Agilent G3510A/G3511A Multimode Inlet

Similar documents
Agilent G4367A Auxiliary Electronic Pressure Control (Aux EPC) Module

Agilent 490 Micro GC Internal and External Mount Genie Filter Brackets

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Agilent G1554A On-Column Capillary Inlet with EPC

Installing a Cool On-Column Inlet

Installing a Purged Packed Inlet

Installing a Purged Packed (PP) Inlet

Installing a PTV Inlet

Installing a PTV Inlet

Agilent 7697A Headspace Sampler

Installation Guide. Volatiles Interface for 6890 Gas Chromatograph Accessory G2319A

Installation Guide. Nitrogen-Phosphorus Detector on 6890 GC Accessories G1575A, G1576A, G1594A, G1599A

Agilent G2738A Upstream Capillary Interface

Installation Guide. Flame Ionization Detector on a 6890 GC. Accessories G1561A, G1562A, G1591A, G1598A

Agilent Multimode Inlet

Installation Guide. Purged Packed Inlet, Electronic Pneumatics Control on 6890 GC Accessory G1551A

Installation Guide. Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A

Installing the CO 2 Cryogenic Oven Cooling Kit

Installing the Flame Ionization Detector EPC Flow Control Manifold

Valve Box and Valve Accessories G4315A, G4316A, G4317A

Installing an Auxiliary Pressure Control Module

Installing an Auxiliary Pressure Control Module

Agilent 7890 Series Gas Chromatograph

Installation Guide. Heater Interface Cable for Mass Sensitive Detector Accessory G1590A

Installation Guide. Flame Ionization Detector on a 6850 GC Accessory G2621

Nickel Catalyst Tube on a 6820 GC, Accessory G4318A

Agilent N2763A Rack Mount Kit for 4000 X-Series Oscilloscopes

1310 Inlets. 1 of 36. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual

Installation Guide. Thermal Conductivity Detector on a 6850 GC Accessory G2623A

Agilent N5470A Series Oscilloscope Rack Mount Kit

Installing the 6850 Direct Valve Column Connector Kit

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Agilent 6890 Gas Chromatograph

Installation Guide Thermal Conductivity Detector Accessory 19232E

260 Solvent Vapor Exit Accessory

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

Agilent G2333B (Single) and G2334B (Dual) Flame Photometric Detectors

Agilent 7820A Gas Chromatograph

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application

INSTALLATION OF OPTIC 3 INJECTOR SYSTEM ONTO AN SHIMADZU GC-2010 GAS CHROMATOGRAPH.

For more information about Agilent Technologies services please visit our web site using the following URL

Using the PSD for Backflushing on the Agilent 8890 GC System

Agilent 1260 Infinity Manual Injector

29 Installation Step 1. Unpacking the GC Cable diagrams Step 2. Placing the GC system on the benchtop Step 3. Turning the power on

P-Series Installation

Agilent G2647A (Single) and G2648A (Dual) Flame Photometric Detector Update Kits

Lamp House with 2.5 kw lamp

Agilent 1200 Series Manual Injector

Agilent G2855A Deans Switching System

Nickel Catalyst Tube Accessory 19205A

Deans Switch Microfluidics

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Agilent G3348B Flame Photometric Detector

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

A g i l e n t A G a s C h r o m a t o g r a p h Service Manual

In addition to this instruction sheet, the following are included in this kit (some additional parts are included for future maintenance):

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

Model 4430 Photoionization Detector (PID) Operator s Manual

Manual No: Revision: F. Mag Probe Assembly Manual

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE

1350 Valves. 1 of 14. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual

Owner s Manual Pelican Whole House Iron and Manganese Filtration System

Choose for Your Instrument. GC Supplies - Agilent Instrument Quick Pick Guide

Agilent 7693A Automated Liquid Sampler

10 Maintenance. Agilent Technologies. This chapter describes the maintenance of the module Infinity Binary Pump User Manual

Agilent 1100 Series Thermostatted Column Compartment. Reference Manual

Application Note. Abstract. Authors. Environmental Analysis

5000TOC Sensor Service Manual

Installation Note. Agilent Technologies E4440A Spectrum Analyzer Option BAB, APC 3.5 Input Connector Retrofit Kit

AssayMAP 96AM Wash Station. Maintenance Guide

6890 Installing ¼ Glass Columns

Air Fuel Ratio Module and AFR-4 Pump Assembly Installation and User Guide.

Installation Note for Solvent Selection Valve Upgrade Kit (G1360A)

Agilent Multimode Inlet for Gas Chromatography

GC Injector Liners. Bruker Varian Agilent (HP) Perkin Elmer Finnigan. Phone: Page 4

AssayMAP 96AM Wash Station. Maintenance Guide

HP Modular Cooling System 200/100 Options Installation Guide

Application. Gas Chromatography June 1995

Drug Testing Labs. Style 2 W/ Ver 4.41 pcb & Pulse Relay pcb(s) Installation, Operations, Start-up and Maintenance Instructions

Procedure Replacing a Cover

OPTIC-4. Multimode GC Inlet. Installation Guide OPTIC-4 Inlet CO 2 Cooling Option.

Torque Module Installation and User Guide for model 250i Motorcycle Dynamometers.

INSTALLATION INSTRUCTIONS FUEL RAIL

PARTS CATALOG MODEL FBD56X SERIES FROZEN BEVERAGE DISPENSERS (INCLUDES AFCTM PARTS)

Owner s Manual Nitrate Series Salt Based Filter System

Dispenser Pan Sensors & Containment Sump Sensors

Installation Instructions for John Deere cotton picker models: 9986 & 2-row and All-row systems included.

Patient Care Facility

Technical Procedure for Gas Chromatography (GC-FID)

Calibration and Performance Check: Agilent Dual Plasma Controller

Drug Testing Labs. Style 2 W/ Ver 4.41 pcb & Pulse Relay pcb(s) Installation, Operations, Start-up and Maintenance Instructions

TOYOTA RAV FOG LIGHT KIT Preparation

GC Inlets. An Overview. Simon Jones GC Applications Engineer

Agilent and ASTM. Update on Recent Activities. Page 1

Upgrading the Sure Coat Modular Gun Control System

Installation Instructions

STATIM CASSETTE AUTOCLAVE TROUBLESHOOTING GUIDE

2-row and All-row systems included.

Installation Manual Stand-Alone Dispenser Pan Sensor

Transcription:

For the Agilent 7890A Gas Chromatograph Installation Instructions Parts Supplied Table Liquid Carbon Dioxide (CO 2 ) Kit G350A Description Liquid CO2 MMI inlet assembly (MMI Inlet weldment assembly, EPC module, Thermocouple board, TC board cable, split vent filter and valve) MMI Inlet Chassis CO 2 Cryo bracket and CO 2 valve Captive screw, M4 x 4 mm long (for CO 2 cryo bracket) 2 Split vent filter retaining bracket Machine screw, M4 x 2 mm long (for Thermocouple board and split vent filter retaining bracket) Column Nut Cup Capillary Cup Insulation Screw M4 x 2 mm long (for column nut cup) 2 MMI ship kit Column hanger Liner, MS Certified, splitless, single taper, deactivated, glass wool mm low bleed septa (5 pk) O-Ring, 2-00, Fluoroelastomer (5 pk) Ferrule, Graphite, 0.5 mm id for 320um columns (0 pk) Agilent 3500A/G350A Multimode Inlet Large Volume InjectionTutorial Wrist strap, disposable 4-LG -W Quantity 3 Agilent Technologies

Table 2 Liquid Nitrogen (N 2 ) Kit G35A Description Liquid N2 MMI inlet assembly (MMI Inlet weldment assembly, EPC module, Thermocouple board, TC board cable, split vent filter and valve) MMI Inlet Chassis CO 2 Cryo bracket Captive screw, M4 x 4 mm long (for CO 2 cryo bracket) 2 Split vent filter retaining bracket Machine screw, M4 x 2 mm long (for Thermocouple board and split vent filter retaining bracket) Column Nut Cup Capillary Cup Insulation Screw M4 x 2 mm long (for column nut cup) 2 Liquid N 2 cryo valve assembly Liquid N 2 insulation cover MMI Liquid N 2 nut plate Screw M4 x 0.7 2mm (for N2 insulation cover, N2 hut plate, N2 cryo assy) 6 MMI ship kit Column form hanger Liner, MS Certified, splitless, single tpr, D, GW mm low bleed septa (5 pk) O-Ring, 2-00, Fluoroelastomer (5 pk) Ferrule Graphite 320um 0.5mm id (0 pk) Agilent 3500A/G350A Multimode Inlet Large Volume InjectionTutorial Wrist strap, disposable 4-LG -W Quantity 3 2 Installation Instructions

Parts Identification Wiring harness EPC module MMI inlet body Split vent filter assembly Split vent valve assembly Figure MMI CO 2 parts EPC module MMI inlet body Split vent filter assembly Split vent valve assembly Thermocouple board N2 cryo valve assembly Figure 2 MMI N 2 parts Installation Instructions 3

CO 2 Cryo assembly CO 2 cryo bracket Figure 3 CO 2 cryo assembly with board bracket Cryo tubing N 2 cryo assembly N 2 cryo insulation cover Figure 4 N 2 cryo assembly 4 Installation Instructions

Tools Required Installation Procedure T- 0 and T- 20 Torx drivers Open- end wrench Diagonal sheet metal cutter This procedure explains how to install the Multimode (MMI) inlet accessory on the Agilent 7890A Gas Chromatograph (GC). This document covers the following Multimode inlet accessories: Table 3 Kit Number Coolant G350-64000 Liquid carbon dioxide (CO 2 ) G35-64000 Liquid nitrogen (N 2 ) Before beginning the installation, read the 7890A Safety Manual found on the Agilent Instrument Utility software DVD. Check the GC firmware revision Prepare the GC Make sure the instrument/component is turned on and not in use. Make sure any sampler or other device is also connected and turned on. 2 Press [Status][Clear] to view the GC firmware version. 3 The MMI requires 7890A GC firmware A.0.0 or higher. If the firmware needs to be updated, use Agilent Instrument Utility B.0.0x or Agilent Lab Advisor B.0.0x or greater to update the firmware. To get the latest firmware, see the Agilent web site. For help in performing the update, see the Instrument Utility or Lab Advisor help. \ Turn off the GC and unplug the power cord. 2 Allow time for the oven and heated zones to cool. 3 Turn off all gas supplies. Installation Instructions 5

4 If there is an Automatic Liquid Sampler (ALS) tray installed on the GC, remove it. See the ALS user documentation for instructions. 5 Remove the detector cover by raising the cover vertically and then firmly lifting up on the right side of the cover to free the lid from the hinge pin. Slide the pin out of the hole on the left side hinge and put the cover aside. 6 Clean out the oven, removing all columns and hardware associated with both inlets. 7 Remove the gray plastic inlet cover by loosening the screws. 8 Remove the left side cover by loosening the top screw, sliding the panel towards the rear, and lifting the cover off. 9 Unsnap the pneumatics cover by pressing the black clips on the sides of the cover. Lift the cover up and off. 0 Remove the EPC module bracket adjacent to the inlet EPC module area by removing the screw and lifting the bracket off. EPC module brackets Figure 5 Locating the EPC module brackets Put on the ESD wrist strap and attach the ground to the GC sheet metal frame for electrostatic protection. Prepare the inlet mounting position Choose the position to install your MMI inlet. The MMI inlet can be installed in the front or back inlet position. If an inlet is already installed in your chosen position, remove the inlet, pneumatics module, and all associated tubing and wiring. 6 Installation Instructions

If your chosen inlet position is not used: Remove the round metal cutout at this location using diagonal cutters. Make the cuts so that the metal nubs remain attached to the discarded metal circle. 2 Remove and discard the circular insulation plug. Replace the inlet chassis If the chassis contains an inlet that you are keeping in addition to the MMI inlet, remove this inlet and temporarily set it beside the chassis. 2 Record how the wiring of the fan is routed but do not unplug the connections. This fan will be installed in the new chassis. 3 Remove the fan from the old inlet chassis by pressing on the clips and lifting the fan out. Fan removed Inlets removed Figure 6 Inlets and fan removed 4 Remove the GC electrical connectors from the inlet chassis by pressing the clips on the connectors and pushing them down, freeing them from the chassis. Record the position and order of the connectors on the chassis. 5 Remove the old inlet chassis by removing the screws in each corner and lifting the chassis off. This chassis will not be used, but the four removed screws with be used to secure the new chassis. 6 Install the new MMI inlet chassis, positioning it similarly to the old chassis alignment with the fan slot to the rear of the GC. Do not screw the chassis down at this time. Installation Instructions 7

7 Align the fan with the sloping rear face of the chassis so that the direction of the air flow matches the direction of the arrow on the fan body. 8 Route the wiring as recorded in step. 9 Press the fan into the chassis until the clips engage. 0 If removed, install the inlet you are keeping. Route the GC electrical connector ends from bottom to top and fit into the slots on the new inlet chassis in the same configuration as before. You may need to lift up the inlet chassis to fit these wires under the GC frame. New inlet chassis GC electrical connectors Figure 7 GC connectors routed properly into inlet chassis frame Install the liquid CO 2 MMI inlet module If you are installing the liquid N 2 MMI inlet module, skip to Install the liquid N 2 MMI inlet module on page 5. 8 Installation Instructions

Prepare the cryo assembly (liquid CO 2 MMI only) The solenoid valve assembly should be installed onto the cryo bracket as shown in Figure 9. If not, assemble as follows: Solenoid valve assembly Bracket Screw positions Figure 8 Installing the cryo valve assembly onto the new bracket Install the solenoid valve on the new bracket using two screws as shown in Figure 8. 2 If necessary, disconnect all electrical connections on the thermocouple board. 3 Mount the thermocouple board onto the new cryo assembly bracket using two screws. Plug the solenoid valve control wire into the connector on the thermocouple board. Route the coolant tube around the back of the thermocouple board bracket to avoid an electrical short. Solenoid valve control wire Thermocouple board Screws (2) Figure 9 CO 2 cryo and board assembly ready for installation Installation Instructions 9

Position the inlet Rest the EPC module in the pneumatics area. Do not install the EPC module at this time. It will be installed at the end of the procedure. 2 Insert the MMI inlet body into the desired inlet position in the chassis so that the tubing and wiring are running to the left side of the chassis. See Figure 3 and Figure 5 for proper routing. Install the split vent valve and filter assemblies Install the split vent filter into the housing, making sure that the flat sides on the split vent filter are vertical. The split vent filter will lock into place when properly installed. See Figure 0. Split vent filter installed in housing Split vent filter retaining bracket Screws securing split vent valve assembly Figure 0 Installing the split vent filter assembly 2 Align the split vent valve assembly over the two screw holes and secure using two captive screws. 3 Install the split vent filter retaining bracket with one screw to secure the split vent filter. 4 Position the tubing from the split vent assembly to the inlet so that it will not be pinched by the left side cover. 0 Installation Instructions

Install the cryo assembly Place the cryo assembly into the position next to the split vent assembly. Inlet tubing and wiring properly routed Cryo assembly installed in position Figure Cryo assembly installed 2 Secure the bracket using two screws. 3 Route the cryo tubing across the left side of the GC. Do not connect the tubing to the inlet until the electrical wires are routed, as described in the following steps. Route the MMI electrical wiring Disconnect the wiring harness from the inlet heater cable. 2 Route the white connector ends (FI and FV) of the wiring harness up through channels on the GC (Figure 2), plugging them into the appropriate connectors (FI and FV or BI and BV) on the inlet chassis. It may be necessary to lift the inlet chassis cover in order to fit the connector ends through the slots. ) shows the routing for a front inlet position. Installation Instructions

Agilent G350A/G35A Multimode Inlet Figure 2 Routing wiring harness through chassis BI connection BV connection FV connection FI connection Thermocouple wire HTR connection Grommet Figure 3 CO2 electrical wire routing (front inlet installed) 2 Installation Instructions

3 Route the thermocouple and heater wires from the inlet down through the GC frame. Again, it may be necessary to lift the inlet chassis cover in order to fit the connector ends through the slots. 4 Route the thermocouple wire across the left side of the GC and up under the split vent line. 5 Plug the thermocouple wire into the thermocouple connector on the Thermocouple board, matching positive and negative connectors. See Figure 4. Wiring harness connection Thermocouple connection Solenoid valve connection Figure 4 CO 2 assembly board connections 6 Plug the heater wire connector into the heater connector labeled HTR on the wiring harness. 7 Route the wiring harness plug labeled PCB across the left side of the GC and up through the channel in the cryo assembly area. 8 Slide the grommet on the left side of the GC out of the sheet metal slot and include all the wires and cables in the grommet. Position the grommet back in the slot. 9 Connect the tubing from the cryo assembly to the open CO 2 tube on the inlet. Installation Instructions 3

Figure 5 Connecting the cryo assembly tubing to the inlet 0 Plug the PCB connector into the front connector on the MMI board. See Figure 4. Complete the installation Secure the MMI inlet into place by tightening the three screws. 2 Secure the inlet chassis onto the GC by tightening the four screws on the inlet chassis. 3 To complete the installation of the liquid CO 2 MMI inlet, skip to Install the EPC module on page 8. 4 Installation Instructions

Install the liquid N 2 MMI inlet module If you are installing the liquid CO 2 MMI inlet module, skip to Install the EPC module on page 8. Prepare the Thermocouple board (liquid N 2 only) If necessary, disconnect all electrical connections on the Thermocouple board. 2 Install the Thermocouple board onto the new board bracket using two screws, as shown below. Thermocouple board Screws (2) Board bracket Figure 6 Thermocouple board assembly ready for installation Position the inlet Rest the EPC module in the pneumatics area. Do not install the EPC module at this time. It will be installed at the end of the procedure. 2 Insert the MMI inlet body into the desired inlet position in the chassis so that the tubing and wiring are running to the left side of the chassis. See Figure 3 and Figure 5 for proper routing. Install the split vent valve and filter assemblies Install the split vent filter into the housing, making sure that the flat sides on the split vent filter are vertical. The split vent filter will lock into place when properly installed. See Figure 0. 2 Align the split vent valve assembly over the two screw holes and secure using two captive screws. 3 Install the split vent filter retaining bracket with one screw to secure the split vent filter. 4 Position the tubing from the split vent assembly to the inlet so that it will not be pinched by the left side cover. Installation Instructions 5

Install the Thermocouple board Install the Thermocouple board in the position next to the split vent assembly and secure using two screws. See Figure 7 for correct bracket and board orientation. PCB connection Thermocouple connection Solenoid valve connection MMI board bracket Figure 7 N 2 assembly board connections Install the cryo assembly and route the MMI electrical wiring Unplug the wiring harness from the inlet heater cable. 2 Route the white connector ends (FI and FV) of the wiring harness up through channels on the GC, plugging them into the appropriate connectors (FI and FV or BI and BV) on the inlet chassis. It may be necessary to lift the inlet chassis cover in order to fit the connector ends through the slots. See Figure 3. 3 Route the cryo line, thermocouple wire, and heater wires from the inlet down through the GC frame. Again, it may be necessary to lift the inlet chassis cover in order to fit the connector ends through the slots. 4 Remove the top round metal cutout on the lower rear cover using the diagonal cutters. Make the cuts so that the metal nubs remain attached to the discarded metal circle. 5 Position the nut plate inside the damper assembly in the rear of the GC. Align nut plate with the screw holes in the left side of the GC and secure with a screw. 6 Installation Instructions

N2 cryo bracket attachment holes Knockout for N2 cryo valve Nut plate secured with captive screw Figure 8 Nut plate installed 6 Slide the nut end of the insulated tubing of the N 2 cryo assembly through the top round cutout on the lower rear cover of the GC. 7 Align the sheet metal plate through holes in the N 2 cryo assembly over the attachment holes in the left side of the GC and secure with three screws. Refer to Figure 8 for the locations of the screw holes. 8 Route the solenoid valve cable from the cryo assembly and the thermocouple wire from the inlet to the MMI board. See Figure 7. 9 Plug the solenoid valve cable into the valve connector on the MMI board. Plug the thermocouple wire into the correct +/- connection on the MMI board. See Figure 7. 0 Route the PCB cable from the wiring harness to the MMI board and plug it into the PCB connector on the board. See Figure 7. Slide the grommet on the left side of the GC out of the sheet metal slot and include the thermocouple wire, solenoid valve cable, and PCB cable in the grommet. Position the grommet back in the slot. See Figure 3. 2 Route the cryo tubing from the inlet through the chassis slot to the cryo valve assembly fitting so that the tubing lies flat against the left side of the GC. See Figure 9. 3 Attach the cryo tubing to the N2 cryo valve. Installation Instructions 7

Cryo tubing routed through slot Figure 9 Connecting the cryo assembly to the inlet 4 Position the sheet metal insulation cover over the insulated tubing of the cryo assembly and secure the cover with two screws. Cryo insulation cover Screws Figure 20 Insulation cover installed Complete the installation Secure the MMI inlet into place by tightening the three screws. 2 Secure the inlet chassis onto the GC by tightening the four screws on the inlet chassis. Install the EPC module Position the EPC module over the selected inlet slot in the pneumatics area. The rear inlet slot is for a rear inlet and the front inlet slot is for a front inlet. 2 Connect the communication cable from the GC to the communication connector on the EPC module, and connect the split vent valve wire from the split vent assembly to the split vent valve connector on the EPC module. 8 Installation Instructions

Split vent valve connector Communication cable connection Figure 2 Split vent valve and communication cable connections on EPC module 3 Slide the EPC module into the slot, being careful not to pinch the wires. 4 Attach the EPC module bracket and secure with a screw. 5 Route the carrier gas tubing from the inlet to the EPC module to the MMI inlet alongside the Inlet chassis. Installation Instructions 9

Install the Insulation Cup Place the insulation in the column nut cup. Foil (Liquid N2 installation only) Insulation Column nut cup Figure 22 Placing the insulation in the column nut cup 2 From inside the oven, install the two heat- resistant screws in the cutouts adjacent to the inlet opening. Do not tighten the screws. Place the column nut cup slots into the screws and tighten the two screws to complete installation. Restore the GC to operating condition Replace all instrument covers in the reverse order that they were removed. 2 Plug in the GC and turn on the power. 3 To unlock the configuration, press [Options], select Keyboard & Display, and press [Enter]. Scroll to Hard Configuration Lock and press [Off/No]. 4 On the GC keypad, press [Configure] then [Front Inlet] or [Back Inlet]. 5 On the unconfigured parameter, press [Mode/Type]. 6 For N 2 or CO 2 Cryo scroll to Install Frnt Inlet (FVCRYO) or Install Back Inlet (BVCRYO). Press [Enter]. A caution message will appear instructing you to reboot. 7 Reboot the GC. a Press [Options]. b Scroll to Communications and press [Enter]. c Scroll to Reboot the GC? and press [On/Yes] twice to reboot the GC and have the changes take effect. 8 Zero the pressure sensors. a On the GC keypad, press [Options]. b c Scroll to Calibration and press [Enter]. Scroll to select the front or back inlet and press [Enter]. Scroll to [Pressure Zero] and press [On/Yes]. 20 Installation Instructions

9 Connect the source gas lines to the EPC module using the provided /8- inch nuts and ferrules. 0 Turn on gas pressure and leak- check all fittings Zero the specific flow sensor. a Press [Options], scroll to Calibration, and press [Enter]. b c d Scroll to the new MMI module to be zeroed and press [Enter]. Scroll to a zero line and press [Info]. The GC will remind you of the conditions necessary for zeroing that specific sensor. Press [On/Yes] to zero or [Clear] to cancel. Installation Instructions 2

Agilent Technologies, Inc. 2009 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. First Edition, May 2009 This information is subject to change without notice. Printed in USA and China Agilent Technologies, Inc. 2850 Centerville Road, Wilmington, DE 9808-60 USA 安捷伦科技 ( 上海 ) 有限公司上海市浦东新区外高桥保税区英伦路 42 号联系电话 :(800)820 32 Manual Part Number G350-9000 Warranty The material contained in this document is provided as is, and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control. G350-9000 22 Installation Instructions