Passive Check Valve Installation Instructions for H-Class Systems / Revision A Copyright Waters Corporation 2014 All rights reserved

Similar documents
ACQUITY UPLC Column Manager - Active and Column Manager - Auxiliary

ACQUITY Advanced Polymer Chromatography

StepWave ion guide: cleaning instructions addendum

Agilent 210/218 - Binary Pump

Back Pressure Regulator

Auxiliary Solvent Manager Overview and Maintenance Guide

PATROL UPLC Process Analysis System for Production

Agilent 210/218 - Isocratic Pump

Paint/Solvent/Dump Valve and Flow-Through Valve

Nickel Catalyst Tube on a 6820 GC, Accessory G4318A

Inverted High Pressure Fluid Filters

User Manual for the ProFlow 1K Release: User Manual for the ProFlow 1K_Rev_A PVA Support Hub. User Manual for the ProFlow 1K

Spray Pressure Control Manifold

XLR-8 60 Extended Rack Autosampler

Spray Pressure Control Manifold

ACQUITY UPLC I-Class System with 2D Technology

TB-690 Conveyor/Stacker

EGR heater plate removal and installation procedure has been added.

DISPLACEMENT PUMP INSTRUCTIONS-PARTS LIST Rev. K. Model , Series A Model , Series B Model , Series A

Installing the Flame Ionization Detector EPC Flow Control Manifold

BETTIS ACTUATOR & CONTROLS SERVICE INSTRUCTIONS FIELD CONVERSION OF SPRING CARTRIDGE FAIL DIRECTION CLOCKWISE TO COUNTER-CLOCKWISE OR THE INVERSE

Varian Dual Chamber Dynamic Mixer Instruction Manual

Replacing the Column Stabilizer Tube with Finger-tight Fittings

PREVIOUS GENERATION. Encore Manual Powder Spray Gun Trigger Switch Replacement. Introduction. Preparation. Gun Disassembly

Exhaust and Exhaust Gas Recirculation September Testing of the DD13. Exhaust Gas Recirculation Cooler Water Manifold Assembly - In Chassis

Installation Operation Maintenance ADDENDUM

O-Ring/Backup Ring Kit

PrimeLine TM HPLC Pumps

Operation and Maintenance Manual

EVOS M5000 Imaging System

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER CRUDE OIL COLLECTION ASSEMBLY HIGH PRESSURE

Installing the CO 2 Cryogenic Oven Cooling Kit

Installation Instructions for the Valor Lightbar Low Hook-On Mounting Kit

SUBJECT: While In Bi-level Mode, Air Flow From Panel Vents Is Less Than That Of The Floor Vents

E-200 Module Service Kit

Agilent G1554A On-Column Capillary Inlet with EPC

FRP SystemOne GELCOATER OPERATION MANUAL. MAGNUM VENUS PRODUCTS Operation Manual. Part No. M Revision

Addendum to the Waters Micromass Q-Tof Premier Mass Spectrometer Operator s Guide

General Specifications

Installing a Purged Packed (PP) Inlet

ON DEMAND INSTALLATION GUIDE

Using Plates and Vials with ACQUITY UPLC and ACQUITY UPLC H-Class Systems

SPE-DEX 3100 Solid Phase Extraction System

Universal Splice Tray Installation Instructions

Loctite Spray Valve Controller Part Number EQUIPMENT Operation Manual

General Specifications

ALITA LINEAR AIR PUMP OPERATION & MAINTENANCE MANUAL. AL- Model Number Date Code / Serial Number Date of Purchase

781Mini Series Spray Valve

Hard Seat Fluid Pressure Regulator

Encore Manual Powder Spray Gun Trigger Switch Replacement

CR4 Discrete Valve and Manifold Valve Repair or Replacement Procedure

18SP653 EPA04 and EPA98 MBE 4000 High Pressure Fuel Line and Transfer Tube Installation

Note: Please read through the entire guide before attempting any kind of installation.

High pressure fuel line and transfer tube removal and installation procedures have been updated.

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER SAMPLE / INJECTION PUMP

Agilent SD-1 Purification System ml/min

DP Series Diaphragm Valve Installation and Maintenance Instructions

Hydrocarbon Fuel Doser Injector Valve Fuel Supply Line December Publication Number / Title Platform Section Title Change

Bruker BioSpin NMR CASE. Installation Manual. Version. NMR Spectroscopy. think forward

INOCULATION TANKS 15 GALLON # GALLON #

ACQUITY QDa Detector Overview and Maintenance Guide

Connectors & Connectors with valves

ASSEMBLY and OPERATING INSTRUCTIONS Mission Oaks Blvd. / Camarillo, CA 93011

Fitting Systems. Leak-free, ee, simple-to-use 1/4-28 and M6 fitting types in

Valve Box and Valve Accessories G4315A, G4316A, G4317A

3 FT CORD Models , , , FT CORD W/GFCI Models , , ,

Battery Powered Sensor Operated Lavatory Faucets. Plug-in Transformer Powered Sensor Operated Lavatory Faucets PARTS LIST

18SP589* Series 60 Right Side EGR Bolt-Together System with Hydraulic Actuator Valve

DAKOTA INSTRUMENTS OPERATING MANUAL ACR ACRYLIC FLOW METERS

NUMBER: S.M. REF.: Listed in Table ENGINE: EPA07 Series 60 DATE: June 2012 SUBJECT: E3 FUEL INJECTOR TUBE ADDITIONS, REVISIONS, OR UPDATES

RA-20 Paint/Solvent/Dump Valve

Section 7.3 Water Manifold has been updated.

DODGE CUMMINS Air-Boss CR Intake Plenum

Agilent 1200 Series Manual Injector

Preventive Maintenance for Pail Melters

Pellicon 2 and 3 Mini Holder. User Guide

XPS-ProFeed Shuttle SERVICE MANUAL. Revised:

Hydro-Sync Slide-Out System

Study Guide MaxxForce TM 5 Engine Update TMT

INSTRUCTIONS INSTALLATION OF LOAD-SHARE TERMINALS FOR VLT DRIVES

Model MF100ACC Accumulator Module

3 FT CORD UL/NEC Models , , , FT CORD w/gfci UL Models , , ,

Waters Reagent Manager Operator s Guide

Welker Gas Sampler Model MPS-2

MDP 1000 OPERATING INSTRUCTIONS PUMP MDP 1000 OPERATING INSTRUCTIONS

MUELLER GAS. No-Blo Valve. Changers. Reliable Connections. General Information 2-3. Plugging Unit Assembly 4-5. Removing Old Valve 5-6

XPS-ProFeed Shuttle SERVICE MANUAL. Revised:

3 Scotch-Weld. User s Manual. Polyurethane Reactive (PUR) Easy Adhesive Applicator

Nylon Fluid Regulators

NUMBER: S.M. REF.: Listed in Table 1 ENGINE: DD13 DATE: July 2009

Automated Liquid Handling Station

Product Features. Easy installation. Pump adds only 3.3 to the overall assembly height. 60 GPH (227 LPH) flow rate while in priming mode.

Ferrari 355 Oil Change Procedure

Installing the Dionex ICS-1100/1600/2100 Auxiliary Valve

Instructions for Use. A-dec 4631 Duo Delivery System and A-dec 4635 Assistant s

PTEX extruder pumps Pneumatic extruder pump for drums of grease up to NLGI grade 2

MicroCoat System Operating Manual MC4000 Series MC785M, MC785M-WF Spray Valves

EPA10 DOC Inlet Pressure Sensor November Removal of the EPA10 Diesel Oxidation Catalyst Inlet Pressure Sensor. without Sensor Box

0-800 psi Carbide Seat Fluid Pressure Regulator

Operator s Safety and Service Manual

Transcription:

Passive Check Valve Installation Instructions for H-Class Systems 75004569 / Revision A Copyright Waters Corporation 04 All rights reserved

General information Copyright notice 04 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND IN IRELAND. ALL RIGHTS RESERVED. THIS DOCUMENT OR PARTS THEREOF MAY NOT BE REPRODUCED IN ANY FORM WITHOUT THE WRITTEN PERMISSION OF THE PUBLISHER. Trademarks ACQUITY UPLC is a registered trademark of Waters Corporation. PEEK TM is a trademark of Victrex plc. The Science of What's Possible is a registered trademark of Waters Corporation. TORX is a registered trademark of Camcar Division, Textron Inc. Waters is a registered trademark of Waters Corporation. Other trademarks or registered trademarks are the sole property of their respective owners. Page ii

Table of contents General information...ii Copyright notice... ii Trademarks... ii Passive Check Valve Installation Instructions...4 Removing the i Valve actuator... 4 Installing the passive check valve... 0 Page iii

Passive Check Valve Installation Instructions This document provides instructions to replace the i Valve actuator in a quaternary solvent manager included in an ACQUITY UPLC H-Class system with a passive check valve. For safety advisory information, see Appendix A in the ACQUITY UPLC H-Class System Guide. Warning: Observe Good Laboratory Practice (GLP) at all times, particularly when working with hazardous materials. Consult the Material Safety Data Sheets regarding the solvents you use. Additionally, consult the safety representative for your organization regarding its protocols for handling such materials. Notice: To avoid contaminating system components, wear clean, chemicalresistant, powder-free gloves when replacing the i Valve actuator. Required materials: Chemical-resistant, powder-free gloves /4-inch open-end wrench 5/8-inch open-end wrench T8 TORX driver Passive inlet check valve assembly i Valve jumper connector Tubing assembly, extending from the mixer manifold to passive check valve Storage bag (for the i Valve following its removal) Removing the i Valve actuator Notice: To avoid damaging the i Valve actuator, do not attempt to force or draw liquid or gas through the valve s inlet or outlet ports. Page 4

To remove the i Valve actuator: Notice: To avoid damaging electrical components and circuitry, do not disconnect an electrical assembly while electrical power is applied to an instrument or device. To completely interrupt power, set the on/off switch to the "off" position, and then disconnect the power cord from the ac source. Wait 0 seconds thereafter before disconnecting an assembly.. Flush the quaternary solvent manager using a nonhazardous solvent.. Power-off the quaternary solvent manager. Tip: The quaternary solvent manager is referred to as pump on the warning label affixed to the i Valve actuator. Notice: To avoid damaging the connector or cable, grasp the i Valve connector by its knurled diameter. 3. Grasp the i Valve connector by the knurled diameter, and pull it toward you, disconnecting it from its receptacle. 3 i Valve connector Connector receptacle 3 Knurled diameter Page 5

4. Loosen the fitting that secures the outlet tubing to the inlet manifold outlet. Outlet-tubing fitting Inlet manifold 5. Use the /4-inch open-end wrench to disconnect the fitting that secures the tubing extending to the inlet manifold from the side of the i Valve, and then remove the tubing. 3 i Valve Outlet-tubing fitting 3 Inlet manifold Page 6

6. Use the 5/6-inch open-end wrench to loosen the shell nut, and then fully unscrew it. Place 5/6-inch open-end wrench here. Notice: To avoid shock or equipment damage, never place the electrical connector in a wet drip tray. 7. Remove the i Valve actuator from the bottom of the primary pump head. Important: Ensure that when you remove the valve assembly, you also remove the PEEK washer, which is normally on the top face of the i Valve cartridge. If the washer remains in the head, the passive check valve you are installing will leak. Page 7

Bottom of primary pump head i Valve actuator 8. Locate the receptacle for the i Valve connector, from which the i Valve actuator was removed. i Valve connector receptacle Page 8

9. Align the white markings on the i Valve jumper connector, and then push the jumper connector into the i Valve electrical receptacle to secure it. i Valve connector receptacle i Valve jumper connector 0. Store the i Valve, cartridge, PEEK washer, and the tubing removed in step 5 in the bag included in the kit. Tip: Store these components for potential use in the future. Page 9

Installing the passive check valve To install the passive check valve:. Unpack the new check valve.. Ensure that the new PEEK washer is inserted into the new check valve so that its chamfered edge faces away from the check valve. 3 4 5 6 Chamfered edge facing away from check valve PEEK washer 3 Check valve 4 Check-valve housing 5 /-inch hex nut 6 5/6-inch open-end wrench flat 3. Insert the check-valve assembly into the head, and use the /-inch wrench to tighten the check-valve nut /8-turn beyond finger-tight. Page 0

4. Attach the outlet-tubing fitting of the tube assembly to the inlet manifold, and then finger-tighten it to the maximum extent. Outlet-tubing fitting Inlet manifold 5. Using the 5/6-inch open-end wrench to hold the check valve in place, attach the compression fitting to the check valve. 6. Use the /4-inch wrench to tighten the compression screw as follows: For a new stainless steel tubing assembly, 3/4-turn beyond finger-tight For an existing stainless steel tubing assembly, as much as /6-turn beyond finger-tight 7. Power-on the quaternary solvent manager. 8. Prime the quaternary solvent manager. Page