SIGNET 2750 ph/orp Sensor ENGLISH

Size: px
Start display at page:

Download "SIGNET 2750 ph/orp Sensor ENGLISH"

Transcription

1 ENGLISH B-8/01 English SAFETY INSTRUCTIONS FOR IN-LINE ELECTRODE INSTALLATION 1. Do not remove from pressurized lines. 2. Do not exceed maximum temperature/pressure specifications. 3. Wear safety goggles or face shield during installation/service. 4. Do not alter product construction. Failure to follow safety instructions may result in severe personal injury! Description: +GF+ SIGNET 2750 ph/orp Sensors offer two output options: If provided with VDC power, the 2750 will output a two-wire 4 to 20mA loop. If provided with 5 VDC power, the 2750 will output digital serial data (the format is S 3 L : Signet Sensor Serial Link, a proprietary digital output for use with +GF+ SIGNET equipment). The 2750 automatically recognizes +GF+ SIGNET 275X ph or ORP electrodes and scales the current output for the appropriate range. The range is factory set for full scale measurement: 4 to20 ma = 0 to 14 ph 4 to20 ma = to mv ORP. Custom scaling is available from the factory. Three versions are available to accommodate different application requirements: In-line installations In-line installations with Local EasyCal Submersible installations, 3/4 in. NPT threads Submersible installations, ISO 7-1/R 3/4 threads Technical Data General Compatible Electrodes: +GF+ SIGNET 275X ph/orp Electrodes Temperature sensor: PT1000 Operational Range: ph: 0.00 to ORP: ±2000 mv Temp. 0 C to 85 C (32ºF to 185ºF) Response Time (includes electrode response): ph: <6 s for 95% of change ORP: application dependent Temp τ: 140 seconds (2754), 196 seconds (2756) Materials: &2 (in-line): PBT (thermal plastic polyester) (submersible): CPVC Weight: &2 (in-line): 0.75 kg (1.75 lb.) (submersible): 0.64 kg (1.4 lb.) Electrical Input Impedance: >10 11 Ω Output update rate: 0.5 s Current output: Power: 12 to 24 VDC ± 10% regulated, 20 ma max. ph scale: Fixed 4 to 20 ma, isolated, 0.00 to (custom scaling available) ORP scale: Fixed 4 to 20 ma, isolated, to 2000 mv (custom scaling available, to 2000 mv) Max Loop Resistance: 50Ω 12V 325Ω 18V 600Ω 24V Accuracy: 25ºC Temperature drift: ±1µA per ºC Resolution: ± 5µA Error indication: 3.6 ma S3L output: Description: Serial ASCII, TTL level 9600 bps Power: 5VDC ±5% regulated, 3 ma max Accuracy: ph: ± C ORP: ± 2 25 C Resolution: ph: 0.02 ph ORP: 1 mv Temp. 0.2 C Error indication: Temp output "+999.9" Environmental Ambient Temp.: -10 C to 85 C Storage Temp.: -20 C to 85 C Immunity: EN Emissions: EN Enclosure Rating: &2 (in-line): NEMA 4X/IP65 (with electrode connected) (submersible): NEMA 6P/IP68 (with electrode and watertight extension pipe connected) Standards and Approvals CE Manufactured under ISO 9001 & ISO page 1 of 8

2 Technical Data General Compatibility: +GF+ SIGNET 2750 ph/orp sensor Pressure/Temperature rating (ph and ORP electrodes) Pipe Size Range: 1/2 and up. Use +GF+ SIGNET installation fittings to 4 in. Use pipe adapters in pipes over 4 in. Submersion with 2750 sensor requires 3/4 in. NPT or ISO 7-1/R 3/4 male threaded extensions Efficiency: 25 C (77 F) Response Time: ph: < 5 s for 95% of signal change ORP: Application dependent BAR PSI Primary Functions: 2754, 2755: Flat surface resists fouling 2756, 2757: Bulb surface for general use 2754-HF: Extended use in applications with trace hydrofluoric acid (<2%) 2756-DI: Extended use in pure waters (<100 µs) C F 2754 and 2756 ph Electrodes Operating Range: 0 to 14 ph Wetted Materials: CPVC Body Glass FPM O-rings Porous UHMW Polyethylene reference junctions ph Reference: Electrolyte: Solidified Acrylamide Gel 3.5M KCI (2754, 2754-HF, 2756) 0.1 M KCl (2756-DI) Element: Ag/AgCl Temperature Sensor: PT-1000 (1096.8Ω = 25 C) Response Time, τ: 140 secs. (2754), 196 secs. (2756) /2757 ORP Electrodes Operating Range: to mv Wetted Materials: CPVC Body Glass FPM 0-rings Porous UHMW Polyethylene reference junctions Platinum sensing surface ORP Reference: Solidified Acrlyamide Gel:3.5M KCL saturated with AgCl Element: Ag/AgCl Temperature and pressure specifications ORP electrode operating Temperature/Pressure C to 65ºC ( to 149 F) ºC to 85 C (58 to 185 F) 2756 and 2756-DI ph electrode operating Temperature/Pressure C to 65ºC ( to 149 F) ºC to 85 C (58 to 185 F) 2754 and 2754-HF ph electrode operating Temperature/Pressure C to 65ºC ( to 149 F) ºC to 85 C (58 to 185 F) Storage Temperature: > -12 C (11 F) Cable supplied: 15 ft. 3 conductor shielded, 22 AWG. May be extended up to 122 m (400 ft) Submersible installation Flat ph Flat ORP 1.06 in./ 27 mm Bulb ph Bulb ORP 1.06 in./ 27 mm In-Line installation 3.7 in./ 94 mm 1.4 in./ 35.3 mm 4.3 in./ 109 mm 1.4 in./ 35.3 mm 4.6 in./ 116 mm 7.1 in./ 181 mm 8.2 in./ 208 mm 2.9 in./ 74 mm 3.2 in./ 81 mm 0.72 in./ 18 mm 0.72 in./ 18 mm 2.16 in./ 55 mm page 2 of 8

3 2750 In-line ph Sensor Installation The in-line sensor (without Local EasyCal) is designed for applications where electrode calibration is facilitated by remote equipment. The In-Line sensor includes Local EasyCal. A +GF+ SIGNET installation fitting (1/2 in. to 4 in.) or a +GF+ SIGNET pipe adapter is required to secure the electrode in the pipe. The +GF+ SIGNET Measurement and Instrumentation catalog offers a complete selection of fittings. Type Description Type Description Plastic tees Available in 1/2 in. to 4 in. sizes PVC, CPVC w/solvent cement socket PVDF and PP w/union end fittings Carbon steel weldolets Available in 2 in. to 4 in. sizes Requires 1-7/16 in. hole in pipe Install by certified welder only PVC saddles Available in 2 in. to 4 in. sizes Requires 1-7/16 in. hole in pipe Carbon steel threaded tees Available in 1/2 in. to 2 in. sizes Female NPT ends Iron strap-on saddles Available in 2 in. to 4 in. sizes Requires 1-7/16 in. hole in pipe Universal pipe adapters Adapt any fitting with 1-1/4 in. outlet PVC, CPVC, or PVDF versions Specify socket or 1-1/4 inch NPT male threads NPT Junction box 4.2 in. 107 mm YES YES 3.7 in./ 94 mm 30º Universal Junction box 4.2 in./ 106 mm NO NO 3.7 in./ 94 mm NO Vertical (0 ) position optimum Do not install within 30º of horizontal page 3 of 8

4 2750 In-line ph Sensor Assembly 1. Insert electrode into Signet installation fitting. Seat the electrode tabs into the alignment notches in the fitting. 2. Thread the in-line retaining cap onto fitting to secure the electrode into place. Hand-tighten the cap. DO NOT use any tools to avoid damaging the threads. Note: Keep the electrical interconnection between electrode and sensor electronics dry and clean at all times. 3. Disengage the locking ring on base of 2750 (The ring is disengaged when the lines on the ring and body of the 2750 are not aligned.) Unlocked 4. Place 2750 onto top of electrode and turn until the assembly drops into position. 5. Turn locking ring 1/4 turn to secure the 2750 assembly. (The ring is locked when the line on the ring and the line on the body of the 2750 are aligned.) Locked In-Line Sensor and Electrode Removal To remove electronics assembly only: Turn locking ring 1/4 turn; lift assembly straight up. To remove electrode from the pipe: Remove electronics assembly, then unthread retaining cap; pull electrode straight up. To remove the complete 2750 system: Hold locking ring in place. Unthread retaining cap and pull electrode straight up. page 4 of 8 CAUTION! Do not remove the electrode from a pressurized pipe. Wear appropriate protective clothing when working with the system.

5 Submersible ph Sensor Assembly and Installation 1. Insert electrode into base of the and turn until keyed contacts are seated. 2. Thread retaining cap over electrode and hand-tighten onto /4 in. NPT threads or ISO 7-1/R 3/4 threads 3. Attach 3/4 in. watertight pipe to the top of the Secure the threaded connection to prevent any leakage. For additional defense against possible accumulation of condensation at the back seal area of the sensor, fill the lower 3-4 inches ( mm) of conduit or extension pipe with a flexible sealant such as silicone. A light coat of non-petroluem based lubricant on the o-ring with will ease installation. Fill with 3 to 4 in. of sealant Note: Keep the electrical interconnection between electrode and sensor electronics dry and clean at all times. Use standard installation hardware to connect the submersible directly to external equipment. OR The 8052 NPT Mount Junction box (or with Local EasyCal) connects to 3/4 in. pipe or conduit and provides convenient wiring termination. OR The 8050 Universal Mount junction box (or with Local EasyCal) mounts flat onto a wall or can be strapped to a post or pipe. 1/2 in. NPT threads Fill with 3 to 4 in. of sealant page 5 of 8

6 Wiring the submersible 2750 direct to external equipment Use the wiring diagram below for systems without a +GF+ SIGNET junction box. Wiring the submersible 2750 with a junction box +GF+ SIGNET junction boxes include a terminal block inside the yellow cover. For Loop output, use 2 cond w/shield, AWG For S 3 L output, use 3 cond w/shield, AWG Out to external equipment + SW1 Run SW2 Cal BLK D1 Fill with 3 to 4 in. of sealant Output to external equipment - S RED WHT SHLD Input from wire color Terminal block label 4-20 ma Output Function S 3 L Output Function BLK + Loop + (12-24 VDC) V+ (5 VDC ±1%) RED WHT - No connection Loop - Data Out/Calibrate Signal Ground SHLD S Earth Ground Earth Ground SW1 Run SW2 Cal + BLK D1 Output to external equipment - S RED WHT SHLD The sensor input for the in-line 2750 is factory-connected Wiring the in-line 2750 page 6 of 8

7 Calibration The 2750 ph/orp Sensor is factory calibrated for out-of-the-box accuracy when used with new +GF+ SIGNET 275x ph and ORP electrodes. Once in service, periodic cleaning and calibration is necessary to compensate for changes in electrode output as a result of coating and aging. The required frequency is application-dependent, but once/week for cleaning and twice/month for calibration is good for a start. The 2750 ph/orp Sensor can also be used as a blind transmitter (provides 4 to 20 ma output but has no local display of ph or ORP value) so Local EasyCal is available as an option to facilitate calibration in the field. This accessory includes a single push-button and an LED, which are located beneath the yellow cap inside the conduit adapter or junction box. There is also a two-position switch, labeled Run and Cal. This switch must be in the Run position during normal operation and during calibration if the 4 to 20mA output of the 2750 is being utilized. If the S 3 L output is being used, and if the operator chooses to perform system calibrations using the Local EasyCal accessory, then the switch must be placed in the Cal position during calibration only. Return the switch to the Run position once calibration is complete. SW1 Run SW2 Cal SW1 SW2 D1 Local EasyCal button RUN/CALIBRATE switch for S 3 L system Local EasyCal Indicator (green LED) + BLK RED D1 Calibration Tips: As electrodes age their reaction time becomes slower. And if there is no local display of the measured values to help the operator determine whether an electrode s output is settled enough for calibration, it may be prudent to allow extra time before and between the steps of the procedure below. For example, changing temperatures can have a significant effect on electrode output, so it is recommended to perform calibration at room temperature. Therefore, the operator should allow enough time for an electrode coming out of a hot or cold process to achieve room temperature prior to calibration. This could take several minutes, depending on temperature, and should be done in tap water. Some suitable cleaning agents such as 5% HCl or even dishwashing detergent can temporarily shock an electrode. It is good practice to rinse an electrode thoroughly after each cleaning and then allow it to soak in a calibration solution or tap water for several minutes prior to calibration. The electrical interconnection between electrode and sensor electronics must remain dry and clean at all times. Local EasyCal Calibration Procedure: The first step (Reset) is recommended each time an electrode is replaced, but is not necessary upon initial installation or periodic calibration. In fact, for periodic calibration it is best for the electrode/sensor assembly to remain intact to minimize the possibility of moisture or other contamination from entering the electrical interconnection area. 1. Reset the 2750 ph/orp Sensor to factory calibration. With no electrode connected to the Sensor, press and hold the button until the LED comes on steady then goes off again (approx. 10 seconds). When the LED goes off, release the button; reset is complete. 2. Connect an electrode to the 2750 ph/orp Sensor. 3. Place the electrode/sensor assembly into a calibration solution as follows: (If the electrode is healthy, then the 2750 will automatically recognize the solution, so the order in which the operator selects solutions for calibration is of no consequence.) For ph calibration, use any two of these international standards: ph 4.0, 7.0 or 10.0 buffer solutions. To produce standards for ORP calibration, mix the chemical Quinhydrone into ph 7.0 and 4.0 buffers to saturation (1/8g per 50ml). Regardless of the size of the container used for calibration, one inch of solution is adequate to completely submerge the tip of the electrode. Allow at least 30 seconds for the electrode response to stabilize. 4. Press and hold the button for approximately 8 to 10 seconds. During this time, the LED will come on steady then go back off. (If the LED blinks several times rapidly, the calibration was not successful. See the troubleshooting section.) 5. Remove the electrode/sensor assembly from the first calibration solution, rinse the electrode with tap water or DI water, and place it in a second solution. Allow at least 30 seconds for the electrode response to stabilize. 6. Press and hold the button for approximately 8 to 10 seconds. During this time, the LED will come on steady then go back off. (If the LED blinks several times rapidly, the calibration was not successful. See the troubleshooting section.) 7. Calibration is complete; return the unit to service. page 7 of 8

8 Troubleshooting LED and Output Condition Possible Causes Suggested Solutions Current Out: LED off, current output is 3.6 ma S3L: Temp out is No electrode installed Bad/dirty contacts between electrode and 2750 Install electrode Check interconnection between electrode and 2750, clean contacts During Local EasyCal, the LED blinks rapidly for 4 seconds. The current output is frozen at some value. The buffer solution is outside of the accepted tolerance for the 2750 The electrode is depleted (> 1.1 ph or 65 mv offset) ph system: Use fresh 4 ph, 7 ph, or 10 ph buffer and restart the calibration. ORP system: Use fresh 4pH and 7 ph buffer solution saturated with quinhydrone. Replace the electrode. Ordering Information Mfr. Part No. Code Description In-line Sensor Electronics w/junction Box In-line Sensor Electronics w/jct. Box & Local EasyCal Submersible Sensor Electronics with 15 ft. cable, 3/4 in. NPT threads Submersible Sensor Electronics with 15 ft. cable, ISO 7-1/R 3/4 threads Flat ph Electrode HF Flat ph Electrode, HF Resistant Flat ORP Electrode Bulb ph Electrode DI Bulb ph Electrode, DI Resistant Bulb ORP Electrode Accessories Mfr. Part No. Code Description Universal Mount Junction Box Universal Mount Jct. Box w/local EasyCal /4 NPT Mount Junction Box /4 NPT Mount Jct. Box w/local EasyCal ph/orp Simulator/System tester Adapter cable for use with Adapter cable for use with ph Buffer Kit P P Universal Pipe Adapter PVC P C Universal Pipe Adapter CPVC P V Universal Pipe Adapter PVDF Cable, 3-conductor shielded 22AWG (per ft) SIGNET Signet Scientific Company, 3401 Aerojet Avenue, El Monte, CA U.S.A. Tel. (626) Fax (626) For Worldwide Sales and Service, visit our website: Or call (in the U.S.): (800) GEORGE FISCHER +GF+ Piping Systems /(B-8/01) English Scientific Company 2001 Printed in U.S.A. on Recycled Paper page 8 of 8

SIGNET 2760 DryLoc ph/orp Preamplifier

SIGNET 2760 DryLoc ph/orp Preamplifier SIGNET 2760 DryLoc ph/orp Preamplifier 3-2760.090 3-2760.090 (A-3/04) English SAFETY INSTRUCTIONS 1. Depressurize and vent system prior to installation or removal. 2. Confirm chemical compatibility before

More information

Signet 2760 DryLoc ph/orp Preamplifier

Signet 2760 DryLoc ph/orp Preamplifier Signet 2760 DryLoc ph/orp Preamplifier *3-2760.090* 3-2760.090 Rev. H 05/11 English English SAFETY INSTRUCTIONS 1. Depressurize and vent system prior to installation or removal. 2. Confi rm chemical compatibility

More information

* * Rev. 9 01/18 Product Manual

* * Rev. 9 01/18 Product Manual English Signet 2760 DryLoc ph/orp Preamplifier *3-2760.090* 3-2760.090 Rev. 9 01/18 Product Manual Description The Signet 2760 ph/orp Preamplifiers with DryLoc connectors are part of Signet's growing line

More information

Signet ph/orp Electrodes

Signet ph/orp Electrodes Signet 2724-2726 ph/orp Electrodes Compatible with ALL Signet ph/orp Instruments Features Flat Glass Protected Bulb Patented reference design for exceptional performance * Mounts in Signet standard fittings

More information

Signet ph/orp Electrodes

Signet ph/orp Electrodes Signet 2724-2726 ph/orp Electrodes Compatible with ALL Signet ph/orp Instruments Features Patented reference design for exceptional performance * Mounts in Signet standard fittings from DN15 to DN100 (½

More information

Signet ph/orp Electrodes

Signet ph/orp Electrodes Signet 2734-2736 ph/orp Electrodes Compatible with Signet 8900/9900 Instruments Features Patented reference design for exceptional performance * Mounts in Signet standard fittings from DN15 to DN100 (½

More information

Signet 2760 DryLoc ph/orp Preamplifiers & Connectors

Signet 2760 DryLoc ph/orp Preamplifiers & Connectors Signet 2760 DryLoc ph/orp Preamplifiers & Connectors Features In-line integral mount and submersible installation versions Automatic temperature compensation Auto configuration for ph or ORP operation

More information

Signet 3719 ph/orp Wet-Tap

Signet 3719 ph/orp Wet-Tap Signet 3719 ph/orp Wet-Tap *3-3719.090* 3-3719.090 Rev H 09/07 English Do not exceed maximum temperature/pressure specifications. Wear safety goggles or face shield during installation/service. Do not

More information

SIGNET 3719 ph/orp Wet-Tap

SIGNET 3719 ph/orp Wet-Tap ENGLISH 3-3719.090 B-6/02 English Do not exceed maximum temperature/pressure specifications. Wear safety goggles or face shield during installation/service. Do not attempt to disassemble the retraction

More information

Signet DryLoc ph/orp Electrodes

Signet DryLoc ph/orp Electrodes Signet 2774-2777 DryLoc ph/orp Electrodes Features Patented DryLoc connector with gold plated contacts* Special design allows for installation at any angle, even inverted or horizontal Quick temperature

More information

Signet DryLoc ph/orp Electrodes

Signet DryLoc ph/orp Electrodes Signet 2774-2777 DryLoc ph/orp Electrodes Features Patented DryLoc connector with gold plated contacts* Special design allows for installation at any angle, even inverted or horizontal Quick temperature

More information

Signet 2850 Conductivity/Resistivity Sensor Electronics and Integral Systems with PVDF Sensor

Signet 2850 Conductivity/Resistivity Sensor Electronics and Integral Systems with PVDF Sensor Signet 2850 Conductivity/Resistivity Sensor Electronics and Integral Systems with PVDF Sensor Features Test certificate supplied with all sensors Universal Mount Custom cell constant programmed into the

More information

* * Signet 2450 Pressure Sensor. Product Manual. English. Description. Dimensions Rev /18

* * Signet 2450 Pressure Sensor. Product Manual. English. Description. Dimensions Rev /18 English Signet 2450 Pressure Sensor *32450.0901* 32450.0901 Rev. 13 03/18 Product Manual Description The 2450 Pressure Sensor has a onepiece injection molded PVDF body and ceramic diaphragm for superior

More information

Signet ph/orp Transmitter

Signet ph/orp Transmitter Signet 87501 ph/orp English *38750.0901* 38750.0901 Rev. F 3/06 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury.

More information

Signet 3719 ph/orp Wet-Tap Assembly

Signet 3719 ph/orp Wet-Tap Assembly Signet 3719 ph/p Wet-Tap Assembly Features Electrode removal without process shutdown Space saving 45 mm (1.75 in.) short-stroke design Sealed pneumatic dampening for smooth and safe operation SafeLoc

More information

SIGNET Temperature Transmitter

SIGNET Temperature Transmitter SIGNET 83501 Temperature *38350.0901* English 38350.0901 Rev. G 3/06 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury.

More information

SIGNET 7002 Vortex Flow Sensor ENGLISH

SIGNET 7002 Vortex Flow Sensor ENGLISH SIGNET 7002 Vortex Flow Sensor ENGLISH 3-7002.090 C-7/02 English SAFETY INSTRUCTIONS 1. Do not remove from pressurized lines. 2. Never install sensor without O-Rings. 3. Confirm chemical compatibility

More information

Signet Conductivity/Resistivity Electrodes

Signet Conductivity/Resistivity Electrodes Signet 2818-2823 Conductivity/Resistivity Electrodes 2818 2819 2820 2821 2822 2823 Sanitary Features Standard process connections - Polypro - SS on 10 and 20 cell - Tri-clamp 1-1½ in., 2 in. - Opt. ½ in.

More information

Signet Temperature Transmitter

Signet Temperature Transmitter Signet 80 Temperature English 80.090 80.090 Rev. G /0 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents.

More information

Chapter Page. 1 Introduction Description 1

Chapter Page. 1 Introduction Description 1 Chapter Page 1 Introduction 1 1.1 Description 1 Table of Contents 2 Calibration 3 2.1 Wet Calibration 3 2.2 Calibration Procedure 4 2.3 Installing Optional Display Board 5 2.4 Display Board Range Adjustment

More information

Signet Pressure Transmitter

Signet Pressure Transmitter Signet 8450 Pressure English 8450.090 8450.090 Rev. G /06 English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents

More information

Signet Conductivity/Resistivity Electrodes

Signet Conductivity/Resistivity Electrodes Signet 2818-2823 Conductivity/Resistivity Electrodes 2818 2819 2820 2821 2822 2823 Sanitary Features Standard process connections - Polypro - SS on 10 and 20 cell - Tri-clamp 1-1½ in., 2 - Opt. ½ in. NPT

More information

Signet 2536 Rotor-X Paddlewheel Flow Sensors

Signet 2536 Rotor-X Paddlewheel Flow Sensors Signet 2536 Rotor-X Paddlewheel Flow s Features Operating range 0.1 to 6 m/s (0.3 to 20 ft/s) Wide turndown ratio of 66:1 Open-collector output Highly repeatable output Simple, economical design PVC (gray

More information

SIGNET Vortex Flow Sensors ENGLISH

SIGNET Vortex Flow Sensors ENGLISH DN X DN X DN X DN X DN X DN X DN X SIGNET Vortex Flow Sensors ENGLISH 37000.090 C12/00 English 1. Mounting Location... 1 2. Installation... 1 3. Wiring Preparation... 2 4. Wiring Options Frequency Output

More information

Signet 4150 Turbidimeter

Signet 4150 Turbidimeter C L I S T E D 113783 US 4150 Turbidimeter Features Description The 4150 Turbidimeter system provides accurate and reliable compliant water quality monitoring for municipal and industrial applications.

More information

Signet 515 Rotor-X Paddlewheel Flow Sensors

Signet 515 Rotor-X Paddlewheel Flow Sensors Signet 515 Rotor-X Paddlewheel Flow Sensors Features Operating range 0.3 to 6 m/s (1 to 20 ft/s) Wide turndown ratio of 20:1 Highly repeatable output Simple, economical design Installs into pipe sizes

More information

Signet 2517 Brass Paddlewheel Flow Sensor * *

Signet 2517 Brass Paddlewheel Flow Sensor * * Signet 2517 Brass Paddlewheel Flow Sensor -250.090 English -2517.090 Rev. F 11/06 English SAFETY INSTRUCTIONS 1. Do not remove from pressurized lines. 2. Do not exceed maximum temperature/pressure specifi

More information

Pipe fittings MUST be installed by a certified welder only. +GF+ SIGNET will not assume liability of any kind for improper fitting installations.

Pipe fittings MUST be installed by a certified welder only. +GF+ SIGNET will not assume liability of any kind for improper fitting installations. +GF+ SIGNET 0 High Performance Flow Sensor (English) -0.090- F-/99 English SAFETY INSTRUCTIONS. Do not remove from pressurized lines.. Do not exceed maximum temperature/pressure specifications.. Do not

More information

Signet V(D) to V(D) PVDF Conductivity Electrodes

Signet V(D) to V(D) PVDF Conductivity Electrodes Signet 2839-1V(D) to 2842-1V(D) PVDF Conductivity Electrodes Features ± 2% accuracy - Custom calibration certificate provided Dual-threaded Compact electrode length for easy in-line installation in small

More information

Signet 515 Rotor-X Paddlewheel Flow Sensors

Signet 515 Rotor-X Paddlewheel Flow Sensors Signet 515 Rotor-X Paddlewheel Flow Sensors Features Operating range 0.3 to 6 m/s (1 to 20 ft/s) Wide turndown ratio of 20:1 Highly repeatable output Simple, economical design Installs into pipe sizes

More information

Signet 7000/7001 Vortex Flow Sensors * *

Signet 7000/7001 Vortex Flow Sensors * * DN X DN X DN X DN X DN X DN X DN X Signet 7000/7001 Vortex Flow Sensors *3-7000.090* 3-7000.090 Rev. F 10/05 English English 1. Mounting Location... 1 2. Installation... 1 3. Wiring Preparation... 2 4.

More information

+GF+ SIGNET Conductivity Transmitter. Instruction Manual GEORGE FISCHER +GF+

+GF+ SIGNET Conductivity Transmitter. Instruction Manual GEORGE FISCHER +GF+ +GF+ SIGNET 3-8800 Conductivity Transmitter Instruction Manual GEORGE FISCHER +GF+ SAFETY INSTRUCTIONS 1. Do not remove from pressurized lines. 2. Do not exceed maximum temperature/pressure specifications.

More information

Pipe fi ttings MUST be installed by a certifi ed welder only. Signet will not assume liability of any kind for improper fi tting installations.

Pipe fi ttings MUST be installed by a certifi ed welder only. Signet will not assume liability of any kind for improper fi tting installations. Signet 50 High Performance Flow Sensor English *3-50.090* 3-50.090 C 05/06 English SAFETY INSTRUCTIONS 1. Do not remove from pressurized lines.. Do not exceed maximum temperature/pressure specifi cations.

More information

Signet 2551 Magmeter Flow Sensor

Signet 2551 Magmeter Flow Sensor Signet 2551 Magmeter Flow Sensor Available in a variety of wetted materials and ideal for pipe sizes up to DN900 (36 in.) Features Test certificate included for -X0, -X1 Patented Magmeter technology No

More information

* * Rev. F 03/16 Operating Instructions

* * Rev. F 03/16 Operating Instructions English Signet 2552 Metal Magmeter *3-2552.090* 3-2552.090 Rev. F 03/16 Operating Instructions 3-2552-2 3-2552-3 Description The Signet 2552 Metal Magmeter from Georg Fischer features all-stainless steel

More information

Signet 2552 Metal Magmeter

Signet 2552 Metal Magmeter Signet 2552 Metal Magmeter *3-2552.090* 3-2552.090 Rev D 06/08 English English SAFETY INSTRUCTIONS 1. Depressurize and vent systems without Hot-tap valve prior to installation or removal. 2. Confi rm chemical

More information

OPERATING INSTRUCTION MANUAL

OPERATING INSTRUCTION MANUAL OPERATING INSTRUCTION MANUAL MODEL P/R65 DIFFERENTIAL 2-WIRE (4-20 ma) ph & ORP PROBES N116-139 REV. 0 Aquametrix, by Water Analytics Inc. 100 School Street Andover, MA 01810 Tel: (978) 749-9949 Toll Free:

More information

Signet 2552 Metal Magmeter

Signet 2552 Metal Magmeter Signet 2552 Metal Magmeter *3-2552.090* 3-2552.090 Rev E 07/10 English English SAFETY INSTRUCTIONS 1. Depressurize and vent systems without Hot-tap valve prior to installation or removal. 2. Confirm chemical

More information

NEW QUICK CABLE-TO-SENSOR RELEASE,

NEW QUICK CABLE-TO-SENSOR RELEASE, Models 399 and 399VP Product Data Sheet PDS 71-399 October 2000 ph/orp Sensors NEW OPTIONS to meet a variety of application requirements, including: * Low ionic strength water * Processes with hydrofluoric

More information

* * Rev M 10/14 Operating Instructions

* * Rev M 10/14 Operating Instructions English Signet 55 Blind Magmeter *-55.090* -55.090 Rev M 0/ Operating Instructions Description The Signet 55 Magmeter is an insertion-style magnetic flow sensor. The patented sensor design is available

More information

Signet Pressure Transmitter

Signet Pressure Transmitter Signet 80 Pressure 80.090 80.090 Rev. J 0/ English CAUTION! Remove power to unit before wiring input and output connections. Follow instructions carefully to avoid personal injury. Contents. Installation.

More information

General Purpose ph/orp Sensors

General Purpose ph/orp Sensors General Purpose ph/orp Sensors Product Data Sheet LIQ_PDS_3900 September 2013 General Purpose ph/orp Sensors SMART preamplifier-stores calibration and other data. Extended sensor life provided by double

More information

OPERATING INSTRUCTION MANUAL MS SERIES CONDUCTIVITY CELLS

OPERATING INSTRUCTION MANUAL MS SERIES CONDUCTIVITY CELLS OPERATING INSTRUCTION MANUAL MS SERIES CONDUCTIVITY CELLS REV. 3.1 Water Analytics, Inc. 100 School Street Andover, MA 01810 978-749-9949 Toll free: 855-747-7623 Fax: 978-749-9961 www.wateranalytics.net

More information

ph/orp Sensor Model 399 ph/orp

ph/orp Sensor Model 399 ph/orp Model 399 ph/orp Product Data Sheet PDS 71-399 September 1999 ph/orp Sensor DOUBLE JUNCTION REFERENCE CELL provides longer sensor life in process solutions containing poisoning ions. ONE-PIECE CONSTRUCTION

More information

Pipe fi ttings MUST be installed by a certifi ed welder only. Signet will not assume liability of any kind for improper fi tting installations.

Pipe fi ttings MUST be installed by a certifi ed welder only. Signet will not assume liability of any kind for improper fi tting installations. Signet 50 High Performance Flow Sensor English *3-50.090* 3-50.090 Rev. F 11/1 English For the most up-to-date information, please refer to our website at www.gfsignet.com SAFETY INSTRUCTIONS 1. Do not

More information

Measurement and Instrumentation

Measurement and Instrumentation 15 Measurement and Instrumentation Use the postcard included in this handbook or call (800) 854-4090 to order the complete Instrumentation Catalog or visit www.gfsignet.com for detailed product information.

More information

Encapsulated Differential ph and ORP Sensors (available in LCP, foreground, or Ryton body materials)

Encapsulated Differential ph and ORP Sensors (available in LCP, foreground, or Ryton body materials) Technology For Solutions Data Sheet LRE/102 Supersedes LRE/400 Encapsulated Differential ph and ORP Sensors (available in LCP, foreground, or Ryton body materials) Common Features Differential Electrode

More information

Models 396/396P/396R. Retraction/Submersion/Insertion ph/orp Sensor

Models 396/396P/396R. Retraction/Submersion/Insertion ph/orp Sensor Models 396/396P/396R Product Data Sheet PDS 71-TUpH September 1999 Retraction/Submersion/Insertion ph/orp Sensor MINIMUM SENSOR MAINTENANCE due to patented TUpH reference technology 1 which ensures steady

More information

User s Guide Shop online at omega.com SM For latest product manuals: CDE-285X SERIES Conductivity Sensors

User s Guide Shop online at omega.com SM   For latest product manuals:   CDE-285X SERIES Conductivity Sensors User s Guide Shop online at omega.com SM e-mail: info@omega.com For latest product manuals: www.omegamanual.info CDE-285X SERIES Conductivity Sensors omega.com info@omega.com U.S.A. Headquarters: Servicing

More information

Signet 515/2536 Rotor-X Flow Sensors * * Operating Instructions. English. Description Rev. X 05/17

Signet 515/2536 Rotor-X Flow Sensors * * Operating Instructions. English. Description Rev. X 05/17 English Signet 515/2536 Rotor-X Flow Sensors *3-0515.090* 3-0515.090 Rev. X 05/17 Operating Instructions 515 Sensor (with red cap) 2536 Sensor (with blue cap) 8510 Integral Sensor 8512 Integral Sensor

More information

General Purpose ph/orp Sensors

General Purpose ph/orp Sensors Product Data Sheet PDS 71-3900/rev. B December 2010 Models 3900 and 3900VP General Purpose ph/orp Sensors EXTENDED SENSOR LIFE provided by double junction reference. ENHANCED PERFORMANCE AND INCREASED

More information

Table of Contents 1. OVERVIEW OPERATION FUNCTION SPECIFICATIONS INSTRUMENT LAYOUT INSTALLATION...

Table of Contents 1. OVERVIEW OPERATION FUNCTION SPECIFICATIONS INSTRUMENT LAYOUT INSTALLATION... Table of Contents 1. OVERVIEW... 2 2. OPERATION FUNCTION... 2 3. SPECIFICATIONS... 3 4. INSTRUM LAYOUT... 4 4.1 OPERATION KEY...4 4.2 INDICATORS...4 5. INSTALLATION... 5 5.1. ENVIRONM CONDITIONS...5 5.2

More information

Combination ph/redox (ORP) Sensors

Combination ph/redox (ORP) Sensors Data Sheet Combination ph/redox (ORP) s Series AP00 User-customizable for Insertion, Flow-through, Immersion, Hot-tap Retractor ideal for Original Equipment Manufacturers and general applications Economical

More information

Combination ph/redox (ORP) Sensors

Combination ph/redox (ORP) Sensors Data Sheet SS/AP00_ Combination ph/redox (ORP) s AP00 Series User-customizable for Insertion, Flow-through, Immersion and Hot-tap Retractor ideal for OEMs (Original Equipment Manufacturers) and general

More information

Digital ph transmitter

Digital ph transmitter Digital ph transmitter Compact, remote versions for DN 15 to 200 Fully programmable ph transmitter functions for all kinds of ph-measurement tasks Large range of process connections with various fittings

More information

AP300 Series Combination ph/redox (ORP) Sensors. Measurement made easy. A reliable economic solution for a wide range of applications

AP300 Series Combination ph/redox (ORP) Sensors. Measurement made easy. A reliable economic solution for a wide range of applications Data sheet DS/AP00 EN Rev. F AP00 Series Measurement made easy A reliable economic solution for a wide range of applications Long life large PTFE junction provides resistance to fouling double junction

More information

* * Rev /18 Operating Instructions

* * Rev /18 Operating Instructions English Signet 255 Blind Magmeter *-2552.090* -255.090 Rev. 6 07/8 Operating Instructions Description The Signet 255 Magmeter is an insertion-style magnetic flow sensor. The patented sensor design is available

More information

Signet 2552 Magmeter. English * * Specifications. Safe to operate Hot-tap

Signet 2552 Magmeter. English * * Specifications. Safe to operate Hot-tap Signet 2552 Magmeter *3-2552.090* 3-2552.090 Rev. A 11/06 English SAFETY INSTRUCTIONS 1. Depressurize and vent systems without Hot-tap valve prior to installation or removal. 2. Confi rm chemical compatibility

More information

2100 Turbine Flow Sensor

2100 Turbine Flow Sensor 200 Turbine Flow Sensor Description Engineered specifically for small pipe diameter applications, the Signet 200 Turbine Flow Sensor provides accurate readings in two flow ranges: 0.3 to 3.8 lpm and 3

More information

Digital ph transmitter

Digital ph transmitter Digital ph transmitter Compact, remote versions for DN 15 to 200 Fully programmable ph transmitter functions for all kinds of ph-measurement tasks Large range of process connections with various fittings

More information

Hydro-Flow Model 2300 Installation and Operation Guide B

Hydro-Flow Model 2300 Installation and Operation Guide B Hydro-Flow Model 2300 Installation and Operation Guide 5700508-B Contents: Understanding and Identifying Your Flow Meter 1 Installing Your Flow Meter 3 Electrical Connections.. 7 Mechanical and Electrical

More information

User s Guide CDCE-90-X SERIES. Conductivity Sensors. Shop online at omega.com SM

User s Guide CDCE-90-X SERIES. Conductivity Sensors. Shop online at omega.com SM User s Guide Shop online at omega.com SM e-mail: info@omega.com For latest product manuals: www.omegamanual.info CDCE-90-X SERIES Conductivity Sensors omega.com info@omega.com U.S.A. Headquarters: Servicing

More information

Hydro-Flow 2300 Fixed Insertion Vortex Flow Meter

Hydro-Flow 2300 Fixed Insertion Vortex Flow Meter Hydro-Flow 2300 Fixed Insertion Vortex Flow Meter Description The Hydro-Flow Model 2300 insertion vortex flow meter was designed specifically for water flow measurement, containing no moving parts like

More information

8000 Series ph/orp Electrode Platform Product Instructions

8000 Series ph/orp Electrode Platform Product Instructions 8000 Series ph/orp Electrode Platform Product Instructions Introduction Congratulations on your purchase of your Sensorex S8000 series electrode platform. The platform is designed in a modular fashion

More information

Analytical Instruments Two-Electrode Conductivity Sensors

Analytical Instruments Two-Electrode Conductivity Sensors Data Sheet Analytical Instruments Two-Electrode Conductivity Sensors Model TB2 Superior accuracy in low conductivity measurement ranges. Measurement ranges from 0 to 199.9 µs/cm to 0 to 19.99 ms/cm. Resolution

More information

OPERATING INSTRUCTION MANUAL

OPERATING INSTRUCTION MANUAL OPERATING INSTRUCTION MANUAL High Purity Water ph Sensor System Worldwide Sales: GLI International, Inc. 9020 West Dean Road Milwaukee, Wisconsin 53224, U.S.A. phone: [414] 355-3601 fax: [414] 355-8346

More information

+GF+ SIGNET 2550 Insertion Magmeter

+GF+ SIGNET 2550 Insertion Magmeter +GF+ SIGNET 2550 Insertion Magmeter Instruction Manual GEORGE FISCHER +GF+ Important Safety Information! CAUTION: Do not remove from pressurized lines! Do not remove safety ring from sensor. Unit may become

More information

Hydro-Flow 2200 Fixed Insertion Vortex Flow Meter

Hydro-Flow 2200 Fixed Insertion Vortex Flow Meter A trademark of Spirax Sarco, Inc. Hydro-Flow 2200 Fixed Insertion Vortex Flow Meter Description The Hydro-Flow Model 2200 fixed insertion vortex flow meter was designed specifically for water flow measurement,

More information

User s Guide PH-2720-PA. Preamplifier. Shop online at.

User s Guide PH-2720-PA. Preamplifier. Shop online at. User s Guide Shop online at www.omega.com e-mail: info@omega.com Preamplifier OMEGAnet Online Service www.omega.com Internet e-mail info@omega.com Servicing North America: USA: One Omega Drive, P.O. Box

More information

AD592/592-10K Non-Room Sensors in BAPI-Box, BAPI-Box 2 or BAPI-Box 4 BA/592-x-(BB, BB2, BB4) Temperature Sensor Instructions & Operation

AD592/592-10K Non-Room Sensors in BAPI-Box, BAPI-Box 2 or BAPI-Box 4 BA/592-x-(BB, BB2, BB4) Temperature Sensor Instructions & Operation 592 Theory of Operation The 592 sensor is a two terminal integrated-circuit temperature sensor. The 592 provides an output current directly proportional to absolute temperature. When the temperature of

More information

Conductivity meter. PLC Solenoid valve. Type Type On/Off Diaphragm valve. Valve islands

Conductivity meter. PLC Solenoid valve. Type Type On/Off Diaphragm valve. Valve islands Conductivity meter Configurable outputs: two transistor and single or dual analog 4... 20 ma Removable backlighted display Universal process connection Type can be combined with... Three cell constants

More information

For additional information, please visit our website at SENSOR SPECIFICATIONS

For additional information, please visit our website at  SENSOR SPECIFICATIONS Instruction Sheet PN 51A-140/rev.E December 2010 Models 140, 141, and 142 Conductivity Sensors For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. SENSOR/PROCESS

More information

HI 8510-HI 8512 HI 8710-HI 8711 HI 8720-HI HI HI

HI 8510-HI 8512 HI 8710-HI 8711 HI 8720-HI HI HI Instruction Manual HI 8510-HI 8512 HI 8710-HI 8711 HI 8720-HI 931500 HI931501-HI 932500 Panel - Mounted ORP - ph Indicators and Controllers ORP HI 8720E mv DOSAGE SLOPE OXID. REDUC. MEA SURE AL COARSE

More information

Instrument Enclosure Assemblies

Instrument Enclosure Assemblies Instrument Enclosure Assemblies Standard instrument enclosure Instrument enclosures offer a simple solution for the customer who doesn t have the time to install Signet meters into panels. Whether it s

More information

Bante220 Portable ph Meter Instruction Manual

Bante220 Portable ph Meter Instruction Manual Bante220 Portable ph Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante220 Portable ph Meter 1 Introduction Thank you for selecting the Bante220 portable ph meter. This manual provides a step-by-step

More information

Figure 1: WPR2 Dimensions and Hardware

Figure 1: WPR2 Dimensions and Hardware Installation and Operation Instructions WPR2 Series (Max. Line Pressure < 300 PSI) Wet to Wet Differential Pressure Ø0.200" 2.60" 4.47" 5.47" 5.28" 3.00" Precautions Figure 1: WPR2 Dimensions and Hardware

More information

CONTROLLERS & SPECIAL VALVES

CONTROLLERS & SPECIAL VALVES CONTROLLERS & SPECIAL VALVES G75-C 1, 2 & 4 Valve Controllers.... 70 G75-C-1W & 1W+ Window Controller... 72 G75-DP Differential Pressure Switch.....................................................................

More information

Hydro-Flow Model 2200 Installation and Operation Guide B

Hydro-Flow Model 2200 Installation and Operation Guide B Hydro-Flow Model 2200 Installation and Operation Guide 5700507-B Contents: Understanding and Identifying Your Flow Meter 1 Installing Your Flow Meter 3 Making Electrical Connections. 7 Mechanical and Electrical

More information

Product Specifications

Product Specifications PE 100 Metric Piping System Specifications 1. Scope This specification covers requirements for the +GF+ Georg Fischer (later named only +GF+) PE 100 Piping System intended for a wide range of applications

More information

Automation Made Easy 2013 Signet Measurement and Control Product Catalog

Automation Made Easy 2013 Signet Measurement and Control Product Catalog Multi-Parameter Chlorine Dissolved Oxygen Turbidity Flow ph/orp Conductivity/Resistivity Temperature Pressure Level Automation Made Easy 2013 Signet Measurement and Control Product Catalog Automation Made

More information

MS Application. Features. Proportional Valve Actuator General Instructions

MS Application. Features. Proportional Valve Actuator General Instructions Proportional Valve Actuator Application The Proportional Valve Actuator is a non-spring return actuator used with proportional to 10 Vdc or 4 to 0 ma controllers and standard 1/" to 1-1/4" two-way and

More information

ORPscan20 Pocket ORP Tester Instruction Manual

ORPscan20 Pocket ORP Tester Instruction Manual ORPscan20 Pocket ORP Tester Instruction Manual BANTE INSTRUMENTS CO., LTD ORPscan20 Pocket ORP Tester 1 Thank you for selecting the ORPscan20 pocket ORP tester. This manual provides a step-by-step guide

More information

General Specifications

General Specifications General Specifications Model PH20 and FU20 ph/redox and Temperature sensor The PH20 is nicknamed "Tempress" because of the patented compensation for changes in the process temperature and pressure. This

More information

Portable pressure calibrator, Wally Box III Model CPH7600

Portable pressure calibrator, Wally Box III Model CPH7600 Calibration technology Portable pressure calibrator, Wally Box III Model CPH7600 WIKA data sheet CT 17.01 Applications Calibration service companies and service industry Measurement and control laboratories

More information

ACTIVE ELECTRODELESS SENSORS Instruction Manual

ACTIVE ELECTRODELESS SENSORS Instruction Manual W A L C H E M An Iwaki America Company Active Electrodeless Sensors ACTIVE ELECTRODELESS SENSORS Instruction Manual Five Boynton Road Hopping Brook Park Holliston, MA 01746 USA TEL: 508-429-1110 FAX: 508-429-7433

More information

Installation Fittings

Installation Fittings PVC TEES SCH 80 - Fitting only Part No. Code Size L H i.d. PV8T005F 159 000 527 0.50 in. 3.75 3.6 0.85 PV8T007F 159 000 529 0.75 in. 3.75 3.8 1.06 PV8T010F 159 000 531 1.00 in. 4.26 4.0 1.33 PV8T012F 159

More information

Sanitary Sensors. High Quality, Precision. Sanitary Sensors. THORNTON Leading Pure Water Analytics

Sanitary Sensors. High Quality, Precision. Sanitary Sensors. THORNTON Leading Pure Water Analytics THORNTON Leading Pure Water Analytics Sanitary Sensors Sanitary Sensors Temperature Sensors 3A Sanitary Turbine Flow Sensors Pressure Sensors Tank Level Sensors High Quality, Precision Sanitary Sensors

More information

Measurement & Instrumentation 2005 Product Catalog on CD. Creating Quality by Connecting Technology to Life. Click on image to begin.

Measurement & Instrumentation 2005 Product Catalog on CD. Creating Quality by Connecting Technology to Life. Click on image to begin. Measurement & Instrumentation 2005 Product Catalog on CD Creating Quality by Connecting Technology to Life Click on image to begin. Main Contents Measurement & Instrumentation 2005 Product Catalog Flow

More information

Retractable ph/orp Sensors

Retractable ph/orp Sensors Model 396R and 396RVP Instruction Manual PN 5100396R January 2001 Retractable ph/orp Sensors ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests

More information

Visit us at

Visit us at Instruction Manual testo 205 ph/ Temperature Measuring Instrument 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com 2 General Information

More information

User s Guide FP-5060 SERIES. Micro-Flow Sensors. Shop online at omega.com SM For latest product manuals:

User s Guide FP-5060 SERIES. Micro-Flow Sensors. Shop online at omega.com SM   For latest product manuals: User s Guide Shop online at omega.com SM e-mail: info@omega.com For latest product manuals: www.omegamanual.info FP-5060 SERIES Micro-Flow Sensors omega.com info@omega.com U.S.A. Headquarters: Servicing

More information

PRESSURE TRANSDUCERS. Specifications. Individual Specifications

PRESSURE TRANSDUCERS. Specifications. Individual Specifications 00 Series and 00 Heavy Duty Series Compact OEM Pressure Transmitters 0 50 psi to 0-0,000 psi ranges (0-.5 bar to 0-,00 bar) High Proof Pressures Broad Choice of Outputs RoHS Compliant For OEMs that need

More information

7777D In-Line Mounting for Durafet II ph Electrodes Installation and Maintenance

7777D In-Line Mounting for Durafet II ph Electrodes Installation and Maintenance 7777D In-Line Mounting for Durafet II ph Electrodes Installation and Maintenance 70-82-25-121 Rev. 0 3/07 Honeywell Process Solutions Industrial Measurement and Control Copyright, Notices, and Trademarks

More information

OPERATING MANUAL. 3700E-series Electrodeless Conductivity Sensors. Represented By: Fax: [414] Web:

OPERATING MANUAL. 3700E-series Electrodeless Conductivity Sensors. Represented By: Fax: [414] Web: OPERATING MANUAL Model 3700E-series Electrodeless Conductivity Sensors Worldwide Headquarters and Sales: Represented By: GLI International, Inc. 9020 West Dean Road Milwaukee, Wisconsin 53224 U.S.A. Phone:

More information

General Purpose Conductivity Sensors ENDURANCE. Model 400 and 400VP Series. Product Data Sheet PDS /rev.K February 2013

General Purpose Conductivity Sensors ENDURANCE. Model 400 and 400VP Series. Product Data Sheet PDS /rev.K February 2013 Product Data Sheet PDS 71-400/rev.K February 2013 Model 400 and 400VP Series ENDURANCE General Purpose Conductivity Sensors INITIAL CALIBRATION NOT REQUIRED. Factory-measured cell constant ensures out-of-the-box

More information

Signet 8150 Flow Totalizer

Signet 8150 Flow Totalizer English Signet 8150 Flow Totalizer *3-8150.090-1* 3-8150.090-1 Rev. K 9/14 English Operating Instructions Description The Signet 8150 Flow Totalizer is a battery-powered instrument capable of providing

More information

RESISTIVITY MONITOR/CONTROLLERS

RESISTIVITY MONITOR/CONTROLLERS RESISTIVITY MONITOR/CONTROLLERS Installation Operation Maintenance User Manual for Models: 750, 752, 753, 762 2450 Impala Drive Carlsbad, CA 92010-7226 USA Tel: 1-760-438-2021 Fax: 1-800-869-7668 / 1-760-931-9189

More information

Inductive conductivity meter

Inductive conductivity meter Inductive conductivity meter Type 8228 can be combined with... Configurable outputs: up to 2 transistor and up to 2 analog 4... 20 ma outputs Removable backlighted display Simulation of process values

More information

Hydro-Flow Model 2200 Fixed Insertion Vortex Flowmeter Installation and Operation Manual

Hydro-Flow Model 2200 Fixed Insertion Vortex Flowmeter Installation and Operation Manual Hydro-Flow Model 00 Fixed Insertion Vortex Flowmeter Installation and Operation Manual Table of Contents Model and Suffix Codes. Theory and Identification.. Installation Guidelines 4 Mechanical Drawing:

More information

HI Portable Waterproof Skin ph Meter INSTRUCTION MANUAL

HI Portable Waterproof Skin ph Meter INSTRUCTION MANUAL HI99181 Portable Waterproof Skin ph Meter INSTRUCTION MANUAL Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using this meter. This

More information