Radius Cuspidor (for Cascade 1040 or Decade 1011/1021 Chairs)

Similar documents
CASCADE CHAIR-MOUNTED HANDPIECE CONTROL WITH CASCADE MASTER SERIES TOUCH PAD

Performer Delivery System (Performer I Chair - International, and Performer III Chair)

INSTALLATION INSTRUCTIONS CASCADE 3181 DUO DELIVERY SYSTEM TO PREFERENCECOLLECTION 5580

INSTRUCTIONS FOR USE CASCADE 2615 CART AND 3175 WALL MOUNT

A-dec 351 Assistant s Instrumentation on an A-dec 311/411/511 Dental Chair INSTALLATION GUIDE

Owner s Guide PERFORMER III

Asepsis 21 Delivery Unit

Owner's Guide CASCADE 2671 DUO CART AND CASCADE 3171 DUO WALL-MOUNT

Procenter Delivery, Console Mounted on an Elevance Chair Installation Guide

Simplicity Magellan Unit

A-dec O Clock Delivery System

A-dec 541 Delivery System and 545 Assistant s Instrumentation

Installation Instructions: Swing Mounted Units BDS-2575

axcs Chair Mounted Unit Installation, Operation and Care Manual

Installation Instructions for Swing Mount System

Simplicity Hygiene Unit

Evogue Dental Unit IMPORTANT

Instructions for Use. A-dec 4631 Duo Delivery System and A-dec 4635 Assistant s

Simplicity Chair Mounted Unit

Pre-Installation Guide Performer

axcs Magellan CMU Installation, Operation and Care Manual

Cascade 3072 Delivery System INSTALLATION GUIDE

DECADE 1021 SOLO CHAIR

A-dec 2122/2132 Radius System on A-dec 511 Dental Chair I NSTALLATION GUIDE

Installation Instructions: Over-The-Patient Units (& Cuspidors) BDS-2500 BDS-2505 BDS-2510 BDS-2515 BDS-2520

Procenter Delivery Mounted on a Free Standing Treatment Console (FTC) Installation

A-dec 200 P RE-INSTALLATION GUIDE

A-dec 311 Dental Chair and Related Systems

Service Guide A-DEC SIMULATORS. Mobile (4810) and Stationary (4820)

Assistant s Instrumentation T

A-dec 336 Delivery System on an A-dec 411 or 511 Dental Chair INSTALLATION GUIDE

Loose Components. VetPro 5000 Wall / Cabinet Mount Installation. Applies to Models:

Product Name Chair Conversion Kit (to LR Delivery Unit) Installation [Kit # xx]

Installation Instructions: Side Delivery Units BDS-2555

A-dec 371L or 571L Dental Light on an A-dec 361, 362, or 363 Support Center INSTALLATION GUIDE

Mounted Instrument Head

NextGen Cart Delivery Unit

Wall / Casework Mounted Delivery Systems (Procenter)

Instructions for Use. Cascade 2615 Cart

DECADE 1021 VAC BACK CHAIR

Product Name Chair Conversion Kit (to Console Delivery Unit) Installation [Kit # xx]

A-dec Dental Lights on an A-dec 300 Base Mount Post INSTALLATION GUIDE

Simplicity Cart. Installation, Operation & Care Manual. Duo. Doctor s

Cascade Master Series

Table of Contents. Section I Introduction. Section V Operation. Section II Pre-installation. Section VI Care. Section III Installation

Instruction Sheet. Subject: Installations Instructions. Air Compressor Kits. Importan

Flex Work Station Delivery Units Installation Guide

Installation Operation Maintenance Troubleshooting Version 1, Oct/09. Palm Beach

A-dec Service Guide Edition , 2003

Troubleshooting Guide: Elevance Delivery Systems

MEDICAL IRRIGATION K PUMP OPERATING MANUAL: Manufactured for: K.M.I 3185 Palisades Dr Corona CA USA Model No.

1416 Cabinet Unit. Installation Operation Maintenance Troubleshooting Version Dec/17

Galaxy Wall / Cabinet Mounted Unit

Spirit 2000 Series Installation Instructions

Chair Mounted Unit. Installation, Operation and Care Manual. DentalEZ Group 2500 Highway 31 South Bay Minette, AL 36507

A-dec 542 Side Delivery System I NSTALLATION GUIDE

PowerAir Compressor Installation

ClassicSeries Compressor Installation

Installation Kitchen Faucet Filtration System Instructions GXK285JBL

MODEL NUMBER: MEDIUM DUTY ONBOARD AIR SYSTEM

Mirage Dental Delivery Unit MSP3500/3550 Installation Instructions

Table of Contents Next FLO-SAFE Manifold System Pre-Installation Kit Matrx Toll-Free: Technical Support: Fax: Rev.

A-dec 200 Service Guide Contents

forest dental australia affordable excellence + designer friendly

CHEMTROL TRIPLE SPACE SAVER CLEANING STATION #U16242 INSTALLATION INSTRUCTIONS

Junction Boxes. Common Components Air / Water Control Valves. Contoured Junction Box. Small Junction Box. Elevance Junction Box

AEU-525CF AEU-525CFH

A-dec 570L Dental Light on a Non-DCS System INSTALLATION GUIDE

Southwest Windpower Instruction Sheet AIR-X Circuit Replacement Kit

INTRODUCTION TO SERVICING THE A-dec Handpiece Control Blocks

Procenter Delivery Systems

The Belmont Warranty does not cover damage to equipment and upholstery caused by disinfectant solutions. WARRANTY

Procharger Stage II Intercooled Supercharger System (11-14 GT)

User s and Installation Manual

Track Light with Monitor Installation Guide

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-12DCE Task Force Deluxe Portable Pneumatic Dental System

Installation Manual 354 / 355 Combination Lighting System

Dry Vacuum System. Part Numbers: V15, 2V15, 3V15 and 4V15 USER S MANUAL. Recommend Daily use of Monarch CleanStream

HEAVY DUTY ONBOARD AIR SYSTEM

INSTRUCTIONS FOR USE. Operation and Maintenance manual. Light Industrial Compressor. Hi-Tech Diagnostics Service

Instructions for Use. Performer. Dental Chair, Delivery System, Assistant s Instrumentation, Support Center with Cuspidor, and Light

HEAVY DUTY ONBOARD AIR SYSTEM PART NO

VALVE AND PLUMBING KIT INSTRUCTIONS SMC 84Q & 2408 LOADERS NEW HOLLAND TRACTORS MODEL 2WD 4WD LESS CAB WITH CAB 1720 X X X 1920 X X X

Quolis Series. Quolis 5000 Chair IMPORTANT

MEDIUM DUTY ONBOARD AIR SYSTEM

Tubing / Wiring Diagrams: Elevance Delivery Systems (Console / LR Mtd: Elevance Chair)

VALVE AND PLUMBING KIT NEW HOLLAND 7310 LOADER NEW HOLLAND TRACTORS

Installation Guide. Nitrogen-Phosphorus Detector on 6890 GC Accessories G1575A, G1576A, G1594A, G1599A

L5000 Plus. Installation & Setup Guide

INSTALLATION and OPERATION INSTRUCTIONS

SECTION 4 - FUEL SYSTEMS AND CARBURETION

SALEEN SPEEDLAB BOOST AND WATER TEMPERATURE GAUGE POD KIT

Air Conditioner for M915 A0/A1 Truck

08-18 STI Flex Fuel Bluetooth Mk2 Kit Install Instructions For Cobb Tuning Access Port

VHM-P (Non-Locking) and VHM-PL (Locking) Variable Height Arm with Slide-In Mounting Plate

PRO RATCHET UNIVERSAL SHIFTER

ADVANTAGE Annual Preventive Maintenance Instructions

6.4L EGR Delete With Intake Elbow

SHOP PRO FXP 95 UPGRADE KIT (REGULATED)

Procenter Cuspidor. Common Components Console Mounted Cuspidor. for Ultra Chairs for Elevance Chairs. for European Assistant s Unit

Transcription:

INSTALLATION INSTRUCTIONS Radius Cuspidor (for Cascade 1040 or Decade 1011/1021 Chairs) You Will Need: Hex key set Level Large flat head screwdriver Pliers 1/2" and 9/16" wrench Cutters Before Beginning this Install 1. Install the air and water manual shutoff valves. Refer to the Floor Box Template (A-dec Publication No. 85.0472.00). Purge the air and water supplies of debris. Chair footswitch Figure 1 Touchpad 2. Install the dental chair. Refer to Cascade 1040 chair Installation Instructions (A-dec Publication No. 86.0157.00) or Decade 1011/1021 Chair Installation Instructions (A-dec Publication No. 86.0156.00). WARNING The chair must be anchored to the floor to meet agency requirements. Refer to the Radius Floor Box Installation Instructions (A-dec Publication No. 85.2885.00) or the Radius 2122/2132 Delivery System Installation Instructions (A-dec Publication No. 85.2877.00) for chair anchoring instructions. Figure 2 3. If a delivery system is being installed, do so now. Refer to Radius Delivery System Installation Instructions (A-dec Publication No. 85.2877.00). ELECTRICAL WARNING Failure to disconnect the chair from its power source can result in serious injury. Raise the Chair Raise the chair base completely up using the footswitch (see Figure 1) or the touchpad on the control head (see Figure 2). Unplug the chair. Page 1 86.0104.00

Remove the Chair Covers 1. Remove the chair stop plate (see Figure 3). If you have a non-radius chair, discard the stop plate. Go to step 2. If you have a Radius chair, remove and discard the cover on the stop plate. Go to step 2. Chair stop plate, non-radius chair (discard) Install the Arm Mount 1. Remove the arm mount from the shipping carton. Loosen the bolt almost all the way (see Figure 6). 2. Place a small amount of blue Loctite on the threads of the bolt (see Figure 6). Loctite here (place on bolt threads) Bolt (loosen) Chair stop plate, Radius chair (keep) Arm mount Figure 6 Stop plate cover, Radius chair (discard) Figure 3 2. Remove and discard the chair upper lift arm cover (see Figure 4). 3. Move the brake handle to the right. Place the arm mount onto the chair swivel casting. Center the arm mount within the grooves of the chair swivel casting and under the lip of the swivel casting (see Figure 7). Be sure that the limit switch cable is beneath the arm mount. Chair upper lift arm cover (discard) CAUTION Be careful not to damage the limit switch cable while installing the arm mount. Chair swivel brake handle Figure 4 3. Remove the support bracket using a large screwdriver (see Figure 5). Set it aside. Screws (discard) Chair swivel casting Arm mount Limit switch cable Figure 7 Support bracket (keep) Figure 5 Screwdriver 4. Tighten the bolt securing the mounting block to the swivel casting (see Figure 6 and Figure 8). Page 2 86.0104.00

Screw (place blue Loctite on threads) Thrust washers Thrust bearing Bolt (tighten) Link arm Washer Figure 8 Cuspidor arm Install the Mounting Arm 1. Remove the link arm from the shipping carton. 2. Install the link arm to the arm mount as shown in Figure 9. Screw (place blue Loctite on threads) Figure 10 Thrust washers Thrust bearing Link arm Rotation stop cover (remove to install mount cover, then reinstall.) Washer Arm mount cover Arm mount cover screw Arm mount Figure 9 3. Install the Radius cuspidor arm as shown in Figure 10. Prop the packaging insert under the cuspidor to help support it while installing. Page 3 86.0104.00

Connect the Cuspidor Umbilical Position the cuspidor to the right or left of the chair according to the doctor s requests before connecting the tubings. Connect the Limit Switch and Touchpad Cable 1. Remove the jumper from the cuspidor limit switch terminal. Tape the jumper inside the chair stop plate in case of any future need. 2. Connect the cuspidor limit switch terminal to the chair circuit board. 3. Connect the cuspidor touchpad cable to the Y-adapter circuit board (see Figure 11). If a delivery system is not present, connect the cuspidor touchpad cable to the chair circuit board (see Figure 11). 4. Peel the covering from the Y-adapter circuit board and adhere it to the power supply. Connect the Tubing Municipal Water Connections 1. Go to Install the Utilities, if you are installing municipal water connections. Connect the Self-Contained Water 1. Cut the red tubing from the floor box to the junction box. Inline the red cuspidor tubing to the Ground wire (connect to the chair structure) Cuspidor umbilical Red tubing (1/8" OD) Radius Cuspidor Umbilical Connections (For clarity, delivery system tubing has been omitted.) Inline cuspidor tubing to junction box tubing. Optional low voltage water heater cable Junction box tubing (Cut) To the delivery system (if present) Cuspidor limit switch and touchpad cable Delivery system touchpad cable To the cuspidor Vacuum tubing Cuspidor drain tubing Yellow with green dashes tubing (1/8" OD) Yellow tubing (1/8" OD) Blue with white dashes tubing (3/8" OD) Cuspidor limit switch terminal connects to chair circuit board Y-adapter circuit board Floor box side Tubing clamp Cuspidor side To the floor box To chair circuit board Page 4 86.0104.00

tubing going to the junction box (see Figure 11). Use the tubing clamps provided to secure the cuspidor drain and vacuum drain tubing. NOTE: The self-contained water option is not available on Radius cuspidors without a delivery system. 2. Connect the red tubing (1/8" OD) to the matching red tubing in the junction box using an inline barb and uni-clamps (see Figure 12) 3. Remove the water bottle bracket from the Radius delivery system post. Cut the red tubing (1/8" (see Figure 15). 5. Connect the three lengths of red tubing (1/8" OD) using a tee and uni-clamps (see Figure 14). Push Delivery system post Bracket mounting screw and spacer Bracket Red tubing Cascade Radius junction box Red tubing (1/8" OD) Tee Red tubing Figure 14 the tee into the water bottle bracket. 6. Reinstall the bracket to the delivery system post. Figure 12 Decade Radius junction box 7. Remove the thumbscrew underneath the cuspidor and slide the cover down (see Figure 15). OD) connected to the bracket (see Figure 13). 4. Pull out the end of the red tubing (1/8" OD) Delivery system post Tubing hole Bracket Bracket mounting screw and spacer Cuspidor cover Red tubing (1/8" OD) cut Thumbscrew Red tubing (1/8" OD) coming from chairmount end of post Figure 13 Figure 15 coming from the chair-mount end of the delivery system post (see Figure 14) and route the tubing through the tubing hole in the bracket Page 5 86.0104.00

8. Disconnect the red syringe tubing from the water manifold. Install a barb plug onto the barb and secure with a uni-clamp (see Figure 16). CAUTION If the water manifold barb is not plugged, flooding will occur. In-line barb Red tubing (1/8" OD) Sleeve Red syringe tubing Barb cap Figure 16 9. Connect the red syringe tubing to the red tubing using an in-line barb and uni-clamps (see Figure 16). 10.Slide up the cover and tighten the thumb screw (see Figure 17). Pull on the saliva ejector and syringe tubing while sliding up the cover to make sure they do not get pinched. Figure 17 Page 6 86.0104.00

Install the Utilities Refer to the appropriate Plumbing Diagram (Figure 18 or 19). Also refer to Radius Floor Box Installation Instructions (A-dec Publication No. 85.2885.00) or Radius 2122/2132 Delivery System Installation Instructions (A-dec Publication No. 85.2877.00) for utility connections. Test the Cuspidor 1. Plug in the chair. 2. Open the manual air and water shutoff valves. For air and water specifications, refer to the Floor Box Owner s Guide (A-dec Publication No. 85.2611.00). 3. Plug in the power supply if applicable. Install barb with nylon washer for red tubing (remove the hex plug) Red tubing (oral water, 1/8" OD) Water filter/ regulator assembly (black body) 40 psi pre-regulator (non-adjustable) Blue with white dashes tubing (water supply, 3/8" OD) Yellow with green dashes tubing (40psi regulated air, 1/8" OD) Yellow tubing (unregulated air, 1/8" OD) Tubing from the umbilical 40 psi regulator Radius Cuspidor (without Radius Delivery System) Figure 18 Air filter/ regulator assembly Water filter/ regulator assembly Yellow tubing (air supply, 1/4" OD) Yellow with red stripe tubing (1/8" OD) Yellow tubing (1/8" OD) Yellow tubing (1/8" OD) Yellow with red dashes tubing (1/8" OD) Red tubing (oral water, 1/8" OD) Do not connect Blue w/white dashes tubing (water supply, 3/8" OD) Tubing from the umbilical 40 psi regulator 40 psi preset regulator (non-adjustable) Yellow with green dashes tubing (1/8" OD) Yellow with green dashes tubing (1/8" OD) Radius Cuspidor Utility Plumbing Diagram (with Radius Delivery System) Figure 19 Page 7 86.0104.00

4. Move the master on/off toggle on the delivery system or floor box to the ON position. 5. Check for air and water leakage at all connections while operating the syringe (see Figure 20). Secure the cuspidor umbilical to the arm mount with a tie wrap (see Figure 22). Arm mount Air Water Spray Figure 20 Umbilical Tie wrap 6. Check for vacuum leaks while operating the HVE. To operate the HVE, simply turn the control valve. 7. Test the cuspidor and check for water leakage at all connections. For operating instructions, refer to the Cascade 7284 and Radius 7285 Cuspidor Owner s Guide (A-dec Publication No. 85.2609.00). Figure 22 Install the Support Bracket Reinstall the support bracket by snapping the bracket onto the chair link arms (see Figure 23). 8. Test the touchpad operation. For operating instructions, refer to the Decade 1011/ 1021 Chair Owner s Guide (A-dec Publication No. 85.2635.00) or the Cascade 1040 Chair Owner s Guide (A-dec Publication No. 85.2605.00). For adjustments to the cuspidor and its components, refer to the appropriate Owner s Guide. Adjust the Tension Adjust the tension of the cuspidor arm to the desired resistance (see Figure 21). The cuspidor arm should rotate easily but not drift. Install caps over the screws after the tension has been adjusted. Support bracket Cuspidor arm Tension adjustment Caps Figure 21 Figure 23 Before Leaving the Facility Reinstall the floor box cover and stop plate if you have a Radius chair. If not, install the new stop plate onto the chair. Test the operation of the stop plate. Distribute the Owner s Guides and instruct the dental team on the operation and maintenance of the equipment. Page 8 86.0104.00

Page 9 86.0104.00

Identification of Symbols SYMBOL DESCRIPTION Classification of Equipment (EN-60601-1) Recognized by Underwriters Laboratories Inc. with respect to electric shock, fire and mechanical hazards only in accordance with UL 60601-1 (2601-1) and under mutual recognition agreement with CAN/CSA C22.2, No. 601.1. Classified by Underwriters Laboratories Inc. with respect to electric shock, fire and mechanical hazards only in accordance with UL 60601-1 (2601-1) and under mutual recognition agreement with CAN/CSA C22.2, No. 601.1. UL listed to UL 61010A-1, BS EN 61010-2-010 and Canadian (CAN/CSA C22.2, No. 1010.1-92) safety standards. Conforms to applicable European Directives (refer to Declaration of Conformity). Protective earth (ground). TYPES/MODE Types of shock protection Degree of shock protection Degree of protection against water ingress CLASSIFICATION CLASS I EQUIPMENT: Dental chairs, dental lights, and power supplies CLASS II EQUIPMENT: Chair, wall, and cart-mounted delivery systems TYPE B APPLIED PART: Delivery systems only ORDINARY EQUIPMENT: All products Functional earth (ground). Attention, consult accompanying documents. No user serviceable parts. Attention, line voltage. Only licensed electrician should remove cover. Type B applied part. Class II equipment. Mode of operation Flammable gasses: CONTINUOUS OPERATION: All models except dental chairs CONTINUOUS OPERATION WITH INTERMITTENT LOAD- ING: Dental chairs - 5% duty cycle Not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide, where such gasses may accumulate in concentration (closed space). Caution: Metal surfaces can be hot during and following the dry cycle. Electrical Rating Environmental Specifications TEMPERATURE/ HUMIDITY Storage/ Transportation SPECIFICATION Temperature: -40C to 70 C (-40 F to 158 F) Relative humidity: 80% for up to 31 C, decreasing linearly to 50% at 40 C. TYPE Volts: Frequency: Current: SPECIFICATION 100/110-120/220-240 VAC 50-60 Hz As configured and specified in equipment manual (products labeled 15A or greater require dedicated circuit, identified in distribution panel). Operating Termperature: 10 C to 40 C (40 F to 104 F) Relative humidity: 80% for up to 31 C, decreasing linearly to 50% at 40 C. Indoor Use: Altitude up to 2,000M (6,563 ft.), installation category II, pollution degree 2. (UL 61010A-1 and CAN/CSA C22.2, No. 1010.1-92 only) Printed in USA. Copyright 2006, All Rights Reserved. Page 10 86.0104.00