VIBRATION AND MECHANICAL SHOCK TEST REPORT

Size: px
Start display at page:

Download "VIBRATION AND MECHANICAL SHOCK TEST REPORT"

Transcription

1 Report No. :DH A-1 Version:A VIBRATION AND MECHANICAL SHOCK TEST REPORT Company :OTO Photonics Inc. Address :9F-5,No. 27, Guanxin Rd., East Dist., Hsinchu City 30072, Taiwan(R.O.C.) Sample Name :Ultra Micro Spectrometer Date Received :JAN 19, 2017 Date Tested :JAN 19, 2017 ~ JAN 20, 2017 TESTING LABORATORY IS ACCREDITED BY: IEC/IECQ certificate of independent test laboratory approval Certificate No.: ISO accredited in respect of laboratory is approved by TAF Certificate No.:L0835 ISO 9001 certificate is approved by TUV CERT certification body of TUV NORD Cert GmbH WE HEREBY CERTIFY THAT: The test(s) shown in the attachment were conducted according to the indicating procedures. We assume full responsibility for the accuracy and completeness of these tests and vouch for the qualifications of all personnel performing them. Name Signature Date Test Engineer Marco Yu Jan 23, 2017 Manager Allan Tseng Jan 24, 2017 NOTE: 1. This report will be invalid if reproduced in part or altered in any way. 2. This report refers only to the specimen(s) submitted to test, and is invalid if used otherwise. 3. This report is ONLY valid with the examination seal and signature of this institute. 4. The tested specimen(s) will only be preserved for thirty days from the date issued, if not collected by the applicant.

2 Reliability Testing Laboratory Page 1 of 15 TABLE OF CONTENTS 1. GENERAL INFORMATION DESCRIPTION OF UNIT UNIT OPERATING CONDITION VIBRATION TEST TEST EQUIPMENT LABORATORY AMBIENCE CONDITION REFERENCE DOCUMENT TEST CONDITION SUMMARY OF TEST PHOTO OF TEST GRAPH OF TEST MECHANICAL SHOCK TEST TEST EQUIPMENT LABORATORY AMBIENCE CONDITION REFERENCE DOCUMENT TEST CONDITION SUMMARY OF TEST PHOTO OF TEST GRAPH OF TEST... 13

3 Reliability Testing Laboratory Page 2 of GENERAL INFORMATION 1.1 DESCRIPTION OF UNIT Manufacturer:OTO Photonics Inc. Sample name:ultra Micro Spectrometer Sample quantity:8 units 1.2 UNIT OPERATING CONDITION None

4 Reliability Testing Laboratory Page 3 of VIBRATION TEST 2.1 TEST EQUIPMENT Model Serial Number Calibration Date King Design EM-600F2K-40N120 UW JUN 02, LABORATORY AMBIENCE CONDITION Temperature:25 C ± 10 C Relative humidity:50 % ± 25 % (RH) 2.3 REFERENCE DOCUMENT The test is based on customer specification 2.4 TEST CONDITION Units are non-operating. Vibration waveform:sine waveform Vibration frequency:10 to 200 Hz Vibration acceleration:1.5 g Cycle time:30 minutes Number of cycle:1 cycle for each axis Vibration axes:x, Y and Z 2.5 SUMMARY OF TEST After testing, visual inspection showed no physical defect or functional degradation of units.

5 Reliability Testing Laboratory Page 4 of PHOTO OF TEST X axis Y axis Z axis

6 Reliability Testing Laboratory Page 5 of 15 After testing Appearance was normal Function was normal

7 Reliability Testing Laboratory Page 6 of GRAPH OF TEST X axis

8 Reliability Testing Laboratory Page 7 of 15 Y axis

9 Reliability Testing Laboratory Page 8 of 15 Z axis

10 Reliability Testing Laboratory Page 9 of MECHANICAL SHOCK TEST 3.1 TEST EQUIPMENT Model Serial Number Calibration Date LAB ASII JUL 01, LABORATORY AMBIENCE CONDITION Temperature:25 C ± 10 C Relative humidity:50 % ± 25 % (RH) 3.3 REFERENCE DOCUMENT The test is based on customer specification 3.4 TEST CONDITION Units are non-operating. Pulse shape:half-sine waveform Impact acceleration:50 g Pulse duration:12 ms Number of shocks:6 shocks (1 shock for each face) Orientation:Bottom, top, left, right, front and rear faces 3.5 SUMMARY OF TEST After testing, visual inspection showed no physical defect or functional degradation of units.

11 Reliability Testing Laboratory Page 10 of PHOTO OF TEST Units

12 Reliability Testing Laboratory Page 11 of 15 Bottom Top Left Right Front Rear

13 Reliability Testing Laboratory Page 12 of 15 After testing Appearance was normal Function was normal

14 Reliability Testing Laboratory Page 13 of GRAPH OF TEST Bottom face Top face

15 Reliability Testing Laboratory Page 14 of 15 Left face Right face

16 Reliability Testing Laboratory Page 15 of 15 Front face Rear face << The Following Blank >>

VIBRATION TEST REPORT

VIBRATION TEST REPORT Report No. :HS1505160024A-1 Version:B VIBRATION TEST REPORT Company Address :Cincon Electronics Co., Ltd. :No.8-1 Fu Kung Rd., Fu Hsing ind. Park, Fu Hsing Hsiang, Chang Hua Hsien, Taiwan, R.O.C. Model

More information

SHOCK TEST REPORT. :Cincon Electronics Co., Ltd. Hua Hsien, Taiwan, R.O.C. :EC7BW-110S15, ECLB40W-110S05, CQB60W-110S24

SHOCK TEST REPORT. :Cincon Electronics Co., Ltd. Hua Hsien, Taiwan, R.O.C. :EC7BW-110S15, ECLB40W-110S05, CQB60W-110S24 Report No. :HS1505160024A-2 Version:B SHOCK TEST REPORT Company Address :Cincon Electronics Co., Ltd. :No.8-1 Fu Kung Rd., Fu Hsing ind. Park, Fu Hsing Hsiang, Chang Hua Hsien, Taiwan, R.O.C. Model Name

More information

THERMAL SHOCK TEST REPORT

THERMAL SHOCK TEST REPORT Report No. :DH1509300024A Version:A THERMAL SHOCK TEST REPORT Company :FLIP CHIP OPTO. INC. Address :No.185, Kewang Rd., Longtan Dist., Taoyuan City 325, Taiwan (R.O.C.) Model Name :Apollo 600-B Lot Number

More information

CONTINUOUS IMMERSION AND DUST TEST REPORT

CONTINUOUS IMMERSION AND DUST TEST REPORT Report No. :HS1404020081A Version:A CONTINUOUS IMMERSION AND DUST TEST REPORT Company Address :Taiway Electronics Components Co. :4F., No. 4, 6, 8, Lane 359, Sec. 2, Zhongshan Rd., Zhonghe Dist., New Taipei

More information

SHOCK TEST REPORT. Report No. : ENVA S9 Model Number : S9

SHOCK TEST REPORT. Report No. : ENVA S9 Model Number : S9 Report No. : ENVA10072605 S9 Model Number : S9 Page: 1 of 6 Product Name : S9 Model Number : S9 Date of Receipt : Jul. 26, 2010 Finished date of Test : Jul. 26, 2010 We, Spectrum Research & Testing Laboratory

More information

QSFP Copper Module Direct Attach Cable Assembly &Cage

QSFP Copper Module Direct Attach Cable Assembly &Cage Qualification Test Report 501-60067 12Mar. 2012 Rev. 1 QSFP Copper Module Direct Attach Cable Assembly& Qualification Test Report QSFP Copper Module Direct Attach Cable Assembly & Tyco Electronics. (Shanghai)

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 DEKRA IST RELIABILITY SEVICES INC. No. 22 Puding Rd., East Dist. Hsinchu City 300, Taiwan, R.O.C. Phone: 886 3 579 5766 ext 7100 Mr. General Lee general.lee@dekra-ist.com

More information

600 Amp Surge Arrester

600 Amp Surge Arrester R&D Center 497 Schooley s Mountain Road Hackettstown, NJ 07840-3906 Tele: 908-852-1122 Fax: 908-813-2091 TEST REPORT NUMBER: 379-17-12164 DATE: 4-3-2012 REPORT OF TEST ON 600 Amp Surge Arrester Statements

More information

SGS TEST REPORT. Page: 1 of 6 Date: April 20, itech Company LLC Christy Street, Fremont CA USA

SGS TEST REPORT. Page: 1 of 6 Date: April 20, itech Company LLC Christy Street, Fremont CA USA Page: 1 of 6 Date: April 20, 2010 itech Company LLC 41758 Christy Street, Fremont CA 94538 USA The following merchandise was submitted and identified by the vendor as: Product Description: 20.1 Multimedia

More information

Our test laboratory - Apparatebau Kirchheim-Teck GmbH

Our test laboratory - Apparatebau Kirchheim-Teck GmbH 1 / 16 Test laboratory equipment 1 Microsection lab for crimp contacts Tensile, pressure and bending test machine: Static + dynamic Vibration test: Vibration + shock Climate tests: Temperature + humidity

More information

MEASUREMENT AND TEST REPORT

MEASUREMENT AND TEST REPORT IES LM-80-2008 MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT For Guangzhou Hongli Opto-Electronic Co., Ltd. West side of Dongfeng Highway, Automobile industrial Base, Huadu

More information

AUGUST 28, 2008 TEST REPORT # REV.1.1 MIXED FLOWING GAS TESTING CONNECTOR SERIES CLP S-D-A FTSH S-DV-A SAMTEC, INC.

AUGUST 28, 2008 TEST REPORT # REV.1.1 MIXED FLOWING GAS TESTING CONNECTOR SERIES CLP S-D-A FTSH S-DV-A SAMTEC, INC. AUGUST 28, 2008 TEST REPORT #208383-3 REV.1.1 MIXED FLOWING GAS TESTING CONNECTOR SERIES CLP-130-02-S-D-A FTSH-130-02-S-DV-A SAMTEC, INC. APPROVED BY: DOMINIC ARPINO PROGRAM MANAGER CONTECH RESEARCH, INC.

More information

AUGUST 28, 2008 TEST REPORT # REV.1.1 MIXED FLOWING GAS TESTING CONNECTOR SERIES SEM H-D-WT TEM H-D-WT SAMTEC, INC.

AUGUST 28, 2008 TEST REPORT # REV.1.1 MIXED FLOWING GAS TESTING CONNECTOR SERIES SEM H-D-WT TEM H-D-WT SAMTEC, INC. AUGUST 28, 2008 TEST REPORT #208383-5 REV.1.1 MIXED FLOWING GAS TESTING CONNECTOR SERIES SEM-125-02-03.0-H-D-WT TEM-125-02-03.0-H-D-WT SAMTEC, INC. APPROVED BY: DOMINIC ARPINO PROGRAM MANAGER CONTECH RESEARCH,

More information

EUMOS SYMPOSIUM 2017

EUMOS SYMPOSIUM 2017 EUMOS SYMPOSIUM 2017 CONTACT Martin Keller TÜV NORD Mobilität GmbH & Co. KG IFM Whole Vehicles Safety of Loads Group Phone: 0049 (0) 511/ 9986-1274 Fax: 0049 (0) 511/ 9986-1998 ladungssicherung@tuev-nord.de

More information

Jun06 Rev D

Jun06 Rev D Product Specification High Speed Serial Data 2 Connector 108-1965 27Jun06 Rev D 1. SCOPE 1.1. Content This specification covers performance, tests and quality requirements for the Tyco Electronics High

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 1, 2, 3 CONTINENTAL AUTOMOTIVE SYSTEMS 2400 Executive Hills Blvd. Auburn Hills, MI 48326 Mr. David E. Swift Phone: 248 209 3813 Email: david.swift@continental-corporation.com

More information

NATIONAL TECHNICAL SYSTEMS

NATIONAL TECHNICAL SYSTEMS Test Report Number 41418-09.PPC Testing of Pancake Alternator & PMU PREPARED FOR: The Perfect Parts Company Sullivan Products 1 North Haven Street Baltimore, MD 21224 PREPARED BY: National Technical Systems

More information

SEP 2016 Rev D2

SEP 2016 Rev D2 Product Specification 108-2112 28 SEP 2016 Rev D2 VAL-U-LOK* Connectors 1. SCOPE 1.1. Content This specification covers performance, tests and quality requirements for VAL-U-LOK* Connectors. 1.2. Qualification

More information

Serial Attached SCSI (SAS) Connector

Serial Attached SCSI (SAS) Connector Product 108-51053 Specification DR: Stanley Huang 24 Oct 2013 Rev G APVD: Vincent Peng Serial Attached SCSI (SAS) Connector 1.0 SCOPE This specification covers the requirements for product performance,

More information

Annex to the Accreditation Certificate D-PL according to DIN EN ISO/IEC 17025:2005

Annex to the Accreditation Certificate D-PL according to DIN EN ISO/IEC 17025:2005 Deutsche Akkreditierungsstelle GmbH Annex to the Accreditation Certificate D-PL-20258-01-00 according to DIN EN ISO/IEC 17025:2005 Period of validity: 2016-02-02 to 2021-02-01 Holder of certificate: Amphenol-Tuchel

More information

PTC Thermistor for Automotive:TPM-C Series

PTC Thermistor for Automotive:TPM-C Series Features 1. Qualification based on AEC-Q200 Rev-C 2. RoHS & Halogen-free compliant 3. EIA size 0603,0805 4. Fast and reliable response Recommended Applications 1. Automotive electronics Part Number Code

More information

Jun 20,2014 Rev A

Jun 20,2014 Rev A Product Specification 108-115067 Jun 20,2014 Rev A DDR4 DIMM Through-hole Memory Socket 1. SCOPE 1.1. Content This specification covers performance, tests and quality requirements for the TE Connectivity

More information

Product Specification Rev A

Product Specification Rev A Product Specification 108-6116 1. Scope :.5 Signal DBL Lock Connector 18P 1.1 Contents This specification covers the requirements for product performance, test methods and quality assurance provisions

More information

Dispensette. Testing Instructions (SOP) 1. Introduction. May 2009

Dispensette. Testing Instructions (SOP) 1. Introduction. May 2009 Dispensette Testing Instructions (SOP) May 2009 1. Introduction The standard ISO DIS 8655 describes both the design and the testing of the bottle-top dispenser. The following Testing Instructions describe

More information

VIA Platform Environmental Qualification Testing Standards

VIA Platform Environmental Qualification Testing Standards VIA Platform Environmental Qualification Testing Standards Revision: 1 Report Dated: September 7 th 2016 Report Prepared By: Noel Joyce Reliability Lab Manager Page 1 of 16 Table of Contents 1. Purpose...

More information

VITLAB Dispenser. Standard Operating Procedure (SOP)

VITLAB Dispenser. Standard Operating Procedure (SOP) VITLAB Dispenser Standard Operating Procedure (SOP) VITLAB GmbH Linus-Pauling-Str. 63762 Grossostheim Germany tel: +49 6026 97799-0 fax: +49 6026 97799-30 info@vitlab.com www.vitlab.com Table of contents.

More information

Aug07 Rev A All Paragraphs Revised

Aug07 Rev A All Paragraphs Revised Qualification Test Report Universal MATE-N-LOK* Connector 110-213 28Aug07 Rev A All Paragraphs Revised 1. INTRODUCTION 1.1. Purpose 1.2. Scope Testing was performed on Universal MATE-N-LOK* connectors

More information

IES LM MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES

IES LM MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES IES LM-80-2008 MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT For Guangzhou Hongli Opto-Electronic Co., Ltd. West side of Dongfeng Highway, Automobile industrial Base, Huadu

More information

2.0 TEST PROCEDURES AND RESULTS Vibration Test TEST EQUIPMENT AND INSTRUMENTATION QUALITY ASSURANCE PROGRAM 7 ATTACHMENTS

2.0 TEST PROCEDURES AND RESULTS Vibration Test TEST EQUIPMENT AND INSTRUMENTATION QUALITY ASSURANCE PROGRAM 7 ATTACHMENTS Page No. 2 TABLE OF CONTENTS Page No. 1.0 INTRODUCTION 3 1.1 Scope 3 1.2 References 3 1.3 Test Specimen Description 3 1.4 Summary 3 2.0 TEST PROCEDURES AND RESULTS 4 2.1 Vibration Test 4 3.0 TEST EQUIPMENT

More information

PTC Thermistor for Automotive:TPM-C Series

PTC Thermistor for Automotive:TPM-C Series SMD PTC Thermistor for Features 1. Qualification based on AEC-Q200 Rev-C 2. RoHS & Halogen-free compliant 3. EIA size 0603,0805 4. Fast and reliable response Recommended Applications 1. Automotive electronics

More information

Product Specification QSL RF Connectors

Product Specification QSL RF Connectors Product Specification QSL RF Connectors 108-2324 04Feb09 Rev A 1. SCOPE 1.1. Content This specification covers performance, tests and quality requirements for the Tyco Electronics QSL RF Connectors which

More information

Test Group Description. Total Number. Figure 1. Unless otherwise stated, the following environmental conditions prevailed during testing:

Test Group Description. Total Number. Figure 1. Unless otherwise stated, the following environmental conditions prevailed during testing: Qualification Test Report 26 November 12 Rev A1 SL Series Jacks, Category 6 1. INTRODUCTION 1.1. Purpose 1.2. Scope Testing was performed on AMP NETCONNECT* SL Series Category 6 Jacks to determine their

More information

PRODUCT SPECIFICATION

PRODUCT SPECIFICATION of 3 C Section : For the 200222, 2002323, 2002444, 20025, 200252, 200252, 20026, 200262, 200262, 20028, 200283, 200284, 2002822, 2002824, 200283, 2002832, 2002833, 2002834 and 2002835 series parts..0 GENERAL

More information

ASTM Consensus Standards The Latest in Packaging Testing. Presented by: Dhuanne Dodrill President Rollprint Packaging Products, Inc.

ASTM Consensus Standards The Latest in Packaging Testing. Presented by: Dhuanne Dodrill President Rollprint Packaging Products, Inc. ASTM Consensus Standards The Latest in Packaging Testing Presented by: Dhuanne Dodrill President Rollprint Packaging Products, Inc. ASTM Overview Established in 1898 Voluntary, Full-Consensus Standards

More information

973 Dew Point Mirror. Portable Industrial Chilled Mirror Hygrometer. Typical applications: REFLECTING YOUR STANDARDS

973 Dew Point Mirror. Portable Industrial Chilled Mirror Hygrometer. Typical applications: REFLECTING YOUR STANDARDS Portable Industrial Chilled Mirror Hygrometer Fundamental drift free humidity measurement Integral frost/dew point measuring head Internal sample pump Optimal Response Injection System for fast measurement

More information

«CHARPY» Computerized Pendulum IMPACT TESTERS

«CHARPY» Computerized Pendulum IMPACT TESTERS «CHARPY» Computerized Pendulum IMPACT TESTERS DIGITAL & COMPUTERIZED models of 300, 450 or 750 Joules capacity 300 J PIC-300/C model 450 J PIC-450/C model 750 J PIC-750/C model ASTM E 23 DIN UNE EN ISO

More information

SGS TEST REPORT. Page: 1 of 8 Date: April 15,2010. itech Company LLC Christy Street, Fremont CA USA

SGS TEST REPORT. Page: 1 of 8 Date: April 15,2010. itech Company LLC Christy Street, Fremont CA USA Page: 1 of 8 Date: April 15,2010 itech Company LLC 41758 Christy Street, Fremont CA 94538 USA The following merchandise was submitted and identified by the vendor as: Product Description: 20.1 Multimedia

More information

Report Forms Test Method D 7038 L-33-1 Version Conducted For. N = Results Cannot Be Interpreted (See Comment Section)

Report Forms Test Method D 7038 L-33-1 Version Conducted For. N = Results Cannot Be Interpreted (See Comment Section) Report Forms Test Method D 7038 Version Conducted For V = Valid I = Invalid N = Results Cannot Be Interpreted (See Comment Section) NR = Non-Reference Test Oil RO = Reference Oil Result Test Number Motoring

More information

TEST REPORT Report No.: HCO0127/2006 Page: 1 of 9 Date: November 1, 2006

TEST REPORT Report No.: HCO0127/2006 Page: 1 of 9 Date: November 1, 2006 Page: 1 of 9 Date: November 1, 2006 ZIBO MICRO COMMERCIAL COMPONENTS CORP. ZHANG LIU ROAD, ZHANGDIAN DISTRICT, ZIBO, SHANDONG, P.R. CHINA The following merchandise was submitted and identified by the vendor

More information

SPECIFICATION FOR BATTERY, NON-RECHARGEABLE, ALKALINE MANGANESE DIOXIDE, 1.5V, NATO STOCK NO XXX-XXXX (NATO TYPE DESIGNATION NBA 3058)

SPECIFICATION FOR BATTERY, NON-RECHARGEABLE, ALKALINE MANGANESE DIOXIDE, 1.5V, NATO STOCK NO XXX-XXXX (NATO TYPE DESIGNATION NBA 3058) 4 JUNE 1984 SPECIFICATION FOR BATTERY, NON-RECHARGEABLE, ALKALINE MANGANESE DIOXIDE, 1.5V, NATO STOCK NO 6135-99-XXX-XXXX (NATO TYPE DESIGNATION NBA 3058) 1 This Supplement is to be read in conjunction

More information

IESNA LM MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT. For

IESNA LM MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT. For IESNA LM-80-2008 MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT For SHENZHEN MTC LIGHTING CO., LTD. NO.65,Chuangyi Road,Dalang Street,Baoan District,Shenzhen,GuangDong Province,CHINA

More information

Flash Point Process Analyzer

Flash Point Process Analyzer Norms and standards ASTM D 56 ASTM D 93; DIN EN ISO 2719; IP 34 DIN EN ISO 13736; IP170 DIN 51755 Application The BARTEC BENKE Flash Point Process Analyzer (FPA-4) is a continuously measuring analyzer,

More information

Lithium Battery UN38.3 Test Report

Lithium Battery UN38.3 Test Report Lithium Battery UN38.3 Test Report Prepared For : Samples Name: Model : Prepared By : Report No.: Allfine Lithium Energy (Hefei) Co., Ltd. Building 12, Mechatronics Industrial Park, Fengle Road, Hefei

More information

Certificate of Calibration

Certificate of Calibration Model Number Manufacturer Description Serial Number DSOX2014A Keysight Technologies Inc Oscilloscope, 4-channel, 100MHz MY52161639 Date of Calibration 5 Dec 2014 Procedure STE-50114571-A.02.00 Temperature

More information

Lithium Battery UN38.3 Test Report

Lithium Battery UN38.3 Test Report Lithium Battery UN38.3 Test Report Prepared For : Samples Name: Model : Prepared By : Report No.: Shenzhen Haomingshe Electronic Technology Co., Ltd. 2F, Building E, Huachuangda Industrial Park, Hangcheng

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 TUV SUD AMERICA - AUBURN HILLS 1755 Atlantic Blvd. Auburn Hills, MI 48326 William (Mac) Elliott Phone: 813 284 2736 MECHANICAL Valid To: August 31, 2018 Certificate

More information

NTC Thermistor for Automotive: TSM-C Series

NTC Thermistor for Automotive: TSM-C Series Features 1. Qualification based on AEC-Q200 Rev-C 2. Operating temperature range: -50 ~ +150 3. Superior stability in high-temperature and high-humidity environment 4. RoHS & Halogen Free (HF) compliant

More information

Standard Calibration Certificate

Standard Calibration Certificate Mettler Toledo Service Business Unit Laboratory 1900 Polaris Parkway Columbus, OH 43240 1-800-METTLER Customer Standard Calibration Certificate Company: ABC Pharma Incorporated Address: 1432 East 9th Street

More information

TEST REPORT. The following merchandise was submitted and identified by the vendor as:

TEST REPORT. The following merchandise was submitted and identified by the vendor as: Page: 1 of 13 Date: Feb 17, 2016 E-ONE MOLI ENERGY CORP. Tainan Science-Based Industry Park No.10 Dail 2nd Rd., Shan-Hwa, Tainan City, Taiwan R.O.C. Tel: 886-6-505-0666, Fax: 886-6-505-0777 http://www.molicel.com.

More information

Verification/Report Number(s): HEARTWAY MEDICAL PRODUCTS Co., Ltd. Applicant Name/Address:

Verification/Report Number(s): HEARTWAY MEDICAL PRODUCTS Co., Ltd. Applicant Name/Address: Test Verification Of Conformity On the basis of the tests undertaken, the sample(s) of the below product have been found to comply with the essential requirements of the referenced specifications at the

More information

Lithium Battery UN38.3 Test Report

Lithium Battery UN38.3 Test Report Lithium Battery UN38.3 Test Report Prepared For : Samples Name: Shenzhen Pu Sheng Power Supply Technology Co., LTD. Rm501, 5F, Jinsheng Industrial Zone, No.109, HuDiPai, DaFu, GuanLan Town, LongHua New

More information

IESNA LM MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT. For

IESNA LM MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT. For IESNA LM-80-2008 MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT For Room 316, Building 2, No.1, Xianke Yi Road, Huadong Town, Huadu District, Guangzhou, China Model:100B10C10(Ra2)

More information

IQLab Pack UNDOT Test Report

IQLab Pack UNDOT Test Report IQ Laboratories 508 Chundong Road Shanghai 201108 China IQLab Pack UNDOT Test Report Tested According To UN Manual of Tests and Criteria, Part III, subsection 38.3, Rev 5 Name of Sample: Rechargeable Lithium

More information

Anti-Surge Thick Film Chip Resistors

Anti-Surge Thick Film Chip Resistors FEATURES EXCELLENT ANTI-SURGE CHARACTERISTICS AEC-Q200 QUALIFIED RATED POWER UPGRADE IN SMALLER PACKAGE SIZE MEETS +85 C/85%RH TEST 000 HOURS MEETS CLIMATE CATEGORY (IEC 60068): 55/55/56 AVAILABLE IN ±

More information

PRODUCT SPECIFICATION

PRODUCT SPECIFICATION 1.0mm Pitch ZIF FPC Page 1 Top Entry Side Entry 1.0mm Pitch ZIF FPC Page 2 1.0 SCOPE. This specification covers performance, tests and quality requirements for 1.0mm Pitch ZIF FPC Connector 2.0 APPLICABLE

More information

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE PRESSURE

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE PRESSURE CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE co 2 PRESSURE www.epluse.com calibration AT the highest level The National Metrological Institute (NMI) or its Designated Institutes

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 INDEPENDENT ENGINEERING LABORATORIES, INC. 1 2400 E. South Street Jackson, MI 49201 Ms. Jessica Stremich Phone: 517 788 6590 E-mail: IELaccounting@ielinc.com

More information

PENTAGON PROTECTION USA

PENTAGON PROTECTION USA PENTAGON PROTECTION USA Shock Tube Testing for Window Systems March 2009 San Antonio, Texas ABS Consulting Project Number 2127174 ABSG Consulting, Inc. 14607 San Pedro, Suite 215 San Antonio, TX 78232

More information

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009 Transferpette Testing Instructions (SOP) May 2009 1. Introduction The standards DIN EN ISO 8655 describe both the design and the testing of the piston operated pipette Transferpette. The following Testing

More information

TV-5, TV-8 rated 10mm flat power relays Silent type is available

TV-5, TV-8 rated 10mm flat power relays Silent type is available TV-, TV- rated 1mm flat power relays Silent type is available LK-F RELAYS TV standards compatible: TV- and TV- TV- type: 7 A inrush current and switching possible at A rated current. : 11 A inrush current

More information

EU RO Mutual Recognition Technical Requirements

EU RO Mutual Recognition Technical Requirements 1. PRODUCT DESCRIPTION... 1 2. DESIGN EVALUATION... 2 3. PRODUCTION REQUIREMENTS... 6 4. MARKING REQUIREMENTS... 6 5. TYPE APPROVAL CERTIFICATE CONTENT... 7 6. APPROVAL DATE AND REVISION NUMBER... 7 7.

More information

IESNA LM MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT. For

IESNA LM MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT. For IESNA LM-80-2008 MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT For Guangzhou Hongli Opto-Electronic Co., Ltd. No.1, Xianke Yi Road, Huadong Town, Huadu District, Guangzhou,

More information

Report of Inspection

Report of Inspection P. O. Box 8238 Searcy, AR 72145 866.811.5237 / 501.279.1166 Fax: 501.279.1225 As required by NFPA 1911 19.3: CFS Inspections is accredited to the requirements of ISO/IEC 17020. Report of Inspection SAFET

More information

EMC Research Institute

EMC Research Institute EMC Research Institute 66-6, Jeil-Ri, Yangji-Myeon, Cheoin-Gu, Yongin-Si, Gyeonggi-Do, Korea Tel : 031-679-9687, Fax : 031-335-9786 Certificate No: ERI-0803070810 Page ( 1 ) of ( 15 ) 1. Client Name :

More information

Annex to the Accreditation Certificate D-PL according to DIN EN ISO/IEC 17025:2005

Annex to the Accreditation Certificate D-PL according to DIN EN ISO/IEC 17025:2005 Deutsche Akkreditierungsstelle GmbH Annex to the Accreditation Certificate D-PL-11272-01-00 according to DIN EN ISO/IEC 17025:2005 Period of validity: 18.10.2017 to 17.10.2022 Holder of certificate: VDZ

More information

Off-grid lighting testing and certification processes. Pradeep Kumar

Off-grid lighting testing and certification processes. Pradeep Kumar Off-grid lighting testing and certification processes Pradeep Kumar Centre for Research on Sustainable Building Science The Energy and Resources Institute (TERI) Indian Habitat Centre, New Delhi, India

More information

TEST METHOD D6121 L-37 Load Evaluation VERSION CONDUCTED FOR: I = Invalid N = Results cannot be interpreted(refer to comment section)

TEST METHOD D6121 L-37 Load Evaluation VERSION CONDUCTED FOR: I = Invalid N = Results cannot be interpreted(refer to comment section) Load Evaluation VERSION CONDUCTED FOR: V = Valid I = Invalid N = Results cannot be interpreted(refer to comment section) NR = Non-Reference Test Oil RO = Reference Oil Result Test Stand: Date Completed:

More information

ULTRASONIC FLOW SENSOR QALCOSONIC FLOW 2

ULTRASONIC FLOW SENSOR QALCOSONIC FLOW 2 AB AXIS INDUSTRIES ULTRASONIC FLOW SENSOR QALCOSONIC FLOW 2 TECHNICAL DESCRIPTION, INSTALLATION AND USER INSTRUCTIONS PESF2V01 KAUNAS Contents Page SAFETY INFORMATION... 1. APPLICATION FIELD... 2. TECHNICAL

More information

Test certificate General construction inspection test certificate Nr. 3666/4421 -a - Fe/Wi 1. Issue ( )

Test certificate General construction inspection test certificate Nr. 3666/4421 -a - Fe/Wi 1. Issue ( ) Test certificate Nr. 3666/4421-a -Fe/Wi- page 1 TESTING LABORATORY FOR CIVIL ENGINEERING MATERIALS INSTITUTE FOR REASEARCH OF CIVIL ENGINEERING MATERIALS, STONE- AND CONCRETE CONSTRUCTION AND FIRE INTEGRITY

More information

FEBRRUARY 22, 2008 TEST REPORT # QSH/QTH/QTE SERIES CONNECTOR SHEAR TESTING SAMTEC, INC.

FEBRRUARY 22, 2008 TEST REPORT # QSH/QTH/QTE SERIES CONNECTOR SHEAR TESTING SAMTEC, INC. FEBRRUARY 22, 2008 TEST REPORT #207908 QSH/QTH/QTE SERIES CONNECTOR SHEAR TESTING SAMTEC, INC. APPROVED BY: DOMINIC ARPINO PROGRAM MANAGER CONTECH RESEARCH, INC. Contech Research REVISION HISTORY DATE

More information

Anti-Surge Thick Film Chip Resistors

Anti-Surge Thick Film Chip Resistors FEATURES EXCELLENT ANTI-SURGE & ANTI-SULFUR CHARACTERISTICS AEC-Q200 QUALIFIED RATED POWER UPGRADE IN SMALLER PACKAGE SIZE AVAILABLE IN ± TOLERANCE BOTH FLOW SOLDER AND REFLOW SOLDERING ARE APPLICABLE

More information

Contact person Date Reference Page Mikael Videby P08110A 1 (6) SP Structural and Solid Mechanics

Contact person Date Reference Page Mikael Videby P08110A 1 (6) SP Structural and Solid Mechanics issued by an Accredited Testing Laboratory Contact person Mikael Videby 2015-11-13 5P08110A 1 (6) SP Structural and Solid Mechanics +46 10 516 50 36 Mikael.Videby@sp.se Accred. No. 1002 Testing ISO/IEC

More information

TABLE OF CONTENTS 1. FEATURES SPECIFICATIONS... 1

TABLE OF CONTENTS 1. FEATURES SPECIFICATIONS... 1 TABLE OF CONTENTS 1. FEATURES... 1 2. SPECIFICATIONS... 1 3. FRONT PANEL DESCRIPTIONS... 3 3-1 Display...3 3-2 Power Switch... 3 3-3 ma ( Red ) Input Terminals...3 3-4 COM ( Black ) Input Terminals...3

More information

IESNA LM MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT. For

IESNA LM MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT. For IESNA LM-80-2008 MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT For Fujian Lightning Optoelectronic Co., Ltd.Shenzhen Branch 5F, Building B, second phase of Chuangjian Industrial

More information

Recordkeeping Requirements of the Federal Motor Carrier Safety Regulations

Recordkeeping Requirements of the Federal Motor Carrier Safety Regulations Recordkeeping Requirements of the Federal Motor Carrier Safety Regulations The following table summarizes the recordkeeping requirements of the Federal Motor Carrier Safety Regulations under 49 CFR Parts

More information

l-tech C O M P A N Y Issue by Engineering Department Product 8.4"Rugged LCD : Model WRD0840 Product Description Test Reason

l-tech C O M P A N Y Issue by Engineering Department Product 8.4Rugged LCD : Model WRD0840 Product Description Test Reason l-tech Thermal Test Report Issue by Engineering Department Product Model Product Description Report No:0807TT0239 8.4"Rugged LCD : WRD0840 New product Rugged Display A/D Board :i-tech / R5W-310 Rugged

More information

Report Test Report

Report Test Report ÖTI Institut für Ökologie, Technik und Innovation GmbH Report 70535 Test Report Applicant FIDIVI Tessitura Vegnano S.p.A. Regione Masio 19 / bis 10046 Poirino (TO) ITALIEN Reference Mr. Alessandro Avenati

More information

NIJ s Body Armor Testing Program 8 th Annual Technologies for Critical Incident Preparedness September 7, 2006

NIJ s Body Armor Testing Program 8 th Annual Technologies for Critical Incident Preparedness September 7, 2006 NIJ s Body Armor Testing Program 8 th Annual Technologies for Critical Incident Preparedness September 7, 2006 Debra Stoe Physical Scientist Standards and Testing Program Office of Science & Technology

More information

TEST REPORT Reference No. : TRHZ _R1 Date : Nov. 27, 2012 Page No.: 1 of 5

TEST REPORT Reference No. : TRHZ _R1 Date : Nov. 27, 2012 Page No.: 1 of 5 Reference. : TRHZ0630_R Date : v. 27, 202 Page.: of pplicant : ddress : HINING LOFTY XUN BNNER TEXTILE CO., LTD. NO.70-,LONGXING ROD, HICHNG STREET, HINING CITY, ZHEJING, CHIN. The following merchandise

More information

SmartSensor 105 & Advance NEMA Salt Fog

SmartSensor 105 & Advance NEMA Salt Fog SmartSensor 105 & Advance NEMA Salt Fog Test Report Wavetronix products mentioned: SmartSensor 105 SmartSensor Advance As contracted by Wavetronix, Environ Laboratories LLC performed a NEMA salt fog test

More information

SPECIFICATION FOR REWIRABLE FUSES (KITKAT FUSES) 100A & 200A SPECIFICATION NO.: DIST/ A- KIT /

SPECIFICATION FOR REWIRABLE FUSES (KITKAT FUSES) 100A & 200A SPECIFICATION NO.: DIST/ A- KIT / SPECIFICATION FOR REWIRABLE FUSES (KITKAT FUSES) 100A & 200A SPECIFICATION NO.: DIST/ A- KIT / 2005-06 SR.NO ITEM PAGE 1. SCOPE 2 2. WEATHER CONDITIONS 2 3. RATED VOLTAGE & FREQUENCY 2 4. GENERAL REQUIREMENT

More information

1/11 表單編號 QS-3Q B

1/11 表單編號 QS-3Q B Battery Pack Test Report UN38.3 Customer: Lenovo Pack Model: L15C4A02 Nominal voltage: 14.4V Nominal capacity: 32Wh Configuration: 4S1P Customer P/N: 5B10L04163 Celxpert P/N: 921300080 Cell Type: LG INR18650S3

More information

SECULIFE PS100 PATIENT SIMULATOR

SECULIFE PS100 PATIENT SIMULATOR Operating Instructions SECULIFE PS100 PATIENT SIMULATOR 3-349-714-03 1/05.13 TABLE OF CONTENTS Page WARNINGS, CAUTIONS, NOTICES... 4 DESCRIPTION SECULIFE PS100 Patient Simulator... 7 LAYOUT... 9 LIMITED

More information

GCE AS and A Level. Physics A. AS exams 2009 onwards A2 exams 2010 onwards. Unit 5C: Approved specimen question paper. Version 1.1

GCE AS and A Level. Physics A. AS exams 2009 onwards A2 exams 2010 onwards. Unit 5C: Approved specimen question paper. Version 1.1 GCE AS and A Level Physics A AS exams 2009 onwards A2 exams 2010 onwards Unit 5C: Approved specimen question paper Version 1.1 Surname Other Names Leave blank Centre Number Candidate Number Candidate Signature

More information

Test Report Current (Gen 2) Pendant Armor and Gen 3 Pendant Armor Polymer A Hanger and Polymer B Hanger March 16, 2016 (Q022516U1R2) Testing Location

Test Report Current (Gen 2) Pendant Armor and Gen 3 Pendant Armor Polymer A Hanger and Polymer B Hanger March 16, 2016 (Q022516U1R2) Testing Location Client Roboworld Molded Products, LLC 8216 Pricinceton-Glendale Road, Suite 246 West Chester, OH 45069 Test Report Current (Gen 2) Pendant Armor and Gen 3 Pendant Armor Polymer A Hanger and Polymer B Hanger

More information

mga research corporation

mga research corporation mga research corporation UNDERGROUND TECHNOLOGIES MAN-HOLE CONSTRUCTION VERTICAL STATIC LOAD TEST SERIES (PO# 11521) 446 executive drive troy, mi 48083 248 / 577-5001 fax 248 / 577-5025 www.mgaresearch.com

More information

IESNA LM MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT. For

IESNA LM MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT. For IESNA LM-80-2008 MEASURING LUMEN MAINTENANCE OF LED LIGHT SOURCES MEASUREMENT AND TEST REPORT For No.1, Xianke Yi Road, Huadong Town, Huadu District, Guangzhou, China Model:40B18C12(Ra2) Report Type: 10000

More information

No , Vers.1

No , Vers.1 Pag. 1 of 4 TEST REPORT No. Test performed in accordance with the standard ASTM B 117 (2003) issued by TÜV Rheinland Italia S.r.l. - Via Gavardina Trav. II, 42 I-25010 Ponte S. Marco (BS) 1. General information

More information

TECHNICAL SPECIFICATION FOR 11 KV CAPACITOR BANK

TECHNICAL SPECIFICATION FOR 11 KV CAPACITOR BANK TECHNICAL SPECIFICATION FOR 11 KV CAPACITOR BANK MAHARASHTRA STATE ELECTRICITY DISTRIBUTION COMPANY LTD. PAGE 1 OF 7 TECHNICAL SPECIFICATION FOR 11 KV.CAPACITOR BANK 1.0 SCOPE: This specification covers

More information

Beamex MB METROLOGY TEMPERATURE BLOCK. Portable temperature dry block delivering bath-level accuracy for industrial applications

Beamex MB METROLOGY TEMPERATURE BLOCK. Portable temperature dry block delivering bath-level accuracy for industrial applications Beamex MB METROLOGY TEMPERATURE BLOCK 87 Portable temperature dry block delivering bath-level accuracy for industrial applications Highly accurate temperature dry block 88 Portable temperature dry block

More information

Heating chamber with mechanical convection

Heating chamber with mechanical convection FED series 720 Drying ovens / Heating ovens Heating chamber with mechanical convection The multi-talented BINDER heating chamber of the FED series provides almost unlimited capacity and is particularly

More information

Electrical Stability Meter. Part No Instruction Manual. Updated 12/10/2015 Ver. 3.0

Electrical Stability Meter. Part No Instruction Manual. Updated 12/10/2015 Ver. 3.0 Electrical Stability Meter Part No. 131-50 Instruction Manual Updated 12/10/2015 Ver. 3.0 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. Houston, Texas 77065 U.S.A. Tele: 832.320.7300 Fax: 713.880.9886

More information

United States Stove Company Project # Model: Breckwell SP1000 Type: Pellet-Fired Room Heater July 25, 2018

United States Stove Company Project # Model: Breckwell SP1000 Type: Pellet-Fired Room Heater July 25, 2018 United States Stove Company Project # 18-415 Model: Breckwell SP1000 Type: Pellet-Fired Room Heater July 25, 2018 ASTM E2779 Standard Test Method for Determining Particulate Matter Emissions from Pellet

More information

INVITATION FOR QUOTATION. TEQIP-II/2015/GJ1G02/Shopping/M-EE-09

INVITATION FOR QUOTATION. TEQIP-II/2015/GJ1G02/Shopping/M-EE-09 INVITATION FOR QUOTATION TEQIP-II/2015/GJ1G02/Shopping/M-EE-09 02-Dec-2015 Sub: Invitation for Quotations for supply of Goods Dear Sir, 1. You are invited to submit your most competitive quotation for

More information

BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA Ph ; Fax

BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA Ph ; Fax BERKELEY ANALYTICAL 815 Harbour Way South, Suite 6 Richmond, CA 94804-3614 Ph. 510-236-2325; Fax 510-236-2335 E-mail info@berkeleyanalytical.com Product Sample Formaldehyde Emissions Customer & Building

More information

Pursuing high precision, speed, and ease of use

Pursuing high precision, speed, and ease of use Carbon/Sulfur Analyzers for Solid Materials EMIA series Pursuing high precision, speed, and ease of use Carbon / Sulfur Analyzer High-frequency induction furnace type EMIA-920V2/320V2 Series New EMIA-V2/FA

More information

Guaranteed Technical Particulars(GTP) Deviation/Remarks Specify if any. Computerized Motor testing system suitable to test LV 3 phase motors

Guaranteed Technical Particulars(GTP) Deviation/Remarks Specify if any. Computerized Motor testing system suitable to test LV 3 phase motors Section II - Technical Specification Tender Enquiry No: BUY/EATD(BEE)-01/12-13 Name of the Equipment: MOTOR TESTING SYSTEM Note: The Offers should be submitted only in this format otherwise the offer will

More information

TYPE APPROVAL CERTIFICATE

TYPE APPROVAL CERTIFICATE TYPE APPROVAL CERTIFICATE Certificate No: A-14136 File No: 899.60 Job Id: 262.1-018818-1 This is to certify: That the Peripheral Equipment with type designation(s) Ha-VIS econ 2000 series, econ 3000 series

More information

United States Stove Company Project # Model: KP60 AKA: AP60, VG60 Type: Pellet-Fired Room Heater May 31, 2018

United States Stove Company Project # Model: KP60 AKA: AP60, VG60 Type: Pellet-Fired Room Heater May 31, 2018 United States Stove Company Project # 18-409 AKA: AP60, VG60 Type: Pellet-Fired Room Heater May 31, 2018 ASTM E2779 Standard Test Method for Determining Particulate Matter Emissions from Pellet Heaters

More information

SFI SPECIFICATION 35.2 EFFECTIVE: DECEMBER 29, 2014 *

SFI SPECIFICATION 35.2 EFFECTIVE: DECEMBER 29, 2014 * SFI SPECIFICATION 35.2 EFFECTIVE: DECEMBER 29, 2014 * PRODUCT: Heavy Duty Stock Car Steel Wheels 1.0 GENERAL INFORMATION 1.1 This SFI Specification establishes uniform test procedures and minimum standards

More information