AppNote 1/2010 KEYWORDS ABSTRACT. Biodiesel, Automation, ASTM D

Similar documents
AppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

Application Note. Author. Introduction. Energy and Fuels

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

MET-Biodiesel Capillary GC Columns

Analysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction

Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105

GC-101. Testing of Biodiesel Using Gas Chromatography. by Bob Armantrout. Presented at: Local Biodiesel: A Biodiesel Coop Conference

Determination of Free and Total Glycerin in B100 Biodiesel

Performing ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Going Green: The Analysis of BioFuels

Application Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector

Application Note. Authors. Abstract

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Analysis of Petroleum Fractions by ASTM D2887

Application Note. Abstract. Authors. Environmental Analysis

Agilent 7693A Automated Liquid Sampler

Agilent 7693A Automated Liquid Sampler

ASTM D Standard Specification for Biodiesel Fuel (B 100) Blend Stock for Distillate Fuels

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

Using the PSD for Backflushing on the Agilent 8890 GC System

High-Temperature Simulated Distillation System Based on the 6890N GC Application

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

Technical Procedure for Gas Chromatography (GC-FID)

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

A Single Method for the Direct Determination of Total Glycerols in All Biodiesels Using Liquid Chromatography and Charged Aerosol Detection

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC

ASTM D for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software

Application Note. Authors. Abstract. Energy & Chemicals

Large Volume Injection of Polycyclic Aromatic Hydrocarbons

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas

Analysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector

Phase Distribution of Ethanol, and Water in Ethyl Esters at K and K

Biodiesel Analysis Toolkit

This document is a preview generated by EVS

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

4001 Transesterification of castor oil to ricinoleic acid methyl ester

Agilent and ASTM. Update on Recent Activities. Page 1

Agilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass. James D. McCurry, Ph.D. Senior Scientist

6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR

economics and performance Spark Holland introduces OPTIMAS The optimum balance of BETTER SAMPLE CARE

Simultaneous Determination of Fatty Acid Methyl Esters Contents in the Biodiesel by HPLC-DAD Method

Application. Gas Chromatography June 1995

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

Sample preparation Methods Analysis. Biodiesel Analysis guide.

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

S 1125 HPLC Pump System

Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems

Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography

Biodiesel Analysis Utilizing Mini-Scan - Handheld Analyzer V.C. Gordon PhD, Bonanza Labs

Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution

Setting up SilFlow for BackFlush in your GC

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE

Meeting the Requirements of ASTM D ( IP548/06 ) with Agilent 1200 LC Systems

Characterization of Tiki Torch Fuels

The next level of True Blue Performance

Study of viscosity - temperature characteristics of rapeseed oil biodiesel and its blends

Agilent 6850 Series II Network GC System G2630A Performance Specifications

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC

GC Analysis of PCBs in Fish Oil by Direct Injection Without Any Sample Cleanup. Philip L. Wylie, Ph.D. Sr. Applications Chemist Agilent Technologies

A Comparison of Boat Introduction and Direct Injection using the Thermo Scientific ipro 5000 Series Combustion Analyzer

Fausto Munari e Andrea Cadoppi ThermoFisher - Italy

High-throughput protein aggregate analysis of monoclonal antibodies using a novel dual-channel UHPLC instrument

Agilent Multimode Inlet for Gas Chromatography

Productivity Improvements for Dual Column Applications: USP <467> and Blood Alcohol Analysis

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

VERITY Purification Systems Build the perfect system to fit your research

Keywords: Simarouba Glauca, Heterogeneous base catalyst, Ultrasonic Processor, Phytochemicals.

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

Kinetics of palm kernel oil and ethanol transesterification

BIODIESEL PRODUCTION IN A BATCH REACTOR 1. THEORY

V.Venkatakranthi Teja. N S Raju Institute of Technology (NSRIT), Sontyam, Visakhapatnam, Andhra Pradesh , India.

Simulated Distillation Analyzers, Software, Standards, Consumables, Training

The Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over

Online sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve

Agilent 1290 Infinity Autosampler

Environmental Testing Solutions from SRI

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

Towards a Biodiesel-based Biorefinery: Chemical and Physical Properties of Reactively Extracted Rapeseed (Canola)

MB3600-CH30 Laboratory FT-NIR analyzer for biodiesel applications Suitable for production optimization and product quality assessment

Analysis of nitrogen isotope ratios in amino acids by GC/C/IRMS

Agilent Multimode Inlet

Spark Holland introduces ALIAS for ÄKTA pure All-round, state-of-the-art

The Pennsylvania State University. Earth and Mineral Sciences. College of Energy and Mineral Engineering A COMPREHENSIVE EVALUATION AND DEVELOPMENT OF

Determination of Sudan Dyes I IV in Curry Paste

DANI Transformer Oil Gas Analyzer

PRODUCT SPECIFICATION. Product Specification Crystal 9000 Gas Chromatograph. Chromatec Crystal Laboratory Gas Chromatography System

Transcription:

AppNote 1/2010 Full Automation of ASTM Method D6584-07 Standard Test Method for the Determination of Free and Total Glycerin in B-100 Biodiesel Methyl Esters by Gas Chromatography using a GERSTEL Dual Rail PrepStation John R. Stuff, Jacqueline A. Whitecavage Gerstel, Inc., 701 Digital Dr. Suite J, Linthicum, MD 21090, USA KEYWORDS Biodiesel, Automation, ASTM D6584-07 ABSTRACT Biodiesel is defined as mono-alkyl esters of long chain fatty acids derived from vegetable oils or animal fats which conform to ASTM D6751 specifications for use in diesel engines. Biodiesel is produced from the fat or oil by transesterification. Glycerin is an unwanted byproduct of the reaction and must be removed from the final product. ASTM Method D6584-07 measures the amount of residual free and bonded (mono-, di-, and triglycerides) glycerin in the biodiesel fuel. Standard and sample preparation, which requires a derivatization step, can be laborious and time consuming. Using the GERSTEL MPS 2 Dual Rail PrepStation and MAESTRO control software, the entire method, including standards and sample preparation, derivatization, dilution, and analysis can be fully automated. This paper shows the details of the automation.

INTRODUCTION A dual rail PrepStation was set up to fully automate the procedure outlined in ASTM D6584-07 for the determination of free and total glycerides in B- 100 biodiesel fuel. The goal is to provide complete automation of standard preparation, automate next sample preparation within the time frame of the GC run, and provide other user-friendly features of the Maestro software control for the ASTM method. Full automation frees up the analyst s time for other laboratory operations. EXPERIMENTAL Instrumentation. Analyses were performed on an Agilent Technologies 6890 GC with flame ionization detector (FID), PTV inlet (CIS 4, GERSTEL) and Dual Rail MPS 2 robotic sampler with 10 μl on-column syringe and an 80 μl sideport syringe with diluter module (GERSTEL). 5 of the VT-32-10 tray. Biodiesel B-100 samples, approximately 100 mg, were weighed directly into 10 ml screw cap vials and placed on the VT-32-10 tray. The samples and standards were prepared using a GERSTEL MAESTRO PrepSequence. Sample introduction. The samples were prepared in vials placed in the VT32-10 tray. One microliter of the standard or sample was injected into the CIS 4 using the MPS 2 autosampler with a 10 μl liquid syringe with replaceable 26 gauge needle for on-column injection RESULTS AND DISCUSSION The GC system used for this study is shown in Figure 1. The method calls for the preparation of a five point calibration curve for glycerin, mono-, di-, and triolein from the stock standards. The required volumes are shown in Table 1. Analysis conditions. PTV: On-Column 60 C (0.05 min); 0.2 C/sec; 230 C (2 min); 0.5 C/sec; 380 C (10 min) Column: 10 m Rtx-Biodiesel TG (Restek) d i = 0.32 mm d f = 0.1 μm with 2m integrated guard column, d i = 0.53 mm Pneumatics: He, constant flow = 3 ml/min Oven: 50 C (1 min); 15 C/min; 180 C; 7 C/min; 230 C; 30 C/min; 380 C (10 min) FID: 380 C Standard description. A biodiesel standards kit (#D6584-SS), containing separate 5 ml ampoules of 500 ng/μl glycerin, 5000 ng/μl monoolein, 500 ng/μl diolein, 5000 ng/μl triolein, 1000 ng/μl butanetriol, and 8000 ng/μl tricaprin, all in pyridine, was purchased from Analytical Services, Inc. (The Woodlands, TX). The kit also contained MSTFA and heptane. Biodiesel B-100 was purchased locally. Sample preparation. The glycerin, mono-, di- and triolein standards were placed in 2 ml screw cap vials in Positions 1-4, respectively, of a VT-98 tray. The butanetriol, tricaprin and MSTFA were put in 10 ml vials with PTFE snap caps and placed in Positions 3, 4, and 5 in the 5 position 10 ml wash tray. Empty 10 ml screw cap vials were placed in Positions 1- Figure 1. System setup. Table 1. Required stock standard volumes. Standard Number 1 2 3 4 5 µl Glycerin Stock 10 30 50 70 100 µl Monoolein Stock 20 50 100 150 200 µl Diolein Stock 10 20 40 70 100 µl Triolein Stock 10 20 40 70 100 AN/2010/1-2

The GERSTEL MAESTRO PrepBuilder function is used to automate the preparation and injection of standards and samples. An example of a standard preparation loop for the addition of the diolein stock to the five standard vials is shown below. The individual steps are selected by mouse-click from a drop-down menue. ACTION METHOD / VALUE SOURCE DESTINATION PREP Vials 3-3 Extensive (R) ADD Add 10 ul Std 3 @ Tray2,VT98 1 @ Tray1,VT32-10 (R) ADD Add 20 ul Std 3 @ Tray2,VT98 2 @ Tray1,VT32-10 (R) ADD Add 40 ul Std 3 @ Tray2,VT98 3 @ Tray1,VT32-10 (R) ADD Add 70 ul Std 3 @ Tray2,VT98 4 @ Tray1,VT32-10 (R) ADD Add 50 ul Std 3 @ Tray2,VT98 5 @ Tray1,VT32-10 (R) ADD Add 50 ul Std 3 @ Tray2,VT98 5 @ Tray1,VT32-10 (R) DILUTE Dilute Wash Syringe Waste END Samples are prepared by weighing approximately 100 mg into an 10 ml empty vial. The preparation of standards and samples is performed as follows: 1. Add 100 μl each of 2 Internal Standards 2. Add 100 μl Derivatizing Reagent, Mix, Wait 15 min 3. Dilute with 8 ml heptane, Mix 4. Analyze using a 1 μl on-column Injection An example for the preparation and analysis loop of the five standards and five samples is shown below. ACTION METHOD / VALUE SOURCE DESTINATION PREP Vials 1-10 Extensive (R) ADD ADD First 50 ul SolvRes1 Tray1,VT32-10 (R) ADD ADD Second 50 ul SolvRes1 Tray1,VT32-10 (R) DILUTE Dilute Wash Syringe Waste (R) ADD ADD First 50 ul SolvRes2 Tray1,VT32-10 (R) ADD ADD Second 50 ul SolvRes2 Tray1,VT32-10 (R) DILUTE Dilute Wash Syringe Waste (R) ADD ADD First 50 ul SolvRes3 Tray1,VT32-10 (R) ADD ADD Second 50 ul SolvRes3 Tray1,VT32-10 (R) DILUTE Dilute Wash Syringe Waste (R) MOVE Tray1,VT32-10 Agitator,AgiTray (R) MIX Mix 1 Minute (R) MOVE Agitator,AgiTray Tray1,VT32-10 (R) WAIT 15.00 min (R) DILUTE Dilute 8 ml Heptane Tray1,VT32-10 (R) MOVE Tray1,VT32-10 Agitator,AgiTray (R) MIX Mix 1 Minute (R) MOVE Agitator,AgiTray Tray1,VT32-10 (L) INJECT ASTM_D6584.m\maestro.mth Tray1,VT32-10 Rear END The first 2 ADD steps correspond to the addition of 100 μl of the butanetriol internal standard. After the syringe is washed, 100 μl of tricaprin internal standard is added, followed by another wash and 100 μl of the MSTFA to derivatize the sample/standard. The sample or standard is then mixed, a 15 minute wait step is inserted to allow the derivatization to be completed, heptane is added for dilution, followed by mixing and GC analysis. The inject action step contains all the Agilent Chemstation and GERSTEL MPS 2 functions required for analysis of the samples. When integrated into Agilent ChemStation or MS ChemStation, one method and one sequence table runs the complete system. AN/2010/1-3

Figure 2 shows a typical chromatogram obtained for Standard #3. The standard and internal standard peaks are labeled in the chromatogram. Table 2 shows the linear regression results for the standard curves. Good correlation is achieved for all standards. Figure 3 shows the calibration curve for diolein. pa 700 600 Tricaprin 500 400 300 200 100 Glycerin Butanetriol Monoolein Diolein Triolein Time--> 5.00 10.00 15.00 20.00 Figure 2. Chromatogram for standard #3. 25.00 Table 2. Regression data for standards. Compound r 2 m b Glycerin 0.9999 1.181-0.0008 Monoolein 0.9997 1.247-0.0122 Diolein 0.9987 1.0203-0.0067 Triolein 0.9985 0.8946-0.0120 0.7 0.6 0.5 04 0.4 0.3 0.2 0.1 0 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 Figure 3. Diolein calibration curve. AN/2010/1-4

Table 3 shows the results for the sample replicates (n=5). Reproducibilities are good with %RSDs ranging from 2.1-3.4%. A typical chromatogram for this sample type is shown in Figure 4. Table 3. Mass percentages for B-100 Biodiesel. Glycerin Monoglycerides Diglycerides Triglycerides Sample 1 0.02608 1.107 0.1898 0.08909 Sample 2 0.02530 1.055 0.1864 0.09379 Sample 3 0.02662 1.059 0.1890 0.09398 Sample 4 0.02706 1.056 0.1858 0.09565 Sample 5 0.02670 1.071 0.1791 0.08861 Average 0.02635 1.070 0.1860 0.09220 Std Dev 0.000683 0.0220 0.00421 0.00316 %RSD 2.59 2.06 2.26 3.42 pa 700 600 Tricaprin 500 400 300 200 100 Glycerin Butanetriol Monoglycerides Diglycerides Triglycerides Time--> 5.00 10.00 15.00 20.00 Figure 4. Chromatogram for biodiesel sample. 25.00 AN/2010/1-5

The preparation of each sample was completed in approximately 27 minutes using MAESTRO PrepBuilder functions, well within the total GC cycle time of 38 minutes (31 minute GC run + 7 minute cooldown). This means that samples can be prepared during the GC run of the preceding sample without adding to the overall analysis time, a fact which is clearly illustrated by the MAESTRO Sequence Scheduler window (Figure 5). The scheduler graphically illustrates that the sample prep and GC run segments of the analysis are performed in parallel, while giving a clear overview of the total analysis time needed in order to facilitate laboratory workflow planning. Figure 5. MAESTRO sequence scheduler CONCLUSIONS A dual rail PrepStation was programmed to prepare standards and samples, and inject samples using GERSTEL MAESTRO software. The results showed good linearity for the standards and good reproducibility (%RSDs from 2.1-3.4%, n=5) for the B-100 biodiesel samples. Complete automation of ASTM D6584-07 was accomplished with next sample preparation within the timeframe of the GC run. AN/2010/1-6

AN/2010/1-7

GERSTEL GmbH & Co. KG Eberhard-Gerstel-Platz 1 45473 Mülheim an der Ruhr Germany +49 (0) 208-7 65 03-0 +49 (0) 208-7 65 03 33 gerstel@gerstel.com www.gerstel.com GERSTEL Worldwide GERSTEL, Inc. 701 Digital Drive, Suite J Linthicum, MD 21090 USA +1 (410) 247 5885 +1 (410) 247 5887 sales@gerstelus.com www.gerstelus.com GERSTEL AG Wassergrabe 27 CH-6210 Sursee Switzerland +41 (41) 9 21 97 23 gerstelag@ch.gerstel.com www.gerstel.ch GERSTEL K.K. 1-3-1 Nakane, Meguro-ku Tokyo 152-0031 SMBC Toritsudai Ekimae Bldg 4F Japan +81 3 5731 5321 +81 3 5731 5322 info@gerstel.co.jp www.gerstel.co.jp GERSTEL LLP Level 25, North Tower One Raffles Quay Singapore 048583 +65 6622 5486 +65 6622 5999 SEA@gerstel.com www.gerstel.com GERSTEL Brasil Av. Pascoal da Rocha Falcão, 367 04785-000 São Paulo - SP Brasil +55 (11)5665-8931 +55 (11)5666-9084 gerstel-brasil@gerstel.com www.gerstel.com.br Information, descriptions and specifications in this Publication are subject to change without notice. GERSTEL, GRAPHPACK and TWISTER are registered trademarks of GERSTEL GmbH & Co. KG. Copyright by GERSTEL GmbH & Co. KG Awarded for the active pursuit of environmental sustainability