Perfusor Space. Service Manual

Similar documents
Operation and Maintenance manual

Cole-Parmer Pro-Spense

INSTRUCTION MANUAL. Battery powered crimping tool SMART-LINE. Battery powered crimping tool SO-6M. Art /M. convincing solutions.

CADD-Solis Infusion Pump

Troubleshooting. This section outlines procedures for troubleshooting problems with the operation of the system:

Service Bulletin No. 2905

B-RAD Select USER MANUAL TABLE OF CONTENTS

Electro-hydraulic crimping tools SD300-6M / SD400-6M. Art /M / /M. convincing solutions. Video /M Video /M

MOOG Medical Devices Group Curlin 6000 Series Ambulatory Infusion Pump PATIENT GUIDE

Repair instruction for HYDRAULIC DRIVE UNITS HA1ES / HA \r04eng 0697

PRESSURE GOVERNOR, ENGINE MONITORING, AND MASTER PRESSURE DISPLAY MODEL: TCA100

Electro-hydraulic crimping tool AH-12. Art convincing solutions. Video

Syringe Driver Introduction. Christine Hull & Anita Webb Staff Nurses, Hospice in the Home 2015

CADD Amulatory Ambulatory Infusion Systems CADD. Ordering Information. CADD Tubing Sets with Flow Stop Free-Flow Protection.

Ambulatory. Infusion. Pumps

IVTM Installation Manual

Battery powered cutting tool. AS6ST-20 Art convincing solutions. Video

CADD Ambulatory Infusion Systems

Quick Reference Guide

NIKI T34 pump. Knowing your way around the NIKI T34 pump

CADD Ambulatory Infusion Systems

Accumate 2300 Volumetric Infusion Pump

TOYOTA TACOMA TVIP V5

DIGITAL BATTERY TORQUE WRENCH (BC-RAD SELECT) USER GUIDE

TOYOTA FJ CRUISER TVIP V5 Preparation

Product Manual. Part Number: MVP-193 Mefi-6 ECU Revision: 1.0

Battary powered cutting tool. AS6-S32 Art convincing solutions. Video

Intermittent Delivery Mode

CADD -Solis VIP Ambulatory Infusion Pump Model 2120

Syringe Infusion Solutions

TOYOTA TUNDRA TVIP V5

Warmbier Germany, Wrist Strap & Footwear Combo Tester

Syringe Infusion Solutions

IVAC PCAM Syringe Pump. Technical Service Manual

TOYOTA YARIS HATCHBACK TVIP V3 (RS3200) Preparation

Installation and operating instructions. Solar charge controller MPPT 10 A / 20 A Z Z

TOYOTA RAV TVIP V5

WARRANTY...1 NOTICE INTRODUCTION USE...8

CTC 400 Cab Trailer Communication Power Line Carrier Technology Communications for Trucks and Trailers

SCION xa LIMITED EDITION SECURITY WITH RKE (V3)

Engineering Standards. Engineering Standard Number C A. Name. Digital Oxygen Remaining System O2 REMAINING

AS AS SAS6-85

TOYOTA PRIUS 2010 TVIP V5 PREPARATION

Electronic Limit Switch Type

User s Manual. Versaflo 200 Syringe Pump. For Veterinary Use Only

english Die holder Pushbutton to advance Pushbutton for return movement

Pivotell Advance Automatic Pill Dispenser Instructions

MANN+HUMMEL ENTARON XD. Installation and Maintenance Manual. Version 0709

TOYOTA TUNDRA TVIP V5

CONTROLLER DIAGNOSTIC GUIDE

HD [ GB ] Tipping bucket rain gauge

evaluation Syringe Pump Codan green stream SY-P ARGUS 600 D 36 (Free to the NHS) evaluation April 2004 NUMBER MAIN FEATURES Power supply

TOYOTA TACOMA TVIP V5

TOYOTA TUNDRA TVIP V3 (RS3000) Section I - Installation Preparation

LIGHT AND POWER SOURCES POWER SOURCES

SR7 Micro-infusion Syringe Pump. Contents

SCION xa 2004 SECURITY WITH RKE (V3) Section I Installation Preparation

TOYOTA RAV4 w/smart 2009 TVIP V4 PREPARATION

ELECTRIC PALLET TRUCK

TOYOTA VENZA 2009 TVIP V2 PREPARATION

STRESS-TEK, INC S. 194 th Street Kent, WA June 2011 Stress-Tek, Inc. Doc Rev A

Installation and operating instructions. Solar charge controller 10 A / 15 A / 0 A / 30 A PHOTOVOLTAIK - PHOTOVOLTAICS - PHOTOVOLTAIQUE - FOTOVOLTAICA

PRX-5A-HV. Operating manual. Stabilised receiver power supply. PRX-5A-HV Power for Receiver - High Voltage. Order No. 4176

Service Information Letter

KaVo Surgery Portfolio Current Questions and Answers

Leica GKL221 Charger User Manual

TOYOTA PRIUS 2010 TVIP V4 PREPARATION

TOYOTA HIGHLANDER w/smart Key TVIP V4 REMOTE ENGINE STARTER (RES) Part Number: PT

LEXUS RX TVIP V2 PREPARATION

Table of Contents. Timer Identification Timer ID BLU-U Features: 1K 6K BLU-U Features 1K 6K

ELECTRIC PALLET TRUCK

Locks LOCKING AND UNLOCKING. Automatic unlocking. Interior door handles and door locking levers. Master lock and unlock switches. Drive-away locking

Original Operating Manual

TOYOTA COROLLA 2010 TVIP V4 PREPARATION

EMERGENCY EGRESS LIGHTING POWER SUPPLY BATTERY

ISO 9001, ISO1400, ISO13485, CE and WHO-GMP certified.

User s Manual. Automatic Switch-Mode Battery Charger

INTRODUCTION. Specifications. Operating voltage range:

Function description of roll-over bar

Going home with a CADD Pump - Chemotherapy Medication

TOYOTA VENZA w/smart 2009 TVIP V4 PREPARATION

FRESENIUS 2008K 2 HEMODIALYSIS SYSTEM PREVENTIVE MAINTENANCE PROCEDURES. Part Number Rev. B

HEMODIALYSIS SYSTEM PREVENTIVE MAINTENANCE PROCEDURES. Part Number Rev. I

2008K 2 HEMODIALYSIS SYSTEM PREVENTIVE MAINTENANCE PROCEDURES. Part Number Rev. J

4L PROVIDER MATERIALS

LoadMaxx. Installation Guide. For Trailers with Mechanical Suspensions. Air-Weigh Customer Support: PN R0


TOYOTA TACOMA 2005 TVIP V5 (RS3200 PLUS) Preparation

BODYGUARD 595 REGIONAL ANALGESIA INFUSION PUMP

Shades Technics. 2050RS Toilet Cubicle. Manual

BodyGuard 121 Twins Infusion System Operator Manual

OPERATOR S MANUAL. Ambulatory Infusion Pump Model 6300 PATIENT CONTROLLED ANALGESIA

1. INTRODUCTION AND SYSTEM DESCRIPTION BLOCK DIAGRAM GENERAL INFORMATION GLOSSARY... 7

TOYOTA RAV4/HV INTERIOR LIGHT KIT Preparation

WEK. Wasserentkeimung METASYS WEK Light Gerätedokument Einbau, Betrieb und Wartung

Ambulatory Syringe Pump Operation Manual

CRONO S-PID 50 Ambulatory infusion pump

BodyGuardTM 323. BodyGuard 323 Multi-Therapy Ambulatory Infusion Pump. Patient Guide

Maintenance notice proceeding

TOYOTA CAMRY HYBRID TVIP V5 (RS3200 PLUS) Preparation

Transcription:

0-12 Perfusor Space Service Manual Version 3.0 English This document is not a complete Service Manual. It only encompasses the pages that describe how to determine the age and how to check the condition of batteries as well as the accordingly adapted template for Technical Safety Check. 0 -

Disassembly / Assembly 3 3.2 Battery Module Designation Ord. No. 1 Battery compartment cover PSP, cpl.............. 3452 0872 Battery module SP (NiMH) without fixing pin......................... 3452 0856 with fixing pin........................... 3452 0856A Battery module SP (Li-Ion) with WLAN and fixing pin................... 8713182A 2 Note Users have been informed as follows: If a Perfusor Space pump used on battery only and showing full battery charge stops during syringe change and alarms, users should remove the pump from use and return it to the Medical Engineering Department for replacement of the battery. In this case the battery has to be replaced regardless of its age or any further test. Fig.: 3-8 Legend of fig. 3-8: ItemDesignation 1 Battery compartment cover 2 Battery compartment cover lock Disassembly Note Move the drive head to the extended end position before starting disassembly (see General pg. 3-1). 1. Press the lock (Fig.: 3-8 / Item 2) on the battery compartment cover (Fig.: 3-8 / Item 1) and remove the battery compartment cover. Note The battery compartment cover of more recent devices is shorter and can only be unlocked using a pointed object. Perfusor Space, 3.0 3-9

5-4 Perfusor Space 5-Technical Safety Check (TSC) Index g (Master to be added to the documentation) Checklist for Technical Safety Check Every 24 Months Unit: Perfusor Space Manufacturer: B. Braun Melsungen AG User Observe the Service Manual and the instructions for use. All measured values are to be documented. Accessories used should be included in testing. Make exclusive use of calibrated measuring equipment. Article No. Unit No. Year of Procurement Stock No. Visual Inspection Electrical Safety Functional Inspection Perfusor Space: Cleanliness, completeness, damage and defects affecting safety, damage and readability of the labels. Particularly: Syringe holder with blade Syringe fastening Membrane in drive head Axial clearance of drive Screw covers Connectors P2 and P3 Protection cap at drive head Accessories: Cleanliness, completeness, damage and faults affecting safety, damage and readability of the label Check the unit and the accessories for compatibility According to IEC / 60601-1 or VDE 0750 and VDE 0751 The patient and housing leakage current of the Perfusor Space is caused exclusively by the operating voltage supply (Power Supply SP or SpaceStation). The Technical Safety Checks of the power supply SP (drawing No. M001321005 F04) or of the SpaceStation (drawing No. M690000046 F04) serve to check whether both limit values are met. Locking with second unit Operating unit magnets Battery compartment cover Determine age of battery If older than 4 years: recommend replacement of battery or perform battery check (refer to Procedural Instructions on the TSC ) Switch on unit with power supply Self-test Indicator lamps (LEDs) Audible alarm Visual alarm Status display Lighting Battery alarm M688 70 01 05 F04 / 3891 7319 (GB) Sheet 1 of 3 Perfusor Space, 3.0 5-1

1-Technical Safety Check (TSC) Index g (Master to be added to the documentation) Visual Inspection Electrical Safety According to IEC / 60601-1 or VDE 0750 and VDE 0751 Functional Inspection Operation Syringe fastening Syringe recognition Infusion Buttons on the operating unit Staff call Trigger bolus at the device Trigger bolus by pressing the PCA button PCA lock (bottom side of pump) with inserted 50/60 ml syringe PCA lock Syringe Anti Removal CAP PSP with inserted 50/60 ml syringe Switch on unit without power supply Self-test Magnetic function of the battery compartment cover Pressure cut-off Syringe type: #LehreOPS50 Delivery rate: 200 ml/h - Strain gauge pressure measurement Pressure stage 1 (9 15N) N Pressure stage 3 (25 33N) N Pressure stage 8 (63 76N) N Error message Alarm / pressure too high at every pressure stage - Motor power limitation Pressure stage 1 (11 28N) N Pressure stage 3 (30 49N) N Pressure stage 6 (58 80N) N Error message Alarm / drive blocked at every pressure stage M688 70 01 05 F04 / 3891 7319 (GB) Sheet 2 of 3 5-2 Perfusor Space, 3.0

1-Technical Safety Check (TSC) Index g (Master to be added to the documentation) Mechanical Aids and Measuring Equipment Used Accessories Used Syringe gauge, serial No. Calibrated until Power Supply SP Battery module Service connector SP Staff call lead Space PCA Kit (PCA button) Syringe Anti Removal Cap PSP Test result: Defects found which could endanger patients, users or third parties: Yes No Measures to be taken: Repair Replacement of battery recommended Special features / documentation: Inspection performed by: Unit handed over on: To: Date / Signature: Next deadline for TSC: M688 70 01 05 F04 / 3891 7319 (GB) Sheet 3 of 3 Perfusor Space, 3.0 5-3

Procedural Instructions on the TSC 6 Functional Inspection Perfusor Space Mechanical Inspection 1. Fit the unit to be tested on top of another Space device and check the proper functioning of the lock. 2. Fit the unit to be tested under another Space device and check the proper functioning of the lock. 3. Hold the device with the operating unit downwards. The operating unit must not open. 4. Check the battery compartment cover lock for proper operation. Functional Check 1. Determine age of battery Open battery compartment cover and remove battery. Read the batch number printed on the type plate of the battery and determine the age of the battery. a) For NiMH batteries, see Fig.: 6-2 b) For Li-Ion batteries, see Fig.: 6-3 Example: Produced in week 42/2008 Example: Produced in week 41/2011 D 2003 E 2004 F 2005 G 2006 H 2007 I 2008... N 2013 Two digits: calender week Letter: year of production according to left table Second two digits: year of production First two digits: calender week Fig.: 6-2 NiMH batteries Fig.: 6-3 Li-Ion batteries Note If the battery is older than 4 years recommend the replacement of the battery or perform the battery check. Perfusor Space, 3.0 6-3

6 Procedural Instructions on the TSC 2. Battery check Use the battery maintenance mode to discharge the battery to battery pre-alarm (the battery status indicator displays no bar). Perform 5 syringe changes with emptied syringe. - Replace the battery if a battery alarm occurs during the syringe changes. 3. Switch on unit with the power supply connected and check the following details: - Self-test Display on the LC display - Indicator lamps LEDs (yellow, green, blue) light up for a short moment LED changes from green to red Changed order for units from SW J on: LEDs (yellow, red, blue) light up for a short moment LED changes from red to green - Audible alarm A deep and a high sound - Status display Battery capacity, mains operation (service connector SP) - Lighting Illumination of syringe compartment, LC display and buttons 4. Syringe fastening a) Axial fastening holds syringe wing b) Syringe holder blade fixes syringe piston c) Both claws in drive head catch the syringe piston plate 5. Check syringe recognition. a) Insert approved 2 ml / 3 ml syringe. The syringe size is recognized. b) Insert approved 50 ml / 3 ml syringe. The syringe size is recognized. 6. Carry out infusion and bolus with any syringe and press all buttons at least once. Infusion and bolus are performed and all buttons trigger the function desired. 6-4 Perfusor Space, 3.0

Procedural Instructions on the TSC 6 7. Trigger bolus by pressing the PCA button. For this purpose, select a PCA drug from the database, fit the PCA button to P2, and trigger bolus. Confirm that the green LED on the PCA button lights up. Bolus is delivered. 8. Staff call a) Plug service connector SP on connector P2. b) Open syringe holder while the infusion is administered. The red LED on the service connector SP lights up. 9. PCA lock on the Perfusor Space a) Check the PCA lock with an approved 50/60ml syringe. 10. PCA lock of the Syringe Anti Removal Kit PSP a) Place on the drive head and lock. b) Check the Syringe Anti Removal Kit PSP with an approved 50/60ml syringe. c) Unlock and remove from the drive head. Battery Alarm 1. Switch device off. 2. Pull off the power supply. 3. Switch on unit. Self-test is carried out. 4. Open battery compartment cover during operation. An alarm is activated. 5. Remove battery. An alarm sounds with the piezo buzzer for at least 3 minutes. Perfusor Space, 3.0 6-5

9 Revision Documentation Version 3.0 - Changed TSC of the unit - Revised procedural instructions on the TSC - Release date: 25.10.2013 9-2 Perfusor Space, 3.0

B. Braun Melsungen AG Carl-Braun-Strasse 1 34209 Melsungen Germany Tel +49 (0) 5661 71-0 Fax +49 (0) 5661 71-4567 www.bbraun.de Mat. No. M688_001152, Doc. No. 8713 9020, Version 3.0, Effective 25.10.2013