Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems

Similar documents
Meeting the Requirements of ASTM D ( IP548/06 ) with Agilent 1200 LC Systems

Application Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Authors. Abstract. Energy & Chemicals

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method

Application Note. Author. Introduction. Energy and Fuels

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

Application Note. Abstract. Authors. Environmental Analysis

Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Using the PSD for Backflushing on the Agilent 8890 GC System

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector

Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution

Group-Type Analysis (PiPNA) in Diesel and Jet Fuel by Flow Modulated GCxGC FID.

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Analysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction

High-Temperature Simulated Distillation System Based on the 6890N GC Application

Application. Gas Chromatography June 1995

This document is a preview generated by EVS

Agilent Multimode Inlet for Gas Chromatography

High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Determination of Free and Total Glycerin in B100 Biodiesel

Analysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID

Large Volume Injection of Polycyclic Aromatic Hydrocarbons

Complete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105

Analysis of Petroleum Fractions by ASTM D2887

The Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over

Online sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve

Technical Procedure for Gas Chromatography (GC-FID)

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

Restek Corporation 110 Benner Circle, Bellefonte, Pa Barry L. Burger, Neil Johansen, Valerie Gamble, Donald Rhoades

Analysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES

TECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure

Method Development for Capillary GC Systems. Slide 1

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

Agilent and ASTM. Update on Recent Activities. Page 1

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels

Sulfur Detection at ppb Levels in Light Hydrocarbon Streams

Technical Overview. Introduction

Agilent Solutions for the Petrochemical and Oleochemical Industries

Simultaneous Determination of Fatty Acid Methyl Esters Contents in the Biodiesel by HPLC-DAD Method

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE

CONSTITUTION OF COAL TAR PITCH AND ITS EFFECT ON PROPERTIES

MET-Biodiesel Capillary GC Columns

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

Module8:Engine Fuels and Their Effects on Emissions Lecture 36:Hydrocarbon Fuels and Quality Requirements FUELS AND EFFECTS ON ENGINE EMISSIONS

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC

AppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS

Agilent 7693A Automated Liquid Sampler

Prediction of Physical Properties and Cetane Number of Diesel Fuels and the Effect of Aromatic Hydrocarbons on These Entities

ANNEX 2, REFERENCE FUELS

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

Agilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass. James D. McCurry, Ph.D. Senior Scientist

Performing ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software

Simulated Distillation Analyzers, Software, Standards, Consumables, Training

a review of analytical methods for the quantification of aromatics in diesel fuels

ASTM D 6730 Detailed Hydrocarbon Analysis

Agilent G5632 Valve Kit Instructions

ANNEX 3 REFERENCE FUELS. Parameter Unit Limits (1) Test method Minimum Maximum Research octane number, RON

Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut

Optimized Supercritical Fluid Chromatographic Instrumentation for the Analysis of Petroleum Fractions

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

White Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information

AppNote 1/2010 KEYWORDS ABSTRACT. Biodiesel, Automation, ASTM D

High-throughput protein aggregate analysis of monoclonal antibodies using a novel dual-channel UHPLC instrument

performance productivity reliability

Determination of Phenolic Antioxidant DBPC and DBP Levels in Electrical Insulating Oil

Application Note. Authors. Abstract

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

Application Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER

Paragon Scientific Ltd Proficiency Testing Scheme Schedule

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve

Emissions from Heavy-Duty Diesel Engine with EGR using Oil Sands Derived Fuels

Rapid Qualitative GC-TOFMS Analysis of a Petroleum Refinery Reformate Standard

C2, C3, C4 Monomer Analysis

The Analysis of Biodiesel for Inorganic Contaminants, including Sulfur, by ICP-OES

Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography

APPLICATION OF SOLID PHASE MICROEXTRACTION (SPME) IN PROFILING HYDROCARBONS IN OIL SPILL CASES

A Single Method for the Direct Determination of Total Glycerols in All Biodiesels Using Liquid Chromatography and Charged Aerosol Detection

Analysis and. Separation of Oxygenates in Hydrocarbon Matrices. Simon Jones Application Engineer Folsom, CA

Fausto Munari e Andrea Cadoppi ThermoFisher - Italy

Practical Steps in GC Troubleshooting

State of Texas UST Monitoring visit us online at The latest TNRCC 1005/1006 UST methods used by Texas.

S 1125 HPLC Pump System

ASTM D Standard Specification for Biodiesel Fuel (B 100) Blend Stock for Distillate Fuels

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

A Comparison of Boat Introduction and Direct Injection using the Thermo Scientific ipro 5000 Series Combustion Analyzer

Transcription:

Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems Application Note Hydrocarbons Authors Michael Woodman Agilent Technologies, Inc. Chemical Analysis Solutions 2850 Centerville Road Wilmington, DE 19808 USA Malgorzata Sierocinska Agilent Technologies European Field Support Centre Waldbronn Germany Important Note: For reliable column performance, it is essential that Agilent publication 5990-9202EN be accessed and followed. Abstract The performance of diesel fuel is predominantly determined by its ignition quality. This parameter is known as the Cetane number. The Cetane number describes the volume % Cetane (hexadecane) present in a mixture of Cetane and 1-Methylnaphthalene. Generally, in order to provide the best performance and lifetime of an engine, the amount of aromatics in diesel should be as low as possible. For the analysis of nonaromatics and aromatics in diesel fuel and petroleum distillates boiling in the range 150 C to 400 C, there exists an IP Method (391/07), which uses HPLC with refractive index detection. The two compound classes (aromatics and non-aromatics) are separated using normal phase HPLC and a column which has little affinity for non-aromatic but pronounced selectivity for aromatic hydrocarbons [1]. Recent growth in biodiesel production created a demand for analysis of petrodiesel and petrodiesel/ biodiesel blends. In this method revision, fatty acid methyl esters (FAME) originating from biodiesel sources must elute after a tetra-aromatic marker peak, chrysene, which facilitates improved accuracy of large PAH molecules without interference from FAME. The refractive index detector is used because this detector responds to both non-aromatic and aromatic hydrocarbons.

About Standard Method IP391/07 This European Standard specifies a test method for the determination of the concentration of mono-aromatic, di-aromatic and tri+-aromatic hydrocarbons in diesel fuels that may contain fatty acid methyl esters (FAME) up to 5 % (v/v) and petroleum distillates in the boiling range from 150 C to 400 C. The polycyclic aromatic hydrocarbon content is calculated from the sum of di-aromatic and tri+-aromatic hydrocarbons and the total content of aromatic compounds is calculated from the sum of the individual aromatic hydrocarbon types. Compounds containing sulfur, nitrogen and oxygen may interfere in the determination; mono-alkenes do not interfere, but conjugated di-alkenes and polyalkenes, if present, may do so. The precision statement of the test method has been established for diesel fuels with and without FAME blending components, with a mono-aromatic content in the range from 6 % (m/m) to 30 % (m/m), a di-aromatic content from 1 % (m/m) to 10 % (m/m), a tri+-aromatic content from 0 % (m/m) to 2 % (m/m), a polycyclic aromatic content from 1 % (m/m) to 12 % (m/m), and a total aromatic content from 7 % (m/m) to 42 % (m/m)." [2] This method, also known as EN12916:2006, is an official method of the Energy Institute (United Kingdom, www.energyinst.org.uk ) which maintains IP (Institute of Petroleum) standards since their acquisition of the IP. Earlier IP391 revisions are similar to ASTM D-6591-06 and include a column backflush-capable instrument configuration and analysis scheme. This requirement was discontinued in the current IP391/07 revision due to erroneous reporting of tri+aromatic hydrocarbons when FAME were present. The main IP391 changes from earlier revisions include the elimination of the backflushing valve, allowing compatibility with biodiesel/ petrodiesel fuel blends (up to 5% v/v FAME) and modifications to calibrants to improve data accuracy. The various methods associated with middle distillate fuel analysis are shown in Table 1. Equipment and Conditions LC: G1312B: G1367C: G1316C: G1362A: Software: Agilent 1200 Series LC including Binary pump, used isocratically with pump head seals for normal phase, Agilent p/n 0905-1420 Autosampler with needle wash Thermostatted column compartment Refractive index detector Agilent ChemStation with version B.04.01 software Columns: ZORBAX NH 2 4.6 mm 150 mm, 5 µm (p/n 883952-708) and ZORBAX SB-CN 4.6 mm 150 mm, 5 µm (p/n 883975-905 ) connected in series using 0.12 70 mm ss connector tubing, Agilent p/n G1316-87303 Mobile phase: n-heptane, HPLC grade Flow rate: 1 ml/min Injection volume: 10 µl Oven temperature: 25 ºC Detection: Refractive index Sample preparation Samples and standards were prepared according to Standard Method IP391/07, using heptane as the diluent. Final quantitative results were reported using Agilent IP391/07 standard mixtures (p/n 5190-0485 system calibration standards SCS1 and SCS2, and p/n 5190-0484 quantitative calibration standards A-D). Results and Discussion The first step in implementing of IP391/07 is the analysis of calibrants that establish overall separation selectivity and resolution, and confirmation of the elution order of the calibrant components. (Sections 8.6, 8.7, 8.9 IP391/07). Figure 1 shows the results of running these calibrants on the Agilent system. Table 1. Middle Distillate Fuel Analysis Methods IP method and revision Method overview Special parameters ASTM method Comments IP391/07 150-400 C diesel fuel No backflush, amino and/or No current equivalent Same as method EN12916:2006 petro/bio blends up to B-5 cyano column available *MAH, DAH, Tri+AH are reported IP436/01 50-300 C No backflush, amino and/or D-6379-04 MAH and DAH reported aviation fuel, kerosene cyano column not for samples with Tri+AH IP548/06 150-400 C diesel fuel Backflush required, amino D-6591-06 MAH, DAH, Tri+AH reported and/or cyano column FAME interferes with result *MAH monoaromatic hydrocarbon, DAH - diaromatic hydrocarbon, Tri+AH tri and higher ring aromatic hydrocarbons 2

SCS1 1. Cyclohexane 2. 1-phenyldodecane 3. o-xylene (dimethylbenzene) 4. Hexamethylbenzene 5. Naphthalene 6. Dibenzothiophene 7. 9-methyl anthracene 3 1 2 4 5 6 7 Figure 1. Standard chromatogram of SCS1. The system calibration standard 1 (SCS1) determines selectivity and retention data for the saturate and aromatic markers used for method acceptance criteria. The SCS1 also determines retention time grouping parameters for sample reporting. Resolution between cyclohexane and o-xylene (1,2-dimethylbenzene) is part of the method specification and must conform to a minimum and maximum value. System calibration standard 2 (SCS2) establishes the selectivity for components present in petro/bio diesel blends, demonstrating that there is no interference with tri+aromatic components by FAME. Petro/bio fuel blends require longer analysis time to elute all FAME peaks before the next analysis is begun. This applies whether there is an interest in quantifying the FAME or not. When FAME is present in the sample, but does not need to be quantified, it is possible to reduce analysis time by programming the flow rate through the column to increase with time. This will rapidly wash FAME components from the column. The method requirements state that chrysene, a tetra-aromatic marker peak, must elute with or before the first FAME peak. As shown in Figure 2, the selected operating conditions provide ample separation of the chrysene from the first FAME peak (C16:0 and C18:0 partially resolved). 3

SCS2 1. Chrysene 2-6. FAME 1 2, 3 4 5 6 Figure 2. Standard chromatogram of SCS2. With a genuine fuel sample, in this case retail quality petrodiesel, we can see greater complexity and overlapping of the various compound class regions (Figure 3). Within the method definitions there are specific "cut" points defining the grouping to be performed in the quantitative reports. These are calculated from retention and peak width data in SCS1. 1. Saturates 2. mono-aromatics 3. di-aromatics 4. tri+aromatics Ultra Low Sulfur Diesel (ULSD), retail pump 1 2 3 4 t = <15 minutes Figure 3. Petroleum diesel sample showing cut points for the various compound groups typically present in these samples. 4

Results and Discussion Method Performance As with most official methods, there are specific performance criteria that allow qualification of the separation system and its subsequent use for reporting quantitative results of diesel fuel analysis. 6.4 Column system, consisting of a stainless steel HPLC column(s) packed with a commercial 3 µm, 5 µm or 10 µm amino-bonded (or amino/cyanobonded) silica stationary phase meeting the resolution requirements given in 8.6, 8.7 and 8.9. 8.6 Ensure the components of the SCS1 are eluted in the order: cyclohexane, phenyldodecane, 1,2, dimethylbenzene, hexamethylbenzene, naphthalene, dibenzothiophene and 9-methylanthracene. 8.7 Ensure that baseline separation is obtained between all components of the SCS1. 8.9 Ensure that the resolution between cyclohexane and 1,2 dimethylbenzene is between 5.7 and 10. [calculated as described in 11.2] 11.2 Column Resolution Calculate the resolution, R, between cyclohexane and 1,2 dimethylbenzene using the following equation. R = 2(t3-t1) 1.699(y1+y3) (difference in retention time) (averaging of peak widths) Ret Time W hh Name [min] Resolution [min] Cut ref 1. Cyclohexane 3.738 0.0558 t1 2. 1-phenyldodecane 4.288 0.0648 t2 3. 1,2-dimethylbenzene 4.695 8.57 0.0756 t3 8.11Ensure the retention time peak of chrysene is higher than the 9-methylanthracene peak.and check that the chrysene peak elutes just before or with the first peak of FAME. In Figure 4, we see that there is distinct separation between the markers specified in section 8.11 of the method. With this information in hand, it is possible to proceed to the evaluation of calibration standards. 9.4 R = >0.999, Intercept <0.01 g/100 ml) 7 7. 9-methyl anthracene 8. Chrysene 9-13. FAME 8 9 13 Figure 4. Stack plot of SCS1 and SCS2 showing proof for section 8.11. 5

RID peak area vs Std conc, o-xylene RID peak area vs Std conc, fluorene RID peak area vs Std conc, phenanthrene RID peak area 400000 350000 300000 250000 200000 150000 100000 500000 0 0.000 10.000 20.000 30.000 40.000 Conc (mg/ml) 3500000 900000 R 2 = 1 R 2 = 1 800000 R 2 = 1 3000000 700000 2500000 600000 2000000 500000 Series 1 Linear (Series 1) RID peak area 1500000 1000000 500000 Series 1 Linear (Series 1) 0 0.000 5.000 10.000 15.000 20.000 25.000 Conc (mg/ml) RID peak area 400000 300000 200000 100000 Series 1 Linear (Series 1) 0 0.000 1.000 2.000 3.000 4.000 5.000 Conc (mg/ml) Figure 5. Calibration plots for o-xylene, fluorene and phenanthrene, the three components of the four calibration levels specified in the method. In the calculated results, all calibration plots exceed linearity of 0.9999 and have calculated intercepts well below 0.01 g/ 100 ml, the method specification of section 9.4. Retention time and peak area precision can be found in Table 2, in which we see that the overall performance of the method is excellent. Table 2. Retention Time and Peak Area Precision Calibrant A Analyte R.T. Avg, n=3 R.T. Std dev R.T. RSD% Area Avg, n=3 Area Std dev Area RSD% Xylene 4.57 0.003 0.06 3.54E+06 5829.9 0.16 Fluorene 6.82 0.004 0.05 3.38E+06 2500.5 0.07 Phenanthrene 8.32 0.004 0.04 8.05E+05 594.03 0.07 Calibrant B Analyte R.T. Avg, n=3 R.T. Std dev R.T. RSD% Area Avg, n=3 Area Std dev Area RSD% Xylene 4.65 0.001 0.02 9.23E+05 636.28 0.07 Fluorene 6.95 0.001 0.02 1.70E+06 1731.17 0.10 Phenanthrene 8.44 0 0.00 4.00E+05 473.79 0.12 Calibrant C Analyte R.T. Avg, n=3 R.T. Std dev R.T. RSD% Area Avg, n=3 Area Std dev Area RSD% Xylene 4.70 0.002 0.03 2.24E+05 474.36 0.21 Fluorene 7.15 0.002 0.03 4.29E+05 507.38 0.12 Phenanthrene 8.62 0.002 0.02 1.00E+05 291.04 0.29 Calibrant D Analyte R.T. Avg, n=3 R.T. Std dev R.T. RSD% Area Avg, n=3 Area Std dev Area RSD% Xylene 4.72 0.001 0.02 4.45E+04 321.90 0.72 Fluorene 7.36 0.002 0.02 1.70E+04 145.91 0.86 Phenanthrene 8.70 0.002 0.02 3.80E+04 551.43 1.45 Average RSD% All Runs 0.029 0.355 6

Results for Specific Petrodiesel and Petro/ Biodiesel Blends Various samples were collected from local commercial and retail fuel delivery points. An overlay of four samples is shown in Figure 6. a retained sample, winter blend fuel 1 2 3 4 newly acquired summer blend samples Figure 6. Overlay of four samples. Results (n=3 each sample) and precision data are reported in Table 3, where vendor 4 results are n=3 based on one sampling of each of the three B-11 biodiesel delivery points (commercial heavy truck, commercial auto/light truck and retail auto/light truck). 7

Table 3. Results (n=3 each sample) and Precision Data for Four Vendors. Vendor Group Avg, n=3 RSD% Std dev 1 MAH 36.0 g/100 ml 0.06 0.022 DAH 8.4 g/100 ml 0.09 0.008 Tri+AH 1.2 g/100 ml 2.78 0.033 2 MAH 28.5 g/100 ml 0.05 0.016 DAH 4.8 g/100 ml 0.48 0.023 Tri+AH 0.6 g/100 ml 3.24 0.020 3 MAH 29.0 g/100 ml 0.07 0.021 DAH 5.3 g/100 ml 0.37 0.020 Tri+AH 0.7 g/100 ml 6.81 0.047 4 MAH 24.7 g/100 ml 5.06 1.252 DAH 5.1 g/100 ml 3.38 0.174 Tri+AH 0.7 g/100 ml 4.58 0.033 Because of very low level response and broadenend peaks, it is much more difficult to get high precision for the tri+aromatic hydrocarbons group of components. If no biodiesel components are present, it would be practical to consider using Method IP548/01, which uses a backflush valve to elute the tri+aromatic hydrocarbons group as a single peak via the backflush configuration. Ruggedness and Stability of the IP391/07 Method As with most normal phase methods the column is susceptible to the adsorption of highly polar components which can affect overall separation performance. Water present in samples or the mobile phase also adsorbs into the column and somewhat predictably causes reduced elution times for all sample components. Using a high quality HPLC grade mobile phase is essential, and the user may consider using a drying agent such as molecular sieve to dehydrate the mobile phase. While this is often done by adding molecular sieve to the solvent container, it is also possible and preferable to prepare a high pressure compatible column with pre-washed drying agent and placing it inline between the pump and injector. Conclusion The performance of the Agilent 1200 Series HPLC with normal phase separation and refractive index detection meets or exceeds the requirements of IP391/07 within the range of samples defined in the method. The user should take care identifying samples of petrodiesel that may contain biodiesel components to ensure adequate analysis time before proceeding to the next analytical run. References 1. ASTM 6591-06 and ASTM 6379-04, www.astm.org 2. IP391/07, Energy Institute (formerly Institute of Petroleum Test Methods) 3. EN12916/06, Energy Institute (formerly Institute of Petroleum Test Methods) 4. Agilent Technologies publication 5965-9044EN, (1997) For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2009 Printed in the USA December 2, 2011 5990-9534EN