Operation and Maintenance Plan for the Vapor Mitigation System Building 040-H

Size: px
Start display at page:

Download "Operation and Maintenance Plan for the Vapor Mitigation System Building 040-H"

Transcription

1 U.S. Army Corps of Engineers Kansas City District Final Operation and Maintenance Plan for the Vapor Mitigation System Building 040-H Accelerated Operable Unit (AOU) 1 Time Critical Removal Action and Mitigation System Installation for the 700 South 1600 East PCE Plume Superfund Site VA Salt Lake City Health Care System Salt Lake City, Utah MARCH 15, 2017

2 This page intentionally left blank

3 U.S. Army Corps of Engineers Kansas City District Final Operation and Maintenance Plan for the Vapor Mitigation System Building 040-H Accelerated Operable Unit (AOU) 1 Time Critical Removal Action and Mitigation System Installation for the 700 South 1600 East PCE Plume Superfund Site VA Salt Lake City Health Care System Salt Lake City, Utah Submitted to: Prepared by: U.S. Army Corps of Engineers - Kansas City District 635 Federal Building 601 East 12 th Street Kansas City, Missouri CTI and Associates, Inc NW 87th Terrace, Ste. 210 Kansas City, Missouri CONTRACT NO. W912DQ-15-D MARCH 15, 2017 REVISION: 00

4 This page intentionally left blank

5 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH 2017 TABLE OF CONTENTS 1.0 INTRODUCTION SITE LOCATION PURPOSE OPERATIONS AND MAINTENANCE INSPECTIONS WITH ROUTINE MAINTENANCE ANNUAL MAINTENANCE INDOOR AIR SAMPLING SPARE PARTS INVENTORY FIGURES Figure 1 Site Location APPENDICES Appendix A Appendix B Appendix C Appendix D Appendix E Site Forms F1 - Inspection Form F2 - Annual Maintenance Form F3 Sample Collection Form Annotated Amaircare Homeowners Manual & Dealer Installation Instructions Damper Information Dwyer Anemometer Manual Photoionization Detector Manual CONTRACT NO. W912DQ-15-D i

6 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH 2017 This page intentionally left blank CONTRACT NO. W912DQ-15-D ii

7 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH 2017 ABBREVIATIONS AND ACRONYMS AOU COC CQCP CQCSM CTI DCN DoD DQCR ESS EDD ER ITR MCL PCE PE PG PGM PM QAPP SOW SQCM SSHP UFP SVOCs TCLP TO UDEQ USEPA USACE-NWK VIMS VHA VOCs Accelerated Operable Unit Contaminants of Concern Contractor Quality Control Plan Contractor Quality Control System Manager CTI and Associates, Inc. Document Control Number Department of Defense Daily Quality Control Report East Side Springs Electronic Data Deliverable Engineering Regulation Independent Technical Review maximum contaminate level Perchloroethylene Professional Engineer Professional Geologist Program Manager Project Manager Quality Assurance Project Plan Statement of Work Site Quality Control Manager Site Safety and Health Plan Uniform Federal Policy Semi-Volatile Organic Compounds Toxicity Characterization Leaching Procedure Task Order Utah Department of Environmental Quality U.S. Environmental Protection Agency USACE Kansas City District Vapor Intrusion Mitigation System Veteran s Healthcare Administration Volatile Organic Compounds CONTRACT NO. W912DQ-15-D iii

8 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH 2017 This page intentionally left blank CONTRACT NO. W912DQ-15-D iv

9 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH 2017 APPROVAL PAGE Operation and Maintenance Plan Accelerated Operable Unit (AOU) 1 Time Critical Removal Action and Mitigation System Installation for the 700 South 1600 East PCE Plume Superfund Site VA Salt Lake City Health Care System Salt Lake City, Utah CONTRACT NO.: W912DQ-15-D Submitted to: Department of Army Kansas City, Corps of Engineers 635 Federal Building 601 East 12 th Street Kansas City, Missouri Program Manager Christopher Winkeljohn, P.E. March 15, 2017 Date Ph: (816) Project Manager Robert Stenson, P.G. Date March 15, 2017 Ph: (920) March 15, 2017 Plan Preparer James R. Dickson, P.E. Date Ph: (920) CONTRACT NO. W912DQ-15-D v

10 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH 2017 This page intentionally left blank CONTRACT NO. W912DQ-15-D vi

11 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH INTRODUCTION Under Contract No. W912DQ-15-D-3010 with the U.S. Army Corps of Engineers Kansas City District (USACE), CTI and Associates, Inc. (CTI) has prepared the following Operation and Maintenance (O&M) Plan for the routine operations of the 040-H vapor mitigation system associated with the Accelerated Operable Unit (AOU) 1 Time Critical Removal Action and Mitigation System Installation for the 700 South 1600 East PCE Plume Superfund Site, VA Salt Lake City Health Care System Salt Lake City, Utah. This Operation and Maintenance (O&M) Plan addresses the scope of work requirement for the O&M of the vapor mitigation system in accordance with the Performance Work Statement (PWS) issued August 3, Site Location The Site, which consists of Building 040-H, is located in Salt Lake City, Utah (See Figure 1). The 040-H location is a privately owned residence and is situated within a westward sloping residential lot of the East Side Springs (ESS) area with visible seeps and springs located in the eastern and northern portions of the lot. The eastern portion of the lowest level of the house is below grade, and the western portion of the lowest level is partially below grade, with a partially below grade living space in the northwest corner of the lowest level and a walkout garage at grade in the southwest corner of the lowest level. The resident reported that a clay pipe traverses from east to west within the foundation to convey spring water and groundwater from the wetter eastern portion of the lot to the western portion of the lot where the clay pipe discharges to a small pond. Residence 040-H is located within a PCE plume that is thought to be the western extent of the 700 South 1600 East PCE Plume. Residence 040-H was identified by VA as having the potential of being adversely affected by the PCE plume through the subsurface to indoor air VI pathway (Figure 1). Seeps, springs, and groundwater in this area with detections of PCE and TCE are most likely associated with the 700 South 1600 East PCE Plume. The plume is located within the Salt Lake City alluvial basin near the Wasatch Mountain front. The Salt Lake City alluvial basin is bounded to east by the Wasatch Fault and a seismically independent segment of the Wasatch fault called East Bench Segment Fault. A drop in surface elevation due to the Wasatch fault scarp has resulted in the development of seeps and springs from groundwater daylighting between 800 South and Michigan Avenue (approximately 1300 South) and between 1000 East and 1300 East. This area is within the ESS area. The focus area for vapor intrusion (VI) lies CONTRACT NO. W912DQ-15-D

12 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH 2017 within the ESS. Residence 040-H is located in the central portion of the ESS area, south of 900 South and east of 1200 East (Figure 1). Current land use near residence 040-H includes single family residential properties with several schools, churches, and small businesses located along the major streets. The few commercial operations near the residence include an auto repair shop/gasoline station, convenience store, and coffee shop located near the intersections of 800 South and 900 South with 1300 East. CTI designed a vapor intrusion mitigation system (VIMS) that consists of an Amaircare whole house air purifier installed on the suction side of the house furnace. The unit is equipped with 36 pounds of active carbon using installed optional carbon filters available from the manufacturer which are required for use as VIMS. This equipment installation was completed on November 18, 2016 in accordance with the approved work plan and design change notice 1 (DCN-01). DCN-01 added a vacuum induced damper to the suction of the furnace to allow for additional untreated air to be drawn into the furnace if for some reason the air purifier supply is insufficient. The damper closes automatically when the ducting is not under a vacuum to allow for the air purifier to circulate treated air thought out the house supply air ducting. 1.2 Purpose The installed VIMS needs to be operated and maintained in accordance with this document to continue to provide adequate capture of contaminants that appear to intrude into the residence. The purpose of this document is to present specific details on the proper operation, maintenance and monitoring of the installed VIMS. Proper operation of the VIMS is necessary to maintain performance of the system to capture potential indoor air contaminants that may enter the basement from the 700 South 1600 East PCE plume in the vicinity of residence 040-H. A summary of work includes the following: Routine Inspections: o Visually inspect equipment for any damage/leaks or other problems; o Record the electrical meter reading for the air purifier supply; o Measure air flow at the air purifier intake, discharge to furnace, and furnace discharge; and o Measure total volatile organic compounds (VOCs) with ppbrae Photoionization Detector with 10.6 ev bulb (PID). CONTRACT NO. W912DQ-15-D

13 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH 2017 Maintenance for the Amaircare Air Purifier: o Change pre-filters -when particle buildup on the filter is observed (typically every 3-4 months based on initial year quarterly inspections); and o Change HEPA filters and carbon filters; (Every 12 months based on manufacturers recommendations for the optional carbon canisters installed). Indoor Air Sampling: o One 24-hour TO15 select ion method (SIM) sample for target analytes will be collected from each floor of residence 040-H, a duplicate sample will be collected during each sample event from a random location. Samples events will occur during each quarter during year one and semiannual for years 2 and 3 to monitor the performance of the vapor mitigation system during operations; o All analytical work shall be performed by a National Environmental Laboratory Accreditation Program (NELAP) accredited laboratory in accordance with the applicable portions of the AOU1 Quality Assurance Project Plan (QAPP) to the extent practical; and o The results will be provided to USACE PM and VA Program Manager. CONTRACT NO. W912DQ-15-D

14 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH OPERATIONS AND MAINTENANCE The Amaircare installed for vapor mitigation operates continuously by running three fans that draw cold return air from the plenum closet through pre-filters, high-efficiency particulate air (HEPA) filters, and high capacity carbon filters. The treated air is discharged to the suction of the house furnace suction which directs the treated air throughout the residence s supply ducts. The furnace/ac will continue to operate as normal during heating and cooling months and the Amaircare will act as an additional filter to the furnace. If for some reason the furnace controls draw more air than is fed to the intake by the air purifier forming a suction, additional make up air will be drawn automatically through the static pressure damper. This condition could occur if the pre-filters begin to foul over time or the furnace variable speed fan draws more air than the Amaircare can supply. 2.1 Inspections with Routine Maintenance The system will be inspected quarterly throughout the initial year of operation. The inspections will determine the actual pre-filter change out requirements. It has been assumed to be quarterly at this time. The O&M Inspection Checklist (Form 1) will be completed during every inspection event. Sequence of events: 1. Calibrate the PID and complete the instrument calibration section of the form (Field or Office) 2. Open the air purification disconnect panel located in the utility room on the wall between the furnace and the hot water heater. Record the value on the LCD of the power meter located inside. (See images above) CONTRACT NO. W912DQ-15-D

15 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH Measure the average airflow and PID readings on the inlet of the air purifier. Record the readings on the form. (See image at right) 4. Remove sample port plug and measure the average airflow and PID reading on the discharge of the Amaircare Record readings on the form. Replace the sample port plug once readings have been collected. Ensure plenum closet door is shut before proceeding to Step 5. (See image at right) 5. Remove sample port plug and measure the average airflow and PID readings on the discharge of the furnace. Record readings on the form. Replace the sample port plug once readings have been collected. (See image at right) CONTRACT NO. W912DQ-15-D

16 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH Measure PID readings near the floor in the utility room and record on the form. Ensure the plenum closet door is closed before taking measurements. 7. Measure PID Readings near the floor in the plenum closet and record on the form. 8. Return to the air purification disconnect panel and open the breaker that powers the air purifier as shown. (See image at right) 9. Proceed to the air purifier and open the filter access cover located on the front of the unit. First remove and set aside the safety screw on top. Then place fingers in the finger tabs shown and lift to remove access panel and set aside. (See image at right) CONTRACT NO. W912DQ-15-D

17 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH Carefully remove and visually inspect the prefilter of each of the three filter units for particle accumulation and fouling. (See image at right) 11. If pre-filter replacement is required, remove the filters one at a time. Turn the filter counter clockwise until the filter is released. Hold the filter by the edge, pull out and lift to remove. (See image at right) CONTRACT NO. W912DQ-15-D

18 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH Remove clips shown to remove old prefilter from the filter frame. Either dispose of used filter offsite or wash for future reuse. Place new/washed filter around the filter frame and re-affix with clips. 13. Install filter back in the air purification unit by inserting and applying pressure while turning clockwise until you feel resistance and stop. (See image at right) 14. Repeat steps 11 and 12 until filters requiring change-out have been changed. 15. Replace the access panel using the finger tabs and secure with the safety screw. 16. Return to the air purification disconnect panel and turn the breaker back on to re-energize the air purification system. 17. Visually inspect the system to ensure operations have been restored. Complete the forms to document observations made and actions taken. In addition, a sticker will be affixed to the front of the unit during the first inspection with lines and columns for date, work performed, performed by and signature to allow for onsite documentation of work. Additional stickers will be added as necessary. CONTRACT NO. W912DQ-15-D

19 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH 2017 Following the inspection, the completed forms shall be transmitted to CTI for review and approval. Once approved, the forms will be forwarded to the designated VHA Representative for records retention purposes, with a copy provided to the USACE PM. 2.2 Annual Maintenance In accordance with the manufacturers recommendations, the entire composite filter will be changed on an annual basis. This includes the pre-filter, HEPA filter and carbon filter units. This filter change-out package is available from Allergy Buyers Club ( The specific package is Amaircare Whole House VOC Annual Filter Replacement Kit Part AM AP VOCC. This kit contains 9 pre-filters and 3 VOC canisters (each containing 12 pounds of carbon to be used as a set). The nine pre-filters will allow for up to two additional pre-filter change outs as routine maintenance between annual maintenance events. Sequence of events: 1. Open the air purification disconnect panel located in the utility room on the wall between the furnace and the hot water heater. Record the value on the LCD of the power meter located inside. (See images above). Once the value has been recorded, open the disconnect on the left side of the meter shown to power off the unit. CONTRACT NO. W912DQ-15-D

20 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH Proceed to the air purifier and open the filter access cover located on the front of the unit. First remove and set aside the safety screw on top. Then place fingers in the finger tabs shown and lift to remove access panel and set aside. (See image at right) 3. The filters are located inside as show to right. CONTRACT NO. W912DQ-15-D

21 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH Remove the filters one at a time. Turn the filter counter clockwise until the filter is released. Hold the filter by the edge, pull out and lift to remove. (See image at right) 5. Remove clips and remove old pre-filter from filter frame and dispose of offsite. Carefully remove the endcap shown at right. Place endcap on new HEPA filter with new O-ring provided. Dispose of old O-rings off site with old filters. A diagram of the filter and installed carbon canister is shown below: CONTRACT NO. W912DQ-15-D

22 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH The HEPA filer come with a small cloth carbon filter installed. This inner carbon filter must be removed and dispose of properly before installing the optional carbon canister required for VOC removal. The optional carbon canisters are required for this system. Remove the shrink wrap from the new optional carbon canister and insert it into the new HEPA filter as shown (the soft inner carbon filter discussed above must be removed or it will jam). 7. When properly installed the carbon canister should fit flush as shown at the right. 8. Place new pre-filter around the filter and affix with clips as shown at right. Place the end cap on using the new o-rings provided. 9. Install filter back in the air purification unit by inserting and applying pressure while turning clockwise until you feel resistance and stop. 10. Repeat until all three filter units have been changed. Replace the access panel using the finger tabs and secure with the safety screw. 11. Return to the air purification disconnect panel and turn the breaker back on to re-energize the air purification system. CONTRACT NO. W912DQ-15-D

23 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH Visually inspect the system to ensure operations have been restored. Complete the forms to document observations made and actions taken. Documentation of work performed will be completed on the sticker installed on the front of the unit. Additional stickers will be added as necessary. Following the filter change, the completed form(s) shall be transmitted to CTI for review. The completed form(s) will be reviewed for accuracy and completeness by CTI and then forwarded to the designated VHA representative for records retention purposes, with a copy provided to the USACE PM. 2.3 Indoor Air Sampling System performance monitoring will be completed using indoor air sampling to confirm that the VIMS is meeting the Tier 1 removal action levels (RALs) for PCE and its daughter product TCE throughout the 040-H residence. Indoor air sampling frequency will consist of quarterly samples during year one and semi-annual sampling for years 2 and 3. Samples will be collected from each floor of 040-H, with a random duplicate sample for quality control purposes. The indoor air samples will be collected in the two locations of the highest values observed during the initial characterization of the house. These two identified locations were the lower level bedroom (TCE concentration up to 5.4J ug/m 3 [RAL 2.1 ug/m 3 ] and PCE concentrations up to 78J ug/m 3 [RAL 41 ug/m 3 ]), and the breakfast nook in the kitchen located above the crawlspace on the main floor (TCE concentration of 4.3 ug/m 3 [RAL 2.1 ug/m 3 ] and PCE concentrations of 59J ug/m 3 [RAL 41 ug/m 3 ]). An additional monitoring location was established for the third floor during preliminary sampling event and will be used moving forward to be consistent. The third floor location is the southern bedroom. All locations are shown in the following images. All samples will be collected in individually cleaned and certified 6 liter SUMMA canisters equipped with 24 hour regulators from NELAP accredited laboratory in accordance with the 2016 quality assurance project plan (QAPP) to the extent practical. Each sample will be analyzed by EPA Method TO-15 SIM with a target analyte list of tetrachloroethylene (PCE), trichloroethylene (TCE), cis-1,2-dichloroethylene CONTRACT NO. W912DQ-15-D

24 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH 2017 (CDCE), vinyl chloride (VC), and 1,4-dioxane. The sample canisters will be picked up and delivered to ALS Environmental Laboratory with chain of custody completed by Sage Environmental. Basement Floor Sample Location (F1): F1 Main Floor Sample Lopcation (F2): Front F2 Front Top Floor Sample Locations (F3): CONTRACT NO. W912DQ-15-D

25 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH 2017 F3 Front As identified in the QAPP, one of the three pre-set locations identified above will be duplicated per sample event and labeled as F4 to provide a blind duplicate to the laboratory. The attached air sampling form will be completed for each sampling event to document the following: Sample ID; Location/Description, including which location sample F4 duplicates; Summa Canister Number; Start Date; Start Time; Starting Vacuum; Stop Date; Stop Time; CONTRACT NO. W912DQ-15-D

26 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH 2017 Ending Vacuum; and Comments if Applicable. Sample nomenclature shall be as follows: First Floor Sample location F1: A 040H IA 001 mmddyy F1 (where mm is two-digit month of sample collection, dd is the two-digit day of sample retrieval and yy is two-digit year of sample collection); Second Floor Sample location F2: A 040H IA 002 mmddyy F2 (as above); Third Floor Sample location F3: A 040H 003 mmddyy F3 (as above); and The duplicate collected from a random floor adjacent to normal sample is F4: A 040H 004 mmddyy F4 (as above and actual location recorded on the air sample form). When the data is provided by the laboratory, CTI will review and compare to the identified Tier 1 RALs for the target analytes discussed above. In the unlikely event that an exceedance is observed the USACE PM and VHA will be notified immediately and a plan of action will be developed and implemented to remedy the situation. The most likely action anticipated is change out of the carbon filter package with a follow-up indoor air sample event. The sample results will be provided in a brief letter report that summarizes the sampling activities, provides tabulated data compared to applicable tier 1 RALs, and provides recommendations as appropriate. The actual laboratory level IV report and a validated Lab Spec 7 electronic data deliverable (EDD) will be appended to the letter report in electronic format.. A letter report will be provided to USACE PM and VA Program Manager within 20 calendar days after sample results are received from the laboratory. CONTRACT NO. W912DQ-15-D

27 PRE-DRAFT O&M PLAN VA SALT LAKE CITY HEALTH CARE SYSTEM SALT LAKE CITY, UTAH MARCH Spare Parts Inventory In order to have the ability to rapidly respond to system issues, the following spare parts are to be maintained at the local support contractor s office/storage area. Item 1: Amaircare VOC Filter Kit with a minimum single complete set of filters: o o o 3-Pre-Filters (new or washed); 3- HEPA Filters with O-rings (new); and 3- Optional VOC (12 pound) Carbon Filters (new). CONTRACT NO. W912DQ-15-D

28 This page intentionally left blank

29 FIGURES

30 This page intentionally left blank

31 Based on FIGURE 1 SITE MAP from the Action Memorandum for Residence 0040-H within Accelerated Operable Unit 1: East Side Springs 700 South 1600 East PCE Plume, Salt Lake City, Utah By: Figure 1 Date: JULY 13, 2017 Site Location and Groundwater Plume Map CTI and Associates, Inc. Vapor Mitigation System Design Building 40-H AOU-1 Salt Lake City, Utah

32 This page intentionally left blank

33 APPENDIX A O&M FORMS F1- INSPECTION LOG F2- ANNUAL MAINTENNCE LOG F-3 SAMPLE COLLECTION LOG

34 This page intentionally left blank

35 F1- Inspection Form Residence 040-H Vapor Mitigation System Operation and Maintenance Instrument Calibration Fresh Air Zero: 0 ppm Span Gas: 10 ppm Isobutalyene Zero/Span Calibration Completed: Date: Time: Performed By: Print: Sign: Current Conditions Weather: Outside Temperature: Notes: SYSTEM LOG Air Purifier Electrical Use Meter Reading: Airflow (Air Purifier Inlet): PID (Air Purifier Inlet): Airflow (Air Purifier Discharge): PID (Air Purifier Discharge): Airflow (Furnace Discharge): PID (Furnace Discharge): PID (Utility Room Floor): PID (Plenum Closet Floor): KW-HR CFM ppb CFM ppb CFM ppb ppb ppb Unit De-enegized: Unit Re-enegized: Comments: Time: Top Pre-FilterInspected: Middle Pre-Filter Inspected: Bottom Pre-Filter Inspected: Time: Changed Clean Notes Yes / No Yes / No Yes / No Yes / No Yes / No Yes / No

36 This page intentionally left blank

37 F2-Annual Maintenance Form Residence 040-H Vapor Mitigation System Operation and Maintenance SYSTEM LOG Air Purifier Electrical Use Meter Reading: KW-HR Unit De-enegized: Time: Top Pre-Filter: Top Pre-Filter HEPA: Top Large (Optional) Carbon: Top Filter O-Ring: Middle Pre-Filter: Middle Pre-Filter HEPA: Middle Large (Optional) Carbon: Middle Filter O-Ring: Bottom Pre-Filter: Bottom Pre-Filter HEPA: Bottom Large (Optional) Carbon: Bottom Filter O-Ring: Changed Yes / No Yes / No Yes / No Yes / No Yes / No Yes / No Yes / No Yes / No Yes / No Yes / No Yes / No Yes / No Notes Unit Re-enegized: Time: Comments: Performed By: Print: Sign:

38 This page intentionally left blank

39 SAGE Environmental, LLC Air Sampling Form Project: Vapor Mitigation System Operations and Sampling Location: Bldg 40 H Sampler Name: Sampling Method: Summa cannisters/ TO-15 SIM 5 TAL Sample ID Location/Description Summa # Regulator # Start Date Start Time Starting Vacuum (in. Hg) Stop Date Stop Time Ending Vacuum (in. Hg) Comments / Notes

40 This page intentionally left blank

41 APPENDIX B ANNOTATED FOR AMAIRCARE HOMEWOWNERS MANUAL & DEALER INSTALATION INSTRUCTIONS

42 This page intentionally left blank

43 2004 Americair Corporation, Mississauga, Ontario, Canada AIR CLEANERS/ FILTERS AMAIRCARE HEPA BYPASS AIR FILTRATION SYSTEM HOMEOWNERS MANUAL & DEALER INSTALLATION INSTRUCTIONS FOR MODELS AWW-350, AWW-675, 8500 & Rules for Safe Installation and Operation READ AND SAVE THESE INSTRUCTIONS! Please read instructions before installing and using the HEPA Bypass Air Filtration System (HEPA system). This will help you obtain the full benefit of the HEPA system you have selected. It will also help you to avoid needless service costs. 1. Read this manual carefully. Failure to follow these rules and instructions could cause a malfunction of the air filter or unsatisfactory service and could void your warranty. 2. Follow a regular service and maintenance schedule to ensure efficient operation. 3. For safety and optimized performance of your HEPA system, all installation and maintenance must be performed by a professional heating and ventilation contractor. The installer should be made aware of your indoor air quality situation and be familiar with your heating, ventilation and air conditioning equipment. 4. High particulate distribution may occur during initial start-up of this product after installation or after scheduled filter changes. Individuals who are highly sensitive to airborne particulates should not be in the building and allow 24 hours of operation for removal of particulates from occupied spaces before re-entry.! WARNING Risk of property damage, injury or death. Installation, adjustments, alterations, service and maintenance must be performed by a qualified technician. Risk of Carbon Monoxide Poisoning. Can cause injury or death.! WARNING Do not operate equipment without access panel in place. Operation of this equipment without all access panels in place may cause gas fumes from the heating system to be drawn into occupied spaces. Shipping and Packing List Package 1 of 1 contains: 1 - HEPA System 1 - Complete Filter Set (Packed inside Unit) 1 - Installation Instructions (this manual) 1 - Registration Card 1 - Filter Changes Reminder Label 4 - Self Adhesive Rubber Feet (AWW Models only) HEPA System Models Model Airwash Whisper-350 (AWW-350) is designed to filter air up to a rate of 350 cfm. Model Airwash Whisper-675 (AWW-675) is designed to filter air up to a rate of 675 cfm. Model 8500 is designed to filter air up to a rate of 850 cfm. Model is designed to filter air up to a rate of 1000 cfm.! WARNING Electrical Shock Hazard. Can cause injury or death. Disconnect all electrical power supplies before servicing. Do not operate equipment without access panels in place

44 Table of Contents Rules for Safe Installation and Operation...1 Shipping and Packing List...1 Specifications...2 Introduction...2 General Information...3 Product Application Guidelines...3 Parts Identification Physical Dimensions of Units Dealer Installation Instructions Operation...11 Maintenance...12 Filter Change Schedule...12 Filter Changing Guidelines...12 Dealer Filter Change Instructions Dealer Motor Assembly Replacement Instructions...15 Electrical Diagrams Replacement Parts...18 Specifications AWW-350 AWW Nom. Air 0.0 E.S.P. 350 cfm 675 cfm 850 cfm 1000 cfm Weight (max - unpackaged) 30 lbs. 33 lbs. 125 lbs. 130 lbs. Supply Voltage 120 VAC 120 VAC 120 VAC 120 VAC Nom. Measured Power Consumption 125 Watts 225 Watts 360 Watts 600 Watts Motor Current Draw - Total 1.0 amps 1.8 amps 3.6 amps 4.8 amps Air Intake Collar (max) 8 -round 10 -round 10 x12 10 x12 Air Outflow Collar (max) 6 -round 8 -round 10 x12 10 x12 Operating Temp. Range - Return air (ºF) 30 to to to to 95 HEPA filter efficiency = 0.3 micron particles Ambient temperature range: -40ºF to 130ºF For 220 VAC Units (only sold outside N. America) Nom. Measured Power Consumption 75 Watts 150 Watts 255 Watts 300 Watts Motor Current Draw - Total.34 amps.7 amps 2.0 amps 1.2 amps Introduction Congratulations! You will quickly realize that you have purchased a very effective air cleaning system. It incorporates state of the art HEPA (High Efficiency Particulate Air) technology. Your Amaircare HEPA system comes with a limited warranty. With proper attention to its care and maintenance, you will receive optimum performance. If your indoor air has abnormally high concentrations of particulates, the life span of the filter media may be shortened. Excessive particulates in the air will reduce the expected life of the HEPA filter. Under normal conditions the HEPA filter will last from two to five years. The optional activated carbon canister has a finite limit as to the amount of odor or other gaseous volatile organic compounds (V.O.C. s) that it can adsorb. The higher the concentrations, the shorter the expected life. Higher humidity may shorten the life of the carbon canister. Under normal conditions, the carbon canister should last up to 12 months. Initially, the HEPA filter s particulate removal efficiency improves with use. However, the air flow through the HEPA filter media will decrease, as dust builds up on the filter, thus lowering its ability to circulate and clean as much air as when it was new. If the air flow through your unit is noticeably reduced, you can inspect the pre-filter and HEPA filter to see if they should be replaced. Replacement (or washing if foam) of the prefilter every 3 months will help extend the life of the HEPA filter. Please contact your local Amaircare dealer regarding replacement of filter media, warranty information or if you have any questions or concerns about the performance of your HEPA system. NOTE: This filtration system is an ADDITIONAL filter, and does NOT replace the existing air handler/furnace system filter. 2

45 General Information Media Filters Media filters strain particulates from the air. The filter media needs to have tiny holes to allow air to pass through, but not particulates. Filter types vary for all sorts of purposes. The most effective and proven filter media is HEPA. HEPA filter media is 99.97% efficient at capturing particles which are 0.3 micron in size or larger. Carbon Filters Carbon media is used to capture chemicals and odors, also referred to as V.O.C. s. Chemicals and odors cannot be captured with media type filters. Activated carbon collects chemicals and odors in a process called adsorption. If air passes through the carbon filter before particulates are removed, the surface of the carbon quickly gets covered with particulates, rendering it ineffective at capturing chemicals and odors. If particulates are removed from the air with a HEPA filter, virtually the entire surface area of the carbon can be used to capture chemicals and odors. This increases the efficiency and filter life of the carbon filter. Separate HEPA and Carbon Filters HEPA and carbon filters have different life spans. It is important for a filtration system to keep the filter components independent from each other, so that the filter that is dirty/saturated can be changed. This is more economical than a system where you need to throw out two or three filters when only one needs changing. The HEPA System s 3 Stage Filtration Process The 3 stage filtration process is used to create a very effective filtration system. Each filter is independent and can be changed individually. Stage 1: Pre-filter - The inexpensive pre-filter (foam or carbon) removes larger particulates from the air, thus prolonging the life of the HEPA filter. Stage 2: HEPA - The HEPA filter removes 99.97% of particulates 0.3 micron and larger. The cleaned air then passes through the third stage filter. Stage 3: Carbon - The carbon filter is about 1/2 an inch thick to give it plenty of surface area for removing chemicals and odors from the air. This filter may be replaced by an optional heavy duty granular carbon canister available for maximum removal of chemicals and odors. The clean air is then reintroduced into the air you breathe. Product Application Guidelines * Chart based on homes with 8 ft. ceilings. Size of House vs. Air Changes per Hour Size of House* Model 1,000 ft 2 1,200 ft 2 1,500 ft 2 1,800 ft 2 2,000 ft 2 2,500 ft 2 3,000 ft 2 3,500 ft 2 Unit (8,000 ft 3 ) (9,600 ft 3 ) (12,00 ft 3 ) (14,400 ft 3 ) (16,000 ft 3 ) (20,000 ft 3 ) (24,000 ft 3 ) (28,000 ft 3 ) Model AWW Model AWW Model Model Notes: Industry experience indicates that one (1) air change per hour generally provides adequate air cleaning. Actual results will depend on multiple factors such as outdoor particulate levels, infiltration rate, indoor activities etc. Generally speaking, the more air changes per hour provided, the more effective a HEPA system will be. People with sensitivities may desire a higher number of air changes per hour for cleaner air. 3

46 Parts Identification AWW-350 & AWW-675 Cabinet Parts Filter Section Access Panel Inside Cabinet (Filter Section) Retaining Clips (2) Intake Collar Cabinet Outflow Collar Power Module Model CM71 Motor/Impeller Filter Adapter (threaded) O Ring Motor Section Access Panel Power Cord figure 1. figure 2. Parts Identification 8500 & Cabinet Parts Inside Cabinet (Filter Section) 10 x12 Outflow Duct Variable Speed Switch 10 x12 Intake Duct Filter Section Access Panel Motor Section Access Panel Finger Tabs (4) Cabinet Filter Adapter (3) Motor/Impeller (3) Filter Retaining Rod Clip (6) HEPA Cartridge (3) Filter Cap (3) Filter Retaining Rod (3) figure 3. figure 4. 4

47 Filter Parts Identification HEPA Cartridge Parts - AWW-350 & AWW-675 Optional Carbon Canister - AWW-350 & AWW-675 HEPA Filter Pre-Filter - Foam HEPA Filter Mesh (Protects HEPA Filter) Inner Carbon Filter For third stage increased removal of chemicals and odors. Located inside the HEPA filter. Discard inner carbon filter when using the optional carbon canister. figure 5. HEPA Cartridge Parts & figure 6. Optional Carbon Canister & Inner Carbon Filter HEPA Filter Pre-Filter - Carbon HEPA Filter Mesh (Protects HEPA Filter) For third stage increased removal of chemicals and odors. Located inside the HEPA filter. Discard inner carbon filter when using the optional carbon canister. figure 7. figure 8. 5

48 Physical Dimensions of Unit Model AWW-350 & AWW-675: Filter Section Access Panel Air Intake Air Outflow A F H Motor Section Access Panel A C B I B A E D figure 9. G J MODEL NO. A B C D E F G H I J Model AWW (368) (692) (200) (149) (406) (32) (32) (76) (216) (51) Model AWW (368) (692) (251) (200) (406) (32) (32) (51) (165) (25) Dimensions in inches (mm) 6

49 Physical Dimensions of Unit Models 8500 & 10000: A E H D I D J G G Air Intake Air Outflow A C F B Filter Section Access Panel figure 10. Motor Section Access Panel MODEL NO. A B C D E F G H I J Models 8500 & (438) (711) (1213) (254) (254) (1276) (64) (66) (135) (3) Dimensions in inches (mm) 7

50 Dealer Installation Instructions Forced air handler/furnace systems: The HEPA system should be installed as a bypass system, with part of the return ducted into the HEPA system. The filtered air is then rerouted back into the return air, and continues through the system to be heated/cooled. Typical Return to Return Application For homes with upflow forced air handler/furnace systems.(air handler/furnace is shown in a typical basement. See A Figure 6.) Distance between A and B should be 6' to 16' for best results B 8 Typical Return to Return Application For homes with horizontal forced air handler/furnace systems. (Air handler/furnace is shown in an typical attic. See figure 7.) Air Handler/ Furnace Air Handler/Furnace Filter Distance between A and B should be 6' to 16' for best results HEPA System Figure 11. Figure 12. NOTE: This filtration system is an ADDITIONAL filter, and does NOT replace the existing air handler/furnace system filter. Preparation: Here are some things to consider as you decide where to install the HEPA system. Location: Make sure there is room to open the HEPA filter access panel for filter changes/inspections. Keep the HEPA system in a location where you can still access the air handler/furnace filter. Keep the HEPA system away from possible water damage. Vibration pads will reduce vibration for installations where the unit is placed on the floor. Install HEPA System on floor or suspended platform. If the unit is suspended, screws must not penetrate through the cabinet. Make sure that you have the proper chains/straps/joists and equipment to keep unit secure. Intake (Marked as Air In on unit): Intake ducts should be installed upstream of any humidifiers and be installed on the main return. Intake duct should be installed at least 6 ft. away from the outflow duct on the main return. Outflow (Marked as Clean Air Out on unit): Outflow duct should be installed as close to the air handler/furnace inlet as possible but not directly into the return air elbow of the main return. If the unit is being installed independently of any other system, room diffusers are recommended to help distribute airflow evenly in the occupied space. Ducting: If HEPA system is installed where inlet and outflow collars face down, metal elbows must be connected to both inlet and outflow collars. Each connection must be sealed with aluminum tape or mastic, including all take offs. Installed duct runs should be as straight as possible (if the duct runs are too long, reduced CFM may result). If duct is exposed to unconditioned air, externally insulated flex duct is highly recommended. Externally insulated flex duct can also be used for noise reduction purposes. For best indoor air quality, do not use ductboard or fiberglass inside of ducts. Electricity: The unit must be plugged into a grounded 120V, 60Hz outlet. Required Materials for Installation of Unit: Items for AWW-350 Flex or rigid duct 8 round & 6 round (length as required) Takeoffs One 8 & One 6 Items for AWW-675 Flex or rigid duct 10 round & 8 round (length as required) Takeoffs One 10 & One 8 Items for 8500 & Flex or rigid duct 10 x 12 (length as required) Takeoffs Two 10 x12 All Models Aluminum tape or mastic as required Misc. hanging materials - field provided NOTE: Be sure to review Rules for Safe Installation and Operation on page 1 of this document before start-up of this unit. B A

51 Dealer Installation Instructions Forced air handler/furnace system with an HRV/ERV This application provides filtration of all Outdoor Air (OA) that is brought into the home through the HRV/ERV, thus reducing the introduction of dust, pollen and mold from the outdoor air. For systems that have HRV/ERV units installed, we recommend ducting the fresh air outflow from the HRV/ERV into the HEPA system. If the HRV/ERV CFM (cubic feet per minute) rating is lower than that of the HEPA system, an additional return needs to be installed into the HEPA system (See figure 8.) The outflow air from the HEPA system then needs to be installed into the main return of the air handler/furnace system. Ensure that you follow the proper installation instructions as outlined in the HRV/ERV Installation manual(s). Preparation: Here are some things to consider as you decide where to install the HEPA system with an HRV/ERV. Location: Make sure there is room to open the HEPA filter access panel for filter changes/inspections. Keep the HEPA system in a location where you can still access the air handler/furnace filter. Keep the HEPA system away from possible water damage Vibration pads will reduce vibration for installations where the unit is placed on the floor. Install HEPA System on floor or suspended platform. If the unit is suspended, screws must not penetrate through the cabinet. Make sure that you have the proper chains/straps/joists and equipment to keep unit secure. Intake (Marked as Air In on unit): Most HRV/ERV systems will not move as much air as the HEPA system. For these systems, install an additional return from another treated air source into the HEPA system. If using an additional return duct, it should be installed upstream of any humidifiers. The fresh air outflow of the HRV/ERV should be "Y" connected to the additional return duct then connected to the inlet of the HEPA system. The additional return duct (if any) should be installed at least 6 ft. away from the outflow duct on the main return. Duct both the HRV/ERV and the additional return into the intake of the HEPA system. Outflow (Marked as Clean Air Out on unit): Outflow duct should be installed as close to the air handler/furnace inlet as possible but not directly into the return air elbow of the main return. Ducting: If HEPA system is installed where inlet and outflow collars face down, metal elbows must be connected to both inlet and outflow collars. Each connection must be sealed with aluminum tape or mastic, including all take offs. Installed duct runs should be as straight as possible (if the duct runs are too long, reduced CFM may result). If duct is exposed to unconditioned air, externally insulated flex duct is highly recommended. Externally insulated flex duct can also be used for noise reduction purposes. For best indoor air quality, do not use ductboard or fiberglass inside of ducts. Electricity: The unit must be plugged into a grounded 120V, 60Hz outlet. Required Materials for Installation of Unit: Items for AWW-350 Flex or rigid duct 8 round & 6 round (length as required) Takeoffs One 8 & One 6 Items for AWW-675 Flex or rigid duct 10 round & 8 round (length as required) Takeoffs One 10 & One 8 Items for 8500 & Flex or rigid duct 12 round or 10 x 12 (length as required) 9 Figure 13. Example of Return to Return Installation with an HRV/ERV For homes with a forced air handler/furnace system and an HRV/ERV system. Additional Return Air Handler/Furnace Air Handler/Furnace Filter HEPA System NOTE: This filtration system is an ADDITIONAL filter, and does NOT replace the existing air handler/furnace system filter. Takeoffs Two 10 x12 All Models Aluminum tape or mastic as required Misc. hanging materials - field provided NOTE: Be sure to review Rules for Safe Installation and Operation on page 1 of this document before start-up of this unit.

52 Dealer Installation Instructions Independent Operation: The HEPA systems can be used independently of any other equipment! The intake and outflow of the filtration system can be ducted into the same room to create a cleaner environment almost anywhere. The intake or outflow can also be ducted elsewhere. The intake and outflow should be installed on opposite sides of the room; however, this varies according to your specific needs. Preparation: Here are some things to consider as you decide where to install the HEPA system independently of other systems. Location: Make sure there is room to open the HEPA filter access panel for filter changes/inspections. Keep the HEPA system away from possible water damage Vibration pads will reduce vibration for installations where the unit is placed on the floor. B Install HEPA System on floor or suspended platform. If the unit is suspended, screws must not penetrate through the cabinet. Make sure that you have the proper chains/straps/joists and equipment to keep unit secure. Intake (Marked as Air In on unit): Intake ducts should be installed near the floor for optimum airflow (see figure 11). If space does not allow, then the inlet can be installed in the ceiling. Intake duct should be installed at least 6 ft. away from the outflow duct. Intake duct should be installed at opposite end of the room from the outflow duct(s) if in the same room. Diffusers are recommended to help distribute airflow evenly. Outflow (Marked as Clean Air Out on unit): Outflow(s) should be installed in the ceiling away from any other air inlet(s) Room diffusers are recommended to help distribute airflow evenly in the occupied space. Ducting: If HEPA system is installed where inlet and outflow collars face down, metal elbows must be connected to both inlet and outflow collars. Each connection must be sealed with aluminum tape or mastic, including all vent connections. Installed duct runs should be as straight as possible (if the duct runs are too long, reduced CFM may result). If duct is exposed to unconditioned air, insulated flex duct is highly recommended. Externally insulated flex duct can also be used for noise reduction purposes. For best indoor air quality, do not use ductboard or fiberglass inside of ducts. Example of Single Room Stand Alone System Installation For single rooms where increased filtration is desired such as a dedicated smoking room. HEPA System A Figure 14. Air should be drawn from location B if space is available, otherwise, use location A. Electricity: The unit must be plugged into a grounded 120V, 60Hz. outlet. Required Materials for Installation of Unit: Items for AWW-350 Flex or rigid duct 8 round & 6 round (length as required) Takeoffs One 8 & One 6 Diffusers One 8 & One 6 Items for AWW-675 Flex or rigid duct 10 round & 8 round (length as required) Takeoffs One 10 & One 8 Diffusers One 10 & One 8 Items for 8500 & Flex or rigid duct 10 x 12 (length as required) Takeoffs Two 10 x12 All Models Aluminum tape or mastic as required Misc. hanging materials - field provided NOTE: Be sure to review Rules for Safe Installation and Operation on page 1 of this document before start-up of this unit. 10

53 Operation (8500, 10,000 & AWW Units with on/off toggle switch) 1. Make sure that the unit is plugged into a grounded outflow (120 Volt, 60 Hz). 2. For optimum performance, the HEPA system should operate when the indoor air handler/furnace blower is on. 3. Turn the unit on by rotating the variable speed dial clockwise. The unit starts in High speed, turn all the way clockwise for low speed. For units with a n on/off toggle switch, press the switch to the on (1) position. 4. To turn the unit off, rotate the variable speed dial counter-clockwise until you feel or hear it click into the off position. For units with an on/off toggle switch, press the switch to the off (0) position. Operation (AWW-350 & AWW-675) 1. Make sure that the unit is plugged into a grounded power source (120 Volt, 60 Hz). 2. For optimum performance, the HEPA system should operate with the indoor air handler/furnace blower on. 3. Turn the unit on by pressing the on/off button and the green power light will illuminate. For units with a n on/off toggle switch, press the switch to the on (1) position. 4. To turn the unit off, press the on/off button and the power indicator light will go out. For units with an on/off toggle switch, press the switch to the off (0) position. Filter Light Indicators Your HEPA system is equipped with an electronic filter timer. This timer monitors the life expectancy of each filter in the unit. By default, the Pre-Filter, HEPA Filter and VOC Blanket lights will turn on when the unit is powered on. 5. If you have upgraded your VOC Blanket to a VOC Canister, you must activate the corresponding timer. Press and hold the VOC Canister button until the light comes on. The VOC Blanket light will automatically go off. The control panel will now monitor the life expectancy of the VOC Canister. 6. If you are switching your VOC Canister to a VOC Blanket, you must activate the corresponding timer. Press and hold the VOC Blanket button until the light comes on. The VOC Canister light will automatically go off. The control panel will now monitor the life expectancy of the VOC Blanket. When a filter indicator light flashes, it is time to check and replace that filter. 7. Turn the unit off and see instructions for filter removal on pg. 14. to determine if filter replacement is needed. See Maintenance on next page. 8. Once the filters have been checked or replaced, turn the unit back on and reset the control panel. To reset, press and hold the flashing filter button until the light is constant. 11

54 Maintenance Proper care and maintenance of your HEPA system will ensure years of service. The unit must be turned off during service/maintenance or when filters are being changed. It is recommended that gloves and a filtered breathing mask be worn during filter replacement. Risk of Sharp Edges Hazard.! WARNING Electrical Shock Hazard. Can cause injury or death. Disconnect all electrical power supplies before servicing. Do not operate equipment without access panels in place! CAUTION Equipment sharp edges can cause injuries. Avoid grasping equipment edges without protective gloves. Filter Change Schedule Note: Note: Failure to properly maintain your HEPA system will decrease the efficiency and air flow. Pre-Filter: 3 to 4 months* HEPA Filter: 2 to 5 years Inner Carbon Filter: 6 months Optional Carbon Canister: 12 months * Foam pre-filters (AWW-350 & AWW-675) can be washed and reused NOTE: Filter life is based on average air content. Some filters may need to be changed more often due to higher amounts of dust, humidity, or chemicals found in your ambient air. Additionally, people who are more sensitive to these airborne contaminates may desire more frequent filter changes. Filter Changing Guidelines Pre-Filter: Dust and other large particles will collect on the pre-filter over time. The color of the filter will change as particulates build up on the pre-filter. Change the pre-filter when you can see the particulate build up start to clog up the pre-filter. HEPA Filter: As the HEPA filter captures particulates, it will darken over time. Replace the HEPA filter when it darkens to the level seen in example D. A. New B. Used C. Used D. Replace Inner Carbon Filter: The inner carbon filter will rarely look used. This filter captures odors and gasses, yet the filter s appearance will not change. When this filter has reached it s maximum adsorbancy of odors and gasses, it will no longer work. Replace this filter when it no longer seems to capture odors, or every 3 months (12 months for optional Carbon Canister), which ever occurs first. Pre-Filter HEPA Filter Inner Carbon Filter Figure

55 Dealer Filter Change Instructions - AWW-350 & AWW-675 Risk of Sharp Edges Hazard.! WARNING Electrical Shock Hazard. Can cause injury or death. Disconnect all electrical power supplies before servicing. Do no operate equipment without access panels in place! CAUTION Equipment sharp edges can cause injuries. Avoid grasping equipment edges without protective gloves. It is recommended that gloves and a filtered breathing mask be worn during filter replacement to avoid breathing particulates (dust, mold, pollen, etc.) captured on the filter that become airborne during the filter(s) changeout. The old filters should be wrapped and sealed in plastic bags immediately upon removal from the unit to avoid distributing particles throughout the house during the process of disposal. 1. Accessing the filters a. Remove safety screw(s) from HEPA filter access panel. b. Unlatch the two retaining clips and lift off HEPA filter access panel. c. Bracing the unit so it does not move, turn the HEPA cartridge counter-clockwise and lift/pull out. 2. Pre-filter Replacement a. Pull the pre-filter up and off the unit. NOTE: The pre-filter may contain contaminants, remove it slowly to avoid releasing particles back into the air. b. Foam pre-filters can be washed several times, wash by hand in warm water. Let it dry completely before placing it back onto the HEPA cartridge. b. If replacing the filter, remove plastic shrink wrap from the new pre-filter. c. Stretch the new/washed filter around the top of the HEPA cartridge and slide it down into place 3. Inner Carbon Filter Replacement a. Look inside the HEPA cartridge to locate the two ends of the inner carbon filter. b. Pull one end of the old inner carbon filter in and bend it into a loose roll so it can be removed. c. Remove the inner carbon filter from the HEPA cartridge. d. Remove plastic shrink wrap from the new inner 13 carbon filter. e. Unroll the inner carbon filter and roll it up in the opposite direction (this makes the filter follow a more contoured profile against the inner HEPA filter surfaces and helps keep it in place), place the rolled inner carbon filter inside the HEPA cartridge and gently unroll it until the ends butt together and the filter is snug against the HEPA filter. 4. HEPA Filter Replacement a. If replacing the HEPA filter with a new filter, discard old HEPA filter and use new when replacing the HEPA filter into the unit. b. With each annual filter replacement kit, a new o ring is provided. The old one is removed by pinching it between two fingers and pulling it off the collar on the blower deck. c. Discard old o ring. d. Place the new o ring onto the collar and slide it down to the base of the blower deck. 5. Optional Carbon Canister a. Remove old carbon canister (if installed) by pulling it out from the inside of the HEPA filter. b. If replacing an inner carbon filter with the carbon canister, remove inner carbon filter by following the steps a. to c. in section 3. c. Remove the plastic shrink wrap from the new carbon canister. d. Slide the carbon canister into the HEPA cartridge, smaller end first. The carbon canister should slide all the way in until the metal edges at the base meet the HEPA filter. e. Support the carbon canister with your fingers so it does not slide out when replacing the HEPA cartridge assembly into the unit. 7. Installing the HEPA Filter Cartridge a. With the filters changed or inspected, all 3 filters are ready to be placed back into the unit. Place the HEPA cartridge gently into the unit (if a carbon canister is being used, take care not to let it slide out as it is heavy and could damage the unit) b. When the HEPA cartridge is in place, brace the unit, press down and gently turn it clockwise to lock it into place. If too much force is used, the cartridge may be difficult to remove! c. Replace the HEPA filter access panel. For AWW-350 & AWW-675, latch it with the two retaining clips. d. Re-install safety screw(s) into HEPA filter access panel. e. Plug the unit back into a power outlet and turn it on.

56 Dealer Filter Change Instructions & Risk of Sharp Edges Hazard.! WARNING Electrical Shock Hazard. Can cause injury or death. Disconnect all electrical power supplies before servicing. Do no operate equipment without access panels in place! CAUTION Equipment sharp edges can cause injuries. Avoid grasping equipment edges without protective gloves. It is recommended that gloves and a filtered breathing mask be worn during filter replacement to avoid breathing particulates (dust, mold, pollen, etc.) captured on the filter that become airborne during the filter(s) changeout. 3. Inner Carbon Filter Replacement a. Look inside the HEPA cartridge to locate the two ends of the inner carbon filter. b. Pull one end of the old inner carbon filter in and bend it into a loose roll so it can be removed. c. Remove the inner carbon filter from the HEPA cartridge. d. Remove plastic shrink wrap from the new inner carbon filter. e. Unroll the inner carbon filter and roll it up in the opposite direction (this makes the filter follow a more contoured profile against the inner HEPA filter surfaces and helps keep it in place), place the rolled inner carbon filter inside the HEPA cartridge and gently unroll it until the ends butt together and the filter is snug against the HEPA filter screen. 4. HEPA Filter Replacement a. If replacing the HEPA filter with a new filter, discard old HEPA filter and use new when replacing the HEPA filter into the unit. The old filters should be wrapped and sealed in plastic bags immediately upon removal from the unit to avoid distributing particles throughout the house during the process of disposal. 1. Accessing the filters a. Remove safety screw(s) from HEPA filter access panel. b. Pull the HEPA filter access panel up by the finger tabs and lift out to remove. c. Lift the filter retaining rod out from the two clips holding it into position (some force may be necessary as it is a tight fit to ensure complete seal), ensure that pressure is kept on the filter cap as it is the only thing keeping the HEPA cartridge in place. d. With the retaining rod removed, pull the HEPA cartridge and filter cap out of the unit. 2. Pre-filter Replacement a. Locate the clips holding the pre-filter in place. Remove them and pull the pre-filter off. NOTE: The pre-filter may contain contaminants, remove it slowly to avoid releasing particles back into the air. b. Remove plastic shrink wrap from the new pre-filter. c. Wrap the new pre-filter around the HEPA filter, making sure that the ends overlap. d. With the clips provided with the new pre-filter, secure the pre-filter ends together tightly so it keeps the pre-filter secure against to the HEPA filter. e. Be sure that all clips are in place. 5. Optional Carbon Canister a. Remove old carbon canister (if installed) by pulling it out from the inside of the HEPA filter. b. If replacing an inner carbon filter with the carbon canister, remove inner carbon filter by following the steps a. to c. in section 3. c. Remove the plastic shrink wrap from the new carbon canister. d. Slide the carbon canister into the HEPA cartridge. The carbon canister should slide all the way in meeting the edges of the HEPA cartridge evenly on both ends. e. Keep the HEPA cartridge horizontal and support it with your fingers to keep it from sliding out when replacing the HEPA cartridge back into the unit. 7. Installing the HEPA Filter Cartridge a. With the filters changed or inspected, all 3 filters are ready to be placed back into the unit. Place the HEPA cartridge gently into the unit (if a carbon canister is being used, take care not to let it slide out as it is heavy and could damage the unit) b. When the HEPA cartridge is in place, place the filter cap onto the HEPA cartridge and replace the filter retaining rod by sliding it into it s two clips. (some force may be necessary as it is a tight fit to ensure complete seal) c. Replace the HEPA filter access panel. d. Re-install safety screw(s) into HEPA filter access panel. e. Plug the unit back into a power outflow and turn it on. 14

57 Dealer Motor Assembly Replacement Instructions Risk of Sharp Edges Hazard.! WARNING Electrical Shock Hazard. Can cause injury or death. Disconnect all electrical power supplies before servicing. Do no operate equipment without access panels in place Do not use this fan with any solid-state speed control device! CAUTION Equipment sharp edges can cause injuries. Avoid grasping equipment edges without protective gloves. 1. Accessing the motor assembly a. Turn the unit off and unplug it from any electrical source before opening the cabinet. b. Remove the safety screw(s) from the motor section access panel. c. Lift the door up by the finger tabs and pull out to remove it from the unit. 3. Installing the new motor assembly a. Slide the new motor into the motor mount making sure that the wires go through the smaller hole offset from the center of the bracket. b. Secure the new motor to the motor mount with the four screws removed in step 2c. d. Connect the wires as follows: - Blue wire from motor to the on/off switch. - Yellow/green wire from motor to the ground post. - Brown wire from the motor to a capacitor post. - Black wire from the motor to the other capacitor post. - White wire from the on/off switch to the capacitor via the black wire piggyback post. - White wire from the power cord to the capacitor via the other white wire s piggyback post. 4. Closing the unit a. Replace the motor section access panel and secure it with the screws removed in step 1b. b. Plug the unit into it s electrical source and turn it on. 2. Removing the old motor assembly a. Disconnect all four motor wires from switch, ground post and capacitor. b. Disconnect the two white wires from the capacitor. c. Separate the motor from the motor mount by removing the four screws found in figure 19. d. Slide the motor out from under the motor mount to remove it from the unit. Figure

58 Electrical Diagram (AWW-350 & AWW-675) Figure 17. Electrical Diagram (AWW-350 & AWW-675 with on/off toggle switch) Figure

59 Electrical Diagram (8500 & 10,000) Figure

60 Replacement Parts Replacement Parts for AWW-350 & AWW-675 (Qty.) Canadian U.S. International Standard Annual Filter Kit (1 Pre-filter, 2 Carbon) Plus Annual Filter Kit (1 Pre-filter, 1 VOC Canister) HEPA, Easy Twist Filter Motor Assembly - AWW-350 (120V) Motor Assembly - AWW-350 (220V) Motor Assembly - AWW-675 (120V) Motor Assembly - AWW-675 (220V) Replacement Parts for 8500 & 10,000 (Qty.) Canadian U.S. International Standard Annual Filter Kit (9 Pre-filters, 6 Carbon) Plus Annual Filter Kit (9 Pre-filters, 3 VOC Canisters) HEPA, Moulded Filter (1) Motor Assembly (120V) (1) Motor Assembly (220V) (1) Motor Assembly - 10,000 (120V) (1) Motor Assembly - 10,000 (220V) (1) Use this unit only in the manner intended by the manufacturer. If you have questions, contact Amaircare. Contact your local Amaircare dealer to order replacement parts. For the Amaircare dealer near you, dial or visit us at 18

61 APPENDIX C DAMPER INFORMATION

62 This page intentionally left blank

63

64

65

66

67 APPENDIX D INSTRUMENTATION MANUALS DWYER ANEMEOMETER INFORMATION

68 This page intentionally left blank

69 Model 471B Digital Thermo Anemometer Specifications - Installation and Operating Instructions Bulletin TE-471B 11/32 [8.74] PS/2 INPUT 29/32 [23.00] SLIDING COVER PS/2 CONNECTOR 2-13/16 [71.59] 16-11/64 [410.61] 6-11/64 [156.92] 6-9/64 [155.82] 6-31/32 [177.13] PROBE CABLE APPROX. 2 LG COILED APPROX. 6 LG EXTENDED 1-5/32 [29.36] The MODEL 471B Digital Thermo Anemometers are versatile dual function instruments that quickly and easily measure air velocity or volumetric flow plus air temperature in imperial or metric units. High contrast LCD display shows both selected readings simultaneously. Convenient backlight provides perfect visibility in low light conditions. Light automatically shuts off after 2-1/2 minutes to prolong battery life. Low battery warning is included. Stainless steel probe with comfortable hand grip is etched with insertion depth marks from 0 to 8 inches and 0 to 20 cm. Extruded aluminum housing fully protects electronics, yet is lightweight and comfortable to hold even when taking multiple readings as part of duct traverses. Up to 99 readings may be stored for later retrieval. An integral sliding cover protects sensors when not in use. Items included with the 471B are 9 volt alkaline battery, sensing probe, wrist strap and custom carrying case. Battery Installation The unit is shipped with a separate 9 volt alkaline battery which must be installed before operation. Remove the two screws holding the bottom endcap in place and remove it. Connect the battery to the enclosed battery clip, observing correct polarity. Be careful not to trap the wires between the case or foam pad which retains the battery. This could make it difficult to install the battery or remove it later for replacement. Be sure the rubber gasket is properly seated in the gasket channel and replace the endcap. Note that the endcap will only fit one way because the holes are slightly offcenter. Place the Z shaped wrist strap clip in one of the screw recesses and replace the screws. Do not overtighten. Attach the wrist strap to the clip. When battery replacement becomes necessary, use only 9 volt alkaline type batteries such as a Duracell MN1604, or Eveready 522 or equivalent. Zinc-carbon types, often labeled Heavy-Duty, are not recommended because of their shorter life and an increased potential for leakage. Alkaline batteries are a better value because they last typically up to three times longer in this device. Low Battery Indicator A weak battery may cause improper operation or inaccurate measurements. A low battery indicator ( ) is provided on the display to show when the battery needs to be replaced. Although the unit may appear to function and read properly, the accuracy of the readings cannot be guaranteed when the ( ) indicator is empty of power segments. Replace the battery with a fresh one. Do not leave an exhausted battery in the unit due to potential battery leakage. NOTICE The side of the probe with dot should face the flow. SPECIFICATIONS Service: Air velocity and temperature of clean, dry air. Temperature Limits: Process Air Velocity: -20 to 212 F (-29 to 100 C); Process Temperature: -40 to 212 F (-40 to 100 C); Ambient: 5 to 125 F (-15 to 51 C). Display: 4.5 digit LCD. Resolution: 0.1%, 0.1 F/ C. Range Air Velocity: 0 to 6000 FPM (0 to 30 m/s). Accuracy Air Velocity: ±3% FS within temperature range of 40 to 90 F (4 to 32 C). Range Volumetric Air Flow: 19,999 in selected flow units. Range Temperature: -40 to 212 F (-40 to 100 C). Accuracy Temperature: ±0.5 F (±0.28 C) from 32 to 122 F (0 to 50 C); ±1.5 F (±0.83 C) from -40 F to 32 F (-40 to 0 C) & 122 to 212 F (50 to 100 C). Probe Length: 8 (203 mm) insertion. Cable Length: 28 (71 cm) retracted, 6 ft (183 cm) extended. Power Requirements: 9V alkaline battery included. Weight: 16 oz (454 g). Agency Approvals: CE. OPERATING INSTRUCTIONS ON/OFF/LIGHT A single press turns unit on, and a second press turns unit off. While unit is on, hold for 2 seconds to turn on the backlight. If the light is active, press once to turn the light off and another to turn the unit off. HOLD A single press freezes the current reading and a second press resumes live process reading. MENU A single press will display the menu. Hit MENU while in the main menu to go back to the home screen or while in the sub-menus to back up one level. ENTER When in the home screen, press to cycle through displayed values. While in the menu, press to enter sub-menus or values. Press and hold for 4 seconds to clear MAX, MIN, or MEM in their respective menus. MEMORY/UP A singles press will retrieve stored data points in the memory. Each successive press will scroll to the next saved memory location ST01, ST02, etc. While in the menu, it becomes the UP button. STORE/DOWN A single press will store the current reading on the screen. ST01, ST02, etc. shows in the display to indicate the memory location where the reading was saved. While in the menu, it becomes the DOWN button. DWYER INSTRUMENTS, INC. P.O. BOX 373 MICHIGAN CITY, INDIANA 46360, U.S.A. Duracell is a registered trademark of The Gillette Company Eveready is a registered trademark of Eveready Battery Company, Inc. Phone: 219/ Fax: 219/ info@dwyermail.com

70 HOLD Engineering Units HOLD will appear on the home screen when the HOLD button is pressed, indicating the process reading has been frozen. Memory Storage and Retrieval Press ENTER to see current engineering units. Press MEMORY/UP or STORE/ DOWN to select another unit. Area From the home screen, push MEMORY/UP to view all previously stored values. Cycle up or down through the stored data with the MEMORY/UP or STORE/DOWN buttons. From the home screen, press STORE/DOWN to store data points to memory. For both data storage and data memory retrieval, press MENU to return to the home screen. MENUS Maximum Press ENTER to view the current duct shape setting. rec for rectangular or CIrCL for circular will appear. To alter between the two, push the MEMORY/UP button or STORE/DOWN buttons. If a circle duct shape is selected, press ENTER to see DIAM the diameter value. If rectangle is chosen, ENTER will allow adjustment of the XdIM and YdIM of the duct. A numeral position will blink indicating it can be adjusted with the MEMORY/UPor STORE/DOWN buttons. Hit ENTER to store a value. Note: In metric mode dimensions will be displayed in mm. Inches will be the value when using imperial units. Power Management Push ENTER to see the maximum recorded reading. When viewing the maximum recorded reading, hold ENTER for 4 seconds to clear. Press MEMORY/UP or STORE/ DOWN to cycle through parameters. Pu sh ENTER to see current time off setting. Select 30 or 10 values indicating minutes for power to automatically shut-off. By selecting 0, the auto shut-off is disengaged and the unit will only turn off manually by the ON/OFF/LIGHT button. Minimum Factory Restore Push ENTER to see the minimum recorded reading. When viewing the minimum recorded reading, hold ENTER for 4 seconds to clear. Press MEMORY/UP or STORE/ DOWN to cycle through parameters. This function allows the user to restore the unit to factory settings. Press ENTER to view the restore function. CLr will be shown on the display. Press and hold ENTER for 4 seconds until CLrd is displayed. The unit will then be restored to factory defaults. Average Press ENTER to view the AVG function. This function will adjust the process reading on the home screen to report the average reading over the period of selected time. Press MEMORY/UP or STORE/DOWN to select between 0, 30, or 60 second values. Push ENTER to store the desired value. MAINTENANCE/REPAIR Upon final installation of the Model 471B, no routine maintenance is required. The Model 471B is not field serviceable and should be returned if repair is needed. Field repair should not be attempted and may void warranty. WARRANTY/RETURN Refer to Terms and Conditions of Sales in our catalog and on our website. Contact customer service to receive a Return Goods Authorization number before shipping the product back for repair. Be sure to include a brief description of the problem plus any additional application notes. Memory This product uses FreeRTOS ( version A copy of the original Free RTOS source shall be provided upon request. Press ENTER to see CLr. Hold ENTER for 4 seconds to clear all stored memory locations. Copyright 2015 Dwyer Instruments, Inc. Printed in U.S.A. 10/15 FR# Rev. 5 DWYER INSTRUMENTS, INC. P.O. BOX 373 MICHIGAN CITY, INDIANA 46360, U.S.A. Phone: 219/ Fax: 219/ info@dwyermail.com

71 APPENDIX E INSTRUMENTATION MANUALS PHOTOIONIZATION DETECTOR (PID) MANUAL

72 This page intentionally left blank

73 ppbrae 3000 User s Guide Rev. E February 2016 P/N

74 FCC Information Contains FCC ID: PI4411B or SU3RM900 The enclosed device complies with part 15 of the FCC rules. Operation is subject to the following conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation. Wireless Approval For UAE In Middle East TRA REGISTERED No: ER36153/14 or ER36153/15 DEALER No.: HONEYWELL INTERNATIONAL MIDDLE EAST LTD DUBAI BR Wireless Approval For QATAR In Middle East ictqatar Type Approval Reg. No.: R-4466 or R-4635 Copyright 2016 RAE Systems by Honeywell

75 ppbrae 3000 User s Guide Contents Read Before Operating... 5 Special Notes... 6 Warnings... 7 Standard Contents... 9 General Information... 9 Physical Description Specifications Charging The Battery Charging A Spare Rechargeable Battery Low Voltage Warning Clock Battery Data Protection While Power Is Off User Interface Display Operating The Instrument Turning The Instrument On Turning The Instrument Off Operating The Built-In Flashlight Pump Status Calibration Status Bump Status Policy Enforcement Operating Modes Basic User Level/Hygiene Mode (Default Settings) Alarm Signals Alarm Signal Summary Preset Alarm Limits & Calibration Testing The Alarm Integrated Sampling Pump Backlight Datalogging Datalogging event Datalogging sample Auto/Manual/Snapshot Datalogging Standard Kit & Accessories AC Adapter

76 ppbrae 3000 User s Guide Alkaline Battery Adapter External Filter Optional Accessories Standard Two-Point Calibration (Zero & Span) Entering Calibration Zero (Fresh Air) Calibration Span Calibration Exiting Two-Point Calibration In Basic User Level Three-Point Calibration Span 2 Calibration Exiting Three-Point Calibration Bump Test Programming Mode Entering Programming Mode Programming Mode Menus Calibration Zero Calibration Span Calibration Measurement Meas. Gas Meas. Unit Alarm Setting High Alarm Low Alarm STEL Alarm TWA Alarm Alarm Type Buzzer & Light Datalog Clear Datalog Interval Data Selection Datalog Type Manual Datalog Snapshot Datalog Monitor Setup Op Mode Site ID User ID

77 ppbrae 3000 User s Guide User Mode Basic Advanced Date Time Duty Cycle Temperature Unit Pump Speed Language Radio Power Real Time Protocol Power On Zero Unit ID LCD Contrast Lamp ID PAN ID Mesh Channel Mesh Interval Hygiene Mode Basic User Level & Hygiene Mode Entering Search Mode From Hygiene Mode Advanced User Level (Search & Hygiene Mode) Advanced User Level & Hygiene Mode Basic User Level & Search Mode Advanced User Level & Search Mode Diagnostic Mode Entering Diagnostic Mode Adjusting The Pump Stall Threshold Pump High Pump Low Exiting Diagnostic Mode Transferring Data To & From A Computer Downloading The Datalog To A PC Uploading Firmware To The instrument From A PC Maintenance Battery Charging & Replacement Replacing The Li-ion Battery Replacing The Alkaline Battery Adapter PID Sensor & Lamp Cleaning/Replacement

78 ppbrae 3000 User s Guide Sensor Components Cleaning The PID Sensor Cleaning The Lamp Housing Or Changing The Lamp Determining The Lamp Type Sampling Pump Cleaning The Instrument Ordering Replacement Parts Special Servicing Note Troubleshooting Technical Support RAE Systems Contacts Controlled Part Of Manual

79 ppbrae 3000 User s Guide Read Before Operating This manual must be carefully read by all individuals who have or will have the responsibility of using, maintaining, or servicing this product. The product will perform as designed only if it is used, maintained, and serviced in accordance with the manufacturer s instructions. The user should understand how to set the correct parameters and interpret the obtained results. CAUTION! To reduce the risk of electric shock, turn the power off before removing the instrument cover. Disconnect the battery before removing sensor module for service. Never operate the instrument when the cover is removed. Remove instrument cover and sensor module only in an area known to be non-hazardous. 5

80 ppbrae 3000 User s Guide Special Notes When the instrument is taken out of the transport case and turned on for the first time, there may be some residual organic or inorganic vapor trapped inside the detector chamber. The initial PID sensor reading may indicate a few ppm. Enter an area known to be free of any organic vapor and turn on the instrument. After running for several minutes, the residual vapor in the detector chamber will be cleared and the reading should return to zero. The battery of the instrument discharges slowly even if it is turned off. If the instrument has not been charged for 5 to 7 days, the battery voltage will be low. Therefore, it is a good practice to always charge the instrument before using it. It is also recommended to fully charge the instrument for at least 10 hours before first use. Refer to this User Guide s section on battery charging for more information on battery charging and replacement. 6

81 ppbrae 3000 User s Guide WARNINGS STATIC HAZARD: Clean only with damp cloth. For safety reasons, this equipment must be operated and serviced by qualified personnel only. Read and understand instruction manual completely before operating or servicing. Use only RAE Systems battery packs, part numbers , , and This instrument has not been tested in an explosive gas/air atmosphere having an oxygen concentration greater than 21%. Substitution of components may impair intrinsic safety. Recharge batteries only in non-hazardous locations. Do not mix old and new batteries or batteries from different manufacturers. The calibration of all newly purchased RAE Systems instruments should be tested by exposing the sensor(s) to known concentration calibration gas before the instrument is put into service. For maximum safety, the accuracy of the instrument should be checked by exposing it to a known concentration calibration gas before each day s use. Do not use USB/PC communication in hazardous locations. 7

82 ppbrae 3000 User s Guide AVERTISSEMENT DANGER RISQUE D'ORIGINE ELECTROSTATIQUE: Nettoyer uniquement avec un chiffon humide. Pour des raisons de sécurité, cet équipment doit être utilisé, entretenu et réparé uniquement par un personnel qualifié. Étudier le manuel d instructions en entier avant d utiliser, d entretenir ou de réparer l équipement. Utiliser seulement l'ensemble de batterie RAE Systems, la reference au au Cet instrument n a pas été essayé dans une atmosphère de gaz/air explosive ayant une concentration d oxygène plus élevée que 21%. La substitution de composants peut compromettre la sécurité intrinsique. Ne charger les batteries que dans emplacements désignés non-dangereuse. Ne pas melanger les anciennes et les nouvelles batteries, ou bien encore les batteries de differents fabriquants. La calibration de toute instruments de RAE Systems doivent être testé en exposant l instrument a une concentration de gaz connue par une procédure diétalonnage avant de mettre en service l instrument pour la première fois. Pour une securite maximale, la sensibilité du l instrument doit être verifier en exposant l instrument a une concentration de gaz connue par une procédure diétalonnage avant chaque utilisation journalière. Ne pas utiliser de connection USB/PC en zone dangereuse. 8

83 ppbrae 3000 User s Guide Standard Contents Instrument Calibration Kit Charging Cradle AC/DC Adapter Alkaline Battery Adapter Data Cable CD-ROM With User s Guide, Quick Start Guide, and related materials General Information The compact instrument is designed as a broadband VOC gas monitor and datalogger for work in hazardous environments. It monitors Volatile Organic Compounds (VOC) using a photoionization detector (PID) with a 9.8 ev or 10.6 ev gas-discharge lamp. Features are: Lightweight and Compact Compact, lightweight, rugged design Built-in sample draw pump Dependable and Accurate Up to 16 hours of continuous monitoring with rechargeable battery pack Designed to continuously monitor VOC vapor at parts-perbillion (ppb) levels User-friendly Preset alarm thresholds for STEL, TWA, low- and high-level peak values. Audio buzzer and flashing LED display are activated when the limits are exceeded. Datalogging Capabilities 260,000-point datalogging storage capacity for data download to PC 9

84 ppbrae 3000 User s Guide The instrument consists of a PID with associated microcomputer and electronic circuit. The unit is housed in a rugged case with a backlit LCD and 3 keys to provide easy user interface. It also has a built-in flashlight for operational ease in dark locations. 10

85 ppbrae 3000 User s Guide Physical Description The main components of the portable VOC monitoring instrument include: Three keys for user to interact with the instrument: 3 operation/programming keys for normal operation or programming LCD display with back light for direct readout and calculated measurements Built-in flashlight for illuminating testing points in dark environments Buzzer and red LEDs for alarm signaling whenever exposures exceed preset limits Charge contacts for plugging directly to its charging station Gas entry and exit ports USB communication port for PC interface Protective rubber cover Specifications Size: Weight: Detector: Battery: Battery Charging: Operating Hours: Display: 9.25" L x 3.6" W x 2.9" H 28 oz with battery pack Photoionization sensor with 9.8eV or 10.6eV UV lamp A 3.7V rechargeable Lithium-Ion battery pack (snap in, field replaceable, at nonhazardous location only) Alkaline battery holder (for 4 AA batteries) Less than 8 hours to full charge Up to 16 hours continuous operation Large dot matrix screen with backlight 11

86 ppbrae 3000 User s Guide Measurement range & resolution Lamp Range Resolution 10.6 ev 1 ppb to 10,000 ppm 1 ppb 9.8 ev 0.01 ppm to 5,000 ppm 10 ppb Response time (T 90 ): Accuracy (Isobutylene): PID Detector: Correction Factors: Calibration: Calibration Reference: Inlet Probe: Radio module: 2 seconds 10 to 2000 ppm: 3% at calibration point. Easy access to lamp and sensor for cleaning and replacement Over 200 VOC gases built in (based on RAE Systems Technical Note TN-106) Two-point field calibration of zero and standard reference gases Store up to 8 sets of calibration data, alarm limits and span values Flexible 5" tubing Bluetooth (2.4GHz) or RF module (433MHz, 868MHz, 915MHz, or 2.4GHz) Keypad: 1 operation key and 2 programming keys; 1 flashlight switch Direct Readout: Intrinsic Safety: Instantaneous, average, STEL, TWA and peak value, and battery voltage US and Canada: Class I, Division 1, Groups A, B, C, D Europe: ATEX (0575 Ex II 2G Ex ia IIC/IIB T4 Gb) KEMA 07 ATEX 0127 Complies with EN :2009, EN :2007 IECEx CSA Ex ia IIC/IIB T4 Gb Complies with IEC :2007, 12

87 ppbrae 3000 User s Guide EM Interference: Alarm Setting: Operating Mode: Alarm: Alarm Type: Real-time Clock: Datalogging: Communication: Wireless Network: IEC :2006 (IIC: Li-ion bat pack or NiMH bat pack; IIB: alkaline bat pack) Highly resistant to EMI/RFI. Compliant with EMC R&TTE (RF Modules) Separate alarm limit settings for Low, High, STEL and TWA alarm Hygiene or Search mode Buzzer (95dB at 30cm) and flashing red LEDs to indicate exceeded preset limits, low battery voltage, or sensor failure Latching or automatic reset Automatic date and time stamps on datalogged information 260,000 points with time stamp, serial number, user ID, site ID, etc. Upload data to PC and download instrument setup from PC via USB on charging station. Mesh RAE Systems Dedicated Wireless Network (or WiFi network for WiFiequipped instruments) Sampling Pump: Internally integrated. Flow rate: 450 to 550 cc/min. Wireless Frequency: Modulation: RF Power (Tx): ISM license-free band, 902 to MHz and 915 to 928 MHz, FCC Part 15, CE R&TTE, IEEE b/g bands (2.4 GHz) DSSS BPSK 10dBm Temperature: -20º C to 50º C (-4º to 122º F) Humidity: 0% to 95% relative humidity (noncondensing) 13

88 ppbrae 3000 User s Guide Housing (including rubber boot): Polycarbonate, splashproof and dustproof Battery can be changed without removing rubber boot. 14

89 ppbrae 3000 User s Guide Charging The Battery Always fully charge the battery before using the instrument. The instrument s Li-ion battery is charged by placing the instrument in its cradle. (The battery can also be charged by placing the instrument in an AutoRAE 2 Cradle.) Contacts on the bottom of the instrument meet the cradle s contacts, transferring power without other connections. Note: Before setting the instrument into its charging cradle, visually inspect the contacts to make sure they are clean. If they are not, wipe them with a soft cloth. Do not use solvents or cleaners. Follow this procedure to charge the instrument: 1. Plug the AC/DC adapter s barrel connector into the instrument s cradle. 2. Plug the AC/DC adapter into the wall outlet. 3. Place the instrument into the cradle, press down, and lean it back. It locks in place and the LED in the cradle glow The instrument begins charging automatically. The Primary LED in the cradle blinks green to indicate charging. During charging, the diagonal lines in the battery icon on the instrument s display are animated and you see the message Charging... When the instrument s battery is fully charged, the battery icon is no longer animated and shows a full battery. The message Fully charged! is shown. The cradle s LED glows continuously green. 15

90 ppbrae 3000 User s Guide Note: If you see the Battery Charging Error icon (a battery outline with an exclamation mark inside), check that the instrument or rechargeable battery has been set into the cradle properly. If you still receive the message, check the Troubleshooting section of this guide. Note: If the instrument or battery has been in the cradle for more than 10 hours and you see the Battery Charging Error icon and a message that says, Charging Too Long, this indicates that the battery is not reaching a full charge. Try changing the battery and make sure the contacts between the instrument (or battery) are meeting the cradle. If the message is still shown, consult your distributor or RAE Systems Technical Services. Charging A Spare Rechargeable Battery A rechargeable Li-ion battery can be charged when it is not inside the monitor. The charging cradle is designed to accommodate both types of charging. Contacts on the bottom of the battery meet the contacts on the cradle, transferring power without other connections, and a spring-loaded capture holds the battery in place during charging. 1. Plug the AC/DC adapter into the monitor s cradle. 2. Place the battery into the cradle, with the gold-plated contacts on top of the six matching charging pins. 3. Plug the AC/DC adapter into the wall outlet. The battery begins charging automatically. During charging, the Secondary LED in the cradle blinks green. When charging is complete, it glows steady green. Release the battery from the cradle by pulling it back toward the rear of the cradle and tilting it out of its slot. Note: If you need to replace the Li-ion battery pack, replacements are available from RAE Systems. The part number is

91 ppbrae 3000 User s Guide Note: An Alkaline Battery Adapter (part number ), which uses four AA alkaline batteries (Duracell MN1500), may be substituted for the Li-Ion battery. WARNING! To reduce the risk of ignition of hazardous atmospheres, recharge and replace batteries only in areas known to be non-hazardous. Remove and replace batteries only in areas known to be nonhazardous. Low Voltage Warning When the battery s charge falls below a preset voltage, the instrument warns you by beeping once and flashing once every minute, and the empty battery icon blinks on and off once per second. You should turn off the instrument within 10 minutes and either recharge the battery by placing the instrument in its cradle, or replace the battery with a fresh one with a full charge. Clock Battery An internal clock battery is mounted on one of the instrument s printed circuit boards. This long-life battery keeps settings in memory from being lost whenever the Li-ion battery or alkaline batteries are removed. This backup battery should last approximately five years, and must be replaced by an authorized RAE Systems service technician. It is not user-replaceable. Data Protection While Power Is Off When the instrument is turned off, all the current real-time data including last measured values are erased. However, the datalog data is preserved in non-volatile memory. Even if the battery is disconnected, the datalog data will not be lost. 17

92 ppbrae 3000 User s Guide User Interface The instrument s user interface consists of the display, LEDs, an alarm transducer, and four keys. The keys are: Y/+ MODE N/- Flashlight on/off The LCD display provides visual feedback that includes the reading, time, battery condition, and other functions. Display LEDs and flashlight Y/+ key N/- key MODE key Flashlight on/off key In addition to their labeled functions, the keys labeled Y/+, MODE, and N/- act as soft keys that control different parameters and make different selections within the instrument s menus. From menu to menu, each key controls a different parameter or makes a different selection. 18

93 ppbrae 3000 User s Guide Three panes along the bottom of the display are mapped to the keys. These change as menus change, but at all times the left pane corresponds to the [Y/+] key, the center pane corresponds to the [MODE] key, and the right pane corresponds to the [N/-] key. Here are three examples of different menus with the relationships of the keys clearly shown: 19

94 ppbrae 3000 User s Guide Display The display shows the following information: Gas info Tells the Correction Factor and type of calibration gas Reading Concentration of gas as measured by the instrument Calibration needed Indicates that calibration should be performed Calibration (or bump) Dark icon indicates that calibration should be needed performed; light icon indicates bump should be performed Radio power Indicates whether radio (Mesh wireless or Bluetooth) connection is on or off Radio signal Indicates signal strength in 5-bar bargraph Battery Indicates battery level in 3 bars Pump Indicates that pump is working Datalog Indicates whether datalog is on or off Y/+ Y/+ key s function for this screen MODE MODE key s function for this screen N/- N/- key s function for this screen 20

95 ppbrae 3000 User s Guide Operating The Instrument The instrument is designed as a broadband VOC gas monitor and datalogger for work in hazardous environments. It gives real-time measurements and activates alarm signals whenever the exposure exceeds preset limits. Prior to factory shipment, the instrument is preset with default alarm limits and the sensor is pre-calibrated with standard calibration gas. However, you should test the instrument and verify the calibration before the first use. After the instrument is fully charged and calibrated, it is ready for immediate operation. Turning The Instrument On 1. With the instrument turned off, press and hold [MODE]. 2. When the display turns on, release the [MODE] key. The RAE Systems logo should appear first. (If the logo does not appear, there is likely a problem and you should contact your distributor or RAE Systems Technical Support.) The instrument is now operating and performs self tests. If any tests (including sensor and memory tests fail), refer to the Troubleshooting section of this guide. Once the startup procedure is complete, the instrument shows a numerical reading screen with icons. This indicates that the instrument is fully functional and ready to use. 21

96 ppbrae 3000 User s Guide Turning The Instrument Off 1. Press and hold the Mode key for 3 seconds. A 5-second countdown to shutoff begins. 2. Once the countdown stops, the instrument is off. Release the Mode key. 3. When you see Unit off... release your finger from the [MODE] key. The instrument is now off. Note: You must hold your finger on the key for the entire shutoff process. If you remove your finger from the key during the countdown, the shutoff operation is canceled and the instrument continues normal operation. Operating The Built-In Flashlight The instrument has a built-in flashlight that helps you point the probe in dark places. Press the flashlight key to turn it on. Press it again to turn it off. Note: Using the flashlight for extended periods shortens the battery s operating time before it needs recharging. Pump Status IMPORTANT! During operation, make sure the probe inlet and the gas outlet are free of obstructions. Obstructions can cause premature wear on the pump, false readings, or pump stalling. During normal operation, the pump icon alternately shows inflow and outflow as shown here: 22

97 ppbrae 3000 User s Guide During duty cycling (PID lamp cleaning), the display shows these icons in alternation: If there is a pump failure or obstruction that disrupts the pump, you will see this icon blinking on and off: If you see this blinking icon, consult the Troubleshooting section of this guide. Calibration Status The instrument displays this icon if it requires calibration: Calibration is required (and indicated by this icon) if: The lamp type has been changed (for example, from 10.6 ev to 9.8 ev). The sensor has been replaced. It has been 30 days or more since the instrument was last calibrated. If you have changed the calibration gas type without recalibrating the instrument. Bump Status The instrument displays this icon if it requires a bump test: 23

98 ppbrae 3000 User s Guide A bump test is required (and indicated by this icon) if: The defined period of time between bump tests has been exceeded (bump test overdue). The sensor has failed a previous bump test. The sensor(s) should be challenged on a periodic basis. Policy Enforcement The ppbrae 3000 can be configured to enforce a facility/company s requirements that calibration and/or bump testing be performed at specified intervals, and to explicitly prompt the user that calibration/bump testing is required. Depending on how Policy Enforcement features are configured, the user may be required to perform a bump test or calibration prior to being able to use the instrument. That is, it can be set to not allow normal operation of the instrument unless calibration or bump testing is performed. If the instrument has been bump tested and calibrated in compliance with the policy settings, a check-mark icon is included along the top of the ppbrae screen: If Policy Enforcement is enabled, then after startup the ppbrae displays a screen that informs the user that the instrument requires either a bump test or a calibration. If both are required, then they are shown in sequence. Note: Policy enforcement features are disabled by default. Setting Policy Enforcement You must use ProRAE Studio II to make changes to Policy Enforcement settings. You must use an AutoRAE 2 Cradle, a ppbrae 3000 Travel Charger, or a ppbrae 3000 Desktop Cradle. Policy violations are captured in the datalog. 24

99 ppbrae 3000 User s Guide Using The Travel Charger, Desktop Charger, or AutoRAE 2 Automatic Test And Calibration System To program a ppbrae 3000 via an AutoRAE 2, you need ProRAE Studio II Instrument Configuration and Data Management Software, the AutoRAE 2 connected to a power source, and a USB PC communications cable. 1. Connect a USB cable between a PC with ProRAE Studio II and the AutoRAE 2 Cradle, Travel Charger, or Desktop Cradle. 2. Apply power to the AutoRAE 2 Cradle, Travel Charger, or Desktop Cradle. 3. Turn off the ppbrae 3000 (or put the ppbrae 3000 into AutoRAE 2 Mode or Communication Mode) and set it in the cradle. 4. Start ProRAE Studio II software on the PC. 5. Select Administrator and input the password (the default is rae ). 6. Click Detect the instruments automatically (the magnifying glass icon with the letter A in it). After a few seconds, the AutoRAE 2 Cradle is found and it is shown, along with its serial number. 7. Click on the icon to highlight it, and then click Select. 8. In ProRAE Studio II, the instrument or AutoRAE 2 Cradle is shown, including its Serial Number, under Online. 9. Expand the view to show the instrument or to show the instrument in the AutoRAE 2 Cradle by clicking the + to the left of the image of the AutoRAE 2 Cradle. 10. Double-click on the icon representing the PpbRAE Click Setup. 12. In the menu that now appears on the left side, click Policy Enforcement. It is highlighted, and the Policy Enforcement pane is shown. For Must Calibrate and Must Bump, you have the options of no enforcement or enforcement (including Can t Bypass, and Can Bypass ). 25

100 ppbrae 3000 User s Guide Must Calibrate. The user is prompted to calibrate the instrument when calibration is due (as set by the calibration interval). There are two programmable options: Can t Bypass. Unless calibration is performed, the instrument cannot be used, and the only option is to turn off the instrument. Can Bypass. If calibration is due but the user does not want to perform a calibration, the instrument can still be used. In this case, the instrument records that the user has bypassed the calibration requirement in a Policy Violation report. Must Bump. The user is prompted to bump test the instrument when a bump test is due (as set by the bump test interval). There are two programmable options: Can t Bypass. Unless a bump test is performed, the instrument cannot be used, and the only option is to turn off the instrument. Can Bypass. If a bump test is due but the user does not want to perform one, the instrument can still be used. In this case, the instrument records that the user has bypassed the bump testing requirement in a Policy Violation report. These are the screens that are shown on a ppbrae 3000 after startup if Can Bypass is selected: 26

101 ppbrae 3000 User s Guide If Can t Bypass is selected, the display looks like this, and only allows the options of performing the test or shutting down: 16. Once you have made your selections in ProRAE Studio II, you must upload the changes to the instrument. Click the icon labeled Upload all settings to the instrument. 17. A confirmation screen is shown. Click Yes to perform the upload, or No to abort. Uploading takes a few seconds, and a progress bar is shown. You can abort the upload by clicking Cancel. 18. Exit ProRAE Studio II. 19. Press [Y/+] on the ppbrae 3000 to exit Communication Mode. 27

102 ppbrae 3000 User s Guide Operating Modes Your instrument operates in different modes, depending on the model and its factory default settings. In some cases, you can change modes using a password and using the instrument s navigation. In other cases, you must use ProRAE Studio software. The default setting for your instrument is: User Level: Basic Operation Mode: Hygiene This is outlined in detail on page 83. The other options, covered later in this guide, are: User Level: Advanced (page 86) Operation Mode: Hygiene User Level: Advanced (page 90) Operation Mode: Search Using ProRAE Studio allows access to other options. In addition, Diagnostic Mode (page 92) is available for service technicians. 28

103 ppbrae 3000 User s Guide Basic User Level/Hygiene Mode (Default Settings) The instrument is programmed to operate in Basic User Level/Hygiene Mode as its default. This gives you the most commonly needed features while requiring the fewest parameter adjustments. Pressing [N/-] steps you from one screen to the next, and eventually return to the main display. If you do not press a key within 60 seconds after entering a display, the instrument reverts to its main display. Note: While viewing any of these screens, you can shut off your instrument by pressing [MODE]. 29

104 Note: Dashed line indicates automatic progression. ppbrae 3000 User s Guide 30

105 ppbrae 3000 User s Guide After the instrument is turned on, it runs through the start-up menu. Then the message Please apply zero gas is displayed. At this point, you can perform a zero air (fresh air) calibration. If the ambient air is clean, you can use that. Otherwise, use a cylinder of zero air. Refer to Zero Calibration on page 42 for a more detailed description of zero calibration. Start zero calibration by pressing Start. You see the message Zeroing followed by a 30-second countdown. Note: You can press [MODE] to quit, bypassing the zero air calibration. When zero calibration is complete, you see the message: Zeroing is done! Reading = 0 ppb The instrument is now sampling and collecting data. Note: At the Average & Peak, Date & Time & Temperature, Calibration Gas & Measurement Gas & Correction Factor, and PC Communications screens, the instrument automatically goes to the main display after 60 seconds if you do not push a key to make a selection. Alarm Signals During each measurement period, the gas concentration is compared with the programmed alarm limits (gas concentration alarm limit settings). If the concentration exceeds any of the preset limits, the loud buzzer and red flashing LED are activated immediately to warn you of the alarm condition. In addition, the instrument alarms if one of the following conditions occurs: battery voltage falls below a preset voltage level, failure of the UV lamp, or pump stall. 31

106 Alarm Signal Summary ppbrae 3000 User s Guide Message Condition Alarm Signal HIGH OVR MAX LOW TWA STEL Pump icon flashes Gas exceeds High Alarm limit Gas exceeds measurement range Gas exceeds electronics maximum range Gas exceeds Low Alarm limit Gas exceeds TWA limit Gas exceeds STEL limit Pump failure 3 beeps/flashes per second* 3 beeps/flashes per second* 3 beeps/flashes per second* 2 beeps/flashes per second* 1 Beep/flash per second* 1 Beep/flash per second* 3 beeps/flashes per second Lamp PID lamp failure 3 beeps/flashes per second plus Lamp message on display Battery icon flashes CAL NEG Low battery Calibration failed, or needs calibration Gas reading measures less than number stored in calibration 1 flash, 1 beep per minute plus battery icon flashes on display 1 beep/flash per second 1 beep/flash per second * Hygiene mode only. In Search mode, the number of beeps per second (1 to 7) depends upon the concentration of the sampled gas. Faster rates indicate higher concentrations. 32

107 ppbrae 3000 User s Guide Preset Alarm Limits & Calibration The instrument is factory calibrated with standard calibration gas, and is programmed with default alarm limits. Cal Gas (Isobutylene) ppbrae 3000 Cal Span unit Low High TWA STEL 10 ppm Testing The Alarm You can test the alarm whenever the main (Reading) display is shown. Press [Y/+], and the audible and visible alarms are tested. Integrated Sampling Pump The instrument includes an integrated sampling pump. This diaphragm-type pump that provides a 450 to 550 cc per minute flow rate. Connecting a Teflon or metal tubing with 1/8" inside diameter to the gas inlet port of the instrument, this pump can pull in air samples from 100' (30 m) horizontally or vertically. Note: In Search Mode, the pump turns on when a sample measurement is started, and turns off when the sample is manually stopped. If liquid or other objects are pulled into the inlet port filter, the instrument detects the obstruction and immediately shuts down the pump. The alarm is activated and a flashing pump icon is displayed. You should acknowledge the pump shutoff condition by clearing the obstruction and pressing the [Y/+] key while in the main reading display to restart the pump. 33

108 ppbrae 3000 User s Guide Backlight The LCD display is equipped with an LED backlight to assist in reading the display under poor lighting conditions. Datalogging During datalogging, the instrument displays a disk icon to indicate that datalogging is enabled. The instrument stores the measured gas concentration at the end of every sample period (when data logging is enabled). In addition, the following information is stored: user ID, site ID, serial number, last calibration date, and alarm limits. All data are retained (even after the unit is turned off) in non-volatile memory so that it can be down- loaded at a later time to a PC. Datalogging event When Datalogging is enabled, measurement readings are being saved. These data are stored in groups or events. A new event is created and stored each time the instrument is turned on and is set to automatic datalogging, or a configuration parameter is changed, or datalogging is interrupted. The maximum time for one event is 24 hours or 28,800 points. If an event exceeds 24 hours, a new event is automatically created. Information, such as start time, user ID, site ID, gas name, serial number, last calibration date, and alarm limits are recorded. Datalogging sample After an event is recorded, the unit records a shorter form of the data. When transferred to a PC running ProRAE Studio, this data is arranged with a sample number, time, date, gas concentration, and other related information. 34

109 ppbrae 3000 User s Guide Auto/Manual/Snapshot Datalogging The instrument has three datalog types: Auto Manual Snapshot Default mode. Collects datalog information when the instrument is sampling. Datalogging occurs only when the instrument s datalogging is manually started (see page 69 for details). Datalogs only during snapshot (single-event capture, initiated by pressing [MODE]) sampling. See page 69 for details. Note: You can only choose one datalog type to be active at a time. 35

110 ppbrae 3000 User s Guide Accessories The following accessories are included with the instrument: An AC Adapter (Battery Charger) Alkaline battery adapter External Filter Organic Vapor Zeroing kit Hard-case kits also include these accessories: Calibration adapter Calibration regulator and Flow controller Standard Kit & Accessories AC Adapter (Battery Charger) WARNING To reduce the risk of ignition of hazardous atmospheres, recharge battery only in area known to be non-hazardous. Remove and replace battery only in area known to be non-hazardous. Ne charger les batteries que dans emplacements designés nondangereuses. A battery charging circuit is built into the instrument cradle. It only needs a regular AC to 12 VDC adapter (wall-mount transformer, part number ) to charge the instrument. To charge the battery inside the instrument: 1. Power off the instrument. 2. Connect the AC adapter to the DC jack on the instrument s cradle. If the instrument is off, it automatically turns on. 3. While charging, the display message shows Charging. The Primary LED on the cradle flashes green when charging. 4. When the battery is fully charged, the LED changes to glowing green continuously, and the message Fully charged appears on 36

111 ppbrae 3000 User s Guide the display. If there is a charging error, the LED glows red continuously. A completely discharged instrument can be charged to full capacity within 8 hours. Batteries drain slowly even if an instrument is off. Therefore, if the instrument has been in storage or has not been charged for several days or longer, check the charge before using it. The factory-supplied battery is designed to last for 16 hours of normal operation (no alarm), for a new battery under the optimum circumstances. As the battery becomes older or is subject to adverse conditions (such as cold ambient temperature), its capacity will be significantly reduced. Alkaline Battery Adapter An alkaline battery adapter is supplied with each instrument. The adapter (part number ) accepts four AA alkaline batteries (use only Duracell MN1500) and provides approximately 12 hours of operation. The adapter is intended to be used in emergency situations when there is no time to charge the Li-ion battery pack. To insert batteries into the adapter: 1. Remove the three Philips-head screws to open the compartment in the adapter. 2. Insert four fresh AA batteries as indicated by the polarity (+/-) markings. 3. Replace the cover. Replace the three screws. To install the adapter in the instrument: 1. Remove the Li-ion battery pack from the instrument by sliding the tab and tilting out the battery. 2. Replace it with the alkaline battery adapter 3. Slide the tab back into place to secure the battery adapter. IMPORTANT! Alkaline batteries cannot be recharged. The instrument s internal circuit detects alkaline batteries and will not allow recharging. If you place the instrument in its cradle, the alkaline battery will not be 37

112 ppbrae 3000 User s Guide recharged. The internal charging circuit is designed to prevent damage to alkaline batteries and the charging circuit when alkaline batteries are installed inside the instrument. If you try to charge an alkaline batteries installed in the instrument, the instrument s display will say, Alkaline Battery, indicating that it will not charge the alkaline batteries. Note: When replacing alkaline batteries, dispose of old ones properly. WARNING! To reduce the risk of ignition of hazardous atmospheres, recharge the battery only in areas known to be non-hazardous. Remove and replace the battery only in areas known to be non-hazardous. External Filter The external filter is made of PTFE (Teflon ) membrane with a 0.45 micron pore size to prevent dust or other particles from being sucked into the sensor manifold, which would cause extensive damage to the instrument. It prolongs the operating life of the sensor. To install the external filter, simply connect it to the instrument s inlet tube. 38

113 ppbrae 3000 User s Guide Optional Accessories Calibration Adapter The calibration adapter for the instrument is a simple 6-inch Tygon tubing with a metal adapter on one end. During calibration, simply insert the metal adapter into the regular gas inlet probe of the instrument and the tubing to the gas regulator on the gas bottle. Calibration Regulator The Calibration Regulator is used in the calibration process. It regulates the gas flow rate from the Span gas cylinder into the gas inlet of the instrument during calibration process. The maximum flow rate allowed by the flow controller is about 0.5L/min (500 cc per min.). Alternatively, a demand-flow regulator or a Tedlar gas bag may be used to match the pump flow precisely. Organic Vapor Zeroing Kit The Organic Vapor Zeroing Kit is used for filtering organic air contaminants that may affect the zero calibration reading. To use the Organic Vapor Zeroing Kit, simply connect the filter to the inlet port of the instrument. AutoRAE 2 Automatic Test & Calibration System The AutoRAE 2 Automatic Test and Calibration System for RAE Systems portable gas monitors makes compliance with monitor test and calibration requirements as easy as pressing a button. Simply cradle the monitor and the system will take care of all calibration, testing, and recharging. The AutoRAE 2 is a flexible, modular system that can be configured to meet your calibration requirements effectively and efficiently. An AutoRAE 2 system can be as simple as a single cradle deployed in standalone mode to calibrate one instrument at a time, or as powerful as a networked, controller-based system supporting ten monitors and five distinct calibration gas cylinders. 39

114 Note: Dashed line indicates automatic progression. ppbrae 3000 User s Guide Standard Two-Point Calibration (Zero & Span) The following diagram shows the instrument s calibrations in Basic/Hygiene mode. 40

115 Entering Calibration ppbrae 3000 User s Guide 1. Press and hold [MODE] and [N/-] until you see the Password screen. 2. In Basic User Level, you do not need a password to perform calibrations. Instead of inputting a password, enter calibration by pressing [MODE]. Note: If you inadvertently press [Y/+] and change any of the numbers, simply press [MODE] and you will be directed to the calibration menu. The Calibration screen is now visible with Zero Calibration highlighted. These are your options: Press [Y/+] to select the highlighted calibration (Zero Calib or Span Calib). Press [MODE] to exit calibration and return to the main display and resume measurement. Press [N/-] to toggle the highlighted calibration type. 41

116 ppbrae 3000 User s Guide Zero (Fresh Air) Calibration This procedure determines the zero point of the sensor calibration curve. To perform a fresh air calibration, use the calibration adapter to connect the instrument to a fresh air source such as from a cylinder or Tedlar bag (optional accessory). The fresh air is clean, dry air without organic impurities and an oxygen value of 20.9%. If such an air cylinder is not available, any clean ambient air without detectable contaminants or a charcoal filter can be used. At the Zero Calibration menu, you can proceed to perform a Zero calibration or bypass Zero calibration and perform a Span calibration. You may also go back to the initial Calibration menu if you want to exit calibration. Press [Y/+] to start calibration. Press [MODE] to quit and return to the main calibration display. If you have pressed [Y/+] to enter Zero calibration, then you will see this message: 1. Turn on your Zero calibration gas. 2. Press [Y/+] to start calibration. Note: At this point, you may press [MODE] if you decide that you do not want to initiate calibration. This will take you directly to the Calibration menu, highlighted for Span calibration. 42

117 ppbrae 3000 User s Guide 3. Zero calibration starts a 30-second countdown and displays this message: Zeroing... During the zeroing process, the instrument performs the Zero calibration automatically and does not require any action on your part. Note: To abort the zeroing process at any time and proceed to Span calibration, press [N/-] at any time while zeroing is being performed. You will see a confirmation message that says Zero aborted! and then the Span calibration menu appears. When Zero calibration is complete, you see this message: Zeroing is done! Reading = 0 ppb The instrument will then show the Calibration menu on its display, with Span Calib highlighted. 43

118 ppbrae 3000 User s Guide Span Calibration This procedure determines the second point of the sensor calibration curve for the sensor. A cylinder of standard reference gas (span gas) fitted with a 500 cc/min. flow-limiting regulator or a flow-matching regulator is the simplest way to perform this procedure. Choose the 500 cc/min. regulator only if the flow rate matches or slightly exceeds the flow rate of the instrument pump. Alternatively, the span gas can first be filled into a Tedlar bag or delivered through a demand-flow regulator. Connect the calibration adapter to the inlet port of the instrument, and connect the tubing to the regulator or Tedlar bag. Another alternative is to use a regulator with >500 cc/min flow but allow the excess flow to escape through a T or an open tube. In the latter method, the span gas flows out through an open tube slightly wider than the probe, and the probe is inserted into the calibration tube. At the Span Calibration menu, you perform a Span calibration. You may also go back to the Zero calibration menu or to the initial Calibration menu if you want to exit calibration. Press [Y/+] to enter Span calibration. Press [N/-] to skip Span calibration and return to Zero calibration. Press [MODE] to exit Span calibration and return to the top calibration menu. If you have pressed [Y/+] to enter Span calibration, then you will see the name of your Span gas (the default is isobutylene) and the span value in parts per million (ppm). You will also see this message that prompts you: 1. Turn on your span calibration gas. 44

119 ppbrae 3000 User s Guide 2. Press [Y/+] to initiate calibration. Note: You may press [MODE] if you decide that you do not want to initiate calibration. This will abort the span calibration and take you directly to the Calibration menu for Zero calibration. 3. Span calibration starts and displays this message: Calibrating... During the Span calibration process, there is a 30-second countdown and the instrument performs the Span calibration automatically. It requires no actions on your part. Note: If you want to abort the Span calibration process, press [N/-] at any time during the process. You will see a confirmation message that says Span is aborted! and then the Zero calibration menu appears. You can then proceed to perform a Zero calibration, perform a Span calibration, or exit to the topmost Calibration menu. When Span calibration is complete, you see a message similar to this (the value is an example only): Span 1 is done! Reading = 10.0 ppm The instrument then exits Span calibration and shows the Zero calibration menu on its display. Note: The reading should be very close to the span gas value. 45

120 ppbrae 3000 User s Guide Exiting Two-Point Calibration In Basic User Level When you are done performing calibrations, press [MODE], which corresponds with Back on the display. You will see the following message: Updating settings The instrument updates its settings and then returns to the main display. It begins or resumes monitoring. 46

121 ppbrae 3000 User s Guide Three-Point Calibration For enhanced accuracy, it is possible to perform a second Span calibration in addition to the Zero and Span calibrations outlined in the previous section. Your instrument first must be set to allow this third calibration. This requires using ProRAE Studio software and a PC, as well as a higher concentration of calibration gas. Note: Once the third calibration is set, you do not need to use ProRAE Studio to allow future 3-point calibrations. Also, you can only disable 3-point calibration capability by using ProRAE Studio again. Perform the Zero and Span calibrations. After the first Span calibration (Span 1) is completed, the display a second Span calibration (Span 2) can be performed. The process is identical to the first calibration. As in the Span 1 calibration, you may exit and return to the Zero calibration screen if you choose not to perform this calibration or to abort it. Note: If a bump test is available, it appears after the last calibration in the menu. See Two-Point Calibration, page 38, for details. Also, refer to page 49 for details on how to perform a bump test. 47

122 Note: Dashed line indicates automatic progression. ppbrae 3000 User s Guide 48

123 ppbrae 3000 User s Guide Span 2 Calibration A cylinder of standard reference gas (span gas) fitted with a 500 cc/min. flow-limiting regulator or a flow-matching regulator is the simplest way to perform this procedure. Note: This gas should be of a higher concentration than the gas used for Span 1 calibration. Choose the 500 cc/min. regulator only if the flow rate matches or slightly exceeds the flow rate of the instrument pump. Alternatively, the span gas can first be filled into a Tedlar bag or delivered through a demand-flow regulator. Connect the calibration adapter to the inlet port of the instrument, and connect the tubing to the regulator or Tedlar bag. Another alternative is to use a regulator with >500 cc/min flow but allow the excess flow to escape through a T or an open tube. In the latter method, the span gas flows out through an open tube slightly wider than the probe, and the probe is inserted into the calibration tube. At the Span Calibration menu, you perform a Span calibration. You may also go back to the Zero calibration menu or to the initial Calibration menu if you want to exit calibration. Press [Y/+] to enter Span 2 calibration. Press [N/-] to skip Span calibration and return to Zero calibration. Press [MODE] to exit Span calibration and return to the top calibration menu. If you have pressed [Y/+] to enter Span calibration, then you will see the name of your Span gas (the default is isobutylene) and the span value in parts per million (ppm). You will also see this message that prompts you: Please apply gas Turn on your span calibration gas. 49

124 ppbrae 3000 User s Guide 5. Press [Y/+] to initiate calibration. Note: You may press [MODE] if you decide that you do not want to initiate calibration. This will take you directly to the Calibration menu for Zero calibration. 6. Span calibration starts a 30-second countdown and displays this message: Calibrating... During the Span calibration process, the instrument performs the Span calibration automatically and does not require any action on your part. Note: If you want to abort the Span calibration process, press [N/-] at any time during the process. You will see a confirmation message that says Span is aborted! and then the Zero calibration menu will appear. You can then proceed to perform a Zero calibration, perform a Span calibration, or exit to the topmost Calibration menu. When Span calibration is complete, you will see a message similar to this (the value shown here is for example only): Span 2 is done! Reading = 1000 ppm The instrument then exits Span calibration and shows the Zero calibration menu on its display. Note: The reading should be very close to the span gas value. 50

125 ppbrae 3000 User s Guide Exiting Three-Point Calibration When you are done performing calibrations, press [MODE], which corresponds with Back on the display. You will see the following message: Updating settings The instrument updates its settings and then returns to the main display. It begins or resumes monitoring. Bump Test RAE Systems recommends that a bump test be conducted prior to each day s use. The purpose of a bump test is to ensure that the instrument s sensors respond to gas and all the alarms are enabled and functional. The ppbrae 3000 must be calibrated if it does not pass a bump test when a new sensor is installed, after sensor maintenance has been performed, or at least once every 180 days, depending on use and sensor exposure to poisons and contaminants. Calibration and bump test intervals and procedures may vary due to national legislation and company policy. To perform a bump test (functional challenge), follow these steps: 1. Select Bump. 2. Install the calibration adapter and connect it to a source of calibration gas. 3. Verify that the displayed calibration value meets the concentration specified on the gas cylinder. 4. Start the flow of calibration gas. 5. Press [Y/+] to start the bump test. 51

126 ppbrae 3000 User s Guide 6. You can abort the calibration at any time during the countdown by pressing [N/-]. 7. If the calibration is not aborted, the display shows reading and then tells you whether the bump test passed or failed. If the bump test failed, then it automatically advances to the Calibration screen. A bump test can be performed either manually or using the AutoRAE 2 Automatic Test and Calibration System. When a bump test is done manually, the instrument makes a pass/fail decision based on sensor performance, but the user still has the responsibility to make sure all the alarms are enabled and functional. Note: Bump testing and calibration can be performed using an AutoRAE 2 Automatic Test & Calibration System. An AutoRAE 2 bump test takes care of both the sensor and alarm tests. Consult the AutoRAE 2 User s guide for details. The same gas is used for a bump test as for calibration. The instrument must be connected to a cylinder of calibration gas with supplied tubing. IMPORTANT! If the instrument does not pass a bump test, perform a full calibration. If calibration also fails, the PID sensor or lamp may require cleaning or replacement. If the instrument repeatedly fails to calibrate, turn it off and refer it for servicing. Important! Anytime a bump test fails, you should perform a full calibration of the instrument. 52

127 ppbrae 3000 User s Guide Programming Mode Programming Mode can be entered from either Hygiene Mode or Search Mode. If the current user mode is Basic, you must provide a 4- digit password to enter. Entering Programming Mode 1. Press and hold [MODE] and [N/-] until you see the Password screen. 2. Input the 4-digit password: Increase the number from 0 through 9 by pressing [Y/+]. Step from digit to digit using [N/-]. Press [MODE] when you are done. If you make a mistake, you can cycle through the digits by pressing [N/-] and then using [Y/+] to change the number in each position. Note: The default password is When you have successfully entered Programming Mode, you see this screen: 53

128 ppbrae 3000 User s Guide Note: The password can only be changed by connecting the instrument to a PC running ProRAE Studio software. Follow the instructions in ProRAE Studio to change it.the Calibration label is shown and its icon is highlighted, but you can press [N/-] to step from one programming menu to the next, with the name of the menu shown at the top of the display and the corresponding icon highlighted. As you repeatedly press [N/-], the selection moves from left to right, and you see these screens: Note: When you reach Monitor Setup and press [N/-], the menu cycles back to Calibration. Programming Mode Menus The Programming Mode allows anyone with the password to change the instrument s settings, calibrate the instrument, modify the sensor configuration, enter user information, etc. Programming Mode has five menus. Each menu includes several sub-menus to perform additional programming functions. The table on the next shows the menus and sub-menus. 54

129 ppbrae 3000 User s Guide Calibration Zero Calibration Span Calibration Bump Test Measurement Alarm Setting Meas. Gas High Alarm Meas. Unit Low Alarm STEL Alarm TWA Alarm Alarm Type Buzzer & Light Datalog Clear Datalog Interval Data Selection Datalog Type Monitor Setup Op Mode Site ID User ID User Mode Date Time Pump Duty Cycle Pump Speed Temperature Unit Language Radio Power Real Time Protocol Power On Zero Unit ID LCD Contrast Lamp ID PAN ID Mesh Channel Mesh Interval 55

130 ppbrae 3000 User s Guide Once you enter Programming Mode, the LCD displays the first menu, Calibration. Each subsequent menu is accessed by pressing [N/-] repeatedly until the desired menu is displayed. To enter a sub-menu of a menu, press [Y/+]. Exiting Programming Mode To exit Programming Mode and return to normal operation, press [MODE] once at any of the programming menu displays. You will see Updating Settings as changes are registered and the mode changes. Navigating Programming Mode Menus Navigating through the Programming Mode menus is easy and consistent, using a single interface format of Select, Back and Next at the top level. The three control buttons correspond to these choices as shown: Note: Pressing [MODE] in the Programming Mode s top level causes the instrument to exit Programming Mode and return to monitoring. 56

131 ppbrae 3000 User s Guide The three keys perform the following functions in Programming Mode: Key [MODE]: [Y/+]: [N/-]: Function in Programming Mode Exit menu when pressed momentarily or exit data entry mode Increase alphanumerical value for data entry or confirm (yes) for a question Provides a no response to a question 57

132 ppbrae 3000 User s Guide Calibration Two types of calibration are available: Zero (fresh air) and Span. Select Zero or Span Calibration by pressing [N/+]. Once your choice is highlighted, press [Y/+]. Zero Calibration The procedure for performing a zero calibration is covered on page 40. Span Calibration The procedure for performing a basic span calibration is covered on page

133 Measurement ppbrae 3000 User s Guide The sub-menus for Measurement are Measurement Gas and Measurement Unit. Meas. Gas Measurement gases are organized in four lists: My List is a customized list of gases that you create. It contains a maximum of 10 gases and can only be built in ProRAE Studio on a PC and transferred to the instrument. Note: The first gas in the list is always isobutylene (it cannot be removed from the list). Last Ten is a list of the last ten gases used by your instrument. The list is built automatically and is only updated if the gas selected from Custom Gases or Library is not already in the Last Ten. This ensures that there is no repetition. Gas Library is a library that consists of all the gases found in RAE Systems Technical Note TN-106 (available online at Custom Gases are gases with user-modified parameters. Using ProRAE Studio, all parameters defining a gas can be modified, including the name, span value(s), correction factor, and default alarm limits. 1. Scroll through each list by pressing [N/-]. 2. Press [Y/+] to select one (My List, Last Ten, Gas Library, or Custom Gases). 59

134 ppbrae 3000 User s Guide 3. Once you are in one of the categories, press [N/-] to scroll through its list of options and [Y/+] to select one. (If you press [MODE], you exit to the next submenu.) 4. Press [Y/+] to save your choice or [N/-] to undo your selection. Leave the sub-menu and return to the Programming Mode menus by pressing [MODE]. Meas. Unit Standard available measurement units include: Abbreviation Unit ppm parts per million ppb parts per billion mg/m3 milligrams per cubic meter ug/m3 micrograms per cubic meter Scroll through the list by pressing [N/-]. Select by pressing [Y/+]. Save your selection by pressing [Y/+] or undo your selection by pressing [N/-]. Leave the sub-menu and return to the Programming Mode menus by pressing [MODE]. 60

135 ppbrae 3000 User s Guide Alarm Setting During each measurement period, the gas concentration is compared with the programmed alarm limits (gas concentration alarm limit settings: Low, High, TWA and STEL). If the concentration exceeds any of the preset limits, the loud buzzer and red flashing LED are activated immediately to warn of the alarm condition. An alarm signal summary is shown on page 32. In this menu, you can change the High and Low alarm limits, the STEL limit, and the TWA. Press [Y/+] to to enter the Alarm Setting menu. Note: All settings are shown in ppm (parts per million), or mg/m 3 (milligrams per cubic meter), depending on your setting. 1. Scroll through the Alarm Limit sub-menu using the [N/-] key until the display shows the desired limit to be changed (High Alarm, Low Alarm, STEL Alarm, and TWA Alarm) 2. Press [Y/+] to select one of the alarm types. The display shows a flashing cursor on the left-most digit of the previously stored alarm limit. 3. Press [Y/+] to increase each digit s value. 4. Press [N/-] to advance to the next digit. 5. Again, use [Y/+] to increase the number. Repeat this process until all numbers are entered. Press [MODE] when you are done. Press [Y/+] to save the changes. Press [N/-] to undo the changes and revert to the previous settings. When all alarm types have been changed or bypassed, press [MODE] to exit to the Programming Menu. 61

136 ppbrae 3000 User s Guide High Alarm You can change the High Alarm limit value. The value is typically set by the instrument to match the value for the current calibration gas. It is expressed in parts per billion (ppb). Note: The default value depends on the measurement gas. To change the High Alarm value: 1. Press [Y/+] to increase each digit s value. 2. Press [N/-] to advance to the next digit. 3. Again, use [Y/+] to increase the number. Repeat this process until all numbers are entered. When you have completed your selections, press [MODE]. You will see two choices: Save and Undo. You have the opportunity to register the new settings or to change your mind and revert to your previous settings. Press [Y/+] to save the changes. Press [N/-] to undo the changes and revert to the previous settings. Low Alarm You can change the Low Alarm limit value. The value is typically set by the instrument to match the value for the current calibration gas. It is expressed in parts per billion (ppb). Note: The default value depends on the measurement gas. To change the Low Alarm value: 1. Press [Y/+] to increase each digit s value. 2. Press [N/-] to advance to the next digit. 3. Again, use [Y/+] to increase the number. Repeat this process until all numbers are entered. 62

137 ppbrae 3000 User s Guide When you have completed your selections, press [MODE]. You will see two choices: Save and Undo. You have the opportunity to register the new settings or to change your mind and revert to your previous settings. Press [Y/+] to save the changes. Press [N/-] to undo the changes and revert to the previous settings. STEL Alarm You can change the STEL Alarm limit value. The value is typically set by the instrument to match the value for the calibration gas. It is expressed in parts per billion (ppb). Note: The default value depends on the measurement gas. To change the STEL Alarm value: 1. Press [Y/+] to increase each digit s value. 2. Press [N/-] to advance to the next digit. 3. Again, use [Y/+] to increase the number. Repeat this process until all numbers are entered. When you have completed your selections, press [MODE]. You will see two choices: Save and Undo. You have the opportunity to register the new settings or to change your mind and revert to your previous settings. Press [Y/+] to save the changes. Press [N/-] to undo the changes and revert to the previous settings. 63

138 ppbrae 3000 User s Guide TWA Alarm You can change the TWA (time-weighted average) Alarm limit value. The value is typically set by the instrument to match the value for the calibration gas. It is expressed in parts per billion (ppb). Note: The default value depends on the measurement gas. To change the TWA Alarm value: 1. Press [Y/+] to increase each digit s value. 2. Press [N/-] to advance to the next digit. 3. Again, use [Y/+] to increase the number. Repeat this process until all numbers are entered. When you have completed your selections, press [MODE]. You will see two choices: Save Undo You have the opportunity to register the new settings or to change your mind and revert to your previous settings. Press [Y/+] to save the changes. Press [N/-] to undo the changes and revert to the previous settings. 64

139 ppbrae 3000 User s Guide Alarm Type There are two selectable alarm types: Latched Automatic Reset When the alarm is triggered, you can manually stop the alarm. The latched setting only controls alarms for High Alarm, Low Alarm, STEL Alarm, and TWA alarm. Note: To clear an alarm when the instrument is set to Latched, press [Y/+] when the main (Reading) display is shown. When the alarm condition is no longer present, the alarm stops and resets itself. 1. Press [N/-] to step from one alarm type to the other. 2. Press [Y/+] to select an alarm type. When you have completed your selections, press [MODE]. You will see two choices: Save and Undo. You have the opportunity to register the new settings or to change your mind and revert to your previous settings. Press [Y/+] to save the changes. Press [N/-] to undo the changes and revert to the previous settings. 65

140 ppbrae 3000 User s Guide Buzzer & Light The buzzer and light alarms can be programmed to be on or off individually or in combination. Your choices are: Both on Light only Buzzer only Both off 1. Press [N/-] to step from one option to the next. 2. Press [Y/+] to make your selection (the dark circle in the radio button indicates your selection). 3. When you have completed your selections, press [MODE]. You will see two choices: Save and Undo. You have the opportunity to register the new settings or to change your mind and revert to your previous settings. Press [Y/+] to save the changes. Press [N/-] to undo the changes and revert to the previous settings. Datalog The instrument calculates and stores the concentration and ID of each sample taken. In the datalog sub-menu, a user can perform the tasks and functions shown below. 66

141 ppbrae 3000 User s Guide 1. Scroll through the Datalog sub-menu using the [N/-] key until the display shows the desired parameter to be changed: Clear Datalog Interval Data Selection Datalog Type 2. Press [Y/+] to make your selection. Exit by pressing [MODE] for Back. Clear Datalog This erases all the data stored in the datalog. Note: Once the datalog is cleared, the data cannot be recovered. Press [Y/+] to clear the datalog. The display asks, Are you sure? Press [Y/+] if you want to clear the datalog. When it has been cleared, the display shows Datalog Cleared! Press [N/-] if you do not want to clear the datalog. The display changes, and you are taken to the next sub-menu, Interval. Interval Intervals are shown in seconds. The default value is 60 seconds. The maximum interval is 3600 seconds. 1. Press [Y/+] to increase each digit s value. 2. Press [N/-] to advance to the next digit. 3. Again, use [Y/+] to increase the number. 67

142 ppbrae 3000 User s Guide Repeat this process until all numbers are entered. When you have completed your selections, press [MODE]. You will see two choices: Save and Undo. You have the opportunity to register the new settings or to change your mind and revert to your previous settings. Press [Y/+] to save the changes. Press [N/-] to undo the changes and revert to the previous settings. Data Selection Data Selection allows you to select which types of data are stored and made available when you offload your datalog to a computer via ProRAE Studio software. You can choose any or all of three types of data (you must choose at least one): Average Maximum Minimum 1. Press [N/-] to step from one option to the next. The highlighter indicates your choice. 2. Press [Y/+] to toggle your selection on or off (the check box indicates on with an X ). 3. When you have completed your selections, press [MODE]. You will see two choices: Save and Undo. You have the opportunity to register the new settings or to change your mind and revert to your previous settings. Press [Y/+] to save the changes. Press [N/-] to undo the changes and revert to the previous settings. 68

143 ppbrae 3000 User s Guide Datalog Type The instrument has three datalog types: Auto Default mode. Collects datalog information when the instrument is sampling. Manual Datalogging occurs only when the instrument s datalogging is manually started (see below for details). Snapshot Datalogs only during single-event capture sampling. Note: You can only choose one datalog type to be active at a time. 1. Press [N/-] to step from one option to the next. 2. Press [Y/+] to make your selection (the dark circle in the radio button indicates on ). 3. When you have completed your selection, press [MODE]. You will see two choices: Save and Undo. You have the opportunity to register the new settings or to change your mind and revert to your previous settings. Press [Y/+] to save the changes. Press [N/-] to undo the changes and revert to the previous settings. Manual Datalog When the instrument is set to Manual Datalog, you turn datalogging on and off by stepping through the displays from the Main Display, and then pressing the keys to select datalog on/off functions. When you reach the screen that says Start Datalog? press [Y/+] to start it. You see Datalog Started, confirming that datalogging is now on. When you reach the screen that says Stop Datalog? press [Y/+] to stop it. You see Datalog Stopped, confirming that datalogging is now off. 69

144 Note: Dashed line indicates automatic progression. ppbrae 3000 User s Guide 70

145 Snapshot Datalog ppbrae 3000 User s Guide When the instrument is in Snapshot datalogging mode, it captures a single snapshot of the data at the moment of your choosing. Whenever the instrument is on and it is set to Snapshot, all you have to do is press [MODE] each time you want to capture a snapshot of the data at that instant. When you send the data to a computer using ProRAE Studio, the data snapshots are uniquely identified by time and other parameters. Monitor Setup Many settings can be accessed in this menu, including setting the date and time and adjusting the pump s on/off duty cycle. Op Mode Under Monitor Setup is Op Mode. Press [Y/+] to select. You see two options (one is highlighted): Hygiene Search 71

146 ppbrae 3000 User s Guide The current mode is indicated by a dark circle within the circle in front of either Hygiene or Search. 1. Select Hygiene or Search by pressing [N/-]. The highlighting changes from one to the other each time you press [N/-]. 2. Press [Y/+] to select that mode for the instrument. 3. Press [MODE] when you want to register your selection to place the instrument in the selected mode. 4. Press [Y/+] to commit the change and exit to the Monitor Setup screen, or press [N/-] to Undo (exit to the Monitor Setup screen without changing the Mode). Site ID Enter an 8-digit alphanumeric/character Site ID in the programming mode. This Site ID is included in the datalog report. 1. Press [Y/+] and the display shows the current site ID. Example: RAE Note that the left-most digit flashes to indicate it is the selected one. 2. Press [Y/+] to step through all 26 letters (A to Z) and 10 numerals (0 to 9). Note: The last four digits must be numerals. 3. Press [N/-] to advance to the next digit. The next digit to the right flashes. Repeat this process until all eight digits of the new site ID are entered. Press [MODE] to exit. If there is any change to the existing site ID, the display shows Save? Press [Y/+] to accept the new site ID. Press [N/-] to discard the change and move to the next sub-menu. 72

147 ppbrae 3000 User s Guide User ID Enter an 8-digit alphanumeric User ID in the programming mode. This User ID is included in the datalog report. 1. Press [Y/+] and the display shows the current User ID. Example: RAE Note that the left-most digit flashes to indicate it is the selected one. 2. Press [Y/+] to step through all 26 letters (A to Z) and 10 numerals (0 to 9). 3. Press [N/-] to advance to the next digit. The next digit to the right flashes. Repeat this process until all eight digits of the new User ID are entered. Press [MODE] to exit. If there is any change to the existing User ID, the display shows Save Press [Y/+] to accept the new site ID. Press [N/-] to discard (undo) the change and move to the next sub-menu. 73

148 ppbrae 3000 User s Guide User Mode The instrument has two user modes: Basic Advanced Basic users can only see and use a basic set of functions. Advanced users can see all screens and perform all available functions. Note: The default value for User Mode is Basic. To change the User Mode: 1. Press [N/-] to step from one option to the next. The highlighting changes each time you press [N/-]. 2. Press [Y/+] to make your selection (the dark circle in the radio button indicates on ). 3. When you have completed your selection, press [MODE]. 4. Press [Y/+] to accept the new User Mode. Press [N/-] to discard the change and move to the next sub-menu. 74

149 ppbrae 3000 User s Guide Date The Date is expressed as Month/Day/Year, with two digits for each. 1. Press [Y/+] and the display shows the current date. Note that the left-most digit flashes to indicate it is selected. 2. Press [Y/+] to step through all 10 numerals (0 to 9). 3. Press [N/-] to advance to the next digit. The next digit to the right flashes. Repeat this process until all six digits of the new date are entered. Press [MODE] to exit. Press [Y/+] to save the new date. Press [N/-] to undo the change and move to the next submenu. Time The Time is expressed as Hours/Minutes/Seconds, with two digits for each. The time is in 24-hour (military) format. 1. Press [Y/+] and the display shows the current time. Note that the left-most digit flashes to indicate it is selected. 2. Press [Y/+] to step through all 10 numerals (0 to 9). 3. Press [N/-] to advance to the next digit. The next digit to the right flashes. Repeat this process until all six digits of the new time are entered. Press [MODE] to exit. Press [Y/+] to save the new date. Press [N/-] to undo the change and move to the next submenu. 75

150 ppbrae 3000 User s Guide Duty Cycle The pump s duty cycle is the ratio of its on time to off time. The duty cycle ranges from 50% to 100% (always on), and the period is 10 seconds. Therefore, a duty cycle of 60% means that the pump is on for 6 seconds and off for four seconds. Duty cycling is employed by the instrument to clean the PID. A lower duty cycle has a greater effect on keeping the PID clean than a higher duty cycle. Important! Pump duty cycling is interrupted when the instrument senses a gas. The pump s duty cycle is disabled when the measurement is greater than the 2ppm threshold and is re-enabled when the reading falls below 90% of the threshold (1.8 ppm). 1. Press [Y/+] to increase the value. 2. When you have completed your selection, press [MODE]. Press [Y/+] to save the new duty cycle value. Press [N/-] to undo the change and move to the next submenu. 76

151 ppbrae 3000 User s Guide Temperature Unit The temperature display can be switched between Fahrenheit and Celsius units. 1. Press [N/-] to step from one option to the next. 2. Press [Y/+] to make your selection (the dark circle in the radio button indicates on ). 3. When you have completed your selection, press [MODE]. Press [Y/+] to save the new temperature unit. Press [N/-] to undo the change and move to the next submenu. Pump Speed The pump can operate at two speeds, high and low. Running at low speed is quieter and conserves a small amount of power. There is almost no difference in sampling accuracy. 1. Press [N/-] to step from one option to the next. 2. Press [Y/+] to make your selection (the dark circle in the radio button indicates on ). 3. When you have completed your selection, press [MODE]. Press [Y/+] to save the new temperature unit. Press [N/-] to undo the change and move to the next submenu. 77

152 ppbrae 3000 User s Guide Language English is the default language, but other languages can be selected for the instrument. 1. Press [N/-] to step from one option to the next. 2. Press [Y/+] to make your selection (the dark circle in the radio button indicates on ). 3. When you have completed your selection, press [MODE]. Press [Y/+] to save your new language choice. Press [N/-] to undo it and return to the previous language selection. Radio Power The radio connection can be turned on or off. 1. Press [N/-] to step from one option to the next (on or off). 2. Press [Y/+] to make your selection (the dark circle in the radio button indicates that the option is selected). 3. When you have completed your selection, press [MODE]. Press [Y/+] to accept the new radio setting (on or off). Press [N/-] to discard the change and move to the next submenu. 78

153 ppbrae 3000 User s Guide Real Time Protocol Real Time Protocol is the setting for data transmission. The choices are: P2M (cable) P2P (cable) P2M (wireless) Point to multipoint. Data is transferred from the instrument to multiple locations using a wired connection. Default data rate: bps. Point to point. Data is transferred only between the instrument and one other location, such as a computer. Default data rate: 9600 bps. Point to multipoint, wireless. Data is transferred wirelessly and can be received by multiple receivers. 1. Press [N/-] to step from one option to the next. 2. Press [Y/+] to make your selection (the dark circle in the radio button indicates on ). 3. When you have completed your selection, press [MODE]. Press [Y/+] to save the new real-time communications protocol. Press [N/-] to undo the change and move to the next sub-menu. Power On Zero When Power On Zero is on, the instrument performs a zero calibration when it is turned on. 1. Press [N/-] to step from one option to the next. 2. Press [Y/+] to make your selection (the dark circle in the radio button indicates your selection). 3. When you have completed your selection, press [MODE]. Press [Y/+] to save the change. Press [N/-] to discard the change and move to the next submenu. 79

154 ppbrae 3000 User s Guide Unit ID This three-digit number keeps data separated by instrument when more than one instrument is used in a network. If multiple sensing units are attempting to communicate with the same Host, then the units must all have a different Unit ID. 1. Press [Y/+] to step through all 10 numerals (0 to 9). If you pass the numeral you want, keep pressing [Y/+]. After it counts up to 9, it starts counting up from 0 again. 2. Press [N/-] to advance to the next digit. The next digit to the right flashes. Repeat this process until all three digits of the Unit ID are entered. 3. Press [MODE] when you are done. Press [Y/+] to save the change. Press [N/-] to discard the change and move to the next submenu. LCD Contrast The display s contrast can be increased or decreased from its default setting. You may not need to ever change the default setting, but sometimes you can optimize the display to suit extreme temperature and ambient brightness/darkness conditions. The minimum value is 20. The maximum value is Press [Y/+] to increase the value or [N/-] to decrease the value. 2. Press [MODE] to save your selection. Press [Y/+] to save your new contrast value. Press [N/-] to undo it and return to the previous value. 80

155 ppbrae 3000 User s Guide Lamp ID The ppbrae 3000 does not automatically identify the type of lamp, so you must select a lamp type manually. 1. Scroll through the menu by pressing [N/-]. 2. Press [Y/+] to make a selection. 3. Press [MODE] when you are done. 4. Press [Y/+] to save your choice or [N/-] to undo your selection. PAN ID The ppbrae 3000 and any other devices that you want to interconnect wirelessly must have the same PAN ID. You can set the PAN ID in the instrument or through ProRAE Studio II. 1. Press [N/-] to advance through the digits from left to right. 2. Press [Y/+] to ] to advance through the numbers (1, 2, 3, etc.). 3. Press [MODE] to register your choice when you are done. Mesh Channel Note: For mesh radio modems operating at 868MHz, only channel 0 is available. For other frequencies, channels 1 through 10 are allowed. 1. Press [Y/+] to increase the number and [N/-] to advance to the next digit. 2. After moving to the last digit and making changes, press [MODE]. Press [Y/+] to save the change. Press [N/-] to undo the change. Mesh Interval Set the time interval at which the instrument s mesh radio sends out a signal. This can range from once every 10 seconds to once every four 81

156 ppbrae 3000 User s Guide minutes (240 seconds). The transmission frequency is user-adjustable, but a rate of at least once every 30 seconds is recommended. Note: Shorter intervals reduce battery life. 1. Press [N/-] to step from one option to the next. 2. Press [Y/+] to make a selection. 3. When you are done, press [MODE]. Hygiene Mode The instrument usually operates in Hygiene Mode, which provides basic functionality. However, it is possible to operate it in a second mode called Search Mode. Here are the primary differences: Hygiene Mode: Automatic measurements, continuously running and datalogging, and calculates additional exposure values. Search Mode: Manual start/stop of measurements and display of certain exposure values. 82

157 Note: Dashed line indicates automatic progression. ppbrae 3000 User s Guide Basic User Level & Hygiene Mode The default setting is navigated in the following way: 83

158 ppbrae 3000 User s Guide Pressing [N/-] steps you from screen to screen. Options include clearing the Peak value and turning on the instrument s PC Communications for data transfer to a PC. 84

159 ppbrae 3000 User s Guide Entering Search Mode From Hygiene Mode In order to change the instrument s operational mode from Hygiene Mode to Search Mode, you must enter the password-protected Programming Mode: 1. Hold [MODE] and [N/-] until you see the password screen. 2. Use [Y/+] to increment to the number you want for the first digit. (If you pass by the desired number, press [Y/+] until it cycles through to 0 again. Then press [Y/+] until you reach the desired number.) 3. Press [N/-] to advance to the next digit. 4. Again press [Y/+] to increment the number. 5. Press [N/-] to advance to the next digit. Continue the process until all four numbers of the password have been input. Then press [MODE] to proceed. The screen changes to icons with the label Calibration. 1. Press [N/-] to advance to Monitor Setup. 2. Press [Y/+] to select Monitor Setup. Under Monitor Setup, you will see Op Mode. Press [Y/+] to select. You will see: Hygiene Search The current mode is indicated by a dark circle within the circle in front of either Hygiene or Search. 1. Select Hygiene or Search by pressing [N/-]. 2. Press [Y/+] to place the instrument into the selected mode. 85

160 ppbrae 3000 User s Guide 3. Press [MODE] when you want to register your selection to place the instrument in the selected mode. 4. Press [Y/+] to commit the change and exit to the Monitor Setup screen, or press [N/-] to Undo (exit to the Monitor Setup screen without changing the Mode). Advanced User Level (Hygiene Mode Or Search Mode) The User Mode called Advanced User Level allows a greater number of parameters to be changed than Basic User Level. It can be used with either of the Operation Modes, Hygiene Mode or Search Mode. Advanced User Level & Hygiene Mode With the instrument in Operation Mode: Hygiene Mode, enter User Mode: Advanced User Level (refer to the section called Monitor Mode for instructions). Once you are in Advanced User Level and Hygiene Mode together, you can change the calibration reference and measurement gas, in addition to performing normal monitoring functions. Pressing [N/-] progresses through the screens, while pressing [Y/+] selects options. Pressing [MODE] makes menu choices when it is shown for Done or Back. Pressing and holding [Mode] whenever the circle with a vertical line in the middle is shown activates the countdown to shutoff. 86

161 Note: Dashed line indicates automatic progression. ppbrae 3000 User s Guide 87

162 ppbrae 3000 User s Guide Basic User Level & Search Mode With the instrument in Operation Mode: Search Mode, enter User Mode and select Basic User Level (refer to the section called User Mode for instructions). When the instrument is in Search Mode, it only samples when you activate sampling. When you see the display that says, Ready Start sampling? press [Y/+] to start. The pump turns on and the instrument begins collecting data. To stop sampling, press [N/-] while the main display is showing. You will see a new screen that says, Stop sampling? Press [Y/+] to stop sampling. Press [N/-] if you want sampling to continue. 88

163 Note: Dashed line indicates automatic progression. ppbrae 3000 User s Guide 89

164 ppbrae 3000 User s Guide Advanced User Level & Search Mode With the instrument in Operation Mode: Search Mode, enter User Mode and select Advanced User Level (refer to the section called Monitor Mode for instructions). Operation is similar to Basic User Level & Sampling Mode, but now allows you to change calibration and measurement reference gases. Refer to the section on measurement gases on page 59 for more details. 90

165 Note: Dashed line indicates automatic progression. ppbrae 3000 User s Guide 91

166 Diagnostic Mode ppbrae 3000 User s Guide IMPORTANT! Diagnostic Mode is designed for servicing and manufacturing, and therefore is not intended for everyday use, even by advanced users. It provides raw data from sensors and about settings, but only allows adjustment of pump stall parameters, which should only be changed by qualified personnel. Note: If the instrument is turned on in Diagnostic Mode and you switch to User Mode, datalog data remains in raw count form. To change to standard readings, you must restart the instrument. Entering Diagnostic Mode Note: To enter Diagnostic Mode, you must begin with the instrument turned off. Press and hold [Y/+] and [MODE] until the instrument starts. The instrument goes through a brief startup, and then displays raw data for the PID sensor. These numbers are raw sensor readings without calibration. The instrument is now in Diagnostic Mode. Note: In Diagnostic Mode, the pump and lamp are normally on. You can enter Programming Mode and calibrate the instrument as usual by pressing both [MODE] and [N/-] for three seconds. You can enter Monitoring Mode by pressing [MODE] and [Y/+] together for three seconds. Once the instrument is started up in Diagnostic Mode, you can switch between Diagnostic Mode and Monitoring Mode by pressing and holding [MODE] and [Y/+] simultaneously for two seconds. In Diagnostic mode, you can step through parameter screens by pressing [MODE]. 92

167 ppbrae 3000 User s Guide Adjusting The Pump Stall Threshold If the gas inlet is blocked but the pump does not shut down, or the pump shuts down too easily with a slight blockage, the pump stall threshold value may be set too high or too low. Use the following steps to adjust the pump stall threshold: Pump High In Diagnostic Mode, press the [MODE] key until Pump High is displayed. The display shows the maximum, minimum, and stall values for the pump at its high speed. Write down the Max reading. Block the gas inlet and watch the pump current reading (labeled I ) increase. Write down its blocked reading. Note: If the pump current reading does not increase significantly (less than 10 counts), then there may be a leak in the gas inlet or the pump is weak or defective. Add the two readings you wrote down. This is the average of the maximum block count and the maximum idle count. Divide that number by 2. Use the [Y/+] or [N/-] key to increase or decrease the stall value to equal that number. Press the [MODE] key to exit this display. Pump Low In Diagnostic Mode, press the [MODE] key until Pump Low is displayed. The display shows the maximum, minimum, and stall values for the pump at its low speed. Write down the Max reading. Block the gas inlet and watch the pump current reading (labeled I ) increase. Write down its blocked reading. Note: If the pump current reading does not increase significantly (less than 10 counts), then there may be a leak in the gas inlet or the pump is weak or defective. Add the two readings you wrote down. This is the average of the maximum block count and the maximum idle count. Divide that 93

168 ppbrae 3000 User s Guide number by 2. Use the [Y/+] or [N/-] key to increase or decrease the stall value to equal that number. Press the [MODE] key to exit this display. Exiting Diagnostic Mode You can exit Diagnostic Mode and go directly to Programming Mode or Monitor Mode as outlined above, or you can exit Diagnostic Mode completely. To exit Diagnostic Mode so that it cannot be re-entered without a restart: Shut down the instrument. When it is off, restart it by holding the [MODE] key. Diagnostic Mode cannot be entered until the instrument is restarted as outlined in Entering Diagnostic Mode. 94

169 ppbrae 3000 User s Guide Transferring Data To & From A Computer Once you have connected your instrument cradle to the PC, you can can transfer data, including a download of the datalog to the computer and updates of firmware to the instrument (should this ever be necessary). Downloading The Datalog To A PC 1. Connect the data cable to the PC and the cradle. 2. Place the instrument into its cradle. The charging LED should be illuminated. 3. Start ProRAE Studio on your PC. 4. From ProRAE Studio, select Operation and select Setup Connection. 5. Select the COM port to establish a communication link between the PC and the instrument. 6. To receive the datalog in the PC, select Downlog Datalog. 7. When you see Unit Information, click OK. During the data transfer, the display shows a progress bar. When the transfer is done, you will see a screen with the datalog information. You can now export this datalog for other use or printing. 95

170 ppbrae 3000 User s Guide Uploading Firmware To The instrument From A PC Uploading new firmware to your instrument requires connecting the instrument and PC. Follow these steps to make the connection: 1. Connect the data cable to the PC and the cradle. 2. Place the instrument into its cradle. The charging LED should be illuminated. 3. Start RAEProgrammer 7000 on your PC. 4. From RAEProgrammer 7000, select Operation and select Setup Connection. 5. Select the COM port to establish a communication link between the PC and the instrument. 6. Select Operation Download Firmware. Once communication is established, follow the instructions that accompany RAEProgrammer 7000 and the firmware to upload the new firmware to your instrument. Note: Check for the latest updates to ProRAEProgrammer 7000 at 96

171 ppbrae 3000 User s Guide Maintenance The major maintenance items of the instrument are: Battery pack Sensor module PID lamp Sampling pump Inlet connectors and filters Note: Maintenance should be performed by qualified personnel only. NOTE: The printed circuit board of the instrument is connected to the battery pack even if the power is turned off. Therefore, it is very important to disconnect the battery pack before servicing or replacing any components inside the instrument. Severe damage to the printed circuit board or battery may occur if the battery pack is not disconnected before servicing the unit. Battery Charging & Replacement When the display shows a flashing empty battery icon, the battery requires recharging. It is recommended to recharge the instrument upon returning from fieldwork. A fully charged battery runs a instrument for 16 hours continuously. The charging time is less than 8 hours for a fully discharged battery. The battery may be replaced in the field (in areas known to be non-hazardous), if required. WARNING! To reduce the risk of ignition of hazardous atmospheres, recharge battery only in area known to be non-hazardous. Remove and replace battery only in areas known to be non-hazardous. 97

172 ppbrae 3000 User s Guide Replacing The Li-ion Battery 1. Turn off the instrument. 2. Located on the rear of the instrument is a battery tab. Slide it down to unlock the battery. 3. Remove the battery pack from the battery compartment by tilting it out. 4. Replace a fully charged spare battery pack inside the battery compartment. Make sure the battery pack is oriented properly inside the compartment. 5. Slide the capture tab back up to its locked position. Replacing The Alkaline Battery Adapter An alkaline battery adapter is supplied with each instrument. The adapter (part number ) accepts four AA alkaline batteries (use only Duracell MN1500) and provides approximately 12 hours of operation. The adapter is intended to be used in emergency situations when there is no time to charge the Li-ion battery pack. To insert batteries into the adapter: 1. Remove the three Philips-head screws to open the compartment. 2. Insert four fresh AA batteries as indicated by the polarity (+/-) markings. 3. Replace the cover. Replace the three screws. 98

173 ppbrae 3000 User s Guide To install the adapter in the instrument: 1. Remove the Li-ion battery pack from the battery compartment by sliding the tab and tilting out the battery. 2. Replace it with the alkaline battery adapter 3. Slide the tab back into place to secure the battery adapter. IMPORTANT! Alkaline batteries cannot be recharged. The instrument s internal circuit detects alkaline batteries and will not allow recharging. If you place the instrument in its cradle, the alkaline battery will not be recharged. The internal charging circuit is designed to prevent damage to alkaline batteries and the charging circuit when alkaline batteries are installed inside the instrument. Note: When replacing alkaline batteries, dispose of old ones properly. WARNING! To reduce the risk of ignition of hazardous atmospheres, recharge the battery only in areas known to be non-hazardous. Remove and replace the battery only in areas known to be non-hazardous. Note: The internal charging circuit is designed to prevent charging to alkaline batteries. 99

174 ppbrae 3000 User s Guide PID Sensor & Lamp Cleaning/Replacement The sensor module is made of several components and is attached to the lamp-housing unit as shown below. Sensor Cover Assembly O-ring, 35mm x 2mm Filter Fixer Lamp 1/2" (10.6eV) PN Seal Nut Inlet Probe Assembly PN Sensor Components Note: The cleaning procedure is not normally needed. Clean the PID sensor module, the lamp and the lamp housing only when one of the following has happened: 1. The reading is inaccurate even after calibration. 2. The reading is very sensitive to air moisture. 3. A chemical liquid has been sucked into the unit and damaged the unit. Use of the external filter helps to prevent contamination of the sensor. To access the sensor components and lamp, gently unscrew the lamphousing cap, remove the sensor adapter with the gas inlet probe and the metal filter all together. Then hold the PID sensor and pull it straight out. A slight, gentle rocking motion helps release the sensor. 100 Porous Metal Filter Sensor Detector PN Teflon O-ring Stainless Steel Washer O-ring, 36.5mm x 2.65mm Sensor Module Assembly PN FRU THP (temperature, humidity) Sensor Module PN FRU

175 ppbrae 3000 User s Guide Cleaning The PID Sensor Place the entire PID sensor module into GC grade methanol. It is highly recommended that an ultrasound bath to be used to clean the sensor for at least 15 minutes. Then dry the sensor thoroughly. Never touch the electrodes of the sensor by hand. Also use a methanol-soaked cotton swab to wipe off the lamp housing where it contacts the sensor when the sensor is installed. Turn over the sensor so that the pins point up and the sensor cavity is visible. Examine the sensor electrodes for any corrosion, damage, or bending out of alignment. The metal sensor electrode fingers should be flat and straight. If necessary, carefully bend the sensor fingers to ensure that they do not touch the Teflon portions and that they are parallel to each other. Make sure that the nuts on the sensor pins are snug but not overtight. If the sensor is corroded or otherwise damaged, it should be replaced. Cleaning The Lamp Housing Or Changing The Lamp If the lamp does not turn on, the instrument will display an error message to indicate replacement of the lamp may be required. 1. If the lamp is operational, clean the lamp window surface and the lamp housing by wiping it with GC grade methanol using a cotton swab using moderate pressure. After cleaning, hold the lamp up to the light at an angle to detect any remaining film. Repeat the process until the lamp window is clean. Never use water solutions to clean the lamp. Dry the lamp and the lamp housing thoroughly after cleaning. CAUTION: Never touch the window surface with the fingers or anything else that may leave a film. Never use acetone or aqueous solutions. 2. If the lamp does not turn on, remove the lamp from the lamp housing. Place the lamp O-ring onto the new lamp. Insert the new lamp, avoiding contact with the flat window surface. 3. Reinstall the PID sensor module. 4. Tighten the Lamp Housing Cap. 101

176 ppbrae 3000 User s Guide Determining The Lamp Type The monitor can accommodate two lamp values: 10.6eV (standard) and 9.8eV. There are two ways to determine the lamp type. 1. Turn off the instrument and remove the lamp. Now look at the serial number. The following identify the lamp type: 10.6eV SN: 106 2Nxxxxx 9.8eV SN: 098 2Nxxxxx 2. When the monitor is running, the lamp type is shown along with the calibration and measurement gas and Correction Factor: Note: This screen can be accessed from the reading screen by pressing [N/-] four times. Programming The Lamp ID The correct measurement gas library is used by the instrument when you ensure that the right lamp value is programmed. To manually select the Lamp ID: 1. Enter the Programming menu. 2. Select Monitor Setup. 3. Scroll down and select the Lamp ID sub-menu. 4. Press [N/-] to scroll down to the desired Lamp ID. 5. Press [Y/+] to select. 6. Press [MODE] to select Done. 7. Select Save. 8. Return to the main menu. Recalibrate the instrument before returning it to service. 102

177 ppbrae 3000 User s Guide Sampling Pump When approaching the end of the specified lifetime of the pump, it will consume higher amount of energy and reduce its sample draw capability significantly. When this occurs, it is necessary to replace or rebuild the pump. When checking the pump flow, make sure that the inlet connector is tight and the inlet tubing is in good condition. Connect a flow meter to the gas inlet probe. The flow rate should be above 450 cc/min when there is no air leakage. If the pump is not working properly, refer the instrument to qualified service personnel for further testing and, if necessary, pump repair or replacement. Cleaning The Instrument Occasional cleaning with a soft cloth is recommended. Do not use detergents or chemicals. Visually inspect the contacts at the base of the instrument, on the battery, and on the charging cradle to make sure they are clean. If they are not, wipe them with a soft, dry cloth. Never use solvents or cleaners. Ordering Replacement Parts If you need replacement parts, contact your local RAE Systems distributor. A list is available online: In the U.S., you can order sensors, replacement batteries, and other accessories online at: 103

178 ppbrae 3000 User s Guide Special Servicing Note If the instrument needs to be serviced, contact either: 1. The RAE Systems distributor from whom the instrument was purchased; they will return the instrument on your behalf. or 2. The RAE Systems Technical Service Department. Before returning the instrument for service or repair, obtain a Returned Material Authorization (RMA) number for proper tracking of your equipment. This number needs to be on all documentation and posted on the outside of the box in which the instrument is returned for service or upgrade. Packages without RMA Numbers will be refused at the factory. 104

179 Troubleshooting Problem Cannot turn on power after charging the battery ppbrae 3000 User s Guide Possible Reasons & Solutions Reasons: Discharged battery. Defective battery. Solutions: Charge or replace battery. Lost password Solutions: Call Technical Support at or toll-free at Reading abnormally High Reasons: Dirty filter. Dirty sensor module. Excessive moisture and water condensation. Incorrect calibration. Reading abnormally Low Buzzer Inoperative Solutions: Reasons: Solutions: Reasons: Solutions: Replace filter. Blow-dry the sensor module. Calibrate the unit. Dirty filter. Dirty sensor module. Weak or dirty lamp. Incorrect calibration. Replace filter. Remove Calibration Adapter. Calibrate the unit. Check for air leakage. Bad buzzer. Check that buzzer is not turned off. Call authorized service center. 105

180 ppbrae 3000 User s Guide Inlet flow too low Reasons: Pump diaphragm damaged or has debris. Flow path leaks. Lamp message during operation Solutions: Reasons: Solutions: Check flow path for leaks; sensor module O-ring, tube connectors, Teflon tube compression fitting. Call Technical Support at or toll-free at Lamp drive circuit. Weak or defective PID lamp, defective. Turn the unit off and back on. Replace UV lamp Technical Support To contact RAE Systems Technical Support Team: Monday through Friday, 7:00AM to 5:00PM Pacific (US) Time Phone (toll-free): Phone: tech@raesystems.com 106

181 ppbrae 3000 User s Guide RAE Systems Contacts RAE Systems by Honeywell World Headquarters 3775 N. First St. San Jose, CA USA Phone: RAE-tech@honeywell.com Web Site: RAE Systems Technical Support Monday through Friday, 7:00AM to 5:00PM Pacific (US) Time Phone (toll-free): Phone: Fax: RAE-tech@honeywell.com 107

182 ppbrae 3000 User s Guide Controlled Part Of Manual Intrinsic Safety: US and Canada: Class I, Division 1, Groups A,B,C,D T4 Europe: ATEX (0575 Ex II 2G Ex ia IIC/IIB T4 Gb) KEMA 07 ATEX 0127 Complies with EN :2009, EN :2007 IECEx CSA Ex ia IIC/IIB T4 Gb Complies with IEC :2007, IEC :2006 Temperature: -20º C to 50º C (-4º to 122º F) Humidity: 0% to 95% relative humidity (noncondensing) Basic Operation Turning The Instrument On 1. With the instrument turned off, press and hold [MODE]. 2. When the display turns on, release the [MODE] key. The instrument is now operating and performs self tests. Once the self tests are complete, the display shows a graph or numerical gas reading. This indicates that the instrument is fully functional and ready to use. Turning The Instrument Off 1. Press and hold the Mode key for 3 seconds. A 5-second countdown to shutoff begins. 2. When you see Unit off... release your finger from the [MODE] key. The instrument is now off. 108

183 ppbrae 3000 User s Guide Note: You must hold your finger on the key for the entire shutoff process. If you remove your finger from the key during the countdown, the shutoff operation is canceled and the instrument continues normal operation. 109

184 ppbrae 3000 User s Guide Alarm Signals During each measurement period, the gas concentration is compared with the programmed alarm limits (gas concentration alarm limit settings). If the concentration exceeds any of the preset limits, the loud buzzer and red flashing LED are activated immediately to warn you of the alarm condition. In addition, the instrument alarms if one of the following conditions occurs: battery voltage falls below a preset voltage level, failure of the UV lamp, pump stall, or when the datalog memory is full. Alarm Signal Summary Message Condition Alarm Signal HIGH OVR MAX LOW TWA STEL Pump icon flashes Gas exceeds High Alarm limit Gas exceeds measurement range Gas exceeds electronics maximum range Gas exceeds Low Alarm limit Gas exceeds TWA limit Gas exceeds STEL limit Pump failure 3 beeps/flashes per second* 3 beeps/flashes per second* 3 beeps/flashes per second* 2 beeps/flashes per second* 1 Beep/flash per second* 1 Beep/flash per second* 3 beeps/flashes per second Lamp PID lamp failure 3 beeps/flashes per second plus Lamp message on display 110

185 ppbrae 3000 User s Guide Battery icon flashes CAL NEG Low battery Calibration failed, or needs calibration Gas reading measures less than number stored in calibration 1 flash, 1 beep per minute plus battery icon flashes on display 1 beep/flash per second 1 beep/flash per second Preset Alarm Limits & Calibration The instrument is factory calibrated with standard calibration gas, and is programmed with default alarm limits. Cal Gas (Isobutylene) ppbrae 3000 MiniRAE 3000 MiniRAE Lite UltraRAE 3000 Cal Span unit Low High TWA STEL 10 ppm ppm ppm ppm

186 ppbrae 3000 User s Guide Charging The Battery Always fully charge the battery before using the instrument. The instrument s Li-ion/NiMH battery is charged by placing the instrument in its cradle. Contacts on the bottom of the instrument meet the cradle s contacts, transferring power without other connections. Note: Before setting the instrument into its charging cradle, visually inspect the contacts to make sure they are clean. If they are not, wipe them with a soft cloth. Do not use solvents or cleaners. Follow this procedure to charge the instrument: 1. Plug the AC/DC adapter s barrel connector into the instrument s cradle. 2. Plug the AC/DC adapter into the wall outlet. 3. Place the instrument into the cradle, press down, and lean it back. It locks in place and the LED in the cradle glows. Note: To release the instrument, press down and tilt the top out of the cradle and lift up. The instrument begins charging automatically. The LED on the front of the cradle marked Primary blinks during charging. During charging, the diagonal lines in the battery icon on the instrument s display are animated and you see the message Charging... When the instrument s battery is fully charged, the battery icon is no longer animated and shows a full battery. The message Fully charged! is shown and the Primary LED on the cradle glows continuously green. 112

187 ppbrae 3000 User s Guide Note: A spare Li-ion battery ( ) or NiMH( ) can be charged by placing it directly in the charging port on the back of the cradle. It can be charged at the same time as the instrument. Press the battery in place, sliding it slightly toward the front of the cradle. This locks it in the cradle. To release the battery, slide it forward again and tilt it up. Note: An Alkaline Battery Adapter (part number ), which uses four AA alkaline batteries (Duracell MN1500), may be substituted for the Li-Ion battery. WARNING! To reduce the risk of ignition of hazardous atmospheres, recharge and replace batteries only in areas known to be non-hazardous. Remove and replace batteries only in areas known to be nonhazardous. Low Voltage Warning When the battery s charge falls below a preset voltage, the instrument warns you by beeping once and flashing once every minute, and the battery icon blinks once per second. You should turn off the instrument within 10 minutes and either recharge the battery by placing the instrument in its cradle, or replace the battery with a fresh one with a full charge. Clock Battery An internal clock battery is mounted on one of the instrument s printed circuit boards. This long-life battery keeps settings in memory from being lost whenever the Li-ion, NiMH, or alkaline batteries are removed. This backup battery should last approximately five years, and must be replaced by an authorized RAE Systems service technician. It is not user-replaceable. 113

188 ppbrae 3000 User s Guide WARNING To reduce the risk of ignition of hazardous atmospheres, recharge battery only in area known to be non-hazardous. Remove and replace battery only in an area known to be non-hazardous. Replacing Rechargeable Li-Ion or NiMH Battery Caution: Turn off the instrument before removing or replacing the battery. Alkaline Battery Adapter An alkaline battery adapter is supplied with each instrument. The adapter (part number ) accepts four AA alkaline batteries (use only Duracell MN1500). Do not mix old and new batteries or different type batteries. 114

189 ppbrae 3000 User s Guide Troubleshooting Problem Cannot turn on power after charging the battery Possible Reasons & Solutions Reasons: Discharged battery. Defective battery. Solutions: Charge or replace battery. Lost password Solutions: Call Technical Support at or tollfree at Reading abnormally High Reasons: Dirty filter. Dirty sensor module. Excessive moisture and water condensation. Incorrect calibration. Reading abnormally Low Buzzer Inoperative Solutions: Reasons: Solutions: Reasons: Solutions: Replace filter. Blow-dry the sensor module. Calibrate the unit. Dirty filter. Dirty sensor module. Weak or dirty lamp. Incorrect calibration. Replace filter. Remove Calibration Adapter. Calibrate the unit. Check for air leakage. Bad buzzer. Check that buzzer is not turned off. Call authorized service center. 115

190 ppbrae 3000 User s Guide Inlet flow too low Reasons: Pump diaphragm damaged or has debris. Flow path leaks. Lamp message during operation Solutions: Reasons: Solutions: Check flow path for leaks; sensor module O-ring, tube connectors, Teflon tube compression fitting. Call Technical Support at or toll-free at Lamp drive circuit. Weak or defective PID lamp, defective. Turn the unit off and back on. Replace UV lamp 116

191

192 RAE Systems by Honeywell World Headquarters 3775 N. First St. San Jose, CA USA Phone: Toll-Free: Fax: (technical support): Web Site: WORLDWIDE SALES OFFICES USA/Canada Europe / Middle East China Asia Pacific Rev. E February 2016 P/N

193 Technical Note TN /16/VK A GUIDELINE FOR PID INSTRUMENT RESPONSE CORRECTION FACTORS AND IONIZATION ENERGIES* RAE Systems PIDs can be used for the detection of a wide variety of gases that exhibit different responses. In general, any compound with ionization energy (IE) lower than that of the lamp photons can be measured.* The best way to calibrate a PID to different compounds is to use a standard of the gas of interest. However, correction factors have been determined that enable the user to quantify a large number of chemicals using only a single calibration gas, typically isobutylene. In our PIDs, correction factors can be used in one of three ways: 1. Calibrate the monitor with isobutylene in the usual fashion to read in isobutylene equivalents. Manually multiply the reading by the correction factor (CF) to obtain the concentration of the gas being measured. 2. Calibrate the unit with isobutylene in the usual fashion to read in isobutylene equivalents. Call up the correction factor from the instrument memory or download it from a personal computer and then call it up. The monitor will then read directly in units of the gas of interest. 3. Calibrate the unit with isobutylene, but input an equivalent, corrected span gas concentration when prompted for this value. The unit will then read directly in units of the gas of interest. * The term ionization energy is more scientifically correct and replaces the old term ionization potential. High-boiling ( heavy ) compounds may not vaporize enough to give a response even when their ionization energies are below the lamp photon energy. Some inorganic compounds like H 2 O 2 and NO 2 give weak response even when their ionization energies are well below the lamp photon energy. Example 1: With the unit calibrated to read isobutylene equivalents, the reading is 10 ppm with a 10.6 ev lamp. The gas being measured is butyl acetate, which has a correction factor of 2.6. Multiplying 10 by 2.6 gives an adjusted butyl acetate value of 26 ppm. Similarly, if the gas being measured were trichloroethylene (CF = 0.54), the adjusted value with a 10 ppm reading would be 5.4 ppm. Example 2: With the unit calibrated to read isobutylene equivalents, the reading is 100 ppm with a 10.6 ev lamp. The gas measured is m-xylene (CF = 0.43). After downloading this factor, the unit should read about 43 ppm when exposed to the same gas, and thus read directly in m-xylene values. Example 3: The desired gas to measure is ethylene dichloride (EDC). The CF is 0.6 with an 11.7 ev lamp. During calibration with 100 ppm isobutylene, insert 0.6 times 100, or 60 at the prompt for the calibration gas concentration. The unit then reads directly in EDC values. Conversion to mg/m3 To convert from ppm to mg/m3, use the following formula: For air at 25 C (77 F), the molar gas volume is 24.4 L/mole and the formula reduces to: RAE Systems by Honeywell raesystems.com 1

AMAIRCARE HEPA BYPASS AIR FILTRATION SYSTEM READ AND SAVE THESE INSTRUCTIONS! AIR CLEANERS/ FILTERS. Rules for Safe Installation and Operation

AMAIRCARE HEPA BYPASS AIR FILTRATION SYSTEM READ AND SAVE THESE INSTRUCTIONS! AIR CLEANERS/ FILTERS. Rules for Safe Installation and Operation 2004 Americair Corporation, Mississauga, Ontario, Canada AIR CLEANERS/ FILTERS AMAIRCARE HEPA BYPASS AIR FILTRATION SYSTEM HOMEOWNERS MANUAL & DEALER INSTALLATION INSTRUCTIONS FOR MODELS AWW-350, AWW-675,

More information

HEPA AIR FILTRATION SYSTEM

HEPA AIR FILTRATION SYSTEM PRODUCT MANUAL FOR Airwash PRO HEPA AIR FILTRATI SYSTEM Table of Contents Rules for Safe Operation...1 Shipping and Packing List...1 Parts Identification/Dimensions...2 Operation...3/4 Product Application

More information

Safety & Installation Instructions

Safety & Installation Instructions Model 8120A & 8126A Digital Ventilation Controller Safety & Installation Instructions READ AND SAVE THESE INSTRUCTIONS Table of contents Safety Instructions... 3 Specifications... 4 Overview... 4 Mounting

More information

we make clean air Air Filtration Solutions Industrial Commercial Residential quatroair.com AIR TECHNOLOGIES INC.

we make clean air Air Filtration Solutions Industrial Commercial Residential quatroair.com AIR TECHNOLOGIES INC. we make clean air Air Filtration Solutions Industrial Commercial Residential Che mic al F ilte r HE PA Filt er Filter Although very effective against very fine particles, filters do not filter out gasses

More information

OWNER S MANUAL MODEL FSHEPA375 WHOLE-HOUSE TRUE HEPA AIR CLEANER IMPORTANT: PLEASE READ MANUAL BEFORE OPERATING UNIT

OWNER S MANUAL MODEL FSHEPA375 WHOLE-HOUSE TRUE HEPA AIR CLEANER IMPORTANT: PLEASE READ MANUAL BEFORE OPERATING UNIT OWNER S MANUAL MODEL FSHEPA375 WHOLE-HOUSE TRUE HEPA AIR CLEANER IMPORTANT: PLEASE READ MANUAL BEFORE OPERATING UNIT Features Long-life, Certified True HEPA (High Efficiency Particulate Air) Filter is

More information

PUREAIR AIR PURIFICATION SYSTEM PHOTOCATALYTIC OXIDATION (PCO) TECHNOLOGY

PUREAIR AIR PURIFICATION SYSTEM PHOTOCATALYTIC OXIDATION (PCO) TECHNOLOGY 507477-01 4/2016 Supersedes 4/2015 AIR CLEANERS / FILTERS Litho U.S.A. PUREAIR AIR PURIFICATION SYSTEM PHOTOCATALYTIC OXIDATION (PCO) TECHLOGY INSTALLATION INSTRUCTIONS FOR PUREAIR AIR PURIFICATION SYSTEM

More information

KenkoAir_Purifier manual ENG-FR_FDA 10/12_Air Filter manual 8/15/13 10:44 AM Page 1 OPERATING MANUAL

KenkoAir_Purifier manual ENG-FR_FDA 10/12_Air Filter manual 8/15/13 10:44 AM Page 1 OPERATING MANUAL KenkoAir_Purifier manual ENG-FR_FDA 10/12_Air Filter manual 8/15/13 10:44 AM Page 1 OPERATING MANUAL KenkoAir_Purifier manual ENG-FR_FDA 10/12_Air Filter manual 8/15/13 10:44 AM Page 2 PRODUCT INTRODUCTION

More information

Cleanshield HEPA 550 Owner s Manual

Cleanshield HEPA 550 Owner s Manual Cleanshield HEPA 550 Owner s Manual READ AND SAVE THESE INSTRUCTIONS E-mail: sales@alorair.com Add: North Carolina USA Caution! Don t adjust or alter the Cleanshield HEPA 550 in any way.use the parts approved

More information

INSTALLATION, OPERATION, AND MAINTENANCE INSTRUCTIONS. Part Number: P DM

INSTALLATION, OPERATION, AND MAINTENANCE INSTRUCTIONS. Part Number: P DM INSTALLATION, OPERATION, AND MAINTENANCE INSTRUCTIONS Whole House Duct-Mount HEPA Air Cleaner Part Number: P102-400DM TABLE OF CONTENTS Page SAFETY CONSIDERATIONS...1 INTRODUCTION...1,2 INSTALLATION...2,3

More information

UV-AIRE AIR PURIFYING SYSTEM

UV-AIRE AIR PURIFYING SYSTEM UV-AIRE AIR PURIFYING SYSTEM Model: UV-16/24 The UV-Aire unit is designed to emit powerful UVC Band light rays, which sterilize and reduce airborne microorganisms as they pass through a heating or air

More information

INSTALLATION GUIDE. Greentek. Residential use only MODEL HS 3.0

INSTALLATION GUIDE. Greentek. Residential use only MODEL HS 3.0 INSTALLATION GUIDE Residential use only MODEL HS 3.0 ITEM# 463625 OCT2018 Product of Greentek Greentek reserves the rights to modify a product, without prior notice, whether in price, design, color or

More information

AeraMax Pro AM IV. Please read these instructions before use. Do not discard: keep for future reference.

AeraMax Pro AM IV. Please read these instructions before use. Do not discard: keep for future reference. AeraMax Pro AM IV Please read these instructions before use. Do not discard: keep for future reference. ENGLISH AeraMax PRO Air Quality Control System: AM IV SAFETY INSTRUCTIONS/WARNING Thank you for purchasing

More information

Model: UV-16R/24 READ THESE INSTRUCTIONS CAREFULLY AND COMPLETELY BEFORE PROCEEDING WITH THE INSTALLATION.

Model: UV-16R/24 READ THESE INSTRUCTIONS CAREFULLY AND COMPLETELY BEFORE PROCEEDING WITH THE INSTALLATION. U V - AIRE AIR PURIFYING S YSTEM Model: UV-16R/24 ITEMS INCLUDED IN KIT: 1-24Volt UV-Aire Ballast Assembly 1-Mounting Base 1-Ultraviolet Lamp 6-Sheet Metal Mounting Screws 1-Instruction Sheet 1-Fiberglass

More information

TOTALINE ELECTRONIC AIR CLEANERS

TOTALINE ELECTRONIC AIR CLEANERS TOTALINE ELECTRONIC AIR CLEANERS TOTALINE GOLD SERIES ELECTRONIC AIR CLEANERS Totaline Gold is our premium line of Electronic Air Cleaners. Designed with state-of-the-art technology, they are easy to install

More information

WSB clean: Installation Manual

WSB clean: Installation Manual WSB clean: Installation Manual Models: WSB 400 WSB 1250 WSB 500 WSB 1500 WSB 600 WSB 1600 WSB 750 WSB 1800 WSB 1000 TABLE OF CONTENTS 1 INTRODUCTION...3 2 SAFETY INFORMATION...3 3 WSB CLEAN SYSTEM OPERATION...4

More information

User s Guide. Oreck Air Purifier with HEPA Filtration. Important! Read this manual carefully, and keep for future reference.

User s Guide. Oreck Air Purifier with HEPA Filtration. Important! Read this manual carefully, and keep for future reference. User s Guide Oreck Air Purifier with HEPA Filtration Important! Read this manual carefully, and keep for future reference. Enjoy Congratulations on your purchase of the Oreck Air Purifier with HEPA Filtration.

More information

READ AND SAVE THESE INSTRUCTIONS. Air Boss MP600M Vertical Air Flow Mist Precipitator Industrial Applications. TRION

READ AND SAVE THESE INSTRUCTIONS. Air Boss MP600M Vertical Air Flow Mist Precipitator Industrial Applications. TRION READ AND SAVE THESE INSTRUCTIONS Vertical Air Flow Mist Precipitator Industrial Applications TRION Vertical Air Flow Mist Precipitator for Industrial Applications Table of Contents Design...2 Installation...2

More information

Air Purifiers Product Guide

Air Purifiers Product Guide Air Purifiers Product Guide Trusted Solutions for a Healthy Home A high performance filter for better indoor air and system protection One-inch filters have been the standard on system installs for years

More information

Model 8144NC Fresh Air Ventilator

Model 8144NC Fresh Air Ventilator Model 8144NC Fresh Air Ventilator Installation and Operating Instructions MOUNTING BRACKETS WIRE ENTRY LOCATION INTEGRAL PRESSURE PORTS (PORT ON INLET SIDE NOT SHOWN) OVAL OUTLET COLLAR FOR 6" DIAMETER

More information

INSTALLATION / OPERATION / MAINTENANCE GUIDE

INSTALLATION / OPERATION / MAINTENANCE GUIDE INSTALLATION / OPERATION / MAINTENANCE GUIDE WHOLE HOUSE HEPA FILTRATION SYSTEMS MODEL HS 3.0 Table of contents About Us Imperial Air Technologies Inc.is the only manufacturer that offers you a complete

More information

Science Applications International Corporation

Science Applications International Corporation February 7, 2011 Ms. Sharon R. Fisher 1425 Eden Road York, PA 17402 Re: 2010 Summary of Operation for North Building 4 Soil Vapor Extraction System SAIC Project 4501020172 / 5000 / 100 Dear Sharon: Science

More information

PHOENIX GUARDIAN H. E. P. A. SYSTEM

PHOENIX GUARDIAN H. E. P. A. SYSTEM PHOENIX GUARDIAN H. E. P. A. SYSTEM Therma-Stor Products Box 8050 Madison, WI 53708 Toll Free 1-800-533-7533 Local 1-608-222-5301 PHOENIX GUARDIAN Operation & Service Instructions Table of Contents 1.

More information

A. Each cabinet must be constructed and installed in accordance with EN12469:2000.

A. Each cabinet must be constructed and installed in accordance with EN12469:2000. BIOLOGICAL SAFETY CABINETS Class II, Biological Safety Cabinet PART 1 GENERAL 1.1 REFERENCES The publications listed below form a part of this section to the extent referenced. The publications are referred

More information

Charcoal Filter Installation NOTE: The charcoal filters are preinstalled if you purchased the range hood with re-circulating kit from us. 1. Remove st

Charcoal Filter Installation NOTE: The charcoal filters are preinstalled if you purchased the range hood with re-circulating kit from us. 1. Remove st Venting Methods This range hood is factory set for venting through the roof or wall. For non-vented (re-circulating) installations, see Ductless Conversion on Page 7. Vent work can terminate either through

More information

Ducting & Ventilation Components

Ducting & Ventilation Components Ducting & Ventilation Components 111 Powell Road Independence, MO 64056 visit www.hemcocorp.com Phone: 816-796-2900 Fax : 816-796-3333 E-Mail : info@hemcocorp.com Exhaust Blowers are constructed of chemical

More information

VWR Zero Air Generator

VWR Zero Air Generator Technical Information Installation, Operation, and Maintenance Manual VWR 89237-564 Zero Air Generator Figure 1 - VWR Zero Air Generator These instructions must be thoroughly read and understood before

More information

Miracle State Of The Art Air Filtration System!

Miracle State Of The Art Air Filtration System! I N D O O R A I R Q U A L I T Y S O L U T I O N S Miracle State Of The Art Air Filtration System! High efficiency particulate (HEPA) filter combined with optimized sorbent media provides the ultimate solution

More information

PHOENIX MINI-GUARDIAN

PHOENIX MINI-GUARDIAN PHOENIX MINI-GUARDIAN H. E. P. A. SYSTEM Therma-Stor Products A Division of DEC International Box 8050 Madison, WI 53708 Toll Free 1-800-533-7533 Local 1-608-222-5301 Table of Contents PHOENIX MINI-GUARDIAN

More information

Operations & Maintenance Manual DFH-1000 DUCTLESS FUME HOOD

Operations & Maintenance Manual DFH-1000 DUCTLESS FUME HOOD Operations & Maintenance Manual DFH-1000 DUCTLESS FUME HOOD ----------------------------------------------------------------We Make Clean Air----------------------------------------------------------------

More information

SPECIAL REVERSE HORIZONTAL FLOW ENCLOSURE OPERATING AND MAINTENANCE MANUAL. Issue No: 1

SPECIAL REVERSE HORIZONTAL FLOW ENCLOSURE OPERATING AND MAINTENANCE MANUAL. Issue No: 1 SPECIAL REVERSE HORIZONTAL FLOW ENCLOSURE OPERATING AND MAINTENANCE MANUAL Issue No: 1 Certificate No: 0954499 CE Marking certifies that this equipment conforms to the following EEC directives: - Low Voltage

More information

IFD AIR CLEANER SERVICE MANUAL

IFD AIR CLEANER SERVICE MANUAL IFD AIR ER SERVICE MANUAL PUB NO. 34-3446-01 1 4 COMPONENTS OF THE AIR ER 5 3 6 3 2 7 8 1 2 3 AIR FLOW 1 1) Pre-filter - traps large particles such as hair and lint before they can enter the cell section.

More information

ELITE SERIES R-410A Ready - Multi-Position

ELITE SERIES R-410A Ready - Multi-Position PRODUCT SPECIFICATIONS AIR HANDLERS CBXM ELITE SERIES R-0A Ready - Multi-Position Bulletin No. 0 October 07 Supersedes June 0 Nominal Capacity -. to Tons Optional Electric Heat -. to 0 kw MODEL NUMBER

More information

Users Manual Certified Series Direct Drive Pump 1-7 LPM

Users Manual Certified Series Direct Drive Pump 1-7 LPM Users Manual Certified Series Direct Drive Pump 1-7 LPM Safety, Operating, Installation, and Maintenance Instructions 600 S 56 th Street #9 Chandler, AZ 85226 Phone: 480-507-6478 Fax: 480-838-2232 www.fogco.com

More information

Air Boss M-Series Media & Adsorption Air Cleaner

Air Boss M-Series Media & Adsorption Air Cleaner Air Boss M-Series Media & Adsorption Air Cleaner A i r P u r i f i c a t i o n S y s te m s w w w. t r i o n i a q. c o m Engineered Solutions For Clean Air TRION s Air Boss media units are industrial

More information

FA1 Series Air Filtration System. Operations Manual

FA1 Series Air Filtration System. Operations Manual FA1 Series Air Filtration System Operations Manual READ AND SAVE THESE INSTRUCTIONS WARNING: Disconnect from power supply before performing maintenance procedures such as cabinet cleaning, changing filters

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS MODELS 600HS, 1200HS, 2000HS PLUS HVAC 600HS L Housing 600HS S Housing 1200/2000HS S Housing 1200/2000HS L Housing 1 ASSEMBLY INSTRUCTIONS HVAC VERSION STEP 1. STEP 2. STEP 3.

More information

Operation and Maintenance

Operation and Maintenance E-Z Set Company requires regular inspection and maintenance of the E-Z Treat Re- Circulating Sand/Media Filter, as a condition of purchase and ongoing operation compliance. The mandatory service contract

More information

United States Stove Company Project # Model: Breckwell SP1000 Type: Pellet-Fired Room Heater July 25, 2018

United States Stove Company Project # Model: Breckwell SP1000 Type: Pellet-Fired Room Heater July 25, 2018 United States Stove Company Project # 18-415 Model: Breckwell SP1000 Type: Pellet-Fired Room Heater July 25, 2018 ASTM E2779 Standard Test Method for Determining Particulate Matter Emissions from Pellet

More information

Complete Whole House Air Cleaning System Photronic Base Media Air Cleaner OWNERS GUIDE

Complete Whole House Air Cleaning System Photronic Base Media Air Cleaner OWNERS GUIDE Complete Whole House Air Cleaning System Photronic Base Media Air Cleaner OWNERS GUIDE Air Handler Base Models: 92-052, 92-053, 92-054 (small) 24VAC- 92-055, 92-056, 92-057 Models: 92-046, 92-047, 92-048

More information

Installation, Operations and Maintenance Guide

Installation, Operations and Maintenance Guide Installation, Operations and Maintenance Guide AtmosAir 508 Unit MCC Tubes V E R S I O N 1 4. 0 AK P A G E 2 Table of Contents 1. Product Overview 3 1.1 Product Diagram 3 2. Installation 4 2.1 Mechanical

More information

X-3400A, X-3500, X-3700 OWNER S MANUAL PLEASE READ AND SAVE THESE INSTRUCTIONS

X-3400A, X-3500, X-3700 OWNER S MANUAL PLEASE READ AND SAVE THESE INSTRUCTIONS X-3400A, X-3500, X-3700 OWNER S MANUAL XPOWER MANUFACTURE, INC. 220 Clary Avenue San Gabriel, CA 91776, USA Tel: 626-285-3301, Fax: 626-285-330 PLEASE READ AND SAVE THESE INSTRUCTIONS 1 Index Introduction...2

More information

WORKSHEET FOR RESIDENTIAL AIR SYSTEM DESIGN page 1

WORKSHEET FOR RESIDENTIAL AIR SYSTEM DESIGN page 1 WORKSHEET FOR RESIDENTIAL AIR SYSTEM DESIGN page 1 Wallet card photocopy Inspection Authority: Signature: Date: / / Submitted For: ( Owner ) Name Address City _ Prov Postal code Phone ( ) Fax ( ) Designer/Signature:

More information

PLANNING THE INSTALLATION

PLANNING THE INSTALLATION a Same for Single Speed Models (Series 1 & 2).. Table 2. F111 Air Volume Specifications for 3 Speed Models (Series 3). Particulates at Gas, V.O.C. and High a Medium Low 0.3 Micron Odor Control in cfm in

More information

1 st Quarter Summary of Meteorological and Ambient Air Quality Data Kennecott Utah Copper Monitoring Stations. Prepared for:

1 st Quarter Summary of Meteorological and Ambient Air Quality Data Kennecott Utah Copper Monitoring Stations. Prepared for: 1 st Quarter 2018 Summary of Meteorological and Ambient Air Quality Data Kennecott Utah Copper Monitoring Stations Prepared for: Prepared by: Mr. Bryce C. Bird Director Division of Air Quality 195 North

More information

Source Capture Air Purification System

Source Capture Air Purification System Source Capture Air Purification System HA-SCP-G3 Owner s Manual Table of Contents HealthyAir HA-IFM-1111 Filter 1 Important Safety Instructions 2 Technical Specifications 3 Packaging Reference 4 Packing

More information

Portable Air Purifiers

Portable Air Purifiers O 0 FLTER 0 HEPA FLTER Portable Air Purifiers READ AND SAVE THESE NSTRUCTONS NSTALL PRE-FLTER BEFORE USE PRE-FLTER PRE-FLTER Distributed by: PREFLTER HEPA PREFLTER T M 0 HEPA FLTER PREFLTER.c om (888)

More information

OWNER S MANUAL C O M M E R C I A L I N D O O R A I R Q U A L I T Y S O L U T I O N S. Total High Efficiency Particulate & Gaseous Filtration System

OWNER S MANUAL C O M M E R C I A L I N D O O R A I R Q U A L I T Y S O L U T I O N S. Total High Efficiency Particulate & Gaseous Filtration System OWNER S MANUAL C O M M E R C I A L I N D O O R A I R Q U A L I T Y S O L U T I O N S Ever Total High Efficiency Particulate & Gaseous Filtration System State-of-the-art air cleaner is the ultimate solution

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS 2005 Lennox Industries Inc. Dallas, Texas, USA INSTALLATION INSTRUCTIONS CB29M & CB30M Elite Series Units MULTI POSITION BLOWER COILS 504,79M 09/05 Supersedes 04/05 Litho U.S.A. RETAIN THESE INSTRUCTIONS

More information

United States Stove Company Project # Model: KP60 AKA: AP60, VG60 Type: Pellet-Fired Room Heater May 31, 2018

United States Stove Company Project # Model: KP60 AKA: AP60, VG60 Type: Pellet-Fired Room Heater May 31, 2018 United States Stove Company Project # 18-409 AKA: AP60, VG60 Type: Pellet-Fired Room Heater May 31, 2018 ASTM E2779 Standard Test Method for Determining Particulate Matter Emissions from Pellet Heaters

More information

3 rd Quarter Summary of Meteorological and Ambient Air Quality Data Kennecott Utah Copper Monitoring Stations. Prepared for:

3 rd Quarter Summary of Meteorological and Ambient Air Quality Data Kennecott Utah Copper Monitoring Stations. Prepared for: 3 rd Quarter 2018 Summary of Meteorological and Ambient Air Quality Data Kennecott Utah Copper Monitoring Stations Prepared for: Prepared by: Mr. Bryce C. Bird Director Division of Air Quality 195 North

More information

Air Bear Supreme and Air Bear Cub. Media Air Cleaners READ AND SAVE THESE INSTRUCTIONS UNITS NOT FOR COMMERCIAL USE. TRION

Air Bear Supreme and Air Bear Cub. Media Air Cleaners READ AND SAVE THESE INSTRUCTIONS UNITS NOT FOR COMMERCIAL USE. TRION READ AND SAVE THESE INSTRUCTIONS UNITS NOT FOR COMMERCIAL USE Air Bear Supreme and Air Bear Cub Media Air Cleaners TRION www.trioniaq.com Table of Contents Introduction... 1 Unit Selection Guide... 2

More information

Maintenance Instructions

Maintenance Instructions Maintenance Instructions CX1000 Portable Air Purifier PRODUCT SPECIFICATIONS Manufacturer: Model: Unit Dimensions: Unit Weight: Max Airflow: Power Consumption: Voltage: Continental Fan Manufacturing Inc.

More information

EV Charging Station Submittal Requirements

EV Charging Station Submittal Requirements EV Charging Station Submittal Requirements To obtain a permit for an Electric Vehicle Charging Station, please provide the following: Plan and details of the charging system installation o Print applicable

More information

Page Number. Contents: Product Features 2. Physical Data 2. Nomenclature. 3. Blower Performance 3. Electrical Data 4. Warranty.. 6

Page Number. Contents: Product Features 2. Physical Data 2. Nomenclature. 3. Blower Performance 3. Electrical Data 4. Warranty.. 6 TC-SGRAH1-1005 October 05 R Series - High Efficiency Multi-position Air Handler Engineering and Specification Guide (Electric and No Heat) With optional field conversion downflow kit Contents: Page Number

More information

Fume Extraction Planning Guide

Fume Extraction Planning Guide Fume Extraction Planning Guide Which System Should I Buy?... 2 BVX-100 Configurations... 3 BVX-200 Configurations... 4 MFX-2200 Configurations... 5 Selecting Control of MFX-2200... 6 Selecting Exhaust

More information

MODEL FA1 Mini Air Filtration System. Operations Manual

MODEL FA1 Mini Air Filtration System. Operations Manual MODEL FA1 Mini Air Filtration System Operations Manual READ AND SAVE THESE INSTRUCTIONS WARNING: Disconnect from power supply before performing maintenance procedures such as cabinet cleaning, changing

More information

1st Quarter Summary of Meteorological and Ambient Air Quality Data Kennecott Utah Copper Monitoring Stations

1st Quarter Summary of Meteorological and Ambient Air Quality Data Kennecott Utah Copper Monitoring Stations 1st Quarter 2014 Summary of Meteorological and Ambient Air Quality Data Kennecott Utah Copper Monitoring Stations Steven A. Root Digitally signed by Steven A. Root DN: cn=steven A. Root, o=weatherbank,

More information

Electrical Safety Program

Electrical Safety Program Electrical Safety Program Medford Area Public School District 124 West State St School District: Medford, WI 54451 715 748 4620 Review Date(s): January 2018 Program Dave Makovsky Coordinator: Purpose and

More information

HOWORTH EXFLOW 28 PREVENTS AIRBORNE INFECTION TECHNICAL SPECIFICATION. Exflow 28 Technical Specification. Page 1 of 7

HOWORTH EXFLOW 28 PREVENTS AIRBORNE INFECTION TECHNICAL SPECIFICATION. Exflow 28 Technical Specification. Page 1 of 7 HOWORTH EXFLOW 28 PREVENTS AIRBORNE INFECTION TECHNICAL SPECIFICATION Exflow 28 Technical Specification Page 1 of 7 TECHNICAL SUMMARY Terminal Unit size 12-3 ¼ x 12-3 ¼ x 23 ¾ Terminal Unit weight 3527

More information

January 2018 PRICE SCHEDULE A 2.0 Vehicular Traffic Control

January 2018 PRICE SCHEDULE A 2.0 Vehicular Traffic Control Section A1 This Price Schedule is effective January 1, 2018. Date Page Comment 1-1-18 All January, 2018 Price Schedule update. 1-26-18 6, 7 Added 1601-180, 1601-181 Gray Housing and 1601-283 Gray arm hub

More information

Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by:

Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by: Storage System Ultrasound Probe Cabinet Operator's Manual Manufactured by: CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA 319.248.6757 / 800.445.6741 WWW.CIVCO.COM Copyright 2018 All

More information

3.5 Exposure Control Measures

3.5 Exposure Control Measures 3.5 Exposure Control Measures 3.5.1 Chemical Fume Hoods Chemical fume hoods are the most reliable and primary engineering control used in the laboratory to protect against the inhalation of hazardous vapors

More information

SECTION HVAC AIR CLEANING DEVICES

SECTION HVAC AIR CLEANING DEVICES SECTION 23 40 00 HVAC AIR CLEANING DEVICES PART 1 - GENERAL 1.1 SUMMARY A. Section includes activated carbon filters, automatic renewable media filters, disposable, extended area panel filters, disposable

More information

CLEAN AIR WORKSTATION REVERSE FLOW. Installation and Operations Manual

CLEAN AIR WORKSTATION REVERSE FLOW. Installation and Operations Manual CLEAN AIR WORKSTATION REVERSE FLOW Installation and Operations Manual CLEAN AIR WORKSTATION REVERSE FLOW Installation and Operations Manual Revision: 16-Jan-17 Page 1 of 12 CLEAN AIR WORKSTATION REVERSE

More information

Omniaire 1000V Instruction Manual-8/05

Omniaire 1000V Instruction Manual-8/05 The OMNIAIRE 1000V HEPA Air Filtration system is a variable speed HEPAfiltered negative air machine designed for removal of toxic mold, asbestos or other airborne contaminants during construction, renovation

More information

CRYSTAL-AIRE. concealed commercial air cleaners

CRYSTAL-AIRE. concealed commercial air cleaners CRYSTAL-AIRE concealed commercial air cleaners CRYSTAL-AIRE A Customized CONCEPT IN Concealed AIR CLEANING. Using a variety of accessories, Crystal-Aire can be adapted to most building configurations and

More information

NSGV PT-1000 PORTABLE WELDING STATION I, O & M MANUAL

NSGV PT-1000 PORTABLE WELDING STATION I, O & M MANUAL APPLICATION OF DUST CONTROL EQUIPMENT: CAUTION - Warning Improper operation of dust control system may contribute to conditions in the work area or facility that could result in severe personal injury

More information

WELCOME. Important Safety Instructions...2. Technical Specifications...2. Parts Description...3. Operating Instructions...4

WELCOME. Important Safety Instructions...2. Technical Specifications...2. Parts Description...3. Operating Instructions...4 WELCOME Congratulations! You are now the proud owner of one of the most advanced and effective room air cleaners available. Your new Defender helps reduce the many problem-causing pollutants and allergens

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS MiniCOREVentilator ENERGY RECOVERY CORE INSTALLATION INSTRUCTIONS MC500-1ERV JANUARY 10, 2018 SUPERSEDES: NONE INSTALLATION INSTRUCTIONS FOR MINICORE VENTILATOR (MCV) WITH FACTORY INSTALLED OPTIONS USED

More information

quatroair.com AIR TECHNOLOGIES INC.

quatroair.com AIR TECHNOLOGIES INC. quatroair.com AIR TECHNOLOGIES INC. LaserBlue Gel Zero Filter Bypass. Guaranteed! Designed to meet the most stringent cleanroom requirements, the LaserBlue gel forms an impenetrable barrier against the

More information

TXH TOTAL EXHAUST HOOD. Installation and Operations Manual

TXH TOTAL EXHAUST HOOD. Installation and Operations Manual TXH TOTAL EXHAUST HOOD Installation and Operations Manual TXH TOTAL EXHAUST HOOD Installation and Operations Manual Revision: 25-Nov-16 Page 2 of 12 TXH TOTAL EXHAUST HOOD Installation and Operations Manual

More information

Alen BreatheSmart HEPA Air Purifier USER MANUAL

Alen BreatheSmart HEPA Air Purifier USER MANUAL Alen BreatheSmart HEPA Air Purifier USER MANUAL Table of Contents Air Purifier Safety Using Your Air Purifier Operating Your Unit Identification Of Parts Know Your Air Purifier Care & Maintenance Replacing

More information

January 2018 PRICE SCHEDULE A 2.0 Vehicular Traffic Control

January 2018 PRICE SCHEDULE A 2.0 Vehicular Traffic Control Section A1 This Price Schedule is effective January 1, 2018. Date Page Comment 1-1-18 All January, 2018 Price Schedule update. 1-26-18 6, 7 Added 1601-180, 1601-181 Gray Housing and 1601-283 Gray arm hub

More information

Decision D ATCO Electric Ltd. Decommissioning of Transmission Line 6L82

Decision D ATCO Electric Ltd. Decommissioning of Transmission Line 6L82 Decision 21447-D01-2016 August 23, 2016 Decision 21447-D01-2016 Proceeding 21447 Application 21447-A001 August 23, 2016 Published by the: Fifth Avenue Place, Fourth Floor, 425 First Street S.W. Calgary,

More information

This is a new permit condition titled, "2D.1111 Subpart ZZZZ, Part 63 (Existing Non-Emergency nonblack start CI > 500 brake HP)"

This is a new permit condition titled, 2D.1111 Subpart ZZZZ, Part 63 (Existing Non-Emergency nonblack start CI > 500 brake HP) This is a new permit condition titled, "2D.1111 Subpart ZZZZ, Part 63 (Existing Non-Emergency nonblack start CI > 500 brake HP)" Note to Permit Writer: This condition is for existing engines (commenced

More information

Nilfisk-CFM SS Vapor Vacuum Instructions for Use and Spare Parts Manual

Nilfisk-CFM SS Vapor Vacuum Instructions for Use and Spare Parts Manual Nilfisk-CFM SS Vapor Vacuum Instructions for Use and Spare Parts Manual Caution: This Nilfisk-CFM vacuum cleaner is not to be used in explosion hazardous areas as serious injury could result. Your new

More information

City of Walnut Creek Development Review Services 1666 N. Main Street, Walnut Creek, CA (925) phone (925) fax

City of Walnut Creek Development Review Services 1666 N. Main Street, Walnut Creek, CA (925) phone (925) fax City of Walnut Creek Development Review Services 1666 N. Main Street, Walnut Creek, CA 94596 (925) 943-5834 phone (925) 256-3500 fax Issued September 30, 2017 Policy Bulletin No. PB-72 Background Residential

More information

MERIT Series R-410A - Upflow / Horizontal - Variable Speed

MERIT Series R-410A - Upflow / Horizontal - Variable Speed PRODUCT SPECIFICATIONS AIR HANDLERS CBX5UHV (-01) MERIT Series R-10A - Upflow / Horizontal - Variable Bulletin No. 10611 March 015 Supersedes November 01 MODEL NUMBER IDENTIFICATION CB X 5 UH V - 00-0

More information

The Premcor Refining Group, Inc. Delaware City Refinery 4550 Wrangle Hill Rd. EXHIBIT A Delaware City, DE 19706

The Premcor Refining Group, Inc. Delaware City Refinery 4550 Wrangle Hill Rd. EXHIBIT A Delaware City, DE 19706 DRAFT Permit: APC-2004/0721-CONSTRUCTION (NSPS) Two Package Boilers The Premcor Refining Group, Inc. Delaware City Refinery 4550 Wrangle Hill Rd. EXHIBIT A Delaware City, DE 19706 ATTENTION: Andrew Kenner

More information

A I R H A N D L E R S CB26UH-R. MERIT Series R-22 - Upflow / Horizontal P R O D U C T S P E C I F I C AT I O N S

A I R H A N D L E R S CB26UH-R. MERIT Series R-22 - Upflow / Horizontal P R O D U C T S P E C I F I C AT I O N S P R O D U C T S P E C I F I C AT I O N S A I R H A N D L E R S CB6UH-R MERIT Series R- - Upflow / Horizontal Bulletin No. 1067 April 01 Supersedes January 01 Nominal Capacity - 1.5 to 5 Tons Optional Electric

More information

Prepared for: Prepared by: 4005 Port Chicago Highway Concord, California

Prepared for: Prepared by: 4005 Port Chicago Highway Concord, California AIR MONITORING SUMMARY REPORT Time-Critical Removal Action for the Experimental Ship Shielding Range, Parcel E-2 Hunters Point Naval Shipyard San Francisco, California Contract Number N62473-10-D-0807

More information

Electric Vehicle Charging Station Article California Electrical code (CEC) General Requirements:

Electric Vehicle Charging Station Article California Electrical code (CEC) General Requirements: Checklist for Multi-Family Residential Electric Vehicle Charging Station Article 625 2016 California Electrical code (CEC) General Requirements: Level 1 Charger: 110V dedicated 20 amp circuit No electrical

More information

For Mini Split Units. Cleaner, Fresher Air in 24 Hours New Advanced Environmental Air Treatment System

For Mini Split Units. Cleaner, Fresher Air in 24 Hours New Advanced Environmental Air Treatment System Cleaner, Fresher Air in 24 Hours New Advanced Environmental Air Treatment System For Mini Split Units NOTE: Please take time to read the attached assembly instructions. All instructions and precautions

More information

PHOTO-CATALYTIC AIR PURIFIER SYSTEMS

PHOTO-CATALYTIC AIR PURIFIER SYSTEMS PHOTO-CATALYTIC AIR PURIFIER SYSTEMS Models: DUO-11/24, DUO-11/120, DUO-14/24, DUO-14/120, PCUV-16UA, DUO-16/24, DUO-16/120 *includes carbon filter ITEMS INCLUDED IN KIT: 1-24 Volt or 120 Volt UV-Aire

More information

TECHNICAL SPECIFICATIONS. FEATURES and BENEFITS WARRANTY

TECHNICAL SPECIFICATIONS. FEATURES and BENEFITS WARRANTY TECHNICAL SPECIFICATIONS KG7S (C and L Series) High Efficiency / Direct Vent or Non Direct Vent Condensing Upflow/Horizontal and Downflow Gas Furnaces Induced Draft - 92. AFUE Input 38,000-20,000 Btuh

More information

Snapshot LX5 USER MANUAL. OK on Dimmer Outdoor OK Sound Activated DMX512 Master/Slave 115V/230V Switch Replaceable Fuse User Serviceable Duty Cycle

Snapshot LX5 USER MANUAL. OK on Dimmer Outdoor OK Sound Activated DMX512 Master/Slave 115V/230V Switch Replaceable Fuse User Serviceable Duty Cycle LX5 Snapshot OK on Dimmer Outdoor OK Sound Activated DMX512 Master/Slave 115V/230V Switch Replaceable Fuse User Serviceable Duty Cycle USER MANUAL Chauvet, 5200 NW 108th Avenue, Sunrise, FL 33351 U.S.A.

More information

OmniAire Portable Air Cleaner OA1200PAC

OmniAire Portable Air Cleaner OA1200PAC OmniAire Portable Air Cleaner OA1200PAC The OmniAire Portable Air Purification system purifies the air of airborne contaminants which could be biological, chemical or particulates. Connected to an isolation

More information

SUPERIOR PERFORMANCE

SUPERIOR PERFORMANCE REVERSE PULSE DUST COLLECTORS OPERATION & MAINTENANCE MANUAL M600-2C, M900-2C & M1200-2C SUPERIOR PERFORMANCE And Quality In Blast Cleaning Equipment! WARNING Read Manual Failure to read, understand &

More information

Portable HEPA System

Portable HEPA System Operation and Maintenance Manual Portable HEPA System Series - PHS MONOXIVENT - SOURCE CAPTURE SYSTEMS - info@ Oct. - 2015 PORTABLE HEPA SYSTEM GENERAL INFORMATION SAFETY PRECAUTIONS: Keep all paper work

More information

4th Quarter Summary of Meteorological and Ambient Air Quality Data Kennecott Utah Copper Monitoring Stations

4th Quarter Summary of Meteorological and Ambient Air Quality Data Kennecott Utah Copper Monitoring Stations 4th Quarter 2016 Summary of Meteorological and Ambient Air Quality Data Kennecott Utah Copper Monitoring Stations Steven Root Digitally signed by Steven Root DN: cn=steven Root, o=weatherbank, Inc., ou,

More information

Contents. Overview. Safety Instructions. Before Installation. Installation. Operation. Maintenance. Replacement Parts. System Specifications 8 01

Contents. Overview. Safety Instructions. Before Installation. Installation. Operation. Maintenance. Replacement Parts. System Specifications 8 01 PrintPRO Universal Contents Overview 1 01 02 03 Overview of your extraction system (front) Overview of your extraction system (back) Overview of the Control Panel Safety Instructions 2 01 Important safety

More information

SUMMARY. St-jean-sur-Richelieu, 2016, January 12th Client : J.A. Roby Project :PI model : Polaris; Sirius; Véga; Antares; Centauri; Rigel

SUMMARY. St-jean-sur-Richelieu, 2016, January 12th Client : J.A. Roby Project :PI model : Polaris; Sirius; Véga; Antares; Centauri; Rigel SUMMARY St-jean-sur-Richelieu, 2016, January 12th 1 Introduction... 4 1.1 General... 4 1.2 Test unit information... 4 1.3 Results... 4 1.4 Pretest information... 5 2 Summary of test results... 5 2.1 Emissions...

More information

KG7T (C and L Series)

KG7T (C and L Series) TECHNICAL SPECIFICATIONS KG7T (C and L Series) Two Stage, Fixed Speed ECM, Condensing Upflow/Horizontal and Downflow Gas Furnaces Induced Draft - 95.1 AFUE Input 60,000-120,000 Btuh The high efficiency

More information

ERV ENERGY RECOVERY VENTILATOR 60 HZ. Indoor / Outdoor 500 to 10,000 cfm Capacity ERV I - D - D - P - 1 MODEL NUMBER IDENTIFICATION

ERV ENERGY RECOVERY VENTILATOR 60 HZ. Indoor / Outdoor 500 to 10,000 cfm Capacity ERV I - D - D - P - 1 MODEL NUMBER IDENTIFICATION P R O D U C T S P E C I F I C AT I O N S I N D O O R A I R Q U A L I T Y ERV ENERGY RECOVERY VENTILATOR 60 HZ Bulletin No. 20245 March 200 Supersedes April 2009 Indoor / Outdoor 500 to 0,000 cfm Capacity

More information

SPH General Manual 400 Series

SPH General Manual 400 Series Operations & Maintenance Manual SPH General Manual 400 Series (SPH-415,416,425,426,4U6,4U5,4B5,4B6) Filename: Quatro/SPH-400 General 12/28/04 R6 ------------------------------------------------------------------------Engineering

More information

Fuel Systems GASOLINE FUEL SYSTEMS

Fuel Systems GASOLINE FUEL SYSTEMS Fuel Systems I - 1 GASOLINE FUEL SYSTEMS Gasoline fuel systems used in Four Winns boats are designed to meet or exceed the requirements of the US Coast Guard, the National Marine Manufacturers Association,

More information

Table of Contents. Safety Information Main Functions Set-Up Operation Maintenance Troubleshooting Specifications Customer Support Warranty Information

Table of Contents. Safety Information Main Functions Set-Up Operation Maintenance Troubleshooting Specifications Customer Support Warranty Information Model: LV-PUR131 True HEPA Air Purifier Questions or Concerns? Please contact us Mon-Fri 9:00AM-5:00PM PST at (888)726-8520 or email support@levoit.com Thank You. Thank you for purchasing the LV-PUR131

More information

TSTAT0101SC Recommended (sold separately)

TSTAT0101SC Recommended (sold separately) Up to 96% AFUE, Communicating, Two Stage Gas Furnace EASIER TO SELL Up to 96% AFUE, all models, all positions California NOx approved All Models have earned the ENERGY STAR Two stage heating operation

More information

Former Fort Ord Sites 2 and 12 Data and Status

Former Fort Ord Sites 2 and 12 Data and Status Sites 2 and 12 Data and Status Meeting, Table 1: GWTP and SVTU Statistics as of October 31, 2016 COC Mass Monthly Statistics Volume Treated Average Flow Percent of Time Online Removed (pounds) Oct 2016

More information

FAN POWERED HEPA FILTER UNIT

FAN POWERED HEPA FILTER UNIT READ COMPLETELY AND SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE FAN POWERED HEPA FILTER UNIT (MOTOR ACCESSIBLE) SSLFHFD-FP-MA INSTALLATION AND SERVICE MANUAL 2 Critical operation conditions of the Fan

More information