AssayMAP 96AM Wash Station. Maintenance Guide

Similar documents
AssayMAP 96AM Wash Station. Maintenance Guide

Agilent N2763A Rack Mount Kit for 4000 X-Series Oscilloscopes

Agilent N5470A Series Oscilloscope Rack Mount Kit

Agilent 1200 Series Manual Injector

Valve Box and Valve Accessories G4315A, G4316A, G4317A

Agilent 1260 Infinity Manual Injector

Agilent G2738A Upstream Capillary Interface

Agilent G1554A On-Column Capillary Inlet with EPC

Nickel Catalyst Tube on a 6820 GC, Accessory G4318A

Agilent 1200 Series Valves

Repair Manual. 3.5 mm and Type-N Slotless Contact. For use with Slotless Repair Kits

REV F2.0. User's Manual. Hydraulic ABS (HABS) Hydraulic Power Brake (HPB) Page 1 of 28

Automated Liquid Handling Station

Model LA 4100 Time Delay OFF Controller

Model LA 4300 Time Delay OFF Controller

Dynojet Research, Inc. All Rights Reserved. Optical RPM Sensor Installation Guide.

Model LA 4400 Time Delay OFF Controller

OnBoard Drum Major Podium

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Simscape Getting Started Guide. R2014a

Deans Switch Microfluidics

Techcon Systems TS5000 Series Rotary Valve

Manual No: Revision: B. CCVP Rain Shield. Installation Guide

BETTIS SERVICE INSTRUCTIONS DISASSEMBLY AND REASSEMBLY FOR CB-SR-S SEISMIC SPRING RETURN SERIES PNEUMATIC ACTUATORS

SCION FR-S FOG LIGHTS

Dynojet Research, Inc. All Rights Reserved. Air Fuel Ratio Module Installation and User Guide.

Agilent G3510A/G3511A Multimode Inlet

SmarTire TPMS Maintenance Hand Tool. Revision User Manual

EMC VMAX Securing Kit Installation Guide

Two-way cable interlock kit for Magnum drawout circuit breakers

LoadMaxx. Installation Guide. For Trailers with Mechanical Suspensions. Air-Weigh Customer Support: PN R0

Model C230 Pump Controller

Two-way cable interlock kit for Magnum fixed circuit breakers

Patient Care Facility

Flow Module. Instruction Manual. Flow Module

Drug Testing Labs. Style 2 W/ Ver 4.41 pcb & Pulse Relay pcb(s) Installation, Operations, Start-up and Maintenance Instructions

Air Fuel Ratio Module and AFR-4 Pump Assembly Installation and User Guide.

GPS Steering System Installation Manual

Installation instructions

SmarTire TPMS Maintenance Hand Tool. Revision User Manual

Patient Care Facility

Installation Instructions PowerBoard Automatic Retracting Running Board

Drug Testing Labs. Style 2 W/ Ver 4.41 pcb & Pulse Relay pcb(s) Installation, Operations, Start-up and Maintenance Instructions

Adjustable Angled Incline Conveyor Owners Manual with Operating Instructions

SUPER PUMP OUT SYSTEM

Autoplex Specifications

Torque Module Installation and User Guide for model 250i Motorcycle Dynamometers.

Pellicon 2 and 3 Mini Holder. User Guide

RIGID ENTRY BOOT INSTALLATION GUIDE

GPS AutoSteer System Installation Manual

Full Length Bed Rail - Model G29

Troubleshooting. This section outlines procedures for troubleshooting problems with the operation of the system:

XLR-8 60 Extended Rack Autosampler

GPS AutoSteer System Installation Manual

3M No Polish Connector 8800-APC/AS SM SC/APC, Angle Splice, 250/900-µm. Instructions. October C

MetaXpress PowerCore System Installation and User Guide

Preparation Part Number: PT Kit Contents Item # Quantity Reqd. Description 1 1 Rear Spoiler 2 1 Hardware Kit. Hardware Bag Contents

Installation instructions

GPS AutoSteer System Installation Manual

P/N Instr Rev Page 1 of 11

Fabric Replacement Top and Doors with Tinted Side and Rear Curtains Installation Instructions

User s Guide. EZ-Bar Livestock Scale WEIGH BARS AND DIGITAL INDICATOR. WB-40 Weigh Bars TS-700 WB Digital Indicator. Guide version 1.

Manual No: Revision: F. Mag Probe Assembly Manual

INSTALLATION AND OPERATING

Installation instructions

Solera Power Awning TROUBLESHOOTING AND SERVICE MANUAL

LEXUS RC 350/RC-F ILLUMINATED DOOR SILLS Preparation

Installation Manual. AutoSteer. Gleaner Combine. AutoGuide 2 Steer Ready. Supported Models A66 A76 R66 R76 S67 S77 PN: A

SIDE-WIND, A-FRAME TRAILER JACK. Model Due to continuing improvements, actual product may differ slightly from the product described herein.

Service Guide JATCO Environmental Protection Tank Model J-7000

SENTRY ISOBUS Tip Flow Monitor. Software Version 1.00

Dispenser Pan Sensors & Containment Sump Sensors

LU27 Series Quick Start

Lamp House with 2.5 kw lamp

Belt Installation, Tracking, and Maintenance Guide

HOW TO REPLACE THE ICHEMVELOCITY WASTE FILTER AND PERFORM THE WASTE LINE FLUSH FOR BECKMAN UA CALL CENTER USE

GUI Customization with Abaqus. Abaqus 2017

OWNER S MANUAL EVOLUTION 3500, 4500, 5500, & 8500 SERIES PUMPS

p.t.o. Slip clutch Read this material before using this product. Failure to do so can result in serious injury. Save this manual.

Dispenser Pan Sensors & Containment Sump Sensors

LU23, LU28 & LU29 Series Quick Start

INSTALLATION INSTRUCTIONS FORD F-150 2WD & 4WD RETAINS FACTORY TOW HOOKS PART #P3063

P-Series Installation

Geometry Translator User s Guide

Haas Hydraulic Power Unit - Troubleshooting Guide

Periodic Maintenance 7-1

Installation Manual TWM Performance Short Shifter Cobalt SS/SC, SS/TC, HHR SS, Ion Redline and Saab 9-3

Installation instructions

6.0 SPECIFICATIONS CONTENTS. Calibration. According to factory procedureeeeeeeeeeeeeee Accuracy*

Techcon Systems TS7000 Series Interchangeable Material Path Rotary Valve

Bauer NE Lock OWNER'S MANUAL

Installation Guide. Installing the FRK Hydraulic Compact Unit (HCU) with Parking Brake Valve on International School Buses

BETTIS SERVICE INSTRUCTIONS DISASSEMBLY AND REASSEMBLY FOR CB DOUBLE ACTING SERIES PNEUMATIC ACTUATORS

Installation & Set-Up Instructions

Installation instructions

Bimini Top, Bimini Top Plus for Jeep Wrangler (YJ) Vehicles Items #141031XX and #143031XX

Toro Sprayer Calibration Tool

Installation Instructions PowerBoard Automatic Retracting Running Board

SlimRack Bed Lift System OEM INSTALLATION MANUAL

CALIBRATION PROCEDURE NI Channel, ±0.5 V, 24-Bit Simultaneous, Channel-to-Channel Isolated Analog Input Module

Transcription:

AssayMAP 96AM Wash Station Maintenance Guide

Notices Agilent Technologies, Inc. 2015 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. Installation Guide Part Number G5542-90004 Edition Revision A, October 2015 Contact Information Agilent Technologies Inc. Automation Solutions 5301 Stevens Creek Blvd. Santa Clara, CA 95051 USA Technical Support: 1.800.979.4811 (US only) or +1.408.345.8011 (rest of the world) service.automation@agilent.com Customer Service: 1.866.428.9811 or +1.408.345.8356 orders.automation@agilent.com Customer Service by country: http://www.chem.agilent.com/en-us/ Contact-US/Pages/ContactUs.aspx Documentation feedback: documentation.automation@agilent.com Web: www.agilent.com/lifesciences/ automation Acknowledgements Microsoft and Windows are either registered trademarks or trademarks of the Microsoft Corporation in the United States and other countries. Warranty The material contained in this document is provided as is, and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control. Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license. Restricted Rights Legend If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as Commercial computer software as defined in DFAR 252.227-7014 (June 1995), or as a commercial item as defined in FAR 2.101(a) or as Restricted computer software as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies standard commercial license terms, and non-dod Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data. Safety Notices A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.

AssayMAP 96AM Wash Station Maintenance Guide 96AM Wash Station Maintenance Guide This guide contains the following topics: About this guide on page 2 About the 96AM Wash Station on page 4 Required resources and workflow on page 4 Determining if the chimneys are in good condition on page 5 Emptying the 96AM Wash Station on page 6 Uninstalling the 96AM Wash Station on page 10 Replacing damaged chimneys on page 11 Installing the wash station and connecting the tubing on page 12 Testing the 96AM Wash Station on page 17 1

About this guide About this guide Who should read this guide The instructions in this guide assume that you have the following: An understanding of the AssayMAP Bravo Platform Familiarity with the Protein Sample Prep Workbench Experience in working on mechanical equipment What this guide covers This guide describes how to inspect and replace chimneys in the 96AM Wash Station. 96AM Wash Station The 96AM Wash Station is a component of the AssayMAP autofilling station, as the following figure shows. AssayMAP autofilling station, including 96AM Wash Station, Pump Module, tubing, and carboys 2 AssayMAP 96AM Wash Station Maintenance Guide

About this guide Related user guides and where to find them For information on Potential safety hazards and how to avoid them Using the Protein Sample Prep Workbench Setting up, operating, maintaining, and troubleshooting the Bravo Platform Setting teachpoints for the AssayMAP Bravo Platform Setting up the Pump Module and the autofilling reservoirs and wash stations Creating and editing VWorks protocols and application interfaces Running a syringe test of the Bravo 96AM Head Replacing a defective syringe in the Bravo 96AM Head See Bravo Automated Liquid Handling Platform Safety and Installation Guide Literature Library in the Protein Sample Prep Workbench Bravo Automated Liquid Handling Platform User Guide Bravo Automated Liquid Handling Platform User Guide Pump Module User Guide VWorks Automation Control User Guide AssayMAP Syringe Test Utility User Guide in the Protein Sample Prep Workbench AssayMAP Bravo Syringe Replacement User Guide To access the product user guides, do one of the following: Go to the Agilent Automation Solutions Knowledge Base web pages at www.agilent.com/chem/askb and click the link for PDF user guides. From within VWorks software, select Help > Knowledge Base or press F1. From the Windows desktop, select Start > All Programs > Agilent Technologies > VWorks > User Guides > Knowledge Base. Accessing the user documentation You can access the user documentation in the Protein Sample Prep Workbench as follows: Application user guides. In the Protein Sample Prep Workbench, click Utility Library, locate the utility of interest, and then click Instructions. Utility instructions. In the Protein Sample Prep Workbench, click Literature Library, locate the topic of interest, and then click Open. Alternatively, you can access the user documentation in the VWorks Knowledge Base using either of the following methods: On the Windows desktop, click Start > All Programs > Agilent Technologies > VWorks > User Guide > VWorks Knowledge Base. Go to http://www.agilent.com/chem/askb. AssayMAP 96AM Wash Station Maintenance Guide 3

About the 96AM Wash Station About the 96AM Wash Station CAUTION Ensure that you use only the 96AM Wash Station on the AssayMAP Bravo Platform. Using a model of the Tip Wash Station other than the 96AM Wash Station can result in a potential crash with the Bravo 96AM Head. The 96AM Wash Station has 96 chimneys and is scaled to work with the syringe probes in the Bravo 96AM Head. The wash station is used for washing the syringes and washing cartridges or pipette tips that are seated on the syringe probes. Chimney detail in the 96AM Wash Station Fresh wash liquid enters the 96AM Wash Station through two inlet ports (1), shown in the following figure. The wash liquid flows up through the chimneys, and the waste overflows from the chimneys and is removed through two outlet ports (2). Inlet ports (1) and outlet ports (2) in the 96AM Wash Station Required resources and workflow WARNING The probes of the Bravo 96AM Head are sharp and can scratch you if they brush across your hand. A probe scratch can expose you to any contaminants remaining on the probes. Wear gloves and use extreme caution to avoid brushing against the probes. Ensure that you have the following tools and materials: Replacement Wash Station Chimneys: large internal diameter, pack 100, Agilent part number 15586-002 Hex driver to uninstall and install 2- mm hex screws 1 needle- nose pliers to remove the bad chimneys from the 96AM Wash Station Small rubber pad or foam to protect your thumb when seating a new chimney in the 96AM Wash Station. 4 AssayMAP 96AM Wash Station Maintenance Guide

Determining if the chimneys are in good condition Perform the following steps in the order given. Step For this task... See... 1 Inspect the chimneys for damage and irregularities. Determining if the chimneys are in good condition on page 5 2 Empty the wash station. Emptying the 96AM Wash Station on page 6 3 Uninstall the wash station. Uninstalling the 96AM Wash Station on page 10 4 Replace the damaged chimneys. Replacing damaged chimneys on page 11 5 Install the wash station. Installing the wash station and connecting the tubing on page 12 6 Verify the wash station teachpoint and adjust it, as necessary: a In the VWorks software, start the AssayMAP Teachpoint Update protocol. b In the AssayMAP 1 profile, adjust the teachpoint for the 96AM Wash Station. c In the AssayMAP Teachpoint Update protocol, apply the teachpoint update to the remaining profiles in the Protein Sample Prep Workbench automatically. 7 Start up the AssayMAP Bravo Platform and verify that all the chimneys work consistently. Bravo Automated Liquid Handling Platform User Guide Note: Agilent field service can refer to the AssayMAP Bravo Platform Installation Guide. Testing the 96AM Wash Station on page 17 Determining if the chimneys are in good condition During normal operation of the 96AM Wash Station, the fluid flow out of the top of each chimney and the shape of the fluid dome at the top of each chimney must consistent. To ensure consistency in the fluid flow from each chimney, the chimneys must be: Clean and free of any debris. Free of deformities or irregularities. AssayMAP 96AM Wash Station Maintenance Guide 5

Emptying the 96AM Wash Station See the good condition chimneys in the following figure. The chimneys must not be damaged, such as those shown in the figure (possibly from a head collision). Equal in height and installed in the same orientation. Close up views of 96AM Wash Station chimneys Chimney irregularities or damaged chimneys can cause large droplets to cling to the syringe probes, which can introduce variability in the assay. Emptying the 96AM Wash Station This topic describes how to use the Pump Reagent task in Bravo Diagnostics to empty any fluid from the wash station. Before you start WARNING The probes of the Bravo 96AM Head are sharp and can scratch you if they brush across your hand. A probe scratch can expose you to any contaminants remaining on the probes. Wear gloves and use extreme caution to avoid brushing against the probes. WARNING When you initialize the Bravo Platform, the head and tie bar can move. To prevent potential injury, keep clear of the device while it is in motion. CAUTION To prevent potential equipment damage, ensure that the deck is clear of any obstacles before initializing the AssayMAP Bravo Platform. CAUTION Using an incorrect profile can cause equipment damage. Ensure that the profile is correct for the head type and deck configuration. 6 AssayMAP 96AM Wash Station Maintenance Guide

Emptying the 96AM Wash Station Running the Pump Reagent task to empty the wash station To empty the wash station using the Pump Reagent task: 1 At the carboy that contains the wash solution, remove the lid and lift it so that the tubing is above the liquid level. 2 Ensure that the Protein Sample Prep Workbench is closed, and then start the VWorks software. You can use the VWorks desktop icon or the Start menu on the Windows desktop to start the VWorks software. 3 Ensure that Simulation is off. 4 In the VWorks window, open the AssayMAP Bravo device file (AssayMAP Bravo.dev). To do this, click File > Open. In the Open dialog box, ensure the file type is Device Files (*.dev), select the file, and then click Open. The device files are installed in the following default location: C:\VWorks Workspace\Workbench\Device Files 5 Click No when the Would you like to initialize devices now? message appears. 6 In the AssayMAP Bravo.dev tab, click Device diagnostics. The Bravo Diagnostics dialog opens. AssayMAP 96AM Wash Station Maintenance Guide 7

Emptying the 96AM Wash Station AssayMAP Bravo.dev tab and Profile setting for AssayMAP 1 7 Click the Profiles tab, select the Profile name AssayMAP 1, and then click Initialize this profile. 8 If the There appears to be a plate present error message appears, click Ignore and Continue to continue the homing process. 9 When the Verify that it is safe to home the W-axis message appears, click Retry to continue homing the w-axis. 8 AssayMAP 96AM Wash Station Maintenance Guide

Emptying the 96AM Wash Station 10 Click the Processes tab and run the Pump Reagent task as follows: a Select the following: Setting Value Location 1 Labware at selected location 96AM Tip Wash Station Command to execute Pump Reagent Pump Reagent properties Reservoir mode Fill Pump speed (0 100%) 50% Pump on time (1 600 s) 60 seconds How often 1 Run second pump Yes Second pump speed (0 100%) 60% Use weigh station No b Click Execute command. The wash station empties. AssayMAP 96AM Wash Station Maintenance Guide 9

Uninstalling the 96AM Wash Station Uninstalling the 96AM Wash Station WARNING To prevent potential injury, turn off the Bravo Platform before you install or remove any accessory. To uninstall the 96AM Wash Station: 1 On the side of the Bravo Platform, press the power switch to the off (o) position. 2 Gently, move the Bravo head manually to position it over deck location 6, which will provide unobstructed access to the wash station at deck location 1. AssayMAP Bravo Platform with head positioned at deck location 6 3 On the left side as you face the front of the AssayMAP Bravo Platform, loosen the two 2- mm screws using a hex driver. Uninstall the bracket that secures the wash station to the platepad. 96AM Wash Station secured with platepad brackets 4 On the opposite side of the platepad, guide the wash station horizontal slot off the platepad inner bracket. 5 Lift the wash station off of the platepad. For ease of handling, you may disconnect the tubing from the wash station. 10 AssayMAP 96AM Wash Station Maintenance Guide

Replacing damaged chimneys Replacing damaged chimneys To replace the damaged chimneys: 1 Using needle- nose pliers, pull out the defective chimneys from the wash station. IMPORTANT Ensure that you discard any chimneys that you remove. Using pliers on a chimney will permanently damage the chimney. 2 Inspect the wash station hole in which the chimney was seated. Ensure that the hole is empty and clean. If necessary, clean the hole. 3 Orient the new plastic chimney so that the end with the little dot is at the top, and then insert the chimney into the hole in the wash station. Dot on chimney oriented at top end for installation 4 Using your thumb, press down on the chimney so that the chimney touches the bottom surface of the wash station and is fully seated. You can insert a rubber pad between your thumb and the chimney to protect your thumb. AssayMAP 96AM Wash Station Maintenance Guide 11

Installing the wash station and connecting the tubing 5 Ensure that all the chimneys in the wash station are oriented so that the dots are at the top end of the wash station. Ensure that the tops of the chimneys are level with each other. Chimney dots oriented at top and tops of chimneys aligned 6 Repeat steps 1 to 5 to replace any other defective chimneys. Installing the wash station and connecting the tubing Installing the 96AM Wash Station Before you start WARNING To prevent potential injury, turn off the Bravo Platform before you install or remove any accessory. Ensure that you have the following: 96AM Wash Station and the platepad with brackets Pump Module 2.0 Tubing kit with inline filter Hex driver for 2- mm screws Ensure that you install the 96AM Wash Station at deck location 1. 12 AssayMAP 96AM Wash Station Maintenance Guide

Installing the wash station and connecting the tubing 96AM Wash Station installed at deck location 1 To install the wash station: 1 At deck location 1, ensure the bracketed platepad is installed, and ensure the platepad crosshairs are oriented towards the northwest corner of the deck, as the following figure shows. Deck location 1 (top view) showing crosshairs orientation 1 2 3 00240 Bravo Plate Orientation 4 5 6 7 8 9 00240 2 On the platepad side that is next to the crosshairs, use a hex driver to uninstall the two 2- mm screws, and then remove the bracket. Note: The wash station is easier to install if you remove the outer bracket, and then replace the bracket after the wash station is in position. AssayMAP 96AM Wash Station Maintenance Guide 13

Installing the wash station and connecting the tubing Bracket next to platepad crosshairs 3 Install the wash station on the platepad as follows: a b c d Ensuring that the inlet and outlet ports are towards the rear of the deck, place the wash station on the platepad. Guide the platepad s inner bracket into the wash station s horizontal slot. Re- install the bracket that you removed in step 2 on the outer side of the platepad. Ensure that you guide the bracket into the corresponding horizontal slot on the wash station. Ensure that both brackets remain engaged as you tighten the bracket screws. Ensure that the wash station sits level on the platepad and that both brackets are holding the wash station securely in place. 96AM Wash Station secured with platepad brackets. Tubing connections for 96AM Wash Station The following figure shows the AssayMAP Bravo tubing connections for the 96AM Wash Station. As the figure shows, the upper pump (A) is for the fill line and the lower pump (B) is for the empty line. 14 AssayMAP 96AM Wash Station Maintenance Guide

Installing the wash station and connecting the tubing 96AM Wash Station fill line (A) and empty line (B) A B To ensure proper chimney filling and waste emptying characteristics for the 96AM Wash Station, ensure that you use the proper tubing combination for the Pump Module. The following table gives two recommendations: one for conservative buffer usage, and one for inducing higher volume flow through the chimneys. Fill-and-empty characteristics High volume flow in chimneys (recommended) Conservative buffer usage Tubing recommendation Fill line (A) Marprene tubing, 6.4 mm (1/4- in) ID, 1.6 wt Emptying line (B) Marprene tubing, 8 mm (5/16- in) ID, 1.6 wt The 5/16- in diameter tubing is too large for the quickdisconnect fittings and the wash station ports. You use this larger tubing only in the pump head and splice it to the 1/4- in tubing to enable connections with the other fittings and wash station ports. See the following figure and table for a tubing example and a description of the tubing components. Fill line (A) Marprene tubing, 4.8 mm (3/16- in) ID, 1.6 wt Emptying line (B) Marprene tubing, 6.4 mm (1/4- in) ID, 1.6 wt AssayMAP 96AM Wash Station Maintenance Guide 15

Installing the wash station and connecting the tubing Tubing example and components for high volume flow in the chimneys 6 7 8 1 2 4 5 3 3 5 4 6 4 8 Item Component Description 1 Source bottle Supplies the fill liquid for the wash station. 2 Waste bottle Collects the waste liquid that is pumped from the reservoir. 3 Tubing, 8- mm (5/16- in) For high volume flow in the chimneys, use the 5/16- in diameter Marprene tubing for the length of tubing in the pump head. 4 Tubing, 6.4- mm (1/4- in) The 1.4- in Marprene tubing is for the input and output tubing connections to the wash station and the source and waste bottles. The 5/16- in diameter tubing is too large for the quickdisconnect fittings and the wash station ports. So, you use a union connector (5) to splice the 5/16- in tubing to the 1/4- in tubing. 5 Connector, union Joins two sizes of tubing, such as the 5/16- in tubing to the 1/4- in tubing. 6 Quick- disconnect fitting Enables easy removal of the wash station. The quick- disconnect fittings include an automatic- close valve. Fluid can flow only if a positive connection is made. 7 Inline pump filter Removes the particulates that can clog the chimneys in the wash station. 8 3- way connector Enables one tube to branch into two tubes at the input and output ports on the wash station. 16 AssayMAP 96AM Wash Station Maintenance Guide

Testing the 96AM Wash Station Testing the 96AM Wash Station Before you start CAUTION Any time that you uninstall and reinstall the wash station, the wash station teachpoint can shift. To prevent a potential crash, verify the wash station teachpoint in all the profiles any time you reinstall the wash station. For the teachpoint verification workflow, see Required resources and workflow on page 4. Starting up the AssayMAP Bravo Platform To test the performance of the wash station, you run the Startup utility in the Protein Sample Prep Workbench. During the Startup protocol, carefully observe the fluid as it flows out of the top of the chimneys in the wash station. Ensure that the fluid flow out of the top of each chimney and the shape of the fluid dome at the top of each chimney is consistent. If fluid does not flow out of a chimney, run the Startup protocol again to remove any bubble that might be blocking the flow. If a fluid dome shape is different than the others, carefully examine the top of the chimney for irregularities. Note: Chimney irregularities can result from a collision between the head and the wash station. To start up the system: 1 Check the levels of the wash station source and waste carboys, and fill or empty as required. 2 Turn on the AssayMAP Bravo Platform, Pump Module, and the Peltier Thermal Station Controller, if included. 3 Start the Protein Sample Preparation Workbench, and open the Utility Library. 4 Open the System Startup/Shutdown utility. 5 Click Run Startup to initialize the AssayMAP Bravo Platform and accessories. WARNING When you initialize the Bravo Platform, the head and tie bar can move. To prevent injury, keep clear of the device while it is in motion. 6 During the Startup protocol, verify that all the wash station chimneys have liquid flowing through them. AssayMAP 96AM Wash Station Maintenance Guide 17

Testing the 96AM Wash Station Troubleshooting problems with the wash station Locate your problem in the following table and try the solution. If the problem persists after you try the solutions, contact Automation Solutions Technical Support. Problem Potential cause Solution Liquid flow is constricted or insufficient in all the chimneys of the wash station. Liquid is not flowing through a subset of the chimneys in the wash station. Waste is accumulating in the wash station. Tubing creeps through the pump head instead of being secured in place. The tubing is crushed or damaged in the pump head. The tubing is kinked or blocked, or the pump head clamp is too tight for the tube size. The inline filter is clogged. The tubing diameter of the fill line is too small. The chimneys are blocked or damaged. The tubing is kinked or blocked, or the pump head clamp is too tight for the tube size. The tubing diameter is too small for the required flow. The pump head clamp is too loose for the tube size. The pump head clamp is too tight for the tube size. Ensure the tubing is not kinked or blocked. Replace the tubing as necessary. If necessary adjust the pump head clamp, see the Agilent Pump Module User Guide. Replace the inline filter. See Tubing connections for 96AM Wash Station on page 14. Replace the tubing on the fill line using larger diameter tubing. See Tubing connections for 96AM Wash Station on page 14. To resolve the problem chimneys: 1 Use a plastic Pasteur pipette to quickly aspirate liquid through the chimneys. 2 Run the Pump Reagent task in Bravo Diagnostics to pump liquid through the chimneys at a maximum rate to dislodge any clogs. 3 If the problem persists, replace the problematic chimneys. See Replacing damaged chimneys on page 11 Ensure the tubing is not kinked or blocked. Replace the tubing as necessary. If necessary adjust the pump head clamp, see the Agilent Pump Module User Guide. Replace the tubing on the empty line, using larger diameter tubing for the section that routes through the pump head (active section). See Tubing connections for 96AM Wash Station on page 14. Adjust the tubing clamp for the tube size. See the Agilent Pump Module User Guide. Adjust the tubing clamp for the tube size. See the Agilent Pump Module User Guide. 18 AssayMAP 96AM Wash Station Maintenance Guide

Agilent Technologies Maintenance Guide G5542-90004 Revision A, October 2015