Electrical Safety in Medical Equipment. D. J. McMahon rev cewood

Size: px
Start display at page:

Download "Electrical Safety in Medical Equipment. D. J. McMahon rev cewood"

Transcription

1 Electrical Safety in Medical Equipment D. J. McMahon rev cewood

2 Key Points Electrical Safety in Medical Equipment: - know the significance of NFPA-99 and the annual editions - define 'the patient care vicinity' (PCV) - know about the 'wet location' and LI vs GFCI - know the definitions AND SYMBOLS for applied parts B, BF, CF - recognize the symbols for 'defibrillation proof' - know the methods for electrical signal isolation & power isolation - be able to define leakage current; what contributes to it? - recognize the electrical safety standards and know which one is commonly used in the US - MEMORIZE the 2015 NFPA-99 standards; 3 values, including the units - do these values apply to non patient care devices? - know the easy fix for high leakage, often built-in to devices - when do we do electrical safety testing? - understand the problem of finding an actual chassis ground point - how do these standards apply to two-prong plug devices? - know about testing 'systems' - know about Line Isolation & LIMs - are extension cords and power strips allowed? what are the requirements?

3 Birth of the HTM Profession : (March 1971)

4 NFPA 70 : National Electrical Code (NEC ) Sets requirements for building design, construction, operation, and maintenance. Applies to all buildings. Adopted as code in 46 states, the VA system, and the Centers for Medicare and Medicaid Services (CMS).

5 National Electrical Code (NEC) Section 517: Health Care Facilities Discusses electrical service for hospitals and clinics > General electrical distribution > Power supply backups > Defines emergency power distribution > Specifies electrical concerns for specialty areas: - anesthetizing locations - isolated power systems and their monitors - pediatric areas - X-ray

6 NFPA 99 Sets criteria for health care services based on risk to the patients, staff, or visitors in health care facilities to minimize the hazards of fire, explosion, and electricity. Includes installation, inspection, testing, and maintenance of facilities, equipment, and appliances, including medical gas and vacuum systems.

7 Per NFPA 99: The Patient Care Vicinity A space within the location intended for patient care, extending 1.5 m (~ 6 feet) beyond the normal location of the bed, table, etc that supports the patient during examination or treatment, and extending vertically 2.5 m (~7.5 feet) above the floor. - Clinical Engineering is responsible for all patient-impact equipment that is within the Patient Care Vicinity. Any equipment in that area must meet NRTL or NFPA-99 electrical safety standards. - Equipment in non-patient care areas (nurses stations, lounges, etc) does not need to meet those standards but may be checked by Facilities Engineering or Clinical Engineering, depending on hospital policy.

8 The Patient Care Vicinity (the bubble )

9 Patient Care Vicinity: a crowded place!

10 Wet Location The area in a patient care area where a procedure is performed that is normally subject to wet conditions while patients are present, including standing fluids on the floor or drenching of the work area, either if which is intimate to the patient or staff. - NFPA-99 (2005) > typically applies to surgical suites and other areas > controversial topic lately - architects tending away from Line Isolation in O.R.s - GFCI s being used instead

11

12 Applied Part of medical equipment as defined by the IEC standard : Type B: Applied parts that are generally not conductive and can be immediately released from the patient. May be grounded. eg: Non-invasive BP monitors [think: Body] Type BF: Devices that have direct contact with the patient, or parts that have long term contact with the patient. eg: ECG monitors [think: Body, Floating (ground)] Type CF: Applied parts that have direct contact with the heart. eg: Invasive pressure monitors, defib paddles [think: Cardiac Floating (ground)]

13 ONE of these ratings should be present on a medical device see the little defib paddles? If Defib Proof, ONE of these

14 IEC Definitions: Class I Equipment Device is protected against shock by use of a grounding wire connected to the exposed conductive parts of the device. Class II Equipment Device is protected by additional insulation in addition to the basic design insulation. Usually called Double Insulated.

15 Why aren t these birds electrocuted?

16 because they re not grounded. Electrical safety is all about current path to ground.

17 How to electrocute homo sapiens

18

19 (through skin) I = 120V / 100KW = 1.2 ma = 1200 µa (under skin) I = 120V / 25KW = 4.8 ma = 4800 µa

20 Ungrounded (2-wire) wiring

21 Grounded (3-wire) wiring

22 Chassis Ground (at power entry)

23 Electrical Isolation: transformer isolation for power opto-isolator for electrical signal:

24 Leakage current: or distributed capacitance Leakage current is both capacitive, caused by intrinsic capacitance between conductors; and resistive, caused by imperfect insulation.

25 RF filters at power entries will add to capacitive leakage:

26 Where does the leakage current go?

27 Open ground! Where does the leakage current go now?

28 Fatality by microshock scenario:

29 NEMA / AAMI (US): Electrical Safety Testing Terminology Hot Neutral Ground Line Voltage Patient Contacts (leads, paddles) Chassis Ground Wire Leakage Current in Ground Wire Chassis leakage - - Lead leakage Leakage between Patient Leads Lead Isolation Ground Wire Resistance - Insulation Resistance between Hot and Neutral to Ground IEC (international): L1 L2 Earth Mains Applied Parts Enclosure Protective Earth (PE) Earth Leakage Current Enclosure Leakage Patient Leakage Patient Auxiliary Mains on Applied Parts (MAP) Earth Resistance Insulation Resistance

30 What are the electrical safety standards? AAMI ES-1 (same as NFPA99) - most commonly used in the U.S. IEC (same as UL in the U.S. CSA C22.2 in Canada) - most widely used, but written for manufacturers IEC written for end-users; a bit looser requirements AN/NZS 3551 (modified )

31 The NFPA-99 Electrical Safety Test: 1) ground integrity 2) chassis leakage 3) patient contact lead-to-ground leakages 4) patient contact lead-to-lead leakages 5) patient contact lead leakage while in contact with the line voltage (MAP) Know your analyzer!!

32 NFPA99 Chassis Leakage (touch) Current 1999 edition 2005 edition 2012 edition 2015 edition * Touch Current Portable Equipment. The touch current for cordconnected equipment shall not exceed 500 μa with normal polarity and the ground wire disconnected (if a ground wire is provided).

33 still valid:.5 Ω 1999 rev; PRE-2012! standard has changed!

34 Electrical Safety Testing Values (per NFPA-99, 2012 edition) Resistance of Ground wire of power cord: 500 milliohms (500mΩ or.5ω) Leakage current at the chassis: NC (normal condition) 100 microamps (100 μa) SFC (Single Fault Condition, e.g. open ground) 500 microamps (500 μa) Leakage current for any one patient contact lead to ground: 10 μa with ground closed, 50 μa with ground open. Leakage current for each patient contact lead to any other lead: 10 μa with ground closed, 50 μa with ground open. Leakage current for patient leads exposed to line voltage: 50 μa

35 Standard circuit described in NFPA 99 for measurement of leakage current: Applied Part = any component of a device that is directly applied to the patient. (Electrodes, paddles, sensors, etc)

36 Test device measuring chassis leakage current: note: no switch to open neutral leg; grounding switch shown in SFC (single fault condition)

37 Test device measuring patient lead leakage current: (DUT)

38 Test device measuring isolation current with MAP (Mains on Applied Parts): (DUT)

39 What about non-patient care devices? For Non-Patient Care devices, within the PCV: Leakage current at the chassis: 500 microamps (0.5 ma) e.g. a mobile computer terminal For Non-Patient Care devices, not in the PCV: the PCV Leakage current at the chassis: 3500 microamps (3.5 ma) e.g. a vacuum cleaner

40 Non-Medical Grade equipment - esp Medical computers, TV s, etc - can have high leakage current due to capacitance in line filters - corrective action for these if used in patient care areas: - connect a redundant ground wire, or - go thru an isolation transformer in the power input

41

42 What about plastic-enclosed, (sometimes battery-driven) medical devices? Many medical devices have no electrically conductive surfaces to access as the enclosure for the purposes of connecting a test device to check leakage or ground integrity. NFPA-99 has a specific way to establish a temporary reference point using a 10x10 cm sheet of foil, but it is not very practical. You will often need to find any possible metallic point that is continuous with the device s ground. In some cases, the device simply can t be checked for leakage.

43 e.g. an irrigation pump used in surgery:

44 When to we do Electrical Safety Testing? 1) on any incoming inspection of new equipment, including rented, borrowed, or donated equipment 2) after any repair or modification 3) as part of scheduled Performance Verifications (depending on department policy) - may also be useful as part of troubleshooting

45 Typical Electrical Safety Analyzers

46 In the NSCC Lab: Rigel SafeTest 50 Dale 544D+ Fluke ESA612

47 The resistance measurement problem:

48

49 One solution: Kelvin (4-wire) resistance leads

50 Kelvin Lead set on an electrical safety tester:

51 Chassis Ground Lug (or POAG: Point of Actual Ground)

52

53 Grounding wire POAG connector

54 Where do we find chassis ground?

55 What about testing multiple medical devices (or systems )? System in the context of medical equipment inventory, refers to a number of medical devices commonly used together, which share a common power strip. Generally, the electrical safety check of a system can be done at the common power cord that feeds it. - be certain that the total current draw of the system, when at maximum, does not exceed ~75% of the power strip s rating.

56 Example: typical medical video tower

57 What about Fixed Equipment? i.e. equipment that is permanently installed to the building structure. e.g.: imaging systems, treadmills, large centrifuges. Point-to-Point testing for leakage current and resistance is done with ES analyzers on equipment that is fixed. Acceptable leakage limits are higher.

58 Standard power distribution grounding:

59 Isolated power branch ( Line Isolation ):

60 Isolated Power System - Generally used in Operating Rooms

61 Line Isolation Monitor (LIM)

62 LIM s alternately look at each side of the line to check for excessive leakage relative to ground:

63

64 Circuit Breaker Panel

65 Line isolation transformer Line isolation monitor (LIM)

66

67 Typical setting and reading for a LIM

68

69 Extension Cords: Generally allowed for use in the patient care vicinity, IF: > They use wire that exceeds current capacity of the power outlet being extended. (usually 14 gauge or 12 gauge, + ground) > Designed with medical-grade components.

70

71 Multioutlet Extension Cords Multiple Outlet Strips Power Strips Relocatable Power Taps Standardized in UL #1363. Allowed by NFPA 99, IF: - they are made of medical grade components - they are of adequate ampacity for the need - they are inventoried and maintained

72

73 Not Hospital Grade Hospital Grade

74 Relocatable Power Taps are now acceptable IF: -- Are permanently attached to equipment. -- The total current drawn is <75% of the cord. -- The cord complies with NFPA The power tap is checked as part of PM s. -- Added devices aren t attached to exceed the allowed leakage current. (CMS Waiver, Sept 25, 2014)

75 Relocatable Power Taps - keep em short! Good practice to keep Relocatable Power Taps short Longer ground wire increases ground wire resistance Longer hot & neutral wires add voltage drop Longer power cables add clutter to the room Longer power cables increase tripping hazard

76

77 PowerBall PowerMate

78 2-prong to 3-prong adapters or Cheaters : Not a safe solution to the ungrounded receptacle, and NOT ALLOWED in healthcare facilities. (These assume that the electrical box on the wall is a good, reliable ground point, tied to earth ground through metal conduit. Even though the building wiring is not grounded? Not a good bet anywhere!)

79 Surge protection power strips: NOT for medical devices!

80

81 AC/DC adapters (plug-in power supplies) Inspect routinely, and replace as needed

82 Plug-in power supplies - Allow manufacturers to internationalize easily. - Requirements for meeting NRTL certification of the product is reduced by isolating the line-level circuitry. - Lets the power supply manufacturer deal with the expense and liability of certification. - Makes repair by the customer possible. Choosing (or replacing) a plug-in supply "only" requires attention to these details: - input voltage and frequency (your national standard) - input connector (NEMA, IEC, or 17 other national standard plugs) - output voltage - output current (maximum available) - whether the output is DC or AC - for DC, whether the supply is: linear un-regulated (must match load) linear regulated switching - if the power supply will fit physically where it needs to go AND... - the output connector!

83 Package insert with ECG wires, before 1985:

84 Any signal connector for patient contact, (eg for ECG) must either be female, or be must be protected by a shield so that it cannot be connected to a source of lethal voltage. - change in ANSI design standards, effective in 1987 after a fatal incident in early 80s

85 Female ( ) Recessed Male ( )

86

CHAPTER 2: ELECTRICAL SAFETY

CHAPTER 2: ELECTRICAL SAFETY CHAPTER 2: ELECTRICAL SAFETY 2.1 Physiological Effect of Electricity 2.2 Macroshock and Microshock Macroshock Microshock 2.3 Approaches toward Protection against Shock 2.4 Medical Safety Test Electrical

More information

Date received into service; / /.

Date received into service; / /. Innovating Together RIGEL SafeTest 50 electrical safety analyzer Copyright 2016 SEAWARD GROUP Last Update: April 20th 2016 Instruction Manual 410A570 Revision 1.2 Limited Warranty & Limitation of Liability

More information

SafeTest 60. A simple, robust and cost-effective medical safety analyser for general electrical safety testing. Key Features

SafeTest 60. A simple, robust and cost-effective medical safety analyser for general electrical safety testing. Key Features SafeTest 60 A simple, robust and cost-effective medical safety analyser for general electrical safety testing The Rigel SafeTest 60 is a robust, reliable and dedicated medical safety analyser. It is ideal

More information

Testing Power Supplies for Medical Electronics Applications

Testing Power Supplies for Medical Electronics Applications white pƒper Testing Power Supplies for Medical Electronics Applications g Introduction The international stard IEC 60601 specifies testing performance for electrical devices used in medical applications.

More information

GET GROUNDED. Renewable Energy System Grounding Basics

GET GROUNDED. Renewable Energy System Grounding Basics GET GROUNDED Renewable Energy System ing Basics by Christopher Freitas ASK TEN RENEWABLE ENERGY INSTALLERS ABOUT SYSTEM GROUNDING and you ll likely get ten different opinions as to what the National Electrical

More information

1 AMP CURRENT SOURCE

1 AMP CURRENT SOURCE 1 AMP CURRENT SOURCE CS-2000-U CS-2000-E USER MANUAL BC BIOMEDICAL CS-2000 SERIES TABLE OF CONTENTS WARNINGS, CAUTIONS, NOTICES... ii DESCRIPTION... 1 LAYOUT... 2 OPERATION... 4 TESTING... 6 MANUAL REVISIONS...

More information

INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS.

INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS. INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS. Before installing and/or using this product, please check for any visual damage of the enclosure, power

More information

Arc Fault Circuit Interrupter (AFCI) FACT SHEET

Arc Fault Circuit Interrupter (AFCI) FACT SHEET Arc Fault Circuit Interrupter (AFCI) FACT SHEET THE AFCI The AFCI is an arc fault circuit interrupter. AFCIs are newly-developed electrical devices designed to protect against fires caused by arcing faults

More information

INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS.

INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS. INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS. Before installing and/or using this product, please check for any visual damage of the enclosure, power

More information

Ensuring the Safety Of Medical Electronics

Ensuring the Safety Of Medical Electronics Chroma Systems Solutions, Inc. Ensuring the Safety Of Medical Electronics James Richards, Marketing Engineer Keywords: 19032 Safety Analyzer, Medical Products, Ground Bond/Continuity Testing, Hipot Testing,

More information

Electrical Safety. Electrical Safety Webinar. Electrical. Printing Industries Alliance Printing Industries Alliance 1

Electrical Safety. Electrical Safety Webinar. Electrical. Printing Industries Alliance Printing Industries Alliance 1 Webinar 1 Electrical 2 1 Webinar Introduction An average of one worker is electrocuted on the job every day There are four main types of electrical injuries: Electrocution (death due to electrical shock)

More information

Learning Module 10: Loadcenters. 101 Basic Series

Learning Module 10: Loadcenters. 101 Basic Series Learning Module 10: Loadcenters 101 Basic Series What You Will Learn We ll step through each of these topics in detail: What Does a Loadcenter Do? 4 Applications 4 Basic Circuitry and Wiring 5 Residential

More information

Does Your Electrical Safety Testing Routine Need a Boost? Exploring the Benefits of the Line Leakage and Functional Run Tests

Does Your Electrical Safety Testing Routine Need a Boost? Exploring the Benefits of the Line Leakage and Functional Run Tests Does Your Electrical Safety Testing Routine Need a Boost? Exploring the Benefits of the Line Leakage and Functional Run Tests Introduction Most electrical product manufacturers perform safety tests on

More information

Electrical Testing Company Pty Ltd / / AS/NZS 3003:2011 Body (Rev E)

Electrical Testing Company Pty Ltd /  / AS/NZS 3003:2011 Body (Rev E) AS/NZS 3003:2011 Electrical Testing Company Pty Ltd / www.electricaltestingcompany.com.au / AS/NZS 3003:2011 Body (Rev E) Body Protected Electrical Areas (shall comply with Section 2 and Section 3 of the

More information

An average of one worker is electrocuted on the job every day There are four main types of electrical injuries:

An average of one worker is electrocuted on the job every day There are four main types of electrical injuries: Electrical Safety Introduction An average of one worker is electrocuted on the job every day There are four main types of electrical injuries: Electrocution (death due to electrical shock) Electrical shock

More information

Language specific user manuals are available on

Language specific user manuals are available on User Manual WEB Charger for LiFePO 4 Batteries EN User Manual Language specific user manuals are available on www.mascot.no/downloads/usermanuals Bruksanvisning Käyttöohjeet Bedienungsanleitung Mode d

More information

Why Electric Safety Standards

Why Electric Safety Standards Paul Haggarty Why Electric Safety Standards To protect patients and caregivers from electrical shock To prevent electrical interference between instrument sub-systems/circuits To prevent electrical interference

More information

Electrical Hazards in Construction

Electrical Hazards in Construction Electrical Hazards in Construction WELCOME Tim Dawson Presentation developed by Dawson Safety Services - 2013 Dangers of Electrical Shock Currents above 10 ma* can paralyze or freeze muscles. Currents

More information

ELECTRICAL. 60 Minutes

ELECTRICAL. 60 Minutes ELECTRICAL 60 Minutes AGENDA Electrical definitions Electrical shocks, burns and secondary injuries Electrical hazards in the workplace Safety devices and prevention steps Do s and Don ts INTRODUCTORY

More information

ECET 211 Electric Machines & Controls Lecture 1-3 (Part 2) Electrical Safety in the Workplace Electrical Safety in the Workplace

ECET 211 Electric Machines & Controls Lecture 1-3 (Part 2) Electrical Safety in the Workplace Electrical Safety in the Workplace ECET 211 Electric Machines & Controls Lecture 1-3 (Part 2) Electrical Safety in the Workplace Text Book: Electric Motors and Control Systems, by Frank D. Petruzella, published by McGraw Hill, 2015. Other

More information

ELECTRICAL ENGINEERING & COMPUTER SCIENCE NEWSLETTER Volume

ELECTRICAL ENGINEERING & COMPUTER SCIENCE NEWSLETTER Volume ELECTRICAL ENGINEERING & COMPUTER SCIENCE NEWSLETTER Volume 4 2016 A Treatise on Grounding of AC Electrical Systems, and an Atypical Electrocution Case Study i Author: Noshirwan K. Medora, SMEE, P.E. Reviewer:

More information

FACT SHEET Standard: Electrical Safety

FACT SHEET Standard: Electrical Safety What is a Ground Fault Circuit Interrupter? FACT SHEET The ground-fault circuit interrupter, or GFCI, is a fast-acting circuit breaker designed to shut off electric power in the event of a ground-fault

More information

ELECTRICAL CONSISTENCY MEETING

ELECTRICAL CONSISTENCY MEETING ELECTRICAL CONSISTENCY MEETING Date: 10/14/2015 Our Goal is FOR ALL STAKEHOLDERS TO HAVE A SUCCESSFUL PROJECT with us and that together we keep people and structures safe and economically viable to serve

More information

Go to to register your product and find more information.

Go to   to register your product and find more information. ESA609 Electrical Safety Analyzer Safety Sheet Go to www.flukebiomedical.com to register your product and find more information. Safety Information A Warning identifies hazardous conditions and actions

More information

AVOIDING ELECTROCUTION HAZARDS

AVOIDING ELECTROCUTION HAZARDS Print AVOIDING ELECTROCUTION HAZARDS MODULE DESCRIPTION This module will provide information that helps you recognize and avoid common electrocution hazards in your work environment. OBJECTIVES After completing

More information

Electrical Safety. Recognizing & Controlling Hazards

Electrical Safety. Recognizing & Controlling Hazards Electrical Safety Recognizing & Controlling Hazards Introduction Healthcare facilities, including hospitals, physician offices, laboratoratories, dental offices and beyond are vulnerable to a variety of

More information

Electrical Testing Company Pty Ltd / download at electricaltestingcompany.com.au / AS/NZS 3003:2011 Cardiac (Rev C)

Electrical Testing Company Pty Ltd / download at electricaltestingcompany.com.au / AS/NZS 3003:2011 Cardiac (Rev C) AS/NZS 3003:2011 Cardiac Protected Electrical Areas (shall comply with Section 2 and Section 4 of the standard) The following is a guide for electricians involved with electrical installations for patient

More information

2.7. IT system floor-standing distribution cabinet Series -IPS-F

2.7. IT system floor-standing distribution cabinet Series -IPS-F Dipl.-Ing. W. Bender GmbH & Co. KG Londorfer Str. 65 35305 Grünberg Tel.: 0640 07-0 Fax: 0640 07-25 IT system floor-standing distribution cabinet Series -IPS-F for supplying power to medical locations

More information

Power Supplies and Overcurrent Protection (fuses)

Power Supplies and Overcurrent Protection (fuses) Power Supplies and Overcurrent Protection (fuses) Key Points Power Supplies & Overcurrent Protection: - know the 5 steps in a linear power supply - know what the waveform is at each stage - be familiar

More information

Understanding National Electric Code (NEC) tap rules How do they apply to circuit breaker terminals?

Understanding National Electric Code (NEC) tap rules How do they apply to circuit breaker terminals? White paper Understanding National Electric Code (NEC) tap rules How do they apply to circuit breaker terminals? Darryl Moser, Business Development Manager, DEM Sales, ABB, Electrification Products Division

More information

Installing Power Components

Installing Power Components This chapter provides instructions on how to install and reinstall power components in the Cisco NCS 4016 chassis. It also covers connecting and disconnecting power and powering on the chassis. The Cisco

More information

Electrical Wiring Practices

Electrical Wiring Practices Chapter 2 Electrical Wiring Practices and Diagrams MElec-Ch2-1 Overview Safety Standards Wiring Considerations Wire Terminations Coaxial Cable Wiring Installations Wiring Diagrams MElec-Ch2-2 Safety Lethal

More information

Language specific user manuals are available on

Language specific user manuals are available on User Manual WEB Charger for Lead Acid Batteries EN User Manual Language specific user manuals are available on www.mascot.no/downloads/usermanuals Bruksanvisning Käyttöohjeet Bedienungsanleitung Mode d

More information

Eaton Corp. - Marina Power and Lighting, Inc. General Specifications for Lighthouse Power Pedestal

Eaton Corp. - Marina Power and Lighting, Inc. General Specifications for Lighthouse Power Pedestal DRILL \ PUNCH OUT TO SIZE REQUIRED DRILL \ PUNCH OUT TO SIZE REQUIRED Lighthouse Power Pedestal - Specifications Lighthouse Power Pedestal Dimensions Height: 44.00" (1117.60 mm) Width: 13.67" (347.22 mm)

More information

WIRING DESIGN & PROTECTION REQUIREMENTS CHECKLIST

WIRING DESIGN & PROTECTION REQUIREMENTS CHECKLIST WIRING DESIGN & PROTECTION REQUIREMENTS CHECKLIST Use & Identification of Grounded and Grounding Conductors YES NO N/A Grounded conductors are identifiable and distinguishable from all other conductors.

More information

TEMPORARY ELECTRIC WIRING FOR CARNIVALS, CONVENTIONS, EXHIBITIONS, FAIRS AND SIMILAR USES

TEMPORARY ELECTRIC WIRING FOR CARNIVALS, CONVENTIONS, EXHIBITIONS, FAIRS AND SIMILAR USES INFORMATION BULLETIN / PUBLIC - ELECTRICAL CODE REFERENCE NO.: LAMC 93.0230 Effective: 3-24-69 DOCUMENT NO. P/EC 2002-006 Revised: 11-17-00 Previously Issued As: RGA #7-69 TEMPORARY ELECTRIC WIRING FOR

More information

Basic. Electrical Safety. January 2009

Basic. Electrical Safety. January 2009 Basic Electrical Safety January 2009 Sunday, 10 January 2010 The complete circuit A complete Circuit or loop is necessary for current to flow Sunday, 10 January 2010 2 A complete circuit complete Circuit

More information

User Manual. Charger for LTO Batteries. User Manual. Language specific user manuals are available on

User Manual. Charger for LTO Batteries. User Manual. Language specific user manuals are available on User Manual Charger for LTO Batteries EN User Manual Language specific user manuals are available on www.mascot.no/downloads/usermanuals Bruksanvisning Käyttöohjeet Bedienungsanleitung Mode d emploi Manual

More information

Power Supplies and Overcurrent Protection (fuses) D. J. McMahon rev cewood

Power Supplies and Overcurrent Protection (fuses) D. J. McMahon rev cewood Power Supplies and Overcurrent Protection (fuses) D. J. McMahon 141005 rev cewood 2017-10-10 Key Points Power Supplies & Overcurrent Protection: - know the 5 steps in a linear power supply - know what

More information

Installation and Construction Notes for EVSE4

Installation and Construction Notes for EVSE4 Installation and Construction Notes for EVSE4 You need to read and understand this if you want to build an EVSE that will be safe and need to pass a building inspectors review. Before beginning this process

More information

SUPPLEMENTAL CORRECTION SHEET FOR SOLAR PHOTOVOLTAIC SYSTEMS - ELECTRICAL

SUPPLEMENTAL CORRECTION SHEET FOR SOLAR PHOTOVOLTAIC SYSTEMS - ELECTRICAL SUPPLEMENTAL CORRECTION SHEET FOR SOLAR PHOTOVOLTAIC SYSTEMS - ELECTRICAL This is intended to provide uniform application of the codes by the plan check staff and to help the public apply the codes correctly.

More information

Electrical Safety Products for Construction, Manufacturing and Industrial Maintenance

Electrical Safety Products for Construction, Manufacturing and Industrial Maintenance Electrical Safety Products for Construction, Manufacturing and Industrial Maintenance 2010 Table of Contents Ground Fault Circuit Interrupters (GFCI) Shock Shield Plug-In Portables..........................................

More information

WF-5110R True Sine Wave Inverter

WF-5110R True Sine Wave Inverter Operator s Manual WF-5110R True Sine Wave Inverter WF-9900 Series WF-5110R ( The Inverter model number is located on the label on top of the enclosure) Distributed in the U.S.A. and Canada by ARTERRA DISTRIBUTION

More information

Service Entrance Methods

Service Entrance Methods Service Section Typical switchboards consist of a service section, also referred to as the main section, and one or more distribution sections. The service section can be fed directly from the utility

More information

Small Motor Control (1/4 HP HP)

Small Motor Control (1/4 HP HP) PDHonline Course E110 (2 PDH) Small Motor Control (1/4 HP - 200 HP) Instructor: Thomas Mason, P.E. 2012 PDH Online PDH Center 5272 Meadow Estates Drive Fairfax, VA 22030-6658 Phone & Fax: 703-988-0088

More information

Electrical Hazard Three factors determine the resistance of a substance to the flow of electricity: conductors insulators extreme caution

Electrical Hazard Three factors determine the resistance of a substance to the flow of electricity: conductors insulators extreme caution Electrical Safety Electrical Hazard Three factors determine the resistance of a substance to the flow of electricity: What it is made of. Its size. Its temperature. Substances with very little resistance

More information

Generator Fire Safety: Generator assemblies should be located outside the building.

Generator Fire Safety: Generator assemblies should be located outside the building. SECTION 33 70 00 - ELECTRICAL DISTRIBUTION PACKAGED GENERATOR ASSEMBLIES Generator Fire Safety: Generator assemblies should be located outside the building. All fuel piping from the outside of the building

More information

ELECTRICAL GROUNDING SAFETY PROGRAM

ELECTRICAL GROUNDING SAFETY PROGRAM ELECTRICAL GROUNDING SAFETY PROGRAM PURPOSE / SCOPE The purpose of this program is to ensure the proper installation, maintenance, inspection, and testing of equipment grounding conductors on construction

More information

Section Index Portable GFCI Right Angles In-Lines Duplex & Quad Boxes Right Angle & In-Line Quad Boxes User Attachables ELCIs & PRCDs

Section Index Portable GFCI Right Angles In-Lines Duplex & Quad Boxes Right Angle & In-Line Quad Boxes User Attachables ELCIs & PRCDs Coleman Cable, along with its subsidiary Technology Research Corporation (TRC), has an extensive line of electrical safety products that ensure protection from dangerous current leakage that can cause

More information

This is intended to provide uniform application of the codes by the plan check staff and to help the public apply the codes correctly.

This is intended to provide uniform application of the codes by the plan check staff and to help the public apply the codes correctly. SUPPLEMENTAL CORRECTION SHEET FOR SOLAR PHOTOVOLTAIC SYSTEMS (ELEC) This is intended to provide uniform application of the codes by the plan check staff and to help the public apply the codes correctly.

More information

DC-AC Power Inverter SAM Manual. Please read this manual before installing your inverter

DC-AC Power Inverter SAM Manual. Please read this manual before installing your inverter DC-AC Power Inverter SAM-100-12 Owner's Manual Please read this manual before installing your inverter WARNINGS TO REDUCE THE RISK OF FIRE, ELECTRIC SHOCK, EXPLOSION OR INJURY 1. Do not connect to AC distribution

More information

Pretest Module 24 Three-phase Service Entrance

Pretest Module 24 Three-phase Service Entrance Pretest Module 24 Three-phase Service Entrance 1. What is the most widely used three-phase service entrance system? 2. What are the three most common voltage combinations for three-phase, four-wire systems?

More information

P80074 (150 amp connectors) P80071 (300 amp connectors)

P80074 (150 amp connectors) P80071 (300 amp connectors) Overview: The Restore-Lite Temporary Power Restoration System is a compact mobile transformer unit that contains a single-phase, floating-tap transformer designed to restore power to customers who have

More information

2000 Cooper Bussmann, Inc. Page 1 of 9 10/04/00

2000 Cooper Bussmann, Inc. Page 1 of 9 10/04/00 DO YOU KNOW THE FACTS ABOUT SINGLE-POLE INTERRUPTING RATINGS? YOU MAY BE IN TROUBLE! Typical plant electrical systems use three-phase distribution schemes. As an industry practice, short-circuit calculations

More information

Please use the Q & A utility to ask us any questions concerning the material being presented.

Please use the Q & A utility to ask us any questions concerning the material being presented. Please use the Q & A utility to ask us any questions concerning the material being presented. Please contact Brittany Socha on the chat line or email Brittany.socha@ikonixusa.com if you would like a copy

More information

Integrated Power Center 2 (IPC2) Space saving designs that save time and reduce costs

Integrated Power Center 2 (IPC2) Space saving designs that save time and reduce costs Integrated Power Center 2 (IPC2) Space saving designs that save time and reduce costs What does IPC2 stand for? The Integrated Power Center 2 (IPC2) is a family of Square D products offered by Integrated

More information

Eaton Corp. - Marina Power and Lighting, Inc. General Specifications for Lighthouse-SS Power Pedestals

Eaton Corp. - Marina Power and Lighting, Inc. General Specifications for Lighthouse-SS Power Pedestals Lighthouse-SS Power Pedestal Dimensions Height: 44.00" (1117.60 mm) Width: 9.565" (242.95 mm) Depth: 9.875" (250.83 mm) Lighthouse-SS Base Diagram Ø0.50" MOUNTING HOLE (TYP.) 9.5625" 8.0625" 7.3125" C:\Documents

More information

Microinverters and AC PV modules are becoming. Microinverters and AC PV Modules. Different Beasts. Perspectives on PV.

Microinverters and AC PV modules are becoming. Microinverters and AC PV Modules. Different Beasts. Perspectives on PV. Perspectives on PV Microinverters and AC PV Modules Are Different Beasts by John Wiles Microinverters and AC PV modules are becoming very common in residential and small commercial PV systems. See photos

More information

INSTALLATION AND OPERATING INSTRUCTIONS

INSTALLATION AND OPERATING INSTRUCTIONS INSTALLATION AND OPERATING INSTRUCTIONS MANUAL TRANSFER SWITCHES FROM 0 Residential Wattage Requirements Appliance Running Watts Add watts for starting Furnace blower, gas or fuel 1/8 hp 300 500 1/8 hp

More information

OWNER S MANUAL AND INSTALLATION INSTRUCTIONS

OWNER S MANUAL AND INSTALLATION INSTRUCTIONS EmerGen Switch Manual Transfer Switch OWNER S MANUAL AND INSTALLATION INSTRUCTIONS For A Series Models 6-5001, 6-7501, 10-7501, 10-12K1 PLEASE READ THIS MANUAL IN ITS ENTIRETY BEFORE INSTALLING AND/OR

More information

SECTION LOW VOLTAGE DISTRIBUTION EQUIPMENT

SECTION LOW VOLTAGE DISTRIBUTION EQUIPMENT SECTION 16400 LOW VOLTAGE DISTRIBUTION EQUIPMENT A. General 1. The University does not accept Series-Rated equipment for power distribution switchboards, distribution panels and branch circuit panelboards.

More information

2016 Photovoltaic Solar System Plan Review List

2016 Photovoltaic Solar System Plan Review List Building Division 555 Santa Clara Street Vallejo CA 94590 707.648.4374 2016 Photovoltaic Solar System Plan Review List GENERAL PROJECT INFORMATION PLAN CHECK NO DATE JOB ADDRESS CITY ZIP REVIEWED BY PHONE

More information

Anchorage SS Center Model AN3050. Installation, Maintenance, and Operation Manual

Anchorage SS Center Model AN3050. Installation, Maintenance, and Operation Manual Anchorage SS Center Model AN3050 Installation, Maintenance, and Operation Manual Rev.3 06-29-16 Marina Electrical Equipment, Inc. 100 Warwick Court Williamsburg, VA 23185 Toll Free: 1-855-258-3939 Fax:

More information

Transfer Switch TS-50. Owner s Manual

Transfer Switch TS-50. Owner s Manual Transfer Switch TS-50 Owner s Manual Table of Contents Introduction 2 Installation 2 Operational Testing 7 Troubleshooting 7 Hi-Pot Testing 8 Generator Note 9 Medical Appliances 10 Caution 10 Disclaimer

More information

2012 Life Safety Code Requirements. James D. Scott Jeffrey Wortley

2012 Life Safety Code Requirements. James D. Scott Jeffrey Wortley 2012 Life Safety Code Requirements James D. Scott Jeffrey Wortley Regulation On May 4, 2016, CMS adopted the 2012 LSC and the 2012 HCFC by final rule. The final rule was published in the Federal Register

More information

ENGINEERING SPECIFICATION

ENGINEERING SPECIFICATION December 206 ENGINEERING SPECIFICATION No. of 6 DATE: 2-9-6 CATEGORY SUBJECT TABLE OF CONTENTS. Overview... 2 2. General Requirements for Service... 3 3. Definitions... 3 4. Abbreviations... 5 5. References

More information

The Evolution of Arc Fault Circuit Interruption

The Evolution of Arc Fault Circuit Interruption 51 st IEEE HOLM Conference on Electrical Contacts The Evolution of Arc Fault Circuit Interruption John A. Wafer Electrical Group Eaton Corporation 2003 Eaton Corporation. All rights reserved. 2 Residential

More information

Design Standards NEMA

Design Standards NEMA Design Standards Although several organizations are involved in establishing standards for the design, construction, and application of motor control centers, the primary standards are established by UL,

More information

ITS-50R TRANSFER SWITCH OWNER S MANUAL

ITS-50R TRANSFER SWITCH OWNER S MANUAL ITS-50R OWNER S MANUAL IOTA Engineering Transfer Switches provide automatic power switching between two or three separate 120/240 volt AC input sources, including powercords, onboard generators, onboard

More information

Corrections most seen on plan review October 18, 2017 David Rankin Seattle Department of Construction and Inspections

Corrections most seen on plan review October 18, 2017 David Rankin Seattle Department of Construction and Inspections Corrections most seen on plan review October 18, 2017 David Rankin Seattle Department of Construction and Inspections One-Line / Riser Diagrams Drawings are not reviewed prior to submission. Because of

More information

Understanding RV Electrical Hazards. Do you really want to plug in here?

Understanding RV Electrical Hazards. Do you really want to plug in here? Understanding RV Electrical Hazards Do you really want to plug in here? John Slaughter RV owner for 12 years, 150,000 miles of travel. Electrical Engineer Full disclosure, I am not a safety expert. Takeaways

More information

Bussmann series Quik-Spec Coordination Panelboard

Bussmann series Quik-Spec Coordination Panelboard Quik-Spec electrical gear 12 Contents Bussmann series Quik-Spec Coordination Panelboard Section page Description Quik-Spec Coordination Panelboards Up to 400 A 2 600 to 1200 A 3 Quik-Spec elevator disconnects

More information

ELECTRICAL INSTALLER EXAMINATION 18 November 2017 QUESTION AND ANSWER BOOKLET

ELECTRICAL INSTALLER EXAMINATION 18 November 2017 QUESTION AND ANSWER BOOKLET Candidate Code No. EIN15 For Board Use Only Result Date Int Result Date Int ELECTRICAL INSTALLER EXAMINATION 18 November 2017 QUESTION AND ANSWER BOOKLET INSTRUCTIONS READ CAREFULLY Time Allowed: Three

More information

Power InverterTM Watt. Continuous. User's Manual. WAGAN Corp. Limited Warranty Registration Form. Item no

Power InverterTM Watt. Continuous. User's Manual. WAGAN Corp. Limited Warranty Registration Form. Item no WAGAN Corp. Limited Warranty Registration Form All WAGAN Corporation products are warranted to the original purchaser of this product. Warranty Duration: This product is warranted to the original purchaser

More information

General Specifications for Lighthouse Power Pedestal

General Specifications for Lighthouse Power Pedestal Dimensions Height: 44.00 11 (1117.60 mm) Width: 13.67" (347.22 mm) Depth: 13.75" (349.25 mm) Lighthou98 Base Diagram ---13,75 1 --- Approximate Weight: 30 lbs. (13.6 kg) Multiply base dimensions by 25.4

More information

Increased requirements on external DC-breakers for transformerless PV inverters in Australia

Increased requirements on external DC-breakers for transformerless PV inverters in Australia Increased requirements on external DC-breakers for transformerless PV inverters in Australia Are the requirements also expedient for PV plants with transformerless SMA PV inverters? Content The standard

More information

Battery Power Inverters

Battery Power Inverters Battery Power Inverters Renogy 500W 1000W 2000W Pure Sine Wave Inverter Manual 2775 E. Philadelphia St., Ontario, CA 91761 1-800-330-8678 1 Version 1.4 Important Safety Instructions Please save these instructions.

More information

SINGLE PHASE WIRING SPECIFICATIONS

SINGLE PHASE WIRING SPECIFICATIONS SINGLE PHASE WIRING SPECIFICATIONS 1-866-MEC-ELEC (1-866-632-3532) Office Locations: Hondo Office 237 Hwy 173 N Hondo, TX 78661-0370 Fax 830.426.3335 Dilley Office 1718 W. FM 117 Dilley, TX 78017 Fax 830.965.1425

More information

90.2 Scope. The installation of electrical conductors, equipment and raceways for:

90.2 Scope. The installation of electrical conductors, equipment and raceways for: NEC Generator Primer Rules on the installation of generators and transfer switches 1 90.2 Scope The installation of electrical conductors, equipment and raceways for: public and private premises Conductors

More information

DIY 15 Amp Portable Generator Transfer Switch

DIY 15 Amp Portable Generator Transfer Switch DIY 15 Amp Portable Generator Transfer Switch ------------------------------------------------------------------------------------------------- WARNING You MUST read these warnings before proceeding. This

More information

Photovoltaic Solar Plan Review

Photovoltaic Solar Plan Review PAIGE B. VAUGHAN, CBO Director of Building and Safety Phone (310) 605-5509 Fax Line (310) 605-5598 E-mail:lbutler@comptoncity.org Building & Safety Department Photovoltaic Solar Plan Review Plan Check

More information

Have You Been Grounded Lately

Have You Been Grounded Lately Have You Been Grounded Lately Thurs, Oct. 26, 2017 9:15-10:15 Phil Simmons Author, Electrical Grounding and Bonding Author Cengage Learning Phil Simmons Have You Been Grounded Lately 2017 NEC Changes Relative

More information

THREE PHASE WIRING SPECIFICATIONS

THREE PHASE WIRING SPECIFICATIONS THREE PHASE WIRING SPECIFICATIONS 1-866-MEC-ELEC (1-866-632-3532) Office Locations: Hondo Office 237 Hwy 173 N Hondo, TX 78661-0370 Fax 830.426.3335 Dilley Office 1718 W. FM 117 Dilley, TX 78017 Fax 830.965.1425

More information

Owner's/Installation Manual

Owner's/Installation Manual Owner's/Installation Manual Power Management Module (PMM) and Starter Kit NOTE: The starter kit must be purchased and installed prior to individual PMM usage. Model Numbers: 00686-0 PMM 00699-0 PMM WITH

More information

Effective discrimination of protective devices

Effective discrimination of protective devices Effective discrimination of protective devices In the event of a fault occurring on an electrical installation only the protective device nearest to the fault should operate,leaving other healthy circuits

More information

Backfeed, Safety, and Work Practices

Backfeed, Safety, and Work Practices Backfeed, Safety, and Work Practices Michael T. Sheehan, P.E. March, 2015 NREL is a national laboratory of the U.S. Department of Energy, Office of Energy Efficiency and Renewable Energy, operated by the

More information

INDUSTRY WIDE LABOR-MANAGEMENT SAFETY COMMITTEE

INDUSTRY WIDE LABOR-MANAGEMENT SAFETY COMMITTEE INDUSTRY WIDE LABOR-MANAGEMENT SAFETY COMMITTEE SAFETY BULLETIN #23 GUIDELINES FOR WORKING WITH LIGHTING SYSTEMS AND OTHER ELECTRICAL EQUIPMENT All electrical systems and electrically energized equipment

More information

MEDICAL IRRIGATION K PUMP OPERATING MANUAL: Manufactured for: K.M.I 3185 Palisades Dr Corona CA USA Model No.

MEDICAL IRRIGATION K PUMP OPERATING MANUAL: Manufactured for: K.M.I 3185 Palisades Dr Corona CA USA Model No. OPERATING MANUAL: MEDICAL IRRIGATION K PUMP Manufactured for: K.M.I 3185 Palisades Dr Corona CA 92880-9432 USA 1-866-412-7867 Model No. 1000-0031 A-1299-5091 Edition 01 TABLE OF CONTENTS Title SAFETY PRECAUTIONS

More information

Beginner's Guide to Marinas, RV Parks, and NEC Requirements for Both

Beginner's Guide to Marinas, RV Parks, and NEC Requirements for Both Eaton's Beginner's Guide to Marinas, RV Parks, and NEC Requirements for Both Beginner's Guide to Marinas, RV Parks, and NEC Requirements for Both Introduction This guide is intended to give you quick,

More information

Preventing Electrocutions and Injury with Industrial GFCIs. White Paper

Preventing Electrocutions and Injury with Industrial GFCIs. White Paper Preventing Electrocutions and Injury with Industrial GFCIs White Paper Introduction Hundreds of people in the United States are killed by electricity every year, and several times that number are shocked

More information

Emergency Power Systems 101 An Overview of Emergency Power Systems. Craig Ayers, MHA, CHFM, CHSP, CPO

Emergency Power Systems 101 An Overview of Emergency Power Systems. Craig Ayers, MHA, CHFM, CHSP, CPO Emergency Power Systems 101 An Overview of Emergency Power Systems Craig Ayers, MHA, CHFM, CHSP, CPO Emergency Power Systems Purpose Generators Distribution Transfer switches Maintenance and testing Contingency

More information

SAM Series Power Inverter

SAM Series Power Inverter SAM Series Power Inverter OWNER S MANUAL For model: (12 V DC INPUT, 115V AC OUTPUT) SAM-100-12 THIS MANUAL CONTAINS IMPORTANT INFORMATION REGARDING SAFETY, OPERATION, MAINTENANCE AND STORAGE OF THIS PRODUCT.

More information

SmartPro Medical-Grade UPS, Line Interactive, Lithium Battery, 6 Outlets - 230V, 1kVA, 750W, Full Isolation

SmartPro Medical-Grade UPS, Line Interactive, Lithium Battery, 6 Outlets - 230V, 1kVA, 750W, Full Isolation SmartPro Medical-Grade UPS, Line Interactive, Lithium Battery, 6 Outlets - 230V, 1kVA, 750W, Full Isolation MODEL NUMBER: SMX1200XLHGL Highlights Lithium battery backup allows time to save files and safely

More information

Unrestricted. S.Mukhopadhyay

Unrestricted. S.Mukhopadhyay Unrestricted S.Mukhopadhyay 10.09.15 Electrical safety A report says that every year 5500 Indians die due to electric shock and 800 due to electric fire. 12% work place death due to electrocution ELECTRICAL

More information

DISCONNECT SWITCH REQUIREMENTS FOR DISTRIBUTED GENERATION CUSTOMERS

DISCONNECT SWITCH REQUIREMENTS FOR DISTRIBUTED GENERATION CUSTOMERS Prepared by: SXZO DISCONNECT SWITCH REQUIREENTS FOR DISTRIBUTED GENERATION CUSTOERS 060559 Department: Electric Distribution Section: Design and Construction Approved by: D.Jantz (DWJ7) Date: 8/15/17 Rev.

More information

WF-5110R True Sine Wave Inverter

WF-5110R True Sine Wave Inverter Operator s Manual WF-5110R True Sine Wave Inverter WF-9900 Series WF-5110R ( The Inverter model number is located on the label on top of the enclosure) Distributed in the U.S.A. and Canada by ARTERRA DISTRIBUTION

More information

Power and Cooling. Chassis Power System

Power and Cooling. Chassis Power System This chapter describes the Cisco NCS 6000 Series Routers power and cooling systems. It also provides the power, grounding, and cooling requirements for the installation site to help you plan the site facilities

More information

ELECTRICAL SYSTEMS AND SAFETY

ELECTRICAL SYSTEMS AND SAFETY ACCREDITATION STANDA RDS ELECTRICAL SYSTEMS AND SAFETY Definitions emergency electrical power supply generator medical electrical equipment normal electrical power supply The local generation of electrical

More information

OPERATION MANUAL. Automatic Endoscope Leak Tester. Model# ZUTR-10003

OPERATION MANUAL. Automatic Endoscope Leak Tester. Model# ZUTR-10003 Automatic Endoscope Leak Tester OPERATION MANUAL ZUT-Lab-004-10003 REV. F Table of Contents Page 2 Introduction 1. Intended Use 2. Labels, Symbols and Signal Words 3. Environmental Conditions 4. Supply

More information

ELECTRICIAN S THEORY EXAMINATION 19 June 2010 QUESTION AND ANSWER BOOKLET

ELECTRICIAN S THEORY EXAMINATION 19 June 2010 QUESTION AND ANSWER BOOKLET Candidate Code No. ET34 For Board Use Only Result Date Int Result Date Int ELECTRICIAN S THEORY EXAMINATION 19 June 2010 QUESTION AND ANSWER BOOKLET INSTRUCTIONS READ CAREFULLY Time Allowed: Three hours

More information