USER MANUAL LVC-200TC DIGITAL THERMOCOUPLE VACUUM GAUGE

Similar documents
Harris IRT Enterprises Digital Resistance Tester Model XP

Vacuum Gauge Catalog

Myers Vacuum RD # 2 Box 247A Kittanning, PA Phone: Fax: Instruction Manual For Thermistor Gauge GT-340A

Myers Vacuum 1155 Myers Lane Kittanning, PA Phone: Fax: Instruction Manual For Thermocouple Vacuum Gauge GTC-390

Terranova Model 924A Thermocouple Vacuum Gauge Controller

TELEDYNE HASTINGS INSTRUMENTS

FAIRCHILD T5420 DIGITAL-PNEUMATIC TRANSDUCER Installation, Operation and Maintenance Instructions

Instruction Manual For Thermistor Vacuum Gauge GT-430

THERMAL CONTROL SYSTEM OPERATION AND MAINTENANCE MANUAL

Mid-West Instrument. Series 700 "Wet/Wet" Installation and Operating Instructions. Differential Pressure Transmitter

Pirani Vacuum Gauges. Digital and Analog Vacuum Gauge Systems. Features and Benefits. HPS Products (800) (303)

Product Guide: Series III Pump Control Board Set (RoHS)

Instruction Manual. Instruction Manual Thermocouple Gauge Control Units Models TCG-531 & TCG-531/2 Models TCG-06M & TCG-06M/2

MODEL 905V OPERATING INSTRUCTIONS

Accurate to 0.1 NIST Traceable Advanced PID Algorighm USB Port Free datalogging/control software. Controllers I N S C O N TA G N

B BASIC HEAT TREATMENT CONSOLE

Model 809 Diaphragm Gauge Controller

Instruction Manual. rev092401sr. Copyright 2001 by Duniway Stockroom Corp.. 1 of 24

Physical Property Analyzer PPA4100 Series. On-Line RVP Analyzer

Model 2500 Horsepower Computer System User Manual

MM1/MM2 INSTALLATION & MAINTENANCE INSTRUCTIONS

AUTOMATIC 12 WAY HEAT TREATMENT CONSOLE

Dynojet Research, Inc. All Rights Reserved. Air Fuel Ratio Module Installation and User Guide.

Varian, Inc. Vacuum Technologies. Vacuum Measurement. Transducers. Gauge Controllers. Active Gauges. Calibration & Service.

OPERATION MANUAL. TZ Two Zone Controller

Series 275 Mini-Convectron Vacuum Gauge Module with Nonlinear Analog Outputs and Process Control Relays Instruction Manual

MPV1 INSTALLATION & MAINTENANCE INSTRUCTIONS

MPV1 INSTALLATION & MAINTENANCE INSTRUCTIONS

SAC SERIES CONTENTS TRIPLE-INTERVAL HIGH PRECISION COUNTING SCALE OPERATION MANUAL 1. INSTALLATION 2. SPECIFICATIONS

IOM IR-110-X. Solenoid Control Valve

LTX RF LEVEL SENSOR. Instruction Manual

MultiCal. Operation Manual for MultiCal mmhga/psia Pressure Module

APSX WIDEBAND D2 AFR CONTROLLER GAUGE. Installation and User Manual. APSX WIDEBAND D2 Manual V1.0

Accurate to 0.1 NIST Traceable Advanced PID Algorithm USB Port Free datalogging/control software

TELEDYNE HASTINGS INSTRUMENTS

Terranova Model 926A

PDR 9000 OPERATION AND MAINTENANCE MANUAL

PRESSURE BENCH OPERATING INSTRUCTIONS

AMI. American Magnetics, Inc. Excellence in Magnetics and Cryogenics

HIGH VACUUM GAUGE / CONTROLLER - analogue

KJL4500. Ionization Gauge Controller Instruction Manual Worthington Ave., Clairton, PA (412)

Observe all necessary safety precautions when controlling the soft starter remotely. Alert personnel that machinery may start without warning.

Rosemount 402 and 402VP

P200 P/I Transducer. Installation, Operation, and Maintenance Instructions INTRODUCTION

INDIAN INSTITUTE OF TECHNOLOGY KANPUR Kanpur , Uttar Pradesh, India Centre for Lasers and Photonics

BAYLAN AUTOMATED TEST BENCH

ECT Display Driver Installation for AP2 Module

DL 01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

CAUTION: CAREFULLY READ INSTRUCTIONS BEFORE PROCEEDING

EMaSM. Principles Of Sensors & transducers

TABLE OF CONTENTS. Page 2 of 12 DataStream 4 User Manual Rev. 2

T100 Vector Impedance Analyzer. timestechnology.com.hk. User Manual Ver. 1.1

SSI Technologies Application Note PS-AN7 MediaGauge (Model MG-MD) Digital Pressure Gauge Product Overview

1291BL Series Technical Specification Single Axis Position and Rate Table System

Type 2000 Transducer Product Instructions

INSTRUCTION MANUAL AUTO-PUMP 220/240 VAC MODEL #091-9B-220. INPUT :220/240 volt, 50/60 Hz, 1.9/2.2 AMPS OUTPUT: 100 PSI MAXIMUM 1 YEAR WARRANTY

IOM IR-910M0-X. Hydrometer Magnetic Drive with solenoid control

MICROPROCESSOR CONTROL WATER SUPPLY SYSTEM (MPR-01AW)

Fluke 750P Series Pressure Modules

LNW-5. Installation... 2 Lift Truck Conditions Indicator Installation Hose Installation

AMETEK Model PPC Series Electronic Pressure Calibrator

CAUTION: CAREFULLY READ INSTRUCTIONS BEFORE PROCEEDING

Instruction Manual For Hot Cathode Ionization Gauge Dual thermocouple GITC-425

4, 6, 8 & 11 Su p e r Ev a c Vacuum Pumps

Safe & Secure Series Electric Actuator with Internal Battery Back-up Installation, Operation & Maintenance Manual

GE Sensing & Inspection Technologies

1291BL Series Technical Specification Single-Axis Rate and Positioning Table System

Snapshot LX5 USER MANUAL. OK on Dimmer Outdoor OK Sound Activated DMX512 Master/Slave 115V/230V Switch Replaceable Fuse User Serviceable Duty Cycle

Industrial Pressure Indicator

CIRCLE SEAL CONTROLS

ODYR-25-1 & SLX-25-1 rev. A VACUUM/BOOST PRESSURE, TEMP, and EGT GAUGE

CLA-VAL e-drive-34. User Manual. Motorised Pilots. CLA-VAL Europe LIN072UE - 04/16

CIRCLE SEAL CONTROLS

SSP-7080 USER MANUAL. 80W Constant Power Switching Mode Power Supply with Master & Slave Remote Sensing. Rev /

QPV1 Electronic Pressure Regulator INSTALLATION AND MAINTENANCE INSTRUCTIONS

R & D SPECIALTIES ROTROL I USER'S MANUAL

IG4400 Ionization Gauge Controller Instruction Manual

Idle Timer Controller - A-ITC620-A Chevrolet Express/GMC Savana

HD and ED Series. Hysteresis Dynamometers. User s Manual

Product Manual MNX10010 / REV B MODEL PS03

Model 102B15. ICP Pressure Sensor. Installation and Operating Manual

Specifications, Applications, Service Instructions & Parts VPIF VALVE POSITION INDICATOR. for use with Sealed Motor Valves VPIF INTRODUCTION

HM860 FLOWMETER OPERATOR S MANUAL DO NOT USE OR OPERATE THIS EQUIPMENT UNTIL THIS MANUAL HAS BEEN READ AND THOROUGHLY UNDERSTOOD

MILWAUKEE ELECTRIC TOOL CORPORATION TEST & MEASUREMENT PRODUCT. REPAIR SERVICE INSTRUCTIONS CALIBRATION Milliamp Clamp Meter

Varian, Inc. Vacuum Technologies. TPS Turbo Pumping Systems. Features and Benefits 2-3. Mini-TASK AG TPS-compact 6-7

Fluke 750P Series Pressure Modules

Instruction Manual Model 4265/4265-PLUS Ultrasonic Cement Analyzer (Original Instructions)

Controlled Ferroresonant Ballast (CFB) Generation #3. Manual

Model 116 DPI Switch Installation and Operating Instructions

PRESSURE TRANSMITTER PYRP-28 INDUSTRIAL & INTRINSICALLY SAFE PRESSURE TRANSMITTER PRESSURE TRANSMITTER

CGD8800X CGD8800X220 Combustible Gas Detector

INSTALLATION DATA MANUAL MODEL 148/248/348

Digital Torquemeters

Nilfisk-CFM SS Vapor Vacuum Instructions for Use and Spare Parts Manual

HASTINGS INSTRUMENTS. AVG and AVC MODULE

Pressure & Vacuum Measurement. Series 356. Solutions MICRO-ION PLUS MODULES

FAIRCHILD T5220 SERIES ELECTRO-PNEUMATIC TRANSDUCER Installation, Operation and Maintenance Instructions

VACUUM PUMP SYSTEMS. 99 Washington Street Melrose, MA Phone Toll Free Visit us at

DigiTemp. Digital Thermometer

Warnings and Cautions. Precision Flow Insertion Mass Flow Meter. Instruction Manual BULLETIN F-41-PFS FR# R

Transcription:

USER MANUAL LVC-200TC DIGITAL THERMOCOUPLE VACUUM GAUGE

CONTACT US PHONE/FAX Toll Free: 800.465.1004 Phone: 801.486.1004 Fax: 801.486.1007 ADDRESS LACO Technologies, Inc. 3085 West Directors Row Salt Lake City, UT 84104 WEB www.lacotech.com info@lacotech.com SMT-07-1007 Rev A2 ii

CONTENTS 1. SCOPE... 1 2. GAUGE ACCURACY... 1 3. TUBE INSTALLATION... 2 4. CALIBRATION... 2 5. RS232 OUTPUT... 4 6. OPTIONAL SOFTWARE... 5 7. ACCESSORIES AND SPARES... 5 8. TROUBLESHOOTING... 6 iii

1. SCOPE This manual contains operation, maintenance, and troubleshooting information for the LVG-200TC Vacuum Gauge. LACO vacuum gauges are designed to ensure safety when used properly. It is the responsibility of the user to follow safety-related warnings, cautions, notes, and other requirements described in this manual. Returned equipment will not be accepted without prior authorization. Prior to shipping please call for a returned material authorization (RMA) number. LACO Technologies reserves the right to cancel the warranty if the gauge is disassembled without authorization. 2. GAUGE ACCURACY The LVG-200TC uses a thermocouple gauge tube made for the range of 1 Torr (1000 microns) to 1 millitorr (1 micron). The gauge electronics can also provide general reference readings above 1 Torr to 760 Torr (1 atm). NOTE The general reference readings from 1 Torr to atmosphere are less accurate than those below 1 Torr. Readings Below 1 Torr (1000 microns) 0 to 9 microns +/- 1 micron 10 to 99 microns +/- 5% of reading +/- 1 micron 100 to 999 microns +/- 10% of reading +/- 5 micron 1000 to 1999 microns +/- 15% of reading +/- 50 micron Readings Above 1 Torr to 760 Torr 1.0 to 9.9 Torr +/- 15% of reading +/-.1 Torr 10 to 99 Torr +/- 20% of reading +/-.2 Torr 100 to 760 Torr +/- 25% of reading +/- 10 Torr NOTE Wide-range thermocouple and Pirani gauges rely on convection (a form of thermal conduction) to achieve their wide range with a thermal sensor. As such, they are sensitive to the gas species. For filling applications, transducer capacitance diaphragm type gauges are recommended. 1

3. TUBE INSTALLATION The tube must be mounted in a stem down position within 10 degrees of vertical. The connection is one-eighth inch male pipe thread. Use Teflon thread tape for installation. Be careful not to twist the cable. NOTE The tube should be installed in a clean part of the vacuum system. It is recommended that the tube is not mounted directly on vacuum pump suction as the tube life may be shortened by backstreaming of pump oil vapors. Inserting a loop of tubing or an optical baffle can be helpful in protecting the tube from backstreaming. After energization, the unit will initialize and operate. The gauge reaches best thermocouple accuracy after a few minutes of warm-up. The gauge has a switching power supply that can be connected to either 115 or 220 VAC, 50/60 Hz. They are powered by 115 to 230 Volts, AC or 5 volts DC. At atmosphere, the unit will read 760 Torr. A visible decimal point indicates that the unit is reading in torr. As pressures are decreased, the readings will shift to 99.0, 9.99,.999, and finally to.001 Torr. A reading of.001 Torr is equal to one micron which is equal to one millitorr. The reading of the gauge will be influenced by the composition of the gasses remaining. All wide range and Pirani vacuum gauges are subject to this phenomenon. For applications in which excellent accuracy is required in the low torr ranges, such as the filling of neon and laser tubes, we offer a transducer based gauge which gives precise readings regardless of gas composition. Transducer gauges are not able to read in the micron ranges. 4. CALIBRATION If the tube is replaced, it is recommended that the unit be recalibrated. It is recommended that this be done at LACO Technologies but field calibration is available. If field calibration is done, the zero and span adjustments are available on two potentiometers which are accessible by removing the front panel of the instrument. The zero is on the left and the span is on the right. Make slight adjustments and wait a few seconds for the effects of the adjustments to be seen. 4.1 ATMOSPHERE 1. Energize the unit at atmospheric pressure. Remove the front panel. 2. Make sure that the gauge tube is in a vertical orientation, as it is position sensitive. 2

3. Connect a Voltmeter to the two pads available from the right side of the printed circuit board. 4. These are called Zero and Atm.420 5. Adjust the potentiometer labeled Atm (leftmost) until the voltage above equals.420 Figure 1 Atmos. Adg. Zero Adg. 4.2 ZERO 1. Operate the unit at the lowest attainable pressure with at least a two stage pump capable of less than 10 microns. Ideally, use a turbo or diffusion pump to calibrate. 2. Connect Voltmeter leads to Com and Vac pins. The voltage should be approximately.16 Volts dc. 3. Make slight adjustments, if necessary, so the low end vacuum reading is correct. If the above readings cannot be attained, the instrument, the gauge tube, or the cable is defective. Gauge tubes frequently become contaminated with oil. Cleaning the gauge tube with a strong solvent, such as a carburetor cleaner, can often improve or restore its calibration. For an Accredited NIST traceable calibration with data, (ref: P/N LSGC0003), call LACO for pricing. 3

Figure 2: LVG-200TC with red lens removed For repair or recalibration, complete an RMA Form (located at www. lacotech.com) and return gauge to LACO. 5. RS232 OUTPUT The LVG-200TC is supplied with an RS-232 output for external data logging. The output can be seen on any computer with a serial port, for example, a PC running Windows Terminal or Hyperterm will show the output. 5.1 PARAMETERS AND DETAILS FOR RS232 The current value on the screen is sent out the RS232 port approximately once per second. The parameters are: 9600 baud, 8 data bits, 1 stop bit, no parity. No prompting or handshaking by the user is required. The data is presented as ASCII text for the numbers (with decimal points as required), followed by a carriage return and line feed. This data may be observed using Windows Terminal, Hyperterm, or any similar terminal program. Data may be saved using Terminal s Transfers / Receive Text File feature. The saved data files may be read with Microsoft Excel or any text editor. Microsoft Excel, or any spreadsheet program can also plot the data. The RS232 connections are made on the rear female 9 pin dsubminiature connector, and the data appears on pin 2, and the ground is on pin 5. 4

6. OPTIONAL SOFTWARE For more convenient logging and plotting, we offer plotting software. The software program gives a visual presentation of the data which is coming out of the vacuum gauge. Update rate is approximately once per second. Logarithmic axes clearly show vacuum from atmosphere to 10-3 or 10-9 (ultra wide range version). 6.1 OPTIONAL RECORDER We offer a special vacuum recorder which displays from atmosphere to hard vacuum. Both the plot and the numbers are displayed. This recorder is often used by rebuilders to document the performance of vacuum pumps, but there are many other uses. Contact LACO Technologies for further information. 6.2 CUSTOM FEATURES AND PROGRAMMING The LVG-200TC gauges are microprocessor controlled, making it possible to provide custom control features at reasonable prices. The software is burned into the microprocessors, which are socketed and easily replaced by a user. Contact LACO to discuss your needs. If the gauge is to be used in a neon sign processing facility, it must be installed with a stopcock or a solenoid valve to protect it from exposure to the system and possible damage during high voltage bombarding. Battery operated gauges are also offered and are particularly suited to neon applications. 7. ACCESSORIES AND SPARES Table 1: Related Accessories and Spares PART NUMBER VA0531 LVFAN25FP12S LVFAN40FP12S LVFAN16FP12S DESCRIPTION Spare gauge tube, 1/8 NPT Adaptor, NW25 1/8 FNPT, SS Adaptor, NW40 1/8 FNPT, SS Adaptor, NW16 1/8 FNPT, SS 5

8. TROUBLE- SHOOTING 1. PROBLEM: The gauge reads 760 Torr and does not respond to vacuum, it is possible the thermocouple voltage is not reaching the gauge. Cause 1: Bad thermocouple Remedy: Replace gauge tube Cause 2: Bad cable connection Remedy: Check red, black, green and white wires on 9 pin dsub connector and check soldered connections on octal connector 2. PROBLEM: The gauge reads 0 Cause 1: Bad thermocouple Remedy: Replace gauge tube Cause 2: Bad cable connections Remedy: Check red, black, green and white wires on 9 pin dsub connector and check soldered connections on octal connector 6