Model 2110 Fraction Collector Instruction Manual

Size: px
Start display at page:

Download "Model 2110 Fraction Collector Instruction Manual"

Transcription

1 Model 2110 Fraction Collector Instruction Manual Catalog Numbers

2 Warranty Model Serial No. Date of Delivery Warranty Period Unless otherwise specified, instruments sold by Bio-Rad Laboratories are under warranty for 1 year against defects in materials and workmanship. If any defects should occur during this warranty period, Bio-Rad will replace the defective parts without charge. However, the following defects are specifically excluded: 1. Defects caused by improper operation 2. Repair or modification done by anyone other than Bio-Rad Laboratories or their authorized agent. 3. Use of fittings or other spare parts supplied by anyone other than Bio- Rad Laboratories. 4. Damage caused by deliberate or accidental misuse. 5. Damage caused by disaster. 6. Damage due to use of improper solvent or sample. This warranty does not apply to tubing and fuses. For inquiry or request for repair service, contact Bio-Rad Laboratories after confirming the model and serial number of your instrument. For Technical Service Call Your Local Bio-Rad Office or in the U.S. call BIORAD( )

3 Table of Contents Section 1 Safety... 1 Section 2 Introduction... 2 Section 3 Unpacking and Voltage Conversion Unpacking Instructions Voltage Conversion... 3 Section 4 Physical Description and Control Features Physical Description Front Panel Functions Rear Panel Connections... 8 Section 5 Instrument Set-Up Tubing Installation Carousel Installation Collection Tube Installation Waste Collection Port Section 6 Stand Alone Operation Stop (Standby) and Run Collection Modes Operation without Direct Connection to a System Manual Advance Reset/End of Run Section 7 Remote Operation of the Model 2110 Fraction Collector Remote Operation with Model EP-1 Econo Pump Remote Operation with BioLogic LP Remote Operation with BioLogic system Remote Operation with other Instruments Section 8 Accessories Instrument Dust Cover Micro Tube Adaptor Section 9 Cleaning and Maintenance Drop Sensor Instrument Case, Carousel, and Drainage Tube Section 10 Troubleshooting Appendix A Technical Specifications... 21

4 Appendix B Model 2110 Fraction Collector Procedure for Adjusting Drop Detector Signal Level Appendix C Ordering Information... 24

5 !! Section 1 Safety Disconnect supply before servicing. No user serviceable parts inside, refer servicing to Bio-Rad service personnel. High leakage current, ensure proper grounding. This instrument is intended for laboratory use only. This product conforms to the Class A standards for electromagnetic emmissions intended for laboratory equipment applications. It is possible that emissions from this product may interfere with some sensitive appliances when placed nearby or in the same circuit as those appliances. The user should be aware of this potential and take appropriate measures to avoid interference. 1

6 Section 2 Introduction Fig Model 2110 Fraction Collector. The Model 2110 Fraction Collector (Figure 2.1), shown with the optional instrument dust cover, is designed to be used with gravity feed columns, preparative electrophoresis and liquid chromatography systems. The Model 2110 Fraction Collector will accurately collect up to 80 fractions, at a flow rate of up to 40 ml/min. The Model 2110 Fraction Collector accepts standard 13 x 100 mm test tubes or, with an optional adaptor, 1.5 ml micro test tubes. Tubes are advanced by a motor-driven 2

7 carousel where tube changes are initiated upon reaching a predetermined drop count (1-999 drops) or time interval ( minutes), or upon command from an external source. The silicone drop former insures uniform drop size for accurate drop count measurement. The drop former will accommodate either 1 ¼8-inch or 1 ¼16-inch inlet tubing, without the need for additional fittings. The Model 2110 Fraction Collector is spill-proof and solvent resistant. A drain tube at the rear of the case channels all spills to waste. The membrane front panel is sealed to the case and the carousel drive is positioned in such a way that it will not be affected by spills. Section 3 Unpacking and Voltage Conversion 3.1 Unpacking Instructions Lift the instrument out of its packing by grasping the instrument case. Do not lift by the drop arm. Remove the power cord and bag of accessories. Remove the fraction collector from its protective bag. Check off all parts against the supplied packing list. You may want to retain the original packaging material in the event the instrument needs to be returned for service. If repackaging is necessary, be careful to position the carousel so that it is not resting against the motor drive gear. Warning: The Model 2110 Fraction Collector is shipped in its 120 V version. For 220 V operation, refer to the procedure below. 3.2 Voltage Conversion Prior to connecting the power cord to the power entry module and power mains, make sure the voltage indicated on the fraction collector power entry module matches your line voltage. If it does not, use the following procedure to make the conversion. (See Figure 3.1.) 1. Insure that the unit is disconnected from the wall socket. 2. Remove the fuse drawer with a small-blade screwdriver or similar tool. 3. Pull the fuse holder out of the fuse drawer and ensure that the voltage shown in the window corresponds to the voltage you are using (100, 120, 220, 240 V ). To change the voltage, grasp the fuse drawer and withdraw it (see Figure 3.1). Rotate the fuse drawer so that the voltage shown is correct, and reinsert it. The fuse holder is keyed and cannot be installed improperly. The Model 2110 Fraction Collector uses 0.20 A fuses for 100/120 V operation, and 0.10 A fuses for 220/240 V operation. 3

8 4. Reinsert the fuse drawer in the power entry module, with the locking tab to the left. The voltage indicator will read right-side-up if the drawer is oriented correctly. Push the fuse drawer in gently until it snaps into place. 5. Connect the power cord to the power entry module and power mains. 6. There is no power switch on the fraction collector. When connected to an AC power source, the Model 2110 Fraction Collector will come on in the Stop or Set mode. Fuse drawer (110 V configuration) Fig Voltage conversion. 4 Fuse drawer (220 V configuration)

9 Section 4 Physical Description and Control Features 4.1 Physical Description Inlet tubing Arm cap Drop former Drop arm Carousel positioning guides Drive seal Drive gear Pivot shaft Slider Fig 4.1. Model 2110 Fraction Collector features. Feature Drop arm Function Directs flow to the collection tubes. Collection position, rotated forward, the drop arm is directly above each collection tube. Standby position, rotated rearward, the drop arm allows installation of the carousel or allows flow to drip onto the case and to waste. To change positions, gently lift and rotate the drop arm. 5

10 Feature Carousel Drop former Arm cap Drive gear Drive seal Waste line Inlet tubing Function Holds the collection tubes. The tube positions are numbered 1-80 on the flange of the carousel, and line up with the center of each tube. The carousel is designed so that any liquid spilled in it will flow to the center and out the rear of the base. The carousel may also be used for storage of test tubes. Seals against the inlet tubing ( 1 ¼16" or 1 ¼8" OD) to form uniform size drops. Holds the drop former in place. Engages the carousel and advances it. Protects the electric drive from direct spills. Collects drainage or spills and directs it to waste. See Section 5.5 for connection. Inlet tubing (not included) can be any rigid or flexible tubing with 1 ¼16" or 1 ¼8" OD, usually coming directly from a detector. 6

11 4.2 Front Panel Functions MODEL 2110 FRACTION COLLECTOR Fig Front panel controls. Feature Run/Stop key Time/Drop key Function The Run/Stop key is used to select standby or active mode. In the active mode, the collector will collect by time or drops, or accept an external signal. When the collector is controlled by a system, both the standby and active mode lights will be lit, and the LED display will read ---. The Time/Drop key is used to select collection by time or by drops. ml min ml/min % Arrow keys LED multi display Advance key The Arrow keys are used for increasing or decreasing the length of time for collection in each tube, or changing the number of drops to be collected in each tube. This three-digit display will show decimal minutes or drop count that has progressed for the tube currently collecting. When the collector is controlled by a system, both the standby and active mode lights will be lit, and the LED display will read ---. Pressing the advance key will advance the carousel one position. 7

12 Fig. 4.3 Rear Panel connections. 4.3 Rear Panel Functions Connection Power entry module Function Grounded 3-pin receptacle for the power cord. Also contained within is a four-position 100 V/120 V/ 220 V/ 240 V line voltage selector. The collector is shipped in its 120 V or 220 V version. For operation at other voltages, refer to Section pin D connector For remote operation of the collector. The Model 2110 can be controlled by Bio-Rad s Model EP-1 Econo Pump, the BioLogic LP low pressure chromatography system, or the BioLogic System, a high resolution liquid chromatography system. The Model 2110 can also be controlled by other non-bio-rad equipment. See Section 7 for additional information. 8

13 Section 5 Instrument Set-Up 5.1 Tubing Installation 1. Insert the drop former, narrow end downward, into the drop arm, pressing it firmly in the hole (see Figure 4.1). 2. Using a twisting motion, press the arm cap onto the drop arm to hold the drop former in place. 3. Insure that the end of the inlet tubing is cleanly cut and not flattened or deformed. Insert tubing into the drop former, pushing down on the tubing until the drop former starts to stretch. Do not use excessive force. Then, gently pull out on the tubing until the drop former returns to its original shape. 4. The inlet tubing may be routed through the tubing holder fitting at the rear of the movable portion of the drop arm. 9

14 a b 2 1 c Fig Carousel installation. 10

15 5.2 Carousel Installation The Model 2110 Fraction Collector fills tubes from the outside of the carousel inward. Load as many tubes as desired into the carousel, starting at the outside end of the spiral and moving inward. The carousel need not be fully loaded, since any eluant flowing after the last tube will drain through the carousel into the well in the case and out the central drain tube located at the rear of the instrument. 1. Swing the drop arm to the waste (rear) position by gently lifting and rotating (Figure 5.1a). 2. Hold the carousel so that the notch in its outline is toward the rear of the instrument. 3. Place the carousel on the pivot shaft, holding it about 1 ¼2" above its seated position. 4. Push the carousel and the pivot shaft toward the rear of the instrument, settling it down over the drive gear so that the gear is trapped near the start of the spiral. 5. Holding the gear teeth free from engagement, turn the carousel counterclockwise until it stops against the gear (Figure 5.1b). Release the carousel and it will spring into engagement on the drive gear. Engagement may be confirmed by trying to move the carousel gently (Figure 5.1c). 6. Check the alignment with the two positioning guides located on the top front of the instrument case. The number corresponds to the first tube that will be filled. Swing the drop arm into the collect (front) position. Make sure the drive gear is fully seated into the grooved underside of the carousel. This automatically aligns the drop former over the first tube. 5.3 Collection Tube Installation Load the carousel with up to eighty 13 x 100 mm test tubes. You may find a tube which fits too tightly into the carousel. It is best to discard these rather than risk breaking the tube upon installation. Tubes 39 through 53 are raised above tubes 1-38 to make tube removal easier. To locate an individual tube in a carousel, find its number on the angled flange. The tube is in line with the number and the center of the carousel. The top row of numbers corresponds to the outermost row of tubes, the middle row corresponds to the next row of tubes, and the bottom row to the innermost row of tubes. To divert a large amount of eluant to waste at the beginning of a run, you can simply remove the first 11

16 collection tube, and the eluant will flow through the carousel to waste out the rear of the collector. 5.4 Waste Collection Port All spills on the top of the case and carousel, as well as waste eluant within the carousel, drain to a central waste-collection port at the rear of the instrument case. From this port, a drain tube passes out through the rear of the case. This tube may be directed into a waste-collection vessel, or an extension tube may be attached to the provided barbed union fitting. Note: To provide proper drainage, the waste-collection vessel must be positioned so that the exit of the drain tube is no higher than the bottom of the fraction collector case. Section 6 Stand Alone Operation This section describes the stand-alone operation of the Model 2110 Fraction Collector. Refer to Section 7 for information about connecting the Model 2110 to the Model EP-1 Econo Pump, the BioLogic LP, the BioLogic System, or Bio-Rad s preparative electrophoresis systems Stop (Standby) and Run The Model 2110 Fraction Collector powers up automatically in the Stop or Standby mode, with the LED lit. When in the Stop mode, the collection mode, either Time or Drops can be selected, and the time interval or number of drops to be collected per tube can be set. When in the Run mode, the Arrow keys are inactivated. 6.2 Collection Modes The Model 2110 Fraction Collector will collect fractions by counting the number of drops or by a specified time interval per tube. The collection mode is selected using the key on the front panel. In either mode, time and drop count are continuously monitored, so that you can switch between the two collection modes, monitoring time and drop count since the last tube advance. 12

17 Drop Collection Mode The fraction collector counts drops at flow rates up to ~5 ml/min. Drop counting is performed by the sensor in the drop arm. If the flow rate exceeds 5 ml/ min., counting will not be accurate, and use of the Time mode is recommended. 1. Plug the Model 2110 Fraction Collector into a grounded wall outlet. The collector will power up in the Stop, or Standby mode, with the LED lit. 2. Press the Time/Drop key until the LED is lit. 3. Select the number of drops to be collected per collection tube by pressing the keys. The LED display will display the number of drops to be collected. 4. Insure that the carousel is in the desired position, and press the Run/Stop key to begin collection. The LED will be lit. 5. The fraction collector will reset the display to 0 and begin counting drops up to the drop limit set, and then advance. The display reads A during each carousel advance. 13

18 Time Mode The fraction collector collects fractions at intervals of.01 minutes, for times per tube in the range of.05 to 9.99 minutes, and at.10 minute intervals for times per tube in the range of 10.0 to 99.9 minutes. 1. Press the Run/Stop key until the LED is lit. 2. Press the Time/Drop key until the LED is lit. ml min ml/min % 3. Select the time interval per tube by pressing the keys. The LED multi-display will display the time interval selected. 4. Insure that the carousel is in the desired position. Press the key to begin the program. The LED will be lit. 5. The fraction collector will reset the display to 0, and begin counting time up to the set time interval, and then advance. Note: When the Model 2110 Fraction Collector reaches the preset time interval in a time mode collection scheme, it will collect one extra drop prior to advancing. This minimizes the chance of a drop falling between tubes during carousel advance. If the delay between drops is greater than 5% of the preset time interval, the collector will advance to the next fraction instead of waiting for one extra drop. 6.3 Operation without Direct Connection to a System The Model 2110 can be used in stand-alone mode with any system. Simply set up the collector in time mode or drop mode and press start at the initiation of the method. Alternatively, a light-weight gravity feed column (such as 2 ml Poly-Prep columns, Econo-Pac columns, and small Econo-Column chromatography columns) can be connected directly to the Model 2110 Fraction Collector, avoiding the need for interconnecting tubing. A two-way stopcock is used to make the connection and control eluant flow. 14

19 1. Attach the stopcock into the hole in the top of the drop arm. Position the stopcock so that its handle overhangs the end of the arm and is free to turn. 2. Attach the column to the stopcock, making sure both stopcock and column are firmly connected. 6.4 Manual Advance The Model 2110 Fraction Collector may be manually advanced to the next tube by pressing the key. This key is functional in both Stop/Standby and Run modes. When pressed during a run, the drop and time counts will automatically reset to 0, and restart the time or drop count. 6.5 Reset/End of Run If the Model 2110 Fraction Collector is allowed to complete 80 advances, the display will read END. To reset, push twice. (Note: Manual advances are not counted.) Reposition the carousel to begin collection again. Section 7 Remote Operation of Model 2110 The Model 2110 Fraction Collector may be controlled remotely. 7.1 Remote Operation with Model EP-1 Econo Pump Use System Cable 1 (catalog number ) to connect the Econo Pump to the fraction collector. Connect the 9-pin D connector to rear of the Model 2110, and the mini-din plug to the rear port of the Econo Pump with the collector icon. (See Econo Pump manual if further explanation is required.) In order for the Econo Pump to remotely operate the Model 2110, a method must be programmed into the Econo Pump that calls for fraction collection. You will be prompted to program the fraction size (volume), the void volume, and the total volume collected. Please note that to correctly use the void volume function, a Model SV-3 Diverter/Bypass valve (catalog number ) must be connected. When programming is complete, the LED display underneath the fraction collector icon on the front panel of the Econo Pump will light and the Model 2110 display will read --- indicating that the collector is controlled by a system. 15

20 7.2 Remote Operation with BioLogic LP Use System Cable 1 (catalog number ) to connect the BioLogic LP to the fraction collector. Connect the 9-pin D connector to rear of the Model 2110, and the mini-din plug to the rear port of the BioLogic LP controller in the port marked FRACTION COLLECTOR. (See BioLogic LP manual if further explanation is required.) The BioLogic LP enables sophisticated collection, including programmed collection windows, peak detection, and a combination of peak detection and collection windows. The BioLogic LP sends advance commands to the fraction collector. 7.3 Remote Operation with BioLogic System Use System Cable 5 (catalog number ) to connect the BioLogic System to the fraction collector. Connect the 9-pin D connector to rear of the Model Connect the black wire to the FC ADV port of the AUX connector, which is port #5. Connect the white wire to the GND port of the AUX connector, which is port #9. Refer to the separate instruction sheet for the wiring diagram and color code for System Cable 5. To connect the wires to the AUX port of the BioLogic System, turn the small flathead screwdriver counterclockwise to back out the screw. Insert the bare wire into the port and turn the screwdriver clockwise to secure the connection. The BioLogic System enables sophisticated collection, including programmed collection windows, peak detection, and a combination of peak detection and collection windows. The BioLogic System sends advance commands to the fraction collector. 7.4 Remote Operation with other Instruments Use System Cable 5 (catalog number ) to connect the Model 2110 Fraction Collector to other instruments. Signals are accessible via the 9-pin D connector on the back of the instrument. All of the input and output signals, with the exception of the Advanced Contact pins 7 and 8, are TTL active low. Drop counting is performed by the sensor in the drop arm. Pin Function Signal Type 1 External Advance Input 2 Drops Output 3 Advanced Output 4 Ready Output 5 Alarm Output 16

21 Pin Function Signal Type 6 External Start Input 7 Advanced Contact 1 Output 8 Advanced Contact 2 Output 9 Logic Ground External Advance (Pin 1) The External Advance line performs two functions. It causes the instrument to enter the Remote mode when held low for longer than two seconds, and it causes the carousel to advance one position each time a high-going pulse is applied. To use the control line to advance the carousel, connect pin 1 of the 9-pin D connector to pin 9 and hold for at least 100 msec, then release. Alternatively, the carousel will advance by applying an external signal (TTL low logic level ) to pin 1 for at least 100 msec (with a return to TTL high logic level state). Be sure to connect the controlling instrument s ground to the fraction collector s logic ground, pin 9. When remote mode is entered (by holding pin 1 low for greater than 2 seconds) the time and drop mode features of the collector are disabled and the front panel control function is lost, with the exception of the manual tubing advance key, which will function normally. The remote mode is indicated by three horizontal bars on the fraction collector display. Drops The fraction collector will generate a 50 msec TTL pulse on pin 2 every time a drop is detected. Advanced Each time the carousel is advanced, a low-going TTL pulse is generated on pin 3. This signal is useful for triggering the event mark input of a chart recorder to indicate fraction intervals. If the recorder being used does not support TTL signals, then see below under Advanced Relay Contacts to trigger the event input. Ready A low level signal on pin 4 indicates that the fraction collector is functioning properly and that it is ready to accept input from a controlling device. Alarm If the fraction collector is unable to carry out an external advance command due to a malfunction (such as a jammed carousel), a low level will be held on pin 5 until the problem is cleared. 17

22 External Start The external start line is used to allow another instrument to remotely start and stop the fraction collector s programmed functions. Applying a low pulse of at least 100 msec to pin 6 has the same effect as pressing the Run/Stop key on the front panel. That is, the first pulse will start the previously selected mode (drops or time) and the second pulse will stop the program. Advanced Relay Contacts Pins 7 and 8 provide a normally open set of contacts that are momentarily closed each time the carousel is advanced. This signal is a convenient way to interface with equipment that does not support TTL level signals. The maximum current and voltage which can be put across these pins is 24 V DC and.25 A. Section 8 Accessories 8.1 Instrument Dust Cover The dust cover (catalog number ) is intended for use only when the drop arm is in the collect (forward) position. The inlet tube may be directed out the top opening or through the small opening at the bottom/rear of the instrument. 8.2 Micro Tube Adaptor The micro tube adaptor is intended to convert a carousel for use with 1.5 ml micro test tubes (catalog number ). Align the step in the outline of the adaptor with the step of the outline of the carousel. Position the adaptor so that its four legs fit into the four openings in the inner wall of the carousel. Push the adaptor downward until it is fully seated on the carousel. Operation with micro tubes is identical to operation with standard test tubes. To prevent splashing, collect no more than 1 ml of eluant in each micro tube. 18

23 Section 9 Cleaning and Maintenance The Model 2110 Fraction Collector requires very little maintenance to insure reliable operation. Two procedures are outlined below. 9.1 Drop Sensor The Model 2110 Fraction Collector counts drops through an optical sensor located in the drop area inside the drop former arm. Precipitates or other foreign debris may obscure the optical path and cause erroneous drop counts. To prevent this problem, the drop sensor should be cleaned periodically. An ordinary cotton swab moistened with water can be used to clean the sensor. 9.2 Instrument Case, Carousel, and Drainage Tube During normal operation, spills and splashes may result in precipitates forming on the case and carousel. Use a squirt bottle and soapy water to wash down carousel and case. The water will drain out the back of the unit through the drainage tube. Prior to squirting the instrument case, make sure the rubber seal around the gear device is in place. Insure that the waste tube is draining properly to avoid instrument damage. Section 10 Troubleshooting Problem Cause Remedy Drops miss tubes Carousel misaligned Reposition carousel Drop former missing Replace drop former Drop head misaligned Align Unit not level Place unit on level surface Drops miss tube Flow rate greater Reduce flow rate during advance than 5 ml/min Drops not counted Drop sensor is dirty Clean drop sensor or E08 error code Flow rate greater than Use Time collection mode 5 ml/min Operating in a windy Try relocating unit, or put environment a piece of tape over slot in drop arm No power Fuses Check fuses, replace if necessary 19

24 Fraction Collector Error Codes There are several self-diagnostics built into the software of the Model 2110 Fraction Collector. Below are the principal error numbers and codes which occur. If other error codes appear, try unplugging and replugging the instrument, or call Bio-Rad Technical Service. 20

25 Error Number Cause Remedy E00 Hardware failure Unplug and replug to E01 E04 mains power. If error persists, contact your local Bio-Rad representative. E02 Motor jam Reposition carousel; unplug and replug to mains power. If problem persists, apply WD-40 to the motor shaft. If motor must be replaced, order part number E08 Drop sensor If problem persists after E10 cleaning the drop sensor with a wetted cotton swab, follow the Procedure for Adjusting Drop Detector Signal Level. See Appendix B. If drop sensor must be replaced, order part number Appendix A Technical Specifications Capacity Collection basis Materials of construction Eighty 13 x 100 mm glass or plastic test tubes, or 1.5 ml micro test tubes, when using adaptor (catalog number ) 1 to 999 drops in 1 drop increments 0.05 to 9.99 minutes in 0.01 minute increments 10.0 to 99.9 minutes in 0.1 minute increments See Econo Pump manual for other collection options when connected to the Econo System Case and carousel: polypropylene Front panel: hard-coated polycarbonate Drop former: silicone rubber Waste outlet tube: polyurethane 21

26 Operating temperature + 4 to + 40 C Storage temperature - 5 to + 71 C Line voltage VAC or VAC, 47 Hz to 63 Hz Power consumption 25 VA max, 15 W max, fused Fuse rating 100,120 V Operation 200 ma, 250 V, 5 x 20 mm, Fast Acting, UL/CSA approved Ref: Schuter or Bussmann GMA200mA 220,240 V Operation 100 ma,250 V, 5 x 20 mm, Time Lag (FST), VDE approved Ref: Schuter Size 24.0 x 33.0 x 25.0 cm (W x D x H) Carousel does not overhang base at any point in operation Weight Output signals 2.3 kg See Section 6.5 of the manual Regulatory CSA: C22.2 No. 151-M1986 Compliance TÜV: EN : 1993 CE: EN55011: 1991 CLASS A, EN : 1995 Appendix B Model 2110 Fraction Collector Procedure for Adjusting Drop Detector Signal Level This procedure should be performed: 1. After replacing the drop arm on the Model 2110 Fraction Collector. 2. Whenever "EO8" is displayed. In this situation, clean the optical window and light source prior to performing this procedure. Safety: The Model 2110 Fraction Collector should be purged of all hazardous chemicals and be disconnected from the main power supply prior to conducting this procedure. Ensure the case is dry. Perform this procedure on a dry surface. During the actual adjustment of the signal level (step 10), potentially dangerous voltages are present inside the case - do not touch exposed wires. Use a plastic-handled screwdriver to minimize the risk of electric shock. 22

27 Tools required: driver. Small phillips-head screwdriver, small slot-head screw Procedure: 1. Disconnect the Model 2110 collector from the power source. 2. Disconnect any tubing leading to the fraction collector, remove the yellow plastic arm cap, white rubber drop former, and carousel. 3. Clean the optical window and light source (located in drop arm), using the a cotton swab moistened with water. 4. Set fraction collector on its side with underside of the instrument facing toward you. 5. Remove the four phillips-head screws from the underside of the instrument. Pull gently on the rubber feet to remove the underside panel. 6. Locate the large circuit board with the power transformer. On this board, locate a small rectangular potentiometer, identified as "R3" on the circuit board. R3 is located near where the ribbon cable connects to the circuit board. 7. Read safety warning above. Use a plastic-handled screwdriver to minimize the risk of electric shock. 8. Press and hold down both the "up" and "down" arrow keys on the front panel. 9. While continuing to press down arrow keys, connect the fraction collector to the power source. Continue to press down both keys for five seconds, then release.. The "time" LED will glow steadily, and the display will show a decimal point, followed by three characters (. _). The fraction collector is now in a "diagnostic" mode. 10. Using a small flat-bladed screwdriver, adjust R3 by turning the small screw until the display reads between "0A0" and "0A5". If a reading between "0A0" and "0A5" cannot be obtained, contact your local Bio-Rad representative. 11. Disconnect unit from the power source. 12. Replace underside panel. 13. Connect fraction collector to a power source. The Model 2110 Fraction Collector is now ready for operation. 23

28 Appendix C Ordering Information Model 2110 Fraction Collector Catalog Number Product Description Model 2110 Fraction Collector, 100/120 VAC (USA power cord), includes 80-tube carousel and tubing connectors Model 2110 Fraction Collector, 220/240 VAC (no power cord), includes 80 tube carousel and tubing connectors Carousel, 80 tube capacity Replacement Drop Former Kit, includes 2 drop and 2 drop caps Micro Tube Adaptor Instrument Dust Cover Econo System Cable 1, 8-pin Mini-DIN to DB-9 connector, to connect the Model 2110 Fraction Collector to the Econo Pump or BioLogic LP Econo System Cable 5, DB-9 connector to bare wires, to connect the Model 2110 Fraction Collector to the BioLogic System or other external instruments Econo System Cable 8, 8-pin standard DIN to DB-9 connector, for connecting the Model 2110 Fraction Collector to Model 1327 Recorder Model SV-3 Diverter/Bypass Valve Model EP-1 Econo Pump Model EP-1 Econo Pump, 100/120 VAC (USA power cord), includes tubing set and starter fittings kit Model EP-1 Econo Pump, 220/240 VAC (no power cord), includes tubing set and starter fittings kit Bio-Rad sells a complete line of integrated chromatography systems for protein purification. For information on the BioLogic System family of products, contact your local Bio-Rad representative. 24

29 Bio-Rad Laboratories Life Science Group Web site USA (800) 4BIORAD Australia Austria (01) Belgium Brazil Canada (905) China /68 Denmark Finland 358 (0) France Germany Hong Kong India (91-124) /113/114 Israel Italy Japan Korea Latin America Mexico to 54 The Netherlands New Zealand Norway Russia Singapore Spain Sweden 46 (0) Switzerland United Kingdom Bulletin 0000 US/EG Rev A Sig 1200 M Rev G

Cooling Module. Instruction Manual

Cooling Module. Instruction Manual Cooling Module Instruction Manual Catalog Numbers 170-3654, 170-3655, and 170-3688 For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-800-424-6723) Warranty Bio-Rad

More information

UView Mini Transilluminator

UView Mini Transilluminator UView Mini Transilluminator Instruction Manual Catalog #166-0531 Hardware Instruction Manual The UView mini transilluminator is suitable for research use only. It must be used by specialized personnel

More information

Valving Schemes for the BioLogic Duo-Flow Chromatography System

Valving Schemes for the BioLogic Duo-Flow Chromatography System Valving Schemes for the iologic Duo-Flow Chromatography System table of contents pplication asic ion Valve Operation using,, and Purge Positions position is for filling a loop with sample or for running

More information

Runwithit! The Mini-PROTEAN. 3 Mini Vertical Electrophoresis System is So Easy to Use, You Won t Want to Run With Anything Else.

Runwithit! The Mini-PROTEAN. 3 Mini Vertical Electrophoresis System is So Easy to Use, You Won t Want to Run With Anything Else. Runwithit! The Mini-PROTEAN 3 Mini Vertical Electrophoresis System is So Easy to Use, You Won t Want to Run With Anything Else. The User-Friendly 3 Mini-PROTEAN A Simple Approach to Gel Electrophoresis.

More information

Eldex Universal Fraction Collector Operator s Manual

Eldex Universal Fraction Collector Operator s Manual Eldex Eldex Universal Fraction Collector Operator s Manual Eldex Laboratories, Inc. 30 Executive Court Napa, CA 94558 Tel: (707) 224-8800 Fax: (707) 224-0688 www.eldex.com Rev. G: 020416 2016 Eldex Laboratories,

More information

Electrophoresis + Blotting + Imaging. V3 Western Workflow Visualize. Verify. Validate.

Electrophoresis + Blotting + Imaging. V3 Western Workflow Visualize. Verify. Validate. Electrophoresis + Blotting + Imaging V3 Western Workflow Visualize. Verify. Validate. Why choose the V3 Western Workflow from Bio-Rad? 1 2 Separate Proteins Visualize Separation Transfer Proteins Speed

More information

Plus Dodeca Cell System

Plus Dodeca Cell System Expression Proteomics PROTEAN Plus Dodeca Cell System 12 High-Resolution, Reproducible Second-Dimension Gels in an Incredible 6 Hours PROTEAN Plus Dodeca Cell Second-Dimension Gels in only 6 Hours The

More information

DynaLoop 25 and DynaLoop 90. Instruction Manual

DynaLoop 25 and DynaLoop 90. Instruction Manual 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 0 5 10 15 20 25 DynaLoop 25 and DynaLoop 90 Instruction Manual Catalog Numbers 750-0451 (25 ml) 750-0452 (90 ml) For technical support, call your

More information

BioLogic DuoFlow Valving Guide

BioLogic DuoFlow Valving Guide Protein Purification iologic DuoFlow Valving Guide Valving Schemes for the iologic DuoFlow Chromatography System Protein Purification iologic DuoFlow Valving Guide Introduction The iologic DuoFlow Chromatography

More information

NI PS-15/16/17 Side Mount Brackets

NI PS-15/16/17 Side Mount Brackets INSTALLATION GUIDE NI PS-15/16/17 Side Mount Brackets This document provides the installation procedure for the NI PS-15/16/17 Side Mount Brackets. This accessory is used to mount NI PS-15/16/17 power

More information

Fiber Optic Illuminator. Model 21AC Model 21DC

Fiber Optic Illuminator. Model 21AC Model 21DC Operation Manual Fiber Optic Illuminator Model 21AC Model 21DC Warning This symbol is intended to alert the user to the presence of uninsulated dangerous voltage within the product s enclosure that may

More information

NANOPAC-300 & 500 Power Supply. Instruction manual NANOPAC-300 & NANOPAC-500

NANOPAC-300 & 500 Power Supply. Instruction manual NANOPAC-300 & NANOPAC-500 NANOPAC-300 & 500 Power Supply Instruction manual NANOPAC-300 & NANOPAC-500 Version 01C Feb 5th, 2014 1 Packing list NANOPAC-300 or 500-1x NANOPAC-300 Power Supply or NANOPAC-500-1x Power Cord - 1x Instruction

More information

Packing Data Installation Instructions

Packing Data Installation Instructions Packing Data PK Installation Instructions This document describes how to install a Dataliner DL10 display in a panel or enclosure and connect power. European Union Directive Compliance wiring and safety

More information

RENA AF371Feeder Operating Manual. Feeder. Operating Manual. Manual Part #: M AF371 Operations Rev

RENA AF371Feeder Operating Manual. Feeder. Operating Manual. Manual Part #: M AF371 Operations Rev Manual Part #: M-3022 Feeder AF371 Operations Rev. 3-16-04 1 RENA AF371 Feeder YOUR RENA AF371 IS DISTRIBUTED BY RENA SYSTEMS INC. SERVICE AND SUPPORT FOR THIS PRODUCT IS PROVIDED BY YOUR RENA DEALER.

More information

Centrifuge Operator / Service Manual

Centrifuge Operator / Service Manual 3000 Centrifuge Centrifuge Operator / Service Manual cat.# 26230 & 26231 The Q-sep 3000 centrifuge complies with all requirements of UL standard 3101 20, Can/CSA C22.2 No. 1010.1, and Can/CSA C22.2 No.

More information

ASSEMBLY AND INSTALLATION INSTRUCTIONS

ASSEMBLY AND INSTALLATION INSTRUCTIONS ASSEMBLY AND INSTALLATION INSTRUCTIONS 356-9231 / 356-9232 WARNING: NOTES: 1. Before installing, consult local electrical codes for wiring and grounding requirements. 2. Customer Service: 1-800-887-6326

More information

Product Guide: Series III Pump Control Board Set (RoHS)

Product Guide: Series III Pump Control Board Set (RoHS) revised 04/08/10 Description: The Series III Pump Control Board Set provides motor drive and pump control for a wide assortment of pumps from Scientific Systems, Inc. The assembly consists of two circuit

More information

PL900 ILLUMINATOR. Fiber-Lite. Operation Manual. Setup. Fiber Optic Connection A PRODUCT OF DOLAN-JENNER INDUSTRIES. Voltage Selection:

PL900 ILLUMINATOR. Fiber-Lite. Operation Manual. Setup. Fiber Optic Connection A PRODUCT OF DOLAN-JENNER INDUSTRIES. Voltage Selection: PL900 ILLUMINATOR A PRODUCT OF DOLAN-JENNER INDUSTRIES Setup Voltage Selection: Operation Manual Read and review the instructions on the label covering the power entry module and the instructions in this

More information

Section 7 - Troubleshooting Guide

Section 7 - Troubleshooting Guide Section 7 - Troubleshooting Guide Section 7 - Troubleshooting Guide IMPORTANT While this troubleshooting guide provides information to aid in troubleshooting problems with the range, it does not contain

More information

Paraffin Dispenser Catalog # PD-120, PD-220 Operator s Manual Version 1.5b, Jan 2007

Paraffin Dispenser Catalog # PD-120, PD-220 Operator s Manual Version 1.5b, Jan 2007 Paraffin Dispenser Catalog # PD-120, PD-220 Operator s Manual Version 1.5b, Jan 2007 Be certain to read this manual thoroughly before proceeding with unpacking and installation. TABLE OF CONTENTS Topic

More information

M9106-AGx-2N0x Series Electric Non-spring Return Actuators

M9106-AGx-2N0x Series Electric Non-spring Return Actuators Installation Instructions M9106AGx-2N0x Issue Date 08/31/01 M9106-AGx-2N0x Series Electric Non-spring Return Actuators Installation IMPORTANT: The M9106-AGx-2N0x Series actuator is intended to control

More information

Amplification: T100 Thermal Cycler. T100 Thermal Cycler

Amplification: T100 Thermal Cycler. T100 Thermal Cycler Amplification: T100 Thermal Cycler T100 Thermal Cycler The Smart PCR Choice The T100 Thermal Cycler s intuitive touch screen makes running PCR easier than ever before. The T100 Thermal Cycler s performance,

More information

BT403. A Geno Technology, Inc. (USA) brand name. BT-300 Power Supply. Cat. No. BT

BT403. A Geno Technology, Inc. (USA) brand name. BT-300 Power Supply. Cat. No. BT BT403 A Geno Technology, Inc. (USA) brand name BT-300 Power Supply Cat. No. BT403 1-800-628-7730 1-314-991-6034 info@btlabsystems.com WARNING... 3 SAFETY INFORMATION... 3 ENVIRONMENTAL CONDITIONS... 4

More information

Lifting Instructions for 1336 PLUS and FORCE D Frame Drives

Lifting Instructions for 1336 PLUS and FORCE D Frame Drives Installation Instructions Lifting Instructions for 1336 PLUS and FORCE D Frame Drives This publication will guide you through the steps needed to properly lift and mount the following drives: 1336 PLUS

More information

M9206-Bxx-2S Series On/Off Electric Spring Return Actuators

M9206-Bxx-2S Series On/Off Electric Spring Return Actuators Installation Instructions M9206-Bxx-2S Issue Date May 14, 2003 M9206-Bxx-2S Series On/Off Electric Spring Return Actuators Installation IMPORTANT: Use this M9206-Bxx-2S Series actuator to control equipment

More information

Instruction Manual AVTM for. Strip Chart Recorder Catalog Nos and

Instruction Manual AVTM for. Strip Chart Recorder Catalog Nos and AVTM220003 Rev. B January 2003 Instruction Manual AVTM220003 for DC µa Strip Chart Recorder Catalog Nos. 220003 and 220003-47 PO Box 9007 Valley Forge, PA 19485-1007 U.S.A. 610-676-8500 Shipping Address:

More information

FD 150 Document Signer

FD 150 Document Signer FD 150 Document Signer OPERATOR MANUAL FIRST EDITION Thank you for purchasing the Formax FD 150 Document Signer The FD 150 has been tested and found to comply with the limits for a Class A digital device

More information

Plastic Tube Flar ing Syste m

Plastic Tube Flar ing Syste m www.swagelok.com Plastic Tube Flar ing Syste m Model F TF 1000 Safety Instructions.......................... 2 General Information......................... 2 Setup Tubing Selection and Preparation..............

More information

DH-2000 Deuterium-Halogen Light Source Installation and Operation Manual

DH-2000 Deuterium-Halogen Light Source Installation and Operation Manual DH-2000 Deuterium-Halogen Light Source Installation and Operation Manual For Products: DH-2000 / DH-2000-BAL / DH-2000-DUV / DH-2000-S / DH-2000-S-DUV / DH-2000-FHS, DH-2000-FHS-DUV /DH-2000-S-DUV-TTL

More information

Instruction Manual. Vortex Mixers

Instruction Manual. Vortex Mixers Instruction Manual Vortex Mixers TABLE OF CONTENTS Package Contents............... 1 Warranty............... 1 Installation............... 2 Maintenance & Servicing............... 2 Environmental Conditions...............

More information

Record the serial number and date of purchase in your manual for future reference.

Record the serial number and date of purchase in your manual for future reference. Air Filtration System Model: 61-200 Owner s Manual Record the serial number and date of purchase in your manual for future reference. Serial number: Date of purchase: For more information: www.rikontools.com

More information

INSTALLATION GUIDE NI 9921/9922 Outdoor IP Enclosure For NI WLS-9000 Series Devices

INSTALLATION GUIDE NI 9921/9922 Outdoor IP Enclosure For NI WLS-9000 Series Devices INSTALLATION GUIDE NI 99/99 Outdoor IP Enclosure For NI WLS-9000 Series Devices This guide describes how to install and use the National Instruments 99 and National Instruments 99 outdoor IP enclosures.

More information

INSTALLATION GUIDE NI WSN-3291 Outdoor IP Enclosure For NI WSN-32xx Series Devices

INSTALLATION GUIDE NI WSN-3291 Outdoor IP Enclosure For NI WSN-32xx Series Devices INSTALLATION GUIDE NI WSN-3291 Outdoor IP Enclosure For NI WSN-32xx Series Devices This installation guide describes how to install and use the National Instruments WSN-3291 outdoor IP enclosure. The NI

More information

Vetlab CombiSpin Centrifuge

Vetlab CombiSpin Centrifuge Vetlab CombiSpin Centrifuge User Manual 26.10.2017 Vetlab Supplies Ltd Unit 13 Broomers Hill Park Broomers Hill Lane Pulborough RH20 2RY Telephone: 01798 874567 Fax: 01798 874787 Email: info@vetlabsupplies.co.uk

More information

Ford Mustang V6 OEM-Style Fog Light Kit Parts List: Quantity: Tool List:

Ford Mustang V6 OEM-Style Fog Light Kit Parts List: Quantity: Tool List: 2015-2017 Ford Mustang V6 OEM-Style Fog Light Kit Parts List: Quantity: Tool List: LED Foglights/ Bezels 2 Flat head & Phillips screwdriver (if you ordered part#3600) Ratchet & Socket set OR Wiring harness

More information

FEATURES AND OPERATIONS

FEATURES AND OPERATIONS REMOTE STARTER - ANTI-START SYSTEM SECURITY SYSTEM For An Automatic Or Manual Transmission Vehicles Table Of Contents Welcome Important Information Remote starter with Automatic Transmission Keyless running

More information

TBI /2012 TRAUMATIC BRAIN INJURY DEVICE

TBI /2012 TRAUMATIC BRAIN INJURY DEVICE USER MANUAL TBI 0310 6/2012 TRAUMATIC BRAIN INJURY DEVICE Page 1 of 26 Setting up the TBI 0310 Head Impactor The TBI 0310 Head Impactor when fully assembled has the following components: 1. Control box

More information

300 Volt Electrophoresis Power Supply

300 Volt Electrophoresis Power Supply 300 Volt Electrophoresis Power Supply Catalog Number E0303 E303-230V 110Volt 230Volt Table of Contents FCC Warning 3 Safety Information 3 Package Contents 4 Product Specifications 5 Overview 6 Description

More information

GX-271 ASPEC Gilson, Inc. All Rights Reserved. LT

GX-271 ASPEC Gilson, Inc. All Rights Reserved. LT GX-271 ASPEC GX-271 ASPEC World Headquarters Gilson, Inc. 3000 Parmenter Street P.O. Box 620027 Middleton, WI 53562-0027 USA Telephone: 608-836-1551 Fax: 608-831-4451 Gilson S.A.S. 19, avenue des Entrepreneurs

More information

Be Strong. Assembly Manual Commercial Series Bike U916. Model: U916. P/N: Rev D (03/28/07)

Be Strong. Assembly Manual Commercial Series Bike U916. Model: U916. P/N: Rev D (03/28/07) Be Strong. Assembly Manual Commercial Series Bike U916 Model: U916 P/N: 001-6883 Rev D (03/28/07) preface Thank you for purchasing the Nautilus Commercial Series U916 bike. For more than 30 years Nautilus,

More information

ValveMate 7000 Controller Operating Manual

ValveMate 7000 Controller Operating Manual A NORDSON COMPANY ValveMate 7000 Controller Operating Manual Steady Test Clear Time Set Fast Slow POWER RUN SETUP CYCLE A NORDSON COMPANY VALVEMATE 7000 Run Setup Purge Program Fast Slow Time Set Pressure

More information

Warner Instruments 1125 Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

Warner Instruments 1125 Dixwell Avenue, Hamden, CT (800) / (203) (203) fax VC-8 Manual, Rev. 050817.1 1 Perfusion Valve Controller Model VC-8 1125 Dixwell Avenue, Hamden, CT 06514 (800) 599-4203 / (203) 776-0664 (203) 776-1278 - fax VC-8 Manual, Rev. 050817.1 2 Table of Contents

More information

Model 1100B CHG Terminator. Installation Instructions

Model 1100B CHG Terminator. Installation Instructions Model 1100B CHG Terminator Installation Instructions 1 Contents: 1.0 Safety Information... 3 2.0 Set-up and Adjustments... 3 3.0 Ram Adjustments... 10 4.0 Wire Termination Quality... 12 5.0 General Maintenance...

More information

Low Profile 8 150watt Amplified Subwoofer Installation Instructions

Low Profile 8 150watt Amplified Subwoofer Installation Instructions Low Profile 8 150watt Amplified Subwoofer Installation Instructions Radio Engineering Industries, Inc. www.radioeng.com Rev A 5/12/14 Thank you for purchasing the 150Watt Amplified Subwoofer from REI.

More information

MODEL 422 Submersible Pump Controller

MODEL 422 Submersible Pump Controller MODEL 422 Submersible Pump Controller Monitors True Motor Power (volts x current x power factor) Detects Motor Overload or Underload Operates on 120 or 240VAC, Single-phase or 3-phase Built-in Trip and

More information

INSTALLATION and OPERATION BALL WASHER MODEL NO: BW-001N

INSTALLATION and OPERATION BALL WASHER MODEL NO: BW-001N Easy Picker Golf Products, Inc. 415 Leonard Blvd. N., Lehigh Acres, FL 33971 PH: 239-368-6600 FAX: 239-369-1579 Service: 800-982-4653 SALES: 800-641-4653 www.easypicker.com salesdept@easypicker.com INSTALLATION

More information

Total credit to the non-financial sector (core debt), % of GDP Table F1.1

Total credit to the non-financial sector (core debt), % of GDP Table F1.1 Total credit to the non-financial sector (core debt), % of GDP Table F1.1 2012 2013 2014 2015 2016 Q2 16 Q3 16 Q4 16 Q1 17 Q2 17 Argentina 62.6 66.4 64.6 75.5 75.3 81.7 80.1 75.3 71.6 72.7 Australia 208.0

More information

TurfDefender Electronic Leak Detector Kit Reelmaster 5000, 6000 and 5010 Series Traction Units

TurfDefender Electronic Leak Detector Kit Reelmaster 5000, 6000 and 5010 Series Traction Units Form No. 56 586 Rev A TurfDefender Electronic Leak Detector Kit Reelmaster 5000, 6000 and 500 Series Traction Units Model No. 05 Installation Instructions The Installation Instructions for Reelmaster 5000/6000

More information

Never charge the batteries with a charger other than that supplied by the manufacturer.

Never charge the batteries with a charger other than that supplied by the manufacturer. INTRODUCTION The Diamond SeroLogic portable pipette controller is a rechargeable electronic pipetting aid designed for use with glass or plastic serological pipettes in the 0.1 to 100mL volume range. The

More information

Preparing The Printer For Maintenance

Preparing The Printer For Maintenance Chapter 4 Preparing The Printer For Maintenance Preparing The Printer For Maintenance WARNING IMPORTANT Unplug the printer power cord from the printer or power outlet before you do any maintenance procedure.

More information

BIGLA30-T/BIELA14-T Event Codes Quick Reference EXPLANATION CORRECTIVE ACTION PARTS TO CARRY ON SERVICE CALL

BIGLA30-T/BIELA14-T Event Codes Quick Reference EXPLANATION CORRECTIVE ACTION PARTS TO CARRY ON SERVICE CALL E13 TEMPERATURE PROBE FAILURE E16 HIGH LIMIT 1 EXCEEDED A. TEMP Probe reading out of range. B. Bad Connection. C. Problem with the temperatur e measuring circuitry including the probe. High limit temperature

More information

JEEVES. JEEVES Installation Manual. Installation Manual The Easiest Do-It-Yourself Dumbwaiter on the Market

JEEVES. JEEVES Installation Manual. Installation Manual The Easiest Do-It-Yourself Dumbwaiter on the Market 1 888-323-8755 www.nwlifts.com JEEVES Installation Manual The Easiest Do-It-Yourself Dumbwaiter on the Market This manual will cover the installation procedure step-by-step. The installation of this dumbwaiter

More information

Information Technology and Economic Development: An Introduction to the Research Issues

Information Technology and Economic Development: An Introduction to the Research Issues November 1998 Working Papers No. 153 Information Technology and Economic Development: An Introduction to the Research Issues TABLE 1 WORLD-WIDE INFORMATION TECHNOLOGY MARKET BREAKDOWN, 1995

More information

User s Manual Power Supply IM E. 5th Edition

User s Manual Power Supply IM E. 5th Edition User s Manual 700938 Power Supply 5th Edition Thank you for purchasing the 700938 Power Supply. This user s manual contains useful information about the instrument s functions and operating procedures

More information

OWNER S MANUAL HC1/HC2 INSTRUCTIONS EN FRANÇAIS AU VERSO

OWNER S MANUAL HC1/HC2 INSTRUCTIONS EN FRANÇAIS AU VERSO /MD OWNER S MANUAL HC1/HC2 INSTRUCTIONS EN FRANÇAIS AU VERSO FCC COMPLIANCE This device complies with FCC rules, and is subject to the following conditions: (1) This device may not cause harmful interference.

More information

Installation. Aerada 1200 Series CS Faucet S Battery-Operated Capacitive Sensing Gooseneck Faucet IMPORTANT

Installation. Aerada 1200 Series CS Faucet S Battery-Operated Capacitive Sensing Gooseneck Faucet IMPORTANT Installation erada 1200 Series CS Faucet S53-325 Battery-Operated Capacitive Sensing Gooseneck Faucet Centershank with 4-inch Trim Plate (Optional: 8-inch or No Trim Plate) Table of Contents Supplies Required..................

More information

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com. Instruction Manual. Model 1672 Triple Output Power Supply

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com. Instruction Manual. Model 1672 Triple Output Power Supply 99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com Instruction Manual Model 1672 Triple Output Power Supply Contents Section Description Page No. CONTENTS 1 1 TEST INSTRUMENT

More information

Troubleshooting Guide: 355 Lights (24V)

Troubleshooting Guide: 355 Lights (24V) Troubleshooting Guide: 355 Lights (24V) Contents Description Refer To: Troubleshooting - Troubleshooting Chart Adjustments / Repair Procedures Bulb Replacing the Bulb Fuse(s) Replacing the Fuse (Ceiling)

More information

Operation Care Service

Operation Care Service HAWK ENTERPRISES OF ELKHART, INC Standard Floor Machine Owner s Manual Operation Care Service 13 175 RPM 1.5 HP 15 175 RPM 1.5 HP 17 175 RPM 1.5 HP 20 175 RPM 1.5 HP 13 160 RPM 1.5 HP XHD 15 160 RPM 1.5

More information

Troubleshooting Guide: 355 Lights (12V)

Troubleshooting Guide: 355 Lights (12V) Troubleshooting Guide: 355 Lights (12V) Contents Description Refer To: Troubleshooting - Troubleshooting Chart Adjustments / Repair Procedures Bulb Replacing the Bulb Fuse(s) Replacing the Fuse (Ceiling)

More information

ALUMAREEL. Electrical Cord Reel. Model EC6 100 Model EC Operational, Maintenance and Installation Manual

ALUMAREEL. Electrical Cord Reel. Model EC6 100 Model EC Operational, Maintenance and Installation Manual ALUMAREEL TM Electrical Cord Reel Model EC6 100 Model EC3 100 Operational, Maintenance and Installation Manual READ AND UNDERSTAND THIS MANUAL BEFORE USING THIS PRODUCT. RETAIN FOR FUTURE REFERENCE 1 P

More information

Installing the Switch

Installing the Switch CHAPTER 3 Revised: January 4, 2012 This chapter describes how to install the Catalyst 4900M switch. For first-time installations, perform the procedures in the following sections in the order listed. Checking

More information

TYPE 5867 PORTABLE Q METER

TYPE 5867 PORTABLE Q METER IMPORTANT: Please read these instructions carefully before operating instrument SUBJECT TO CHANGE WITHOUT NOTICE This manual superseded all previous versions please keep for future reference TINSLEY, A

More information

AL625 & AL625HD INSTALLATION & OWNER S MANUAL

AL625 & AL625HD INSTALLATION & OWNER S MANUAL AL625 & AL625HD INSTALLATION & OWNER S MANUAL These instructions are provided to assist you in the installation of the AL625. If you require further assistance, our trained staff is ready to provide you

More information

UV Transilluminator Dual Wavelength (Large)

UV Transilluminator Dual Wavelength (Large) BT505 A Geno Technology, Inc. (USA) brand name UV Transilluminator Dual Wavelength (Large) Cat. No. BT505 1-800-628-7730 1-314-991-6034 info@btlabsystems.com INTRODUCTION BT Lab Systems UV Transilluminators

More information

home ELITE CURVE CUSTOM CURVED RAIL STAIRLIFT Operator s Manual Keep this manual in a safe and readily accessible place for future reference.

home ELITE CURVE CUSTOM CURVED RAIL STAIRLIFT Operator s Manual Keep this manual in a safe and readily accessible place for future reference. home ELITE CURVE CUSTOM CURVED RAIL STAIRLIFT Operator s Manual Keep this manual in a safe and readily accessible place for future reference. REV. 23 January 2015 PRODUCT REGISTRATION FORM home Product

More information

SIGNETMARINE. Livewell Bait Pump Alarm. No Dead Bait.

SIGNETMARINE. Livewell Bait Pump Alarm. No Dead Bait. B A I T WAT C H SIGNETMARINE Livewell Bait Pump Alarm No Dead Bait. The SIGNETMARINE BaitWatch System alerts you to a livewell pump problem before your fishing trip turns into one. The loud alarm and dual-color

More information

Uplift Premium Power Lifting Seat

Uplift Premium Power Lifting Seat Uplift Premium Power Lifting Seat User Guide Read all instructions before using the Uplift Power Seat DANGER - To reduce the risk of electric shock: IMPORTANT SAFETY INSTRUCTIONS Always unplug the Uplift

More information

Amplification: T100 Thermal Cycler. T100 Thermal Cycler

Amplification: T100 Thermal Cycler. T100 Thermal Cycler Amplification: T100 Thermal Cycler T100 Thermal Cycler The Smart PCR Choice The T100 thermal cycler s intuitive touch screen makes running PCR easier than ever before. The T100 thermal cycler s performance,

More information

IMPORTANT! READ BEFORE INSTALLING!

IMPORTANT! READ BEFORE INSTALLING! 1360 Redwood Way, Suite A Petaluma, CA. 94954 TEL: (707) 763-7799 Toll free in the USA (800) 767-9543 FAX (707) 763-2631 www.xandexsemi.com EMAIL: info@xandex.com Instructions for Xandex Pneumatic Controller

More information

Z-Truck (Vertical Moving) Z-truck Flag. Y-Truck (Horizontal Moving) FIGURE 1: VIEW OF THE Z-TRUCK. Flexshaft Assembly

Z-Truck (Vertical Moving) Z-truck Flag. Y-Truck (Horizontal Moving) FIGURE 1: VIEW OF THE Z-TRUCK. Flexshaft Assembly Checking and Replacing the AC Motor To remove and replace the AC Motor you will need the following tools: #2 Phillips screwdriver (magnetic tip preferred) Removing the AC Motor 1. Ready the machine by

More information

IMPORTANT! DO NOT THROW AWAY THE SHIPPING CARTON AND PACKING MATERIAL

IMPORTANT! DO NOT THROW AWAY THE SHIPPING CARTON AND PACKING MATERIAL Operator s Manual IMPORTANT! DO NOT THROW AWAY THE SHIPPING CARTON AND PACKING MATERIAL ii Table of Contents Operator Safety... 1 Introduction... 2 Unpacking and Setup... 3 Unpacking... 3 Setup... 4 ROCKET

More information

Product Name. Paraffin Dispenser. Operator s Manual. Histology Innovation for a NEW Generation PD-120; PD-220. A Division of General Data Healthcare

Product Name. Paraffin Dispenser. Operator s Manual. Histology Innovation for a NEW Generation PD-120; PD-220. A Division of General Data Healthcare A Division of General Data Healthcare Histology Innovation for a NEW Generation Pub No.: OM PD-1 June 2013, Version 1.8 PD-120; PD-220 Operator s Manual Product Name Here Catalog #s Paraffin Dispenser

More information

Operation Care Service

Operation Care Service HAWK ENTERPRISES OF ELKHART, INC Glide Floor Machine Owner s Manual Operation Care Service Hawk Enterprises 52744 Park Six Court Elkhart, IN 46514 Phone: 574-294-1910 Toll Free: 888-289-4295 Fax: 574-970-0289

More information

Tooling Assistance Center

Tooling Assistance Center Safeguards are designed into this application equipment to protect operators and maintenance personnel from most hazards during equipment operation. However, certain safety precautions must be taken by

More information

Damper Actuator Installation

Damper Actuator Installation FANs 268.1, 977, 1628.3 Installation Bulletin EPP-1000 Issue Date 0800 EPP-1000 Series Electro-Pneumatic Positioners A pplication The EPP-1000 is designed for applications where electronic control of pneumatic

More information

FD Conveyor for FD 260 Tabber

FD Conveyor for FD 260 Tabber FD 260-20 Conveyor for FD 260 Tabber Operator Manual 1/2011 First Edition TABLE OF CONTENTS 1. INTRODUCTION... 1 1.1 FD 260-20 DESCRIPTION... 1 1.2 ITEMS INCLUDED... 1 1.3 OPERATING MANUAL SAFETY TERMS...

More information

D24-53 Series Electric Non-Spring Return Actuators 53 lb in Models (6 N m)

D24-53 Series Electric Non-Spring Return Actuators 53 lb in Models (6 N m) Installation Instructions D24-53 Issue Date November 4, 23 D24-53 Series Electric Non-Spring Return Actuators 53 lb in Models (6 N m) Installation IMPORTANT: The D24-53 Series actuator is intended to control

More information

Hawk Enterprises of Elkhart THE BRUTE Owner s Manual

Hawk Enterprises of Elkhart THE BRUTE Owner s Manual Hawk Enterprises of Elkhart THE BRUTE Owner s Manual Operation Care Service COMMERICAL/INDUSTRIAL DUTY STONE RESTORATION MACHINE 13, 15, 17, 20 165 RPM 1.5 HP 180 FRAME TOTALLY ENCLOSED FAN COVERED MOTOR

More information

Pressure Transducer Field Installation Instructions for the LI-6252 CO 2 and LI-6262 CO 2 /H 2 O Analyzers

Pressure Transducer Field Installation Instructions for the LI-6252 CO 2 and LI-6262 CO 2 /H 2 O Analyzers 6262-05 6262-03 Pressure Transducer Field Installation Instructions for the LI-6252 CO 2 and LI-6262 CO 2 /H 2 O Analyzers Please read these instructions before beginning installation. Please note that

More information

HD 7700 Setup & Operator Manual

HD 7700 Setup & Operator Manual HD 7700 Setup & Operator Manual Issue 1 December, 01 Performance Design Inc. The Heavy Duty Ultima (HD 7700) electric punch has been designed to punch most any job that may pass through your bindery or

More information

WARRANTY...1 NOTICE INTRODUCTION USE...8

WARRANTY...1 NOTICE INTRODUCTION USE...8 Contents WARRANTY...1 NOTICE...2 1 INTRODUCTION...3 1.1 ABOUT...3 1.2 UNPACKING...3 1.3 INCLUDED ITEMS...4 1.4 COMPONENTS...4 2 USE...8 2.1 POWER ON...8 2.2 SETTINGS...9 2.2.1 Setup Wizard...9 2.2.2 Set

More information

L-K INDUSTRIES BENCHMARK S MANUAL

L-K INDUSTRIES BENCHMARK S MANUAL L-K INDUSTRIES Demand a higher standard. BENCHMARK S MANUAL Heated Laboratory Centrifuge To set the standard for MANUFACTURING QUALITY and customer service in the PETROLEUM TESTING industry. Petroleum

More information

A U T O M A T I C T R A N S M I S S I O N M U L T I - C H A N N E L T W O - W A Y L C D R E M O T E S T A R T E R AS-2510 TW.

A U T O M A T I C T R A N S M I S S I O N M U L T I - C H A N N E L T W O - W A Y L C D R E M O T E S T A R T E R AS-2510 TW. A U T O M A T I C T R A N S M I S S I O N M U L T I - C H A N N E L T W O - W A Y L C D R E M O T E S T A R T E R S Y S T E M AS-2510 TW User Guide Transmitter Part Number and Module Serial Number...2

More information

MM184 Rev. 12 (9 94)

MM184 Rev. 12 (9 94) 465 480 490 MM184 Rev. 12 (9 94) This manual is furnished with each new TENNANT Model 465, 480, and 490. It provides necessary operating and preventive maintenance instructions. Read this manual completely

More information

SMART TICKET INTEGRATION GUIDE

SMART TICKET INTEGRATION GUIDE SMART TICKET INTEGRATION GUIDE SMART Ticket Integration Guide MANUAL AMENDMENTS Rev. Date Amendment Details Issued by 18-01-13 First Issue AB SMART Ticket Integration Guide SMART TICKET MANUAL - INTRODUCTION

More information

Series 957, 957N, 957Z, 957RPDA, 957NRPDA, 957ZRPDA

Series 957, 957N, 957Z, 957RPDA, 957NRPDA, 957ZRPDA Series 957, 957N, 957Z, 957RPDA, 957NRPDA, 957ZRPDA Reduced Pressure Zone Assemblies Reduced Pressure Detector Assemblies Sizes: 2 1 2" 10" (65 250mm) Installation Service Repair Kits Maintenance RP/IS-957/957RPDA

More information

OPERATORS MANUAL FOR USE WITH MAGNETIC/HEATED STIRRER MG200

OPERATORS MANUAL FOR USE WITH MAGNETIC/HEATED STIRRER MG200 OPERATORS MANUAL FOR USE WITH MAGNETIC/HEATED STIRRER MG200 ISSUE 4 April 2007 TABLE OF CONTENTS Page 1.0. GENERAL INFORMATION 2 1.1. DEFINITION 2 1.2. PRINCIPE OF OPERATION 2 2.0. TECHNICAL DESCRIPTION

More information

SBS-Control. Table of contents. Operating Instructions BKS24-9A. Operating Instructions

SBS-Control. Table of contents. Operating Instructions BKS24-9A. Operating Instructions Operating Instructions BKS24-9A Operating Instructions BKS24-9A Table of contents Operating Instructions 1 Operating controls and indicators 2 2 Preparations for faultless commissioning and operating 3

More information

Warner Instruments 1125 Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

Warner Instruments 1125 Dixwell Avenue, Hamden, CT (800) / (203) (203) fax Perfusion Valve Controller Model VC-8M 1125 Dixwell Avenue, Hamden, CT 06514 (800) 599-4203 / (203) 776-0664 (203) 776-1278 - fax Table of Contents OPERATING DIRECTIONS...4 Setup...4 Overflow sensor...6

More information

User Manual. T6 Tachometer. Online: Telephone: P.O. Box St. Petersburg, Florida 33736

User Manual. T6 Tachometer. Online:   Telephone: P.O. Box St. Petersburg, Florida 33736 User Manual T6 Tachometer Online: www.phareselectronics.com Telephone: 727-623-0894 P.O. Box 67251 St. Petersburg, Florida 33736 Table of Contents Overview... 1 Description... 1 Wiring... 1 T6 Tachometer

More information

One- Touch Installation Instructions

One- Touch Installation Instructions One- Touch Installation Instructions 1 1 Height Adjustable Pivot w/ screws 9 Upper Work Surface 2 Rail Mount Knobs 10 Back Cover 3 Transformer 11 Center Pivot w/ screws 4 Support Legs 12 Left Monitor Arm

More information

Parts and Diagrams7. Printer. Stacker. HP DesignJet 4500_4500ps_4500mfp

Parts and Diagrams7. Printer. Stacker. HP DesignJet 4500_4500ps_4500mfp 7 7 Printer Printer Support 7- Center Covers 7-4 Electronics Module 7-6 Right Hand Cover 7-8 Left Hand Cover 7-0 Right Hand Assemblies 7- Left Hand Assemblies 7-4 Carriage Assembly 7-6 Scan-Axis Assemblies

More information

Colt Series C400, C500

Colt Series C400, C500 Colt Series C400, C500 RP/IS-A-C400/C500 C400 OSY Reduced Pressure Zone Assemblies Reduced Pressure Detector Assemblies Sizes: 2 1 2" 10" (65 250mm) Installation Service Repair Kits Maintenance For other

More information

Remove 4 circled pins. Route wiring along dashed line. Remove the 2 9mm nuts and black retaining plate that secure extractor.

Remove 4 circled pins. Route wiring along dashed line. Remove the 2 9mm nuts and black retaining plate that secure extractor. 2015 Ford Mustang Turn Signal Hood Kit Parts List: Quantity: Tool List: Bracket & pre-installed lamp 2 Flat head screwdriver Wiring harness 1 Phillips screwdriver PB-3660 Parts Bag 1 Ratchet & Socket set

More information

ML7421A,B Electric Linear Valve Actuator

ML7421A,B Electric Linear Valve Actuator ML7421A,B Electric Linear Valve Actuator FEATURES PRODUCT DATA Quick and easy installation. No separate linkage required. 3/4 in. (20 mm) stroke on ML7421A; 1-1/2 in. (38 mm) stroke on ML7421B. No adjustments

More information

DIN-A-MITE C. SCR Power Controller Delivers Up To 80 Amperes in a Compact Package. TOTAL CUSTOMER SATISFACTION 3 Year Warranty. Features and Benefits

DIN-A-MITE C. SCR Power Controller Delivers Up To 80 Amperes in a Compact Package. TOTAL CUSTOMER SATISFACTION 3 Year Warranty. Features and Benefits DIN-A-MITE C TOTAL CUSTOMER SATISFACTION 3 Year Warranty SCR Power Controller Delivers Up To 80 Amperes in a Compact Package The DIN-A-MITE C silicon controlled rectifier (SCR) power controller provides

More information

Encore Manual Powder Spray Gun Trigger Switch Replacement

Encore Manual Powder Spray Gun Trigger Switch Replacement Instruction Sheet P/N 1096694A Encore Manual Powder Spray Gun Trigger Switch Replacement Preparation 1. Remove the nozzle nut, nozzle, and electrode assembly. 2. Disconnect the powder feed hose adapter,

More information

8. Do not install this product on the exterior of buildings. 9. Do not use as a doorstop. This will void warranty. Installation Instructions

8. Do not install this product on the exterior of buildings. 9. Do not use as a doorstop. This will void warranty. Installation Instructions INSTALLATION 8310-IQ/8320-IQ IntelliMag Electromagnetic Locks Pre-Installation Instructions 1. This product must be installed according to all applicable building and life safety codes. 2. Due to the variety

More information

12V PROGRAMMABLE POWER OUT

12V PROGRAMMABLE POWER OUT Page 1 ACCESSORIES STARTER IGNITION BATTERY WIRES SIDE VIEW BLUE RED YELLOW 30 A 10 A BLUE / WHITE YELLOW WHITE / BLUE WHITE / DOOR TRIGGER See opt. 16 DOOR TRIGGER (input positive) See opt. 16 PARKING

More information