APC SERIES MODELS AND SURGI-CHAIRS ELECTRIC POWERED MODELS: EPC, EPD, ESC, & ESD OPERATING MANUAL

Size: px
Start display at page:

Download "APC SERIES MODELS AND SURGI-CHAIRS ELECTRIC POWERED MODELS: EPC, EPD, ESC, & ESD OPERATING MANUAL"

Transcription

1 APC SERIES MODELS AND SURGI-CHAIRS ELECTRIC POWERED MODELS: EPC, EPD, ESC, & ESD OPERATING MANUAL READ THIS MANUAL BEFORE OPERATING YOUR APC / SURGI-CHAIR. SAVE THIS MANUAL FOR FUTURE USE. THE MOST CURRENT VERSION OF THIS MANUAL CAN BE FOUND ONLINE AT APC-US-INS-LAB-REVB14

2 CONTENTS INTRODUCTION A WORD FROM GF HEALTH PRODUCTS, INC... 4 INDICATIONS FOR USE... 4 SERVICE INFORMATION... 4 ADVISORY LIST OF WARNINGS AND CAUTIONS... 5 SIGNIFICANCE OF SAFETY STATEMENTS... 5 DANGER / WARNING / CAUTION / NOTICE SUMMARY... 5 WARNING: TO REDUCE THE RISK OF BURNS, FIRE, ELECTRIC SHOCK, OR PERSONAL INJURY... 5 WARNING CAUTIONS AND PROPER OPERATION UNCRATING INSTRUCTIONS... 8 IMPORTANT REPORT ANY SHIPPING DAMAGE IMMEDIATELY:... 8 NOTICE POSSIBLE EQUIPMENT DAMAGE:... 8 WARNING PERSONAL INJURY HAZARD:... 8 ENVIRONMENTAL CONDITIONS... 8 OPERATING... 8 STORAGE AND TRANSPORT... 8 UNPACKING INSTRUCTIONS OPERATING INSTRUCTIONS APC / SURGI-CHAIR SPECIFICATIONS... 9 ELECTRICAL SPECIFICATIONS (APPLIES ONLY TO EPC AND ESC MODELS)... 9 ELECTRICAL SPECIFICATIONS (APPLIES ONLY TO EPD AND ESD MODELS) FEATURES, WARNINGS AND PROPER OPERATION OPERATING INSTRUCTIONS WARNINGS CAUTIONS AND PROPER OPERATION (SEE DIAGRAM ON FOLLOWING PAGE) FEATURES (SHOWN IN ILLUSTRATION)...11 WARNINGS CAUTIONS AND PROPER OPERATION (SEE LIST ON PREVIOUS PAGE) BRAKING AND STEERING OPERATION APPLYING THE BRAKES RELEASING THE BRAKES APPLYING THE STEERING BRAKES RELEASING THE STEERING BRAKES ELECTRIC CONTROL LOCATIONS PENDANT CONTROL LOCATION PLUG LOCATION LOW BATTERY LIGHT (LED) TOP ADJUSTMENTS HEIGHT ADJUSTMENT RAISING THE HEIGHT LOWERING THE HEIGHT FOOT SECTION ADJUSTMENT RAISING THE FOOT SECTION LOWERING THE FOOT SECTION BACK SECTION ADJUSTMENT RAISING THE BACK SECTION LOWERING THE BACK SECTION ELECTRICAL CONTROL LOCATIONS QUICK DROP ACTIVATION ADJUSTABLE FOOTREST REPOSITIONING THE FOOTREST HINGING RAILS REPOSITIONING THE RAIL ESCEYE / ESDEYE HEADREST PRE-OP / POST-OP HEAD EXTENSION ADJUSTING THE HEADREST COMMON OPTIONAL ACCESSORIES MOUNTING THE WRIST REST APC-US-INS-LAB-RevB14 2

3 3.9.2 ADJUSTING THE WRIST REST INSTALLING #18 IV ROD INSTALLING THE SIDE TRAY USING RESTRAINT STRAPS INSTALLING THE PATIENT TRAY TROUBLESHOOTING GUIDE ELECTRIC POWERED CHAIRS BATTERY REPLACEMENT INSTRUCTIONS PREVENTIVE MAINTENANCE FOR THE USER OPTIONAL ACCESSORIES GF HEALTH PRODUCTS, INC. LIMITED WARRANTY FOR HAUSTED BRAND PRODUCT LINE WITHIN THE U.S DISPOSAL AND KEY TO SYMBOLS DISPOSAL KEY TO SYMBOLS APPENDIX GUIDANCE AND MANUFACTURER S DECLARATION ELECTROMAGNETIC EMISSIONS GUIDANCE AND MANUFACTURER S DECLARATION GUIDANCE AND MANUFACTURER S DECLARATION ELECTROMAGNETIC IMMUNITY RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT AND HAUSTED APC SERIES MODELS AND SURGI-CHAIRS INDEX GF Health Products, Inc. is not responsible for typographical errors. All illustrations, specifications, packaging and warranties contained in this document are based on the latest product information available at the time of printing. The most current product information, including the most current revision of this manual, can be found online at Graham-Field and Hausted are registered trademarks of GF Health Products, Inc. COPYING PROHIBITED This manual is protected by Federal Copyright Law, which provides for damages of up to USD $20,000, as well as criminal fines and imprisonment, for unauthorized copying GF Health Products, Inc. All Rights Reserved Printed in U.S.A. APC-US-INS-LAB-RevB14 3

4 INTRODUCTION A WORD FROM GF HEALTH PRODUCTS, INC. This manual contains important information on proper use and maintenance of the Hausted APC Series models and Surgi-Chairs. All personnel involved in the use and maintenance of this equipment must carefully review and comply with the warnings, cautions and instructions contained in this manual. These instructions are important to protect the health and safety of personnel operating models EPC, EPD, ESC, and ESC and should be retained in a conveniently accessible area for quick reference. Complete instructions for uncrating and putting your new equipment in service, as well as equipment drawings, have been furnished. If missing, contact GF Health Products, Inc. for replacement copies, giving the serial number and model number of the unit. GF Health Products, Inc. carries a complete line of accessories for use with these chairs; your representative will gladly review these with you. Indications for Use The Hausted APC Series models and Surgi-Chairs are intended for use in patient treatment, transport or recovery. The chair s backrest can be positioned from sitting to supine. Height positioning, as well as backrest and leg section adjustment, is electric/battery powered and is activated with a handheld control. Four easy-rolling, steerable casters allow maximum mobility and maneuverability. Service Information A thorough preventive maintenance program is essential to safe and proper unit operation. This manual contains maintenance schedules and procedures which should be followed for satisfactory equipment performance. We encourage you to contact GF Health Products, Inc. with maintenance concerns. Advisory A listing of the safety precautions to be observed when operating and servicing this equipment can be found in Section 1 of this manual. Do not operate or service the equipment until you have become familiar with this information. Any alteration of this equipment not authorized or performed by GF Health Products, Inc., could affect its operation, will void the warranty, could violate national, state, and local regulations, and could jeopardize your insurance coverage. Manufactured by: GF Health Products, Inc. 336 Trowbridge Drive Fond du Lac, WI Main Fax Class 1 Equipment Type B Equipment Equipment not suitable for use in the presence of flammable anesthetic mixture with air or oxygen or nitrous oxide. IPX4 (Splash-proof equipment) Not suitable for continuous operation (Duty Cycle: 5% 1 Min in 19 Min) Info: The base language of this document is ENGLISH. Any translations must be made from the base language document. APC-US-INS-LAB-RevB14 4

5 1 LIST OF WARNINGS AND CAUTIONS IMPORTANT: Before using the Hausted APC / Surgi-Chair, please read and adhere to the following safety precautions and warnings. Failure to do so could result in serious personal injury or damage to the APC / Surgi-Chair. Always consult your healthcare professional to determine safe methods most suitable for your individual abilities. Protect yourself, your attendant, and the Hausted APC / Surgi-Chair by having it serviced regularly. If you experience any malfunction, contact your Graham-Field authorized distributor immediately, as a hazardous condition could result, causing personal injury or damage to the Hausted APC / Surgi-Chair. Periodic inspection, adjustment and replacement of worn parts are necessary to provide years of excellent service. Maintenance MUST be performed by qualified personnel ONLY. SAVE THESE INSTRUCTIONS. SIGNIFICANCE OF SAFETY STATEMENTS Please note the following special statements, used throughout this manual, and their significance: DANGER: Indicates a potential hazard situation or unsafe practice that, if not avoided, will result in death or serious personal injury. WARNING: Indicates a potential hazard situation or unsafe practice that, if not avoided, could result in death or serious personal injury. CAUTION: Indicates a potential hazard situation or unsafe practice that, if not avoided, could result in minor or moderate personal injury. s NOTICE: Indicates a potential hazard situation or unsafe practice that, if not avoided, could result in product or property damage. Info: Provides application recommendations or other useful information to ensure that you get the most from your product. DANGER / WARNING / CAUTION / NOTICE Summary The following is a listing of the safety precautions which must be observed when operating and servicing this equipment. These precautions are repeated (in whole or in part), where applicable, throughout the manual. WARNING: To Reduce the Risk of Burns, Fire, Electric Shock, or Personal Injury DANGER: SHOCK HAZARD To reduce the risk of electric shock, do not remove the cover. Unit is to be serviced by qualified personnel only. DANGER: SHOCK HAZARD To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth. WARNING: LACERATION HAZARD When cutting bands always us a tool specifically designed for that purpose. This will help to avoid personal injuries frequently incurred when bands are cut and tension released. WARNING: PERSONAL INJURY HAZARD The batteries are wired in series; failure to connect the same way can cause batteries to explode. APC-US-INS-LAB-RevB14 5

6 WARNING CAUTIONS AND PROPER OPERATION WARNING: The chair has a warning label located on the back section stating: Maximum patient weight 227 kilograms (500 lbs.) WARNING: The chair is not intended to replace a stretcher or gurney. WARNING: The chair has warning labels on both the head and foot end stating: Do not sit on end as tipping may occur. WARNING: Do not stand on footrest tipping may occur. WARNING: When not in use, do not leave the chair in a recline position. WARNING: When patient is seated in the chair, ensure the side rails are up and the patient is secured with patient restraint straps. WARNING: Patient entry, egress and transfer from the chair should always be from the center side rail location with the side rail in the down position and the brakes locked. WARNING: At no time should the patient be permitted to enter or exit from the ends of the chair when in partial or total recline position. WARNING: Ensure the brakes are locked when the patient is not being transported. WARNING: The patient pendant has a warning label on it stating: Place pendant in holder when not in use keep cord clear of moving parts. WARNING: All electric powered chairs are equipped with a built in battery back-up system: nevertheless, the unit should remain plugged into wall receptacle during normal use. The battery back-up is recommended for transport and emergency only. WARNING: The backrest quick drop handle is intended to be used during emergency situations only WARNING: To turn electric controls on, plug into wall receptacle. To turn off, remove plug from wall receptacle. The electric powered chairs do not have a separate on/off switch. WARNING: The chair has a warning label located on the control box cover stating: To reduce the risk of electrical shock do not remove the cover. Service by qualified personnel only. WARNING: Always disconnect the power source whenever troubleshooting or servicing any electric powered chair. WARNING: Do not use a sharp instrument to remove fuse, as it may scratch the circuit board. WARNING: The batteries are wired in series; failure to install or rewire the same way may cause the batteries to explode. WARNING: Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this manual. WARNING: Portable and mobile RF communications equipment can affect Medical Electrical Equipment. WARNING: The use of accessories, transducers, and cables other than those specified by the manufacturer may result in increased Emissions or decreased Immunity of the Hausted Equipment. APC-US-INS-LAB-RevB14 6

7 WARNING: The Hausted Equipment should not be used adjacent to or stacked with other equipment; if adjacent or stacked use is necessary, the Hausted equipment should be observed to verify normal operation in the configuration in which it will be used. WARNING: Steam cleaning and pressure washing of chair is not recommended and can void warranty. WARNING: Rail release handle will not release unless force is applied laterally to outside top of rail. WARNING: The chair has a warning label located on both side rails indicating a pinch point between seat section and side rail. WARNING: NOTICE FOR CALIFORNIA CUSTOMERS- California Proposition 65 Warning: This product contains a chemical known to the State of California to cause cancer and reproductive or developmental harm. APC-US-INS-LAB-RevB14 7

8 2 UNCRATING INSTRUCTIONS IMPORTANT REPORT ANY SHIPPING DAMAGE IMMEDIATELY: WARNING: Inform shipper of any damages leave carton intact. Leave equipment in the receiving area until inspection is complete. NOTICE POSSIBLE EQUIPMENT DAMAGE: s NOTICE: The crate contains fragile, expensive medical equipment. Uncrate and handle carefully. If after uncrating the equipment you find any damage (no matter how slight), report the damage to GF Health Products, Inc. WARNING PERSONAL INJURY HAZARD: WARNING: When cutting bands, always use tool specifically designed for that purpose. This will help avoid personal injuries possibly incurred when bands are cut and tension is released. ENVIRONMENTAL CONDITIONS Operating Temperature 5 C to 40 C Relative Humidity 20% to 30 C Atmospheric Pressure 700 to 1060 hpa Storage and Transport Temperature -10 C to 50 C Relative Humidity 20% to 30 C Atmospheric Pressure 700 to 1060 hpa IMPORTANT: Follow each step in the order shown in these instructions. UNPACKING INSTRUCTIONS Your Hausted equipment has been carefully packed at our manufacturing plant to ensure safe shipment to your medical facility. There are several procedures you must follow to put your new equipment in service. These procedures only take a few minutes to complete and all are required to ensure proper operation of the equipment. 1. Cut the two bands around the shipping carton. 2. Remove the top half of the carton and cut one side of the bottom half. 3. Remove the equipment from the carton. 4. Check to see if all features of the equipment work properly. If all the features work, advance to step 5. If any of the features do not work properly, call GF Health Products, Inc. at Info: Although the equipment has been fully charged prior to shipment, plug the unit into a wall socket prior to checking any electric features. 5. Clean the equipment using mild detergent to remove any dirt accumulated during shipment, and place the equipment into service. APC-US-INS-LAB-RevB14 8

9 3 OPERATING INSTRUCTIONS 3.1 APC / SURGI-CHAIR SPECIFICATIONS Info: All dimensions are specified in inches. Unless otherwise noted, all dimensions are ±.375. Stacked dimensions are minimum (left) and maximum (right). GF Health Products, Inc. reserves the right to change specifications without notice. B DIMENSION MODEL A B C D E EPC250/ EPD250 ESC250/ ESD / / C 6 TRENDELENBERG D (TO BASE COVER) E A ±1.00 / 825 / 32.5 ± Electrical Specifications (Applies only to EPC and ESC Models) Product Classification: 1 Input Voltage: 120V~, ±5%, 60Hz Amperage: Max. 3.5A Duty Cycle: 5% 1 Min in 19 Min IP Rating: IPX 4 Grounding Protection: Type B Electrical Specifications (Applies only to EPD and ESD Models) Product Classification: 1 Input Voltage: 230V~, ±10%, 50Hz Amperage: Max. 1.9A Duty Cycle: 5% 1 Min in 19 Min IP Rating: IPX 4 Grounding Protection: Type B APC-US-INS-LAB-RevB14 9

10 3.2 FEATURES, WARNINGS AND PROPER OPERATION OPERATING INSTRUCTIONS WARNINGS CAUTIONS AND PROPER OPERATION (See Diagram on following page) A. WARNING: The chair has a warning label located on the back section stating: Maximum patient weight 227 kilograms (500 lbs.) B. WARNING: The chair is not intended to replace a stretcher or gurney. C. WARNING: The chair has warning labels on both the head and foot end stating: Do not sit on end - as tipping may occur. D. WARNING: Do not stand on footrest tipping may occur. E. WARNING: When not in use, do not leave the chair in a recline position. F. WARNING: When patient is seated in the chair, ensure the side rails are up and the patient is secured with patient restraint straps. G. WARNING: Patient entry, egress and transfer from the chair should always be from the center side rail location with the side rail in the down position and the brakes locked. H. WARNING: At no time should the patient be permitted to enter or exit from the ends of the chair when in partial or total recline position. I. WARNING: Ensure the brakes are locked when the patient is not being transported. J. WARNING: The patient pendant has a warning label on it stating: Place pendant in holder when not in use keep cord clear of moving parts. K. WARNING: All electric-powered chairs are equipped with a built in battery back-up system, but it is recommended that the unit remain plugged in wall receptacle during normal use. The battery back-up is recommended for transport and emergency only. L. WARNING: The backrest quick drop handle is intended to be used during emergency situations only. M. WARNING: To turn electric controls on, plug into wall receptacle. To turn off, remove plug from wall receptacle. The electric powered chairs do not have a separate on/off switch. N. WARNING: The chair has a warning label located on the control box cover stating: To reduce the risk of electrical shock do not remove the cover. Service by qualified personnel only. O. WARNING: Always disconnect the power source whenever servicing any electric powered chair. P. WARNING: Do not use a sharp instrument to remove fuse, as it may scratch the circuit board. Q. WARNING: The batteries are wired in series; failure to install or rewire the same way may cause the batteries to explode. R. WARNING: Rail release handle will not release unless force is applied laterally to outside top of rail. S. WARNING: The chair has a warning label located on both side rails indicating a pinch point between seat section and side rail. APC-US-INS-LAB-RevB14 10

11 Features (Shown in Illustration) PATIENT COMFORT CUSHIONS HEAD SECTION TUBE WARNING LABEL See Warnings C & H BACK OF CHAIR WARNING LABEL See Warning A!! PATIENT ENTRY See Warnings G & H! HINGING RAIL See Warning F EMERGENCY QUICK DROP HANDLE See Warning L PATIENT PENDANT WARNING LABEL See Warning K! RESTRAINT ROD See Warning F! ADJUSTABLE FOOT REST WARNING LABEL See Warning D RAIL LATCH See Warnings F, R, & S POWER CORD See Warning F BRAKE PEDAL See Warnings G thru I! FOOT SECTION TUBE WARNING LABEL See Warnings C & H ELECTRIC CONTROL BOX & CIRCUITRY See Warnings M thru Q! WARNINGS CAUTIONS AND PROPER OPERATION (See List on Previous Page) APC-US-INS-LAB-RevB14 11

12 3.3 BRAKING AND STEERING OPERATION Applying the Brakes Activate the four wheel central braking system by depressing the red pedal at any of the four corners of the unit (Figure 3-1). Engage the brakes fully by depressing the pedal to approximately 45. All four caster wheels should then be locked from swiveling and rotating Releasing the Brakes Release the brakes by depressing either green pedal at the head end of the unit until the pedal is in a horizontal position (Figure 3-2). All four wheels should then rotate and swivel freely. WARNING: Ensure the brakes are locked when the patient is not being transported Applying the Steering Brakes From the head end of the chair, depress the green pedal downward into locked position (Figure 3-3). Push the chair forward. Both casters at the foot end will lock into nonswivel mode; the steering lock will then guide the chair along a straight path with minimal steering effort by the attendant. FIGURE 3-1 FIGURE 3-2 FIGURE Releasing the Steering Brakes Release the steering brakes by depressing either red pedal at the head end of the unit until the pedal is in a horizontal position (Figure 3-2). All four casters should then rotate and swivel freely. APC-US-INS-LAB-RevB14 12

13 3.4 ELECTRIC CONTROL LOCATIONS Pendant Control Location The pendant is located in the pendant holder on the back of the chair (Figure 3-4). This chair is equipped with a battery back up for transport but the unit should be plugged into a wall receptacle when not in transport. The plug is located on the back of the base cover (Figure 3-5). Do not position the unit so that it is difficult to disconnect the plug. s NOTICE: Place pendant in holder when not in use. Keep cord clear of moving parts Plug Location This chair is equipped with a battery back up for transport but the unit should be plugged into a wall receptacle when not in transport. The plug is located on back of the base cover (Figure 3-5). Do not position the unit so that it is difficult to disconnect the plug. Important: It is recommended that unit remain plugged in wall receptacle during normal use. Battery back-up is recommended for transport and emergency only. FIGURE 3-4 FIGURE 3-5 FIGURE Low Battery Light (LED) This chair is equipped with a low battery indicator light, located on top of the base cover of the patient s left (Figure 3-6). When a button on the pendant is pressed, and the LED light illuminates, it indicates that the backup batteries require charging; plug the unit into the wall receptacle. APC-US-INS-LAB-RevB14 13

14 3.5 TOP ADJUSTMENTS Height Adjustment Remove the pendant from pendant holder, (see Section 3.4.1, Pendant Control ). Raising the Height Press the first button on the third row of buttons on the pendant (Figure 3-7); hold until the desired height is achieved. Lowering the Height Press the second button on the third row of buttons on the pendant (Figure 3-8); hold until the desired height is achieved. s NOTICE: Place pendant in holder when not in use. Keep cord clear of moving parts Foot Section Adjustment Raising the Foot Section Press the first button on the second row of buttons on the pendant (Figure 3-9); hold until the desired angle is achieved. Lowering the Foot Section Press the second button on the second row of buttons on the pendant (Figure 3-10); hold until the desired angle is achieved Back Section Adjustment Raising the Back Section Press the first button on the first row of buttons on the pendant (Figure 3-11); hold until the desired incline is achieved. Lowering the Back Section Press the second button on the first row of buttons on the pendant (Figure 3-12); hold until the desired recline is achieved. FIGURE 3-7 FIGURE 3-8 FIGURE 3-9 FIGURE 3-10 FIGURE 3-11 FIGURE 3-12 APC-US-INS-LAB-RevB14 14

15 3.5 ELECTRICAL CONTROL LOCATIONS Quick Drop Activation This chair is equipped with a manual override function for the back section of the chair. This option should only be used in an emergency situation. To activate the quick drop, support the back section and pull upward on the red activating handle located on the back of the chair (Figure 3-13). FIGURE 3-13 APC-US-INS-LAB-RevB14 15

16 3.6 ADJUSTABLE FOOTREST Repositioning the Footrest The footrest has three positions: retracted, lower, and upper. With the pan in the retracted position, pull out on the handles on either side of pan (Figure 3-14). The footrest will drop into the lower position. To place the footrest into the upper position, grasp the handles on both sides of the pan. Tilt the pan up while lifting (Figure 3-14). Once the pan is in the fully up position, tilt the pan out until it locks into the upper position (Figure 3-15). To return the footrest to the retracted position, tilt the pan up while letting it slide down (Figure 3-14) into the lower standard position. Continue tilting the footrest until it is in the retracted position. WARNING: TIPPING HAZARD Do not stand on footrest tipping may occur. FIGURE 3-14 FIGURE 3-15 APC-US-INS-LAB-RevB14 16

17 3.7 HINGING RAILS Repositioning the Rail The chair rail has two positions, up and down. Both positions lock the rail into place. To lower the rail, apply a lateral force to the top of the rail (Figure 3-16) and pull on the red rail handle (Figure 3-17). Allow the rail to hinge all the way down until it locks into place in the rail catch (Figure 3-18). To raise the rail, take hold of the rail and lift it until the latching mechanism locks into place. (Figure 3-16). WARNING: When patient is seated in the chair, ensure the side rails are up and the patient is secured with restraint straps. WARNING: Rail release handle will not release unless force is applied laterally to outside top of rail. WARNING: The chair has a warning label located on both side rails indicating a pinch point between seat section and side rail. FIGURE 3-16 FIGURE 3-17 FIGURE 3-18 APC-US-INS-LAB-RevB14 17

18 3.8 ESCEYE / ESDEYE HEADREST Pre-op / Post-op Head Extension To remove the head extension, rotate the knobs on the two ends of the extension counterclockwise (Figure 3-19). With knobs loose, pull out on the extension. To install the extension, slide the extension over the surgical mounting bars using the guides on the extension. Slide the extension fully down the surgical side bars, then tighten the knobs on the end of the extension. s NOTICE: Ensure the head extension is secure before applying any pressure Adjusting the Headrest Remove head extension (see Pre-op / Postop Head extension, Section 3.8.1). Grasp the right ball style knob (Figure 3-20), and rotate the knob counterclockwise to articulate the head section upward (Figure 3-21). Rotate the knob clockwise to articulate the head section downward (Figure 3-22). Once the upward articulation has been set, grasp the left ball style knob (Figure 3-23), and rotate the knob counterclockwise to adjust the height of the headrest (Figure 3-24). Info: After understanding which knob creates which action, quick and smooth adjustment can be achieved by rotating the knobs simultaneously (Figure 3-25). FIGURE 3-19 FIGURE 3-20 FIGURE 3-21 FIGURE 3-22 FIGURE 3-23 FIGURE 3-24 FIGURE 3-25 APC-US-INS-LAB-RevB14 18

19 3.9 COMMON OPTIONAL ACCESSORIES Mounting the Wrist Rest Place the wrist rest into one of the three square sockets under the headrest (Figure 3-26). Rotate the T knob on the back of the wrist rest (Figure 3-27) clockwise. s NOTICE: Ensure the wrist rest is secure before applying any pressure Adjusting the Wrist Rest Properly install the wrist rest (see the eye headrest instructions). Support the wrist rest and loosen the black star knob on the side of the support post (Figure 3-28). Position the wrist rest to desired height and rotation. Tighten the star knob located on the side of the support post (Figure 3-28). s NOTICE: Ensure the wrist rest is secure before applying any pressure Installing #18 IV Rod Remove the IV Rod from the clips located on the base (Figure 3-29). Insert IV Rod into preferred socket (Figure 3-30). Place IV Rod back into storage clips when not in use (Figure 3-29) Installing the Side Tray Insert pin end of Side Tray assembly into preferred socket on the seat section. Align hole in pin with hole in seat section socket and insert locking pin to prevent movement of tray (Figure 3-31). FIGURE 3-26 FIGURE 3-27 FIGURE 3-28 FIGURE 3-29 FIGURE 3-30 FIGURE 3-31 APC-US-INS-LAB-RevB14 19

20 3.9.5 Using Restraint Straps Locate both ends of the restraint strap, on either side of the chair. One will have a square loop on the end, and the other will have hook and loop fasteners (Figure 3-32). Slide the end of the strap with the hook and loop fasteners through the loop on the other strap (Figure 3-33). Pull out on the strap until the patient is secure. Fold the hook and loop fastener end of the strap back until it makes contact with the mating hook and loop fastener on the strap (Figure 3-34) Installing the Patient Tray Using the push buttons underneath the tray, lock the legs 90 to the tray prior to installing (Figure 3-35). Check to make sure that the legs are located in the same hole (Figure 3-36). Insert the first leg into preferred socket on one side of the patient, and insert second leg into corresponding socket on other side (Figure 3-47). Adjust as necessary to best fit patient. FIGURE 3-32 FIGURE 3-33 FIGURE 3-34 FIGURE 3-35 FIGURE 3-36 FIGURE 3-37 APC-US-INS-LAB-RevB14 20

21 4 TROUBLESHOOTING GUIDE 4.1 ELECTRIC POWERED CHAIRS DANGER: SHOCK HAZARD To reduce the risk of electric shock, do not remove the cover. Unit is to be serviced by qualified service personnel (minimum 1 year medical equipment service and repair experience) only. DANGER: SHOCK HAZARD Always disconnect the power source whenever troubleshooting or servicing any electric powered chair. If One motor or one column does not move but all others are operating correctly. Nothing moves. The unit runs when plugged into wall receptacle, but does not run on battery backup. Then Step 1: Check all motor and column plug connections at the controller box. Step 2: If a column does not move: Check the connection at the column. Step 3: Plug a connector from the faulty component into a different socket: If the component does not run: Replace that component. Info: Refer to the appropriate Illustrated Part breakdown to replace components as follows Foot Actuator (Figure 7.3), Back Actuator (Figure 7.5), Column (Figure 7.1 and 7.2). If the component does run: Test pendant by plugging a functioning component into the non-functioning socket on the controller. If this component does not run, replace the pendant (Figure. 7.8) If replacing the pendant does not fix the problem, then the replace the controller (Figure 7.1). Step 1: Plug unit into a mains supply wall receptacle, observe the pilot light on the controller: If the pilot light is off: Replace the controller (Figure 7.1). If the pilot light is on: Check the pendant control plug connection at the controller. Step 1: Plug unit into a wall receptacle overnight. If the batteries do not hold a charge, replace the batteries (section 4.2). GF Health Products, Inc. may be contacted at for additional information required to service or repair the equipment. 4.2 BATTERY REPLACEMENT INSTRUCTIONS DANGER: SHOCK HAZARD To reduce the risk of electric shock, do not remove the cover. Unit is to be serviced by qualified service personnel (minimum 1 year medical equipment service and repair experience) only. DANGER: SHOCK HAZARD Always disconnect the power source whenever troubleshooting or servicing any electric powered chair. WARNING: PERSONAL INJURY HAZARD The batteries are wired in series; failure to connect the same way can cause batteries to explode. The batteries are the only field serviceable components do not attempt to repair the circuit boards. Battery replacement Step 1: Remove the controller. Step 2: Remove the 4 screws from the right half, located on top of controller. Step 3: (After 4/11/00) replace (2) batteries with YUSHA #NP , 12 volt, 1.2 Ah batteries (Hausted P/N required, for CB-12 controllers). Step 4: Ensure battery connector in place on left side of batteries. Step 5: Replace cover. APC-US-INS-LAB-RevB14 21

22 5 PREVENTIVE MAINTENANCE FOR THE USER Component Cleaning Procedure Schedule Cleaning Agent * Special Notes Pads/ Mattresses Chair Electrical components Mechanical chair components Mechanical accessories Wipe with damp cloth to remove any foreign material Wipe with damp cloth to remove any foreign material Wipe external surfaces ONLY with damp cloth to remove any foreign material Wipe with damp cloth to remove any foreign material Wipe with damp cloth to remove any foreign material After each use After each use After each use After each use After each use Routine hospital grade disinfectants, soap and water Routine hospital grade disinfectants, soap and water Routine hospital grade disinfectants, soap and water Routine hospital grade disinfectants, soap and water Routine hospital grade disinfectants, soap and water Use only medium strength cleaners Do not steam clean or pressure wash Lubricate pivot points after cleaning Use only medium strength cleaners Lubricate pivot points after cleaning Lubricate pivot points after cleaning Procedure Schedule Material Lubricate all moving and sliding parts and hinge points Every 3 months Lubricating oil, light-duty grease, wax stick lubricant or Never-Seez lubricant NEVER LUBRICATE GAS SPRING, MOTOR OR MECH-LOCK SHAFTS Inspect all fasteners to ensure proper fit, position and tightness (including nuts, bolts, etc.) Inspect all surfaces and remove any sharp or burred areas; apply touch-up paint where required Every 3 months Every 3 months Proper size wrench and screwdriver Metal file, proper color paint (specify color when ordering) * Disinfectants, vinyl products: Phenolic disinfectants are the best choice for vinyl products; properly diluted quaternaries are also acceptable. Quaternary/Isopropyl disinfectants are not recommended for vinyl. Disinfectants, urethane products: Quaternary disinfectants are recommended for urethane products. Quaternary/ Isopropyl disinfectants are recommended for urethane. Phenolics should be avoided on urethane. Disinfectants, all products: All fabrics may be cleaned with a 1:10 dilution of household bleaches containing 5.25% sodium hypochlorite as recommended by the Centers for Disease Control in Atlanta, Georgia. There is no harmful effect on the fabric. Disinfectants applied at full concentration or in highly concentrated solutions will decrease the useful life of fabric. Iodophor type disinfectants used on fabrics may result in staining. Soils or Stains: Use neutral soapsuds and lukewarm water. Do not use harsh cleansers, solvents or detergents. Hard-To -Clean Spots: Use standard household/vinyl cleansers and a soft bristle brush on troublesome spots or stains. Presoak heavy, dried-on soil. Laundering Vinyl-laminated or Rubber-coated fabric is not recommended. Laundering may substantially decrease the useful life of the fabric. Laundering Polyurethane-coated fabric: Machine wash with detergent up to 120 F. Some surface wrinkling may occur; this wrinkling has no adverse effect on the fabric properties. Hang or tumble dry thoroughly at 140 F before storage. s NOTICE POSSIBLE EQUIPMENT DAMAGE HAZARD: Steam cleaning and pressure washing of chair is not recommended and can void warranty. Info: For more detailed information, please contact GF Health Products, Inc. at Info: In addition to the User Preventive Maintenance, a more detailed Preventive Maintenance Program is also required to keep the equipment in good working order. This Preventive Maintenance Program is found on our website, at APC-US-INS-LAB-RevB14 22

23 6 OPTIONAL ACCESSORIES Universal Accessories APC Restraint straps with hook and loop fasteners Telescoping IV Rod, stainless 27" to 50" 2-section with holder APC Heel stirrups with mounting adaptors Oxygen tank holder Flip-down side table Side tray Adjustable patient tray SA Foot pedal controls Paper roll holder SA SA ESC250 Only Armboard with 2" pad and pre-attached clamp Surgical accessory mounting rail IV well mount P Pad, 1/2" head boltaflex black (convoluted) SA SA EPC250 Only Dual side lateral access wrist rest Full U over the brow wrist rest Armboard with pad for sitting position WARNING: It is recommended that only accessories approved by GF Health Products, Inc. be used with this device. The use of accessories, transducers, and cables other than those specified by the manufacturer may result in increased emissions or decreased immunity of the Hausted equipment. Info: To order accessories, or for more detailed information on accessories, please contact GF Health Products, Inc. at APC-US-INS-LAB-RevB14 23

24 7 GF HEALTH PRODUCTS, INC. LIMITED WARRANTY FOR HAUSTED BRAND PRODUCT LINE WITHIN THE U.S. SCOPE OF WARRANTY GF Health Products, Inc. ( GF ) warrants to the original purchaser only that it will replace or repair components, at GF s sole discretion, that are defective in material or workmanship under normal use and service. All warranties are conditioned upon the proper use of the products strictly in accordance with good commercial practice and applicable GF instructions and manuals, including proper use and maintenance. To the extent that a component is warranted by a third party, GF conveys all of its rights under that warranty to the original purchaser, to the extent permitted. This limited warranty shall only apply to defects that are reported to GF s customer service team within the applicable warranty period and which, upon examination by GF or its authorized representative, prove to be a warranty item. This limited warranty is not transferable. The warranted components and time period are set forth below: COMPONENT PARTS WARRANTY LABOR WARRANTY Frame years years Hydraulics year year Casters year year Pneumatic gas springs year year Pads and mattresses year year Patient weighing system year year Electrical components year year Optional accessories year year Replacement parts * days * The warranty period is as designated above. If a part is replaced under warranty, the original warranty period will not be affected. All other replacement parts will be subject to the warranty period specified. The applicable warranty period shall commence from date of shipment to the original customer, unless there is an expiration date on the component in which case the warranty shall expire on the earlier of warranty period or the expiration date. OBTAINING WARRANTY SERVICE A GF Customer Service Representative must authorize warranty service. Please contact the GF Customer Service department by calling , sending a fax request to or by ing a request to cs@grahamfield.com. Specific directions will be provided by the Customer Service Representative. Failure to abide by the specific directions will result in denial of the warranty claim. EXCLUSIONS The warranty does not cover and GF shall not be liable for the following: 1) Defects, damage, or other conditions caused, in whole or in part, by misuse, abuse, negligence, alteration, accident, freight damage, tampering or failure to seek and obtain repair or replacement in a timely manner; 2) Products which are not installed, used, or properly cleaned and maintained as required in the official manual for the applicable product; 3) Products considered to be of a consumable nature including, but not limited to: filters, fuses, gaskets, lubricants, and charts; 4) Accessories or parts not provided by GF; 5) Charges by anyone for adjustments, repairs, replacement parts, installation or other work performed upon or in connection with such products which are not expressly authorized in writing, in advance, by GF; 6) Any shipping charges incurred in the replacement part installation or repair; 7) Costs and expenses of regular maintenance and cleaning; and 8) Representations and warranties made by any person or entity other than GF. ENTIRE WARRANTY, EXCLUSIVE REMEDY AND CONSEQUENTIAL DAMAGES DISCLAIMER THIS WARRANTY IS GF S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED. GF MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IF ANY MODEL OR SAMPLE WAS SHOWN TO THE CUSTOMER, SUCH MODEL OR SAMPLE WAS USED MERELY TO ILLUSTRATE THE GENERAL TYPE AND QUALITY OF THE PRODUCT AND NOT TO REPRESENT THAT THE PRODUCT WOULD NECESSARILY CONFORM TO THE MODEL OR SAMPLE IN ALL RESPECTS. THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT OF THE DEFECTIVE PARTS. GF SHALL NOT BE LIABLE FOR AND HEREBY DISCLAIMS ANY DIRECT, SPECIAL, INDIRECT, INCIDENTAL, EXEMPLARY OR CONSEQUENTIAL DAMAGES, INCLUDING, BUT NOT LIMITED TO: DAMAGES FOR LOSS OF PROFITS OR INCOME, LOSS OF USE, DOWNTIME, COVER, OR EMPLOYEE OR INDEPENDENT CONTRACTOR WAGES, PAYMENTS AND BENEFITS. The warranties contained herein contain all the representations and warranties with respect to the subject matter of this document, and supersede all prior negotiations, agreements and understandings with respect thereto. The recipient of this document hereby acknowledges and represents that it has not relied on any representation, assertion, guarantee, warranty, collateral contract or other assurance, except those set out in this document. For additional information on this Hausted product or this warranty, please contact a GF Customer Service Representative. NOTES: 1) Additional terms and conditions may apply. 2) Freight claims must be notated on the appropriate shipping documents and must be made with immediacy. International, federal and state regulations govern specific requirements for freight claims. Failure to abide by those regulations may result in a denial of the freight claim. GF will assist you in filing the freight claim. 3) Claims for any short shipment must be made within thirty (30) days of the invoice date. APC-US-INS-LAB-RevB14 24

25 8 DISPOSAL AND KEY TO SYMBOLS DISPOSAL Hausted equipment and accessories can be disposed of. We recommend disassembling and dividing the equipment and components into different waste groups such as: metal, cable, electronic, recoverable resource and plastic for recycling or combustion. Most plastic components are provided with a plastic types code and fiber content to aid sorting of plastic parts. Product Metal Scrap Cable Scrap Electronic Scrap Plastic Recycling or Combustion ESCEYE X X X X ESDEYE X X X X EPC250 X X X X EPD250 X X X X Info: Lead-acid batteries contained in the controller are recoverable resources and should be recycled. KEY TO SYMBOLS The following symbols are used on Hausted product labels. Protective Earth Manufacturer Earth Ground Keep Dry General Warning Sign Fragile, Handle with Care CE Mark Electrical and Electronic Equipment ETL Consult Instructions for Use EC REP European Authorized Representative Caution Disconnect before Service Pinch Point APC-US-INS-LAB-RevB14 25

26 9 APPENDIX 9.1 GUIDANCE AND MANUFACTURER S DECLARATION ELECTROMAGNETIC EMISSIONS The Hausted APC Series Models and Surgi-Chairs are intended for use in the electromagnetic environment specified below. The customer or the user of the Hausted APC Series Models and Surgi-Chairs should assure that it is used in such an environment. Emissions Test Compliance Electromagnetic Environment Guidance RF emissions CISPR 11 RF emissions CISPR 11 Harmonic emissions IEC Voltage fluctuations/flicker emissions IEC Group 1 Class A Class A Complies The Hausted APC Series Models and Surgi-Chairs uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. The Hausted APC Series Models and Surgi-Chairs are suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. 9.2 GUIDANCE AND MANUFACTURER S DECLARATION The Hausted APC Series Models and Surgi-Chairs are intended for use in the electromagnetic environment specified below. The customer or the user of the Hausted APC Series Models and Surgi-Chairs should assure that it is used in such an environment. Immunity Test IEC Test Level Electrostatic Discharge (ESD) IEC Electrical fast transient/burst IEC Surge IEC Voltage dips, short interruptions and voltage variations on power supply input lines IEC Power frequency (50/60 Hz) magnetic field IEC ±6 kv contact ±8 kv air ±2 kv for power supply lines ±1 kv for input/ output lines ±1 kv line(s) to line(s) ±2 kv line(s) to earth <5 % UT (>95 % dip in UT) for 0,5 cycle 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dip in UT) for 25 cycles <5 % UT (> 95% dip in UT) for 5 sec Compliance Level Electromagnetic Environment Guidance ±6 kv contact ±8 kv air ±2 kv for power supply lines ±1 kv for input/ output lines ±1 kv line(s) to line(s) ±2 kv line(s) to earth <5 % UT (>95 % dip in UT) for 0,5 cycle 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dip in UT) for 25 cycles <5 % UT (> 95% dip in UT) for 5 sec Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. Mains power quality should be that of a typical commercial or hospital environment. Mains power quality should be that of a typical commercial or hospital environment. Mains power quality should be that of a typical commercial or hospital environment. If the user of the Hausted APC Series Models and Surgi-Chairs requires continued operation during power mains interruptions, it is recommended that the Hausted APC Series Models and Surgi-Chairs be powered from an uninterrupted power supply or a battery. 3 A / m 3 A / m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. NOTE: UT is the a.c. mains voltage prior to application of the test level. APC-US-INS-LAB-RevB14 26

27 9.3 GUIDANCE AND MANUFACTURER S DECLARATION ELECTROMAGNETIC IMMUNITY The Hausted APC Series Models and Surgi-Chairs are intended for use in the electromagnetic environment specified below. The customer or the user of the Hausted APC Series Models and Surgi-Chairs should assure that it is used in such an environment. Immunity Test Conducted RF IEC Radiated RF IEC IEC Test Level 3 Vrms 150 khz to 80MHz 3 V/m 80 MHz to 2,5 GHz Compliance Level 3 vrms 3 V/m Electromagnetic Environment Guidance Portable and mobile RF communications equipment should be used no closer to any part of the Hausted APC Series Models and Surgi-Chairs, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance d = 1.2 P d = 1.2 P 80 MHz to 800 MHz d = 2.3 P 800 MHz to 2,5 ghz where P is the maximum output power rating of the transmitter in watts (w) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. b Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Hausted APC Series Models and Surgi-Chairs are used exceeds the applicable RF compliance level above, the Hausted APC Series Models and Surgi-Chairs should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocation the Hausted APC Series Models and Surgi-Chairs. b Over the frequency range 150 khz to 80 MHz, field strengths should be less than 3 V/m. APC-US-INS-LAB-RevB14 27

28 9.4 RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT AND HAUSTED APC SERIES MODELS AND SURGI-CHAIRS The Hausted APC Series Models and Surgi-Chairs are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Hausted APC Series Models and Surgi-Chairs can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Hausted APC Series Models and Surgi-Chairs as recommended below, according to the maximum output power of the communications equipment. Rated maximum output power of transmitter (W) Separation distance according to frequency of transmitter (m) 150 khz to 80 MHz d = 1.2 P 80 MHz to 800 MHz d = 1.2 P 800 MHz to 2,5 GHz d = 2.3 P 0, , For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. APC-US-INS-LAB-RevB14 28

29 10 INDEX A Accessories, optional 19 Accessories, optional, list 23 Advisory 4 APC / Surgi-Chair specifications 9 A word from GF Health Products, Inc. 4 B Back section, lowering 14 Back section, raising 14 Battery replacement instructions 21 Brakes, releasing 12 Brakes, steering, applying 12 Brakes, steering, releasing 12 Braking and steering operation 12 C CAUTION labels 11 CAUTION statement, significance 5 Cleaning 22 D DANGER statement, significance 5 DANGER / WARNING / CAUTION / NOTICE Summary 5 Disposal 25 E Electrical specifications (EPC / ESC) 9 Electrical specifications (EPD / ESD) 9 Electric control locations 13 Electromagnetic emissions, guidance and manufacturer s declaration 26 Electromagnetic immunity, guidance and manufacturer s declaration 27 Environmental conditions 8 F Features 11 Footrest, adjustable 16 Footrest, repositioning 16 Foot section adjustment 14 Foot section, lowering 14 Foot section, raising 14 H Headrest, adjusting 18 Headrest, ESCEYE/ESDEYE 18 Height, lowering 14 Height, raising 14 I Indications for Use 4 Info statement, significance 5 IV rod, #18 (optional accessory), installing 19 L Limited warranty 24 Low battery light 13 Lubrication 22 N NOTICE statement, significance 5 O Operating instructions 9 P Patient tray (optional accessory), installing 20 Pendant control location 13 Plug location 13 Preventive maintenance for the user 22 PROPER OPERATION labels 11 Q Quick drop activation 15 R Rail, repositioning 17 Rails, hinging 17 Restraint straps (optional accessory), using 20 S Safety statements, significance of 5 Service information 4 Side tray, optional accessory, installing 19 Symbol key, Hausted product labels 25 T Troubleshooting guide 21 U Uncrating instructions 8 Unpacking instructions 8 W WARNING labels 11 Warnings 5 WARNINGS and CAUTIONS, list 5 WARNING statements 5 WARNING statement, significance 5 Wrist rest (optional accessory), adjusting 19 Wrist rest (optional accessory), mounting 19 APC-US-INS-LAB-RevB14 29

30

Charging Station - Instructions For Use

Charging Station - Instructions For Use Charging Station - Instructions For Use READ THE HYBRESIS CONTROLLER AND PATCH INSTRUCTIONS FOR USE FOR ADDITIONAL IMPORTANT INFORMATION. REF: 199586 GLOSSARY OF SYMBOLS This device may contain one or

More information

Podiatry Procedures Chair. For Models: Barrier-Free. User s Guide

Podiatry Procedures Chair. For Models: Barrier-Free. User s Guide Podiatry Procedures Chair For Models: 646 647 Barrier-Free User s Guide Product Information (The information below is required when calling for service.) Dealer : Date of Purchase: Model / Serial Number:

More information

LF1600 Stand Assist Assembly and Operation Manual

LF1600 Stand Assist Assembly and Operation Manual LF1600 Stand Assist Assembly and Operation Manual LF1600-INS-LAB-RevH18 LF1600 Stand Assist user and attendant: Read this manual before operating the Stand Assist. Save this manual for future reference.

More information

Power Blade Charger Users Guide

Power Blade Charger Users Guide Power Blade Charger Users Guide Desktop Configuration Wall Mount Configuration Desktop Configuration (Displayed with Jaco Power Blade Batteries and 300 Series Laptop and LCD carts. Sold separately) 25-0180

More information

Installation Operation Maintenance Troubleshooting Version Dec/ EL / Biscayne E.L.

Installation Operation Maintenance Troubleshooting Version Dec/ EL / Biscayne E.L. Installation Operation Maintenance Troubleshooting Version Dec/17 6000EL / Biscayne E.L. Table of Contents DEAR CUSTOMER... 3 PRODUCT DESCRIPTION... 3 CLASSIFICATIONS... 4 GENERAL DATA... 8 TECHNICAL SPECIFICATIONS...

More information

LF1600 Stand Assist Assembly and Operation Manual

LF1600 Stand Assist Assembly and Operation Manual LF1600 Stand Assist Assembly and Operation Manual LF1600-INS-LAB-RevD13 2013 GF Health Products, Inc. LF1600 Stand Assist user and attendant: Read this manual before operating the Stand Assist. Save this

More information

Laboratory Centrifuge Model GF614B

Laboratory Centrifuge Model GF614B Laboratory Centrifuge Model GF614B User Manual Read this manual before operating the GF614B Laboratory Centrifuge. Save this manual for future reference. GF614B-INS-LAB-RevC15 Info: The Grafco GF614B Laboratory

More information

2100 Childbearing Bed OPERATIONS MANUAL. For Parts or Technical Assistance

2100 Childbearing Bed OPERATIONS MANUAL. For Parts or Technical Assistance 2100 Childbearing Bed OPERATIONS MANUAL For Parts or Technical Assistance 1 800 327 0770 Table of Contents INTRODUCTION........................................................................... 2 SPECIFICATIONS.........................................................................

More information

Renaissance Series Full Length X Ray Stretcher OPERATIONS MANUAL. For Parts or Technical Assistance

Renaissance Series Full Length X Ray Stretcher OPERATIONS MANUAL. For Parts or Technical Assistance Renaissance Series 1026 Full Length X Ray Stretcher OPERATIONS MANUAL For Parts or Technical Assistance 1 800 327 0770 Table of Contents INTRODUCTION.......................................................................

More information

ELORA BATTERY INTERFACE. Owner s Manual

ELORA BATTERY INTERFACE. Owner s Manual ELORA BATTERY INTERFACE Owner s Manual CONTENTS 3 4 7 8 12 19 IMPORTANT READ AND UNDERSTAND THIS MANUAL BEFORE OPERATION INTRODUCTION KEY FEATURES AND BENEFITS ORDERING INFORMATION MOUNTING FUEL GAUGE

More information

1000 Emergency Care Stretcher 1500 PACU Stretcher OPERATIONS MANUAL. For Parts or Technical Assistance

1000 Emergency Care Stretcher 1500 PACU Stretcher OPERATIONS MANUAL. For Parts or Technical Assistance 1000 Emergency Care Stretcher 1500 PACU Stretcher OPERATIONS MANUAL For Parts or Technical Assistance 1 800 327 0770 Table of Contents INTRODUCTION........................................................................

More information

USER MANUAL. EC+ Extra Care Bariatric Trauma Stretcher Model ConvaQuip Ind., Inc. P.O. Box 3417 Abilene, TX

USER MANUAL. EC+ Extra Care Bariatric Trauma Stretcher Model ConvaQuip Ind., Inc. P.O. Box 3417 Abilene, TX USER MANUAL EC+ Extra Care Bariatric Trauma Stretcher Model 1600 ConvaQuip Ind., Inc. P.O. Box 3417 Abilene, TX 79604 www.convaquip.com Table of Contents Parts Locator 3 Safety Notes 5 Weight Limit 6

More information

LF2020 EASY LIFT SIT-TO-STAND LF2090 BARIATRIC EASY LIFT SIT-TO-STAND

LF2020 EASY LIFT SIT-TO-STAND LF2090 BARIATRIC EASY LIFT SIT-TO-STAND LF2020 EASY LIFT SIT-TO-STAND LF2090 BARIATRIC EASY LIFT SIT-TO-STAND USER MANUAL LF2020-INS-LAB-RevE15 2005 GF Health Products, Inc. Read this manual before operating your Easy Lift Sit-to-Stand. Save

More information

OAKWORKS 300 Series Procedure Chair

OAKWORKS 300 Series Procedure Chair USER MANUAL OAKWORKS 300 Series Procedure Chair www.oakworksmed.com 717.235.6807 made in the USA with US & imported parts Copyright 2013 OAKWORKS, Inc. Notice This manual was originally drafted and approved

More information

UNIVERSAL BATTERY CHARGER II

UNIVERSAL BATTERY CHARGER II UNIVERSAL BATTERY CHARGER II USER S MANUAL TABLE OF CONTENTS INTRODUCTION General Information 2 Intended use 2 Compatibility 2 Abbreviations 2 Safety precautions 3 Scope of delivery 3 Storage and transport

More information

OAKWORKS 300 Series Procedure Chair

OAKWORKS 300 Series Procedure Chair User MANUAL OAKWORKS 300 Series Procedure Chair www.oakworksmed.com 717.235.6807 made in the USA with US & imported parts Copyright 2013 OAKWORKS, Inc. Notice This manual was originally drafted and approved

More information

Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by:

Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by: Storage System Ultrasound Probe Cabinet Operator's Manual Manufactured by: CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA 319.248.6757 / 800.445.6741 WWW.CIVCO.COM Copyright 2018 All

More information

Heavy Duty Four Wheeled Walker

Heavy Duty Four Wheeled Walker Heavy Duty Four Wheeled Walker Weight Capacity: 500 lbs. ITEM # W1802 Made in China 2011 ESSENTIAL MEDICAL SUPPLY, INC. Manufactured for Orlando, FL 32822 -- SAVE THESE INSTRUCTIONS -- Do not attempt to

More information

Owner s Manual EchoBed X VasScan Table X

Owner s Manual EchoBed X VasScan Table X Owner s Manual EchoBed X VasScan Table X Owner Model Serial # Date Echo/Vascular User Manual (15381-C) Page 1 of 46 DCR-00172 Table of Contents Symbols and Definitions... 3 Safety Warnings & Cautions...

More information

25 GALLON PORTABLE OIL LIFT

25 GALLON PORTABLE OIL LIFT 25 GALLON PORTABLE OIL LIFT Model 92859 SET UP AND OPERATING INSTRUCTIONS Diagrams within this manual may not be drawn proportionally. Due to continuing improvements, actual product may differ slightly

More information

IV STANDS. Assembly, Usage and Maintenance INSTRUCTION MANUAL

IV STANDS. Assembly, Usage and Maintenance INSTRUCTION MANUAL INSTRUCTION MANUAL IV STANDS Assembly, Usage and Maintenance Read and understand all of the instructions and safety information in this manual before operating this product. MAN-045 REV A 2018 MAC Medical,

More information

1. Contents 1. CONTENTS INTRODUCTION SAFETY WARNINGS POWER TILT STANDARD OPERATING INSTRUCTIONS... 5.

1. Contents 1. CONTENTS INTRODUCTION SAFETY WARNINGS POWER TILT STANDARD OPERATING INSTRUCTIONS... 5. 1. Contents 1. CONTENTS..... 2. INTRODUCTION..... 3. SAFETY WARNINGS..... 4. POWER TILT STANDARD OPERATING INSTRUCTIONS....... 5. POWER SEAT LIFT STANDARD OPERATING INSTRUCTIONS...... 6. POWER TILT / SEAT

More information

Owner s Manual Ultra DBI TM Model: 9400

Owner s Manual Ultra DBI TM Model: 9400 Owner s Manual Ultra DBI TM Model: 9400 Owner Model Serial # Date Medical Positioning, Inc. 1146 Booth Street, Kansas City, Kansas 66103 T: 816-474-1555 1-800-593-3246 www.medicalpositioning.com Table

More information

1250 LB. CAPACITY MECHANICAL WHEEL DOLLY

1250 LB. CAPACITY MECHANICAL WHEEL DOLLY 1250 LB. CAPACITY MECHANICAL WHEEL DOLLY 67287 SET-UP AND OPERATING INSTRUCTIONS Visit our website at: http://www.harborfreight.com Read this material before using this product. Failure to do so can result

More information

Simplicity Light Post Mount

Simplicity Light Post Mount Simplicity Light Post Mount Table of Contents Section I Introduction Specifications... 2 Classifications...2-3 Packaging... 3 Section II Installation Post Mount with Service Console...4-5 Section III Operation

More information

Veterinary Anesthesia Gas Scavenging System (Active)

Veterinary Anesthesia Gas Scavenging System (Active) Veterinary Anesthesia Gas Scavenging System (Active) Installation and Operation Manual MX103200i 10434800 Rev. AA1 5/5/17 Table of Contents Section 1 UNPACKING...2 1.1 Examining Packaged Components...2

More information

Operating Manual. Hausted Horizon Series Multi-Purpose Stretchers Electric Powered Models: 4E1 & 4E2 4D1 & 4D2 V

Operating Manual. Hausted Horizon Series Multi-Purpose Stretchers Electric Powered Models: 4E1 & 4E2 4D1 & 4D2 V Operating Manual Hausted Horizon Series Multi-Purpose Stretchers Electric Powered Models: 4E1 & 4E2 4D1 & 4D2 V1 05.11 HAUSTED Patient Handling Systems, LLC WARNING - COPYING PROHIBITED This manual is

More information

User Manual SONESTA S2 & S3

User Manual SONESTA S2 & S3 User Manual SONESTA S2 & S3 Sonesta manufactures and sells tables and chairs for urology, gynecology, urodynamic diagnostics and upper GI procedures Copyright 2018 SONESTA MEDICAL AB. All rights reserved.

More information

Magician Comfy. Owner s Manual

Magician Comfy. Owner s Manual Magician Comfy Owner s Manual Table Of Contents Introduction...................................... 2 Safety Precautions................................ 3 Assembly........................................4-9

More information

User MANUAL. PG Series. Ultrasound Series. Ultrasound series. made in the USA with US & imported parts

User MANUAL. PG Series. Ultrasound Series. Ultrasound series. made in the USA with US & imported parts User MANUAL PG Series & Ultrasound Series PG Series Ultrasound series made in the USA with US & imported parts www.oakworksmed.com 717.235.6807 Copyright 2010 OAKWORKS, Inc. Notice This manual was originally

More information

Owner s Manual. Powered Dandy Lifts PLM-100, PLM100W, PLM150, PLM-150W. (See separate manual for PLM-250) Model # Serial #

Owner s Manual. Powered Dandy Lifts PLM-100, PLM100W, PLM150, PLM-150W. (See separate manual for PLM-250) Model # Serial # Owner s Manual Powered Dandy Lifts PLM-00, PLM00W, PLM50, PLM-50W (See separate manual for PLM-250) Model # Serial # Southworth Products Corp P.O. Box 380/Portland, Maine 0404-380 Phone 800-743-000 FAX

More information

Tri-Wheel Wheeled Walker

Tri-Wheel Wheeled Walker Tri-Wheel Wheeled Walker Handle with lever brake Brake cable Handle height adjustment knob Removable basket and tray Large vinyl bag 8 inch (203 mm) wheels user guide Prior to use please read all instructions.

More information

Swing Arm Magnifying Lamp

Swing Arm Magnifying Lamp Owner s Manual & Safety Instructions Save This Manual Keep this manual for the safety warnings and precautions, assembly, operating, inspection, maintenance and cleaning procedures. Write the product s

More information

Euro Lightweight Wheeled Walker

Euro Lightweight Wheeled Walker Euro Lightweight Wheeled Walker Handle with lever brake Backrest Brake cable Handle height adjustment button Removable shopping bag 8 inch (200 mm) wheels Cane holder user guide Prior to use please read

More information

OPERATION and SERVICE MANUAL EC TRANS. CHAIR

OPERATION and SERVICE MANUAL EC TRANS. CHAIR OPERATION and SERVICE MANUAL EC TRANS. CHAIR MODEL 900EC THE SERIAL NUMBER WILL BE FOUND ON THE LOWER FRAME AT THE HEAD END. RECORD THE NUMBER BELOW: SERIAL NUMBER: DATE OF PURCHASE ConvaQuip Ind., Inc.

More information

SAVE THESE INSTRUCTIONS DANGER

SAVE THESE INSTRUCTIONS DANGER Contoura Back and Seat Cushions Assembly, Installation and Operating Instructions SAVE THESE INSTRUCTIONS To ensure the safe operation and use of the Contoura Back and Cushion, these instructions MUST

More information

Operating Instructions & Parts Manual. Fuel Tank Adapter

Operating Instructions & Parts Manual. Fuel Tank Adapter Operating Instructions & Parts Manual Fuel Tank Adapter Model Number 40080 Capacity 80 lb.! This is the safety alert symbol. It is used to alert you to potential personal injury hazards. Obey all safety

More information

Megohmmeter Test Probe

Megohmmeter Test Probe Megohmmeter Test Probe User Manual ENGLISH www.aemc.com CHAUVIN ARNOUX GROUP Copyright Chauvin Arnoux, Inc. d.b.a. AEMC Instruments. All rights reserved. No part of this documentation may be reproduced

More information

RENA AF371Feeder Operating Manual. Feeder. Operating Manual. Manual Part #: M AF371 Operations Rev

RENA AF371Feeder Operating Manual. Feeder. Operating Manual. Manual Part #: M AF371 Operations Rev Manual Part #: M-3022 Feeder AF371 Operations Rev. 3-16-04 1 RENA AF371 Feeder YOUR RENA AF371 IS DISTRIBUTED BY RENA SYSTEMS INC. SERVICE AND SUPPORT FOR THIS PRODUCT IS PROVIDED BY YOUR RENA DEALER.

More information

Printed: Doc-Nr: PUB / / 000 / 00

Printed: Doc-Nr: PUB / / 000 / 00 ORIGINAL OPERATING INSTRUCTIONS Hilti HTE-P 33 dispenser It is essential that the operating instructions are read before the tool is operated for the first time. Always keep these operating instructions

More information

KidWalk KidWalk II Dynamic Mobility System

KidWalk KidWalk II Dynamic Mobility System OWNER S MANUAL KidWalk KidWalk II Dynamic Mobility System Manufactured By Prime Engineering A Division of Axiom Industries, Inc. Supplier Info 70111KWOM 2 TABLE OF CONTENTS This owner s manual is organized

More information

QA4 Surgery Trolley System Manual Function Operating Instructions

QA4 Surgery Trolley System Manual Function Operating Instructions QA4 Surgery Trolley System Manual Function Operating Instructions Catalogue No. 21310 Anetic Aid Ltd. Queensway Guiseley West Yorkshire, LS20 9JE United Kingdom T +44 (0) 1943 878647 F +44 (0) 1943 870455

More information

Users Guide for Ac-sync

Users Guide for Ac-sync Problem solved. Users Guide for Ac-sync Thank you for choosing Anywhere Cart! The AC-SYNC is designed to sync, charge and store 1-36 ipads or tablets. Adjustable device divider bays allow fitment of any

More information

THE PHOENIX MODEL 1800 KNEE CPM OPERATOR'S MANUAL SERIES Seeds Road Grove City, OH Fax

THE PHOENIX MODEL 1800 KNEE CPM OPERATOR'S MANUAL SERIES Seeds Road Grove City, OH Fax TM THE PHOENIX SERIES MODEL 1800 KNEE CPM OPERATOR'S MANUAL www.mckelor.com 6312 Seeds Road Grove City, OH 43123 800.273.5233 Fax 614.871.9353 Phoenix Model 1800 Knee CPM Operator's Manual Section 1 TABLE

More information

3000-Lb. Vehicle Positioning Jacks. Owner s Manual

3000-Lb. Vehicle Positioning Jacks. Owner s Manual 3000-Lb. Vehicle Positioning Jacks Owner s Manual WARNING: Read carefully and understand all ASSEMBLY AND OPERATION INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety

More information

User s Manual. ClipperCreek, Inc. Innovative Infrastructure for Electric and Hybrid Vehicles. Model PCS-15

User s Manual. ClipperCreek, Inc. Innovative Infrastructure for Electric and Hybrid Vehicles. Model PCS-15 ClipperCreek, Inc. Innovative Infrastructure for Electric and Hybrid Vehicles User s Manual ClipperCreek, Inc. 11850 Kemper Rd., Suite E Auburn, CA 95603 www.clippercreek.net Model PCS-15 THIS PAGE INTENTIONALLY

More information

User s Manual and Operating Instructions

User s Manual and Operating Instructions User s Manual and Operating Instructions Model Numbers: CL-36-BDF-A, CL-42-BDF-A, CL-48-BDF-A E355088 READ AND SAVE THESE INSTRUCTIONS IMPORTANT: Read and understand all of the instructions in this manual

More information

Blades with the MatrixPRO Driver may result in damage to the MatrixPRO Driver and/or screwdriver blades, and should

Blades with the MatrixPRO Driver may result in damage to the MatrixPRO Driver and/or screwdriver blades, and should Description: The MatrixPRO Driver (05.000.020) is designed to insert DePuy Synthes MatrixNEURO TM Self Drilling Screws (3 4mm lengths) and Low Profile Neuro Self Drilling Screws (3 4mm lengths) and features

More information

540/545 Universal Procedures Stretchers

540/545 Universal Procedures Stretchers Go To Table Of Contents 540/545 Universal Procedures Stretchers Installation and Manual 503-027-00 Rev. H (/05) Important Information Page 2 Installation Page 4 Description Page 5 Components Overview Page

More information

Xenon Fiberoptic Lightsource

Xenon Fiberoptic Lightsource Xenon Fiberoptic Lightsource Operating Instructions for ASB-XE-175W Lightsource 2659A Pan American Frw N.E Albuquerque, New Mexico 87107 USA (English) Page 1 of 12 TABLE OF CONTENTS PAGE 1. INTRODUCTION

More information

MODEL 7400 STRUT SPRING COMPRESSOR

MODEL 7400 STRUT SPRING COMPRESSOR MODEL 7400 STRUT SPRING COMPRESSOR Installation, Operation & Repair Parts Information Branick Industries, Inc. 4245 Main Avenue P.O. Box 1937 Fargo, North Dakota 58103 REV112712 P/N: 81-0103A TABLE OF

More information

Installation / Operation Instructions Sunnex ORION Series Exam Lights

Installation / Operation Instructions Sunnex ORION Series Exam Lights Installation / Operation Instructions Sunnex ORION Series Exam Lights OR-120 OR-127 OR-220 OR-227 Models: OR-300 OR-400 OR-500 OR-600 1. APPLICATIONS The Sunnex ORION Series light was designed specifically

More information

Transit Wheelchair. user guide. Profiled grip attendant handles. Padded seat and backrest. Lift up, padded armrests. Park brakes

Transit Wheelchair. user guide. Profiled grip attendant handles. Padded seat and backrest. Lift up, padded armrests. Park brakes Transit Wheelchair Profiled grip attendant handles Padded seat and backrest Lift up, padded armrests Park brakes Pneumatic 12 (305 mm) rear wheels Solid 7 (180 mm) front wheels Detachable, swing away footrests

More information

Operator s Manual. Pre-Clinical Blood Sampling Tool

Operator s Manual. Pre-Clinical Blood Sampling Tool Operator s Manual Pre-Clinical Blood Sampling Tool + Read the entire Operator s Manual before using GentleSharp Model 101. Disregarding the cautions and instructions presented in this manual constitutes

More information

Apollo Tilt Wheelchair 18" and 20" Instruction Manual

Apollo Tilt Wheelchair 18 and 20 Instruction Manual Apollo Tilt Wheelchair 18" and 20" Instruction Manual Table of Contents Contents Table of Contents... 2 Introduction... 3 Warning... 4 Safety Guidelines... 5 Parts of the Tilt Wheelchair... 6 Setup & Operation...

More information

Adjustable Angled Incline Conveyor Owners Manual with Operating Instructions

Adjustable Angled Incline Conveyor Owners Manual with Operating Instructions Adjustable Angled Incline Conveyor Owners Manual with Operating Instructions Revision 012211 Table of Contents Basic Conveyor Features 3 Getting Started 4 Setting Up the Incline Conveyor 5 Belt Removal

More information

Gives useful tips, recommendations and information for efficient, trouble-free use.

Gives useful tips, recommendations and information for efficient, trouble-free use. User Manual DEALER: This manual MUST be given to the user of the product. USER: BEFORE using this product, read this manual and save for future reference. 1 General 1.1 Symbols Invacare Matrx Stabilite

More information

LifeGuardLift. LifeGuard Power Lift Model #100287A OWNERS MANUAL. Rev: 2/14/11

LifeGuardLift. LifeGuard Power Lift Model #100287A OWNERS MANUAL. Rev: 2/14/11 LifeGuardLift OWNERS MANUAL LifeGuard Power Lift Model #100287A Rev: 2/14/11 Table of Contents 1. ASSEMBLY INSTRUCTIONS A. Lift Assembly B. Setup C. Disassembly 2. CONTROL SYSTEM A. Batteries B. Battery

More information

DenLite Illuminated Dental Mirror

DenLite Illuminated Dental Mirror DenLite Illuminated Dental Mirror Directions for Use DenLite Illuminated Dental Mirror Miltex, Inc. assumes no responsibility for any injury to anyone, for any illegal or improper use of the product that

More information

900 PEAK AMP PORTABLE JUMP STARTER

900 PEAK AMP PORTABLE JUMP STARTER 900 PEAK AMP PORTABLE JUMP STARTER Item Number W1665 OWNER S MANUAL WARNING It is the owner and/or operators responsibility to study all WARNINGS, operating, and maintenance instructions contained on the

More information

OPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-707A & AEU-707AV2 Implant / Surgery Systems

OPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-707A & AEU-707AV2 Implant / Surgery Systems OPERATION AND MAINTENANCE INSTRUCTION MANUAL AEU-707A & AEU-707AV2 Implant / Surgery Systems TABLE OF CONTENTS: Introduction.........................1 Package Contents....................1 Setting Up the

More information

OAKWORKS PF Series Tables

OAKWORKS PF Series Tables User MANUAL OAKWORKS PF Series Tables PF400 Shown with optional Adjustable Side Arm Rests and Universal Table Extender PF250 PF200 Shown with optional QuickLock Face rest platform and Boiance crescent

More information

Renaissance Series Critical Care Bed OPERATIONS MANUAL. For Parts or Technical Assistance

Renaissance Series Critical Care Bed OPERATIONS MANUAL. For Parts or Technical Assistance Renaissance Series 2020 Critical Care Bed OPERATIONS MANUAL For Parts or Technical Assistance 1 800 327 0770 Table of Contents Introduction...............................................................................

More information

Installation Guide Document N Cleanroom Mirrors Copyright 2016 Terra Universal Inc. All rights reserved. Revised November 2016

Installation Guide Document N Cleanroom Mirrors Copyright 2016 Terra Universal Inc. All rights reserved. Revised November 2016 Document N0. 1800-09 Copyright 2016 Terra Universal Inc. All rights reserved. Revised November 2016 Terra Universal, Inc. TerraUniversal.com 800 S. Raymond Ave. Fullerton, CA 92831 TEL: (714) 578-6000

More information

Cordless Sweeper. OWNER S MANUAL Model: CFT Volt DC. Made Exclusively for ALDI Inc. Batavia, IL.,

Cordless Sweeper. OWNER S MANUAL Model: CFT Volt DC. Made Exclusively for ALDI Inc. Batavia, IL., Cordless Sweeper OWNER S MANUAL Model: CFT25 4.8 Volt DC Made Exclusively for ALDI Inc. Batavia, IL., 60510-1477 CFT25 ONE (1) YEAR LIMITED WARRANTY This product is warranted to be free from defects in

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Thank you for purchasing a LOMAX TM Hard Tri-Fold or Professional Series Cover. Agri-Cover, Inc. proudly manufactured this cover using superior quality materials and workmanship.

More information

OAKWORKS 100 Series Procedure Chair

OAKWORKS 100 Series Procedure Chair USER MANUAL OAKWORKS 100 Series Procedure Chair Model 121A Model 100 www.oakworks.com 717.235.6807 Copyright 2018 OAKWORKS, Inc. Notice Printed in U.S.A. All rights are reserved. No part of this document

More information

PUSH BUTTON KEY CABINET

PUSH BUTTON KEY CABINET PUSH BUTTON KEY CABINET Model 95689 INSTALLATION And Operation Instructions Due to continuing improvements, actual product may differ slightly from the product described herein. 3491 Mission Oaks Blvd.,

More information

User Manual. Invacare Matrx Flovair. Models ITFM and ITFG. 1 General

User Manual. Invacare Matrx Flovair. Models ITFM and ITFG. 1 General User Manual DEALER: This manual MUST be given to the user of the product. USER: BEFORE using this product, read this manual and save for future reference. 1 General Invacare Matrx Flovair Models ITFM and

More information

APCO CRF-100A RUBBER FLAPPER SWING CHECK VALVES

APCO CRF-100A RUBBER FLAPPER SWING CHECK VALVES APCO CRF-100A RUBBER FLAPPER SWING CHECK VALVES Instruction D12043 June 2016 DeZURIK Instructions These instructions provide installation, operation and maintenance information for APCO CRF-100A Rubber

More information

Owner s Manual & Safety Instructions

Owner s Manual & Safety Instructions Owner s Manual & Safety Instructions Save Save This This Manual Keep Keep this this manual manual for for the the safety safety warnings warnings and and precautions, assembly, assembly, operating, inspection,

More information

USER MANUAL. Tides La Mer Treatment Table Rev A 3/28/19. *all tables shown with optional accessories and/or finishes

USER MANUAL. Tides La Mer Treatment Table Rev A 3/28/19. *all tables shown with optional accessories and/or finishes USER MANUAL Tides La Mer Treatment Table *all tables shown with optional accessories and/or finishes 1-400-0041 Rev A 3/28/19 Living Earth Crafts 2019, All Rights Reserved Legal Notice The information

More information

MODEL 7600 STRUT SPRING COMPRESSOR

MODEL 7600 STRUT SPRING COMPRESSOR MODEL 7600 STRUT SPRING COMPRESSOR Installation, Operation & Repair Parts Information Branick Industries, Inc. 4245 Main Avenue P.O. Box 1937 Fargo, North Dakota 58103 REV6162014 P/N: 81-0246 TABLE OF

More information

Push Down Wheeled Walker

Push Down Wheeled Walker Push Down Wheeled Walker Push down hand grip Handle height adjustment knob Backrest Padded seat with concealed pouch Side brace Bag with shopping basket inside Push down brake stopper 6 inch (150 mm) wheels

More information

DSC250 SCALE for LF1050 and LF1090 Lifts USER MANUAL

DSC250 SCALE for LF1050 and LF1090 Lifts USER MANUAL DSC250 SCALE for LF1050 and LF1090 Lifts USER MANUAL Read these instructions before using the Lumex Lift Scale. Save these instructions for future reference. The most current version of these instructions

More information

Linear Actuator. Installation Manual. warranty installation parts list. Linear Actuator Installation Manual Page 1

Linear Actuator. Installation Manual. warranty installation parts list. Linear Actuator Installation Manual Page 1 Linear Actuator Installation Manual warranty installation parts list January 2004 Linear Actuator Installation Manual Page 1 MA1221B12 Warranty Information Chore-Time Equipment ( Chore-Time ) warrants

More information

Bimini Top, Bimini Top Plus for Jeep Wrangler (YJ) Vehicles Items #141031XX and #143031XX

Bimini Top, Bimini Top Plus for Jeep Wrangler (YJ) Vehicles Items #141031XX and #143031XX MT_YJ_BIMINI_INST_LTR_6.qxp_Layout 1 4/20/17 4:21 PM Page 1 Installation Instructions Bimini Top, Bimini Top Plus for 1987-1995 Jeep Wrangler (YJ) Vehicles Items #141031XX and #143031XX Bimini Top Plus

More information

HTR-202 INSTRUCTION MANUAL. Portable Digital Height Rod. Befour, Inc. 102 Progress Drive Saukville, WI 53080

HTR-202 INSTRUCTION MANUAL. Portable Digital Height Rod. Befour, Inc. 102 Progress Drive Saukville, WI 53080 Befour, Inc. 102 Progress Drive Saukville, WI 53080 Phone: 262-284-5150 Toll Free: 1-800-367-7109 Fax: 262-284-5966 Email: mail@befour.com www.befour.com 2014 Befour, Inc. Rev.1 HTR-202 Portable Digital

More information

CT4880 Collaboration Huddle Table Assembly Instructions

CT4880 Collaboration Huddle Table Assembly Instructions CT4880 Collaboration Huddle Table Assembly Instructions Monitor, monitor stand and chairs are not included. Thank you for choosing the CT4880 Collaboration Huddle Table from AmpliVox Portable Sound Systems.

More information

Universal Battery Charger II.

Universal Battery Charger II. Universal Battery Charger II. Instructions for Use This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation Table of Contents Introduction General

More information

AM-134, AM-140, AM-234, AM-240, AM-334, AM-340, AM-368

AM-134, AM-140, AM-234, AM-240, AM-334, AM-340, AM-368 This manual has been prepared for the owners and operators of the ARMEDICA AM-Series Hi-Lo Treatment Tables. It contains installation instructions, precautionary instructions and maintenance procedures

More information

Mighty Mack Wheeled Walker

Mighty Mack Wheeled Walker Mighty Mack Wheeled Walker Handle with lever brake Backrest Handle height adjustment knob Brake cable Extra wide padded seat Shopping basket Side brace 7 inch (180 mm) wheels user guide Prior to use please

More information

1 Intended Use. 2 Precautions for handling and operation. English

1 Intended Use. 2 Precautions for handling and operation. English Thank you for purchasing the NSK icare. Please read this Operation Manual carefully before use to become familiar with operation instructions and care & maintenance. Keep this Operation Manual for future

More information

Lubricator Gun: 10,000 psi (700 bar) Maximum Delivery Pressure when disconnected from Dispenser

Lubricator Gun: 10,000 psi (700 bar) Maximum Delivery Pressure when disconnected from Dispenser INSTRUCTIONS-PARTS LIST 30 455 INSTRUCTIONS This manual contains important warnings and information. READ AND KEEP FOR REFERENCE. Rev. C Supercedes B Hand-Operated Portable Grease Dispenser Buckshot Luber

More information

Owners Manual. LifeGuard Power Lift Model # Rev. 2/1/13

Owners Manual. LifeGuard Power Lift Model # Rev. 2/1/13 Owners Manual LifeGuard Power Lift Model #100287 Rev. 2/1/13 Table of Contents 1. ASSEMBLY INSTRUCTIONS 3-5 A. Lift Assembly 3 B. Setup 3 1. Clinch Pin Location Drawings 4 2. Down Tube and Seat Assembly

More information

Hematocrit Centrifuge Model 410E User Manual

Hematocrit Centrifuge Model 410E User Manual Hematocrit Centrifuge Model 410E User Manual Read this manual before operating the 410E Hematocrit Centrifuge. Save this manual for future reference. 410E-INS-LAB-RevC15 CONTENTS INTRODUCTION... 2 PRODUCT

More information

Air-Operated Waste Oil Drainer

Air-Operated Waste Oil Drainer Air-Operated Waste Oil Drainer 20-Gallon Tank Owner s Manual WARNING: Read carefully and understand all ASSEMBLY AND OPERATION INSTRUCTIONS before operating. Failure to follow the safety rules and other

More information

DeZURIK 2 20" BOS BUTTERFLY VALVES

DeZURIK 2 20 BOS BUTTERFLY VALVES 2 20" BOS BUTTERFLY VALVES Instruction D10459 October 2013 2-20 BOS Butterfly Valves Instructions These instructions provide information about BOS Butterfly Valves. They are for use by personnel who are

More information

Automatic Emergency Light

Automatic Emergency Light Automatic Emergency Light Item 38013 Read this material before using this product. Failure to do so can result in serious injury. Save this manual. When unpacking, make sure that the product is intact

More information

OPERATING MANUAL POWER DRIVE SYSTEM. Cat. No RevH

OPERATING MANUAL POWER DRIVE SYSTEM. Cat. No RevH OPERATING MANUAL POWER DRIVE SYSTEM Cat. No. 400005 525-1300-00 RevH System components Symbol Definitions Qty Description Catalog Number 1 Power Drive System 300106 1 Controller 300108 1 Cutter Drive

More information

Owner s Manual. Bariatric Manual Wheelchairs. Specializing in Bariatric Patient Aids

Owner s Manual. Bariatric Manual Wheelchairs. Specializing in Bariatric Patient Aids Owner s Manual Bariatric Manual Wheelchairs 700 LB CAPACITY XL Series Wheelchair by ConvaQuip Specializing in Bariatric Patient Aids P.O. Box 3417 * Abilene, TX 79604 U.S.A. Toll 800/637-8436 * FAX 325/677-7217

More information

HEINE NT4 MED V

HEINE NT4 MED V HEINE NT4 MED 113594 2018-01-23 V-200.00.601 HEINE NT4 Please read and follow these instructions for use and keep them for future reference. Intended Use The HEINE charger NT4 is to be used exclusively

More information

User Manual S Line Bed

User Manual S Line Bed User Manual S Line Bed Important Please read before operation Model SLLS001 2 Contents 1 Essential Information 5 1.1 Technical Specifications 5 1.2 Intended Use 6 1.3 Unintended Use 7 1.4 Transportation

More information

User Manual. Posture+ Adjustable Base. For customer service call:

User Manual. Posture+ Adjustable Base. For customer service call: User Manual Posture+ Adjustable Base For customer service call: 1-877-707-7533 1 IMPORTANT INFORMATION PLEASE READ THESE INSTRUCTIONS THOROUGHLY BEFORE USING THIS PRODUCT. PROPER OPERATION OF YOUR ADJUSTABLE

More information

Air Curtain. Installation, Operating and Maintenance Instructions

Air Curtain. Installation, Operating and Maintenance Instructions Installation, Operating and Maintenance Instructions Save this manual for future reference. Air Curtain Model Numbers: ES026, ES036, ES042, ES048, ES060, ES072 READ THIS OWNER S MANUAL CAREFULLY BEFORE

More information

BK BK A BK MA BK MA BK J93 MN93-BK MN193-BK MR193-BK SL261-BK SR193-BK

BK BK A BK MA BK MA BK J93 MN93-BK MN193-BK MR193-BK SL261-BK SR193-BK n COMPATIBLE CURRENT PROBES & SENSORS FOR POWER QUALITY METERS 193-24-BK 193-36-BK A196-24-BK MA193-10-BK MA193-14-BK J93 MN93-BK MN193-BK MR193-BK SL261-BK SR193-BK E N G L I S H User Manual Copyright

More information

Lumina 28 Traction Drive Model: Battery Burnisher M28036TDQP OPERATION SERVICE PARTS CARE

Lumina 28 Traction Drive Model: Battery Burnisher M28036TDQP OPERATION SERVICE PARTS CARE Lumina 28 Traction Drive Model: Battery Burnisher M28036TDQP OPERATION SERVICE PARTS CARE Table of Contents IMPORTANT SAFETY INSTRUCTIONS...1 OPERATING INSTRUCTIONS...2 INSPECTION...2 ELECTRICAL...2 BATTERIES...2

More information

User Manual Stealth s Power Tilt

User Manual Stealth s Power Tilt User Manual Stealth s Power Tilt Copyright 2014 Stealth Products, Inc. Copyright 2014 Stealth Products, Inc. All rights reserved. Published by Stealth Products, Inc. November 7, 2014 P45D61R1 Customer

More information

Adjustable Base. CONTENTS Advisory Setup and Installation Innova Wired Hand Control year Warranty 1-10

Adjustable Base. CONTENTS Advisory Setup and Installation Innova Wired Hand Control year Warranty 1-10 Adjustable Base #201 Wired Owner Manual CONTENTS Advisory Setup and Installation Innova Wired Hand Control 1-2-10 year Warranty 1-10 WARNING Attention: Read the following information before using this

More information

Dispenser & Warmer RIC-1909 RIC-1909EXP

Dispenser & Warmer RIC-1909 RIC-1909EXP Dispenser & Warmer RIC-1909 RIC-1909EXP Safety Precautions CAUTION This equipment is designed and sold for commercial use only. This equipment is not to be used by the consumer in home use. Do not allow

More information