1 World Leaders in Diesel Fuel Injection Test Equipment. HM1000 Microdiesel Clean Cabinet Installation and Operation Manual Ref. No. HL022 (EN), Issue 1
2
3 HARTRIDGE LIMITED HM1000 MICRODIESEL CLEAN CABINET CONTENTS SAFETY INFORMATION INTRODUCTION AND SPECIFICATION OVERVIEW SPECIFICATION INSTALLATION AND OPERATION MAINTENANCE AIR SPEED CHECKING AND ADJUSTMENT PRE FILTER REPLACEMENT HEPA FILTER REPLACEMENT FLUORESCENT TUBE REPLACEMENT SPARES CONTACTING HARTRIDGE APPENDIX A - ELECTRICAL SCHEMATIC...11 HL022 01/03 1
4 HM1000 MICRODIESEL CLEAN CABINET HARTRIDGE LIMITED This page is intentionally blank 2 HL022 01/03
5 HARTRIDGE LIMITED HM1000 MICRODIESEL CLEAN CABINET SAFETY INFORMATION Most accidents involving equipment operation and maintenance are caused by failure to observe basic safety rules or precautions. An accident can often be avoided by recognising potentially hazardous situations beforehand. The equipment is designed to be intrinsically safe whenever possible. Guards are fitted to afford optimum operator protection. Read and understand all warning, caution, prohibition and mandatory notices and labels on the equipment before operation or carrying out any maintenance or servicing. If in doubt - ASK! Warnings, Cautions and Notes The precautionary notes in this publication, indicated by the words WARNING, CAUTION, or NOTE provide information about potential hazards to personnel or equipment. Ignoring these notes may lead to serious injury to personnel and/or damage to equipment. These notes appear as follows: WARNING! INDICATES THAT A SITUATION MAY BE HAZARDOUS TO PERSONNEL. INSTRUCTIONS ARE PROVIDED FOR AVOIDING PERSONAL INJURY. CAUTION! Indicates that conditions exist that could result in damage to equipment. Instructions are provided to prevent equipment damage. NOTE: Indicates additional information for clarification where there may be confusion. Warning Symbols Isolate the electrical supply before performing any maintenance operations. Do not work on electrical equipment while voltage is supplied. If such cannot be avoided, e.g. for measurements, tests or adjustments, have the action carried out by qualified personnel only. Impervious gloves and overalls should be worn if regular contact with ISO4113 test fluid is likely. Refer to the Health & Safety Data Sheets. Ensure that the servicing requirements and intervals as set out in the Maintenance section are adhered to. Operate and service this equipment only if competent to do so. Carry out regular inspections to make sure all high pressure connections are tight and safe. Ensure good levels of lighting for safe, efficient equipment operation. HL022 01/03 3
6 HM1000 MICRODIESEL CLEAN CABINET HARTRIDGE LIMITED This page is intentionally blank 4 HL022 01/03
7 HARTRIDGE LIMITED HM1000 MICRODIESEL CLEAN CABINET 1. INTRODUCTION AND SPECIFICATION This document describes the installation of equipment, both electrical and mechanical 1.1. Overview The Microdiesel Cabinet (HM1000) is an industrialised clean air booth designed specifically for the assembly of diesel fuel injection equipment. It enables diesel workshops to set up and maintain a clean assembly area to accepted clean room standards in order to meet the stringent demands of FIE manufacturers. Figure Specification Size 1273 mm (w) x 784 mm (d) x 1371 (h) (Refer to Fig 1). Weights (approx) Complete Unit 155 kg Filter 22.5 kg Power requirements 220VAC, 50 Hz, 10A. Filtration A pair of prefilter mats and a HEPA filter cartridge capable of particle filtration down to 0.5 um. Equipment 4 off 55W Fluorescent lights (Standard) Hydraclamp, HM910 (Optional) Door (Optional) Tool Rack (Optional) HL022 01/03 5
8 HM1000 MICRODIESEL CLEAN CABINET HARTRIDGE LIMITED 2. INSTALLATION AND OPERATION Isolate the electrical supply before performing any maintenance operations. Do not work on electrical equipment while voltage is supplied. If such cannot be avoided, e.g. for measurements, tests or adjustments, have the action carried out by qualified personnel only. To prolong the filter life, it is recommended that the Microdiesel Cabinet is installed in an enclosed assembly area. The cabinet is designed for bench top use, the recommended dimensions of which are shown in figure 2. Figure 2. The unit should be positioned on a suitable work surface/bench, located to a convenient single phase supply 50Hz). When supplied, the air flow is set to a nominal 0.7m/s at the factory and as the filters gradually block the air speed will decrease. For optimum performance it is necessary to maintain the air flow above 0.4m/s, details of which are described in Section 3 - Maintenance. 6 HL022 01/03
9 HARTRIDGE LIMITED HM1000 MICRODIESEL CLEAN CABINET 3. MAINTENANCE Isolate the electrical supply before performing any maintenance operations. Do not work on electrical equipment while voltage is supplied. If such cannot be avoided, e.g. for measurements, tests or adjustments, have the action carried out by qualified personnel only. The unit is nominally set to 0.7m/s at the factory and as the filters become gradually blocked the air speed will decrease. To maintain optimum performance the air flow must be maintained above a minimum of 0.4m/s. This can be achieved by adjusting the fan controller to increase the fan speed, but once the maximum speed setting is achieved it indicates that the main HEPA filter has come to the end of its life Air Speed Checking and Adjustment This maintenance procedure requires the front cover to be removed and the motor to be running. Some rotating parts of the fan motor are exposed. Take extra care when carrying out this procedure to minimise risk of entanglement. To monitor and adjust the air speed it is recommended that a suitable Rotating Vane Anemometer is used, capable of reading between 1.0 and 0.3m/s with an accuracy of +/-5% of reading and a resolution of 0.05 m/s. 1. To monitor the air flow, power on the unit and monitor the flow at various positions across the filter face with the anemometer held approximately 100 mm from the filter face. 2. Ensure that the minimum flow recorded is not below 0.4 m/s (+/- 20%). 3. If the minimum flow is too low it will be necessary to remove the front panel (secured by six bolts) to gain access to the speed control (Fig 3). 4. Measure the flow where it was at its lowest whilst adjusting the speed controller to obtain a minimum air velocity of 0.4 m/s Note: The minimum performance acceptable is 0.4 m/s. The preferred speed is between 0.6 and 0.7 m/s. If the flow cannot be achieved it will be necessary to replace the HEPA filter (refer to section 3.3) 5. Recheck the flow at various points across the filter face and ensure the typical velocity is above 0.4 m/s. 6. Once set refit the front panel Pre Filter Replacement To extend the life of the main HEPA filter, it is recommended that the pre-filters are changed at regular 8 week intervals. These simply lie in trays accessed from the top of the unit (refer to figure 3). HL022 01/03 7
10 HM1000 MICRODIESEL CLEAN CABINET HARTRIDGE LIMITED 3.3. HEPA Filter Replacement The HEPA assembly weighs 22.5 kg. A minimum of 2 people are required to remove the unit. It is necessary to change the HEPA filter once the air speed cannot be maintained above 0.4 m/s. 1. Switch off the unit and isolate from the mains electricity supply. 2. Remove the front panel to gain access to the fan assembly jacking points (see figure 3). 3. Using a 13 mm socket, jack up the fan assembly until all pressure is removed from the HEPA filter assembly. Take care to adjust the screws alternately to keep the assembly level. 4. Slide out the HEPA filter assembly. 5. Replace with the new unit making sure that the unit is the correct way up (refer to figure 3). Note: The HEPA filter assembly has foam sealing strips around the edges. Take care when removing and refitting the assembly not to damage these. 6. Undo the jacking screws alternately until the full weight of the fan assembly is on the HEPA filter. 7. Replace the front cover. 8. Check and set the airflow as described in section Air Flow Direction 2 Figure 3 Fan/Plenum Assembly 1 - Pre-Filter Mat Positions 2 - Fan Speed Controller 3 - Fan Assembly Jacking Points 8 HL022 01/03
11 HARTRIDGE LIMITED HM1000 MICRODIESEL CLEAN CABINET 3.4. Fluorescent Tube Replacement Isolate the electrical supply before performing any maintenance operations. Do not work on electrical equipment while voltage is supplied. If such cannot be avoided, e.g. for measurements, tests or adjustments, have the action carried out by qualified personnel only. 1. There are four fluorescent tubes fitted to the unit, two mounted on either side panel (one upper and one forward). 2. The tubes are a push fit into the holder with an additional clip to hold the fluorescent tube. To remove a fluorescent tube unclip the body of the tube first then withdraw it from the holder. 3. To replace an upper tube it will be necessary to remove the diffuser which is secured by two bolts. 4. Fit the new tube in the reverse order. HL022 01/03 9
12 HM1000 MICRODIESEL CLEAN CABINET HARTRIDGE LIMITED 4. Spares Prefilter Kit (4 pairs) HEPA Filter Kit HS214 HS215 Tubes Fan Motor Fan Motor Controller Contacting Hartridge To contact Hartridge for technical or spares information: Customer Support Hartridge Limited The Hartridge Building Network 421 Radclive Road Buckingham MK18 4FD United Kingdom Tel: +44 (0) FAX: +44 (0) or via: In correspondence with Hartridge, please supply the details on the identification plate, fitted at the rear of the unit, quoting the Part Number and Serial Number 10 HL022 01/03
13 HARTRIDGE LIMITED HM1000 MICRODIESEL CLEAN CABINET 6. APPENDIX A - Electrical Schematic HL022 01/03 11
14 HM1000 MICRODIESEL CLEAN CABINET HARTRIDGE LIMITED This page is intentionally blank 12 HL022 01/03
15
16 World Leaders in Diesel Fuel Injection Test Equipment The Hartridge Building, Network 421, Radclive Road Buckingham, MK18 4FD United Kingdom Hartridge Limited, 2005 Tel: +44 (0) Fax +44 (0)
World Leaders in Diesel Fuel Injection Test Equipment. IFT-c. Injector Function Tester Controller. Operating and Servicing Manual
World Leaders in Diesel Fuel Injection Test Equipment. IFT-c Injector Function Tester Controller Operating and Servicing Manual HL048 (EN) Issue 4, H1852, August 2011 HARTRIDGE LIMITED IFT-c Operating
More informationSPECIAL REVERSE HORIZONTAL FLOW ENCLOSURE OPERATING AND MAINTENANCE MANUAL. Issue No: 1
SPECIAL REVERSE HORIZONTAL FLOW ENCLOSURE OPERATING AND MAINTENANCE MANUAL Issue No: 1 Certificate No: 0954499 CE Marking certifies that this equipment conforms to the following EEC directives: - Low Voltage
More informationFILTRATION FUME CABINET MODELS T800 T1800 OPERATING AND MAINTENANCE MANUAL
FILTRATION FUME CABINET MODELS T800 T1800 OPERATING AND MAINTENANCE MANUAL ISSUE A 6 TH December 2013 2 CONTENTS SECTION 1... 5 DESCRIPTION OF THE CABINET... 5 SECTION 2... 6 INSTALLATION... 6 TESTING
More informationMOBILE FILTRATION FUME CABINET
MOBILE FILTRATION FUME CABINET MODELS INCLUDE: MFC1000 MFC1000E MFC1000ES WITH MONITORING SYSTEM Thermoline Scientific Equipment Pty. Ltd. T/A Thermoline Scientific ABN 80 000 859 129 10-12 Ross Place.
More informationWarning. Document ref: A October 2015
Sentry Door Curtain Installation, Operation & Maintainance Instructions Dunham-Bush Ltd, Downley Road, Havant, Hampshire, P09 2JD Tel. 023 9247 7700 Email. info@dunham-bush.co.uk Document ref: 128-000-001-A
More informationKatharometer Power Supply Units
Instruction Manual IM/4234600 Rev. D Katharometer Power Supply Units 4234 600 & 4234 601 ABB The Company EN ISO 9001:1994 We are an established world force in the design and manufacture of instrumentation
More informationWorld Leaders in Diesel Fuel Injection Test Equipment HK900. All Makes Common Rail Injector Test Kit (AVM2) Operating Manual
World Leaders in Diesel Fuel Injection Test Equipment HK900 All Makes Common Rail Injector Test Kit (AVM2) HL030(EN), Issue 1, June 2005 CONTENTS FOREWORD...3 1. INTRODUCTION...7 1.1 SPECIFICATION...7
More informationWorld Leaders in Diesel Fuel Injection Test Equipment. IFR-50. All Makes CR Injector Flushing Rig. Operating and Servicing Manual
World Leaders in Diesel Fuel Injection Test Equipment. IFR-50 All Makes CR Injector Flushing Rig Operating and Servicing Manual IFR-50 CR Injector Flushing Rig This page intentionally left blank CONTENTS
More informationContents. Overview. Safety Instructions. Before Installation. Installation. Operation. Maintenance. Replacement Parts. System Specifications 8 01
PrintPRO Universal Contents Overview 1 01 02 03 Overview of your extraction system (front) Overview of your extraction system (back) Overview of the Control Panel Safety Instructions 2 01 Important safety
More informationINSTRUCTION MANUAL. AMS 4000 One Piece. Negative Pressure Unit (AMS-4000 NPU 230V)
INSTRUCTION MANUAL AMS 4000 One Piece Negative Pressure Unit (AMS-4000 NPU 230V) Page 1 of 19 WARNING Asbestos exposure can cause severe and fatal diseases. Approved Code of Practices should be applied
More informationINSTRUCTION MANUAL. AMS 1500 One Piece. Negative Pressure Unit (AMS-1500 NPU 230V)
INSTRUCTION MANUAL AMS 1500 One Piece Negative Pressure Unit (AMS-1500 NPU 230V) Page 1 of 19 WARNING Asbestos exposure can cause severe and fatal diseases. Approved Code of Practices should be applied
More informationHS LINEAR ACTUATOR
HS667-24 LINEAR ACTUATOR General Installation Notes: Make sure that power is disconnected from system prior to servicing. Installation of this equipment and related OEM equipment should be in accordance
More informationOPERATING MANUAL C-1350 CABLE PUSHER. Copyright 2003 by CBS Products (KT), Ltd
CBS Products (KT), Ltd, Pillings Road, Oakham, Rutland, LE15 6QF. UK Telephone: +44(0)1572723665 Fax: +44(0)1572 756006 E-Mail: sales@cbsproducts.com Website:www.cbsproducts.com OPERATING MANUAL C-1350
More informationOPERATIONAL INSTRUCTIONS
OPERATIONAL INSTRUCTIONS PRINTPro Universal FUME EXTRACTION UNITS July 2009 Page 1 TABLE OF CONTENTS TABLE OF CONTENTS... 2 SAFETY INSTRUCTIONS... 3 Symbols used... 3 Electrical safety... 3 Dangers to
More informationInstruction Manual FOLDABLE CRANE. NOTE: Owner / Operator must read and understand this. instruction manual before using the foldable crane.
Instruction Manual FOLDABLE CRANE NOTE: Owner / Operator must read and understand this instruction manual before using the foldable crane. - 1 - Specifications Model SC500 SC1000 SC2000 Capacity (kg) 500
More informationgeneral operation and maintenance instructions
general Installation and maintenance work is to be carried out by qualified personnel only. WÖHWA cannot be held responsible for any damages caused by improper installation or maintenance work. Before
More informationIMPORTANT INSTRUCTIONS FOR OPERATION & MAINTENANCE OF
IMPORTANT INSTRUCTIONS FOR OPERATION & MAINTENANCE OF CONVEYORS EASIKIT 300 EASIKIT 450 EASIKIT 600, 900, 1200 & 1500 The manufacturer does not accept responsibility for any loss, damage to other equipment,
More informationInlet Controller TC5-ITA USER'S MANUAL. M rev. 02 K rev. 00
Inlet Controller TC5-ITA USER'S MANUAL M 890-00047 rev. 02 K 895-00458 rev. 00 TABLE OF CONTENTS PRECAUTIONS... 3 FEATURES... 4 LOCATION OF THE CONTROLS... 5 Status Leds...5 Internal Switches...6 INSTALLATION
More informationUV-cabinet DNA/RNA UVC/T-M-AR
UV-cabinet DNA/RNA For version V.4GD Contents 1 Safety...4 2 General Information...5 3 Getting Started...6 4 Operation of...7 5 Specifications...9 6 Guarantee and service...10 7 Declaration of Conformity...11
More information(UK) Date : January 2014
Sentinel Door Curtain - Model 270 Installation, Operation & Maintenance Dunham-Bush Ltd, Downley Road, Havant, Hants PO9 2JD Tel. 023 9247 7700 Fax. 023 9245 3601 (Global) www.dunham-bush.com Document
More informationGL1200T Cut & Seal. (GL1200 Seal only)
GL INSTRUMENTS Technical Description & Operating Instructions Safe Seal Cut & Seal Welder for Filter Sacks GL1200T Cut & Seal (GL1200 Seal only) For industrial use only Important Safety & Operating Instructions
More informationHLS Watt Halogen Illuminator Service Manual
HLS-150 150-Watt Halogen Illuminator Service Manual 6018 Bowdendale Avenue Jacksonville, FL 32216 Customer Service: 904 737 7611 FAX 904 733 4832 Toll Free 877 677 2832 (English) TABLE OF CONTENTS Page
More informationBM kv Digital Insulation Tester USER MANUAL
M BM5200 5 kv Digital Insulation Tester USER MANUAL CONTENTS Safety warnings...3 Symbols used on instrument...4 Cleaning...4 General description...5 Insulation resistance test modes...5 Automatic discharge...5
More informationRescue Pac. Please read and fully understand the instructions in this manual before operation. Keep this manual safe for future reference
Please dispose of Packaging for the product in a responsible manner. It is suitable for recycling. Help to protect the environment, take the packaging to the local amenity tip and place into the appropriate
More informationHEPA AIR FILTRATION SYSTEM
PRODUCT MANUAL FOR Airwash PRO HEPA AIR FILTRATI SYSTEM Table of Contents Rules for Safe Operation...1 Shipping and Packing List...1 Parts Identification/Dimensions...2 Operation...3/4 Product Application
More informationInstallation Instructions Electric Heaters 5 20 kw
Small Packaged Products 2 to 5 Tons Accessory Electric Heaters Cancels: IIK 564A-24-2 IIK 564A-24- -02 Installation Instructions Electric Heaters 5 20 kw NOTE: Read the entire instruction manual before
More informationSOLAR-150HD. 150-Watt Dual Halogen Illuminator. Service Manual Bowdendale Avenue Jacksonville, FL 32216
SOLAR-150HD 150-Watt Dual Halogen Illuminator Service Manual 6018 Bowdendale Avenue Jacksonville, FL 32216 Customer Service: 904 731-5869 FAX 904 733 0012 Toll Free 800-684-6404 TABLE OF CONTENTS Page
More informationOPERATING AND MAINTENANCE MANUAL. Primary Current Injection Test Set. 750ADM-H mk2
OPERATING AND MAINTENANCE MANUAL Product: Type: Primary Current Injection Test Set 750ADM mk2 750ADM-H mk2 DESIGNED AND MANUFACTURED BY: T & R Test Equipment Limited 15-16 Woodbridge Meadows, Guildford,
More informationOPERATIONAL INSTRUCTIONS
OPERATIONAL INSTRUCTIONS AD ORACLE FUME EXTRACTION UNIT 1 Version 1 Oct 2009 TABLE OF CONTENTS SAFETY INSTRUCTIONS... 3 Symbols used... 3 Electrical safety... 3 Dangers to eyes, breathing and skin... 3
More informationInstallation Instructions Electric Heaters 5 20 kw
Small Packaged Products to 5 Tons Accessory Electric Heaters Cancels: IIK 564A--1 IIK 564A-- 11-01 Installation Instructions Electric Heaters 5 0 kw NOTE: Read the entire instruction manual before starting
More informationSHOT BLAST CABINET MODEL No. CSB20B Part No OPERATING & MAINTENANCE INSTRUCTIONS
SHOT BLAST CABINET MODEL No. CSB20B Part No. 7640110 1001 OPERATING & MAINTENANCE INSTRUCTIONS Item Description Part No. Item Description Part No. SPARE PARTS AND SERVICING Please contact your nearest
More informationAbrasive Blast Cabinet Assembly & Operating Instructions
3951103 Abrasive Blast Cabinet Assembly & Operating Instructions READ ALL INSTRUCTIONS AND WARNINGS BEFORE USING THIS PRODUCT. SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE. This manual provides important
More informationMODEL 905V OPERATING INSTRUCTIONS
MODEL 905V OPERATING INSTRUCTIONS Quantek Instruments 183 Magill Drive Grafton, MA 01519 Tel: (508) 839-3940 Fax: (508) 819-3444 Email: sales@quantekinstruments.com GENERAL DESCRIPTION These instructions
More informationHIGH FREQUENCY AUTOMATIC BATTERY CHARGER
HIGH FREQUENCY AUTOMATIC BATTERY CHARGER MODEL NO: HFBC12 PART NO: 6267000 OPERATION & MAINTENANCE INSTRUCTIONS LS0814 INTRODUCTION Thank you for purchasing this CLARKE High Frequency Automatic Battery
More informationDCE 100 Series Dust Filter
Installation, Operation and Maintenance Manual DCE 100 Series Dust Filter Publication 2638B (GB) 0702 1419-8000A Donaldson Torit DCE reserve the right to alter design without notice. Freedom from patent
More informationModel G14SD Vapor Management System
Instruction Manual Model G14SD Vapor Management System (For Models: DSD and SSD 110/115v) Manufactured in USA for Medivators Reprocessing Systems by: PCI Medical Inc. 12 Bridge Street, PO Box 188 Deep
More informationOwner s Information Manual
50ES---A and 50VL---C Comfort 13SEERThreePhase2½---5NominalTons (Sizes 30---60) Comfort 14 SEER Single and Three Phase 2---5 Nominal Tons (Sizes 24---60) Single Packaged Air Conditioner System With Puronr
More informationMODEL MC1500 Installation and Operation Manual Important:
MODEL MC1500 Installation and Operation Manual Important: This manual contains specific cautionary statements relative to worker safety. Read this manual thoroughly and follow as directed. It is impossible
More informationSERVICE MANUAL. Koala
SERVICE MANUAL Koala US How to contact Permobil Permobil Inc. USA 6961 Eastgate Blvd. Lebanon, TN 37090 USA Phone: 800-736-0925 Fax: 800-231-3256 Email: info@permobilusa.com Head Office of the Permobil
More informationF-4600 INLINE ULTRASONIC FLOW METER Installation and Operation Guide
F-4600 INLINE ULTRASONIC FLOW METER Installation and Operation Guide 11451 Belcher Road South, Largo, FL 33773 USA Tel +1 (727) 447-6140 Fax +1 (727) 442-5699 1054-7 / 34405 www.onicon.com sales@onicon.com
More informationPCR 6. Operators Manual. Vertical Laminar Airflow Cabinet with U.V. Sterilisation. Issue 05 September 2013
Vertical Laminar Airflow Cabinet with U.V. Sterilisation Issue 05 September 2013 Puricore International Ltd 175 Kenn Road Clevedon North Somerset BS21 6LH Tel +44 1275 793000 Fax +44 1275 341313 labcaire@puricore.com
More informationOperating Instructions. International English. CCS Dust Extractor
Operating Instructions International English CCS Dust Extractor Electric Cast Saw Dust Removal System Warnings & Safety Rules WARNING!: THIS EQUIPMENT MUST BE EARTHED! Danger!: Possible explosion hazard
More informationADVANCED SAFETY ENGINEERING FOR EFFECTIVE HANDLING OF HAZARDOUS PARTICULATE DRUGS AND POWDERS IN A CLASS II, TYPE A2 BIOLOGICAL SAFETY CABINET
ADVANCED SAFETY ENGINEERING FOR EFFECTIVE HANDLING OF HAZARDOUS PARTICULATE DRUGS AND POWDERS IN A CLASS II, TYPE A2 BIOLOGICAL SAFETY CABINET NuAire, Inc. 2100 Fernbrook Lane Plymouth, MN 55447 U.S.A.
More informationTriple Output Power Supply
Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 TestEquipmentDepot.com Model 1672, 1673 Triple Output Power Supply INSTRUCTION MANUAL 1 Safety Summary The following safety precautions
More informationKV4-300 mk2. KV5-200 mk2. KV5-100T mk2. KV3-250 mk2
OPERATING AND MAINTENANCE MANUAL Product: Type: High Voltage AC Test Set KV3-250 mk2 KV4-300 mk2 KV5-100 mk2 KV5-100T mk2 KV5-200 mk2 DESIGNED AND MANUFACTURED BY: T & R Test Equipment Limited 15-16 Woodbridge
More informationQuest DRY 150. Quest PowerHEPA 500. quest. Read and Save These Instructions. Quest PowerHEPA 500:
Quest DRY 150 Quest PowerHEPA 500 Read and Save These Instructions This manual is provided to acquaint you with the air scrubber so that installation, operation and maintenance can proceed successfully.
More informationSwing Check Valve ASAHI AV VALVES. Contents. User s Manual. (1) Be sure to read the following warranty clauses of our product 1
Serial No. H-V013-E-13 Swing Check Valve User s Manual Contents (1) Be sure to read the following warranty clauses of our product 1 (2) General operating instructions 2 (3) General instructions for transportation,
More informationEndura AZ10 Oxygen probe Sensor sub-assembly tested spares kit
Information INF15/166 EN Endura AZ10 Oxygen probe Sensor sub-assembly tested spares kit Measurement made easy 1 Introduction This publication details how to replace the sensor sub-assembly on an AZ10 oxygen
More informationSupplemental Information & Instructions for Rack & Pinion Assembly with Installation Kit Rack & Pinion Assembly
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 Safety Warning Supplemental Information & Instructions for 453-618 Rack & Pinion Assembly with Installation Kit 453-619 Rack & Pinion Assembly 453-628
More informationForeword
Foreword Content 1. Summary 2. Safety instruction Safety Symbol Indication Attention Static discharge sensitive Electric shock Dangerous Indicate risk of serious injury or death or seriously damage the
More informationMPS14 Rack Mount Power Supply
MPS14 Rack Mount Power Supply USER GUIDE Publication AP3898 Limited One Year Warranty This product has been manufactured in the UK by ALLEN & HEATH and is warranted to be free from defects in materials
More informationOPERATOR S MANUAL Please read all the information in this booklet before using the unit.
TECHNE UCAL400 + Calibrator OPERATOR S MANUAL Please read all the information in this booklet before using the unit. Unit Serial number: 4/02 Issue 1 TECHNE is a registered Trademark of Techne Incorporated.
More informationInstalling the Switch Chassis
Safety, page 1 Installation Options with Racks and Cabinets, page 2 Airflow Considerations, page 2 Installation Guidelines, page 2 Unpacking and Inspecting the Switch, page 3 Installing the Switch, page
More informationBiological Safety Cabinet
Biological Safety Cabinet Biological safety cabinets are used to provide primary containment in the laboratory when the investigator is using potentially infectious materials. There are three types of
More informationTIDALFLUX 2300 F Quick Start
Quick Start Electromagnetic flow sensor for partially filled pipes The documentation is only complete when used in combination with the relevant documentation for the signal converter. KROHNE CONTENTS
More informationInstallation Instructions
NOTE: Read the entire instruction manual before starting the installation. This symbol indicates a change since the last issue. SAFETY CONSIDERATIONS Installing and servicing air conditioning equipment
More informationDC Master 24/ A
USERS MANUAL DC Master 24/12 50-60A DC-DC converter MASTERVOLT Snijdersbergweg 93, 1105 AN Amsterdam The Netherlands Tel.: +31-20-3422100 Fax.: +31-20-6971006 www.mastervolt.com ENGLISH Copyright 2015
More informationAffordable portable filtration unit
AQC source capture equipment Affordable portable filtration unit 750 to 800 CFM at the hood High efficiency cartridge Ideal for welding applications MAXAIR 6 X 10 fume arm included Filter pressure status
More informationTable of Contents. Safety Assembly Pre-operation / Starting. 7. Operation.. 8. Maintenance. 9. Storage 10
Table of Contents Safety... 3 Assembly... 6 Pre-operation / Starting. 7 Operation.. 8 Maintenance. 9 Storage 10 Parts drawings..11 Parts list by number..12 Notes.13 2 Safety Information Attention; this
More informationOPERATING AND MAINTENANCE INSTRUCTIONS HYDRAULIC ELECTRICAL PUMPS HAM (Manual control) HAE (Electrical control)
OPERATING AND MAINTENANCE INSTRUCTIONS HYDRAULIC ELECTRICAL PUMPS HAM (Manual control) HAE (Electrical control) Part Nr : HA M 4 6 2 1 B C 1. Essential safety requirements. 2. Technical Characteristics.
More informationLight Source User Guide
Light Source User Guide Compact Light Source Range Models covered by this manual: UFO 70/150 CG Glass, White Light - 240V UFO 70/150 CP Plastic, White Light - 240V UFO 70/150 CGC Glass, Colour Wheel (continuous)
More informationInstallation Instructions
Installation Instructions Accessory Ultraviolet (UV) Germicidal Lamp 1 & 2 Lamp Models 115-v & 208/230-v UV Lamp C03009 Fig. 1 Two Lamp Model UV Lamp NOTE: Read the entire instruction manual before starting
More informationModels: EP EP EP EP EP EP USER MANUAL ENGLISH
P R O F E S S I O N A L P O W E R A M P L I F I E R Models: EP - 300 EP - 500 EP - 800 EP - 1000 EP - 1300 EP - 1800 USER MANUAL ENGLISH WARNING FOR YOUR OWN SAFETY, PLEASE READ THIS USER MANUAL CAREFULLY
More informationKD1E Range Potential Indicators
KD1E Range Potential Indicators Operating Instructions 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Limited Warranty & Limitation
More informationPro 400 OPERATION MANUAL. Pro 400 Lacer Pictured.
Pro 400 OPERATION MANUAL Pro 400 Lacer Pictured. Table of Contents OPERATING INSTRUCTIONS Machine Preparation - Pro 400...1 Belt Preparation...3 Selecting and Installing the Proper Adapter/Comb...3 Selecting
More informationINSTALLATION INSTRUCTIONS
INSTALLATION INSTRUCTIONS MODELS 600HS, 1200HS, 2000HS PLUS HVAC 600HS L Housing 600HS S Housing 1200/2000HS S Housing 1200/2000HS L Housing 1 ASSEMBLY INSTRUCTIONS HVAC VERSION STEP 1. STEP 2. STEP 3.
More informationInstallation and Operating Guidelines Battery Back-up Pump System
Installation and Operating Guidelines Battery Back-up Pump System www.tritonsystems.co.uk 1 Contents 1.0 Product Summary... 3 2.0 Installation Guidelines... 3 3.0 Technical Specification... 4 4.0 Dimensions...
More informationAuthorised By: Health and Safety Committee TITLE : ELECTRICAL SAFETY GUIDELINES IN-STORE DISPLAYS
Original Issue: May 2012 Document No.: SMS.10.2 Revision No.: v4 (May) 2014 Page: 1 of 11 Revision Date: June 2015 Authorised By: Health and Safety Committee TITLE : ELECTRICAL SAFETY GUIDELINES IN-STORE
More informationMC2500. Installation and Operation Manual
MC2500 Installation and Operation Manual Important: This manual contains specific cautionary statements relative to worker safety. Read this manual thoroughly and follow as directed. It is impossible to
More informationSEAUPS-III Uninterruptible Power Conditioning System. System Manual
SEAUPS-III Uninterruptible Power Conditioning System System Manual N9246 Hwy 80 South, Suite 7, Necedah, WI 54646 Tel: 608-565-5877 Fax: 608-565-5879 Email: sales@lpc-ups.com Website: www.lpc-ups.com BILS-10012
More informationPortable filtration unit / 650 CFM (305 L/s) Portable filtration unit / 1500 CFM (710 L/s) For 1 or 2 arms configuration
-MC -MC Diversified Air Systems, Inc. pantone 2756 Portable filtration unit / 650 CFM (305 L/s) pantone 2756 Portable filtration unit / 1500 CFM (710 L/s) For 1 or 2 arms configuration 99.9% efficiency
More informationFE-WS6 Downflow Ductless Workstation. Operator s Manual
FE-WS6 Downflow Ductless Workstation Operator s Manual MISONIX, Inc. 1938 New Highway Farmingdale, NY 11735 Phone: 631-694-9555 Fax: 631-694-1320 www.misonix.com 0 FE-WS6 DOWNFLOW WORKSTATION CONSTRUCTION
More information3 PHASE TESTER & MOTOR ROTATION TESTER
3 PHASE TESTER & MOTOR ROTATION TESTER R S T TEST CAT. III 500V 3 PHASE TESTER & MOTOR ROTATION TESTER CCW CW BAT. L2 L1 L3 L1 L2 L3 MOTOR ROTATION INSTRUCTION MANUAL INDEX Page 1. INTRODUCTION... 1 2.
More informationOperating & Maintenance Manual
Operating & Maintenance Manual CONTENTS CONTENTS INTRODUCTION Introduction 2 Operating Specifications 3 Working Envelope Diagram 3 Do s and Don ts 4 Primary Components 5 Operating Procedures (incl. Emergency
More informationSELF PRIMING CHEMICAL SERVICE PUMPS
SELF PRIMING CHEMICAL SERVICE PUMPS INSTALLATION AND OPERATING INSTRUCTIONS This Manual covers: SELF PRIMING MODEL RANGE J50ECX TO J250ECX STAINLESS STEEL*, and NON METALLIC SEAL PUMP MODEL: SERIAL NO:
More informationPro Booster 802Li. Please read and fully understand the instructions in this manual before operation. Keep this manual safe for future reference.
Please dispose of packaging for the product in a responsible manner. It is suitable for recycling. Help to protect the environment, take the packaging to the local amenity tip and place into the appropriate
More informationAquaLift Bath Lift System MODEL #8100
AquaLift Bath Lift System MODEL #8100 Thank you for Choosing Columbia! Your AquaLift Bath Lift has been built to the highest standards of quality to ensure years of superior service. Please take the time
More informationVane pump, direct operated
Vane pump, direct operated RE 1/8.8 Replaces: 1. 1/18 Type PV7 A Nominal sizes 1 to Series 1X / X Maximum operating pressure 1 bar Displacement volume 1 to cm 3 H17 H9 Overview of contents Features Contents
More informationIE3. BOGE C series. Motor. Premium Efficiency. BOGE Competence
BOGE C series IE3 Motor Premium Efficiency BOGE Competence Over 100,000 compressed air users expect more when it comes to their compressed air supply. BOGE Air provides them with the air to work. Screw
More information153M -48V POWER PLANT 5000 AMP
Critical Power Product Manual 153M -48V POWER PLANT 5000 AMP Service and Assistance - +1 877 546 3243 2015 General Electric Company. All rights reserved. 642153MP Issue 68 May 2015 PE.TechSupport@ge.com
More informationMODEL Nos: CVAC20SS CVAC25SS CVAC30SSR OPERATING & MAINTENANCE INSTRUCTIONS. Distributed by CLARKE International Ltd.
MODEL Nos: CVAC20SS CVAC25SS CVAC30SSR OPERATING & MAINTENANCE INSTRUCTIONS Distributed by CLARKE International Ltd. 1006 Thank you for selecting this VAC KING Stainless Steel Vacuum Cleaner, which will
More informationSERVICE MANUAL. Chairman. Playman/Robo
SERVICE MANUAL Chairman Playman/Robo US How to contact Permobil Permobil Inc. USA 6961 Eastgate Blvd. Lebanon, TN 37090 USA Phone: 800-736-0925 Fax: 800-231-3256 Email: info@permobilusa.com Head Office
More informationMULTI-LANE HOT FOIL PRINTER FOR CFS
MULTI-LANE HOT FOIL PRINTER FOR CFS OPERATOR INSTRUCTIONS PARTS LISTING Designed and manufactured by: OPEN DATE EQUIPMENT LIMITED PUMA TRADE PARK, 145 MORDEN ROAD, MITCHAM, SURREY, CR4 4DG. UNITED KINGDOM.
More informationAssembly instructions PRORUNNER mk1 2
Assembly instructions PRORUNNER mk1 Version 0.1 / 01-JUN-2013 Copyright Qimarox B.V. All rights reserved. No part of this document may be copied, stored in a database and/or published by means of printing,
More informationTurbocharger / VTR..0, VTR..1 Original assembly instructions English
Assembly Instructions Turbocharger / VTR..0, VTR..1 Original assembly instructions English This document is valid for the VTR..0/..1 series: VTR160, VTR200, VTR250, VTR320, VTR400 VTR161, VTR201, VTR251,
More informationIB PowlVac ITE-HK Remote Racking Device
IB-51802 PowlVac ITE-HK Remote Racking Device for use with ITE-HK 5kV & 15kV Circuit Breakers and PowlVac ITE-HK 5kV & 15kV Replacement Circuit Breakers Powered by Safety PowlVac ITE-HK Remote Racking
More informationBSC-1100IIA2-X is designed to meet the operating principles:
Introduction BSC-1100IIA2-X: Function and Operating Principle The Class II Biological Safety Cabinet is designed with inward air flow at a velocity to protect personnel, HEPA-filtered vertical laminar
More informationRE-PR1-F 1-Phase Din-Rail Mount 1.5, 3 & 6kW
Page 1 of 5 RE-PR1-F 1-Phase Din-Rail Mount 1.5, 3 & Features: Benefits: 0-10Vdc or 0-5Vdc control input Over temperature protection with auto reset Din-rail mounting Efficient electronic switching No
More informationINSTALLATION GUIDE. Greentek. Residential use only MODEL HS 3.0
INSTALLATION GUIDE Residential use only MODEL HS 3.0 ITEM# 463625 OCT2018 Product of Greentek Greentek reserves the rights to modify a product, without prior notice, whether in price, design, color or
More informationMULTIVOLTAGE PORTABLE BATTERY CHARGER MVM
_ M MULTIVOLTAGE PORTABLE BATTERY CHARGER MVM User's MANUAL Code: MVM Version: 01-BF Date: OCT 2005 Page 1/10 _ 1. INTRODUCTION Before starting to use your Energic plus MVM battery charger, please take
More informationINSTRUCTION MANUAL FOR THE THERMOCOUPLE ATTACHMENT UNIT (TAU)
INSTRUCTION MANUAL FOR THE THERMOCOUPLE ATTACHMENT UNIT (TAU) Model Number: 41756 (100 125Vac) 41757 (220 240Vac) Manufactured in the United Kingdom CONTENTS Chapter Page Specification 3 What the TAU does
More informationMonicon Instruments Co., Ltd. CHR-1285/2485 CHR-1285/2485 BATTERY CHARGER
CHR-1285/2485 BATTERY CHARGER TEL:886-4-2238-0698 FAX:886-4-2238-0891 Web Site:http://www.monicon.com.tw E-mail:sales@monicon.com.tw Copyright 2007 Monicon Instruments Co., Ltd. All right reserved. Contents
More informationDisc Grinder Model G 18MR G 23MR G 23MRU
Disc Grinder Model G 18MR G 23MR G 23MRU Handling instructions G23MR NOTE: Before using this Electric Power Tool, carefully read through these HANDLING INSTRUCTIONS to ensure efficient, safe operation.
More information04979, 04980, 04981, 04982, 04983, 04984, 39054, 39055, 39056, 39057, 39225, 81459,
Hydraulic Bottle Jacks 04979, 04980, 04981, 04982, 04983, 04984, 39054, 39055, 39056, 39057, 39225, 81459, 82132. These instructions accompanying the product are the original instructions. This document
More information12v / 24v Diesel Transfer Pump Kit
Please dispose of packaging for the product in a responsible manner. It is suitable for recycling. Help to protect the environment, take the packaging to the local amenity tip and place into the appropriate
More informationThyristor power controller TE10A. Burst-firing Single-cycle Advanced Single-cycle
Thyristor power controller TE10A Burst-firing Single-cycle Advanced Single-cycle Control of short-wave infrared elements and loads of constant resistance up to 20kW User Manual Copyright Eurotherm Automation
More informationGUIDELINES ON ELECTRICAL SAFETY AT WORKPLACES
MINISTRY OF LABOUR, INDUSTRIAL RELATIONS AND EMPLOYMENT GUIDELINES ON ELECTRICAL SAFETY AT WORKPLACES Occupational Safety & Health Inspectorate These guidelines should be used with approved standards:
More informationArm Fume Hood FHA-1. Operation Manual PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION
Arm Fume Hood FHA-1 Operation Manual PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC.9.16 Preface Dear users: Thank
More informationSM-1700 Secondary Shunt Trip Coil Assembly. Powered by Safety
SM-1700 Secondary Shunt Trip Coil Assembly Powered by Safety SM-1700 Secondary Shunt Trip Coil Assembly SM-1700 Contact Information Powell Electrical Systems, Inc. www.powellind.com info@powellind.com
More informationCONTENTS DIVERSE. Preface SQUEEZE ANALYSER SQA2 RESISTANCE WELDING FORCE MONITOR. Resistance Welding. Squeeze Analyser OPERATING INSTRUCTIONS
DIVERSE SQUEEZE ANALYSER SQA2 RESISTANCE WELDING FORCE MONITOR OPERATING INSTRUCTIONS UK/US VERSION 2012 CONTENTS DIVERSE type SQA2 DIVERSE type SQA2 2 DESIGNED & MANUFACTURED BY:- DIVERSE TECHNOLOGIES
More information