Axial Turbine Flow Sensor Series Turbotron VTH 25 / VTI 25 / VTM 25

Similar documents
FTB 372 Series Axial Turbine Flow Meter

Turbotron // VTH15, VTM15, VTP15, VTI15

Turbine Flow Meters for Liquids

SM-08. Features. Miniature Turbine Flowmeter for Fluid Media. Description: Application:

Positive displacement flowmeter for continuous flow measurement

Flow Transmitter / Switch OMNI-RR..32

Flow Transmitter / Switch OMNI-RR..32

Flow Transmitter / Switch OMNI-RR..32

INSERTION flow sensor for continuous flow measurement

Force, Displacement, Flow, Speed

Flow switches for insertion installation

FLOWX3 F3.20 High Pressure Paddlewheel Flow Sensor MAIN FEATURES APPLICATIONS OPERATING PRINCIPLE

INLINE Flow sensor for hazardous area II 1 G/D - II 2 D - II 3 GD - I M1

Flow - rotor, inline model

DMP 331. Industrial Pressure Transmitter for Low Pressure. Stainless Steel Sensor

Paddle-wheel flow controller for On/Off control

Paddle-wheel flow controller for On/Off control

SIKA Flow Switches -1-

DMP 331 DMP 331. Industrial Pressure Transmitter for Low Pressure. Stainless Steel Sensor. Pressure Transmitter. Industrial

Flow Meters & Instruments. Insertion Paddle Wheel & Magnetic Flow Meters Flow & Anaytical Instruments. Technical Catalogue

Assembly and operating Instructions Bypass Flow Meter DST

Series Turbotron. Series VTY Push-in flow sensors. Turbine flow sensors

Type 2712 (8630) Continuous TopControl System. General data Compatibility Materials of wetted parts Body Rotor Shaft Seal

Product. Information. A.u.K. Müller. Flow meter turbine, with shut-off device. Series Characteristics. Applications.

PSF50 Paddle flow switch

Type Operating Instructions. Bedienungsanleitung Manuel utilisateur

Flow - dynamic diaphragm

Insertion turbine INSTRUCTION SHEET. TECHNICAL PRODUCT

Reliable - Safe - Proven!

F111.X Flow Sensor for Hot Tap Installation INSTRUCTION MANUAL. 1. Introduction Safety Instructions Unpacking Description.

BI-DIRECTIONAL INSERTION FLOW TRANSDUCER

VHS... Screw Flow Meter

Positive displacement batch controller

FLOW SWITCHES. With pipe tees (inline) Direct installation (insertion) For OEM applications Flow monitors. Including products with:

Steering column and sensor

DMP 331P DMP 331P. Industrial Pressure Transmitter. Pressure Ports And Process Connections With Flush Welded Stainless Steel Diaphragm

INSERTION flowmeter with paddle wheel, ELEMENT design

DCT 533. Industrial Pressure Transmitter with IO-Link Interface. Stainless Steel Sensor

RUN ACCUM. TOTAL STOP BAT LOW HIGH

DMK 331P. Industrial Pressure Transmitter. Pressure Ports With Flush Welded Stainless Steel Diaphragm

Flowmeter Ultrasonic Flow Metering / Dosing Device. Operating Manual

Conductivity meter. PLC Solenoid valve. Type Type On/Off Diaphragm valve. Valve islands

Pressure transmitters for heavy duty applications MBS 3200 and 3250

1. Preface For best results, please refer to this manual before using this product. 2. Safety precautions Model Precautions DFS-1 The maximum pressure

Octave Water Meter Installation Manual

DMP 331P. Industrial Pressure Transmitter. Pressure Ports And Process Connections With Flush Welded Stainless Steel Diaphragm

Industrial OG meter series IOG Inline and flanged oval gear meter

Paddle-wheel flow controller with optical principle for On/Off control

DMP 333. Industrial Pressure Transmitter For High Pressure. Stainless Steel Sensor

Assembly and operating Instructions. Bypass Flow Meter DST

Inductive conductivity meter

Flow meter Type Recordall (RCDL)

Positive displacement flowmeter

DOR Insertion Paddle Wheel Flowmeter

Manual Turbine Flow Meters HM Series

BFS-10-O G1/4: BFS-10-O G1/2: BFS-10-O G1: Contact rating: BFS-10-O G1/4

Liquid Turbine Flow Meter OwnerÊs Manual

RPFS. ROTA PULSE FLOW SENSOR (Slip Insertion Paddlewheel) DATASHEET FEATURES

flow x3 flow monitoring

Magnetic-inductive flow meter EPS / UMF2. Technical data sheet

Type Operating Instructions. Bedienungsanleitung Manuel d utilisation

Pressure transmitter for general industrial purposes MBS 3000 and MBS 3050

DTK. OEM Pelton Wheel Flow Sensor

ProMinent. Operating Instructions Turbine Meter turbodos

ROTA PULSE FLOW SENSOR Slip Insertion Paddlewheel (with inductive coil pulse output & built in op-amp circuit)

Contactless Encoder with Stainless Steel Housing Incremental: ppr RI360P0-EQR24M0- INCRX2-H1181

flow switches J with pipe tees (inline) J Direct installation (insertion) J for hvac applications J with interchangeable paddles J For air flow

RotorFlow Sensors Provide Visual Indication, Continuous Sensing and Accurate Switching

Pressure transmitter for general industrial purposes Type MBS 3000 and MBS 3050

Type Operating Instructions. Bedienungsanleitung Manuel d utilisation

TL-W. Standard Flat Sensors in Many Different Variations. Flat Inductive Proximity Sensor. Ordering Information

Flow - magnetic inductive - probe form

Operating Instructions. Angle Seat Control Valve. Type 7020

Insertion magnetic inductive flowmeter

Plastic tube variable area flowmeter for liquids and gases

JUMO MIDAS S06. OEM-Pressure Transmitter Low Pressure. Applications. Brief description. Customer benefits. Special features

Volumeter TRZ 03-K PRODUCT INFORMATION. Reliable Measurements of Gas

Digital Pressure Regulator Sentronic PLUS Series 614

SIKA Flow Switches -1-

Diaphragm valve MV 309

Fixed Displacement Bent Axis Axial Piston Motors TMB 700

ELADOS EMP IV E00 Diaphragm Pump type

magnetic flow meters Johannesburg (Head Office) Technical Data Safmag Wafer Type

F-4600 INLINE ULTRASONIC FLOW METER Installation and Operation Guide

High temperature pressure transmitters for heavy-duty applications MBS 2200 and MBS 2250

RMG Meßtechnik GmbH. Publication No E

Temperature Sensor without thread

Reed, Electronic. MOVEMENT > Series CST-CSV-CSH proximity switches GENERAL DATA

BI-DIRECTIONAL INSERTION METER

SM-00. Features. Impeller Flowmeter for Small Volumes of Fluid. Description: Application:

Mounting and Operating Instructions EB 5868/5869 EN

Pressure transmitters for heavy duty applications Type MBS 3250

DMK 331P. Industrial Pressure Transmitter. Pressure Ports With Flush Welded Stainless Steel Diaphragm. accuracy according to IEC 60770: 0.

IP800-Series FLOW SENSOR INSTRUCTIONS IP810 IP :2008 ISO CERTIFIED COMPANY IP80-SERIES FLOW SENSOR INSTRUCTIONS

Orifice plate flowmeters Series PR By-pass flowmeter for liquids, gases and steam

RPFS. ManuFlo Page 1 MANU ELECTRONICS PTY LTD Flow Measurement Products 41 Carter Road, Brookvale NSW 2100

Data Sheet for Linear Sensors

TX 115/215 Hot-tap Insertion Turbine Instructions

TX80-Series FLOW SENSOR INSTRUCTIONS TX81 TX :2008 ISO CERTIFIED COMPANY TX80-SERIES FLOW SENSOR INSTRUCTIONS

Operating Instructions. Pneumatic Control Valve Low Temperature. Type Series GS3

Transcription:

Installation Instruction Axial Turbine Flow Sensor Series Turbotron VTH 25 / VTI 25 / VTM 25 Table of contents Page 1 Function of Turbotron 1 2 Safety instructions 2 3 Important notes and requirements to installation and operation 3 4 Installation in piping 3 4.1 Installation of flow sensor made of brass or stainless steel using connecting adaptors (recommended kind of installation) 3 4.2 Installation of flow sensor made of brass or stainless steel without connecting adaptors 4 4.3 Installation of flow sensor made of plastics using hose barbs 4 4.4 Installation of flow sensor made of plastics using gluing or welding adaptors 4 4.5 Installation of flow sensor made of plastics without fittings 5 5 Electrical connection 6 5.1 Turbine flow sensor with pulse output 6 5.2 Turbine flow transmitter with analog output, version AI 6 5.3 Integrated temperature sensor (option) Pt 100/3-wire or Pt 1000/3-wire 6 Replacement of turbine insert 6 7 Cleaning of Turbotron 7 8 Shut-down and disposal 7 9 Materials table 8 10 Technical data 8 10.1 Turbine flow sensor with pulse output 10.2 Turbine flow transmitter with analog output, version AI 11 Dimensions 9 1 Function of Turbotron The turbine flow sensor of the series Turbotron is a transducer for flow rate and total flow measurement. It has an almost unlimited application through its exceptionally compact design, its very wide measurement range and its convincing measurement accuracy. The liquid flowing through the flow sensor makes the turbine wheel to rotate. Based on high-quality sapphire bearings and low rotational speed, the turbine assures an exceptionally long life expectancy. The rotor speed is transmitted to an electrical pulse signal (frequency): - VTH and VTM are equipped with magnets on the rotor. A Hall-Effect sensor detects the rotation of the rotor. - VTI has stainless steel pins in the rotor. An inductive proximity switch detects the rotation of the rotor. In both cases, a flow-proportional frequency signal (square wave signal) is provided. Technical changes reserved 1

2 Safety instructions Before you install the product, please read the relevant chapters of the installation instruction carefully. The turbine flow sensor is only suitable for measuring liquid flows, on no account for measuring gas flows. Check before installation, whether the material of the turbine flow sensor is suitable for the medium to be monitored (see Materials Table, Chapter 9)! You can fit the turbine flow control instrument in any position. If it is installed into vertical pipes, the flow direction is preferably upwards. You must avoid a free outlet. The arrow which is placed on the flow sensor ( ) shows the only permitted flow direction. For precise measurement, the length of the straight in - and outlet tubes must be observed (see Chap. 3: Requirements on Installation and Operation) The internal diameter of the in - and outlet tube must correspond with the internal diameter of the flow sensor. The flow medium to be monitored should preferably contain as few solid particles as possible. Present particles must not exceed a diameter of 0.63 mm. If necessary, install a filter! Avoid absolutely the formation of gas bubbles or cavitation in the medium by taking proper measures. The material of the instrument is not suitable for monitoring oils. The strength of the used plastic parts would be considerably reduced. In order to clean the flow sensor of contaminations, flush the unit with water reverse to the flow direction (see Chap. 7). The instrument must not be blown out with compressed air. The turbine can be damaged. We recommend to use only screened connection cables. Connect the shield on one side (the wire ends) on ground. Attention: The pick-off (Hall-Effect-Sensor or inductive proximity switch) is sealed and must not be opened. There are special customer designs that may differ from the standard data listed in these instructions always consider the specifications noted on the type plate. Technical changes reserved 2

3 Important notes and requirements to installation and operation Observe the following instructions in order to achieve highest-possible measurement accuracy and specified output signal. Before installing the turbine flow sensor flush the pipe carefully. You avoid a blocking of the turbine caused by particles from the pipe installation. The installation position of the flow sensor is unreserved. If it is installed into vertical pipes, the flow direction is preferably from below upward. You must avoid a free outlet. The arrow which is placed on the flow sensor ( ) shows the only permitted flow direction. A straight tube in front (upstream) of the flow sensor must be retained, min 10 x DN, i.e. 25 cm. Behind the flow sensor (downstream), a straight outlet tube of 5 x DN, i.e. 12.5 cm, must be kept. The internal diameter of the in- and outlet tubes must correspond with the internal diameter of the flow sensor = 25 mm. Before and behind the stabilization tubes, the line may be contracted or enlarged. In practice these instructions often can not be observed. Then the pulse rate and the measurement accuracy can be affected. The flow medium to be measured should preferably contain as few solid particles as possible. Present particles must not exceed a diameter of 0.63 mm. If necessary, install a filter! The materials of the instrument are not suitable for monitoring oils. The strength of the used plastic parts would be considerably reduced. Attention: The pick-off (Hall-Effect-Sensor or inductive proximity switch) is sealed and must not be opened. 4 Installation in piping Now you can install the Turbotron in the piping system which was prepared according to chapter 3. 4.1 Installation of flow sensor made of brass or stainless steel using connecting adaptors (recommended kind of installation) At first screw-in the connecting adaptors into the tube. Use only a suitable compound for sealing. Take care that no fibrous sealing compounds get into the turbine (hemp or Teflon strip). Now install the turbine. Make sure that the provided seals fit properly and tighten the union nuts. Technical changes reserved 3

4.2 Installation of flow sensor made of brass or stainless steel without connecting adaptors The turbine system fits into the Turbotron casing and is correctly positioned by the two connecting adaptors. If you do not want to use connecting adaptors, you must observe the following instructions: Your piping system must have a collar at the outlet side of the Turbotron which prevents slipping of the turbine insert. Recommended dimensions are: Internal diameter: 25-26 mm External diameter: 38-38,5 mm When installing the Turbotron, make sure that the seals fit properly. 4.3 Installation of flow sensor made of plastics using hose nozzles In the outlet of the Turbotron, there is a spacer. Remove it before installation of the hose barb. Push the hose barb with a rotational movement into the Turbotron casing. Make sure that the O-ring is not displaced and tighten the union nut. Put the hose barb into the piping and fix it with suitable means, e.g. with hose band clips. 4.4 Installation of flow sensor made of plastics using gluing or welding adaptors Glue or weld at first the connecting adaptors to the piping. Now install the turbine. Make sure that the provided seals fit properly and tighten the union nuts. Technical changes reserved 4

4.5 Installation of flow sensor made of plastics without fittings Install the Turbotron without removing the spacer from the outlet. Your piping system must have a collar at the outlet side of the Turbotron which prevents slipping of the turbine insert. Recommended dimensions are: Internal diameter: 25-26 mm External diameter: 38 mm When installing the Turbotron, make sure that the seals fit properly. Technical changes reserved 5

5 Electrical connection Attention: We recommend to use only screened connection cables. Connect the shield on one side (the wire ends) on ground. 5.1 Turbine flow sensor with pulse output The output signal of Turbotron is a flow-proportional frequency signal. The shape of the signal is a square wave and its amplitude corresponds approximately with the supply voltage. It is an open collector signal, NPN- or PNP-switching. The connected electronic instrument should have a loading resistance (pull-up or pull-down resistor) of 5 kω in the inlet. Schematic representation A connection is made with three leads, the supply voltage must be connected between +U and GND (ground), the output signal can be tapped between and GND. The color assignment of the supply cables can be taken from the sketch on the type plate. VTH VTM BN = brown, GN = green, WH = white, R = resistor VTI 5.2 Turbine flow transmitter with analog output, version AI 5.3 Integrated temperature sensor Pt 100/3-wire or Pt 1000/3-wire (option) BN = brown GN = green WH = white Technical changes reserved 6

6 Replacement of turbine insert Dismount the flow sensor. The pick-off (Hall-Effect-Sensor or inductive proximity switch) is sealed and must not be opened. Press the turbine insert out of the tube piece in flow direction using a flat tool. The insert fits very tight in the tube piece. You should not use your fingers and never use a pointed tool to press it out of the tube. The turbine insert consists of two cylinders of different diameters which must never be dismounted. Push the new insert with the small diameter to the front into the pipe section against the flow direction. Turn the insert in such a way that the webs are not directly beneath the Hall sensor or the proximity switch. Press the insert into the pipe section up to the stop. The position will be correct, if the face of the inserts is flush with the pipe section (applies only for metallic version). Plastic version: push the insert up to the stop, now do the same with the spacer. The spacer must be flush with the tube piece. Reinstall the Turbotron in the piping. Make sure that the provided seals fit properly. Technical changes reserved 7

7 Cleaning of Turbotron To remove dirt from the flow sensor, you should flush it with water reverse to the flow direction. Attention: The pick-off (Hall-Effect-Sensor or inductive proximity switch) is sealed and must not be opened. Warning: You must not blow-out the instrument with compressed air. The turbine bearings can be damaged. 8 Shut-down and disposal Remove the electrical connection and carry out the flow sensor. The Turbotron unit is made of different materials (see technical data). Don't dispose the Turbotron unit together with household waste. The official rules of your area have to be observed at disposing of the Turbotron unit. Technical changes reserved 8

9 Materials Table Materials Type VTH 25 MS-180 VTH 25 K6-180 VTM 25 MS-180 VTM 25 VA-180 VTI 25 MS-180 VTI 25 K6-180 Pipe section Brass, DZR CuZn36Pb2As CW602N PP Brass, DZR CuZn36Pb2As CW602N Stainless steel 1.4571 Brass, DZR CuZn36Pb2As CW602N PP Turbine cage PPO Noryl GFN 3V 960 Rotor PPO Noryl GFN 2V 73701 Rotor assembly Permanent magnets, Recona 28 nickel plated Stainless steel 1.4305 Shaft Stainless steel 1.4436 Bearing Pick-off PPO Noryl GFN 1630 V Brass, DZR CuZn36Pb2As CW602N Sapphire / PA O-ring 72 NBR 872 Filtering screen (option) matching O- ring Stainless steel 1.4301 70 EPDM 281 --- Housing transducer (version AI) Stainless steel 1.4571 Stainless steel 1.4301 70 EPDM 281 PA POM Delrin 100 P Spacer --- PP --- --- --- PP --- 10 Technical Data 10.1 Turbine flow sensor with pulse output** Type VTH VTM VTI Material pipe section Brass Plastics Brass Stainless steel Brass Plastics Pick-off Hall sensor Inductive proximity switch Nominal diameter DN 25 Measurement range (1) 4...160 l/min, continuous operation max. 80 l/min Start of signal output < 1 l/min Measurement accuracy +/- 3% of reading Reproducibility +/- 0.5% Output signal - Signal shape Square signal NPN open collector - Pulse rate / K-factor - Resolution - Frequency range Amplitude Signal current Pull-up-resistance Supply voltage Electrical connection * additional fitting required 65 pulses / liter 15,4 ml / pulse (1,08) 4,33...173,3 Hz ~ applied power supply max. 20 ma 5 K (recommendation) 10... 30 VDC 2 m PVC cable, screened 4-pin plug M12x1 (T max = 75 C) Protection class IP 54 Max. medium temperature 85 C 80 C (2 bar) 60 C (5 bar) 30 C (10 bar) Square signal PNP open collector 65 pulses / liter 15,4 ml / pulse (1,08) 4,33...173,3 Hz ~ applied power supply max. 50 ma 85 C 60 C 60 C (5 bar) 30 C (10 bar) Nominal pressure PN10 PN50 PN10 Max. size of foreign particles < 0.63 mm Process connection 1¼ BSP male* 1¼ BSP male 1¼ BSP male* 1¼ BSP male **There are special customer designs that may differ from the standard data listed in these instructions always consider the specifications noted on the type plate. Technical changes reserved 9

10.2 Turbine flow transmitter with analog output, version AI** Type VTH VTM VTI Material pipe section Brass Plastics Brass Stainless steel Brass Plastics Pick-off Hall sensor Inductive proximity switch Nominal diameter DN 25 Measurement range (scaling see type plate 4...20 ma corresponding to (continuous operation max. 80 l/min) 0...60 l/min, 0...100 l/min 0...160 l/min Start of signal output < 1 l/min Output signal 4...20 ma Current limit approx. 26 ma Max. current consumption 30 ma Max. resistance 250 against GND Residual ripple 0,2 ma SS over the entire range Type 3 wire, galvanically not separated, common GND of power supply and output signal Power supply 18...30 VDC Electrical connection 4-pin plug M12x1 Protection class IP 54 Max. medium temperature 85 C 80 C (2 bar) 60 C (5 bar) 30 C (10 bar) * additional fitting required **There are special customer designs that may differ from the standard data listed in these instructions always consider the specifications noted on the type plate. 11 Dimensions (selection) VTH 25 MS-180 with connecting adaptor 85 C 60 C 60 C (5 bar) 30 C (10 bar) Nominal pressure PN10 PN50 PN10 Max. size of foreign particles < 0.63 mm Process connection 1¼ BSP male* 1¼ BSP male 1¼ BSP male* 1¼ BSP male Technical changes reserved 10

VTM 25 MS-180 VTM 25 VA-180 VTH 25 K6-180 Turbine flow transmitter, version AI Technical changes reserved 11