1 Rock Compressors Model Numbers: Twin R22 Triple R42 Service and Parts Manual R42 Shown FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY SF-1862 Part No Rev. C ( /2013)

2 Table Of Contents General Information Section A GENERAL INFORMATION Symbols... iii Ordering Parts... iii Model / Serial Number Location... iii Weights, Dimensions, Electrical Specifications... iv Warranty Information... v OPERATION & TROUBLESHOOTING Compression Cycle...A-2 Purge Cycle... A-4 Section B Section C Section E Section Section E D FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY * Indicates multiple pages due to a serial number break for the parts illustration COMPONENT TESTING & REPLACEMENT Pressure Switch... B-2 On/Off Switch... B-7 Capacitor... B-10 Compressor Heads... B-13 Low Voltage - MCP... B-17 Intake Filter... B-19 Checking for Leaks... B-21 Coalescing Filter... B-22 Solenoid... B-24 Desiccant Tank... B-27 Fuse Replacement... B-29 Fan Replacement... B-31 Check Valve... B-36 ACCESS PROCEDURES Quiet Cover... C-2 WIRING DIAGRAMS R22... D-2 R42... D-3 (Motor & Pressure Switch) R42... D-4 (On & Off Switch) EXPLODED VIEWS / PARTS LISTS R22... E-2 R42... E-3

3 General Information Symbols Caution Indicates a potentially hazardous situation which could result in injury if not avoided. Equipment Alert Indicates a potentially hazardous situation which could result in equipment damage if not avoided. Note Amplifies a procedure, practice, or condition. Indicates that the component the check mark appears beside should be tested before replacing it. In Section A, test the components in the order indicated. (ex. 1st then, 2nd ) Refer to Section B for component testing procedures. These symbols are used throughout this manual to represent the operational status of table functions and components. Indicates the function / component is working properly. No action required. Indicates the function / component is working, but a problem exists. Indicates the function / component is not working at all. Ordering Parts General Information The following information is required when ordering parts: Serial number & model number Part number for desired part. [Refer to Exploded Views / Parts Lists section] Non-warranty parts orders may be faxed to Midmark using the Fax Order Form in the back of this manual. For warranty parts orders, call Midmark's Technical Service Department with the required information. Hours: 8:00 am until 5:00 pm EST [Monday - Friday] Phone: 1-(800)-Midmark Model / Serial Number Location Serial / Model Number iii

4 General Information Weights, Dimensions, Electrical Specifications Built-In Low Voltage Contactor Condensate Drain Canister (Provided) Fresh Air Intake Kit (Provided) 5 High Pressure Flexible Air Hose Terminated in 3/8 MNPT (Provided) Leveling Feet AA143400i 32 in. (81.3 cm) 32 in. (81.3 cm) Model Users No. Breaker 80PSI db-a Volts HP Total Amps Tank Size Width Depth Height Weight Heads (Gal.) (In.) (In.) (In.) (lbs.) R R iv

5 General Information Warranty All Apollo products are thoroughly inspected and tested in accordance with rigid specifications and standards. Our products are guaranteed against any defective material and workmanship from the date of shipment; provided, that the installation, operation, and maintenance is done in accordance with Apollo procedures as outlined in our Installation and Maintenance Guides. No other warranties or guarantees, expressed or implied are made. Midmark s obligation under the warranty is to provide parts to the dealer for repairs or, at its option, to provide the replacement product (excluding labor and shipping charges). All special, incidental and/or consequential damages are excluded. We will not issue credit for product without first attempting to correct the problem in the field. Written notice of breach of warranty must be given to Midmark within the warranty period. The warranty does not cover damage resulting from improper installation or maintenance, accident or misuse. The warranty does not cover damage resulting from the use of cleaning, disinfecting or sterilizing chemicals and processes. The warranty does not cover vacuum failures due to hard water deposits. Failure to follow instructions provided in Midmark s Installation and Maintenance Guides will void the warranty. * All warranty repairs must be repaired by authorized dealers and/or service dealers only - no exceptions. * Labor charges are NOT included. * Warranty period starts from the installation date. * Shipping damage or damage caused by equipment handlers and installers is not covered. Warranty Periods from Date of Install RevolutionTM Scroll Oil-Less Air Compressor 3 Years UltraVac Vacuum 3 Years Classic BronzeTM Vacuum 2 Years Gold Series Vacuum 2 Years Dry Vacuum (DryVacTM) 2 Years The Rock Oil-less Air Compressors 2 Years Oil-less Air Compressors 2 Years Lubricated Air Compressors 2 Years Replacement Parts and Accessories 90 Days** ** Replacement parts and repairs are covered under the remaining warranty of the unit they are replacing, or 90 days from the date of shipment, which ever is the longest. Customer Service/Technical Support Department hours are from 8:00 AM through 5:00 PM EST. Midmark Customer Service/Technical Support is available for order entry and technical support by phone MIDMARK or (937) or FAX (877) , Monday through Friday. We are closed on major holidays. Additional product documentation and specifications are available at your request or at our website at v

6 General Information vi

7 Section A Operation & Troubleshooting Operation & Troubleshooting Function / System Compression Cycle... A-2 Purge Cycle... A-4 A-1

8 Operation & Troubleshooting Intake Air PreCooler EQUIPMENT ALERT Room temperature should be between 40 and 100 F (4 C to 38 C). Lack of proper ventilation is the main reason for compressors failures. Whenever possible, use exhaust fan to bring cool air from dental office to the equipment room, or run fresh air intake lines to the outside so the compressor can breathe in fresh, outside air. Exhaust the stale, hot air out through the roof or side wall. Intake Filter Compressor (Running) White Red Purge Tank (Pressurizing) Atmospheric Air Pressurized Air Troubleshooting Compressor will not Start... A-6 Motor runs Intermittently or "Chugs"... A-7 Compressor Heads run but will not Pressurize to 100 PSI... A-8 Cycles with no Air in use... A-9 Moisture Indicator is Pink... A-10 Element Indicator is Red... A-11 High Pressure... A-12 Excessive Noise... A-13 Overheating... A-14 Compressor runs to Frequently... A-15 Coalescing Filter Dessicant Tank Check Valve (Open) Storage Air Moisture Indicator Pressure Relief Valve (Closed) (Opens at 125 PSI) Compression Cycle Atmospheric air is drawn in through intake filters to the compressor head(s) where air is compressed and exhausted. Bottom Port (Closed) Normally Open Solenoid Valve (Closed) Pressure Switch (Contacts Closed) Compressor Head Unloader Valve (Closed) Manual Needle Purge Valve (Closed) Ball Valve (Open) 100 PSI Supply Air to Operatory Air travels through the pre-cooler where it is cooled 10 to 15 F (-12.2 to -9.4 Celsius) within ambient room temperature. Cool air enters the coalescent filter, which removes % of all particulate matter. As the clean air travels through the desiccant tank it is dried to -25 F (-31.7 Celsius) dew point. Condensate Drain Can Assembly Storage Tank (Pressurizing) The clean dry air flows into the main storage tank through a check valve. Some air is directed to the purge tank. As both tanks fill, the pressure switch opens at preset pressure and stops the compressor. AA150401i A-2 Compression Cycle Rev 7/23/08 All

9 Operation & Troubleshooting Compression Cycle Electrical Schematic Supply Line Voltage is always present on one side of the Contactor contacts. Line Voltage is also being supplied to the 240 / 24 VAC transformer energizing it. The Transformer supplies 24 VAC thru the fuse to the On/Off switch on the wall and to one side of the 24 VAC Contactor Coil. When the On/Off switch is turned to the on position the Contactor Coil is energized closing the contacts. With the contacts closed, line voltage is supplied to the Fan Motor energizing it. At the same time, line voltage is supplied to one side on the Vac Unit On/Off switch. Line VAC 24 VAC Transformer ENERGIZED 24 VAC Fuse On Unit Wall On / Off Switch ON Fan Motor RUNNING LINE VAC Solenoid Valve (Normally Open when not energized) ENERGIZED (CLOSED) When the Vac Unit On/Off switch is turned on, line voltage is supplied to one side of the Pressure Switch. If the Pressure Switch is at 80 PSI (+ / - 5 PSI) or below it's contacts are closed. Line voltage is then supplied to the Compressor energizing it. Supply Line Voltage 24 VAC Contactor w/ 24 VAC Coil CLOSED LINE VAC Unit On / Off Switch ON LINE VAC Pressure Switch LINE VAC CLOSED (Below Pressure Setting) LINE VAC Compressor RUNNING AA150700i It is also supplied to the normally open Solenoid Valve, closing the valve, preventing the compressed air from being discharged to the Condensate Drain Can Assembly. The system will continue to run until the pressure inside the system reaches 100 PSI ( + / - 5 PSI) Note: The compression and purge cycles will continue as the pressure inside the system fluctuates below the 80 PSI and up the 100 PSI pressure settings. All Compression Cycle A-3

10 Operation & Troubleshooting Purge Cycle The pressure switch will open, shutting off electrical current to compressor head(s) and the solenoid/purge valve. A mechanical valve on the pressure switch opens up to unload the compressor heads. At the same time the unloader solenoid opens, starting the purge cycle. Intake Air Intake Filter Troubleshooting Motor runs Intermittently or "Chugs"... A-7 Moisture Indicator is Pink... A-10 Element Indicator is Red... A-11 High Pressure... A-12 Excessive Noise... A-13 Overheating... A-14 Compressor runs to Frequently... A-15 Purge Tank (Releasing Pressure) Air from the purge tank flows back through the gauge cluster in through the metered orifice. The clean dry air drives out the moisture from the desiccant beads and into the condensate drain canister.. PreCooler Coalescing Filter Compressor (Not Running) Dessicant Tank Purge Air Check Valve (Closed) Storage Air White Red Atmospheric & Purge Tank Pressure Pressurized Air Moisture Indicator Pressure Relief Valve (Closed) (Opens at 125 PSI) Note: Bottom Port (Open) Ball Valve (Open) Manual Needle Purge Valve (Closed) Under normal operation conditions, a small amount of moisture with some oil may be under the unloader exhaust muffler. This is not a malfunction. It is evidence of what the filter on the Desiccant Drying System captured the moisture, and did not permit it to enter the operatory air supply. Condensate Drain Can Assembly Normally Open Solenoid Valve (Open) Pressure Switch (Contacts Open) Compressor Head Unloader Valve (Open) 100 PSI Supply Air to Operatory Storage Tank (Pressure At Pressure Switch Setting) AA150501i A-4 Purge Cycle All

11 Operation & Troubleshooting Purge Cycle Electrical Schematic Pressure inside the system reaches 100 PSI (+ / - 5 PSI) causing the Pressure Switch Contacts to open. Power is removed from the Compressor stopping it. At the same time, power is removed from the Normally Open Solenoid Valve, Opening the valve, allowing pressure to be released from the top of the compressor head to the Condensate Drain Can Assembly. It also allows the Purge Tank to release its pressure thru the Desiccant Tank to the Condensate Drain Can Assembly. The Fan Motor continues to run as voltage is present as long as the Wall On / Off Switch is On, providing power to the contactor Coil closing its contacts. Line VAC Supply Line Voltage 24 VAC Transformer ENERGIZED 24 VAC Fuse On Unit 24 VAC Contactor w/ 24 VAC Coil CLOSED LINE VAC Wall On / Off Switch ON Fan Motor RUNNING Unit On / Off Switch ON LINE VAC Pressure Switch NO VAC OPEN (At or Above Pressure Setting) NO VAC NO VAC Solenoid Valve (Normally Open when not energized) NOT ENERGIZED (OPEN) Compressor OFF AA150800i Note: The compression and purge cycles will continue as the pressure inside the system fluctuates below the 80 PSI and up the 100 PSI pressure settings. All Purge Cycle A-5

12 Operation & Troubleshooting Compression Cycle Refer To: Pressure Switch... B-2 ON/OFF Switch... B-7 Capacitor... B-10 Compressor Heads... B-13 Low Voltage Contactor... B-17 3rd Check Capacitor Refer to: Section B Capacitor Problem: Compressor will not Start 1st Check Low Voltage Contactor Refer to: Section B Low Voltage Contactor 5th Motor/Compressor Frozen Turn off electrical power. Verify compressor fan will rotate by hand. If not Replace compressor head(s). Refer to: Section B Compressor Heads 4th Thermal Overload Cut Allow compressor to cool, then restart. 2nd Check voltage at Breaker Was proper voltage is measured at the breaker and not at motor terminals? If yes, check the in the following order : Check for broken or loose wires. If there is Pressure, Perform Pressure Switch Adjustment. (Refer to: B Pressure Switch) Check voltage at On/Off Switch. (R42 Model ONLY) (Refer to: B On/Off Switch) A-6 Compression Cycle All

13 Compression Cycle & Purge Cycle- continued Operation & Troubleshooting Problem: Motor runs Intermittently or "Chugs" Refer To: Line Voltage... B-35 Perform Line Voltage Test Refer to: Section B Line Voltage 1st 2nd Purge System Manually purge compressor. AA orfice 3rd Check for Blockage in Air Line Remove discharge hose from head and test run. If unit runs correctly check for obstruction in air line to storage tank. Obstruction will be at an orifice or other restriction. All Compression & Purge Cycle A-7

14 Operation & Troubleshooting Compression Cycle - continued Refer To: Solenoid (Exhaust Valve)... B-24 Pressure Switch... B-2 Intake Filter... B-19 Checking for Leaks... B-21 Compressor Heads... B-13 Check for Blockage in Air Line Inspect all air lines for restrictions. 6th Problem: Compressor Heads run but will not Pressurize to 100 PSI 9th Check for Worn Piston Cup Refer to: Section B Compressor Heads 5th Check for Air Leaks Refer to: Section B Checking for Leaks. Check Coalescing Filter 7th If indicator is red, change filter. Refer to: Section B Coalescing Filter 8th Check Solenoid Refer to: Section B Solenoid Clean or Replace Intake Filter Refer to: Section B Intake Filter. 4th Check Exhaust Valve Refer to: Section B Solenoid. 2nd Verify Shut-Off valve is Closed. 1st Clean or Replace Pressure Switch Unloader Valve Refer to: Section B Pressure Switch. 3rd AA A-8 Compression Cycle All

15 Compression Cycle - continued Operation & Troubleshooting Problem: Compressor cycles with no air being used Refer To: Checking for Leaks... B-21 3rd Check for Air Leaks Refer to: Section B Checking for Leaks. 1st Check Manual Purge Valve Should be in "OFF" position (Horizontal) AA Check for Leak in Office Air System Close the storage tank shut off. Pump up storage tank to 100 PSI. If pressure is maintained at 100 PSI for minutes, leak is not in compressor. 2nd All Compression Cycle A-9

16 Operation & Troubleshooting Refer To: Desiccant Tank... B-27 Solenoid... B-24 Compressor runs to frequently... A-15 Compression Cycle & Purge Cycle- continued Problem: Moisture Indicator is Pink Is Compressor running too frequently? Refer to: Section A Compressor runs to Frequently. 3rd Perform Unloading System Function Test Refer to: Section B - Solenoid 1st Check Desiccant Drying Chamber Refer to: Section B Desiccant Tank 2nd Purge Air Lines 5th Preform a Manual Purge. Note: Under normal operation conditions, a small amount of moisture with some oil may be under the unloader exhaust muffler. This is not a malfunction. It is evidence that the filter on the Desiccant Drying System captured the moisture, and did not permit it to enter the operatory air supply. AA Check for Condensation in Air Lines Is there moisture in operatory air? 4th A-10 Compression & Purge Cycle All

17 Compression Cycle & Purge Cycle- continued Problem: Element Indicator is Red Operation & Troubleshooting Refer To: Intake Filter... B-19 Coalescing Filter... B-22 Verify Float is Working Check for clogs. 2nd 1st Change Intake Filter Refer to: Section B Intake Filter 3rd Change Coalescing Filter Refer to: Section B Coalescing Filter AA All Compression & Purge Cycle A-11

18 Operation & Troubleshooting Refer To: Check Valves... B-36 Compression Cycle & Purge Cycle- continued Problem: High Pressure 1st Check Exhaust Remove hoses one at a time between Heads and Pressure Relief Valve. Verify exhaust air flow is not obstructed. Clean or Replace Check Valve 2nd Refer to: Section B Check Valve 3rd Check Pressure Switch Settings Check contacts and operating range, psi. AA A-12 Compression & Purge Cycle Rev 6/06 All

19 Operation & Troubleshooting Compression Cycle & Purge Cycle- continued Problem: Excessive Noise Refer To: Compressor Heads... B-13 Checking for Leaks... B-21 Intake Filters... B-19 1st Check for Worn Piston Cup Refer to: Section B Compressor Heads 2nd Check for Leaky Hose Refer to: Section B Checking for Leaks 3rd Replace Intake Filters Refer to: Section B Intake Filters. AA All Compression & Purge Cycle A-13

20 Operation & Troubleshooting Refer To: Intake Filter... B-19 Solenoid (Exhaust Valve)... B-24 Pressure Switch... B-2 Line Voltage... B-35 Fans... B-31 Compression Cycle & Purge Cycle- continued Problem: Overheating 4th Check Exhaust Remove hoses one at a time between Heads and Pressure Relief Valve. Verify exhaust air flow is not obstructed. 1st Clean Intake Filters Refer to: Section B Intake Filter 2nd Clean or Replace Exhaust Valve Refer to: Section B Solenoid. Check fan operation Refer to: Section B Fans 6th Clean or Replace Pressure Switch Unloader Valve Refer to: Section B Pressure Switch. 3rd AA Perform Line Voltage Test Refer to: Section B Line Voltage 5th A-14 Compression & Purge Cycle All

21 Operation & Troubleshooting Compression Cycle & Purge Cycle- continued Problem: Compressor runs too Frequently Refer To: Solenoid... B-24 Pressure Switch... B-2 Coalescing Filter... B-23 1st Check for Air Leaks Refer to: Section B Checking for Leaks Change Element Indicator 2nd Refer to: Section B Coalescing Filter Check Solenoid Refer to: Section B Solenoid 3rd 4th Check Pressure Switch Refer to: Section B Pressure Switch AA Note: More than the 35% duty cycle the Rock Compressors are rated for would be "Running to Frequently". All Compression & Purge Cycle A-15

22

23 Component Section BTesting & Repair Component Testing & Repair Components / Procedure Pressure Switch... B-2 On/Off Switch... B-7 Capacitor... B-10 Compressor Heads... B-13 Low Voltage - Master Control Panel... B-17 Intake Filter... B-19 Checking for Leaks... B-21 Coalescing Filter... B-22 Solenoid... B-24 Desiccant Tank... B-27 Fuse Replacement... B-29 Fan Replacement... B-31 Line Voltage... B-35 Check Valve... B-36 B-1

24 Component Testing & Repair Pressure Switch Location and Function During the Compression Cycle... When the storage tank is filled to the preset pressure the pressure switch will open and stop the compressor. Pressure Switch Testing... B-3 Adjustment... B-4 Replacement... B-5 Wiring Diagrams... D-1 Exploded Views / Part Numbers... E-2 During the Purge Cycle... The pressure switch shuts off current to the compressor head(s) and the solenoid / purge valve. A mechanical valve on the pressure switch opens up to unload the compressor headlines. At the same time the unloader solenoid opens, starting the purge cycle. Cut-In / Cut-Out time Cut-In / Cut-Out is the minimum and maximum discharge pressures at which the compressor will switch from unload to load operation (cut-in) or from load to unload (cut-out). The pressure switch controls Cut-In / Cut-Out times. AA AA NOTE: Pressure Switches on models up to 2HP have a Preset Cut-In adjuster. Models 2HP and over have Manual adjuster for Cut-In setting. Both Models have Manual Cut-Out Adjusters. B-2 Pressure Switch All

25 Component Testing & Repair Pressure Switch Testing NOTE: Step 1: Perform Adjustment. Refer to Section B Pressure Switch Adjustment. If Adjustment can not be made, then perform voltage test to the On/Off Switch before checking voltage on the Pressure Switch. You will need the readings from the On/Off Switch check to verify the test on the Pressure Switch. Refer to: Section B On/Off Switch Test Pressure Switch Function / Location... B-2 Adjustment... B-4 Replacement... B-5 On/Off Switch Test... B-8 Wiring Diagrams... D-1 Exploded Views / Part Numbers... E-2 Step 2: Turn power off. GRN W/YLW STRIPE Step 3: Remove cover and verify there are no black/burnt melted wires. If there are replace pressure switch. WHT GRY GRY BLK Step 6: Place meter probes on black wire and white wire. Step 4: Turn power on. Step 5: Set meter to V. WHT GRN W/YLW STRIPE GRY GRY BLK Step 7: Place meter probes on both inner gray wires. Note: Verify pressure is less than "cut in". Refer to:sectin B: Adjustment. Meter Reading No Pressure on Gauges. Not same Volt Readings as on/off switch. Less than 230 Volts on inner wire check. Status Required Action Replace Pressure Switch. NOTE: If pressure is above cut-in and voltage is read during testing when the contacts are supposed to be open, then contacts are melted and pressure switch needs replacement. Same Voltage as on/off Switch. 230 Volt Readings on inner wires. Pressure Switch - OK AA153600i All Pressure Switch B-3

26 Component Testing & Repair Pressure Switch Adjustment Pressure Switch Function / Location... B-2 Testing... B-3 Replacement... B-5 Wiring Diagrams... D-1 Exploded Views / Part Numbers... E-2 Cut-In Adjustment Step 1: Run compressor until pressure reaches 100 lbs. Note: Compressor should shut off close to 100 lbs. Cut-Out Adjustment - Models over 2HP Step 1: Run compressor until pressure reaches 100 lbs. Note: Compressor should shut off close to 100 lbs. If not perform a Cut-In adjustment Cut-In Adjuster Cut-Out Adjuster Step 2: If compressor didn't shut down at 100 lbs, turn nut on adjuster to reset gauge. Clockwise will increase reading. Counter Clockwise will decrease reading. AA Step 2: Compressor should start back up after a 20 lb drop on the gauge. If needed, turn nut to adjust. Clockwise will increase reading. Counter Clockwise will decrease reading. NOTE: Pressure Switches on models up to 2 HP have a Preset Cut-In adjuster. Models 2HP and over have Manual adjuster for Cut-In setting. Both Models have Manual Cut-Out Adjusters. B-4 Pressure Switch All

27 Component Testing & Repair Pressure Switch Replacement of Unloader Valve Twin Rock Valve installed in Opposite Direction Removal Caution Line voltage is present at the power inlet Step 1: Disconnect power to compressor. Remove cover. Pressure Switch Function / Location... B-2 Testing... B-3 Adjustment... B-4 Wiring Diagrams... D-1 Exploded Views / Part Numbers... E-2 Installation Step 3: Install cover. Unloader Valve Removal Step 3: Remove valve from pressure switch. AA153700i Installation Step 1: Install new unloader valve in pressure switch. Removal Step 2: Remove air line from pressure switch. Installation Step 2: Reconnect air line to unloader valve. All Pressure Switch B-5

28 Component Testing & Repair Pressure Switch Replacement of Pressure Switch Removal Caution Line voltage is present at the power inlet Step 1: Disconnect power to compressor. Remove cover. AA155800i Pressure Switch Function / Location... B-2 Testing... B-3 Adjustment... B-4 Replacement... B-5 Wiring Diagrams... D-1 Exploded Views / Part Numbers... E-2 Removal Step 4: Unscrew pressure switch from storage tank fitting. Installation Step 1: Screw new pressure switch into storage tank. Removal Step 3: Disconnect wires. Remove conduits from pressure switch box. Installation Step 2: Install conduits. Connect wiring. Refer to: Sec D Wiring Diagrams Installation Step 4: Install cover. AA154100i Removal Step 2: Remove air line from pressure switch. Installation Step 3: Install air line. B-6 Pressure Switch All

29 Component Testing & Repair On / Off Switch Function / Location The on/off switch sends power to the compressor heads. On / Off Switch Testing... B-8 Replacement... B-9 Wiring Diagrams... D-1 Exploded Views / Part Numbers... E-9 R22 - Twin Rock R42 - Triple Rock All On / Off Switch B-7

30 Component Testing & Repair On / Off Switch Testing NOTE: Verify system is not pressurized. (System pressure below 80 PSI.) Step 3: Place meter probes on rear terminals on relay. Reading = 230 volts. On / Off Switch Function / Location... B-7 Replacement... B-9 Wiring Diagrams... D-1 Exploded Views / Part Numbers... E-9 Step 1: Remove electrical cover and on/off switch cover. Relay Step 2: Set meter to V. Step 5: Place meter probes on two gray wires on right side of terminal block. Reading should show no voltage. To Motor To Motor Step 8: Place meter probes on top right and top left side. Reading = 230 volts. Top Step 4: Turn switch off. Step 7: Turn switch on. Step 9: Place meter probes on bottom right and bottom left side. Reading = 230 volts. Bottom Meter Reading Status Required Action AA154600i Relay, Switch readings < 230 volts. If Voltage is present on Terminal Test. Replace On / Off Switch. Caution Do not allow wires to touch together. Relay, Switch readings = 230 volts. If Voltage is not present on Terminal Test. On / Off Switch - OK Step 6: Pull switch out of housing. B-8 On / Off Switch R42

31 Component Testing & Repair On / Off Switch Replacement Removal Caution Line voltage is present at the power inlet Step 1: Disconnect power to compressor. AA155800i On / Off Switch Function / Location... B-7 Testing... B-8 Wiring Diagrams... D-1 Exploded Views / Part Numbers... E-9 Twin Model Triple Model Removal Twin Model - R22 Step 2: Remove electrical box cover. Disconnect wires and remove switch. Removal Triple Model - R42 Step 2: Remove switch cover. Replace switch. Refer to: Section D for wire diagram Installation Twin Model - R22 Step 1: Connect wires and install switch. Install electrical box cover. Installation Triple Model - R42 Step 1: Connect wires and install switch. Install switch cover. AA154300i Installation Step 3: Reconnect power to compressor. All On / Off Switch B-9

32 Component Testing & Repair Capacitor Function / Location The capacitor s function is to store electricity, or electrical energy. The capacitor also functions as a filter, passing alternating current (AC), and blocking direct current (DC). Capacitor Testing... B-11 Replacement... B-12 Wiring Diagrams... D-1 Exploded Views / Part Numbers... E-5 B-10 Capacitor All

33 Component Testing & Repair Capacitor Testing Step 2: Disconnect wire connections on capacitor. Capacitor Function / Location... B-10 Replacement... B-12 Wiring Diagrams... D-1 Exploded Views / Part Numbers... E-5 Step 3: Discharge capacitors by touching two screwdrivers to two contacts on capacitor and each other. Electrical Box Cover Step 1: Remove electrical box cover. Step 5: Place meter probes on capacitor connectors. Meter Reading Status Required Action "O" or "Open" Reading Replace Capacitor Start Low and Increases Capacitor - OK Step 4: Set meter to Ω. All Capacitor B-11

34 Component Testing & Repair Capacitor Replacement Removal Caution Line voltage is present at the power inlet Step 1: Disconnect power to compressor. AA155800i Capacitor Function / Location... B-10 Testing... B-11 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 Removal Step 2: Remove quiet cover if applicable. Refer to: Section C Quiet Cover Removal Step 3: Remove electrical box cover. Disconnect wires Remove mounting screws from capacitor and replace. Installation Step 2: Install Quiet Cover if applicable. Refer to: Section C Quiet Cover. Capacitors Installation Step 1: Install capacitor mounting screws. Connect wires. Install electrical box cover. Electrical Box Cover B-12 Capacitor All

35 Component Testing & Repair Compressor Heads Function and Location Atmospheric air is drawn in through intake filters to the compressor head(s) where air is compressed and exhausted to the resorvoir (tank). Heat is generated from the compression process. Compressor Heads Rebuild - Removal... B-14 Rebuild - Installation...B-15 Replacement... B-16 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 All Compressor Heads B-13

36 Component Testing & Repair Compressor Heads Rebuild - Removal Removal Caution Line voltage is present at the power inlet Vent air lines to the compressor to release pressure. Step 1: Disconnect power to compressor. Removal Step 2: Remove head bolts and head. Note: Mark orientation of ports so head covers are reassembled correctly. Exhaust Side Compressor Heads Function / Location... B-13 Rebuild - Installation...B-15 Replacement... B-16 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 Removal Step 4: Remove cross tubes. Intake Side Removal Step 3: Clean and inspect valves and valve plates. If discolored or worn, replace. Verify "X" notch on corner is facing up. Torque in. lbs. Removal Step 5: Remove the retainer plate, piston cup and cylinder. Discard piston cup and o-rings on cylinder and cross tube. NOTE: Verify proper orientation of the ports, incorrect placement of the head on the valve plate will result in misalignment of the valve ports and valve damage. Damaged valves must be replaced before pump can operate. AA154900i Removal Step 6: Clean all parts with water based solvent cleaner. B-14 Compressor Heads All

37 Compressor Heads Rebuild - Installation Installation Step 9: Reconnect power supply to pump and check compressor for leaks. Refer To: Section B Checking for Leaks. Installation Step 8: Install head over valve plates and torque to in. lbs. Component Testing & Repair Compressor Heads Function / Location... B-13 Rebuild - Removal... B-14 Replacement... B-16 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 Installation Step 7: Install o-ring on top of valve plate. Installation Step 4: Install valves. Installation Step 6: Install new cylinder o-rings in the bottom of the valve plates. Position valve plates over cylinders. Installation Step 5: Install new o-rings on cross tubes and reinstall tubes into valve plate. Installation Step 3: Install cylinders at an angle over cups to avoid damage. Installation Step 1: Place new piston cup on rod. Installation Step 2: Apply medium thread locking compound to retainer screws and install retainer plate. AA154800i All Compressor Heads B-15

38 Component Testing & Repair Compressor Heads Replacement Removal Installation Caution Line voltage is present at the power inlet Vent air lines to the compressor to release pressure. Step 1: Disconnect power to compressor. Equipment Alert Do not over torque Inlet/Outlet fittings. Head will crack and leak. Compressor Heads Function / Location... B-13 Rebuild - Removal... B-14 Rebuild - Installation...B-15 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 Removal Step 2: Disconnect the wiring, tubing and mounting nuts from heads. Note: On Model R22 Loose platform bolts and raise platform slightly for more room to work. Step 1: Install new heads on platform. Connect the following to heads... Wiring - Refer to: Section D Wire Diagrams Tubing Mounting Nuts under platform Removal Step 3: Remove heads from platform. AA155000i Installation Step 2: Reconnect power supply to pump and check compressor for leaks. Refer To: Section B Checking for Leaks. B-16 Compressor Heads All

39 Component Testing & Repair Low Voltage - Master Control Panel Function and Location The Master Control Panel will allow the user to control the dental equipment from the office area. Supplies power to the fans and relay switch. Low Voltage Control Testing... B-18 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 *Note: Relay Switch is always live unless Low Voltage is turned off. AA156700i Equipment Room Operatory All Low Voltage Contactor B-17

40 Component Testing & Repair Low Voltage - Master Control Panel Testing Step 1: Remove electrical box cover. Check for broken or loose wiring. Step 2: Check high voltage on contactor relay. Set meter to V. Place meter probes on front, black and white wires. Note: Verify reading is 230 volts. Relay Low Voltage Control Function and Location... B-17 Fuse... B-29 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 Step 4: Insert screwdriver into relay to start manually. Note: if system starts, replace contactor relay. 230 VAC Primary Relay Transformer 230 VAC Secondary RED LOW VOLTAGE WHITE CONTROL BLUE BLU WHT BLU YLW Caution When testing components with power on use care to prevent electrical shock. 1/2 A FUSE Step 5: Check fuse. Visually check to see if blown. Refer To: Section B Fuse Step 3: Check Low Voltage from transformer to low voltage wires. Place meter probes on Yellow and Red wire connections at Relay. Note: Verify reading is 24 volts. Meter Reading Status Required Action High Voltage = < 230 Volts Low Voltage = < 24 Volts High Voltage = 230 Volts Low Voltage = 24 Volts Replace Relay Replace Transformer Relay - OK Transformer - OK AA158701i Step 6: Bypass the remote switch to verify it is not defective. Disconnect Red, White and Blue wires from switch. Connect the blue and red wires together. If the relay and fans work, then replace remote switch or wiring. B-18 Low Voltage Contactor Rev 5/07 All

41 Component Testing & Repair Intake Filters Function and Location Atmospheric air is drawn in through intake filters before flowing to the compressor heads. Intake filters protect the heads from dust and debris and should be changed annually. Intake Filters Clean and Replacement... B-20 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 All Intake Filter B-19

42 Component Testing & Repair Intake Filter Clean and Replacement Cleaning Filter Step 1: Unclamp filter from hose. Remove barbed fitting on motor tray. Pop off one side of filter housing. Intake Filters Function and Location... B-19 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 Removal Step 1: Remove filter from hose and barbed fitting on motor tray. Cleaning Filter Step 2: Remove filter and clean with a nonflammable solvent. Wipe out housing and allow filter and housing to dry. Do Not Remove Cap Installation Step 1: Install filter onto barbed fitting. Cleaning Filter Step 3: Install filter back into housing. Assemble housing. Connect hose and install onto barbed fitting. AA156800i Installation Step 2: Clamp hose onto filter housing. B-20 Intake Filter All

43 Component Testing & Repair Checking for Leaks Checking for Leaks Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 Check Step 1: Verify shut-off valve is closed. Check Step 2: Turn power on. Note: Compressor should run quietly and storage tank will begin to pressurize. Check Step 3: Use soapy water to check for compressor plumbing joint leaks. AA Soapy Water All Checking for Leaks B-21

44 Component Testing & Repair Coalescing Filter Function and Location The Coalescing Filter removes water, oil and other contaminates from compressed air. Coalesced liquids gravitate into the sump bowl where they are discharged through the drain. Coalescing Filter Replacement... B-23 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 AA B-22 Coalescing Filter All

45 Coalescing Filter Replacement Removal Step 1: Disconnect power to compressor. Drain storage tank of all air and liquid. Removal - Element Indicator Only Step 2: Remove mounting screws. Remove Element Indicator by cutting ties. Installation - Element Indicator Only Step 1: Install new indicator and mounting screws. Element Component Testing & Repair Coalescing Filter Function and Location... B-22 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 Removal - Entire Assembly Only Step 2: Disconnect coalescing housing from tube fitting on bottom. Disconnect top of coalescing houseing at fittings. AA155800i Removal - O-ring and Filter Only Step 2: Push on black release button on front of filter housing, turn slightly and pull off lower half of coalescing assembly. Filter O-ring Removal - Entire Assembly Only Step 3: Remove fittings and tube from bottom of housing. Installation - Entire Assembly Only Step 1: Install fittings and tube onto bottom of housing. Removal - O-ring and Filter Only Step 3: Remove filter and o-ring. Release Button Installation - Entire Assembly Only Step 2: Connect power to compressor and check for leaks. Refer to: Section B Checking for Leaks Installation - O-ring and Filter Only Step 1: Install new filter and o-ring. Old Style Fittings (Threaded) Old Fittings can be Replaced by New Style Fittings (Compression) Installation - O-ring and Filter Only Step 2: Connect housing back to coalescing assembly and lock into place. All other Apollo Models New Style Fitting R22 Model New Style Fitting Soapy Water AA157300i All Coalescing Filter Rev 3/07 B-23

46 Component Testing & Repair Solenoid Function and Location During the compression cycle... The normally open Solenoid Valve, closes, preventing the compressed air from being discharged to the Condensate Drain Can Assembly. Solenoid Testing... B-25 Replacement... B-26 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 During the purge cycle... Power is removed from the Normally Open Solenoid Valve, Opening the valve, allowing pressure to be released from the top of the compressor head to the Condensate Drain Can Assembly. It also allows the Purge Tank to release its pressure thru the Desiccant Tank to the Condensate Drain Can Assembly, regenerating the desiccant drying system. B-24 Solenoid All

47 Component Testing & Repair Solenoid Testing Step 1: Check high voltage on pressure switch. Set meter to V Step 3: Verify no air is coming out of bottom of coalescing filter. Solenoid Function and Location... E-24 Replacement... E-26 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 GRN W/YLW STRIPE BLK GRY GRY WHT Step 2: Place meter probes on both inner gray wires. Meter Reading Status Required Action Less than 230 Volts on inner wire check. Air is coming out of coalescing filter. Doesn't work after cleaning. 230 Volt Readings on inner wires. NO air coming out of coalescing filter. Works after Cleaning. Replace Pressure Switch or ON/OFF Switch Pressure Switch - OK IN AA158900i Step 4: Clean solenoid. Loosen nut on top of solenoid and lift top off. Wipe stem off with a clean rag. Reassemble. All Solenoid B-25

48 Component Testing & Repair Solenoid Replacement Removal Step 1: Disconnect power to compressor. Drain storage tank of all air and liquid. Installation Step 1: Install new solenoid assembly to fittings. Connect tubing to coalescing filter. Solenoid Function and Location... B-24 Testing... B-25 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 AA155800i Model R22 Installation Step 2: Reconnect wiring and conduit to pressure switch. Refer To: Section D Wire Diagrams Installation Step 4: Connect power. Perform Pressure Switch adjustment to verify system is working correctly. Refer To: Section B Pressure Switch Removal Step 3: Disconnect tubing from solenoid assembly. Unscrew solenoid assembly from fittings. Model R42 Removal Step 2: Remove pressure switch cover. Disconnect wiring and conduit from pressure switch. Refer To: Section D wire Diagrams Installation Step 3: Install pressure switch cover. AA159000i B-26 Solenoid All

49 Component Testing & Repair Desiccant Tank Function and Location Moisture vapor drying takes place as the air is directed through the silica and alumina gel (desiccant material) from bottom to top in the drying chamber of the desiccant tank. Desiccant Tank Replacement... E-28 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 When the desiccant becomes so saturated with water and oil that the manual purge cycle fails to regenerate it, the desiccant tank will need to be replaced. The drying chamber is internally coated with an anticorrosive material. A stainless steel wire mesh screen supports the air drying media, consisting of an activated alumina buffer bed and silica gel (desiccant material). Twin Model R22 Triple Model R42 All Desiccant Tank B-27

50 Component Testing & Repair Desiccant Tank Replacement Removal Step 1: Disconnect power to compressor. Open storage tank drain valve to release all air pressure, and then close valve. Installation Step 4: Connect power to the compressor. AA155800i Removal Step 3: Remove mounting bolts. Installation Step 2: Install tank to saddle with mounting bolts. Desiccant Tank Location and Function... E-27 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 Installation Step 5: Test pump-up and purge operations. Check for and seal all leaks. Triple Model R42 Twin Model R22 Removal Step 2: Loosen compression nuts and separate copper tubing from end of the desiccant tank. Installation Step 3: Install copper tubing to desiccant tank. Removal Step 4: Unscrew tank from solenoid assembly. Installation Step 1: Screw tank into solenoid assembly. AA159200i B-28 Desiccant Tank All

51 Component Testing & Repair Fuse Function and Location The fuse protests the low voltage control system against overload failure, mainly short circuits. Only the Triple Model R42 has a fuse. It is located on the front panel. Fuse Replacement... E-30 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 Note: The Twin Model R22, uses a low voltage circuit breaker switch instead. R42 Fuse B-29

52 Component Testing & Repair Fuse Replacement Removal Step 1: Unscrew fuse holder out of front panel. Remove fuse from holder. Fuse Function and Location... E-29 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 Installation Step 1: Insert fuse into fuse holder and screw into front panel. Fuse NOTE: Fuse needs replaced if it is burnt or broken. AA159500i B-30 Fuse R42

53 Component Testing & Repair Fan Function and Location Fans cool down the heads and also cause air to flow over the after cooler which helps cool the system air after compression. Fan Testing... E-32 Replacement (R22-Twin)... E-33 Replacement (R42-Triple)...E-34 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 Twin Model R22 (Two fans are located underneath Heads) Triple Model R42 All Fan B-31

54 Component Testing & Repair Fan Testing Step 2: Set meter to V. Fan Function and Location... E-31 Replacement (R22-Twin)... E-33 Replacement (R42-Triple)... E-34 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 Step 1: Disconnect wires from fan. TO 3 FANS BLK BLK BLK WHT WHT WHT WHT BLK Step 3: Place meter probes on fan wires. Reading = 230 volts. Meter Reading Status Required Action Step 4: Place meter probes on rear terminals on relay. Reading = 230 volts. Relay One Fan < 230 Volts Replace Fan All Fans < 230 Volts 230 Volts Replace Relay Fan and Relay OK AA159900i B-32 Fan All

55 Component Testing & Repair Fan Replacement Removal Pre-Cooler Fan Step 1: Remove electrical box cover. Disconnect wires between fan and relay. Installation Pre-Cooler Fan Step 1: Install screws into fan guard. Connect fan wires. Fan Function and Location... E-31 Testing... E-32 Replacement (R42-Triple)... E-34 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 Removal Pre-Cooler Fan Step 2: Remove screws from pre-cooler fan and replace fan. Installation Pre-Cooler Fan Step 2: Install electrical box cover. Installation Fans under Platform Step 3: Install screws into fan guard on top and bottom of platform. Removal Fans under Platform Step 3: Remove intake filters. Remove tubing from platform. Unbolt motors. AA160200i Removal Fans under Platform Step 4: Move motor up to remove screws out of fan guard on platform. Remove screws from fan guard underneath platform. Replace fan from underneath. Installation Fans under Platform Step 4: Bolt Motors into platform. Install tubing from platform to motors. Install intake filters to tubing. Reconnect wires between fan and relay. R22 Fan B-33

56 Component Testing & Repair Fan Replacement Removal Step 1: Disconnect fan wires. Fan Function and Location... E-31 Testing... E-32 Replacement (R22-Twin)...E-33 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 Removal Step 3: Remove screws from front of fan and remove fan. Removal Step 2: Remove screws from behind platform. Installation Step 2: Connect wires between fan and relay. Installation Step 1: Install screws into front and back of fan guard. AA160300i B-34 Fan R42

57 Component Testing & Repair Line Voltage Testing Components to check for line voltage include the High Voltage Breaker, Low Voltage Contactor, Relay, On/Off Switch and the Pressure Switch. Step 1: Check high voltage in Breaker. Set meter to V Line Voltage Pressure Switch (Testing)... B-3 On/Off Switch (Testing)... B-8 Low Voltage Contactor (Testing)... B-18 Wiring Diagrams... D-1 Exploded Views (R22 - Twin)... E-2 Exploded Views (R42 - Triple)... E-3 Step 2: Check voltage in Low Voltage Contactor and Relay. Refer to: Section B Low Voltage Contactor Meter Reading Status Required Action < 230 Volts in Breaker Replace Breaker < 24 Volts - Low Voltage Contactor Replace Low Voltage Contactor < 230 Volts - Relay Replace Relay < 230 Volts - On/Off Switch Replace On/Off Switch < 230 Pressure Switch Replace Pressure Switch AA160400i < 230 Volts Terminal Block Replace Terminal Block 230 Volts = Breaker Breaker 24 Volts = Low Voltage Contactor Low Voltage Contactor Step 3: Check voltage in On/Off Switch and high voltage terminal connections. Refer to: Section B On/Off Switch 230 Volts = Relay Relay 230 Volts = On/Off Switch On/Off Switch 230 = Pressure Switch Pressure Switch Step 4: Check voltage in Pressure Switch. Refer to: Section B Pressure Switch 230 = Volts Terminal Block Terminal Block All Low Voltage Contactor B-35

58 Component Testing & Repair Check Valve Replacement Removal Step 1: Disconnect power to compressor. NOTE: Installation Step 12: Connect power to compressor. Use teflon tape and thread adhesive on fitting threads when reinstalling. Removal Step 4: Remove tube from Desiccant tank to elbow. Remove elbow from block fitting. Removal R42 Models Only Step 2: Remove shroud. Installation R42 Models Only Step 11: Install shroud. Removal Step 5: Unscrew gauge from block fitting. Installation Step 8: Install gauge. Block R22 Model Removal Step 6: Unscrew block. Unscrew check valve. R42 Model Installation Step 9: Install elbow into block. Install tube from elbow to Desiccant tanks. Check Valve Installation Step 7: Install new check valve. Install block. Removal Step 3: Remove tube from side of cluster block to below moisture gage. Installation Step 10: Install tube from side of cluster block to below moisture gage. AA171100i B-36 Check Valve Rev 7/06 ALL

59 Section C Access Procedures Access Procedures Removing & Installing: Quiet Cover... C-2 Midmark Corporation 2005 SF-1864 C-1

60 Access Procedures Quiet Cover Removal / Installation Refer To: Exploded Views / Part Numbers: R22 - Twin... E-2 R42- Triple...E-3 Removal Step 1: Turn both quiet cover fastener counter-clock wise. Lift quiet cover off of compressor. Fastener AA160500i Installation Step 1: Place quiet cover on top of compressor. Turn both quiet cover fastener clockwise. C-2 Quiet Cover Midmark Corporation 2005 SF-1864 All

61 Section D Wiring Diagrams Wiring Diagrams Model R22... D-2 R42 (Motor & Pressure Switch).. D-4 R42 (On/Off Switch)... D-5 D-1

62 Wiring Diagrams LEFT MOTOR WHT BLK GRN W/YLW BRN BRN/WHT BRN BRN/WHT RIGHT MOTOR BLK WHT GRN W/YLW Refer To: Operation & Troubleshooting... A-1 Component Testing / Repair... B-1 Access Procedures... C-1 Wiring Diagrams... D-1 Exploded Views / Part Numbers... E-1 WHT WHT BLK BLK YLW BLK L1 115 BLK or 230 ORG WHT 230 VAC PRIMARY 24 VAC SECONDARY BLU GRN W/YLW GRY GRY WHT T1 T2 L2 115V or 230V WHT BLK GRN W/YLW RED WHT BLK WHT BLK WHT BLK BLK WHT GRN W/YLW GRY GRY FAN ON A/C WHT BLK WHT (L) FAN BLK WHT WHT BLK (R) FAN GRY GRY ON/OFF - 1/2 AMP CIRCUIT BREAKER SWITCH BLU RED WHT LOW VOLTAGE MPM60622 D-2 Motor & Pressure Switch Rev 2/22/06 R22

63 Wiring Diagrams TO LEFT MOTOR # 1 BRN BRN WHT BLK GRN W/YLW STRIPE TO 3 FANS BLK BLK BLK WHT WHT WHT TO TERMINAL BLOCK 230 VAC Primary Transformer WHT Relay VAC Secondary BLK YLW BLU BLK TO LT MTR BLK TO MID MTR WHT TO LT MTR WHT TO MID MTR WHT BLK BRN BRN TO WHT TERMINAL BLOCK GRY GRY BLK GRN W/YLW STRIPE TO MIDDLE MOTOR # 2 BRN BRN TO WHT RIGHT BLK MOTOR GRN W/YLW STRIPE # 3 PRESSURE SWITCH GRN W/YLW STRIPE BLK GRY GRY GRY WHT ON/OFF SWITCH L1 T1 T2 L2 Refer To: Operation & Troubleshooting... A-1 Component Testing / Repair... B-1 Access Procedures... C-1 Wiring Diagrams... D-1 Exploded Views / Part Numbers... E-1 WHT ORG WHT RED BLU 230V BLK BLK GRN W/YLW STRIPE RED BLUE WHITE LOW VOLTAGE CONTROL 1/2 A FUSE MPM60623 R42 Motor & Pressure Switch Rev 2/22/06 D-3

64 Wiring Diagrams GRN BLK WHT BLK LOAD WHT GRN WHT BLK A D Refer To: Operation & Troubleshooting... A-1 Component Testing / Repair... B-1 Access Procedures... C-1 Wiring Diagrams... D-1 Exploded Views / Part Numbers... E-1 BLK WHT B C LINE E Item Description Part Number A B Switch Box, DBL, W/GND STUD Elec. Wiring to Switch F C Press Switch Wiring D Ground Label E Nut KEPS 10/32 BNT20210 F Switch ON/Off DPST ECS10417 AA145100i D-4 On / Off Switch R42

PowerMax PM-1 PM-3 PM-4. Service and Parts Manual FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY. Model Numbers:

PowerMax PM-1 PM-3 PM-4. Service and Parts Manual FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY. Model Numbers: PowerMax Model Numbers: PM-1 PM-3 PM-4 Service and Parts Manual FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY SF-1920 Part No. 10559400 Rev AA1 (7/14/17) Table Of Contents General Information Section A Section

More information

PowerVac. Service and Parts Manual P3 P5 P7. Single Vac Units. Twin Vac Units P6 P10 P14 FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY.

PowerVac. Service and Parts Manual P3 P5 P7. Single Vac Units. Twin Vac Units P6 P10 P14 FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY. PowerVac Model Numbers: Single Vac Units P3 P5 P7 Earlier Version Current Version Service and Parts Manual Twin Vac Units P6 P10 P1 FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY SF-1873 Part No. 00-09-00

More information

PowerAir Compressor Installation

PowerAir Compressor Installation PowerAir Compressor Installation Applies to Models: P21-115 Volts P22, P32-230 Volts P52, P72-230 Volts Equipment Alert Compressor system must be installed per local plumbing and electrical codes. Model

More information

ClassicSeries Compressor Installation

ClassicSeries Compressor Installation ClassicSeries Compressor Installation Applies to Models: CL21-115 Volts CL22, CL32, CL52-208- 230 Volts Equipment Alert Compressor system must be installed per local plumbing and electrical codes. Note:

More information

ClassicSeries Wet-Ring Vacuums Single Model Numbers: CV3 CV3R CV5 CV5R CV6 CV6R CV10 CV10R. Service and Parts Manual

ClassicSeries Wet-Ring Vacuums Single Model Numbers: CV3 CV3R CV5 CV5R CV6 CV6R CV10 CV10R. Service and Parts Manual ClassicSeries Wet-Ring Vacuums Single Model Numbers: CV3 CV3R CV5 CV5R Service and Parts Manual Twin Model Numbers: CV6 CV6R CV10 CV10R FOR USE BY MIDMARK TRAINED TECHNICIANS ONLY SF-1915 Part No. 10546800

More information

TECH WEST INC. Manufacturers of Dental Vacuum and Air Systems N. Argyle Ave. Fresno, CA (559) (800) FAX (559)

TECH WEST INC. Manufacturers of Dental Vacuum and Air Systems N. Argyle Ave. Fresno, CA (559) (800) FAX (559) TECH WEST INC. Manufacturers of Dental Vacuum and Air Systems 2625 N. Argyle Ave. Fresno, CA 93727 (559) 291-1650 (800) 428-7139 FAX (559) 348-9677 ULTRA CLEAN OILLESS COMPRESSOR INSTALLATION AND SERVICE

More information

ULTRA CLEAN OILLESS COMPRESSOR

ULTRA CLEAN OILLESS COMPRESSOR ULTRA CLEAN OILLESS COMPRESSOR INSTALLATION AND SERVICE MANUAL Revised 8-08 TECH WEST INC. Manufacturers of Dental Vacuum and Air Systems 2625 N. Argyle Ave. Fresno, CA 93727 (559) 291-1650 (800) 428-7139

More information

PowerMax Surgical Suction Producers

PowerMax Surgical Suction Producers PowerMax Surgical Suction Producers Models: PM-1, PM-3, PM-4 Installation & Operation 10496800 Rev. B Table of Contents Section 1 About PowerMax...1 Section 2 Specifications & Model Selection...3 2.1

More information

FTIR PG28 PURGE GAS GENERATOR OPERATIONS & MAINTENANCE

FTIR PG28 PURGE GAS GENERATOR OPERATIONS & MAINTENANCE FTIR PG28 PURGE GAS GENERATOR OPERATIONS & MAINTENANCE TABLE OF CONTENTS SECTION 1 General Information: Model Description Pg 2 Serial Numbers Description of Operation SECTION 2 Safety Instructions: Pg

More information

Model 8144NC Fresh Air Ventilator

Model 8144NC Fresh Air Ventilator Model 8144NC Fresh Air Ventilator Installation and Operating Instructions MOUNTING BRACKETS WIRE ENTRY LOCATION INTEGRAL PRESSURE PORTS (PORT ON INLET SIDE NOT SHOWN) OVAL OUTLET COLLAR FOR 6" DIAMETER

More information

Installation & Operation

Installation & Operation 2" PVC Vent Pipe (Provided By Plumber) 2" PVC Vent Connection (Vent To Outdoors) 1/4" x 8' High Pressure Water Inlet Hose with Water Control Valves Located On Platform (1/2" or 1/4" MPT Fittings Provided)

More information

HEAVY DUTY ONBOARD AIR SYSTEM PART NO

HEAVY DUTY ONBOARD AIR SYSTEM PART NO IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this manual before installing and using this product. SAVE THIS MANUAL FOR FUTURE REFERENCE

More information

SECOND GENERATION Use this guide with unit serial number prefix beginning with BWF using Terra Power separator.

SECOND GENERATION Use this guide with unit serial number prefix beginning with BWF using Terra Power separator. Technical Information and Diagnostic Guide for SECOND GENERATION Use this guide with unit serial number prefix beginning with BWF using Terra Power separator. This guide will assist you in becoming more

More information

Installation and Service Manual for SRC25, SRC252, SRC50, SRC502, SRC75, SRC752

Installation and Service Manual for SRC25, SRC252, SRC50, SRC502, SRC75, SRC752 Rocking Piston Compressors Installation and Service Manual for SRC25, SRC252, SRC50, SRC502, SRC75, SRC752 Thank you for purchasing the Stratus SRC series rocking piston compressor. This instruction manual

More information

MODEL NUMBER: MEDIUM DUTY ONBOARD AIR SYSTEM

MODEL NUMBER: MEDIUM DUTY ONBOARD AIR SYSTEM MODEL NUMBER: 10003 MEDIUM DUTY ONBOARD AIR SYSTEM IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this manual before installing and using

More information

HEAVY DUTY ONBOARD AIR SYSTEM

HEAVY DUTY ONBOARD AIR SYSTEM HEAVY DUTY ONBOARD AIR SYSTEM PART NO. 10005 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this manual before installing and using this

More information

Instruction Sheet. 1/2 HP Portable Electric Pumps SAFETY FIRST. L2062 Rev. F 02/ IMPORTANT RECEIVING INSTRUCTIONS 2.

Instruction Sheet. 1/2 HP Portable Electric Pumps SAFETY FIRST. L2062 Rev. F 02/ IMPORTANT RECEIVING INSTRUCTIONS 2. Instruction Sheet 1/2 HP Portable Electric Pumps L2062 Rev. F 02/12 Index: English:...................................... 1-7 Français:.................................... 8-14 Deutsch:...................................

More information

Maintenance Manual WITH ONE SUBMERSIBLE PUMP. Automated Fuel Maintenance System FTI-5A FUEL TECHNOLOGIES INTERNATIONAL LLC

Maintenance Manual WITH ONE SUBMERSIBLE PUMP. Automated Fuel Maintenance System FTI-5A FUEL TECHNOLOGIES INTERNATIONAL LLC Maintenance Manual WITH ONE SUBMERSIBLE PUMP Automated Fuel Maintenance System FTI-5A FUEL TECHNOLOGIES INTERNATIONAL LLC Replacement Manuals Available on Website: www.fueltechnologiesinternational.com

More information

HALLMARK INDUSTRIES INC

HALLMARK INDUSTRIES INC Performance Part No. HP. CONVERTIBLE JET PUMP USER S MANUAL GPH of Water @ Total Discharge Pressure of 40 psi Max. Pressure Max suction (shallow well) Max Suction (deep well) Max GPM (@0 head) Max Discharge

More information

VT Oil-Free Range Operators Handbook

VT Oil-Free Range Operators Handbook VT Oil-Free Range Operators Handbook Covering Models:- VT7 / VT7D VT10 / VT10D VT200 / VT200D VT300 / VT300D VT00 / VT00D BAMBI AIR COMPRESSORS LTD 12 Thimble Mill Lane Heartlands Birmingham B7 HT Tel:

More information

40041 Heavy Duty ADA System with Booster Bracket for JK Heavy Duty ADA System with Booster Bracket for 2012 to Current JK

40041 Heavy Duty ADA System with Booster Bracket for JK Heavy Duty ADA System with Booster Bracket for 2012 to Current JK 40041 Heavy Duty ADA System with Booster Bracket for 2007-2011 JK 40044 Heavy Duty ADA System with Booster Bracket for 2012 to Current JK 40049 Heavy Duty ADA System Universal for all vehicles (booster

More information

Electric Reciprocating Compressor. Installation Guide

Electric Reciprocating Compressor. Installation Guide Electric Reciprocating Compressor Installation Guide Air compressors should only be installed trained installation personnel call 800-531-9656 to find a local trained. Warning: Read all installation steps,

More information

Enclosed Electric Rotary Screw Compressor Installation Guide

Enclosed Electric Rotary Screw Compressor Installation Guide Enclosed Electric Rotary Screw Compressor Installation Guide Air compressors should only be installed trained installation personnel call 800-531-9656 to find a local trained. Warning: Read all installation

More information

6L Oil-less Air Compressor 53103

6L Oil-less Air Compressor 53103 6L Oil-less Air Compressor 53103 Operating Instructions Please read and save these instructions before attempting to assemble, install, operate or maintain the product. Protect yourself and others by observing

More information

Dental Compressed Air System

Dental Compressed Air System Part No. CI-101A (Rev B) Dental Compressed Air System Installation/User Guide Operation & Maintenance Manual Registered Trademark/ Trademark of JUN-AIR Inc. Copyright 2011 JUN-AIR Manufacturing Inc. All

More information

MEDIUM DUTY ONBOARD AIR SYSTEM

MEDIUM DUTY ONBOARD AIR SYSTEM MEDIUM DUTY ONBOARD AIR SYSTEM PART NO. 10003 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of this manual before installing and using this

More information

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST BELT TYPE IMPORTANT PLEASE MAKE CERTAIN THAT THE PERSON WHO IS TO USE THIS EQUIPMENT CAREFULLY READS AND UNDERSTANDS THESE INSTRUCTIONS BEFORE STARTING

More information

SERVICE MANUAL (INTERNATIONAL)

SERVICE MANUAL (INTERNATIONAL) SERVICE MANUAL (INTERNATIONAL) IMPINGER CONVEYOR OVENS MODEL 1421-000-E, 1454, 1455 WITH PUSH BUTTON CONTROLS Lincoln Foodservice Products, LLC 1111 North Hadley Road Fort Wayne, Indiana 46804 United States

More information

LABORATORY ZERO AIR GENERATOR MODEL N-GC1500 USER S MANUAL

LABORATORY ZERO AIR GENERATOR MODEL N-GC1500 USER S MANUAL LABORATORY ZERO AIR GENERATOR MODEL N-GC1500 USER S MANUAL Content 1. Introduction...2 2. Important safety instruction...3 3. System component...4 4. Engineering design overview...5 5. Installation...6

More information

DECADE 1021 SOLO CHAIR

DECADE 1021 SOLO CHAIR Owner's Guide DECADE 1021 SOLO CHAIR 85-2626-00 Warranty A-dec warrants its products against defects in material or workmanship for one year from time of delivery (except for handpieces which have a warranty

More information

UltraClave Automatic Sterilizers

UltraClave Automatic Sterilizers UltraClave Automatic Sterilizers Model Numbers: M9 M9D M11 M11D -020 thru -022-020 & -022-020 thru -022-020 & -022 Serial Number Prefixes: RN, RP, RR, V RW, RX, V RS, RT, RV, V RY, RZ, V Service and Parts

More information

Maintenance Manual. Automated. Fuel Maintenance System FTI-5A. FUEL TECHNOLOGIES INTERNATIONAL LLC

Maintenance Manual. Automated. Fuel Maintenance System FTI-5A. FUEL TECHNOLOGIES INTERNATIONAL LLC Maintenance Manual Automated Fuel Maintenance System FTI-5A FUEL TECHNOLOGIES INTERNATIONAL LLC www.fueltechnologiesinternational.com 03/01/2011 - Fuel Technologies FTI-5A Maintenance Section FTI - Fuel

More information

IMPORTANT SAFETY INSTRUCTIONS

IMPORTANT SAFETY INSTRUCTIONS OWNER S MANUAL FLO-MASTER XP2 SERIES PUMPS IMPORTANT SAFETY INSTRUCTIONS When installing and using this electrical equipment, basic safety precautions should always be followed, including the following:

More information

348002K/348012K Manifold Block Style Service Manual 12/2000

348002K/348012K Manifold Block Style Service Manual 12/2000 348002K/348012K Manifold Block Style Service Manual 12/2000 Service Manual 348002K/348012K Manifold Block Style Recovery/Recycling/Recharging Unit For R-12 or R-134a Only TABLE OF CONTENTS: Theory of Operation

More information

PAGE 1. TES Operation & Testing Guidelines: Tes Trouble shooting

PAGE 1. TES Operation & Testing Guidelines: Tes Trouble shooting PAGE 1 This document outlines questions to ask and components to check during TES troubleshooting. More detailed troubleshooting procedures are available in the TES Troubleshooting Guide. 1. Flow Light

More information

Flow Max Fluid Pump. by Duraself OEM INSTALLATION MANUAL

Flow Max Fluid Pump. by Duraself OEM INSTALLATION MANUAL Flow Max Fluid Pump by Duraself OEM INSTALLATION MANUAL TABLE OF CONTENTS System and Safety Information 2 Resources Required 3 Installation 3 General Installation Instructions 3 Mounting Instructions 3

More information

Maintenance Manual. Automated Fuel Maintenance System FTI-5A SINGLE TANK FUEL TECHNOLOGIES INTERNATIONAL

Maintenance Manual. Automated Fuel Maintenance System FTI-5A SINGLE TANK FUEL TECHNOLOGIES INTERNATIONAL Maintenance Manual Automated Fuel Maintenance System FTI-5A SINGLE TANK FUEL TECHNOLOGIES INTERNATIONAL 05/01/2016 Rev A Fuel Technologies FTI-5A Single Tank Maintenance Section FTI - Fuel Maintenance

More information

OWNER S MANUAL EVOLUTION 3500, 4500, 5500, & 8500 SERIES PUMPS

OWNER S MANUAL EVOLUTION 3500, 4500, 5500, & 8500 SERIES PUMPS OWNER S MANUAL EVOLUTION 3500, 4500, 5500, & 8500 SERIES PUMPS IMPORTANT SAFETY INSTRUCTIONS When installing and using this electrical equipment, basic safety precautions should always be followed, including

More information

Troubleshooting 3Z8 038 Rev B

Troubleshooting 3Z8 038 Rev B Troubleshooting 3Z8 038 Rev B INSTRUCTIONS WARNING INJECTION HAZARD This form is only a quick reference for troubleshooting Graco sprayers. To reduce the risk of serious injury, including fluid injection,

More information

Installation Operation Maintenance Troubleshooting Version 1, Oct/09. Palm Beach

Installation Operation Maintenance Troubleshooting Version 1, Oct/09. Palm Beach Installation Operation Maintenance Troubleshooting Version 1, Oct/09 1655 Palm Beach Congratulations! All of us at Summit Dental Systems want you to know that your 1655 Palm Beach Cuspidor has been built

More information

Flow Max. by Duraself. Flow Max by Duraself. Fluid Pump. Installation and Owner s Manual (For Aftermarket Applications)

Flow Max. by Duraself. Flow Max by Duraself. Fluid Pump. Installation and Owner s Manual (For Aftermarket Applications) Flow Max Installation and Owner s Manual (For Aftermarket Applications) Table of Contents Introduction... 2 Parts List... 2 Resources Required... 3 Installation... 3 General Installation Instructions...

More information

PRODUCT NUMBERING SYSTEM SERIES PHASE. 1: Single Phase 3: Three Phase

PRODUCT NUMBERING SYSTEM SERIES PHASE. 1: Single Phase 3: Three Phase TABLE OF CONTENTS Product Numbering System and Specifications... Safety... Receiving and Inspection... Installation... Electrical...6 Start-up...7 INTRODUCTION The compressor you have purchased is a combination

More information

RESPA -SD or Gideon HVAC Fresh Air Filtration System Installation Guidelines for John Deere 450J, 550J, 650J, 700J, 750J, 850J Crawlers

RESPA -SD or Gideon HVAC Fresh Air Filtration System Installation Guidelines for John Deere 450J, 550J, 650J, 700J, 750J, 850J Crawlers RESPA -SD or Gideon HVAC Fresh Air Filtration System Installation Guidelines for John Deere 450J, 550J, 650J, 700J, 750J, 850J Crawlers Cab Air Quality Systems Filter Options: HEPA-level (EU P3) MERV 16

More information

SYSTEM SAVER 318 AIR COMPRESSOR FOR MACK E-TECH AND ASET ENGINES MAINTENANCE MANUAL

SYSTEM SAVER 318 AIR COMPRESSOR FOR MACK E-TECH AND ASET ENGINES MAINTENANCE MANUAL SYSTEM SAVER 318 AIR COMPRESSOR FOR MACK E-TECH AND ASET ENGINES MAINTENANCE MANUAL NON-THROUGH DRIVE THROUGH DRIVE Service Notes About This Manual This manual provides service and repair procedures for

More information

WARNING This manual should only be used by a qualified Service Technician. FinishPro 390/395 Airless/Air-Assisted Sprayer Repair Electrical Manual

WARNING This manual should only be used by a qualified Service Technician. FinishPro 390/395 Airless/Air-Assisted Sprayer Repair Electrical Manual FinishPro 390/395 Airless/AirAssisted Sprayer Repair Electrical Manual First choice when quality counts. Rev. B 10/11 /07 FinishPro 395 3.00 ti9026a Red Yel Yel Air Hose Connection FinishPro 390 Exhaust

More information

6 Litre Oil-Less Air Compressor

6 Litre Oil-Less Air Compressor Operator s Manual 6 Litre Oil-Less Air Compressor WARNING! Before using this appliance, read the Operator s manual and follow all its safety rules and instructions. SPECIFICATION HWKAC1 1.1 kw / 1.5 HP

More information

SHURFLO HIGH FLOW SYSTEM II 10.0 GPM MARINE FRESH WATER SYSTEM PRODUCT TECHNICAL DATA SHEET

SHURFLO HIGH FLOW SYSTEM II 10.0 GPM MARINE FRESH WATER SYSTEM PRODUCT TECHNICAL DATA SHEET SHURFLO HIGH FLOW SYSTEM II 10.0 GPM MARINE FRESH WATER SYSTEM PRODUCT TECHNICAL DATA SHEET OEM: AFTERMARKET: 4558-153-E75 4558-153-E75 APPLICATION: Multi-fixture Marine fresh water system installation.

More information

SPECIFICATIONS Horsepower: 1.5 HP Running Maximum PSI: 125 PSI Tank Capacity: 15 Gallons CFM: 6 40 PSI 5 90 PSI

SPECIFICATIONS Horsepower: 1.5 HP Running Maximum PSI: 125 PSI Tank Capacity: 15 Gallons CFM: 6 40 PSI 5 90 PSI 15 GALLON AIR COMPRESSOR Model: 7678 DO NOT RETURN TO STORE Please call 800-348-5004 for parts and service CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials

More information

INSTALLATION INSTRUCTIONS For Model 855 Wolo Express TM

INSTALLATION INSTRUCTIONS For Model 855 Wolo Express TM INSTALLATION INSTRUCTIONS For Model 855 Wolo Express TM 12-Volt Train Horn Your purchase of a WOLO EXPRESS TRAIN HORN is the choice that will complement your vehicle. Wolo s products are manufactured with

More information

AA-75CF Portable Compressor NSN:

AA-75CF Portable Compressor NSN: AA-CF Portable Compressor NSN: 0-0--0 COMMAND l AIR SERVICE MANUAL & PARTS LIST TABLE OF CONTENTS Maintenance and Servicing Instructions... Cleaning and Lubrication................ Performance Verification................

More information

Installation Instructions

Installation Instructions Instructions Aerada 900 Series Futura Faucet With Battery (BIR) Infrared Control S53-284 4" Centerset/Centershank S53-289 Centerset/Centershank w/plate Table of Contents Pre- Information............2 900

More information

Pump Installation and Service Manual HRS Hydromatic Retractable System

Pump Installation and Service Manual HRS Hydromatic Retractable System Pump Installation and Service Manual HRS Hydromatic Retractable System NOTE! To the installer: Please make sure you provide this manual to the owner of the pumping equipment or to the responsible party

More information

ASVC2 OWNER S MANUAL. Vane Cutter PART # MACHINERY DIVISION

ASVC2 OWNER S MANUAL. Vane Cutter PART # MACHINERY DIVISION A S ASVC2 Vane Cutter PART # 43013 MACHINERY DIVISION OWNER S MANUAL -1- INTRODUCTION The ASVC2 Hollow Vane Cutting Machine was designed utilizing the best current technology to provide greater reliability.

More information

Rev: Flow Max 115V AC Fluid Pump

Rev: Flow Max 115V AC Fluid Pump Flow Max 115V AC Fluid Pump by Duraself Installation and Owner s Manual (For Aftermarket Applications) Table of Contents Introduction... 2 Parts List... 2 Resources Required... 3 Installation... 3 General

More information

Model and Series 115 VAC INDUSTRIAL DIAPHRAGM PUMPS. PumpAgents.com - buy pumps and parts online INDUSTRIAL DIAPHRAGM PUMPS

Model and Series 115 VAC INDUSTRIAL DIAPHRAGM PUMPS. PumpAgents.com - buy pumps and parts online INDUSTRIAL DIAPHRAGM PUMPS Model 31801 and 31800 Series 115 VAC INDUSTRIAL DIAPHRAGM PUMPS INDUSTRIAL DIAPHRAGM PUMPS FEATURES Run Dry Ability Self-Priming Thermal Overload Protected Motor Easy Installation Low Amp Draw Compact

More information

Source Capture Air Purification System

Source Capture Air Purification System Source Capture Air Purification System HA-SCP-G3 Owner s Manual Table of Contents HealthyAir HA-IFM-1111 Filter 1 Important Safety Instructions 2 Technical Specifications 3 Packaging Reference 4 Packing

More information

FHC-1D user manual. Features

FHC-1D user manual. Features Phason The Fan and Heater Control (FHC-1D) automatically controls the temperature in a room by adjusting the speed of variable speed fans and controlling a heater interlock. When the temperature is at

More information

STOP CITY PRESSURE BOOSTER PUMP INSTRUCTION MANUAL

STOP CITY PRESSURE BOOSTER PUMP INSTRUCTION MANUAL CITY PRESSURE BOOSTER PUMP INSTRUCTION MANUAL MODEL #VP05, VP10 C US NSF/ANSI 372 255405 For loose, missing or damaged parts, or if the unit does not seem to be operating properly, please call before returning

More information

INSTALLATION & OPERATING INSTRUCTIONS

INSTALLATION & OPERATING INSTRUCTIONS INSTALLATION & OPERATING INSTRUCTIONS WARNING RISK OF ELECTRIC SHOCK. CONNECT ONLY TO A CIRCUIT PROTECTED BY A GROUND-FAULT CIRCUIT-INTERRUPTER. THE UNIT SHOULD BE INSTALLED BY A QUALIFIED SERVICE REPRESENTATIVE.

More information

OXYGEN GENERATING SYSTEMS INTL. Oxygen System Start-up, Overview and Maintenance Guide

OXYGEN GENERATING SYSTEMS INTL. Oxygen System Start-up, Overview and Maintenance Guide OXYGEN GENERATING SYSTEMS INTL. Oxygen System Start-up, Overview and Maintenance Guide OXYGEN GENERATOR SYSTEM START-UP STEP 1 Walk around the skid and inspect for damage. Then check that all electrical

More information

Technical Information and Diagnostic Guide

Technical Information and Diagnostic Guide Technical Information and Diagnostic Guide This guide will assist you in becoming more familiar with the working components of the NITE System and the proper steps and procedures to completely diagnose

More information

Dental Dry Suction. Installation/User Guide

Dental Dry Suction. Installation/User Guide Dental Dry Suction System Installation/User Guide Operation & Maintenance Manual Registered Trademark/ EcoPowerVac Trademark of Omxie Corp. Copyright 2017. All Rights Reserved. WWW.SMTMAX.com. 1 P a g

More information

Portable Oil Free Silent Series Compressor Operating Instructions

Portable Oil Free Silent Series Compressor Operating Instructions Portable Oil Free Silent Series Compressor Operating Instructions NOTICE Carefully read this instruction manual before attempting to operate this compressor. MODEL # SERIAL # 1-800-551-2406 www.eaglecompressor.com

More information

CLEAN ROOM DEVICES, LLC "WHERE TUBING AND FITTINGS COME TOGETHER"

CLEAN ROOM DEVICES, LLC WHERE TUBING AND FITTINGS COME TOGETHER CLEAN ROOM DEVICES, LLC "WHERE TUBING AND FITTINGS COME TOGETHER" CRD600 Automatic Fitting Inserter OPERATIONS MANUAL VERSION 2.1 LAST EDITED 7.25.14 DOCUMENT NUMBER 001 cleanroomdevices.com 1 Table of

More information

SPC-PANEL Simplex, Single Phase Pump Control Panel

SPC-PANEL Simplex, Single Phase Pump Control Panel Pump Installation and Service Manual SPC-PANEL Simplex, Single Phase Pump Control Panel Pump Controls for 2 HP Grinder Pumps NOTE! To the installer: Please make sure you provide this manual to the owner

More information

Val-Matic Air / Oil Hydraulic Panel Pump Control System. Operation, Maintenance and Installation Manual

Val-Matic Air / Oil Hydraulic Panel Pump Control System. Operation, Maintenance and Installation Manual Manual No. 5AOP-OM1-2 Val-Matic Air / Oil Hydraulic Panel Pump Control System Operation, Maintenance and Installation Manual INTRODUCTION... 1 RECEIVING AND STORAGE... 1 DESCRIPTION OF OPERATION... 1 INSTALLATION...

More information

Application Engineering

Application Engineering Application Engineering February, 2009 Copeland Digital Compressor Controller Introduction The Digital Compressor Controller is the electronics interface between the Copeland Scroll Digital Compressor

More information

HVF110, 210, 310, 410HD

HVF110, 210, 310, 410HD 342 N. Co. Rd. 400 East Valparaiso, IN 46383 219-464-8818 Fax 219-462-7985 www.heatwagon.com Installation and Maintenance Manual Please retain this manual for future reference. HVF110, 210, 310, 410HD

More information

PNEUMATICS CONTENTS HST-TAMPER

PNEUMATICS CONTENTS HST-TAMPER CONTENTS Tamper Pneumatic Component Locations...2 Manually Releasing Air Brakes... 3 Releasing Air Pressure... 4 Removing Water from Air Tanks... 4 Adjusting Air Compressor Governor... 4 Check/Replace

More information

1416 Cabinet Unit. Installation Operation Maintenance Troubleshooting Version Dec/17

1416 Cabinet Unit. Installation Operation Maintenance Troubleshooting Version Dec/17 Installation Operation Maintenance Troubleshooting Version Dec/17 1416 Cabinet Unit Page 1 of 16 Table of Contents Important Information... 3 Weight & Dimensions... 4 Operator s Instructions... 5 Flow

More information

XENON POWER SUPPLY 4000 Watt Gladiator IV

XENON POWER SUPPLY 4000 Watt Gladiator IV XENON POWER SUPPLY 4000 Watt Gladiator IV 220 Volt Equipment Type 62-00049 Rev. February 2003 STRONG INTERNATIONAL a division of Ballantyne of Omaha, Inc. 4350 McKinley Street Omaha, Nebraska 68112 USA

More information

Third Generation NITE Phoenix

Third Generation NITE Phoenix Technical Information and Diagnostic Guide for Third Generation NITE Phoenix Use this guide with unit serial number prefix beginning with BYC, CAI built after 2-10-2012 and CCA, CDJ, CIA units after 6/25/2012

More information

XCITE Owner s Manual. Reso-not TM Damping System XCITE 1502C HYDRAULIC POWER SUPPLY

XCITE Owner s Manual. Reso-not TM Damping System XCITE 1502C HYDRAULIC POWER SUPPLY Reso-not TM Damping System XCITE Owner s Manual 1502C HYDRAULIC POWER SUPPLY Xcite Systems Corporation 675 Cincinnati RDS Batavia - 1 Pike Cincinnati, Ohio 45245 Tel: (239) 980-9093 Fax: (239) 985-0074

More information

IMAGING UTILITY ROOM ACCESSORIES

IMAGING UTILITY ROOM ACCESSORIES IMAGING UTILITY ROOM ACCESSORIES PARTS CATALOG 2014 Equipped for Life with Air Techniques Dear Valued Customer: At Air Techniques, we are focused on high quality products, innovative new technology and

More information

Wiring diagrams on page 29 are for reference only. For detailed vehicle wiring refer to Navistar documents.

Wiring diagrams on page 29 are for reference only. For detailed vehicle wiring refer to Navistar documents. 1 10/2014 REV 7 !!Attention!! Before performing diagnostics: Wiring diagrams on page 29 are for reference only. For detailed vehicle wiring refer to Navistar documents. Check for Fault Codes using the

More information

Users Manual Professional Series Direct Drive Pump.25-2 GPM Series Safety, Operating, Installation, and Maintenance Instructions

Users Manual Professional Series Direct Drive Pump.25-2 GPM Series Safety, Operating, Installation, and Maintenance Instructions Users Manual Professional Series Direct Drive Pump.25-2 GPM Series Safety, Operating, Installation, and Maintenance Instructions 600 S 56 th Street #9 Chandler, AZ 85226 Phone: 480-507-6478 Fax: 480-838-2232

More information

Positive Displacement Pump

Positive Displacement Pump www.conairgroup.com U S E R G U I D E UGC028-1105 Positive Displacement Pump Models PD 3. 5, 7.5, 10, 15 and 25 Corporate Office: 724.584.5500 l Instant Access 24/7 (Parts and Service): 800.458.1960 l

More information

SEWAGE PUMP MODEL # Zoeller is a registered trademark of Zoeller Co. All Rights Reserved. Español p. 14

SEWAGE PUMP MODEL # Zoeller is a registered trademark of Zoeller Co. All Rights Reserved. Español p. 14 SEWAGE PUMP Zoeller is a registered trademark of Zoeller Co. All Rights Reserved. MODEL #1261-0001 Español p. 14 ATTACH YOUR RECEIPT HERE Serial Number Purchase Date Questions, problems, missing parts?

More information

GRINDER PUMP MODEL # Zoeller is a registered trademark of Zoeller Co. All Rights Reserved. Español p. 13

GRINDER PUMP MODEL # Zoeller is a registered trademark of Zoeller Co. All Rights Reserved. Español p. 13 GRINDER PUMP Zoeller is a registered trademark of Zoeller Co. All Rights Reserved. MODEL #2701-0005 Español p. 13 ATTACH YOUR RECEIPT HERE Serial Number Purchase Date Questions, problems, missing parts?

More information

Troubleshooting Manual

Troubleshooting Manual Troubleshooting Manual NOTICE: DO NOT DISCARD THIS MANUAL Models: LEGACY42-IFT PHOENIX42-IFT 1 TABLE OF CONTENTS A. Normal Operation...3 B. Wiring Diagram...4 C. Troubleshooting IntelliFire Touch...5 D.

More information

4" ENVIRONMENTAL E-SERIES PUMPS OWNER'S MANUAL. DANGER warns about hazards that will cause. WARNING warns about hazards that can cause

4 ENVIRONMENTAL E-SERIES PUMPS OWNER'S MANUAL. DANGER warns about hazards that will cause. WARNING warns about hazards that can cause 4" ENVIRONMENTAL E-SERIES PUMPS OWNER'S MANUAL BEFORE INSTALLING PUMP, BE SURE TO READ THIS OWNER S MANUAL CAREFULLY. CAUTION Fill pump with water before starting or pump will be damaged. The motor on

More information

Read this entire manual before operation begins.

Read this entire manual before operation begins. Read this entire manual before operation begins. Record below the following information which is located on the serial number data plate. Serial No. Model No. Date of Installation Contents Specifications.............

More information

Service and Parts Manual. NO LONGER IN PRODUCTION Some service parts may not be available for this product. Otolaryngology Chair.

Service and Parts Manual. NO LONGER IN PRODUCTION Some service parts may not be available for this product. Otolaryngology Chair. thru 391-001 -002 Otolaryngology Chair Serial Number Prefixes: EN, PD & V Service and Parts Manual NO LONGER IN PRODUCTION Some service parts may not be available for this product. 391-001 thru -002 NOTE:

More information

SERVICE MANUAL (INTERNATIONAL)

SERVICE MANUAL (INTERNATIONAL) SERVICE MANUAL (INTERNATIONAL) IMPINGER CONVEYOR OVENS MODEL 1433-000-E, 1434-000-E, 1456, 1457 WITH PUSH BUTTON CONTROLS Lincoln Foodservice Products, LLC 1111 North Hadley Road Fort Wayne, Indiana 46804

More information

Reproduction or other use of this Manual, without the express written consent of Vulcan, is prohibited.

Reproduction or other use of this Manual, without the express written consent of Vulcan, is prohibited. SERVICE MANUAL ELECTRIC BRAISING PANS (30 & 40 GALLON) VE30 VE40 ML-126849 ML-126850 VE40 SHOWN - NOTICE - This Manual is prepared for the use of trained Vulcan Service Technicians and should not be used

More information

C T h e A d va n t a g e

C T h e A d va n t a g e C The Advantage TABLE OF CONTENTS Introduction...1 Product Numbering System...1 Safety...2 Receiving and Inspection...2 Installation...2 Electrical...3 Parts Identification...7 Lubrication...9 Start-up...10

More information

COOL DRAFT INSTRUCTIONS OWNER S MANUAL CONTENTS. Cool Draft MiniMister 14 Misting Fan CDMP1420BLU

COOL DRAFT INSTRUCTIONS OWNER S MANUAL CONTENTS. Cool Draft MiniMister 14 Misting Fan CDMP1420BLU COOL DRAFT BY INSTRUCTIONS OWNER S MANUAL! Keep for future reference. Common problems and solutions are listed within this booklet. IMPORTANT INFORMATION INSIDE: Read, understand, and follow all safety

More information

CIRCLE SEAL CONTROLS

CIRCLE SEAL CONTROLS CIRCLE SEAL CONTROLS ATKOMATIC SOLENOID VALVES INSTALLATION, MAINTENANCE, AND OPERATION INSTRUCTIONS 13000 SERIES Stainless, Normally Closed, Direct Lift, 3-Way Valve Installation Instructions WARNING:

More information

LESTRONIC II BATTERY CHARGER TAYLOR-DUNN MODEL TYPE 24LC25-8ET

LESTRONIC II BATTERY CHARGER TAYLOR-DUNN MODEL TYPE 24LC25-8ET LESTRONIC II BATTERY CHARGER TAYLOR-DUNN 79-301-10 MODEL 13110-32 TYPE 24LC25-8ET AC Supply: DC Output: Battery Capacity: Specifications 120 volts, 60 Hertz, single-phase 24 volts, 32 amps Use only on

More information

DISCONTINUED. Installation. Aerada 900 Series Futura Faucet. With Accu-Zone (AZ) Infrared Control

DISCONTINUED. Installation. Aerada 900 Series Futura Faucet. With Accu-Zone (AZ) Infrared Control Aerada 900 Series Futura Faucet With Accu-Zone (AZ) Infrared Control BRADLEY SC A S53-141 4" Centerset S53-148 4" Centerset, no Solenoid Valve S53-186 4" Centerset with 8" Trim Plate S53-285 Centershank

More information

PARTS & SERVICE MANUAL

PARTS & SERVICE MANUAL PARTS & SERVICE MANUAL Impinger Low Profile Advantage Digital Series (Electric) International Models MODELS: Please note that the model numbering system changed March 2007. The chart below shows the old

More information

FPS- 80. Manual. Reverso Pumps, Inc. Ph: (954)

FPS- 80. Manual. Reverso Pumps, Inc.   Ph: (954) FPS- 80 Manual Table of Contents System Overview... 1 Control Panel Overview... 2 Technical Specifications... 3 Electrical and Installation... Tank Diagrams... 5 Initial Setup... 6 Digital Timer Instructions:

More information

Page 1 of 29 Section 04-05: Suspension, Computer Controlled 1997 Town Car Workshop Manual DIAGNOSIS AND TESTING Procedure revision date: 05/16/2000 Suspension, Computer Controlled Inspection and Verification

More information

Compressor Installation Guide

Compressor Installation Guide Compressor Installation Guide C Series COMPRESSOR C1103 C1203 C2106 C2206 C3210 Copyright 2006 DCI. All Rights Reserved. 92094, Rev. D, 08/13 1 Compressor Model Specification C - Series Compressor Oil-Less

More information

SPECIFICATIONS: Tank Size: 80 gallons PUMP RPMs: 1050 CFM: 40PSI; 90 PSI Max Pressure: 150 PSI Thermal overload protection

SPECIFICATIONS: Tank Size: 80 gallons PUMP RPMs: 1050 CFM: 40PSI; 90 PSI Max Pressure: 150 PSI Thermal overload protection 5HP 80 GALLON TWO STAGE COMPRESSOR Models: 51866, 51870 CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials such as wood, paint, metal, concrete, cement,

More information

6. Document tasks performed during visit and report any observations to the appropriate

6. Document tasks performed during visit and report any observations to the appropriate Chiller Annual Maintenance RFP Attachment B Services Required Tasking: Vibration Analysis 1. All work shall be performed in accordance with safety policies 2. Check with appropriate customer representative

More information

Value Super Filter Max Installation Manual

Value Super Filter Max Installation Manual Value Super Filter Max Installation Manual Barrie, Ontario, Canada, L4N 4Y8 www.excaliburwater.com EXCALIBUR VALUE SUPER FILTER MAX INSTALLATION MANUAL INSTALLATION PROCEDURES: The Value Super Filter Max

More information

3 FT CORD Models , , , FT CORD W/GFCI Models , , ,

3 FT CORD Models , , , FT CORD W/GFCI Models , , , POWERLINE XP II Swimming Pool Pump (Light Oak color model) With Automatic Timer 3 FT CORD Models 0-1395-226, 0-1396-226, 0-1397-226, 0-1398-226 25 FT CORD W/GFCI Models 0-1395-220, 0-1396-220, 0-1397-220,

More information

Model G14SD Vapor Management System

Model G14SD Vapor Management System Instruction Manual Model G14SD Vapor Management System (For Models: DSD and SSD 110/115v) Manufactured in USA for Medivators Reprocessing Systems by: PCI Medical Inc. 12 Bridge Street, PO Box 188 Deep

More information

ESE Series Cast Iron Sewage Pumps

ESE Series Cast Iron Sewage Pumps Owner s Manual ESE Series Cast Iron Sewage Pumps TABLE OF CONTENTS General Safety.................... 2 Specifications..................... 3 Installation.................... 4 & 5 Troubleshooting...................

More information