Bruker BioSpin NMR CASE. Installation Manual. Version. NMR Spectroscopy. think forward

Similar documents
Bruker BioSpin NMR CASE. Troubleshooting Guide. Version. NMR Spectroscopy. think forward

Bruker BioSpin MATCH. Multiple Adjustable Tube Clamp Holder User Manual - Version 005. NMR Solutions. think forward

minispec Plus Release Letter Innovation with Integrity Version 001 AIC

AUTO REWIND AIR HOSE REEL

TB-690 Conveyor/Stacker

MAS Rotor Test Station

COMPONENT WORK SAMPLE 8 Simulated Assembly MAINTENANCE MANUAL. From 1974 to March 15, 2003

Installation Instructions Right Hand Drive Megashifter

Installation Instructions Z-Gate Shifter

Hurst VMATIC3 INSTALLATION

Installation Instructions Megashifter

SERIES 6000 FILM OPACITY METER MODEL RT REV A.1 INSTALLATION & OPERATION MANUAL

Light Truck MegaShifter

Installation Instructions. QuickSilver Shifter. Fits: GM, Ford, Chrysler Transmissions See Application Guide for Specific Applications Part # 80683

Installation & Programming Manual. Quick Reference

Installation Instructions Sport Shifter

Prerequisites: Shop Manual (recommended) pages 3-9 through 3-13.

VECTRIX VX-2 SERVICE MANUAL. Version 1.0/May 2011 VECTRIX, LLC

OPERATOR'S MANUAL AND MAINTENANCE INFORMATION MODEL T-2000 TEXTURE TEST SYSTEM

Instruction Manual. Coating Thickness Gauge. Optionally with Memory

<THESE INSTRUCTIONS MUST BE GIVEN TO THE END USER> BASE BOLT BAG (RVB3500) ITEM DESCRIPTION QTY. 1 2 " X 1 1 2" Hex Cap Screws 16

Installation Instructions Table of Contents

COMPONENT WORK SAMPLE 14 Integrated Peer Performance MAINTENANCE MANUAL

IOTA ONE OPERATING MANUAL AND SETUP INSTRUCTIONS IOTA ONE. Please Read this Instruction Manual COMPLETELY Before Operating the System IOTA ONE

Z-Gate Universal Shifter

A B C D E F. Tools Required (supplied by others)

HD4170 SETUP & OPERATOR MANUAL. Performance Design Inc East Braniff St. Boise, Idaho 83716

RAAS Troubleshooting Guide. Confirm that the power switch to the unit is in the on position.

Agilent G1554A On-Column Capillary Inlet with EPC

<THESE INSTRUCTIONS MUST BE GIVEN TO THE END USER> 1 2 " X 1 1 2" Hex Cap Screws " Split Lock Washers " Threaded block 4

Welker Sampler. Installation, Operation, & Maintenance Manual. Model GSS-4HP

Items # and #

Setup and Programming Manual

Upgrading the Sure Coat Modular Gun Control System

iphone 5s Front Panel Replacement

EP1022-U1 MASSAGE LOUNGER SPECIFICATIONS ORDER NO.HPD0104U00C1. Power consumption. : Approx. 500 times/min.(per side)

If You Are Experiencing This, Click Thru:

METROLOGIC INSTRUMENTS, INC. MX001 Industrial Control Interface Installation and User s Guide

A B C D E F. b.tools Required (supplied by others) 3/16" Drill Bit 3/8" Wrench Phillips Head Screwdriver

SECTION B1: MACHINE COMPONENTS

Installation Instructions

BOVIBOOSTER WDone & WDtwo

1 Green Pressure Regulator Spring Automatic transmissions operate at temperatures between 150ºF and

All images and other materials are protected by United States copyright law and may not be reproduced, distributed, transmitted, displayed, published

with Touch & Go Feature

SECTION B: SYSTEM OPERATING COMPONENTS

Retractable System Operation & Maintenance Manual

Installation Instructions Console Megashifter

IMPORTANT! DO NOT THROW AWAY THE SHIPPING CARTON AND PACKING MATERIAL

Section 7 - Troubleshooting Guide

R800-16PCPS DUAL REDUNDANT POWER SUPPLY FOR THE R RACK USER MANUAL

GENERAL <ELECTRICAL>

GENERAL <ELECTRICAL>

VTCM Installation Manual Table of Contents

The M7 Direct Flow Intercooler (DFIC) Installation Guide

Uplift Power Seat Users Guide

MP18 Stacking Valve System Technical Information Manual

Easy Tool Retracting System SR2159

CARE AND USE OF 384 BUBBLE PADDLE RESERVOIRS AND MOTOR/MAGNETIC CLUTCH UNIT. Setup & operation

Electropneumatic Converters i/p Converters Type 6111 Mounting and Operating Instructions EB 6111 EN

For additional information, please visit our website at SENSOR SPECIFICATIONS

AL625 & AL625HD INSTALLATION & OWNER S MANUAL

WEBER CARBURETOR TROUBLESHOOTING GUIDE

GPS Steering System Installation Manual

Cellular Shades MOTORIZED SHADE Top Down - Bottom Up. Installation & Care Instructions

Installation Guide Current Ford F-250 & Ford F-350 Super Duty. Product Code: 109 & 119

Serena Remote Controlled Shades

Adjustable Angled Incline Conveyor Owners Manual with Operating Instructions

Roller Shades MOTORIZED SHADE. Simplicity. Installation & Care Instructions

Emergency Power Supply for the OMEGA ACTIVE-System. Operating Instructions

Assisted Steering Hydraulic Installation Manual for Vehicle Kit Number Fits Only AGCO Gleaner R62/R65 (after S/N 66000) and R72/R75 (after

OLYMPIAN MODEL 740 Operation and Service Manual

TCI FastGate Shifter Installation Instructions

SL-10/20/30, ST-10/10Y - Motor Coupling - Alignment

Pellicon 2 and 3 Mini Holder. User Guide

Installation Manual TWM Performance Short Shifter Cobalt SS/SC, SS/TC, HHR SS, Ion Redline and Saab 9-3

Ideal Installation. I & M Mark 67 (1/2 6 ) Control Line. Installation & Maintenance Instructions for Mark 67 Pressure Regulators

Installation Instructions Unimatic Shifter

TC62D Installation Instructions

THANK YOU FOR CHOOSING BATHERBOX GETTING STARTED WHAT S IN THE BOX? N. Northview Ave Sioux Falls, SD 57107

Be Strong. NS 200X. Assembly Manual. Model: NS 200X. P/N: Rev A (10/04/2006)

INSTALLATION AND SERVICE MANUAL FOR TURBO-CARB HIGH CAPACITY CARBONATOR

Installation and Service Instructions for 87,000 Series Manual Adjust Brakes (rev. B)

SERVICE MANUAL HOT DELI 3/4/5/7 MERCHANDISERS

WARM ENGINE STARTING PROCEDURE

PROBLEM SOLVING GUIDE ELMECO SLUSH MACHINE MODEL FCM

PET UNIT DOSE TABLE INSTALLATION MANUAL FN: Rev A 1/18

Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by:

PET UNIT DOSE CABINET

Please contact BrakeAway Products tech support for additional (503) or

Model 2300JL Installation Guide

MAX FLOW FUEL KIT KIA FORTE SX HYUNDAI VELOSTER TURBO INSTALLATION INSTRUCTIONS

HM860 FLOWMETER OPERATOR S MANUAL DO NOT USE OR OPERATE THIS EQUIPMENT UNTIL THIS MANUAL HAS BEEN READ AND THOROUGHLY UNDERSTOOD

MODEL 660 AUTOMATIC FASTENING CENTER OPERATOR S MANUAL

IL500 Sierra Inclined Platform Lift

OPERATION & MAINTENANCE INSTRUCTIONS

Thank you for purchasing the Craven Speed FlexPod Complete Gauge Pod Kit

Section 3 Technical Information

Model 2300DL Installation Guide

CORPORATION OPERATOR S MANUAL HWH HYDRAULIC SPACEMAKER ROOM EXTENSION SYSTEMS. FEATURING: Single / Multiple Room Extensions / Generator Slide

Transcription:

Bruker BioSpin NMR CASE Installation Manual Version 011 think forward NMR Spectroscopy

Copyright by Bruker BioSpin NMR GmbH All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form, or by any means without the prior consent of the publisher. Product names used are trademarks or registered trademarks of their respective holders. This manual was written by Martin Rindlisbacher January 29, 2004: Bruker Biospin Corporation Billerica, MA, USA This manual was desktop published by Stanely J. Nles January 8, 2009: Bruker Biospin GmbH Rheinstetten, Germany P/N: B2926 DWG-Nr.: 0000011 For further technical assistance on the NMR CASE unit, please do not hesitate to contact your nearest BRUKER dealer or contact us directly at: BRUKER BioSpin GMBH am Silberstreifen D-76287 Rheinstetten Germany Phone: + 49 721 5161 0 FAX: + 49 721 5171 01 Email: service@bruker.de Internet: www.bruker.com

Contents Contents... 3 1 Introduction... 5 1.1 Warnings... 5 1.2 Disclaimer... 5 1.3 System Requirements... 5 1.4 Non-magnetic Tool Set P/N B2852... 6 1.5 Non-magnetic Ladder... 6 2 Parts List... 7 2.1 NMR CASE Parts List... 7 3 Installation Instructions... 9 3.1 Short Installation Description... 9 3.2 Mount the Front Legs to the Base... 9 3.3 Mount the Base to the Shim Stack... 9 3.4 Connect the Pneumatic Hose Assembly... 10 3.5 Installation of the Motion Controller Unit... 11 Placement... 11 Main Air Connection... 11 Power... 11 Pneumatic Hose Assembly...11 Test Connector (9 pins)...12 3.6 Switch ON the Motion Controller... 12 3.7 Mount the Sample Holder Tray... 12 3.8 First Manual Test Without Samples... 13 3.9 First Manual Test With Samples... 13 3.10 Adjustments for Reliable Long Term Operation... 14 4 Operation... 15 4.1 Automatic Operation... 15 4.2 Manual Operation... 15 4.3 Second Sample Holder Tray... 16 4.4 Sample Diameters... 17 4.5 Temperature Range... 17 4.6 Troubleshooting... 18 A Illustrations... 19 Figures... 29 Installation Manual Version 011 BRUKER BIOSPIN 3

Contents 4 BRUKER BIOSPIN Installation Manual Version 011

Introduction 1 1 Warnings 1.1 WARNING! The NMR CASE has to be installed on top of a NMR magnet cryostat. DANGER! This magnet creates a very STRONG MAGNETIC FIELD. Objects and tools can be attracted to the magnet with great force and may cause damage. Use only NON-MAGNETIC tools and equipment. A non-magnetic TOOL SET is supplied with the NMR Case. Read and follow carefully this installation manual.product description, usage, scope of manual, etc. Disclaimer 1.2 BRUKER accepts NO responsibility for damages resulting from: 1. Installations by untrained personnel. 2. Not following the manuals provided. System Requirements 1.3 Before installing the NMR Case, read and follow the System Requirement manual P/N B2928. Installation Manual Version 011 BRUKER BIOSPIN 5 (31)

Introduction Non-magnetic Tool Set P/N B2852 1.4 Parts list: One non-magnetic Philips screwdriver #2, P/N 92437 One collar, mounted to the screwdriver, P/N B2817 One non-magnetic wrench 10 mm, P/N B2438 One circular level, P/N 92438. If you cannot find this non-magnetic tool set, call BRUKER and order one (The Bruker part number is B2852). WARNING! To avoid damage to the magnet cryostat, use only NON - MAG- NETIC tools. Non-magnetic Ladder 1.5 To install the NMR case on top of the magnet cryostat, a non-magnetic ladder made from wood or aluminum must be available. Bruker does not supply nonmagnetic ladders. WARNING! To avoid damage to the magnet cryostat, use only NON - MAG- NETIC equipment. 6 (31) BRUKER BIOSPIN Installation Manual Version 011

Parts List 2 2 NMR CASE Parts List 2.1 One NMR Case base assembly One gray security ring (mounted to base) Two sets of telescopic front legs One (or several) exchangeable sample holder trays One BST adapter collar (mounted to base) One motion controller unit Two mounting brackets One power cord One spare fuse (inside the fuse holder) One pneumatic hose Ø 4 mm One pneumatic cable assembly One non-magnetic tool set One pneumatic hose Ø 6 mm One adapter Ø 6 to Ø 8 mm One T piece Ø 8 mm One BSMS sample control modifier cable One System requirement manual One Trouble shooting guide One instruction sheet Interface tube for new BST This manual One instructions for modifier cable P/N B2436 P/N B2383 P/N B3165 P/N B3015 P/N B2874 P/N B1428 P/N B2928 P/N B2927 P/N B3696 P/N B2926 P/N B2300 Installation Manual Version 011 BRUKER BIOSPIN 7 (31)

Parts List 8 (31) BRUKER BIOSPIN Installation Manual Version 011

Installation Instructions 3 3 Short Installation Description 3.1 Mount the front legs to the NMR Case base assembly. Mount the base assembly to the shim upper stack. Align the base to the magnet cryostat. Adjust the length of both telescopic front legs and level the base. Connect the pneumatic hoses. Install the motion controller unit Switch on the motion controller. Install a sample holder tray. First tests. Operation. Refer to Appendix "Illustrations" on page 19 for figures and illustrations supporting the installation and operation. Mount the Front Legs to the Base 3.2 Locate the two sets of telescopic front legs. Mount the legs using the plastic hex screws. Adjust the legs into the shortest position. Tighten all eight plastic hex screws with the non-magnetic wrench. Mount the Base to the Shim Stack 3.3 The shim stack has to have a specific orientation. Read and follow section 1.5 of the System requirement manual P/N B2928. The interface tube has to be selected to match the BST. Read and follow the instruction sheet Interface tube for new BST P/N B3696. 1. On the base plate, locate the adapter collar and remove the warning label from the gray safety ring. 2. Rotate the gray safety ring until the three screws for the collar are visible through the openings. Installation Manual Version 011 BRUKER BIOSPIN 9 (31)

Installation Instructions 3. Using the non-magnetic Philips screwdriver, loosen these three screws, but do not remove them. 4. Check that the collar can now be rotated. 5. Use a non-magnetic ladder and remove all labels from the shim stack top head. Refasten the labels to the lower, narrow part of the shim stack. 6. Lift up the NMR Case base and line up the openings of the adapter collar with the corresponding nipples of the shim stack. 7. Push the collar down over the top head of the shim stack. It should sit firmly and one should feel the engagement of the O-ring inside the adapter collar with the shim stack top head. See Figure A.3. to Figure A.5. for details. 8. Using the non-magnetic wrench, loosen the white hex screws for the front legs and pull the legs down until they rest on the magnet cryostat. Rotate the lower leg piece until it is lined up with the surface of the magnet cryostat. 9. Rotate the NMR Case base plate until there is equal distances from both nitrogen (and helium) towers. 10. Place the circular level (from the tools set) on top of the NMR Case base plate. 11. Observing the bubble of the circular level, adjust the length of both front legs until the base plate is level. The bubble must be inside the innermost ring. Tighten all white hex screws of the front legs with the non-magnetic wrench. 12. Using the non-magnetic Phillips screwdriver, tighten all three screws on the adapter collar. 13. Check the circular level. If necessary, readjust the length of the front legs until the base plate is level. 14. Unscrew the screw for the base cover and lift up the base cover slightly. 15. Remove the gray safety ring. Check that the seal ring and the gasket are still in place. 16. Place back the base cover and fasten the screw. Connect the Pneumatic Hose Assembly 3.4 The sample lift air hose going to the shim stack needs to be rerouted through the NMR Case. Remove the sample lift air hose (white color) from the shim stack and push it into the corresponding connector located on the manifold below the NMR Case base plate. With a short piece of 8-mm hose, connect the outlet on the other side of the manifold back to the shim stack sample lift air inlet. Connect the pneumatic hose assembly to the NMR Case manifold. The hoses are labeled in sequence # 5, 1, 2, 3, 4, 6, 7. Push the hoses # 5, 1, 2, 3, 4, 6 all the way into the pneumatic connectors located on the manifold below the NMR Case base plate. Hose # 7 goes into the pneumatic connector located at the adapter collar. 10 (31) BRUKER BIOSPIN Installation Manual Version 011

Installation Instructions Installation of the Motion Controller Unit 3.5 Placement 3.5.1 Place the motion controller unit on top of a rack or some other convenient place. It must be at least 2 meters (6 feet) away from the magnet cryostat dewar. The environment temperature range must be 0 C to + 50 C. It can also be mounted inside the NMR console. For console mounting, remove the black rails of the motion controller and mount two flat mounting brackets instead (B3015). Remove the back panel of the NMR console and mount the motion controller in a free spot (usually in the top left corner). WARNING! The motion controller unit is magnetic and does not work in the magnetic field. It can be attracted to the magnet with great force. It must be 2 meters (6 feet) away from the magnet cryostat dewar. Main Air Connection 3.5.2 Connect the motion controller inlet marked MAIN AIR to a compressed air supply (for specifications, see the manual System Requirements P/N B2928, section 1.7). To make this connection, use a length of 6 mm hose, a 6 to 8 mm adapter and a Tee piece (all supplied with the NMR Case). Power 3.5.3 Using the power cable, connect the inlet marked 110/220 VAC to a power source, either 110 or 220 V and 50 or 60 Hz. The unit selects the voltage automatically, there is no voltage selector switch. Pneumatic Hose Assembly 3.5.4 Connect the pneumatic hose assembly to the motion controller as follows: Controller I/O Hose label Hose NMR Case CYL A, L-IN L-IN # 1 Manifold # 1 CYL A, L-OUT L-OUT # 2 Manifold # 2 CYL B, EXT EXTEND # 3 Manifold # 3 Installation Manual Version 011 BRUKER BIOSPIN 11 (31)

Installation Instructions CYL B, RET RETRACT # 4 Manifold # 4 SENSOR 1 SENSOR # 5 Manifold # 5 SL-SENSE SL-SENS # 6 Manifold # 6 SENSOR 2 SENSOR 2 # 7 Adapter Collar Test Connector (9 pins) 3.5.5 Do not connect anything to the nine pin connector. It is for test purposes only. Switch ON the Motion Controller 3.6 At this time the motion controller can be turned on. Observe the black push lever on the side of the case body. It should move to the right and come out all the way. Test that the push lever is powered by air by grabbing it hand and trying to move it to the left and in. There should be a firm resistance. WARNING! Leave the motion controller ON at all times during operation. Switch it off only for test purposes or if it is not used anymore. Press the reset button. This clears the sample tracking memory of the controller. Mount the Sample Holder Tray 3.7 For the first test, leave the sample holder tray empty without samples. 1. Locate the sample holder tray. 2. Make sure that all samples are removed from the tray. 3. Locate the green START position. 4. While pushing the tray lock lever inside the tray to the position Unlock, rotate the sample stop head to the location of the START position. 5. Let go the tray lock lever. Check that it snaps back to the locked position. 6. Grab and lift the sample holder tray at the handle. 7. Line up and lower the sample holder tray over the NMR Case base cover. 12 (31) BRUKER BIOSPIN Installation Manual Version 011

Installation Instructions 8. Line up the sample stop head with the black off-center pin and engage it. 9. Line up the center with the black on-center pin and engage it. 10. Engage the tray into the side pin. 11. The sample holder tray settles down. Make sure that it seated correctly all the way down. 12. Check that the tray lock lever is in the unlocked position. First Manual Test Without Samples 3.8 1. Manually switch the sample lift ON. 2. After about 6 seconds the sample holder tray should rotate away from position 0 and settle on position 1. 3. Manually switch OFF the sample lift. 4. Wait until the sample lift airflow is completely OFF (no noise anymore), then switch it on again. 5. The NMR Case motion controller should advance (rotate) the sample holder tray to the next sample position. 6. This sequence can be continued as necessary by switching the sample lift airflow manually OFF and ON. If this test is NOT successful, then read and follow the NMR Case trouble shooting guide, P/N B2927 Important: The sample holder tray advancing movement needs to be adjusted. Read and follow section 1.2.5 and 6 of the troubleshooting guide. First Manual Test With Samples 3.9 For the first tests, use inexpensive and empty sample tubes only. 1. Place two (or more) spinners with empty sample tubes into the next available positions of the sample holder tray on top of the magnet. 2. Manually switch the sample lift ON. 3. The NMR Case motion controller should sense the sample lift (after 6 seconds) and advance (rotate) the sample holder tray to the next sample position containing the first sample. 4. For about 2 seconds the sample tube should float on the sample lift air, before it is inserted rapidly (1 sec) down into the BST shim stack. There it should float again. Installation Manual Version 011 BRUKER BIOSPIN 13 (31)

Installation Instructions 5. Manually switch the sample lift OFF. 6. With the decreasing sample lift air-flow, the sample should slowly go down all the way into the shim coil. 7. The NMR case motion controller should sense the absence of sample lift airflow and retract the advance lever (but not rotate the sample holder tray yet). 8. Manually switch the sample lift ON. 9. With the increasing sample lift airflow, the sample should come up all the way into the sample holder tray within 10 seconds. If it does not come up all the way or if it takes longer than 10 seconds, the sample lift airflow must be increased. Read and follow section 1.1 and 1.3 of the troubleshooting guide. 10. The motion controller should sense the sample top position and advance (rotate) the sample holder tray to the next position. 11. The next sample should be inserted down into the shim stack and float on the sample lift air. 12. This sequence can be continued as necessary by switching the sample lift airflow manually OFF and ON. If this test is NOT successful, then read and follow the NMR Case trouble shooting guide, P/N B2927 Adjustments for Reliable Long Term Operation 3.10 The following 3 items need to be properly adjusted for a reliable long-term operation of the NMR Case. Read and follow the NMR Case troubleshooting guide, P/ N B2927, as indicated: Item Necessary Adjustments Troubleshooting Guide 1 Level of the NMR Case Section 2.1 2 Sample lift air flow Section 1.1 and Section 2.4 3 Sample holder tray movement Section 2.5 and Section 2.6 14 (31) BRUKER BIOSPIN Installation Manual Version 011

Operation 4 4 Automatic Operation 4.1 1. If the manual test is successful, then the NMR Case is ready for automatic operation. 2. Read and follow the manual for the automation software: How to set up Icon NMR P/N B0951. 3. The NMR Case sample tray does not need to be filled sequentially with samples. There can be one or several empty positions between samples. The NMR Case will jump over these empty positions. However it does not jump on it s own, the sample lift needs to be switched ON and OFF the usual way and the sample tray will then advance to the next position. This feature is useful when measuring several small batches of samples, adding them to the already loaded sample holder tray without removing the tray from the magnet. 4. The NMR Case will stop advancing when the sample lift is not activated anymore. 5. The NMR Case will remember if a sample is still inside the shim coil and has not been ejected yet. If this sample is being removed manually (by removing the sample tray, activating the sample lift manually and removing the sample by hand), then the sample holder tray will not advance anymore. To solve this situation, either insert another sample manually, or press the reset switch located at the right side of the motion controller. Switching off the power to the motion controller unit will NOT erase the sample tracking memory. 6. If a sample jams and does not get inserted all the way into the shim coil, then the sample changer does not advance the sample holder tray. Therefore the next sample lift sample insert sequence will be done with the same sample, until this sample it is successfully inserted. The software will not track this repetition; therefore some samples at the end of the queue will not be measured. 7. Built in safety features will prevent sample damage in case of power and air failures. Manual Operation 4.2 1. The NMR Case can be used for manual operation by simply removing the sample holder tray and inserting the sample through the adapter collar into the shim stack. It is not necessary to remove the NMR Case base assembly and it is not necessary to switch off the motion controller. 2. In addition, it is possible to fill the sample holder tray with just one sample and then operating the sample lift manually in the usual way, without the automation software. Installation Manual Version 011 BRUKER BIOSPIN 15 (31)

Operation 3. To use the NMR Case in automatic mode again, the sample tracking memory has to be cleared. While the sample lift is OFF, press the RESET button on the side of the motion controller. 4. Note: pressing the RESET button while the sample lift is ON does not work and the sample holder tray will not advance away from the position 0. Second Sample Holder Tray 4.3 A second (or more) sample holder tray can be purchased under part number B2436 and be used to load fresh samples while the first tray is in use on top of the magnet. A simple swap of the trays keeps the spectrometer running, while the first tray can be conveniently loaded with new samples again. 1. Grab the sample holder tray at the handle and lift it straight up and away from the NMR Case base. WARNING! Do not swing the sample holder tray. Watch out for the glass tubes close to the wrist. handle with care. 2. Place the sample holder tray on a secure table surface. 3. Remove all samples from the tray. 4. Locate the green START position. 5. While pushing the tray lock lever inside the tray to the position Unlock, rotate the sample stop head to the location of the START position. 6. Let go the tray lock lever. Check that it snaps back to the locked position. 7. Insert a sample tube mounted in a spinner into sample position 1. 8. Insert more samples in sequential order. Any number of samples (up to 24) can be used. 9. Grab and lift up the sample holder tray at the handle. 10. Line up and lower the sample holder tray over the NMR Case base cover. 11. Line up the sample stop head with the black off-center pin and engage it. 12. Line up the center with the black on-center pin and engage it. 13. Engage the tray into the side pin. 14. The sample holder tray settles down. Make sure that it seated correctly all the way down 15. Check that the tray lock lever is in the unlocked position. 16. If accidentally there is a sample left inside the shim coil, it will be inserted into the start position of the sample holder tray. 16 (31) BRUKER BIOSPIN Installation Manual Version 011

Operation Sample Diameters 4.4 The NMR Case can handle sample tubes with diameters from 2.5 mm up to 10 mm. WARNING! Do not use sample tubes with diameters larger than 10 mm, the glass tubes may break. The sample lift airflow needs to be adjusted depending on the weight of the sample and the spinner. Heavier sample / spinners need a larger sample lift airflow. See the trouble shooting guide Section 1.1 and Section 2.4 on how to adjust the sample lift airflow. For reliable long-term operation it is not recommended to mix lightweight and heavy sample /spinners in one sample holder tray. Sample Diameter Spinner Material Spinner Color Approved for NMR Case Max Temp Weight 5+2.5 mm POM Blue Yes 50 ºC lightweight 8 mm POM Blue Yes 50 ºC lightweight 10 mm POM Blue Yes 50 ºC Medium weight 15 mm PPO NO Do not use 5+2.5 mm PCTFE Gray Yes 80 ºC Medium weight 8 mm PCTFE Gray Yes 80 ºC Medium weight 10 mm PCTFE Gray Yes 80 ºC Medium weight 15 mm PCTFE Gray NO Do not use 5+2.5 mm Ceramic White Yes 80 ºC heavy 8 mm Ceramic White Yes 80 ºC heavy 10 mm Ceramic White Yes 80 ºC heavy 15 mm Ceramic White NO Do not use Temperature Range 4.5 The NMR Case sample changer can handle a temperature range of -20 C to + 50 C. Installation Manual Version 011 BRUKER BIOSPIN 17 (31)

Operation However it is acceptable to measure the NMR samples inside the probe with a temperature range of 50 C to + 80 C, because the sample lift air flow will warm up or cool down the sample and the spinner by the time they reach the NMR Case sample holder tray. But a larger temperature range is not possible even using the higher rated white ceramic spinners, because the material of the NMR Case will get distorted by the hot shim upper stack and the hot spinners (Ceramic spinners hold the temperature for a long time). The motion controller itself has a temperature range of 0 C to + 50 C. Troubleshooting 4.6 For all troubleshooting issues, read and follow the NMR Case Troubleshooting Guide, P/N B2927. 18 (31) BRUKER BIOSPIN Installation Manual Version 011

Illustrations A A Figure A.1. NMR CASE Base Installation Manual Version 011 BRUKER BIOSPIN 19 (31)

Figure A.2. NMR CASE Base Mounted 20 (31) BRUKER BIOSPIN Installation Manual Version 011

Figure A.3. Adapter Collar for the BST Installation Manual Version 011 BRUKER BIOSPIN 21 (31)

Figure A.4. Adapter Collar Installation 22 (31) BRUKER BIOSPIN Installation Manual Version 011

Figure A.5. Incorrect Installation of the BST Collar Installation Manual Version 011 BRUKER BIOSPIN 23 (31)

Figure A.6. Pneumatic Hose Connections 24 (31) BRUKER BIOSPIN Installation Manual Version 011

Figure A.7. Sample Holder Tray Installation Manual Version 011 BRUKER BIOSPIN 25 (31)

Figure A.8. Sample Holder Tray Line Up 26 (31) BRUKER BIOSPIN Installation Manual Version 011

Figure A.9. Sample Holder Tray Mounted Installation Manual Version 011 BRUKER BIOSPIN 27 (31)

28 (31) BRUKER BIOSPIN Installation Manual Version 011

Figures 1 Introduction 5 2 Parts List 7 3 Installation Instructions 9 4 Operation 15 A Illustrations 19 Figure A.1. NMR CASE Base... 19 Figure A.2. NMR CASE Base Mounted... 20 Figure A.3. Adapter Collar for the BST... 21 Figure A.4. Adapter Collar Installation... 22 Figure A.5. Incorrect Installation of the BST Collar... 23 Figure A.6. Pneumatic Hose Connections... 24 Figure A.7. Sample Holder Tray... 25 Figure A.8. Sample Holder Tray Line Up... 26 Figure A.9. Sample Holder Tray Mounted... 27 Installation Manual Version 011 BRUKER BIOSPIN 29 (31)

Figures 30 (31) BRUKER BIOSPIN Installation Manual Version 011

End of Document Installation Manual Version 011 BRUKER BIOSPIN 31 (31)

Bruker BioSpin your solution partner Bruker BioSpin provides a world class, market-leading range of analysis solutions for your life and materials science needs. Bruker BioSpin Group Bruker BioSpin B2926 info@bruker-biospin.com www.bruker-biospin.com