Inserter Kit - 120mm Inserter Kit - 130mm MR Lead Insertion Tube with Stylet WayPoint Surgical Kit

Similar documents
WARNING. Filter Service. Section Refer to the following pages for timing and procedures of fi lter element replacement.

USE, CARE & STERILIZATION INSTRUCTIONS. Consumable Accessories Burs Saw Blades K-Wires Steinmann Pins Twist Drills Rasps Ortho Pins/Drills Guide Wires

Installation Instructions Supertop NX Twill

Installation Instructions Cable Top Twill Replace-a-top with Tinted Windows Upper Door Skins not included

Installation Instructions Twill Replace-a-top

Installation Instructions Sunrider

Instructions for use TKR POSITIONERS TKR TKR. Contents

Installation Instructions Supertop NX

Installation Instructions Supertop NX

Installation Instructions Cable Top Twill Replace-a-top with Tinted Windows

Installation Instructions Twill Replace-a-top with Tinted Windows

Installation Instructions Cable Top Twill Replace-a-top with Tinted Windows

Installation Instructions Twill Replace-a-top with Tinted Windows

2 Piece Soft Door Installation Instructions

Installation Instructions Trektop NX Twill

USE, CARE & STERILIZATION INSTRUCTIONS. Consumable Accessories Burs Saw Blades K-Wires Steinmann Pins Twist Drills Rasps Ortho Pins/Drills Guide Wires

Installation Instructions Trektop

Installation Instructions Supertop

Installation Instructions Sailcloth Replace-a-top with Tinted Windows

Snowden-Pencer. Single-Piece Laparoscopic Devices

Supertop NX. Installation Instructions INSTALLATION TIME SKILL LEVEL TOOLS. Vehicle Application:

Installation Instructions Fabric Replacement Top with Tinted Side

Installation Instructions Sailcloth Replace-a-top with Tinted Windows

Installation Instructions Twill Replace-a-top

Vercise PC Implantable Pulse Generator

Installation Instructions

Installation Instructions Sailcloth Replace-a-top with Tinted Windows

Fabric Replacement Top and Doors with Tinted Side and Rear Curtains Installation Instructions

Installation Instructions Trektop NX

Installation Instructions Supertop for Truck

Installation Instructions Trektop NX

Installation Instructions Supertop for Truck

Installation Instructions Trektop NX

Installation Instructions Supertop for Truck

Installation Instructions Halftop

Consensus Orthopedics Inc. Bespoke ACO Polyamide Acetabular Cup Orientation (ACO) guides and ACO bone models

OPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-707A & AEU-707AV2 Implant / Surgery Systems

OPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-17B & AEU-17BV2 Implant / Surgery Systems

Installation Instructions Supertop with Tinted Side and Rear Windows

Automatic Roof Hatch Opener

Top Down Rollstar Shade Installation Instructions

PFadvantage MF 6850/6855; Ideal 9080/9090

Replay Fabric Replacement Top

Installation Instructions Supertop with Tinted Side and Rear Windows

Full Top Deck Kit. Installation Instructions

Tools, Equipment and Supplies Needed:

Fabric Replacement Top Installation Instructions

Table of Contents pg Rev. A

HYDRAULICS. TX420 & & lower. Hydraulic Tandem Pump Removal. 4. Remove the LH side panel (Fig. 0388).

Installation Instructions Supertop for Truck

Installation Instructions Specialty Tigertop

ALWAYS DISCONNECT DISPENSER FROM WATER SOURCE WHEN DISPENSER IS NOT IN USE.

Replace - A - Top Installation Instructions

Trackstar Motorized Folding Shade Installation Instructions

INSTALLATION & OWNER S MANUAL

FLOWSERVE CORPORATION NOBLE ALLOY VALVE BALL VALVE REPAIR INSTRUCTION

Vercise TM DBS Charging Handbook

COYOTE Dome Closure 9.5" x 28" with Transition Tray for High Density Splice Applications

Heavy Duty Miniature Quick-Change Applicator (Side-Feed Type) with Mechanical or Air Feed Systems

710 Splicing Rig. Instructions. January E

Convertible Top Installation Instructions

SPC-P Smart Power Connection with Temperature Monitoring Transmitter Pipe Mounted Installation Instructions

2001 Dodge RAM 3500 PICKUP

Installation Instructions Specialty Tigertop Includes Doors

INSTALLATION & OWNER S MANUAL

NORTHSTAR TRAILERS. Assembly Guide for SPORTSTAR II Trailer

ASSEMBLY. Transmission Automatic Transaxle/Transmission. Special Tool(s) Alignment Set, Fluid Pump 307-S039 (T74P X) Special Tool(s)

Installation Instructions TJ Tinted Window Kit

Installation Instructions For: Suzuki Samurai, All Years Part Number: TM

Installation Instructions Tigertop

Rollstar Shade Installation Instructions

Installation Instructions Headliner

3M Cold Shrink QT-III Silicone Rubber Three-Core Termination Kits

Installation Instructions PowerBoard Automatic Retracting Running Board

Installation Instructions PowerBoard Automatic Retracting Running Board

INSTALLATION & OWNER S MANUAL

Cleaning and Rotor Replacement Instructions

1.1 Quad Valve and Tri-Valve. Contents WATER DUMP BODY O-RING EXHAUST VALVE QUAD VALVE TIE WRAP EXHAUST MAIN BODY O-RING

SERVICE PARTS LIST 1-1/8" ROTARY HAMMER MILWAUKEE ELECTRIC TOOL CORPORATION W. LISBON RD., BROOKFIELD, WI Drwg.

665 Series Dispenser. Metering tip kit

Wrangler Replay Top with Tinted Windows Installation Instructions For: Jeep Wrangler Unlimited (JK) 4 Door 2007 and Newer Part Number: 51201

Maintenance Information

JBS-100-STB Single Entry Power Connection with Junction Box Installation Instructions

Installation Instructions Front 2 Piece Door with Rotary Latch

Installation Instructions Sunrider

8 Breast Biopsy. 8.1 SENSE Breast 7 coil. Breast biopsy can be performed using the SENSE Breast 7 coil.

AutoDose Automatic Dispensing System

Tooling Assistance Center

Read this document carefully before implementing the upgrade!

Maintenance Information

C15C C15C. Page 1 of 20

Maintenance Information

Replace-a-Top. Installation Instructions. 60th Anniversary Top INSTALLATION TIME SKILL LEVEL TOOLS. Vehicle Application:

THE GLIDER 5th Wheel Attachment

Installation Guide 3x1 and 4x2 Cable

Installation Instructions Tigertop Includes 1-Piece Full Soft Doors

For: Model Year Jeep Wrangler/TJ Part Number: For: Model Year Jeep Wrangler/TJ Part Number: 58709

JUL 17 Rev M

Electric motor testing

2013 RT / 2014RT / 2015 RT - Shock Spring Adjuster Installation Instructions

Transcription:

STarFix Platform System Directions for Use Including: 66-WP-ID 66-WP-ID-01 66-IT-MR1 66-WP-MR Inserter Kit - 120mm Inserter Kit - 130mm MR Lead Insertion Tube with Stylet WayPoint Surgical Kit L011-61 (Rev C0 April 2015) FHC, Inc. 1201 Main Street Bowdoin, ME 04287 Fax +1-207-666-8292 www.fh-co.com 24 hour technical service: 1-800-326-2905 (US & Can) +1-207-666-8190 FHC Europe (TERMOBIT PROD srl) 42A Barbu Vacarescu Str, 3rd Fl Bucharest 020281 Sector 2 Romania FHC Latin America Calle 6 Sur Cra 43 A-200 Edificio LUGO ofi cina 1406 Medellín-Colombia L011-61 (Rev C0 April 2015) 1

2 STarFix Platform System Directions for Use

Contents 4 4 4 5 6 7 8 9 10-11 11-18 Indications for use Warnings and Cautions Storage Components Handling the 66-IT-MR1 Cleaning of sterilizable components (Manual/Automated) Sterilization Platform Inserter Height Chart Pre-Use Check Procedure for MRI Implantation of Lead L011-61 (Rev C0 April 2015) 3

Indications for use: The STarFix Platform System is intended to be used with commercially available stereotactic systems for neurosurgical procedures which require the accurate positioning of microelectrodes, stimulating electrodes, or other instruments in the brain or nervous system. Warnings WARNING: Federal law restricts this device to sale by or on the order of a physician. WARNING: Handle all components with extreme care. They may be damaged if excessive force or incorrect handling occurs. WARNING: User must examine unit prior to use to ensure it has no shipping damage. Cautions CAUTION: Insertion tubes and all SLS components are for Single Patient use only. Do not reuse any single use components. CAUTION: Insertion tubes are not intended for implantation. Storage Store all components at temperatures between 34 C (-29 F) and 57 C (135 F). Do not exceed 135 F for long-term storage. Specifi cations Array spacing: 2.00 mm from center Array guide hole diameter: 1.88mm Bushing configuration: center hole on stereotactic axis with 4 holes offset by 2.00 mm on center and orthogonal to the center hole. Lead Electrode IT and stylet: Dimensions/material: 1.8 mm OD, 1.6 mm ID/ Polyimide, Ceramic coated with epoxy STar SteriSuite: Material: Electropolished Stainless Steel Insert: Nylabond coated Stainless Steel 4 STarFix Platform System Directions for Use

5 6 7 8 9 10 11 4 3 2 12 1 13 1. Center Positioner 2. Offset Positioner 3. Inserter 4. Lead Holder 5. Center Hub 6. Entry Offset Hub 7. Target Offset Hub 8. Thumbknobs 9. LeadLoc Depth Stops 10. Lead Measurement Fixture 11. Burr Hole Marker 12. Drape Support 13. Hex Wrench 14. STar SteriSuite Case 15. Standoff Wrench 16. Manual Handle 17. Combination Driver 18. Burr Hole Marker Bushing 19. Sterilization Tray 18 17 16 15 14 19 66-WP-MR includes: 11, 13, 14, 15, 16, 17, 18, 19 66-WP-ID / 66-WP-ID-01 includes: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 12 WARNING: These items are single use. Do not reuse. 66-IT-MR1 WARNING: Metal items should not be used in the MRI environment or in the presence of an MRI magnet. L011-61 (Rev C0 April 2015) 5

Handling the 66-IT-MR1 WARNING: Stylet should not be subjected to forces that cause it to bend more than 1mm. Insertion tubes and stylets shown in shipping package 3. Place insertion tube/stylet into the support slots of the sterilization tray, as shown. To secure, apply enough pressure on the side of the silicone support to deflect 1. Keep stylet in insertion tube during all handling operations. about 1cm and gently push the tube down 1mm. Release the silicone which will secure the tube in the slot. Repeat at the silicone support at the other end. 4. Secure the lid to the base using the 2 latches. 5. Sterilize following recommended sterilization parameters. 2. Remove insertion tube and stylet from the shipping package prior to sterilization by first spreading the blue foam with fi rst and middle fi ngers and lifting the assembly to the foam surface. Then repeat at the other end. 6. To remove the tube/stylet for use, apply enough pressure on the side of the silicone support to deflect about 1cm and gently lift the tube about 1mm to free it from the slot. Repeat at the silicone support at the other end. 6 STarFix Platform System Directions for Use

Cleaning Refer to system diagram on page 5 for the components identifi ed with letters, listed below. Note that insertion tubes and SLS components should not be cleaned prior to use. These components should only be sterilized following the protocol on page 8. Manual Cleaning Items Covered Protocol STar SteriSuite case needed (14) Wrenches (13, 15) Tools (11, 18) Manual Handle (16) Combination Bit (17) 1. Prepare the detergent according to manufacturer s recommendations: Asepti Wash Plus liquid (2.5 ml per liter or 1/4 oz per gal), using warm tap water. 2. Separate the Wrenches, Tools, Manual Handle, and Combination Bit, and immerse them in the wash solution for a minimum of 5 minutes. Actuate the devices during the soak. 3. Using a clean soft cloth that has been soaked in the detergent, wipe the tray, and its insert, to remove any visible soil. Use a soft bristle brush and syringe to reach hard-to-clean areas. 4. Place the components back in the tray. 5. Prepare the detergent in a sonication unit according to manufacturer s recommendations: Asepti Wash Plus liquid (2.5 ml per liter or 1/4 oz per gal). 6. Immerse the tray in the sonication unit and sonicate for a minimum of 10 minutes. 7. Rinse all components with running reverse osmosis/de-ionized water to remove any residual detergent. 8. Dry components using a clean soft cloth. 9. Visually inspect to ensure all visible soil is removed. Automated Cleaning Items Covered Protocol STar SteriSuite case needed (14) Wrenches (13, 15) Tools (11, 18) Manual Handle (16) Combination Bit (17) Phase Pre-Wash 1 Recirculation Time (min) 2:00 Water Temperature Cold Tap Water (16 C maximum) Ecolab Inc. detergent (2.5 ml/l or 1/4 oz/gal) N/A Ecolab GmbH detergent (2.5 ml/l or 1/4 oz/gal) N/A Enzyme Wash 2:00 Hot Tap Water (43 C minimum) Asepti Wash Plus Sekusept AR Wash 1 2:00 65.5 C (Set Point) Asepti Wash Plus Sekusept AR Rinse 1 2:00 Heated Water (66.0 C) N/A N/A Pure Water Rinse 0:10 Heated (66.0 C) Asepti Rinse Sekusept FNZ or Sekumatic Multiclean Dry Phase 7:00 115 C N/A N/A L011-61 (Rev C0 April 2015) 7

Sterilization Method Platform/ Drape Support Platform Adapter Kit Insertion Tubes Protocol Preconditioning Set Points: Wrapped Sterilization Parameters: EO Sterrad Temperature: 54 C Relative Humidity: 40% Vacuum Set Point: 1.5 psia Preconditioning Time: 60 min Sterrad 100S full cycle Temperature: 54 C Relative Humidity: 40% EO Concentration: 725 mg/l Gas Exposure Time- (full cycle SAL 10-6): 120 minutes Aeration Time: 12 hours Aeration Temperature: 55±4 C Steam Gravity Steam Sterilization (Flash): Minimum Temperature, Unwrapped: 132 C Full Cycle Time: 10 minutes Minimum Dry Time: none Prevacuum Steam Sterilization: Preconditioning Pulses: 3 Minimum Temperature, Wrapped: 132 C Full Cycle Time: 4 minutes Minimum Dry Time: 40 minutes Steam Gravity Steam Sterilization: Minimum Temperature, Wrapped: 132 C Full Cycle Time: 30 minutes Minimum Dry Time: 35 minutes Prevacuum Steam Sterilization: Preconditioning Pulses: 3 Minimum Temperature, Wrapped: 132 C Full Cycle Time: 12 minutes Minimum Dry Time: 30 minutes Steam Gravity Steam Sterilization: Minimum Temperature, Wrapped: 132 C Full Cycle Time: 30 minutes Minimum Dry Time: 35 minutes Prevacuum Steam Sterilization: Preconditioning Pulses: 3 Minimum Temperature, Wrapped: 132 C Full Cycle Time: 4 minutes Minimum Dry Time: 20 minutes WARNING: Users should be aware that the effects of unvalidated sterilization protocols could result in damage to the components and affect their functioning or performance. The components of this system are not validated for use with alternative sterilization protocols, and FHC does not recommend or endorse their use. Users with questions regarding this safety issue should contact FHC s Technical Service Department at 1-207-666-8190. WARNING: DO NOT attempt to sterilize the insertion tubes using the package they were shipped in. Remove insertion tubes from the package prior to sterilization. FHC recommends use of the FHC sterilization tray. WARNING: The STarFix Platform is shipped non-sterile and must be sterilized before use. Invert the Platform on its hub surface with the legs pointing up during sterilization. 8 STarFix Platform System Directions for Use

Platform Inserter Height Chart 13 mm Target Target 40CM Lead Implant 319mm 215.5mm 130mm Trg - 120mm Trg - 130mm 120mm

Pre-Use Check 1. Confi rm that there are no contaminants or debris on any of the parts. Lettering has been darkened for visual representation 15mm 25mm 8mm 8mm 25mm 25mm 4. Pull the stylet up 2cm out of the insertion tube. Ensure that the stylet tip retracts into the insertion tube. 15mm 25mm 2. Confi rm all thumbknobs are present. Note that a sterile marking pen may be used to mark the screw heads prior to use, making it easier to differentiate between them during the surgical procedure. WARNING: Delicate insertion tubes should always be carefully inspected before use because damage to them can cause targeting errors and impact patient safety. Handle insertion tubes carefully to prevent bending. 3. Test mount the lead holder on the inserter, confi rm there are no stripped threads or looseness. 5. Place the center hub into the integrated indexing ring. Secure the hub with the two indexing ring screws. Ensure correct fit and seat of the hub within the ring. Repeat for the second side for a bilateral platform. 10 STarFix Platform System Directions for Use

Procedure for MRI Implantation of Lead 6. Place the two 8mm thumbknobs in the free indexing ring tapped holes and thread them in lightly. Mount the drape support as shown, tightening the thumbknobs until the drape support is held securely. 1. Attach and handtighten standoffs to anchors. 2. Mount platform on standoffs using thumbknobs. WARNING: Do not overtighten thumbknobs. Knobs should be tightened until they make contact with the mating surface and then turned an additional 1/4 for security. 7. Secure the center positioner to the inserter. 8. Install the insertion tube with stylet into the inserter. When the inserter and positioner are mounted on the platform in the center hub, the tip of the stylet will be exactly at the predicted target. 9. Disassemble all parts and lay them out in preparation of the start of the surgery. 3. Install center hub in indexing ring (built into platform). Tighten both indexing ring screws, which should be positioned 90º apart. L011-61 (Rev C0 April 2015) 11

4. Insert burr hole marker bushing into center hub. 7. Remove burr hole marker bushing. 8. Remove platform and create burr hole. 5. Insert the burr hole marking tool down through the bushing. 6. Create indentation in skin/skull for location of burr hole and remove marking tool. 9. Reattach Platform. 12 STarFix Platform System Directions for Use

WARNING: If any error or erratic function is observed, discontinue use of the Platform System immediately and evaluate the potential impact to patient safety before continuing its unmitigated use. WARNING: Always confirm the tightness of thumbknobs before beginning the procedure. WARNING: While often snug, all tubes used with the microtargeting Inserter System have been designed to be inserted and removed by hand. Any other tool should be used only as a last resort. 10. Install center positioner in center hub. Ensure that hub thumbknob is positioned 135 from indexing ring thumbknobs. The tab on the center positioner should be positioned opposite the thumbknob. Examine burr hole to ensure entry area is clear. If necessary, remove the center hub and install the entry offset hub. WARNING: The insertion tube will enter the brain at this stage. WARNING: Never move the insertion tube in the brain without a stylet or lead inside. WARNING: The stylet should not be removed until tube is inserted in brain. WARNING: The insertion tube and stylet are fragile and should be handled with care. WARNING: Incorrect thumbknob orientation may affect targeting accuracy. WARNING: The Entry Offset Hub creates a new trajectory which is orthogonally offset 3mm from original at approximate entry depth, and coincides with the original at the target. 12. Insert the MRI safe insertion tube with the stylet in through the center hole. Ensure that the insertion tube enters the brain without hitting the edges of the burr hole. 11. Mount inserter on the center positioner and tighten the thumbknob. The thumbknob of the inserter should be in alignment with the tab of the positioner. WARNING: When there is an insertion tube in the brain, every effort should be made to minimize lateral forces to the microtargeting Inserter System as it can translate into significant lateral movements of the tube in the brain. WARNING: The insertion tube has no electrical conductivity and cannot be used as an electrical common. L011-61 (Rev C0 April 2015) 13

In the MRI Environment: 16. Take MRI image to confi rm location for DBS implant. 17. In WayPoint Navigator, load MRI image and merge with platform plan. 13. Place the two 8mm thumbknobs in the free indexing ring tapped holes and thread them in lightly. 18. If the position is not correct, reposition using offset positioner and/or hub. Alternately, use one of the four available tracks in the ben-gun array. Repeat steps 10-11. front view back view WARNING: The Target Offset Hub creates a new trajectory with target area orthogonally offset 6mm from the original at the target. 14. Mount the drape support as shown. The center positioner allows for investigation of nine separate tracks. The selected, or center, trajectory remains constant regardless of orientation. Four parallel tracks at 2mm orthogonal offset are achieved with alignment to locations A,C,E and G (designated as the black dots below) and four additional 45 degree 2mm Euclidean tracks at locations B,D,F and H (designated as red). The offset positioner can be oriented in any of eight directions yielding an additional forty tracks. Locations A,C,E and G offset the entire fi ve hole matrix 3mm orthogonal to the origin (orange, green, yellow and blue). While locations B,D,F and H (magenta, gray, cyan, turquoise) offset the matrix 45 degrees at 3mm. 15. Tighten the thumbknobs until the drape support is held securely. The thumbknobs should be fl ush with the outer edge of the drape support. The dimensioned grid can be utilized to determine the precise location of all available track changes. Choose the position and track closest to the desired grid move. 14 STarFix Platform System Directions for Use

In the non-mri Environment: 21. Release pressure on the LeadLoc. 19. Secure lead holder to the DBS measurement fi xture. LeadLoc fl ats 20. Pinch the tabs of the LeadLoc together slightly and slide the distal end of the lead through the top of the LeadLoc to approximately 25 mm from the lead stylet handle, making sure the lead is in the center notch of the LeadLoc. 22. Insert the lead and the LeadLoc into the lead holder making sure the fl ats of the LeadLoc are aligned with the opening on the lead holder. Gently tighten the thumbknob to secure the LeadLoc into the holder then loosen slightly, ensuring that the LeadLoc cannot move vertically. WARNING: Do NOT over-pinch the tabs. Over-pinching the tabs could result in the LeadLoc not holding the lead. WARNING: When sliding the lead inside the LeadLoc, ensure the lead is inserted from the top hole down through the bottom hole of the LeadLoc and does not come out of the center slot. 23. While pinching the LeadLoc, move the lead so that the contacts are in the desired position relative to target depth. 24. Release pressure on the LeadLoc to secure the lead. To confi rm the lead does not move during the insertion process, use an appropriate (sterile) pen to make a mark on the lead just above where it comes out of the LeadLoc. L011-61 (Rev C0 April 2015) 15

25. Loosen the lead holder and remove from the measuring fi xture. 27. Remove drape support by removing the two 8mm thumbknobs and then lifting the drape support straight up. 26. Remove the LeadLoc with lead from lead holder. 28. Remove stylet. Be sure to hold the collar of the insertion tube down to prevent movement. 16 STarFix Platform System Directions for Use

LeadLoc fl ats 29. Secure the lead holder to the inserter. 31. Carefully secure the LeadLoc (and lead) in the lead holder making sure the fl ats of the LeadLoc align with the opening on the lead holder. 32. Confi rm the lead location. 30. Insert the lead through the insertion tube and into the brain. If the center track is not being used, the lead should be placed down through the selected track. WARNING: Observe the exposed segment of the lead while advancing the inserter and ensure it advances into the insertion tube without binding or bending. 33. Loosen and raise the lead insertion tube out of the skull. 34. Secure the lead next to the skull with smooth tweezers or the selected securing device like Medtronic Navigus cap. 35. Remove the stylet from the lead. L011-61 (Rev C0 April 2015) 17

36. Loosen the lead holder thumbknob only slightly, ensuring that it still retains the LeadLoc and that the LeadLoc cannot move vertically. 37. Pinch the LeadLoc and pull the lead down through the drive. WARNING: Trying to pull the LeadLoc over the lead may dislodge the lead from its implantation location. 40. Insert the combination driver into the manual handle. 41. Place the hex wrench over the fi rst anchor to be removed. Drop the combination driver down into the hex wrench until it engages with the anchor. Twist the hex wrench counterclockwise to remove the anchor. 42. Repeat the step above for all remaining anchors. 38. Remove the mounted equipment and the platform. 39. Remove the standoffs using the standoff wrench. 18 STarFix Platform System Directions for Use