Baxter 6201 and 6301 Tips to Avoid Nuisance Alarms Call x 335 for troubleshooting assistance.

Similar documents
CADD -Solis VIP Ambulatory Infusion Pump Model 2120

NIKI T34 pump. Knowing your way around the NIKI T34 pump

BodyGuardTM 323. BodyGuard 323 Multi-Therapy Ambulatory Infusion Pump. Patient Guide

MOOG Medical Devices Group Curlin 6000 Series Ambulatory Infusion Pump PATIENT GUIDE

If You Are Experiencing This, Click Thru:

Troubleshooting: Installation 8. Troubleshooting: Door Jamming and Door Handing 9. Troubleshooting: Keypad 13. Troubleshooting: Smart Home Systems 14

Troubleshooting: Door Jamming and Door Handing 6. Troubleshooting: Keypad 11. Troubleshooting: Smart Home Systems 12. Troubleshooting: Battery 14

Maintenance Information

Troubleshooting: Installation 10. Troubleshooting: Door Jamming and Door Handing 11. Troubleshooting: Touchscreen 14

CADD-Solis Infusion Pump

Installation and Maintenance Instructions. World Leader in Modular Torque Limiters. PTM-4 Load Monitor

CurveMaker DFS v2.0 Dyna FS Ignition Programming Software

Disco 3 Clock Spring / Rotary Coupler replacement

All of the control valves that we install the Mk. 16IQ positioner on are factory calibrated for proper operation prior to shipment.

Operations Manual. Automated Fuel Maintenance System FUEL TECHNOLOGIES INTERNATIONAL

OP-COM IMMOBILISER-II Guide

Acoustic Door Closer Instructions

USER GUIDE KAS-840-DEN KAS-850-RDEN

Ambulatory. Infusion. Pumps

OVERVIEW. Table of Contents. Bikes with a H4 or H7 (with P43t expander) headlamps

Acoustic Door Closer Instructions

Going home with a CADD Pump - Chemotherapy Medication

USER GUIDE KAS-840-DEN KAS-850-RDEN

High technology theft-prevention systems

Troubleshooting Guide for Limoss Systems

TIP SHEET. Installation Tips for your RS IB-MUX / PKUMUX (D) + SPDT T1205 v1.2 4/3/14. 1 P a g e

Operator Manual of Guangzhou Huaxi Medical Technology Co., Ltd. JORVET INFUSION PUMP J1060 OPERATOR MANUAL

Wireless Tire Pressure and Temperature Monitoring System Instruction Manual Model #: TM-507 SCE 507 Commercial Cap Sensors with Monochrome Display

Solar & Roller Shades

1. Connect the Tester to the L300 FS Cuff plug holes. See Figure Figure 11-1: Tester connected to the L300 FS Cuff.

Ford Gasoline Speedo-Pro Programmer. Reprogram. Power

Locks LOCKING AND UNLOCKING. Automatic unlocking. Interior door handles and door locking levers. Master lock and unlock switches. Drive-away locking

ECT Display Driver Installation for AP2 Module

Quick Reference Guide

Operating Guide TSFS-300 Air Parcel Conveyor Scale With automatic totalizing

HGM-MZ Multi-Zone Monitor Annual Maintenance And Troubleshooting Guide

Power. Reprogram. Ford Gasoline E-Z TUNE Programmer

Smart meter Instructions

SECTION G2: CABLE PROCESSOR MODULE MAINTENANCE

REV F2.0. User's Manual. Hydraulic ABS (HABS) Hydraulic Power Brake (HPB) Page 1 of 28

HOW TO USE A MULTIMETER, PART 4: MEASURING CURRENT (AMPERAGE)

SAVANNAH EI & DELUXE SYSTEM TROUBLE SHOOTING GUIDE

Troubleshooting of the LubeTech Grease System

Spray Height Controller

30A BLDC ESC. Figure 1: 30A BLDC ESC

CADD-Legacy 1 Ambulatory Infusion Pump Model 6400

Manifold QF. Contents. Description. Installation & Setup Guide. Safety 3. Introduction 3. Installation Standards 4. Specifications 5.

BodyGuard 121 Twins Infusion System Operator Manual

TIP SHEET T0937. Installation Tips For RS00/PS00 + ADS-TBSL-PL + SPDT

Method statement: Setting the Vibration Cutout Switch

KANGAROO JOEY Feeding Pump Instructions for use

N62 Secondary Air Cleaner Part #: AGA-N62-SEC-AIR

Syringe Driver Introduction. Christine Hull & Anita Webb Staff Nurses, Hospice in the Home 2015

RETAINS STEERING WHEEL CONTROLS, ONSTAR TM, ONSTAR TM BLUETOOTH, XM TM SATELLITE AND MORE!

Sentinel Enteral Feeding Pump. Model: S-1000-SI. Operating Manual

Bunya Solar. Renewable Energy Systems. Remote area power supply system manual

CAUTION-ELECTRICALLY OPERATED PRODUCT

Solar & Roller Shades

IMPORTANT SAFEGUARDS READ THIS MANUAL AND FOLLOW ALL SAFETY INSTRUCTIONS THOROUGHLY BEFORE OPERATING THE EMIU INVERTER SYSTEM SAVE THESE INSTRUCTIONS

PowerLevel s e r i e s

TBI /2012 TRAUMATIC BRAIN INJURY DEVICE

PRESSURE SENSOR INSTRUCTIONS

DynoTune Wideband Gauge

Centrifuge Operator / Service Manual

Active Head Restraint - Removal

Owner s guide. AVGM5 - Rear Seat Video Retention Harness SPGM5 - Optional Speaker for GM5 Vehicles

Solar & Roller Shades

PRODUCT MANUAL Gecko Wireless One Zone LED Dimmer and Receiver

7500 Greensand Filter Installation & Start Up Guide

EMIU 720W Mini Inverter

(2) Remove the customer's Waste Ink Cartridge and replace to a new one. (3) Make sure GT-541 is not in the middle of any operation.

2002 Dodge Intrepid ES ACCESSORIES & EQUIPMENT Anti-Theft Systems - Concorde, Intrepid & 300M

Automated Control Electronics (ACE ) System Operation and Diagnostics

INSTALLATION INSTRUCTIONS

Z-Truck (Vertical Moving) Z-truck Flag. Y-Truck (Horizontal Moving) FIGURE 1: VIEW OF THE Z-TRUCK. Flexshaft Assembly

Statim G4 Error Codes

Wireless Tire Pressure and Temperature Monitoring System Color Display Manual. Wide Screen Color Display Model #: TST-507-D-C

Complete Home Water Protection

DIGITAL OIL PRESET METER OWNER S MANUAL

Amerec T100 Steamer Control For use with AT and 3T steam Generators

THUNDERBIRD. Agricultural Weighing Systems SS1000 Instructions

CurveMaker HD v1.0 2Ki Programmable Ignition programming software

SAFETY PRECAUTIONS AND WARNINGS...

RH800 & 2000 MAINTENANCE GUIDE

TEL:1-866-XANATOS INSTALLATION INSTRUCTIONS PART#: SR-FOR SPARTAN RUNNING BOARD FORD ECONOLINE 99-14

Wallbox Commander. User Guide WBCM-UG-002-EN 1/11

ELD DRIVER GUIDE June 21, 2018

Service Bulletin Immobilizer System (Type 6) (Supersedes , dated February 20, 2009; see REVISION SUMMARY) February 22, 2013

NWA3D A5 User Manual

Amerec T100 Steamer Control For use with AT and 3T steam Generators

U S A 1 WAY. User's Guide E P

Patient Information CADD-Legacy 1 Pump

Section 8. MAINTENANCE & TROUBLESHOOTING

USER MANUAL LEVER LOCK DIGITAL TOUCHPAD TEMPLATE. 120x165(mm)

Arrow Shark TS760-Marine Onboard E-Starter. Owner Manual

Milano AUTOMATIC FAUCET

WIRELESS BLOCKAGE MONITOR OPERATOR S MANUAL

Vehicle Security / Remote Start / Remote Access System Installation

Power. Reprogram. JET Performance Products Apex Circle, Huntington Beach, CA (714) Fax: (714)

ZoneTouch V2 Zone Control System Installation Manual

Transcription:

P a g e 1 Baxter 6201 and 6301 Tips to Avoid Nuisance Alarms Call 1-708-449-7600 x 335 for troubleshooting assistance. Tip #1: To Ensure Door is Closed Correctly You should hear 2 clicks when closing the door. 1.) The metal pin pushing in (part of the slide clamp assembly) 2.) The blue clip popping out Tip #2: To Avoid Occlusion Alarm/Pinched Tubing Tip #3: To Avoid Nuisance Occlusion Alarms If you heard 2 clicks after closing the door (so you know the clip popped out successfully), you can also gently tug the tubing that is coming out at the bottom, down and away from you, toward the back of the pump. (See arrow for where to tug tubing.) - This will ensure an occlusion was not created when closing the door and will also make sure the tubing isn t being pinched at all by the clip. If you ve been using the tubing for a while, try using a fresh section of the tubing by moving the clip down and loading this new section of tubing. Tip #4: To Avoid Nuisance Occlusion Alarms Occasionally the occlusion sensors get sticky. Pushing on the occlusion sensors often loosens them back up. (See illustrations 1 & 2) Tip #5: To Avoid Tubing Misload Alarms and Potential for Cut Tubing When installing the tubing, run your finger from the point where the clip is installed all the way up to the top. This will ensure the tubing is fully inserted into the sensors. See illustration 1

P a g e 2 Proper Loading Instructions 1. Plug the pump in. (The plug icon lights whenever the pump is plugged in. When the pump is not plugged in and is battery-powered, the battery icon will light.) 2. Prepare solution. Prime the set. Ensure all air is expelled from the set. 3. Close set regulating clamp. 4. Open pump door. Raise pump door latch to horizontal position and pull door open. 5. Load the set. a. Press SAFETY CLAMP to open position. b. If your hospital has enabled the slide clamp loading feature, insert the slide clamp on the I.V. set fully into the slot, so that it is flush with the pump housing. (See Illustration 1.) If the slide clamp loading feature is not enabled, a black spring retainer occupies the slide clamp slot and the slide clamp cannot be inserted. c. Load the tubing through the guide channel from bottom to top as shown in Illustration 1. d. Ensure that the tubing is loaded straight through the pump mechanism tubing guides and safety clamp. e. Ensure that the tubing is touching the pumping fingers before closing the pump door. 6. Close pump door. If resistance is felt when closing the door, check for misloaded I.V. set. 7. Open set regulating clamp completely. a. Verify that no drops are falling in drip chamber. If flow is observed, close regulating clamp, recheck I.V. set loading and verify that the proper administration set is being used. b. If flow is again observed, do not use the pump. Have it inspected by service personnel. Note: Always close the administration set regulating clamp(s) before opening pump door and removing set. 8. Attach set to I.V. access site. 9. Turn pump on by pressing the ON/OFF CHARGE key. Verify that the pump performs the following self-test: a. All segments of both displays appear momentarily. b. Three separate audible tones sound. c. If a Hospital Area Designator (HAD) has been programmed into the pump, it appears for 3 seconds. d. The occlusion detection level is momentarily displayed in the message display (LEVEL 1, 2, or 3), followed by AUDIBLE SWITCHOVER if the Audible Switchover option is enabled. e. If Auto Restart and Flow Check are both enabled, the message AUTO RESTART appears for one second following the occlusion detection level display. f. If the pump is plugged into an AC outlet, the plug icon is lit. If the pump is running on battery power, the battery icon is lit. g. If the message INSERT SLIDE CLAMP appears when you close the door, the slide clamp loading option is enabled. Close the I.V. set's regulating clamp, open the door, and insert the I.V. set's slide clamp fully into the slide clamp slot located below the safety clamp. Close the door and open the regulating clamp. 10. Set VOLUME knob on the rear of the pump to the desired level.

P a g e 3 Proper set loading: Make sure blue clamp is inserted into slot firmly. Then run your finger all the way up the line, making sure it is sitting in the correct channel. (Illustration 1) Error Codes If you have a Baxter 6201 or Baxter 6301 with an error code F49 or F94, you can usually reset that error code quickly and easily with the process detailed on pages 5 & 6 of this document - so you won t have to send in your pump! In some cases, the F49 or F94 error code is caused by a bad battery. If after completing this process you believe that this is the case, please contact us so we may help you verify this and if so we will assist you with returning your pump for correction. You can reach our technicians for troubleshooting by calling 800-658-5582 and asking for our Service Department.

P a g e 4 Pump Head Features ITEM 1. Upstream occlusion sensor Pump Head Features FUNCTION Detects a complete tubing restriction upstream of the pump. (Illustration 2) 2. Pumping fingers Cam-driven pumping fingers which manipulate the tubing against the back plate, resulting in fluid movement in a downward direction. 3. Downstream occlusion sensor Detects tubing restrictions downstream of the pump. The sensitivity level can be adjusted to suit the needs of the hospital. 4. Air sensor Detects air bubbles in the tubing. The sensitivity is adjustable via the pump's configuration. 5. Safety clamp Prevents accidental fluid flow by automatically occluding the tubing whenever the pump door is opened. Provides Free Flow Protection. 6. Slide clamp slot When the slide clamp loading option is enabled, the I.V. set's slide clamp must be inserted into this slot. The operator must push the slide clamp into the slot to occlude the tubing before the I.V. set can be removed from the pump. The slide clamp loading option is selectable through the configuration.

P a g e 5 Why do these errors occur? How to Correct Common Failures F-94 and F-49 These two error codes are interrelated in the sense that the method for correcting them both is identical. These errors are typically caused by a total power loss to the pump that has forced the programing to revert back to the manufacturer s default settings and will require a reprograming in order to resume normal operation. The key part of the programing that needs to be reset is the time and date. The following process describes how to accomplish this task quickly and easily. Resetting the time and date to correct F-94 & F-49: 1. Plug the unit in and ensure that it is powered off 2. On the back side of the pump locate the Panel Lock Out button. It will be a small black push button. 3. Press and hold the Panel Lock Out button then turn to the front side of the pump 4. On the front keypad, locate the Stop and On/Off buttons. 5. While continuing to hold the Panel lock out button also press the Stop button. 6. With these two buttons pressed and held, tap the On/Off button and continue holding the two previously mentioned buttons. 7. Monitor the top screen of the pump. You are looking for it to say Modify Config. Once it says this you can release the two buttons. 8. Now that you are in the Modify Config mode, locate the 3 rd key down from the top of the keypad that says Tot Vol/ Status and has the text Next lit above it. 9. This key will be used to cycle through the screens of the Modify Config mode. 10.Press this key approx. 18 times until you see the top screen say Time with some numbers below it. 11.Don t worry about what time the pump is currently saying we are now going to program in the current time at your location in 24 hour or Military format. I.E. if it s 2:34 pm where you are then the time you ll input is 1434. 12.Utilize the number keys to input the current time in the format listed, you ll see that time show up in the bottom screen. 13.Once the time is inputted correctly in the bottom screen, press the Pri Start key and you should see the time move from the bottom screen to the top screen. 14.This indicates that your setting has been accepted. If you don t see this move then you ll need to hit the clear button and repeat this process. 15.If you enter a wrong number at this step or the next simply use the Tot Vol/Status or Next key to cycle through the menus back around to the option you are trying to change. 16.After the time is properly inputted, press the Tot Vol/Status or Next key once more and the screen should display Date followed by numbers for the date. 17.Don t worry about what date the pump is currently saying. We are now going to program in the current date in MM/DD/YY format. I.E. if it is Nov. 19 th, 2011 we ll type in 111911. 18.Utilize the number keys again to input the current date in the format listed, You ll see that date show up in the bottom screen in the following format: 11 19 2011

P a g e 6 19.Once the date is inputted correctly in the bottom screen, press the Pri Start key and you should see the date move from the bottom screen to the top screen and will display as: 11/19/11. 20.You are now done resetting the pump and can use the On/Off button to power down the pump, then power it back up. 21.Upon power up you may see one of two things. It will either come up in the normal mode and be ready for use or it may come up with an additional Common Failure code of F-91 22.F-91 error is a common occurrence after this process and simply needs to have the pump powered off then back on again using the On/Off key. 23.Your pump should now be ready for use. Although they can be caused by other factors, typically these errors are caused by the main battery of the pump being completely drained. After completing this process the pump should be allowed to charge for a solid 12 hours. If after a full charge this problem repeats, please contact UnitedRx IV Department at 708-449-7600 x 335.