TRACEABILITY OF THE CALIBRATION OF TEST CAR FOR ROLL BRAKE TESTER Aimo Pusa Raute-MIKES Finland, by Raute Precision Oy, Force and Mass Laboratory

Similar documents
A PORTABLE SYSTEM FOR THE CALIBRATIONOF TRANSDUCERS AND TORQUE WRENCHES: THE CALIBRATION BELL

PERFORMANCE OF THE NEW PRIMARY TORQUE STANDARD MACHINE OF INMETRO, BRAZIL

Hybrid torque standard machine for 1 kn m developed in CENAM

The analysis of the accuracy of the wheel alignment inspection method on the side-slip plate stand

LEVER OPTIMIZATION FOR TORQUE STANDARD MACHINES

Vehicle Dynamic Simulation Using A Non-Linear Finite Element Simulation Program (LS-DYNA)

USER MANUAL FOR AREX DIGI+ SYSTEMS

Contents. Page. 1. Product description. 2. The AXC line of linear axes. 3. AXLT line of linear tables. AXC and AXS product overview...

Development of Dynamic Calibration Machine for Pressure Transducers

Shock tube based dynamic calibration of pressure sensors

FEASIBILITY STYDY OF CHAIN DRIVE IN WATER HYDRAULIC ROTARY JOINT

Skid against Curb simulation using Abaqus/Explicit

ANALYTICAL STUDY ON THE UNCERTAINTY OF LOAD CELLS CALIBRATED WITH DEADWEIGHT-FORCE-STANDARD MACHINE AND FORCE-COMPARATOR MACHINE

A dream? Dr. Jürgen Bredenbeck Tire Technology Expo, February 2012 Cologne

METHOD FOR TESTING STEERABILITY AND STABILITY OF MILITARY VEHICLES MOTION USING SR60E STEERING ROBOT

RETROFIT OF A ROLLER BRAKE TESTER AT FAMENA

RENOVATION OF NMIJ'S 5 MN HYDRAULIC AMPLIFICATION TYPE FORCE STANDARD MACHINE

Certificate of Calibration

Calibration Laboratory Assessment Service CLAS Certificate Number Page 1 of 5

Wheeled Mobile Robots

IMGC-CNR Activities to Improve the Force Chain in Italy. C. Ferrero and C. Marinari Istituto di Metrologia G. Colonnetti (IMGC-CNR) Torino - ITALY

NEW INNOVATION. Shock Absorber Tester. Model: MAHA-Shock-Diagnostic MSD 3000

Fatigue properties of railway axles: new results of full-scale specimens

SPECIFICATION SHEET Pulse tool / Impact Wrench Testing System MODEL : IWT-101

a) Calculate the overall aerodynamic coefficient for the same temperature at altitude of 1000 m.

Cornering & Traction Test Rig MTS Flat-Trac IV CT plus

INITIAL DISTRIBUTION. Chief, Bureau of Ships, Project Records (324) Commander, U.S. Naval Ordnance Laboratory, White Oak, Silver Spring 19, Md.

CIRCOGRAPH Sensor systems Rotating Head Ro 35 P Rotating Head Ro 35 L

What is model validation? Overview about DynoTRAIN WP5. O. Polach Final Meeting Frankfurt am Main, September 27, 2013

Air Bearing Shaker for Precision Calibration of Accelerometers

AVL SHOT TO SHOT PLU 131 UREA FLOW MEASURING SYSTEM

Industry 4.0 Is Unthinkable Without Calibration!

THE DESIGN OF 1 N M TORQUE STANDARD MACHINE AT NIM

Study on System Dynamics of Long and Heavy-Haul Train

pressure transducer Miniature pressure transducer model 81530

Design of A New Non-Contact Screw Seal and Determination of Performance Characteristics

2012 IECEE CTL PTP Workshop. Ingrid Flemming IFM Quality Services Pty Ltd

Introduction to Torque Calibration

DATUM ELECTRONICS RS425 NON-CONTACT TORQUE TRANSDUCER PRODUCT OVERVIEW

Accessories for the measurement of tightening torque. Digital and electronics dynamometric wrenches

Innovative Testing Equipment. Series DF torque sensors

Analysis and control of vehicle steering wheel angular vibrations

SPMM OUTLINE SPECIFICATION - SP20016 issue 2 WHAT IS THE SPMM 5000?

Simulation of a Narrow Gauge Vehicle using SIMPACK, Model Validation using Scaled Prototypes on Roller-Rig

PNEUMATIC HIGH SPEED SPINDLE WITH AIR BEARINGS

Defining Parameters for Ballistic High Pressure Sensors. Transducer Workshop June 20-21, 2000 Aberdeen, MD

Enhanced Heat Transfer Surface Development for Exterior Tube Surfaces

PARALLEL INDEX DRIVES TP Series

EMS ELONGATION MEASUREMENT SYSTEM. Strain measurement system for wind turbines optimizing the control & condition monitoring

6-2. Traceability. (1) Traceability system. Chapter 6-2

AVL FUEL BALANCE & FUEL TEMPERATURE CONTROL

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF TORQUE METROLOGY (Torque Generating Device Calibration)

WFT-C x. precise robust quick setup. Flexible 6-component wheel force transducer for road testing and test stands. productive testing

NUMERICAL ANALYSIS OF IMPACT BETWEEN SHUNTING LOCOMOTIVE AND SELECTED ROAD VEHICLE

Booming Noise Optimization on an All Wheel Drive Vehicle

Metal forming machines: a new market for laser interferometers O. Beltrami STANIMUC Ente Federate UNI, via A. Vespucci 8, Tbrmo,

NUMERICAL ANALYSIS OF LOAD DISTRIBUTION IN RAILWAY TRACK UNDER WHEELSET

EMC-HD. C 01_2 Subheadline_15pt/7.2mm

mga research corporation

Headlight Test and Rating Protocol (Version I)

New Capabilities on Hybrid & Electric Drives

IMPACT REGISTER, INC. PRECISION BUILT RECORDERS SINCE 1914

The Mechanical Equivalent of Heat

GENERIC EPS MODEL Generic Modeling and Control of an Electromechanical Power Steering System for Virtual Prototypes

MODELING SUSPENSION DAMPER MODULES USING LS-DYNA

Small hydrocarbon flow calibration facilities at NMIJ

ALWAYS ON THE SAFE SIDE

Handtorque Multipliers What is a Torque Multiplier? How Handtorque Torque Multipliers Work Why use a Handtorque Torque Multiplier?

AVL PLU 131U FLOW METER

DaimlerChrysler Alternative Particulate Measurement page 1/8

NCAT Report EFFECT OF FRICTION AGGREGATE ON HOT MIX ASPHALT SURFACE FRICTION. By Pamela Turner Michael Heitzman

New Concept for Higher Speed on Existing Catenary System: Auxiliary Pantograph Operation

Vertical Tensile Machine

Electromechanical Steering with Parallel-axis Drive

The INDOT Friction Testing Program: Calibration, Testing, Data Management, and Application

MEASUREMENT TECHNOLOGY FOR A CALIBRATING VEHICLE FOR MULTIPLE SENSOR WEIGH-IN-MOTION SYSTEMS

Config file is loaded in controller; parameters are shown in tuning tab of SMAC control center

ROUGHSCAN SURFACE ROUGHNESS TESTER OPERATIONAL MANUAL

Hydraulic Deadweight Tester Model CPB5800

Legal Metrology Control of a Liquefied Petroleum Gas Transfer Standard and Field Practice

Cost-Benefit Analysis of Options for Certification, Validation and Monitoring and Reporting of HDVs

Config file is loaded in controller; parameters are shown in tuning tab of SMAC control center

Global Flow Measurement Calibration Consistency

Technical Guide No. 7. Dimensioning of a Drive system

Camshaft Torque Analysis of Diesel Engine

New generation vehicle test lanes

Standard Calibration Certificate

STIFFNESS CHARACTERISTICS OF MAIN BEARINGS FOUNDATION OF MARINE ENGINE

Article: Sulfur Testing VPS Quality Approach By Dr Sunil Kumar Laboratory Manager Fujairah, UAE

Temperature Field in Torque Converter Clutch

Is Low Friction Efficient?

Injection Moulding Kit - Section: Situation / Brief

Axial-radial cylindrical roller bearings

Hydrostatic drives HYDROTRAC GFT for mobile applications

III B.Tech I Semester Supplementary Examinations, May/June

of ROAD Abstract Keywords: acceleration Jussi Seppä control of a terminal box wirelessly via Bluetooth to shown in real surveyors.

Round robin tests in the Netherlands

Test bed for hydrostatic bearing gap

FUNCTIONS AND TECHNOLOGY OF DIGITAL TACHOGRAPHS SYSTEMS WITH CONTEMPORARY TELEMATIC SYSTEM IN ROAD TRANSPORT, ON BOARD INFORMATICS NET AND GPS SYSTEM

BM14200 Roller Brake Tester

Transcription:

TRACEABILITY OF THE CALIBRATION OF TEST CAR FOR ROLL BRAKE TESTER Aimo Pusa Raute-MIKES Finland, by Raute Precision Oy, Force and Mass Laboratory Abstract The calibration of the roll brake tester for trucks has been made general by direct loading of the force transducer. That means to ignore the complete force transmission to rollers, which are in contact to trucks wheel. To be able calibrate exactly the braking force the Finnish institute YTOL has developed a test car to measure direct this braking force. The car has an extra axle with free rotating wheels and braking equipments as well the torque measuring devices. The axle can be moved in vertical position to have the weight maximum 4 kg on the axle. Raute-MIKES had the task to realise the calibration of this measurement device. The calibration has been made with a 2 knm reference torque transducer, which is calibrated static by PTB. The calibration of the brake measurement system has been made as dynamic with the speed of braking as static. Introduction The test of braking force by wheels in a truck is made with brake roll tester. The axle of car is located on two rolls, which are driven by an electromotor. By braking the wheels the electromotor tries to hold the speed with increasing power. The stator of electromotor is fitted with bearings, which give it freedom to rotate as well. This rotation is hindered by a force transducer, which measured the rotating force, the torque. Wheel of the car to test Rolls of the test device Chain Electromotor Supporting force against the rotating force of the electromotor, this force is measured as function of the braking force Figure 1. Principe of the roll brake tester Traditionally testers are calibrated with a beam and mass fixed directly to the force transducers. This method does not take in to account the influence of the mechanical system, bearings and diameter of rolls. Therefore YTOL-institute in Finland has developed calibration trail car to do the calibration dynamically. The trail car has a third axle in middle, figure 1. This axle is hanged free

with bearings and it has possibility to rotate longitudinal around the bearings center. This rotation is hindered with two beams supported on the force transducers, which measure the torque created by braking. It is possible to calibrate the force transducers individually but there are other factors, like length of the beams, friction of the bearings and so on, which are influencing the measurement uncertainty. Figure 2. Construction of the test car. Traceability Figure 3. One of the two beams with force transducer. Due the construction of the car, it was not possible to do the calibration in the laboratory. The calibration has been done as field calibration with a reference transducer of 2 knm. Dynamic calibration The aim of the dynamic calibration was to find rotation power of the no loaded axle, rolling friction of the wheels and finally the quasi-static calibration. The rotation of the axle was realized by an

electromotor and gearbox, the power of electromotor was 55 kw. The reference torque transducer was connected between the gearbox and the wheel of trail car, showed in figure 4. Figure 4. Rotating the wheel by quasi-static calibration. By rotating the axle in air (wheels did not contact the rollers) the needed torque was ca. 25 35 Nm (the friction of the bearings without vertical load). When the wheels were on the rolls (rolls free rotating) without braking, the measured torque was about 15 Nm. In normal measurement this torque will reduced from results by roll brake tester. Figure 5 shows how the force measurement follows the used torque. The force transducers are measuring the poor braking force, whereas torque transducer is measuring all needed torque. The speed of the wheels is by the measurement 2 km/h or 3 km/h. The measured signal shows that however the measuring is slowly dynamical the change of signal can have high frequency and the amplifier must have good dynamical behaviors. The dynamic character of signals is occurring from the wheel it self, ovality and roughness, and from the surface of the rolls. Test 42 7 6 5 4 N ; Nm 3 2 1,5 1 1,5 2 2,5 3 3,5 4 4,5 5 5,5 6-1 Time [s] Force transducers Reference transducer Figure 5. Braking with speed of 2, km/h, braking pressure 2 bar, used torque and measured braking force.

Static calibration By static calibration the reference transducer was located similar on the end of the axle and the torque was build by a beam and a screw, figure 6. Figure 6. Static calibration. By loading the torque both signals, torque transducer and force transducers, has been read synchronous with a high accuracy two channel amplifier. Figure 7 shows measured signals. The calibration has been measured several times to get the repeatability. The results show as well the friction of the system, the mechanical reason for that was not investigated. It has been assumed to come from disalignment error of the additional axle bearings. Calibration 11.8.1 Deviation to reference [Nm] 3 2 1-1 2 4 6 8 1 12-2 -3-4 -5-6 kal66 kal67 kal68 kal69 kal7 kal71 kal72 kal73 kal74 Torque [Nm] Figure 7. Results of the static calibration of the force measuring systems.

Length of the beams The length of the beams has been investigated by static calibration separate for each beam. Figures 8 and 9 show the results of these measurements. Right beam Signal of transducer [V] 3,E+ 2,5E+ 2,E+ 1,5E+ 1,E+ 5,E-1,E+.18 12.4 24.89 36.79 48.63 6.39 72.11 84.111 Time [s] 12 1 8 6 4 2 Torque [Nm] Right transducer Torque Figure 8. Calibration of the length by right beam. Left beam 3,E+ 12 Signal of transducer [V] 2,5E+ 2,E+ 1,5E+ 1,E+ 5,E-1 1 8 6 4 2 Torque [Nm],E+ 12 24 36 48 6 72 84 Time [s] Left transducer Torque Figure 9. Calibration of the length by left beam. From results calculated difference of the beam lengths is 7 mm which means relative error of,7 %. Mechanical measurement of the beam lengths was not easy, because construction did not have uniform reference points.

Uncertainty of the calibration The uncertainty of the calibration has been calculated according the guide EA-4/2 Calculation of the measurement uncertainty with following uncertainty components. Uncertainty of the reference transducer The transducer was calibrated by PTB in the dead weight calibration machine before and after the measurement, time between the calibrations was about 6 months. Calibration gave as uncertainty for the transducer < 8 1-5 (k=2) and shift of the sensitivity < 1 1-4. Uncertainty of the amplifier is based on the calibration certificate of the manufacturer, which is given as value 94µV/V (k=2). This component has very low influence to the final uncertainty, but it is taken into account by the calculation. Uncertainty of the force transducers is based as well on the calibration certificate, the relative value is below,1 %, which is not significant compared to the final uncertainty. Difference in the length by the beams is token into account with the value of,7 %, the real length is not corrected, which would be as well possible. Vertical possible deviation between the wheel axel and supporting points (bearings points) has been calculated from the mechanical dimensions and gives a value up to 1 Nm. Repeatability of the measurement has been evaluated from the results of the measurement with the value 15 Nm. Hysterese of the equipment has been evaluated as well from measurements results. In practical measurement the hysterese has very few influence to the result, because the load is continuously increasing. The uncertainty has been calculated with and without hysterese. The numerical value of the hysterese is 3 Nm. The change of the sensitivity by beams has been evaluated as well from the calibration results, the value is ca 35 Nm. Torque Uncertainty Uncertainty Torque Uncertainty Uncertainty with without with without hysterese hysterese hysterese hysterese knm % % knm % % 1 2,3 1,6 6 3,9 2,7 2 1,3 5,6 7 3,5 2,5 3 7, 4, 8 3,2 2,4 4 5,4 3,3 9 3, 2,3 5 5, 2,9 1 2,8 2,3 Figure 1. Uncertainty of the calibration

Summary The expectation value for the uncertainty of the calibration was 2... 3 %, which did not completely achieved. The normal, practical used measurement range is from 5 knm up to 12 knm and for this range the reached uncertainty is sufficient. For lower range the characteristic should be improve. The work has shown some weakness in the construction and it will renew for final routine checking of roll brake testers in Finland. Aimo Pusa MIKES-Raute by Raute Precision Oy Force and Mass Laboratory PL 22 FIN-1581 Lahti Finland Phone: +358 4 357 87 Fax: +358 3 829 416 E-mail: aimo.pusa@rauteprecision.fi