7777D In-Line Mounting for Durafet II ph Electrodes Installation and Maintenance

Size: px
Start display at page:

Download "7777D In-Line Mounting for Durafet II ph Electrodes Installation and Maintenance"

Transcription

1 7777D In-Line Mounting for Durafet II ph Electrodes Installation and Maintenance Rev. 0 3/07 Honeywell Process Solutions Industrial Measurement and Control

2 Copyright, Notices, and Trademarks Printed in U.S.A. Copyright 2007 by Honeywell Revision 0 March 2007 WARRANTY/REMEDY Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Contact your local sales office for warranty information. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace without charge those items it finds defective. The foregoing is Buyer's sole remedy and is in lieu of all other warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use. While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the application. Honeywell Process Solutions Industrial Measurement and Control 512 Virginia Drive Fort Washington, PA Durafet II is a trademark of Honeywell Hastelloy is a trademark of Haynes International, Inc Teflon is a trademark of E. I. DuPont De Nemours & Company, Inc. Other brand or product names are trademarks of their respective owners. ii 7777D In-Line Mounting Installation and Maintenance 3/07

3 About This Document Abstract This manual provides in-line installation instructions for Durafet II ph Electrodes as part of a 7777D electrode assembly. Contacts World Wide Web The following lists Honeywell s World Wide Web sites that will be of interest to our customers. Honeywell Organization Corporate Industrial Measurement and Control Technical Tips WWW Address (URL) Telephone Contact us by telephone at the numbers listed below. Organization Phone Number United States and Canada Honeywell Tech. Support Service 3/ D In-Line Mounting Installation and Maintenance iii

4 Symbol Definitions The following table lists those symbols used in this document to denote certain conditions. Symbol Definition This CAUTION symbol on the equipment refers the user to the Product Manual for additional information. This symbol appears next to required information in the manual. WARNING PERSONAL INJURY: Risk of electrical shock. This symbol warns the user of a potential shock hazard where HAZARDOUS LIVE voltages greater than 30 Vrms, 42.4 Vpeak, or 60 Vdc may be accessible. Failure to comply with these instructions could result in death or serious injury. ATTENTION, Electrostatic Discharge (ESD) hazards. Observe precautions for handling electrostatic sensitive devices Protective Earth (PE) terminal. Provided for connection of the protective earth (green or green/yellow) supply system conductor. Functional earth terminal. Used for non-safety purposes such as noise immunity improvement. NOTE: This connection shall be bonded to protective earth at the source of supply in accordance with national local electrical code requirements. Earth Ground. Functional earth connection. NOTE: This connection shall be bonded to Protective earth at the source of supply in accordance with national and local electrical code requirements. Chassis Ground. Identifies a connection to the chassis or frame of the equipment shall be bonded to Protective Earth at the source of supply in accordance with national and local electrical code requirements. iv 7777D In-Line Mounting Installation and Maintenance 3/07

5 Contents 1. INTRODUCTION Overview Description About This Manual SPECIFICATIONS Physical Pressure and Temperature Limits Model Selection Guide Series 777D Durafet II Electrode GENERAL ASSEMBLY INSTRUCTIONS Selecting Materials of Construction Sealing Pipe Joints Durafet II Electrode Preparation Cap Adapter Option for Durafet II Electrodes Overview of Assembly ELECTRODE CABLE CONNECTIONS Possible System Configurations Electrode Cables Cabling Procedure CONNECTING A DURAFET II ELECTRODE TO A UDA2182 ANALYZER (DIRECT CONNECTION) CONNECTING A CAP ADAPTER Cap Adapter Connection to a UDA2182 Analyzer Connection to an APT Series ph Transmitters and Analyzers IN-LINE MOUNTING General Information MAINTENANCE, STANDARDIZATION, AND REPLACEMENT PARTS Maintenance Standardization Replacement and Accessory Parts / D In-Line Mounting Installation and Maintenance v

6 Tables Table 3-1 Installation Tasks for In-Line Electrode 5 Figures Figure 5-1 Terminal Designations for Durafet II Electrode 8 Figure 6-1 Terminal Designations for Durafet II Electrode with Cap Adapter 9 Figure 6-2 Connecting to an APT2000 ph Transmitter 10 Figure 6-3 Connecting to an APT4000 ph Analyzer 10 Figure 7-1 Proper Mounting Angle for Electrode 12 vi 7777D In-Line Mounting Installation and Maintenance 3/07

7 1. Introduction 1.1 Overview The 7777D Series of electrode mountings enables a user to interface a Durafet II ph electrode to a process by mounting it into a ¾ threaded tee connection. Honeywell's Durafet II Series of industrial electrodes have been designed to provide accurate and stable ph measurements for a wide variety of industrial applications. The Durafet II ph electrode is a non-glass ph electrode. The ph measurement is based in ISFET (Ion Selective Field Effect Transistor) technology, which results in a solid state ph sensing electrode. The Durafet ph electrode is much more rugged than the traditional glass ph electrode. The sturdy ph electrode reduces replacement and inventory costs. The solid state sensing element is packaged in a durable Ryton body that results in a ph electrode that has extended life in a wide variety of process conditions. The ISFET technology also produces an electrode that is up to 10 times faster than glass electrodes. This fast response improves product quality and provides better process control to optimize chemical usage. For Immersion applications use the Durafet III ph electrodes. 1.2 Description Durafet II ph electrodes for the 7777D Series are designed to meet various customer application and installation needs. The electrode body is 1 in diameter with ¾ threads at both ends of the electrode. For in-line applications the electrode is threaded into one leg of a pipe tee. A temperature sensor is mounted internal to the electrode to measure process temperature and provide a temperature signal for automatic (Nernstian) temperature compensation. Two temperature sensors are available: 8550 ohm thermistor and 1000 ohm RTD. The electrodes come with a connector with a separate cable for in-line installations. Various cable length options are selected based on installation requirements. The Durafet II ph electrode is compatible with various ph instrumentation: Honeywell UDA2182 Universal Dual Analyzers. (Direct connection or with Cap Adapter) APT2000/4000 Transmitter/Analyzer with Cap Adapter. Non-Honeywell instruments with Cap Adapter. Consult your Honeywell sales representative for a list of instruments with this compatibility. 1.3 About This Manual This manual is part of a set documenting installation and use of the 7777D Durafet II ph Electrodes. The set consists of the following manuals. This manual Instruction Manual Durafet II Industrial ph Electrode Series Instruction manual for the ph instrument 3/ D In-Line Mounting Installation and Maintenance 1

8 2..Specifications 2.1 Physical Materials in contact with process solution Dimensions Electrode: Polyphenylene sulfide (PPS) body, silicon sensor, Viton seal, ceramic junction, EPM O Rings. Electrode Models: 2.5 cm (1 in.) diameter, 17.1 cm (6.75 in.) long 2.2 Pressure and Temperature Limits Temperature Pressure Electrodes: See manual supplied with the electrodes. Cap Adapter: 0 C to 70 C (32 F to 158 F) Electrodes: See manual supplied with the electrodes. Cap Adapter: 10 psig D In-Line Mounting Installation and Maintenance 3/07

9 2.3 Model Selection Guide Series 7777D Durafet II Electrode Instructions Select the desired Key Number. The arrow to the right marks the selection available. Make one selection each per table using the column below the proper arrow. A dot ( ) denotes unrestricted availability. Key Number I II III _, KEY NUMBER Selection Availability Description 7777 Durafet II Electrode Mounting 07777D TABLE I - Mounting Type, Temperature Sensor, Part No. Mounting Temperature Sensor Reference Fill Electrode P/N Immersion Use Durafet III (MSG# ) for all immersion applications. In-Line 8550 Ohm Thermistor KCl Ohm RTD KCl TABLE II - Instrumentation to Sensor Cables Selection Cable Length Cable Type Part Number None feet (6,10 meters) Cap Adapter feet (15,24 meters) Cap Adapter feet (6,10 meters) UDA2182 (Direct Input) feet (15,24 meters) UDA2182 (Direct Input) TABLE III - Options Tagging Linen Customer I.D. Tag L (limit to 3 lines, 22 spaces per line) Stainless Steel Customer I.D. Tag S (limit to 3 lines, 22 spaces per line) Certificate of Yes 10 Calibration 3/ D In-Line Mounting Installation and Maintenance 3

10 3. General Assembly Instructions 3.1 Selecting Materials of Construction Preamplifier and electrode The materials used in the construction of the electrode are listed in the specifications. Materials of wetted parts must be compatible with the process temperature and corrosion conditions. User-supplied components Pipe couplings, and tees must be supplied by the user. Select materials that are compatible with the process temperature and corrosion conditions. 3.2 Sealing Pipe Joints When making pipe joints apply Teflon tape pipe sealant to male threads. Wrap the threads with the tape overlapping by 50% on each wrap. Start the wrap at the end of the pipe and wrap in the direction of the thread at least two turns. When installing an electrode, apply Teflon tape to the threads, then hand-tighten the electrode in the fitting only until snug. If necessary to stop a leak, a wrench may be applied to the electrode s wrenching flats to further tighten the joint only until the leak has stopped. Always leave at least one thread on the electrode showing outside the fitting. CAUTION When tightening an electrode never exceed 15 ft-lb applied torque. Applying too much torque can result in damage to the electrode. 3.3 Durafet II Electrode Preparation Prepare Durafet II electrode as described in the electrode manual, part number CAUTION Before using the Durafet II ph electrode, the cable must be connected to the electrode and it must be allowed to warm up for at least 10 minutes. 3.4 Cap Adapter Option for Durafet II Electrodes The Cap Adapter option for Durafet II electrodes eliminates the need to separately mount a preamplifier. The Cap Adapter is a preamplifier electronics module that is integral with the electrode-to-instrument cable. The preamplifier PCA is molded into the cap of the quick-disconnect connector. The cable has tinned leads that are ready to be wired directly to a UDA2182 Analyzer or APT2000/4000 Transmitter. The Cap Adapter mounts directly on the Durafet II electrode. The electrode has a key that mates with the keyway in the connector on the cable. The cap is then threaded onto the electrode. The cap should be handtightened and engage the O-ring at the top of the electrode. The knurled fitting at the top of the cap should be hand-tightened to seal an internal O-ring around the cable. When properly threaded onto the electrode the connection is waterproof D In-Line Mounting Installation and Maintenance 3/07

11 3.5 Overview of Assembly Instructions for unpacking, preparing, and maintaining the electrodes are in the manual supplied with the electrodes. Instructions for cabling the electrode directly to the instrument are in Section 5 of this manual. Instructions for connecting a Cap Adapter to an instrument are in Section 6. The recommended configuration for in-line mounting of electrodes is described in Section 7 of this manual. Installation tasks Table 3-1 lists the tasks required for in-line mounting of electrodes. Step Table 3-1 Installation Tasks for In-Line Electrode Action 1 Read the general information about in-line mounting at the beginning of Section 7. 2 Select an appropriate location and install the required pipe tee. 3 Obtain the required materials listed in Section Thread electrode into pipe tee. 5 Connect the electrode cable to the electrode as described in Section 4. 6 Connect the electrode cable directly to the analyzer as described in Section 5 or Section 6 for the Cap Adapter. 3/ D In-Line Mounting Installation and Maintenance 5

12 4. Electrode Cable Connections 4.1 Possible System Configurations Variety of applications The Durafet II electrode can be cabled to: Directly to the input terminals of a UDA2182 analyzer To the UDA2182 (preamp input option) or APT2000/4000 transmitter when the Cap adapter is specified. 4.2 Electrode Cables The Durafet II ph electrode are supplied with a connector at the electrode top to provide a quick disconnect for easy electrode replacement. Cable connectors The electrode cable terminates with an edge metal connector on one end and tinned leads at the UDA2182 analyzer end. Cap Adapter cables have tinned leads at the analyzer end. The edge connector mates with the Durafet II electrode. Avoid contamination The connectors are constructed with high resistance insulating material which can be contaminated if exposed to oil and salts from bare hands. Avoid contaminating the internal area of the connectors. CAUTION Do not allow liquids or other foreign matter to contact the cable connectors. Save the protective cap supplied with electrodes with connectors and put the cap on the connector whenever the cable is not installed on an electrode. How to clean the connectors The connectors can be cleaned by wiping with a cotton swab moistened with isopropyl or grain alcohol. Allow connectors to dry completely before making connections. 4.3 Cabling Procedure How to attach the cable to the electrode Remove the protective cap from the cable. Loosen the ferrule and slide back the ferrule, small o-ring, and plastic cover to expose the cable connector. Align the cable connector keyway with the electrode key (small black tab on the top of the electrode). Slide the plastic cover over the electrode end and hand-tighten the cover on to the electrode D In-Line Mounting Installation and Maintenance 3/07

13 ATTENTION Make sure the o-ring on the electrode is seated on the shoulder of the electrode, below the threads. CAUTION Be careful not to cross the threads on the connector. Slide the o-ring and ferrule down the cable to the back of the cover cap. Hand-tighten the ferrule onto the cover cap. How to remove cable from electrode Loosen the ferrule from the cover cap and slide the ferrule and o-ring back on the cable. Loosen the cover cap and slide back on the cable. Pull the cable connector off the electrode. CAUTION Keep cable connector and edge connector of the electrode clean and free from contamination. Put protective caps that are supplied with the electrode and cable back in place while they are disconnected. 3/ D In-Line Mounting Installation and Maintenance 7

14 5. Connecting a Durafet II Electrode Directly to a UDA2182 Analyzer (Direct Input No Cap Adapter) Cable shield (yellow) to chassis ground screw Wire Color Green Green with Black stripe Blue Orange Signal Name R KO res- (Low) R KO res- (High) Drain Source Red Black Substrate Reference 9 8 White with Black stripe Orange with Black stripe White Red with Black stripe Counter RTH 3 rd Wire RTH Low RTH High Remove pre-wired jumper at terminals 5 & Figure 5-1 Terminal Designations for Durafet II Electrode D In-Line Mounting Installation and Maintenance 3/07

15 6. Connecting a Cap Adapter 6.1 Cap Adapter The Cap Adapter is an integral part of the electrode cable. It is essentially a preamplifier that does not require separate mounting. The output from the Cap Adapter can be connected directly to a ph instrument. The Cap Adapter and cable for Durafet II electrodes are available in lengths of 20', and 50'. One end of the Cap Adapter cable is the mating connector to the edge connector of the Durafet II electrode. The other end of the cable terminates with tinned leads. The tinned leads connect to the input terminals of the ph instrument as described below. 6.2 Connection to a UDA2182 Analyzer The Cap Adapter is connected to a UDA2182 analyzer. The UDA2182 input card must be a preamp input type. Connect the tinned leads of the Cap Adapter cable as follows: Wire Color Blue Green Black Signal Name (+) Volt Supply (-) Volt Supply Supply Common 12 Orange 11 ph Input Signal Red Red with Black stripe White RTH 3 rd Wire RTH Low RTH High Remove pre-wired jumper at terminals 5 & 6 3 Cable shield (yellow) to chassis ground screw 2 1 Figure 6-1 Terminal Designations for Durafet II Electrode with Cap Adapter 3/ D In-Line Mounting Installation and Maintenance 9

16 6.3 Connection to an APT Series ph Transmitters and Analyzers The Cap Adapter can be connected to the APT2000 ph Transmitter. (See the APT2000 ph Transmitter Manual, part number , for more detailed description of the APT2000.) +3 V/0.5 ma GND -3 V/0.4 ma Do not connect + 4 to 20 ma _ T1 T2 N.C. Aux. el. Reference el. Meas. el BLUE BLACK GREEN WHITE RED/BK YELLOW WT/BK NOTE: Orange and Red wires are not typically connected. These should be clipped and electrically sealed to avoid possible contact with other conductors. Figure 6-2 Connecting to an APT2000 ph Transmitter The Cap Adapter can be connected to the APT4000 ph Analyzer. (See the APT4000 ph Analyzer Manual, part number , for more detailed description of the APT4000.) Figure 6-3 Connecting to an APT4000 ph Analyzer D In-Line Mounting Installation and Maintenance 3/07

17 7. In-Line Mounting 7.1 General Information Choice of pipe tee This mounting allows the electrode body to be inserted directly into a Schedule 40, 3/4 in. NPT metal pipe tee in a pipe line. The electrode will NOT fit into standard PLASTIC pipe tees. Use a metal tee or the special Honeywell schedule 80 CPVC 3/4 in. pipe tee, Honeywell part number Avoiding damage to the electrode The Durafet II electrode for in-line mounting has an exposed sensor. Exercise care when inserting or removing the electrode from the pipe tee to prevent damage to the sensor surface. Ensuring accurate temperature sensing When process temperature varies considerably from ambient temperature, insulate the entire Durafet II electrode body to ensure accurate process temperature sensing. Included in assembly Assemblies vary according to catalog number configuration ordered, but may include a Durafet II combination electrode, a cap adapter module, and connection cables. A description of the Durafet II catalog number is provided in Section 2. If no cap adapter is used, the electrode would be cabled directly to the instrument as described in Section 5. Orientation of electrode For reliable measurement, the electrode must be immersed in the process fluid. Therefore, orientation of the electrode vertical to the horizon is not recommended, as this orientation may prevent sufficient depth penetration to reliably immerse the sensor. The electrode mounting angle should be at least 45 degrees from vertical as shown in Figure 7-1. If solids are present in the process fluid, avoid angles exceeding 90 degrees from vertical to minimize accumulation of solids around the sensor. ATTENTION The ISFET die orientation in the pipe tee is determined by using the H in Honeywell on the electrode label. The ISFET die is lined up with the H on the label. 3/ D In-Line Mounting Installation and Maintenance 11

18 Figure 7-1 Proper Mounting Angle for Electrode D In-Line Mounting Installation and Maintenance 3/07

19 8. Maintenance, Standardization, and Replacement Parts 8.1 Maintenance Keeping electrode moist When an electrode is removed from the process for any reason, assure that it does not become dry and remain dry for more than a short period of time. The electrode may require more frequent maintenance if used in a batch treatment installation which leaves the electrode dry between batches, or if it is exposed to process fluids that leave a deposit on the surface of the sensor chip (measuring electrode) and reference electrode. The manual supplied with the electrode contains instructions for cleaning the electrode. Thoroughly rinse the electrode with water after any type of cleaning. In addition to periodic cleaning, other electrode maintenance includes electrode performance checks, and treatment for a clogged junction or severely dry electrode. CAUTION Read the manual (part number ) supplied with the electrode before attempting any maintenance procedures. Improper handling of the electrode can result in damage that will affect accuracy. 8.2 Standardization In addition to electrode maintenance, successful ph measurement relies on periodic standardizing of the measurement instrument and its electrode system. (This is required because all electrodes do not produce exactly the same potential in a solution of known ph. A periodic corrective adjustment eliminates any deviation from the standard value.) Establish regular intervals for standardizing according to conditions and experience. Procedures for standardizing are given in the measuring instrument instructions. CAUTION If the cable connection has to be disconnected to get the Durafet ph electrode out of its mounting, then it is necessary to allow the Durafet ph electrode to warm up for at least 10 minutes prior to beginning a calibration. 3/ D In-Line Mounting Installation and Maintenance 13

20 8.3 Replacement and Accessory Parts Durafet II electrode Description Standard Durafet II cables for UDA2182 Analyzer electrode cable, 6.10 m (20 ft) length electrode cable, m (50 ft) length Cap Adapter Durafet II cables 6.10 m (20 ft) length m (50 ft) length Teflon tape, 1/2 in. x 260 in. roll CPVC pipe tee, special dimension 3/4 in. NPT for in-line mounting ph Buffer Solutions (NIST Traceable) 4.01 ph 6.86 ph 9.18 ph Part Number See the electrode manual (part number ) (provided with every Catalog 7777) Gel Re-fill Kit Electrode Tip (Smooth) D In-Line Mounting Installation and Maintenance 3/07

21

22 Honeywell Process Solutions Industrial Measurement and Control 512 Virginia Drive Fort Washington, PA Rev Printed in USA

Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes Installation and Maintenance Rev. 1 3/04

Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes Installation and Maintenance Rev. 1 3/04 Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes 70-82-25-116 Rev. 1 3/04 Industrial Measurement and Control Copyright, Notices, and Trademarks Printed in U.S.A. Copyright 2004 by

More information

7777 Immersion/In-Line Mounting for Durafet II ph Electrodes Installation and Maintenance

7777 Immersion/In-Line Mounting for Durafet II ph Electrodes Installation and Maintenance 7777 Immersion/In-Line Mounting for Durafet II ph Electrodes Installation and Maintenance 70-82-25-86 Rev. 1 4/00 Sensing and Control Copyright, Notices, and Trademarks Printed in U.S.A. Copyright 2000

More information

7777 Immersion/In-Line Mounting for Meredian II ph/orp Electrodes Installation and Maintenance Manual

7777 Immersion/In-Line Mounting for Meredian II ph/orp Electrodes Installation and Maintenance Manual 7777 Immersion/In-Line Mounting for Meredian II ph/orp Electrodes Installation and Maintenance Manual 70-82-25-05 Rev. 1 3/08 Honeywell Process Solutions Copyright, Notices, and Trademarks Printed in U.S.A.

More information

7774 Series Stainless Steel Insertion/Removal Assembly for Use with Meredian Electrodes

7774 Series Stainless Steel Insertion/Removal Assembly for Use with Meredian Electrodes 7774 Series Stainless Steel Insertion/Removal Assembly for Use with Meredian Electrodes 70-82-25-44 Rev. 1 10/07 Honeywell Process Solutions Copyright, Notices, and Trademarks Copyright 2007 by Honeywell

More information

7774 Series Stainless Steel Insertion/Removal Assembly for Use with Durafet II ph Electrodes Rev. 0 5/99

7774 Series Stainless Steel Insertion/Removal Assembly for Use with Durafet II ph Electrodes Rev. 0 5/99 7774 Series Stainless Steel Insertion/Removal Assembly for Use with Durafet II ph Electrodes 70-82-25-88 Rev. 0 5/99 Copyright, Notices, and Trademarks Copyright 1999 by Honeywell Inc. Revision 0 5/99

More information

HercuLine 2000 Series Gear Kit Replacement Instructions

HercuLine 2000 Series Gear Kit Replacement Instructions HercuLine 2000 Series Gear Kit Replacement Instructions 51452443-501 Document Information Form: 62-86-33-35 Effective: 5/04 Supersedes: None Summary This kit contains all the gears, idler shafts and hardware

More information

U00X ULTRASONIC LEVEL SWITCH. Ultrasonic Liquid Level Switches INSTALLATION AND OPERATIONS MANUAL. For Models: U002, U003 & U004

U00X ULTRASONIC LEVEL SWITCH. Ultrasonic Liquid Level Switches INSTALLATION AND OPERATIONS MANUAL. For Models: U002, U003 & U004 U00X ULTRASONIC LEVEL SWITCH INSTALLATION AND OPERATIONS MANUAL Ultrasonic Liquid Level Switches For Non-Hazardous Locations For Models: U002, U003 & U004 READ THIS MANUAL PRIOR TO INSTALLATION This manual

More information

EQUIVALENT BASIC CIRCUIT

EQUIVALENT BASIC CIRCUIT FEATURES Reliable semiconductor technology Calibrated and temperature compensated Rugged, stainless steel package NEMA 4 design Small size Absolute, gage, sealed gage, vacuum gage pressures 0 psi to 3

More information

10260S Series Actuators Auto-Manual Switch Replacement Instruction

10260S Series Actuators Auto-Manual Switch Replacement Instruction 10260S Series Actuators Auto-Manual Switch Replacement Instruction Document Number Form: 62-86-33-26 Effective: 06/01 Supersedes: none Overview These instructions apply to the following kit: 51500581-501

More information

11280A HercuLine Series Actuator. Specification. Non-Contact Position Sensing and Variable Speed Motor Control. Overview. Features. Actuator Operation

11280A HercuLine Series Actuator. Specification. Non-Contact Position Sensing and Variable Speed Motor Control. Overview. Features. Actuator Operation 11280A HercuLine Series Actuator Non-Contact Position Sensing and Variable Speed Motor Control 10/00 Page 1 of 12 Specification Overview To operate with maximum efficiency and improve process uptime, state-of-the-art

More information

General Purpose ph/orp Sensors

General Purpose ph/orp Sensors Product Data Sheet PDS 71-3900/rev. B December 2010 Models 3900 and 3900VP General Purpose ph/orp Sensors EXTENDED SENSOR LIFE provided by double junction reference. ENHANCED PERFORMANCE AND INCREASED

More information

NEW QUICK CABLE-TO-SENSOR RELEASE,

NEW QUICK CABLE-TO-SENSOR RELEASE, Models 399 and 399VP Product Data Sheet PDS 71-399 October 2000 ph/orp Sensors NEW OPTIONS to meet a variety of application requirements, including: * Low ionic strength water * Processes with hydrofluoric

More information

General Purpose ph/orp Sensors

General Purpose ph/orp Sensors General Purpose ph/orp Sensors Product Data Sheet LIQ_PDS_3900 September 2013 General Purpose ph/orp Sensors SMART preamplifier-stores calibration and other data. Extended sensor life provided by double

More information

Instruction Manual. Model HC1 Series Electrodless Conductivity Sensors

Instruction Manual. Model HC1 Series Electrodless Conductivity Sensors Instruction Manual Anderson Instrument Co. Inc. 156 Auriesville Road Fultonville, NY 12072 1-800-833-0081 Fax 518-922-8997 www.anderson-negele.com Instrument Model Number Instrument Serial Number Model

More information

U00X ULTRASONIC LEVEL SWITCH. Ultrasonic Liquid Level Switches INSTALLATION AND OPERATIONS MANUAL. For Models: U002, U003 & U004

U00X ULTRASONIC LEVEL SWITCH. Ultrasonic Liquid Level Switches INSTALLATION AND OPERATIONS MANUAL. For Models: U002, U003 & U004 U00X ULTRASONIC LEVEL SWITCH INSTALLATION AND OPERATIONS MANUAL Ultrasonic Liquid Level Switches For Non-Hazardous Locations For Models: U002, U003 & U004 READ THIS MANUAL PRIOR TO INSTALLATION This manual

More information

* * Rev. 9 01/18 Product Manual

* * Rev. 9 01/18 Product Manual English Signet 2760 DryLoc ph/orp Preamplifier *3-2760.090* 3-2760.090 Rev. 9 01/18 Product Manual Description The Signet 2760 ph/orp Preamplifiers with DryLoc connectors are part of Signet's growing line

More information

Installation and Operating Manual

Installation and Operating Manual Model u002 & u003 Installation and Operating Manual Ultrasonic Level Switch 60 Great Hill Road Naugatuck, CT 06770 ph: 203-729-6434 fax: 203-729-6696 www.innovativesensing.com Read this Manual Before Installing

More information

Microcontroller Unit (MCU) LCD Hour Meters/Counters. LM Series. Datasheet

Microcontroller Unit (MCU) LCD Hour Meters/Counters. LM Series. Datasheet Microcontroller Unit (MCU) LCD Hour Meters/Counters LM Series Datasheet LM Series Microcontroller Unit (MCU) LCD Hour Meters/Counters Honeywell s MCU (microcontroller unit) LCD (liquid crystal display)

More information

ph/orp Sensor Model 399 ph/orp

ph/orp Sensor Model 399 ph/orp Model 399 ph/orp Product Data Sheet PDS 71-399 September 1999 ph/orp Sensor DOUBLE JUNCTION REFERENCE CELL provides longer sensor life in process solutions containing poisoning ions. ONE-PIECE CONSTRUCTION

More information

For additional information, please visit our website at SENSOR SPECIFICATIONS

For additional information, please visit our website at  SENSOR SPECIFICATIONS Instruction Sheet PN 51A-140/rev.E December 2010 Models 140, 141, and 142 Conductivity Sensors For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. SENSOR/PROCESS

More information

Accessories Guide ph/orp/conductivity Sensors

Accessories Guide ph/orp/conductivity Sensors Jet Cleaners and Scrubber Accessories Isolation Ball Valves Assemblies Compression Fittings and Sheath Sets High Pressure (2500 PSIG) Sensor Housings In-line Flow Cells Nut Lock Union Style Adapters Sensor

More information

SIGNET 2760 DryLoc ph/orp Preamplifier

SIGNET 2760 DryLoc ph/orp Preamplifier SIGNET 2760 DryLoc ph/orp Preamplifier 3-2760.090 3-2760.090 (A-3/04) English SAFETY INSTRUCTIONS 1. Depressurize and vent system prior to installation or removal. 2. Confirm chemical compatibility before

More information

3.5mm Coaxial Calibration Kit

3.5mm Coaxial Calibration Kit User Guide 3.5mm Coaxial Calibration Kit DC to 26.5 GHz Models: 8050CK10/11 8050CK20/21 8050-511 (A) 2/15 User Guide 3.5mm Coaxial Calibration Kit DC to 26.5 GHz Models: 8050CK10/11 8050CK20/21 2900 Inland

More information

10260 Retrofit to 10260A Non-Contact Sensor Instructions Kits through -512

10260 Retrofit to 10260A Non-Contact Sensor Instructions Kits through -512 10260 Retrofit to 10260A Non-Contact Sensor Instructions Kits 51404974-501 through -512 Document Number Form: 62-86-33-16D Effective: 3-1-01 Supersedes: 11-00 Summary Retrofitting your 10260 actuator from

More information

Models 396/396P/396R. Retraction/Submersion/Insertion ph/orp Sensor

Models 396/396P/396R. Retraction/Submersion/Insertion ph/orp Sensor Models 396/396P/396R Product Data Sheet PDS 71-TUpH September 1999 Retraction/Submersion/Insertion ph/orp Sensor MINIMUM SENSOR MAINTENANCE due to patented TUpH reference technology 1 which ensures steady

More information

Insertion/Submersion Flow Through Sensor

Insertion/Submersion Flow Through Sensor Product Data Sheet PDS 71-381/rev.B September 2004 Model 381 ph/orp Insertion/Submersion Flow Through Sensor ADVANCED ON-LINE SENSOR DIAGNOSTICS 1 (Model 381+ only), for use with the Models 54e ph/orp

More information

VSP SERIES Installation, Operation and Maintenance Manual

VSP SERIES Installation, Operation and Maintenance Manual VSP SERIES Installation, Operation and Maintenance Manual VARIABLE SPEED PERISTALTIC METERING PUMPS VSP-12 and VSP-20 Models READ ALL WARNINGS CAREFULLY BEFORE INSTALLING PUMP PUMP DATA/SPECIFICATIONS

More information

Signet DryLoc ph/orp Electrodes

Signet DryLoc ph/orp Electrodes Signet 2774-2777 DryLoc ph/orp Electrodes Features Patented DryLoc connector with gold plated contacts* Special design allows for installation at any angle, even inverted or horizontal Quick temperature

More information

9142-B-P-156/AF. User s Guide. Field Metrology Well. Distribution is limited to DoD and U.S. DoD contractors only.

9142-B-P-156/AF. User s Guide. Field Metrology Well. Distribution is limited to DoD and U.S. DoD contractors only. 9142-B-P-156/AF Field Metrology Well User s Guide Distribution is limited to DoD and U.S. DoD contractors only. May 2013 2013 Fluke Corporation. All rights reserved. Specifications are subject to change

More information

OPERATING INSTRUCTIONS PLEASE READ CAREFULLY

OPERATING INSTRUCTIONS PLEASE READ CAREFULLY OPERATING INSTRUCTIONS PLEASE READ CAREFULLY 925-0330 Rev 0 0416 TABLE OF CONTENTS SAFETY SUMMARY... 3 SPECIFICATIONS... 4 1.0 INTRODUCTION/DESCRIPTION.... 5 2.0 LOCATION AND MOUNTING... 5 3.0 CONNECTIONS

More information

Signet 2760 DryLoc ph/orp Preamplifier

Signet 2760 DryLoc ph/orp Preamplifier Signet 2760 DryLoc ph/orp Preamplifier *3-2760.090* 3-2760.090 Rev. H 05/11 English English SAFETY INSTRUCTIONS 1. Depressurize and vent system prior to installation or removal. 2. Confi rm chemical compatibility

More information

Signet DryLoc ph/orp Electrodes

Signet DryLoc ph/orp Electrodes Signet 2774-2777 DryLoc ph/orp Electrodes Features Patented DryLoc connector with gold plated contacts* Special design allows for installation at any angle, even inverted or horizontal Quick temperature

More information

Defender 3000 Series Base Instruction Manual

Defender 3000 Series Base Instruction Manual Defender 3000 Series Base Instruction Manual 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Compliance This product conforms to

More information

1 AMP CURRENT SOURCE

1 AMP CURRENT SOURCE 1 AMP CURRENT SOURCE CS-2000-U CS-2000-E USER MANUAL BC BIOMEDICAL CS-2000 SERIES TABLE OF CONTENTS WARNINGS, CAUTIONS, NOTICES... ii DESCRIPTION... 1 LAYOUT... 2 OPERATION... 4 TESTING... 6 MANUAL REVISIONS...

More information

Installation and Service Instructions for EX Series Explosion-proof Switches

Installation and Service Instructions for EX Series Explosion-proof Switches Installation and Service Instructions for EX Series Explosion-proof Switches ISSUE 4 50009662 WARNING PERSONAL INJURY DO NOT USE these products as safety or emergency stop devices or in any other application

More information

Rosemount 402 and 402VP

Rosemount 402 and 402VP Rosemount 402 and 402VP Contacting Conductivity Sensors Instruction Manual LIQ-MAN-402 Rev. M May 2017 hasgkas Essential Instructions Read this page before proceeding! Emerson designs, manufactures and

More information

Signet 3719 ph/orp Wet-Tap

Signet 3719 ph/orp Wet-Tap Signet 3719 ph/orp Wet-Tap *3-3719.090* 3-3719.090 Rev H 09/07 English Do not exceed maximum temperature/pressure specifications. Wear safety goggles or face shield during installation/service. Do not

More information

INSTRUCTIONS PARTS LIST This manual contains important warnings and information. READ AND RETAIN FOR REFERENCE

INSTRUCTIONS PARTS LIST This manual contains important warnings and information. READ AND RETAIN FOR REFERENCE INSTRUCTIONS PARTS LIST 308 493 This manual contains important warnings and information. READ AND RETAIN FOR REFERENCE Rev. A Second Gun Hose Kit 100 psi (6.9 bar) Maximum Working Pressure These kits include

More information

Distributed by: www.jameco.com -800-83-22 The content and copyrights of the attached material are the property of its owner. 0-5 psi Through 0-300 psi FEATURES Low Cost Sensor Element Internal Temperature

More information

CONTACTING CONDUCTIVITY GENERAL PURPOSE SENSORS Instruction Manual

CONTACTING CONDUCTIVITY GENERAL PURPOSE SENSORS Instruction Manual CONTACTING CONDUCTIVITY GENERAL PURPOSE SENSORS Instruction Manual Notice 2018 WALCHEM, Iwaki America Inc. (hereinafter Walchem ) 5 Boynton Road, Holliston, MA 01746 USA (508) 429-1110 All Rights Reserved

More information

4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual Rev. 1 7/99

4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual Rev. 1 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 70-82-25-19 Rev. 1 7/99 Copyright, Notices, and Trademarks Printed in U.S.A. Copyright 1999 by Honeywell Inc. Revision 1 7/99 While

More information

Load Accessories. Test & Measurement Sensors DESCRIPTION TONGUE SHACKLE FEATURES PULL PLATE FEATURES YOKE SHACKLE FEATURES LOAD BUTTON FEATURES

Load Accessories. Test & Measurement Sensors DESCRIPTION TONGUE SHACKLE FEATURES PULL PLATE FEATURES YOKE SHACKLE FEATURES LOAD BUTTON FEATURES Test & Measurement Sensors DESCRIPTION Honeywell provides a wide range of accessories for our test and measurement sensor products, which are designed to meet additional requirements for most customer

More information

INSTALLATION INSTRUCTIONS SWITCH HOUSING SUPPLIED ONLY AS AN ASSEMBLY OF ENCAPSULATED SWITCH & LEADS. FLOAT SUPPLIED ONLY WITH ENCAPSULATED MAGNET.

INSTALLATION INSTRUCTIONS SWITCH HOUSING SUPPLIED ONLY AS AN ASSEMBLY OF ENCAPSULATED SWITCH & LEADS. FLOAT SUPPLIED ONLY WITH ENCAPSULATED MAGNET. INSTALLATION INSTRUCTIONS SWITCH HOUSING, L-40N NORYL (PPO) SWITCH HOUSING, L-40VCR FORTRON (PPS) CERAMIC MAGNET BUSHING, L-40N NORYL (PPO) L-40VCR FORTRON (PPS) LEADS 24 LONG OR LONGER REED SWITCH FLOAT,

More information

AEROMOTIVE Part # L Mustang Digital FMU INSTALLATION INSTRUCTIONS

AEROMOTIVE Part # L Mustang Digital FMU INSTALLATION INSTRUCTIONS AEROMOTIVE Part # 17113 4.6L Mustang Digital FMU INSTALLATION INSTRUCTIONS CAUTION: Installation of this product requires detailed knowledge of automotive systems and repair procedures. We recommend that

More information

SETRA Operating Instructions Series DPT 230 Pressure Transducers

SETRA Operating Instructions Series DPT 230 Pressure Transducers SETRA Operating Instructions Series DPT 230 Pressure Transducers GENERAL INFORMATION Your Setra Pressure Transducer has been carefully tested and calibrated before shipment. Specific performance information

More information

Retractable ph/orp Sensors

Retractable ph/orp Sensors Model 396R and 396RVP Instruction Manual PN 5100396R January 2001 Retractable ph/orp Sensors ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests

More information

Model NTX7 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006

Model NTX7 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006 B R A N D Model NTX7 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006 For Sales, Support and Service phone: 407-331-4793 fax: 407-331-4708 website: www.xenotronix.com email: information@xenotronix.com

More information

INSTALLATION, CALIBRATION & TROUBLESHOOTING MANUAL

INSTALLATION, CALIBRATION & TROUBLESHOOTING MANUAL INSTALLATION, CALIBRATION & TROUBLESHOOTING MANUAL Model 7200 Transducers GP:50 2770 Long Road, Grand Island, NY 14072 Tel: 716-773-9300 Fax: 716-773-5019 Email: sales@gp50.com Website: http://www.gp50.com

More information

Model 616 Rotating Disk Electrode Instruction Manual

Model 616 Rotating Disk Electrode Instruction Manual Model 616 Rotating Disk Electrode Instruction Manual 219303 C / 1202 Printed in USA Advanced Measurement Technology, Inc. a/k/a Princeton Applied Research, a subsidiary of AMETEK, Inc. WARRANTY Princeton

More information

2.4mm Coaxial Calibration Kit

2.4mm Coaxial Calibration Kit User Guide 2.4mm Coaxial Calibration Kit DC to 50 GHz Models: 7950CK10/11 7950CK20/21 7950-511 (A) 2/15 User Guide 2.4mm Coaxial Calibration Kit DC to 50 GHz Models: 7950CK10/11 7950CK20/21 2900 Inland

More information

OPERATING INSTRUCTION MANUAL

OPERATING INSTRUCTION MANUAL OPERATING INSTRUCTION MANUAL MODEL P/R65 DIFFERENTIAL 2-WIRE (4-20 ma) ph & ORP PROBES N116-139 REV. 0 Aquametrix, by Water Analytics Inc. 100 School Street Andover, MA 01810 Tel: (978) 749-9949 Toll Free:

More information

Model GD-K77D4X-XX Supplement to the GD-K77D Operator s Manual

Model GD-K77D4X-XX Supplement to the GD-K77D Operator s Manual Model GD-K77D4X-XX Supplement to the GD-K77D Operator s Manual Part Number: 71-0217RK Revision: 0 Released: 3/1/11 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information

Conductivity Sensors. Model 140, 141, 142. Product Data Sheet series/rev.b February 2011

Conductivity Sensors. Model 140, 141, 142. Product Data Sheet series/rev.b February 2011 Product Data Sheet 71-140 series/rev.b February 2011 Model 140, 141, 142 Conductivity Sensors RUGGED CONSTRUCTION withstands high temperature and pressure TWO MOUNTING CONFIGURATIONS: screw-in and retractable

More information

Signet ph/orp Electrodes

Signet ph/orp Electrodes Signet 2734-2736 ph/orp Electrodes Compatible with Signet 8900/9900 Instruments Features Patented reference design for exceptional performance * Mounts in Signet standard fittings from DN15 to DN100 (½

More information

COOPER POWER SERIES. Screened Separable Connectors MN650023EN. Effective October 2016 Supersedes March 2012 (IS )

COOPER POWER SERIES. Screened Separable Connectors MN650023EN. Effective October 2016 Supersedes March 2012 (IS ) Screened Separable Connectors MN650023EN Effective October 2016 Supersedes March 2012 (IS550-60-1) COOPER POWER SERIES DTS1242 Deadbreak Bolted Tee Connector: Interface C for Single Core Cable with Copper

More information

BlueFin User's Manual. Version 1.1

BlueFin User's Manual. Version 1.1 BlueFin User's Manual Version 1.1 10 August 2005 1 Safety Precautions Battery charging The BlueFin unit comes with a power adapter. Please use this power adapter for operating the unit and charging the

More information

COOPER POWER SERIES. Screened Separable Connectors MN650024EN. Effective October 2016 Supersedes March 2012 (IS )

COOPER POWER SERIES. Screened Separable Connectors MN650024EN. Effective October 2016 Supersedes March 2012 (IS ) Screened Separable Connectors MN650024EN Effective October 2016 Supersedes March 2012 (IS550-61-1) COOPER POWER SERIES DTB1242 Deadbreak Bolted Companion Tee Connector: Interface C for Single Core Cable

More information

Switching DC Power Supply

Switching DC Power Supply 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Model 1693, 1694 Switching DC Power Supply INSTRUCTION MANUAL 1 Safety Summary

More information

OWNER S MANUAL. Model 9880 Intrinsically Safe Air Velocity Meter

OWNER S MANUAL. Model 9880 Intrinsically Safe Air Velocity Meter OWNER S MANUAL Model 9880 Intrinsically Safe Air Velocity Meter LIMITATION OF WARRANTY AND LIABILITY Seller warrants the goods sold hereunder, under normal use and service as described in the operator's

More information

Retractable ph/orp Sensors

Retractable ph/orp Sensors Instruction Manual PN 51-396R/rev.J February 2011 Model 396R and 396RVP Retractable ph/orp Sensors ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and

More information

OPERATING INSTRUCTION MANUAL MS SERIES CONDUCTIVITY CELLS

OPERATING INSTRUCTION MANUAL MS SERIES CONDUCTIVITY CELLS OPERATING INSTRUCTION MANUAL MS SERIES CONDUCTIVITY CELLS REV. 3.1 Water Analytics, Inc. 100 School Street Andover, MA 01810 978-749-9949 Toll free: 855-747-7623 Fax: 978-749-9961 www.wateranalytics.net

More information

PRG 350 Series Stainless Steel Regulators

PRG 350 Series Stainless Steel Regulators PRG 350 Series Stainless Steel Regulators omega.com info@omega.com U.S.A. Headquarters: Servicing North America: Omega Engineering, Inc. Toll-Free: 1-800-826-6342 (USA & Canada only) Customer Service:

More information

For Hazardous Locations

For Hazardous Locations L500E LIQUID LEVEL FLOAT SWITCH INSTALLATION AND OPERATIONS MANUAL Single or Multi-Point, Vertically-Mounted Liquid Level Float Switch For Hazardous Locations READ THIS MANUAL PRIOR TO INSTALLATION This

More information

Signet 3719 ph/orp Wet-Tap Assembly

Signet 3719 ph/orp Wet-Tap Assembly Signet 3719 ph/p Wet-Tap Assembly Features Electrode removal without process shutdown Space saving 45 mm (1.75 in.) short-stroke design Sealed pneumatic dampening for smooth and safe operation SafeLoc

More information

Signet ph/orp Electrodes

Signet ph/orp Electrodes Signet 2724-2726 ph/orp Electrodes Compatible with ALL Signet ph/orp Instruments Features Flat Glass Protected Bulb Patented reference design for exceptional performance * Mounts in Signet standard fittings

More information

Switch-Tek Horizontal and Vertical And Float-Point Float Switches LV20, LV21, LV35, LV36, LH25, LH29, LH35, AV_6 Series 22AUG 08 Rev A

Switch-Tek Horizontal and Vertical And Float-Point Float Switches LV20, LV21, LV35, LV36, LH25, LH29, LH35, AV_6 Series 22AUG 08 Rev A Switch-Tek Horizontal and Vertical And Float-Point Float Switches LV20, LV21, LV35, LV36, LH25, LH29, LH35, AV_6 Series 22AUG 08 Flowline, Inc. 10500 Humbolt Street Los Alamitos, CA 90720 Tel: (562) 598-3015

More information

Signet ph/orp Electrodes

Signet ph/orp Electrodes Signet 2724-2726 ph/orp Electrodes Compatible with ALL Signet ph/orp Instruments Features Patented reference design for exceptional performance * Mounts in Signet standard fittings from DN15 to DN100 (½

More information

Liquid Conductivity Two & Four Electrode Sensors

Liquid Conductivity Two & Four Electrode Sensors Industrial Process Control & Monitoring Rugged Two Electrode Conductivity Sensors Unique Four Electrode Conductivity Sensors Versatile Installation Options (Threaded, Quick Change, Retractable) Specialized

More information

Rosemount 3300HT/3300HTVP/3400HT/ 3400HTVP/3500P/3500VP

Rosemount 3300HT/3300HTVP/3400HT/ 3400HTVP/3500P/3500VP Instruction Manual LIQ-MAN-PERpH-X Rev. J April 2017 Rosemount 3300HT/3300HTVP/3400HT/ 3400HTVP/3500P/3500VP PERpH-X High Performance ph/orp Sensors hasgkas Essential Instructions Read this page before

More information

Component Test Fixture

Component Test Fixture PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT 1865-52 Component Test Fixture User and Service Manual 1865-52 im/march 2010 Copyright 2010 IET Labs, Inc. IET LABS, INC. 534 Main Street, Westbury, NY 11590

More information

FAIRCHILD T5420 DIGITAL-PNEUMATIC TRANSDUCER Installation, Operation and Maintenance Instructions

FAIRCHILD T5420 DIGITAL-PNEUMATIC TRANSDUCER Installation, Operation and Maintenance Instructions FAIRCHILD T5420 DIGITAL-PNEUMATIC TRANSDUCER Installation, Operation and Maintenance Instructions Figure 1. T5420 Outline Dimensions. INSTALLATION The Model T5420 can be mounted directly onto a flat surface

More information

BULLETIN NO.ELEC IM121/10A Replaces IM121/09A

BULLETIN NO.ELEC IM121/10A Replaces IM121/09A Mid-West Instrument BULLETIN NO.ELEC IM121/10A Replaces IM121/09A INSPECTION Model 121 Indicating Differential Pressure Switch / Transmitter Electrical: Installation and Operating Instructions Upon receipt

More information

User's Manual. High Voltage Megohmmeter. Analog Insulation Tester plus AC Voltage and Continuity Tests. Model

User's Manual. High Voltage Megohmmeter. Analog Insulation Tester plus AC Voltage and Continuity Tests. Model User's Manual High Voltage Megohmmeter Analog Insulation Tester plus AC Voltage and Continuity Tests Model 380353 Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects

More information

DUSTTRAK AEROSOL MONITOR SOLAR POWER KIT MODEL

DUSTTRAK AEROSOL MONITOR SOLAR POWER KIT MODEL DUSTTRAK AEROSOL MONITOR SOLAR POWER KIT MODEL 854060 (USED FOR POWERING ENVIRONMENTAL ENCLOSURE MODELS MODELS 854030, 8535 AND 8537) OPERATION AND MAINTENANCE MANUAL P/N 6008416, REVISION C JUNE 2017

More information

IMPORTANT SAFETY INSTRUCTIONS

IMPORTANT SAFETY INSTRUCTIONS OWNER S MANUAL FLO-MASTER XP2 SERIES PUMPS IMPORTANT SAFETY INSTRUCTIONS When installing and using this electrical equipment, basic safety precautions should always be followed, including the following:

More information

Installation Instructions for Eaton Surge Protective

Installation Instructions for Eaton Surge Protective Contents Description Page PH Model Introduction.... 2 Safety Precautions.... 2 Operation.... 2 Preparation... 2 Installation Procedures... 3 Enclosure Mounting (New Applications - Eaton PH and PV Models)....

More information

BMK-30. Heavy-Duty By-Pass Filtration System Installation and Servicing Instructions

BMK-30. Heavy-Duty By-Pass Filtration System Installation and Servicing Instructions BMK-30 Heavy-Duty By-Pass Filtration System Installation and Servicing Instructions IMPORTANT NOTICE Read all instructions completely before attempting to install this unit. Improper installation could

More information

MICRO SWITCH Solenoid Safety Interlock Switches GKL/R Series Issue 2. Datasheet

MICRO SWITCH Solenoid Safety Interlock Switches GKL/R Series Issue 2. Datasheet MICRO SWITCH Solenoid Safety Interlock Switches GKL/R Series DESCRIPTION MICRO SWITCH GKL/R Series switches are a heavy-duty key operated solenoid safety interlock switch incorporating a key trapping mechanism.

More information

Model 8340 Intrinsically Safe VELOCICHECK Air Velocity Meter

Model 8340 Intrinsically Safe VELOCICHECK Air Velocity Meter Ventilation Testing/Balancing Model 8340 Intrinsically Safe VELOCICHECK Air Velocity Meter Operation and Service Manual 1980076, Revision E October 2002 Model 8340 Intrinsically Safe VELOCICHECK Air Velocity

More information

007 Tank Monitor. WARNING Never use with gasoline or highly flammable liquids.

007 Tank Monitor. WARNING Never use with gasoline or highly flammable liquids. Model #7575 Recommended Installation, Maintenance and Inspection Instructions 007 Tank Monitor 7575 Important safety instructions - save these instructions in a readily accessible location. WARNING Never

More information

User s Guide CDCE-90-X SERIES. Conductivity Sensors. Shop online at omega.com SM

User s Guide CDCE-90-X SERIES. Conductivity Sensors. Shop online at omega.com SM User s Guide Shop online at omega.com SM e-mail: info@omega.com For latest product manuals: www.omegamanual.info CDCE-90-X SERIES Conductivity Sensors omega.com info@omega.com U.S.A. Headquarters: Servicing

More information

CIRCLE SEAL CONTROLS

CIRCLE SEAL CONTROLS CIRCLE SEAL CONTROLS ATKOMATIC SOLENOID VALVES INSTALLATION, MAINTENANCE, AND OPERATION INSTRUCTIONS 13000 SERIES Stainless, Normally Closed, Direct Lift, 3-Way Valve Installation Instructions WARNING:

More information

HYGRODYNAMICS DIGITAL DEW POINT MONITOR PC BOARD MODEL 6392N & 6392N2 TABLE OF CONTENTS

HYGRODYNAMICS DIGITAL DEW POINT MONITOR PC BOARD MODEL 6392N & 6392N2 TABLE OF CONTENTS HYGRODYNAMICS DIGITAL DEW POINT MONITOR PC BOARD MODEL 6392N & 6392N2 TABLE OF CONTENTS INTRODUCTION... 2 SPECIFICATIONS... 2 INSTALLATION... 3 PC Board Mounting...3 Sensor Installation...3 Manifold Assembly...4

More information

MS Application. Features. Proportional Valve Actuator General Instructions

MS Application. Features. Proportional Valve Actuator General Instructions Proportional Valve Actuator Application The Proportional Valve Actuator is a non-spring return actuator used with proportional to 10 Vdc or 4 to 0 ma controllers and standard 1/" to 1-1/4" two-way and

More information

F-4600 INLINE ULTRASONIC FLOW METER Installation and Operation Guide

F-4600 INLINE ULTRASONIC FLOW METER Installation and Operation Guide F-4600 INLINE ULTRASONIC FLOW METER Installation and Operation Guide 11451 Belcher Road South, Largo, FL 33773 USA Tel +1 (727) 447-6140 Fax +1 (727) 442-5699 1054-7 / 34405 www.onicon.com sales@onicon.com

More information

OPERATING INSTRUCTION MANUAL

OPERATING INSTRUCTION MANUAL OPERATING INSTRUCTION MANUAL Model 3422-series Compression Fitting Style Contacting Conductivity/Resistivity Sensors (Enhanced Performance Design) Worldwide Sales: GLI International, Inc. 9020 West Dean

More information

Model 385+ Instruction Manual PN February ph/orp Sensor

Model 385+ Instruction Manual PN February ph/orp Sensor Model 385+ Instruction Manual PN 5100385+ February 2001 ph/orp Sensor ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests its products to meet

More information

VL Series. General Purpose Compact Limit Switches. Sensing and Control

VL Series. General Purpose Compact Limit Switches. Sensing and Control FEATURES Gold plated silver contacts Compact design for small mounting space. Special design for easy wiring with ample wiring space Two-circuit double-break Diecast base and plastic cover Standard mounting

More information

onlinecomponents.com VL Series General Purpose Compact Limit Switches Sensing and Control

onlinecomponents.com VL Series General Purpose Compact Limit Switches Sensing and Control FEATURES Gold plated silver contacts Compact design for small mounting space. Special design for easy wiring with ample wiring space Two-circuit double-break Diecast base and plastic cover Standard mounting

More information

Encapsulated Differential ph and ORP Sensors (available in LCP, foreground, or Ryton body materials)

Encapsulated Differential ph and ORP Sensors (available in LCP, foreground, or Ryton body materials) Technology For Solutions Data Sheet LRE/102 Supersedes LRE/400 Encapsulated Differential ph and ORP Sensors (available in LCP, foreground, or Ryton body materials) Common Features Differential Electrode

More information

660R-RAC Series. Gasoline Fuel Filter/Water Separators. Overview: Contact Information: Product Features: Instruction Part Number Rev F

660R-RAC Series. Gasoline Fuel Filter/Water Separators. Overview: Contact Information: Product Features: Instruction Part Number Rev F 660R-RAC Series Gasoline Fuel Filter/Water Separators Instruction Part Number 1385 Rev F Overview: Don t be caught in the water without one of these Racor gasoline spin on series filters. These filters

More information

Universal Battery Holder Operator s Manual

Universal Battery Holder Operator s Manual Universal Battery Holder Operator s Manual Copyright 2018 Gamry Instruments, Inc. Revision 1.0 August 13, 2018 988-00068 Limited Warranty Gamry Instruments, Inc. warrants to the original user of this product

More information

Digital ph transmitter

Digital ph transmitter Digital ph transmitter Compact, remote versions for DN 15 to 200 Fully programmable ph transmitter functions for all kinds of ph-measurement tasks Large range of process connections with various fittings

More information

SIGNET 2750 ph/orp Sensor ENGLISH

SIGNET 2750 ph/orp Sensor ENGLISH ENGLISH 3-2750.090 B-8/01 English SAFETY INSTRUCTIONS FOR IN-LINE ELECTRODE INSTALLATION 1. Do not remove from pressurized lines. 2. Do not exceed maximum temperature/pressure specifications. 3. Wear safety

More information

User s Guide FP-5060 SERIES. Micro-Flow Sensors. Shop online at omega.com SM For latest product manuals:

User s Guide FP-5060 SERIES. Micro-Flow Sensors. Shop online at omega.com SM   For latest product manuals: User s Guide Shop online at omega.com SM e-mail: info@omega.com For latest product manuals: www.omegamanual.info FP-5060 SERIES Micro-Flow Sensors omega.com info@omega.com U.S.A. Headquarters: Servicing

More information

TBM Series 3-Way Ball Valve

TBM Series 3-Way Ball Valve www.simtechusa.com TBM Series 3-Way Ball Valve Operating, Installation, & Maintenance Manual Corrosion Resistant Fluid and Air Handling Systems. Dated 06-26-13 TBM Series Ball Valves SIMTECHRECOMMENDSREADINGTHEFOLLOWINGINFORMATIONPRIORTOINSTALLINGANDUSING

More information

Digital Thermometers Instruction Manual MODELS 868 and 869

Digital Thermometers Instruction Manual MODELS 868 and 869 Digital Thermometers Instruction Manual MODELS 868 and 869 Instruction Manual PN# 869-901-01CD Publication Date: May 2017 REV. G NOTE: This User s Manual was as current as possible when this product was

More information

AV16, AV26, AV36, AV46, AV56, LH25, LH29, LH35, LV20, LV21, LV35 & LV36 Series Manual

AV16, AV26, AV36, AV46, AV56, LH25, LH29, LH35, LV20, LV21, LV35 & LV36 Series Manual Switch-Tek Horizontal, Vertical & Float-Point Float Switches AV16, AV26, AV36, AV46, AV56, LH25, LH29, LH35, LV20, LV21, LV35 & LV36 Series Manual Flowline, Inc. 10500 Humbolt Street, Los Alamitos, CA

More information

KOBOLD DF-MA Flow Transmitter. User Instructions

KOBOLD DF-MA Flow Transmitter. User Instructions KOBOLD DF-MA Flow Transmitter User Instructions KOBOLD Instruments Inc. 1801 Parkway View Drive Pittsburgh PA 15205 Phone (412) 788-2830 Fax (412)-788-4890 www.koboldusa.com Manual-DF-MA_12-13 DF-MA Table

More information

Filter Holder Manifold User Guide. 3- and 6-Place Stainless Steel Manifolds 3- and 6-Place Sterility Test Manifolds 3-Place PVC Manifold

Filter Holder Manifold User Guide. 3- and 6-Place Stainless Steel Manifolds 3- and 6-Place Sterility Test Manifolds 3-Place PVC Manifold Filter Holder Manifold User Guide 3- and 6-Place Stainless Steel Manifolds 3- and 6-Place Sterility Test Manifolds 3-Place PVC Manifold Notice The information in this document is subject to change without

More information