Ltd / / AS/NZS 3003:2011 Body (Rev E)
AS/NZS 3003:2011 Electrical Testing Company Pty Ltd / www.electricaltestingcompany.com.au / AS/NZS 3003:2011 Body (Rev E) Body Protected Electrical Areas (shall comply with Section 2 and Section 3 of the
More informationWhite Paper # 211 IEC 601. What are the Implications for Power Quality?
White Paper # 211 IEC 601 What are the Implications for Power Quality? White Papers are produced as an educational and informational tool by POWERVAR, Inc., 1450 Lakeside Drive, Waukegan, IL 60085. Comments
More informationS&C Fault Tamer Fuse Limiter. Outdoor Distribution (15 kv through 25 kv)
S&C Fault Tamer Fuse Limiter Outdoor Distribution (5 kv through 25 kv) Introducing S&C s new generation of pole-top transformer protection... Application Although the S&C Fault Tamer Fuse Limiter handles
More informationUL TEST REPORT AND PROCEDURE
Issue Date: 2012-03-23 Page 1 of 16 Report Reference # E186249-A202-UL UL TEST REPORT AND PROCEDURE Standard: Certification Type: CCN: Product: Model: Rating: UL 60950-1, 2nd Edition, 2011-12-19 (Information
More informationLoad (2) Single or Dual
Features Regulated Converter Reinforced insulation for 25VAC working voltage Clearance and creepage distance >8.mm 5kVAC I/P to O/P 2MOPP isolation 2µA patient leakage current Industry standard pinout
More information2019 Medical Power Solutions Product Portfolio
09 Medical Power Solutions Product Portfolio Company Profile TRACO Electronic AG is a Swiss company with headquarters based in Baar, Switzerland. As a leading power supply specialist with more than 35
More informationDoes Your Electrical Safety Testing Routine Need a Boost? Exploring the Benefits of the Line Leakage and Functional Run Tests
Does Your Electrical Safety Testing Routine Need a Boost? Exploring the Benefits of the Line Leakage and Functional Run Tests Introduction Most electrical product manufacturers perform safety tests on
More informationSafeTest 60. A simple, robust and cost-effective medical safety analyser for general electrical safety testing. Key Features
SafeTest 60 A simple, robust and cost-effective medical safety analyser for general electrical safety testing The Rigel SafeTest 60 is a robust, reliable and dedicated medical safety analyser. It is ideal
More information013 : 2009 CEB SPECIFICATION MOULDED CASE CIRCUIT BREAKERS
013 : 2009 CEB SPECIFICATION MOULDED CASE CIRCUIT BREAKERS FOR OVERHEAD NETWOKS CEYLON ELECTRICITY BOARD SRI LANKA Specification for MOULDED CASE CIRCUIT BREAKERS FOR OVERHEAD NETWOKS CEB Specification
More informationInstallation Manual for D154X Series Multi-Indicator, General Purpose Power Supplies
Installation Manual for D154X Series Multi-Indicator, General Purpose Power Supplies D1541 D1542 D1543 D1545 D1543-X-8 D1545-X-8 D1545-4BNC D154X Series 1A 2A 3A 5A 3A fitted with 8-way output splitter
More informationInvestigation into UK socket-outlets incorporating USB charging points
Investigation into UK socket-outlets incorporating USB charging points Electrical Safety First investigated a number of commercially available UK socket-outlets incorporating USB ports, as a repeat of
More informationTEST REPORT IEC Part 1: General requirements for basic safety and essential performance
Test Report issued under the responsibility of: TEST REPORT IEC 60601-1 Part 1: General requirements for basic safety and essential performance Report Number.... : T223-0347/14 Date of issue... : 2014-12-05
More informationUnified requirements for systems with voltages above 1 kv up to 15 kv
(1991) (Rev.1 May 2001) (Rev.2 July 2003) (Rev.3 Feb 2015) (Corr.1 June 2018) Unified requirements for systems with voltages above 1 kv up to 15 kv 1. General 1.1 Field of application The following requirements
More information2.7. IT system floor-standing distribution cabinet Series -IPS-F
Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str. 65 35305 Grünberg Tel.: 0640 07-0 Fax: 0640 07-25 IT system floor-standing distribution cabinet Series -IPS-F for supplying power to medical locations
More informationTechnical information No. 01. IT systems. The basis for reliable power supply
IT systems The basis for reliable power supply FA01en/01.2004 IT systems The basis for reliable power supply in critical areas The advantages of sophisticated industrial systems can only be of use, if
More informationSpecification for 70mm pole pitch Air circuit breaker up to 1600 A
Specification for 70mm pole pitch Air circuit breaker up to 1600 A Protective device for low voltage electrical installation Last update :2011-07-08-1 - Table of contents: 1 General...3 2 Compliance with
More informationElectrical Testing Company Pty Ltd / download at electricaltestingcompany.com.au / AS/NZS 3003:2011 Cardiac (Rev C)
AS/NZS 3003:2011 Cardiac Protected Electrical Areas (shall comply with Section 2 and Section 4 of the standard) The following is a guide for electricians involved with electrical installations for patient
More informationINSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS.
INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS. Before installing and/or using this product, please check for any visual damage of the enclosure, power
More informationPRODUCT / TEST MANUAL 2L9K2 BATTERY EARTH FAULT RELAY
Sheet 1 of 5 Order Number Serial Number PRODUCT / TEST MANUAL 2L9K2 BATTERY EARTH FAULT RELAY Issue Date Level B 27/05/1999 Initial issue. Summary of changes Due to RMS continuous product improvement policy
More informationINSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS.
INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS. Before installing and/or using this product, please check for any visual damage of the enclosure, power
More informationProduct Manual & Installation Guide Vertical Rack Mount 5.0 to 8.6 kw
Product Manual & Installation Guide Vertical Rack Mount 5.0 to 8.6 kw 30A 208V 20A 208V 3 Phase 30A 208V 3 Phase Important Information Specifications Installation & Operation Outline & Wiring Drawings
More informationCompany Overview. Product Offering & Attributes
Company Overview TDK-Lambda is one of the oldest and most trusted manufacturers of high quality medical power supply solutions. We offer a broad product line with over 5,000 models to choose from. Backed
More informationRE-PR3-E-86&105 3-Phase Panel Mount 86 and 105kW
Page 1 of 6 3-Phase Panel Mount 86 and 105kW Features: Benefits: 0-10Vdc, 0-5Vdc, 4-20mA or manual via potentiometer control input Over temperature protection with auto reset Enclosed panel mounting Efficient
More informationContactors. CL1115/02, CL1215/02, CL1315/02 1, 2 and 3 pole AC and DC NO contactors for voltages up to 1,500 V Catalogue C25.en
Connect - Contact - Control 3 Contactors CL1115/02, CL1215/02, CL1315/02 1, 2 and 3 pole AC and DC NO contactors for voltages up to 1,500 V Catalogue C25.en 2 CL1115/02, CL1215/02, CL1315/02 1, 2 and 3
More informationCopy of marking plate. Page 2 of 50 Report No. TRS
Copy of marking plate Page 2 of 50 Report No. TRS09030062 Page 3 of 50 Report No. TRS09030062 GENERAL INFORMATION Test item particulars (see also clause 5): Classification of installation and use... :
More informationEnsuring the Safety Of Medical Electronics
Chroma Systems Solutions, Inc. Ensuring the Safety Of Medical Electronics James Richards, Marketing Engineer Keywords: 19032 Safety Analyzer, Medical Products, Ground Bond/Continuity Testing, Hipot Testing,
More informationIT 28 ELECTRICAL INSPECTOR EXAMINATION ANSWER SCHEDULE. Question 1 Marks Reference Marking notes. (1 mark) ESR 57(3)(e)(ii)
IT 28 ELECTRICAL INSPECTOR EXAMINATION ANSWER SCHEDULE Notes:1. means that the preceding statement/answer earns 1 mark. 2. This schedule sets out the expected answers to the examination questions. The
More informationSwitching DC Power Supply
99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Model 1693, 1694 Switching DC Power Supply INSTRUCTION MANUAL 1 Safety Summary
More informationDIGITAL POWER. Product catalogue. roallivingenergy.com
DIGITAL POWER Product catalogue roallivingenergy.com DIGITAL POWER Ultra compact Switching Power Supplies DIGITAL DISPLAY IT NETWORKING MAJOR APPLIANCES REALLY COOL EFFICIENT POWER SOLUTIONS TELECOM INDUSTRIAL
More informationElectric Vehicle Charging Safety Guidelines Part 2: Selection and Installation Edition DRAFT
Date of publication: 1 November 2016Not yet published Issued by: Mark Wogan, Manager Energy Safety WorkSafe New ZealandNot yet issued Electric Vehicle Charging Safety Guidelines Part 2: Selection and Installation
More informationORION PLUS three-phase kVA
ORION PLUS three-phase 30-1250kVA Orion Plus stabilizers are available for different ranges of input voltage fluctuation. In the ±15%/ ±20% and ±25%/ ±30% types, the change of input range is obtained through
More informationReliability with the most essential intellect. Philips HF-Essential Electronic ballasts for TL5 lamps
Reliability with the most essential intellect Philips HF-Essential Electronic ballasts for TL5 lamps HF-Essential - Reliability with the most essential intellect Electronic ballasts for TL5 lamps The cost-saving
More informationXiria the smart solution
Xiria 24 kv Ring Main Unit Xiria the smart solution High operational safety Maintenance-free Safe visible earthing and position indication Environmentally friendly Compact Suitable for substation automation
More informationشركة كهرباء محافظة القدس المساهمة المحدودة
Our Ref: Date: Messrs. Dear Sir, Tender 22/2017 (re-bidding) 11& 33Kv Fuse Cutouts and Links You are kindly requested to bid for supply and delivery (DDP) for the medium voltage overhead line accessories
More informationSD, SE, SDE, SCO series Multipole switchgear for rail vehicles: Disconnecting switches, Earthing Switches, Disconnector with Earthing Switches,
4 Electrics for Rolling Stock Multipole switchgear for rail vehicles: Disconnecting switches, Earthing Switches, Disconnector with Earthing Switches, Change-Over-Switches 600 V to 3 kv Catalogue F84.en
More informationRegulatory Compliance Engineering News
UPCOMING REGULATORY REQUIREMENTS 1. IEC / EN60950-1 will soon be replaced by IEC / EN62368-1 2015 2016 2017 2018 2019 June 20 IEC / EN60950-1 (Safety Standard for ITE) is to be withdrawn by June 20, 2019.
More informationApplications. Overview. Technical Specifications
FLLD4 TH, 53/35 VAC, 8 A High Performance Chassis Mount Three-Phase and Neutral Filters Overview These are compact, general purpose, three-phase + N filters, with terminal blocks for quick installation
More informationXiria the smart solution
Xiria 24 kv Ring Main Unit in IEC and ENATS 41-36 design Xiria the smart solution High operational safety Maintenance-free Safe visible earthing and position indication Environmentally friendly Compact
More information1. The term "withstand" means "the unit will remain in place without separation of any parts from the device when subjected to the seismic forces.
SECTION 262816 - ENCLOSED SWITCHES AND CIRCUIT BREAKERS PART 1 - GENERAL 1.1 SUMMARY A. Section Includes: 1. Fusible switches. 2. Nonfusible switches. 3. Receptacle switches. 4. Shunt trip switches. 5.
More informationSOLID STATE POWER CONTROLLERS ACL-series AC1-series AC3-series ACF3-series DAC3- series
H & C SOLID STATE POWER CONTROLLERS ACL-series AC1-series AC3-series ACF3-series DAC3- series OPERATION This power controller is designed to regulate a resistive load by switching the load on and off in
More informationELECTRICIANS REGULATIONS EXAMINATION 25 June 2011
ER42 Candidate Code No. For Board Use Only Result Date Result Date Int ELECTRICIANS REGULATIONS EXAMINATION 25 June 2011 Int QUESTION AND ANSWER BOOKLET INSTRUCTIONS READ CAREFULLY Time Allowed: Three
More informationTEST REPORT IEC Information technology equipment Safety Part 1: General requirements
Test Report issued under the responsibility of: TEST REPORT IEC 60950-1 Information technology equipment Safety Part 1: General requirements Report Number.... : 31282706.003 Date of issue... : 22 nd March,
More informationSmartPro Medical-Grade UPS, Line Interactive, Lithium Battery, 6 Outlets - 230V, 1kVA, 750W, Full Isolation
SmartPro Medical-Grade UPS, Line Interactive, Lithium Battery, 6 Outlets - 230V, 1kVA, 750W, Full Isolation MODEL NUMBER: SMX1200XLHGL Highlights Lithium battery backup allows time to save files and safely
More informationburners with or without fan. It conforms to EN C). APPROVAL EU type test approval as per EU Gas Appliance Directive.
LAST UPDATE 29/08/12 INSTALLATION INSTRUCTIONS v. 3 UNIT SPECIFICATIONS Main Gas Valve, Pilot Gas Valve Ionisation, Ignition and Flame Monitoring with Common Electrode or Remote Sensor Integrated Cyclic
More informationLED Driver Linear / area fixed output
Driver LC 65W 4-7mA flexc lp ANCED series Product description Built-in constant current LED Driver Adjustable output current between 4 and 7 ma Max. output power 65 W Up to 94 % efficiency Nominal life-time
More informationXiria 12 kv Ring Main Unit in IEC and ENATS design
Xiria 12 kv Ring Main Unit in IEC and ENATS 41-36 design Product Focus High operational safety Maintenance-free Safe visible earthing and position indication Environmentally friendly Compact Suitable for
More informationISO-TECH IPS1810H and IPS1603D DC power supply ISO-TECH IPS 1810H et IPS 1603D Alimentation électrique c.c.
ISO-TECH IPS1810H and IPS1603D DC power supply ISO-TECH IPS 1810H et IPS 1603D Alimentation électrique c.c. ISO-TECH IPS 1810H 204-713 ISO-TECH IPS 1603D 204-729 82IP-1810HME Ei Statement of Compliance
More informationModel Number Structure
Solid State Relays (Single-phase) G3PB New Single-phase Solid State Relays with Compact Size for Heater Control Slim models with a thickness of only 22.5 mm are also available. Compact design achieved
More informationElectric Vehicle Charging Safety Guidelines Part 2: Selection and Installation
86 Customhouse Quay PO Box 165, Wellington 6140 New Zealand Date of publication: 1 November 2016 Issued by: Mark Wogan, Manager Energy Safety WorkSafe New Zealand Electric Vehicle Charging Safety Guidelines
More informationEaton 93PS 8-40kW UPS Technical Specification Manufacturer s declaration in accordance with IEC
Manufacturer s declaration in accordance with 8- Frame Model catalogue reference 93PS-XX(20)-YY- 93PS-XX(40)-YY- Number of power modules (UPM) 1 1 or 2 Number of internal batteries 0 to 2 x 32 blocks 0
More informationFeatures. LED Driver. RACT09 9 Watt. TRIAC Dimmable. Single Output RACT09- AC/DC Converter
Features TRIAC Dimmable LED Driver Triac dimmable with leading or trailing edge dimmers Class II with SELV output (no earth required) Extra-large screw terminals and integrated cable clamps for easy installation
More informationIndependent Testing Laboratory for High Voltage Equipment
T E S T I N G L A B O R A T O R Y Independent Testing Laboratory for High Voltage Equipment 1 CHALLENGING ENVIRONMENT The primary concern for utilities is to ensure grid reliability, efficiency, and security.
More informationLA Motor Protection. A South African Company to be Proud of
LA Motor Protection Relay Physical Address: 298 Soutter Street, Pretoria West Tel: 083 454 6949, +27 12 327 1729 Fax: +27 (0)12 327 1733 Toll Assist: 0860 10 30 41 www.newelec.co.za sales@newelec.co.za
More informationROLF JANSSEN GMBH ELEKTROTECHNISCHE WERKE LOW VOLTAGE SWITCH PANEL
ROLF JANSSEN GMBH ELEKTROTECHNISCHE WERKE LOW VOLTAGE SWITCH PANEL TopStick Fixed installation technology LOW VOLTAGE SWITCH PANEL TopDraw plus The modular system solution Quality, Conformity Increased
More informationThe University of New South Wales. School of Electrical Engineering and Telecommunications. Industrial and Commercial Power Systems Topic 6
The University of New South Wales School of Electrical Engineering and Telecommunications Industrial and Commercial Power Systems Topic 6 PROTECTIONS 1 FUNCTION OF ELECTRICAL PROTECTION SYSTEMS Problems:
More informationIndoor Linear 50W Driver SI-CU WW
Product Data Sheet Rev.2.0 2016. 5. 27 1# LED Driver Indoor Linear 50W Driver SI-CU1425001WW Constant Current LED Driver Features & Benefits Output Current Range: 500~1400 ma (adjustable via R-set) Output
More informationHigh voltage products PASS M0S 252 kv Innovative solution for transmission substation up to 252 kv
High voltage products PASS M0S 252 kv Innovative solution for transmission substation up to 252 kv PASS M0S 252 kv Innovation In today s market, the HV substation is increasingly becoming a key element
More informationMV Air Insulated Switchgear TAP17. Technical Data TGOOD
MV Air Insulated Switchgear TAP17 Technical Data TGOOD 2017.1 Operating environmental conditions Place of installation: Indoor or outdoor Ambient temperature: 25 ~ +40 (higher or lower temperature optional)
More informationTechnical Specification. for. Supply, installation and commissioning of Diesel Generators for UNDP
United Nations Development Programme Recovery and Crisis Prevention برنامج األمم المتحدة اإلنمائي Iraq Technical Specification for Supply, installation and commissioning of Diesel Generators for UNDP February
More informationMARTINDALE INSTRUCTIONS ELITE FUSE FINDER KIT ELECTRIC. Trusted by professionals. 4.4 Storage Conditions
4.4 Storage Conditions The FD650/R and FD500/T or FD600/T should be kept in warm dry conditions away from direct sources of heat or sunlight, with the battery removed and in such a manner as to preserve
More informationThe High Power Stack Horizon
Voltage Supply 480-600-690V Current Rating from 150 to 2100A Designed to drive 1-2-3 Phase Loads Internal Fuse with Micro for Fuse Failure Stall Fan Protection for 1100 to 2100A Control Board with Plug
More informationOPERATING AND MAINTENANCE MANUAL. Primary Current Injection Test Set. 750ADM-H mk2
OPERATING AND MAINTENANCE MANUAL Product: Type: Primary Current Injection Test Set 750ADM mk2 750ADM-H mk2 DESIGNED AND MANUFACTURED BY: T & R Test Equipment Limited 15-16 Woodbridge Meadows, Guildford,
More informationUser Manual. Charger for LTO Batteries. User Manual. Language specific user manuals are available on
User Manual Charger for LTO Batteries EN User Manual Language specific user manuals are available on www.mascot.no/downloads/usermanuals Bruksanvisning Käyttöohjeet Bedienungsanleitung Mode d emploi Manual
More informationUL TEST REPORT AND PROCEDURE
Issue Date: 2012-08-17 Page 1 of 21 Report Reference # E186249-A203-UL UL TEST REPORT AND PROCEDURE Standard: Certification Type: CCN: Product: Model: Rating: UL 60950-1, 2nd Edition, 2011-12-19 (Information
More informationInterruption Technology of Breakers for High-voltage Direct Current
Interruption Technology of Breakers for High-voltage Direct MORIAI Hiroshi ABSTRT Applications for direct current () electric distribution have been spreading along with the increase of data centers and
More informationSize and Configurations
!!"#!""!"$%!!" &'()&('& AK Axial Flow Fans the high efficiency direct-driven fan with adjustable pitch impeller specially designed to operate with low noise emission and vibration that meets the stringent
More informationR-MAG. Vacuum Circuit Breaker with Magnetic Actuator Mechanism
R-MAG Vacuum Circuit Breaker with Magnetic Actuator Mechanism R-MAG Features: Low maintenance 10,000 mechanical operations (five times ANSI requirements) Simple magnetic actuator Vacuum interruption Definite
More informationTransformer bushings, type GOH. Technical guide
Transformer bushings, type GOH Technical guide This Technical Guide has been produced to allow transformer manufacturers, and their designers and engineers, access to all the technical information required
More informationThree-phase 200 V series: FRN0025 to 0060C2S-2U Three-phase 400 V series: FRN0013 to 0030C2S-4U
Instruction Manual Supplement for FRN-C2S-2U/4U (7.5 to 20 HP) Compact Inverter 200 V series: FRN0025 to 0060C2S-2U series: FRN0013 to 0030C2S-4U Thank you for purchasing our FRENIC-Mini series of inverters.
More informationELECTRICIANS REGULATIONS EXAMINATION 23 November 2013 QUESTION AND ANSWER BOOKLET
ER54 Candidate Code No. For Board Use Only Result Date Result Date Int ELECTRICIANS REGULATIONS EXAMINATION 23 November 2013 QUESTION AND ANSWER BOOKLET Time Allowed: Three hours INSTRUCTIONS READ CAREFULLY
More informationBATTERY CHARGERS NO. 3
Date: January 10, 1992 BATTERY CHARGERS NO. 3 To: Subject: Manufacturers of Commercial/Industrial Battery Chargers Expansion of CSA Certification Service for Power Supplies to Include Low Voltage, Outdoor
More informationSelection Guide Motor Control Device Solutions
Selection Guide Motor Control Device Solutions Expect more and get it from c3controls. Our portfolio of Motor Control Devices consists of worldclass products designed and manufactured to meet your requirements
More informationA Caspian Technology Company all materials presented are trade mark and copyright protected by C.M.TECHNOLOGY Pty Ltd
PPS12 230v AC to 12 VDC POWER SUPPLY WITH BATTERY MANAGER OPTION 2 Table of Contents Introduction:... 3 1.0: Input Voltage:... 4 1.1: Mains Hold Up (missing mains cycles):... 4 1.2: Mains Input:... 4 1.3:
More informationCORCOM /// EMI FILTERS FOR MEDICAL APPLICATION - PRODUCT OVERVIEW. CORCOM EMI Filters for Medical Application - Product Overview
CORCOM /// EMI FILTERS FOR MEDICAL APPLICATION - PRODUCT OVERVIEW CORCOM EMI Filters for Medical Application - Product Overview CORCOM /// EMI FILTERS FOR MEDICAL APPLICATION - PRODUCT OVERVIEW WHAT S
More informationE-15 Uninterruptible Power Systems (UPS)
Guideline No.E-15 (201510) E-15 Uninterruptible Power Systems (UPS) Issued date:20 October, 2015 China Classification Society Foreword This Guide is a part of CCS Rules, which contains technical requirements,
More informationMolded Vacuum Reclosers The recloser you want, all in one package. Single- and Three-Phase Molded Vacuum Reclosers, 15 38kV
The recloser you want, all in one package. Single- and Three-Phase, 15 38kV The need for automated reclosers has never been greater, but many of today s reclosers come with penalties. They weigh too much,
More informationUL TEST REPORT AND PROCEDURE
Issue Date: 2016-04-08 Page 1 of 24 Report Reference # E186249-A263-UL UL TEST REPORT AND PROCEDURE Standard: Certification Type: CCN: Product: Model: Rating: UL 60950-1, 2nd Edition, 2014-10-14 (Information
More informationProduct Portfolio 2017 Medical
Product Portfolio 2017 Medical PAGE 2 TRACO POWER products for applied versus non-applied medical requirements For non-applied parts sections of medical equipment, power and safety requirements can be
More informationFile E47705 Project 97ME August 21, 1997 REPORT. MOTOR CONTROLLERS, MANUAL Amberg, FED REP Germany
File E47705 Project 97ME18400 August 21, 1997 REPORT on MOTOR CONTROLLERS, MANUAL Amberg, FED REP Germany Copyright 1997 Underwriters Laboratories Inc. Underwriters Laboratories Inc. authorizes the above
More informationINTER PLANT STANDARD - STEEL INDUSTRY THYRISTOR CONVERTORS FOR ARMATURE AND FIELD OF MAIN DRIVE MOTORS. Corresponding IS does not exist
INTER PLANT STANDARD - STEEL INDUSTRY IPSS THYRISTOR CONVERTORS FOR ARMATURE AND FIELD OF MAIN DRIVE MOTORS Corresponding IS does not exist IPSS: 1-10-038-08 0 FOREWORD 0.1 This Inter Plant Standard has
More informationZHEJIANG MAXGE ELECTRIC TECHNOLOGY CO., LTD. Address... :
Test Report issued under the responsibility of: TEST REORT IEC 60898-1 Circuit-breakers for over current protection for household and similar installations art 1 - Circuit-breakers for a.c. operation Report
More informationLoad Com/Diff(KV) (Kg) Rating * V 240V 277V V 240V 277V V 240V 277V > / 4 0.
Approbations IS 15885 ( art 2 / Sec 13 ) Output Output Output Efficiency Max Case Input Max Input Inrush THD @ ower Surge Weight Envir. ower Voltage Current at Max Load Temp Current ower Current Max Load
More informationPD-Series Ground Fault Circuit Protection
PD-Series Ground Fault Circuit Protection SmartGuard is an equipment ground fault protection device that functions as a standard high-quality Carling hydraulic/ magnetic circuit breaker, offering customized
More informationCertificate of Compliance
Certificate of Compliance Certificate: 1588199 (LR 12730-233) Master Contract: 165071 Project: 1588199 Date Issued: 2004/11/03 Issued to: Siemens AG A&D CD CC TS2 Werner-von-Siemens-Strasse 48 Amberg,
More informationuni-cart THE RIGHT CHOICE WHERE SPACE IS AT A PREMIUM uni-cart
THE RIGHT CHOICE WHERE SPACE IS AT A PREMIUM is a compact and safe equipment cart with a loading capacity of up to 65 kg / 143 lbs. With its 4 or 5-foot-base frames and lockable wheels this trolley is
More informationINTER PLANT STANDARD STEEL INDUSTRY SPECIFICATION FOR STATIC EXCITATION CONVERTORS FOR SYNCHRONOUS MOTORS (FIRST REVISION)
INTER PLANT STANDARD STEEL INDUSTRY I P S S SPECIFICATION FOR STATIC EXCITATION CONVERTORS FOR SYNCHRONOUS MOTORS (FIRST REVISION) Corresponding IS does not exist IPSS: 1-10-036-12 0 FOREWORD 0.1 This
More informationTEST REPORT. NO.: A001T Date: Jul 29, 2014 Page 1 of 7
TEST REORT NO.: A001T140610025001-2 Date: Jul 29, 2014 age 1 of 7 Applicant: EASY WAY(HONG KONG) CO.,LTD Address: Flat D, 13F Sing Kong Building 233 Lockhard Wanchai Hong Kong Report on the submitted sample
More informationSINAMICS GM150 IGCT version
/2 Overview /2 Benefits /2 Design /6 Function /8 Selection and ordering data /8 Options Technical data /14 General technical data /15 Control properties /15 Ambient conditions /16 Installation conditions
More informationTriple offset butterfly valve meets the stringent requirements of TA-Luft and ISO Advanced environmental protection the ZETRIX way!
Triple offset butterfly valve meets the stringent requirements of TA-Luft and ISO 15848 Author: Michael Vullhorst (Laboratory Manager ARI-Armaturen) Introduction Both TA-Luft [1] and DIN EN ISO 15848-1
More information