INSTALLATION OF OPTIC 3 INJECTOR SYSTEM ONTO AN SHIMADZU GC-2010 GAS CHROMATOGRAPH.

Size: px
Start display at page:

Download "INSTALLATION OF OPTIC 3 INJECTOR SYSTEM ONTO AN SHIMADZU GC-2010 GAS CHROMATOGRAPH."

Transcription

1 INSTALLATION OF OPTIC 3 INJECTOR SYSTEM ONTO AN SHIMADZU GC-2010 GAS CHROMATOGRAPH. Please read the manual first before installation. ATAS GL International B.V. De Run LS Veldhoven Tel: +31 (0) Fax: +31 (0) Info@atasgl.com Page : 1 of 13

2 Unpacking the installation kit Check for the following parts in the installation kit Description Quantity Part Number Cable for OPTIC 3 Shimatzu H Mounting plate 1 H Spacer with internal thread, 3 M4x30x9 25 pin male sund connector 1 Page : 2 of 13

3 Pre-installation P1 Gases P1.1 Helium supply: % purity for the carrier gas on OPTIC 3. The carrier supply should be made available using 1/8 o.d. copper tubing. The maximum input pressure for the OPTIC 3 is 700 kpa (100 PSI). P1.2 A compressed air supply will be required for cooling the OPTIC 3 injector (or for driving the liquid CO 2 cooling valve when a liquid CO 2 cooling kit, Pt.No.H has been purchased). The compressed air supply should be free of oil, water mist and particulate to ensure reliable operation of the cooling system. The supply should be made available using 1/8 o.d. tubing. The operating pressure is 500 kpa (72 PSI) and the maximum inletpressure is 700 kpa (100 PSI). P1.3 When an OPTIC 3 is to be used with a liquid CO 2 sub ambient kit, please refer to the CO 2 kit installation manual. P2 Electrical Supply One mains supply socket will be required. The OPTIC 3 has a universal electronic power supply and can be used with 110 V 60 Hz and 230 V 50Hz. P3 Mounting Considerations P3.1 OPTIC 3 Injector Positioning. The has two injector positions available. The OPTIC 3 injector will fit into both the injector positions. It is advisable with a DTD unit to install the injector in the front position. P3.2 Control Module Positioning. The OPTIC 3 control module is usually placed on the left hand side of a Shimadzu GC The dimensions of the control module are 43cm x 15cm x 43cm (h x w x d). P4 GC and Autosampler I/O Requirements P4.1 If the OPTIC 3 is being mounted on a stand-alone chromatograph the Cable for OPTIC 3 - is required. (H400013) P4.2 If the installation is with a CombiPal, the Cable for OPTIC 3 - (H400013) is required and an extra cable between CombiPal and OPTIC 3 is required. (H400001) P4.3 The combination of a Shimadzu autosampler and OPTIC 3 is not supported from ATAS GL at the moment. Page : 3 of 13

4 1 Safety Before removing any cover from the gas chromatograph Disconnect the mains supply Disconnect or de-pressurise any gas connections Allow all ovens and heated zones to cool to a safe temperature Read the relevant sections of the gas chromatograph manual 2 Removal of Autosampler and/or MSD 2.1 If fitted, remove the autosampler and the autosampler mounting plate from the top the GC oven, referring to the autosampler and/or GC manuals if required. 2.2 If fitted, the MSD may be removed if it obstructs the installation, refer to the GC and/or manuals for information. (Usually no necessary) 3. Removing the inlet, pneumatics, and manifold When no free position is available one of the injectors should be removed. If there is a free position available for the OPTIC 3, skip section Turn off the oven and the inlet and allow them to cool. WARNING Turn off the oven and the inlet and allow them to cool. Turn off all flows at the initial gas supply. Then turn off the main power switch and unplug the power cord. 3.2 Remove the top cover, left side panel and the pneumatics chassis cover. 3.3 Remove the column. 3.4 Loosen the two screws, which secure the inlet to the oven, and lift the inlet out of the oven. 3.5 Remove the heater and temperature sensor connector. 3.6 Remover the gasline blok attached at the regulator and remove the complete injector. 3.7 In order to prevent the G.C. reporting an electronic fault condition on the injector removed above, a replacement plug should be placed. 4 Placing the GC thermocouple Page : 4 of 13

5 In order to have an accurate constant column flow, it is necessary to know the GC column oven temperature. This is measured with a thermocouple supplied with the OPTIC Find a hole in the GC oven top and guide the thermocouple trough the wall inside the GC oven. Or place the thermocouple near the injector and guide it inside the GC oven. 4.2 Fix the thermocouple end some where in the GC oven so it s measuring the oven temperature in a good way. Take care that the end of the thermocouple is not in contact with metal. 4.3 Route the thermocouple cable to the back of the host chromatograph. 4.4 Connect the thermocouple plug to the thermocouple socket on the rear panel of the OPTIC 3 control module. Note that it is important to ensure that the correct connector is used. Warning: Do not use the connector for the Injector thermocouple. 5 Mounting the OPTIC 3 Injector 5.1 Place the insulation material supplied with the installation kit in the hole where the injector should be installed. 5.2 Mount the installation bracket and place the tree 30mm spacers. 5.3 Place the injector in the position prepared and fix it using the 3 M4x10 screws supplied with the injector kit. 6 Gas Control Connection to the Injector 6.1 The pipe work from the OPTIC 3 injector will be connected to the ports on the rear of the control module. 6.2 The injector top should be located and the pipe work uncoiled. Re-fit the injector top. 6.3 Form the pipe work neatly on the oven top and route the pipe work such that it may be connected to the ports on the rear of the control module. 6.4 The carrier out, split, pressure and septum purge ports on the rear of the control module have been fitted with appropriate reducing ferrules to enable direct connection to 1/16 o.d. pipe work. 6.5 Connect the split line from the OPTIC 3 injector (with an Split identifier) to the Control module Split port using the PTFE reducing ferrule supplied with the injector. Page : 5 of 13

6 Do not push the split line to much inside the ferrule so it is touching the metal of the coupling. Make sure there is an electrical insulation between the control module and the OPTIC 3 injector split line. 6.6 Connect the carrier line (with a Carrier identifier) to the Carrier Out port on the control module. 6.7 Connect the septum purge line (with Septum Purge identifier) to the Aux Gas In / Purge In port on the control module. As the standard, there is a fixed restrictor installed allowing a few ml/min septum purge flow. 7 Cooling Air 7.1 Connect the 1/8 o.d. Nylon pipe supplied with the injector to the To Injector port on the cooling solenoid on the control module rear panel. 7.2 Route the cooling air line to the injector. It is suggested that a similar route to the carrier, pressure, split and purge lines is chosen Fit the cooling tube to the cooling pipe. Take care not to brake the cooling pipe on the welding. 7.4 The cooling pipe should be positioned such that the outlet projects down into the cup of the OPTIC 3 injector and is approximately level with the injector central tube when the injector is viewed from above. Ensure that the end of the cooling tube does not come into contact with the central tube of the injector when the cooling pipe is secured in place. The cooling air should not blow direct on the thermocouple. 7.5 The cooling air line can be cut to length using a sharp knife to improve the neatness of the installation before the outlet end of the Nylon line is pushed over the inlet end of the cooling pipe. A secure connection will be ensured if the nylon pipe overlaps the cooling pipe by mm. 9 Connections of Injector Power and Thermocouple Cables 9.1 Route the injector thermocouple cable under the back cover and of the host chromatograph. 9.2 Connect the thermocouple plug to the thermocouple socket on the rear panel of the OPTIC 3 control module. Note that it is important to ensure that the correct connector is used. Warning: Do not use the connector for the GC thermocouple. Page : 6 of 13

7 Figure Route the injector power cable (flat blue cable sleeved with a braided glass fibre sheath) to the injector and connect the connectors to the OPTIC 3 injector. See figure Take care that the heater cable clamps are not in contact with the top-boss clamp nut or the central tube. If this happens, there is a short cut and the injector will not perform well. 9.5 Secure connections between the power cable and the injector are necessary for reliable operation of the system. (Torque settings 5lb/inches [58grms/meter]) Connections that are insufficiently tight will cause increased contact resistance and may cause the connections to overheat. Note, however, that overtightening the power cable fixing nuts may cause damage to the threads on the mounting studs which may require replacement of the injector. 10 Connection of Helium and Cooling Air Supplies 10.1 Connect a supply of helium of a suitable quality at a pressure between 200 and 700 kpa to the Carrier IN port on the rear of the control module using the copper tube supplied. Note that the maximum pressure that can be set using the OPTIC 3 gas control will be approximately 5 psi less than the carrier gas supply pressure. As a result a low carrier supply pressure may limit the flexibility of the equipment Connect a supply of cooling air to the Air In port on the cooling solenoid on the rear of the control module at a pressure between 400 and 500 kpa. Lower pressure can be used if decreased cooling performance is acceptable. Page : 7 of 13

8 11 Logic Cable Connection 11.1 If the host chromatograph is either a stand alone unit or if fitted with an autosampler, connect the 15 way D connector of the logic cable (P/No. H400035) to the Chromatograph connector on the rear of the OPTIC 3 control module and connect the other end to the host chromatograph. Connect the wires according to the figure 2. White to 1 Brown to 2 Yellow to 3 Green to 4 12 Software installation (Evolution Workstation) 12.1 Refer to the OPTIC 3 user s guide appendix B. Figure Communication configuration 13.1 Refer to the OPTIC 3 user s guide appendix B. Page : 8 of 13

9 14 Configuration of the OPTIC The OPTIC 3 IO menu should be set as follows: OPTIC-GC interface: OPTIC Master Ready Out: Disabled Closed Ready In: Closed on Ready Autosampler output: Closed on Ready Evolution software Figure 3. I/O Settings. 15 Configuration settings of the GC 15.1 The should give a closed signal on ready. This can be configured in the following way: Press FUNC button go to 6, GC Config and press enter. Go to 9, other configurations and press enter. Set the Polarity in Ready to Close. See figure 4. Page : 9 of 13

10 Figure 4. Cable when AOC 20i is used Use the start and ready calbe from Shimadzu supplied with the sampler. Solder this cable to the 25 pin male subd connector supplied with this kit. Pin number (OPTIC) Colour autosampler cable 1 White 2 Black 6 Yellow 5 Red Connect the 25 pin connector on the back of the OPTIC control box, using the autosampler port. 16 Testing 16.1 Fit an O ring to a liner and place the liner in the injector, ensuring that the notch at the end of the liner is towards the bottom of the injector Fit the white PTFE insulating washer to the top of the injector base before securing the injector top to the injector base. Note that excessive tightening of the top boss clamp nut is unnecessary to form a seal between the top and bottom halves of the injector Refit the covers to the host chromatograph, routing the OPTIC 3 cables and pipework under the injector mouldings and covers if appropriate. Tywraps are provided to allow the cables and pipework to be clipped together to ensure a tidy installation. Refit the injector moulding and cover or autosampler mounting plate. Page : 10 of 13

11 16.4 Fit a capillary column to the injector and check the pipework for leaks. An electronic Leakseeker or 50/50 isopropanol/water should be used for leak checking. Under no circumstances should soap solution or other proprietary leak detector fluid be applied to couplings in the piping system Run a test sample of known composition to test the system chromatographically and to check logic connections between the OPTIC 3, host chromatograph and auto sampler, where fitted. Page : 11 of 13

12 17 Testing the OPTIC 3 communication in a system. (Ready Input & Ready Output) In principal there are two system possibilities, depending on the hardware The OPTIC 3 is Master: The auto sampler waits for the OPTIC to be Ready and the OPTIC waits for the GC to be Ready. Auto sampler starts the OPTIC; OPTIC starts the GC; GC starts data-acquisition. Testing: Turn the GC oven to OFF; set the data-acquisition software to ready; start a CombiPal method; Start an OPTIC method. The CombiPal should wait; the OPTIC should wait (Initialising on the display of the OPTIC). Turn the GC oven back to on. When the GC gets Ready, the OPTIC gets Ready and the CombiPal will start. Check the start sequence as described above When the original Shimadzu sampler is used there is the same sequence as with a CombiPal. To check the communication with the original Shimadzu sampler see The installation is now complete. If you have suggestions for this manual please mail to info@atasgl.com Page : 12 of 13

13 Shimadzu GC-2010 Bare Threads Interface cable for OPTIC 3 Cable P/No: H ,5 meter 4 core shielded OPTIC 3 15 pin female Chromatograph Shimadzu 2010 OPTIC 3 remote Bare threads chromatograph 15 pin female Yellow 3 TTL Ready out 1 Logic Ready Input Green 4 gnd 9 0V White 1 Start Start in 7 Relay Run Output Brown 2 gnd 14 Relay Run Output Page : 13 of 13

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory For the Agilent Intuvo 9000 Gas Chromatograph Installation Instructions Parts Supplied Table 1 Parts supplied with an Agilent G3969A

More information

Hot Injection Adapter Kit - Installation Guide

Hot Injection Adapter Kit - Installation Guide Hot Injection Adapter Kit - Installation Guide This guide provides the instruction for installing the Hot Injection Adapter kit (P/N 19050733) on the SSL/SSLBKF injector for the use with an external device

More information

Installing the 6850 Direct Valve Column Connector Kit

Installing the 6850 Direct Valve Column Connector Kit Use this kit to bypass the inlet and connect your 6850 gas or liquid sampling valve directly to a column. To use this kit, your 6850 GC must have a purged packed inlet. This kit contains: Installing the

More information

OPTIC-4. Multimode GC Inlet. Installation Guide OPTIC-4 Inlet CO 2 Cooling Option.

OPTIC-4. Multimode GC Inlet. Installation Guide OPTIC-4 Inlet CO 2 Cooling Option. OPTIC-4 Multimode GC Inlet Installation Guide OPTIC-4 Inlet CO 2 Cooling Option www.atasgl.com :D Disclaimer ATAS GL International B.V. makes no representations or warranties with respect to the contents

More information

Installing a Purged Packed Inlet

Installing a Purged Packed Inlet Installing a Purged Packed Inlet Agilent 6850 Gas Chromatograph This kit contains: Accessory G2651A Description Purged packed column inlet system 1 Top block insulation 1 Bottom block insulation 1 Insulation

More information

29 Installation Step 1. Unpacking the GC Cable diagrams Step 2. Placing the GC system on the benchtop Step 3. Turning the power on

29 Installation Step 1. Unpacking the GC Cable diagrams Step 2. Placing the GC system on the benchtop Step 3. Turning the power on 29 Installation Step 1. Unpacking the GC Step 2. Placing the GC system on the benchtop Step 3. Turning the power on Step 4. Connecting tubing to the gas supply tank Step 5. Attaching traps to the gas supply

More information

260 Solvent Vapor Exit Accessory

260 Solvent Vapor Exit Accessory Theory of operation The Solvent Vapor Exit (SVE) is a GC accessory for performing large volume injections with a Cool On-Column inlet (COC). At the start of the run, the SVE solenoid valve is open as the

More information

Installing a PTV Inlet

Installing a PTV Inlet Agilent 6850 Series II Network GC System Accessories G3345B (Septumless) and G3346B (Septum) These instructions are divided into two parts: Part 1 to prepare the Electronic Pressure Control ("EPC") module

More information

Installing a Cool On-Column Inlet

Installing a Cool On-Column Inlet Agilent 6850 Series II Network GC System Accessory G3344B This kit contains: Description Quantity Machine screws, M4 x 0.7 12 mm 6 Cable ties,.062.625 diameter 6 Ship kit* 1 T-20 Torx screw, M4 x 8 mm

More information

Installing a PTV Inlet

Installing a PTV Inlet Agilent 6850 Series II Network GC System Accessories G3345B (Septumless) and G3346B (Septum) There are kits for installing both septum and septumless PTV inlets. This document describes both installations.

More information

Installation Guide. Volatiles Interface for 6890 Gas Chromatograph Accessory G2319A

Installation Guide. Volatiles Interface for 6890 Gas Chromatograph Accessory G2319A Guide Volatiles Interface for 6890 Gas Chromatograph Accessory G2319A Agilent Technologies, Inc. 2001 All Rights Reserved. Reproduction, adaptation, or translation without permission is prohibited, except

More information

Agilent G1554A On-Column Capillary Inlet with EPC

Agilent G1554A On-Column Capillary Inlet with EPC Agilent G1554A On-Column Capillary Inlet with EPC 6890 Gas Chromatograph Installation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2002 No part of this manual may be reproduced in any

More information

Agilent G3510A/G3511A Multimode Inlet

Agilent G3510A/G3511A Multimode Inlet For the Agilent 7890A Gas Chromatograph Installation Instructions Parts Supplied Table Liquid Carbon Dioxide (CO 2 ) Kit G350A Description Liquid CO2 MMI inlet assembly (MMI Inlet weldment assembly, EPC

More information

Valve Box and Valve Accessories G4315A, G4316A, G4317A

Valve Box and Valve Accessories G4315A, G4316A, G4317A Valve Box and Valve Accessories G4315A, G4316A, G4317A Installation Guide The G4315A, G4316A, and G4317A accessories consist of separate kits depending on the custom configuration ordered. These kits may

More information

Installing the CO 2 Cryogenic Oven Cooling Kit

Installing the CO 2 Cryogenic Oven Cooling Kit Installing the CO 2 Cryogenic Oven Cooling Kit Agilent 6850 Series II Network GC System Accessory G2625B This kit contains: Description Chassis 1 Chassis Cover 1 Ship kit: 1 Cooling coil 1 Coil bracket

More information

Installing a Purged Packed (PP) Inlet

Installing a Purged Packed (PP) Inlet Installing a Purged Packed (PP) Inlet Agilent 6850 Gas Chromatograph Accessory G2651A These instructions are divided into two parts: Part 1 to prepare the Electronic Pressure Control ("EPC") module for

More information

Nickel Catalyst Tube Accessory 19205A

Nickel Catalyst Tube Accessory 19205A Installation and Operating Guide Nickel Catalyst Tube Accessory 19205A Agilent Technologies 2000 All Rights Reserved. Reproduction, adaptation, or translation without permission is prohibited, except as

More information

1310 Inlets. 1 of 36. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual

1310 Inlets. 1 of 36. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual 30 Inlets This section contains illustrated parts breakdowns for 6890 GC inlets and related components. Split/Splitless Inlet Split/Splitless Inlet Column Liners Split/Splitless Inlet EPC Pneumatic Module

More information

Deans Switch Microfluidics

Deans Switch Microfluidics TRACE 1300 and TRACE 1310 Gas Chromatographs Deans Switch Microfluidics Installation Guide 31709740 Revision A June 2014 2014 Thermo Fisher Scientific Inc. All rights reserved. TRACE 1300, and TRACE 1310

More information

ENGINE DISPLACEMENT 2000 NUMBER OF VALVES 16 ENGINE SET NUMBER

ENGINE DISPLACEMENT 2000 NUMBER OF VALVES 16 ENGINE SET NUMBER MANUFACTURER Subaru TYPE Forester ENGINE DISPLACEMENT 2000 NUMBER OF VALVES 16 ENGINE CODE / NUMBER FA20 DIT VEHICLE CATEGORIES M TRANSMISSION AT VERSION Direct LiquiMax-2.1 PETROL ECU MANUFACTURER / CODE

More information

BLOC U A-12G BLOC U C-12G

BLOC U A-12G BLOC U C-12G 3.3- INJECTOR BLOCKS AND ORIFICES 3.3.1- INJECTOR BLOCKS WITHOUT IGNITION AND WITHOUT FLAME SAFETY BLOC U-0-000-000-00-A-12G...3.3.1.1 BLOC U-0-000-000-00-C-12G...3.3.1.2 Related sections : Burners...3.2

More information

Installation Guide. Flame Ionization Detector on a 6890 GC. Accessories G1561A, G1562A, G1591A, G1598A

Installation Guide. Flame Ionization Detector on a 6890 GC. Accessories G1561A, G1562A, G1591A, G1598A Installation Guide Flame Ionization Detector on a 6890 GC Accessories G1561A, G1562A, G1591A, G1598A Agilent Technologies 2001 All Rights Reserved. Reproduction, adaptation, or translation without permission

More information

C FORD F250 / F L POWERSTROKE DIESEL WITH AUTOMATIC TRANSMISSIONS ONLY

C FORD F250 / F L POWERSTROKE DIESEL WITH AUTOMATIC TRANSMISSIONS ONLY EXHAUST BRAKES C40019 1999-2003 FORD F250 / F350 7.3L POWERSTROKE DIESEL WITH AUTOMATIC TRANSMISSIONS ONLY Getting Started Thank you and congratulations on your purchase of a Pacbrake exhaust retarder.

More information

Agilent G2855A Deans Switching System

Agilent G2855A Deans Switching System Agilent G2855A Deans Switching System Installation and Operation Agilent Technologies Notices Agilent Technologies, Inc. 2003 No part of this manual may be reproduced in any form or by any means (including

More information

Agilent SD-1 Purification System ml/min

Agilent SD-1 Purification System ml/min Agilent SD-1 Purification System - 200 ml/min Introduction Installation of the Capillary Kit This technical note shows you how to install the Stainless Steel Capillary Kit (SD- 1 Purification System -

More information

Nickel Catalyst Tube on a 6820 GC, Accessory G4318A

Nickel Catalyst Tube on a 6820 GC, Accessory G4318A Nickel Catalyst Tube on a 6820 GC, Accessory G4318A Installation Guide The Nickel Catalyst tube (NCT) is used with the 6820 Gas Chromatograph (GC) for trace analysis of CO and CO 2 using a Flame Ionization

More information

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors Model 8610C Gas Chromatograph Mounts up to Six Detectors and Five Injectors Ambient to 400 C Temperature Programmable Column Ovens Dimensions: 19 wide x 13.5 high x 14.5 deep Implement virtually any EPA

More information

Agilent 6850 Series II Network GC System G2630A Performance Specifications

Agilent 6850 Series II Network GC System G2630A Performance Specifications IET International Equipment Trading Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Agilent 6850 Series II Network GC

More information

Installation Guide. Purged Packed Inlet, Electronic Pneumatics Control on 6890 GC Accessory G1551A

Installation Guide. Purged Packed Inlet, Electronic Pneumatics Control on 6890 GC Accessory G1551A Installation Guide Purged Packed Inlet, Electronic Pneumatics Control on 6890 GC Accessory G1551A Agilent Technologies 2001 All Rights Reserved. Reproduction, adaptation, or translation without permission

More information

Installation Guide. Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A

Installation Guide. Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A Installation Guide Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A Agilent Technologies 2001 All Rights Reserved. Reproduction, adaptation, or translation without permission is prohibited,

More information

Agilent 7697A Headspace Sampler

Agilent 7697A Headspace Sampler Agilent 7697A Headspace Sampler Installation and First Startup Agilent Technologies Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any means (including

More information

Installation Guide. Nitrogen-Phosphorus Detector on 6890 GC Accessories G1575A, G1576A, G1594A, G1599A

Installation Guide. Nitrogen-Phosphorus Detector on 6890 GC Accessories G1575A, G1576A, G1594A, G1599A Installation Guide Nitrogen-Phosphorus Detector on 6890 GC Accessories G1575A, G1576A, G1594A, G1599A Agilent Technologies 2001 All Rights Reserved. Reproduction, adaptation, or translation without permission

More information

Installing an Auxiliary Pressure Control Module

Installing an Auxiliary Pressure Control Module Installing an Auxiliary Pressure Control Module Agilent 6850 GCs Accessory G3349B This kit contains: Description Quantity Cable tie.062-.625 diameter 2 Hex nut w/lockwasher 1 Disposable wrist strap 1 Minifold

More information

Agilent G2738A Upstream Capillary Interface

Agilent G2738A Upstream Capillary Interface Agilent G2738A Upstream Capillary Interface 4890, 5890, and 6890 Gas Chromatographs Installation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2003 No part of this manual may be reproduced

More information

AZ Chrom s.r.o.

AZ Chrom s.r.o. AZ Chrom s.r.o. azetchrom@hplc.sk www.azetchrom.sk Í Ì» Ú ²¹» Ì» ÙÝ Ú» æ Ì» ͳ ß» ² ª» ÛßÍÇ ÎÛÔ ßÞÔÛ ÔÛßÕ ÚÎÛÛ ÒÑ ÌÑÑÔÍ ÎÛÏË ÎÛÜ Í Ì» Ú ²¹» Ì»» ²» ¹»²» ±² º» »³ º± ¹ ½ ±³ ±¹»³ ¼» ª» ²¹ ²» ô» µ º»» ² ±²

More information

Installing the Flame Ionization Detector EPC Flow Control Manifold

Installing the Flame Ionization Detector EPC Flow Control Manifold Installing the Flame Ionization Detector EPC Flow Control Manifold The FID EPC Flow Control Manifold kit can be used to replace any HP 6890 Series FID EPC flow control manifold. This kit contains: Kit

More information

Choose for Your Instrument. GC Supplies - Agilent Instrument Quick Pick Guide

Choose for Your Instrument. GC Supplies - Agilent Instrument Quick Pick Guide Choose for Your Instrument GC Supplies - Agilent Instrument Quick Pick Guide Syringes Septa Inlet Liners O-rings and Sealing Rings SilTite FingerTite Ferrules Ferrules Your chromatography analysis does

More information

PRODUCT SPECIFICATION. Product Specification Crystal 9000 Gas Chromatograph. Chromatec Crystal Laboratory Gas Chromatography System

PRODUCT SPECIFICATION. Product Specification Crystal 9000 Gas Chromatograph. Chromatec Crystal Laboratory Gas Chromatography System Chromatec Crystal 9000 Laboratory Gas Chromatography System GC Chromatec-Crystal 9000 Crystal 9000 is a leading gas chromatograph in Chromatec GC product line. Crystal 9000 has highest performances and

More information

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS) Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph-mass spectrometer (GC-MS). 2.0 Scope

More information

1350 Valves. 1 of 14. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual

1350 Valves. 1 of 14. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual This section contains illustrated parts breakdowns for the following 6890 GC valves and related components: Valve Box Assembly Valve Driver Assembly Valve Actuator Assembly Valco W-series Minivalve Nickel

More information

C40008 & C40009 EXHAUST BRAKES

C40008 & C40009 EXHAUST BRAKES EXHAUST BRAKES C40008 & C40009 1995 2003 Ford F250 / F350 7.3 L Powerstroke Diesel with manual transmissions 1995 1998 Ford F250 / F350 7.3 L Powerstroke Diesel with automatic transmission* *Requires the

More information

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis Page 1 Alternate Carrier Gas Considerations and Faster GC Analysis Faster GC Total Analytical Cycle Times A Variety of Approaches Pre-Run ALS Set-Up Chromatographic Run Post-Run Bake-Out Post-Run Cool-Down

More information

Installing an Auxiliary Pressure Control Module

Installing an Auxiliary Pressure Control Module Installing an Auxiliary Pressure Control Module Agilent 6850 GCs Accessory G3349B These instructions are divided into two parts: Part 1 to prepare the Auxiliary Electronic Pressure Control ("AuxEPC") module

More information

Installation Guide Thermal Conductivity Detector Accessory 19232E

Installation Guide Thermal Conductivity Detector Accessory 19232E Installation Guide Thermal Conductivity Detector Accessory 19232E Agilent Technologies 2000 All Rights Reserved. Reproduction, adaptation, or translation without permission is prohibited, except as allowed

More information

ENGINE DISPLACEMENT NUMBER OF VALVES 16 VEHICLE CATEGORIES

ENGINE DISPLACEMENT NUMBER OF VALVES 16 VEHICLE CATEGORIES MANUFACTURER Dacia TYPE Duster ENGINE DISPLACEMENT 1200cc NUMBER OF VALVES 16 ENGINE CODE / NUMBER H5F (TCe125) VEHICLE CATEGORIES M TRANSMISSION MT (6-speed) VERSION AFC-2.1 PETROL ECU MANUFACTURER /

More information

5. Engine Control Module (ECM) I/O Signal

5. Engine Control Module (ECM) I/O Signal 5. Engine Control Module (ECM) I/O Signal A: ELECTRICAL SPECIFICATION B134 B135 B136 B137 17 16 15 14 13 12 11 10 9 8 27 26 25 24 23 22 21 20 19 18 34 33 32 31 30 29 28 19 18 17 16 15 14 13 12 11 10 9

More information

Choose for Your Instrument

Choose for Your Instrument Choose for Your Instrument GC Supplies - Instrument Quick Pick Guide Agilent Technologies Bruker/Scion (Varian) PerkinElmer Shimadzu Thermo Scientific GC Supplies - Instrument Quick Pick Guide Choose for

More information

NPD to FID conversion on SRI 8610C or 310C GC July2012

NPD to FID conversion on SRI 8610C or 310C GC July2012 July2012 Unplug GC from wall power The SRI 8610C GC is shown at right. The NPD detector will be located at the right side of the column oven covered with an insulated aluminum box. Remove the single screw

More information

Installation Instructions General Motors 8.1 Sequential Vapor Injection (S.V.I.) System 7500/6500 Series Trucks model year.

Installation Instructions General Motors 8.1 Sequential Vapor Injection (S.V.I.) System 7500/6500 Series Trucks model year. Installation Instructions General Motors 8.1 Sequential Vapor Injection (S.V.I.) System 7500/6500 Series Trucks 2003-2005 model year. Technocarb Equipment (2004) Ltd. 4-30435 Progressive Way Abbotsford,

More information

6890 Installing ¼ Glass Columns

6890 Installing ¼ Glass Columns 6890 Installing ¼ Glass Columns A15891 This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this material. The information

More information

The Quality of Light TM. MODEL: SSI-600-PDC Controller Operation Manual. Metaphase Technologies Inc. JAN-09

The Quality of Light TM. MODEL: SSI-600-PDC Controller Operation Manual. Metaphase Technologies Inc. JAN-09 The Quality of Light TM MODEL: SSI-600-PDC Controller Operation Manual For packaging purposes the line light and controller are shipped separately. The user must make the connection currently according

More information

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application 7890A/7890B GC Overview October 29, 2014 1 Definitions Carrier Gas Pressurized gas used to transport the sample through the system.

More information

Technical Procedure for Gas Chromatography (GC-FID)

Technical Procedure for Gas Chromatography (GC-FID) Technical Procedure for Gas Chromatography (GC-FID) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph (GC- FID). 2.0 Scope This procedure applies to all

More information

Using the PSD for Backflushing on the Agilent 8890 GC System

Using the PSD for Backflushing on the Agilent 8890 GC System Application Note Petrochemicals Using the PSD for Backflushing on the Agilent 889 GC System Author Brian Fitz Agilent Technologies, Inc. Wilmington, DE, USA. Abstract An Agilent 889 series GC equipped

More information

ENGINE DISPLACEMENT NUMBER OF VALVES ENGINE SET NUMBER VEHICLE CATEGORIES. Direct LiquiMax-2.0 PETROL ECU MANUFACTURER / CODE Bosch MED 17.

ENGINE DISPLACEMENT NUMBER OF VALVES ENGINE SET NUMBER VEHICLE CATEGORIES. Direct LiquiMax-2.0 PETROL ECU MANUFACTURER / CODE Bosch MED 17. MANUFACTURER VOLVO TYPE V60 / V70 T4F ENGINE DISPLACEMENT 1600cc NUMBER OF VALVES 16v ENGINE CODE / NUMBER B4164T2 VEHICLE CATEGORIES M TRANSMISSION MT/AT VERSION Direct LiquiMax-2.0 PETROL ECU MANUFACTURER

More information

MAKE OF AUTOMOBILE: ENGINE SET NUMBER 366/

MAKE OF AUTOMOBILE: ENGINE SET NUMBER 366/ MAKE OF AUTOMOBILE: TYPE: SORENTO PISTON DISPLACEMENT: 3300 NUMBER OF VALVES: 24 ENGINE NUMBER: G6DB TRANSMISSION TYPE ( MT / AT ) MT VEHICLE CATEGORIES M or N M TYPE VSI INJECTOR ( NUMBER + COLOR ) 180/30340

More information

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector Authors: The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector Introduction Specification of the hydrocarbon composition of LPG is required as traces

More information

Welker Sampler. Installation, Operation, & Maintenance Manual. Model GSS-4HP

Welker Sampler. Installation, Operation, & Maintenance Manual. Model GSS-4HP Installation, Operation, & Maintenance Manual Welker Sampler Model GSS-4HP The information in this manual has been carefully checked for accuracy and is intended to be used as a guide to operations. Correct

More information

Choose for Your Instrument. GC Supplies - Bruker (Varian) Instrument Quick Pick Guide

Choose for Your Instrument. GC Supplies - Bruker (Varian) Instrument Quick Pick Guide Choose for Your Instrument GC Supplies - Bruker (Varian) Instrument Quick Pick Guide Syringes Septa Inlet Liners O-rings and Sealing Rings SilTite FingerTite Ferrules Ferrules Your chromatography analysis

More information

LABORATORY ZERO AIR GENERATOR MODEL N-GC1500 USER S MANUAL

LABORATORY ZERO AIR GENERATOR MODEL N-GC1500 USER S MANUAL LABORATORY ZERO AIR GENERATOR MODEL N-GC1500 USER S MANUAL Content 1. Introduction...2 2. Important safety instruction...3 3. System component...4 4. Engineering design overview...5 5. Installation...6

More information

Model DS1000/DT20 System

Model DS1000/DT20 System Model DS1000/DT20 System Installation, Commissioning and Operation Manual Alpha Moisture Systems Alpha House 96 City Road Bradford BD8 8ES England Tel: +44 1274 733100 Fax: +44 1274 733200 Email: mail@amsytems.co.uk

More information

Setting up SilFlow for BackFlush in your GC

Setting up SilFlow for BackFlush in your GC Setting up SilFlow for BackFlush in your GC What is backflush and why use it? The BackFlush system eliminates the need to bake heavy sample fractions off the capillary column. Oils, tars and other semivolatile

More information

AZ Chrom s.r.o.

AZ Chrom s.r.o. AZ Chrom s.r.o. azetchrom@hplc.sk www.azetchrom.sk ÙÝ Í ó Ê ²ñÞ «µ» ² «³»² Ï«½µ Ð ½µ Ù«¼» Í ²¹ ²¹» Í» ²» Ô ²» Ñó ²¹ Ú» Ú ²¹» Ì» Ú» Ako nás možno kontaktova : AZ Chrom s.r.o. Robotnícka 10 831 03 Bratislava

More information

VDF G G VDF G G VDF G G METALLIC HOSE 12GF FITTING FOR SOLDERING 12 12G

VDF G G VDF G G VDF G G METALLIC HOSE 12GF FITTING FOR SOLDERING 12 12G 3.4- ACCESSORIES 3.4.1- GAS ACCESSORIES VDF G31-37 12G...3.4.1.1 VDF G20-20 12G...3.4.1.2 VDF G25-25 12G...3.4.1.3 METALLIC HOSE 12GF 700...3.4.1.4 FITTING FOR SOLDERING 12 12G.3.4.1.5 FITTING FOR SOLDERING

More information

NUMBER: S.M. REF.: Listed in Table ENGINE: EPA07 Series 60 DATE: October 2012 SUBJECT: REFERENCES TO THE TURBOCHARGER PURGE ROUTINE

NUMBER: S.M. REF.: Listed in Table ENGINE: EPA07 Series 60 DATE: October 2012 SUBJECT: REFERENCES TO THE TURBOCHARGER PURGE ROUTINE NUMBER: 10 01 12 S.M. REF.: Listed in Table ENGINE: EPA07 Series 60 DATE: October 2012 SUBJECT: REFERENCES TO THE TURBOCHARGER PURGE ROUTINE ADDITIONS, REVISIONS, OR UPDATES Publication Number Platform

More information

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D759 Application Petrochemical Author James D. McCurry Agilent Technologies 285 Centerville

More information

Agilent 218 Purification Solution

Agilent 218 Purification Solution Agilent 218 Purification Solution System User Guide Purification Solution System User Guide Agilent Technologies Notices Agilent Technologies, Inc. 2012, 2013 No part of this manual may be reproduced in

More information

TUNE-UP KIT INSTRUCTIONS

TUNE-UP KIT INSTRUCTIONS Jacobs Engine Brake Tune-Up Kit 790/795 Series P/N 29013 Information in this manual was current at time of printing and is subject to change without notice or liability. TUNE-UP KIT INSTRUCTIONS Refer

More information

Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut

Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut Application Note Environmental Author Ken Lynam Agilent Technologies, Inc. Abstract Specially designed self-tightening

More information

Units with pedestal and busbars on the short side (+H360, flat mounting)

Units with pedestal and busbars on the short side (+H360, flat mounting) 91 Units with pedestal and busbars on the short side (+H360, flat mounting) Delivery check Check that there are no signs of damage. Before attempting installation and operation, check the information on

More information

MODEL YEAR: SYSTEM APPROVAL NUMBER ( R115 ) R ENGINE SET NUMBER 354/

MODEL YEAR: SYSTEM APPROVAL NUMBER ( R115 ) R ENGINE SET NUMBER 354/ MAKE OF AUTOMOBILE: MERECEDES TYPE: E200 W211 PISTON DISPLACEMENT: 1796 NUMBER OF VALVES: 16 ENGINE NUMBER: M271 TRANSMISSION TYPE ( MT / AT ) AT VEHICLE CATEGORIES M or N M TYPE VSI INJECTOR ( NUMBER

More information

Copyright 2004 Alltech Associates, Inc.UNP. brochure #487K. accessories

Copyright 2004 Alltech Associates, Inc.UNP. brochure #487K. accessories Copyright 2004 Alltech Associates, Inc.UNP brochure #487K accessories GC Organizers 5-Column Storage Unit Plastic Storage Box 10-Column Storage Unit Utility Cabinet Lab Storage Bin Unit GC Organizers

More information

5. Engine Control Module (ECM) I/O Signal

5. Engine Control Module (ECM) I/O Signal 5. A: ELECTRICAL SPECIFICATION B134 B135 B136 B137 17 16 15 14 13 12 11 10 9 8 27 26 25 24 23 22 21 20 19 18 34 33 32 31 30 29 28 19 18 17 16 15 14 13 12 11 10 9 8 27 26 25 24 23 22 21 20 35 34 33 32 31

More information

Agilent 6890 Gas Chromatograph

Agilent 6890 Gas Chromatograph Agilent 6890 Gas Chromatograph Maintaining Your GC Agilent Technologies Notices Agilent Technologies, Inc. 2007 No part of this manual may be reproduced in any form or by any means (including electronic

More information

Installation, Operation, and Maintenance Manual

Installation, Operation, and Maintenance Manual Installation, Operation, and Maintenance Manual Welker Flanged Liquid Sampler Model LSS-1F Drawing No.: AD478BF.1 Manual No.: IOM-140 The information in this manual has been carefully checked for accuracy

More information

Welker Gas Sampler Model MPS-2

Welker Gas Sampler Model MPS-2 Installation, Operation, and Maintenance Manual Welker Gas Sampler Model MPS-2 The information in this manual has been carefully checked for accuracy and is intended to be used as a guide to operations.

More information

MONGOOSE. Introduction. < blueroomelectronics > Assembly Instructions. Mongoose was designed as an introduction to Mechatronics.

MONGOOSE. Introduction. < blueroomelectronics > Assembly Instructions. Mongoose was designed as an introduction to Mechatronics. MONGOOSE Assembly Instructions Introduction Mongoose was designed as an introduction to Mechatronics Page 1 of 12 Before you start The Mongoose is a complex kit and skipping instructions is not advised,

More information

Product Guide: Series III Pump Control Board Set (RoHS)

Product Guide: Series III Pump Control Board Set (RoHS) revised 04/08/10 Description: The Series III Pump Control Board Set provides motor drive and pump control for a wide assortment of pumps from Scientific Systems, Inc. The assembly consists of two circuit

More information

We recommend that when you need spare parts, you should use only original parts.

We recommend that when you need spare parts, you should use only original parts. Tanco Auto wrap 1300 SPARE PARTS LIST We recommend that when you need spare parts, you should use only original parts. When ordering spare Parts use the following: - Drawings are marked with a position

More information

HOLLEY VR2 FUEL PUMPS P/N & Installation Instructions 199R10975

HOLLEY VR2 FUEL PUMPS P/N & Installation Instructions 199R10975 HOLLEY VR2 FUEL PUMPS P/N 12-3000 & 12-3000-2 Installation Instructions 199R10975 WARNING! THESE INSTRUCTIONS MUST BE READ AND FULLY UNDERSTOOD BEFORE BEGINNING INSTALLATION. FAILURE TO FOLLOW THESE INSTRUCTIONS

More information

GC Injector Liners. Bruker Varian Agilent (HP) Perkin Elmer Finnigan. Phone: Page 4

GC Injector Liners. Bruker Varian Agilent (HP) Perkin Elmer Finnigan.   Phone: Page 4 Bruker Varian Agilent (HP) Perkin Elmer Finnigan www.excelcomponent.com sales@excelcomponent.com Phone: 909-229-6363 Page 4 Deactivated Split/Splitless Injector We offer a full line of Split/Splitless

More information

SE RV I CE MANUAL. Page 1 of 13

SE RV I CE MANUAL. Page 1 of 13 SE RV I CE MANUAL Technical Data Mechanical Repair Electrical Repairs Calibration Fault Diagnosis Tools and Test Equipment Parts List Page 1 of 13 Technical Data Continuous operation, with intermittent

More information

Microsep Low-Bleed PTFE-Lined Septa. XLB Septa

Microsep Low-Bleed PTFE-Lined Septa. XLB Septa Septa Chromatogram Cross Reference Guide Actual Size Drawing Measurement 1/4" (6mm) Shimadzu*, SRI* Chromatographs 3/8" (9.5mm) 7/16" (11mm) Antek, Finnigan 9100, 9600, Gow- Mac, Agilent 5700, 5880 & 5900,

More information

In addition to this instruction sheet, the following are included in this kit (some additional parts are included for future maintenance):

In addition to this instruction sheet, the following are included in this kit (some additional parts are included for future maintenance): Instruction Sheet 62068-89 Overview The components in this kit directly replace components in the existing analyzer. Some of these components have safety and/or electrostatic discharge considerations associated

More information

MODEL YEAR: 2007 SYSTEM APPROVAL NUMBER ( R115 ) R ENGINE SET NUMBER 350/

MODEL YEAR: 2007 SYSTEM APPROVAL NUMBER ( R115 ) R ENGINE SET NUMBER 350/ MAKE OF AUTOMOBILE: TYPE: KALINA 1117/ 1118 / 1119 PISTON DISPLACEMENT: 1600 NUMBER OF VALVES: 8 ENGINE NUMBER: 21114 TRANSMISSION TYPE ( MT / AT ) MT VEHICLE CATEGORIES M or N M TYPE VSI INJECTOR ( NUMBER

More information

Prins autogassystemen b.v. Veldhoven

Prins autogassystemen b.v. Veldhoven Prins autogassystemen b.v. Veldhoven MOUNTING INSTRUCTION ENGINE CONVERSION SET MAKE OF AUTOMOBILE: TYPE: ASTRA PISTON DISPLACEMENT: 2000 cc MT NUMBER OF VALVES: 16V ENGINE NUMBER: X20XEV INJECTION SYSTEM:

More information

Table of Contents Next FLO-SAFE Manifold System Pre-Installation Kit Matrx Toll-Free: Technical Support: Fax: Rev.

Table of Contents Next FLO-SAFE Manifold System Pre-Installation Kit Matrx Toll-Free: Technical Support: Fax: Rev. FLO-SAFE Manifold System Pre-Installation Kit Instruction Manual Matrx 145 Mid County Drive Orchard Park, New York 14127 716-662-6650 Toll-Free: 800-847-1000 Technical Support: 888-279-1260 Fax: 716-662-8440

More information

Agilent G2647A (Single) and G2648A (Dual) Flame Photometric Detector Update Kits

Agilent G2647A (Single) and G2648A (Dual) Flame Photometric Detector Update Kits Agilent G2647A (Single) and G2648A (Dual) Flame Photometric Detector Update Kits For Agilent 6850 and 6890 GCs Update, Operation and Maintenance Agilent Technologies Notices Agilent Technologies, Inc.

More information

ASSEMBLY INSTRUCTIONS FOR 10L, 15L, & 20L JACKETED PROCESS REACTOR SYSTEMS

ASSEMBLY INSTRUCTIONS FOR 10L, 15L, & 20L JACKETED PROCESS REACTOR SYSTEMS 3800 N. Mill Road Vineland, NJ 08360 Phone: 800.843.1794 Fax: 800.922.4361 ASSEMBLY INSTRUCTIONS FOR 10L, 15L, & 20L JACKETED PROCESS REACTOR SYSTEMS New & Improved Open Frame Design Provides Excellent

More information

INSTALLATION INSTRUCTIONS FUEL RAIL

INSTALLATION INSTRUCTIONS FUEL RAIL INSTALLATION INSTRUCTIONS FUEL RAIL MITSUBISHI EVO X Document# 19-0067 Support: info@radiumauto.com WARNING: DON'T SMOKE OR WORK WITH OPEN SPARKS WHILE WORKING ON THE FUEL SYSTEM PREPARING THE VEHICLE:

More information

Agilent G2333B (Single) and G2334B (Dual) Flame Photometric Detectors

Agilent G2333B (Single) and G2334B (Dual) Flame Photometric Detectors Agilent G2333B (Single) and G2334B (Dual) Flame Photometric Detectors For Agilent 6890 GC Installation and Operation Agilent Technologies Notices Agilent Technologies, Inc.2004, 2005 No part of this manual

More information

Product manual Oil Streak Sensor INTRODUCTION CONSTRUCTION. Master Sensor

Product manual Oil Streak Sensor INTRODUCTION CONSTRUCTION. Master Sensor Product manual Oil Streak Sensor INTRODUCTION Oil streak sensors are designed to detect traces of oil travelling through air tubes, down to flows as low as 5mm 3 /min. The product utilizes a master and

More information

Chapter 11. Engine & Associated Systems Contents INTRODUCTION. Tools Required

Chapter 11. Engine & Associated Systems Contents INTRODUCTION. Tools Required Chapter 11 Engine & Associated Systems Contents INTRODUCTION...1 Tools Required...1 Parts Required...2 FITTING OF ENGINE AND SYSTEMS...3 1 Fitting Chassis Mount...3 2 Fitting Radiator Brackets...3 3 Preparing

More information

Included parts: 1 - New Bosch CP3 Pump 1 - HSM Pulley 1 - Serpentine Belt 1 - Pump Bracket/ Hardware STEP 1

Included parts: 1 - New Bosch CP3 Pump 1 - HSM Pulley 1 - Serpentine Belt 1 - Pump Bracket/ Hardware STEP 1 TROUBLESHOOTING: Please read and understand all installation instructions before proceeding with the installation. If you have questions during the installation of this product, please contact H&S Motorsports

More information

Engine Emission Control 6.7L Diesel

Engine Emission Control 6.7L Diesel Page 1 of 6 SECTION 303-08: Engine Emission Control 2011 F-250, 350, 450, 550 Super Duty Workshop Manual DESCRIPTION AND OPERATION Procedure revision date: 03/12/2010 Engine Emission Control 6.7L Diesel

More information

Installation Guide. Heater Interface Cable for Mass Sensitive Detector Accessory G1590A

Installation Guide. Heater Interface Cable for Mass Sensitive Detector Accessory G1590A Installation Guide Heater Interface Cable for Mass Sensitive Detector Accessory G1590A Agilent Technologies, Inc. 2001 All Rights Reserved. Reproduction, adaptation, or translation without permission is

More information

Part# PLEASE READ ALL INSTRUCTIONS BEFORE INSTALLATION.

Part# PLEASE READ ALL INSTRUCTIONS BEFORE INSTALLATION. 16 November 2009 HPCR Dodge Twin Turbo Kit #1045430 1 BD Twin Turbo R700 Kit 2003-2007 Dodge HPCR ISBe Installation Instructions Part# 1045430 PLEASE READ ALL INSTRUCTIONS BEFORE INSTALLATION. UNLESS AN

More information

Agilent 7820A Gas Chromatograph

Agilent 7820A Gas Chromatograph Agilent 7820A Gas Chromatograph Illustrated Parts Breakdown 15 Illustrated Parts Breakdown Overview of the IPB 297 Inlets 298 Detectors 316 EPC modules 357 Nickel catalyst accessory 365 Covers 369 Automatic

More information

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI) Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI) Application Note All Industries Authors Brian Fitz and Bill Wilson Agilent Technologies, Inc. 285 Centerville Road Wilmington,

More information

Pressure Transducer Field Installation Instructions for the LI-6252 CO 2 and LI-6262 CO 2 /H 2 O Analyzers

Pressure Transducer Field Installation Instructions for the LI-6252 CO 2 and LI-6262 CO 2 /H 2 O Analyzers 6262-05 6262-03 Pressure Transducer Field Installation Instructions for the LI-6252 CO 2 and LI-6262 CO 2 /H 2 O Analyzers Please read these instructions before beginning installation. Please note that

More information