StepWave ion guide: cleaning instructions addendum

Similar documents
ACQUITY QDa Detector Overview and Maintenance Guide

Addendum to the Waters Micromass Q-Tof Premier Mass Spectrometer Operator s Guide

DART-Waters XEVO Interface Manual

CAUTION Fuel Lift Pump. Initial Check. Resistance Check.

Passive Check Valve Installation Instructions for H-Class Systems / Revision A Copyright Waters Corporation 2014 All rights reserved

TCU Static Mixer Installation

CROSBY SERIES 800 AND 900 OMNI-TRIM PRESSURE RELIEF VALVES INSTALLATION AND MAINTENANCE INSTRUCTIONS

Spectrum VT Inline Powder Feed Pump

Reasons for reissue of this instruction sheet are provided in Section 8, REVISION SUMMARY.

Instruction Sheet Simplex and Duplex. LightCrimp Plus LC. Fiber Optic Connector Kits Feb 2017 Rev E

PREVIOUS GENERATION. Encore Manual Powder Spray Gun Trigger Switch Replacement. Introduction. Preparation. Gun Disassembly

INSTALLATION INSTRUCTIONS South Highway 11 Westminster, SC Toll Free (888) (864) FAX (864)

Sure Coat 90 Degree Spray Extension

Disassembly and Reassembly for CBB (Double Acting) Series Pneumatic Actuators

I-317. AWWA Check Valves WARNING INSTALLATION AND MAINTENANCE INSTRUCTIONS SERIES 317 WARNING

Encore Manual Powder Spray Gun Trigger Switch Replacement

Installing the Dionex ICS-1100/1600/2100 Auxiliary Valve

Install. (/qs3/pubsys2/xml/en/manual/ / titlepage.html) Immerse the piston in a container of clean 15W-40 oil.

Additions, Revisions, or Updates

N) manual. Register your instrument! Eppendorf Reference 2. Operating manual

Waters Xevo TQ-S IVD Mass Spectrometry System Operator s Overview and Maintenance Guide

Oreck Magnesium Series Service Manual. The Oreck Manufacturing Company

COOPER POWER. SERIES CMU Medium voltage power fuse installation instructions. Fusing Equipment MN132032EN

SAFETY THIS PRODUCT IS FOR OFFROAD USE ONLY. ALL LIABILITY FOR INSTALLATION AND USE RESTS WITH THE OWNER.

Installation, Operation and Maintenance Instructions Series 608 Ball Valve

E-200 O-rings Service Kit

2003 Dodge Pickup R DRIVE AXLES' 'Axle Shafts - Front - Ram Pickup WD DRIVE AXLES

N) manual. Register your instrument! Eppendorf Research plus. Operating manual

Register your instrument! Eppendorf Reference 2. Operating manual

Pressure Relief Valve Maintenance Manual

SD Bendix DD-3 & SD-3 Safety Actuators PUSH PLATE & SHAFT ASSY. LOCKPORT SERVICE DIAPHRAGM SEPARATOR LOCKING PISTON O-RING LOCKING PISTON

Disassembly and Reassembly for CBB-SR (Spring Return) Series Pneumatic Actuators

Table of Contents Visual Inspection and Neutralizing... 3 Disassembly

AIRTEK for Mack Vehicles Single to Dual Height Control Valve Conversion Kit

Winters Instruments: Corporate Engineering White Paper: Waste Water Treatment Application: Isolation Ring Assembly. Principle of Operation

Fuel Pump ( ) Table of Contents. Summary. View Related Topic. Summary. General Information Test. Preparatory Steps. Remove.

BETTIS SERVICE INSTRUCTIONS DISASSEMBLY AND REASSEMBLY FOR CB-SR-S SEISMIC SPRING RETURN SERIES PNEUMATIC ACTUATORS

User s Manual Power Supply IM E. 5th Edition

Maintenance Information

AIRTEK for Volvo Vehicles Single to Dual Height Control Valve Conversion Kit

Injector. General Information CAUTION. Use only the specified injector for the engine.

2001 Dodge RAM 3500 PICKUP

Operation Manual for the Drummond Hood Mate Pipet-Aid

Throttling Pendulum Gate O-ring Replacement Procedure Rev. A, May 2012

GHG14 DD15 Asymmetrical Turbocharger May Removal of the GHG14 DD15 Asymmetrical Turbocharger

Ultra Elite Facepiece

INSTALLATION INSTRUCTIONS

PET UNIT DOSE TABLE INSTALLATION MANUAL FN: Rev A 1/18

CTC Battery, Inc. Lithium Iron Phosphate Battery Specification

MALCOM Table-Top Digital Viscometer Model PC-1TL Series

Maintenance Information

Brake Shoe: Service and Repair Removal Procedure

Input Adjusting Ring Leak Repair Procedure for Meritor MT-14X Series Forward Differential Carriers

Fitting Systems. Leak-free, ee, simple-to-use 1/4-28 and M6 fitting types in

Cleaning and Rotor Replacement Instructions

TOYOTA TRUCKS / SUVs COLD AIR INTAKE Section I Installation Preparation. 4.0L V6 (1GR-FE) Part Number(s): PTR

Iso-Flo Coupling Repair

Backpack Sprayer. Use and Care Manual

Normally Open H200 and 90_ CF200 Module Seal Rebuild Kit

FLOWSERVE CORPORATION NOBLE ALLOY VALVE BALL VALVE REPAIR INSTRUCTION

Product Features. Easy installation. Pump adds only 3.3 to the overall assembly height. 60 GPH (227 LPH) flow rate while in priming mode.

A-320 Balanced 1st Stage

Integra Pneumatically Operated Diaphragm Valves with Position Indication Option

CROSBY SERIES BP OMNI-TRIM PRESSURE RELIEF VALVES INSTALLATION, MAINTENANCE AND ADJUSTMENT INSTRUCTIONS

DART-AB Sciex Interface Manual

Operating Manual for. Dispenser. Read this manual thoroughly before operating the Labmax Bottle Top Dispenser!

STERNDRIVE UNIT 3 A DRIVE SHAFT HOUSING

Installation Instructions

INSTALLATION GUIDE. Clutch Cable Actuated Models Manual Revision:

Additions, Revisions, or Updates

Service Manual Air Plus Second Stage

Drive Axles SHAIS171. The Dana LMS Hub is available on the following Steer and Drive Axle Models:

User s Manual Power Supply. IM E 3rd Edition. Yokogawa Electric Corporation

Navigator ADS551 Low level dissolved oxygen Replacement / Upgrade procedures wet section spares

General Safety and Maintenance Manual. Bore and Stroke. Throttle Type. Blows per Minute. (L) Lever Inches (564 mm)

Installing the Low Level Indicator (P/N or P/N )

250L Cartridge Dual Seal

DT35 and DT55 Series Rotary Sprinklers Installation & Service Instructions

BETTIS SERVICE INSTRUCTIONS DISASSEMBLY AND REASSEMBLY FOR CB DOUBLE ACTING SERIES PNEUMATIC ACTUATORS

Seattle (564) Bungalow & Tree of Life Face

INSTALLATION INSTRUCTIONS South Highway 11 Westminster, SC Toll Free (888) (864) FAX (864)

18SP589* Series 60 Right Side EGR Bolt-Together System with Hydraulic Actuator Valve

LMT Lens/Optics Cleaning Procedure. Lens Cleaning Procedure

66 CHAPTER FOUR. Spark Plug Removal Refer to Figure 28 for spark plug wive routing according to engine.

Tillotson Tc3A Carburator

OCTOPUS SERVICE PROCEDURE - SP007 REV NEW 15-FEB-2008

WARNING Carefully Read These Instructions Before Use

SM64052 Maintenance & Repair Manual Commercial 2" Unisex Coupling - Valved Model 64052

Service Manual. Climate Control Inc.

1 of 13 2/10/2016 6:46 PM

X-Tractor Series. Cutters INSTRUCTION MANUAL

GATOR ESP710 Battery Stud Punch

DT34 and DT54 Series Rotary Sprinklers

Additions, Revisions, or Updates

Waters SQ Detector Operator s Guide

GL Ludemann Y-Strainers

Tool-less Hinged Closure Installation, Operation, & Maintenance

FireHawk M7XT Ultra Elite XT Facepiece

1. Get fork mounted in stand. You can leave it in the bike, but you must remove the wheel and front brake.

INSTALLATION GUIDE. Kawasaki KLR Manual Revision:

Transcription:

StepWave ion guide: cleaning instructions addendum StepWave ion guide: cleaning instructions addendum This document, an addendum to the following documents, describes an improved cleaning procedure for the StepWave ion guide: Waters Xevo TQ-S Mass Spectrometry System Operator s Overview and Maintenance Guide, Revision A; replaces Cleaning the StepWave ion guide assembly on pages 4-59 to 4-78. Waters Xevo G2-S Tof Operator s Overview and Maintenance Guide, Revision A (part number 715003597); replaces Cleaning the StepWave ion guide assembly on pages 5-58 to 5-78. Waters Xevo G2-S QTof Operator s Overview and Maintenance Guide, Revision A (part number 715003596); replaces Cleaning the StepWave ion guide assembly on pages 5-58 to 5-78. Waters SYNAPT G2-S Mass Spectrometer Operator s Overview and Maintenance Guide, Revision B; replaces Cleaning the StepWave ion guide assembly on pages 5-40 to 5-58. Waters SYNAPT G2-S HDMS Mass Spectrometer Operator s Overview and Maintenance Guide, Revision B; replaces Cleaning the StepWave ion guide assembly on pages 5-40 to 5-58. For safety advisory information, see Appendix A in the operator s overview and maintenance guide for your instrument. Contents: Topic Page Cleaning the StepWave ion guide assembly 2 Waters is a registered trademark, and StepWave and THE SCIENCE OF WHAT S POSSIBLE. are trademarks of Waters Corporation. PEEK is a registered trademark of Victrex plc. All other trademarks are the sole property of their respective owners. Copyright 2012 Waters Corporation. *715003880* *Rev.A* November 2012, 715003880 Rev. A 1

Cleaning the StepWave ion guide assembly Clean the StepWave ion guide assembly if cleaning the ion block and isolation valve fails to increase signal sensitivity. Handling the StepWave ion guide assembly Caution: To avoid damaging the StepWave ion guide assembly, handle it and its components carefully throughout the cleaning procedure. In particular, to avoid damaging the wiring on assemblies fitted with an externally-wired printed circuit board (PCB), do not touch the wiring. To avoid damaging assemblies fitted with a flexible PCB, which is not externally wired, do not touch the PCB. Removing the ion block support from the source assembly Required materials Chemical-resistant, powder-free gloves 3-mm Allen wrench O-ring removal kit To remove the ion block support from the source assembly: Warning: To avoid personal contamination with biologically hazardous, toxic, or corrosive materials, and to avoid spreading contamination to uncontaminated surfaces, wear clean, chemical-resistant, powder-free gloves when working with the probe and source components. 1. Remove the source enclosure from the instrument (see Removing the source enclosure from the instrument in your instrument s Operator s Overview and Maintenance Guide). 2. Remove the ion block assembly from the PEEK ion block support (see Removing the ion block assembly from the source assembly in your instrument s Operator s Overview and Maintenance Guide). 2 November 2012, 715003880 Rev. A

Cleaning the StepWave ion guide assembly 3. Use the 3-mm Allen wrench to unscrew and remove the 4 screws securing the PEEK ion block support to the adaptor housing. Adaptor housing PEEK ion block support Securing screws StepWave assembly 4. Remove the PEEK ion block support from the adaptor housing. 5. Use the O-ring removal kit to carefully remove all the O-rings from the PEEK ion block support (see Removing O-rings and seals in your instrument s Operator s Overview and Maintenance Guide). Warning: To avoid contaminating uncontaminated surfaces with biologically hazardous, toxic, or corrosive materials, dispose of the O-rings according to local environmental regulations. 6. If any of the O-rings show signs of deterioration or damage, dispose of them in accordance with local environmental regulations. November 2012, 715003880 Rev. A 3

Removing the StepWave assembly from the source assembly Required materials Chemical-resistant, powder-free gloves Seal breaker and locator tool StepWave assembly removal and insertion tool Recommendation: When not in use, store the seal breaker and locator tool on the end of the StepWave assembly removal and insertion tool. To remove the StepWave assembly from the source assembly: Warning: To avoid personal contamination with biologically hazardous, toxic, or corrosive materials, and to avoid spreading contamination to uncontaminated surfaces, wear clean, chemical-resistant, powder-free gloves when working with the probe and source components. Caution: To avoid damaging the StepWave ion guide assembly when removing it from the source assembly, use only these tools: Seal breaker and locator tool StepWave assembly removal and insertion tool Do not use any other tool. 1. Position the seal breaker and locator tool so that it engages as follows: Its pin with the hole on the adaptor housing directly under the end of the StepWave ion guide assembly. Its lip behind the StepWave ion guide assembly s ion guide cap. 4 November 2012, 715003880 Rev. A

Cleaning the StepWave ion guide assembly Seal break and locator tool: Lip Pin Handle Seal breaker and locator tool positioned on the adaptor housing: Ion guide cap Adaptor housing Seal breaker and locator tool November 2012, 715003880 Rev. A 5

2. Push firmly on the seal breaker and locator tool s handle, to lever the StepWave assembly slightly out of the adaptor housing. Rationale: Moving the assembly in this manner releases it from a seal located inside the instrument. 3. With the StepWave removal and insertion tool s cutout uppermost, insert the tool s pins into the ion block support screw holes above and below the aperture in the pumping block assembly. Ion block support screw holes Cutout StepWave assembly removal and insertion tool Pins Brown PEEK ion guide cap Slot Caution: To avoid damage when removing the StepWave ion guide assembly from the adaptor housing, handle only the brown PEEK ion guide cap. 4. Inserting your thumbs through the slots in the StepWave removal and insertion tool, pull the StepWave ion guide assembly from the pumping block assembly and into the StepWave removal and insertion tool. 6 November 2012, 715003880 Rev. A

Cleaning the StepWave ion guide assembly Caution: To avoid damaging the StepWave ion guide assembly, handle it and its components carefully throughout the cleaning procedure. 5. Remove the StepWave ion guide assembly from the StepWave removal and insertion tool. 6. Using both hands, fit the source enclosure to the two supporting studs on the source adaptor housing. 7. Close the source enclosure. Rationale: Fitting and closing the source enclosure prevents debris entering the instrument while you are working on the StepWave ion guide assembly. November 2012, 715003880 Rev. A 7

Disassembling the StepWave ion guide assembly Required materials Warning: To avoid personal contamination with biologically hazardous, toxic, or corrosive materials, and to avoid spreading contamination to uncontaminated surfaces, wear clean, chemical-resistant, powder-free gloves when working with the probe and source components. Chemical-resistant, powder-free gloves Combined 2.5-mm Allen wrench and cone extraction tool O-ring removal kit To disassemble the StepWave ion guide assembly: 1. Use the combined 2.5-mm Allen wrench and cone extraction tool to remove the 2 screws securing the first ion guide assembly to the second ion guide assembly. First ion guide assembly Second ion guide assembly Securing screw 8 November 2012, 715003880 Rev. A

Cleaning the StepWave ion guide assembly 2. Separate the first and second ion guide assemblies. First ion guide assembly Second ion guide assembly 3. Remove the brown PEEK gasket from the second ion guide assembly. Brown PEEK gasket Second ion guide assembly November 2012, 715003880 Rev. A 9

4. Remove the O-ring from the differential pumping aperture on the second ion guide assembly (see the section Removing O-rings and seals in your instrument s operator s overview and maintenance guide). Differential pumping aperture O-ring Second ion guide assembly Warning: To avoid contaminating uncontaminated surfaces with biologically hazardous, toxic, or corrosive materials, dispose of the O-ring according to local environmental regulations. 5. If the O-ring shows signs of deterioration or damage, dispose of it in accordance with local environmental regulations. Caution: To avoid misalignment of the StepWave ion guide, do not remove the differential pumping aperture or any further components from the assembly. 10 November 2012, 715003880 Rev. A

Cleaning the StepWave ion guide assembly Cleaning the StepWave ion guide assembly Warning: To avoid personal contamination with biologically hazardous, toxic, or corrosive materials, and to avoid spreading contamination to uncontaminated surfaces, wear clean, chemical-resistant, powder-free gloves when working with the first ion guide and second ion guide assemblies. Caution: To avoid damaging the first ion guide and second ion guide PCB assemblies, handle them carefully throughout this procedure. To avoid damaging the StepWave ion guide assembly, do not use acetone, chlorinated solvents, or acid as solvents when cleaning it. Use only the solvents specified in this procedure. Required materials Chemical-resistant, powder-free gloves. Two 250-mL measuring cylinders in which to completely immerse the first ion guide and second ion guide assemblies when cleaning. Two lengths of small-diameter PEEK or PTFE tubing appropriately sized for suspending the first ion guide and second ion guide assemblies in the glass vessels when cleaning. HPLC-grade deionized water. Waters MS Cleaning Solution. Holding container for used Waters MS Cleaning Solution. HPLC-grade isopropyl alcohol. Ultrasonic bath. Source of oil-free, inert gas (for example, nitrogen) for drying. November 2012, 715003880 Rev. A 11

To clean the first ion guide and second ion guide PCB assemblies: 1. Bend a PEEK or PTFE tube into a hook shape. Caution: To avoid damage to the ion guide caused by vibration, ensure that the bottom of the ion guide does not touch the bottom of the glass vessel. 2. Use the hook to carefully suspend the first ion guide PCB assembly in the glass vessel so that the bottom of the assembly does not touch the bottom of the vessel. Hook First ion guide PCB assembly 3. Add Waters MS Cleaning Solution to the glass vessel until the first ion guide PCB assembly is immersed completely. 12 November 2012, 715003880 Rev. A

Cleaning the StepWave ion guide assembly 4. Repeat step 1 through step 3 for the second ion guide PCB assembly, placing the hook through one of the support rod holes. Hook Second ion guide PCB assembly 5. Place the vessels containing the first ion guide and second ion guide PCB assemblies in the ultrasonic bath for 20 minutes. 6. Carefully pour the cleaning solution from the vessel holding the first ion guide PCB assembly into the holding container, retaining the ion guide in the vessel. Tip: You can reuse the cleaning solution for one subsequent cleaning. 7. Fill the vessel with deionized water, to rinse the first ion guide PCB assembly, then discard the water. 8. Refill the vessel with deionized water, to rinse the first ion guide PCB assembly a second time, and then discard the water. 9. Repeat step 6 through step 8 for the second ion guide PCB assembly. 10. Fill both vessels with deionized water, ensuring that each ion guide PCB assembly is immersed completely. 11. Place the vessels containing the first ion guide and second ion guide PCB assemblies in the ultrasonic bath for 20 minutes. 12. Carefully pour away and discard the deionized water from both vessels, retaining the ion guide PCB assemblies in each vessel. November 2012, 715003880 Rev. A 13

13. Fill both vessels with isopropyl alcohol, ensuring that each ion guide PCB assembly is immersed completely. 14. Place the vessels containing the first ion guide and second ion guide PCB assemblies in the ultrasonic bath for 20 minutes. 15. Carefully remove each ion guide PCB assembly from its vessel, and blow-dry each assembly using inert, oil-free gas. 16. Discard the used isopropyl alcohol, using an appropriate waste container. 14 November 2012, 715003880 Rev. A

Cleaning the StepWave ion guide assembly Assembling the StepWave ion guide assembly Required materials Chemical-resistant, powder-free gloves Combined 2.5-mm Allen wrench and cone extraction tool New O-ring PEEK gasket StepWave assembly securing screws (2) To assemble the StepWave ion guide assembly: Caution: To avoid damaging the StepWave ion guide assembly, handle it and its components carefully when performing this procedure. 1. Fit the new O-ring to the differential pumping aperture on the second ion guide assembly. Differential pumping aperture O-ring Second ion guide assembly November 2012, 715003880 Rev. A 15

2. Fit the brown PEEK gasket to the second ion guide assembly. Important: Ensure that the gasket is orientated correctly. Brown PEEK gasket Second ion guide assembly 3. Align the first ion guide assembly with the second ion guide assembly. 4. Use the combined, 2.5-mm, Allen wrench and cone extraction tool to fit and tighten the 2 screws securing the first ion guide assembly to the second ion guide assembly. 16 November 2012, 715003880 Rev. A

Cleaning the StepWave ion guide assembly Fitting the StepWave assembly to the source assembly Required materials Chemical-resistant, powder-free gloves Seal breaker and locator tool StepWave assembly removal and insertion tool To fit the StepWave assembly to the source assembly: Warning: To avoid personal contamination with biologically hazardous, toxic, or corrosive materials, and to avoid spreading contamination to uncontaminated surfaces, wear clean, chemical-resistant, powder-free gloves when working with the probe and source components. Caution: To avoid damage to the StepWave ion guide assembly when fitting it to the source assembly, use only the seal breaker and locator, and the assembly s removal and insertion tools. 1. Disconnect the probe adjuster and options cables from the instrument s connectors. 2. Pull the source enclosure release (located at the bottom, right-hand side) outwards, and swing open the enclosure. 3. Using both hands, grasp the source enclosure, and lift it vertically off the two supporting studs on the source adaptor housing. 4. Slide the first ion guide assembly end of the StepWave assembly into the StepWave removal and insertion tool. Tip: The StepWave assembly can only be inserted in the correct orientation in the StepWave removal and insertion tool. November 2012, 715003880 Rev. A 17

StepWave assembly Cutout Pins StepWave assembly removal and insertion tool 5. With the StepWave removal and insertion tool s cutout uppermost, insert the StepWave removal and insertion tool s pins into the ion block support screw holes above and below the aperture in the pumping block assembly. Caution: To avoid damage when fitting the StepWave ion guide assembly to the adaptor housing, handle only the brown PEEK ion guide cap. 6. Inserting your thumbs through the slots in the StepWave removal and insertion tool, firmly push the StepWave ion guide assembly into the pumping block assembly. Tip: You will detect some resistance to motion when the StepWave assembly encounters the seal inside the instrument; continue pushing until you detect further resistance. 7. Remove the StepWave removal and insertion tool. 18 November 2012, 715003880 Rev. A

Cleaning the StepWave ion guide assembly 8. Invert the seal breaker and locator tool, and locate it over the end of the StepWave assembly. Inverted seal breaker and locator tool StepWave assembly Adaptor housing 9. Push firmly on the seal breaker and locator tool until the tool s face contacts the adaptor housing. Rationale: This fully locates the StepWave assembly in the adaptor housing. 10. Remove the seal breaker and locator tool. November 2012, 715003880 Rev. A 19

Fitting the ion block support to the source Required materials Chemical-resistant, powder-free gloves 3-mm Allen wrench New seals and O-rings To fit the PEEK ion block support to the source: Warning: To avoid personal contamination with biologically hazardous, toxic, or corrosive materials, and to avoid spreading contamination to uncontaminated surfaces, wear clean, chemical-resistant, powder-free gloves when working with the probe and source components. 1. Ensure that the grooves for the PEEK ion block support O-rings are free from dirt and debris. Tip: If contamination is present, use 1:1 methanol/water, applied to a lint-free cloth, to carefully clean the grooves. 2. Fit the O-rings (new ones if you disposed of the old) to the PEEK ion block support. Tip: To fit an O-ring in its groove, start fitting the O-ring at the notch in the groove, and then progressively work the ring into the groove, in either direction from the notch. 3. Fit the PEEK ion block support to the instrument s housing. 4. Use the 3-mm Allen wrench to fit and tighten the 4 PEEK ion block support securing screws. 5. Fit the ion block assembly to the PEEK ion block support (see Fitting the ion block assembly to the source assembly in your instrument s Operator s Overview and Maintenance Guide). 6. Fit the source enclosure to the instrument (see Fitting the source enclosure to the instrument in your instrument s Operator s Overview and Maintenance Guide). 20 November 2012, 715003880 Rev. A