Octave Water Meter Installation Manual

Similar documents
ENGLISH. A Member of the

A Member of the. Integrated Metering Technologies

ENGLISH. A Member of the

WATER METERS OCTAVE ULTRASONIC WATER METER INSTALLATION & USER GUIDE

A Member of the. Integrated Metering Technologies

Electromagnetic flowmeters Series FLOMID Electromagnetic flowmeter for conductive liquids

F-4600 INLINE ULTRASONIC FLOW METER Installation and Operation Guide

Electromagnetic flowmeters Series FLOMAT Insertion electromagnetic flowmeter for conductive liquids

Paddle-wheel flow controller for On/Off control

Paddle-wheel flow controller for On/Off control

Liquid Turbine Flow Meter OwnerÊs Manual

F-2600 SERIES INLINE VORTEX FLOW METER

The full WeatherTRAK Flow3 Installation and Maintenance Manual is available online at

ULTRASONIC FLOW SENSOR IFX-M4-01

F-2700 SERIES INSERTION VORTEX FLOW METER

Appendix. Clearance. Skewed profiles. Application Schematics

imag 4700 SERIES FLANGED MAGMETER

Multivariable Vortex Flow Meter

WATERFLUX 3070 Quick Start

ISTEC MAGNETIC METERS

SIGNET 7002 Vortex Flow Sensor ENGLISH

INLINE Flow sensor for hazardous area II 1 G/D - II 2 D - II 3 GD - I M1

Ultra Mag SUBMITTAL PACKAGE

Range of Ultrasonic Flowmeters.

Multivariable Vortex Flow Meter

F-1500 SERIES INSERTION TURBINE FLOW METER

Model TUR-200D Turbine Flowmeter

Turbine flowmeters Series TM

INSERTION flowmeter with paddle wheel, ELEMENT design

Hydro-Flow Model 2200 Installation and Operation Guide B

mass flowmeter for compressed air

WMX-series MUNICIPAL/INDUSTRIAL MAGMETER INSTRUCTIONS

Options guide. Electromagnetic flowmeters. M9xx

MINI-NBK ECONOMICAL LEVEL INDICATOR

F-2600 SERIES INLINE VORTEX FLOW METER

Armstrong In-Line and Insertion Vortex Flow Meters

VORTEX FLOW METTER DVH. Despriction. Advantages of the DVH Vortex Flow Meter

RCM - Direct Reading Flowmeters

ManuFlo Page 1 MANU ELECTRONICS PTY LTD Flow Measurement Products Rev PK KMS501W & KMS502F- TW. Electromagnetic TRADEWASTE Flowmeters FEATURES:

Axial Turbine Flow Sensor Series Turbotron VTH 25 / VTI 25 / VTM 25

AG2000 IRRIGATION MAGMETER INSTRUCTIONS

Magnetic-inductive flow meter EPS / UMF2. Technical data sheet

VLM20 In-line Vortex Flowmeter

Electromagnetic flowmeters

VALVCHEQ BACKFLOW PREVENTERS FIGURE DC03

Hydro-Flow Model 2200 Fixed Insertion Vortex Flowmeter Installation and Operation Manual

Temperature sensor TE2

Signet Pressure Transmitter

Glass tube flowmeters Series 6000 Variable area flowmeter for liquids and gases

SIGNET Temperature Transmitter

Conductivity meter. PLC Solenoid valve. Type Type On/Off Diaphragm valve. Valve islands

EX80 Series EX81 EX82. E X 8 0 S E R I E S E l e c t r o m a g n e t i C F l o w s e n s o r i n s t r u c t i o n s. The Leader in Flow Meter Value.

Digital ph transmitter

SHARKY 775 COMPACT ENERGY METER ULTRASONIC

INSERTION flow sensor for continuous flow measurement

Mounting and Operating Instructions EB 8135/8136 EN. Series V2001 Valves Type 3535 Three-way Valve for Heat Transfer Oil

Technical Information Proline Promag 10L

ULTRASONIC FLOWMETER / HEATMETER

NO ENTRY No entry for Models 605, 610 or 620. Enter maximum flow rate for Model 600 with external flow sensor

Oscillating piston flowmeters Series COVOL Positive displacement flowmeter for liquids

Operating Manual Model- P282 Digiflow Controller

SERIES FLMG MUNICIPAL/INDUSTRIAL MAGMETER INSTRUCTIONS

FLS F6.60 MAGMETER FLOW SENSOR INSERTION FLOW SENSORS APPLICATIONS MAIN FEATURES

ARCHIVED. EX80 Series. (Includes Dates 6/28/05 to 10/06) EX81 EX82

UFM 800 W, C and UFM 800 W Hot Tapping. Ultrasonic flowmeters. for water and wastewater

EGM1100C MAGMAX. Compact Electromagnetic Flowmeter GENERAL FEATURES STANDARD SPECIFICATION. General Specification

Flow meter Type Recordall (RCDL)

Paddle-wheel flow controller with optical principle for On/Off control

List of Contents. Agrimag User Guide 1 V V

Signet Temperature Transmitter

display on the wall or in a panel away from the inline sensor SE30 for easy meter reading.

To avoid injury, death of personnel ordamage to instrument, the operator must refer to an explanation in the User smanual.

Signet 7000/7001 Vortex Flow Sensors * *

Electronic Pressure Switches

DN15 DN250 PN 16. Corrosion Protection

ULTRASONIC FLOW SENSOR QALCOSONIC FLOW 2

Hydro-Flow Model 3100 Retractable Insertion Vortex Flowmeter Installation and Operation Manual

Observe all necessary safety precautions when controlling the soft starter remotely. Alert personnel that machinery may start without warning.

Electronic Pressure Switches

Signet Pressure Transmitter

Rosemount 8750WA Magnetic Flowmeter System For Water and Wastewater Industries

Pressure transmitters for heavy duty applications MBS 3200 and 3250

Electromagnetic Flow Meter Owner s Manual HARDWARE

Type Operating Instructions. Bedienungsanleitung Manuel utilisateur

Figure DC03. Double check valves for medium hazard rated applications BSP screwed connections. Features

Hydro-Flow Model 2300 Installation and Operation Guide B

MEASURE TODAY. ENSURE TOMORROW. SpireMag Series T-MAG Electromagnetic BTU Meter

FPI Mag FULL PROFILE INSERTION ELECTROMAGNETIC FLOW METER MODELS 394L AND 395L MAG METER SUBMITTAL

Digital ph transmitter

Isoil Industria a worldwide supplier of electromagnetic flowmeters Total flow control. Accurate to the drop.

FlowCon Energy FIT System

Positive displacement flowmeter for continuous flow measurement

Compact Heat Meters. Features. -A Pulsed output. -B M-Bus output. Accessories. UK Sales Tel: International Tel:

Ultrasonic heat meter

WATERFLUX 3070 Technical Datasheet

VIM20 Vortex Insertion Flowmeter

CONTROLLERS & SPECIAL VALVES

Vortex Flow Meter Model DVH

F-2700 SERIES INSERTION VORTEX FLOW METER

PanaFlow MV80. In-line Multivariable Flowmeter for Mass, Temperature, and Pressure. GE Sensing. Applications. Key Benefits

Transcription:

Octave Water Meter Installation Manual

1.0 Notes For proper flow measurements, the OCTAVE s measuring tube should be completely full at all times. Non-wetted sensors show loss of signal. Though this will not cause damage to the meter, it will, however, not measure flow and display zero. Flow direction: The OCTAVE is a bidirectional flow meter. Note the indicating arrow on the OCTAVE s display for forward and backward flows. In case of direct sunlight exposure, it is recommended to keep the lid closed, though no direct damage will occur with the lid open. Do not expose the OCTAVE to excessive vibration. To avoid vibration, support the pipeline on both side of the meter Ambient working temperature: -25 to +55 C. Water working temperature: 0.1 to +50 C. To avoid measuring errors due to air in the flow tube, observe the following precautions: Since air collects at the highest point of the system, installation of the flowmeter should be at the lowest point. Always install control valves downstream of the meter in order to avoid cavitation. Never install the meter on a pump suction side in order to avoid cavitation. 2.0 Start-up Check that the meter has been installed correctly. Check that the flow rate and volume units are correctly preprogrammed on the display. Check that the output module is correctly attached. 3.0 Digital Display Flow direction System error Flow rate units Water temperature Accumulator mode Output mode Communication mode Volume units Low battery alert Pulse resolution 1

4.0 Mechanical Data Maximum Working Pressure 16 bar Liquid Temperature 0.1 up to 50 º C Precision Class ISO 4064 rev.2005, Accuracy class 2 Configuration Power Source Environmental Protection Volume Display Options Data Logger Connections Severity levels Pressure Loss Compact - The display is built in to the unit 2 D size Li-battery: up to 15 years life time IP 68, Ambient operation temp. -25 C up to +55 C 1. Net (Forward less reverse) 2. Forward only 3. Reverse only 4. Forward & reverse alternating Volumes and alarms data (48KB) 1½-2 threaded: with couplings to NPT/ BSP 2-12 flanged: flanges according to ISO, BS 10 and ANSI 150 Mechanical class M1 Electromagnetic environment class E1 ΔP 0.16 bar Outputs Analog Output Digital (pulse) Output Dry Contact Output Encoder Output Modbus Protocol Output Output Extension The Analog Output shows the currently measured flow rate. The Analog Output is a 4-20 ma current loop (the end user must supply power to the unit). The Analog Output is programmable for forward and reverse flow (see Operation Manual for more details). The 20mA point is programmable per customer request (To any flow lower than the max flow of the meter). The Digital (pulse) Output is an open drain transistor output that provides pulse per quantity with these options: 1. Two scaled forward, reverse or net mode pulses 2. One scaled forward, reverse or net mode pulse and one alarm frequency output 3. Measuring units of the output can be programmed different than displayed units Pulse resolution will be shown on the display for each pulse separately The Dry Contact Output is a dual relay output that provides pulse per quantity with these options: 1. Two scaled forward, reverse or net mode pulses 2. Measuring units of the output can be programmed different than displayed units Pulse resolution will be shown on the display for each pulse separately Onsite power supply of 12-24 VDC is needed. The Encoder Output is a serial communication collector utilizing UI1203 or UI1204 communication protocol. (Sensus protocol) Pulse output included. The Modbus Protocol Output has the following available functions: 1. Alarms (battery, empty pipe) 2. AMR serial number 3. Real Time Clock (RTC) 4. Volume units 5. Flow rate units 6. Current flow 7. Flow direction 8. Forward and reverse volumes 9. Flow and volume resolution 5m extension cable for installation in pits and vaults Dimensions Model Octave Nominal size (mm) 40 Threaded (inch) 1½ Threaded 50 Threaded 50 65 80 100 150 200 250 300 2 Threaded 2 2.5 3 4 6 8 10 12 L Length without couplings (mm) 300 300 200 200 225 250 300 350 449 499 W Width (mm) 113 113 165 185 200 220 285 340 406 489 H Height (mm) 155 155 194 210 210 223 282 332 383 456 h Height (mm) 35 35 40 90 90 103 140 165 203 245 Weight (kg) - cast iron body 8 9 11.5 13 15 32 45 68 96 Weight (kg) - polymer body 1.4 1.45 H h W L W 2

5.0 Mechanical Installation 5.1 Handling the Flow Meter Important handling information: Do not lift the Octave by the electronic housing. Do not carry the Octave by its lid. Do not place the Octave on the electronic housing. When handling the Octave, avoid hard blows, jolts or impacts. 5.2 Installation: Location and Position Recommended installation Conditional installation Wrong installation Recommended: If this is not the highest point in the system or a hydraulic jump has been installed to keep the flow meter full. Not recommended: If this is the highest point in a system or if pipeline and/or flow meter is subject to being emptied between uses avoid this installation. The Following Examples are Arad s Recommendations For Achieving Top Performance When installing the Octave downstream of any hydraulic component (valve, pump) the recommended installation requirements are no less than the drawings recommendations. For upstream & downstream straight pipes please use as much as installation site will allow ( the longer the better) When installing Pressure Breaker after the meter - the pipe length should be at least 2 pipe diameter (the longer the better). 2 DN 2 DN 2 DN 2 DN 2 DN 10 DN 2 pipe diameters before and after elbows (90 ) Minimum of ten (10) pipe diameters after pumps Minimum of two (2) pipe diameters before and after 2 DN 2 DN 2 DN 2 DN Minimum of two (2) pipe diameters before T connections Minimum of two (2) pipe diameters after strainers 2 pipe diameters before and after elbows (90 ) in vertical installations 3

6.0 Polymeric Octave Installation - General Instructions Please follow the general instructions for water meters (Check Valve, upstream & down stream and system flushing on new installations). Existing and new installations: 1. It is recommended that the meter will NOT be installed in the middle of the system, so the meter will not suffer from the load of all the installed fixtures. 2. It is recommended that at least one side of meter will be connected to a PVC (or plastic) pipe. 3. Please make sure that the end connections are parallel and inline to each other. 4. When using standard tail piece, please use rubber gaskets only (not fiberglass)! After adding the gaskets the gap between the end connection and the meter should not exceed 1mm. Please unscrew the end connection in order to keep the recommended distance. Please do not use force in order to close the gap. Please don t do any welding while the meter is connected to the pipe. Rubber gasket 1mm max 4

5. If it is not possible to use plastic connections on one side of the meter, please consider the use of our flexible couplings. These couplings were especially developed for the polymeric Octave meters. You will need to use this coupling only on one side of the meter - please install the coupling on the outlet, if possible - as described in the below pictures. 5mm max 3mm max 6. Please start to tighten the end connections by hand. Final tightening will be with no more than 100Nm torque. 7. Please don t do any welding while the meter is connected to the pipe. 8. In case it is not possible to use plastic end connections (at least on one side), please support the meter as shown in the drawing below. Support Plastic End Connection 5

7.0 Electrical Outputs The OCTAVE has several outputs options. 1. Dual pulse output for volumes 2. Dry Contact output for volume 3. Analog output (4-20mA) for flow 7.1.1 Pulse V5.2 Output Module Specification Pulse types: 1. Open Drain (N-channel FET transistor output) that allows current loading of 200mA, and up to 35volt 2. Dry contact 3. SSR Outputs Diagram OCTAVE Pulse Long RED 1 Pulse#1 EMI /RFI / OverVoltage PROTECTION GREEN BLACK 1 1 Pulse#2 GND Read-out instrument SHEILD 1 Ground Ground 1 Short s Warning - Signal connection polarity is mandatory! Wire Function Long Red Pulse Out#1 Green Pulse Out#2 Black GND Open Ground / Sheild Short Ring Terminal Ground Output Characteristics Outputs Type Length - supplied Maximum Length* Maximum Applied Voltage Open Drain 1.5/5 [meter] 500 [meter] 35 [Vdc] * The maximum length depends on: type, controller and electrical noise level. Teldoor PN 8005003101 or similar 6

7.1.2 Dry Contact V1.1 Module Specification Outputs Diagram Long Octave Dry Contact v1.1 Out#1 Out#2 Read-out instrument 5-35 Vdc Short Earth s Wire Function Long Red + Orange Out#1 Black + Brown Out#2 Short Red 5-35V + Black 5-35V - Yellow Earth * Signal connection polarity is mandatory! Output Characteristics Outputs Type Length - supplied Maximum Length* Maximum Applied Voltage Switching Power max. Life Expectancy Dry Contact 1.5/5 [meter] 500 [meter] 35 [Vdc] 15 [Watt] 10^9 [Cycles] * cable Teldoor PN 8005003101 or similar 7

7.1.3 SSR v1.0 Module Specification Outputs Diagram Long Octave SSR v1.0 Out#1 Out#2 Read-out instrument 5-35 Vdc Short Earth s Wire Function Long Red + Orange Out#1 Black + Brown Out#2 Short Red 5-35V + Black 5-35V - Yellow Earth * Signal connection polarity is mandatory! Output Characteristics Outputs Type On-Resistance max. Output current max. Output Voltage max. Supply Voltage 3-35 Vdc Length - supplied Maximum Length* Dry Contact 25 [Ω] 120 [ma] ±400 [V] 5-35 [Vdc] 1.5 [meter] 500 [meter] * cable Teldoor PN 8005003101 or similar 8

7.1.4 4-20mA Module (Analog) The current output is a passive 4-20mA. 4mA is always 0 (zero) flow and the 20mA is factory programmable according to the customer requirements.. (If the customer did not specify, the 20mA will be the maximum flow rate). Outputs Diagram Long (+) Octave 4/20mA module Read-out + - instrument (-) Shield Ground Ground Short s Signal connection polarity is mandatory! Wire Function Long Red current loop + Black current loop - Open GROUND / SHIELD Short Ring Terminal GROUND Output Characteristics Outputs Type Supplied Length Maximum Length * Loop supply voltage Output Impedance 4-20mA current output 1.5/5 [meter] 500 [meter] 12-24 [Vdc] 25 [M] typ * cable Teldoor PN 8005003101 or similar 9

8.0 Module Replacement / Mounting Manual 1. Properly dry the area of the connector. 2. Remove seal cover from the screw using tool with sharp edge. 3. Remove the screws using Allen key 3mm. 4. Remove the module/cover. 5. Properly dry again the area of the connector. 6. Make sure the module o-ring is in position. If not, Insert new o-ring into the module. The o-ring must be lubricated (silicone grease) 7. Attach module to the connector. 8. Tight both screws by hand only, using Allen key 3 mm to achieve symmetric o-ring pressure, then apply the torque 2 N*m with torque wrench. 9. Insert seal cover onto the screw. 10

9.0 GND Mounting Manual 1. Insert M5 screw to ring terminal lug 6. Slide the fork terminal lug between the flat washer and serrated washer 2. Insert flat washer on top of ring terminal lug 7. Add serrated washer on the other side of the bolt (On meter s flange side) 3. Insert serrated washer on flat washer 8. Add flat washer on the serrated washer 4. Attach to the fork terminal lug and tight 9. Add nut and tight. Make sure fork terminal lug is in position. 5. Insert flat washer and serrated washer on the bolt. Use the correct washers according to bolt size (M16 or M20 respectively) Insert the bolt to the hole on the pipe s flange. (Do not install on meter s flange). 11

9.1 Special Parts Set for ring terminal lug. Fork terminal lug Serrated washer Flat washer M5 screw Set for connecting fork terminal lug to pipe. M16 or M20 bolt. Not included in kit Flat washer M16&M20 (2 units each size) Included in kit Serrated washer M16&M20 (2 units each size). Included in kit M16 or M20 nut Not included in kit 12