Meeting the Requirements of ASTM D ( IP548/06 ) with Agilent 1200 LC Systems

Similar documents
Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems

Application Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC

Application Note. Authors. Abstract. Energy & Chemicals

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE

Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

Agilent and ASTM. Update on Recent Activities. Page 1

Application Note. Abstract. Authors. Environmental Analysis

Using the PSD for Backflushing on the Agilent 8890 GC System

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell

Application Note. Author. Introduction. Energy and Fuels

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Application. Gas Chromatography June 1995

Agilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass. James D. McCurry, Ph.D. Senior Scientist

Agilent G5632 Valve Kit Instructions

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Group-Type Analysis (PiPNA) in Diesel and Jet Fuel by Flow Modulated GCxGC FID.

Analysis of Petroleum Fractions by ASTM D2887

Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution

Analysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

Application Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method

Determination of Free and Total Glycerin in B100 Biodiesel

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

Complete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges

Simulated Distillation Analyzers, Software, Standards, Consumables, Training

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

Agilent Solutions for the Petrochemical and Oleochemical Industries

ASTM D 6730 Detailed Hydrocarbon Analysis

Emissions from Heavy-Duty Diesel Engine with EGR using Oil Sands Derived Fuels

White Paper.

TECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure

High-Temperature Simulated Distillation System Based on the 6890N GC Application

Module8:Engine Fuels and Their Effects on Emissions Lecture 36:Hydrocarbon Fuels and Quality Requirements FUELS AND EFFECTS ON ENGINE EMISSIONS

Restek Corporation 110 Benner Circle, Bellefonte, Pa Barry L. Burger, Neil Johansen, Valerie Gamble, Donald Rhoades

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography

Method Development for Capillary GC Systems. Slide 1

Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

Determination of Sudan Dyes I IV in Curry Paste

Tennessee Department of Agriculture

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE

Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

Agilent 1290 Infinity II Ultra Low Dispersion Kit

Agilent Valve Kit G4231A/B/C and G4232B/C/D

White Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information

Agilent Valve Kit G4231A/B and G4232A/B/C

Voting Draft Standard

GC Method Compliance and Large Valve Oven Application

Large Volume Injection of Polycyclic Aromatic Hydrocarbons

SUPPRESSION IN ANALYSIS OF ARTIFICIAL SWEETENERS WITH

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector

The Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC

AppNote 1/2010 KEYWORDS ABSTRACT. Biodiesel, Automation, ASTM D

WET GRIP TEST METHOD IMPROVEMENT for Passenger Car Tyres (C1) GRBP 68 th session

C2, C3, C4 Monomer Analysis

a review of analytical methods for the quantification of aromatics in diesel fuels

Performing ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software

Technical Procedure for Gas Chromatography (GC-FID)

On-Line Process Analyzers: Potential Uses and Applications

Online sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve

Zürich Testing on Fuel Effects and Future Work Programme

Refill Capacities and Recommendations

Sulfur Detection at ppb Levels in Light Hydrocarbon Streams

3.0 FUEL INJECTION AND INDUCTION SYSTEM CLEANER CONCENTRATE

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

Analysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES

AppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS

MET-Biodiesel Capillary GC Columns

Determination of chlorine/chlorides using MWD- XRF

Analysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID

ISBN SANS 342:2006 Edition 4 SOUTH AFRICAN NATIONAL STANDARD Automotive diesel fuel Published by Standards South Africa 1 dr lategan roa

GC Inlets. An Overview. Simon Jones GC Applications Engineer

DxC 700 AU Job Aid Booklet

Installation of the 2 position/10 port valve into the 1100 series thermostatted column compartment

ElvaX Sulfur in Oil analyzer

Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology

Synthetic Fuel Formulation from Natural Gas via GTL: A Synopsis and the Path Forward

Comparison of IG Testing Standard. IGMA Summer Meeting August 8, 2004

Operation and Applications of Differential Flow Modulation

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

S 1125 HPLC Pump System

Ultra Low Sulfur Diesel Fuel For Use In Locomotive Marine

Quality-by-Design in Method Development. Mijo Stanic General Manager and Technical Director

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

Transcription:

Meeting the Requirements of ASTM D 6591-06 ( IP548/06 ) with Agilent 1200 LC Systems M.Woodman, Chemical Analysis Solutions Unit, Wilmington, Delaware, USA M.Sierocinska, European Field Support Centre, Waldbronn, Germany

Agenda What is the source and application for ASTM D 6591-06 (and IP548/06)? Description of Recent ASTM D 6591-06 Method Updates New D 6591-06 / IP548/06 Method Requirements Alternate methods for biodiesel blends Recommendations for the best configuration Results from locally collected diesel samples Summary Page 2

Usage and Source of ASTM/IP/EN methods used ASTM D 6591-06 and IP548/06 measure aromatic hydrocarbon groups in petroleum distillates having specified boiling point ranges of 150-400 o C. These methods are mainly applied to fuel analysis, particularly diesel motor fuel. Kerosene based fuels for aviation are covered by another related method ASTM (www.astm.org) is the source of ASTM methods. The Energy Institute (www.energyinst.org.uk) is the source of the IP (sometimes referenced as EI) methods IP (or EI) and EN methods (European Committee for Standardization http://www.cen.eu/cen/pages/default.aspx ) often have parallel ASTM methods The method revision designation is very important Page 3

Related IP, EN and ASTM Methods IP Method and Revision Method Overview Special Parameters ASTM Method? Comments IP391/07 150-400 o C diesel fuel petro/bio blends up to B-5 no backflush amino and/or cyano column no equivalent available at this time same as method EN12916:2006 MAH, DAH, Tri+AH reported IP548/06 150-400 o C diesel fuel backflush required D-6591-06 MAH, DAH, Tri+AH reported amino and/or cyano column FAME interferes with result IP436/01 50-300 o C aviation fuel, kerosene no backflush amino and/or cyano column D-6379-04 MAH and DAH reported not for samples with Tri+AH Page 4

General Conditions and Schematic of ASTM D 6591 Mobile phase: n-heptane at 1ml/min flow rate Column: Amino-propyl and/or Amino-Cyano-propyl bonded phases. Amino-propyl 4.6x250mm 5um is commonly used Column Temperature: approx. 20-25 o C Detection: Refractive Index Special Feature: 6 port valve for backflushing the column to elute tri+ aromatics as a single peak Page 5

General Schematic of ASTM D 6591-06 Figure 1, Agilent publication 5965-9044E Page 6

Recent HGA Method Updates: ASTM D6591-06 Terms redefined more specific description of polyaromatics adopted Designate a new IP method correlation IP391-95 (used with ASTM D 6591-00) replaced by IP548/06 Exclusion or warning regarding Bio/Petrodiesel Blends the method recognizes difficulties with biodiesel fuel blends and refers the analyst to IP391/07 (EN12916:2006) for samples containing biodiesel (FAME) components due to inclusion of FAME components in the tri+ah backflush peak, erroneously high tri+ aromatic results are reported if D6591-06 is used with biodiesel blends unusually high tri+ah results with this method should be investigated to determine if unsuspected FAME is present due to fuel contamination or mislabeling Page 7

General Conditions and Schematic of D 6591-06 System Performance Standard 1. cyclohexane 2. o-xylene (dimethylbenzene) 3. dibenzothiophene 4. 9-methyl anthracene 1 2 3 4 ASTM D 6591-06 System Performance Std., Agilent p/n 5190-0483 Mobile phase: n-heptane at 1ml/min flow rate Column: ZORBAX NH2 4.6x250mm, 5um Column Temperature: 20 o C Detection: Refractive Index, 35 o C Page 8

General Conditions and Schematic of D 6591-06 Calibration Standards 1. cyclohexane 2. o-xylene (1,2-dimethylbenzene) 3. 1-methyl-naphthalene 4. phenanthrene with backflush at ~9 min. 1 2 3 4 ASTM D 6591-06 Calibration Standard A-D, Agilent kit p/n 5190-0482 Mobile phase: n-heptane at 1ml/min flow rate Column: ZORBAX NH2 4.6x250mm, 5um Column Temperature: 20 o C Detection: Refractive Index, 35 o C Page 9

Analysis of Petrodiesel, overlay of 3 injections 1. saturates 2. mono-aromatics 3. di-aromatics 4. tri+ aromatics t= <30 minutes Ultra Low Sulfur Diesel (ULSD), retail pump 1 2 3 4 Page 10

Analysis of Petrodiesel, overlay of 3 brands 1. saturates 2. mono-aromatics 3. di-aromatics 4. tri+ aromatics t= <30 minutes Ultra Low Sulfur Diesel (ULSD), retail pumps a winter blend, old retained sample summer blends 1 2 3 4 Page 11

ASTM D 6591-06 Separation Criteria The official method clearly describes the appearance and statistical evaluation of method conditions for the new revision Resolution, selectivity, precision and calibration linearity are clearly stated In the following slides we review the performance of the Agilent approach to this method. indicates that Agilent meets requirements

ASTM D 6591-06 Criteria Column Selection Column Descriptions: 4.6 x 250 mm, stainless steel column packed with 5um ZORBAX NH2 (p/n 880952-708) SPS 1. cyclohexane 2. o-xylene (1,2-dimethylbenzene) 3. dibenzothiophene 4. 9-methyl anthracene 1 2 3 4 1ml/min heptane 20C column temp. 35C RID temp. 20ul SPS no backflush. 6.4 Column System Any stainless steel HPLC column(s) packed with an approved amino-bonded (or polar amino/cyano-bonded) silica stationary phase is suitable

SPS Separation and Resolution Criteria SPS 1. cyclohexane 2. o-xylene (1,2-dimethylbenzene) 3. dibenzothiophene 4. 9-methyl anthracene 1 2 3 4 1ml/min heptane 20C column temp. 35C RID temp. 20ul SPS backflush 9.4 min... 6.4 provided it meets the resolution requirements laid down in 9.4.3. 9.4.1 Ensure that baseline separation is obtained between all components of the SPS. [without backflush]. 9.6 Calculate the backflush time, B, in seconds, using the following equation: B = ta + 0.4(tB ta)

Calculation of Resolution (Cyclohexane to 1,2 dimethylbenzene (o-xylene)) 9.4.3 Ensure that the resolution between cyclohexane and 1,2 dimethylbenzene is not less than 5. 9.4.3.1 Column Resolution -- Calculate the resolution, R, between cyclohexane and 1,2 dimethylbenzene using the following equation. 2(t 2 -t 1 ) (difference in retention time) R = ------------------------ 1.699(y 1 +y 2 ) (averaging of peak widths). Name R. Time width (hh) Rsln. 1. cyclohexane 3.307 0.059 2. 1,2-dimethylbenzene) 4.477 0.097 8.79 3. dibenzothiophene 8.907 0.186 4. 9-methyl anthracene 18.905 0.282

Linearity (section 9.4: R = >0.999, Intercept <0.01g/100ml) RID peak area vs. Std.Conc., o-xylene 3500000 3000000 y = 812489.767010x + 963.112364 R 2 = 0.999995 RID Peak Area 2500000 2000000 1500000 1000000 Series1 Linear (Series1) 500000 0 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 Conc. (g/100 ml). correlation coefficients all >0.999 (fits are nominally 1), linear fit of the calibration points Intercepts all =<0.005g/100ml

Reproducibility for Retention and Area, Cal. A-D Calibrant A Analyte R.T. Avg., n=3 R.T. Stdev R.T.RSD% Area Avg., n=3 Area Stdev Area.RSD% xylene 4.44 0.0005774 0.01% 3.29E+06 755.932096 0.02% 1-Me-naphthalene 5.96 0.001 0.02% 5.76E+06 2299.268 0.04% phenanthrene 20.14 0.0020817 0.01% 7.12E+05 8351.82533 1.17% Calibrant B Analyte R.T. Avg., n=3 R.T. Stdev R.T.RSD% Area Avg., n=3 Area Stdev Area.RSD% xylene 4.55 0.0020817 0.05% 8.33E+05 5263.96922 0.63% 1-Me-naphthalene 6.24 0.0041633 0.07% 1.46E+06 14197.7087 0.97% phenanthrene 20.13 0.0023094 0.01% 3.55E+05 849.500049 0.24% Calibrant C Analyte R.T. Avg., n=3 R.T. Stdev R.T.RSD% Area Avg., n=3 Area Stdev Area.RSD% xylene 4.63 0.0017321 0.04% 2.06E+05 536.395687 0.26% 1-Me-naphthalene 6.44 0.0036056 0.06% 3.66E+05 1830.77315 0.50% phenanthrene 20.12 0.0040415 0.02% 8.87E+04 139.006199 0.16% Calibrant D Analyte R.T. Avg., n=3 R.T. Stdev R.T.RSD% Area Avg., n=3 Area Stdev Area.RSD% xylene 4.67 0.0005774 0.01% 4.03E+04 214.794072 0.53% 1-Me-naphthalene 6.65 0.0020817 0.03% 2.96E+04 334.193512 1.13% phenanthrene 20.10 0.0025166 0.01% 1.76E+04 176.541808 1.00% Average RSD% All Runs 0.028% 0.555%

Recommendations for the best configuration Agilent pump, minimum isocratic G1311A Agilent sampler, minimum G1329A Agilent thermostatted column compartment (TCC) with 6 port 2 position switching/backflush valve Agilent RID G1362A Agilent ChemStation B.04.02 or EZChrom software, or third party compatible control/data software Agilent ZORBAX NH2 4.6x250mm 5um column equilibrated to heptane mobile phase Agilent calibrants for system performance and calibration HPLC grade n-heptane, 2-propanol Page 18

What to order, for consumables, ASTM D 6591-06 ASTM D 6591-06 (IP548/06) System Performance Standard (SPS) Agilent p/n 5190-0483 ASTM D 6591-06 (IP548/06) System Calibration Solutions A-D Agilent p/n 5190-0482 Column(s): ZORBAX NH2 4.6x250mm, 5um, Agilent p/n 880952-708 guard cartridge (pk/4) 4.6x12.5mm, 5um, Agilent p/n 820950-908 guard hardware (holder) kit, 4.6mm, Agilent p/n 820888-901 Normal phase pump seals, (pk/2), Agilent p/n 0905-1420 Mobile phase: n-heptane, HPLC grade, minimum 4 Liter or 2.5 Liter bottles, 2-propanol, HPLC grade, 1 Liter (installation, maintenance) Page 19

Performance with Selected Samples Samples were collected from retail and commercial diesel and biodiesel distribution centers around the Chicago area One older sample was collected in the cold winter months, all others in recent warm summer months. Fuel blends vary seasonally in this area, with change occurring around early September and April Sample preparation: Accurately weigh ~1g fuel and dilute to 10ml with n-heptane. Mix thoroughly and inject 10ul. Calculations: Determine sample concentrations from calibration curves using 1,2 dimethylbenzene for MAH, 1-methyl-naphthalene for DAH and phenanthrene for Tri+AH. Correct for sample weight and dilution variables. Conditions: as described in previous slides Page 20

Analysis of Petrodiesel, overlay of 3 brands 1. saturates 2. mono-aromatics 3. di-aromatics 4. tri+ aromatics t= <30 minutes Ultra Low Sulfur Diesel (ULSD), retail pumps a winter blend, old retained sample summer blends 1 2 3 4 Page 21

Analysis of Petrodiesel, n=3, Vendor 1 Group Avg., n=3 RSD% MAH 29.26 g/100ml 0.68% DAH 4.74 g/100ml 2.77% Tri+AH 0.70 g/100ml 6.63% Vendor 1, Retail pump, summer MAH DAH Tri+AH Mobil Page 22

Analysis of Petrodiesel, n=3, Vendor 2 Group Avg., n=3 RSD% MAH 29.38 g/100ml 0.12% DAH 5.10 g/100ml 0.60% Tri+AH 0.77 g/100ml 1.52% Vendor 2, Retail pump, summer MAH DAH Tri+AH Marathon Page 23

Analysis of BioPetrodiesel with ASTM D 6591-06 Amount units Group 21.75 g/100ml MAH 4.12 g/100ml DAH 3.59 g/100ml Tri+AH Vendor 3, consumer retail pump, summer, dispensing pump labeled diesel erroneous high Tri+AH due to FAME in sample Citgo Page 24

Analysis of BioPetrodiesel with ASTM D 6591-06 Amount units Group 26.21 g/100ml MAH 5.12 g/100ml DAH 4.12 g/100ml Tri+AH Vendor 4, commercial biodiesel blend, dispensing pump labeled diesel erroneous high Tri+AH due to FAME in sample RR Page 25

Analysis of Biodiesel B-11 with Method IP391/07 Group Avg., n=3 RSD% MAH 24.74 g/100ml 5.06% DAH 5.15 g/100ml 3.38% Tri+AH 0.72 g/100ml 4.58% Vendor 4, ULSD-based biodiesel blend retail auto/light truck Results are n=3 by IP391/07 method approved for diesel and biodiesel blends Tri+AH region FAME resolved RR Page 26

ASTM D 6591-06 Observations The method retention time stability is very sensitive to column activity, as noted in the official method. This is typical for many normal phase separation methods Moisture in solvents or samples is problematic and affects column activity Variable retention is primarily due to moisture and temperature, though good system maintenance is essential Use of normal phase seals in the pump is very important Page 27

Summary ASTM D 6591-06 Status Specialized all-agilent column arrangement meets SPS resolution requirements for cyclohexane and 1,2-dimethylbenzene (o-xylene) Adding the 6-port valve to the system configuration is required, however it is possible to easily switch between current IP391/07 (no backflush, for biodiesel blends) and ASTM D 6591 (with backflush, no FAME content) Page 28

References 1. IP391/07 / EN12916:2006, Energy Institute, UK 2. Agilent application note 5965-9044E 3. ASTM Methods D6591-06, D6379-04 Page 29

Appendix and supplemental slides may be used depending on audience and need. Page 30

Acknowledgements Cynthia guidance, resources and encouragement Maggie continued technical support and for persistence in keeping this development on the road map Melanie/Georgia/Phil standards packaging and work with the Ultra Scientific team Andre creating the opportunity and motivation Ultra Scientific too many to name, once again committed to a successful development with Agilent Page 31