TECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure

Similar documents
The Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over

Boosting Trace Detection Performance with the Vanquish Diode Array Detector and High-Sensitivity LightPipe Flow Cell

Technical Overview. Introduction

HIGH PRESSURE SWITCHING VALVES

Agilent 1290 Infinity II Ultra Low Dispersion Kit

QuickSplit TM Flow Splitters

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve

HyperShear TM HPLC and UHPLC Mixers

Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution

HyperShear TM HPLC and UHPLC Mixers

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

Application Note. General Use Inlet Solvent Filters. Inlet Solvent Filters. Filters and Degassers. New!

S 1125 HPLC Pump System

Online sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve

HPLC SYSTEM S 500 SERIES

Fixed Flow Splitter *

HyperShear TM HPLC Mixers

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC

NEW HPLC - SYSTEM LINE

Thermo Scientific Dionex UltiMate 3000 Pumps

HPLC System S 600 Series

Performance Characteristics of the Agilent 1260 Infinity II Bio-Inert LC

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell

HyperShear TM HPLC Mixers

STI OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

C196-E081C. Ultra High Performance Liquid Chromatograph Nexera X2. Specifications

Ultra High Performance Fittings. Fittings Connectors Kits

Filters & Frits. IDEX Health & Science introduces Upchurch Scientific precolumn and inline very high pressure (VHP) filters on page 159

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

Agilent G5632 Valve Kit Instructions

Using the PSD for Backflushing on the Agilent 8890 GC System

Installation of the 2 position/10 port valve into the 1100 series thermostatted column compartment

Determination of Sudan Dyes I IV in Curry Paste

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

High-throughput protein aggregate analysis of monoclonal antibodies using a novel dual-channel UHPLC instrument

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Application Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC

Setting up SilFlow for BackFlush in your GC

VERITY Purification Systems Build the perfect system to fit your research

Adjustable Flow Splitter *

IDEX Health & Science introduces Upchurch Scientific precolumn and inline very high pressure (VHP) filters on page 159

ACQUITY UPLC I-Class System with 2D Technology

System Manual. Agilent 1260 Infinity Capillary LC System

Simplify your search for. You will find both PEEK-lined stainless steel hardware as well as standard ss column hardware at the below stockingstations.

Analyte Focusing in Elevated Temperature HPLC

Agilent 1100 Series Capillary LC System. System Manual

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

Technical Procedure for Gas Chromatography (GC-FID)

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE

Application Note. Abstract. Authors. Environmental Analysis

Agilent InfinityLab Quick Change Valve G4237A

Chromatography. Liquid. HPLC Accessories SupelPRO Valves SUPELCO

Application. Gas Chromatography June 1995

Agilent 1290 Infinity Autosampler

Agilent Valve Kit G4231A/B and G4232A/B/C

Liquid Chromatography

ASTM D for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software

Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems

Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D

MET-Biodiesel Capillary GC Columns

Agilent G4239C Valve Kit - Instructions

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter

HPLC Tubing and Fittings Hints & Tips for the Chromatographer

GlycoChrom TM HPLC Analyzer. Hardware Installation and System Test

Agilent InfinityLab LC Purification Solutions SELECTION GUIDE

PATROL UPLC Process Analysis System for Production

High-Temperature Simulated Distillation System Based on the 6890N GC Application

Stability Axial Compression Technology Application Ordering Information

Application Note. Authors. Abstract. Energy & Chemicals

Thermo Scientific Dionex UltiMate 3000 Thermostatted Column Compartment

QuickSplit TM Adjustable Flow Splitters Analytical Splitters Analytical range, 0.1 ml/min. to 5 ml/min. input flow, calibration flow rate 1.0 ml/min.

Agilent 1200 Series Thermostatted Column Compartment G1316A/G1316B/G1316C

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method

Agilent 1200 Series Valves

Method Development for Capillary GC Systems. Slide 1

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

Agilent Valve Kit G4231A/B/C and G4232B/C/D

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

AppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS

Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology

HPLC Model 210D. UV detector lamp switch. Rheodyne 7725i injection valve. Prime/purge valve. Column. compartment. thumbscrews. Storage.

ÄKTA system Training Guide. Purifier 100 / 10. amersham bio science

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Agilent Multimode Inlet for Gas Chromatography

PRODUCT MANUAL. CARBONATE REMOVAL DEVICE 200 CRD 200 (4 mm) (P/N ) CRD 200 (2 mm) (P/N )

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

STABILITY AND RELIABILITY: NEW APPROACHES IN PREPARATIVE HPLC COLUMN DESIGN

Dionex CRD Carbonate Removal Device

Instrumentation of HPLC. HPLC Solvent Pumping Systems

Combustion ion chromatography with a Dionex Integrion HPIC System

ACQUITY UPLC I-Class System Specifications

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

The Only No Compromise

Transcription:

TECHNICAL REPORT Modifying Agilent 1100 HPLC Systems to Achieve UHPLC-like performance with HALO Fused-Core Columns Agilent 1100 HPLC system With a few modifications, an Agilent 1100 HPLC can produce UHPLC-like performance using HALO Fused-Core columns. Introduction The Agilent 1100 HPLC systems are found in many labs, and their performance in their as-shipped configuration is excellent when used with conventional HPLC columns. However, the recent introduction of columns packed with Fused- Core particles offers Agilent 1100 users the opportunity to obtain even higher performance from their existing HPLC equipment. HALO columns packed with Fused-Core particles deliver resolution and speed similar to sub--µm UHPLC columns, but generate only 40% to 50% of the back pressure produced by sub--µm columns. (See Figure 1.) The high efficiency and lower back pressure of HALO columns make it possible to obtain UHPLC-like performance from conventional 400-bar equipment, like the Agilent 1100. HALO columns, like their sub--µm column counterparts, produce smaller peak volumes (50% to 75% smaller) than conventional columns of the same dimensions. Because of the smaller peak volumes, it is important to keep extra column band broadening appropriate for the particular column geometry used. How your Agilent 1100 instrument is configured, and the method settings you choose will determine the amount of extra column band broadening, and, ultimately the efficiency and resolution you can obtain with a given HALO column geometry. Many HALO columns can be used quite effectively with the Agilent 1100 in its standard configuration, but much more performance can be obtained by modifying the equipment and settings for optimum results. For example, a 4.6 x 100 mm HALO column can reduce analysis time by 67% compared to a 4.6 x 50 mm, 5 µm conventional column using a standard Agilent 1100 quaternary system under identical conditions. However, by making a few modifications to the Agilent 1100, the same HALO column will yield 46% more plates. The moderate back pressure and high flow velocity characteristics of HALO columns facilitate a 85% reduction in analysis time compared to a conventional column, while maintaining resolution. (see Figure ). Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure 1.7 µm 1.8 µm. µm.7 µm Fused-Core 3.4 3 µm 3.5 µm 5 µm 1.8 5. 7.7 8.7 0 4 6 8 10 RELATIVE BACK PRESSURE In HPLC, peak volume decreases as column ID or length decreases. Small peak volumes are more vulnerable to extra column band broadening, thus preventing maximum column efficiency from being achieved. While very good results can be obtained with many of the larger volume HALO columns using the Agilent 1100 in its standard configuration, you will likely find it worthwhile to reduce the amount of extra column volume and adjust detector settings to achieve much better results with some of the smaller volume HALO columns. In this technical report we first discuss extra column band broadening and the instrument parameters that contribute to it. Next, we recommend how an Agilent 1100 can be configured to get the most performance from specific HALO columns. Finally, we provide step-by-step instructions for making those recommended modifications to the Agilent 1100. 5 µm 3.5 µm 3 µm. µm Sub- µm HALO 0 50k 100k 150k 00k 50k EFFICIENCY: N/METER HALO columns packed with Fused-Core particles provide over 80% of the efficiency (theoretical plates, N) and 90% of the resolving power of sub--µm columns, but require less than half the back pressure. This lower pressure permits HALO columns to be used with conventional 400 bar-limit HPLC equipment and achieve speed and resolution very similar to UHPLC.

Figure : An Agilent 1100 modified to reduce extra column band broadening permits significantly better performance to be obtained from HALO columns. Standard configured Agilent 1100 Conventional column: 4.6 x 50 mm, 5 μm Flow rate: 1.0 ml/min Pressure: 91 bar R s = 6.5 0 minutes 1 Standard configured Agilent 1100 HALO column: 4.6 x 100 mm Flow rate: 1.0 ml/min Pressure: 14 bar R s = 5. R s = 6.3 The HALO column delivers a much faster separation compared to a conventional column when using a standard configured Agilent 1100, but the resolution is less than expected. When the Agilent 1100 is modified to reduce extra column band broadening, 46% more plates and higher resolution are obtained from the same HALO column. The Fused-Core particles packed in the HALO column permit a higher flow velocity to be used without sacrificing resolution or exceeding the pressure limit of the Agilent 1100. The result is a reduction in analysis time of 85% while maintaining similar resolution compared to the conventional column. Extra Column Band Broadening N = 5,900 N = 16,700 0 minutes 4 Agilent 1100 modified to reduce extra column band broadening HALO column: 4.6 x 100 mm Flow rate:.0 ml/min Pressure: 47 bar N = 4,500 0 minutes In liquid chromatography the observed peak width (w obs ) is comprised of the dispersion (or band broadening) from the chromatographic process itself (w col ) and from dispersion elsewhere in the HPLC system (w ec ). The relationship can be described by this equation: w obs =w col +w ec The observed, column-related, and extra columnrelated variances are related similarly, σ = σ +σ obs col ec where w = 4σ, σ is the standard deviation of the peak and σ is the peak variance. The total extra column variance (σ ec ) is equal to the sum of the variances from the various contributors to band broadening (assuming all variables are independent): σ ec = σ injection + σ tubing + σ flow cell + variance from detector response time and data rate The importance of extra column band broadening can be assessed relative to the peak width or volume at the retention time or volume for each analyte. In general, the contributions to band broadening and peak volume are inversely proportional to the retention time or volume of the analyte, so that extra column band broadening is most important for the least retained analytes in a separation. For a given HPLC system configuration, the extra column volume, or ECV, is the total volume outside of the column that contributes to the observed peak volume. For an Agilent 1100 the ECV includes the volumes associated with the following items: the injection volume, the flow path in the autosampler and valve, the tubing from the autosampler to the precolumn heat exchanger, the precolumn heat exchanger, the tubing from precolumn heat exchanger to the column, the tubing from the column to the detector flow cell, the detector flow cell, and any volume added by in-line filters, unions, guard columns, etc. NOTE: For simplicity, any contributions to band broadening due to fittings and connections are considered to be negligible with proper choice of hardware and good practices, and will be ignored. The total extra column band broadening also includes the dispersion due to electronic signal filtering from the detector. With Agilent UV-VIS detectors the ChemStation software Peak Width setting is directly related to detector response time, and inversely related to data acquisition rate for a particular detector type and model. Typically, you can find the relationships among Peak Width setting, response time, and data rate summarized in a table in the appropriate detector manual. The larger the Peak Width setting is, the slower the data rate, and the greater the amount of signal filtering. Conversely, smaller Peak Width settings are associated with faster data rates, less signal filtering, and more short-term noise. A general recommendation is to

Figure 3: Extra Column Volume, ECV Pump Sample Injection Flow Cell Column Detector TABLE A: Standard configured Agilent 1100 (as typically shipped from Agilent) Acceptable configuration for the following HALO columns: 4.6 x 150 mm, 4.6 x 100 mm, and 4.6 x 75 mm. Description Part No. Volume (μl) Connecting Tubing and Fittings Waste Needle seat capillary for standard autosampler 0.17 x 100 mm, Green Or, needle seat capillary for well-plate autosampler 0.17 x 100 mm, Green G1313-87101.5 G1367-8730.5 Solvent Reservoir Data Collection In this schematic of an HPLC system, the gold color indicates where extra column volume, ECV, is located. Stainless steel tubing from autosampler to column compartment heat exchanger 0.17 x 180 mm, Green Column compartment heat exchanger 3 ul Stainless steel tubing from column compartment heat exchanger to column 0.17 x 90 mm, Green Stainless steel tubing from column to detector flow cell 0.17 x 380 mm, Green G1313-87305 4.5 3.0 G1316-87300. G1315-87311 9.4 set the detector Peak Width setting to be equal to or slightly less than the half-height width of the narrowest peak of interest in the chromatogram. In practice, it is wise to compare performance (peak widths, tailing factors, signal-to-noise ratios) from chromatograms acquired with several detector Peak Width settings above and below the value for the narrowest peak, in order to choose the best setting for your needs. If the peak width setting is too small, the observed peak widths will be minimized and peak heights maximized, but the signal-to-noise ratio will suffer due to increased baseline noise. If the peak width setting is too large, baseline noise will be reduced, but the peaks will be shorter and broader and may begin to tail and merge. Several report styles (Performance, Performance and Noise, and Extended Performance) within the Data Analysis View of the ChemStation software can help you find an appropriate or optimum setting for your column and conditions. Fortunately, the data rate is automatically set when you choose the Peak Width setting as just described. Nevertheless, a good rule of thumb is to ensure that the data rate is sufficiently fast to acquire at least 0 points across the narrowest peak. As an example, for a peak width at half height of 0.05 min., the 4σ baseline width (1.7 X PW ½ ) is 5.1 sec., and the data rate should be at least 0 points per 5.1 seconds, or ~ 4 Hz. The detector flow cell (volume and design), the detector response time and the data rate are typically the largest contributors to band broadening. These are the parameters you should optimize first when trying to get the most performance from HALO columns. If you are using HALO columns with ID 3.0 mm, you may also have to minimize connecting tubing volume, heat exchanger volume, and injection volume. Recommended Instrument Configurations to Use with Specific HALO Columns The smaller the peak volume produced by a column, the more important it is to make sure that extra column volume (ECV) does not rob you of the resolving power that the column can deliver. Depending on the dimensions of the HALO column you wish to use, three different configurations of the Agilent 1100 will be recommended; the standard configuration (as shipped from the Standard flow cell for VWD 10 mm, 14 μl, 400 bar Standard flow cell for DAD/MWD 10 mm, 13 µl, 10 bar TABLE B: Low ECV configured Agilent 1100 G1314-6008 14.0 G1315-600 13.0 Total ECV 34.6/35.6 Acceptable configuration for the following HALO columns: 4.6 x 150 mm, 4.6 x 100 mm, 4.6 x 75 mm, 4.6 x 50 mm, 4.6 x 30 mm, 3.0 x 150 mm, 3.0 x 100 mm, and 3.0 x 75 mm. Description Part No. Volume (μl) Needle seat capillary for standard autosampler 0.1 x 100 mm, Red Or, needle seat capillary for well-plate autosampler 0.1 x 150 mm, Red Stainless steel tubing from autosampler to column compartment heat exchanger 0.1 x 180 mm, Red Column compartment heat exchanger 3 ul Stainless steel tubing from column compartment heat exchanger to column 0.1 x 70 mm, Red Stainless steel tubing from column to detector flow cell 0.1 x 180 mm, Red Alternate longer length stainless steel tubing from column to detector flow cell 0.1 x 80 mm, Red Semi-micro flow cell for VWD 6 mm, 5 μl, 10 bar Semi-micro flow cell for DAD/MWD 6 mm, 5 μl, 10 bar G1313-87103 1.3 G1367-87303 1.9 G1313-87304.3 3.0 G1316-87303 0.9 G1313-87304.3 01090-87610 3.6 G1314-60083 5.0 G1315-60011 5.0 Total ECV 14.8/16.7 3

TABLE C: Ultra-low ECV configured Agilent 1100 Acceptable configuration for the following HALO columns: 4.6 x 150 mm, 4.6 x 100 mm, 4.6 x 75 mm, 4.6 x 50 mm, 4.6 x 30 mm, 3.0 x 150 mm, 3.0 x 100 mm, and 3.0 x 75 mm, 3.0 x 50 mm, 3.0 x 30 mm,.1 x 150 mm,.1 x 100 mm,.1 x 75 mm,.1 x 50 mm,.1 x 30 mm. Description Part No. Volume (μl) Needle seat capillary for standard autosampler 0.1 x 100 mm, Red Or, needle seat capillary for well-plate autosampler 0.1 x 150 mm, Red Stainless steel tubing from autosampler to low volume heat exchanger 0.1 x 180 mm, Red Low volume heat exchanger 1.6 ul (includes connecting tubing to column) Stainless steel tubing from column to detector flow cell 0.1 x 180 mm, Red Alternate longer length stainless steel tubing from column to detector flow cell 0.1 x 80 mm, Red Micro flow cell for VWD 1 μl, 6 mm 40 bar Micro flow cell for DAD/MWD μl, 3 mm 40 bar TABLE D: Recommended Agilent 1100 configurations to use with specific HALO columns HALO Column Dimensions 4.6 x 150 mm Standard, Low-ECV, Ultra-low ECV 4.6 x 100 mm Standard, Low-ECV, Ultra-low ECV 4.6 x 75 mm Standard, Low-ECV, Ultra-low ECV 4.6 x 50 mm Low-ECV, Ultra-low ECV 4.6 x 30 mm Low-ECV, Ultra-low ECV 3.0 x 150 mm Low-ECV, Ultra-low ECV 3.0 x 100 mm Low-ECV, Ultra-low ECV 3.0 x 75 mm Low-ECV, Ultra-low ECV 3.0 x 50 mm Ultra-low ECV 3.0 x 30 mm Ultra-low ECV.1 x 150 mm Ultra-low ECV.1 x 100 mm Ultra-low ECV.1 x 75 mm Ultra-low ECV.1 x 50 mm Ultra-low ECV.1 x 30 mm Ultra-low ECV G1313-87103 1.3 G1367-87303 1.9 G1313-87304.3 G1316-8000 1.6 G1313-87304.3 01090-87610 3.6 G1314-60081 1.0 G1315-6004.0 Total ECV 8.5/11.4 Notes: The ChemStation peak width setting is optimized best by carrying out several runs under method conditions using different settings. One can choose the setting that gives the best signal to noise for all analytes, consistent with observed peak width and USP tailing factors using Performance and Noise and Extended Performance reports from the Data Analysis view of the Agilent ChemStation software In the autosampler, replacing the 0.17 x 100 mm needle seat capillary (Agilent part number G1313-87101,.5 µl) with the 0.1 x 100 mm seat capillary (Agilent part number G1313-87103, 1.1 µl) will reduce ECV and improve system efficiency slightly. Agilent 1100 Configuration factory), a low ECV configuration, and an ultra-low ECV configuration. See Tables A, B and C for the three different Agilent 1100 configurations we recommend for specific HALO column dimensions. Modifying Your Agilent 1100 System Modifying your Agilent 1100 system from the standard configuration to either the low ECV or ultra-low ECV configuration is quite simple, once you have the parts on hand. The Agilent part numbers needed to make the modifications are listed in Tables A, B and C. Substitution of similar sized capillary tubing from a source other than Agilent is generally okay, but we recommend that the replacement low volume heat exchanger and detector flow cells come from Agilent to ensure compatibility. Step 1: Replace the standard capillary tubing (0.17 x 180 mm, green, G1313-87305) connecting the autosampler valve to the heat exchanger in the column compartment with the low volume capillary tubing (0.1 x 180 mm, red, G1313-87304). For ultra-low ECV configuration connect the outlet of the capillary tubing to the inlet union of the low volume heat exchanger (G1316-8000). Position the low volume heat exchanger in one of the wells of the column compartment heat exchanger block, and secure it using one of the column clips as needed. See Figure 4. Step : Replace the standard capillary tubing (0.17 x 90 mm, green, G1316-87300) connecting the heat exchanger to the column with the low volume capillary tubing (0.1 x 70 mm, red, G1316-87303). For the ultra-low ECV configuration, connect the outlet tubing of the low volume heat exchanger (G1316-8000) directly to the column inlet using an appropriate stainless steel, PEEK or polyketone fitting. See Figure 4. Step 3: Replace the standard capillary tubing (0.17 x 380 mm, green, G1315-87311) connecting the column to the detector flow cell with the low volume capillary tubing (0.1 x 180 mm, red, G1313-87304) or the alternate longer capillary tubing (0.1 x 80 mm, red, 01090-87610). See Figure 4. Note: Do not connect the tubing to the detector until the column and tubing have been flushed. Step 4: Prime the instrument and flush the tubing and column with either 100% methanol or acetonitrile. Some people find they achieve faster equilibration by first flushing the capillary tubing before connecting the column and flushing it. Step 5: Replace the standard detector flow cell with either a semimicro flow cell (low ECV) or a micro flow cell (ultra-low ECV). See Figure 5. Step 6: Connect the capillary tubing coming from the outlet of the HALO column to the detector flow cell. For diode-array and multiwavelength detectors the capillary tubing is normally connected to a zero dead volume union, which then is connected to the capillary tubing connected into the flow cell body. For.1 mm HALO columns in shorter lengths (30, 50, 75 mm) it is beneficial to connect the column directly to the capillary inlet tubing of the flow cell to minimize extracolumn volume and extra connections. It may be necessary to stack the column compartment directly above the DAD/MWD detector in such situations. Step 7: Pump 100% methanol or acetonitrile through the tubing, column and the detector flow cell for ~ 5 minutes or until you observe stable pressure and baseline. You should observe a flat baseline at 5-50 maufs with a UV detector, and can now proceed to equilibrate your column with your mobile phase and begin running samples. 4

Figure 4: Replace standard capillary tubing (0.17 mm, green) with low volume capillary tubing (0.1 mm, red). Step 1: Replace capillary tubing from sample injector valve to heat exchanger (3 μl) in the column compartment. Step : Replace capillary tubing connecting the heat exchanger to the column. Step 3: Replace capillary tubing connecting the column to the detector flow cell. band after injection and before the column is usually minimized or eliminated, if the injection volume is not too large and the starting mobile phase is not too strong. However, it continues to be important to minimize extra column band broadening after the column (tubing and flow cell volume, and Chemstation Peak Width setting) for gradient separations when using low-volume, high efficiency HALO columns. If you plan to only run gradient separations with your Agilent 1100, you may be able to skip Steps 1 and when modifying the equipment. For gradient separations the delay volume is also important because it imposes an isocratic hold to the start of the gradient program, such that the desired mobile phase composition reaches the column at a time later than programmed by a factor of V d /F, where V d is the delay volume and F is the volumetric flow rate. For a discussion of the translation of gradient methods please see our Quick Tips Guide http://www.mac-mod.com/qtpdf. Conclusion Figure 5: Replace standard detector flow cell with either a semi-micro flow cell (low ECV configuration) or a micro flow cell (ultra-low ECV configuration). Step 5: Remove standard detector flow cell and replace with a semi-micro flow cell or a micro flow cell. Figure 6 illustrates the benefits of modifying an Agilent 1100 to an ultra-low ECV configuration. Compared to a conventional column, analysis time is reduced by over 90% while resolution is maintained. This performance compares favorably with UHPLC, but is accomplished with a conventional Agilent 1100 quaternary system by making a few simple, and inexpensive, modifications to the equipment. Figure 6: An ultra-low ECV configured Agilent 1100 using a HALO column provides performance that compares favorably to UHPLC systems. Gradient Separations All of the considerations and recommendations about reducing extra column band broadening which have been made thus far in this technical report have dealt with isocratic separations. If you are developing a new gradient method or translating an existing gradient method to a HALO column, the most important parameters are the same ones that have been discussed for isocratic separations. However, there are two major differences. 1. The extra column volume before the column is less important than the extra column volume after the column when using a gradient, and,. the delay volume (system volume from the point where solvents A and B are mixed up to the column itself ) is important because it is a major contributor to the hold time at the beginning of a gradient. Extra column volume before the column is less important for gradient separations, because there will be a refocusing of the analytes at the head of the column (unless pre-elution of polar analytes occurs) in a gradient separation. Consequently, any dispersion of the sample Conditions Column: HALO C18, 3.0 x 50 mm Mobile Phase: 35% acetonitrile to 90% acetonitrile in 1 minute Flow Rate:.0 ml/min. Column Temperature: 40º C Pressure: 304 bar > 167 bar Response Time: < 0.15 sec 1 3 4 Sample: 1. Acetophenone. Propiophenone 3. Butyrophenone 4. Benzophenone 5. Valerophenone 6. Hexanophenone 7. Heptanophenone 8. Octanophenone 1.5 minutes 7 8 0 minutes An ultra-low ECV configured Agilent 1100 quarternary system using a high efficiency HALO column (Fused-Core particles) separates these 8 alkylphenones in 90 seconds with ample resolution between all peaks. This speed and resolution compare favorably to the performance reported by chromatographers using UHPLC equipment and sub- μm columns. The pressure generated by the HALO column is well within the acceptable operating limits of the Agilent 1100. 5 6 5

103 Commons Court, Chadds Ford, PA 19317 Toll Free: 800-441-7508 Phone: 610-358-9696 Fax: 610-358-5993 info@mac-mod.com www.mac-mod.com LC501 HALO and Fused-Core are registered trademarks of Advanced Materials Technology, Inc. Agilent and ChemStation are registered trademarks of Agilent Technologies, Inc.