1 Instructions For Use SX4400 Swinging Bucket Rotor For Use in the Beckman Coulter Allegra X-30 Series Centrifuges B01198AC September 2017 Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA U.S.A.

2 PN B01198AC (September 2017) 2017 Beckman Coulter, Inc. All rights reserved. Beckman Coulter, the stylized logo, and the Beckman Coulter product and service marks mentioned herein are trademarks or registered trademarks of Beckman Coulter,Inc. in the United States and other countries. All other trademarks, service marks, products, or services are trademarks or registered trademarks of their respective holders. Contact Us If you have any questions, contact our Customer Support Center. Worldwide, find us via our website at In the USA and Canada, call us at Outside of the USA and Canada, contact your local Beckman Coulter Representative. Beckman Coulter Eurocenter S.A. 22, rue Juste-Olivier Case Postale 1044 CH Nyon 1, Switzerland Tel: +41 (0) Printed in U.S.A.

3 Revision History This document applies to the latest software listed and higher versions. When a subsequent software version changes the information in this document, a new issue will be released. For updates, go to and download the most recent manual or system help for your instrument. Initial Issue B01198AA, 04/2011 Rev AB, 10/2013 Changes or updates were made to the following: Specifications The Rotor Buckets and Accessories Installing the Rotor Yoke Maintenance Rotor Components Rev AC, 09/2017 Changes or updates were made to the following: Table 1, Available Tube Adapters for the SX4400 Rotor. Note: Changes that are part of the most recent revision are indicated in text by a bar in the left margin of the amended page. B01198AC iii

4 Revision History iv B01198AC

5 Safety Notice Read all product manuals and consult with Beckman Coulter-trained personnel before attempting to use this equipment. Do not attempt to perform any procedure before carefully reading all instructions. Always follow product labeling and manufacturer s recommendations. If in doubt as to how to proceed in any situation, contact your Beckman Coulter Representative.! This safety notice summarizes information basic to the safe use of the rotor described in this manual. The international symbol displayed to the left is a reminder to the user that all safety instructions should be read and understood before operation or maintenance of this equipment is attempted. When you see the symbol on other pages of this publication, pay special attention to the safety information presented. Observance of safety precautions will also help to avoid actions that could damage or adversely affect the performance of the rotor. Alerts for Warning, Caution, and Note WARNING WARNING indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. CAUTION CAUTION indicates a potentially hazardous situation, which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices. NOTE NOTE is used to call attention to notable information that should be followed during installation, use, or servicing of this equipment. Safety Information for the SX4400 Rotor Handle body fluids with care because they can transmit disease. No known test offers complete assurance that such fluids are free of micro-organisms. Some of the most virulent Hepatitis (B and C) viruses, HIV (I V), atypical mycobacteria, and certain systemic fungi further emphasize the need for aerosol protection. Handle other infectious samples according to good laboratory procedures and methods to prevent spread of disease. Because spills may generate aerosols, observe proper safety precautions for aerosol containment. Do not run toxic, pathogenic, or radioactive materials in this rotor without taking appropriate safety precautions. Biosafe containment should be used when Risk Group II materials (as identified in the World Health Organization Laboratory Biosafety Manual) are handled; materials of a higher group require more than one level of protection. B01198AC v

6 Safety Notice Safety Information for the SX4400 Rotor The rotor and accessories are not designed for use with materials capable of developing flammable or explosive vapors. Do not centrifuge such materials in nor handle or store them near the centrifuge. This rotor was developed, manufactured, and tested for safety and reliability as part of a Beckman Coulter centrifuge/rotor system. Its safety or reliability cannot be assured if used in a non- Beckman Coulter centrifuge, in a Beckman Coulter centrifuge not specified for use with the rotor, or in a Beckman Coulter centrifuge that has been modified without Beckman Coulter s approval. Although rotor components and accessories made by other manufacturers may fit in the SX4400 rotor, their safety in this rotor cannot be ascertained by Beckman Coulter. Use of other manufacturers components or accessories in the SX4400 rotor may void the rotor warranty and should be prohibited by your laboratory safety officer. Only the components and accessories listed in this publication should be used in this rotor. Hook all four buckets, loaded or empty, to the rotor for every run. Make sure that filled containers are loaded symmetrically into the rotor and that opposing tubes are filled to the same level with liquid of the same density. If disassembly reveals evidence of leakage, you should assume that some fluid escaped the rotor. Apply appropriate decontamination procedures to the centrifuge and accessories. Never exceed the maximum rated speed of the rotor and labware in use. Refer to the section on Run Speeds, and derate the run speed as appropriate. Do not use sharp tools on the rotor that could cause scratches in the rotor surface. Corrosion begins in scratches and may open fissures in the rotor with continued use. vi B01198AC

7 Contents Revision History,iii Safety Notice,v Alerts for Warning, Caution, and Note, v Safety Information for the SX4400 Rotor, v SX4400 Swinging Bucket Rotor, 1 Specifications, 1 Description, 2 The Rotor,2 Buckets and Accessories, 2 Preparation and Use, 5 Tubes, 5 Prerun Safety Checks, 5 Installing the Rotor Yoke, 6 Loading the Buckets, 7 Operation, 9 Removal and Sample Recovery, 10 Run Speeds, 10 Care and Maintenance, 11 Maintenance, 11 Cleaning, 12 Decontamination, 14 Sterilization and Disinfection, 14 Tube Breakage, 15 Storage, 15 Returning a Rotor, 15 Supply List, 16 Replacement Rotor Parts, 16 Other, 16 Beckman Coulter, Inc. Benchtop Rotor Warranty vii

8 Illustrations Illustrations 1 The SX4400 Rotor and Accessories, 3 2 Installing the Rotor Yoke on the Centrifuge Drive Shaft, 6 3 Arranging tubes in the Rotor, 8 4 Seating Buckets on Yoke Pivot Pins, 9 viii

9 Tables Tables 1 Available Tube Adapters for the SX4400 Rotor, 3 ix

10 x Tables

11 Specifications 1 rav (101 mm) rmax (172 mm) 1. Axis of Rotation rmin (71 mm) Maximum speed Allegra X-30R rpm Allegra X rpm Density rating at maximum speed g/ml Critical speed range a to 1200 rpm Relative Centrifugal Field b at maximum speed at r max (172 mm) Allegra X-30R g Allegra X g Conditions requiring speed reductions... see Run Speeds Maximum allowable imbalance of opposing loads grams Number of buckets... 4 Available labware... Table 1 Nominal tube capacity (largest tube) ml Nominal rotor capacity ml Approximate acceleration time to maximum speed (fully loaded) sec Approximate deceleration time from maximum speed (fully loaded) sec Weight of fully loaded rotor kg (12.6 lb) Rotor and bucket material... anodized aluminum Bucket cover material... polysulfone a. The critical speed range is the range of speeds over which the rotor shifts so as to rotate about its center of mass. Passing through the critical speed range is characterized by some vibration. b. Relative Centrifugal Field (RCF) is the ratio of the centrifugal acceleration at a specified radius and speed (rw 2 ) to the standard acceleration of gravity (g) according to the following formula: RCF = r /g where r is the radius in millimeters, is the angular velocity in radians per second (2 RPM /60), and g is the standard acceleration of gravity (9807 mm/s 2 ). After substitution: RCF = 1.12r (RPM/1000) 2 B01198AC 1-1

12 Description Description Bucket Cover (392266) 2. Bucket (B01664) 3. Bucket O-ring (392621) This Beckman Coulter rotor has been manufactured in an ISO 9001 or facility for use with the specified Beckman Coulter centrifuges. The Rotor The SX4400 is a four-place swinging bucket rotor. This rotor carries adapters that allow centrifugation of a wide range of tube and bottle sizes (from 1.5-mL reaction vials to 400-mL bottles). The SX4400 rotor develops centrifugal forces that are suitable for rapidly sedimenting protein precipitates, large particles, cells, and cell debris. The rotor yoke and buckets are made of aluminum and are black-anodized for corrosion protection. The buckets are run by placing them over pivot pins on the arms of the yoke. Buckets swing out to a horizontal position during centrifugation. This rotor was tested * to demonstrate containment of microbiological aerosols under normal operating conditions of the associated Beckman Coulter centrifuge, when used and maintained as instructed. A tie-down screw is used to secure the rotor to the drive shaft during centrifugation. Figure 1 shows the rotor and accessories. The centrifuge identifies rotor speed during the run by means of a magnetic speed sensor system in the centrifuge rotor chamber and magnets imbedded in the rotor. The overspeed system ensures that the rotor does not exceed its permitted speed. Buckets and Accessories Threaded-top buckets with screw-on transparent covers hold several sizes of labware in modular or one-piece adapters. Each cover requires an O-ring that seats in a groove at the top of the bucket. Although the covers are not designed to contain aerosols that may result from tube breakage, they will contain most liquids and broken tube particles, allowing you to take appropriate precautions before opening the covers in the event of tube breakage. * Validation of microbiological containment was done at an independent third-party testing facility (CAMR, Porton Down, UK, or USAMRIID, Ft. Detrick, MD, U.S.A.). Improper use or maintenance may affect seal integrity and thus containment. 1-2 B01198AC

13 Description A variety of tube sizes can be centrifuged in the rotor, supported in tube adapters listed in Table 1. The adapters can also serve as tube racks in the laboratory. Figure 1 The SX4400 Rotor and Accessories Bucket 2. Bucket Cover 3. T-handle Rotor Wrench 4. Tie-down Screw 5. Rotor Yoke NOTE The latest information regarding labware and accessories used in Beckman Coulter rotors can be obtained at Table 1 Available Tube Adapters for the SX4400 Rotor Adapters Tubes Maximum Number Part Number (pkg 2) Tubes per Adapter Description Nominal Volume Tubes per Rotor N/A N/A Bottle, polypropylene (Part Number: B01435) 400 ml mm (flat bottom w/screw-on cap) a 250 ml mm 180 ml mm 125 ml mm (glass) 80/100 ml 4 B01198AC 1-3

14 Description Table 1 Available Tube Adapters for the SX4400 Rotor (Continued) Adapters Tubes Maximum Number Part Number (pkg 2) Tubes per Adapter Description Nominal Volume Tubes per Rotor mm 85/94 ml mm (glass) 50 ml mm (conical) 50 ml mm 50 ml mm (w/screw-on cap) 30 ml mm (glass w/screw-on cap) 25 ml mm (glass, open-top) mm (conical) 15 ml 20 C x 56 mm (conical) 5 ml mm 10 ml mm (round bottom) 10/15 ml mm Monovettes 10 ml mm Vacutainer 10 ml mm Vacutainer b 15 ml mm Vacutainer 5 ml 64 Hemolyse RIA tubes 5 ml 5 ml mm RIA tubes 5 ml mm (glass) 7 ml mm 1.5/2.2 ml mm microfuge/pcr tubes 0.5 ml 96 a. The rotor was tested using Nalge bottle (polycarbonate). b. Bucket covers cannot be used with this tube/adapter combination. The maximum length of tubes used in the SX4400 buckets with covers is 125 mm for 250-mL bottles and 122 mm for all other tubes (when placed near the center of the adapter * ). The 3-mm difference * In multiple-place adapters, the maximum length of tubes that can be placed near the adapter periphery while still fitting under the bucket cover is 115 mm. 1-4 B01198AC

15 Preparation and Use in length (250-mL bottles vs. other tubes) is specified because 250-mL bottle adapter is a bottomless sleeve, while all other adapters have a bottom surface that is approximately 3 mm tall. See the Warranty at the back of this manual for warranty information. Preparation and Use Specific information about the SX4400 rotor is given here. Information about the centrifuge is contained in the centrifuge manual, which should be used together with this manual for complete centrifuge, rotor, and accessory operation. NOTE Although rotor components and accessories made by other manufacturers may fit in the SX4400 rotor, their safety in this rotor cannot be ascertained by Beckman Coulter. Use of other manufacturers components or accessories in the SX4400 rotor may void the rotor warranty and should be prohibited by your laboratory safety officer. Only the components and accessories listed in this publication should be used in this rotor. Tubes Tubes should be pretested under anticipated run conditions (using water instead of valuable samples) if operating below 4 C or above room temperature. Refer to Chemical Resistances (publication IN-175) for information on the chemical resistances of tube and adapter materials. Temperature Limits Plastic containers have been centrifuge tested for use at temperatures between 2 and 25 C. For centrifugation at other temperatures, pretest tubes under anticipated run conditions. If plastic containers are frozen before use, make sure that they are thawed to at least 2 C prior to centrifugation. Prerun Safety Checks Read the Safety Notice section at the front of this manual before using the rotor. 1 Make sure that the rotor, buckets, and bucket covers (if used) are clean and show no signs of corrosion or cracking. If any evidence of damage is present, do not centrifuge the rotor. 2 Check the chemical compatibilities of all materials used (refer to Chemical Resistances, publication IN-175). 3 Verify that the tubes and accessories being used are listed in Table 1. B01198AC 1-5

16 Preparation and Use Installing the Rotor Yoke NOTE Before the first use of a new rotor, lubricate the pivot pins with Spinkote (306812). After the first use, apply Spinkote to the pivot pins approximately once a week and after cleaning or autoclaving. NOTE Refer to the centrifuge manual for proper installation and removal instructions. Failure to follow printed instructions may result in damage to the rotor and the centrifuge. 1 Before installing the yoke in the centrifuge, lightly coat the drive hole with lubricant such as Spinkote (306812) (see instructions under Maintenance). 2 Carefully lower the yoke straight down onto the centrifuge drive shaft (see Figure 2). Be sure the yoke is properly seated on the shaft. Figure 2 Installing the Rotor Yoke on the Centrifuge Drive Shaft Lower the yoke onto the drive shaft 2. Tighten the tie-down screw with the T-handle wrench 3 Fasten the tie-down screw (361367) onto the drive shaft. Use the T-handle rotor wrench (365636) to tighten the tie-down screw firmly on the shaft. CAUTION Never drop the rotor yoke onto the centrifuge drive shaft. The drive shaft can be damaged if the rotor is forced sideways or dropped onto it. NOTE If the yoke is left in the centrifuge between runs, before each run make sure it is properly seated on the drive shaft and that the tie-down screw is tight. 1-6 B01198AC

17 Preparation and Use Loading the Buckets For runs at other than room temperature, refrigerate or warm the rotor and precool the centrifuge beforehand for faster equilibration. WARNING Handle body fluids with care because they can transmit disease. No known test offers complete assurance that such fluids are free of micro-organisms. Some of the most virulent Hepatitis (B and C) viruses, HIV (I V), atypical mycobacteria, and certain systemic fungi further emphasize the need for aerosol protection. Handle other infectious samples according to good laboratory procedures and methods to prevent spread of disease. Because spills may generate aerosols, observe proper safety precautions for aerosol containment. Do not run toxic, pathogenic, or other hazardous materials in this rotor without taking all appropriate safety precautions. Biosafe containment should be used when Risk Group II materials (as identified in the World Health Organization Laboratory Biosafety Manual) are handled; materials of a higher group require more than one level of protection. Symmetrical and Balanced Loading To ensure optimal performance and stability, the rotor must be loaded symmetrically. Two factors affect symmetric loading: 1. The buckets or carriers must be loaded symmetrically with respect to their pivotal axes. 2. The rotor should be loaded symmetrically with respect to its center of rotation. For best results, load opposing buckets or carriers with the same type of labware containing the same amounts of fluid of equal density. Additionally, opposing buckets and their contents should weigh approximately the same (within 10 grams). It is not necessary to completely fill all tubes or positions in buckets; however, partially filled buckets must be balanced with respect to the bucket pivotal axis. Each tube should be placed so that its weight is balanced by a tube in a diametrically opposite position across the pivotal axis in the same bucket. Opposing buckets should also be filled the same way. Using Buckets You can load buckets before or after they are installed on the rotor yoke. In either case, we recommend filling the appropriate labware first and then loading the labware into the buckets to avoid tripping the imbalance detector. B01198AC 1-7

18 Preparation and Use 1 Load the filled containers into the buckets (see Tubes for tube information). NOTE Tubes placed in opposing buckets must be filled to the same level with liquid of the same density. Fewer than maximum number of tubes can be centrifuged in the multi-tube adapters if they are arranged symmetrically in the buckets (see Figure 3). Figure 3 Arranging tubes in the Rotor 1 1. Pivotal Axis NOTE All buckets must be attached to the rotor for every run, whether loaded or empty. Never run the rotor with fewer than four positions filled. 2 Place covers on buckets, if applicable. Screw covers to the right (clockwise) to tighten. 3 Attach each bucket to the yoke by aligning the grooves in the bucket sides with the pivot pins, then sliding the buckets down until the pivot pins are seated in the bucket pockets. CAUTION Attach all four buckets, loaded or empty, to the rotor yoke. If only two buckets are filled, place them in opposite positions on the yoke. All four positions must contain buckets during a run. 4 Gently swing the buckets to ensure that they are properly seated on the pivot pins. 1-8 B01198AC

19 Preparation and Use 5 Make sure buckets are properly seated over the pivot pins on the rotor yoke as shown in Figure 4. Figure 4 Seating Buckets on Yoke Pivot Pins CORRECT 1 INCORRECT 2 Pivot pins are seated in bucket pockets 3 4 Pivot pins are not seated in bucket pockets 1. CORRECT 2. INCORRECT 3. Pivot pins are seated in bucket pockets 4. Pivot pins are not seated in bucket pockets NOTE Buckets must be inserted at an angle, or the bucket bottom will rest on the rotor base and the pivot pins will not seat in the pin pockets. Operation 1 Refer to the centrifuge instruction manual for centrifuge operation. 2 See Run Speeds, page 10, for information about speed limitations. B01198AC 1-9

20 Preparation and Use Removal and Sample Recovery CAUTION If disassembly reveals evidence of leakage, you should assume that some fluid escaped the rotor. Apply appropriate decontamination procedures to the centrifuge and accessories. 1 Remove the buckets from the centrifuge. 2 Remove covers (if used) from buckets, and remove the tubes from the buckets. NOTE If glass tubes break, remove the glass very carefully from the adapter and bucket. If all the glass particles are not contained in the bucket or adapter, be careful when examining or cleaning the centrifuge gasket and chamber as glass particles may be embedded in their surfaces. 3 If removing the rotor yoke, turn the T-handle wrench to the left (counterclockwise) to loosen the tie-down screw. Lift the yoke straight up and off the drive shaft. NOTE Refer to the centrifuge manual for proper installation and removal instructions. Failure to follow printed instructions may result in damage to the rotor and the centrifuge. Run Speeds The centrifugal force at a given radius in a rotor is a function of speed. Comparisons of forces between different rotors are made by comparing the rotors relative centrifugal fields (RCF). When rotational speed is adjusted so that identical samples are subjected to the same RCF in two different rotors, the samples are subjected to the same force. If RPM is entered, the RCF at each speed is automatically calculated by the centrifuge software; if the RCF is entered, the centrifuge calculates the equivalent revolutions per minute (rpm). Do not select rotational speeds that exceed 4700 rpm for refrigerated centrifuges, or 4200 rpm for nonrefrigerated centrifuges. Speed Derating The maximum run speed listed in the rotor specifications is for operation when all conditions are within the standard specifications. Maximum allowable run speed for the rotor must be reduced in some circumstances. 1. The rotor can process solutions at the maximum rated speed if the solution density is 1.2 g/ml or less. When centrifuging solutions of densities greater than 1.2 g/ml, the maximum run 1-10 B01198AC

21 Care and Maintenance speeds must be reduced according to the following equation to protect the rotor from excessive stresses due to the added tube load: reduced maximum speed = (4700 rpm)* 1.2 g/ml density of tube contents *Use 4200 for nonrefrigerated Allegra X-30 centrifuges 2. When some glass tubes are used, reduce rotor speed to prevent tube breakage. Because the strength of tubes can vary from lot to lot, and will depend on handling and usage, Beckman Coulter highly recommends that you pretest these tubes in the appropriate labware in the SX4400 rotor, using water samples, to determine optimal operating conditions. NOTE Scratches (even microscopic ones) significantly weaken glass tubes. Care and Maintenance Maintenance Do not use sharp tools on the rotor, as they can scratch the anodized surface. Corrosion begins in scratches and may open fissures in the rotor with continued use Bucket Carrier 2. Check for Corrosion 3. Bucket O-ring 1 Periodically (at least monthly) inspect the rotor yoke and buckets, especially inside cavities, for rough spots or pitting, white powder deposits frequently aluminum oxide or heavy discoloration. If any of these signs are evident, do not run the rotor. Contact your Beckman Coulter representative for information about the Field Rotor Inspection Program and the rotor repair center. 2 Regularly check the condition of the tie-down screw (361367); if it is worn or damaged, replace it. Regularly check the condition of the bucket O-rings (392621); if they are worn or damaged, replace them. B01198AC 1-11

22 Care and Maintenance 3 Check bucket covers and adapters for crazing or cracks before use. Do not centrifuge cracked covers or adapters. 4 Approximately once a week, and after cleaning and/or autoclaving, lubricate the rotor pins and pin sockets with a lubricant such as Spinkote (306812). Store the rotor in a dry environment (not in the centrifuge). Refer to Chemical Resistances for the chemical compatibilities of rotor and accessory materials. Your Beckman Coulter representative provides contact with the Field Rotor Inspection Program and the rotor repair center. Cleaning Wash rotor components immediately if salts or other corrosive materials are used or if spillage has occurred. Do not allow corrosive materials to dry on the rotor. Under normal use, wash the rotor frequently (at least weekly) to prevent buildup of residues. NOTE Do not wash the rotor components in a dishwasher. Do not soak the rotor in detergent solution for long periods, such as overnight. Rotor Components 1 Wash the rotor yoke and tie-down screw in a mild detergent, such as Solution 555 (339555), that won t damage the rotor. The Rotor Cleaning Kit (339558) contains two plastic-coated brushes and two quarts of Solution 555 for use with rotors and accessories. Dilute the detergent 10 to 1 with water. 2 Thoroughly rinse the cleaned rotor components with distilled water. 3 Air-dry the rotor components upside down. Do not use acetone to dry the rotor. 4 Lubricate the rotor pins with Spinkote (306812) B01198AC

23 Care and Maintenance Before reinstalling the rotor yoke, lightly lubricate the drive hole with Spinkote (306812) to prevent the rotor from sticking, as follows: 1 Apply the lubricant onto a swab. 2 Draw the coated swab through a paper towel to remove excess lubricant. 3 Lightly coat the inside of the drive hole with the lubricant remaining on the swab. Buckets 1 Remove the O-ring from the buckets before cleaning. 2 Wash the buckets, covers, and O-rings with a mild detergent such as Solution 555, diluted 10 to 1 with water, and a soft brush. 3 Thoroughly rinse with water and air-dry. 4 Replace the O-ring in the groove. Adapters 1 Disassemble modular adapters for washing by pulling them apart. 2 Use a mild detergent such as Solution 555, diluted 10 to 1 with water, and a soft brush to scrub the adapters. 3 Rinse thoroughly with water. 4 Air-dry upside down. Reassemble modular adapters. B01198AC 1-13

24 Care and Maintenance Decontamination If aluminum rotor components become contaminated with radioactive material, decontaminate them using a solution that will not damage the anodized surfaces. Beckman Coulter has tested a number of solutions and found two that do not harm anodized aluminum: RadCon Surface Spray or IsoClean Solution (for soaking), * and Radiacwash. While Beckman Coulter has tested these materials and found that they do not damage components, no guarantee of decontamination is expressed or implied. Follow appropriate decontamination procedures as directed by your laboratory safety officer. If the rotor or other components are contaminated with toxic or pathogenic materials, follow appropriate decontamination procedures as directed by your laboratory safety officer Sterilization and Disinfection The rotor and all rotor components (except bucket covers) can be autoclaved at 121 C for up to an hour. Remove the covers from the buckets and place the rotor yoke and buckets in the autoclave upside down. Bucket covers should not be autoclaved for more than 15 minutes. Ethanol (70%) or hydrogen peroxide (6%) may be used on all rotor components, including those made of plastic. Bleach (sodium hypochlorite) may be used, but may cause discoloration of anodized surfaces. Use the minimum immersion time for each solution, per laboratory standards. While Beckman Coulter has tested these methods and found that they do not damage the rotor or components, no guarantee of sterility or disinfection is expressed or implied. When sterilization or disinfection is a concern, consult your laboratory safety officer regarding proper methods to use. Refer to publication IN-192, included in each box of tubes or bottles, for tube sterilization and disinfection procedures. * In U.S., contact Nuclear Associates (New York); in Eastern Europe and Commonwealth States, contact Victoreen GmbH (Munich); in South Pacific, contact Gammasonics Pty. Ltd. (Australia); in Japan, contact Toyo Medic Co. Ltd. (Tokyo). In U.S., contact Biodex Medical Systems (Shirley, New York); internationally, contact the U.S. office to find the dealer closest to you. Flammability hazard. Do not use in or near operating ultracentrifuges B01198AC

25 Returning a Rotor Tube Breakage CAUTION To reduce the potential for corrosion, clean buckets thoroughly immediately following a tube breakage. If a glass tube breaks, remove the glass very carefully from the adapter and bucket. Imbedded glass particles that remain in the bucket or adapters can cause tube failure during subsequent runs. Storage When it is not in use, store the rotor in a dry environment (not in the centrifuge). Returning a Rotor Before returning a rotor or accessory for any reason, prior permission must be obtained from Beckman Coulter, Inc. This form may be obtained from your local Beckman Coulter sales office. The form, entitled Returned Material Authorization (RMA) for United States returns or Returned Goods Authorization (RGA) for international returns, should contain the following information: rotor type and serial number, history of use (approximate frequency of use), reason for the return, original purchase order number, billing number, and shipping number, if possible, name and address of the person to be notified upon receipt of the rotor or accessory at the factory, name and address of the person to be notified about repair costs, etc. To protect our personnel, it is the customer s responsibility to ensure that all parts are free from pathogens and/or radioactivity. Sterilization and decontamination must be done before returning the parts. Smaller items (such as tubes, bottles, etc.) should be enclosed in a sealed plastic bag. All parts must be accompanied by a note, plainly visible on the outside of the box or bag, stating that they are safe to handle and that they are not contaminated with pathogens or radioactivity. Failure to attach this notification will result in return or disposal of the items without review of the reported problem. Use the address label printed on the RGA/RMA form when mailing the rotor and/or accessories. Customers located outside the United States should contact their local Beckman Coulter office. B01198AC 1-15

26 Supply List Supply List NOTE Publications referenced in this manual can be obtained at or by calling Beckman Coulter at in the United States, or by contacting your local Beckman Coulter office. Call Beckman Coulter Sales ( in the United States) for detailed information on ordering parts and supplies. For your convenience, a partial list is given below. Replacement Rotor Parts SX4400 rotor assembly (includes 4 buckets without covers) B01425 Bucket covers (set of 4) Buckets (set of 2) B01426 Rotor yoke B01429 Bucket O-ring (set of 4) Tie-down screw T-handle rotor wrench Other Adapters see Table 1 Spinkote lubricant (2 oz) Silicone vacuum grease (1 oz) Rotor Cleaning Kit Beckman Solution 555 (1 qt) Rotor cleaning brush B01198AC

27 Beckman Coulter, Inc. Benchtop Rotor Warranty Subject to the conditions specified below and the warranty clause of the Beckman Coulter, Inc., terms and conditions in effect at the time of sale, Beckman Coulter agrees to correct either by repair or, at its election, by replacement, any defects of material or workmanship which develop within seven (7) years after delivery of a benchtop centrifuge rotor to the original buyer by Beckman Coulter or by an authorized representative, provided that investigation and factory inspection by Beckman Coulter discloses that such defect developed under normal and proper use. Should a Beckman Coulter centrifuge be damaged due to a failure of a rotor covered by this warranty, Beckman Coulter will supply free of charge all centrifuge parts required for repair. Conditions Except as otherwise specifically provided herein, this warranty covers the rotor only and Beckman Coulter shall not be liable for damage to accessories or ancillary supplies including but not limited to (i) tubes, (ii) tube caps, (iii) tube adapters, or (iv) tube contents. This warranty is void if the rotor has been subjected to customer misuse such as operation or maintenance contrary to the instructions in the Beckman Coulter rotor or centrifuge manual. This warranty is void if the rotor is operated with a rotor drive unit or in a centrifuge unmatched to the rotor characteristics or operated in a Beckman Coulter centrifuge that has been improperly disassembled, repaired, or modified. Thermoplastic rotors or components used in some benchtop centrifuges are warranted for one (1) year from date of purchase. Disclaimer IT IS EXPRESSLY AGREED THAT THE ABOVE WARRANTY SHALL BE IN LIEU OF ALL WARRANTIES OF FITNESS AND OF THE WARRANTY OF MERCHANTABILITY AND THAT NEITHER BECKMAN COULTER, INC. NOR ITS SUPPLIERS SHALL HAVE ANY LIABILITY FOR SPECIAL OR CONSEQUENTIAL DAMAGES OF ANY KIND WHATSOEVER ARISING OUT OF THE MANUFACTURE, USE, SALE, HANDLING, REPAIR, MAINTENANCE, OR REPLACEMENT OF THE PRODUCT. B01198AC Warranty-1

28 Beckman Coulter, Inc. Benchtop Rotor Warranty Warranty-2 B01198AC

29

30 Related Documents Allegra X-30 Series (B01145) Safety Introduction Description Installation Operation Troubleshooting Care and Maintenance Available in hard copy or electronic pdf by request. Additional References Chemical Resistances for Beckman Coulter Centrifugation Products (IN-175) Use and Care of Tubes and Bottles (IN-192) Available in hard copy or electronic pdf by request. Available at Beckman Coulter, Inc. All Rights Reserved

Instructions For Use. FA361.5 Fixed-Angle Rotor. For Use in Beckman Coulter Microfuge 20 Series Centrifuges. PN B30182AA June 2013

Instructions For Use. FA361.5 Fixed-Angle Rotor. For Use in Beckman Coulter Microfuge 20 Series Centrifuges. PN B30182AA June 2013 Instructions For Use FA361.5 Fixed-Angle Rotor For Use in Beckman Coulter Microfuge 20 Series Centrifuges PN B30182AA June 2013 Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA 92821 U.S.A. B30182AA

More information

BS-TB-004EB February TA Rotor. Used in the Beckman Coulter Allegra 25R and TJ-25 Centrifuges Beckman Coulter, Inc.

BS-TB-004EB February TA Rotor. Used in the Beckman Coulter Allegra 25R and TJ-25 Centrifuges Beckman Coulter, Inc. BS-TB-004EB February 2014 TA-14-50 Rotor Used in the Beckman Coulter Allegra 25R and TJ-25 Centrifuges 2008 2014 Beckman Coulter, Inc. SAFETY NOTICE This safety notice summarizes information basic to the

More information

Instructions for Using Micro Plus Multiwell Plate Carriers

Instructions for Using Micro Plus Multiwell Plate Carriers Litpak 361309 GS6-TB-011EA February 2014 Instructions for Using Micro Plus Multiwell Plate Carriers In Beckman Coulter GH-3.8, GH-3.8A, and JS-4.3 Rotors A B C D E F G H SPECIFICATIONS Maximum speed...................................

More information

Instructions For Use. F301.5 and FX301.5 Fixed-Angle Rotors

Instructions For Use. F301.5 and FX301.5 Fixed-Angle Rotors Instructions For Use F301.5 and FX301.5 Fixed-Angle Rotors For Use in the Beckman Coulter Microfuge 22R (F301.5), Allegra X-22 Series (FX301.5), and Allegra X-30 Series (FX301.5) Centrifuges September

More information

JS-13.1 Swinging Bucket Rotor

JS-13.1 Swinging Bucket Rotor J-TB-036MA February 2014 JS-13.1 Swinging Bucket Rotor Used In Beckman Coulter J2, J6, and Avanti J Series Centrifuges 2014 Beckman Coulter, Inc. Trademarks Beckman Coulter, the stylized logo, and Avanti

More information

JA-12 Fixed Angle Rotor

JA-12 Fixed Angle Rotor J-TB-051AG July 2009 JA-12 Fixed Angle Rotor Used In Beckman Coulter J2 and Avanti J Series Centrifuges 2009 Beckman Coulter, Inc. ! SAFETY NOTICE This safety notice summarizes information basic to the

More information

J-Lite JLA Fixed Angle Rotor

J-Lite JLA Fixed Angle Rotor J-TB-072J September 2008 J-Lite JLA-16.250 Fixed Angle Rotor Used In Beckman Coulter J2 and Avanti J Series Centrifuges 2008 2009 Beckman Coulter, Inc. ! SAFETY NOTICE This safety notice summarizes information

More information

JA-20 Fixed Angle Rotor

JA-20 Fixed Angle Rotor J-TB-003AZ November 2008 JA-20 Fixed Angle Rotor Used In Beckman Coulter J2, J6, and Avanti J Series Centrifuges 2008 2009 Beckman Coulter, Inc. ! SAFETY NOTICE This safety notice summarizes information

More information

JA-10 Fixed Angle Rotor

JA-10 Fixed Angle Rotor J-TB-006BB February 2014 JA-10 Fixed Angle Rotor Used In Beckman Coulter J2, J6, and Avanti J Series Centrifuges 2014 Beckman Coulter, Inc. Trademarks Beckman Coulter, the stylized logo. and Avanti, are

More information

JA-20.1 Fixed Angle Rotor

JA-20.1 Fixed Angle Rotor J-TB-022AT February 2014 JA-20.1 Fixed Angle Rotor Used In Beckman Coulter J2, J6, and Avanti J Series Centrifuges 2014 Beckman Coulter, Inc. Trademarks Beckman Coulter, the stylized logo, and Avanti,

More information

Instructions For Use. JA Ti Fixed-Angle Rotor. For Use in Beckman Coulter J2 and and Avanti J Series Centrifuges (except Avanti J-E)

Instructions For Use. JA Ti Fixed-Angle Rotor. For Use in Beckman Coulter J2 and and Avanti J Series Centrifuges (except Avanti J-E) Instructions For Use JA-30.50 Ti Fixed-Angle Rotor For Use in Beckman Coulter J2 and and Avanti J Series Centrifuges (except Avanti J-E) J-TB-070AN February 2014 Beckman Coulter, Inc. 250 S. Kraemer Blvd.

More information

J-TB-089AE February JS-5.3 Rotor. Used in Beckman Coulter Avanti J-20 and J-26 Series, and Avanti J-E Centrifuges Beckman Coulter, Inc.

J-TB-089AE February JS-5.3 Rotor. Used in Beckman Coulter Avanti J-20 and J-26 Series, and Avanti J-E Centrifuges Beckman Coulter, Inc. H G F E D C B A 1 2 3 4 5 6 7 8 9 10 11 12 J-TB-089AE February 2009 JS-5.3 Rotor Used in Beckman Coulter Avanti J-20 and J-26 Series, and Avanti J-E Centrifuges 2009 Beckman Coulter, Inc. ! SAFETY NOTICE

More information

Instructions For Use. JA Fixed-Angle Rotor. For Use in Beckman Coulter J2 and Avanti J Series Centrifuges. PN J-TB-056FB February 2014

Instructions For Use. JA Fixed-Angle Rotor. For Use in Beckman Coulter J2 and Avanti J Series Centrifuges. PN J-TB-056FB February 2014 Instructions For Use JA-25.50 Fixed-Angle Rotor For Use in Beckman Coulter J2 and Avanti J Series Centrifuges PN J-TB-056FB February 2014 Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA 92821 U.S.A.

More information

Instructions For Use. JS-5.3 Swinging-Bucket Rotor

Instructions For Use. JS-5.3 Swinging-Bucket Rotor Instructions For Use JS-5.3 Swinging-Bucket Rotor For Use in Beckman Coulter Avanti J-26S XP Series, Avanti J-26 XP Series, Avanti JXN-26, and Avanti J-E Centrifuges PN J-TB-089AG March 2014 Beckman Coulter,

More information

Instructions For Use. JS-4.2, JS-4.2SM, and JS-3.0 Swinging-Bucket Rotors

Instructions For Use. JS-4.2, JS-4.2SM, and JS-3.0 Swinging-Bucket Rotors Instructions For Use JS-4.2, JS-4.2SM, and JS-3.0 Swinging-Bucket Rotors JS-4.2 Rotor for Use in Beckman Coulter Avanti J-HC and J6 Series Centrifuges JS-4.2SM Rotor for Use in Beckman Coulter J6 Series

More information

Instructions for Using the A-100/18 and A-100/30 Rotors

Instructions for Using the A-100/18 and A-100/30 Rotors AF-TB-009AK February 2014 Instructions for Using the A-100/18 and A-100/30 Rotors In the Beckman Coulter Airfuge Ultracentrifuge 18 A-100/18 rmin rav rmax Axis of Rotation U.S. Pat. No. 3,997,104 SPECIFICATIONS

More information

INSTRUCTIONS FOR USING THE GH-3.8A SWINGING BUCKET ROTOR In Beckman Coulter Allegra 6 Series and Spinchron R and DLX Centrifuges

INSTRUCTIONS FOR USING THE GH-3.8A SWINGING BUCKET ROTOR In Beckman Coulter Allegra 6 Series and Spinchron R and DLX Centrifuges GS6-TB-017EB September 2016 INSTRUCTIONS FOR USING THE GH-3.8A SWINGING BUCKET ROTOR In Beckman Coulter Allegra 6 Series and Spinchron R and DLX Centrifuges This rotor has been manufactured in a registered

More information

Instructions for Using JS-5.2 and JS-4.0 Swinging Bucket Rotors

Instructions for Using JS-5.2 and JS-4.0 Swinging Bucket Rotors J6-TB-006K December 2007 Instructions for Using JS-5.2 and JS-4.0 Swinging Bucket Rotors JS-5.2 Used in Beckman Coulter J6 Series Centrifuges JS-4.0 Used in Beckman Coulter Avanti J-26XP Series and J6

More information

Instructions For Use. J-Lite JLA and JLA Fixed-Angle Rotor Assemblies. For Use in Beckman Coulter Avanti J Series Centrifuges

Instructions For Use. J-Lite JLA and JLA Fixed-Angle Rotor Assemblies. For Use in Beckman Coulter Avanti J Series Centrifuges Instructions For Use J-Lite JLA-8.1000 and JLA-9.1000 Fixed-Angle Rotor Assemblies For Use in Beckman Coulter Avanti J Series Centrifuges PN J-TB-073GA November 2012 Beckman Coulter, Inc. 250 S. Kraemer

More information

L5-TB-036AW July Type 45 Ti Rotor. Used In Beckman Coulter Class H, R, and S Preparative Ultracentrifuges Beckman Coulter, Inc.

L5-TB-036AW July Type 45 Ti Rotor. Used In Beckman Coulter Class H, R, and S Preparative Ultracentrifuges Beckman Coulter, Inc. L5-TB-036AW July 2009 Type 45 Ti Rotor Used In Beckman Coulter Class H, R, and S Preparative Ultracentrifuges 2009 Beckman Coulter, Inc. ! SAFETY NOTICE This safety notice summarizes information basic

More information

JA-18 Fixed Angle Rotor

JA-18 Fixed Angle Rotor FOR USE I J SERIES ISTRUMETS J-TB-035AR February 2014 JA-18 Fixed Angle Rotor BECKMA COULTER SERIAL O. 95U 1077 Made in U.S.A. JA-18 18,000 RPM Used In Beckman Coulter J2, J6, and Avanti J Series Centrifuges

More information

LXL-TB-005AN August Type 90 Ti Rotor. Used In Beckman Coulter Class H, R, and S Preparative Ultracentrifuges Beckman Coulter, Inc.

LXL-TB-005AN August Type 90 Ti Rotor. Used In Beckman Coulter Class H, R, and S Preparative Ultracentrifuges Beckman Coulter, Inc. LXL-TB-005AN August 2009 Type 90 Ti Rotor Used In Beckman Coulter Class H, R, and S Preparative Ultracentrifuges 2009 Beckman Coulter, Inc. !!!!!!!! SAFETY NOTICE This safety notice summarizes information

More information

Instructions For Use. Type 70 Ti Fixed-Angle Rotor. For Use in Beckman Coulter Class H, R, and S Preparative Ultracentrifuges

Instructions For Use. Type 70 Ti Fixed-Angle Rotor. For Use in Beckman Coulter Class H, R, and S Preparative Ultracentrifuges Instructions For Use Type 70 Ti Fixed-Angle Rotor For Use in Beckman Coulter Class H, R, and S Preparative Ultracentrifuges PN L5-TB-030BA February 2014 Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea,

More information

Instructions For Use. TLA-110 Fixed-Angle Rotor

Instructions For Use. TLA-110 Fixed-Angle Rotor Instructions For Use TLA-110 Fixed-Angle Rotor For Use in the Beckman Coulter Optima MAX-XP, MAX-TL, MAX, MAX-E, TL, TLX, and TL-100 Tabletop Ultracentrifuges PN TL-TB-019AG February 2014 Beckman Coulter,

More information

F0850 Rotor. Used in Beckman Coulter Allegra 64R, Avanti 30, Allegra X-22 Series, Allegra 21 Series, Spinchron 15 Series, and GS-15 Series Centrifuges

F0850 Rotor. Used in Beckman Coulter Allegra 64R, Avanti 30, Allegra X-22 Series, Allegra 21 Series, Spinchron 15 Series, and GS-15 Series Centrifuges GS-TB-003J February 2008 F0850 Rotor Used in Beckman Coulter Allegra 64R, Avanti 30, Allegra X-22 Series, Allegra 21 Series, Spinchron 15 Series, and GS-15 Series Centrifuges 2008 Beckman Coulter, Inc.

More information

JS-5.3 ROTOR. Used in Beckman Coulter Avanti J-20 and J-26 Series, and Avanti J-E Centrifuges. J-TB-089D November 2005

JS-5.3 ROTOR. Used in Beckman Coulter Avanti J-20 and J-26 Series, and Avanti J-E Centrifuges. J-TB-089D November 2005 H G F E D C B A 1 2 3 4 5 6 7 8 9 10 11 12 J-TB-089D November 2005 JS-5.3 ROTOR Used in Beckman Coulter Avanti J-20 and J-26 Series, and Avanti J-E Centrifuges 2005 Beckman Coulter, Inc. Printed in the

More information

J-Lite JLA Fixed Angle Rotor

J-Lite JLA Fixed Angle Rotor J-TB-072G March 2005 J-Lite JLA-16.250 Fixed Angle Rotor Used In Beckman Coulter J2 and Avanti J Series Centrifuges 2005 Beckman Coulter, Inc. Printed in the U.S.A. ! SAFETY NOTICE This safety notice summarizes

More information

TLA-55 Rotor. Used in Beckman Coulter Optima MAX, MAX-E, TL, and TLX Series Tabletop Ultracentrifuges

TLA-55 Rotor. Used in Beckman Coulter Optima MAX, MAX-E, TL, and TLX Series Tabletop Ultracentrifuges TL-TB-020C November 2002 TLA-55 Rotor Used in Beckman Coulter Optima MAX, MAX-E, TL, and TLX Series Tabletop Ultracentrifuges 2002 Beckman Coulter, Inc. Published by the Centrifuge Instrument Systems Development

More information

JS-7.5 Swinging Bucket Rotor

JS-7.5 Swinging Bucket Rotor J-TB-007P January 2008 JS-7.5 Swinging Bucket Rotor Used In Beckman Coulter J2, J6, and Avanti J Series Centrifuges 2008 Beckman Coulter, Inc. Printed in the U.S.A. ! SAFETY NOTICE This safety notice summarizes

More information

Tissue Master. User Manual

Tissue Master. User Manual Tissue Master User Manual This page left blank intentionally This manual is a guide for the use of the Omni International Tissue Master and accessories. Data herein has been verified and validated. It

More information

Instruction Manual. Labnet Spectrafuge 16M Microcentrifuge

Instruction Manual. Labnet Spectrafuge 16M Microcentrifuge Instruction Manual Labnet Spectrafuge 16M Microcentrifuge Labnet International PO Box 841 Woodbridge, NJ 07095 Phone: 732 417-0700 Fax: 732 417-1750 email: labnet@labnetlink.com Safety Precautions NEVER

More information

INSTRUCTION MANUAL. Model F14BCI-6x250y High Speed Rotor

INSTRUCTION MANUAL. Model F14BCI-6x250y High Speed Rotor FIBERLite CENTRIFUGE, Inc. Piramoon Technologies INSTRUCTION MANUAL Model F14BCI-6x250y High Speed Rotor For use only in Beckman Coulter Avanti J or HP centrifuges except the Avanti J-HC. Not for use in

More information

Thermo Scientific Fiberlite Rotors. User Reference Manual

Thermo Scientific Fiberlite Rotors. User Reference Manual Thermo Scientific Fiberlite Rotors User Reference Manual Contents Page A. GENERAL INFORMATION 1. Safety Marks and Notices 1 2. Rotor Model Designations 5 3. Characteristics of Composite Materials 5 4.

More information

STANDARD OPERATING PROCEDURES SOP: 1002 PAGE: 1 of 5 REV: 0.0 DATE: 12/23/96

STANDARD OPERATING PROCEDURES SOP: 1002 PAGE: 1 of 5 REV: 0.0 DATE: 12/23/96 PAGE: 1 of 5 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 INTERFERENCES AND POTENTIAL PROBLEMS 4.0 EQUIPMENT/APPARATUS 5.0 PROCEDURE 6.0 QUALITY ASSURANCE/QUALITY CONTROL 7.0 HEALTH AND SAFETY

More information

INSTRUCTION MANUAL. CAT Place Table Top Centrifuge

INSTRUCTION MANUAL. CAT Place Table Top Centrifuge 1 INSTRUCTION MANUAL CAT. 72359-06 6-Place Table Top Centrifuge 2 CONTENTS IMPORTANT SAFETY GUIDELINES PLEASE READ BEFORE USE.. 3 SUPPLIED EQUIPMENT... 5 OPTIONAL ACCESSORIES... 5 SETUP... 5 UNPACKING...

More information

Model : LMHS-60 Please read this manual carefully before using the instrument Labnics Equipment

Model : LMHS-60 Please read this manual carefully before using the instrument Labnics Equipment HIGH SPEED CENTRIFUGES Instruction Manual Model : LMHS-60 Please read this manual carefully before using the instrument Labnics Equipment Before Installation The centrifuge can be damaged if connected

More information

XCM12 Mini-Centrifuge

XCM12 Mini-Centrifuge Multi-speed Digital User Manual v. 1.0 C&A Scientific Disclaimer C & A Scientific owns the copyright of this user manual. No person shall reprint or distribute the contents of this user manual, including

More information

Instructions For Use. Allegra X-30 Series Centrifuges. Rx Only. PN B05529AD November 2015

Instructions For Use. Allegra X-30 Series Centrifuges. Rx Only. PN B05529AD November 2015 Instructions For Use Allegra X-30 Series Centrifuges Rx Only PN B05529AD November 2015 Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA 92821 U.S.A. Allegra X-30 Series Centrifuges PN B05529AD (November

More information

Instructions For Use. SW 28 and SW 28.1 Swinging-Bucket Rotors. For Use in Beckman Coulter Class H, R, and S Preparative Ultracentrifuges

Instructions For Use. SW 28 and SW 28.1 Swinging-Bucket Rotors. For Use in Beckman Coulter Class H, R, and S Preparative Ultracentrifuges Instructions For Use SW 28 and SW 28.1 Swinging-Bucket Rotors For Use in Beckman Coulter Class H, R, and S Preparative Ultracentrifuges L5-TB-069PA October 2011 Beckman Coulter, Inc. 250 S. Kraemer Blvd.

More information

Z100A Clinical Centrifuge

Z100A Clinical Centrifuge Z100A Clinical Centrifuge User Manual C0100-A Rev 2 August 2016 About This Manual This manual is designed to assist you in the optimal usage of your Labnet Z100A Clinical Centrifuge. The manual is available

More information

1250 LB. CAPACITY MECHANICAL WHEEL DOLLY

1250 LB. CAPACITY MECHANICAL WHEEL DOLLY 1250 LB. CAPACITY MECHANICAL WHEEL DOLLY 67287 SET-UP AND OPERATING INSTRUCTIONS Visit our website at: http://www.harborfreight.com Read this material before using this product. Failure to do so can result

More information

An-50 Ti and An-60 Ti Analytical Rotor, Cells, and Counterbalance

An-50 Ti and An-60 Ti Analytical Rotor, Cells, and Counterbalance LXL/A-TB-003F June 2001 TM An-50 Ti and An-60 Ti Analytical Rotor, Cells, and Counterbalance For Use with the Beckman Coulter Optima XL-A and XL-I Analytical Ultracentrifuges 2001 Beckman Coulter, Inc.

More information

Instructions For Use. Allegra X-14 Series. Centrifuges Rx Only in the U.S.A. PN B07493AD January 2018

Instructions For Use. Allegra X-14 Series. Centrifuges Rx Only in the U.S.A. PN B07493AD January 2018 Instructions For Use Allegra X-14 Series Centrifuges Rx Only in the U.S.A. January 2018 Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA 92821 U.S.A. Allegra X-14 Series Centrifuges for In-Vitro Diagnostics

More information

Instructions For Use. Avanti J-15 Series Centrifuges. Rx Only in the U.S.A. PN B80286AC March 2018

Instructions For Use. Avanti J-15 Series Centrifuges. Rx Only in the U.S.A. PN B80286AC March 2018 Instructions For Use Avanti J-15 Series Centrifuges Rx Only in the U.S.A. PN B80286AC March 2018 Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA 92821 U.S.A. Avanti J-15 Series Centrifuges for In Vitro

More information

Owner s Manual & Safety Instructions

Owner s Manual & Safety Instructions Owner s Manual & Safety Instructions Save Save This This Manual Keep Keep this this manual manual for for the the safety safety warnings warnings and and precautions, assembly, assembly, operating, inspection,

More information

VTi 65.2 Rotor. Used In Beckman Coulter Class H, R, and S Preparative Ultracentrifuges. L5-TB-089H December 2007

VTi 65.2 Rotor. Used In Beckman Coulter Class H, R, and S Preparative Ultracentrifuges. L5-TB-089H December 2007 L5-TB-089H December 2007 Used In Beckman Coulter Class H, R, and S Preparative Ultracentrifuges 2007 Beckman Coulter, Inc. Printed in the U.S.A. ! SAFETY NOTICE This safety notice summarizes information

More information

INSTRUCTION MANUAL FOR THE CLINICAL 50 CENTRIFUGE

INSTRUCTION MANUAL FOR THE CLINICAL 50 CENTRIFUGE INSTRUCTION MANUAL FOR THE CLINICAL 50 CENTRIFUGE 82013-800 January 2006 INDEX PAGE 1. General Information... 2 1.1 Description... 2 1.2 Safety precautions to be observed before operating the centrifuge...

More information

CONTENTS. 3 MAINTENANCE 3.1 Service and Maintenance Maintenance and cleaning Glass breakage Disinfection

CONTENTS. 3 MAINTENANCE 3.1 Service and Maintenance Maintenance and cleaning Glass breakage Disinfection 1 PRODUCT DESCRIPTION 1.1 Usage in accordance with safety standards 1.1.1 General Information 1.1.1.1 Hazards and precautions 1.1.1.2 Brief description 1.1.1.3 Safety standards 1.1.1.4 Included items 1.1.1.5

More information

GLO GT416. Benchtop Centrifuge. User s Manual. Glotech Co., Ltd

GLO GT416. Benchtop Centrifuge. User s Manual. Glotech Co., Ltd Benchtop Centrifuge GLO GT416 User s Manual Glotech Co., Ltd Contents 1. GLO GT416G Performance 2. Precautions 3. Technical Specifications 4. Installation 4 1. Delivery Checklist 4 2. Unpacking the Instrument

More information

1580R User s Manual. Multi-Purpose, High Speed Centrifuge. Gyrozen Co., Ltd.

1580R User s Manual. Multi-Purpose, High Speed Centrifuge. Gyrozen Co., Ltd. Multi-Purpose, High Speed Centrifuge 1580R User s Manual Gyrozen Co., Ltd. B Station, 30-12 Gyeryong-ro 141-gil, Yuseoung-gu, Daejeon 305-301, Korea Tel: +82-42-719-8200 Fax: +82-42-826-9848 www.gyrozen.com

More information

FastPrep FP120 Cell Disrupter INSTRUCTION MANUAL

FastPrep FP120 Cell Disrupter INSTRUCTION MANUAL FastPrep FP120 Cell Disrupter INSTRUCTION MANUAL Applies to FP100, FP120, and FP220 Models Qbiogene, Inc. (North America) 2251 Rutherford Road Carlsbad, CA 92008 Phone (800) 424-6101 Fax (760) 918-9313

More information

Nilfisk-CFM SS Vapor Vacuum Instructions for Use and Spare Parts Manual

Nilfisk-CFM SS Vapor Vacuum Instructions for Use and Spare Parts Manual Nilfisk-CFM SS Vapor Vacuum Instructions for Use and Spare Parts Manual Caution: This Nilfisk-CFM vacuum cleaner is not to be used in explosion hazardous areas as serious injury could result. Your new

More information

Manual Operated Floor Jack

Manual Operated Floor Jack Manual Operated Floor Jack OPERATING INSTRUCTIONS Note: There may be some slight differences in the appearance of the various manually-operated floor jacks, however the instructions in this manual apply

More information

Air Operated Diaphragm Pumps Operating and Maintenance Instructions

Air Operated Diaphragm Pumps Operating and Maintenance Instructions Product & Chemical Disclaimer The user must take responsibility in the selection of the products materials of construction. Empire Pumps Ltd will act in an advisory role and offer recommendations; however,

More information

Clinispin CT20 User Manual

Clinispin CT20 User Manual Clinispin CT20 User Manual C0060-WOD-230EU Lit M00031 Rev 1 June, 2012 Safety Precautions NEVER NEVER NEVER NEVER NEVER NEVER NEVER NEVER ALWAYS ALWAYS ALWAYS ALWAYS use the centrifuge in any manner not

More information

SPECTRAFUGE 6C. User Manual C0060 C V

SPECTRAFUGE 6C. User Manual C0060 C V SPECTRAFUGE 6C User Manual C0060 C0060-230V Lit M00031 Rev 2 August 2016 About This Manual This manual is designed to assist you in the optimal usage of your Labnet Spectrafuge 6C Centrifuge. The manual

More information

Assistant s Instrumentation T

Assistant s Instrumentation T Instructions for Use Assistant s Instrumentation 85.2610.00 T Copyright 2015 A-dec Inc. All rights reserved. A-dec Inc. makes no warranty of any kind with regard to this material, including, but not limited

More information

TLS-55 Rotor. Used in Beckman Coulter Optima MAX, MAX-E, TL, and TLX Series Tabletop Ultracentrifuges

TLS-55 Rotor. Used in Beckman Coulter Optima MAX, MAX-E, TL, and TLX Series Tabletop Ultracentrifuges TL-TB-006J November 2001 TM TLS-55 Rotor Used in Beckman Coulter Optima MAX, MAX-E, TL, and TLX Series Tabletop Ultracentrifuges 2001 Beckman Coulter, Inc. PUBLISHED BY THE SPINCO BUSINESS CENTER OF BECKMAN

More information

92831 TEL: (714) FAX:

92831 TEL: (714) FAX: Document No. 1788-40 Copyright 2017 Terra Universal Inc. All rights reserved. Revised June 2017 Terra Universal, Inc. TerraUniversal.com 800 S. Raymond Ave. Fullerton, CA 92831 TEL: (714) 578-6000 FAX:

More information

Lower Polymer Block Cleaning Kit

Lower Polymer Block Cleaning Kit Lower Polymer Block Cleaning Kit For Applied Biosystems 3730/3730xl DNA Analyzers and Applied Biosystems 3130/3130xl Genetic Analyzers Protocol November 15, 2004 2:23 pm, CleanKit_Title.fm Copyright 2004

More information

p.t.o. Slip clutch Read this material before using this product. Failure to do so can result in serious injury. Save this manual.

p.t.o. Slip clutch Read this material before using this product. Failure to do so can result in serious injury. Save this manual. p.t.o. Slip clutch 65517 Installation Instructions Distributed exclusively by Harbor Freight Tools. 3491 Mission Oaks Blvd., Camarillo, CA 93011 Visit our website at: http://www.harborfreight.com Read

More information

materials and workmanship for 2 years. Should the centrifuge require warranty or Phone: or Fax:

materials and workmanship for 2 years. Should the centrifuge require warranty or Phone: or Fax: WARRANTY The Drucker Company warranties that this centrifuge is free from defects in materials and workmanship for 2 years. Should the centrifuge require warranty or out-of-warranty service please contact:

More information

Thermo Scientific Heraeus Primo and Primo R Centrifuges. Flexible and efficient, compact high-performance centrifuges

Thermo Scientific Heraeus Primo and Primo R Centrifuges. Flexible and efficient, compact high-performance centrifuges Thermo Scientific Heraeus Primo and Primo R Centrifuges Flexible and efficient, compact high-performance centrifuges Thermo Scientific Heraeus Primo and Primo R Centrifuges The compact high-performance

More information

OMNI GLH GENERAL LABORATORY HOMOGENIZER USER MANUAL REV. D H258-GLH

OMNI GLH GENERAL LABORATORY HOMOGENIZER USER MANUAL REV. D H258-GLH OMNI GLH GENERAL LABORATORY HOMOGENIZER USER MANUAL REV. D H258-GLH WARRANTY INFORMATION This manual is a guide for the use of the Omni GLH, General Laboratory Homogenizer and accessories. Data herein

More information

Instruction Manual for the Revolutionary Science RS-102 RevSpin Microcentrifuge

Instruction Manual for the Revolutionary Science RS-102 RevSpin Microcentrifuge Instruction Manual for the Revolutionary Science RS-102 RevSpin Microcentrifuge REVOLUTIO NARY SCIENCE Manufacturer of Precision Laboratory Equipment Table of Contents Introduction 2 Recommended Safeguards

More information

Instruction Manual. Labnet Spectrafuge 7M Microcentrifuge

Instruction Manual. Labnet Spectrafuge 7M Microcentrifuge Instruction Manual Labnet Spectrafuge 7M Microcentrifuge Labnet International PO Box 841 Woodbridge, NJ 07095 Phone: 732 417-0700 Fax: 732 417-1750 email: labnet@labnetlink.com IMPORTANT NOTICE The use

More information

LUBRICATOR GUN INSTRUCTIONS-PARTS LIST. 10,000 psi (700 bar) Maximum Delivery Pressure. Detachable-type

LUBRICATOR GUN INSTRUCTIONS-PARTS LIST. 10,000 psi (700 bar) Maximum Delivery Pressure. Detachable-type INSTRUCTIONS-PARTS LIST 306 460 INSTRUCTIONS This manual contains important warnings and information. READ AND KEEP FOR REFERENCE. Rev. E Supercedes D Detachable-type LUBRICATOR GUN 10,000 psi (700 bar)

More information

NEUTRALIZATION SYSTEM FOR OPA & GLUTARALDEHYDE MEDI-NEWT. User Manual. Endoscope reprocessing the the way way it it should be be

NEUTRALIZATION SYSTEM FOR OPA & GLUTARALDEHYDE MEDI-NEWT. User Manual. Endoscope reprocessing the the way way it it should be be NEUTRALIZATION SYSTEM FOR OPA & GLUTARALDEHYDE MEDI-NEWT User Manual Endoscope reprocessing the the way way it it should be be 1 Medi-Newt Neutralization System Instruction Manual 50096-951 Rev C 2011,

More information

L5-TB-104B April NVT 65.2 Rotor. Used in Beckman Coulter Class H, R, and S Preparative Ultracentrifuges Beckman Coulter, Inc.

L5-TB-104B April NVT 65.2 Rotor. Used in Beckman Coulter Class H, R, and S Preparative Ultracentrifuges Beckman Coulter, Inc. L5-TB-104B April 2008 NVT 65.2 Rotor Used in Beckman Coulter Class H, R, and S Preparative Ultracentrifuges 2008 Beckman Coulter, Inc. ! SAFETY NOTICE This safety notice summarizes information basic to

More information

Submersible Pond Pump with UV 800 Gallons Per Hour

Submersible Pond Pump with UV 800 Gallons Per Hour Submersible Pond Pump with UV 800 Gallons Per Hour REMINDER CALL 1-888-755-6750 BEFORE RETURNING TO STORE. PACKAGE CONTENTS ITEM #PP800UV Questions, problems, missing parts? Before returning to your retailer,

More information

Instruction Manual for the Revolutionary Science RS-200 RevSpin Microcentrifuge

Instruction Manual for the Revolutionary Science RS-200 RevSpin Microcentrifuge Instruction Manual for the Revolutionary Science RS-200 RevSpin Microcentrifuge REVOLUTIO NARY SCIENCE Manufacturer of Precision Laboratory Equipment Table of Contents Introduction 2 Recommended Safeguards

More information

TBE-700. Operation, Safety and Maintenance Manual. Stainless Steel Weld Cleaning System. Ensitech Pty Ltd

TBE-700. Operation, Safety and Maintenance Manual. Stainless Steel Weld Cleaning System. Ensitech Pty Ltd Stainless Steel Weld Cleaning System Operation, Safety and Maintenance Manual TBE-700 TBE-700 Instruction Manual Oct 2013 (USA) FOREWORD FOREWORD This manual is a very important tool! Keep it with the

More information

Automatic Emergency Light

Automatic Emergency Light Automatic Emergency Light Item 38013 Read this material before using this product. Failure to do so can result in serious injury. Save this manual. When unpacking, make sure that the product is intact

More information

Swing Arm Magnifying Lamp

Swing Arm Magnifying Lamp Owner s Manual & Safety Instructions Save This Manual Keep this manual for the safety warnings and precautions, assembly, operating, inspection, maintenance and cleaning procedures. Write the product s

More information

Full View Flow Indicator

Full View Flow Indicator Full View Flow Indicator Threaded and Flanged Process Connection Installation / Operation / Maintenance Manual P.O. Box 1116 Twinsburg, OH 44087 Phone: 330/405-3040 Fax: 330/405-3070 E-mail: view@ljstar.com

More information

CAMBELT TENSION GAUGE

CAMBELT TENSION GAUGE CAMBELT TENSION GAUGE Model 96557 Operating Instructions Diagrams within this manual may not be drawn proportionally. Due to continuing improvements, actual product may differ slightly from the product

More information

Model G14SD Vapor Management System

Model G14SD Vapor Management System Instruction Manual Model G14SD Vapor Management System (For Models: DSD and SSD 110/115v) Manufactured in USA for Medivators Reprocessing Systems by: PCI Medical Inc. 12 Bridge Street, PO Box 188 Deep

More information

Max. 22,000 rpm / 48,700 xg, full 3 liter capacity

Max. 22,000 rpm / 48,700 xg, full 3 liter capacity Max. 22,000 rpm / 48,700, full 3 liter capacity Easy-to-use Compact Design (only 50cm wide) Features 3 Liter Capacity 6 x 500mL bottles can be centrifuged at 12,090. Auto Interlock Lid To close lid Push

More information

OPW Installation and Maintenance Instructions OPW Series Primary and Secondary Bucket Replacement Instructions

OPW Installation and Maintenance Instructions OPW Series Primary and Secondary Bucket Replacement Instructions OPW Installation and Maintenance Instructions OPW 1-3100 Series Primary and Secondary Bucket Replacement Instructions IMPORTANT: Please read these warnings and follow the assembly instructions completely

More information

25 GALLON PORTABLE OIL LIFT

25 GALLON PORTABLE OIL LIFT 25 GALLON PORTABLE OIL LIFT Model 92859 SET UP AND OPERATING INSTRUCTIONS Diagrams within this manual may not be drawn proportionally. Due to continuing improvements, actual product may differ slightly

More information

Instructions for Using the A-100/18 and A-100/30 Rotors

Instructions for Using the A-100/18 and A-100/30 Rotors AF-TB-009H October 2007 Instructions for Using the A-100/18 and A-100/30 Rotors In the Beckman Coulter Airfuge Ultracentrifuge 18 A-100/18 rmin rav rmax Axis of Rotation U.S. Pat. No. 3,997,104 SPECIFICATIONS

More information

Viscount I Hydraulic Motor and Displacement Pump

Viscount I Hydraulic Motor and Displacement Pump INSTRUCTIONS-PARTS LIST 308 674 INSTRUCTIONS This manual contains important warnings and information. READ AND KEEP FOR REFERENCE. Rev. C Supersedes Rev. B Viscount I Hydraulic Motor and Displacement Pump

More information

Model P-40 & Model P-25 POWER PUSHER

Model P-40 & Model P-25 POWER PUSHER Power Pusher Description INSTRUCTION MANUAL The Power Pusher provides ram capability by using the spreading power of the POWER HAWK P-16 Rescue Tool. (The Power Pusher may also be used with other spreader

More information

HATCHGRIP Installation Instructions/Operation and Maintenance Manual

HATCHGRIP Installation Instructions/Operation and Maintenance Manual HATCHGRIP Installation Instructions/Operation and Maintenance Manual Models: HTG-PCG Contact Information Table of Contents: Safety Precautions... 2 Product Information... 2 Operation... 3 Installation

More information

TL-TB-023D February MLS-50 Rotor. Used in Beckman Coulter Optima MAX, MAX-XP, and MAX-E Tabletop Ultracentrifuges Beckman Coulter, Inc.

TL-TB-023D February MLS-50 Rotor. Used in Beckman Coulter Optima MAX, MAX-XP, and MAX-E Tabletop Ultracentrifuges Beckman Coulter, Inc. TL-TB-023D February 2008 MLS-50 Rotor Used in Beckman Coulter Optima MAX, MAX-XP, and MAX-E Tabletop Ultracentrifuges 2008 Beckman Coulter, Inc. ! SAFETY NOTICE This safety notice summarizes information

More information

Falcon Type 2 Upgrade Kit

Falcon Type 2 Upgrade Kit 1 Falcon Type 2 Upgrade Kit Kit # 00-10-32-400-160 Universal, Modular Cartridge System, Series 3 / Type 2, All 4, 1.60" Bore Kit # 00-10-32-100-160 Universal, Modular Cartridge System, Series 3 / Type

More information

APCO CRF-100A RUBBER FLAPPER SWING CHECK VALVES

APCO CRF-100A RUBBER FLAPPER SWING CHECK VALVES APCO CRF-100A RUBBER FLAPPER SWING CHECK VALVES Instruction D12043 June 2016 DeZURIK Instructions These instructions provide installation, operation and maintenance information for APCO CRF-100A Rubber

More information

Advanced Calibration Designs, Inc. Release II

Advanced Calibration Designs, Inc. Release II Advanced Calibration Designs, Inc. Release II www.goacd.com WARNING: This instrument generates test gas for toxic gas detectors. The instruction manual should be read and understood prior to operation

More information

Installation Power Management Unit Battery Cables and Battery Harness

Installation Power Management Unit Battery Cables and Battery Harness Installation Power Management Unit Battery Cables and Battery Harness Important Safety Messages SAVE THESE INSTRUCTIONS - This manual contains important instructions that should be followed during installation

More information

INSTALLATION, CALIBRATION & TROUBLESHOOTING MANUAL

INSTALLATION, CALIBRATION & TROUBLESHOOTING MANUAL INSTALLATION, CALIBRATION & TROUBLESHOOTING MANUAL Model 7200 Transducers GP:50 2770 Long Road, Grand Island, NY 14072 Tel: 716-773-9300 Fax: 716-773-5019 Email: sales@gp50.com Website: http://www.gp50.com

More information

GENERAL LABORATORY HOMOGENIZER OWNER S MANUAL

GENERAL LABORATORY HOMOGENIZER OWNER S MANUAL GENERAL LABORATORY HOMOGENIZER OWNER S MANUAL Cole-Parmer Instrument Company 625 East Bunker Court Vernon Hills, IL 60061 USA Toll Free: 1-800-323-4340 Phone: 1-847-549-7600 Fax: 1-847-247-2929 Hearing

More information

Owner s Manual & Safety Instructions

Owner s Manual & Safety Instructions Owner s Manual & Safety Instructions Save This Manual Keep this manual for the safety warnings and precautions, assembly, operating, inspection, maintenance and cleaning procedures. Write the product s

More information

580RK SERVICE GUIDELINES For coupling models: 580 & 580J

580RK SERVICE GUIDELINES For coupling models: 580 & 580J 580 Coupling THE FIRST NAME IN QUALITY COUPLINGS 580RK SERVICE GUIDELINES For coupling models: 580 & 580J BEFORE GETTING STARTED: This procedure should only be performed by a qualified mechanic. Measure

More information

Stealth Power i-tilt

Stealth Power i-tilt Power Tilts OWNERS MANUAL Stealth Power i-tilt User Manual and Maintenance Guide for Stealth s i-tilt Customer Satisfaction 1.0 Stealth Products strives for 100% customer satisfaction. Your complete satisfaction

More information

NECO Pumping Systems

NECO Pumping Systems INSTALLATION OPERATION & MAINTENANCE INSTRUCTIONS For Your NECO Pumping Systems PACKAGED CIRCULATING SYSTEM THIS COMPLETELY ASSEMBLED, TESTED, PACKAGED CIRCULATING SYSTEM IS OF THE HIGHEST QUALITY AND

More information

INSTRUCTION MANUAL FOR BENCH TOP CENTRIFUGE Z 200 A

INSTRUCTION MANUAL FOR BENCH TOP CENTRIFUGE Z 200 A INSTRUCTION MANUAL FOR BENCH TOP CENTRIFUGE Z 200 A Labnet International, Inc. PO Box 841 Woodbridge, NJ 07095 Phone: (732) 417-0700 Fax: (732) 417-1750 1 INDEX...Page 1. General Information 1.1 Precautions

More information

Model NTX7 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006

Model NTX7 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006 B R A N D Model NTX7 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006 For Sales, Support and Service phone: 407-331-4793 fax: 407-331-4708 website: www.xenotronix.com email: information@xenotronix.com

More information

TLA-110 Rotor. Used In Beckman Coulter Optima MAX-XP, MAX, MAX-E, TL, TLX, and TL-100 Tabletop Ultracentrifuges. TL-TB-019C September 2007

TLA-110 Rotor. Used In Beckman Coulter Optima MAX-XP, MAX, MAX-E, TL, TLX, and TL-100 Tabletop Ultracentrifuges. TL-TB-019C September 2007 TL-TB-019C September 2007 TLA-110 Rotor Used In Beckman Coulter Optima MAX-XP, MAX, MAX-E, TL, TLX, and TL-100 Tabletop Ultracentrifuges 2007 Beckman Coulter, Inc. Produced in the U.S.A. ! SAFETY NOTICE

More information

Lower Extremities USER MANUAL RLSHW100. Stealth s User Manual and Maintenance Guide for Residual Limb Support

Lower Extremities USER MANUAL RLSHW100. Stealth s User Manual and Maintenance Guide for Residual Limb Support Lower Extremities USER MANUAL RLSHW100 Stealth s User Manual and Maintenance Guide for Residual Limb Support Customer Satisfaction 1.0 Stealth Products strives for 100% customer satisfaction. Your complete

More information

Sentinel Enteral Feeding Pump. Model: S-1000-SI. Operating Manual

Sentinel Enteral Feeding Pump. Model: S-1000-SI. Operating Manual Sentinel Enteral Feeding Pump Model: S-1000-SI Operating Manual CONTENTS 1. Introduction.. 2 2. Control Panel 5 3. Operating Instructions.. 6 4. Indicators and Alarms.. 9 5. Maintenance.. 10 6. Troubleshooting

More information